ARUN MICROELECTRONICS LTD. PRESSURE GAUGE
Contents
1. LEAD IS DISCONNECTED The mains supply is filtered to prevent conducted electromagnetic interference affecting the operation of this or other equipment nearby To ensure that this filtering is effective and because there is an earth leakage current generated within the instrument it is necessary to return this directly to the vacuum system ground reference star point Ensure that other instruments are directly and separately earthed so that return or fault currents cannot flow in any common ground impedance This is particularly important in cases where there are high voltage power supplies in the system there must be specific low impedance paths for return or flashover currents reliance on frame continuity or sneak paths will cause noises and spikes to be coupled into instrumentation Low resistance connections do not necessarily have low impedance which is most successfully achieved with a Litz conductor of large cross section and number of strands Information on wiring for process control will be found in section 3 5 below 2 4 Fusing Two fuses for instrument protection are provided Spare fuses are provided in the accessory kit Do not use fuses of other types or ratings as this may result in damage to the instrument or gaugeheads under fault conditions 2 4 1 The mains fuse The rating of the mains fuse is dependent on the range of supply voltages from which the instrument is to be operated and is printed on the rear panel adjac
2. 1 Bridge supply Ov red 2 Signal voltage green 3 Bridge supply 2 06v yellow The colours refer to those of the conductors in VG PVG1 leads The blue wire is not used and the end should be cut back and insulated The pin numbering is consistent with that of the connector types formerly used B 5 Bakeout thermocouple connector and cables This is a miniature flat pin K type connector for example Labfacility type FMTC K M The thermocouple should be a type K insulated junction type and must be connected the correct way to the plug The brown wire should be connected to the terminal marked and the blue wire to the terminal marked If a thermocouple with an uninsulated junction is used and allowed to touch any conducting surface serious errors of temperature measurement will occur CONNECTION TO ANY SOURCE OF VOLTAGE MAY DAMAGE THE INSTRUMENT Mineral insulated K thermocouples are recommended and are inexpensive and widely available If the lead supplied is not long enough extend it with type K extension wire and connectors only Using copper or other types of extension cables may give rise to serious errors of temperature measurement Signal voltages from thermocouples are very small so some precautions against induced interference are desirable Mineral insulated thermocouples are well screened but the installation should only earth the screen at one point to avoid currents in the screen Insulated thermocouple wires or extension cables s
3. Bits 3 0 Bit 4 Bit 5 Bit 6 Bit 7 Error byte Bit 0 Bit 1 Bit 2 Bit 3 Bit 4 Bit 5 Bit 6 Instrument type 01002 PGC1 0 local mode 1 remote mode 1 0 0 gauge specific error over temperature trip internal settings were lost and restored to factory default temperature warning auto emission error host command not accepted always set to 1 The value in the error byte is maintained until reset by a lt reset error gt command 3 1 3 Timing of Next Command The host computer should not poll the PGC1 more than 4 time begin transmitting a new command until a CR LF has been received signalling the end of transmission or two instruments may conflict on the serial line It is not necessary to per second and we strongly recommend that a delay of at least 100ms is implemented before the next report request 3 2 PGC1 Response time For the commands lt poll gt lt control gt lt release gt lt long gt lt short gt and lt reset error gt which do not require any extra parameters the instrument addressed will begin transmitting a response within about 200 us The response to other commands is sent as quickly as possible typically 1 5 ms but a minimum response time cannot be specified The transmission time taken for a report to be sent is determined principally by the baud rate 9600 3 3 Local remote control An instrument starts operation in local control i e using the front panel In
4. 2 or 3 Pirani gauges 4 Capacitance Manometer T TSP control B Bakeout control System configuration record 12 Header byte Pirani 1 interlock disabled enabled Relay configuration lon gauge relay de energised when ion gauge is off lon gauge relay is energised when ion gauge is off ae Units of measurement M mbar P pascal or T torr 10 18 Program date 8 character ASCII string i e DD MM YY 19 21 Ambient temperature NNN degrees 22 25 Capacitance manometer NNNX where NNN is 1 10 or 100 and X is either M full scale mbar or T torr 26 28 lon gauge sensitivity NNX where NN is the sensitivity and X is M mbar P pascal or T torr 29 40 Additional system parameters may be defined in future Two byte checksum CR LF 13 Appendix A Example of a host PGC1 dialogue EEE Asci Hex Deseripon o O 2A 50 35 Host polls instrument 5 Instr 5 24 40 PGC1 under local control CR LF OD 0A Host Host m 2A 50 31 Host Host polls instrument instrument 1 Instr 1 4A 34 41 PGC1 under host control Gauge specific error CR LF OD 0A occurred e er 2A 53 31 Host requests short report Instr 1 Status and error bytes d Relays A C D energised GI1 Gauge Bayard Alpert gauge 1 A On A Error over emission 2 7E 03 32 2E 37 45 Pressure 2D 30 33 2C GP2 47 50 32 Gauge record Pirani Gauge 1 A 41 40 running 7 5E 03 37 2E 35 45 2D 30 33 2C 47 50 33 Ga
5. 3 12 Remote operation Control and interrogation of the instrument by a host computer through the Remote connector is possible A demonstration program which will run on any IBM XT AT or compatible computer is available on diskette The host computer must have at least one RS232 serial port Up to eight instruments may be connected to one serial port and each is identified by a number from 1 to 8 which must be selected when remote operation is initiated Information on the interface protocols is given in Appendix C below All instruments connected to the serial interface should be switched on before REMOTE is selected on any instrument Do not switch off any instrument which may be communicating with the host computer The preferred method of disconnecting an instrument from an active interface is to exit REMOTE and to disconnect the Remote connector before switching it off 4 FAULTS AND WARNINGS All faults or potential faults detected by the instrument are announced on the LED display and more detailed information is shown on the LCD display Normally this gives the user the option to cancel the message by pressing the ENTER switch which returns the LCD to the first menu under START Before doing so it is normally worthwhile pressing the HELP switch which will cause the LCD to display further useful information regarding the fault Once ENTER has been pressed this information is no longer available Most of the messages likely
6. The external wiring is connected to the instrument through a two part pluggable terminal block This can be removed and wired independently of the instrument Ensure that external wiring is of adequate cross section for the load current Strip and twist the wires do not tin them and poke into the receptacle in the terminal block Close the leaf on the wire by tightening the screw immediately above Take appropriate action to strain relieve the wiring nearby and to restrict access to the terminals if harmful voltages are to be present 3 5 2 Assigning the relays to gauges or functions Switch on the instrument or return to the initial menu and select SETUP of the configuration parameters Progress to the display which has RELAY A on the top line The content of the bottom line will depend on the options fitted to this particular instrument and defines the gauges or functions to which the relay can be assigned Selecting OTHER will reveal another set of possible assignments Selecting OTHER again may reveal more sets or return to the first set according to the fitted options and software It is suggested that relay A be assiged to the lon Gauge in order to obtain the required displays at later stages of this procedure Assign relays B C and D with reference to the following paragraphs If a relay is assigned to a gauge the relay is energised when the pressure is below the setpoint The hysteresis is fixed and the relay will de energise whe
7. choices available in some subsequent menus Selecting the incorrect type will cause problems Selection of Iridium for a tungsten filament gauge will probably result in insufficient power being available but selection of tungsten for an iridium filament gauge will probably destroy the filament at some time If you do not know the type of filaments in use select Iridium as this will give the greatest protection to the gaugeheads although it may not be possible to operate one which is actually tungsten properly under all conditions You are now asked to adjust the filament current limit If the gaugehead manufacturer has a recommended limit adjust to a little below that limit In the absence of a recommendation select about 1 5 Amps for Iridium and 3 0 Amps for Tungsten This will probably be too low for most gauges under all conditions but will protect the filaments Use an instrument screwdriver with a 3mm blade and insert this in the hole adjacent to the HELP switch on the front panel marked ILIM Using no more pressure than is necessry to keep the screwdriver in the slot turn the single turn adjustment until the desired limit is displayed on the LCD Adjust the lon Gauge overpressure limit to the recommended value or lower The default value of 1 0E 02 is satisfactory for most gauges Skip through the menus till Pirani 1 interlock appears and set it as desired Skip through the remaining menus You are now ready to connect ga
8. degas period Modification of the programs may be necessary to suit the pumping capacity of the system and the condition of the gaugeheads The quick n program allows the operator to apply degas power for one minute at the selected power If a degas power step of 2 watts per minute is selected the first step will be 4 watts because normal emission at 10mA gives a grid dissipation of 2 watts A maximum degas power of only 30 watts is allowed for Iridium filaments in order to avoid evaporating the Thoria coating 3 7 2 Running degas Gauges should have been run for some time at 10mA emission before selecting DEGAS Ensure that the pressure in the system is sufficiently low and that the pumping capacity is adequate for the expected gas load Degas operation at high pressures may result in the production of glow discharges in the gauge space or oxidation or other reactions at the surface of the gauge filament which will reduce its life During degas the power and the remaining degas period are displayed on the LCD together with DEG on the LED display The cursor is left in a position where one press on the ENTER switch will stop emission Degassing the collector electrode is of little practical significance However it can be achieved by making up a special section of lead to connect the collector BNC connector into the grid connection on the ion gauge socket Appendix B contains details of these connectors This lead section must be r
9. is conducted away from the filament by residual gas The thermal conductivity of a gas depends on its pressure below about 1 millibar and the nature of the residual gas The Pirani gauge unbalances the Wheatstone bridge and the voltage across the bridge represents pressure over the range of 0 5 millibar to about 1x10E 3 millibar The lower pressure limit is determined by the heat loss due to radiation becoming significant compared to that due to thermal conductivity The radiant heat loss depends on the emissivity of the filament A new filament is bright but can become blackened by deposits from decomposed rotary pump oils and the lower limit of pressure readings will rise It is possible to clean filaments A 3 Capacitance Manometers Capacitance manometers operate by measuring the deflection of a thin circular radially tensioned membrane between the vacuum space and a reference volume at a pressure substantially below the operating range of the transducer The deflection is measured as a modulation of the electrical capacitance between the membrane and a fixed plate and converted to a voltage proportional to the pressure difference across the membrane APPENDIX B Connectors IN THE SECTIONS BELOW THE MATING CONNECTORS ON THE CABLES ARE DESCRIBED B 1 Mains connector This is a female IEC CEE22 type and is supplied with the instrument in the accessory pack in the form of a moulded lead with integral 13 Amp British domestic style plug I
10. lon Gauge 3 2 4 Measuring pressure with the ion gauge 3 3 Operation at extreme UHV and emission current selection 3 4 Operation of the ion gauge at high pressures 3 5 Process Control and Trips 3 5 1 Use and wiring of the relays 3 5 2 Assigning the relays to gauges or functions 3 5 3 lon gauge trips 3 5 4 Setting trip pressures before operating the ion gauge 3 5 5 Setting trip pressures while operating the ion gauge 3 6 Leak Detection 3 7 lon Bombardment Degas 3 7 1 Degas programs 3 7 2 Running degas 3 7 3 Problems in degas INDEX 3 8 Bakeout Control 3 9 Capacitance Manometer 3 10 Recorder Output 3 11 External inhibit of the lon Gauge 3 12 Remote operation 4 FAULTS AND WARNINGS APPENDIX A Gauge principles A 1 lonisation gauges A 2 Pirani Gauges A 3 Capacitance Manometers APPENDIX B Connectors B 1 Mains connector B 2 lon gauge power connector B 3 lon gauge collector connector B 4 Pirani connectors B 5 Bakeout thermocouple connector and cables B 6 Remote connector B 8 Auxiliary connector 1 INTRODUCTION This intelligent Pressure Gauge Controller is designed to replace earlier generation equipment and provides enhanced performance and facilities at significantly reduced size and with no increase in cost The large easily read LED display allows viewing of pressure in either bargraph or numeric formats selectable with a single keypress A continuous voltage analog of pressure is available
11. lt sound gt n Instr D T Using divisor D to divide a frequency of 921600 Hz produce a tone for T ms 1 these instrument addresses may be global i e X Examples C2 control instrument 2 CX control all instruments SO get a short status report from instrument 0 i01 turn on IG in instrument 0 at 1 mA n1 turn off IG in instrument 1 r6C2 0E 10 setpoint relay C instrument 6 2 0e 10 mbar 3 5 Status reports There are two types of status report The short status report gives the operating status and pressure of each gauge in the instrument The long status report lists the configuration of each gauge each relay and the instrument The information given by the long status report only needs to be read once after taking control of an instrument its information may change during operation but only predictably in response to host computer commands The short report can be requested as often as required Pressures The pressure measurement for each gauge in the instrument is updated 4 times a second Pirani and capacitance manometer gauge pressures are displayed and reported without filtering lon gauge pressures can be filtered with a time constant of 1 2 4 or 8 seconds but filtering can be turned off if desired The pressure field for a non operating gauge is filled with spaces Status report formats All status reports begin with the status and error characters described above and terminate with a checksum and CR LF
12. maintain a usable collector current without creating a pressure burst from increased filament temperature Advanced Electrometer Design An electrometer with a logarithmic characteristic is included as this gives a wide dynamic range which is guaranteed to be monotonic and smooth over the entire range of the ion gauge A novel compensation technique ensures that the error currents in the logging circuits do not cause an optimistically low pressure reading to be given at the bottom of the operating range as is common in log electrometers The settling time at low input currents is minimised by a new non linear frequency compensation technique Conformance and high ambient temperature performance are improved by advanced temperature compensation techniques The electrometer is mounted directly in the path of incoming forced air and is thermally isolated from internal sources of heat 2 INSTALLATION 2 1 Checks on receipt of the instrument On receipt of the instrument remove all packing material and check that all items on the shipping list have been received Report any damage or shortages to the Company or the Agent who supplied the instrument The packing material has been specially designed to protect the instrument and should be retained for possible future use 2 2 lon gaugehead installation Consult the information supplied with the gaugehead for advice on flanges gaskets and adaptors for mechanical fixing Consult section 2 5 and Appen
13. only available in numeric format If the pressure measured by the CM is greater than its fullscale or if it is disconnected the reading is replaced on the LEDs by CM gt FS 3 10 Recorder Output An analog voltage representing pressure is available on the auxiliary connector This is scaled at 0 25 volts per decade of pressure and 0 volts represents 1 0x10E 14 in the selected units of measurement When the ion gauge is operating its filtered pressure measurement is represented otherwise that of Pirani 1 is used as the source The output resistance of the recorder output is 1 kilohm and the output will withstand a continuous short circuit to either 15v Ov or 15v which are all present on the auxiliary connector Consult appendix B 8 for information on connector type and pin numbering If an analog voltage representing pressures above the ion gauge range is required when the ion gauge is operating the voltage output of the capacitance manometer is recommended The output impedance of this is generally such that a recorder may be driven directly 3 11 External inhibit of the lon Gauge Operation of the ion gauge may be inhibited by an external contact closure which prevents starting of the ion gauge by any means The contacts should be connected to pins 7 and 8 of the Auxiliary connector and must be isolated from any external source of voltage Gold plated contacts are recommended as the internal voltage and current source are small
14. procedure 3 5 4 Setting trip pressures before operating the ion gauge Enter the RUN menu and select Other then Rel If you have followed the suggestion in section 3 5 2 the LCD display will show Relay A IGEN DN on the top line IG shows the assignment of relay A Note that the cursor is positioned under IG Pressing ENTER confirms that the trip is to be active whereas moving it to EN or DN and pressing ENTER suspends the trip and energises or de energises the relay regardless of pressure Whatever choice is made you are then required to enter the trip pressure in the normal way You will find that only those relays which have been assigned to a gauge will appear in sequence because there is no pressure setpoint associated with those which have been assigned to other functions and it is not possible to suspend control of the relays 3 5 5 Setting trip pressures while operating the ion gauge Start the ion gauge Note that at the bottom right of the LCD display Relays followed by a four digit number is shown Press HELP for an explanation of this Move the cursor under relays and press ENTER The relays their status and setpoints are then displayed The display will revert to the previous display after a few seconds Before this happens move the cursor and note that it moves between sections relating to individual relays Press ENTER when the cursor is on the relay of interest The resulting display is exactly analogou
15. provided by rack manufacturers for this purpose Support brackets may be mounted on the tapped M3 fixing holes on the sides near the rear If these or other arrangements are attached to these holes ensure that the screws used are steel and penetrate the case between 6 and 10 millimetres 2 3 2 Ventilation The instrument is forced air ventilated through grilles on the right side and a vent in the rear panel Mount it in a location where there is an adequate supply of air as close as possible to cool room ambient temperature The instrument is tolerant of and is compensated for operation at elevated ambient temperatures and will give a warning when these are likely to affect measurements or safe operation If the temperature of the incoming air is measured at 45 Celsius or greater the instrument is placed in a low power mode of operation and a recommendation to switch off is given Long term accuracy and reliability will be enhanced by operation at the lowest possible temperature If there are other instruments in the rack which generate significant amounts of waste heat try to ensure that this is deflected away from this instrument 2 3 3 Connection to the mains supply The mains is connected via an IEC CE22 pattern connector THE INSTRUMENT MUST ALSO BE CONNECTED TO EARTH BY THE STUD PROVIDED FAILURE TO PROVIDE THIS CONNECTION MAY RESULT IN A SHOCK HAZARD FOR THE OPERATOR IF HIGH VOLTAGES ARE CONNECTED TO THE GAUGE OR SIGNAL LEADS WHEN THE MAINS
16. show a green light Those marked INC and DEC are used mainly to define choices from a menu or to change numbers displayed on the LCD After making the changes they are confirmed by pressing the ENTER switch this usually produces a new display on the LCD The HELP switch calls up an advisory message on the LCD and sometimes there is a related sequence of these In these cases the last character displayed is an arrow and the next part of the sequence is called by pressing HELP The LCD reverts to displaying the current menu after the last help message All displays of help messages are transient and will change to the next in sequence or be replaced by the previously displayed menu after a few seconds If you have pressed any switches recall the initial menu by switching the power off for a few seconds You are asked to select from RUN SETUP REMOTE Press the HELP key to show an explanation of these terms When the RUN SETUP REMOTE menu returns press INC a few times and observe the effect on the cursor shown on the LCD then do the same with DEC Note that the cursor jumps between the ends of the display Leave the cursor under SETUP and press ENTER 3 1 2 The SETUP menu The security code and numeric data entry Selection of RUN or REMOTE will result eventually in a FAULT message being displayed because of the absence of gauges or a host computer Although this is not harmful it is more instructive to view the menus und
17. status Bit 6 always set to 1 Bit 5 gauge externally inhibited Bit 4 gauge in leak detect Bit 3 gauge in degas Bit 2 gauge controlling bakeout Bit 1 gauge starting Bit 0 gauge operating can combine with bits 2 4 5 Gauge error Bayard tig Gauge error Bit 0 gauge filament open circuit Bit 1 gauge overemission Bit2 gauge underemission Bit3 maximum pressure exceeded Bit4 pirani interlock prevents starting Bit6 1 Bit7 0 Pirani Gauge error Bit0 Pirani gauge open circuit Bit6 1 Bit7 0 6 13 Pressure Comma delimited string in scientific notation e g 1 3E 07 If the gauge is not operating the string consists of spaces only i e Two byte checksum CR LF 11 3 5 3 Long status report Status byte Error byte For each gauge a gauge configuration record mpm o Gauge type l Bayard Alpert P Pirani M capacitance manometer LP filter TC 0 1 2 4 or 8 seconds BA gauge only Filament number 1 or 2 BA gauge only Filament type 0 Iridium 1 Tungsten BA gauge only 0 100uA 1 1mA 2 10mA 3 Auto BA gauge only Maximum pressure N Ne NN mbar cold cathode and BA gauges For each relay a relay configuration record Details 3 Status 0 associated with gauge 2 inhibited de energised 1 over ride energised Setpoint Pressure in scientific notation N NE NN Associated gauge or 1 BA gauge function
18. to be shown in normal service have been described in the appropriate section above The remaining ones concern dangerous overtemperature conditions and fault diagnosis and are adequately amplified by HELP messages where required Environments where there is a large amount of electrical noise may cause occasional erratic behaviour This will show as a blank or unintelligible LCD display or a spontaneous return to the RUN SETUP REMOTE menu In the former case pressing the HELP key will restore the display The second symptom shows that the instrument s program has been disturbed in some way Normally the stored operating and setup parameters will not have been altered if they were then a warning message will be displayed Operation may be restored in the normal way If this occurs more than extremely infrequently you should attempt to cure the problem at the source of the noise The most common cause is noise coupling via inadequate earthing arrangements Refer to section 2 3 for general advice on this point APPENDIX A A 1 ionisation gauges lonisation gauges are thermionic triode devices The appropriate choice for UHV use is the Bayard Alpert type This consists of a very thin collector wire mounted along the axis of a cylindrical mesh grid The filament is outside the grid and usually parallel to it The grid is voltage biased positively with respect to the filament and the collector negatively A stabilised emission current is establish
19. 76 and similar transceivers do not guarantee that the output state of a receiver is defined when the inputs are terminated but undriven because of the 200 millivolt offset of the receiver The claim on the data sheet that The receiver incorporates a fail safe feature which guarantees a high output state when the inputs are left open refers to a condition without termination Connection of a termination external cables or a driver with its outputs in the TRI STATE condition will give unpredictable results and result in transmission errors Normally a 1 kilohm pullup resistor connected between the RXD line and the receiver 5 volt supply and a 1 kilohm pulldown from RXD to the O volt supply line will ensure a high output state under properly terminated conditions For RS232 receivers an open circuit input normally ensures a high output state If this is not the case then adding a 10 kilohm pulldown resistor between the received data line and the receiver negative supply voltage will correct this Because the number of PGC1 instruments driving an interface may be between 1 and 8 it is not possible to fit these additional components on the drivers in the PGC1 2 6 Demonstration software A demonstration programs is available for IBM PC XT AT or compatible computers The computer requires DOS 3 0 or higher a colour display and a free serial port Further information is contained in files included on the demonstration disc 3 INTERFACE PROTO
20. ARUN MICROELECTRONICS LTD PRESSURE GAUGE CONTROLLER MODEL PGC1 USER MANUAL ISSUE 1 31 For use with Program Version 2 00 onward COPYRIGHT RESERVED 1989 The information contained in this manual has been carefully checked and is believed to be correct No responsibility is assumed for any errors User comment and criticism is welcome please write quoting the serial number of the instrument and the version number of the software as displayed by the instrument to Customer Services Arun Microelectronics Ltd Fitzalan Road ARUNDEL West Sussex BN18 9JP England Please direct other enquiries to the distributor or agent from whom you purchased the instrument INDEX 1 INTRODUCTION 2 INSTALLATION 2 1 Checks on receipt of the instrument 2 2 lon gaugehead installation 2 3 Instrument installation 2 3 1 Mounting 2 3 2 Ventilation 2 3 3 Connection to the mains supply 2 4 Fusing 2 4 1 The mains fuse 2 4 2 The grid fuse 2 5 lon Gauge Cables 2 5 1 The ion gauge power cable 2 5 2 The ion gauge collector cable 3 OPERATION 3 1 Familiarisation 3 1 1 Power on the LCD switch functions and HELP 3 1 2 The SETUP menu The security code and numeric data entry 3 1 3 Exploring the menus 3 1 4 Preparing to measure pressure Setting the lon Gauge parameters 3 2 Pressure Measurement 3 2 1 Pressure measurement using Pirani Gauges 3 2 2 Starting the lon Gauge 3 2 3 Problems with starting or running the
21. COL 9600 baud 8 data bits 1 stop bit no parity no handshaking Up to eight instruments can be connected to a single RS232 or RS422 RS485 party line Each instrument maintains its transmit line in a high impedance state except when transmitting data in response to a request from the host computer Each instrument has a unique address set from the setup menu 3 1 General Command and Response formats 3 1 1 Host Computer Commands The host computer sends commands to instruments in the following format First byte ASCII 47 Second byte Command character All commands are represented by a single character Third byte Instrument address Instruments are identified by a single character from 0 8 corresponding to addresses 0 8 Some commands can be addressed to all connected instruments by X Optional ASCII Additional command parameters may be single ASCII characters or character parameters strings a series of ASCII characters followed by a delimiting character any of ASCII 0 13 or 3 1 2 PGC1 response If the command was addressed to all instruments there is no response Otherwise the instrument addressed responds with a status byte and an error byte see next page followed by a CR LF ASCII 13 10 If a status report was requested the status and error bytes are followed by the report a checksum and then the CR LF All responses terminate in CR LF Instrument status byte coding Status byte
22. Four mains power rated changeover relays assignable to any gauge or function are provided Ease of Use The instrument may be configured or set up by selection from menus displayed on a 24 character 2 line liquid crystal display and such setups are maintained in the instrument s memory even in the absence of mains power Information of secondary importance and extensive operator assistance in the form of help screens are also available on this display in unambiguous plain English An experienced operator should be able to use the full facilities without referring to the Manual After setup certain functions may be locked out to prevent unauthorised operation High Accuracy The two most significant factors which determine the performance of ion gauge controllers are the accuracy of emission control and the quality of the electrometer measuring the collector current This unit offers improvements in both of these critical areas Precise Emission Control The filament power is provided by a smooth direct current supply controlled by a fast acting control circuit instead of the normal transformed and chopped mains supply This allows control of emission current to an accuracy of about 1 besides eliminating a source of potential electromagnetic interference Emission current may be selected manually but to allow optimal extension of the dynamic range at extreme UHV Autoemission modulates the electron current in a smoothly varying fashion to
23. ade with conductors of cross section adequate to carry the filament current at the highest ambient temperature expected The cross sectional area of the wiring should be a minimum 1 square millimetre The grid wire carries a very small current and its cross sectional area is relatively unimportant The insulation on all wiring should be rated for at least 1000 volts RMS and should be adequately protected against mechanical damage A further consideration is the voltage drop in the cable at maximum filament current This is again a function of the gauge filament operating voltage and current and is not well defined as it changes as the filament ages As a general guide a VIG17 or VIG22 gauge with tungsten filaments requires about 7 volts at just over 4 amperes during ion bombardment degas This may increase up to about 10 volts at the end of the filaments life Iridium filaments require about half the voltage and current The total drop in the filament and cable which the instrument will accommodate is about 13 5 volts Voltages over this threshold will cause the instrument to report a filament or wiring fault It is good practice to minimise the drop in the cables by making them as short and of as large a cross section as is convenient as this minimises the temperature rise in the instrument 2 5 2 The ion gauge collector cable The ion gauge collector cable should also be as short as convenient Although this cable is screened the amount of in
24. and confirm it is satisfactory If it was faulty the problem should be rectified the cursor moved to OK and ENTER pressed If the fault cannot be rectified moving the cursor to Not OK and pressing ENTER aborts the bakeout An open circuit thermocouple gives a temperature close to 0 degrees As it is possible that the chamber is at this temperature or colder at the start of bakeout the reading may be valid In this case pressing ENTER with the cursor on OK will allow the bakeout to run for 30 minutes after which time a temperature reading of below 5 degrees Celsius again gives a fault indication and stops bakeout 3 9 Capacitance Manometer The capacitance manometer CM is connected via the Auxiliary Connector For details on this connector refer to Appendix B 8 The instrument caters for unheated capacitance manometers with full scale output voltages of 10v representing 10 100 or 1000 millibars or Torr 15v and 15v power is available to drive the CM at up to 35mA if your CM requires more power than this an external power supply will be required this will also allow heated CMs to be used These have better resolution and stability Ensure that the full scale selected in the SETUP configuration parameters matches that of the CM Although this has to be set in millibars or torr the readout is scaled in the selected display units The CM can be interrogated by pressing the DISP switch until the CM pressure reading appears This is
25. ary to use an external relay or contactor to switch the heater current as the rating of the internal relays is inadequate for this purpose An internal relay should be used to energise the coil of the contactor as this presents an inductive load a snubber should be fitted Progress through the SETUP configuration parameters and assign a relay to the Bakeout function Continue until the Bakeout menu is reached You are asked to confirm whether the preset program is satisfactory and you may change the total time including warmup the temperature setpoint and the overpressure trip point The operation of the ion gauge at the end of bakeout may be specified Emission of 10mA is recommended as this runs the gauge hotter and discourages material from condensing on it as the system cools Continue on to the RUN menu and select Bk During bakeout the relay controlling the heaters is energised when the temperature is below the setpoint provided the overpressure trip is not exceeded The hysteresis on temperature switching is 5 degrees Celsius and that on pressure is half the trip pressure The status of the heaters is shown on the LCD together with current temperature pressure and bakeout time remaining The LED display shows BAKEOUT The cursor is left in a position such that a single press on the ENTER switch stops bakeout If the thermocouple is disconnected reversed open circuit FAULT is shown on the LEDs and you are asked to check
26. at low pressures without manual intervention The emission current is progressively increased as the collector current decreases This is done in small increments to avoid pressure bursts caused by increased evolution of adsorbed gas at the higher filament temperature necessary to support the increased emission If the gauge requires degassing or if the photocurrent in the collector generated by soft X rays from the grid is large compared to the ion current the instrument detects this and recommends a return to manual selection of emission This instrument contains an advanced electrometer circuit which is well compensated for environmental changes It is located directly in the path of a cooling airflow just where this enters the instrument and it is thermally isolated from sources of heat Its ambient temperature is monitored and if this approaches a level at which pressure readings may be degraded an advisory warning is given stating the measured temperature This will occur only at extreme UHV and very hot ambient temperatures Increasing the emission current to 10mA will improve the accuracy of the readings If warnings are regularly given with 10mA emission you are advised to improve the cooling arrangements to improve the life and reliability of the instrument Operation of the instrument in conditions of high humidity may affect extreme UHV measurements if condensation is allowed to form within the case The most common cause of this is in
27. combination The checksum is formed by adding all the preceding bytes starting with the status byte taking the two s complement of the least significant 8 bits It is sent in hexadecimal as two characters the most significant 4 bits first Status reports can be read in two ways a because the CR LF terminator can only occur at the end of a report the report can be read into a buffer until the terminator is seen The report is then decoded into gauge relay or system records This is the best approach on slower machines such as IBM PCs b alternatively the report can be read intelligently and decoded record by record As each record has a header character G R or S and the records are of fixed length the first character of the checksum signals the end of the report 3 5 1 Short status report Status byte Error byte A relay status byte The relay status byte is of the form 0100XXXXz where the least significant 4 bits indicate the state of each relay 1 energised The relay status byte indicates the status of relays A to D with relay A indicated by the least significant bit The status of a relay assigned to the ion gauge when it is not in emission is defined in the setup menu this setting is given in the long status report see below An unused byte 10 Then for each gauge in the instrument a gauge record e OOO 2 Gauge type I Bayard Alpert P Pirani M capacitance manometer 4 Gauge
28. d Data lines Disconnect the power connector and remove the smaller screws at the sides of the instrument Note that the ventilation slots are at the right side when viewed from the front Remove the cover in an upward direction Locate the interface PCB by following the wires from the remote connector It is a small PCB inserted into an IC socket near the rear of the instrument Fitting a termination is accomplished by moving the link fitted to the interface PCB from the OFF or NO position to the ON or 85 position only one of these alternative legends will be present according to the issue of the PCB Only one PGC1 receiver termination should be fitted per serial interface all others should have links in the OFF position For compliance with RS485 conditions lines should be terminated at both extremities Termination of the most remote PGC1 transmitter must be done outside the instrument If a PGC1 fitted with a RS485 interface is to be used without being connected to a host computer it is recommended that a termination be fitted in order to decrease its sensitivity to radiated noise 2 4 Cable construction and installation Twisted pair cables are recommended for both RS232 and RS422 RS485 lines Pairs of 7 x 0 2mm 0 22mm PVC equipment wire twisted about 1 revolution per 1 to 1 cm are suitable This results in a characteristic impedance of around 100 ohms and is a reasonable match to the RS422 RS485 terminations An ov
29. dix B for more information on making or adapting cables to connect between the gaugehead and instrument Mount the gaugehead in a position where the free electrons generated in its vicinity will not affect other equipment The performance of the ion gauge may be affected by other electron or ion generating processes within the vacuum chamber should shielding of the gaugehead be necessary ensure that the conductance between the gaugehead and volume of interest is not significantly decreased by its presence The orientation of the gaugehead should be such that the filament is to the side of or below the grid structure This will ensure that if the filament should sag or break it will not short circuit to the grid The gauge and controller are protected from all normal failure modes of either Users should be aware of potential hazards from other equipment however particularly those introducing high voltages into the vacuum chamber X ray sources for example As a direct discharge from one of these at high pressure may cause extensive damage shielding should always be introduced in such cases 2 3 Instrument installation 2 3 1 Mounting The instrument is suitable for mounting in a standard 19 rack and occupies 1U 1 75 44 5mm of the rack The mounting holes in the front panel are intended for retaining the instrument in the rack and will not support its weight Additional support is required toward the rear and various arrangements are
30. e instrument to Customer Services Arun Microelectronics Ltd Fitzalan Road ARUNDEL West Sussex BN18 9JS England Please direct other enquiries to the distributor or agent from whom you purchased the instrument CONTENTS 1 INTRODUCTION 2 HARDWARE 2 1 PGC1 Line Drivers and receivers 2 2 2 3 RS485 422 Line termination 2 4 Cable construction and installation 2 5 Host computer line drivers and receivers 2 6 Demonstration software 3 INTERFACE PROTOCOL 3 1 General Command and Response formats 3 1 1 Host Computer Commands 3 1 2 PGC1 response 3 1 3 Timing of Next Command 3 2 PGC1 Response time 3 3 Local remote control 3 4 Host Computer Command Format 3 5 Status reports 3 5 1 Short status report 3 5 2 Long status report Appendix A Example of a host PGC1 dialogue Appendix B Remote connector 1 INTRODUCTION This manual is intended to be used in conjunction with the PGC1 manual issue 1 3 It describes the serial interface changes in program version 2 2 Some changes to the menus in the PGC1 have been made to accommodate the new interface protocol of V2 2 Up to 8 PGC1 instruments can be connected to a single serial line Line conditions are in electrical accordance with RS232C for PGC1 or RS485 for PGC1F Instruments compatible with RS485 are downwardly compatible with RS422 since the drivers and receivers are connected to separate balanced lines i e no transceivers are used Each instrume
31. e with the high voltage supplies switched off 2 HARDWARE 2 1 PGC1 Line Drivers and receivers The type of line drivers and receivers fitted to any particular PGC1 instrument may be determined by inspecting the model number inscribed on the rear panel If this contains the suffix F then the line interface circuits are compatible with RS422 RS485 Otherwise they are RS232 types Although the Remote Connectors fitted to the rear panel are the same type for the two interfaces ALL INSTRUMENTS AND HOST COMPUTER CONNECTED TOGETHER MUST HAVE THE SAME TYPES OF LINE DRIVERS AND RECEIVERS Driving RS485 receivers with RS232 drivers may cause damage to one or the other The type of line drivers selected will depend on the distance between the host computer and the most remote PGC1 the number of PGC1s sharing the same interface lines and the speed of response desired The speed of response will normally be dominated by the transmission times the latency of the PGC1 and host program should be small in comparison Generally speaking if the maximum distance is over 15 metres or there are more than six instruments sharing the same interface then RS485 drivers should be used 2 2 This section is deliberately left blank 2 3 RS485 422 Line termination If the instrument has option F fitted it has RS422 RS485 drivers and receivers The instrument which is furthest from the host computer should have a termination resistor connected across the Receive
32. earning in a logical sequence which will not result in damage to the instrument or gaugeheads As you progress the amount of detail given is reduced to avoid tedium you should therefore not skip sections and only progress to the next when you understand the current one 3 1 Familiarisation Operation of the instrument is designed to be easy for users familiar with vacuum system operation and with the types of gauges used Brief descriptions of the principles of operation of various types of gauge will be found in appendix A 3 1 1 Power on the LCD switch functions and HELP Connect the instrument to the mains supply Do not connect any gauges or other equipment to the connectors on the rear panel at this stage Switch on The liquid crystal display LCD will show a short sequence of messages and then wait for a response KKK KK KKK KKK RK k k k k kk k k RE k k kk k k kk k k kk k k kk k k kk k kk k k kk kk kk kk kk kk kk kkk 7 WARNING IT IS RECOMMENDED THAT YOU READ THE NEXT SECTIONS AS YOU MAY OTHERWISE INADVERTENTLY CHANGE CONTROL PARAMETERS WHICH WILL BE STORED INTHE INSTRUMENT S MEMORY EVEN WHEN THE POWER IS DISCONNECTED 7 DO NOT CHANGE THE SECURITY CODE WITHOUT j RECORDING THE NEW CODE Si KKK KK KKK KERR k k k k KERR kk k k kk k k kk k k kk k k RE k k kk k kk k k kk kk kk kk kk kk kk kkk There are four switches adjacent to the LCD and those to which the instrument will respond at any time
33. ed between the incandescent filament and the grid structure Electrons oscillate on long paths through the open grid structure being repelled from the central collector and attracted to the grid A proportion of the electrons encounter gas molecules before reaching the grid These molecules are ionised by the collision and are attracted to the collector to form a current which is proportional to the concentration of gas molecules over a very wide range Pressure may be derived from the ion current by solving the equation lon current Pressure Sensitivity x Emission Current where the units for the two currents are the same and the sensitivity is a quoted constant for a particular gaugehead and gas species The impact of electrons on the grid stucture generates soft X rays some of these impinge on the collector and release photo electrons These form a small current in the same direction as the ion current When this photocurrent becomes significant in relation to the true ion current the gauge ceases to function as a reliable pressure transducer and is said to have reached its X Ray limit A 2 Pirani Gauges The Pirani Gauge is a thermal conductivity gauge A tungsten filament in the vacuum space is heated from a constant voltage source and is incorporated in a Wheatstone bridge The electrical resistance of the filament depends on its temperature and this in turn depends on the rate at which heat
34. ed to the new system pressure and does not necessarily indicate that the pressure is the same as before probe gas was introduced via the leak A blocking agent may give rise to a deviation first to the right and then to the left before settling A probe gas may give a single deviation in either direction depending on the relative sensitivity of the ion gauge to the residual gas in the chamber and the probe gas and to the relative pumping rates for those gases If the deviation had been significantly greater than the maximum displayable on the LEDs you should reduce the sensitivity The lower gain should be used if significant random deviations unrelated to probing are observed Further localisation of a leak may be attempted after the deviation has been restored 3 7 lon Bombardment Degas lon gauges may be degassed by ion bombardment of the grid structure The grid is biased to a nominal 500 volts with respect to earth The filament operates close to earth potential Emission is regulated to produce dissipations of up to 50 watts in the grid in order to raise it to red heat 3 7 1 Degas programs Preset programs for degas are available and these may be modified to suit gauge requirements in the configuration parameters under SETUP Programs 1 and 2 allow a stepped increase in emission to be applied at minute intervals until the specified power is achieved in order to reduce the pressure burst That power is then maintained for the rest of the
35. emoved after degas is complete 3 7 3 Problems in degas If FAULT IG underemission trip is reported the most likely cause is that the filament current limit is too low because the gauge needs a higher filament temperature to support more emission current Before adjusting the limit check that normal emission still works If emission cannot be established the grid fuse is probably blown Refer to section 2 4 for instruction on how to replace this fuse You should also check for a short circuit between the grid and filament and the grid and chamber earth as these are the most likely causes of excessive grid current Another possible cause of excessive current during degas is operation at high pressure if this cause is suspected reduce the degas power or its rate of increase in order to reduce the outgassing rate from the grid If FAULT IG overemission is reported there was either a discharge caused by a high gas load or a short circuit between grid and filament If no short circuit is found check that emission at 1mA is satisfactory If this is so check the grid fuse and attempt degas again at a lower power or lower power step size 3 8 Bakeout Control The bakeout control function allows switching of heaters in accordance with system temperature measured by a type K thermocouple and pressure indicated by the ion gauge Take care to connect the thermocouple correctly For further details on this refer to Appendix B 5 It will be necess
36. ent to the fuseholder It must be replaced with a 20mm x 5mm anti surge fuse of the same rating Such fuses are marked with a T after the current rating There is provision to house a spare fuse in the fuse drawer which is integral to the mains connector It is necessary to disconnect the mains supply when replacing fuses 2 4 2 The grid fuse The grid fuse protects the instrument and ion gaugehead against short circuits in the gaugehead during electron bombardment degassing It is a 20mmx5mm 125mA fast acting fuse This fuse is mounted inside the instrument to replace it disconnect the power connector and remove the smaller screws at the sides of the instrument Note that the ventilation slots are at the right side when viewed from the front Remove the cover and locate the fuseholder on the right hand edge of the left hand circuit board The fuse is beneath a clear plastic cover and the rating is printed adjacent on the surface of the circuit board Replace the fuse and its cover Replace the instrument cover and screws in the correct locations Do not use screws longer than those supplied 2 5 lon Gauge Cables The use of correctly constructed cables will enable the instrument to meet its specification The following is a general guide to the considerations affecting ion gauge cable construction a specification of connector types and pin connections is given in Appendix B 2 5 1 The ion gauge power cable The ion gauge power cable must be m
37. equately indicate that pumpdown is progressing by making the characteristic noises Badly calibrated or oil contaminated Pirani gaugeheads may even indicate 5 mbar or less at atmospheric pressure Accurate pressure readings in this range are achieved by capacitance manometers If the Pirani filament should fail the circuit is arranged to give a reading of 1 bar 3 2 2 Starting the lon Gauge Start pumping the system When the system pressure is low enough attempt to start ion gauge emission by selecting St on the menu The Pirani interlock will prevent this if Pirani 1 pressure indication is above 1x10E 2 millibars Pirani gauges are generally more repeatable at the lower end of their pressure range although not necessarily more accurate There is sometimes an adjustment for the low pressure end of the scale and this may need attention Alternatively if you are confident that the system pressure is low enough and the Pirani interlock is preventing the ion gauge from being run then enter the SETUP menu and switch off the interlock If you have succesfully started the ion gauge refer next to section 3 2 4 3 2 3 Problems with starting or running the lon Gauge If FAULT IG underemission trip is reported at startup or on increasing emission this is probably due to the filament current limit being set too low However if the grid is not connected or shorted to ground the same message will be given Prudent operators will check the cables befo
38. er SETUP first Successive selections from this menu allow definition of units of pressure ion gauge characteristics degas program parameters assignment of relays to gauges and functions interlocks bakeout parameters and capacitance manometer range If you have followed the procedure above the LCD should be showing on its top line Enter the security code and the bottom line will show a flashing arrow pointing to the number 200 The default value of the security code is 200 Press Enter Another menu should appear If the same menu and security code is displayed then the code has been changed to another value Since there are 500 possible codes you should ask previous users which value they changed it to If this information is not available consult the Company or agent from whom you purchased the instrument to find out how to restore the security code to 200 It will be necessary to quote the serial number and software version number The top line of the display now shows Change the security code and the lower line shows a cursor underlining no Press Enter to confirm that you do not wish to change the security code from 200 The next menu allows you to define the units of pressure measurement This and the next few succeeding menus are not of immediate interest and can be skipped by pressing Enter which confirms the previous or default selection Continue to do so until the top line of the LCD display shows ION GAUGE o
39. erall screen is recommended and should be connected to the signal ground pin on the connectors There is a 100 ohm 1 watt resistor included in series with this connection in order to limit circulating currents due to ground line differences In order to eliminate ground currents completely you may for example connect the incoming screen and leave the outgoing screen disconnected at each connector on the daisy chain The allowable common mode differences in the RS485 specification is 7 volts and the PGC1 complies with this Connector pin numbers are defined in Appendix B Each used pin should be connected to all similarly numbered pins at every PGC1 and host computer connector on the interface The wiring should be daisy chained from the host computer to each instrument in turn no branches or spurs should be allowed The most remote PGC1 should have a termination connected as described in section 2 3 above 2 5 Host computer line drivers and receivers It is advisable to check certain conditions in the host computer line drivers and receivers even if these are guaranteed to comply with the relevant standards If facilities exist for terminating lines at the receiver this should be done For compliance with RS485 the line driver should also be terminated Consideration should be given to the receiver output state when the all the PGC1s are silent and their lines are in the TRI STATE condition For RS422 RS485 the commonly used DS75176 SN751
40. f you need to change to an alternative style of connector take care to make the connections correctly to ensure operator safety The wires are colour coded as follows Brown Line Blue Neutral Green and yellow Earth B 2 lon gauge power connector This is supplied with the instrument in the accessory pack in the form of a kit of parts It is a 6 pin ITT Cannon Trident rectangular male type In the USA equivalent types are available from Burndy Inc ITT Burndy Cable mounting plug 229 23593J SMS 6 P 1 Hood 229 23597H SMS 6 H 1 Crimp pin 6 required 229 90010L SM 16 ML 11S6 Pin connections are 1 Safety interlock 2 Grid 3 Safety interlock 4 Filament common 5 Filament 1 6 Filament 2 Pins 1 and 3 must be connected together by a short wire within the housing This forms part of a safety interlock which prevents power being applied to the ion gauge connector when the gauge lead is not mated Information on suitable cable construction and installation will be found in section 2 5 above Suitable crimp tools are available from ITT Cannon and Burndy and many other suppliers The connector hood is closed on the cable by a cable tie which is supplied in the kit B 3 lon gauge collector connector This is a 50 ohm BNC type free plug B 4 Pirani connectors Two of these are supplied with the instrument in the accessory pack They are 3 way DIN audio free plugs with screen Types with or without latches may be used Pin connections are
41. hould be twisted together to avoid magnetic induction and routed away from power cables B 6 Remote connector This is a 25 way D type female connector Used pin connections and signal names are 3 Received data 2 Transmitted data 5 Signal ground The signal levels on pins 2 and 3 are compatible with those of the RS232C standard However the signals may be connected to up to eight of these instruments which is not catered for in the standard B 8 Auxiliary connector External connections should be made to a free 8 pin DIN audio connector with screen as supplied in the accessory kit Types with or without latches may be used Connections for the recorder output are 1 Signal voltage 6 Signal return earth Connections to the capacitance manometer are 2 15 volt supply 3 Signal voltage i e manometer output 4 Signal return and supply 0 volts 5 15 volt supply Connections to the remote inhibit lines are 7 Not Inhibit 8 Signal return logic ground 20 21 ARUN MICROELECTRONICS LTD PRESSURE GAUGE CONTROLLER MODELS PGC1 and PGC1F INTERFACE MANUAL ISSUE 2 For use with Program Version 2 20 onward COPYRIGHT RESERVED 1998 The information contained in this manual has been carefully checked and is believed to be correct No responsibility is assumed for any errors User comment and criticism is welcome please write quoting the serial number of the instrument and the version number of the software as displayed by th
42. local mode the instrument responds only to commands without parameters lt poll gt lt control gt lt long gt lt short gt and lt reset error gt The lt control gt command puts the instrument into remote mode and all the other commands can then be used The front panel can still be used to change the display but not to start gauges or change setpoints When a host takes control of the PGC1 emission is stopped and any current Setup operation is cancelled When the host returns the instrument to local control emission is again stopped 3 4 Host Computer Command Format Gauges within the instrument are addressed by a numeric character found from the status report In a PGC1 instrument the ion gauge is numbered 1 the two pirani gauges 2 and 3 and the capacitance manometer gauge by 4 Relays are addressed by uppercase letters starting with A A PGC1 instrument contains relays A to D Command parameters may be single printable ASCII characters or ASCII strings terminated by a delimiter Numbers in scientific notation must be sent as ASCII strings of the form 9 9E 99 other ASCII strings can be of any length including zero The error byte should be checked after sending a command to ensure that the command has been understood and carried out In the following tables the different parameter types are called Char single character no terminator Value ASCII string with terminator and SN Value ASCII st
43. n the pressure exceeds twice the setpoint pressure Adjustment of setpoints is described in section 3 5 3 If a relay is assigned to TSP that relay is energised momentarily to trigger a titanium sublimation pump controller into renewal of the getter layer The time between energisations is related to current pressure in accordance with the following table pressure exponent mBar delay time gt 6 infinite 7 10 min 8 30 min 9 90 min 10 10 hours 11 infinite Refer to the manual of the TSP controller for information on how to use the momentary contact closure in some cases a voltage source is necessary A relay assigned to Bakeout is energised to switch on the bakeout heaters More information on this is contained in section 3 8 below Relays which are not assigned to a gauge or other function should be assigned to None These relays may be energised or de energised by the operator while measuring pressure and may be useful in various ways 3 5 3 lon gauge trips After assigning the relays you are asked to specify what state the relays should be in when the ion gauge is not operating Normally these relays are only energised when the pressure is below the trip level It may be useful however to specify En when the gauge is not operating taking into account that if more than one relay is assigned to the ion gauge that they will be all be affected After assigning the functions of the relays progress to the end of the SETUP
44. nt is identified by a unique address code selected from the numbers 0 to 7 inclusive The address code is shown momentarily on the main LED display each time the instrument is switched on Instruments connected to a single serial line are allowed to send characters only when addressed by the host computer Either a short report giving current measured pressures or a full report of the status of the PGC1 can be requested Current pressures are reported in filtered form Because the transmission rate is high and the quantity of data low the host computer can poll all instruments faster than any significant event in the vacuum system The response time of an instrument to a host request is generally less than 1 millisecond Remote control is established and relinquished via the serial interface Either an individually addressed instrument or all instruments may be switched to remote or local control by a single command The status of any instrument is unchanged in every other respect after a change in control status The host computer can control Trip setpoints and relay status lon gauge operating conditions in pressure measurement modes The content of the LCD display The sound generator and all commands may be directed to a single gauge or instrument or all instruments At switch on or after a reset due to derangement of the embedded program or an external reset via the Auxiliary connector an instrument is reset into the local operation mod
45. ossible without the presence of a host computer or gauges in order to become familiar with the structure of the control menus help and error messages If you become lost in the menus you can escape either by switching off for a short period or by selecting options such as STOP and EXIT when these are presented Sometimes you will have to complete a sequence of selections which you have inadvertently entered Pressing ENTER in any menu without moving the cursor confirms the previously selected choice and is an easy rapid way of skipping through a long sequence of choices where no change is required For ease of use the menus are arranged so that after swiching power on two presses on the ENTER key will start pressure measurement with the ion gauge Since completion of any sequence returns to the RUN SET REMOTE menu over enthusiastic pressing of this switch will cause the gauge to start 3 1 4 Preparing to measure pressure Setting the lon Gauge parameters Enter the SETUP menu Set and record the required security code and the units of pressure measurement The gauge sensitivity may be entered in units other than the units of pressure measurement press INC until those you desire are displayed and then press ENTER Adjust and enter the numeric value according to the gaugehead manufacturer s specification Select the filament type This adjusts the limits on maximum allowed filament current and degas power and alters the
46. re increasing the allowed filament current Disconnect the mains power connector before checking any cables The filament current limit adjustment can be made at any time but in order to view the limit you must enter the SETUP menu as described in section 3 1 4 If FAULT IG overemission is shown there is probably a short circuit from grid to filament Disconnect the mains power connector before checking any cables If Fault IG filament open circuit is shown the cable or filament is probably open circuit Disconnect the mains power connector before checking any cables In some cases if there is a high resistance connection or the cable is long or of inadequate cross section this fault may be reported after some period of normal operation To change to the other filament return to the initial menu under RUN and select Other and then Fil 3 2 4 Measuring pressure with the ion gauge Emission is initiated at 0 1mA Review the formats of pressure display available by pressing the DISP switch and then increase the emission current to 1mA and then 10mA by the appropriate selection from the menu This may have caused Fault IG underemission trip to be reported In this case the filament current limit may be set too low to support the increased emission The filament current limit adjustment can be made at any time but in order to view the limit you must exit the RUN menus and enter the SETUP menu as described in section 3 1 4 When
47. ring with terminator in scientific notation form All pressure values read from the PGC1 or sent as commands must be in the PGC1 s current display units lt poll gt Instr Poll instrument return status and error byte P Remotely control an instrument lt release gt R inst Return instrument to manual control lt reset error gt Reset all error flags i lt control gt lt short gt Instr Request a short status report from an instrument reporting operating status of all gauges Request a long status report from an instrument lt long gt lt Gauge on gt Instr N Switch on ion gauge Emission current N 0 100uA N 1 1mA N 2 10mA N 3 auto emission control lt Gauge ote fo ins Swichoffion gauge eee Instr SN Set the ion gauge overpressure trip Value lt filament gt cflament gt ft instr Select flament F or 2 0 1 or 2 lt filter time constant gt s Instr T Select filter time constant where T O off 1 2 4 or 8 seconds lt setpoint gt r Instr Relay Change a relay setpoint pressure SN Value lt override gt Instr Relay Permanently energise relay Command lt setpoint gt restores normal operation lt inhibit gt Instr Relay Permanently de energise relay Command lt setpoint gt restores normal operation lt display gt Instr Text Displays message Text up to 24 characters on the PGC1 LCD display A blank message clears the display
48. s to that decribed in section 3 5 4 above 3 6 Leak Detection The leak detector works by indicating rapid changes of pressure on a bargraph and optional frequency modulated tone Such changes are produced with a probe gas or volatile blocking agent It is difficult to exercise this function unless you have a leak valve or a leak on the system although you may be able to produce a pressure burst by switching on other equipment or firing a titanium sublimation pump During leak detection the TSP program is prevented from triggering the TSP but the ion gauge overpressure trip and process control relays function normally Select Lk on the RUN menu and then select which gauge is to be used This will depend on the pressure at which the leak is limiting pumpdown If you select the ion gauge the emission is set to 1mA and the interlock with Pirani 1 still functions The LED display now shows a histogram of rate of pressure deviation with baseline restoration and the LCD menu allows you to change its sensitivity by selecting Gain and to switch on the tone by selecting Sound If you have means to simulate a leak you should do so and observe the effect on the display and tone After a pressure change has been observed turn off the probe gas supply or stop applying the blocking agent and wait until the baseline restorer has caused the histogram and tone to return close to the centre of their range This shows that the leak detector has adjust
49. terference which can be induced is proportional to its length and can become significant at extreme UHV Site the cable run away from other cables carrying high power or high frequency signals The cable installation should be such that movement or flexing is discouraged Mechanical movement of the cable can generate triboelectric charges which may affect UHV measurements It is important to avoid the production of earth loops in sensitive signal return paths as these can have significant currents induced in them at mains frequency The screen of the collector cable should ONLY be earthed at the instrument end If existing cables are to be adapted and are currently connected to earth at the gauge end disconnect the screen but ensure that the guard cylinder around the collector feedthrough on the ceramic gauge base remains connected to earth This is important as it provides a path to earth for leakage on the surfaces of the ceramic from the grid voltage which would otherwise add to the collector current 3 OPERATION The number of ways this instrument can be used is very large but the underlying structure to the control of its operation is very simple and consistent The instrument will offer helpful prompts and explanation if the HELP switch is depressed For these reasons this section of the manual is written in tutorial style Following the procedure below will ensure that the instrument is exercised in a way which will result in the user l
50. the relays which may then be changed by selection of Relays More information on this is contained in section 3 5 below Selection of Q quit will restore the former display 3 3 Operation at extreme UHV and emission current selection You may select 0 1 1 or 10mA of emission current in the ion gauge as the gauge is being operated or allow the instrument to adjust to the optimum current by selecting AUTO For the vast majority of measurements emission of 1mA will be used Although most gauges are substantially linear and the instrument regulates emission current well maintaining 1mA emission wherever possible will eliminate some error in the readings 10mA emission is used at the extremes of UHV and the instrument will advise when this is desirable because the limits of the dynamic range of the electrometer at its current ambient temperature are being approached If there is some fluctuation of the ion gauge pressure readings at very low pressures these may be reduced by increasing the time constant of the filter To do this STOP the ion gauge select Other then FIL Select a time which just exceeds any observed periodicity in the fluctuations or the minimum time which reduces them to an acceptable level The trips and trend indicator operate on unfiltered pressure readings so their response times are not affected by the filter Automatic control of emission current ensures that the dynamic range of the ion gauge is extended
51. troducing a cold instrument into warm conditions If this situation exists it will normally be self correcting after a few hours of operation in non condensing conditions 3 4 Operation of the ion gauge at high pressures For operation at the upper end of the ion gauge pressure range an emission current of 0 1mA is available This should not be used as a means of extending the upper limit of operation as the filament temperature is only slightly reduced and oxidation and other reactions are not significantly retarded Iridium filaments are more suitable for relatively high pressure measurements with lon Gauges 3 5 Process Control and Trips Four process control relays are provided which can be used in many ways The relays and their contacts are described in the following paragraph and their assignment and use in the remainder of this section 3 5 1 Use and wiring of the relays The contacts of the relays are shown diagramatically on the rear panel legend in the de energised condition adjacent to the terminal blocks to which they are connected Switching loads of more than 5 Amps is not recommended Bakeout heaters will require an external contactor for switching Inductive loads including contactor coils should have snubber networks connected in parallel to avoid arc generation which could interfere with the operation of this and other equipment In general avoid introducing noise sources into the instrument via the relay contacts
52. uge record Pirani Gauge 2 41 40 running 1 0E 03 31 2E 3045 at atmosphere 2B 30 33 2C 8D 38 44 Checksum OD 0A End of report riot e 2A 45 31 Host acknowledges and clears Instr 1 error flag Instr 1 4 34 40 Instrument clears its error byte and indicates CR LF readiness to 4 lt anew command Host o 2A 6E 31 Host turns off Host turs offgauge s Instr 1 4 34 40 Instrument 1 status byte No error CR LF OD 0A 14 Appendix B Remote connector This is a 9 way D type male connector Used pin connections and signal names for RS232 are 3 Received Data 2 Transmitted data 5 Signal ground If option F is present as indicated by the suffix F in the model number on the rear panel then the pin connections and signal names are compatible with RS422 or RS485 as follow Received data Received data Transmitted data Transmitted data Signal ground Protective ground 00 ON D 15
53. uges and measure pressure 3 2 Pressure Measurement Before connecting the instrument to any gauges you should perform the suggested familiarisation procedure described above As an absolute minimum you must have completed the procedure in section 3 1 4 as failure to do so may result in destruction of the lon Gaugehead 3 2 1 Pressure measurement using Pirani Gauges Switch off disconnect the mains power lead and connect the gauges It is assumed that Pirani 1 is protecting the ion gauge ensure that the correct lead is connected to the Pirani 1 connector on the rear panel Reconnect the power lead and switch on Pirani gauges and capacitance manometer are powered up immediately The left section of the LED display shows a histogram of Pirani pressure at all times This should be read against the scale at the top or bottom as indicated by the line to the illuminated pressure unit indicator LEDs at the left end of the main section of the LED display Pressing the DISP switch will change the format of the display and the source of the measurement sequentially P1 and P2 are Pirani gauges and C is the capacitance manometer Pirani gauges are not very accurate or repeatable transducers at pressures close to atmospheric For this reason the instrument has no discrimination of pressures between 1 bar and 5 mbar This is of no practical consequence since a typical rotary pump roughed system will traverse this range in a few seconds and will ad
54. verpr trip Note that the lower line shows the overpressure trip in scientific notation in the selected units of pressure measurement The default value is 1 0E 02 millibars A flashing arrow is pointing toward the exponent part of the display This number can be changed by pressing either the INC or DEC switches Note that if you keep the switch depressed the number continues to change and that the rate of change increases after a short while When the allowed limit value has been reached the number stops changing and the green light in the switch is extinguished You must then press the other switch to change the number in the opposite direction Return the exponent to the default value and press the ENTER switch You are now asked to change the mantissa in exactly the same way Note that the mantissa overflows into the exponent at the end of each decade This is the general procedure for entry of all scientific format numeric values into the instrument You should now progress to the end of the setup sequence of menus by pressing ENTER repeatedly until RUN SETUP REMOTE is displayed Alternatively if you have become lost switch off for a few seconds 3 1 3 Exploring the menus By this time you will be able to progress through the menus and use the help messages with facility although you are advised not to change any parameters whose significance you do not fully understand Progress through the RUN and REMOTE menus as far as p
55. you have achieved 10mA emission you should then turn the ILIM adjustment slowly anti clockwise just until underemission is reported View the limit in the SETUP menu and then increase it slightly to give a margin of safety The setting may also need to be adjusted if the gauge requires more power during ion bombardment degas at relatively high powers If the ion gauge or filament is changed the adjustment will need to be repeated It is worthwhile taking the trouble to find the optimum setting of this control as this will improve the protection and longevity of the filament lon gauge operation may also be initiated by selecting Au autostart from the first menu displayed after selection of RUN This starts the ion gauge during pumpdown when Pirani 1 pressure indicates less than 1x10E 2 millibars The autostart function is allowed to function only once after selection and the gauge will not be restarted if the ion gauge overpressure trip is subsequently temporarily exceeded The overpressure trip is independent of the Pirani interlock pressure During emission the Pirani pressure is indicated in numeric format on the top line of the LCD The status of the four relays is shown to the right of Relays on the bottom line Four digits represent relays A B C and D from left to right 0 represents the de energised condition and 1 represents the energised condition The entire display can be replaced by a display of the status and trip pressures of