Home
        User Manual
         Contents
1.      button         p Move to the upper menu  Input MENU       Default 1 5000    b  SCALE input     Shortcut key  W  7999  meee    yes  o  o scum        Enter the scale value of the input sensor by using the buttons      After setting up a value      39999    is changed by pressing  mem  _ser   button           E   Move to the upper menu  Input MENU           Default 1000 0    3 3  Auto calibration   Shortcut key  MB  S ms   gt  mme 7  8  oy rese   gt  aa     After permitting the load  weight  on the sensor  input it by  using the buttons    After setting up the value  BE is changed by pressing    mT  key      E M Move to the upper menu  Input MENU              W    Y Set up the display unit     Moving      Moving   Average    Set up the number of movement average   uo 0e         T I    Diaital    Setup the digita shit   Shift Filter  Analo  inalog                            i      Set up the analog filter        12    4 1  Unit  Unit setup    Shortcut key  ope y ves   p mt   5  gt  Fase     x22  9      Set up the display unit corresponding to the display value      wp o LE o ger  E          Move to the upper menu  Input MENU   EESN EH            Set up the unit by pressing EH Ej keys    Pressing _ser   key  Mmmm is changed        Default NONE  Select 79 kinds     5  Input setup  4   8      Shortcut key  222    gt  m  Lee   DER  sme ao h      iain   io       Designate and display the display range     When it is out the range  display the digital limiter value            Limi
2.     10    2 1  Pattern Select  Shortcut key  ops   Ls   see  5 O Les      Select the Pattern  POO P15  to use     Select the wanted pattern by pressing al E keys     Pressing _ser   key       29      is changed           E I  Move to the upper menu  Input MENU          Default POO   POO P15     2 2  Pattern Copy  Shortcut key  ops   LES   p m     y ALS       Select the Pattern  POO P15  to copy     Select the wanted pattern by pressing EI Ej keys     Pressing sr   key         9    js changed      2M    Move to the upper menu  Input MENU          Default POO  POO P15     3  INPUT Setup  2   8    Shortcut key  ove    gt  zs     gt  meee   BE   EN Pase  gt       i Manual ij       2 al A user enters the characteristic values of the    Zera Auto  Galibration Gallbration    Manual  calibration    corresponding sensor while looking a data sheet        Auto      _gailtration      Put the actual load  weight  which can be    guaranteed  calculate the calibration value  and save it     3 2  Manual calibration   Shortcut key  399   E Sp omm   S  s pases  y aE         S lp Enter the sensor characteristic calibration value     8E   Enter the sensor characteristic scale value          E p Move to the upper menu  Input MENU            3X  Move to the MAIN MENU    11    a  CAL input     Shortcut key  ons   s  s    gt  ee   ES   _    Pase        calibration    gt  CAL h    Enter the output value of the sensor by using the buttons      After setting up a value  is changed by pressing          
3.    ES    n case the comparison is selected as the Reference  the  Replay performs OK and NG according to the reference setting  Relay 1  CE ECI p    Output Type   For the left figure  the Relay 1 outputs ON when it is bigger    17005  905  RefHigh I Ref Low than the setting reference    g      The Type selects High On  Low On  Range  and R Over  Range Over           As the output type  the Output Type sets up Fast  Display  Decide  Graph End  M D etc     Ref High and Ref Low sets the hysteresis value  apply the hysteresis that the each value  makes the relay ON and OFF  and remove the relay ON OFF repetition phenomenon due to the    minimal change of the measured value     4 3  Mode setup menu       EN    n case of the Relay Channel Mode  select the comparison  uem sp method of the input source          Default  Display   Display  Graph  Source     Source    18      Display  The standard value is regarded as the value displayed on the screen after comparing  the relay     Graph  The standard value is regarded as the graphic output value after comparing the relay       Source  The standard value is regarded as the input source after comparing the relay     4 4  Type setup menu      In case of the limit mode  select the output type   The output type selects it between HH  HI  OK  LO  and LL       1   A y  BH Select an item by pressing       keys     Ew  Pressing      key after selecting the item  MEE is    displayed and the setup is completed        Limit mode    SELECT  NL CMM
4.  SET     Pressing           key after selecting the item  ME is       n case of the reference mode  select the output type   The output type selects it between High On  Low On  RANGE     and R OVER       Select an item by pressing     Y  keys        displayed and the setup is completed     Reference mode    4 5  Output type setup      Select the output type               Output type selects it between Fast  Display  Decide  Graph    Pow o End  and M D     Select an item by pressing        keys       Pressing       key after selecting the item  BEEN is  displayed and the setup is completed    Output Type    Fast  Synchronize with the sampling time     19    Display  Synchronize with the screen output    Decide  When inputting the decide signal  output a result    Graph End  When finishing drawing a graph  output the graph end signal and relay at the same  time     M D       1005     905  4 6  Ref High and Ref low setup _ReHigh   RefLow      Input the Ref Hysteresis value     The Ref High sets the bigger value of the hysteresis and the    Low sets the smaller value     HHHigh   HHLow       set up the hysteresis value of HH on the limit          mode   The hysteresis can be set up according to the each item of HH  HI  LO  and LL   5  Relay Ach reference setup  10   11       Input the reference of Ach        T  lw   set up the hysteresis value of HH on the limit       mode       The hysteresis can be set up according to the each item of       HH  HI  LO  and LL     6  Relay Bch r
5.  as 0x00 OxFF      The data comprises the index and actual values of each item  and index  2byte  and 8 byte or       the data of 2 byte     Command Code  R  Request the current value   PC     DN GI200    D  Transmit the current value   DN GI200   gt  PC     7 2  Current Value Request  PC  gt  DN GI200     Example  Device ID  No  1  channel No  1 data request 40102R0100D6 L    Length Channel Checksum ETX       Iu NE ANC  eS ee ee      The ID and channel is displayed as OxOO OxFF which is the Hex value of 07255      When requesting the current value  the data is Index 00      20    7 3  Current Value Transmission  DN GI200     PC     Example  Device ID   No  1  channel  No  1  data value  492 0    010AD0100 00492 0010327 L     Data  channel  and current value  01 00492 0         STX  ID   Length  Code Channel   Index Data Hold Peak  Relay ChecksumETX            10 1 O0 E  D  O0 1 0 0    000 492     0  0 1 0 3   2 744                                                                                 When transmitting the current value  the data comprises Index  00    display value      00492 0    Hold Peak  01    Relay  03     29       Example  DN GI200 graphic mode setting      In case the process result screen is the same as below on the graphic mode  the infection    point and relay setting is proceeding with the below order     1  Test Measure    p n 40 00kgf ZERO MODE  Y    120 00    PEAK mm  4    HOLD  0    X  150 00    0    2  Measured Waveform Analysis       According to t
6.  memory activation key    10 00    0      guason  y cursor display key           E    cursor moving key        GE l all sorts of data input mode conversion key           8IXIT     GRAPH START key    3  Caution screen  Shortcut key  Z        When entering the MODE  it alarms the caution     Pressing   388    enter to the MODE       Pressing  39     return to the MAIN MENU        6 2 MODE Screen    1  Mode screen  1   2   Shortcut key  ove   _5 wes          Pressing  2    move to the MAIN MENU    E cet up the value of Ach     ww e     es    Set up the value related to the relay output     C  Set up the hold function        Page Page    peer        5ew     set up the function of a SD memory   2  Mode screen  Shortcut key 39      x9  p         Se   Set up the basic function     ZI up the graph                 Waveform    u SYStlel     n    EH Set up the basic function 2           6 3 INPUT  1  Selecting INPUT channel  Shortcut key  Mone     gt  ss   y eee       When setting up the INPUT  this is an Ach  Bch selection screen             Move to the upper menu  MODE   _   ENT  Move to the MAIN MENU    u Ach        O   Select the Ach channel        2  INPUT setting  1   8   Shortcut key  2922   gt  m   apum   p 3        ES           Move to the upper menu  select the INPUT channel     cono B oo B     FN Move to the main menu   E       ee      Select the saved pattern   Pattern       PES p Copy the saved pattern     Bensor      ovr      Display the sensor voltage     Pattern  Copy   
7. 2 OUT RY2  A OK  15          3 OUT RY3  A LO  16          4 OUT RY4  B HI  17          5 OUT RY5  B OK  18 INPUT PO  6 OUT RY6  B LO  19 INPUT P1  7 OUT RY7  BUSY  20 INPUT P2  8 OUT RYS GRAPH END  21 INPUT P3  9 OUT COM N24 PLC  22 INPUT      10 INPUT GRAPH 23 INPUT HoldSave SD   11         24 INPUT MORY SD   12         25 INPUT Graph Hold  13         26 INPUT P24 PLC   5 2 DIO HRS 40Pin Pinmap  HRS   TML HRS   TML  I O FUNCTION I O FUNCTION  NO NO NO NO  1 A1   OUT COM N24 PLC  2 B1 INPUT P24 PLC   3 A2 OUT RY8 GRAPH END  4 B2 INPUT Graph Hold  5 A3 OUT RY7  BUSY  6 B3 INPUT MORY SD   7 A4 OUT RY6  B LO  8 B4 INPUT HoldSave SD   9 A5 OUT RY5  B OK  10 B5 INPUT                                                                                                                                         11 A6 OUT RY4  B HI  12 B6 INPUT P3  13 A7 OUT RY3  A LO  14 B7 INPUT P2  15 A8 OUT RY2  A OK  16 B8 INPUT P1  17 A9 OUT RY1  A HI  18 B9 INPUT PO  19 A10         20 B10 INPUT GRAPH  21 A11         22 B11          23 A12         24 B12          25 A13         26 B13          27 A14         28 B14          29 A15         30 B15          31 A16         32 B16          33 A17         34 B17          35 A18         36 B18          37 A19         38 B19          39 A20         40 B20          5 3 DIO HRS 30Pin Pinmap  HRS   TML HRS   TML  I O FUNCTION I O FUNCTION   NO NO NO NO   1 A1 IN COM P24 PLC  2 B1 IN COM P24 PLC    3 A2 INPUT PO 4 B2 INPUT P1   5 A3 INPUT P2 6 B3 INPUT P3   7 A4 IN
8. PUT     8 B4 INPUT HoldSave SD   9 A5 INPUT Memory SD  10 B5 INPUT Graph Hold  11 A6 INPUT GRAPH 12 B6 INPUT Graph Save                                                                                     13 A7 INPUT     14 B7 INPUT      15 A8 OUT RY1  A H H  16 B8 OUT RY2  A HI   17 A9 OUT RY3  A OK  18 B9 OUT RY4  A LO   19 A10 OUT RY5  A L L  20 B10 OUT RY6  B H H   21 A11 OUT RY7  B HI  22 B11 OUT RY8  B OK   23 A12 OUT RY9  B LO  24 B12 OUT RY10  B L L   25 A13 OUT RY11  BUSY  26 B13 OUT RY12 GRAPH END  27 A14         28 B14          29 A15   OUT COM N24 PLC  30 B15   OUT COM N24 PLC   5 3 Time Chart  1        1100ms 0    100ms 0        it    BEES    MEMORY SD                 en E 2    GRAPH      BUSY OUT        5 4 INPUT OUTPUT Circuit Diagram    P24    INPUT COM       INPUT         OUTPUT                                                        A    Inside of meter    me  gt  OUT COM       6  Name and Function of Display Screen    6 1 Main Menu    1  SINGLE display screen    _   mars   display conversion key  GRAPH SINGLE     Display on  amp  off state of  ER     3  20           relay    Hu   output  7    Pressing the corresponding key  it is possible to set up a  value immediately      Ea  digital zero      HOLD PEAK operation status monitor        When changing the operation status Homm  BEHK     2  GRAPH display screen       GRAPH   y     T   wonc       assi display conversion key  GRAPH SINGLE   200 00 Ach   BN   digital zero key  PEAK  HOLD    0     80839  l  data SD
9. User Manual    Model  DN GI200     3   AD ba r 1 L palig ATT a      INDICATI    5      ia    LITE    DACE Li       DACELL CO  LTD     Address   681 1 Cheoksan  Ri  Nami   Myeon   Cheongwon   Gun   Chung Buk Korea   TEL   82 43 260 2242   FAX   82 43 260 2245   nttp   www dacell com   E Mail   info  dacell com    Thank you for choosing our product    The User Manual explains how to use and how to be careful of the product  Wrong handling  may cause safety accidents and damage of the product  Therefore  you must read the user  manual before using the product and use it correctly    If any issues occur while you use the product  refer to the user manual or contact the technical  department of our company      Details of the user manual can be changed without any notice according to a version of the    product     1  Characteristic    The product amplifies the output signals of various sensors  converts them into digital and  then indicates them as graphs on a LCD  the graph data are saved on a SD memory and    analyzed in a PC  Details of the characteristics are as below     1 1  A D Conversion Method  A D device is adapted which is converting the input signals from a sensor over 1000 times per    second     1 2  Calibration Method  Two methods are available that one is an auto calibration by an actual load  standard weight     and the other is a manual calibration by inputting the rated input value of a sensor     1 3  Various Input  amp Output Functions   The internal paramete
10. dication  Comply with the digital scaling    Input calibration rage  0 1  3 5mV V  CAL    Display  load  rage   99999   499999   Nonlinearity  Within  0 02  FS 1digit  When inputting 3mV V   23 C 5 C     Analog filter  Select it between 10  100 and 1000 Hz      3 2  Input Section  B ch     3 3  General Specifications   Backup  The setting data  data zero value  amp  display screen  is saved in a flash ROM   Data setting method  The operation of each setting menu and touch panel   Power  AC100   AC240V   10  50 60Hz     Consumption power  About 50VA  maximum     Dimension  96mm  W  x 96mm  H  x 160mm  D   Used temperature humidity range  0 40 C  35    85  RH  non dew condensation     Weight  About 1 0kg    4  Front and Rear Panel       GRAPHIC INDICATOR       UTWCELL  gr DN GI200       4 1  Front Panel 4 3 Mounting drawing     1  Touch panel LCD  Display  amp  display the data        D SD memory  Save the measured data              4 2 Rear Panel     3  DIO     4  232 communication panel  COM  TXD  RXD            5  Sensor input terminal CUTTING   6  CAL  FILTER SIZE    92     7  Ach digital input   Sensor input terminal  Encoder      9  Bch DIGTAL input   Power input terminal  AC 907220V      11  GND terminal    5  DIO instruction    5 1 DIO D SUB 26Pin HD Pinmap      Add the hold data save input and change the input pin                                                                                        PIN NO I O FUNCTION PIN NO I O FUNCTION  1 OUT RY1  A HI  14          
11. eference setup  10   11       1005     905  ENSEM noh   HHLow    LOHigh         95  omg rcm ora c3 us    Reset   The hysteresis can be set up according to the each item of      Input the reference of B Ch        set up the hysteresis value of HH on the limit          LL High    HH  HI  LO  and LL     20    6 6 Hold    1  Hold  1   3   Shortcut key  MENU   ERU     gt  iin      Pattern       Select        Select the saved pattern     Pattern    Pattern      Copy   Copy the saved pattern   Select Level    Start      Pattern Start     Condition   Set up the starting condition   Copy Condition    _    Type    em   Set up the Hold type        1 1  Pattern Select    Refer to 6 3 Input Pattern Select   1 2  Pattern Copy      Refer to 6 3 Input Pattern Copy       Start    1 3  Start Condition Condition    SELECT  EN CEN  E   This is not used currently     Riging Edge  Trailing Edge      Set up the Hold Start Condition        1 4  Type Type      ES   Set up the Hold Type       The Hold type can be set up as below      Normal    Sample       Area Peak    AreaValley   AreaPeakVall    P P D        Time   Time Valley   Time Peak Vall          Level TimeP   Levels Time V   Lev TimeP V       MaximalValue    MinimalValue    Infection P     zl    2  Hold  2   3     Eg   The menu is not used currently   Detection Detection    Minimal  Value  Detection       3  Hold  3   3     Inflection    pos      Range b Set up the inflection point detection range     Inflection    lecto       B  we      Set 
12. etection  set up the output for a result           n order to distinguish the measured result between OK and NG  use the inflection point  value and X axle maximum value     From the replay setting  decide Ach by the standard of the load value of the inflection point    and decide B Ch by using the deflection maximum value     od    
13. he measured result value  select the detecting method to decide OK or NG  In    case of the above  decide it by using the inflection point detection         r3    Interval ChA Start   Inflection Inflection  ime Level Condition   Range Start    Start Type ChB Start Inflection  Leve  Condition       Before  Inflection          n case of the above  the inflection point detection is done by shown as the above  and the    graph start condition is set as E S  D S      n addition  set up ChB Level on the GRAPH  4 6  screen as 100 00  Start Condition Riging  Edge and begin to measure from the point that the value of X axle is 100 00       For the inflection point and start condition  decide the proper value by repeating 1  Test    Measure several times     30    3  Inflection Detection    Detecting the inflection point according to the setting of 2   the screen pops up as below     GRAPH 40 00kof ZERO MODE      Abch     ch    PEAK    HOLD  0    0  105 200m      10 33kgf  105 20mm on the middle of the bottom screen is the inflection point detection       location  and they are displayed as green X on the graph      120 00mm of the right center indicates X axle maximum value which represents the  maximum movement distance    3  Detection Waveform Screen Setting   In order to make the inflection point and start condition setting more clear  change the start    points of XY axles and end point     E  n 40 00kgf ZERO MODE      Abch     100 00    s       When completing the setup for the d
14. inflammable gas or steam  and dust must not be avoided for installation   e Ground terminal  GND  must be grounded    e Make sure that the wiring of a wattmeter or with lots of noise should be wired separately    e For an analog output  surely connect a load over 10KQ without fail and do not connect an  inductive load    e For a sensor cable  you must use four wire two pair shield cable      In case of the wiring with lots of noise  it must be wired separately      2 2  Cautions for Use  e As the product is a precision instrument to measure physical quantity  be careful of having an  effect of noise     If required  select an optional noise block method  Main noise source     wiring of a  wattmeter  wireless equipment  electric welder  motor  inverter etc    e Do not remodel the product for other purposes other than measuring   e For the no load condition and actual load inputs  wait until it is stable without wobble and  then input them   e Pressing a touch key under unstable condition causes a calibration error   e Do not press the touch key arbitrarily in use  and refer to the function and input method of    the touch key  5  Name and function of display screen     3  Specifications   3 1  Input Section  Ach    Sensor power  DC10V  5V  2 5V  10  output below 120mA   Conformance sensor  Various sensors  four wire system  of a strain gauge type   Parallel connection is available for 3500 strain gauge type sensor up to 4 units    Input signal range   3 5mV V      3 5mV V   In
15. on the figure  it is converted to the key input    D screen and the values of each point can be set up     PEAK PL  HOLDM  0    0       Zo    3  GRAPH  3   6     Interval  al      Time Set up the graph start delay time   Graph Disp       Sete b Graph start method    Interval  Time    Free Run  Repeat the graph continuously     Start Type      EN Ext  Signal  Get started by the external start signal        Start Level  Start the graph by the set level   E S  W S   Start the graph by the external signal and setup level of Ach   E S  D S   Start the graph by the external signal and setup level of B Ch   E S  W S  DZ  Start the graph by the external signal and setup level of Ach and reset it when    starting B Ch distance value     4  GRAPH  4   6     ChA    pn Level     Set up the start level of Ch A   Graph Disp EN j  ChB       Lee   Set up the start level of Ch B   Level Condition Start      Condition    Set up the start condition of each channel     ChB Start  Level Condition      Start condition       Pass  When passing the set start level    Riging Edge  When passing up the set level   Trailing Edge  When passing down the set level   More  In case of bigger than the set level     Less  In case of smaller than the set level     5  GRAPH  5   6     Ey   When the hold type is the area mode on the graphic screen   EJ MENU    set up the comparison range   Area High Area Low   Each item is indicated as the coordination values of the  ChB    rm graphic screen and the digits below 
16. or  Relay keys     E  Pressing       key  MEME is displayed and it is set      e       ul           Move to the upper menu  Input MENU     Ew  SIEF T T I    T        Limit Mode    Relay Mode     3  Relay operation setup  2 9   11     EN nuca   Select the input source  Ach  Bch  Signal  Off   u Mode        ECH ECH     Select the relay output mode  Display  Graph     Source       ciao   Set up the relay comparison method        a es   Set up the relay output type     a OutputType             Set up the output speed      HHHigh   Hilo   Set up the upper and lower limit values for the comparison       Each separate setup for eight relay units is available     3 1  Input setup      Select the input by pressing       keys       Select it by pressing the key between Ach  Bch  LOAD  amp  DIST     E and Signal     Bch     Pressing       key  M is displayed and set      ut            Move to the upper menu  Input MENU     Default Ach  Ach  Bch  LOAD  amp  DIST  Signal     16      In case of selecting Ach or Bch  set up each relay in accordance with 4      LOAD  amp  DIST does not have a setup       In case of the Signal  refer to 3 2  as below     3 2  Input signal    EN   In case of setting up the input as the signal  the relay can  E      gl    output the internal setting signal     The output signal is SD Busy  Graphic End  Sync etc       Select the output signal by selecting the mode menu      Select the input by pressing       keys     Select it by pressing the key between Busy  G
17. r can be set up by using the touch of the front panel and the auto zero  function can be performed    On the rear panel  RS 232C connection terminal  relay contact output terminal   graph memory pattern select hold zero input terminals are available    1 4  Replay output function   Display the measured the analog input  and output the result comparing the set HH  HI values    and LO  LL values to the relay contact signal     1 5  Data Preservation Function  In case of a blackout or the electric power is cut off  the inputted data is preserved because the    setting values such as zero  scale  relay etc  are stored in a flash memory device     1 6  Display Method  amp  Control Function of Response Speed  Two types  digital filter and analog filter for a sensor input are adapted in order to obtain the    required response speed according to the digital display and purpose of the front panel     1 7  Various Data Output  As an option  the analog power output   10V  BCD  BIN output  serial interface  RS 232C  4     20mA  and 0 20mA analog current output can be used     2  Standard built in Product  1  Main body    2  Instruction manual    2 1  Cautions for Installation and Use  For a safe use of the product  please read and be familiar with the following cautions     e Make sure that the installation site has no water     e The site of installation must be free from vibration  impact  high temperature and humidity   dust  and air containing salt and ion    e The sites containing 
18. raph End  and    Sync     Graph End    n  Pressing       key  BEEN is displayed and set    ync V    A          al        Move to the upper menu  Input MENU     Default Busy  Busy  Graph End  Sync     4  Relay channel mode operation    EZ    n case of selecting the relay input as the Ach or Bch  the  detailed operation can be set    E ELS   Each operation is performed differently according to the  choice of the comparison     The comparison is divided as two operations  limit and       reference     Ea   Select the input by pressing    Y keys       Select it by pressing the key between the Limit and       Reference keys     Reference  NENNEN     Pressing           key  HEN is displayed and set            ul        Move to the upper menu  Input MENU       Default Limit  Limit  Reference     17    4 1  Limit comparison       n case the comparison is selected as the Limit  the Replay  EN  NH  lt      performs OK and NG according to the set HH  HI  LO  and LL    values in advance     OutputType   For the left figure  the Relay 1 performs the output as the HH    10 58 905    HHHigh B HH Low Limit       The Type selects HH  HI  OK  LO  and LL          As the output type  the Output Type sets up Fast  Display  Decide  Graph End  M D etc     HH High and HH Low sets the hysteresis value  apply the hysteresis that the each value makes  the relay ON and OFF  and remove the relay ON OFF repetition phenomenon due to the    minimal change of the measured value     4 2  Reference comparison 
19. stopped             Model         When the data exceeds the set value  it is saved  from the front       default    StopWrite  Mode1      2  SD setup  2   2   Shortcut key  aos   ss   gt    ES   ES Pase  gt       EL n  Ei  Setup Late amp Time  File Save setting    Date amp Time       Setting     Set up a date and time     Setup  File Load   Setup S  File Delte       14    2 1  DATE  amp  TIME setting     Shortcut key  M2   gt  xs   Lue y 808m      eese      PS h      Set up the values  year  month  date  hour  minute  amp  second     by using the keys   How to enter  Using      L2 keys  enter the values after    selecting the wanted value  red color        Pressing    key  it is Clear 0    Pressing F  key  mamasa is changed           E   Move to the upper menu  SD Setup MENU     6 5 Relay    1  Selecting the Relay  Shortcut key  _  E     gt   8    gt  Mw       Move to the Relay setting screen     A            Move to the upper menu  MODE   u MENU      Pattern z Move to the MAIN MENU    2  Relay mode setup  1   11     channel     Ere Relay Disp    MENU W    Move to the MAIN MENU  Pattern  E   ta          u            Move to the upper menu  Select the Relay Ach Bch       Select the saved pattern        b Copy the saved pattern       ik b Select the Relay display method     2 1  Pattern Select      Refer to 6 3 Input Pattern Select   2 2  Pattern Copy      Refer to 6 3 Input Pattern Copy     Lo    2 3  Relay Disp       ES   Select the Relay display method by pressing the limit 
20. te by pressing BI Ej keys    Pressing _set   key        313 8     is changed        E p Move to the upper menu  SETUP MENU       Default 10  10 0  5 0  2 5  1 0  0 5        1 2  Display Rate        Shortcut key 298   5 x58    gt  m  p IE  y Sie          1 3  Sampling Rate    Select the rate by pressing a E keys      Pressing T  key  Mmmm js changed            Move to the upper menu  SETUP MENU     2000 CFS OPS    Default 1000CPS  4000  2000  1000  500  200  100  50  20  10     1000 CPE      1000 CPS    SET       X Depending on the choices  the graph data storage maximum    time is changed     2  Setup  2   5   Shortcut key  ons    gt  8    gt   Fase gt   x3     Digtal ZEKO   l    d When turning the power off  the data is set to be  Backup    stored    Whenever pressing somo   key  Maw  Digital zero  value change is stored   Mmmm js changed            _     Move to the upper menu  MENU     20    6 9 System  Basic function setup 2      Shortcut key  MOM  pE  gt   ese   gt   seem     Initialize    u me p All set data is initialized     1  System  3   4        Initialize Initialize  Waveform Setup    Initialize Initialize  Output ALL       1 2  Initialize ALL     Shortcut key  W  E   pws   gt  vtm   eue  o gt  TAE         All stored data is initialized      3X When initializing  all stored data is deleted        Zt    7  Summary of Communication Protocol    7 1  Communication Protocol    eee ee eee  m L    2 Output device ID   00   FF  Indicate ID from 0 to 255  and the Hex value
21. ter HI           sen set up the minimum renewal range of the display          999999 0     value     Limiter LO  5 1  Digital Limiter HI     Shortcut key  woe    xes  S20 meme   5  om   35 ERES h      Enter the Limiter HI value using the buttons         After setting up the value        3399  is changed by pressing  _set   button          E l Move to the upper menu  Input MENU          Default 99999 0   99999   99999     5 2  Digital Limiter LO   Shortcut key  mwj y s  p mee   5  BE   35 Eee h      Enter the Limiter LO value by using the buttons          After setting up the value     99k9     js changed by pressing    E EN fea EN  _ser   button         _      Move to the upper menu  Input MENU       Default  99999 0   99999   99999     lo    5 3  Step  Step width     Shortcut key  OBE  ps  pomme y ease  x 3 57    Select the step width by pressing a E keys    Pressing _ser   key  Mf is changed           E I  Move to the upper menu  Input MENU       Default 1  1  2  5  10        6 4 SD Setup  1  SD Setup  1   2   Shortcut key  W  5     mew y soseem f      eae   Set up the data save method     SD Write Waveform  Mode File Delete       Waveform Available  Data ap Menor  Overvrite  Made       1 2  Overwrite mode  Shortcut key  2    gt  ses   lm    Ea Tee        Set up the mode by pressing EH Ej keys         p  Pressing ser   key  is changed      stoplirite                Move to the upper menu  SD Setup MENU        1  1        _   geste When the data exceeds the set value  it is 
22. the decimal point are    displayed according to the set decimal places        24    oe   woe     n case of the above setup  the graph screen indicates this    100 00 area comparison screen        6  GRAPH  6   6       After completing a graph on the graph mode  check a result  by expanding a characteristic section of the X axle   Graph Mag  Graph Mag  EN JOD    tar Set up the graph expansion start point     SIM MN    n    Set up the graph expansion On Off setup     Set up the graph expansion end point     opem   woe    This figure indicates the result in case of turning the graph    100 00 expansion on under the above setup         Display the previous screen and expanded screen in turn by    HOLD  0         pressing 40m key on the figure        6 8 Setup  1  Setup  1   5   Shortcut key 2992   gt    gt  se  Lm      Set      Select the meter operation       Displav m    _ hate      Select the cycle to renew the display value        Meter   sampling    Setup Rate    sampling      u MN Select the storage maximum time of X axle     Display  Rate    Example  When setting up 4000 CPS  it is 1 sec  and when       setting up CPS  it is 2 sec     Zo    1 1  Meter Setup     Shortcut key  MODE   DES  _5  Page  EBENE  gt  sets        Select the channels by pressing al E keys    When pressing r  key  Me js changed             Move to the upper menu  SETUP MENU       Default Ach   Ach  Bch  2 ch  Wave A  Wave B  W amp D     Shortcut key  39  o0  paseo   gt  se      EXE        Select the ra
23. up the detecting time B   ange        Inflection          Infection   After Set up the detecting time A   etore    Inflection  S          tart H e 2   H  a   Start detecting the inflection point     Inflection      Inflection Range Range    Input the change vale  R on the figure  to detect the    inflection point     Inflection      Inflection Before Before      Input the inflection point comparison standard time  B on    the figure       Inflection      Inflection After After         Input the inflection point detection change time  A on the figure       Inflection      Inflection Start Start      The input level condition to start the inflection detection     2A    6 7  GRAPH    1  GRAPH  1   6   Shortcut key  MENU   eS   gt  itl      z Pattern  Graph Disp EN   Select  Pattern    Pattern a Copy  Select  Pattern   Copy       Pattern    1 1  Pattern Select Select    Refer to 6 3 Input Pattern Select     Pattern    1 2  Pattern Copy Copy      Refer to 6 3 Input Pattern Copy     2  GRAPH  2   6     b Select the saved pattern       Copy the saved pattern     pue   Set up the minimum maximum values of the XY axles on the       _    Y Min  NEL Y Max  _u X Min  _ Mt X Max    graph screen     b Set up the Y axle display start point value   b Set up the Y axle display maximum value     Set up the X axle display start point value         Set up the X axle display maximum value       The value setup is available by using the touch keys on the graph screen     TAE   wok     Touching C  
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
User Manual  取扱説明書  Samsung M632 User Manual  第11回「でんちフェスタ」を開催  簡易支持力試験機 エレフット(PDF:492KB)  カタログ - フジ医療器    MOVIDRIVE® B – 2 Day Operation Start Up and Troubleshooting  RCA RCRP05BR remote control  Canon PowerShot S1 IS    Copyright © All rights reserved. 
   Failed to retrieve file