1. 1 26 Vista basic Contents CONTENTS Warnings amp Cautions i c 2 0cccsceessesesseeeeceteemeenmensscenaatcecneenenens 3 IMPGCOCIUIG TION rires ancenanienceunet EE EEEE 5 Preparing the Infusion suicacustrnrnasemnciivacnnaumadsrcndeadercatacteaines 7 Continuous Mode 22 co cenessoriaasrenhdanisiaideaasnedpahiedsdaseaatanenetedideaeneds 8 Options GlearAllozesessneneniieninoni iaaea 11 Piggyback Senne nen ene ome eer ee Ret meter een mare oe mets 11 Ramp Taper Mode ccccceeeeeeeeeeeeeceeeeeeeeeeeeeees 12 Program Mode ics ccctcrencicsaut crn orteccienmarcenteccigrsgiduce 14 Delay A Mode ceeeeeeeeeeeeeceeeeeeeeeeeeeenaeeeeeeeeeeeeeeeee 16 Dosage CalCulatlOMscxicctic siete ctemietaticeieetis Steen 17 Bolus Mode Interval Bolus Intermittent Mode 18 Total Infused Standby cccccceeeeeeeeeeeeeeeeeeeeeeees 19 Occlusion Pressure Drop Sensor Battery 20 Data Lock Contrast Selection c cceeeeeteeeeeeee 21 Adjust Clock Loudness sssesseeseeeeeeeeeeeees 22 PNAS es twaccaseaaaduniswascuiandnatidvansualavcadvasiovandvantvancsanivantvaeteancrduevantuaees 23 AC 12V Battery amp CC Interface ccccceeeeeeeeeeeeeeeeeeeeeeeees 24 Maintenance Storage amp Service cccccccccccceccceececeeeeeeeeeeeeess 25 Specifications 5 225 ic ceitine cauctcncteapete te nagudcs scaptsagueaguausucaatenceusieaaes 26 The Vista basic is a volumetric
2. 3 26 WARNINGS amp CAUTIONS After starting an infusion make sure drops are falling in the drip chamber If no drops are falling make sure the roller clamp is open If the roller clamp is open and still no drops are falling replace and dispose of the set The air detector cannot recognize the introduction of air at 3 way stopcocks infusion ports and other lines tubes below the pump See section entitled Drop Sensor for additional information on using the device without a drop sensor e To avoid mechanical or electronic damage do not steam autoclave or immerse the pump in any fluids or cleaning solutions Always disconnect electrical power cord from outlet before cleaning to prevent electrical shock CAUTIONS Rxonly Do not attempt to infuse two fluids simultaneously using the Vista basic e Variations in pressure e g caused by change in level of fluid container may affect the accuracy of the device If several pieces of infusion Use only equipment accessories parts and disposables that are intended for use with the Vista basic Do not clean disinfect or sterilize any part of the device by autoclaving or with ethylene oxide gas Doing so may damage the device and void the warranty Only external parts of the pump should be disinfected See Maintenance Storage amp Service for suggested cleaning solutions Technical specifications are not guarante
3. displayed to the right 2 Continue to hold down the first BOL while also pressing the second BOL A bolus will be given as long as both BOL keys are held down The pump will display the amount of bolus being administered and will beep once after each milliliter delivered INTERVAL BOLUS INTERMITTENT MODE The basal KVO rate and entire bag volume must be entered in CONTINUOUS MODE first 1 Press OPTS until display shows BOLUS MODE Press ON Press TIME Enter the time between the start of each dose then press OK Press RATE Enter the rate at which the dose is to be given then press OK Press VTBD Enter the volume of each dose then press OK Press END 0 Press ror The pump will deliver at the basal KVO rate for the interval time then begin delivery of the first dose OMNANARWD Vista basic Instructions for Use 18 26 950983 Rev B 03 04 OPTIONS Total Infused Standby Function TOTAL INFUSED The Total Infused information screen displays the total amount of fluid in ml that has been pumped since the last time the Total Infused was cleared To Clear the Total Infused 1 Press OPTS until display shows Tot Infused 2 Press C 3 Press END STANDBY This function allows the pump to be put on standby while retaining all set infusion data This function is only available in Continuous Mode Press OPTS until display shows STANDBY Press ON Enter time in hours and minutes the pump is to stand by
4. Case Easy upkeep drip proof and resistant to disinfectants Function Keys Allow choices of YES NO OK INFO OPTS etc ON OFF Power Key Alarm Silence Key Silences alarm tone for 2 minutes if a START STOP Key Vista basic Instructions for Use 6 26 950983 Rev B 03 04 PREPARING THE INFUSION Prepare the Pump 1 Check for cleanliness completeness and damage If unit appears damaged do not use it 2 Attach the pump to an IV pole by turning the knob on the pole clamp clockwise or place the pump on a flat stable surface 3 Plug the pump s power cord into an electrical outlet Turn on the Pump 1 Press ON OFF 2 Listen for 3 loud beeps 3 Check screen for completeness of all pixels on the display The Alarm amp and Infusing 6 lights will blink once Load the Set 1 Prepare the solution container and prime the administration set according to the instructions provided on the set package Close the roller clamp 2 Check that the drip chamber is about 1 3 full and there is no air in the set 3 Open the pump door by pressing the Door Open Button on the top right corner of the pump 4 Push and hold open the clamp at the top of the fluid path 5 Insert tubing in clamp and release clamp 6 Thread the tubing down the fluid path making sure the tubing is pressed into the Air in Line detector at the bottom of the fluid path The roller clamp should be below the pump 7 Close the do
5. Press C to clear the current data and re enter new data Press OK to accept new data Press OK The pump will count down to zero and then alarm Pump Stopped PONS gt A To Exit Standby Function 1 Press EXIT to return to Continuous Mode Note If countdown has already started press END to return to Continuous Mode Vista basic Instructions for Use 19 26 950983 Rev B 03 04 OPTIONS Occlusion Pressure Drop Sensor Battery OCCLUSION PRESSURE The pressure at which a downstream occlusion alarm will be triggered may be varied The higher the pressure limit the less sensitive the pump is to changes in fluid resistance 1 Press OPTS until display shows OCCLUSION PRESSURE 2 Press to change pressure to Low Press to change pressure to High 3 Press END to return to Continuous Mode NoTE Low corresponds to approximately 8 PSI and high to approximately 17 PSI High pressure limits may increase the severity of an infiltration without alarming DROP SENSOR Using a drop sensor eliminates the need for pre setting a VTBD in Continuous Mode and senses upstream occlusions and empty solution containers NoTE Turning the Drop Sensor feature off forces volumes to be entered and prevents sensing of upstream occlusions To Turn On the Drop Sensor 1 Press OPTS until display shows DROP SENSOR 2 Press ON 3 Press END to return to Continuous Mode To Turn Off the Drop Sensor 1 Press OPTS until
6. and Options refer to those sections later in this manual There are three ways to enter infusion data in Continuous Mode Vista basic Instructions for Use 950983 Rev B 03 04 8 26 Continuous Mode RATE and VTBD Volume To Be Delivered 1 Enter rate in ml hr 2 Press VTBD Volume To Be Delivered 3 Enter bag volume in ml 4 Press OK To leave screen without saving changes press EXIT 5 Press ss to begin the infusion Check for drops in the drip chamber VTBD and TIME 1 Press VTBD 2 Enter bag volume in ml 3 Press OK To leave screen without saving changes press EXIT 4 Press TIME 5 Enter infusion time in hours and minutes Note For a2 hour infusion enter 2 0 0 The pump will show Time 02h00m 6 Press OK To leave screen without saving changes press EXIT The pump calculates the rate and asks Accept Rate 7 Press OK 8 Press smx to begin the infusion Check for drops in the drip chamber RATE and TIME 1 Enter rate in ml hr 2 Press TIME 3 Enter infusion time in hours and minutes 4 Press OK To leave screen without saving changes press EXIT 5 Press sw to begin the infusion Check for drops in the drip chamber To Change the Rate 1 Press sw to stop the infusion 2 Press C The display will reset to 000 0 mi h 3 Enter a new rate 4 Press to begin the infusion Check for drops in the drip chamber Vista basic Instructions for Use 9 26 950983
7. infusion pump intended to provide accurate and continuous flow of parenteral and enteral fluids Qualified medical staff should decide how the device should be used based on its features and specifications Rx only Vista basic Instructions for Use 2 26 950983 Rev B 03 04 WARNINGS amp CAUTIONS The following should be observed to avoid patient user injury This Instructions for Use manual contains detailed instructions and warnings on the use of the Vista basic Infusion Pump Please read it completely prior to using this device This manual is intended to reinforce the teaching given to the user by a trained health care professional or an authorized B Braun representative WARNINGS B Braun will assume no responsibility for incidents which may occur if the product is not used in accordance with product labeling Make sure the pump is stable either fastened securely to the IV pole or resting on a flat surface Prior to use always check the audible and visual alarms and staff call where applicable Connection of pump tubing to the patient should only be made after the tubing is loaded and the unit is turned on Disconnect tubing from patient when changing solution containers to avoid the danger of incorrect dosage Ensure there are no kinks in the tubing Always read and follow the instructions that accompany the source container and IV administration sets you are using Carefully follow the
8. instructions for loading removing and reloading the set as well as the recommended set change intervals Change pumping section on tubing every 24 hours Do not use tubing for more than 72 hours or per current facility protocol Always check pump data prior to starting infusion Vista basic Instructions for Use 950983 Rev B 03 04 Do not operate this device in environments where there is an explosion hazard A possible explosion hazard exists if the pump is used in the presence of flammable anesthetics The EMC limits Electro magnetic compatibility according to IEC EN 60601 1 2 and IEC EN 60601 2 24 are maintained If the equipment is operated near other equipment which may cause high levels of interference e g HF surgical equipment nuclear spin tomography units mobile telephones etc maintain the recommended protective distances for these devices Under certain conditions malfunctions may occur which lead to a device alarm with permanent alarm tone Interference may occur e g at electro magnetic fields gt 10 V m resp electro magnetic discharges gt 8 kV Monitoring of the patient and infusion status is necessary to ensure the infusion is being delivered as anticipated The Vista basic has been designed to stop fluid flow to the patient under certain alarm conditions Also prior to starting an infusion verify that no drops are falling in the drip chamber and the programmed information is correct
9. than 9 Steps are needed press OK with a rate of zero Press OK to accept the total Program Volume Note Pressing CHNG on this screen returns to Step 1 for review and changes 10 Press OK to accept the total Program Time NoTE PRESSING CHNG ON THIS SCREEN RETURNS TO STEP 1 FOR REVIEW AND CHANGES 11 Press SS to start the infusion Check for drops in the drip chamber OONOARWN To Review or Change Program Mode data 1 Press OPTS until display shows PROGRAM MODE Note If Program Mode is infusing stop the infusion and press EXIT 2 Press ON 3 Review the data on each screen and press OK if data is correct or press C to clear the current data and re enter new data Press OK to accept new data 4 OR press CHNG on Accept Program or Accept Program Time screens see 9 amp 10 Notes above To Clear Program Mode 1 Press OPTS until display shows PROGRAM MODE Note If Program Mode is infusing stop the infusion and press EXIT 2 Press CLR 3 Press YES to clear all the data in Program Mode 4 Press ON and continue to program a new Program Vista basic Instructions for Use 14 26 950983 Rev B 03 04 OPTIONS Program Mode To Exit Program Mode 1 When infusion is complete press S to stop the KVO 2 Press EXIT 3 Press END to return to Continuous Mode Vista basic Instructions for Use 15 26 950983 Rev B 03 04 OPTIONS Delay A Mode DELAY A MODE The start of Continuous Piggyback Ramp Taper and Progra
10. 0 120 VAC 50 60 Hz 13 Watts Leakage Current Max 30uA Battery 7 2V 1 2 Ah NiCd rechargeable Battery Operating Time 3 5 hours at 125 mL hr Rate Range 0 1 to 800 0 mL hr in 0 1 mL increments with Vista tubing Volume to be Infused Range 0 1 to 9999 9 mL Occlusion Pressure Variable Settings Low 8 PSI High 17 PSI KVO Rate 1 mL hr or selected rate whichever is lower Flow Rate Accuracy 5 measurement time 8hr temp 71 6 F Alarms Air in Line Low Battery Plug In Now Bag Empty Downstream Occlusion Door Open Pump Stopped KVO Unit Malfunction Air in Line Alarm Air bubble size of 0 3ml default value The value can be set from 0 01 to 0 3 ml in service program Cumulative air in line alarm Can be set from 0 to 3 5 ml h in service program O cumulative air in line alarm is deactivated Default setting for the cumulative air in line alarm is 0 deactivated Memory User option to retain last Continuous and Piggyback Mode data upon power up Permanent data retention in Program and Ramp Taper Modes after pump is turned off Tubing B Braun Vista administration sets Modes Features Continuous and Piggyback Modes Program Mode for variable infusions Ramp Taper Mode for TPN infusions and Delay A Mode for postponing infusion start and finish Operating Operating 64 4 F to 95 F 10 15 PSI to 15 37PSI Temperatures Non Operating 13 F t
11. Enter the date using two digits for month day and year Example November 5 2001 is entered as 11 05 01 4 Press OK 5 Press TIME to enter the time of day Note The time is set using 24 hour Military Time Example 6 30pm is entered as 18 30 6 Press OK 7 Press END to return to Continuous Mode LOUDNESS This feature allows the volume of the alarm tone to be changed 1 Press OPTS until display shows LOUDNESS 2 Press to increase the alarm volume Press to decrease the volume 3 Press END to return to Continuous Mode Vista basic Instructions for Use 22 26 950983 Rev B 03 04 OPTIONS Alarms In alarm situations press gt to silence the alarm Correct the problem Resume the infusion by pressing lt Alarm Possible Cause Action Air In Line Air bubble in air in line sensor Remove air from system Check Drop Solution container empty Replace solution container Sensor Roller clamp closed Open roller clamp Free Flow Set loaded incorrectly Re load set close roller Not Enough Drops Too Many Drops Kinks closed clamps or blockage Droplets on wall of drip chamber Drop sensor out of position clamp Check for blockage Shake drip chamber Reposition drop sensor Door Open Door opened during infusion Close door Dwnstream Downstream tubing kinked Correct kinks Occlusion Filter blocked Correct blocked filter Closed clamp Open clamps IV positional or infiltrated Check IV si
12. Rev B 03 04 Continuous Mode To Change the Rate While Infusing 1 Enter a new rate 2 Press OK Note If the OK key is not pressed within 10 seconds of the change the infusion will continue at the old rate KEEP VEIN OPEN RATE After an infusion is complete the pump will alarm and begin pumping at 1 0 ml hr or the current infusion rate whichever is less 1 Stop the infusion 2 Turn off the pump or reset the program INFORMATION KEY This key is only available while the pump is infusing 1 Press INFO 2 The Total Infused value and the Total Time of the infusion completed are displayed 3 Press INFO again and the Battery Left and Total Operation time of the pump is displayed 4 The pump will return to the main infusion screen after 10 seconds Vista basic Instructions for Use 10 26 950983 Rev B 03 04 OPTIONS Clear All Piggyback The Vista basic contains many options to allow custom programming To access the various options press OPTS while the pump is stopped Pressing OPTS each time will move you to the next option Eventually the OPTS key will bring you back to the mode from which you began Depending on the model some modes options may have been disabled and will not appear on the display when the OPTS key is pressed CLEAR ALL The Clear All feature allows the user to clear all programming data within pump 1 Press OPTS until the display shows CLEAR ALL 2 Press CLR Note The pump will displ
13. Vista basic Instructions for Use Software IFVB B BRAUN er BBRAUN SHARING EXPERTISE 950983 Rev B 03 04 The CE mark confirms compliance with the Council Directive on Medical Products 93 42 EEC dated the 14 of June 1993 B Braun Melsungen AG INSPECTION ON DELIVERY Despite careful packaging the risk of transport damage cannot be entirely excluded Upon delivery please check that nothing is missing Do not use a damaged device If the device is damaged contact the Service Department See Maintenance Storage amp Service Vista basic Instructions for Use 950983 Rev B 03 04 RESPONSIBILITY OF MANUFACTURER Manufacturer assembly and installation personnel or instructors can only be held responsible for any effects on device safety reliability and performance if Installation expansion work readjustments modifications or repairs are carried out by personnel authorized by the above The electrical wiring for AC power satisfies the requirements of VDE 0100 0107 and or the IEC publications or regional and or national variations e The device is operated in line with the Instructions for Use and Regular technical inspections Preventative Maintenance are carried out every 2 years as outlined in the Service Manual Packaging The pump packaging is reusable Package Contents Vista basic Infusion Pump drop sensor power cable pole clamp Instructions for Use
14. and re enter new data Press OK To calculate the concentration of the solution use lt to move the to the mg field Enter the amount of drug in the solution Then press to move the to the ml field Enter the amount of solution Press OK Enter the body weight if part of the dosing parameters then press OK If body weight is not part of the dosing parameters press OK with 0 in the body weight field If time unit desired is other than the one displayed press button under the h until time unit desired appears in the first line Enter the dose or use gt to move the to the rate field and enter the rate Press OK Press st to begin the infusion To Review Data During Infusion 1 Press INFO twice To Exit Dosage Calculation 1 Stop the infusion 2 Press OPTS until display shows DOSAGE CALC 3 Press OFF 4 Press END Vista basic Instructions for Use 17 26 950983 Rev B 03 04 OPTIONS Bolus Mode Interval Bolus Intermittent Mode BOLUS MODE This feature allows a bolus to be given during a Continuous Mode infusion Press OPTS until display shows BOLUS MODE Press ON Press RATE Enter the rate at which the bolus should infuse then press OK Press END To give a bolus during the infusion press BOL Enter the amount of bolus to be given in ml Press OK ONOaAPPwWN Bolus Without Entering a Volume 1 During the infusion press BOL and hold down until a second BOL is
15. ay This will clear all data in all modes 3 Press YES 4 Press END to return to Continuous Mode PIGGYBACK The Piggyback feature allows programming of a secondary Piggyback infusion The pump automatically switches from the Piggyback to the Continuous Mode infusion when the Piggyback program is complete The option is only available if there is data in the Continuous Mode n LA 1 Connect the Piggyback set to the injection site on the Piggyback 8 tubing above the pump on the primary set Open the aia roller clamp on the Piggyback set 2 Hang the Piggyback solution container at least 8 inches Primary above the primary solution container 3 Program the continuous infusion in the Continuous Mode see Continuous Mode Press OPTS until the display shows PIGGYBACK MODE Press ON to turn the Piggyback feature on Press END PIGY will now be displayed in the lower left corner P Press PIGY Enter the Piggyback infusion data as you would Continuous Mode data Press 8 to start the Piggyback infusion 0 Check that drops are falling in the Piggyback drip chamber ee Pao me Vista basic Instructions for Use 11 26 950983 Rev B 03 04 OPTIONS Piggyback Ramp Taper Mode CAUTION Do not attempt to infuse both Continuous and Piggyback fluids at the same time Use primary sets with check valves NOTE To prevent incorrect delivery the programmed Piggyback volume to be delivered must equal the amount of fluid in the Piggyback c
16. de To Review or Change Ramp Taper Mode Data 1 Press OPTS until display shows RAMP TAPER MODE 2 Press ON 3 Review the data on each screen and press OK if data is correct or press C to clear the current data and re enter new data Press OK to accept new data To Clear Ramp Taper Mode Press OPTS until display shows RAMP TAPER MODE Press CLR Press YES to clear all the data in Ramp Taper Mode Press ON and continue to program a new Ramp Taper program PONS To Exit Ramp Taper Mode 1 When infusion is complete press S to stop the KVO 2 Press EXIT 3 Press END to return to Continuous Mode Fast Taper The infusion can be interrupted to begin an early taper down process during the Ramp Taper Mode Infusion Stop the infusion Press FAST Enter the FAST Taper Time in hours and minutes Press OK Press s to start the infusion Check for drops in the drip chamber al loa Vista basic Instructions for Use 13 26 950983 Rev B 03 04 OPTIONS Program Mode PROGRAM MODE Program Mode is for infusions requiring up to 9 separate rates and volumes to be delivered one after another This is designed to deliver infusions for stepped oncology drugs circadian medication delivery and other custom programs Press OPTS until display shows PROGRAM MODE Press ON Enter the rate of the first Step Press OK Enter the volume of the first Step Press OK Continue to enter the rates and volumes for Steps 2 9 If less
17. display shows DROP SENSOR 2 Press OFF 3 Press END to return to Continuous Mode BATTERY CAPACITY This is an indicator of the remaining battery time in hours and minutes This screen also indicates the amount of time the pump has been in operation 1 Press OPTS until display shows Battery Left 2 Press END to return to Continuous Mode Vista basic Instructions for Use 20 26 950983 Rev B 03 04 OPTIONS Data Lock Contrast Selection DATA LOCK This feature prevents unauthorized use by locking out all keys except s and mee 1 Press OPTS until display shows DATA LOCK 2 Press ON 3 Press END to return to Continuous Mode To Turn Off Data Lock 1 Press OPTS until display shows DATA LOCK 2 Press the decimal key 3 Press OFF 4 Press END to return to Continuous Mode CONTRAST SELECTION This feature allows the contrast of the screen to be changed for best possible viewing 1 Press OPTS until display shows CONTRAST SELECTION 2 Press to increase the contrast make the text darker Press to decrease the contrast make the text lighter 3 Press END to return to Continuous Mode Vista basic Instructions for Use 21 26 950983 Rev B 03 04 OPTIONS Adjust Clock Loudness ADJUST CLOCK This feature allows the time and date to be changed 1 Press OPTS until display shows ADJUST CLOCK 2 Press DAT to enter the date Press C to clear the current data and re enter new data 3
18. ed with any other tubing than that approved for use with the Vista basic Any analog and digital components connected must demonstrate that they meet EN specifications i e EN 60950 for data processing equipment and EN 60601 for electrical medical equipment Only an authorized system configuration technician responsible for meeting the specifications of standard EN 60601 1 1 may connect additional devices to the signal inputs or outputs SAFETY STANDARD The Vista basic meets all safety standards for medical electrical devices equipment are connected together corresponding to IEC 60601 1 and IEC itis possible there may be mutual 60601 2 4 influence Possible incompatibilities Note IEC 60601 1 corresponds to can be found in the instructions for European Standard EN 60601 use for the drug or for the other appliances Vista basic Instructions for Use 950983 Rev B 03 04 4 26 INTRODUCTION Display All essential information is displayed in plain language Green background lighting illuminates only when connected to AC power or in battery operation only when a key is pressed Keyboard Allows data entry Alarm Light Lights up during an alarm Display will also show flashing AAA A Infusing Light Indicates pump is infusing Vista basic Instructions for Use 950983 Rev B 03 04 Handle for carrying Clear Key Resets a value to zero 5 26 INTRODUCTION Door Open Button Aluminum
19. letely discharging it from time to time and then fully recharging it on AC power Vista basic Instructions for Use 950983 Rev B 03 04 CC Interface Interface descriptions available from B Braun Connection to interface input uses MFC plug Documentation All operating data of the Vista basic can be called up and logged via external computer 24 26 MAINTENANCE STORAGE amp SERVICE Maintenance Clean the pump with a soft lint free cloth or swab dampened with soap and water a general non staining chemical disinfectant isopropyl alcohol 15 derivatives of phenylphenol 5 or ammonium chloride 10 As a general recommendation for cleaning and disinfecting this device use a lint free cloth dampened not soaked with a 0 5 bleach Sodium Hypochlorite solution 10 solution with water The solution is most effective when prepared weekly and allowed to remain on the device for approximately 10 minutes prior to wiping off Do not use acetone solutions containing glutaraldehyde or abrasive cleansers on the pump If necessary Betadine or lodine solutions may be used but they will stain the pump Do not use abrasive cleaning tools or pressurized spraying devices as these may damage the device Do not sterilize the pump using ethylene oxide EtO gas Do not steam autoclave or immerse the pump in any fluids or cleaning solutions Always turn off the pump and disconnect it from the elec
20. m Modes may all be delayed The Vista basic also provides an option to extend the Infusion Complete alarm at the end of the infusion PON gt gi 10 Program the infusion Press OPTS until display shows DELAY A MODE Press ON Enter in how many hours and minutes the infusion should begin Note This is NOT a clock time like 12 30 pm Press OK Enter how long in hours and minutes after the end of the infusion the pump should wait before alarming Press OK Press YES if the pump should run at KVO during the delay times Press NO if the pump should not infuse during the delay times Review the data and press OK To change the data press EXIT and begin with 3 above Press to begin the delayed infusion To Exit Delay A Mode 1 Stop the infusion 2 Press OPTS until display shows DELAY A MODE 3 Press OFF 4 Press END to return to the previously active mode Vista basic Instructions for Use 16 26 950983 Rev B 03 04 OPTIONS Dosage Calculation DOSAGE CALCULATION This feature allows a rate to be calculated given the concentration and dose 10 11 Press OPTS until display shows DOSAGE CALC Press ON If concentration unit desired is other than the one displayed press button under the mcg until units desired appear in the CONCENTRAT 0 ml line Enter the concentration of the solution If the concentration of the solution is not known go to 6 below Press C to clear current data
21. o 131 F 10 15 PSI to Vista basic Instructions for Use 950983 Rev B 03 04 15 37 PSI 26 26 2004 B Braun Medical Inc Vista basic Instructions for Use 950983 Rev B 03 04 B BRAUN SHARING EXPERTISE B Braun Medical Inc 824 Twelfth Ave Bethlehem PA 18018 Tel 800 227 2862 www bbraunusa com
22. ontainer Automatic Piggyback is not intended for the infusion of fluids requiring flushing before and after administration NOTE If the Piggyback container is not empty the remaining Piggyback fluid will be delivered at the Continuous Mode rate once the pump transitions back to the Continuous Mode To Exit Piggyback Press so to stop the infusion Press END Press OPTS Press OFF to turn Piggyback off Press END to return to Continuous Mode Oe ON RAMP TAPER MODE The Ramp Taper Mode is designed to deliver infusions such as Total Parenteral Nutrition TPN which require gradual ramp up and taper down rates The pump automatically calculates the rate changes required to match the volume total time and ramp taper time parameters Press OPTS until display shows RAMP TAPER MODE Press ON Enter the total bag VOLUME to be delivered Press OK Enter the RAMP UP time in hours and minutes Note The pump defaults to 01h00m To change press C and enter the time Press OK Enter the TAPER DOWN time in hours and minutes NoTE The pump defaults to 01h00m To change press C and enter the time 8 Press OK 9 Enter the TOTAL Time for the infusion 10 Press OK Note The pump calculates the maximum rate and asks Accept Rate 11 Press OK START 12 Press sto to start the infusion Check for drops in the drip chamber Ji e oN NO Vista basic Instructions for Use 12 26 950983 Rev B 03 04 OPTIONS Ramp Taper Mo
23. or and open the roller clamp Check that no drops are falling in the drip chamber 8 Place drop sensor on drip chamber optional 9 Connect set to patient site Turning off the Pump 1 Press S to stop the infusion 2 Close the roller clamp 3 Press andhold for 2 seconds Vista basic Instructions for Use 7 26 950983 Rev B 03 04 Continuous Mode When the pump is turned on it may display Tubing Type Confirm you are using a Vista Pump Set VISTA Pump Set then press YES The pump may then display Therapy stopped at mm dd yy 00h00m Resume YES Resumes the last therapy with all data retained NO Moves to the next screen The pump may then display Is This a New Therapy YES Begins in Continuous Mode with all data cleared The pump may then display Prime Set YES Press YES to begin priming set Press STOP to end priming Press EXIT to begin Continuous Mode NO Begins in Continuous Mode with all data cleared Press the green key below your answer CONTINUOUS MODE NO Returns to the beginning of the last program with all data retained YES Press YES to begin priming set Press STOP to end priming Press EXIT to return to the beginning of the last program with all data retained NO Returns to the beginning of the last program with all data retained This section explains how to program a Continuous Mode infusion For directions on how to program other Modes
24. te Catheter or vein too small Increase occlusion limit Occlusion limit too low Invalid data Entered data is out of range Press re enter data Keep Vein Open Rate Infusion Complete Programmed infusion is complete Stop infusion Low Battery 30 minutes until battery is fully discharged Silence alarm and connect pump to AC power Plug In Now Pump Stopped Battery is discharged No key panel activity for 2 minutes or Standby time expired Connect pump to AC power Silence alarm begin infusion Unit Malfunction Internal error Press ae until display reads Do not press any key until display is off wait until pump turns off turn device on If alarm repeats return for service Vista basic Instructions for Use 950983 Rev B 03 04 23 26 AC 12 V BATTERY amp CC INTERFACE AC and 12 V Check AC voltage as per rating label Insert screw on power lead 12 V lead in ambulance car on back Battery Operation Switching to internal battery is automatic in the event of power failure or if 12 V or AC is not connected Charge battery on initial use following a battery alarm or after 2 months of non use Charging time approx 16 hours Longer charging has no detrimental effect Full battery provides operation for 3 5 hours at 125 mL hr Battery capacity is approx 50 after 2 years The service life of the battery can be extended by comp
25. trical outlet prior to cleaning Vista basic Instructions for Use 950983 Rev B 03 04 Storage Store the pump away from excessive heat cold or humidity Keep devices plugged into electrical power during storage Service If the pump fails to respond to the operating or troubleshooting procedures listed in this manual and the cause cannot be determined discontinue use and forward it to an authorized B Braun Service Center Should it be necessary to return a pump for repair contact Technical Support at B Braun Customer Service at 800 627 PUMP A Returned Materials Authorization number will be provided Carefully pack the pump preferably in the original packing and ship it prepaid to B Braun Medical Inc Attn Service Manager 1601 Wallace Drive Suite 150 Carrollton TX 75006 USA B Braun cannot assume responsibility for loss or damage to returned pumps while they are in transit Product complaints may be sent to the Quality Manager at the above address With each complaint please include pump serial number and full description of difficulty including all settings types of fluids times and alarm messages Return set used if possible Contact Technical Support for an RMA number prior to return 25 26 SPECIFICATIONS Mechanism Linear peristaltic volumetric infusion pump Dimensions WxHxD 5 5 in 14cm x 9 4 in 24cm x 7 8 in 20cm Weight 6 8 lbs 3 1 kg Power Requirements 10