Home
        LASERSTAR
         Contents
1.     10 1 2 2 Total Accuracy of Calibration    Since the instruments are calibrated against NIST standards  the accuracy is  generally 1  at the power level at which the calibration has been performed  This  accuracy has been verified by checking the scatter of the results when several  instruments are calibrated against the same standard  The maximum error in  measurement will be less than the sum of the specified accuracy and linearity  Since  the linearity is also 1   the maximum error in measurement will generally be less than  2      10 1 3 Calibration Factor    For models equipped with the Ophir CAL resistor  the calibration factor is determined  at the factory by the following procedure  After calibration of the sensor as described  above in Section 10 1 1 2  electrical power is applied to the CAL resistor  The  calibration factor is given by the formula     CAL factor   reading on meter   applied electrical power   If you wish to check and adjust the power reading  apply electrical power to the CAL  resistor  and measure the voltage and current to obtain the applied power  The  correct power reading for the meter is the applied electrical power multiplied by the  CAL factor given above     10 1 4 Recalibration of Thermopile Heads Using the Laserstar    If your measuring head is equipped with the exclusive Ophir CAL resistor then  calibration is performed according to the instructions in Section 10 1 4 2     If you do not have the CAL resistor  then you can recalibrate 
2.     filter is out     Be sure to remove the filter from the detector head    4  When finished press exit and esc to return to the measurement screen     5 4 4 Zoom    The Zoom function causes the scale to expand around the present reading  20    Thus  if the full scale of the bargraph is 20mW and your reading is 15mW  pressing   zoom  will make the bargraph scale range between approximately 12 and 18 mW   Variations in power are more easily seen in this mode  When zoom is engaged  the  legend is highlighted  To cancel the zoom function  press  zoom  again  The zoom  function can be useful for laser power tuning and peaking     To access the zoom function  press  zoom   When it is highlighted it is engaged  Note  that the fiducial marks of the bargraph are further apart  indicating that zoom is  engaged  Pressing  zoom  again will cancel the function     5 4 5 Offset    The PD300 and PD300 3W heads have automatic background subtraction as described  in Section 5 1  In addition  the offset function can be used to subtract the residual  background signal that remains if desired  If the Laserstar shows a nonzero power  reading even when there is no laser  you can subtract the background using the zero  function  For example  the Laserstar display reads 0 1mW when the laser is blocked  and 20 5 mW with laser power applied  In this case  the true power is 20 5   0 1   20 4  mW  To subtract the background  press  offset  while the laser is blocked  The  Laserstar will now read zero 
3.    1100   1600nm          5   1600  1800nm        7    Table 4     Maximum Error as a Function of Wavelength and Filter      Add  2  to error for filter in   5  for PD300 UV from 220 to 300nm      10 2 3 Recalibration at a specific wavelength from a Known Sourceof Laser    Power    From the bargraph power screen press the    menu    button  select    advanced     and press    go     Select    calibrate    and press    go        2  Using the    value    button select the proper laser wavelength    3  Use the    select    and    value    buttons to choose the appropriate laser  wavelength and averaging period    4  Apply the known laser power    5  Adjust up down until the power reading on the screen equals the known power   The power calibration factor will change accordingly    6  Press exit and    permanent save     The new value will be saved    Note 1     The relative readings at different wavelengths are fixed by the wavelength    calibration curve stored in the head EEROM  When changing the calibration at  one wavelength  the calibration at all other wavelengths will change  proportionately     91    10 3 Calibration of Pyroelectric type Heads  Three types of absorber surface are used in Ophir pyroelectric measuring heads     1  Metallic type  The type with no suffix in the name have a partially reflective  multilayer metallic coating which absorbs approximately 50  and whose  absorption graph is shown in graph 2 below  The metallic coating permits very  high repetition 
4.    5 4 3 Operation with Filter In or Out    The PD300 head is equipped with a built in filter so that the photodiode can measure  up to 30mW without saturating the detector  In addition  the PD300 comes with an  additional removable filter for measuring up to 300mW  Other models of the PD300  series also have built in and removable filters  The exact maximum power is reached  when the reading reaches full scale or the output current from the head reaches  1 3mA  whichever comes first  See Table 2 for the exact maximum as a function of  wavelength     Depending on what powers you wish to measure  you should choose to work with the  removable filter installed or not  For this purpose  the Laserstar has a  filter  setting  and uses the proper correction curve depending on whether the filter is installed or  not     Warning   If the PD300 is used in the  filter in  setting and the filter is not installed or vice  versa the readings will be completely incorrect   If the power of your laser exceeds the maximum for filter in  you can purchase a  thermal or integrating sphere head for that wavelength  Consult your Ophir agent  for details     48    To choose the filter setting     1  From the main measurement screen press the menu button and then    go       2  Ifyou wish to work with filter installed  press  change  until display says  filter is  in   Be sure to install removable filter on detector head    3  If you wish to work with the filter out  press  change  until display says
5.    RS232 output    Max real time data logging rate   gt 30Hz  Onboard data logging rate   gt 200Hz   Data transfer from instrument to PC  500 points s  Max points stored onboard 59 400      The above refers to the rate for logging every single point  Above that rate the  instrument will sample points but not log every single point        IEEE 488 1 GPIB    Compliance   EEE488 1   Max real time data logging rate    Onboard data logging rate  1500Hz   Data transfer from instrument to PC  gt 1500 points s     The above refers to the rate for logging every single point  Above that rate  the instrument will sample points but not log every single point      gt 1500Hz       General    Maximum communication rate 38400 baud             Can store up to 10 files for a total of 54 000 points  Max points stored onboard  per power file is 5400   Data is not lost when instrument is turned off        100       11 2 Head Specifications                                                                                                    EAD MAX POWER MAX AVG  POWER ABSORBER    WATTS  DENSITY TYPE  PD300 UV IR 300mW 50W cm  PD  PD300 3W 3W 50W cm  PD  3A IS 3W 200W cm  Int Sph PD  F100A IS 100w 200W cm  Int Sph Ther  2A 2W 200W cm  BB  3A P CAL 3W 50W cm  P  10A P 10W 50W cm  P  30A P 30W 50W cm  P  30 150 A HE 30 150 W 1000W cm  HE HE1  10A 10W 20KW cm  BB  30A 30W 20KW cm  BB  30 150 A 30 100 W 20KW cm  BB  F150A 150W 20KW cm  BB  FL250A 250W 20KW cm  BB  F300A 300W 20KW cm  BB LP  1000W 1000W 6
6.    To save configuration     To save configuration  follow directions in Section 2 2 2 above     2 3 Photodiode Heads    2 3 1 Use of Laserstar with photodiode type heads     1   2     While the Laserstar is off  plug in the head then switch it on again    From the main measurement screen  press  laser  to select the correct laser  wavelength  If the wavelength you want is not listed  see section 0 below    If you want to save this new wavelength as the startup default  follow the  directions on configuration in section 2 3 3 below     To choose manual or automatic ranging or dBm in power measurement     1     Press  range  and then select the appropriate manual range  autorange or dBm   logarithmic scale     Press  exit  to return to the main measurement screen  If you wish to save the  selected range as the startup default  press  save  before exiting     2 3 2 Selecting Chosen Wavelengths    1     w    2 3 3    From the power measurement mode  press  menu  and select  wavelength    Press  go     Select the wavelength you wish to change and press  go    Using the up down buttons  change to the desired wavelength  Now press exit   Repeat steps 2 and 3 for other wavelengths desired  Up to 6 wavelengths may  be selected    When finished  press  save      exit      esc  to return to the main measurement  screen  If you only want a temporary change  Press    esc    without    save        Setting the Startup Configuration    From the power measurement mode  press the menu button 
7.    before pressing    exit        8 3 Dual Head Operation with Two heads Plugged In     1   2     Attach the two heads to the two input connectors   See Figure 4    Press the menu button   Press the arrow keys until    mode    is highlighted  Press       go        78    3  Press    value    to select    BOTH    so both heads are active    4  Now press    select    to select    CHANNEL A    and press value to select if you  want the head to be in power or energy mode  Now press    select    again to  select    CHANNEL B    and press value to select if you want the second head to  be in power or energy mode Now press    select    again until    MODE    is  highlighted  Press    value    to select the mode of operation    5  Now press    exit    and    esc    and the Laserstar will operate in dual head mode   If  you want the default start up mode to be the one you have selected  press     save    before pressing    exit        Note 1    Description of modes    a  Dual  The Laserstar will now operate with the two heads at the same time  independently  as if they are attached to two different Laserstars    b  A B  The Laserstar will display a screen showing the values of each head as  well as the ratio of head A divided by head B    c  B A   The same as above but the ratio is now head B divided by head A    d  A B  The Laserstar will now display the difference head A minus head B    e  B A  The same as above but now the difference is head B minus head A   f  A B  The Laserstar wi
8.    go    button will now  change to    pause     If you wish to pause the measurements without losing data  gathered  press    pause      Press    pause    again to resume measurements  When  the data register is full     FULL    will flash on the screen  If    zoom    is pressed at any  time  the screen display will expand fivefold around the present power value   Press    zoom    again to cancel the expansion  If you want to start a new  measurement and not save the present data  press    reset     If you want to save the  data before making a new graph  see the next section     31    Note   For data logging  the instrument has to be in a manual power range  If it is not  a  message will occur     Warning   Pressing    reset  will erase all previously stored data and start recording new  data     4  If you press the    stats    button  the graphical display will be replaced by a  statistical summary of the data gathered till now  This summary is updated every  2 seconds     5min    go pause reset       Figure 8  Laser Power VS Time Graph    4 4 2 7 Power vs  Time Graph and Data Logging  4 4 2 7 1 Data Storage and Transmission    When using thermal and photodiode heads the data logging period can be set to  anywhere from 12 seconds to 600 hours  The number of points stored in the data  register is up to 5 400 points  where less will be stored for data logging periods of less  than 5 minutes  The data can be stored to file as described below  To store the data do  the followi
9.   aperture  a temperature gradient is created across the thermopile of the enclosed  detector disc  This generates a voltage proportional to the incident power  The  display unit amplifies this signal and indicates the power level received by the head   At the same time  signal processing software causes the display unit to respond faster  than the thermal rise time of the detector disc  thus reducing the response time of the  Laserstar  Energy of a single pulse is measured on the Laserstar by digitally integrating  the pulse power over time     23    4 2 Selecting Settings from the Laserstar Screen    The Laserstar can be set to various chosen settings while operating  In addition  it can  be set so that it will be in the desired configuration when turned on the next time     4 2 1 To Set Type of Laser being Used    Thermopile heads have somewhat different absorption at different wavelengths  In  order to compensate for this  each head has been calibrated by laser at several  wavelengths  When you choose the correct laser wavelength  the correction factor for  that wavelength is automatically introduced  Note that the laser wavelength  correction in use is displayed in the upper left corner of the display     4 2 2 To choose Manual or Automatic Ranging in Power Measurement    Autorange   The Laserstar allows you to choose autorange or manual ranging  In  autorange mode  you do not have to change scales  When the reading of the meter or  bar is more than 100  of full scale  t
10.  0 0   and the 0 1 mW background will be subtracted from  all subsequent readings  The laser power reading will thus be 20 4 mW     49    When  offset  is engaged  the legend is highlighted  To cancel  press  offset  again  If  the  offset  is engaged  and you wish to subtract a new value of the background  press   offset  twice  The first press will cancel the old value  and the second will activate a  new value  If you suspect that the Laserstar has a permanent zero offset  then  disconnect the head while it is in the power measurement mode     5 4 6 Measuring loss using the dB offset function    Since dBm is a logarithmic measurement  the ratio between two measurements will be  the difference between the dBm measurements  For instance  if you want to measure  the loss in a fiber optic cable where the measurement before the cable is 1mW   0dBm  and the measurement after the cable is 0 1mW    10dBm  The ratio is then 1 10   0 1  and the dB loss is 0       10    10dB     The dB offset function allows you to easily measure this  To do so do as follows     1  When measuring the reference value press    dB O     The value changes to 0 dB   note that now the units are dB  a relative value instead of dBm  an absolute  value     2  Now make your second measurement and the value of the difference in dB   ratio  in numerical units will be shown     Note   If there is a zero offset in the reference value  you cannot subtract this using the  dB offset function  Instead  before the star
11.  1 1     633nm   20mw   20mw   Bmw  W fiam   mw    ow fi   w fi        N A  N A  1W  2W  3W  3W  3W      250mW   NA       1060mm   250mw   25w   300mw  na f J      Table 2   Maximum Measurable Laser Power as a Function of Wavelength       N A   W  W  W   1W   1 5W  2 5W    5 2 Setting up the PD300 to display the user s chosen wavelengths    The PD300 series has built in wavelength correction curves for measurements either  with the removable filter installed  filter in  or removed  filter out   These curves are  stored in the head EEROM  The correction curves  with a resolution of Inm  ensure  that the power reading is correct at all laser wavelengths     In order to simplify changing from one laser wavelength to another  the user can    program up to 6 different wavelengths to be available from the screen menu  Please  use the following procedure to set the PD300 to your laser wavelengths     45    From the power measurement mode with the bargraph display  press the menu  button located on the right and select    wavelength     Press    go       Press    select    to select the first wavelength desired  Press    go       Change to the wavelength desired using the up and down buttons  If you desire to  permanently save the new value press    save     Press esc    Repeat steps 2 and 3 for other wavelengths desired  When finished press exit   Those values for which    save    was pressed will be default startup values  the  others will be saved until the instrument is turned of
12.  5 5 IEEE Address  If you are using the Laserstar for IEEE 488 communications  the IEEE adapter should be    plugged in and the instrument switched on  The IEEE address should be set by  selecting  IEEE address   choosing the appropriate number and then pressing save     3 5 6 Zero Adjustments    In the Laserstar  all adjustments  including zeroing internal circuits  are done from the  software  This ensures simple and accurate realignment  It is recommended to re   zero the Laserstar every 2 months for best performance  The simple zeroing  procedure follows     1     w    If a head is connected  disconnect the head  turn off the instrument then back on  again so the Laserstar can identify that no head is connected   Head  Disconnected  will be displayed    Let the Laserstar run for at least 30 seconds before performing zero adjustment   Press  zero     Make sure the instrument is not in an electrically noisy environment and is  undisturbed  Press  go  and wait for message   zeroing completed   Now press   save  and the new zero will be saved  Press  esc  to return to the main screen     Note  for Thermal Heads only      For best results with thermal heads  it may be necessary to do the procedure  once with the head disconnected then afterwards again with the head  connected     After completing steps 1   4 above  Connect the head and make sure it is at room  temperature and well shielded from any stray thermal power  It may be best  advised to lay the head with the absorber 
13.  It measures average power with a high degree of accuracy    3  in general  In addition  there is a fast photodiode detector  which measures the  energy of the laser pulses in real time  The photodiode detector is mounted so that a  fraction of the radiation falling on the thermopile absorber  4  to 15  depending on  head type  is scattered form the absorber  A fraction of that falls on the photodiode   see Figure 17   There is also an additional  larger  fast photodiode that is connected  directly to an oscilloscope via the BNC connector and with this one can measure the  temporal pulse shape of the beam to a resolution of ns      64    BNC connector to fast  oscilloscope for  temporal pulse shape          Si photodiode    Removable filter  Incoming laser beam  Scattered light vA   Volume absorber   Figure 17  Schematic Diagram of RP Head    The RP photodiode and its associated circuitry give a signal proportional to the energy  of arriving pulses  The photodiode signal also shows the time between adjacent  pulses  All of the above information is fed into the Laserstar microprocessor and is  averaged over two seconds  From the above information  the Laserstar can calculate  the average pulse rate  average energy and the energy per pulse  as given by the    formulae below   Average pulse rate    number of pulses in time period     time period   Average energy    average power     pulse rate     energy of pulse  i    average energy  x  photodiode signal i      average  2 sec  
14.  a bargraph or mechanical meter  this  display shows graphically what came before as well as the current reading and the  trend  This allows you to see at a glance if you have reached maximum power     The Laserstar also has the option of graphing the laser power vs  time  or successive  energy points as long as the Laserstar has not been turned off  While measurements  are being taken  the Laserstar will record data until the screen is full  or  reset  or  exit   is pressed     If the laser power is fluctuating  the Laserstar can also display the average power  readings with averaging periods varying from 1s to 1 hour     These functions are the same for thermal type heads and for photodiode heads  See  Section 4 4 2 4 for details     5 5 Special Photodiode Heads   5 5 1 PD300 BB Radiometer Head   In addition to photodiode heads for individual wavelengths  Ophir also supplies heads  for measuring the output of broadband light sources  The PD300 BB head has    spectrally flat response from 400 to 1000nm and therefore can give the true total power  of any broadband light source in that spectral region     51    Note     Because the BB heads are spectrally flat  wavelength selection is disabled  Other  than that  operation is the same as other PD300 type heads     5 5 2 PD300 CIE Photometer Head    The PD300 CIE head has a spectral response similar to that of the human eye and can  therefore make measurements in eye response units of Lux  The PD300 ClE is  designed with a small
15.  a rate from 400 to 800Hz depending on the laser pulse rate  If you choose    1 out  of n     the RP will sample at a rate of 1 n of what it would in the every pulse mode   Now press    log    and the energy log screen will appear  Press    go     The  measurement will immediately begin  When the screen is full  the display will  scroll and the latest points will continue to be displayed  When the data register is  full     FULL    will be displayed  If you wish to see finer variations in the graph  press  the    zoom    button    When    reset    is pressed  the RP Laserstar is ready to begin a new measurement   If you want to store the data before making a new measurement  see section  4 4 2 7 1 on data storage     Warning     Pressing  reset  will erase all previously stored data     When the    stats    button is pressed  a statistical summary of the data on the  screen is shown  If you wish to return to the graphical display  press    graph     If  you wish to permanently store the data in a file  press exit and then press    files     to enter the file system  See section 4 4 2 7 1 for further information     75    7 11 4 Stabilization and recording of initial pulses    As discussed in section 7 6 paragraph 7  The RP allows you to record the absolute  energy of the first pulses of a laser s operation even though the average power is not  yet stabilized     When logging energy pulses  the RP logs the photodiode output from the first pulse   You do not see pulses on the b
16.  also not transmit to PC at higher rates than 180 Hz in regular data log   mode    In order to log data or transmit it to PC at rates  gt 1500Hz  you must be in turbo mode    This will put you into a special mode where the Laserstar can handle data at high   speed while curtailing other functions  To log in turbo mode  do as follows    1  Make sure that the Laserstar is in the correct energy measurement range  If not   press    range     adjust and press    exit       2  From the energy measurement screen press    menu    and then select    data log      Press    go     Now press    turbo       3  With the up down buttons  indicate the laser frequency and then choose the log  size  i e  how many points you want to log  Now press    go     The measurement will  begin and the Laserstar will indicate when the file is full     Note1   The Laserstar does not measure frequency in turbo mode  Therefore you have to  indicate the frequency before starting the measurement  This means you cannot  measure pulses with frequencies which vary during the measurement run     Note2   The Laserstar will not respond to any buttons until the turbo logging session is  over  This applies either for data logging or data transmission in turbo mode     c02  ENERGY  2 875J    READY    go pause reset       Figure 11  Energy Log Screen    39    4 6 Advanced Display Screens  4 6 1 Beam Splitter Attenuator Screen    If you have a measurement setup with a beam splitter or attenuator and wish to  display the
17.  are individually  calibrated against NIST traceable standards over the entire operating range of  wavelengths for both filter out and filter in   The calibration curve is normalized to the  correct absolute calibration at 632 8 nm using a HeNe laser against a reference meter  traceable to NIST     The spectral sensitivity curve of the detector as well as the spectral transmission  curve of the filters is fed into the head EEROM and this information is used to set the  gain to the proper value at wavelengths other than the wavelength the instrument was  calibrated  When the user selects his wavelength on the Laserstar  the correction  factor for that wavelength is applied     10 2 2 Accuracy of Calibration    Since the instruments are calibrated against NIST standards  the accuracy is  generally  2  at the wavelength the calibration has been performed  The maximum  error in measurement will be less than the sum of the calibration accuracy  linearity   inaccuracy due to errors in the wavelength curve and variations in gain with  temperature  The linearity of the photodiode detector is extremely high and errors due  to this factor can be ignored  The maximum error due to the above factors is given in  Table 4 below     90                                                       WAVE ERROR  FILTER OUT   LENGTH PD300   PD300 3W PD300  PD300  3A IS  UV IR   200   250nm  5   250   360nm      3     360   400nm  10   10   3     400   950nm  3   3   3  5   950   1100nm  5   5   5   5  10
18.  as well as the current reading and the  trend  This allows you to determine if you have reached maximum power  The tune  function also has an audio option so you can follow a rising or falling audio tone in  tuning your laser     The screen is completely autoranging  Therefore  as soon as the cursor goes over the  top or under the bottom edge of the display  it re scales to put the cursor back to the  middle of the screen  This allows you to devote all your attention to tuning the laser   without having to worry about the Laserstar settings  You can also use the audio tone  feature so you do not have to look at the Laserstar at all while tuning the laser     4 4 2 3 1 To use the Laserstar to fine tune laser power    1  From the bargraph power measurement screen press the menu button    2  Press       gt     until    tune    is highlighted then press    go     Now press    setup       3  Set the power expansion scale to  50    25    10    100 or  75  as desired   using the    value    soft key   The default value is 50   If the laser power will  change a lot  use a large value  if a little  use a small value    4  Set the horizontal sweep time to the rate desired  The default value is 1 minute    Set the audio to on or off as you desire   See below    6  Now save these settings by pressing    save    and press    exit    to enter the tune  screen  If you want to save the settings only until the instrument is turned off   press exit without save     g1    4 4 2 3 2 Measuring w
19.  detector where the source overfills the detector  It measures  the light intensity per unit area in units of Lux or Foot Candles     Note     Because the CIE heads have a spectral response similar to the human eye  wavelength selection is disabled  Other than that  operation is the same as other  PD300 type heads     52    Chapter 6 Operation with Pyroelectric Heads    Warning   Before using the head for power or energy measurement  check that your laser  power  energy or energy density do not exceed the head ratings as listed in tables  6 and 7 with the head specifications  Otherwise  there is a risk of damaging the  absorber     With the pyroelectric head  you have been supplied a test slide with the same coating  as on your pyroelectric detector  You can also obtain this slide from your dealer  You  should use this slide to test the damage threshold with your laser pulses  If the slide  is damaged  then either enlarge your beam or lower the laser energy until damage is  no longer seen     6 1 Pyroelectric Detector Heads    When a pulsed heat source  such as a laser  is directed at the detector head  a  temperature gradient is created across the pyroelectric crystal mounted in the head   An electric charge is produced which is proportional to the energy absorbed  The  detector head has sophisticated circuitry unique to Ophir  patented  that determines  the baseline before the pulse is received  measures the voltage after a pre determined  interval  amplifies it and holds i
20.  laser power or energy before the splitter or attenuator and not the actual  value impinging on the measurement head the beam splitter function allows you to do  this  For example  if you are splitting off 4 5  of a laser beam into the meter and you  wish to display the full beam  do as follows     1  From the main power screen  press the menu button and select    advanced      Press    go     Select    attenuate    and press    go       2  Press    set    and then    value    to select first number  Press    set    again then     value    again to select second number and so on  For example  in the screen  shown below  4 5  has been selected     If you press    exit  exit  esc    you will return to the standard measurement screen     If you want the instrument to start up in the attenuation screen  follow the instructions  for instrument startup defaults in section 3 5     2 354W     52 31 w     4 5004  set percent exit       Figure 12  Attenuation Screen    Note 1   The attenuator screen only affects the value displayed on that screen and nothing  else  The power and energy scales remain the same  the damage thresholds  remain the same etc  The attenuator screen is only an extra screen to show a  different calculated number introduced by the user     40    Note 2   The attenuation value is stored in the Laserstar and not the head so if it is set  any  head connected will come up with the value chosen     Note 3   If you have a dual channel Laserstar  the function will only 
21.  otherwise  there is a risk of damaging the absorber  With the pyroelectric head   you have been supplied a test slide with the same coating as on your pyroelectric  detector  You can also obtain this slide from your dealer  You should use this slide  to test the damage threshold with your laser pulses  If the slide is damaged  then  either enlarge your beam or lower the laser energy until damage is no longer seen     57    Note   To measure pyroelectric energies properly  it is important that the head is not  grounded to the optical bench  Make sure that the head is isolated electrically  from the ground  The PE head has been supplied with an insulating mounting post  for this purpose     6 4 1 Energy Average Power and Frequency Measurement    Unlike most pyroelectric energy meters  the Laserstar measures the pulse frequency  to a high degree of accuracy and is therefore able to display pulse frequency and  average power as well as pulse energy  The following paragraph will describe how to  set up and measure these functions     6 4 1 1 Zeroing Head Against Laserstar Display    Unlike with the Nova display  there is a slight dependence of pyroelectric reading from  display to display  Therefore  for most accurate pyroelectric energy measurements it is  necessary to zero the pyroelectric head against the Laserstar display it will be used  with  After this is done  the head is    conditioned    to work with the particular Laserstar  the zeroing was done against  It is not nece
22.  save the new calibration value  Note that in the save screen  the previous  value of the calibration is shown for your reference  As soon as you save  this  value is set to the newly saved value     Note 1   For units which have different calibration factors  e g  C02  YAG  or VISIBLE  press   laser  to select correct type of laser before calibration  When recalibrating using  the CAL resistor  the other two settings will change proportionately  If you wish to  change the calibration of only one laser type  use the function  Power with laser        88    Note 2   The unit retains for your information the current sensitivity of the head in A W in  engineering notation  The original value is retained at the factory  You should  record the original setting so you can return to it if you wish     10 1 4 3 Recalibration from a Known Source of Laser Power   Energy    1  Set the instrument to power or energy depending which you wish to recalibrate    2  From the bargraph power screen press the    menu    button and select    advanced      Select    calibrate    and press    go     select    calibrate     press    go    and press     power with laser       3  Using the    value    button select the proper laser wavelength   See note 1 below     4  Press  power energy with laser    5  Usethe    select    and    value    buttons to choose the appropriate laser wavelength  and averaging period  See note 1 below   6  Apply the known laser power energy    7  Adjust up down until the powe
23.  startup with as default and press    save       2  Connect the head and switch on the Laserstar  From the main measurement  screen  press the menu button located on the right side  Select    configuration      Press    go       3  Press  value  to choose  power  or  energy  for the desired startup screen  The  Laserstar will now start up in power or energy measurement mode  respectively   when switched on    4  Press the arrow key until the laser wavelength is highlighted  Now press  value to select the laser wavelength you want to be the default  The Laserstar will  now be correctly calibrated for the absorption of the detector at the wavelength  of the laser type selected  If you wish to choose another wavelength  See Section  6 2 1    5  Select    Max Pulse Length     to choose the default pulse length  See section 6 2 2  for details    6  Select    Energy Range    to select the default energy range    7  Select    Average Over     to select the time period you wish to average over  See  section 6 2 2 2 for details    8  Now press    save    when all selections have been made  and press exit  If you only  want to save the changes until the Laserstar is turned off  exit without saving     1 1 Energy  Power or Exposure Measurement    Warning   Before using the head for power or energy measurement  check that your laser  energy  energy density and average power do not exceed the head ratings as  listed in tables 6 and 7 with the head specifications at the end of the manual  
24.  threshold 3X  more sensitive by choosing LOW and 3X less sensitive by choosing HIGH   Choosing LOW will allow you to measure pulses of low energy more accurately but  may cause the measurement to false trigger from noise also    Single Shot Energy  This determines the range for single shot energy that has to  be chosen independent of the RP energy ranges     7 9 Zeroing    This function pertains to the thermal head of the RP and is described in the section on  thermal heads  chapter 3 5     7 10 Customizing Gain    This function allows you to automatically tailor the gain of the RP to your measurement  situation     The RP photodiode circuit has a factory default gain setting  This gain setting is a  compromise that allows the RP to operate under a wide variety of laser energies   wavelengths  pulse widths and repetition rates  Since the RP must not go into  saturation under any of these circumstances  the gain is set lower than necessary in  many cases and therefore the system noise is higher than necessary     73    The customize function allows you at the push of a button to set the gain to an  optimum value for the laser parameters you are measuring now  It is recommended to  use this function and the user can always return to the factory default if desired  To  customize the gain settings     1  From the main measurement screen  press    menu        more        customize    and     go       2  Wait for the power reading shown to be steady  Press    next       3  When th
25. 0   the reading on the display is blanked  If the user has chosen to activate the  audio warning in the screen that appears with not head attached  an audible warning  will appear when the highest range has been exceeded     Low Battery  When the battery is almost discharged  the flashing message  BAT   appears  When the low battery message appears  the Laserstar should be connected  to the charger  It will operate normally and charge slowly while connected to the  charger  If the user has chosen to activate the audio warning in the screen that  appears with not head attached  an audible beep will sound every minute when the  battery is low  Also  the Backlight will turn off when the battery is low     Sat  When the photodiode current exceeds 1mA  and the detector starts to saturate   or the pyroelectric voltage exceeds the maximum  the message  sat     saturated  is  displayed    10 5 Troubleshooting    10 5 1 Laserstar Display       Problem Cause Remedy   Instrument will not operate after   Check that charger provides 200mA  being completely discharged  and connected to charger           Instrument operates with Battery is low  Recharge overnight with the  charger but not with battery Laserstar turned off for 14 16 hours  If the  alone  Laserstar still doesn t work with battery     then the NiCd battery is probably dead   Replace battery                 95    10 5 2 Thermal Heads  Energy Measurements       Problem    Cause Remedy       Instrument triggers on  background noi
26. 10 6um  5   15   Table 5    Maximum Measurement Error due to Wavelength  10 3 3 Recalibration from a Known Source of Laser Energy    1  Set the instrument to energy    2  Setto the proper energy range    3  From the main screen press the menu button and select    advanced     Press    go     and select    calibrate  Press    go     again    4  Select    Max pulse length and use the    value    button to set to the proper scale  Set  the laser wavelength to the correct value   Using the  value  button  select the time period over which you wish to average   Now press  select  and value  to choose the correct laser wavelength  See Note  1 below  Now press  go     5  Apply the known laser energy    6  Adjust up down until the energy reading on the screen equals the known energy   The energy calibration factor will change accordingly    7  Press exit  The new value will be saved     Note 1   For metallic type heads  when the calibration is changed at one laser wavelength   the overall calibration of all other wavelengths changes proportionately  For BB  type  the calibration is only changed for the wavelength set     94    10 4 Error Messages    The Laserstar displays various error messages when operated outside its normal  range     Over range  When the power or energy being measured exceeds the range of the  measurement scale being used  the over range message Is displayed  but the reading  still appears on the display  If the power or energy exceeds the maximum by more  than 1
27. C using the    Starcom PC package supplied with the Laserstar  The data can be transmitted in real  time as well according to the following table     Power Measurement RS232 GPIB  Power log period 12s to 600hr 12s to 600hr  Max points stored onboard  per file  5400 5400   Max points in StarCom real time log 6000 6000   Max points direct to PC file in real time unlimited unlimited    33    Energy measurement RS232 GPIB    Max real time data logging rate     gt 30Hz  gt 1500Hz     Onboard data logging rate     gt 200Hz  gt 200Hz   Data transfer from instrument to PC  500points s   gt 1500points s     Max points stored onboard 59 400 59 400     1   The above refers to the rate for logging every single point  Above that rate the  instrument will sample points but not log every single point     2   The rate shown is for GPIB with the instrument in turbo mode  In regular data  logging mode  the maximum rate for every single point is 180 points sec   See the section 4 5 6 2 for further details     4 5 Energy Measurement  4 5 1 Setup    Switch the Laserstar on  If it is in power mode  press the left soft key    energy    to put it  into energy mode  Figure 10      The energy mode is manual ranging  press    range    and then use the arrow keys to get  to the proper range  The correct range is the lowest one which is larger than the pulse  energy measured  If you plan to use the same setting often  use the  configure   function as described in Section 4 3  before switching the Laserst
28. KW cm  BB  1500W 1500W 6KW cm  BB LP  5000W 5000W 5KW cm  BB LP  8000W 8000W 5KW cm  BB LP  L30A EX 30W 1KW cm  EX  FL250A EX 250W 1KW cm  EX  PD10 20mW 50W cm  PD  PE10 2W 50W cm  PE  PE25 10W 10W cm  PE  PE50 20W 10W cm  PE  PE50 DIF 40W 500W cm  PE   Table 6     PD   Photodiode    Max Power Specifications of Heads    P   P type volume absorber for short pulse lasers  HE HE1  volume absorber for high energy pulses     EX   Excimer type  volume absorber  PE   Pyroelectric metallic or black absorber    BB   Broadband surface absorber  high power density  LP   Broadband surface absorber for highest power density    101                                           Absorber Max Energy Density J cm   Type Pulse Length   10ns Ips 300ps  P 10 10 10  HE HE1 5 10 100  BB 0 3 0 5 5  EX 0 5 0 6 4  PE  Metallic 0 1 0 2 4  PE  BB 0 3 0 3 1  PE DIF 1 5 3 40  PE BB DIF 3 3 10   Table 7     Maximum Energy Densities for Various Absorbers  Single Pulse     102                  BB Thermal  lt  300w      Pulsed Laser Damage Threshold     10K LP  gt  1500W     BB thermal  gt  1000W    z HL  E 1K   Z  S Pyro    o metallic     gt  100        G  iv Z  w wW  Z O  p  10     HE HET 5 5  1    EX 0 5     BB 0 3  metallic  LP  0 1  HL W cm       0 01     1 i l 1  10   10   10   10   10   10   10  10   107 10  14  1 nsec 1 psec 1 msec 1 sec                                              PULSE WIDTH IN SECONDS    Laserstar User Manual    Ophir P N 1J06020    29 March 2005    Rev 2 45 3    103    
29. Laserstar to normalize the power   1  From the main power screen  press the menu button and select    normalize      2  Press    go    and when ready to normalize  press    apply     From then on  the power    will be proportional to the normalized power value   see Figure 7   3  If you wish to return to the the regular display  press    exit        30    24 50W Normalization    1 021    laser range       Figure 7  Thermopile Normalize Screen    4 4 2 6 Power vs  Time Graph and Data Logging    The Laserstar in datalog mode has the option of graphing the laser power against time  and storing the data in up to 10 permanent storage files with up to 5 400 data points in  each file  While measurements are being taken  the Laserstar will record data until the  screen is full or until the reset button is pressed  With the RS232 or IEEE adaptor  the  data can be read out to PC in real time or later after storage  Power logging has the  following characteristics     To use the Laserstar to graph power vs  time     1     From the bargraph power measurement screen  press the menu button and select     data log     Press    go       Now press    setup    to access the setup screen  Now select the time scale   maximum power and minimum power you wish and press    exit     If you wish to  save these settings as defaults  press save before exiting    Now press    log    and you will be in the power graph screen  See Figure 2   Press     go    and the measurement will begin immediately  The 
30. NIST calibrated standard meter   The meter can be switched to give the exact calibration at the various wavelengths   Argon  YAG  and  where applicable  C02    For other wavelengths  the user should  consult the wavelength graph and use the closest of the above settings    The EX type detector is calibrated by measuring the ratio of absorption in the UV to  that at 515nm  The total absorption is measured in both cases using an integrating  sphere  The detector is then calibrated with an argon laser and given a correction  factor from this ratio     10 1 2 Linearity and Accuracy of Ophir Thermal Heads  10 1 2 1 Linearity    The linearity of most Ophir thermal detectors is specified to be 1  over the specified  power range of each particular instrument and is tested by Ophir from time to time  On  CAL models  the built in CAL resistor can be used to directly measure the linearity over  the range of powers of the CAL resistor  On non CAL models  a CAL resistor is  temporarily affixed to the sensor  For those models for which the linearity is not tested  over their entire range  randomly chosen sample models are tested periodically over  their entire range  The test is performed with a high power laser that can cover the  entire detector range using a rear leak detector for comparison  This rear leak  detector is a low power Ophir detector that has previously been tested for linearity   Thus  in all cases  the linearity of the detectors is traceable to electrical standards     85
31. OPHIR      Laser Measurement Group       LASERSTAR    LASER POWER ENERGY MONITOR  USER MANUAL    OPHIR OPTRONICS       www ophiropt com    Table of Contents                               Chapter 1 Introduction  How to Use This Manual             ccsscsscssssesseeseeseeeseeseeseees 3  Chapter 2 Quick Reference             csscssessessessessessesstessesseesessesseeseeseesessnssueseeseeseesseees  2 1 Getting Started  i inven  2 2 Thermal Headsin a  2 3 Photodiode Heads tes a   2 4 Pyroel ctric HO ads isinisisi innri 11  Chapter 3 The Laserstar Display Unit    cccsssssccscssecsessessessseeseeseesesseesteaseeseesees 15  3 1 General De SCription            sscssssssssssesssesssesssecssscsesecneessessuessneeeuessneesneesneesseeans 15  3 2 Smart Connectors and Multi head Operation              cccssecsecsessesseesseesees 16  3 3 Soft Keys   3 4 Power up and Shut DOWN           cccccssssssssssssssesssscsessseeseeeseteseteseteseeeseteseeeseeens 18  3 5 Laserstar Functions which are independent of Head Type                    18  3 6 Hardware FUNCTIONS           sc ssesseeseecessesseessesseeseeeeceeeeseeseesseeseeasesess   Chapter 4 Operation with Thermopile Absorber Heads   4 1 Thermopile Absorber Heads            ssssssesseseeeeees    4 2 Selecting Settings from the Laserstar Screen  4 3 Startup and Configuration of Defaults            ssssssssssssesssesserrserssrrsserrsersseesss  44 Power Measurement    45 Energy Measurement                ssecsecsesseessesseeseessesseeseeseesseeseee
32. Use this if you want to measure over a specific time period  independent of the number of pulses    Pulses  Use this if you want to measure over a specific number of  pulses    Stop mode  Choose    after time out  if you want to measure over a specific  time period Choose    after pulses    if you wantto measure a  specified number of pulses  Choose manual only    if you want  to stop the measurement manually only   Note that you can  also stop the measurement manually in the other modes as  well     Now press    start    and you will be in the exposure measurement screen     60    PE50 BBH exposure    224 Ts    00  00  49 3 172P  setup reset start stop       Figure 15  Exposure Screen    4  Press    start    again  The Laserstar will start summing laser energy exposure   When you wish to stop measuring  press  stop  or wait for the    time out  selected  in the setup screen  The Laserstar will now display the total energy emitted during  that period  the time elapsed and the number of pulses during the period  If you  wish to reset reading to zero before another reading  press  reset     5  To return to the measurement screen  press exit then esc     Normalize    When you wish to show the readings as compared to some standard reading  the  normalize function allows you to do this  After setting    normalize     all subsequent  readings will be shown as a ratio to the reference reading  For instance  if you  normalize the reading at an energy of 1 5mJ  then subsequent reading
33. ain power measurement  screen  press the menu button and select  advanced   Press  go    and select   calibrate  Press  go   Now press  power with laser     Select the next laser  wavelength sent with the disc  Adjust up down until the CAL factor is the same as  that sent with the disc for the wavelength shown  Now press  exit  and     permanent save    to save that value    Now repeat step 3 for the other laser wavelengths  Each time  select a different  laser wavelength and enter the proper calibration factor sent with the disc for that  wavelength     98    10 6 1 4 Battery Replacement    If the Laserstar battery is defective and does not hold a charge  a replacement can be  ordered from your agent  The new battery is installed as follows     1  Turn the Laserstar display upside down and unscrew the 6 Phillips screws on the  bottom panel of the Laserstar     2  Raise the bottom panel and disconnect the battery connector  red and black leads   from the PC board     3  Insert the new battery into the same location in place of the old battery  When the  upside down Laserstar is placed so the battery is near you  the black lead will be  on the left     4  Reconnect the battery connector to the terminal BT1  Make sure the sloping side of  the connector is away from you  Reclose the panel and replace the 6 screws     sloping  side    red orange       Figure 20  Battery Connection    99    Chapter 11 Laserstar Specifications    11 1 Laserstar System Display Specifications       I
34. an zero the instrument to eliminate any offsets   He also can set the Laserstar to give an audible warning if the head power energy  limits are exceeded and if the battery is low     3 5 1 Startup Screen    The Laserstar can operate in the standard power energy measurement screen or can  display the information in specialized forms  an attenuation screen  a power energy  density screen or a screen with limits  Set the screen to the type you want the  instrument to start up in and press select to go to the next option  For more details on  the various screen options  see section 4 6      3 5 2 Line Frequency    The Laserstar is factory set for 50 Hz  If your country uses 60 Hz you must reset the line  frequency    To set the line frequency  press    select    to get to the line frequency option  then press  value to select the correct line frequency then press save     3 5 3 RS232 Baud Rate    If you are using the Laserstar for RS232 communication with PC  the baud rate can be  set by selecting  baud rate  and selecting the appropriate rate then pressing save     3 5 4 Audio Warning    The audio warning indicates when you are over range on the highest range and also if  the battery is low  When audio warning is activated and the maximum power or energy  rating is exceeded  a beeping tone will sound  If the battery is low  a beep will sound  every 5 seconds  To activate the audio warning  select    audio warning    and then  press  value  to set to  on   Then press  save      3
35. and select     configure       Press  go     Now press  value  to select filter in or out to be the default    Now select    Power Range     and choose the manual power range you wish to be  the default or choose autorange or dBm    Now select    Average Over     and choose what period to average power over or  leave the default    NONE       Now select    Wavelength    to choose the startup laser wavelength    When finished press    save    and    exit     If you wish to save these values only  until the instrument is turned off exit without saving  For further details on  configuration see Section 5 3     Power Measurement    Warning     Do not exceed maximum head limits for power  energy  power density and  energy density as listed in tables 6 and 7 in section 11 2 Head Specifications   Otherwise  there is a risk of damaging the absorber     To expand the bargraph scale  5x about the present reading     1   2     Press the  zoom    button   Press the  zoom    button again to return to the ordinary bargraph     To subtract background and set current reading to zero     1  From the main power measurement screen press  offset    2  Press  offset  again to cancel  See Section 5 4 5 for full details     To use the Laserstar to fine tune Laser power     1  From the main power measurement screen press the menu button then select   tune   Now press  go  then  setup     2  Set the percentage range of the power scale to be displayed by repeatedly  pressing the value key    3  Now se
36. anel       OFF ON    HEAD INPUT B OUTPUT HEAD INPUTA 12VDC BACK    LIGHT    LO O    Ratio Multihead model rear panel showing head input A and head input B       OFF ON    Figure 1  Laserstar Rear Panel View    off on backlight    t    12 34 mw    energy zoom offset range menu       Figure 2  LaserStar Top View    To switch the Laserstar on   1  Flick the toggle switch on the right side of the rear panel up  Note that there are  three positions  down is off  the center is on and the upper position is on with    backlight on   See Figure 2      The unit will switch on  and the display will appear  Note that the head must be  plugged in before the unit is switched on     To switch the Laserstar off   1  Return the toggle switch to its lowest position     2  If you wish to save the current Laserstar configuration  use the  configure   function   See section2 2 2      To set line frequency     1     w    Disconnect the head and switch off then on again   Head Disconnected  will  appear    Select    line frequency       Press  value  to change to the correct line frequency    Press  save  and the line frequency will be saved     To zero instrument     Disconnect the head  turn off then on again   Head Disconnected  will be  displayed    Make sure instrument is not in an electrically noisy environment and is  undisturbed    Press  zero  and then press    go     When zeroing is completed     zeroing  completed will be displayed    For thermal heads  zeroing with the head may also be 
37. ar off the first time   When the Laserstar is switched on again  it will automatically go into the same range  until a different configuration is saved  Note that the present range is displayed on the  lower right of the energy measurement screen as shown in Figure 10     30A CAL energy      J    power laser READY range menu    30J       Figure 10  Thermopile Energy Screen    34    4 5 2 Measurement    To measure energy of a single pulse  set up as above and fire the laser  The display  will go blank while the energy is being integrated  After 2 4 seconds  the correct  energy will be displayed  When the Laserstar is ready to measure a new pulse   ready   will appear on the screen and flash on and off   See Figure 10   The next pulse will  automatically blank the screen  and the new value will be displayed  If you fire  another pulse before  ready  appears  the reading may be inaccurate or may not be  displayed     4 5 3 Minimum Energy Threshold    If the Laserstar is used in a noisy environment or where there is a high level of  background thermal radiation  the instrument may trigger spuriously on the noise or  background radiation  It would then fail to measure the intended pulse  Since there is  always some degree of noise or background radiation  the instrument is designed not  to respond to pulses below some preset minimum size  This  Minimum Energy  Threshold  is typically set to 0 3  of full scale of the selected range  If this level is  found to be too sensitive for t
38. argraph display until after the power stabilization  period  After that period  the RP calculates the absolute energy of the initial pulses on  the assumption that the laser power was the same as after the stabilization period  chosen from the first pulse on  If this assumption is approximately true  then you have  the true pulse energies from the first pulse on recorded in the energy logging data   These values can be seen in the data files recorded and sent to PC as well      lt 800  ENERGY  2 875J    every  pulse  go pause reset       Figure 19  Energy Log Screen    7 11 5 Missing Pulses and Jitter  The RP has the ability to check for missing pulses and record them during data  logging  When this function is activated  the RP will also measure the time jitter    between pulse to pulse     To set for missing pulses recording  from the main energy screen go into the setup  screen as follows     76    menu rp config go and select    Missing Pulses Detection     Using the    value    key   choose  lt 40Hz or  gt 40Hz depending on your laser pulse rate  If you want to save this  setting as the startup default  press    save    before exiting this screen     The RP will now record    missing    instead of the pulse value in the data log screen  when a pulse is missing     Note 1   The missing pulses function will only operate at up to 100Hz     Note 2   The missing pulses function will only operate correctly if the missing pulses are    less than 5  of the total     Note 3   Jit
39. ated fast processor circuitry can handle up to 10 000  pulses per second  sampling pulses      Accuracy  The RP accuracy is determined by the thermopile  which measures the average power  with  3  accuracy or better  This is better than the accuracy typical for pyroelectric  detectors or the accuracy using beam splitters especially for measuring high energies  and average powers  In most cases  standard deviations of fractions of 1  can be  measured     66    High Duty Cycle   Unlike pyroelectric heads  the RP is able to measure the pulse energy of lasers whose  pulse width is a large duty factor  i e  the laser is on most of the time  This means that  you can measure for instance a diode laser pulsing at 3Hz with 100ms pulse width  where the duty cycle is over 30      7 4 Temporal Pulse Shape    The RP head also contains a second fast photodiode with nanosecond rise time  This  can be used to measure the laser pulse shape in addition to other parameters  mentioned above     Connect the BNC output on the RP head  Figure 17  into a fast oscilloscope  The  oscilloscope must have a 50 Q input impedance or be terminated with a 50 Q  impedance  or the response time will be on the order of microseconds rather than  nanoseconds     7 5 Measuring Repetitively Pulsed Lasers with the RP    When you switch on the Laserstar with an RP head  it immediately enters the  measurement screen and is ready for action  For first time use or when changing  measurement parameters  you must access t
40. ation of Power using the Built in CAL Resistor    If your thermal head has the Ophir CAL resistor then you have a traceable absolute  calibration source available just by measuring the voltage and current into the CAL  resistor and multiplying by the calibration factor provided by Ophir     Calibration instructions     1  Set the instrument to power  If you want to recalibrate energy  this can be done  from a known energy source  See section 10 1 4 3    2  Fromthebargraphpowerscreenpress the    menu    button and select    advanced      Select    calibrate    and press    go     select    calibrate    press    go    and press    power  with CAL       3  Using the    value    button select the proper laser wavelength   See note 1 below     4  Using a DC power supply  apply power to the CAL resistor  black and red terminals  at the back of the head   For highest accuracy  use the same power as you intend  to measure  but do not exceed values in Table 3  Section 10 1 4 1    5  Maximum voltage applied should not exceed        resistance of CAL resistor  x  maximum power     6  The voltage should be measured at the input to the head and the current in series  with the current loop    7  Adjust the up and down buttons until the power reading on the screen reads the   voltage in volts  x  current in amps  x  current calibration factor for your model  and laser or as shown on Laserstar     8  When the correct power is shown on the screen  press    exit    and    permanent  save    to
41. cseesseesseesseesseeceteestesseeeneeeatersess  75 Measuring Repetitively Pulsed Lasers with the RP   7 6 Setup for RP   Rapid Pulse measurement           sssssssssssrsssrrssersrrssrrrseesseees  7 7 Choosing Default Pulse Width Setting                sssssessessessesseessesseesstesses  78 Configuration for Power and Single Shot Energy      7 9 ZOOL OUNG sess eaatcats Goer acaat E   7 10 Customizing Gain   fis   7 11 Data Logg in Gurania eee eile deeeiiaes  Chapter 8 Operation with the Dual Channel Laserstar           csscssssseeeseeneenees  8 1 Operation with One Head Plugged in Only                       8 2 Single Head Operation with Two heads Plugged In      8 3 Dual Head Operation with Two heads Plugged In         Chapter 9 Circuit Description          c sccsccscsessessesssesssecessccsesseesteesesseesteseesteeseeseentenses  Chapter 10 Calibration  Maintenance and Troubleshooting           ccscscseceees  10 1 Calibration of Thermopile or RP Heads            ssscssecssesseesseesseessessseesseeesesses  10 2 Calibration of Photodiode type Heads          ccssscssessessessssssessesseesseesseesees  10 3 Calibration of Pyroelectric type Heads   10 4 Error M  ssages erise eea ii  10 5 Troubleshooting  RE  10 6 Man ena E a a aani iE  Chapter 11 Laserstar Specific ations            cccssssssssssssssscssesssessessesscsnesneesseeseeneens  11 1 Laserstar System Display Specifications   11 2 Head Specific ations          cscssccssecssesssessesseessesssessesseesseesseeesnesseesseesseessee
42. cted     25    Select    Average over     and select the period you wish to average power readings  over  This feature is especially useful for lasers with nonsteady output  See  section 4 4 2 4 for details    Select    Power Range     using the arrow keys  Now choose the manual power  range you wish to be the default or autorange  The Laserstar will now be in the  range selected when switched on with the head in question    Select    Energy Range     using the arrow keys  Now choose the energy range you  wish to be the default or autorange  The Laserstar will now be in the range  selected when switched on with the head in question  Note that there are only  manual ranges in energy    Select    Energy Threshold    if you want to change the energy threshold  See  section 4 5 3 for details       Now press    save    and    exit    and the defaults will be saved  If you only want the    changes saved for the present  exit without save and the changes will be saved  until the Laserstar is switched off only     11  Switch the Laserstar off then on by means of the switch on the rear   See Figure 4      The Laserstar will enter the mode of operation last saved with the  configure   operation     4 4 Power Measurement    The next section describes the procedure for basic power measurement as well as  more advanced features with thermal measuring heads     4 4 1 Power Measurement  Basic Operation    1     3     Switch the Laserstar on by means of the switch on the left side   See Figu
43. desired  The default value is    none    which means that the pulses displayed are  individual pulses with no averaging  See section 6 2 2 2 for further information on  averaging     The Laserstar is now ready to measure energy pulses  The energy reading will  operate in the following manner if  none  selected as the averaging period  following  every 1 5th of a second period  the Laserstar will display on the screen the next pulse  which arrives  The Laserstar will hold the display of the latest pulse until a new one  arrives  If an averaging period is selected  the Laserstar will collect pulses at up to  180Hz and will then display the average of those pulses  When a new pulse triggers  the Laserstar     Trig    will appear on the screen  At over 5Hz  the  T  stops flashing and  appears all the time     Note   The Laserstar is able to record each pulse at up to 180Hz even though it is  displaying at a much lower rate  This data can be stored in a data file and  recovered and transmitted later     The Laserstar also displays other information at the same time  see Figure 14   The  additional information displayed is as follows    upper line  the laser wavelength in the left corner  the head type in the center and  whether the instrument is in power or energy mode in the right corner  A or B  signifying which head input is the main one  see the section on multihead inputs     2nd and 3rd line  information about the average period and status    4th line  the energy value and u
44. e message    customization completed    appears  press    save    to save the  new gain value and exit    4  If you want to restore the factory default values  from the main measurement  screen press    menu        more        go        customize        go    and    reset        7 11 Data Logging   7 11 1 Power Logging   As with other type heads  the RP allows you to log data and store it on board and send  it to PC  Data logging for average power is exactly the same as for Ophir thermal  heads  Follow the directions in section 4 4 2 7 for power data logging  For RP energy  data logging  do as follows    7 11 2 RP Energy logging    The RP Laserstar has the option of displaying successive energy points in a scrolling  vertical bargraph form  While measurements are being taken  the Laserstar will    record data  Up to 59 400 points are stored See section 4 4 2 7 1 for more information  on data storage and handling     74    7 11 3 To use the RP for graphical display of energy     2 1    2 2    Make sure that the RP is in the correct energy measurement range    From the main measurement screen  press    menu    and select    data log     Press     go     Press    setup    to set up measurement parameters for head and sample rate   If you have more than one head attached  choose the head you wish for data  logging    When set to    Sample Rate  every pulse   The RP will sample every pulse up to  400Hz in absolute mode and 600Hz in relative mode  Above that rate it will sample  at
45. e voltage is proportional  to the reading on the display and scaled such that full scale equals 1 00 volts  The  output can drive up to 2mA into an external device     CD OUTPUT HEAD INPUT 12VDC    OX Ke O      OFF ON    BACK       Standard model rear panel    HEAD INPUTB LCD OUTPUT HEAD INPUT A 12VDC Back    LIGHT    AEON oy Jo  0    OFF ON       Ratio Multihead model rear panel    Figure 4  Laserstar Rear Panel View    3 6 5 LCD Contrast Adjustment    The LCD contrast can be adjusted with the blue contrast adjustment knob     22    Chapter 4 Operation with Thermopile Absorber Heads    Warning   Before using the head for power or energy measurement  check that your laser  power  energy and energy density do not exceed the head ratings  See tables 6  and7 in section 11 2 Head Specifications     If the head is a water cooled type  ensure that the cooling water is flowing at an  adequate rate  see table below  Also  note that the reflectance from the absorber  could be as much as 10  and with C02 lasers  the reflected beam can be quite  specular  so it is advisable to provide a beam stop for the reflected beam with the  highest power lasers                             HEAD TYPE LITERS MIN US GALLONS  PER PRESSURE PER MINUTE  MINUTE BAR  8000W 7 1 5 2  5000W 4 5 0 8 1 2  1500W 2 5 0 5 0 7  300W 1 0 0 3 0 25          Table 1   Minimum Flow Rates For Water Cooled Heads    4 1 Thermopile Absorber Heads    When a radiant heat source  such as a laser  is directed at the absorber head
46. ed to laser  pulses    Instrument shows frequency  which is too high       Possible electromagnetic interference from   pulsing laser is causing misreading and or   false triggering   Check the following    1 Head is mounted to stand using insulated  plastic rod provided with instrument  and  not metal rod    2  Try keeping cable away from bench    3 Move head display further away from  EMI        10 6 Maintenance    10 6 1 Maintenance of Thermal Heads    10 6 1 1Tracing the Signal from the Head to the Laserstar    1     With the instrument on  apply an approximately known amount of power to the  head  This test can be performed using either a laser or an electrical power  supply and the calibration resistor    Estimate the approximate signal current that should be developed by the head by  multiplying the input power by the head sensitivity shown in the head specification    tables 6 and 7     Using a multimeter set to current  unplug the head from the Laserstar and check    that this current appears between pin 1 and 9 of the D type plug     10 6 1 2 Disc Replacement    If you have ordered a replacement disc  it has been factory calibrated for the  wavelength region shown on the calibration sheet sent with the disc   calibrated at the wavelength requested  or for the 3 wavelengths listed  If it is a CAL  type disc  a CAL Factor is also included to use for electrical calibration checks on your    power energy meter     97       It has been    Please follow the following step
47. eeeateseesesaeeatesseees  46 Advanced Display Screens          ssscssscssscssesssessseessessstesseesstessteeseesstesseeestesssss  47 Special Thermopile Heads              s  ce0e   Chapter 5 Operation with Photodiode Type Heads   5 1 Photodiode  Heads s 2      s ciccsscsee  arsecsersgsennspsspseeiecde eas seeaa can aeetnae aitiaceeen ened  5 2 Setting up the PD300 to display the user s chosen wavelengths           45  5 3 Startup  Default  scoeccz2sccssvcccteczszctveurwesesesssttgen vecteneeteesvoniventenee teeuaeneeenastes  5 4 Selecting Settings from the Laserstar Screen   5 5 Special Photodiode Heads          c sssecssscssscssssescssssesssesssesssesseessneeseeeseeessesses  Chapter 6 Operation with Pyroelectric Heads          c  scssessecssessssseeseessestesteeseeneenees  6 1 Pyroelectric Detector Heads            seccsecsecsesssessessesseessessesesesseeseeseesaeeseesseees  6 2 Selecting Settings from the Laserstar Screen   6 3 Startup and Configuration of Defaults   6 4 Energy  Power or Exposure Measurement            sescsecsssseeseesesseeseeseeeeteees  6 5 Measuring Repeating Pulses of High Energy              ss sssessessessesseeseeseeeee  6 6 Energy LOTin cantante    6 7 Pyro Scope Adapter    Chapter 7 Operation with RP heads           ccsscsssssscssssessessessessseeseesceseeseeseesteeeeseesees 64                   7 1 General Description                7 2 Principle of Operation   7 3 Advantages of the RP   74 Temporal Pulse Shape         csssccsscsseesssessessseessee
48. ent   press  energy    To go from the energy measurement screen to power measurement press the  power  soft key     4 3 Startup and Configuration of Defaults    If you are planning to do measurements of the same type most of the time  you can set  up the Laserstar so that it automatically starts up in the measurement mode desired   Since these settings are saved in the head  smart connector   you can save different  settings for different heads     To set the Laserstar to start up in desired screen and to save the chosen  configuration     1     With the head disconnected  switch on the Laserstar and press    value    to select  the desired startup screen   Note that this selection is saved in the Laserstar and  not the head so will be operative for all heads   For more details on the screen  types see section 4 6    Now switch the Laserstar off  connect the head and switch on again  From the  main power or energy measurement screen press the menu button located on the  right side    Choose  config  with the arrow keys  Press  go     Press  value  to choose  power  or  energy  for the desired startup screen  The  Laserstar will now start up in power or energy measurement mode  respectively   when switched on    Select    Laser      Now press  value  to select the laser type you want to be the  default  The Laserstar will now be correctly calibrated for the absorption of the  detector at the wavelength of the laser type selected and will startup with the  laser wavelength sele
49. ent see chapter Chapter 4      To set up for these functions  do as follows   From the main measurement screen press    menu    and choose    config     Press    go      Press    select    to choose the function you want to set and    value    to choose the    desired option  When you are finished  press    save    and then    exit     If you only want  to save the choices until the instrument is turned off  press    exit    without    save        72       In the following section  we explain the various functions     1     Startup  Choose whether you want the start up default of the RP will be energy or  power measurement    Laser  Choose the laser wavelength   This can be chosen from other menus as  well     Average Over  This function allows you to average the power readings over a  period from 1s to 1 hour  The average is a moving average  i e  the average is from  the present backwards  For instance  if the user chooses average over 10s  The  average will be from the beginning till now for the first 10s and from 10s ago till the  present afterwards    Power  This determines the power range that can be fixed or autorange   This can  also be chosen from the RP config screen     Energy Mode  As mentioned  the RP can measure single shot energy as well as  repetitive pulses  With this function you choose RAPID PULSE that is the RP mode  or SINGLE SHOT  If you choose  singe shot  there are additional menus to choose   Energy Threshold  In the SINGLE SHOT mode  you can make the
50. er    2  If the display shows energy units of J  mJ etc  then press  power  to switch to  the power measurement mode     2 4 5 4 To measure Total Energy Exposure     1  Press the menu button and select    exposure     Press    go       2  Select parameters then press    start     Press    start    again  The Laserstar will  start summing laser energy exposure  If you wish to stop before the chosen  period  press    stop       3  If you wish to reset the reading to zero before another reading  press   reset     4  To return to the main screen  press    exit    and    esc        Chapter 3 The Laserstar Display Unit    3 1 General Description    The model Laserstar laser power energy meter represents a new level of  sophistication  sensitivity  compactness and accuracy  coupled with ease of operation   It can operate with thermal  pyroelectric and photodiode heads  It has smart  connector technology  Simply plugging in the head configures and calibrates the  Laserstar to operate with that head     The Laserstar can graph power or energy vs  time  It displays power measurements in  both digital and analog form at the same time  It will also autorange  so you do not  have to set scales  or it can manual range if you wish  It will remember what mode you  were using before you turned it off and will return to that mode when turned on  You  can zoom in on the present reading  or subtract background  Although the calibration  information is stored in the smart connector  you can recal
51. es the optical trigger  However  in some cases  the optical trigger will    not work and you should select one of the thermal trigger settings   low    med  or   high   You should use thermal triggering on the 3A P in the following cases     The wavelength is outside the range of the of the photodiode  i e  outside of 0 19    1 064um   You are measuring the energy of a shuttered pulse of a continuous laser  In this  case  the peak power is too low to trigger the photodiode    For some other reason  the instrument does not trigger on the  optical  setting     When using the optical trigger  the instrument should be able to measure down to 0 1    0 2 mJ  depending on wavelength  When using the thermal trigger  low  med or high    you should not attempt to measure less than about 1mJ     4 5 4 Measuring Pulses of Very Low Energy    When it is necessary to measure pulses of very low energy  i e   less than 0 5  of the  maximum range of the instrument  the following two alternative methods allow greater  accuracy to be obtained     1     2     A continuous train of pulses may be fired  and the average power measured using     power    mode  The energy per pulse can be calculated by     Energy per pulse   Average power   Pulse Repetition Rate  A train of a known number of pulses may be fired  and the total energy measured    in    energy    mode  This train should not exceed 5 seconds duration  The energy  per pulse can be calculated by     36    Energy per pulse   Total Energy   N
52. esass    Chapter 1 Introduction  How to Use This Manual    The Ophir Laserstar is a microprocessor based Laser Power Energy Meter providing a  broad range of measurements  displays  and data handling options  It operates with  thermopile  pyroelectric and photodiode heads  and uses smart connector technology   Just connecting the head configures and calibrates the instrument  The Laserstar also  is available with the multihead option where it can work with 2 heads at once as well  as other options     This manual tells you what you need to know to make full use of the Laserstar for all  your laser measurement needs  It includes a  Quick Reference    Chapter 2  to allow  you to perform basic measurements immediately  without reading the whole manual     The main measurement sections  Chapters 4  5 and 6 include a general description and  a section detailing operating options     Chapter 2 Quick Reference    2 1 Getting Started    The Laserstar is equipped with  soft keys   That is  the functions of the keys change  as indicated by the legend above each key   See Figure 2      To connect head to the Laserstar Display    Insert the 15 pin D type connector of the measuring head cable into the socket marked   Head Input  on the rear panel of the Laserstar display   See Figure 1  If you are using  the multihead version of the Laserstar and are using only one head  it is preferable to  use head input A     LCD OUTPUT HEAD INPUT 12VDC AC       O s a o A LIGHT    Standard Model Rear P
53. ess the  zoom    button    2  Press the  zoom    button again to return to the ordinary bargraph      To subtract background and set current reading to zero     1  From the main power measurement screen press  offset    2  Press  offset  again to cancel  See Section 4 4 2 2 for full details     To use the Laserstar to fine tune Laser power     1  From the main power measurement screen press the menu button then select   tune   Now press  go  then  setup     2  Set the percentage range of the power scale to be displayed by repeatedly  pressing the value key    3  Now select the horizontal sweep time you desire  Now select if you want an  audio tone for tuning or not    4  Now press  exit  and use the screen to tune the laser power  See Section 4 4 2 3  for full details     To use the Laserstar to measure Laser energy     Verify that the display shows  energy  in the upper right hand side of the display   If the display shows  power  and units of W  mW etc  then press  energy  to  switch to the energy measurement mode    The energy mode is manual ranging  Press  range  then use the arrow soft keys  to select the proper range  Press save if you want this range to be the startup  default  Now press  exit   The correct range is the lowest one that is larger than  the pulse energy measured  Note that the range will be displayed in the lower  right corner of the main display    When the Laserstar screen flashes  ready   on and off  fire the laser  See  Section 4 5 for full details  
54. f     5 3 Startup Defaults    If you are planning to do similar measurements most of the time  you can set up the  Laserstar so that it automatically goes into the measurement mode you want upon  startup  Since these desired settings are saved in the head  smart connector     you  can save different settings for different heads  The following section describes how to  set up the Laserstar with the PD300 head so that it will start up in the desired  configuration     Setting up the startup defaults   See section 5 4 for further details     1     With the head disconnected  turn on the Laserstar  Press    value    to select the  startup screen you wish to startup with as default and press    save       Now turn on the Laserstar with the head attached  From the power measurement  bargraph screen  press the menu button located on the right side and select     configure     Press    go       Press  value  to select filter in or out as the startup default  For details of filter  in out See Section 5 4 3    Now select    Power range    to choose the manual power range  autorange or dBm  you wish to be the default   The Laserstar will now be in the mode selected when  switched on with this head connected    Now select    Average over    to choose the averaging period for power  measurement  If you do not want averaging  choose the value    NONE    See  section 4 4 2 4 for details about averaging    Now select    wavelength    and use the value button to select the default startup  wa
55. face down on the table     Switch the instrument off then on again with the head connected    Press the menu button then select  advanced   Press  go      zero    should be  highlighted    Press  go  and follow steps 2  4 above     20    3 6 Hardware Functions  3 6 1 Backlight    The backlight illuminates the display from the rear and is operated by the same switch  as turns the instrument on and off  See Figure 4   The Laserstar backlight consumes  considerably less power than competing instruments and therefore it can operate from  the battery even when the charger is not plugged in  Nevertheless  since it does  shorten the time between charges  it is recommended to operate the instrument  without the backlight when the instrument is operated without the charger  The electro  luminescent backlight also does have a finite lifetime of about 10 000 hours  Therefore   it is recommended to use only when necessary     3 6 2 Swivel LCD    The Laserstar is equipped with a swivel LCD so the display can be tilted for easy  viewing  The display angle is variable from 0 to 55 degrees  The LCD can fold down flat  for transportation and when the Laserstar is folded flat  it is only 225 x 190 x 52 mm  the  most compact unit of this type on the market     3 6 3 Charging    The Laserstar is operated by a rechargeable battery  To charge the battery  plug the  charger into the jack labeled  12VDC  on the back panel   See Figure 4   Note the  polarity of the charger  The Laserstar charges full
56. from a known source of  power or energy following the instructions in Section 10 1 4 3     On the rear center of the sensor disc is the exclusive Ophir calibration heater  The    disc can be heated by electrical power instead of by radiation by applying electrical  power to the terminals at the rear of the sensor head     86    By accurately measuring the voltage and current applied  the calibration can be  checked and if necessary  readjusted  taking into account the calibration factor  See  Section 10 13   10 1 4 1 Guidelines for using the CAL resistor for calibration    1  Do not exceed the maximum electrical power stated in the table                                            MODEL POWER APROX  MAX  VOLTAGE   3A Series 1 Watt 7 Volts  10A Series 10 Watts 17 Volts  30A Series 30 Watts 35 Volts  150A Series 100 Watts 65 Volts  L250A Series 200 Watts 95 Volts  300W Series 125 Watts 65 Volts  1500W Series 200 Watts 95 Volts  and above   Table 3     Maximum Power to apply to sensor CAL resistors    2  Use the resistor for calibration only    Turn off the power as soon as calibration is done    4  Do not operate the calibration resistor when the sensor disc is under radiation  the  excessive heat may burn out the calibration resistor    5  To eliminate residual non linear effects  it is advisable to perform electrical  calibration at about the same power level as that to be measured  so long as the  rating of the heating resistor is not exceeded     w    87    10 1 4 2 Recalibr
57. g  the energy with and without the beam splitter    3  Ophir has two instruments specifically designed for high energy density pulses   They are the PE50 DIF model pyroelectric heads with a diffuser and the RP  thermal photodiode heads   Both of these heads can measure energy densities up  to several Joules cm 2     Contact your Ophir dealer for details     62    6 6 Energy Log    This is the same as for thermal heads  except you do not have to wait for  reset   See  sections 4 5 6  and 4 4 2 7 1     Note   For Pyroelectric heads logging energy the only setup option is    sample rate     The  user can choose the number of pulses he wishes to log   from    every pulse    to    1  out of 1000    pulses   but not the time period     6 7 Pyro Scope Adapter    A special adapter can be purchased which is plugged in between the head connector  and the Laserstar head socket  This adapter has a cable and BNC connector to attach  to an oscilloscope  It enables the user to see on the scope pulses proportional to  energy up to the maximum pulse rating of the head  Unlike the Laserstar display  which  samples at the maximum rate  the scope adapter shows every single pulse     Activate the scope adapter as follows     1  Plug the adapter into the Laserstar display and then plug the head plug into the  adapter  Plug the BNC connector into a standard 1MQ scope input    2  Press    menu     select    scope    and press    go     The display will now say    in scope  mode     Select the approp
58. han 3 pulses per second  the  RP switches to a ten second sample time  This allows it to collect enough pulses  for a meaningful sample    Choose the correct setting depending on whether your laser pulse rant is greater  than or less than 3Hz     70    9  Laser Pulse Width   The RP is able to measure lasers with a wide range of pulse widths from  nanoseconds or below to hundreds of ms  In order to operate properly  the RP  must be set up for the laser pulse width  This sets up the integration time during  which the laser pulse energy is measured  The shorter the pulse width setting  the  higher the maximum repetition rate available and the less noise in the signal  The  factory default gives 6 choices for the laser pulse width setting that vary from head  to head  Choose the setting that is the smallest one which is larger than the  maximum pulse width of the laser  If the default values given are not appropriate  then you may choose other default values  See section 7 7 for this     Warning   If the pulse width of your laser is longer than the setting in the Laserstar  the  energy reading will be erroneous     10  Scope Mode  As with pyroelectric heads  if you have the oscilloscope adapter  the RP is able to  operate in scope mode where you see every pulse up to high repetition rates on an  oscilloscope   At high repetition rates  the ordinary RP mode only samples a  fraction of the pulses   If you want the RP to start up as default in scope mode  choose    scope  OFF     Fo
59. he range goes to the next higher one  The ranges  are arranged in factors of 1  10  100  etc  When the reading falls below 9  of full scale   the range changes to one range lower  This change only occurs after a few seconds  delay  This provides overlap  hysteresis  to keep the Laserstar from flipping back and  forth when reading close to the end of the scale     Manual Range   There are certain disadvantages to autorange since it changes scale  even if you don t want it to do so  If you want to measure the same range all the time   it is better to use manual range  To select manual range  press  range   select  auto   then press  exit  to return to the measurement screen  The correct range to select is  the lowest one that is larger than the largest expected measurement  If you wish the  Laserstar to be in manual mode when switched on  save this configuration before  switching off  See section 4 3      To choose auto or manual range  follow these steps   1  From the main power measurement screen  press  range      2  Select the appropriate manual range or autorange using the arrow buttons   3  Press  exit  to return to the main power measurement screen     24    4 2 3 To choose Power or Energy Measurement    The Laserstar thermopile heads can measure both power and single shot energy  See  sections 4 4 and 4 5 for details  To change from power to energy measurement or  vice versa do the following     1     To go from the main power measurement screen power to energy measurem
60. he setup screen  Press    menu    then  select    RP configuration    and press    go    and you are ready to set the parameters for  your application  The following section describes the parameter option in detail     67    30A P RP filter  out A    22 69m       10 00Hz Trig 50us  power laser  gt 3Hz range  Figure 18  RP Energy Screen    7 6 Setup for RP   Rapid Pulse measurement    Your RP is capable of measuring rapidly pulsing laser pulses of any energy  from  fractions of a millijoule to 10 Joules or higher  To ensure that the instrument gives you  the best possible results  you must first set up your instrument to match the type of  laser that you use  This is described below   After you have chosen the desired values from the setup screen  press    save    before  exiting to save these as the default value  If you want these values saved only until the  instrument is turned off  exit without saving   In order to choose the desired options from the main measurement screen  press     menu     select    RP config    and press    go     Select the various options by pressing     select    and    value      1  Energy or Power  With    energy    highlighted  press    value    to choose whether you want the startup  default mode of measurement will be    energy    or    power     If you choose     energy     the RP will measure energy per pulse upon startup  If you choose     power     the RP will measure the average power   2  Filter OUT or IN  Now select    Filter     and c
61. he user s particular environment  it may be altered by the  user  The threshold should not  however  be raised higher than necessary  This will  cause a degradation in the accuracy of energy measurements of pulses below about 4  times the threshold level  The factory setting of energy threshold is  med  for medium   If the unit triggers on noise  set the threshold to  high   If you are measuring small  energies and the unit does not trigger  set the threshold to  low      To set the energy threshold     1  From the power or energy measurement screen  press  menu      2  Select    config     Press    go     Select    energy threshold      Energy Threshold      MED     will be displayed    3  Press    value    to select  high    med   or  low  threshold  For the 3A P head   optical  is also available and is the default choice   See below    4  Press    exit    and    esc    to return to main measurement screen  If you wish to save  this value  press  save in the configure screen before exiting   See also section  4 3      35    Note  to users of 3A P type heads      The 3A P type head has special circuitry enabling it to trigger on energy pulses of  very low energy  This is accomplished by a special photodiode trigger which  alerts the Laserstar that a pulse has been received and to start integration  This  allows the 3A P to react to smaller pulses than would be measurable just by  triggering on the rising thermal signal as in other heads  When you select  optical   this enabl
62. hoose filter    OUT or    IN    depending on whether you are  going to work in the highest range which requires the filter to be inserted in the  head     Note   With the filter out setting on the most sensitive energy range  you can sometimes  get triggering from room light  If you want to work on the lowest range and get  triggering from room light  you have to somewhat darken the room     68    3  Absolute or Relative Mode  The next choice is the operation mode    op mode      This can be    absolute    or     relative     The absolute relative function determines whether the energy  displayed will be absolute or relative  Ordinarily     absolute    will be selected  In this  mode  the RP will calculate the actual pulse energy of each pulse based on the  average power  as explained in section 7 2     Principle of Operation      If the average power is very low and you are interested in pulse to pulse variation  rather than absolute energy  then you should select    relative    mode  In this mode   the RP ignores average power and displays a convenient value proportional to  pulse energy  but which is independent of average laser power  For low average  power values  this average could be highly inaccurate  Since the standard  deviation depends only on the pulse to pulse variation  the standard deviation  reading will still be valid in the relative mode  even though it will then be expressed  in relative units   In this manner  you can use the RP to measure the pulse to pul
63. ibrate from the screen and  store the new calibration in the head  You can also zero the Laserstar at the touch of a  button     The main instructions are clearly shown on the screen so you should not have to refer  to this manual very often  Above all  the Laserstar has advanced circuitry and digital  signal processing for excellent sensitivity  signal to noise ratio  accuracy  and  response time  It also has special circuitry to reject electromagnetic interference     The multi head version of the Laserstar also allows you to measure two heads  independently or measure the ratio sum or difference of two heads  The Laserstar is  also available with IEEE 488 GPIB interface as an optional extra  Contact your Ophir  agent for details     3 2 Smart Connectors and Multi head Operation    The Laserstar display is versatile and can operate with either thermal  pyroelectric or  photodiode type laser measuring heads  The head configuration and calibration  information is stored in an EEROM in the head connector plug  This means that when  the head is plugged in  the Laserstar automatically identifies the head type  calibration  and configuration  The user does not have to adjust anything     Note   The Laserstar automatically loads the head information when first turned on  so  when changing heads  the display should be turned off  the new head plugged in  and then be turned on again     When no head is plugged in and the Laserstar is turned on  the display indicates  Head  Disconnec
64. ith the Tune Screen    You can now use the tune screen to adjust and maximize your laser power   Figure 5    Note that you can not only see the trend  but also see if you have already reached a  higher value before  The numbers on the left side show you the   present power reading as well as the highest one reached till now  Press    max    to  reset the maximum reading to the present time  Now measure the power of the laser  and adjust the laser until you determine that the power has reached a maximum     28       Figure 5  Power Tune Screen    4 4 2 3 3 Using the Audio Tune Function    The Laserstar has an exclusive audio tune facility  The instrument can generate a  rising or falling audio tone to indicate higher or lower power  To engage the audio  feature  simply choose    Audio  on    in the setup screen and save if you want this  feature to be the default     4 4 2 4 Average    When a laser output is fluctuating or unstable  it is useful to measure the average  power over a certain period  The Laserstar gives you this exclusive feature  allowing  averaging over periods varying from 1 s to 1 hour     The average mode works as follows     As soon as the main power measurement screen  See Figure 6  is entered and the  instrument is set to average mode  the instrument displays the average of readings  over the period since the screen was entered up to the present  At the same time  the  small vertical bargraph on the right starts to fill up  See Figure 6   When the time pe
65. laser pulses during the first few seconds of operation since the  average power measurement has not yet stabilized and the RP needs this value in  order to calculate the absolute energy  The    stabilize    function allows you to  obtain the absolute energy of the first few seconds of operation as well when  operating in the energy log mode  See section 0 on energy logging    It operates as follows   The user chooses a stabilization time in the RP configuration screen of up to 15  seconds during which the average power measurement value is assumed to reach  a stable value  When the user starts logging energy pulses  the RP then collects  the relative energy values during this initial period   See para  3 above for  explanation of relative vs  absolute   After the initial period  the RP uses the now  stabilized average power value to calculate the absolute energy of the pulses  during the first several seconds of operation     Note   Stabilize will only give correct readings for the absolute energy during the initial  period of operation if the laser average power is approximately constant from the  first pulse until the end of the stabilization period  If this is not the case  then use  relative mode as described in para 3 above     8  Laser Frequency  gt 3Hz or  lt 3Hz   The RP is designed to measure pulses over a wide range of rates  from 0 5Hz to  several KHz  For most rates  the sample time for calculating averages and  standard deviations is two seconds  For rates less t
66. lect the horizontal sweep time you desire    Now choose audio tone to be on or off    5  Now save and exit to use the tune screen  See Section 4 4 2 3 for full details      gt     2 4 Pyroelectric Heads  2 4 1 Zeroing Instrument Against Head    For most accurate calibration  you should zero the pyroelectric head against the  Laserstar it is being used with  Proceed as follows     Make sure the head is in a quiet environment and not subject to pulsed radiation  Plug  head into Laserstar and turn on  Press    menu     select    advanced     press    go    and     go    again  Press    start     When    zeroing completed    appears  press    save        2 4 2 Selecting Chosen Wavelengths  metallic type only     Press  menu  and select  wavelength   Press  go     Select the wavelength you wish to change and press  go    Using the up down buttons  change to the desired wavelength  Press    exit       Repeat steps 2 and 3 for other wavelengths desired  Up to 6 wavelengths may   be selected    5  When finished  press    save        exit    and    esc    to return to the main   measurement screen  If you only want a temporary change  Press    esc    without  save        EOIN    2 4 3 To set type of laser being used    From main measurement screen  press  laser  until the correct laser type or  wavelength is displayed     2 4 4 Setting Startup Configuration    1  With the head disconnected  switch on the Laserstar and press the    value     button to select the type of startup screen 
67. ll now display the sum head A plus head B     Note 2   Allowed combinations   In dual head mode  any combination of Ophir smart heads can be combined  In  ratio or difference mode  the permitted combinations are  thermopile or  photodiode  vs   themopile or photodiode  or else   pyroelectric  vs   pyroelectric    In ratio or difference mode  power vs  power or energy vs  energy is permitted but  power vs  energy is not     Note 3   Data logging   In dual head operation  data logging is only allowed in ratio or difference mode   The data points for each channel are stored inside separately but are displayed by  the Laserstar only in the combined form  If the data is uploaded to PC using the  Ophir Starcom for Windows program  then the data can be displayed either in the  combination form or each channel individually     79    Chapter 9 Circuit Description    The Laserstar has three circuit boards  an analog processor  a digital processor and a  power supply     Analog Processor    The signal from the detector head enters the analog processor board and passes  through EMI protection components to a differential transimpedance preamplifier   From there it is further amplified by a programmable gain voltage amplifier and passes  to a dual slope integrating analog to digital  A D  converter  All calibration data for the  analog processor is stored in a memory chip on the analog processor board  There are  no adjustable components  trimmers etc   in the Laserstar except for the dis
68. llected can be stored for later recovery and  transmission   See section 4 4 7 2 1   However  the display can only display at rates  up to 15Hz  At higher rates  if the user has chosen    NONE     the Laserstar will  display individual pulses sampled at a rate of 15Hz  If the user has chosen to  average over a time period  the instrument will display the average of readings  over the period from the time the screen was entered up to the present  At the  same time  the small vertical bargraph on the right starts to fill up  See Figure 14   When the time period of the average is reached  the bargraph is filled and from  then on  the average is a running average over the average period backward in  time  For instance  if the average period is 30 seconds  at 15 seconds  the average  is over 15 s   at 30 seconds itis over 30 seconds  at 5 minutes  it is over the period  from 4 5 to 5 minutes  30 seconds back from the present etc       6 3 Startup and Configuration of Defaults    If you are planning to do repetitive measurements most of the time  you can set up the  Laserstar so that it automatically goes into the measurement mode you want upon  startup  Since these desired settings are saved in the head  smart connector     you  can save different settings for different heads     56    To set the Laserstar startup mode and to save the chosen Laserstar configuration     1  With the head disconnected  turn on the Laserstar  Press    value    to select the  startup screen you wish to
69. log  processor and converts them into a voltage that can be processed as necessary     80    The digital processor is also responsible for controlling the basic analog processor   the fast analog processor  reading the keypad  driving the display and communicating  via RS232     The CPU obtains calibration and capability data from a memory chip in the plug of the  detector head and configures itself accordingly  Recalibration data and saved settings  are also stored in this memory chip  In addition  the digital processor contains two  extra connectors   one is for connection to a second analog board if the user has  purchased the multihead ratio version of the Laserstar  The other is for the external  IEEE 488 communications package  check with your Ophir representative about  availability of the IEEE option     Power Supply    The power supply provides the internal DC supply voltages for the analog and digital  processors  It also contains the battery charging circuit and the AC supply for the  backlight  The power supplies for the charger and the main digital DC voltage operate  in switch mode     EMI Protection    The digital processor and the whole Laserstar instrument is protected by EMI  protection component on all signals that pass in and out of the box  In addition  EMI  protection is added Internally to prevent disturbances to the normal functioning of the  instrument   With the above mentioned components  the instrument meets the  requirements of the EEC with respec
70. nction subtracts background from the signal  If the ambient environment  has a thermal background  so that the Laserstar shows a nonzero power reading even  when there is no laser  you can subtract the background using the zero function  For  example  the Laserstar display reads 0 1 Watts when the laser is blocked  and 20 5  Watts with laser power applied  In this case  the true power is 20 5   0 1   20 4 Watts   To subtract the background  press the    offset    button while the laser is blocked  The  Laserstar will now read zero  and the 0 1 Watt background will be subtracted from all  subsequent readings  The laser power reading will thus be 20 4 Watts     When    offset    is engaged  the legend is highlighted  To cancel  press    offset    again   If the    offset    is engaged  and you wish to subtract out a new value of the background   press    offset    twice  The first press will cancel the old value  and the second will  activate a new value  If you suspect that the Laserstar has a permanent zero offset   then disconnect the head while it is in the power measurement mode  If the Laserstar  still shows a similar offset even when the head is not connected  the instrument s  internal zero should be reset  See Section 3 5 6     27    4 4 2 3 Tune  see Figure 5     The Tune function  an exclusive Ophir feature  makes adjusting your laser to its  maximum power easier than ever before  Unlike a bargraph or mechanical meter  this  display shows graphically what came before
71. nd from the main  measurement screen  press  menu     3  Using the arrow keys  select  config   Press  go     4  Press  value  to choose  power  or  energy  for the desired startup screen    5  Select    Laser     Now press  value  to select the laser type you wish to be the  default    6  Now select    Average over    to select what period you wish to average power  over or select the default   none   See section 4 4 2 4 for details    7  Now select    Power range    and press  value  to choose the manual power  range you wish to be the default or choose autorange    8  Now select    Energy Range    and choose the default energy range    9  Now select    Energy threshold    if you wish to change the energy threshold  See  section4 5 3 for details    10  Press  save  to save all present settings then  exit  and  esc  to get back to the  main measurement screen     Further details on configuration in section 4 3     2 2 3 Power or Single Shot Energy Measurement    Warning   Do not exceed maximum head limits for power  energy  power density and  energy density as listed in tables 6 and 7 in section 11 2 Otherwise  there is a  risk of damaging the absorber     To use the Laserstar to measure Laser power     1  Verify that the display shows  power  in the upper right hand side of the display   2  If the display shows  energy  and units of J  mJ etc  then press  power  to  switch to the power measurement mode     To expand the bargraph scale  5x about the present reading     1  Pr
72. necessary  See section  3 5 6     2 2 Thermal Heads    2 2 1 Use of Laserstar with thermal type heads    To set type of laser being used     1   2     While the Laserstar is off  plug in the head then switch it on again    From the main measurement screen  press the menu button   laser  will be  highlighted  Press  go  and select the appropriate laser wavelength    If you want to save this new wavelength as the startup default  press  save    Return to main screen by pressing  exit  then  esc      To choose manual or automatic ranging in power measurement     POLDA    From the main measurement screen  press  range     Select the appropriate manual range or autorange    If you want to save this new range as the startup default  press  save     Press  exit  to return to the main measurement screen  The range will be  displayed in the lower right corner of the main measurement screen     To choose power or energy measurement     1  To go from the main power measurement screen to the energy measurement  screen  press  energy     2  To go from the energy measurement screen to power measurement  press     power        2 2 2 Setting and Saving the Startup Configuration    1  Turn on the Laserstar with the head disconnected and press    value    to select  the type of screen you wish to start up with   This choice is saved in the  instrument and not the head so the Laserstar will start up from the screen  selected for all heads     2  Connect the head  turn the Laserstar off and on a
73. ng     1  When you have gathered enough points press    pause    or wait until the data    register is full and    FULL    flashes  Press    exit    and you sill be asked if you want to  enter the file control system  Press    yes        32    FILE CONTROL  Temp  5400     0  1  0   9  0    2  0   10  0  3  0   Free Space  54000  esc       Figure 9  File Control Screen    2  When the screen is entered  the data in the latest power graph will be stored  temporarily in register    Temp     Press    select    to select the permanent register  you wish to use for storing the data and press    save     The data in the    Temp     register will be transferred to the selected register  This data storage is permanent  and the data is not lost when the instrument is turned off  If you try to save into a  file which already has data  a warning will appear that the new save will erase  previous data in the file  If you want to delete the data from all the files  press     delete       3  Pressing    view    will get you into a screen showing a summary of the data in the  selected file  If you press    data     you will enter a screen showing all the data  points collected until now  Press    home    to get to the first page of data  Press     end    to get to the last page and use the arrow keys to go through the data page  by page  Pressing    esc    will get you back to the previous screen     Data Transmission to PC  The data stored in the Laserstar file system can be transmitted to P
74. nits    5th line  The maximum pulse length and current energy scale    6th line  the current soft keys     59    6 4 1 3 Power Measurement    Make sure the instrument is in the power measurement mode indicated by the  legend    power    in the top right of the screen and the units of mW  W  etc  If it is  in energy measurement mode with units of mJ  uJ etc   then press the    power     soft key or alternatively  press the menu button and then the left  power  soft key   The Laserstar can be configured to start up in power measurement mode when  turned on  See section 6 3    Press the range button then use the arrow keys to select the proper range  The  correct range is the lowest one that is larger than the pulse energy measured   Note that the proper range is related to the pulse energy even though we are  measuring average power    Press exit to return to the measurement screen     6 4 2 Measuring Total Energy Exposure    The Laserstar has the ability to sum the total energy of a number of pulses over a given  time period or number of pulses  This gives the total energy  exposure  over that time  period  For example if the laser is pulsing at 30 times s at ImJ per pulse and you  measure the exposure over 20 seconds then the total exposure is 30 x 1 x 20   600mJ    In order to measure exposure  do the following     1   2     Press the menu button and select    exposure     Press    go        Setup the desired measurement parameters in the setup screen as follows    Timeout  
75. nput Specifications  Thermal  Photodiode       Input Ranges    15nA   1 5mA full scale in 16 ranges       Ato D Sampling rate    15Hz       Ato D resolution    17 bits plus sign   0 0009  resolution        Electrical accuracy     0 25    20pA new   0 5   50pA after 1 year       Electrical input noise level    500nV or 1 5pA   0 0015  of input range  3Hz        Dynamic range    9 decades  1 109        Input Specifications  Pyroelectric Heads       Input Ranges    23mV   6V full scale in 9 ranges       Ato D Sampling rate    40KHz maximum       Ato D resolution    12 bits no sign  11 with sign  0 02 0 05 res         Electrical accuracy       2mV new     4mV after 1 year       Electrical input noise level    2mV       Dynamic range    3 5 decades  1 4000        General Specifications       Detector Compatibility    Thermopile  photodiode and pyroelectric       Analog output    0 1 Volt with 3mV  0 03   resolution  100 ohms impedance       Analog output accuracy     0 2   2mV relative to display                         Dimensions 228W x 195D x 54H   Mass 1 45 Kg    Display 240 x 64 pixel super twist LCD  Display digit height 17mm   Backlight EL  Operates from charger or battery  Bargraph segments 240   Battery 3x D size  4 0A h NiCd battery built in        Charger input    DC  12  25V  5W  AC  9   18V RMS  5W  charge time 10   14Hr  15   20Hr if operating       Operation between charges    18 Hrs  thermal and photodiode  14 Hrs  pyroelectric heads        Data Logging and Com     
76. of photodiode detector signals    65    A removable filter is provided in front of the photodiode to increase the dynamic range  of energies measurable by the RP  If the energy per pulse of the laser is within the  highest range  the 10x reduction filter is inserted so as to prevent detector saturation     The thermopile detector also has a built in heater resistor that allows calibration to be  performed by the electrical substitution method  Since the calibration of the RP  depends only on the thermal head power calibration  the calibration procedure for an  RP head is the same as any Ophir thermal head  See section 10 1   Calibration of  Thermal Heads for details of the calibration procedure     Since the photodiode is extremely linear  it is possible to make a fast detector circuit  that will give an output proportional to pulse energy  within fractions of a percent  The  time between pulses is measured to great accuracy   1us   therefore the accuracy of  the pulse energy is limited only by the accuracy with which the thermopile measures  average power     7 3 Advantages of the RP    Because the RP measures pulse energy with two different detectors  it has several  advantages over other types of repetitive detectors     High Energy Density   The RP volume absorber can stand up to as much as 10 Joules cm  for short pulses  and 60J cm  for long pulses  This allows measurement of large energies and high  average powers     High Repetition Rate  The fast photodiode and associ
77. olume absorbers    P  pulse  type absorber   The models with the P suffix  for use with pulsed lasers  have a special absorbing  glass with an absorbance of 95  2  over the operating range  Since the surface is a  glass  its reflectivity does not change even if damaged or melted locally    HE HE1  high energy  absorber   The HE and HE1 types have a particularly high damage threshold for pulsed and  repetitively pulsed lasers of both the short and long pulse variety and are useful where  the highest pulse energies and average powers are used    HL  high energy long pulse  absorber   The HL absorber is the absorber of choice for high energy repetitive pulses longer than    2ms  For these long pulses it has flatter spectral response than the HE HE1 and  withstands higher pulse energy and average power     83       Percent absorption  100                                        0 1 100     E    BB therm      amp     EX        P Wavelength  um                         Percent absorption  100                                     o4 1 10    Wavelength  um                      Graph 2   Absorption of Ophir Absorbers vs  Wavelength    84       10 1 1 2 Method of Calibration    The absorption of the various Ophir thermal absorbers can vary from disc to disc   Therefore  all Ophir absorbers are individually calibrated against NIST traceable  standards     Ophir power energy meters with the broadband or P type absorbers are individually  calibrated by laser at several wavelengths against a 
78. or graphical display of energy   regular mode    1     Make sure that the Laserstar is in the correct energy measurement range  If not   press    range     adjust and press    exit       From the energy measurement screen press    menu    and then select    data log      Press    go     Now press    log    and the energy log screen will appear  Press    go      Now fire a pulse when the  ready  legend flashes  The measurement will  immediately begin  When the screen is full  the display will scroll and the latest  points will continue to be displayed  When the data register is full     FULL    will be  displayed  If you wish to see finer variations in the graph  press the    zoom    button   When    reset    is pressed  the Laserstar is ready to begin a new measurement  If  you want to store the data before making a new measurement  see section  4 4 2 7 1 on data storage    Warning     Pressing  reset  will erase all previously stored data     When the    stats    button is pressed  a statistical summary of the data on the  screen is shown  If you wish to return to the graphical display  press    graph     If  you wish to permanently store the data in a file  press exit and then choose    yes     to enter the file system  See section 4 4 2 7 1 for further information     Note     Zoom can only be pressed after at least one pulse has been recorded     38    4 5 6 2 Turbo mode    In regular data log mode above  you can only log every point at up to 180 points per   sec  You can
79. play  contrast adjustment on the rear panel     Fast Analog Processor    In addition to the above basic analog processor  the Laserstar contains a second fast  analog processor  This supports certain Ophir heads that read energy pulses at higher  rates than can be supported by the basic analog processor  but provides less overall  accuracy than the above processor  The fast processor consists of an EMI filtered  voltage input that is passed to a first stage of mild voltage attenuation  This is then  amplified and passed to an A D converter  Calibration data is stored on the same  memory chip mentioned above     Digital Processor    The digital processor is built around a Motorola MC68332 32 bit central processing unit   CPU  with a 16 bit bus  On the digital processor board there is a pair of sockets  containing the upgradable FLASH chips which can be upgraded in situ by your Ophir  agent  In addition there are a pair of static RAM chips totaling 256Kbytes of RAM  The  large amount of RAM allows storage of relatively large amounts of data permanently  even when the instrument is turned off   See section 4 4 2 7 1 on data storage   The  digital processor receives signals from the basic analog processor and converts them  into a measurement of current in amps  When used with thermopile heads the signal is  then processed by a sophisticated digital filter that speeds up the effective response  time of the head and rejects noise  In addition  it converts signals from the fast ana
80. r Measurement    Warning   Do not exceed maximum head limits for power  energy  power density and  energy density as listed in tables 6 and 7 in section 11 2 Head Specifications   Otherwise  there is a risk of damaging the absorber     With the pyroelectric head  you have been supplied a test slide with the same coating  as on your pyroelectric detector  You can also obtain this slide from your dealer  You  should use this slide to test the damage threshold with your laser pulses  If the slide is  damaged  then either enlarge your beam or lower the laser energy until damage is no  longer seen     2 4 5 1 To choose Energy or Average Power Measurement     1  To go from the main energy measurement screen to power measurement  press   power     2  To go from the main power measurement screen to energy measurement  press   energy      2 4 5 2 To use the Laserstar to measure Laser Energy and Frequency     1  In measurement mode  verify that the units are mJ  pJ etc  and    energy    is  displayed in the upper right corner  If not  press the    energy    key    2  Press  range  then use the arrow keys to select the proper energy range  The  correct range is the lowest one that is larger than the pulse energy measured   See Section 6 4 for details    3  Press  exit  to return to the bargraph measurement screen     2 4 5 3   To use the Laserstar to measure Average Power     1  Verify that the display shows power units in W  mW etc  and    power    is  displayed in the upper right corn
81. r a description of how to operate in scope mode see  section 6 7     7 7 Choosing Default Pulse Width Setting    As described in para  9 below  the RP head has factory default settings of laser pulse  width  If you want to change these to more convenient values  please do as following     1  From the main measurement screen press    menu    and choose    more     Press    go     and choose    pulse width     Press    go     With the arrow key  select a pulse width  you wish to change and press    go     With the    up    and    down    keys choose the  value you desire and press    save    then    esc        exit        esc    to return to the  measurement screen  Note that in the pulse width setup screen  you are shown  the maximum laser pulse rate for each setting of the pulse width  These values are  given in the following chart     71       Maximum Laser Repetition Rate for Given Pulse Width Setting  Maximum Pulse Width Setting  ms     0 01 0 1 1 10 100 1000   100000   10000 T  o  T  ao   1000 g    2     a   100     E  5      x   10                Graph 1  Maximum Laser Repetition Rate for Given Pulse Width Setting    7 8 Configuration for Power and Single Shot Energy    As part of its equipment to measure the energy per pulse of repetitively pulsing lasers   the RP contains a complete Ophir thermal power energy head  As such  it is able to  measure power and single shot energy with the thermal head alone   For further  information on power and single shot energy measurem
82. r allows you to choose autorange or manual ranging when  the reading is in watts  In autorange mode  you do not have to change scales  When  the reading of the meter or bar is more than 100  of full scale  the range goes to the  next higher one  When the reading falls below 9  of full scale  the range changes to  one range lower  This change occurs after a short delay  This provides overlap   hysteresis  to keep the Laserstar from flipping back and forth when reading close to  the end of the scale     Manual Range   There are certain disadvantages to autorange since it changes scale  even if you don t want it to do so  If you want to measure the same range all the time   it is better to use manual range  The correct range to select is the lowest one that is  larger than the largest expected measurement  If you wish the Laserstar to be in  manual mode when switched on  save this configuration before switching off  See  Section 5 3     47    dBm   The Laserstar allows the measurement to be made in units of dBm that is a  logarithmic scale  dBm units are defined as     10 x log  reading in mW   At imW the reading will be 0 dBm  at 100mW it will be 20 dBm etc     To choose auto  manual or dBm range  follow these steps     1  From the bargraph measurement screen  press    range       2  Now select the correct manual  auto or dBm range with the arrow keys  If you  want the selected range to be the startup value press    save     Now press exit to  return to the measurement screen  
83. r energy reading on the screen equals the known  power energy  The power energy calibration factor will change accordingly  See  note 1    8  Press exit and    permanent save     The new value will be saved     Note 1   For units which have different calibration factors  e g  C02  YAG  or VISIBLE  press   laser  to select correct type of laser before calibration  When recalibrating using   power with laser   function  only the calibration of the chosen laser wavelength  will change  If you wish to change the calibration of all wavelengths  proportionately  then use the function  power with CAL        Note 2   When changing power calibration  this will also change energy calibration  proportionately  However  changing energy calibration will not change power  calibration     89    10 2 Calibration of Photodiode type Heads    Photodiode detectors are inherently very linear but also have a large variation in  sensitivity with wavelength  In addition  the Ophir model PD300 is equipped with both a  built in filter and removable filter to allow measurement of higher powers without  detector saturation  These filters also have a transmission that depends on  wavelength  Therefore  the PD300 Laserstar has a built in calibration adjustment for  wavelength which is described in the next paragraph     10 2 1 Method of Factory Calibration    The sensitivity of various Ophir photodiode sensors can vary from one to another as  well as with wavelengths  Therefore  Ophir photodiode detectors
84. rates  up to 5000Hz as well as relatively high damage threshold    2  Broadband type  The type with the BB suffix has a broadband black absorbing  coating to provide high absorptivity from the UV through the IR  This coating can  withstand energy densities  up to 0 3J cm  for short pulses and 2J m  for long  pulses without changing calibration  The absorption of this coating is above 90   for most of its range  as shown in Graph 2 below  This coating is available for the  PE50 and PE25        Absorption  100    P ioni S A ON A OA E aan  BB   Broadband    80                      70    60       50       40          Metallic Coating                30  0 1 1  Wavelength  nm             Graph 3  Absorption of Ophir Pyroelectric Absorbers  Absorption     vs  Wavelength  nm     92    10 3 1 Calibration    The sensitivity of the various Ophir pyroelectric sensors can vary from one to another  as well as with wavelengths  Therefore  Ophir pyroelectric detectors are individually  calibrated against NIST traceable standards  In addition  the calibration is corrected  in the Laserstar for different wavelengths     Ophir pyroelectric detectors are calibrated using a 1 06um repetitively pulsed laser  referenced to a NIST traceable thermal power meter  The average energy is set to the  average power of the standard power meter divided by the laser frequency  The  metallic PE25 and PE50 heads are also calibrated with an excimer laser at 248nm to  correct the rather large absorption variation
85. re 4    The Laserstar will enter the mode of operation last saved with the  configure   operation  For basic operation  you need only set up the parameters for the type of  measurement you wish  Section 4 3 describes the procedure to store your  configuration    and the Laserstar will be ready to perform that type of  measurement each time it is switched on  The normal default mode is the  bargraph autoranging power measurement screen    If the Laserstar is in  energy  mode  pressing the  power  soft key will cause it to  enter the power measurement mode    Center the laser beam carefully on the absorber surface and read the power     26    4 4 2 Additional Power Measurement Features    The following section describes additional Laserstar functions in power mode in  addition to the basic power measurement mode described above     4 4 2 1 Zoom    The Zoom function causes the scale to expand the present scale fivefold  Thus  if the  full scale of the bargraph is 20 watts  and your reading is 15 watts  pressing    zoom     will make the bargraph scale range between approximately 13 and 17 watts   Variations in power are more easily seen in this mode To access the zoom function   simply press the    zoom    button  It will be highlighted to show that it is engaged  Note  that the fiducial marks of the bargraph are further apart  indicating that zoom is  engaged  Now press the    zoom    button again to return to the unexpanded bargraph  display     4 4 2 2 Offset    The offset fu
86. riate measurement range in the usual way    3  You can return to the ordinary measurement mode by pressing    exit    from the  scope mode screen    4  If you wish the instrument to start up in scope mode  press    menu     Select     config    and press    go     Select    mode    and press    value     It will change to     scope     Press    save    and    exit    to save the new configuration     63    Chapter 7 Operation with RP heads    Note   It is highly recommended to use the Customize function every time you work with a  new laser type or setup  See section 7 10 below    7 1 General Description    The exclusive Ophir RP type heads represents a unique way to accurately measure the  energy of high energy repetitively pulsed lasers  while at the same time measuring  average power  pulse shape  standard deviation  average energy  missing pulses and  jitter  As with other Ophir smart heads  you can log the laser energy and store it on  board or send it to PC  In addition to the above  the RP head is also a regular thermal  smart head and will measure power or single shot energy like any thermal head     7 2 Principle of Operation    RP heads incorporate an innovation  patented  that allows measurement of the energy  of high energy repetitively pulsed lasers in a manner superior to previous methods   The basic approach is to incorporate two sensors in one measurement head     The RP measurement head has a standard thermopile type detector to measure  average laser power  
87. riod  of the average is reached  the bargraph is filled and from then on  the average is a  running average over the average period backward in time  For instance  if the  average period is 1 minute  at 30 seconds  the average is over 30 s   at 1 minute it is  over 1 min  at 5 minutes  it is over the period from 4 to 5 minutes  1 minute back from  the present etc       29    To use the Laserstar to display average     1  From the main power screen  press the menu button and select    config      2  Press    go    and select    Average over      Use the value button to select the  average period you want or select    NONE  if you do not want to average   3  If you want to permanently select this average period  press    save     If you only  want to average at the present until the instrument is turned off  exit without  pressing    save        30A CAL    energy zoom       Figure 6  Thermopile Power Screen    4 4 2 5 Normalize    When you wish to show the readings as compared to some standard reading  the  normalize function allows you to do this  After setting    normalize     all subsequent  readings will as a ratio to the reference reading  For instance  if you normalize the  reading at a power of 1 5W  then subsequent readings will be with ratio to that  i e  if  the power is then 2W  the reading  dimensionless  on the screen will be 2 1 5   1 33   The screen will also show in small digits the power at which the normalization was  done and the present power     To use the 
88. s in disc replacement     Remove the screws from the front and rear flanges of the absorber head    Remove the front flange and raise the rear flange  taking care not to tear the wires  attached to the connectors    Unsolder the wires from the faulty absorber disc attached to the small PC board in  the head  remembering which wire is connected to each terminal    Pry the absorber disc free from its seat by using a screwdriver in the hole provided  at the edge of the seat    Smear the absorber disc seat evenly with silicone heat sink compound    Feed the wires through the hole and install the new disc  It is best to turn the disc  back and forth a few times while pressing on it  to ensure good contact between  the disc  grease and head    Reassemble the front flange  making sure the screws are tight    Resolder the two wires to their proper locations on the rear flange and reassemble  the rear flange     10 6 1 3 Recalibrating the Head to a new Disc    1     From the main power measurement screen press the menu button  then select   advanced  and press    go       Select  calibrate  and press  go  then  power with CAL   With the value button   select the first laser wavelength sent with the disc  Now adjust up down until the  sensitivity agrees with that sent with the disc  Now press save  exit and     permanent save    to save the new value    For heads with calibration at several laser wavelengths  you now have to adjust  the calibration for the other wavelengths  From the m
89. s in that spectral region with those heads     The spectral absorption of the detector coating is measured spectroscopically and the  absorption curve is used to correct the calibration for other wavelengths  When the  user selects his wavelength on the Laserstar  the correction factor for that wavelength  is applied     10 3 2 Accuracy of Calibration    Since the instruments are calibrated against NIST standards  the accuracy is  generally 3  at the energy level and wavelength at which the calibration has been  performed  This accuracy has been verified by checking the scatter of the results  when several instruments are calibrated against the same standard  The maximum  error in measurement will be less than the sum of the specified accuracy  linearity and  inaccuracy due to errors in the wavelength curve     The non linearity is approximately 2   and the error due to wavelength is given in  Table 5 below     In addition to the above errors  the reading of pyroelectric heads changes with  frequency  The Laserstar has a built in correction for this error  For frequencies  above 50  of maximum frequency  inaccuracies in this correction can increase the  total error by up to 3      93    The maximum error in measurement will be less than the sum of the above errors and  in general will be considerably less                                         COATING TYPE  BROADBAND METALLIC  WAVELENGTH  ERROR   ERROR   190   350nm  1   4   400   800nm  1   4   1064nm 0 0  2   3um  2   8   
90. s measurement in normal room light     42    The calibration of each measurement head is in the head plug so that heads and  displays can be interchanged without losing calibration  It also allows user  recalibration if desired     The BC20 has advanced circuitry with excellent sensitivity  signal to noise ratio   accuracy  and response time  its special circuitry rejects electromagnetic  interference  The measuring wand swivels 180   providing access to beams hidden in  complicated assemblies  The BC20 has a hold function  allowing user to hold highest  reading for 5s before updating     43    Chapter 5 Operation with Photodiode Type Heads    Warning   Before using the head for power measurement  check that your laser power or  energy and energy density does not exceed the head ratings  See table 6     5 1 Photodiode Heads    When a photon source  such as laser  is directed at one of the PD300 or 3A IS series  photodiode detectors  a current is created proportional to the light intensity and  dependent on the wavelength     The PD300 and PD300 3W heads have a unique dual detector head  patented  in which  the two detectors are identical and connected back to back  When a uniform signal   such as room light background  falls on the detector head the signal from the two  detectors cancels    On the other hand  when a laser beam falls on the head  it illuminates only the first  detector and therefore is detected  Thus the PD300 subtracts most of the background  while detecting 
91. s will be with  ratio to that  i e  if the energy at that point is 2mJ  the reading  dimensionless  on the  screen will be 2 1 5   1 33  The screen will also show in small digits the energy at  which the normalization was done and the present energy     To use the Laserstar to normalize the energy   1  From the main energy screen  press the menu button and select    normalize      2  Press    go    and when ready to normalize  press    apply     From then on  the energy    will be proportional to the normalized energy value  see Figure 16   If you wish to  return to the the regular display  press    exit        61    24 50mJ Normalization A    1 021    24 00mJ  apply laser    Figure 16  Pyroelectric Normalize Screen       6 5 Measuring Repeating Pulses of High Energy    Because of their construction  pyroelectric heads are restricted in the energy density  they can withstand  particularly for short pulses on the order of nanoseconds  If the  energy density of your laser exceeds the rating of the pyroelectric absorber  there are  several options available     1  You can enlarge your laser beam using a negative lens until the energy density is  below damage threshold  You should test this using the test slide  See section  6 4    2  You can use a beam splitter  splitting off typically 8  10  of the light  If you use this  method  note that there may be polarization effects  You can calibrate the beam  splitter by running the laser at energies below damage threshold and measurin
92. se or sometimes  fails to catch large pulse     Increase threshold level  See Section  4 5 3        Instrument does not show ready  for a long while after a reading  is made     Increase threshold level  See Section 4 5 3           Non reproducible results when  measuring very small energy  pulses  or no response to pulses  at low energy        Decrease threshold level  See Section 4 5 3        10 5 3 Thermal Heads  Power       Problem    Cause Remedy       Instrument shows zero reading    Check connections between the head and       in both power and energy   the instrument  See signal tracing  Section   modes  10 5  Check that the sensor disc is  operative  Resistance between the pins 1  and 9 of the head connector should be  about 1 8K  If the sensor is defective  there  will be an open or short circuit    Instrument responds while   Replace sensor disc  See Section 10 6 1 2     head is cold  but suddenly fails  as it heats up           Instrument does not return  completely to zero on power  measurement        If head is very hot  allow it to cool   Disconnect the head from the instrument  If    readout unit does not zero  follow  instructions in section 3 5 4  If the offset  persists  try zeroing with the head    connected as well  as described in Section  3 4 2        96          10 5 4 Pyroelectric Heads       Problem    Cause Remedy          Instrument reads incorrectly or    erratically  especially on  sensitive scale   Instrument triggers even    without being expos
93. se variation of  lasers whose average power is too small to measure thermally   Another advantage of relative mode is that it allows you to directly gather  meaningful data from the first pulse collected   See also the    stabilize    function  below   Since the absolute energy measurement   based on the power reading   and since it takes several seconds for a stable power value to be determined  it is  not always possible to measure the absolute energy during the first few seconds  of operation    4  Laser Wavelength  Since the calibration changes with wavelength  you must select the correct  wavelength band for your laser  Select    Laser     and choose the laser wavelength  band you want to be the default startup    5  RP Energy Range  With this function you choose the correct RP energy range  Choose the lowest  energy range value that is higher than the energy of your laser pulses  If the  highest scale available with filter    OUT    is not high enough  then set filter to    IN      slip the filter into its slot and go back to the RP energy selection to select a higher  range    6  Power Range  With this function you choose the correct power range  Choose the lowest power  range value that is higher than the average power of your laser or choose    AUTO     for autoranging     69    7  Stabilize  This function is connected with energy logging  See section 0 for further  details    As explained above in paragraph 4  the RP is unable to measure the absolute  energy of the 
94. ssary to do this procedure again unless the  head is used with a different Laserstar  If the procedure is not done  errors of 2  or so  can occur  To zero the head against the Laserstar  proceed as follows     Make sure the head is in a quiet environment and not subject to pulsed radiation  Plug  the head into the Laserstar and turn on  Press    menu     select    advanced     press    go     and    go    again  Press    start     When    zeroing completed    appears  press    save        6 4 1 2 Energy Measurement    1  Ordinarily  when the instrument is switched on  it is in the main energy  measurement mode  This can be verified by the units of mJ  pJ etc  and the legend     energy    in the upper right corner of the display  See Figure 10   If it is in power  measurement mode with units of mW  W etc   then press    energy       2  Press the range button then use the arrow keys to select the proper range  The  correct range is the lowest one that is larger than the pulse energy measured     58    3  Press exit to return to the measurement screen and save before exit if you want  this to be the default range    4  Press the menu button and select    configuration     Press    go     Select    Max Pulse  Length    and set to the correct value for your laser   See section 6 2 2 1 for further  details     5  If you wish to display the average of a number of pulses  press the menu button  and select    configuration      Press    go    and set    Average over     to the value  
95. t for a pre determined time     Due to this innovative circuitry  Ophir pyroelectric heads can measure very long pulses  as well as short ones  They can measure low energies as well as high  They can also  measure at higher repetition rates than was possible before     The Laserstar display amplifies this signal and indicates the energy received by the  head as well as the frequency at which the laser is pulsing  Using the energy and  frequency information  the Laserstar is also able to display average power    6 2 Selecting Settings from the Laserstar Screen   The Laserstar can conveniently be set to various chosen parameters from the screen     In addition  it can be set to startup defaults so it will be in the desired configuration  when turned on     53    PE50 BB energy    2  69m    10 00Hz i max  50us 2mJ  power menu       Figure 14  Pyroelectric Energy Screen    6 2 1 Setting up the head to display the user s chosen wavelengths  Metallic  Type Only     The absorption of the detector coating varies somewhat with wavelength  The  correction curve for the absorber is stored in the head EEROM  This correction curve  ensures that the power reading is correct at all laser wavelengths     In order to simplify changing from one laser wavelength to another  the user can  program up to 6 different wavelengths to be available from the screen menu  Please  use the following procedure to set the pyroelectric head to your laser wavelengths     1  From the main display  press the men
96. t of the measurement  press    offset     and subtract the zero offset  Then follow steps 1 and 2 above  The zero offset  subtracted when    offset    was pressed will be saved in the dBm scale and you can  now use the dB 0 setting to measure true ratio without zero offset problems     5 4 7 Average and Measuring Very Low Power Measurement    If the laser power is fluctuating  the Laserstar can display the average power readings  with averaging periods varying from 1s to 1 hour  When measuring very low powers   such as picowatt measurements using the PD300 IRG or PD300 UV  there will be a  rather large zero offset coming from the detector as well as a considerable noise  fluctuation  Nevertheless  you can measure these low values by using the average  function and pressing offset to eliminate the detector zero offset     50    In order to measure very low powers do as follows    1  Press    menu    and select    configure     Choose the appropriate value you want to  average over and save it and then exit    2  Now block the power source you wish to measure  wait for a few measurement  periods and press    offset    to subtract the zero offset    3  Now unblock the power source and measure    For more about the average function See Section 4 4 2 4     Note   The average function does not work with autorange or dBm scales    5 4 8 Tune and Power log    The Tune function  an exclusive Ophir option  makes adjusting your laser to its  maximum power easier than ever before  Unlike
97. t to electromagnetic compatibility and has the CE  mark     81    Chapter 10 Calibration  Maintenance and Troubleshooting    10 1 Calibration of Thermopile or RP Heads   10 1 1 Absorber types and Method of Calibration of Ophir Power Meters   10 1 1 1 Types of Ophir Laser Absorbers   Two types of absorber surface are used in Ophir thermal measuring heads    1 Surface Absorbers    BB  broadband  absorber   On standard  high power density  broadband Ophir power monitor heads  a special  refractory coating is used to provide high absorptivity from the UV through the IR  This  coating can withstand very high power densities  up to 20 KW cm   without changing  calibration  The absorption of this coating is above 90  for most of its range  as shown  in Graph 1 below    EX  excimer  absorber   The EX absorber provides high absorption in the UV  and it can withstand both the  pulse energies and the average power of excimer lasers  These discs also have  excellent absorption for 10 6 um and other wave lengths  They can therefore be used    for other types of lasers as well  The absorption of the various Ophir absorbers as a  function of wavelength is shown in Graph 1 below     LP  long pulse  absorber  This absorber has a particularly high damage threshold for long pulse  ms  or  continuous lasers and is therefore offered for use with high power heads  It is    calibrated for use with YAG laser at 1 064um or CO  laser at 10 6um and absorbs about  90  at these wavelengths     82    2  V
98. ted  and gives the user the opportunity to change the power line frequency  and baud rate  re zero the instrument or other features  See section 3 5     3 3 Soft Keys  3 3 1 Soft Keys  An Instruction    The soft keys accessed by the menu button have functions defined by the legend  above the key  The legend usually indicates what will happen when pressing the key   For example  if  energy  appears above a key  pressing that key will change the  Laserstar into energy mode  Some functions operate when the key is pressed and are  canceled when the key is pressed again  Those keys show reverse highlighting when  operational  Pressing the same key again cancels the operation and the highlighting     3 3 2 Key Functions    The Laserstar has certain conventions as to the meaning of standard keystrokes and  these are as follows     Highlighted item  The highlighted item is the item presently active      lt       gt  These buttons move to the next active item leftwards   rightwards       gt  or select This selects available values of present item choice in acyclical  fashion as button is pressed repeatedly    go This puts into operation the active item    esc This gets you to the previous screen and does not save any  changes made    exit This gets you to the previous screen and saves changes made  until instrument is turned off    save This saves changes permanently  The change is implemented  now and is also saved for future use of the instrument    up down These keys allow continuo
99. ter will be measured only if the missing pulses function is turned on     77    Chapter 8 Operation with the Dual Channel Laserstar    If you have the dual head version of the Laserstar  you can measure with two heads at  the same time  either independently or in combination  measuring the ratio or  difference between them     8 1 Operation with One Head Plugged in Only    When only one head is plugged into the dual channel Laserstar  in either channel  it  operates the same as would a single channel Laserstar  Its operation is described in  the rest of this manual  An    A    or    B    in the upper right corner of the instrument  indicates if channel A or B is active     8 2 Single Head Operation with Two heads Plugged In    1   2     Attach the two heads to the two input connectors   See Figure 4     Press the menu button  Press the arrow keys until    mode    is highlighted  Press     go       Press    value    to select one of the choices for which heads are active  head A   the head plugged into the left side   or head B    Now press    select    to select    CHANNEL    and press value to select if you want  the head to be in power or energy mode  Now press    exit    and    esc    and the  Laserstar will operate in single head mode with the head selected  The Laserstar  will now operate the same as a single channel Laserstar and its operation is  described in the rest of this manual  If you want the default start up mode to be  the one you have selected  press    save 
100. the desired signal  The subtraction is not perfect but usually 98  of  the background signal is eliminated so the detector can usually be used in ordinary  laboratory lighting conditions     The Laserstar display unit amplifies this signal and indicates the power level received  by the head  Due to the superior circuitry of the Laserstar  the noise level is very low   and the PD300  3A IS series heads with the Laserstar display have a large dynamic  range from nanowatts to hundreds of milliwatts     Since many low power lasers have powers on the order of 5 to 30mW  and most  photodiode detectors saturate at about 2mw  most heads of the PD300 series have  been constructed with a built in filter so the basic head can measure to 30mW or more  without saturation  When the additional filter is installed  the maximum power is on  the order of 300mW  or 3W with model PD300 3W   The PD300 saturates when the  output current exceeds 1 3mA so the exact maximum power depends on the sensitivity  of the detector at the wavelength used  When saturated  the legend  SAT  will appear  onthe screen  Table 2 gives the actual maximum power as a function of wavelength     44    FILTER OUT    WAVE  PD300   PD300    PD300 UV   3A IS   WAVE    PD300   ae ove  S Lieven  e      900mm   10mw   tomw   25mw   3w      1060nm   25mw   25mw   smw   ow TCT    FILTER IN    LENGTH UV LENGTH IR    250 350nm_   N A    NA   100mW   NA  f 800nm   200mW_    iw   250mWw   isw   250mwW   NA fi  W   NA      NA     NA  
101. the main measurement screen  press the menu button and select     configure     Press    go       2  Now select    Max Pulse len     and use    value    to select the setting for your laser  pulse length  If your laser pulses are longer than the short pulse length value  then  the long pulse value should be selected and vice versa  Now press    save    if you  want this to be the startup default and press    exit        6 2 2 2 Setup for the Time Period to Average Over  The Laserstar has the ability to measure a number of pulses and display the average  value of the energy of the pulses  This function is useful if the laser energy is not    stable  Then the user can choose to measure a number of pulses and display only the  average     55    To set up for averaging over a number of pulses     1     From the main measurement screen  press the menu button and select     configuration     Press    go        2  Now select    Average over     Now press  value  until the time period you wish to  average over is displayed  If you do not want to average  but collect individual  pulses then select  none     3  Press  exit   If you wish to save the new setting as the default  press    save     before exiting    Note     The Pyroelectric head is capable of measuring pulses up to very high repetition  rates on the order of kilohertz or above  The Laserstar actually captures each  pulse at over 180Hz and will sample pulses at up to 1000 to 5000 pulses depending  on the head  The pulses co
102. ts screen do as follows     1     From the main power or energy screen press    menu     select    advanced     press     go     select    limits    and press    go     You are now in the limits screen    Press    setup    and select the lower and upper limits as follows  With the up down  buttons select the lower limit digits  then with the arrow go to the exponent  screen and with up down select the exponent  For instance  if you want to select  a lower limit of 0 1W  set DIGITS to 1 000 and EXP to    1    Now do the same for the upper limits on the right side of the screen and press     save    to save the values    If you want the limits screen to be the startup default of the Laserstar  follow the  instructions in section 5 3     Now when the power or energy is over or under the limits chosen  the error message    will be shown as in Figure 13  Note that the lower limit is shown on the upper left and  the upper limit on the upper right        Figure13  Limits Screen    4 7 Special Thermopile Heads    4 7 1 BC20 Power Meter for Scanned Beams    T    he BC20 scanned beam laser power meter has become the industry standard for    measuring scanned laser beams aswell as hard to reach static beams  The BC20 uses    a    silicon detector and innovative circuitry to measure scanned as well as static beams    of up to 20 milliwatts with a noise level of microwatts  This instrument has patented  dynamic background subtraction that eliminates over 95  of background light and  allow
103. u button located on the right and select     wavelength     Press    go       2  Press    select    to select the first wavelength desired  Press    go       3  Change to the wavelength desired using the up and down buttons  If you desire to  permanently save the new value press    save     Press    exit       4  Repeat steps 2 and 3 for other wavelengths desired  When finished press exit   Those values for which    save    was pressed will be default startup values  the  others will be saved until the instrument is turned off     Note   The broadband type heads  BB  have less variation with wavelength  and in those  heads  fixed wavelength ranges are selected similar to thermal heads  This section  is not relevant with those heads     54    6 2 2 Setup of Laser Pulse Length and Pulses Sample  6 2 2 1 Laser Pulse Length    As was mentioned before  the Ophir pyroelectric heads can measure long as well as  short pulses  In order to do this  the user must indicate to the Laserstar if the laser  pulses are going to be in the long or short pulse regime  BB models only have the long  pulse setting and measure all pulse lengths on this setting In that case    N A    will  appear      Warning   If the pulse length is incorrectly set to the short setting for pulses longer than that  value  the reading will be erroneously low  If it is set to the longer setting for short  pulses the reading will be correct but noisier     To set up for pulse length  please do the following     1  From 
104. umber of Pulses   In both of the above methods  the pulse repetition rate must exceed 3Hz  Higher rates  will generally give improved accuracy  but care should be taken not to exceed  maximum power ratings    4 5 5 Measuring Energy of Rapidly Repeating Pulses   The standard Laserstar will only measure individual pulses every 5 seconds or so   You can also calculate the average energy of rapidly repeating pulses by measuring  average power on the power setting and using the formula    Average Energy per Pulse   Average Power   Pulse Repetition Rate   For rapidly repeating pulses  you can use one of the Ophir pyroelectric heads as long  as the pulse energies do not exceed the ratings of the pyroelectric absorbers  The  pyroelectric heads are compatible with the Laserstar and just have to be plugged in to    be used     For higher pulse energies  you can use the Ophir model RP to measure both average  power and exact individual pulse energy of the rapidly repeating pulses     See your Ophir agent or sales representative for details     37    4 5 6 Energy Log and Data Storage  See Figure 11     The Laserstar has the option of displaying successive energy points in a scrolling  vertical bargraph form  While measurements are being taken  the Laserstar will  record data  Up to 59 400 points are stored See section 4 4 2 7 1 for more information  on data storage and handling     Note     For Thermopile heads logging energy  there are no setup options     4 5 6 1 To use the Laserstar f
105. us changing of values as in adjusting  calibration     off on backlight switch    12 34 mw    energy zoom offset range       Figure 3  Laserstar Top View    3 4 Power up and Shut Down  To turn the Laserstar on     Flick the toggle switch on the left side of the rear panel up  The unit will switch on  and  the display will appear  If no head is connected  a display showing  Head  Disconnected  will appear  If a head is connected  the appropriate default  measurement screen will appear     Note   There are three positions  the center is on and the upper position is on with the  backlight lit  See Figure 4      To switch the Laserstar off     1  Return the toggle switch to its lowest position    2  If you wish to save the current Laserstar configuration  use the  configure   function before switching off  For measurements of the same type  the  Laserstar does not need to be set up each time  See the appropriate section  in the chapter dealing with the head type you intend to use     3 5 Laserstar Functions which are independent of Head Type    When no head is connected to the Laserstar and it is turned on  the message  Head  Disconnected  appears  In this mode  the user can adjust the startup screen  either  the standard power energy screens or other specialized screens  He can choose the  power line frequency so the instrument can be adjusted for minimum interference at  the power line frequency  He can also adjust the baud rate for optimum  communication with PC  or the user c
106. velength   See section 5 2 for changing to your desired wavelength values    Press    save    when all selections have been made to save all current Laserstar  settings  If you only want to save the last changes made on the previous screen   then exit without pressing save     46    5 4 Selecting Settings from the Laserstar Screen    The Laserstar can conveniently be set to various chosen parameters from the screen   In addition these settings can be saved as startup defaults so the head will be in the  desired configuration when turned on     5 4 1 To Set to the Laser Wavelength Being Used    Photodiode heads have a different sensitivity at different wavelengths  Moreover  the  filters used in the head have a different transmission at different wavelengths  In  order to compensate for this  each head has a built in calibration curve  with Inm  resolution  over the measurement range    When you choose the correct laser wavelength  the correction factor for that  wavelength is automatically introduced  Note that the laser wavelength presently  corrected for is displayed in the upper left corner of the display  In order to define the  laser wavelengths to choose from  See section 5 2     In order to choose the laser wavelength  simply press the    laser    button from the main  power screen until the correct wavelength is displayed  In order to set the default  startup wavelength  see section 5 2     5 4 2 To choose Manual or Automatic Ranging or dBm    Autorange   The Lasersta
107. work in single channel  mode and the value set will be applied irrespective of which channel you use     4 6 2 Power and Energy Density Screens    The Laserstar gives you the option to measure in units of power and energy density  instead of power and energy  You input the beam diameter and the display then  calculates and displays the power or energy density in units of W cm  or J cm2  In  order to activate this screen do as follows     1  From the main power screen press    menu    then select    advanced    and press     go       2  Select    density    and press go  The display will now be in units of W cm  or J cm   depending if you were in the power or energy measurement screen    3  Now press    change    and adjust the four values given to values of spot size that  you desire  Press    save    and    exit       4  Now select the beam diameter you want of the 4 given by repeatedly pressing  next    5  Press    exit    to get back to the standard measurement screen  If you want the  instrument to start up in the density screen  follow the instructions on startup  defaults in section 5 3     4 6 3 Limits Screen  The Laserstar gives you the option to set upper and lower limits of power and energy   When these limits have been set in the limit screen  when the power or energy is    above the maximum limit set the message    OVER    will appear on the screen and when  under the minimum limet set  the message    UNDER    will appear on the screen     41    To set up the limi
108. y in about 14 hours when switched  off and longer when on  It is not recommended to leave the charger plugged in for  much longer than 14 hours when it is off  in order to preserve the battery lifetime     The approximate time between charges is given below for various configurations     One thermal head  no backlight  18 hrs   Same  with backlight 12 hrs   Multi head configuration  two non pyro heads  no backlight  14 hrs   One pyro head  no backlight  14 hrs   Two pyro heads  no backlight  9 hrs   Same  with backlight  7 hrs     21    When the battery is low   BAT  flashes on and off and if the audio warning is activated   a warning beep will sound every 5 seconds  When this happens  the charger should  be plugged in  In order to charge the batteries again  the charger should be left  plugged in with the unit switched off for 14 hours  However  even with  BAT   displayed  the unit will operate properly for about one hour before its accuracy will be  impaired     3 6 4 Analog Output    The instrument provides an analog voltage output via the 25 pin output plug on the rear  panel   See Figure 4   The analog output uses pins 1  high  and 7  low  for channel A  and 14  high  and 7  low  for channel B  The Laserstar is provided with an adapter plug  that connects to these pins and outputs to two BNC plugs  When the Laserstar only  has one channel  the first channel analog output will be active  The analog output is  useful for driving chart recorders and other analog devices  Th
109. you wish   Note that this choice is  stored in the Laserstar and not the head so will be operative for all heads     2  Now switch the Laserstar off  connect the head and switch on again  From the  main energy measurement mode  press    menu    and select    configure     Press     go       3  Press  value  to choose  power  or  energy  for the desired startup screen    4  With the arrow keys select    Laser    and with    value    select the laser type you  wish to be the default    5  Now select    Max PulseLen    and with    value    select the maximum pulse length  you wish to allow  Only choose the long value if your laser pulse is too short for  the short value     Warning   Incorrect readings will result if pulse length is not set up correctly    Note   Some heads have only the long pulse setting for all pulse lengths  In that case  N A appears     6  Now select    Mode    and with    value    select the mode of operation  Select     scope    if you wish to use the scope adapter    7  Now select    Energy Range    and with    value    select the energy range you wish  for the startup default    8  Now select    Average Over    and with    value    select the time period you wish to  average over or leave the default    NONE       9    When finished press    save    and    exit     If you wish to save these values only  until the instrument is turned off  exit without saving  For further details on  configuration see Sections 6 2 and 6 3     2 4 5 Energy or Average Powe
    
Download Pdf Manuals
 
 
    
Related Search
 LASERSTAR  laserstar  laserstar technologies  laserstar welder  laserstar berlin  laserstar academy  laserstar oldenburg  laserstar technologies corporation  laserstar la garde  laserstar regulator  laserstar workstation  laserstar laser welder  laserstar schwerzenbach  laserstar orlando fl 
    
Related Contents
ハイフローフィルター(45°ベント) 取扱説明書    Manuel d`utilisation Lecteur Blu-ray Disc DMP-BD45  Ateca AT217AA015 flat panel floorstand  Eccellenza Café - Service Manual  油圧シリンダ/共通注意事項q    Panasonic KX-FP155 Plain Paper Thermal transfer Fax  Approche silhouette métallique 2012  Nokia 6600i Cell Phone User Manual    Copyright © All rights reserved. 
   Failed to retrieve file