         Contents
1.     Optional equipment consists of a range of warming blocks  Please contact your local  distributor or K SYSTEMS Kivex Biotec for details about the full range of warming blocks     Examples of different types of warming blocks  Warming block BO1  Holds seven tubes 017 mm    Falcon 352001  Order code  26001  Warming block BO6    Holds twelve tubes of  012 mm    Falcon 352003  NUNC 343923    Order code  26006    Warming block B09  Holds two tubes of 017 mm    Falcon 352095    Order code  26018    Warming block B28  Holds one 4 well dish    Falcon 176740    Order code  26029       Instruction Manual     T47    4 User Manual    4 1 Definition of use    The K SYSTEMS Kivex Biotec T47 warming plate is designed for warming the media used  during oocyte retrieval in both human and animal assisted reproduction  It can also be  used for maintaining a stable temperature of cells and specimens in a test tube syringe or  similar     4 2 General description    The aluminium surface maintains a preset temperature in the range ambient to 42 0  C   The temperature is electronically controlled by a PID controller and can be read on the  display in increments of 0 1  C     socket    connector       Instruction Manual     T47    5 Features and operation    5 1 Assembling the T47  No assembly needed  Just follow directions in section 5 2 and the unit is ready for use     5 2 Connection to the mains    The mains voltage is automatically regulated to the correct level by the main adapter   PSU    
2.    T47    11 Liability    Because K SYSTEMS Kivex Biotec has no control or influence over the conditions under  which this device is used  over its method of use or administration  or on handling of the  product after it leaves its possession  K SYSTEMS Kivex Biotec takes no responsibility for  the results  use and or performance of the product  K SYSTEMS Kivex Biotec expects that  use of the product will be confined to trained and expert users     In no event shall K SYSTEMS Kivex Biotec be liable for any direct or indirect damages  including incidental  consequential or special damages  arising out of or in connection with  the use or performance of the product     If K SYSTEMS Kivex Biotec provides you with technical documentation  this does not  authorise you to perform repairs  adjustments or alterations on the device or accessories     No representative of K SYSTEMS Kivex Biotec and no vendor of the product is authorised  to change any of the foregoing terms and conditions  and the purchaser accepts the  product subject to all terms and conditions herein  subject always to any contrary  provisions which are necessarily implied by stature or law notwithstanding the within terms  and conditions     12 Replacement    As mentioned in the Limited Warranty  the decision whether to provide any remedy or  whether to refund any portion of the purchase price shall be at the discretion of K   SYSTEMS Kivex Biotec     Before returning a product for any reason  please contact your n
3.   Press the HEAT key one time     then let go of the setpoint key     The display now shows the 4  temperature     Note The time is shown as a 24 hour clock e g 19 30 not 7 30 PM     6 11 REST    The REST function will restore all factory set values   Any changes made to the displayed temperature  unit readings and controller value will be  reset     Note Calibration value in the TUNE option will remain and not be reset     Follow these steps to navigate in the REST option   Key Display       Press and hold UP DOWN key for   3 seconds to enter the menu    The option    UNIT    will appear in   the display    Press DOWN key until the option      REST    appears in the display    Press and hold the setpoint key to   change values  e    While holding the setpoint key in   press either the arrow UP or arrow  DOWN key     When the display shows                 let  go of the setpoint key     Press and hold UP DOWN key for  3 seconds to exit the menu      This means all factory values have now been restored     20    Instruction Manual     T47    7 User maintenance    7 1 Cleaning    The T47 is a low maintenance unit designed to be both easy to use and clean  plus it is  durable due to its construction     All surfaces should be cleaned with a 7096 alcohol solution on a clean cloth or lint free  paper towel     e Clean all surface components by wiping gently with gauze     e To remove fingerprints or oil smudges  wipe with gauze slightly moistened with a  mixture of ether  70   and
4.  Connect the T47 to the mains supply using the enclosed PSU    For safety reasons the unit has an external PSU which operates at a low voltage     5 3 Start of apparatus    Connect the unit to the PSU and the PSU to the mains wall socket   To power up the unit  press the heat on button on the keyboard  The display will indicate  the current temperature     5 4 Keyboard functions    A   Heat ON Off          LED Display TIT   MEN T     Instruction Manual     T47    5 5 Operating the heated area    Press heat ON OFF key to    show the actual  temperature        5 6 Temperature setting and control    The display indicates the average temperature of the surface     Action   Press setpoint key until   set temperature is   shown  a    Press and hold setpoint  key     While holding the  setpoint key in  press  either the arrow UP or  arrow DOWN key to  raise or lower the set  temperature     When the desired  temperature is  reached  let go of the  setpoint key        Note The setpoint can be adjusted to a reading between ambient and 42 0   C     5 7 Combination keys  Action Key Display          Press alarm and UP The keyboard is locked   Md D    Press alarm and   The keyboard is unlocked  arrow UP key    D    Press Heat ON OFF o owitches between  and setpoint key    temperature and time        Instruction Manual     T47    5 8 Temperature alarm    The temperature alarm indicates too high or too low temperature    The alarm will be activated if the temperature rises or falls more tha
5.  alcohol  30       Note Since solvents such as ether and alcohol are highly flammable  they must be  handled carefully  Be sure to keep these chemicals away from open flames or  potential sources of ignition     electrical equipment must be switched off    Also remember to always use these chemicals only in a well ventilated room     e  f smudges are difficult to clean  wipe them with a soft cloth slightly moistened with  a diluted neutral detergent     e The heated area is able to withstand some water spills  However  precautions are  still necessary if water is spilled on the surface  Unplug the power cord  and then  wipe dry with a dry cloth immediately     Note If water gets inside the unit  contact your K SYSTEMS Kivex Biotec  representative to check electrical safety     7 2 Calibration    For optimal performance the unit should be calibrated once a year by authorized K   SYSTEMS Service provider     21    Instruction Manual     T47    8 Trouble shooting    Temperature alarm   Temperature more   Check the set  is on  than  0 5   C from temperature  Wait  set temperature  for the temperature  to stabilize     Setpoint below Change setpoint  ambient   eg  37  C      when    HEAT    key is connector   not connected  connector   key  keyboard        9 Technical data    DC output 24V 5A  Audible and visible for out of range  temperature       22    Instruction Manual     T47    10 Limited warranty    K SYSTEMS Kivex Biotec warrants to the purchasers of all devices and produc
6.  key for 3 seconds again to  exit the menu        While holding the setpoint   key in  press either the arrow   UP or arrow DOWN key to e  change values within the   options     The menu consists of the 10 options listed below   For further detail see the description on the following pages     6 1 Overview of options                  UNIT    RS232          C     INT T    Ti 5t    51 51    4 al E 4 a a      LI   LJ           HEAT    REST       r   a  LI Li           a      rl   k                   Calib  value             H  1       ff    o         k       n E    11    Instruction Manual     T47    Instruction Manual     T47    6 2 UNIT    In the UNIT option the displayed temperature unit can be set to either degrees Celsius  or Fahrenheit     Follow these steps to navigate in the UNIT option   Key Display  Press and hold UP DOWN key for  3 seconds to enter the menu  Caz  The menu  UNIT  will appear in      the display     Press and hold setpoint key to  change values  e    When the required type has been  selected let go of the setpoint key     Press and hold UP DOWN key for  3 seconds to exit the menu        While holding the setpoint key in   press either the arrow UP or arrow  DOWN key to select the required  type of unit     Instruction Manual     T47    6 3 RS232    The RS232 option is for serial communication and data logging of the temperature and  alarm conditions     Follow these steps to navigate in the RS232 option   Display  Press and hold UP DOWN key for  3 secon
7.  works in conjunction with St St   If St  St is set to 8 00 and heat function is on  the heater will  automatically turn on at 8 00 using 24 hour clock     Follow these steps to navigate in the HEAT menu                   Press and hold UP DOWN key for  3 seconds to enter the menu     The first menu  UNIT  will appear  in the display    Press DOWN key once and the  menu  HEAT  appears in the  display    Press and hold setpoint key to  change values     While holding the setpoint key in   press either arrow UP or arrow  DOWN key to select between   ON  and    OFF        When the required setting is  activated  let go of the setpoint  key    Press and hold UP DOWN key for  3 seconds to exit the menu        Note A dot will turn on in the right side of display  indicating that the Heat function  IS active     Instruction Manual     T47    6 9 A ST    The A St  Automatic Start  option is used to repeat the timer  St St  function every day of  the week     Note This function works in conjunction with St St and HEAT   If St St is set to 8 00 and heat function is on  the heater will automatically turn on at  8 00 every day     Follow these steps to navigate in the A st option     Press and hold UP DOWN key for  3 seconds to enter the menu     The option  UNIT  will appear in  the display    Press DOWN key until the option     A st    appears in the display     Press and hold setpoint key to  change values     While holding the setpoint key in   press either arrow UP or arrow  DOWN ke
8. K SYSTEMS  Instruction Manual    Warming plate 147          Instruction Manual     T47    Table of contents    1 GENERAL INFORMATION  amp  SERVICE 2                    cccccccccccccccccccccccccccccccccccccccccccccccccccccccccccccceccccccccees 3  2 UNPACKING AND INSPECTION                   ccccccccsscccccsccccccccccccccccccccccccccccccccccccccccccccccccccccccccceccccccscccsccece 4  3 ACCESSORIES    v t ae E Deis I VE EV UI VI eM i d mE 4  Sl SPANDABD ACCESSORIES disce ese iene Nennen 4   3   ACP TION ABO UIPBNDEN TI essen een ee ee 5   4 USER MANUAL                  aUe ee eive devised acct vos ed E ade e CE a a Oed eve vc eet                Eee cu edo Peer ee Te 6  Ad  JDEBINDIHONOEUSE  d         nto ee ner een 6   25 SGENERAT DESCRIPTION o acess sea asec stolis ls ro      6   5 FEATURES AND  OPERA TION    eene vedi vo nda Ce eren ed ei 7  IE  ASSEMBLNGTHETA TFT odere iecur e tod aet E vc e dura es nU rare co    7   3 2     CONNECTION TO THE MAINS a a                     7   23 9  SARTO A A A a aaa        a    Tee 7   ca re  Nia E o LARSA EOI ks A ne E TIA E A E AT EES 7   3 3     OPERATING THE HEATED AREA  ss el ee neigen aes a 8   5 6 TEMPERATURE SETTING AND CONTROL            cccccccscccccsccscosccsccsccececcosccscoscceccnscesceccecescaccnscnscsecescaccnsensens 8   A Cosi TION KEY S         SEE TER 8   3 8  TEMPERATURE ALARM feet tase a enh cia ee ade le Deu MC M Dn        A E De MIELE 9   3 9                                                             9   IN D DOE aaa a 
9. ds to enter the menu     The option  UNIT  will appear in  the display    Press DOWN key until the option     RS232    appears in the display     While holding the setpoint key in     press either the arrow UP or arrow  DOWN key to select between   ON  and  OFF         OFF     The RS232 function is  turned off     ON   Data is send once every two  seconds     When the required setting is      let go of the setpoint       Note The RS232 cable must be connected before start     Instruction Manual     T47    6 4 TUNE    The TUNE option is for calibration of the displayed temperature    If there is an offset between the value on the display and any measurements made with a  high precision external temperature sensor  this can be corrected    The new temperature reading will be kept as the displayed value  with temperature control  conducted on this basis     Follow these steps to navigate in the TUNE option        Display  Press and hold UP DOWN key for  3 seconds to enter the menu   The menu  USER  will appear in  the display   Press DOWN key until the option                 appears in the display     Press and hold setpoint key to  change values     While holding the setpoint key in   press either the arrow UP or arrow  DOWN key to select the value  temperature    When the required setting is  activated  let go of the setpoint  key    Press and hold UP DOWN key for  3 seconds to exit the menu        Any change in the factory set calibration should always be based      very  cer
10. earest K SYSTEMS Kivex  Biotec distributor for assistance and instructions     Only for all customers in Denmark  Norway and Sweden  please take direct contact with  our head office at     Tel   45 4599 5600  Fax   45 4599 5619    Email  info k systems dk    24    
11. ld UP DOWN key for  3 seconds to exit the menu     Instruction Manual     T47    6 7 St St    The St St  Start Set  option is the timer function for the heat   By using this option  time can be saved as the T47 can be ready for use before procedures  start in the morning     Note This function will only work correctly if the clock is set     Follow these steps to navigate in the St St option   Action Display  Press and hold UP DOWN key for  3 seconds to enter the menu   The menu  UNIT  will appear in  the display   Press key DOWN until the option   St St  appears in the display        Press and hold setpoint key to  change values     While holding the setpoint key   press either the arrow UP to  change minutes or arrow DOWN  key to change the hour    When the time is set  let go of the  setpoint   Press and hold UP DOWN key for  3 seconds to exit the menu     Note If the function Heat is set  a dot will light up in the right side of display     St St is connected with HEAT  It must be chosen when the timer starts   To turn the heater on  please see section 6 8 HEAT     WARNING Always make sure when the timer function is being used  that the heated area  is clear of any objects that might be damaged by the heat or adversely  affected by it in way    Caution should always be exercised when a heated area is turned on without  any supervision     Instruction Manual     T47    6 8 HEAT  The HEAT option is used for the heater to turn on at a certain time     Note This function only
12. n 0 5  C from the  setpoint    There will be an acoustic signal and the red LED light in the alarm key will be activated     Press alarm key to mute the  acoustic signal     Press alarm key once again to  bring the signal back on        Note The red LED light and the acoustic signal will turn off when the temperature stabilizes  to the setpoint level     5 9 Warming up    Warming up from 20 0 to setpoint takes less than 30 minutes  under normal conditions   When a cold item is placed on the heated area it may need up to 40 minutes to stabilize   depending on the thermal mass of the object and the initial temperature     5 9 1 Safety barrier    There is a greater heat loss from the edges of the warming plate  and the tolerances can  not be kept as precisely in these areas  A high accuracy can only be maintained in the  centre area of the warming plate     Please see picture below     Top view          Our general recommendations   e Dark area only   0 3   Celsius  e Whole area    0 8   Celsius    Note If the nature of your work demands high accuracy of temperature  then only the area  marked above must be used  This area is located 8 cm from the edges     Instruction Manual     T47    6 Menu function    The user can access a number of advanced functions via the menu                     Press and hold UP DOWN  key for 3 seconds to enter the  menu    The first option  UNIT  will  appear in the display     Press UP DOWN key to  navigate within the Menu     Press and hold UP DOWN 
13. t is subject to copyright  All rights reserved  This  manual should not be photocopied  otherwise copied or distributed  completely or in part   without the approval of K SYSTEMS Kivex Biotec     Users of K SYSTEMS Kivex Biotec products should not hesitate to contact us if there are  any unclear points or ambiguities in this manual     K SYSTEMS   Kivex Biotec A S  Klinteho  vaenge 3 5  3460 Birkerod  Denmark    Tel    45 45 99 56 00  Fax    45 45 99 56 19    E mail  inforoK SYS TEMS dk  Internet  www K SYSTEMS dk    Issue 2  Changenote No  424    Service address   Please contact your local K SYSTEMS Kivex Biotec distributor for details of your nearest  authorised service agent     CAUTION If the equipment is used in a manner not specified by this manual  the safety of  the user may be at risk and the equipment may be damaged  Always use the  equipment as outlined in this instruction manual     Instruction Manual     T47    2 Unpacking and inspection    e Unpack the equipment    e Check the packing list to insure all accessories are there    e Examine the packing material carefully for separately packed items   e Check the warming plate for external surface damage     e Check that the display is not broken or damaged     3 Accessories    3 1 Standard accessories    24 volts main adapter  PSU     Order code  51122    Mains Cable    Standard Schuko     Oder code  52768  UK     Order code  53886  US     Order code  52773       Instruction Manual     T47    3 2 Optional equipment
14. t tla re ee er 9   6 NIENUCEUNC TION      eee Ep OIOV Roe aee or Pee oSv                         10       S3OVERVIEW OPFOPTIONS ESS ee incre Reais des                            ee 11   52 UNES S e a t E A Mec ELE M E pA ROI    12   OS TR                                  ens      13   or REINE Se    euren         14   Gros  UNET ete e Hcc ae he Sa      a a a tale 15        Se es acetate cater EL oC auo                    ath gerd                ud 15                       ENE OR RPE rT DE oM lee ae ODE C t nOn Lt    16   0G HEA RSS                   17   275 EMEN s NON rn AIRONET t mon ea MEN ORC NOAA APT OTE RISEN PCR RC 18   6 10               ETT THE RET 19   6 11 inse                                       om 20   7 USER MAINTENANCE    eese VoYvevesveesecev ee culvcrece  vi  ee ca vives Ove ee e Lee eee aUe                             21  JL COODPBANING edente ee nee seen MeL cts ehe nie 21      E AN ET adore aon cn mac cote ars ha ache at a ttc Ke eal AU           LEE a MEC ER      A AA 21   8 TROUBLE SHOOTING            ran 22  9 TECHNICAL DATA                                 ea een      ua was vae             o eas ee pecu be dep ees epa      22  10 TEIMITED WARRANTY             vo EURO EXER AVE ad wie ona 23  IL  LIABILITY essen ann ans 24  12  REPEACEMENT      ea des sieuseldcdbecewdanccssshcbessaSeussevedondeucadectavelsesseudedineussiscanstedaas 24    Instruction Manual     T47    1 General Information  amp  Service  COPYRIGHT    This manual contains information tha
15. tain temperature measurements  K SYSTEMS Kivex Biotec only recommend the  use of a high quality PT 100 sensor that is calibrated with the precision instrument  used for measuring according to the manufacturers specifications  Also ensure an  optimal contact between the sensor and the place measured  Use only a sensor type   that is correct for the purpose and of correct size  If in doubt contact your local  distributor or K SYSTEMS Kivex Biotec     14    Instruction Manual     T47    6 5 INT T    The INT T  Integral Time  option is for changing the base value for the PID controller   This should not be attempted by unauthorized persons  If set at a different level the  controlling principle will be affected    From the factory it is set to a closely calculated value specific for the model     CAUTION Do not change the value here  If by mistake any changes are made or doubts    occur if the value are correct please set the REST function in the menu for the  restore of the defaults     6 6 TI St  The Ti St  Time Set  option is for setting the time     Follow these steps to navigate in the Ti St option        Display  Press and hold UP DOWN key for  3 seconds to enter the menu   The option    UNIT    will appear in  the display   Press DOWN key until the option   Ti St   appears in the display     Press and hold setpoint key to  change values     While holding the setpoint key in   press either the arrow UP to  change minutes or arrow DOWN  key to change the hour    Press and ho
16. ts  manufactured by K SYSTEMS Kivex Biotec  the product was prepared and tested in  accordance with good manufacturing practices and guidelines and are in compliance to  the CE norms issued by the competent authority     This limited warranty does not apply to products subjected to abnormal use or conditions   improper storage  damaged by accident  misuse or abuse  improper line voltage  products  whose serial number has been altered  to products not shipped in accordance with the  recommendations of K SYSTEMS Kivex Biotec  and or to products altered or serviced by  anyone other than K SYSTEMS Kivex Biotec authorised distributors  Distributor is  responsible for the labour and travel costs during this period     This limited warranty is exclusive and in lieu of all other warranties whether written  oral   expressed or implied  In particular  K SYSTEMS Kivex Biotec does not warrant that the  product is suitable for the needs of the purchaser and there are no warranties given as to  merchantability or fitness for a particular purpose other than the one specified in K   SYSTEMS Kivex Biotec literature that accompanies every specific product  K SYSTEMS  Kivex Biotec assumes that the Purchaser is experienced in the use of this device and is  able to judge from his her own expertise the suitability or otherwise of the product for the  intended use     K SYSTEMS Kivex Biotec reserves the right to change or discontinue this product without  prior notice     23    Instruction Manual  
17. y to select between   ON  and    OFF        When the required setting is  activated  let go of the setpoint  key    Press and hold UP DOWN key for  3 seconds to exit the menu        Note If this function is set to  on  the heat function will be repeated every day  but if  A St function is set to  off  the heat function will only be activated one time     18    6 10 HOUR    Instruction Manual     T47    The HOUR option gives the opportunity to show the time on the display when the heat is    off     Follow these steps to navigate in the HOUR option     Press and hold UP DOWN key for  3 Seconds to enter the menu     The option  UNIT will appear in the    display     Key  Press DOWN key until the option   HOUR  appears in the display     While holding the setpoint key in   press the arrow UP or arrow  DOWN key to select between   ON  and    OFF      When the required setting is  activated  let go of the setpoint  key    Press and hold UP DOWN key for   3 seconds to exit the menu     If you want to see the time on the display when the heat is turned on   Follow these steps to switch between time and temperature     Action Key    Press and hold setpoint key           Display    Press and hold setpoint key to  change values  e    Display    While holding the setpoint key in   press the HEAT key one time   then let go of the setpoint key   The display now shows the time     Instruction Manual     T47    To change back to temperature     Press and hold setpoint key  e    Display     
    