1.   The battery voltage can be observed by scrolling the menu key until VBAT is displayed  13 5 volts is  fully charged and 11 5 volts is fully discharged  The analyzer fault led will illuminate at 11 5 volts and  LOW BAT will appear in the current alarm menu  An automatic wall charger is included to charge the  batteries when AC power is available  The supplied vehicle cable can be used to charge the analyzer from  a vehicle     Installation Procedure    Step 1   Step 2   Step 3     Step 4   Step 5     Step 6   Step 7     Step 8   Step 9     Step 10     Unpack the analyzer and check for damage   Ensure that the analyzer power supply and range are suitable for the application   Select an installation location that is close to the sample point    e Ensure that the selected installation site provides adequate room for maintenance and  repair   Wire power  analog outputs and discrete outputs to the analyzer  See section D    Tube the sample inlet  and sample vent lines to the analyzer   e 1 4    316 stainless steel tubing is recommended for the sample tubing   e 1 8    316 stainless steel tubing can also be used if the response time of the analyzer is  of particular concern   e All fittings in the sample and vent lines must be 316 stainless steel   e The vent line should be tubed in 3 8    stainless steel tubing to a maximum of 6      e 1 2    316 stainless steel tubing should be used for vent lines exceeding 6      e The tubing should be installed with a slight downward slope and
2.  and sets the analog 4 20 mA output to 2 mA    Scroll Right       Used to move the cursor to the right  Also used to  SAVE configuration adjustments     Scroll Left   lt   Used to move the cursor to the left  Also used to  CANCEL configuration adjustments   Menu Set Used to cycle through the menu options  Also used to  increase numerical values when making configuration  adjustments        Installation and Start up    Your analyzer was configured  functionally tested and calibrated at the factory  All test and calibration  data is documented in the factory calibration report   See the end of this manual     A    The analyzer should be located in an area in which it is not exposed to vibration and excessive pressure  and temperature variations  The Sulfur Sentry is designed for general area locations  Ensure that the area  can be safely de classified to non hazardous before operating     Envent Engineering is available for installation and start up  if required  See Envent   s pre commissioning  guidelines on our website  http   www envent eng com documents php      Sample Point Selection    The sample to the analyzer must be representative of the process stream and should be taken from a point  as close as possible to the analyzer to avoid lag times and sample degradation in the tubing     A 34    weldelet is required for installation  The probe must be installed vertically on a horizontal section of  pipe ensuring that the sample is drawn from between the middle and the top 
3.  instruction manual is included with the software     Calibration Gas Humidifier    The Calibration Gas Humidifier is used for calibration of Sulfur Sentries and H2S analyzers that do not  use humidifiers  This humidifier should be replaced every six months     13    This document has been continuously improved and revised over time  see the table below for revision   rev  information     Rev Rev Rev Description  No  Date  Sept  2011   Updated range specifications  02 Oct  2011   Included changes required for different humidifier  updated  format of document    03 Nov  2011   Created manual in booklet format  and synchronized with  updated 330 amp 331 user manual  04 Oct  2012   Edited principle of operation figure  removed some  reference to the humidifier from cleaning procedures  Jul 2015   Changed the manual layout        For further information  or a copy of our most recent operating manual  please visit us at www envent   eng com  Envent Engineering Ltd  reserves the right to change product design and specifications at any  time without prior notice    Head Office    7060E Farrell Road SE  Calgary  AB T2H 0T2 Canada    T  1 403 253 4012  F  1 403 253 4016  CAN Toll Free 1 877 936 8368    info  envent eng com       Envent    Engineering Ltd     14    
4.  procedure     Higher or Lower a  Liquid contamination in Sample conditioning system requires  than expecting sample tubing cleaning  Refer to  reading    Cleaning procedures   b  Sample vent either Vent should be 3 8  or larger 316 stainless  blocked or frozen steel tubing on a downward slope  Possible  heat trace required   c  Contaminants in sample Sample conditioning system requires    chamber cleaning  Refer to cleaning  procedures     Fault light a  Sensor Low fault Re zero Sensor block  Refer to Sensor Re   indicated zero procedure   b  Sensor High fault Re zero Sensor block  Refer to sensor re   zero procedure   c  Low Tape Sensing tape requires change  Possible  low tape sensor failure   d  Low pressure Pressure on outlet of regulator is lower  then set point of pressure switch    factory set to 10 psi   Sensor Fault a  Sensor didn t zero on white Re zero Sensor block  Refer to Sensor Re   tape  zero procedure     b  Sensor Wire failure Wire or Sensor requires replacement        12    Appendices    Recommended Spare parts list    Part Qty Description     330079  330138    Rear Window  amp  Gasket    e few  BEE       H2S Concentration Conversion Factors    Original Unit Multiply By Final Unit    0 0626 grain   100 cf       Computer Software    The software for the 330  amp  331 allows the user to change the factory configuration  The software is  available on the Envent website at www envent eng com  Contact our technical support department for  assistance  A separate
5.  should be as short as  possible   e The sample vent line must be tubed to atmospheric pressure and cannot be connected  to a flare line   Ensure there is enough H2S sensing tape   With the sample pressure turned off  sample inlet valve closed    e Apply power to the analyzer  The display will illuminate and the sensing tape will  advance    e Press the menu button until mV is displayed  Check that the mV reading is 1000 mV    100 mV   Turn on sample gas flow  open sample inlet valve    Open the sweep valve slightly and adjust pressure regulator to 15 psig and the flowmeter   to 2 0   Allow twenty minutes for the analyzer to stabilize  The analyzer calibration can be   verified if calibration gas is available  refer to the calibration section   If no calibration   gas 1s available  the analyzer may be operated using the factory calibration settings until   calibration gas is available     Analyzer configuration and calibration was set at the factory  The factory gain and alarm settings can be  found in the factory calibration sheet  found at the end of this manual  Analyzer settings can be read and    adjusted using the operator interface     Calibration Procedure    Ensure a suitable calibration gas and a clean stainless steel regulator with the correct CGA fitting is    available  Check that the regulator is rated for calibration cylinder pressure     e The recommended calibration gas is hydrogen sulfide in a balance of nitrogen certified and    analyzed     e Calibration g
6.  sulfur sentry II     Re Zero Sensor Procedure    Step 1   Step 2   Step 3   Step 4     Step 5   Step 6     Step 7   Step 8     Step 9   Step 10   Step 11     Press the bypass button and verify that the    Bypass    LED illuminates   Ensure the sample calibration gas flow is turned off   Press the    Menu Set    button until    Mtr Run    is displayed   Push the right arrow         to move the H2S sensing tape forward   Ensure there is white tape at the sensor block window   Remove the sensor cover   Press the small pushbutton on the sensor block located on the lower left side next to the  wire connector   The sensor block will implement a    re zero    procedure  indicated by a lit  red LED  When the    re zero    procedure is complete the LED light will turn green   Initiate another motor run  Step 3  amp  4    Press the    Menu Set    button until        mV    is displayed   Value should be between 900  amp  1100 mV Step 8  Put on sensor cover   Turn on sample calibration gas flow   Set the gas pressure to 15 psig   Remove the analyzer from bypass    Maintenance  amp  Troubleshooting    Monthly Checkup    Your analyzer will provide reliable service with very little attention  If the analyzer is kept clean there    should be no requirement to recalibrate from factory gain settings  However  regular check up will ensure  that the analyzer is operating to specifications     e Ensure that the tape take up and feed reels are tight    e Ensure that the flow meters  and sample ch
7. SULFUR SENTRY II    Hydrogen Sulfide Analyzer       Operator   s Manual    Last Updated  July 03  2015    Revision  5    Envent Engineering Ltd  7060 Bay E Farrell Road S E  Calgary  Alberta  Canada  T2H 0T2  Tel   403 253 4012   Fax  403 253 4016   Email  info  envent eng com       Web  www envent eng com  For further information  or a copy of our most recent operating manual  please visit us at www envent  eng com   Envent Engineering Ltd  reserves the right to change product design and specifications at any time without prior  notice    All products carry a one year limited warranty from the date of start up or 18 months from date of shipment   whichever occurs first  F O B  the factory  against defective parts or workmanship     Quick Start Guidelines    Analyzer must be connected to no more than I5psi sample inlet pressure  The analyzer contains a critical orifice on  the sample inlet bulkhead to achieve a flow of 2 into the sample chamber     The analyzer is not equipped with per conditioning filtration for wet dirty test sample  It is recommended that a  minimum sample point selection should be a Genie Probe Once a suitable sample point has been selected  connect  to sample inlet lineon the analyzer  Use Menu centre button to scroll to mV  Once gas is flowing into the analyzer  the mV reading will begin to drop  This will indicate that the analyzer is reading your sample     Important  Once the Mv start to drop  give the analyzer at least 15min to for the unit to stabi
8. amber tubing are free of liquid or particulate    contamination     e Check that the sample conditioning filters every tape change or when sample conditioning system    is saturated with liquid  Replace the inlet filter as required     10    Troubleshooting  For other possible solutions  please visit the    Frequently Asked Questions    section of our website  at    www envent eng com H2SFAQ php     Sulsur sentry II Troubleshooting recommendations     Problem Possible reasons Possible Solution    Erratic H2S a  Trigger slide not seated Ensure trigger slide is seated in the  Readings properly groove of the sample chamber     b  Pressure in building moving Eductor may be plugged or vent blocked   up and down from fan  exhaust Check that all vent tubing and fittings are  or wind 316 stainless steel  sized 3 8  or larger     c  Sample vent either blocked Vent should be 3 8  or larger tubing on  or frozen a downward slope  Possible heat trace  required     d  Liquid carry over in Sample conditioning system requires  sample conditioning cleaning  Refer to  system    Cleaning procedures     e  Regulator not maintaining Pre regulation to 50 psig of sample at   15 psig sample point  Possible regulator requires  repair or replacement of Hydrocarbon liquid  carried over through the sample regulator   Heated regulator may be required     f  Analog Input 2 jumper removed   Re install jumper in Analog Input across     4 20  amp   4 20      g  Sensor block fault Re zero Sensor block  Refer 
9. as concentration should be approximately 2 3 of the full scale range of the analyzer  or at the alarm point of the facility to be monitored    e Check the expiry date on the calibration gas bottle before using    e The recommended Calibration gas supplier is Air Gas    Note that the following instructions apply to analyzers with standard sample systems  however  the basic  principles still apply     Step 1   Step 2   Step 3   Step 4     Step 5   Step 6     Step 7     Step 8   Step 9   Step 10   Step 11     Press the bypass button and verify that the    Bypass    LED illuminates   Ensure there is enough H2S sensing tape  See E 2 2 if tape requires changing   Push the    Menu Set    button until    Mtr Run    is displayed   Push the right arrow         to move the H2S sensing tape forward    e Ensure there is white tape at the sensor block window    Press the    Menu Set    button until the mV reading is displayed         mV       If the mV reading is 1000   100 mV  proceed to the next step  otherwise   Ensure there is no brown stain on the sensing tape at the sensor block window  Ensure the sample chamber is clean   Re zero the sensor block  See section E 2 1  Start at Step 4     Re check that the mV reading is 1000   100 mV   Connect the calibration gas using the Calibration Gas Humidifier  see Appendix   Connect the short end into the calibration port of the analyzer  ensure arrow is  pointing towards the analyzer   Connect the long end into the calibration gas bottle   Set 
10. ccrs lt g0xc0aeseneccadenveusbascatetedcGaesteasenscseQesxbeseabendsdedabancdeaienvstedacuesenadensscedace 13    Introduction    The Sulfur Sentry H2S analyzer designed for temporary monitoring of H2S in natural gas pipelines  The  analyzer will operate from an internal battery for over 12 hours  The H2S concentration is archived  internally in non volatile memory for later retrieval  The archive data can be exported to a spreadsheet for  analysis and storage     Technical Specifications    Sulfur Sentry Technical Specifications    Measurement ASTM D4084   07   Standard Test Method for  method Analysis of Hydrogen Sulfide in Gaseous Fuels   Lead Acetate Reaction Rate Method     Ambient 0 50 Deg C    standard  consult factory for other    requirements  0 to 90  humidity  non     Temperature condensing     Power 12 vdc internal battery   less than 3 watts for   gt 12 hour operation  120 vac wall charger and 12  volt vehicle charge cables included   Output Ranges Standard ranges are between 0 100 ppm    20 sec to 90  of step change    Display 2x16 character LCD with back lighting  Menu is  scrolled by internal button       JK af      i d  POWER    NZ i    e ki K       Sample Inlet     Sample Vent 12 VDC Power Input       Sulfur Sentry component locations    Principle of Operation    The Envent Engineering Ltd  Sulfur Sentry uses ASTM D4084 07  Standard Test Method for Analysis of  Hydrogen Sulfide in Gaseous Fuels  Lead Acetate Reaction Rate Method   This method uses lead acetat
11. e  impregnated paper  referred to by Envent Engineering Ltd  as    H2S sensing tape        The H2S sensing tape reacts when in contact with hydrogen sulfide by the relationship shown below  This  reaction is visibly evident by a brown stain directly on the H2S sensing tape  Considering this reaction   the electronics built into the Models 330  amp  331 have been programmed to measure the rate of darkening  over time which  in turn  gives the hydrogen sulfide concentration     H S   Pb CH  COO           gt  PbS   2CH  COOH           DETECTOR    _   _ PAN    LED     SS    Kal            Ng  if  P  fa    SENSOR BLOCK       Sulfur Sentry principle of operation diagram    The figure above shows a flow regulated  pressure regulated and filtered process sample gas passing  through a membrane humidifier and into the sample chamber   See the following figure  A window in the  sample chamber allows the gas to come in contact with the sensing tape creating the brown stain     Trigger Slide       Sample chamber and parts    For concentration applications over 10 ppm there may be a restricting aperture behind the window     Various sizes of apertures match different measurement ranges  See the table below     10 ppm     20 ppm 1 16    330103  20 ppm     50 ppm 1327 330102  More than 100 ppm Consult Factory 330101    Operation Interface       The Sulfur Sentry is configured by using the push buttons  Refer to table below     Bypass Used to inhibit all analyzer alarms to a non alarm  state 
12. lize  YOU WILL NOT  SEE AN IMMEDIATE RESPNSE FROM THE ANALYZER AND IT MAY NOT DO A TAPE RUN FOR SEVERAL  MINUTES AS THE ELECTRONICS CALCULATE A FIRST READING     This analyzer has been calibrated from 0 20ppm if your sample is higher please consult the factory Select a  suitable vent line that is away from the analyzer     Take analyzer out of bypass using bypass control    Table of Contents    eege ee e coca secenseeecdhessuceoneqsneseheceuecsvesaneseseeduceone neeedseseuecssesanes seed N E l  PT CCHMICAL SC CM O a a E E E E A E ceteaeeacene l  Pane If SONY O e A E A dnd E les E EE 2  Opeon Tor 4  AAS  eV AUTON ana D EE 5  SIE Ponta e ME 5  Sample Volume and Flow E 5  Sample Lae Tim   Vs  TUDIN S126 ssiceciessiseaccereretannte cate pesteaneaiaceaacauwipnmsdousde a EE aa ada naa 6  Sample lag Ume WSs DEE E SE E E raea Eege ege 6  POW CED PNY ee 6  lastalla on Gage 7  Calibration PrOCCCUIC eege ee 8  Mapes e E Tt 9  331 Analyzer Tape change procedure  Similar to sulfur sentry UI     9  Re Zero Sensor ROCCO a ccsa sca ne os ereseser Sacecececesasee socentncatecuteebetentacbenentae seunstacet setae Sea EERST aE ES naea 10  Maintenance  amp  Troubleshooting senesine n E TEAKE EEE RaO 10  Momi kg De 1 E 10  Eege 11  SAP CN SS ee E E ees catoredcandoguseas 13  Recommended Spare parts WSU oiscsccnscascencehesedbstnosactssucdesvansccbetedesedbetnacaetaaleveessanccehadebesed ET ina 13  H2S Concentration Conversion Eege 13  COE ON T Ei E E E 13    Calibration Gas Humidifier osc oscices
13. the calibration gas pressure to 15 psig   Allow the analyzer reading to stabilize  10 to 15 minutes    Press the    Menu Set    button until the gain setting is displayed         Cam  Adjust the gain setting such that the analyzer reads the correct concentration  Calculate gain  calculation should be close to the factory calibration sheet  See  end of this manual     CalGasConcentration l l   x  CurrentGain     NewGain     Current Re ading    Press the right          and or left         arrows until the cursor is underneath the  number you wish to change    e Adjust the number using the    Menu Set    pushbutton  it will increase until    9      then cycle back through to 0       e The new gain can be saved by pressing the right arrow         until    Saved     appears or discarded by pressing the left    lt    arrow until    Cancel    appears    Step 12  Allow the analyzer to complete two cycles with the new gain setting  e The reading should match the calibration gas concentration   Step 13  Disconnect calibration gas bottle   Step 14  Allow the analyzer to return to a no alarm condition   Step 15  Remove the analyzer from bypass   Step 16  Calibration is complete     Tape Change Procedure       STEP 1  Place analyzer into BYPASS STEP 3  Remove  amp  discard used tape  STEP 2  Remove reels covers D       STEP 4  Install new tape STEP 5  Re attach reel covers  STEP 6  Initiate a MTR Run  STEP 7  Take analyzer out of BYPASS       331 Analyzer Tape change procedure  Similar to
14. third of the pipeline     An optional Genie GPR probe regulator may be included  The function of this probe is to ensure a clean  dry sample to the analyzer and to reduce the pressure of the sample  The lower pressure will improve the  response time of the analyzer  For installation instructions  refer to associated documents  Do not install   the Genie probe regulator on a vertical pipe     Sample Volume and Flow Rate    The sample should be supplied to the analyzer at 10 15 psig and at a flow between 100 200 cc min  set  flowmeter at 2 0   A bypass sweep is recommended to reduce sample lag time in the sample line if it is at  high pressure or it is longer than 15 feet  The standard sample tubing material is 1 4    316 stainless steel   however  1 8    stainless steel tubing can be used if the response time is critical  Carbon steel sample line  and or fittings are not acceptable     Sample Lag Time Vs  Tubing Size     Sample lag time vs  tubing size    Flow Flow Pressure Lag Lag  Std    SCFH   cc min   PSIA     Time per Time per  100    Gmin  100     sec        The function of the sample system is to regulate and filter particulates or free liquids from the sample   Consideration must be taken of upset conditions as well as normal conditions when designing the sample  system  The figure below shows the typical sample system used for the Sulfur Sentry     Power Supply    The analyzer operates from 12 volt internal Gel Cell batteries for more than 12 hours when fully charged 
15. to Sensor  Re zero procedure  check for green  status led on sensor block     h  Contaminants in Clean sample chamber  Replace  sample chamber aperture and window if required     Clear grease from window  Envent no  longer uses grease on the chamber  window  Applies to earlier H2S analyzers     i  Contaminants in Sample conditioning system requires  sample conditioning cleaning  Refer to  system    Cleaning procedures   Tape does not a  No tension on take up reel Check setscrew in take up reel    advance collars  Check to see if manual  advance is possible on tape     Continued on next page    11       Problem Possible reasons Possible Solution    Tape breaking a  High liquid content in sample Genie probe and additional filtration may be  gas required    b  Feed wheel not spinning freely   Dust and refuse build up between feed  wheel and chassis  Requires removal and  cleaning of chassis    c  Tape cover wheels Tape cover wheel became warped    pressing against tape  Needs to be flattened to not contact tape  when on feed wheel bolt    d  Trigger slide not seated Ensure trigger slide is seated in groove of   properly sample chamber    Slow Response a  Aperture in chamber not Removal or change of aperture type  optimized for range required  Contact  Envent tech support for ideal setting  b  Liquid contamination in Sample conditioning system requires  sample tubing cleaning  Refer to  Cleaning procedures   c  Sensor block in fault Re zero Sensor block  Refer to Sensor Re   zero
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Manual - Netkiosk - net-kiosk  CONGRESSO NACIONAL  AP SWING FEEDER/ KOPMOBOfi ABTOMAT User guide  Manual de instalación e instrucciones  Comfort Builder Version 6.0 Frequently Asked Questions  Istruzioni per l`uso  Sistema Compacto de Som  Catálogo de la Suzuki Intruder M 800  Comercial Lozano      Copyright © All rights reserved. 