Home
Manual - Test Products International
Contents
1. Displays NOx in ppm NOTE When selecting oil as fuel be sure to use the optional oil filter A773 or readings could become erratic COMBUSTION ANALYSIS Continued 5 3 Combustion Displays continued Combustion Display 4 Displays NO in ppm Displays NOx in ppm Combustion Display 5 Displays CO air free CF Combustion Display 6 Displays Temperature reading of Channel 1 T1 Displays Temperature reading of Channel 2 T2 Displays the Differential Temperature Diff between ch1 and ch2 oPEn will be displayed if no K type probe is connected to the thermocouple socket Press the Up Arrow Key to switch between C and F 5 During a combustion test you can save print and send data to a computer See section 7 through 11 Allow readings to stabilize before saving or printing the data 6 After the test is complete remove the flue probe from the test location and allow the analyzer to purge in a fresh air environment Turn the analyzer off as outlined in section 4 2 COMBUSTION ANALYSIS Continued 4 As seen below use the Scroll Enter Key to move through the various com bustion displays 5 3 Combustion Displays You can move through the following Combustion Analysis Screens by repeat edly pressing Scroll Enter Combustion Display 1 Default Start Up Display Displays Carbon Monoxide CO reading in parts per million ppm Displays calculated C
2. 3 714 Display CAL FLo oPEn Action Press the Func Key to bypass 4 714 Display CAL FLo CLos Action Press the Func Key to bypass 5 714 Display CAL LEAK rEF Action Press the Func Key to bypass 6 714 Display 0ppm Co rEF Action Press the Func Key to bypass 7 714 Display 20 9 o2 rEF Action Press the Func Key to bypass 8 714 Display 0 0mbar 0 0 oFS Action Press the Func Key to bypass 9 714 Display 2000 A r SEt Action If you are not changing the CO alarm point press the Func Key Otherwise press the Up and Down Keys to set the ppm level the alarm will sound at Factory default is 2000ppm Press the Scroll Enter Key to accept the selection 10 714 Display CAL StoP go Action Press the Up Arrow Key to select StoP by making it blink and press the Scroll Enter Key to accept the selection The 714 will return to normal operation Appendix G MANUALLY INITIALIZING SENSORS Manually Initializing the Carbon Monoxide and Oxygen Sensors If the CO or O2 sensor fails to initialize this procedure can be performed to try and manually initialize the sensor s WARNING Changing parameters in field calibration mode can adversely affect the operation of your analyzer Follow the instructions carefully 1 Press and hold the Scroll Enter Key down
3. and bars will change position on the LCD until printing is complete After printing is complete End will be displayed and YES or no will blink Selecting YES will return the 714 to normal operation selecting no will return the 714 to the print selection screen so you can print again Use the Up and Down Arrow Keys to make the selection and press the Scroll Enter Key STOR Print Mode If Stor mode is selected Addr will be displayed on the top line along with SA and a location number from 0 to 9 will be flashing on the screen Select the required address location that you wish to print data from by pressing the Up and Down Arrow Keys and press the Scroll Enter Key The analyzer will display Wait out and bars will change position on the LCD until printing is complete After printing is complete End will be displayed and YES or no will blink Selecting YES will return the 714 to normal operation selecting no will return the 714 to the print selection screen so you can print again Use the Up and Down Arrow Keys to make the selection and press the Scroll Enter Key Combustion and pressure draft printouts must be performed separately Perform a combustion test and save the data Perform a pressure draft test and save the data in a different address Print each address sepa rately 8 RECALLING DATA 1 Press the Recall Key once Addr will be displaye
4. Draft 0 15 inH2O Oxygen 4 to 7 Carbon Monoxide Less than 100ppm air free Stack Temperature 325 F to 625 F Draft 0 15 inH2O Power Burners Gas Fired Power Burners Oil Fired Oxygen 7 to 9 Carbon Monoxide Less than 100ppm air free Stack Temperature 325 F to 500 F Draft 0 15 inH2O to 0 4 inH2O Gas Fired Burners Atmospheric Fan Assist Figure 3 6 2 Function 2 Manometer Press the Func Key repeatedly to access the thermometer function In this mode the 714 functions like a differential manometer The pump will stop running when in this function 1 Ensure you have Pressure Sampling Tube connected to one or both of the Pressure Ports and there are no restriction in the tubing see figure below 2 Zero the display by pressing the Down Arrow Key 3 Connect the tube s to the device under test and read the pressure on the display Other Features Pressing the Up Arrow Key will scroll through mbar kPa and inH2O Pressing the Down Arrow Key will Zero the Pressure reading Pressing the Scroll Enter Key will toggle the ch2 temperature reading ON and OFF The 714 incorporates a differential manometer This means when a single hose is connected to the port and pressure is applied the 714 will read positive pressure If a single hose is connected to the port and pressure is applied the 714 will read a negative pressure If both ports are connected to a pressure the 714
5. Standard A764 PC Software Optional A769 IRDA RS232C Adapter Optional A767 Appendix C GUARANTEE Your TPI 714 Flue Gas Analyzer is guaranteed free from defects in materials and workmanship for 3 Years from the date of purchase This guarantee does not affect your statuary rights For additional information please refer to the included warranty card or contact TPI at 800 368 5719 To obtain warranty performance or maintenance on your analyzer Include with the product your name address phone number written description of the prob lem and proof of purchase date Carefully package and return to TPI Attn Repair 9615 SW Allen Blvd Suite 104 Beaverton OR 97005 Appendix D INSTALLING THE OPTIONAL A773 OIL FILTER When performing combustion tests on oil fired equipment it is important to use the optional A773 oil filter to ensure stable readings The A773 does not have to be removed when working with other types of fuels PROCEDURE 1 Pull the yellow thermocouple out of the cord channel in the tubing up to the cord stop 2 Cut the tubing approximately 1 1 2 inches from the end of the A762 water trap fil ter assembly 3 Insert the A773 oil filter into the tube below the A762 4 Cut the remaining piece of tube into a 1 1 2 inch piece and connect it onto the end of the A773 oil filter 5 Press the thermocouple cord into the cord channel The A773 should be replaced when most of the pellets become discolored Appe
6. until a beep is heard and COMM is no longer displayed Requests for stored and real data can be made from the PC through Hyper Terminal to the 714 Requesting Sored Data 1 Follow steps 1 through 3 on page 21 2 From your computer type QTS All Caps and press ENTER 3 The 714 will send stored data and return to normal operation Requesting Real Data 1 Follow steps 1 through 3 on page 21 2 From your computer type QTR All Caps and press ENTER 3 The 714 will send Real time data and return to normal operation If using the optional PC software part number A772 please refer to the instructions that comes with the software Appendix A SPECIFICATIONS Instrument Operating Temperature Range 14 F to 122 F 10 C to 50 C Battery Rechargeable Ni MH Battery Life gt 6 Hours Charger Input Voltage 115V or 230V 50 60 Hz AC Fuels Natural Gas LPG Light Oil Heavy Oil Pressure Ranges mbar kPa amp inH2O Display Backlit LCD Data Storage 10 sets of readings Data Logging 400 sets of readings on Timed Interval Logging Time amp Date 24 Hour Real Time Clock Dimensions 200mm x 90mm x 60mm Weight 500g Casing Rubber Boot as Standard Switch Off Failsafe Exhaust Safety Spigot Conforms to BS7927 and the draft BS7967 Appendix D INSTALLING THE OPTIONAL A773 OIL FILTER When performing combustion tests on oil fired equipment it is important to use the optional A773 oil filter to ensure stable readings T
7. 100 readings total 1 reading recorded every 1 minute 03 M 100 readings total 1 reading recorded every 3 minutes by pressing either the Up or Down Arrow Keys 2 Press the Scroll Enter Key once PAGE PA will be displayed and a page number from 0 to 3 will be flash ing Select the required page location that you wish to log the saved data to by pressing the Up and Down Arrow Keys 3 Press the Scroll Enter Key once The instrument will return to the previous screen function with Logger flashing on the top line The instrument will continue logging until all 100 readings have been saved or you press and hold down the Store Logger Key WARNING The instrument will not Turn Off if data logging is commenc ing A beep will be heard to warn you of this fact if you try to turn the instrument off Logged data can only be retrieved by sending the information to a com puter This requires the use of the optional A767 IRDA adapter and either Microsoft Hyper Terminal or the optional PC software part num ber A772 11 COMMUNICATING TO A PC This function requires the optional A767 IrDA Adapter Please refer to the A767 manual for complete instructions on using the adapter Optional software part number A772 is available for use with MICROSOFT WINDOWS or communication can be per formed using MICROSOFT HYPER TERMINAL USING MICROSOFT HYPER TERMINAL 1 Open MICROSOFT HYPER TERMINAL and set the COM port as fol lows Baud 960
8. 5 2 Connection for any K type thermocouple probe see 6 1 T2 Socket Connection for any K type thermocouple probe see 5 2 amp 6 1 Gas Sample Port Connection for Gas Sampling Probe see 5 2 P Port Connections for Pressure Tubing see 6 2 P Port Connections for Pressure Tubing see 6 2 3 4 Top View 3 3 Side Views Exhaust Port Port for connection of Exhaust Adapter Infrared Window Window for sending stored data to IR Printer or PC see 9 amp 11 Rubber Boot Protects the instrument from accidental damage Exhaust Port Infrared Window Rubber Boot Charger Socket T1 T2 P P Gas Sampling Port Charger Socket Connection for 220V 115V charger see 4 3 T1 Socket Connection for thermocouple plug on flue probe see 5 2 Connection for any K type thermocouple probe see 6 1 T2 Socket Connection for any K type thermocouple probe see 5 2 amp 6 1 Gas Sample Port Connection for Gas Sampling Probe see 5 2 P Port Connections for Pressure Tubing see 6 2 P Port Connections for Pressure Tubing see 6 2 3 4 Top View 4 Turning On amp Off and Charging 4 1 Turning On Always Before turning on please ensure that the Mini Pump Protection Filter assembly and the Temperature Sampling Probe complete with In Line Filter or the Tubing amp In Line Filter for Leak Detection are not connected to the Gas Sample Port see 2 2 or 2 4 Press and hold down the Power
9. 50 C Battery Rechargeable Ni MH Battery Life gt 6 Hours Charger Input Voltage 115V or 230V 50 60 Hz AC Fuels Natural Gas LPG Light Oil Heavy Oil Pressure Ranges mbar kPa amp inH2O Display Backlit LCD Data Storage 10 sets of readings Data Logging 400 sets of readings on Timed Interval Logging Time amp Date 24 Hour Real Time Clock Dimensions 200mm x 90mm x 60mm Weight 500g Casing Rubber Boot as Standard Switch Off Failsafe Exhaust Safety Spigot Conforms to BS7927 and the draft BS7967 Appendix A SPECIFICATIONS Continued Flue Temperature Probe Construction Pistol Grip with Stainless Steel Shaft Hose Length 2500mm Insertion Length 200mm K Type Thermocouple Accuracy 0 3 1 C Maximum Temperature 800 C Gases Range Resolution Accuracy Oxygen O2 0 25 0 1 0 3 Carbon Monoxide CO 0 10 000 ppm 1 ppm 5 ppm or 5 Whichever is greater Nitric Oxide NO 0 5 000ppm 1 ppm 5 ppm or 5 Whichever is greater Nitrogen Oxides NOx 0 5 250ppm 1 ppm Calculated Carbon Dioxide 0 25 0 1 Calculated CO CO2 Ratio 0 0 999 0 001 Calculated Combustion Efficiency 0 100 0 1 Calculated Pressure Measurement Selectable Ranges mbar kPa and inH2O Range 150 mbar to 150 mbar 15 kPa to 15 kPa 60 inH2O to 60 inH2O Resolution 0 01 mbar Accuracy 0 5 fsd COMMUNICATING TO A PC Continued 5 To return to normal operation press and hold the Print Comm Key
10. A776 x2 Barbed to 1 8 NPT fittings 2 each Gas Valve Adapter A611 Adapter Tubing Instruction Manual Denotes part number Your TPI 714 Flue Gas Analyzer has the following options available PC Software for analysis of Data Logging results see 11 amp Appendix B Infrared PC link for non compatible PCs see 11 amp Appendix B Various Temperature Probes see Appendix B 3 2 Back View Sample and Pressure Inlet Ports Connection for Gas Sampling Probe see 3 4 amp 5 2 Connection for Mini Pump Protection Filter see 3 4 amp 6 2 Connections for Pressure Tubing see 3 4 amp 6 2 Calibration and Information Label Displays calibration information and Displays serial number Battery Compartment Holds rechargeable battery Rubber Boot Protects the instrument Calibration and Information Label Battery Compartment Rubber Boot Connection for Mini Pump Protection Filter Connections for Pressure Tubing Sample and Pressure Inlet Ports Scrolls through selectable fuels see 3 1 amp 4 1 Switches between Gross and Net Efficiency see 5 3 Switches between C and F see 5 3 amp 6 1 Moves up through the Stored Data Addresses see 7 8 9 Increases data logging time intervals see 10 Zeroes pressure reading see 6 2 Moves down through the Stored Data Addresses see 7 8 9 Decreases data logging time intervals see 10 Scrolls through Combustion
11. After the test is complete remove the flue probe from the test location and allow the analyzer to purge in a fresh air environment Turn the analyzer off as outlined in section 4 2 COMBUSTION ANALYSIS Continued 4 As seen below use the Scroll Enter Key to move through the various com bustion displays 5 3 Combustion Displays You can move through the following Combustion Analysis Screens by repeat edly pressing Scroll Enter Combustion Display 1 Default Start Up Display Displays Carbon Monoxide CO reading in parts per million ppm Displays calculated Carbon Dioxide CO2 figure in percentage Displays calculated CO CO2 Ratio figure WARNING Should the CO reading rise above 2 000ppm a continuous series of Alarm Beeps will be heard The Probe should immediately be disconnected from the instrument and the instrument returned to a clean air environment This Alarm alerts the user that there is a high concentration of CO and this procedure will protect the sensors within the instrument Combustion Display 2 Displays Oxygen O2 reading in percentage Displays calculated Excess Air X Air figure in percentage Displays calculated Efficiency Eff figure in percentage Pressing the Up Arrow Key will toggle between Gross amp Net Efficiency Note Gross Efficiency is used in the USA Combustion Display 3 Displays CO in parts per million ppm Displays O2 in percentage
12. Day the Month and the Year 6 The unit will return to normal after the desired Year has been confirmed 6 5 Function 4 Combustion Analysis Press the Func Key repeatedly to return to Combustion Analysis mode and the default display Combustion display 1 Combustion Display 1 Default Start Up Display Displays Carbon Monoxide CO reading in parts per mil lion ppm Displays calculated Carbon Dioxide CO2 figure in per centage Displays calculated CO CO2 Ratio figure Function Display 3 Date Time Time Date Month Year 9 PRINTING DATA WARNING To operate correctly there must be a clear line of sight between the Infrared Window on the instrument see 3 3 and the Infrared Window on the IR Printer see Printer instructions Place the printer so it is no closer than approximately 6 inches from the analyzer 1 Press the Print COM Key once rEAL and Stor will be displayed and rEAL will be flashing Selecting rEAL will cause the 714 to print the test data currently on the display real time data Selecting Stor will cause the 714 to print data stored in memory Use the Up and Down Arrow Keys to select the desired print mode and press the Scroll Enter Key REAL Print Mode If rEAL print mode is selected the analyzer will dis play Wait out and bars will change position on the LCD until printing is complete After printing is complete End will be displayed
13. of something wrong in the charging circuit Please contact TPI technical assistance at 800 368 5719 Alternatively the instrument can be used when plugged into the charger 4 4 Activating the Backlight The display backlight can be activated at any time by pressing and holding down the Func Key for approximately 2 seconds The backlight will automatically shut off after approximately 20 seconds to preserve battery life NOTE When selecting oil as fuel be sure to use the optional oil filter A773 or readings could become erratic See Appendix E for installation instructions 4 Turning On amp Off and Charging 4 1 Turning On Always Before turning on please ensure that the Mini Pump Protection Filter assembly and the Temperature Sampling Probe complete with In Line Filter or the Tubing amp In Line Filter for Leak Detection are not connected to the Gas Sample Port see 2 2 or 2 4 Press and hold down the Power Key and the TPI 714 will start its 30 second countdown PURGE will be displayed The instrument MUST be turned on in a clean air environment as the 30 second purge will set the Carbon Monoxide level to Zero and the Oxygen to 20 9 The 714 may re start the coutdown several times before normal operation This means the sensors are still initializing and is normal Ensure that the filters are clean and dry as dirty or wet filters will result in a loss of flow rate and Lo Flo will be displayed to inform you that filters s
14. purge completely from last sample Worn or defective carbon monoxide sensor Disconnect probe and restart Purge for 20 minutes and restart Return to TPI for sensor replacement or install new sensor Try manually initializing the sensor See Appendix G InIt NO Err Nitric oxide sensor failed to initialize Flue probe connected to 714 prior to power up 714 did not purge completely from last sample Worn or defective carbon monoxide sensor Disconnect probe and restart Purge for 20 minutes and restart Return to TPI for sensor replacement or install new sensor Try manually initializing the sensor See Appendix G Lo bat Low battery Battery needs to be charged Charge or replace battery oFL Overflow indication The tem perature or pressure being measured is outside the maximum measurement capability Temperature or pressure being measured is too high or low Remove temperature or pressure source Problem Possible Cause Corrective Action Efficiency reading incorrect NET efficiency selected Ambient temperature probe not plugged in to T2 Incorrect fuel selected Select GROSS efficiency See section 5 3 Plug ambient probe into T2 See section 5 2 Select the proper fuel for the equipment being tested See section 4 1 Readings are erratic when working on oil fired equip ment Oil filter not installed or installed incorrectly Make sure the opt
15. to the computer and turn it on Place the A767 6 to 8 inches from the 714 with a clear line of sight between the sensor windows 4 Press and hold the Recall key until a beep is heard The 714 will dis play Real Stor and Logg Using the Up and Down Arrow Keys select the desired transfer mode and press the Scroll Enter Key Real This mode will transfer current readings real time to the com puter After communication the 714 will return to normal operation Stor This mode will transfer the 10 save locations to the computer While the 714 is communicating Stor Out will be displayed After communication is complete the 714 will return to normal operation Logg This mode will transfer logged data locations 0 through 99 of the selected page 0 through 3 When the 714 is communicating Logg Out will be displayed After communication is complete the 714 will return to normal operation 10 LOGGING DATA 1 Press and hold down the Store Logger Key for approximately 2 seconds S t will be displayed with 05 S flashing indicating that the instrument will log a set of 100 readings every 5 seconds You can choose from the following logging times 05 S 100 readings total 1 reading recorded every 5 seconds 10 S 100 readings total 1 reading recorded every 10 seconds 20 S 100 readings total 1 reading recorded every 20 seconds 30 S 100 readings total 1 reading recorded every 30 seconds 01 M
16. will read the difference P1 P2 between the two 6 FUNCTION SELECTION Pressing the Func Key enables access to the different functions available on the 714 Available functions are Combustion Analyzer Thermometer Manometer and Date Time display 6 1 Function 1 Thermometer Press the Func Key repeatedly to access the thermometer function In this mode the 714 functions like a K Type thermocouple thermometer The pump will stop running when in this function 1 Ensure you have a K type probe connected to one or both of the thermo couple sockets T1 or T2 refer to figure below WARNING There is ONLY one correct way to connect the K type thermo couple plug into the socket see 4 2 Forcing the plug into the socket the wrong way may result in damage to the instrument 2 Touch the temperature probe to the item under test and read the displayed temperature Other Features Pressing the Up Arrow Key will toggle between C and F Pressing the Scroll Enter Key will toggle the Differential Temperature ON and OFF oPEn will be displayed if no K type probe is connected to the thermocou ple socket The thermometer func tion allows you to per form temperature tests The pressure function allows you to perform draft and pressure tests Function Display 1 Temperature T1 Temperature T2 Temperature Differential Temperature Function Display 2 Pressure Pressure Reading T2 Temperature 6 2 F
17. 0 Data Bits 8 Parity None Stop Bit 1 Flow Control None 2 Press and hold the Print Com Key until a beep is heard and COMM is flashing in the upper right hand of the display 3 Connect the A767 to the computer and turn it on Place the A767 6 to 8 inches from the 714 with a clear line of sight between the sensor windows 4 Press and hold the Recall key until a beep is heard The 714 will dis play Real Stor and Logg Using the Up and Down Arrow Keys select the desired transfer mode and press the Scroll Enter Key Real This mode will transfer current readings real time to the com puter After communication the 714 will return to normal operation Stor This mode will transfer the 10 save locations to the computer While the 714 is communicating Stor Out will be displayed After communication is complete the 714 will return to normal operation Logg This mode will transfer logged data locations 0 through 99 of the selected page 0 through 3 When the 714 is communicating Logg Out will be displayed After communication is complete the 714 will return to normal operation 10 LOGGING DATA 1 Press and hold down the Store Logger Key for approximately 2 seconds S t will be displayed with 05 S flashing indicating that the instrument will log a set of 100 readings every 5 seconds You can choose from the following logging times 05 S 100 readings total 1 reading recorded every 5 seco
18. 5 3 amp 6 1 Moves up through the Stored Data Addresses see 7 8 9 Increases data logging time intervals see 10 Zeroes pressure reading see 6 2 Moves down through the Stored Data Addresses see 7 8 9 Decreases data logging time intervals see 10 Scrolls through Combustion Analysis Displays See 5 3 Scrolls through Function Screens see 6 Turns temperature differential calculation ON OFF see 6 1 Turns ch2 temperature ON OFF see 6 2 Allows you to change the Date and Time see 6 3 Allows you to choose a Stored Data Address see 7 8 9 10 Sends stored data to a separate infrared printer see 9 or IRDA adapter p n A767 See 11 Allows you to view stored data on the display see 8 Stores readings to memory see 7 Starts and Stops data logging see 10 Scrolls through and selects analyzer functions see 6 Turns Backlight ON and OFF see 4 4 Turns the instrument ON and OFF see 4 1 amp 4 2 Print COM Recall Store Logger Func Scroll Enter 3 3 Side Views Exhaust Port Port for connection of Exhaust Adapter Infrared Window Window for sending stored data to IR Printer or PC see 9 amp 11 Rubber Boot Protects the instrument from accidental damage Exhaust Port Infrared Window Rubber Boot Charger Socket T1 T2 P P Gas Sampling Port Charger Socket Connection for 220V 115V charger see 4 3 T1 Socket Connection for thermocouple plug on flue probe see
19. 714 Flue Gas Analyzer The Value Leader TM www tpi thevalueleader com 1 Introduction Thank you for purchasing TPI brand products The TPI 714 Flue Gas Analyzer is a state of the art easy to use analyzer designed not only to display and calculate the required readings from a flue but also to cover most of the other measurements associated with combustion The instrument is ruggedly constructed and comes with a 3 Year unit and 2 Year sensor Guarantee 2 General Overview The TPI 714 combustion analyzer uses state of the art electrochemical sensors This sensor technology provides the longest lasting most accurate and reliable means for performing combustion tests The sen sors in your analyzer will need to be replaced periodically and calibra tion is recommended once every year Electrochemical sensors by nature are always active Therefore the time the analyzer is off and not being used must be taken into account when determining sensor life The sensors in your analyzer are war ranted for two years This warranty does not cover sensors damaged through misuse of the analyzer You should keep the battery of your 714 charged so power is constant ly being supplied to your sensors The following guidelines will help prevent damage to your sensors Always use the mini pump filter when testing flue gases Always periodically check and replace the mini pump filter as needed Always make sure the in line filter water trap is in
20. Analysis Displays See 5 3 Scrolls through Function Screens see 6 Turns temperature differential calculation ON OFF see 6 1 Turns ch2 temperature ON OFF see 6 2 Allows you to change the Date and Time see 6 3 Allows you to choose a Stored Data Address see 7 8 9 10 Sends stored data to a separate infrared printer see 9 or IRDA adapter p n A767 See 11 Allows you to view stored data on the display see 8 Stores readings to memory see 7 Starts and Stops data logging see 10 Scrolls through and selects analyzer functions see 6 Turns Backlight ON and OFF see 4 4 Turns the instrument ON and OFF see 4 1 amp 4 2 Print COM Recall Store Logger Func Scroll Enter 3 2 Back View Sample and Pressure Inlet Ports Connection for Gas Sampling Probe see 3 4 amp 5 2 Connection for Mini Pump Protection Filter see 3 4 amp 6 2 Connections for Pressure Tubing see 3 4 amp 6 2 Calibration and Information Label Displays calibration information and Displays serial number Battery Compartment Holds rechargeable battery Rubber Boot Protects the instrument Calibration and Information Label Battery Compartment Rubber Boot Connection for Mini Pump Protection Filter Connections for Pressure Tubing Sample and Pressure Inlet Ports Scrolls through selectable fuels see 3 1 amp 4 1 Switches between Gross and Net Efficiency see 5 3 Switches between C and F see
21. Key and the TPI 714 will start its 30 second countdown PURGE will be displayed The instrument MUST be turned on in a clean air environment as the 30 second purge will set the Carbon Monoxide level to Zero and the Oxygen to 20 9 The 714 may re start the coutdown several times before normal operation This means the sensors are still initializing and is normal Ensure that the filters are clean and dry as dirty or wet filters will result in a loss of flow rate and Lo Flo will be displayed to inform you that filters should be changed During the last 20 seconds of the 30 second purge time the user can scroll through the following Fuels Natural Gas LPG Light Oil Heavy Oil OPT1 Bituminous Coal OPT 2 Anthracite Coal OPT 3 Coke OPT 4 Butane OPT 5 Wood Dry OPT 6 Bagasse by pressing the Up Arrow Key to select the Fuel they are working with When desired fuel is displayed release the key displayed fuel is now selected After the 30 second countdown the instrument is ready to take Flue Temperature Pressure or Leak Detection readings and Combustion Display 1 will be the first display The 714 will auto power off if no keys have been pressed for 10 minutes and the CO level is below 15ppm The Scroll Enter Key allows you to scroll through the combustion displays as out lined in section 5 3 The Func Key is used to select the 714 function mode and allows you to set it to be a combustion analyzer manometer thermometer or c
22. Probe Tubing 5 COMBUSTION ANALYSIS 5 1 Overview Testing for carbon monoxide oxygen and stack temperature is very important to the overall safety and efficiency of heating equipment The following guidelines and descriptions are generic and meant to provide you with a basic understand ing of combustion testing TPI always recommends you contact the manufacturer of the device under test obtain information specific to the device and follow the procedures and safety guidelines for performing tests and affecting repairs In general for most applications flue gas samples should be taken prior to the draft diverter or any other opening that allows room air to enter the system This prevents room air from mixing with gases in the flue and diluting the test sample To ensure accurate and consistent combustion tests it is important gas and tem perature samples be taken at the same location This is easy with the TPI flue probe because the temperature sensor is an integral part of the probe Prior to taking a sample the device under test should be on and operating Putting the flue probe in the sample area prior to starting the device may cause saturation of the sensors due to the higher initial concentration of carbon monox ide that may be encountered upon start up If this happens allow your analyzer to purge in fresh air until the carbon monoxide level returns to 0 ppm and the oxy gen level returns to 20 9 This may take more than an hour
23. Ratio figure Function Display 3 Date Time Time Date Month Year 7 SAVING DATA During testing data can be stored for later retrieval by following these steps 1 Press the Store Logger Key once Addr will be displayed on the top line along with SA and a location number from 0 to 9 will be flashing on the screen Select the required address location that you wish to save the data to by pressing the Up and Down Arrow Keys 2 Press the Scroll Enter Key once The location number which you have chosen will stop flashing and after about 3 seconds the instrument will return to the screen function you were previous ly on You have just successfully stored a set of readings which can be either reviewed on screen see 6 or sent to the IR printer see 9 When data is saved all data previously in the address will be overwritten When saving pressure draft readings select an address different from the one used to save combustion readings 6 4 Function 3 Date Time Press the Func Key repeatedly to access the date time screen From this func tion Time Date and Year can be changed 1 Once the Time Date and Year screen is displayed press the Scroll Enter Key once to allow you to change the data 2 Press the Up Arrow Key to Increase the Minutes 3 Press the Down Arrow Key to Decrease the Minutes 4 Press the Scroll Enter Key to confirm the desired Minute and move onto the Hours 5 Repeat steps 2 to 4 to change the Hour the
24. a short Beep will be heard The Instrument can only be switched off if the CO is below 15ppm The instrument has an auto shut off after 10 minutes should no keys have been pressed for this period and as mentioned above that the CO is below 15ppm Should the CO be above 15ppm then the 10 minute auto shut off countdown will not begin till the CO has gone below 15ppm 4 3 Charging Plug the Charger supplied into the charger socket on the instrument see 3 4 If the instrument is turned on then a charging symbol will be displayed Should the instrument then be turned off turn off automatically or be turned off when the charger is plugged in then the charging symbol will not be displayed BUT the instrument will still be charging The instrument should be charged overnight for a period of 10 to 12 hours and will give over 6 hours Operating Time If a beeping noise is heard during charging disconnect the charger This is an indication of something wrong in the charging circuit Please contact TPI technical assistance at 800 368 5719 Alternatively the instrument can be used when plugged into the charger 4 4 Activating the Backlight The display backlight can be activated at any time by pressing and holding down the Func Key for approximately 2 seconds The backlight will automatically shut off after approximately 20 seconds to preserve battery life NOTE When selecting oil as fuel be sure to use the optional oil filter A773 or readings could
25. acement or install new sensor Try manually initializing the sensor See Appendix G Lo bat Low battery Battery needs to be charged Charge or replace battery oFL Overflow indication The tem perature or pressure being measured is outside the maximum measurement capability Temperature or pressure being measured is too high or low Remove temperature or pressure source Appendix G MANUALLY INITIALIZING SENSORS Manually Initializing the Carbon Monoxide and Oxygen Sensors If the CO or O2 sensor fails to initialize this procedure can be performed to try and manually initialize the sensor s WARNING Changing parameters in field calibration mode can adversely affect the operation of your analyzer Follow the instructions carefully 1 Press and hold the Scroll Enter Key down until FILD CAL is displayed The 714 will cycle through a 30 second countdown then display CAL Air gAS 2 Use the Up Arrow Key to select Air by making it blink Press the Scroll Enter Key to accept the selection 3 714 Display CAL FLo oPEn Action Press the Func Key to bypass 4 714 Display CAL FLo CLos Action Press the Func Key to bypass 5 714 Display 0ppm Co rEF Action Press the Scroll Enter Key After the 10 second countdown is complete press the Scroll Enter Key twice 6 714 Display 20 9 o2 rEF Action Pres
26. and YES or no will blink Selecting YES will return the 714 to normal operation selecting no will return the 714 to the print selection screen so you can print again Use the Up and Down Arrow Keys to make the selection and press the Scroll Enter Key STOR Print Mode If Stor mode is selected Addr will be displayed on the top line along with SA and a location number from 0 to 9 will be flashing on the screen Select the required address location that you wish to print data from by pressing the Up and Down Arrow Keys and press the Scroll Enter Key The analyzer will display Wait out and bars will change position on the LCD until printing is complete After printing is complete End will be displayed and YES or no will blink Selecting YES will return the 714 to normal operation selecting no will return the 714 to the print selection screen so you can print again Use the Up and Down Arrow Keys to make the selection and press the Scroll Enter Key Combustion and pressure draft printouts must be performed separately Perform a combustion test and save the data Perform a pressure draft test and save the data in a different address Print each address sepa rately 8 RECALLING DATA 1 Press the Recall Key once Addr will be displayed on the top line along with SA and a location number from 0 to 9 will be flashing on the screen Select the required address l
27. arbon Dioxide CO2 figure in percentage Displays calculated CO CO2 Ratio figure WARNING Should the CO reading rise above 2 000ppm a continuous series of Alarm Beeps will be heard The Probe should immediately be disconnected from the instrument and the instrument returned to a clean air environment This Alarm alerts the user that there is a high concentration of CO and this procedure will protect the sensors within the instrument Combustion Display 2 Displays Oxygen O2 reading in percentage Displays calculated Excess Air X Air figure in percentage Displays calculated Efficiency Eff figure in percentage Pressing the Up Arrow Key will toggle between Gross amp Net Efficiency Note Gross Efficiency is used in the USA Combustion Display 3 Displays CO in parts per million ppm Displays O2 in percentage Displays NOx in ppm NOTE When selecting oil as fuel be sure to use the optional oil filter A773 or readings could become erratic 5 4 TYPICAL TEST LOCATIONS Atmospheric Gas Fired Fan Assist Boiler Furnace Typical Test Locations COMBUSTION ANALYSIS Continued COMBUSTION ANALYSIS Continued Condensing Boiler Furnace Typical Test Locations Figure 1 Figure 2 5 4 TYPICAL TEST LOCATIONS Atmospheric Gas Fired Fan Assist Boiler Furnace Typical Test Locations COMBUSTION ANALYSIS Continued COMBUSTION ANALYSIS Continued Condensing Boiler F
28. become erratic See Appendix E for installation instructions COMBUSTION ANALYSIS Continued 5 2 Procedure 1 Turn the 714 on as outlined in section 4 1 At any time you can activate the Backlight by holding down the Func Backlight Key for 2 seconds The back light will automatically shut off after 20 seconds to preserve battery life 2 Ensure you have connected the Mini Pump Protection Filter assembly and Temperature Sampling Probe complete with In Line Filter to the Gas Sample Port see 3 2 or 3 4 and the K Type Thermocouple Plug into Thermocouple T1 Socket see 3 4 The GK11M ambient air temperature probe must be connected to the T2 socket refer to figure below for correct hookup WARNING Ensure that the In Line Filter hangs in a vertical position when readings are being taken particularly if water is visible Failure to comply may result in damage to the instrument WARNING There is ONLY one correct way to connect the K type thermo couple plug into the socket see 2 4 Forcing the plug into the socket the wrong way round may result in damage to the instrument 3 Insert the flue probe into the test location of the device under test The probe tip should be in the middle of the flue pipe or exhaust stream Refer to figures 1 through 3 for typical test locations Mini Pump Protection Filter Assembly Ambient Air Connection T2 GK11M Probe Flue Probe Connection T1 In line Filter Assembly Flue
29. d on the top line along with SA and a location number from 0 to 9 will be flashing on the screen Select the required address location that you wish to save the data to by pressing the Up and Down Arrow Keys 2 Press the Scroll Enter Key once The Time amp Date of the Saved Data from the selected address location will be displayed flashing on the screen The rest of the Saved Data at this address location can be reviewed by pressing the Up and Down Arrow Keys 4 Press the Scroll Enter Key once when you are finished reviewing the data End will be displayed with YES flashing 5 Press the Scroll Enter Key once to EXIT or 6 Press the Up or Down Arrow Keys End will be displayed with no flashing 7 Press the Scroll Enter Key once to CHOOSE another address location to review and repeat steps 2 to 5 11 COMMUNICATING TO A PC This function requires the optional A767 IrDA Adapter Please refer to the A767 manual for complete instructions on using the adapter Optional software part number A772 is available for use with MICROSOFT WINDOWS or communication can be per formed using MICROSOFT HYPER TERMINAL USING MICROSOFT HYPER TERMINAL 1 Open MICROSOFT HYPER TERMINAL and set the COM port as fol lows Baud 9600 Data Bits 8 Parity None Stop Bit 1 Flow Control None 2 Press and hold the Print Com Key until a beep is heard and COMM is flashing in the upper right hand of the display 3 Connect the A767
30. depending on how saturated the sensors are The included figures show locations for performing tests on commonly encoun tered equipment Remember to consult with the manufacturer of the device under test for specific test information NOTE When selecting oil as fuel be sure to use the optional oil filter A773 or readings could become erratic See Appendix E for installation instructions COMBUSTION ANALYSIS Continued 5 2 Procedure 1 Turn the 714 on as outlined in section 4 1 At any time you can activate the Backlight by holding down the Func Backlight Key for 2 seconds The back light will automatically shut off after 20 seconds to preserve battery life 2 Ensure you have connected the Mini Pump Protection Filter assembly and Temperature Sampling Probe complete with In Line Filter to the Gas Sample Port see 3 2 or 3 4 and the K Type Thermocouple Plug into Thermocouple T1 Socket see 3 4 The GK11M ambient air temperature probe must be connected to the T2 socket refer to figure below for correct hookup WARNING Ensure that the In Line Filter hangs in a vertical position when readings are being taken particularly if water is visible Failure to comply may result in damage to the instrument WARNING There is ONLY one correct way to connect the K type thermo couple plug into the socket see 2 4 Forcing the plug into the socket the wrong way round may result in damage to the instrument 3 Insert the flue p
31. easurements associated with combustion The instrument is ruggedly constructed and comes with a 3 Year unit and 2 Year sensor Guarantee 2 General Overview The TPI 714 combustion analyzer uses state of the art electrochemical sensors This sensor technology provides the longest lasting most accurate and reliable means for performing combustion tests The sen sors in your analyzer will need to be replaced periodically and calibra tion is recommended once every year Electrochemical sensors by nature are always active Therefore the time the analyzer is off and not being used must be taken into account when determining sensor life The sensors in your analyzer are war ranted for two years This warranty does not cover sensors damaged through misuse of the analyzer You should keep the battery of your 714 charged so power is constant ly being supplied to your sensors The following guidelines will help prevent damage to your sensors Always use the mini pump filter when testing flue gases Always periodically check and replace the mini pump filter as needed Always make sure the in line filter water trap is installed properly Always periodically check and replace the in line filter as needed Always remove water or condensation from the inside of the in line fil ter water trap assembly prior to performing tests Always use the optional oil filter p n A773 when performing tests on oil burning equipment Contents 1 Introduc
32. ection 10 714 Display CAL StoP go Action Press the Up Arrow Key to select StoP by making it blink and press the Scroll Enter Key to accept the selection The 714 will return to normal operation Notes Notes Notes Notes Notes Notes Notes Notes Test Products International Inc 9615 SW Allen Blvd Ste 104 Beaverton OR 97005 Tel 503 520 9197 Fax 503 520 1225 www tpi thevalueleader com Test Products International Ltd 342 Bronte Road South Unit 9 Milton Ontario Canada L9T5B7 Tel 905 693 8558 Fax 905 693 0888
33. emperature 800 C Gases Range Resolution Accuracy Oxygen O2 0 25 0 1 0 3 Carbon Monoxide CO 0 10 000 ppm 1 ppm 5 ppm or 5 Whichever is greater Nitric Oxide NO 0 5 000ppm 1 ppm 5 ppm or 5 Whichever is greater Nitrogen Oxides NOx 0 5 250ppm 1 ppm Calculated Carbon Dioxide 0 25 0 1 Calculated CO CO2 Ratio 0 0 999 0 001 Calculated Combustion Efficiency 0 100 0 1 Calculated Pressure Measurement Selectable Ranges mbar kPa and inH2O Range 150 mbar to 150 mbar 15 kPa to 15 kPa 60 inH2O to 60 inH2O Resolution 0 01 mbar Accuracy 0 5 fsd COMMUNICATING TO A PC Continued 5 To return to normal operation press and hold the Print Comm Key until a beep is heard and COMM is no longer displayed Requests for stored and real data can be made from the PC through Hyper Terminal to the 714 Requesting Sored Data 1 Follow steps 1 through 3 on page 21 2 From your computer type QTS All Caps and press ENTER 3 The 714 will send stored data and return to normal operation Requesting Real Data 1 Follow steps 1 through 3 on page 21 2 From your computer type QTR All Caps and press ENTER 3 The 714 will send Real time data and return to normal operation If using the optional PC software part number A772 please refer to the instructions that comes with the software Appendix A SPECIFICATIONS Instrument Operating Temperature Range 14 F to 122 F 10 C to
34. ent from accidental damage Display Large 3 Parameter Backlit LCD Display Battery Condition Icon Shows condition of rechargeable batteries Keypad Selects all available functions Selected Fuel Type N GAS LPG LIGHT OIL HEAVY OIL OPT 1 Bituminous Coal OPT 2 Anthracite Coal OPT 3 Coke OPT 4 Butane OPT 5 Wood OPT 6 Bagasse Rubber Boot LCD Display Keypad Battery Condition Icon Selected Fuel Type General Overview Continued Never over saturate your sensors by performing tests on equipment with gas levels beyond the capability of you analyzer Always keep the A762 water trap filter assembly clean and replace the fil ter as necessary Replacement filter part number is A762F This manual will guide you through the functions of the TPI 714 which will give you many years of reliable service Your TPI 714 Flue Gas Analyzer comes complete with the following stan dard accessories TPI 714 Instrument Rubber Boot A765 Infrared Printer A740 model 714A740 comes with the printer Soft Carrying Case A768 Flue Sampling Probe A770 In Line Filter installed on Flue probe A762 Spare In Line Filter one included A762F is a package of 5 filters Temperature Probe GK11M Battery Charger A766 Mini Pump Protection Filter Assembly and spare filters A763 Exhaust Spigot removable A764 Pressure Tubing A774 Static Pressure Tips
35. es to let you know of a malfunction Sensor replacement requires calibration gas Code Displayed Code Definition Possible Causes Corrective Action Flo Err Pump not drawing sample at correct flow rate Blockage kink in flue probe hose Dirty or blocked filter s Worn pump Check and rectify Replace filter s Return to TPI for service InIt O2 Err Oxygen sensor failed to ini tialize Flue probe connected to 714 prior to power up 714 did not purge completely from last sample Worn or defective oxygen sensor Disconnect probe and restart Purge for 20 minutes and restart Return to TPI for sensor replacement or install new sensor Try manually initializing the sensor See Appendix G InIt CO Err Carbon monoxide sensor failed to initialize Flue probe connected to 714 prior to power up 714 did not purge completely from last sample Worn or defective carbon monoxide sensor Disconnect probe and restart Purge for 20 minutes and restart Return to TPI for sensor replacement or install new sensor Try manually initializing the sensor See Appendix G InIt NO Err Nitric oxide sensor failed to initialize Flue probe connected to 714 prior to power up 714 did not purge completely from last sample Worn or defective carbon monoxide sensor Disconnect probe and restart Purge for 20 minutes and restart Return to TPI for sensor repl
36. he A773 does not have to be removed when working with other types of fuels PROCEDURE 1 Pull the yellow thermocouple out of the cord channel in the tubing up to the cord stop 2 Cut the tubing approximately 1 1 2 inches from the end of the A762 water trap fil ter assembly 3 Insert the A773 oil filter into the tube below the A762 4 Cut the remaining piece of tube into a 1 1 2 inch piece and connect it onto the end of the A773 oil filter 5 Press the thermocouple cord into the cord channel The A773 should be replaced when most of the pellets become discolored Appendix B CALIBRATION amp SERVICE It is recommended that the instrument be calibrated every 12 months Please consult Test Products International for further details The following are consumable parts for the instrument In Line Filter Element pkg of 5 User Replaceable A762F Mini Pump Protection Filter Assem User Replaceable A763 Oxygen Sensor User Factory Replaceable A761 Carbon Monoxide Sensor User Factory Replaceable A760 Sensor replacement requires calibration gas The following are accessories for the instrument Flue Temperature Probe w k type Standard A770 K Type Temperature Probe Standard GK11M In Line Filter Complete Standard A762 Mini Pump Protection Filter Assem Standard A763 Battery Charger Standard A766 Rubber Boot Standard A765 Infrared Printer Optional A740 Soft Carrying Case Standard A768 Exhaust Spigot
37. hould be changed During the last 20 seconds of the 30 second purge time the user can scroll through the following Fuels Natural Gas LPG Light Oil Heavy Oil OPT1 Bituminous Coal OPT 2 Anthracite Coal OPT 3 Coke OPT 4 Butane OPT 5 Wood Dry OPT 6 Bagasse by pressing the Up Arrow Key to select the Fuel they are working with When desired fuel is displayed release the key displayed fuel is now selected After the 30 second countdown the instrument is ready to take Flue Temperature Pressure or Leak Detection readings and Combustion Display 1 will be the first display The 714 will auto power off if no keys have been pressed for 10 minutes and the CO level is below 15ppm The Scroll Enter Key allows you to scroll through the combustion displays as out lined in section 5 3 The Func Key is used to select the 714 function mode and allows you to set it to be a combustion analyzer manometer thermometer or combustible leak detector In addition the Func Key is used to access the date and time display See section 6 for more information on the Func Key 4 2 Turning Off Always Before turning off return the instrument to a clean air environment and allow the Carbon Monoxide level to return to below 15ppm and the Oxygen level to return to 20 9 0 3 Press the Power Key to turn the instrument off NOTE Should you attempt to turn the instrument Off and the CO reading is above 15ppm then the instrument will remain On and
38. ion number which you have chosen will stop flashing and after about 3 seconds the instrument will return to the screen function you were previous ly on You have just successfully stored a set of readings which can be either reviewed on screen see 6 or sent to the IR printer see 9 When data is saved all data previously in the address will be overwritten When saving pressure draft readings select an address different from the one used to save combustion readings 6 4 Function 3 Date Time Press the Func Key repeatedly to access the date time screen From this func tion Time Date and Year can be changed 1 Once the Time Date and Year screen is displayed press the Scroll Enter Key once to allow you to change the data 2 Press the Up Arrow Key to Increase the Minutes 3 Press the Down Arrow Key to Decrease the Minutes 4 Press the Scroll Enter Key to confirm the desired Minute and move onto the Hours 5 Repeat steps 2 to 4 to change the Hour the Day the Month and the Year 6 The unit will return to normal after the desired Year has been confirmed 6 5 Function 4 Combustion Analysis Press the Func Key repeatedly to return to Combustion Analysis mode and the default display Combustion display 1 Combustion Display 1 Default Start Up Display Displays Carbon Monoxide CO reading in parts per mil lion ppm Displays calculated Carbon Dioxide CO2 figure in per centage Displays calculated CO CO2
39. ional oil fil ter A773 is installed See Appendix D One or all of the following parameters NOx Ratio CO air free excess air and effi ciency read and print dash es Measured values are such that the calculated values of these parameters are out of range Redo combustion test Since these are calculated values the measure values must be within certain levels for these to display If the measured oxygen level is above 19 9 these parameters won t read These parameters might not display or be applicable in some tests Pressure prints as N A on my combustion analysis print out During combustion analysis the pressure function is dis abled The manometer fea ture is a separate function and must be printed sepa rately Perform the combustion test and then perform a pressure test Save each test in a dif ferent location and print each test Battery will not charge or hold a charge Defective charger or battery Replace the charger or bat tery Send to TPI for service Beeping noise heard during charging Defect in charging circuit or shorted battery Disconnect from the charger and contact TPI at 800 368 5719 Analyzer won t turn off Oxygen and or carbon monoxide levels outside lim its Allow the analyzer to purge longer Appendix E ERROR CODES amp TROUBLESHOOTING Continued Appendix E ERROR CODES amp TROUBLESHOOTING The 714 analyzer will display certain cod
40. ndix B CALIBRATION amp SERVICE It is recommended that the instrument be calibrated every 12 months Please consult Test Products International for further details The following are consumable parts for the instrument In Line Filter Element pkg of 5 User Replaceable A762F Mini Pump Protection Filter Assem User Replaceable A763 Oxygen Sensor User Factory Replaceable A761 Carbon Monoxide Sensor User Factory Replaceable A760 Sensor replacement requires calibration gas The following are accessories for the instrument Flue Temperature Probe w k type Standard A770 K Type Temperature Probe Standard GK11M In Line Filter Complete Standard A762 Mini Pump Protection Filter Assem Standard A763 Battery Charger Standard A766 Rubber Boot Standard A765 Infrared Printer Optional A740 Soft Carrying Case Standard A768 Exhaust Spigot Standard A764 PC Software Optional A769 IRDA RS232C Adapter Optional A767 Appendix C GUARANTEE Your TPI 714 Flue Gas Analyzer is guaranteed free from defects in materials and workmanship for 3 Years from the date of purchase This guarantee does not affect your statuary rights For additional information please refer to the included warranty card or contact TPI at 800 368 5719 To obtain warranty performance or maintenance on your analyzer Include with the product your name address phone number written description of the prob lem and proof of purchase date Carefull
41. nds 10 S 100 readings total 1 reading recorded every 10 seconds 20 S 100 readings total 1 reading recorded every 20 seconds 30 S 100 readings total 1 reading recorded every 30 seconds 01 M 100 readings total 1 reading recorded every 1 minute 03 M 100 readings total 1 reading recorded every 3 minutes by pressing either the Up or Down Arrow Keys 2 Press the Scroll Enter Key once PAGE PA will be displayed and a page number from 0 to 3 will be flash ing Select the required page location that you wish to log the saved data to by pressing the Up and Down Arrow Keys 3 Press the Scroll Enter Key once The instrument will return to the previous screen function with Logger flashing on the top line The instrument will continue logging until all 100 readings have been saved or you press and hold down the Store Logger Key WARNING The instrument will not Turn Off if data logging is commenc ing A beep will be heard to warn you of this fact if you try to turn the instrument off Logged data can only be retrieved by sending the information to a com puter This requires the use of the optional A767 IRDA adapter and either Microsoft Hyper Terminal or the optional PC software part num ber A772 Appendix A SPECIFICATIONS Continued Flue Temperature Probe Construction Pistol Grip with Stainless Steel Shaft Hose Length 2500mm Insertion Length 200mm K Type Thermocouple Accuracy 0 3 1 C Maximum T
42. nt each test Battery will not charge or hold a charge Defective charger or battery Replace the charger or bat tery Send to TPI for service Beeping noise heard during charging Defect in charging circuit or shorted battery Disconnect from the charger and contact TPI at 800 368 5719 Analyzer won t turn off Oxygen and or carbon monoxide levels outside lim its Allow the analyzer to purge longer Appendix E ERROR CODES amp TROUBLESHOOTING Continued Appendix E ERROR CODES amp TROUBLESHOOTING The 714 analyzer will display certain codes to let you know of a malfunction Sensor replacement requires calibration gas Code Displayed Code Definition Possible Causes Corrective Action Flo Err Pump not drawing sample at correct flow rate Blockage kink in flue probe hose Dirty or blocked filter s Worn pump Check and rectify Replace filter s Return to TPI for service InIt O2 Err Oxygen sensor failed to ini tialize Flue probe connected to 714 prior to power up 714 did not purge completely from last sample Worn or defective oxygen sensor Disconnect probe and restart Purge for 20 minutes and restart Return to TPI for sensor replacement or install new sensor Try manually initializing the sensor See Appendix G InIt CO Err Carbon monoxide sensor failed to initialize Flue probe connected to 714 prior to power up 714 did not
43. o starting the device may cause saturation of the sensors due to the higher initial concentration of carbon monox ide that may be encountered upon start up If this happens allow your analyzer to purge in fresh air until the carbon monoxide level returns to 0 ppm and the oxy gen level returns to 20 9 This may take more than an hour depending on how saturated the sensors are The included figures show locations for performing tests on commonly encoun tered equipment Remember to consult with the manufacturer of the device under test for specific test information NOTE When selecting oil as fuel be sure to use the optional oil filter A773 or readings could become erratic See Appendix E for installation instructions COMBUSTION ANALYSIS Continued 5 3 Combustion Displays continued Combustion Display 4 Displays NO in ppm Displays NOx in ppm Combustion Display 5 Displays CO air free CF Combustion Display 6 Displays Temperature reading of Channel 1 T1 Displays Temperature reading of Channel 2 T2 Displays the Differential Temperature Diff between ch1 and ch2 oPEn will be displayed if no K type probe is connected to the thermocouple socket Press the Up Arrow Key to switch between C and F 5 During a combustion test you can save print and send data to a computer See section 7 through 11 Allow readings to stabilize before saving or printing the data 6
44. ocation that you wish to save the data to by pressing the Up and Down Arrow Keys 2 Press the Scroll Enter Key once The Time amp Date of the Saved Data from the selected address location will be displayed flashing on the screen The rest of the Saved Data at this address location can be reviewed by pressing the Up and Down Arrow Keys 4 Press the Scroll Enter Key once when you are finished reviewing the data End will be displayed with YES flashing 5 Press the Scroll Enter Key once to EXIT or 6 Press the Up or Down Arrow Keys End will be displayed with no flashing 7 Press the Scroll Enter Key once to CHOOSE another address location to review and repeat steps 2 to 5 9 PRINTING DATA WARNING To operate correctly there must be a clear line of sight between the Infrared Window on the instrument see 3 3 and the Infrared Window on the IR Printer see Printer instructions Place the printer so it is no closer than approximately 6 inches from the analyzer 1 Press the Print COM Key once rEAL and Stor will be displayed and rEAL will be flashing Selecting rEAL will cause the 714 to print the test data currently on the display real time data Selecting Stor will cause the 714 to print data stored in memory Use the Up and Down Arrow Keys to select the desired print mode and press the Scroll Enter Key REAL Print Mode If rEAL print mode is selected the analyzer will dis play Wait out
45. ocouple thermometer The pump will stop running when in this function 1 Ensure you have a K type probe connected to one or both of the thermo couple sockets T1 or T2 refer to figure below WARNING There is ONLY one correct way to connect the K type thermo couple plug into the socket see 4 2 Forcing the plug into the socket the wrong way may result in damage to the instrument 2 Touch the temperature probe to the item under test and read the displayed temperature Other Features Pressing the Up Arrow Key will toggle between C and F Pressing the Scroll Enter Key will toggle the Differential Temperature ON and OFF oPEn will be displayed if no K type probe is connected to the thermocou ple socket The thermometer func tion allows you to per form temperature tests The pressure function allows you to perform draft and pressure tests Function Display 1 Temperature T1 Temperature T2 Temperature Differential Temperature Function Display 2 Pressure Pressure Reading T2 Temperature 7 SAVING DATA During testing data can be stored for later retrieval by following these steps 1 Press the Store Logger Key once Addr will be displayed on the top line along with SA and a location number from 0 to 9 will be flashing on the screen Select the required address location that you wish to save the data to by pressing the Up and Down Arrow Keys 2 Press the Scroll Enter Key once The locat
46. ombustible leak detector In addition the Func Key is used to access the date and time display See section 6 for more information on the Func Key 4 2 Turning Off Always Before turning off return the instrument to a clean air environment and allow the Carbon Monoxide level to return to below 15ppm and the Oxygen level to return to 20 9 0 3 Press the Power Key to turn the instrument off NOTE Should you attempt to turn the instrument Off and the CO reading is above 15ppm then the instrument will remain On and a short Beep will be heard The Instrument can only be switched off if the CO is below 15ppm The instrument has an auto shut off after 10 minutes should no keys have been pressed for this period and as mentioned above that the CO is below 15ppm Should the CO be above 15ppm then the 10 minute auto shut off countdown will not begin till the CO has gone below 15ppm 4 3 Charging Plug the Charger supplied into the charger socket on the instrument see 3 4 If the instrument is turned on then a charging symbol will be displayed Should the instrument then be turned off turn off automatically or be turned off when the charger is plugged in then the charging symbol will not be displayed BUT the instrument will still be charging The instrument should be charged overnight for a period of 10 to 12 hours and will give over 6 hours Operating Time If a beeping noise is heard during charging disconnect the charger This is an indication
47. onal Appendix F CO ALARM SETTINGS Setting the Carbon Monoxide Alarm Point and or Auto Power Off Mode The CO alarm and auto power off settings can be changed as necessary by per forming the following procedure WARNING Changing parameters in field calibration mode can adversely affect the operation of your analyzer Follow the instructions carefully 1 Press and hold the Scroll Enter Key down until FILD CAL is displayed The 714 will cycle through a 30 second countdown then display CAL Air gAS 2 Use the Up Arrow Key to select Air by making it blink Press the Scroll Enter Key to accept the selection 3 714 Display CAL FLo oPEn Action Press the Func Key to bypass 4 714 Display CAL FLo CLos Action Press the Func Key to bypass 5 714 Display CAL LEAK rEF Action Press the Func Key to bypass 6 714 Display 0ppm Co rEF Action Press the Func Key to bypass 7 714 Display 20 9 o2 rEF Action Press the Func Key to bypass 8 714 Display 0 0mbar 0 0 oFS Action Press the Func Key to bypass 9 714 Display 2000 A r SEt Action If you are not changing the CO alarm point press the Func Key Otherwise press the Up and Down Keys to set the ppm level the alarm will sound at Factory default is 2000ppm Press the Scroll Enter Key to accept the sel
48. ors by performing tests on equipment with gas levels beyond the capability of you analyzer Always keep the A762 water trap filter assembly clean and replace the fil ter as necessary Replacement filter part number is A762F This manual will guide you through the functions of the TPI 714 which will give you many years of reliable service Your TPI 714 Flue Gas Analyzer comes complete with the following stan dard accessories TPI 714 Instrument Rubber Boot A765 Infrared Printer A740 model 714A740 comes with the printer Soft Carrying Case A768 Flue Sampling Probe A770 In Line Filter installed on Flue probe A762 Spare In Line Filter one included A762F is a package of 5 filters Temperature Probe GK11M Battery Charger A766 Mini Pump Protection Filter Assembly and spare filters A763 Exhaust Spigot removable A764 Pressure Tubing A774 Static Pressure Tips A776 x2 Barbed to 1 8 NPT fittings 2 each Gas Valve Adapter A611 Adapter Tubing Instruction Manual Denotes part number Your TPI 714 Flue Gas Analyzer has the following options available PC Software for analysis of Data Logging results see 11 amp Appendix B Infrared PC link for non compatible PCs see 11 amp Appendix B Various Temperature Probes see Appendix B 3 Instrument Overview 3 1 Front View Rubber Boot Protects the instrum
49. robe into the test location of the device under test The probe tip should be in the middle of the flue pipe or exhaust stream Refer to figures 1 through 3 for typical test locations Mini Pump Protection Filter Assembly Ambient Air Connection T2 GK11M Probe Flue Probe Connection T1 In line Filter Assembly Flue Probe Tubing 5 COMBUSTION ANALYSIS 5 1 Overview Testing for carbon monoxide oxygen and stack temperature is very important to the overall safety and efficiency of heating equipment The following guidelines and descriptions are generic and meant to provide you with a basic understand ing of combustion testing TPI always recommends you contact the manufacturer of the device under test obtain information specific to the device and follow the procedures and safety guidelines for performing tests and affecting repairs In general for most applications flue gas samples should be taken prior to the draft diverter or any other opening that allows room air to enter the system This prevents room air from mixing with gases in the flue and diluting the test sample To ensure accurate and consistent combustion tests it is important gas and tem perature samples be taken at the same location This is easy with the TPI flue probe because the temperature sensor is an integral part of the probe Prior to taking a sample the device under test should be on and operating Putting the flue probe in the sample area prior t
50. s the Scroll Enter Key After the 10 second countdown is complete press the Scroll Enter Key twice 7 714 Display 0ppm No rEF Action Press the Scroll Enter Key After the 10 second countdown is complete press the Scroll Enter Key twice 8 714 Display 0 0mbar 0 0 oFS Action Press the Func Key to bypass 9 714 Display 2000 A r SEt Action Press the Func Key to bypass 10 714 Display CAL StoP go Action Press the Up Arrow Key to select StoP by making it blink and press the Scroll Enter Key to accept the selec tion The 714 will return to normal operation If the 714 continues to fail to automatically initialize or this procedure fails to initialize the sensors return your analyzer to Test Products International Appendix F CO ALARM SETTINGS Setting the Carbon Monoxide Alarm Point and or Auto Power Off Mode The CO alarm and auto power off settings can be changed as necessary by per forming the following procedure WARNING Changing parameters in field calibration mode can adversely affect the operation of your analyzer Follow the instructions carefully 1 Press and hold the Scroll Enter Key down until FILD CAL is displayed The 714 will cycle through a 30 second countdown then display CAL Air gAS 2 Use the Up Arrow Key to select Air by making it blink Press the Scroll Enter Key to accept the selection
51. stalled properly Always periodically check and replace the in line filter as needed Always remove water or condensation from the inside of the in line fil ter water trap assembly prior to performing tests Always use the optional oil filter p n A773 when performing tests on oil burning equipment Contents 1 Introduction 2 General Overview 3 Instrument Overview 3 1 Front View 3 2 Back View 3 3 Side Views 3 4 Top View 4 Turning On amp Off and Charging 4 1 Turning On 4 2 Turning Off 4 3 Charging 4 4 Activating The Backlight 5 Combustion Analysis 5 1 Overview 5 2 Procedure 5 3 Combustion Displays 5 4 Typical Test Locations 5 5 Typical Test Results 6 Function Selection 6 1 Function 1 Thermometer 6 2 Function 2 Manometer 6 3 Function 3 Date Time 6 4 Function 4 Combustion Analysis 7 Saving Data 8 Recalling Data 9 Printing Data 10 Logging Data 11 Communicating to a PC Appendix A Specifications Appendix B Calibration amp Service Appendix C Guarantee Appendix D Installing the Optional A773 Filter Appendix E Error Codes and Trouble Shooting Appendix F CO Alarm Settings Appendix G Manually Initializing Sensors 1 Introduction Thank you for purchasing TPI brand products The TPI 714 Flue Gas Analyzer is a state of the art easy to use analyzer designed not only to display and calculate the required readings from a flue but also to cover most of the other m
52. tion 2 General Overview 3 Instrument Overview 3 1 Front View 3 2 Back View 3 3 Side Views 3 4 Top View 4 Turning On amp Off and Charging 4 1 Turning On 4 2 Turning Off 4 3 Charging 4 4 Activating The Backlight 5 Combustion Analysis 5 1 Overview 5 2 Procedure 5 3 Combustion Displays 5 4 Typical Test Locations 5 5 Typical Test Results 6 Function Selection 6 1 Function 1 Thermometer 6 2 Function 2 Manometer 6 3 Function 3 Date Time 6 4 Function 4 Combustion Analysis 7 Saving Data 8 Recalling Data 9 Printing Data 10 Logging Data 11 Communicating to a PC Appendix A Specifications Appendix B Calibration amp Service Appendix C Guarantee Appendix D Installing the Optional A773 Filter Appendix E Error Codes and Trouble Shooting Appendix F CO Alarm Settings Appendix G Manually Initializing Sensors 3 Instrument Overview 3 1 Front View Rubber Boot Protects the instrument from accidental damage Display Large 3 Parameter Backlit LCD Display Battery Condition Icon Shows condition of rechargeable batteries Keypad Selects all available functions Selected Fuel Type N GAS LPG LIGHT OIL HEAVY OIL OPT 1 Bituminous Coal OPT 2 Anthracite Coal OPT 3 Coke OPT 4 Butane OPT 5 Wood OPT 6 Bagasse Rubber Boot LCD Display Keypad Battery Condition Icon Selected Fuel Type General Overview Continued Never over saturate your sens
53. unction 2 Manometer Press the Func Key repeatedly to access the thermometer function In this mode the 714 functions like a differential manometer The pump will stop running when in this function 1 Ensure you have Pressure Sampling Tube connected to one or both of the Pressure Ports and there are no restriction in the tubing see figure below 2 Zero the display by pressing the Down Arrow Key 3 Connect the tube s to the device under test and read the pressure on the display Other Features Pressing the Up Arrow Key will scroll through mbar kPa and inH2O Pressing the Down Arrow Key will Zero the Pressure reading Pressing the Scroll Enter Key will toggle the ch2 temperature reading ON and OFF The 714 incorporates a differential manometer This means when a single hose is connected to the port and pressure is applied the 714 will read positive pressure If a single hose is connected to the port and pressure is applied the 714 will read a negative pressure If both ports are connected to a pressure the 714 will read the difference P1 P2 between the two 6 FUNCTION SELECTION Pressing the Func Key enables access to the different functions available on the 714 Available functions are Combustion Analyzer Thermometer Manometer and Date Time display 6 1 Function 1 Thermometer Press the Func Key repeatedly to access the thermometer function In this mode the 714 functions like a K Type therm
54. until FILD CAL is displayed The 714 will cycle through a 30 second countdown then display CAL Air gAS 2 Use the Up Arrow Key to select Air by making it blink Press the Scroll Enter Key to accept the selection 3 714 Display CAL FLo oPEn Action Press the Func Key to bypass 4 714 Display CAL FLo CLos Action Press the Func Key to bypass 5 714 Display 0ppm Co rEF Action Press the Scroll Enter Key After the 10 second countdown is complete press the Scroll Enter Key twice 6 714 Display 20 9 o2 rEF Action Press the Scroll Enter Key After the 10 second countdown is complete press the Scroll Enter Key twice 7 714 Display 0ppm No rEF Action Press the Scroll Enter Key After the 10 second countdown is complete press the Scroll Enter Key twice 8 714 Display 0 0mbar 0 0 oFS Action Press the Func Key to bypass 9 714 Display 2000 A r SEt Action Press the Func Key to bypass 10 714 Display CAL StoP go Action Press the Up Arrow Key to select StoP by making it blink and press the Scroll Enter Key to accept the selec tion The 714 will return to normal operation If the 714 continues to fail to automatically initialize or this procedure fails to initialize the sensors return your analyzer to Test Products Internati
55. urnace Typical Test Locations Figure 1 Figure 2 COMBUSTION ANALYSIS Continued Test all exhaust ports at the top of the heat exchanger Atmospheric Forced Air Furnace Typical Test Locations COMBUSTION ANALYSIS Continued 5 5 TYPICAL TEST RESULTS Actual test results vary depending on the equip ment under test TPI recommends you check with the manufacturer of the equipment being tested to determine specific acceptable results Oxygen 3 to 6 Carbon Monoxide Less than 100ppm air free Stack Temperature 300 F to 500 F Draft 0 15 inH2O Oxygen 4 to 7 Carbon Monoxide Less than 100ppm air free Stack Temperature 325 F to 625 F Draft 0 15 inH2O Power Burners Gas Fired Power Burners Oil Fired Oxygen 7 to 9 Carbon Monoxide Less than 100ppm air free Stack Temperature 325 F to 500 F Draft 0 15 inH2O to 0 4 inH2O Gas Fired Burners Atmospheric Fan Assist Figure 3 COMBUSTION ANALYSIS Continued Test all exhaust ports at the top of the heat exchanger Atmospheric Forced Air Furnace Typical Test Locations COMBUSTION ANALYSIS Continued 5 5 TYPICAL TEST RESULTS Actual test results vary depending on the equip ment under test TPI recommends you check with the manufacturer of the equipment being tested to determine specific acceptable results Oxygen 3 to 6 Carbon Monoxide Less than 100ppm air free Stack Temperature 300 F to 500 F
56. y package and return to TPI Attn Repair 9615 SW Allen Blvd Suite 104 Beaverton OR 97005 Problem Possible Cause Corrective Action Efficiency reading incorrect NET efficiency selected Ambient temperature probe not plugged in to T2 Incorrect fuel selected Select GROSS efficiency See section 5 3 Plug ambient probe into T2 See section 5 2 Select the proper fuel for the equipment being tested See section 4 1 Readings are erratic when working on oil fired equip ment Oil filter not installed or installed incorrectly Make sure the optional oil fil ter A773 is installed See Appendix D One or all of the following parameters NOx Ratio CO air free excess air and effi ciency read and print dash es Measured values are such that the calculated values of these parameters are out of range Redo combustion test Since these are calculated values the measure values must be within certain levels for these to display If the measured oxygen level is above 19 9 these parameters won t read These parameters might not display or be applicable in some tests Pressure prints as N A on my combustion analysis print out During combustion analysis the pressure function is dis abled The manometer fea ture is a separate function and must be printed sepa rately Perform the combustion test and then perform a pressure test Save each test in a dif ferent location and pri
Download Pdf Manuals
Related Search
Related Contents
Cuisinart PG-29942 User's Manual User Manual Mobile Navigator 5 Symbian Edition BUREAU 19012012 DELL Inspiron 3542 WIRELESS 868 MHz TEMPERATURE STATION Southbend S536D User's Manual Copyright © All rights reserved.
Failed to retrieve file