USER`s Manual CLINOTRONIC 2000
Contents
1. In the area of the sensor address the number of calibrating points is displayed incl zero position In the major display area the total calibration range to be calibrated is seen Result of number of calibration points and the distances chosen By applying RY ZERO SELECT the number of calibration points may be changed The corresponding calibration range will automatically adjust Confirm with ENTER In the display the first angle to set is seen flashing In the window Address Sensor the number 1 is seen for the first calibration point to be set The CLINO 2000 or the external sensor must now be set to the first calibrating position on a suitable calibration gauge rotary table sine bar This angle is displayed flashing on the screen and the figure 1 is seen which stands for the first calibration point Page 31 of 33 If the correct position is installed and a few seconds settling time were allowed the value can be saved by pressing ll ENTER During the reading of the data the display remains stable After successful reading the next calibration point will be displayed and the respective procedure must continue After entering the last calibration point the complete calibration data will be saved in the sensor head During this procedure a moving bar is displayed and when successful finished the CLINO 2000 switches to the measuring mode If something went wrong the following message shows up ERROR 9 It was
2. Remarks Until the input is confirmed by using the E ENTER key every manipulation can be stopped at any time by pressing SEND ESC Result Despite the fact that the CLINO 2000 is still placed at the same angle the display of the instrument now shows the value 0 00 00 This is the so called virtual ZERO DUD UD Bina This new virtual ZERO is the reference for all the future measurements based on this angle originally 0 09 16 OL 0 09 16 Remarks Please see also Relative measurement Chapter 3 9 2 Page 15 of 33 3 3 SELECTION OF THE MEASURING UNIT The desired measuring unit can be selected by repeatedly pressing the key A ON MODE until the selection pointer is below the function UNIT and then the a ENTER is pressed The possible formats may now be chosen by pressing the key RY ZERO SELECT until the desired unit is displayed This selection must be confirmed by pressing lll ENTER The selection remains valid until changed in the same manner as described above The following measuring units can be chosen Xxx Xx DEG Deg Min XX XX xx DEG Deg Min Sec XXXX XX DEG Min Sec XXX XXX DEG Degree 3 decimal XXX XX o o 2 decimal XXX XX Artillerie Promille XX XXXX IREL mm per relative base 4 decimal XXX XX mm REL mm per relative base 2 decimal XXXX XX mRad Milliradian 2 decimal XX XXXX 12 Inch per 12 Inch 4 decimal XX
3. Confirm with ENTER Adjustment of angle at O Pre set value will be displayed The dowel pins are to be inserted in such a way that the instrument would display 0 when two pins are on a horizontal plane E g hold the instrument on the side of a measuring and setting plate the pins on top Press HOLD keep the instrument stable First displayed is oooooo then the measured value is seen Confirm with ENTER Adjustment of angle at 45 Pre set value will be displayed The dowel pins are to be inserted in such a way that the instrument would display 45 when two pins are ona horizontal plane E g hold the instrument on the side of a measuring and setting plate the pins on top Press HOLD keep the instrument stable First displayed is then the measured value is seen Confirm with ENTER During the sampling of the collected values first displayed 1 then at the end it will automatically switch in the measuring mode By applying a reversal measurement set the instrument to absolute zero see point 3 2 1 The CLINO 2000 is now calibrated and can be used for measuring Point 3 9 1 Press A ON MODE several times until the selection pointer is below ZERO The calibration aids dowel pins must be stored grease applied and any damage is to be prevented Also the holes in the instrument must remain free of dust and dirt Page 29 of 33 5 5 2 FINE CALIBRATION OVE
4. Page 30 of 33 For the calibration procedure the selection pointer must be moved to the position CALIB by repeatedly pressing the key A ON MODE If reached it needs to be accepted by pressing ENTER By using the key RY ZERO SELECT the respective port may be chosen Port A for the internal sensor in the CLINO 2000 Port B for a possible external sensor The selected possibility must be confirmed by pressing MI ENTER After this the CLINO 2000 is looking for all the connected sensors respectively their address The address of the sensor will be displayed flashing During the searching procedure the frame of the Sensor address will flash in circular motion below the selected port Are several sensors connected the required one must be selected by the key 2 ZERO SELECT and confirmed with MM ENTER Immediately after this the calibration procedure is starting In the display the last entered distance between two calibration points is seen By using the key 1 9 ZEROISELECT t this distance may be entered from new Two basic possibilities are available l Choose calibration point distance is between 0 5 and 7 5 degrees in steps of 0 5 degrees II Choose calibration point distance is between 1 0 and 15 0 mm m in steps of 1 0 mm m For calibrating the CLINO 2000 it is recommended to use 5 7 5 steps After choosing the distance and confirming with ENTER the procedure continuous as follows
5. Key word A Absolute measurement Absolute measurement Relative measurement Absolute measurement Relative measurement Automatic shut off deactivating B Battery display BATT Battery lifetime Calibration Calibration Calibration Calibration principle Cancelling any task Changing of sensor address Checking the functions Data format of transmission Description of panel functions Description of the CLINO 2000 Description of the individual keys Description of the panel Differential measurement with two CLINO 2000 Differential resp reference measurement with two CLINO 2000 E ENTER Key Error messages Error messages F Filter selection Fine calibration over the whole measuring range Function SEND Print Function General H HOLD key HOLD Function How to use the CLINO 2000 Hyper terminal program Introduction Key functions Limits Limits of error LIMITS measuring with Manual set up of the virtual ZERO Measuring by setting limits Measuring units Measuring units ON MODE key Page 3 of 33 Chapter 2 2 1 5 4 2 1 1 1 2 2 1 3 11 2 2 1 3 7 5 1 2 2 1 2 2 2 2 2 3 10 6 3 10 3 10 Page 22 22 22 29 29 33 8 27 11 18 10 25 25 8 27 27 21 30 18 27 17 11 19 24 24 13 24 10 16 Panel and display Possible configuration Power supply external Quick calibration Reference resp differential measurement with two CLINO 2000 REL zero o
6. ONLINE Na The Zero Offset Basic offset computed by applying the reversal method is the deviation of the instrument s ZERO to the absolute ZERO The display on the CLINO 2000 therefore is the Displayed value Value of the CLINO 2000 minus Zero Offset The reversal measurement to establish the exact Zero Offset Basic offset should be done periodically if high quality measurements are required Especially after longer intervals the method is useful 3 2 2 ZERO VIRTUAL I Manual set up of the virtual ZERO One possibility of using ZERO virtual means to establish an artificial plane respectively angle to be chosen as absolute zero Another possibility is to change the existing Zero Offset Basic offset of the CLINO 2000 e g established by means of using the reversal method Only well skilled and trained user should make use of this feature Example Initial position The display on the CLINO 2000 shows the measured value of 0 07 20 This value is the true inclination of the instrument PortiSenser A 0 07 20 O01 20 m Only one sensor internal sensor of the CLINO 2000 or the external sensor connected to Port B at a time can be set to ZERO virtual This sensor must be selected first The selection pointer must be moved below the marking ZERO in the display of the CLINO 2000 by pressing the key A ONIMODE several times If reached the selection must
7. is calculated by adding the two readings and dividing the result by two Page 12 of 33 2 The ZERO absolute will automatically be set by applying the reversal measurement Taking two measurements on the same spot but in the 180 opposite direction For this procedure a suitable surface must be chosen rigid flat and as horizontal as possible On this surface the CLINO 2000 must be placed the exact position of the instrument is to be marked on the Measurement A Measurement B The ZERO OFFSET is stored in the CLINO 2000 E 1 l Port Sensor DES A SENSOR ABSOLUTE RELZERO UNIT ZERO LIMITS FILTER ADDRESS CALIB ONLINE EP ep Gi EP UND Moms ae a Lu EN EE UNE a GEN Port Sensor DEG SENSOR ABSOLUTE RELZERO UNIT 2 02 L LIMITS FILTER ADDRESS CALIB T4 11 HN e Port Sensor DEG SENSOR ABSOLUTE RELZERO UNIT ZERO LIMITS FILTER ADDRESS CALIB ONLINE If this Zero Offset Basic offset must now be saved the key EE ENTER must be pressed Immediately after this the selection pointer jumps to the measuring mode absolute and the measuring result under consideration of the Zero Offset Basic offset will be displayed AED GE UND LL Eu Port Sensor DEG SENSOR ARSOLLTF RFI ZFRO UNIT 5 LIMITS FILTER ADDRESS GALIB FRO
8. FILTER ADDRESS CAUB ONLINE direction indicator is flashing M 2 14 of 33 C UN The CLINO 2000 must now sit at position one and after a few seconds settling time allowed by pressing the key HOLD ta in the first reading is collected This first value is displayed and the Port Sensor selection pointer below ZERO is flashing which means the A eee second reading is expected et LIMTS FILTER ADDRESS GALIB After few seconds settling time the key HOLD must be pressed again for collecting the second reading The instrument remains at the same position without turning 180 T u The newly computed Zero Offset is now displayed on the Tta i l 7 CLINO 2000 Un DEG The displayed Zero Offset is calculated by adding the two C ums readings and dividing the result by two If this Zero Offset must now be saved the key lal ENTER must be pressed Immediately after this the selection pointer jumps to 4 t It It the measuring mode absolute and the measuring result under consideration of the Zero Offset Basic offset will be displayed 50 Port Sensor SENSOR ABSOLUTE RELZERO UNIT ZERO LIMITS FILTER ADDRESS CALIB ONLINE The ZERO OFFSET is stored in the CLINO 2000 Displayed value Value of the CLINO 2000 minus Zero Offset
9. SENSOR ABSOLUTE REL ZERO UNIT ZERO ly LIMITS FILTER ADDRESS ONLINE j ml Nd The desired mode must be confirmed with ENTER Means limits ON or limits OFF If the limits must not be stored the procedure can be cancelled by using sEND ESC If during a measurement the limits f TENE a OCOLIT L3 ds are reached the display starts E i I2 LI flashing alternatively the measured value and the remark HI or LO depending on the limit reached are OQ lugs monss cue omne seen 1 7 Port Sensor DEG Port Sensor SENSOR ABSOLUTE RELZERO UNIT ZERO mm SENSOR ABSOLUTE RELZERO ZERO mm LI LIMITS FILTER ADDRESS ONLINE E At the same time two signal will be transmitted to the port RS 232 The first signal may be used for an external control action or a relay 12V 500 mA may be activated The second signal transmits the polarity of the signal in relation to the set limitation Page 24 of 33 3 11 DIFFERENTIAL RESPECT REFERENCE MEASUREMENT WITH TWO CLINOTRONIC 2000 Applying differential measurement respectively reference measurement means by definition using two instruments a Measuring Instrument A and a Reference Instrument B whereas the difference between the two angles is of interest A B This difference remains unchanged when the same angular change is applied on both instruments This measurement is a special form of a relative measure
10. XXXX 710 Inch per 10 Inch 4 decimal XXXX XX mm m mm per m 2 decimal XXX XXX GON New degree 3 decimal For a measurement with relative base length the selection Y mn pointer must be moved by repeatedly pressing the A ON MODE Tos key to the position UNIT Confirm with ENTER SENSOR ABSOLUTE RELZERO UNI pos By using the keys 2 ZERO SELECT the desired unit LS Md can now be chosen mm REL or inch REL The selection must be confirmed with the key i ENTER On the display the direction indicator is flashing alternatively plus R7 1nmumun or minus the existing base length is shown Standard length is T hud LR LI LI LI 1000 mm or 10 od By pressing IN ENTER this value may be accepted or with SENSOR ABSOLUTE RELZERO 0 ZERO SELECT the relative base length may be Du E adjusted as required and then finally confirmed with BS ENTER In the relative mode the angle is related to the height X as the elevation over the set relative base length in the chosen x tan o unit in mm or Inch Exception RESET see point 3 1 Relative Base Page 16 of 33 3 4 HOLD FUNCTION This function is available in all measuring modes Place the CLINO 2000 on a flat stable surface Use the key nn HOLD While the CLINO 2000 waits for a valid E uuuuuu measurement two identical values in succession the display Port Sensor PER will sh
11. be confirmed by pressing The last computed or manually entered value Zero Offset Basic offset is displayed In our example 00 00 00 The direction indicator is flashing This displayed value may now be changed manually by applying the keys ZERO SELECT to the desired value e g from 0 00 00 to new 0 07 20 The direction indicator is flashing The value is now the new Zero Offset Page 13 of 33 Port Sensor DEG SENSOR ABSOLUTE REL ZERO UNIT ZERO Lu IMTS FIITFR ADDRFSS CAIR ONLINF WA LI c Port Sensor DEG SENSOR ABSOLUTE REL ZERO UNIT ZERO 1 Lu IMTS FIITFR ANDRESS CAIR ONLINF N If this so manually established Zero Offset must now be nmm mm saved the key Mii ENTER must be pressed The new Zero 4 l Offset is now saved CLINO 2000 Immediately after this Bue m RE SEG the selection pointer jumps to the measuring mode absolute A ABSOLUTE RELZERO UNIT ZERO and the measuring result under consideration of the new anness oN INE Zero Offset will be displayed The following new value will gt now be displayed in the absolute mode ZERO OFFSET is stored in the CLINO 2000 Displayed value Value of the CLINO 2000 minus Zero Offset Remarks Until the input is confirmed by using the i ENTER key every manipulation can b
12. the port RS 232 if the measured value rises above the upper limit or falls below the lower limit For the setting of limits the selection pointer must be moved to the position E l P10 m mm LIMITS by repeatedly pressing the key A ONIMODE If reached it needs ZA m om to be accepted by pressing MISI ENTER In the display of the CLINO 2000 DEG a possible installed lower limit is displayed with a marking LO This value SENSOR ABSOLUTE RELZERO UNIT ZERO may NOW manually be adjusted to the required value by using 410 FILTER ADDRESS CALB ONLINE f M ZEROISELECT If reached it must be accepted by pressing ll ENTER After this the upper value is displayed with the marking HI This value may now manually be adjusted to the required value by using ZERO SELECT If reached it must be accepted by pressing ll ENTER ATTENTION The upper limit HI must always absolutely seen be higher than the lower limit LOW CFR After the confirmation in the display the upper limit is seen remark HI This value can be changed manually by using the keys ae r1 29 ft M ZEROISELECT and after confirming with ENTER i the limit is set SENSOR ABSOLUTE RELZERO UNIT ERO HI IMTS FILTER ADDRESS ONLINE After the saving of the upper limit the display shows OFF or ON By using the key A ONIMODE the display jumps between the two possibilities
13. 00100 Rad Verbunden 00 00 27 Autom Erkenn 9600 8 N 2 Rl 4 resp 100 uRad CLINO 2000 with address 50 is connected to port A one external sensor with address 21 is connected to port B Measuring mode A amp B alternatively measured Meaning of the display LM2000 HyperTerminal Datei Bearbeiten Ansicht Anruf bertragung 031 Continuous number Da als 23 e A050 Port A address 50 031 A050 0 000104 B021 0 001079 0 000104 0 000104 Rad resp 104 uRad B021 Port address 0 Verbunden 00 0417 l amp domEem 08 2 GROSS 7 0 001079 0 001079 Rad resp 1079 Rad CLINO 2000 with address 50 is connected to port A one external sensor with address 21 is connected to port B Measuring mode A B differential measurement Meaning of the display LM2000 HyperT erminal Datei Bearbeiten Ansicht Anruf bertragung 141 Continuous number 141 A050 B021 0 001144 A050 B021 Measuring mode Differential mode Port CLINO 2000 with address 50 minus Port B sensor with address 21 0 001144 Difference Verbunden 0001 12 Atom Erkenn 98008N 2 GROE 7 Sensor A B 0 001144 Rad resp 1144 Rad Page 20 of 33 3 6 TERMINATING A MEASUREMENT OR A SET UP In case of the required canceling of a measuring procedure or the changing of parameters the process may be ended at any time and the previous state wil
14. E mum mL O me A B pe el 5 Ll mm CLINO 2000 connected to a PC RS232 Page 6 of 33 CLINO 2000 connected with a MINILEVEL or LEVELTRONIC for differential mode All the above configurations can be used For specifications of the sensors and the instruments see the respective data sheets 2 STARTING HOW TO USE THE INSTRUMENT 2 1 DESCRIPTION OF PANEL FUNCTIONS 2 1 1 PANEL AND DISPLAY Batterie Installation Batteries installation EEE Querlibelle Setting vial Batteriedeckel Battery cover Tasten Keys Masseinheiten Hauptanzeige Measuring units Main display Messkanal Port xterne Anschl sse Measuring channel port xternal connections Sensoradresse Batterieanzeige Auswahlschalter MODE Sensor address Battery display Selection pointer MODE Page 7 of 33 2 2 KEY FUNCTIONS 2 2 1 DESCRIPTION OF THE INDIVIDUAL KEYS Function 1 Function 2 Function 3 Function 4 Pressing the ON MODE key starts the CLINO 2000 When keeping the key pressed all the LCD elements are lighted After releasing the key the instrument switches to the measuring mode and the calibration data of the internal sensor has been collected After a short period of time the actual angle will be displayed in the last used measuring unit In case of trouble an error message will be displayed See point 4 When pressing the A
15. GE 7 s WYLER AG Tel 0041 0 52 233 66 66 Im Holderli Fax 0041 0 52 233 20 53 CH 8405 WINTERTHUR Switzerland Homepage http www wylerag com E Mail wyler wylerag com USER s Manual CLINOTRONIC 2000 File man_c20e doc 25 7 1999 CONTENT PAGE 1 INTRODUCTION 5 1 1 DESCRIPTION OF THE CLINO 2000 5 1 2 STARTING 5 1 3 POSSIBLE CONFIGURATION 6 2 STARTING HOW TO USE THE INSTRUMENT 7 2 1 DESCRIPTION OF PANEL FUNCTIONS 7 2 2 KEY FUNCTIONS 8 3 HOW TO USE THE CLINO 2000 11 3 1 CHECKING THE FUNCTIONS 11 3 2 ZERO SETTING ZERO ABSOLUTE AND ZERO VIRTUAL 12 3 3 SELECTING THE MEASURING UNIT 16 3 4 HOLD FUNCTION 17 3 5 FUNCTION SEND PRINT FUNCTION 18 3 6 TERMINATING A MEASUREMENT OR A SET UP 21 3 7 SELECTING A SENSOR OR AN INSTRUMENT 21 3 8 SELECTING CORRECT FILTER TYPE FOR VARIOUS APPLICATIONS 21 3 9 ABSOLUTE MEASUREMENT RELATIVE MEASUREMENT 22 3 10 MEASURING BY SETTING LIMITS 24 3 11 DIFFERENTIAL MEASUREMENT WITH TWO CLINO 2000 25 4 ERROR MESSAGES 27 5 SERVICE 27 5 1 GENERAL 2 5 2 STORAGE 27 5 3 SPARE PARTS 2 5 4 CHANGING OF SENSOR ADDRESS 2 5 5 CALIBRATION 29 6 TECHNICAL DATA 33 Modifications Datum Date Ge ndert durch Beschreibung der nderung Modified by Description of modifications 7 of July 2000 HEH New chapter 3 11 Differential measurement with two CLINO 2000 24 of Sept 2001 HEH Units page 14 modified 8 of Jan 2003 HEH MO Index included Page 2 of 33 INDEX
16. ON MODE key for more than 3 seconds the display starts flashing and the automatic shut off is disabled In the standard mode the instrument shut off after about 5 minutes To turn off the instrument press the A ON MODE key for more than 3 seconds until the display disappears Exeption In case the instrument is powered by an external power supply the instrument is never shutting off automatically The ON MODE is also used for moving the selection pointer on the lower display area By pressing this key the selection pointer moves from one position to the next For shutting off the CLINO 2000 the A ON MODE key needs to be pressed for 3 4 seconds until the displayed figures disappear Used for ZERO Setting of saved memory values in the function REL ZERO To set the memory in this function to ZERO press the A ON MODE key the values will become ZERO This has to be confirmed by pressing ENTER The key EE ENTER is used to save an entered value or to accept a selected function Function 1 Function 2 Function 3 The key amp SEND is used for the transmission of a measuring value through the RS232 port or to send the value to a printer or another connected system Through the same port the values may also be transmitted to a PC or Laptop for further treatment of the data Also used for canceling any HOLD function and return to the measuring mode Canceling any task not yet finished Page 8 o
17. R THE WHOLE MEASURING RANGE The fine calibration may be required when the existing calibration was lost the accuracy is not sufficient anymore a connected sensor or the measuring range has changed For such a calibration procedure a precise sine bar or an accurate index table should be available Remarks In order to guarantee the precise functioning of the instrument a minimum of seven calibration points are required over the whole measuring range Principle of the calibration method F1 F2 Hz Calibrating a digital system with the Levelmeter 2000 50 40 Angle degrees The bases of all the measurements with digital technology like the CLINO 2000 and the connected sensors instruments are the calibration data stored in the sensor heads Throughout the whole measuring range a number of reference calibrating points are stored in the sensor heads The linearity of the measuring range depends on the number of these reference calibrating points Each measured angle delivers a signal in form of a frequency and to these reference frequencies the respective calibration points are matched From these values a calibration curve will be established The measured angles between these calibration points are computed by interpolation and displayed on the screen in the desired measuring unit All sensors instruments leaving the factory at WYLER s are calibrated according the measuring range and have undergone a severe quality test procedure
18. confirmed with MB ENTER the position indicator jumps to the position SENSOR select sensor 5 SERVICE 5 1 GENERAL The CLINO 2000 needs no special service other than the regular cleaning 5 2 STORAGE The CLINO 2000 should be stored in a save place best in the transportation case For longer storage periods it is recommended to remove the batteries 5 3 SPARE PARTS The following spares are available Batteries Cables 5 4 CHANGING OF SENSOR ADDRESS Move the position indicator to the position ADDRESS by repeatedly pressing the key A ON MODE and confirm the setting with MB ENTER If only one sensor is connected as described above the respective port and address number is displayed flashing SENSOR ABSOLUTE RELZERO UMT ZERO LIMITS FILTER as CALIE ONLINE Remarks If more than one sensor is connected the error message ERROR 5 is displayed More than one sensor found i101 p va Using the keys MW ZEROISELECT the new address can 4 Ji UL L3 be entered and confirmed by pressing lll ENTER If the PoniSenser DEG procedure was successful the measurement starts with the A SENSOR ABSOLUTE RELZERO UNT ZERO newly given address c 5 UMTS FILTER ADDRESS CALIB ONLINE Page 27 of 33 Addresses of external Zerotronic sensors may be chosen between no 1 and no 254 Addresses of external Measuring instruments Minilevel NT Leveltronic NT may only be chosen between no 1 and no 32 The address no 255 is reserved for
19. cr gt MessMode B 9xy Baaa 0 000000 lt cr gt MessMode A minusB 9xy Aaaa Baaa 0 000000 lt cr gt MessMode A B 9xy 0 000000 Baaa 0 000000 lt cr gt aaa Sensor Address 1 255 resp 0 when address undefined 0 General error 1 Port A error 2 Port B error y 0 Menu active no error 20 Error codes According to manual Format of transmission asynchrony 7Bit 2 Stopbits no parity Page 18 of 33 Example using the Hyper Terminal of Windows NT or Windows Terminal program Example is WIN NT 1 Open the Terminal Program in Windows Accessories and insert a name Beschreibung der Verbindung Confirm with OK Li 2000 2 Enter the serial port definition connected to the CLINO 2000 Verbinden mit Confirm with OK 3 Enter the parameters Bits per Second 9600 Data bits T Parity no Stop bits 2 Protocol no Confirm with OK The HyperTerminal Windows appear Repeatedly pressing the key SEND ESC the actual value will be transmitted in Rad Alternatively the value can be called by pressing the key P on the PC keyboard 19 of 33 CLINO 2000 with address 50 connected to port A Meaning of the LM 2000 HyperT erminal display Datei Bearbeiten Ansicht Anruf bertragung 7 Del 33 soles 068 Continuous number 3 measurements per second are performed 068 A050 0 000100 A050 Port A address 50 0 000100 0 0
20. e stopped any time by pressing e SEND ESC Result Despite the fact that CLINO 2000 is still placed at the same angle the display of the instrument now shows the value 0 00 00 This is the so called virtual ZERO EZ Ta m This new virtual ZERO is the reference for all the future measurements based on this angle originally 0 07 20 OL 070720 Il Automatic set up of the virtual ZERO Practically quite often a measuring surface is not absolutely horizontal but should be the reference plane for other measurements In these cases it is useful to set the basic value for this surface to 0 00 00 This can be done by manually setting the virtual zero as described before or by automatically define the value with the CLINO 2000 set up It is done by using a reversal measurement mode without turning the instrument 180 deg Procedure Initial position The display shows the measured value on the CLINO 2000 T0 09 16 This value is the true inclination of the CLINO 2000 Us of a QL 070916 The selection pointer must be moved below the marking ZERO the x ly mor Fa display of the CLINO 2000 by pressing the key A ON MODE several d an 1 Li times If reached the selection must be confirmed by pressing d Part Sensor Bi ENTER A SENSOR ABSOLUTE RELZERO UNIT ZERO The last computed or manually entered value Zero Offset The ou UMTS
21. es oe oa selection can be accepted by pressing MB ENTER After this X the CLINO 2000 is looking for all the connected sensors respectively their address Up to 255 sensors may be connected The address of the first sensor will be displayed flashing During the searching procedure the frame of the sensor address will flash in circular motion below the selected port LIMTS FILTER ADDRESS GALIB ONLINE If more than one sensor is connected to the port B the one desired can be selected by pressing MW ZEROISELECT and confirmed with I ENTER In case of differential measurement between two sensors the Same procedure must be repeated for port B After pressing MB ENTER the respective measurement starts 3 8 SELECTING CORRECT FILTER TYPE FOR VARIOUS APPLICATIONS A number of different types of filters are incorporated The standard factory setting is filter type no 5 To change this setting the selection pointer must be moved below Filter After pressing MB ENTER the actual setting is displayed The number may be changed from 1 to 10 by using the or arrows and then be confirmed by pressing Ml ENTER The most suitable type of filter is best found out by practically testing as the measuring conditions have quite an influence Generally spoken it can be said that the lowest figure almost no filter sends the collected signals rapidly to the display possibly resulting in an unstable di
22. f 33 The key 4t 9 ZERO can be used to change a number of possible parameters like e g Measuring units Measuring units respectively ports A B Changing of the measuring range Changing of the instruments address Set up the relative base length Set up of the Zero Offset Set up of the REL Zero Offset etc Function 1 Function 2 With the HOLD key a measured value may frozen This value will be displayed until by pressing the key SEND the CLINO 2000 will return to the measuring mode When using the mode REL ZERO and ZERO the actual measuring value can be accepted by pressing the HOLD key Page 9 of 33 2 2 2 DESCRIPTION OF THE PANEL mm declining to the right negative angle a ca inclining to the right positive angle With the selection pointer the intended function can be chosen SENSOR Select the active sensor port ABSOLUTE Absolute measuring mode REL ZERO Relative measuring mode UNIT Selection of measuring units ZERO Set absolute zero LIMITS Output of control signals FILTER Adjustment of filter type ADDRESS Display and changing of sensor addresses CALIB Calibration of sensors ONLINE Display of active communication through port When the battery power is low the sign BATT is constantly on The actual measuring unit is displayed 10 basic units are available and most of them can in addition be changed into
23. ffset Relative base length measurement Relative measurement Relative measurement absolute measurement Reset Resolution Reversal measurement Selecting a sensor or an instrument Selecting correct filter type for various applications Selecting the measuring unit Selection pointer SEND ESC key Sensor instrument selection Sensor Port A B Sensor address Sensor address changing Service Service Settling time Shutting off the instrument Spare parts Spare parts Starting How to use the instrument Storage Storage Technical data Technical data Temperature range Terminating a measurement or a set up Terminating a measurement or a set up V Virtual zero automatic set up 2 ZERO absolute with reversal measurement Zero offset ZERO VIRTUAL ZERO VIRTUAL ZERO SEL key ZERO Setting ZERO absolute and ZERO virtual ZEROTRONIC sensor Page 4 of 33 2 11 551 2 2 1 2 6 3 2 2 29 25 1 22 22 11 33 12 21 21 16 10 21 10 10 27 27 8 27 27 27 27 21 21 14 12 12 13 12 28 1 INTRODUCTION 1 1 DESCRIPTION OF THE CLINO 2000 The CLINO 2000 was developed by WYLER AG as an intelligent display and measuring unit together with the digital instrument s family ZEROTRONIC With this instruments all the sensors and instruments of the ZEROTRONIC family can be used See point 1 3 Configuration The CLINO 2000 is a Measuring instrument Display un
24. he requirement is now to set the instrument s inclination at an angle of zero and use this angle as a reference for future measurements OL 00 10 00 Page 22 of 33 Fd The selection pointer must be moved below the marking l emomm mm REL ZERO in the display of the CLINO 2000 by pressing the pid key A ON MODE several times If reached the selection must P TETEE be confirmed by pressing E ENTER A SENSOR ABSOLUTE RELZERO UNT ZERO The last computed or manually entered value REL ZERO GO fums Aoress cae value in our example 0 00 00 The direction indicator is flashing The CLINO 2000 must now be at the correct position and after a eat To few seconds settling time allowed by pressing the key I i T HOLD the reading is collected This value REL ZERO PortiSensor EG OFFSET is displayed A SENSOR ABSOLUTE ee UNIT ZERO eL LIMITS FILTER ADDRESS GALIB ONLINE Should this value no be stored press lll ENTER This relative zero value is stored in the CLINO 2000 The display jumps into the measuring mode REL ZERO the display is 0700 00 and e mm the measurement continues by considering the REL ZERO 4 HE H EDEL Port Sensor DEG As an alternative the value can also be changed manually by an ni i i using the key f M ZEROISELECT before confirming E with Result Despite the fact that the CLINO 2000 is s
25. ibration data not available The display shows the malfunctioning item port and address also possible when the battery power is not sufficient Communication with sensor not possible For other error messages see point 4 To reinstall the basic set up a RESET may be done at any time The procedure is as follows Both keys A ON MODE and ENTER must be pressed simultaneously for at least one second All the manually installed data is lost and the CLINO 2000 will be set to the following basic standards Basic standards Measuring mode absolute Measuring unit the next possible and useful unit Relative Base 1000 mm ZERO OFFSET absolute remains intact Sensor PORT A no Sensor ERROR 2 Filter No 5 Immediately after RESET the error message 2 is displayed This means the connected sensor must be addressed correctly from new See point 3 7 Selecting a sensor or an instrument Page 11 of 33 3 2 Remarks Using ZERO SETTING two different tasks can be realized ZERO SETTING ZERO ABSOLUTE AND ZERO VIRTUAL a ZERO ABSOLUTE The instrument shows 0 when the angle of the mounting surface is exactly positioned in the direction of the center of the earth b ZERO VIRTUAL The instrument shows O at any predefined angle 3 2 1 ZERO ABSOLUTE WITH REVERSAL MEASUREMENT The ZERO absolute is the basis for all the inclination measurements in the absolute mode For such a measurement if high precision is required it is of u
26. it On the CLINO 2000 a number of parameters can be changed and adjusted Measuring unit Measuring mode Relative base length and many more The CLINO 2000 is compatible with the other instruments of the ZEROTRONIC family by using the RS485 serial communication The measuring values may be transmitted to a printer or a PC Laptop via RS232 output The measuring principle of the ZEROTRONIC family is based on a changing of the capacity of a condenser built of two electrodes and a pendulum in form of a shield installed in between The measured change of capacity is directly influenced by the change of the inclination of the pendulum This capacity change is the primarily used signal for the angle to be measured The measuring system is designed to be completely antimagnetic The basic signal received in the CLINO 2000 will be computed into an angle by comparing the signal with a reference curve stored and then displayed in the required unit 1 2 STARTING Before working with the CLINO 2000 it is strongly recommended to carefully read this manual first This will avoid malfunctioning by making unsuitable manipulations like e g deleting the calibration data in the sensor head etc Page 5 of 33 1 3 POSSIBLE CONFIGURATION Stand alone unit Two units connected for remote display or differential mode max 15m CLINO 2000 connected with ZEROTRONIC Sensor max 15m ww E EL
27. l be installed by pressing SENDI ESC This is true only as long as no change has been accepted by pressing lll ENTER before 3 7 SELECTING A SENSOR OR AN INSTRUMENT With the CLINO 2000 the possibility exists to display the measured values of the instrument itself as well as the difference between the instrument and an additional instrument or a connected sensor ZEROTRONIC In case of the differential measurement the second instrument must be connected to the port B External sensor B It is not possible to measure the difference of two sensors connected to the same port Basically the following set up is possible Measurement of the CLINO 2000 internal sensor A corresponding to Sensor Port A Measurement of one or more instruments connected to port B Differential measurement between the CLINO 2000 and another instrument at port B Alternating display of the CLINO 2000 and an instrument connected to port B For differential measurement respectively measurement with a reference CLINOTRONIC 2000 see chapter 3 11 Differential respect Reference Measurement with two CLINOTRONIC 2000 For choosing the measuring mode and the address of the sensors connected the ON MODE key must be pressed repeatedly until the selection pointer is below SENSOR and accepted by pressing la ENTER The possibilities Port A Port B Port A B or Port A B may be selected by N Fra applying the key M ZERO SELECT The desired Eee acy a
28. ment The differential measurement has the greatest advantages in the following tasks Measurements made on objects under influence of vibration Measurements made on objects and systems not rigid enough unstable flimsy Measuring Instrument A Reference Instrument B Preconditions e Both instruments are powered by either internal batteries in each instrument or by using an external power supply connected to one of the two instrument s free socket RS232 e Both instruments are connected together using the ports EXTERNAL SENSOR with a bus cable max 1000 m Start up Both instruments are placed in position and properly connected and power supply installed as mentioned above First the Reference instrument must be started by pressing the button A ON MODE The ON MODE key must now be pressed R repeatedly until the selection pointer is below ONLINE m and accepted by pressing lll ENTER The display ME M GCSE EIS COME shows now REF what means that the instrument is in LIMTS FILTER ADDRESS CALIB ONLINE the reference mode Page 25 of 33 Secondly the Measuring instrument must be started by pressing the button A ON MODE The ON MODE key must now be pressed repeatedly until the selection pointer is below SENSOR and confirmed by pressing ENTER The Port A B must be selected for choosing differential mode applying the key FK ZERO SELECT until the desired selection is displayed
29. n using external power supply the batteries may remain in the instrument Rechargeable batteries must be charged outside of the instrument Page 33 of 33
30. ng SENDIESC the further procedure may be cancelled immediately The displayed value may now be accepted or the process must be repeated If after 60 seconds no value was accepted the ERROR 7 message is displayed HOLD function has not been successfully completed within 60 seconds The procedure needs to be repeated By using the function RESET the CLINO 2000 will return to the standard settings With the exception of the calibration data all previous adjustments are lost Page 17 of 33 3 5 FUNCTION SEND PRINT FUNCTION Using the key SEND ESC will send the displayed value through the port RS 232 to a connected PC or Laptop via the RS232 port This function can also be used in combination with HOLD in order to transmit the frozen value The function SEND may be activated from a connected PC Laptop by transmitting a letter as letter through the serial port RS 232 DATA FORMAT OUT PORT Measurement active MessMode A sss sn nnnnnn lt cr gt MessMode B sss Baaa sn nnnnnn lt cr gt MessMode A minusB sss Aaaa Baaa sn nnnnnn cr MessMode A B sss sn nnnnnn Baaa sn nnnnnn lt cr gt sss 0 255 Continuous no aaa Sensor Address 1 255 e g 004 ML LT NT 1 32 sn nnnnnn 9 999999 Positive Overrange 9 999999 Negative Overrange other value Angle in rad e g 0 226349 Menue aktive or error 0 000000 means no data available MessMode A 9xy 0 000000 lt
31. not possible to save the calibration data successfully in the sensor head Page 32 of 33 6 TECHNICAL DATA Measuring range 45 Arc deg Messbereich Optional 10 Arc deg 30 Arc deg 60 Arc deg Calibration Built in software and 23 Setting points Kalibrierung calibration aids Interne Kalibriersoftware und 23 Kalibrierpunkte mitgelieferte Kalibrierhilfen Messzeit Anzeige nach Resolution Depending on units set 5 Sec of arc 0 025 mm m Auflosung Abhangig von ausgewahlter Masseinheit Limits of error lt 5 Sec of arc 0 07 R O Fehlergrenze After quick calibration using lt 30 Sec of arc the calibration aids Nach Kurzkalibrierung mit den Kalibrierhilfen Data connection Special cables RS 485 asynchr 7 Bit Anschluss Spezialkabel 2 Stopbits no parity 9600 Baud Power supply with Batteries 2 x Size AA 1 5V Alkaline batteries Lifetime 40 60 hrs Stromversorgung mit Batterien Betriebsdauer 2 x Size AF 1 2 V NiMH rechargeable 30 50 hrs Externe Speisung 200 500 mW Genauso Gewicht Geh use Gewicht Stahl rostfrei Option rechargeable batteries Temp range Operating 0 to 40 C Betriebstemperatur Temp Bereich Storage 20 to 70 C Lagertemperatur Remarks Bemerkungen F S Full Scale R O Read Out Remarks The instruments are delivered with batteries of AA type When using the type AF the installed spacer is to be removed Whe
32. ow 000000 As it is practically impossible for two ABSOLUTE RELZERO UNIT ZERO successive values to be identical while the instrument is Sums FLTER ADDRESS caus ONLINE handled the instrument may be positioned after the key is depressed As soon as the condition for a valid measurement is fulfilled the measuring value is displayed the direction indicator flashing To read the measurement value the CLINO 2000 m tm Imt E ar lw n hd may be removed from its located position The display value is lj 1 3 frozen Cl 7 DEG Use the key SEND ESC Ifa printer is connected the sensor UNT measured value will be transmitted and the CLINO 2000 will be mm ANDRESS CAIR ON NE ready for use in the normal mode If you just want to read the value pressing sEND ESC again will return to the standard mode If HOLD is required for the next measurement it is possible without canceling directly to recall HOLD again Remarks The level of vibration at the measuring location considerably influences the time necessary to collect a valid measurement Severe vibration may even completely prevent the condition two identical values in Succession necessary to register a true measurement In order to regain measuring capability under this condition take the instrument to an object with a lower vibration level and complete the operation in process By key usi
33. service purposes and should not be used The address of the internal sensor of the CLINO 2000 can t be changed The following error messages are possible ERROR 4 No sensor address found ERROR 5 More than one sensor found ERROR 6 Changing of sensor address was not successful Page 28 of 33 5 5 CALIBRATION 5 5 1 QUICK CALIBRATION The CLINO 2000 is equipped with an integrated calibration set up for a quick calibration procedure On the backside of the instrument a number of precisely manufactured and placed holes are available for installing the dowel pins as calibration aids These pins are part of the delivery and can be inserted into the holes With the quick calibration method the values at and 45 as well as the exact zero value can be fixed By this procedure the instrument can be set to a sufficiently high precision for most of the applications The calibration procedure is as follows 1 2 Start the instrument in the mode Absolute Select menu ZERO and confirm with MI ENTER Press shortly less than 5 sec Adjustment of angle at 45 Pre set value will be displayed The dowel pins are to be inserted in such a way that the instrument would display 45 when two pins are on a horizontal plane E g hold the instrument on the side of a measuring and setting plate the pins on top Press HOLD keep the instrument stable First displayed is 000000 then the measured value is seen
34. splay The maximum filter No 10 mathematically treats the signal by collecting data and sending somewhat delayed a signal integrated over a period of time Page 21 of 33 3 9 ABSOLUTE MEASUREMENT RELATIVE MEASUREMENT 3 9 1 ABSOLUTE MEASUREMENT The standard mode by default is the absolute measurement If this is not the case when starting the CLINO 2000 the A ON MODE key must be pressed repeatedly until the selection pointer is in position ABSOLUTE This must be confirmed with lll ENTER and the absolute mode is ready The displayed value is the Value of the CLINO 2000 minus ZERO OFFSET ea I UC 05 TE Parl Ganser DEG A oo A is Ot 1070205 3 9 2 RELATIVE MEASUREMENT Important remark The defined REL ZERO OFFSET is superimposed on the ZERO OFFSET stored in the CLINO 2000 The REL ZERO OFFSET stored temporarily in the CLINO 2000 can be recalled with any time when the same instrument is connected This value must be confirmed or changed for example to zero In case of differential measurement 1 Sensor or instrument connected to Port A an port B each the REL ZERO OFFSET will be stored in the CLINO 2000 Displayed value Value of the instrument minus ZERO OFFSET minus REL ZERO OFFSET Initial position The display shows the measured value on the CLINO 2000 0 10 00 This value is the true inclination of the instrument T
35. subunits Function 1 The connection port is displayed for information concerning the displayed measuring value active sensor instrument for calibration purposes Function 2 Shows the measuring mode e g individual measuring or differential measuring A B The active sensor address is displayed Page 10 of 33 3 HOW TO USE THE CLINO 2000 3 1 CHECKING THE FUNCTIONS The key A ON MODE can be pressed As long as this key is pressed all elements of the LCD display will show up After releasing the key the mode is switched to the measuring mode The active signal is displayed in the last used measuring unit During the reading of the calibration data six small will be displayed The following display variations are correct 1 034 20 834 degrees a OO 29 44023 U EN OVER Over Range ooo00d Levelmeter busy When now the connected instrument or sensor is tilted the displayed value must change accordingly If a second instrument or an additional sensor has been connected to port B a respective message may appear if some errors are noted The following error messages are possible ERROR 0 Instrument is defect to be sent to the service center ERROR 1 No sensor or instrument connected or not properly installed ERROR 2 No measurement has been done so far with the connected sensor instrument The sensor must be addressed correctly See point 3 7 Selecting a sensor or an instrument ERROR 3 Cal
36. then accept by pressing MB ENTER After this the CLINO 2000 is looking for all the connected sensors respectively their address The correct addresses of A and B must be confirmed with M ENTER On the display of the measuring instrument the measuring result value of measuring instrument minus value of reference instrument A B is now seen As long as the installed configuration remains the same only the A ON MODE key must be pressed on the two instruments for a renewed start up h PonSensor Page 26 of 33 5 20 mm m SENSOR ABSOLUTE RELZERO UMIT LIMITS FILTER ADDRESS ZERO ONLINE 4 ERROR MESSAGES ERROR 0 Instrument s fault needs to be sent to the service center ERROR 1 Sensor s Instrument s not connected or not correct address ERROR 2 No measuring has been made with the connected sensor The sensor must be addressed first See point 3 7 Selecting a sensor or an instrument ERROR 3 No calibration data available the respective sensor port is displayed ERROR 4 No sensor address found ERROR 5 More than one sensor found ERROR 6 Changing of sensor address was not successful ERROR 7 HOLD function has not been successfully completed within 60 seconds The procedure needs to be repeated ERROR 8 Battery not correctly installed ERROR 9 It was not possible to save the calibration data successfully in the sensor head Remarks After the error message is displayed and
37. till placed at the same angle the display now shows the value 0 00 00 Fig 1 SENSOR ABSOLUTE RELZERO UN mmm LIMITS FILTER ADDRESS CALIB 0 This new relative zero is the a 01000 reference for all the future measurements based on this angle Fig 1 originally 0 10 00 The selection pointer of the CLINO 2000 can now be placed below the marking ABSOLUTE by pressing the A ON MODE key several times to return to the absolute measuring mode After confirming with lll ENTER the original value is displayed Fig 2 DIQUU Port Sensor SENSOR ZERO UNIT LIMITS FILTER ADDRESS CALIB ONLINE amp 0 10 00 Fig 2 The values stored in the stacks of ZERO and Relative Zero can be deleted as follows Press the ON MODE key repeatedly until the selection pointer is located under REL ZERO or ZERO and confirm with ENTER In the display the existing value is seen By using the key 1 VK ZERO SELECT the value can be set to 0 This can now be confirmed with lll ENTER and the CLINO 2000 starts in the measuring mode The stack is empty Page 23 of 33 3 10 MEASURING BY SETTING LIMITS When your task requires the indication of a surpassed measuring value with an alarm the LIMITS function must be used In this mode it is possible to define an upper and a lower measuring limitation LIMITS After installation of the limits a signal will be seen on the screen and also sent through
38. tmost importance that the object to be measured and the CLINO 2000 are both at the same temperature level The CLINO 2000 should have been turned on a few minutes before the measurement procedure For this procedure a suitable surface must be chosen rigid flat and as horizontal as possible On this surface the CLINO 2000 must be placed the exact position of the instrument is to be marked on the surface Measurement A surface Turning the instrument atthe same spot by 180 ZERO OFFSET Measurement B EXAMPLE The selection pointer of the CLINO 2000 must now be placed below the marking ZERO by pressing the ON MODE key several times If the required position is reached it must be confirmed by pressing MB ENTER On the display the last set Zero Offset Basic offset will be displayed The direction indicator is flashing The CLINO 2000 must now sit at position one and after a few seconds settling time allowed by pressing the key HOLD the first reading is collected This first value is displayed and the selection pointer below ZERO is flashing which means the second reading is expected The CLINO 2000 must now be rotated horizontally by 180 and placed exactly on the same spot After a few seconds settling time the key HOLD must be pressed again for collecting the second reading The computed Zero Offset Basic offset is now displayed on the CLINO 2000 The displayed Zero Offset