         Contents
1.    12    VI  Care and Maintenance of Polisher       Straight Micro motor  Prophy Hand  nner Piece       Knurled Knob Locking Ring    1  The straight handpiece is pushed onto the rod of the micromotor  It will snap into place   2  The prophy angle is next  attach to the top of the straight handpiece   A  The locking ring on the straight handpiece needs to be in the open position  To do  this  push it down and to the right  clockwise    B  The prophy angle is placed on the straight handpiece with the cut out portion  at the  base  placed over the raised bubble on the straight handpiece   C  The locking ring is move left  counter clockwise  so that the dot and the screw on  the handpiece line up  If the locking ring 1s not engaged the prophy angle will not spin   D  To remove the prophy angle  simply reverse the above procedure     The prophy head needs lubrication and cleaning on a regular basis     1  Remove rubber cup   2  Remove cup holder with pliers by turning the knurled knob  The knurled knob is a  reverse thread  turn right not left  Wash out debris from inside  then put on drop of oil  into the gears   3 in I oil works well    3  Make sure all hair is removed from the end of the prophy angle    4  Replace knurled knob     13    To polish the teeth place a rubber cup on the prophy angle and place a little prophy paste in  the cup  Press the micromotor switch so that the LED is lit  Adjust the oscillating dial to a  low speed of about 2 000 rpm  Depress the foot pedal an
2.  90     B  Labels   The meaning of labels printed on the outside of the package box is listed below     FRAGILE    KEEP AWAY FROM WATER       DON   T HOOK    THIS SIDE UP       11 3  Safety Instructions    Grounding   Before any connection to the output connectors 1s made  the unit must be grounded  The  main plug shall be inserted only into a wall outlet with a protective ground     NOTE  The unit should be positioned so that the plug is accessible   Main voltage range and fuse     Before inserting the main plug into the wall outlet  make sure that the unit is compatible with  the voltage supply used     WARNING  The equipment must be disconnected from all voltage sources when  replacing the fuse    The main  line  fuse holder is located on the rear panel below input socket  When the mains   line  fuse needs replacing  proceed as follows   1  Disconnect the unit from the power cord   2  Remove the cover of the fuse holder by means of a small screwdriver   3  Replace a new fuse with the correct rating and put back  the fuse holder cover  The fuse will be 2A 250V delay action type     11 4  Setting Up    Unpack the JorVet Scaler Polisher Combo Unit and connect the components as shown in the  figure below        Figure 3  J841Q setting up diagram    1  Check the machine and make sure that both LED indicators and digital display are off when the  switch is off    2  Plug the unit into a grounded AC power outlet    3  Choose the correct INSERT for desired procedure  and insert it
3.  into the handpiece  Make sure the  insertion is complete    4  Make sure the water line is properly connected    5  Connect the micro motor to the power unit     Several installation suggestions are listed below     Because patients may experience some tissue trauma during the treatment  it is suggested  that the operator uses purified or distilled water  This will significantly reduce the  possibility of infection     After installation  the extra length of the power cord should be neatly arranged to avoid  any accidents     The footswitch should be firmly placed at position where the users doctors can easily  access it  Any extra cord to the footswitch extension should also be neatly arranged to  avoid any accidents     When pushing the insert into the handpiece  do not push the insert all the way into the  handpiece at once  First  push the insert into the handpiece half way  step on the  footswitch to turn on the scaler  and let some water come out from the handpiece  After  some water flow  push the insert all the way into the handpiece  The above procedure will  get rid of air inside the handpiece so that the handpiece will not generate too much heat  and possibly extending the lifespan of the handpiece     Caution  Make sure that new insert is fully seated   Caution  If the insert does not readily slide into the handpiece  lubricate the rubber O ring    with water and slide it with a twisting motion into handpiece  DO NOT FORCE  into place     III  Operation Procedure
4. User Manual    J841Q    Magnetostrictive Scaler   Polisher Combo Unit    Jorgensen Laboratories  Tel  800 525 5614 Fax  970 663 5042    Email  jortech  jorvet com    Warnings  Important Notes   The equipment is only to be used by a qualified veterinarian or other qualified personnel   e No one with a pacemaker should operate this unit     e A grounded AC power cord must be used with this equipment   J841Q unit should be powered from a separate wall outlet with a grounding point     Note before using      amp  The electric power used must be grounded  If this  requirement is not met  it could cause damage to  the unit and possibly to the user       The machine should be placed on a level and stable  platform or surface  Placing the machine on  unstable or tilted surfaces may degrade the  performance and or may accidentally cause damage  to the system      amp  Do not dismantle the machine on your own or by  uncertified technicians  Violation of the  requirement may cause harm to the user and or  damage to the machine  This will also void the  watranty      amp  For electrical safety  the power cord should not be  placed under heavy objects  and should also be kept  away from high temperature heat sources        If you observe any unusual situations when the  machine is in use  unplug the power cord for  precautionary measure                4 e    Contents    Contents EE 3    Descriptions and Functionalities of Component    Descriptions of ComonentS eee eee ccce cece cece ennee
5. and autoclave the insert at  270F for 12 minutes or as recommended by the manufacturer of your autoclave unit     Warning  High room temperature conditions  improper dilutions  or excessive immersion  time in a chemical sterilant can result in damage to the plastic and elastomeric  materials of the insert     Caution  The use of a dry heat oven  incompatible chemical vapor type sterilizers or  quaternary ammonium compounds must be avoided as damage can result to the  plastic and elastomeric materials     3  Main Unit    Since the Main Unit does not have direct contact with the patients  the cleaning 1s simple   Just carefully wipe the Main Unit with alcohol  and keep away from dust   If other  disinfectant is used  choose one that has no chemical effects on the surface of the plastic case  of the Main Unit  Please try it out first 1f not certain      4  Micromotor  Before cleaning  please remove the attached parts from the micromotor  The outer surface of  the micro motor should be cleaned with an antiseptic solution  Never oil the micromotor     Make sure no moisture gets inside the micromotor     5  Straight Handpiece    The straight handpiece should be lubricated at the end of each day of use  Drop 2 3 drops of  mineral oil into the top and allow it to flow down through the handpiece     6  Prophy Angle    The prophy angle should be cleaned of hair and debris after each patient  At the end of the day  it should be cleaned and lubricated as explained on the following page  
6. caler and inserts   5  Handpiece    The handpiece consists of two coils  one called the main coil and the other called the  feedback coil  and an insert tip     6  Micro Motor    This is a DC Micro motor  Micro motor speed will be controlled with the power dial   Please see    Operation Procedure        7  Main Power Switch    Indicates main power switch     Before turning the unit on  be sure the foot  switch is in the    OFF    position and there is no weight on the pedal     8  Forward   Reverse    This controls the micro motor revolving direction  Reverse is useful for removing tangled  hair from end of prophy angle     9  Insert    The insert tip can be divided into two parts  a nickel metal stack and stainless steel insert  head  To operate  the nickel stack needs to be  inserted  into the handpiece  These nickel  stacks then pick up force from the main coil of the handpiece via the electromagnetic effect   The stainless scaler head then moves back and forth at a high speed for tooth calculus  cleaning     O Description of Control Panel    The function of each button or knob on the control panel is detailed below     Ga Scaler Polisher  x100 RPM       Figure 2  Jorvet Front Panel    I  FUNCTION BUTTON    This is the scaling polishing mode change switch  Press the button to turn on the     Polishing    mode  and the indicator  LED   7  will light up  Press it again for    Scaling     mode and  LED   6  will light up     2  DIGITAL DISPLAY  Indicates the scaling output po
7. d begin polishing each tooth  surface  Add prophy paste as needed       Trouble shooting    NO        E ge     If no water 1s flowing from the handpiece  clean or change the water filter disk    The handpiece will overheat if there is not enough water flow  Increase water flow by  turning water up    An insert that is starting to wear out will also cause low water  Replace insert    If you have no power to the unit  change the fuse   If the micromotor handpiece is spinning  replace the prophy angle  the gears are worn   If the prophy angle knurled knob is not spinning  remove hair and debris  then lubricate   The straight handpiece may lose some power 1f not properly cleaned and lubricated     Note  Product requiring any maintenance should be returned to Jorgensen Laboratories   Call for proper procedure to return unit  JorVet will repair or replace any product  under warranty at no charge provided the repair meets the limited terms and  conditions of the warranty     VII  Product codes for accessories and replacement parts     Universal insert J452D1  Periodontal insert J452D1P  opt   Portable water tank  J452D4  opt   Rubber cups J452D6  opt   Diamond Burrs  7  J452D8M opt   Straight Handpiece J452D10  Inline Filter Screen J452D13FS  VIII  Warranty Terms   Main unit  PC Board  6 years  Handpiece I year  Micromotor I year  Straight handpiece I year  Prophy angle 90 days    Scaler inserts    90 days    Spatula Insert  Prophy Angle  Polishing paste    Diamond cutting disk    C
8. eeeeeeeanees 4  Descriptions of Control Panel nce eee ccc cece cece eee e eens 5  Ir te e ES TEE 7  Operation ee EEN H  Ee GC EE 10    Standard Accessories    Optional Accessories    Ce  anns nd SENNA ee 1   Care and Maintenance of Polisher oo    cece ccc cece eee e ec eeee 13  Product OG 14  WL      ver 14  INS DT  9  ENEE 15    I  Descriptions and Functions of Components    e Descriptions of Components       Figure 1  JorVet J841Q Outlook diagram    1  Main Unit    It is the power center of the unit  The main unit generates the needed 25KHz operation   signal and passes it onto the handpiece  and produces the power to vibrate the insert tip  inside the handpiece  Due to pressure variations  a signal from the handpiece will be fed  back to the main unit so that the main unit can track the variation of the loading  and adjust  the intensity of the controlling signal automatically     2 AC Power cord  3  Footswitch  The user friendly footswitch helps operator stop start the unit easily     4  Water Inline Connections    This unit comes with a female quick connect and an 8 foot water line  The dental scaler is  set up to easily connect to a portable water tank   J452D4   You can also order a saddle  valve with quick release  J452D14  to connect inline to existing water supply     NOTE   Inline Water Filter    This unit comes with an inline water filter  The filter disk should be cleaned and least once  weekly and replaced as necessary  This will extend the life of your dental s
9. he MAIN POWER SWITCH to turn on the digital display     NOTE  If no water comes out  please check the water supply   NOTE  If digital display does not appear  please contact Jorgensen Laboratories       Daily shut off    To turn off the unit at the end of the day    Press the POWER SWITCH to turn the unit off    Turn off the water supply shut off valve     The following suggestions are also useful in extending the product life of the unit     e Place the Main Unit in an open area where air can flow freely around it   e If you need to move the Main Unit  handle with care   e Make sure the AC power is turned off and the water source is turned off when not in use     e After six months or if you find the output power of the handpiece is not enough to perform  treatment  it is very possible that the insert is worn out  If the insert is worn out replace it with  a new one     IV  Accessories    Standard Accessories     1  Footswitch x1 J452D18  2  AC Power Line x1 J452DPC       10     3  Waterline Tube x1  4  MICRO MOTOR  x1 J453D 1        5  Insert 1x Universal Internal Flow Plastic Handle Insert J452D1  Ix Wide Spatula Internal Flow Plastic Handle Insert J452D2     6  In Line Water Filter J453D13    Note  This unit comes with a female quick connect to attach to the male  quick release on the portable water tank  J452D4      V  Cleaning and Sterilization    1  Handpiece    Before cleaning  remove the insert from the handpiece  Let the handpiece run for a couple of  seconds to dra
10. in out the water and any possible contamination left inside the handpiece  The  outer surface of the handpiece should be cleaned with an antiseptic solution  Rinse off with  water and wipe or spray with a chemical disinfectant that is compatible with the handpiece  material  A sterile insert is then re attached to the handpiece in preparation for the next  patient     Warning  Do not put the handpiece and the extension wire directly into the disinfectant   Fluid inside the machine will interfere with the normal operations of the system     At the end of the day with the insert removed  the handpiece and cable should be scrubbed  with an antiseptic solution and rinsed off with water  The handpiece should be scrubbed a  second time with an antiseptic soap or solution and rinsed off with water     Warning  The chemical disinfectant should not be allowed to remain on the surface longer  than the recommended time or material damage may result     Note  Cleaning of the handpiece is suggested after each patient use     11    2  Insert    After each usage  there will be saliva  blood and other debris left on the insert  consequently  it  is necessary to clean the insert with a cleaner first  This can be done manually by scrubbing  with a brush or by use of an ultrasonic cleaner with a solution of detergent and water  After  scrubbing  the insert  nozzle should be rinsed thoroughly with water to remove all detergent  and then dried  Then put the insert in a sterilization bag or wrap  
11. in the Dental Environment   Journal of the American Dental  Association  Vol  91  No  6  pp  1224 1229    15    
12. ontra Angle    Inline Water Filter    Micromotor    J452D2  J452D3  J452D5 opt   J452D7  opt   J452D9  opt   J452D13  J453D1    IX  Specifications  JorVet J841Q unit is designed and manufactured to meet the most demanding environment  Its  specifications are listed below        Specifications     Power supply 115V 45   50 60Hz 160VA    230V   5   50 60Hz 160VA     amp  Scaler Working frequency  24 5KHz  5   for 25KHz Insert      amp  Polisher speed 2000 30000 R P M      Dimension 6    L  x 8   W  x 3 5    H       Weight 6 5 Ibs   including handpiece     Handpiece Cable ee    amp  Footswitch Cable 83       Operation environment     Temperature 32 F 80 F     Relative Humidity 10    90   non condensing     Transport and storage conditions     Temperature Same as above     Relative Humidity 10    90   non condensing     Attention   Avoid use of this machine around pacemakers        It has been shown that electronic appliances including razors  hair dryers  microwave ovens  TV  receptors  and some electronic medical equipment may interfere with the normal operations of  pacemakers  It is suggested that patients who have pacemakers avoid treatment with this Unit   For further readings on the subject  please refer to         Advances in Cardiac Pacemaker   The New York Academy of Sciences  Vol  167  Article 2  pp   515 1075        Electromagnetic Radiation Interference with Cardiac Pacemaker   U  S  Department of Health   Education and Welfare        The Individual with a Pacemaker 
13. s     amp  When preparing to treat a patient     Ke    Push the POWER SWITCH to light the ON indicator  LED    Check the water source   Select the needed INSERT  Wet the O ring then place the INSERT into the handpiece with a    clockwise motion until it is fully tightened or seated    Set the POWER KNOB to the suitable level for the insert    Hold the HANDPIECE with the insert end pointing up over a suitable drain  Step on the foot  control and allow water to run from the handpiece for a few seconds until 1t flows without    Spurting     CAUTION  The above procedures should be repeated each time when insert is replaced    into the handpiece     CAUTION  Make sure that the water spray is at the desired temperature and is    reaching the working edge of the tip     Controlling the handpiece with the foot switch     The foot control is designed to control the functions of the handpiece in two ways    1  With the foot held down on the foot control  the handpiece is activated and water  flows from the handpiece     2  With the foot control released  both handpiece and water flow are shut off     e Water and Temperature control  The water knob controls the volume of water flowing from the handpiece by turning the  water knob clockwise or counter clockwise until the desired rate of flow 1s obtained     CAUTION  A continuous flow of water is required to keep the handpiece cool     Daily start up    To turn on the unit at the beginning of the day    Turn on the water supply    Press t
14. wer level or rotational speed of Micro motor  x100 R P M      3  POWER    Scaling mode  Control knob for adjusting output power level  The power level indicates the strength of  impact of the tip of the scaler  Turn the knob clockwise for maximum output power  Turn  knob counter clockwise to decrease the power intensity     Polishing mode  Control knob for adjusting rotational speed of micromotor  Turn knob clockwise for  maximum speed  Turn knob counter clockwise to decrease the speed   NOTE  Speed  x 100 RPM  is on digital display   4  WATER    Control knob for adjusting the water volume     Warning  Do not operate insert without any water     Warning  Do not adjust control knob in    counter clockwise direction over three  circles     Warning  Do not turn water pressure up too much at a time when adjusting water  knob     5  Micro motor Socket Seat   Connect Micro motor plug end into this socket   6  SCALING LED   Indicates Scaling mode   7  POLISHING LED    Indicates Polishing mode     II  Installation Guide    II 1  Unpacking  When unpacking the box  check this unit for any damage  If any damage is found  please contact    Jorgensen Laboratories immediately  Enter the unit serial number on your warranty card and mail it  within 10 days after setting up the machine     II 2  Storage  A  Environment   The unit should be stored in a clean  dry environment  The following environmental limitations  apply to both storage and shipping   Temperature  32   F to 80F  Humidity  10 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Mode d`emploi Eléments de commande Affichage  Manual  Abril 2008 Manual do Utilizador  Mode d? - Roland Central Europe  Front Grille Smart Fortwo 450 User Manual  Accéder à la documentation sur l`AFS  Sony DRX-710UL User's Manual  User Manual  US Robotics 125 Mbps Wireless MAXg PCI Adapter  Service manual - Clark Science Center    Copyright © All rights reserved. 