Home
SOLUTIONS
Contents
1. 12 5 Mechanical design Housing components Protective pane for SCD 21310 DP Ventilation openings Degree of protection Connector panel Weight in kg e SCD 21310 D without stand e SCD 21310 D with stand e SCD 21310 DP without stand e SCD 21310 DP with stand Dimensions W x H x D in mm e SCD 21310 D DP without stand e SCD 21310 D DP with stand 12 6 Climatic conditions Operation Temperature range Temperature gradient During transport and storage packaged Temperature range Temperature gradient SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 12 5 Mechanical design Metal sectional frame Protective glass with non reflective optical coating on both sides In rear panel IP20 in accordance with DIN 40050 On rear panel under cover 7 1 12 1 8 4 13 4 366 x 472 x 110 366 x 511 610 x 240 0 C 40 C ambient temperature Max 5 C h no condensation 20 C 60 C ambient temperature Max 5 C h no condensation 43 Technical specifications 12 7 Mechanical requirements 12 7 Operation Mechanical requirements Vibrations Shock Packaged unit 12 8 44 Vibrations Shock Falls Safety regulations Approvals Protection class Degree of protection Conformity To EN 60068 2 6 10 58 Hz at 0 075 mm oscillation 58 500 Hz at 10 m s To EN 60068 2 29 150 m s 6 ms To EN 60068 2 6 5 8 51 Hz at 3 5 mm oscillat
2. The SCD 21310 High Bright Color Display features a latest generation controller module and is supported by the calibration and quality control software SMfit ACT version 3 4 and higher SMfit ACT Basics The SMfit ACT Basics program is normally included in the software and documentation CD shipped with each display The software is also available as a free download on the Siemens Display Technologies website The SMfit ACT Basics software contains some SMfit ACT Calibration functions The software and documentation CD also contains SMfit ACT Basics documentation Additional information may be contained in a Readme txt file on the installation CD SMfit ACT Calibration and SMfit ACT Remote See also SCD 21310 SMfit ACT Calibration is a powerful calibration software tool that is available separately SMfit ACT Remote is compatible with both SMfit ACT Calibration and SMfit ACT Basics It enables remote access to the displays connected to the network and control of these displays via a central computer Additional information about this software package is available on the Siemens Display Technologies website Information on the display settings Page 28 Color coordinate setting Page 29 Internet http www siemens com displays Instruction Manual 01 2007 A5E00456472N 001 31 Software description 8 2 Display specific software support in SMfit ACT Basics 8 2 Display specific software suppo
3. 1 Front view SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 13 47 Dimensional drawings 13 1 Portrait 13 1 2 View from right with cover 48 SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 Dimensional drawings 13 1 3 View from right without cover o s220 q S SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 73 7 Portrait 49 Rear view Dimensional drawings 13 1 Portrait 13 1 4 o o 0000000000000000000000000000000000 2900000000000000000000000000000000 0000000000000000000000000000000000 0000000000000000000000000000000 000000000000000000000000000000 ll 0000000000000000000000000000000000 000000000000000000000000000000 0000000000000000000 000000000 000000000 000000000 000000000 000000000 000000000 000000000 000000000 000000000 000000000 000000000 000000000 000000000 000000000 000000000000000000000000 000000000000000000000000 000000000000000000000000 er SER EEE TEEN EIER EIER o o ES AA ES SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 50 Dimensional drawings 13 1 Portrait 13 1 5 View from above SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 51
4. 21310 Instruction Manual 01 2007 A5E00456472N 001 37 Service and maintenance 10 2 Maintenance 38 SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 Troubleshooting The LED continuously lights up green when operating normally In the event of a fault localization is possible based on the screen image and the operation LED e Please check the possible reasons and remedies listed in the table before contacting service personnel n More information SMfit ACTO provides additional information concerning the operating status of the display These options are described in detail in the SMfit ACT user manual Screen LED Cause Remedy No picture Dark Switch is off Switch on power switch Power cable is not plugged Check power cable in or incorrectly plugged in Power cable is defective Replace power cable Blown fuse Inform service personnel Green blinking No fault DPMS Display The operating system on the Power Management workstation switches the Signaling active display to power down mode to save energy and extend the life of the backlight unit No DVI signal DVI cable is not connected Incorrect timing is set Correct the timing Red blinking quickly Dangerous temperature Switch off display threshold exceeded Power Check the ventilation and consumption of the device improve if necessary will be reduced to a minimum to reduce the temperature Intern
5. 6 5 Description of connection procedure eee nana aaa anna nen 24 7 Commissioning viii At ii dr 2006 25 7 1 Use of the serial interface nene eee aaa nana mea aaa 25 7 2 Switching between portrait and landscape orientation mmmn nenea nana 25 7 3 Switch on the display and PC aena 0422 a da 26 7 4 Screen o aaa a a LANA d a da d aaa 27 7 5 Graphics controller settings ta aa a a tt ta ia d i i n i i a d i ai 27 7 6 Information on the display settings nenea nea enma nn anna 28 7 7 Check for pixel detects ia ener aie arena A ee 28 7 8 Golor coordi ate SetunG iss aaa MA c aa aaa al tte ae el eel oe aa aa Ia ea 29 8 Software descrita 31 8 1 Program for changing the display settings eee nea cnn ocn cnn nccnnnnnon 31 8 2 Display specific software support in SMfit ACT Basics nana 32 SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 Table of contents 10 11 12 13 Operation NANA NADAAN NANIRAHAN GNG NAAN AGO haa aaa BN dadu Balada du 33 9 1 Switch on the display and PG asniensis eiiean a E ana nn 33 9 2 Measures in the event Of an error ee aaa aaa aaa 33 Service and maintenance 3a aia ra dna d dana na oa iii pa aaa canin is ada a dept s a dan NG 35 10 1 Eleaninge aid aaa 35 10 2 Maintenance teria nio atada BI 36 TROUDIESHOOTING a an ES 39 Technical specifications cir A BAN MANINATA NABA a cae 41 12 1 Display features te ns len lenken a sa a iat 4
6. Dimensional drawings 13 2 Landscape 13 2 Landscape 13 2 1 Front view 52 SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 Dimensional drawings 13 2 Landscape 13 2 2 View from left 315 202 gt 278 SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 53 00000000000000000000090 0000000000000000000000 O 0000000000000000000000 0000000000000000000000 0000000000000000000000 0000000000000000000000 0000000000000000000000 0000000000000000000000 0000000000000000000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 0000000000000000000000 0000000000000000000000 0000000000000000000000 0000000000000000000000 0000000000000000000000 0000000000000000000000 0000000000000000000000 0000000000000000000000 o0o0o0000000000000000000 0000 0000 0000 0000 0000 0000 e O Rear view Dimensional drawings 13 2 Landscape 13 2 3 o View from above 13 2 4 SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 54 Appendix A 1 Guidance and manufacturer s declaration Electromagnetic emissions The SCD 21310 device is intended for use in the environment spec
7. SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 Safety notes See also Installation site Page 17 General connection information Page 21 Cleaning Page 35 SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 11 Safety notes SCD 21310 12 Instruction Manual 01 2007 A5E00456472N 001 Description 3 3 1 Scope of delivery Note Siemens recommends retaining the packaging material for subsequent transport of the display SCD 21310 D e Display with stand SCD 21310 DP 3 2 SCD 21310 e Display with protective glass pane and stand The scope of supply of both the SCD 21310 D and SCD 21310 DP includes the following components e CD with user manual and SMfit ACTO Basics e European power cable e US power cable e Dual link DVI video cable Applications The High Bright Color Flat Panel Display SCD 21310D Order No 6GF6200 4CA23 SCD 21310 DP Order No 6GF6200 4CA25 has been specially designed for medical diagnostics The display is available in a desktop version D desktop either with or without an optional protective glass pane P protective glass Instruction Manual 01 2007 A5E00456472N 001 13 Description 3 3 Important features 3 3 Important features The display has the following features which permit a wide range of applications Flawless picture reproduction with LCD technology Image geometry distortions do not occur with LCD technology The Flat Panel Disp
8. service personnel SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 Commissioning 7 4 Screen saver Switching displays on before the PC 1 Switch on all displays The operation LED blinks green 2 Switch on the PC The operation LED lights green if the connected signal can be shown by the display Switching PC on before displays 1 Switch on the PC 2 Switch on all displays The green operation LED is lit if the connected signal can be shown by the display See also Switching between portrait and landscape orientation Page 25 Troubleshooting Page 39 7 4 Screen saver e A screen saver function should be used in order to reduce image sticking that may occur with TFT displays Image sticking Image sticking is an effect in which a faint image of the previous screen contents can be seen after the display contents have changed By using a screen saver with continuously changing screen contents an unnecessary long depiction of a single image on the screen can be avoided 7 5 Graphics controller settings When the PC is switched on the display properties are read out and the graphics controller is initialized accordingly The graphics controller must support DDC communication After this process the display is recognized for example by Windows as plug and play device The EDID data Extended Display Identification Data are transmitted to the graphic card via the DVI cable DDC channel To oper
9. 1 12 2 POWEr SUpp Vcc e pie dd nee He dada ande 41 12 3 INpUtS OUtPUtS 02 2 cis a ae E AS a tee iat nein 42 12 4 Controls and Connectors oii AA a na NMA A cc a L 42 12 5 Mechanical design AN ta ALIGN NANAMAN PDI BATANG MANG 43 12 6 Climatic conditions na ANAND AA A ete ast 43 12 7 Mechanical requirements eee nea ana ana anna aaa aaa aa 44 12 8 Safety regulations ie a zei iai ce e a d a a a ta ae ta a cc i a a a te 44 12 9 Electromagnetic compatibility mm eee emana emana ana ea nenn 45 Dimensional drawings 2n a oa e era a da 47 13 1 Portrait a IE aaa o e a NAG 47 131 1 Front VIEW zar AA 47 13 1 2 View from right with cover eee nenea eee nenea aa nenea aa 48 13 1 3 View from right without cover aaa ana nana 49 13 14 Rear VIEW are NN 50 13 1 D Viewitfrom abovei i a ta aa til Da pd a 51 13 2 Landscape AK AN At d a at aaa La 52 13 21 RN Ah 52 132 2 Miewitromilellar r ra a aa a a a a Ad a a NA Al can Bl ld du a a Ra a Rail 53 132 9 A a TOTS 54 13 2 4 View from above oiei eiii itai tanane abaid RR RRA aaa nen anna aaa 54 APPONI a a At 55 A 1 Guidance and manufacturer s declaration Electromagnetic emissions mean ere 55 A 2 LOSS Of Wa NV A 08 A a aaa a 55 A 3 IVANA A AARON 56 A 4 BIT GA da 56 A 5 Environmental protection mmm eee nenea nana nana ae 56 A 6 Accessory AVISA a a a 56 A 7 ContactiaddresS id aa ee aa ae rar DAE aE 56 O 3 e Ses tac OA 57 SCD 21310 Instruction Manual 0
10. 1 2007 A5E00456472N 001 Introduction 1 1 1 Contents of this document This document describes the functionality and appropriate use of the SCD 21310 High Bright Color Flat Panel Display To ensure clarity it does not contain all detailed information on this product The contents of this document are neither part of a previous or existing agreement commitment or legal relationship nor does it modify such SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 7 Introduction 1 1 Contents of this document SCD 21310 8 Instruction Manual 01 2007 A5E00456472N 001 Safety notes 2 SCD 21310 To ensure safe and reliable operation the display must be properly transported stored mounted and installed and it must be operated and serviced with the appropriate care The unit must only be used for applications in which displays are normally used The technical specifications of the product must be observed In this document the term users refers to medical personnel e g radiology technician medical technician while the term servicing implies specifically trained and authorized personnel e g hospital technician medical equipment manufacturer For safety reasons the following precautions must be observed Danger Failure to observe the warning information in this document presents danger to life Severe personal injury or damage to property may occur Do not open the unit Certain components within the
11. CTO Version 4 0 and higher if the graphics card used does not support the DDC channel Service connection The service connection is used by service personnel for loading software updates See also Safety notes Page 9 6 3 Power connector Caution Only use the power cable supplied with the display or a cable with a PE conductor and an appliance socket plug in accordance with DIN 49 547 IEC 320 max length 3 m cable e g HO5VV F 3x1 0 mm Furthermore the cable must adhere to all local safety regulations applicable to the specific country in which the display is used Information for installations in the USA and Canada Molded power supply plugs must comply with the requirements for hospital grade attachments CSA Std C22 2 No 21 and UL 498 Power supply The display power supply is connected using an appliance plug cable e Check the permissible power supply range before connecting the display See also Power supply Page 41 6 4 DVI connector The display has a dual link DVI D connector Use the cable supplied or similar high quality cable The picture quality noise immunity and radiated interference of the entire system depend on cable quality and length SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 23 Connecting 6 5 Description of connection procedure 6 5 Description of connection procedure A Caution The angle position and height of the display can be changed on the preinstalled
12. Instruction Manual Version 01 2007 SIEMENS SIEMENS LC displays for medical diagnostics 3MP 21 High Bright Color Flat Panel Display SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 Introduction 1 Safety notes 2 Description 3 Application planning 4 Assembly 5 Connecting 6 Commissioning 7 Software description 8 Operation 9 Service and maintenance 1 0 Troubleshooting 1 1 Technical specifications 1 2 Dimensional drawings 1 3 A Appendix Safety Guidelines This manual contains notices you have to observe in order to ensure your personal safety as well as to prevent damage to property The notices referring to your personal safety are highlighted in the manual by a safety alert symbol notices referring only to property damage have no safety alert symbol These notices shown below are graded according to the degree of danger Danger indicates that death or severe personal injury will result if proper precautions are not taken Warning indicates that death or severe personal injury may result if proper precautions are not taken Caution BEB with a safety alert symbol indicates that minor personal injury can result if proper precautions are not taken Caution without a safety alert symbol indicates that property damage can result if proper precautions are not taken Notice indicates that an unintended result or situation
13. SAB 2110 RC stand During handling the display can tip downwards potentially pinching body parts between the housing and stand The stability of the SCD 21310 display with the supplied SAB 2110 RC stand has been tested If a different stand or wall bracket is used please ensure that the display is secure gt More information Please note the instructions in the user manual Multifunctional base for LC displays SAB 2107 RC SAB 2110 RC Place the display on a stable horizontal surface If fastened loosen the combo torx slotted screw from the connector panel Remove the cover from the connector panel Connect the appliance cable plug to the display s power connector Connect the DVI plug to DVI socket and fasten the connector screws If needed connect USB cable to the display s USB port N Oa oO FP QO N Aaa Feed the appliance cable the DVI cable and if needed the USB cable through one of the three cable openings in the cover Please route the cables with some play through the connector panel to ensure that the plug connections are not detached Select one of the three cable openings in the cover that best suits the orientation of the display Please ensure sufficient play is allocated for the cables 8 Replace the cover on the connector panel 9 If desired the connector panel cover can be fastened using the combo torx slotted screw See also Scope of delivery Page 13 Switching between portrait a
14. al fault Inform service personnel Green Video signal detected but Check settings with the display is setup SMfit ACT e g LUT incorrectly brightness no test image etc Video signal detected but Inform service personnel device defective SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 39 Troubleshooting Screen Picture displayed LED Green Cause No error correct operating status Remedy Green blinking No fault but the refresh rate is not setup optimally for the display Set image refresh rate for the graphics card to 30 Hz if the rate is not optimized flickering effects may be noted in moving images Check the resolution Red blinking slowly Critical temperature threshold exceeded The target value for the backlight unit control system will be halved to significantly reduce the luminance level to help lower the temperature avoid potential damage Switch off display Check the ventilation and improve if necessary Select a lower luminance level for standard operation See also 40 Switch on the display and PC Page 26 SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 Technical specifications 12 1 12 2 SCD 21310 Display features Type Screen size Screen diagonal Resolution Refresh rate Pixel organization Pixel pitch Contrast ratio Horizontal viewing angle Vertical viewing angle Protective
15. ate the display at the desired resolution a driver designed for the graphics controller in use must be installed Once the driver has been successfully installed the resolution can be configured as dictated by the driver and the operating system settings Using these settings the display can for example be changed from landscape to portrait format Please refer to the graphics controller manufacturer s manual for detailed information about installation and configuration of the graphics controller SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 27 Commissioning 7 6 Information on the display settings 7 6 See also 7 7 28 Information on the display settings All displays are optimally preset in the factory such that adjustments are not usually required nevertheless the programs SMfit ACTO Basics and SMfit ACTO Calibration Version 3 4 or higher allow one to modify the settings if so desired Please note the operating system manufacturer s instructions related to power management settings The display supports the so called DPMS mode Display Power Management Signaling which can be used for energy saving For example the display can be switched off automatically after 20 minutes using DPMS Program for changing the display settings Page 31 Check for pixel defects Pixel defects small bright or dark dots can occur in LCD monitors During the manufacturing process all displays are checked for the permitt
16. can occur if the corresponding information is not taken into account If more than one degree of danger is present the warning notice representing the highest degree of danger will be used A notice warning of injury to persons with a safety alert symbol may also include a warning relating to property damage Qualified Personnel The device system may only be set up and used in conjunction with this documentation Commissioning and operation of a device system may only be performed by qualified personnel Within the context of the safety notes in this documentation qualified persons are defined as persons who are authorized to commission ground and label devices systems and circuits in accordance with established safety practices and standards Prescribed Usage Note the following A Warning This device may only be used for the applications described in the catalog or the technical description and only in connection with devices or components from other manufacturers which have been approved or recommended by Siemens Correct reliable operation of the product requires proper transport storage positioning and assembly as well as careful operation and maintenance Trademarks All names identified by O are registered trademarks of the Siemens AG The remaining trademarks in this publication may be trademarks whose use by third parties for their own purposes could violate the rights of the owner Disclaimer of Liability We have
17. e application The display features can be adjusted with the SMfit ACTO Basics and SMfit ACTO Calibration software tools to ensure optimal image quality The display can be adapted to the local lighting conditions at the click of a mouse using the five preset practice oriented look up tables or it can be automatically calibrated to the desired luminance levels using the SMfit ACT calculate LUT function In this way the color coordinates of the display can be easily adapted to the user s preferences and to ensure optimal matching with neighboring displays Dual link DVI input 14 The display is connected to the workstation via dual link DVI SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 Description 3 3 Important features Innovative mechanical design The mechanical properties of the display support quick and easy installation The easily removable cover provides convenient access to the spacious connector panel where the connections are clearly arranged Three openings in the cover provide practical cable routing when changing the orientation of the display or mounting on the wall or ceiling Protective glass pane The SCD 21310 DP has a protective pane with a non reflective optical coating in front of the LCD panel to protect the panel surface from knocks and scratches The space between the protective pane and the panel is sealed to prevent dust from entering thus helping ensure the internal surfaces remain cl
18. ean Operation in portrait or landscape mode The device can be operated in portrait or landscape mode The orientation is automatically detected by the integrated position sensor When the connected workstation is booted the current orientation is recognized and the display is automatically supplied with the approriate video timing A rotation function in the graphics card is not necessary SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 15 Description 3 3 Important features SCD 21310 16 Instruction Manual 01 2007 A5E00456472N 001 Application planning 4 4 1 Installation site Provide adequate ventilation Ventilation holes are located on the rear of the housing Observe the permissible ambient temperature range The permissible ambient temperature range must not be violated Avoid reflections on the screen The display has an anti glare surface which is only effective if the screen is clean and grease free The SCD 21310 DP display is also equipped with a protective glass pane with a non reflective optical coating on both sides and is particularly suitable for use in sterile environments when cleaning agents are used Please note the cleaning instructions Page 35 The display should be positioned such that reflections of lights windows furniture with shiny surfaces or light colored walls do not appear on the screen In order to reduce mirroring on the display only use anti glare lights non da
19. ed number of defective pixels During setup the display can be checked as follows using SMfit ACTO Calibration and SMfit ACTE Basics e Generate a black test screen and examine the screen for luminous pixel errors e Generate a white test screen and examine the screen for non luminous pixel errors Pixel defects on an LCD monitor can not be corrected e Make a note of the position of the defective pixel SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 Commissioning 7 8 Color coordinate setting 7 8 Color coordinate setting CE More information For further details please refer to the SMfit ACTO documentation The color coordinates of the SCD 21310 display are measured during the factory calibration procedure and stored in the display as reference values Users can adjust the color coordinates to their application requirements by selecting one of the various internal LUTs Users can also tailor the display to specified color coordinate values with the SMfit ACTO function Calculate Calibration Data Note Adjusting the color coordinates will generally reduce the maximum luminance levels See also Program for changing the display settings Page 31 SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 29 Commissioning 7 8 Color coordinate setting 30 SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 Software description 8 8 1 Program for changing the display settings Note
20. g the display it is essential that the screws used adhere to the following specifications Fastening screw specifications Number 4 Thread M4 Strength 8 8 in accordance with ISO 898 1 Insertion depth 6 mm minimum 8 mm maximum Torque 3 Nm Switching between portrait and landscape orientation Page 25 Assembly 5 1 Mounting the display SCD 21310 20 Instruction Manual 01 2007 A5E00456472N 001 Connecting 6 6 1 General connection information A Caution All information and warnings related to this product must be observed to ensure danger free operation Caution Observe shielding measures The shielding measures described in the applicable national EMC guideline must be observed If these guidelines are not observed malfunctioning of the display may result Information on cable installation e Only shielded cables are permitted for all signal connections e All plug connections must be screwed tight or locked if the corresponding facility is present on the connector e Signal and power cables must not be routed through the same duct e The display must not share a power supply with motors or valves interference e Externally connected cables can represent a trip hazard Please ensure that all incoming cables are safely routed See also Electromagnetic compatibility Page 45 Safety notes Page 9 Guidance and manufacturer s declaration Electromagnetic emissions Page 55 SCD 21310 Instruc
21. ified below The customer or user of the SCD 21310 should make sure that it is used in such an environment Interference emission tests Compliance Electromagnetic environment guide RF emissions Group 1 The SCD 21310 uses to CISPR 11 RF energy only for its internal functions Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions Class B The SCD 21310 is suitable for to CISPR 11 use in all types of location Harmonic emissions Class A including domestic buildings and to IEC 61000 3 2 those directly connected to the A public power supply network Voltage fluctuations flicker Complies that supplies buildings used for emissions domestic purposes to IEC 61000 3 3 See also Safety notes Page 9 General connection information Page 21 Electromagnetic compatibility Page 45 A 2 Loss of warranty All unauthorized electrical or mechanical alterations on or in the unit will render the warranty null and void SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 55 Appendix A 3 Warranty A 3 Warranty All commitments on the part of Siemens AG are contained in the respective sales contract which also contains the complete and solely applicable warranty conditions These warranty conditions in the contract are neither extended nor limited by the contents of this publication A 4 Repairs Please contact the distrib
22. intenance The illumination properties slowly may change due to aging of the LCD unit and the backlight For this reason routinely check and if necessary correct the display settings to ensure compliance with local guidelines Siemens also recommends checking the displays for pixel errors A Serial Spot Meter Universal Serial Luminance Meter or an Advanced Serial Luminance Meter can be used for taking measurements These units can be connected directly to the display e One can automatically perform all calibration and quality tests using the SMfit ACTO Calibration program e One can also manually adjust display settings using a Serial Spot Meter and the SMfit ACTO Basics program Notice Confirm image quality visually after calibrating the display After calibration the display must be visually inspected to verify successful and correct completion of the calibration procedure This inspection can be performed by reviewing the SMPTE image The gray levels should be displayed correctly and visibly at both 5 and 95 Alternatively the VerLum image can serve as a useful test pattern Successful calibration can be verified if the small squares are displayed correctly in all gray shades SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 Service and maintenance 10 2 Maintenance See also Check for pixel defects Page 28 Program for changing the display settings Page 31 Use of the serial interface Page 25 SCD
23. ion 8 51 500 Hz at 10 m s To EN 60068 2 29 250 m s 6 ms in storage packaging To EN 24180 2 60 cm height EN 60601 1 CAN CSA C 22 2 No 601 1 UL 60601 1 Protection class IP20 CE to MPG 93 42 EEC Class SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 Technical specifications 12 9 See also SCD 21310 Electromagnetic compatibility Interference voltage emitted noise Voltage fluctuations Burst on power cables Burst on signal lines Surge on power cables Electrostatic discharge on casing parts ESD RF irradiation HF on network and signal line Low frequency and static magnetic fields Line reaction to harmonics Line reaction to voltage fluctuations 12 9 Electromagnetic compatibility EN 60601 1 2 Class B FCC Class B EN 61000 4 11 EN 61000 4 4 2 kv EN 61000 4 4 1 kv EN 61000 4 5 1 kV symmetric 2 kV assymmetric EN 61000 4 2 8 kV air 6 kV contact EN 61000 4 3 80 MHz 2500 MHz 3 V m 80 AM 1 kHz EN 61000 4 6 EN 61000 4 8 Magnetic DC fields 4000 A m Magnetic alternating fields 1000 Aeff m EN 61000 3 2 EN 61000 3 3 Guidance and manufacturer s declaration Electromagnetic emissions Page 55 Instruction Manual 01 2007 A5E00456472N 001 45 Technical specifications 12 9 Electromagnetic compatibility 46 SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 Dimensional drawings All dimensions in mm 13 1 Portrait 13 1
24. lay provides a flicker free picture even at low refresh rates The display thus fulfills even the strictest ergonomic requirements The SCD 21310 High Bright Color Flat Panel Display is equipped with an active 21 TFT display module with IPS In Plane Switching technology The module offers maximum contrast and color stability across an extremely wide viewing angle The screen resolution has been optimized to display 1536 x 2048 pixels Fully Automated Stability Compact design The Flat Panel Display is equipped with an automated stability system Fully Automated Stability in accordance with the DICOM standard The integrated stability system ISS ensures consistent luminance levels using an internal light sensor located at the center of the backlight unit Using SMfit ACTO and one of Siemens external photometer options the ISS sensor is automatically adjusted to ensure superior traceable calibration The display is precalibrated in the factory using extremely sensitive instruments and the calibration data is stored in the five internal look up tables The saved calibration data simplify installation and servicing Even if the graphics card or workstation is replaced e g during upgrade the graphics settings are retained and there is no need to recalibrate the monitor The low weight small size minimal power consumption and low heat generation of the display provide advantages over conventional CRT monitors Settings for optimizing th
25. nd landscape orientation Page 25 SCD 21310 24 Instruction Manual 01 2007 A5E00456472N 001 Commissioning 7 1 Use of the serial interface The display has a serial RS 232 interface with a 6 pin mini DIN socket The following units may be connected to the socket e Ambient light sensor for the SCD display series permissible in the patient environment e Universal Serial Luminance Meter with a 9 pin Sub D to 6 pin mini DIN adapter e Advanced Serial Luminance Meter with 9 pin Sub D adapter to 6 pin mini DIN e Serial Spot Meter with 9 pin Sub D to 6 pin mini DIN adapter This service mode is not permissible in the presence of patients 7 2 Switching between portrait and landscape orientation Note Use of a graphics card picture rotation function is not recommended Many graphics cards support image rotation functions The use of such functions is not recommended If such a function is used the setting must be reset upon rebooting the workstation Adjusting orientation prior to rebooting the workstation The SCD 21310 display is shipped in portrait orientation but it can also be operated in landscape orientation The display has an internal position sensor that automatically recognizes it s orientation when the workstation boots up The orientation of the display should be set prior to switching the display on This ensures that the EDID data valid for the set display orientation is available to the graphics ca
26. pane for SCD 21310 DP Backlighting Screen brightness Lifetime of backlight Power supply Power connector Line voltage Line frequency Power consumption Instruction Manual 01 2007 A5E00456472N 001 TFT dual domain IPS amorphous silicon 318 mm x 424 mm 20 8 52 8 cm 1536 x 2048 max pixels in portrait orientation 60 Hz 24 bit 8 bit RGB 3 subpixels per pixel 0 207 mm x 0 207 mm 400 1 typically Typically 170 at CR 10 1 Typically 170 at CR 10 1 Protective glass with non reflective optical coating on both sides CCFL Typically 400 cd m 30 000 hours at 25 C operating temperature Appliance plug socket 100 V 240 V AC 10 50 60 Hz 5 105 W maximum 41 Technical specifications 12 3 Inputs outputs 12 3 Inputs outputs DVI input RS 232 USB RJ 45 12 4 Controls and connectors Rear side 42 Dual link DVI I socket Service and communication with SMfit ACT over DDC channel Mini DIN photometer socket Service and optional communication with SMfit ACT Service or software update e 1x power switch e 1x operation LED e 1x serial interface 6 pin mini DIN socket Beneath protective cover e 1x power connection socket e 1x DVI e 1x RJ45 e 1x USB optional Note Use the SMfit ACTO Basics or SMfit ACTO Calibration software to adjust the display settings SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 Technical specifications
27. patibility Standards guidelines 45 F Fully Automated Stability Fully Automated Stability 14 Image sticking 27 IPS In Plane Switching 14 ISS Integrated stability system 14 M Manually change display settings With Serial Spot Meter and SMfit ACTO Basics 36 Measuring instruments for maintenance 36 N Non heating appliance connector 23 24 Non heating appliance connector Connecting to power input of display 24 Note for installations in the USA and Canada 23 on cable installation 21 O Observe EMC requirements when connecting 21 Order No SCD 21310D 13 57 Index SCD 21310 DP 13 V Ventilation 17 P VESA adapter 19 Packaging material 13 photometers Approved 22 Pixel defects Correcting defective pixels 28 Power cable 9 13 23 Power connector 22 23 41 Power supply 23 Product disposal 56 S Serial interface Connect devices 22 25 Serial Spot Meter Connection to serial interface 22 25 Measuring instrument 36 Service Authorized personnel 9 Service connection 23 Servicing socket See also serial interface 25 Setting the display using SMfit ACT Basics software or SMfit ACT Calibration software 42 settings Application optimizing 14 Shielding measures 21 SMfit ACT Basics Information on 31 SMfit ACT Calibration and SMfit ACT Remote Information on 31 specifications Fastening screw 19 Start up sequence of display and PC 26 U Universal Serial L
28. r cloth and if necessary a cleaning agent Only clean housing parts using the recommended cleaning agents e The entire display must only be disinfected using the tested disinfecting agents Recommended cleaning and disinfection agents Diverseylever Labs Morning Mist Surfanios Fraicheur Citron Anios Labs Agent class Tested cleaning and disinfection Further examples agents Alcohol Ethyl alcohol 96 Hospiset cloth Mikrozid liquid Aldehyde Melsitt Aldasan 2000 Cidex Kohsolin Gigasept FF Chlorine derivatives Terralin Quartamon Med Disinfecting agents TaskiDS5001 Guanidine derivatives Lysoformin Quaternary compounds Incidur spray undiluted SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 35 Service and maintenance 10 2 Maintenance Agent class Tested cleaning and disinfection Further examples agents Standard household washing up Tempo Fairy Ultra Pril Palmolive liquid Petroleum spirit Petroleum ether Pyridine derivatives Activ spray undiluted Water Tap water Distilled water Unauthorized cleaning and disinfection agents 10 2 Agent class Tested cleaning and disinfection agents Peroxide compounds Perform Dismozon pur These cleaning and disinfection agents can bleach the paint after a longer period of application Maintenance Routinely check the display settings 36 The display requires little ma
29. rd SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 25 Commissioning 7 3 Switch on the display and PC Adjusting orientation after rebooting the workstation See also If the orientation of the display is changed during normal operation it may be necessary to restart the workstation to ensure that the EDID data are reread Please observe the instructions for the workstation supplier and or operating system More information Please also note the instructions in the user manual Multifunctional base for LC displays SAB 2107 RC SAB 2110 RC regarding switchover of the orientation of the display Mounting the display Page 19 Description of connection procedure Page 24 7 3 Switch on the display and PC 26 Notice Only switch on the PC after the orientation of the display has been set Please note when changing the orientation of the display between portrait and landscape the PC should only be switched on after the adjustment has been made Page 25 The display and PC can be switched on in any order requirement The PC and display are connected through a DVI cable and the graphics card supports the DDC channel Notice The operation LED does not light green after power has been switched on If the operation LED does not light up green after the equipment has been switched on and a video signal has been applied please check the system for basic connection and operating errors before requesting
30. reviewed the contents of this publication to ensure consistency with the hardware and software described Since variance cannot be precluded entirely we cannot guarantee full consistency However the information in this publication is reviewed regularly and any necessary corrections are included in subsequent editions Siemens AG Order No A5E00456472N 001 Copyright Siemens AG 2007 Automation and Drives 01 2007 Technical data subject to change Postfach 48 48 90437 NURNBERG GERMANY Table of contents 1 Introduction 2 aaa aa cara 7 1 1 Contents of this document eee cnn cnn anne cnn aaa nana aaa aaa 7 Safety Molestar A dad 9 Descripcion AER LARREA oi tdi ce outer E DA oia tau 13 3 1 Sl sesizeaza ta ate re ee at EDER dent 13 3 2 Applications rica a a aa ca iaca a a a ta a AABANG 13 3 3 Important feature 8 za p rea A n Dau NMA NANA AN 14 4 Application planning nain KAAGAD ti 17 4 1 Installation Sitten nA na KABA Rn e AE Kana Pa at 17 4 2 Distances and arrangement of units 00 0 ee eee nenea nana aaa aaa ana 18 5 ASOMAN hae A A Baie ee 19 5 1 Mounting tie display pice iza ci 2 HB a a a cu 19 6 Connecting na ri nn clin oie elven dete event cues teen a 21 6 1 General connection information eee nana ea eee 21 6 2 Connector location iasa taca i ao aduca at aa aa anni da Data ao a du del aaa 22 6 3 Power connector asliii noa see NANANG at ce Da AD dbl a ada de a de d d a 23 6 4 DVI connector akan ce ata i o ii AI ease AG 23
31. rt in SMfit ACTO Basics The SMfit ACTO Basics program is described in the corresponding documentation normally included on the software and documentation CD The SCD 21310 display is supported by SMfit ACTO Basics version 3 4 and higher SCD 21310 32 Instruction Manual 01 2007 A5E00456472N 001 Operation 9 9 1 Switch on the display and PC Once commissioned operation of the display is limited to switching it on and off After switching on the display the operation LED is lit green continuously If the LED lights up with another color the device is not operating within normal parameters See also Troubleshooting Page 39 9 2 Measures in the event of an error Should the display not function correctly Siemens recommends checking all basic connections and operation errors before contacting service personnel See also Troubleshooting Page 39 SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 33 Operation 9 2 Measures in the event of an error 34 SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 Service and maintenance 10 1 Cleaning Caution Device maintenance cleaning and disinfecting e The front panel is extremely sensitive to mechanical damage Avoid all scratches knocks etc e Remove moisture immediately contact with water over a longer period can cause discoloration or allow calcium deposits to form on the surface e When dirty clean the front panel using a soft dry microfibe
32. tion Manual 01 2007 A5E00456472N 001 21 Connecting 6 2 Connector location 6 2 Connector location Power switch Serial interface for photometer Operation LED Service connection DVI connector Power connector USB connector Figure 6 1 Connector panel Power connector and DVI connector The power and DVI connectors are located within the connector panel under the cover on the rear of the display A Caution The connector panel cover must only be opened for servicing Patients must not be present when the cover is open Serial interface for photometer Caution Only photometers tested for calibrating the equipment may be connected to the serial interface Only service personnel or those trained by them may connect or disconnect a device from the unit A photometer must not be connected in the presence of the patient The serial interface is freely accessible next to the cover of the connector panel This 6 pin mini DIN socket is used to connect an ambient light sensor or a photometer The Serial Spot Meter the Universal Serial Luminance Meter and the Advanced Serial Luminance Meter have been approved as photometers SCD 21310 22 Instruction Manual 01 2007 A5E00456472N 001 Connecting 6 3 Power connector USB connection The USB port supports communication between the display and a connected PC This connection can be used for example to communicate with SMfit A
33. uminance Meter Connection to serial interface 22 25 Measuring instrument 36 USB connection 23 Users Health personnel 9 SCD 21310 58 Instruction Manual 01 2007 A5E00456472N 001 Siemens AG Automation and Drives Systems Engineering Display Technologies Siemensallee 84 76187 KARLSRUHE DEUTSCHLAND Order No A5E00456472N 001 www siemens com displays
34. unit operate at high voltages i e touching them presents a danger to life Never use defective power cables A damaged power cable may result in fire or electric shock Always hold the plug when pulling to disconnect the power cable Do not insert any objects into the housing Objects inserted into the housing may result in damage to the unit or personal injury Do not place any objects on top of the units Liquid entering the unit may result in fire or electric shock Do not connect the display in the presence of patients Avoid all possible contact with patients when handling the connection cables Correct calibration reduces the risk of misdiagnosis When used in medical environments for diagnostic purposes improper calibration of the display can result in incorrect diagnoses Instruction Manual 01 2007 A5E00456472N 001 Safety notes 10 Caution Improper installation may result in extensive damage to property Installation must be performed properly 1 To prevent injury to patients and users of a medical electrical system utilizing Siemens products in a patient environment please observe the safety requirements of IEC 60 601 1 1 and its national equivalent for Specifications for the Safety of Medical Electrical Systems when assembling the system Use appropriate measures to ensure that leakage currents in particular remain below the necessary limits Appropriate measures include Use of a separator de
35. utor from whom the product was originally purchased A 5 Environmental protection Please observe all local requirements and laws pertaining to the disposal of displays A 6 Accessory devices Devices connected to the display e g PC must also comply with the relevant safety specifications A 7 Contact address E mail dt aud siemens com See also Internet http www siemens com displays SCD 21310 56 Instruction Manual 01 2007 A5E00456472N 001 Index A Ambient temperature Operation 43 Transport and storage 43 Ambient temperature range 17 C Carry out calibration and quality tests automatically With SMfit ACT Calibration 36 Ceiling mounting 19 Change of environment 17 Changes to the settings via SMfit ACTO Basics and SMfit ACTO Calibration 28 Connect ambient light sensor 22 25 Contact partners 56 D DDC channel 27 Desktop Desktop version 13 Display Avoid reflections on the screen 17 Cleaning and disinfection agents 35 Dimensions 43 Minimum clearance to other devices 18 Performance features 14 Weight 43 Display Power Management Signaling See also DPMS 28 DP Desktop version with protective glass pane 13 DPMS Display Power Management Signaling 28 DVI cable 26 27 39 DVI connector 22 23 DVI plug Connect to DVI socket 24 DVI video cable 13 SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 E EDID Extended Display Identification Data 27 Electromagnetic com
36. vice for the input or output signal component Use of a safety isolating transformer Use of the additional protective conductor terminal 3 During installation only use cables recommended by the manufacturer 4 Use power cables with protective ground contact Only plug the device into sockets with protective grounding 5 For certain applications the video ground can be connected to protective ground via the additional protective conductor terminal in the plug panel IEC 60 601 1 1 must be observed 6 Close the plug panel using the provided cover and lock it with the screw Note to users The closed connector panel must not be opened by the user Servicing information If housing components have to be removed for servicing this must not be carried out in the presence of patients users or other persons not involved with servicing 7 The following applies to installations in the USA and Canada Molded power supply plugs must comply with the requirements for hospital grade attachments CSA Std C22 2 No 21 and UL 498 Caution Failure to observe warnings may result in substantial damage to property Do not subject device to excessive shocks Take care when transporting Use the original packaging The panel in particular should be protected against shock Notice If the panel surface is touched the physical contact or an electrostatic discharge may cause a brief disturbance in the picture quality
37. zzling for the ceiling lighting Change of environment See also SCD 21310 If brought from the cold into a warm environment water may condense inside the unit The unit should not be switched on until all condensed water has evaporated including that within the unit This may take several hours depending on the conditions Cleaning Page 35 Climatic conditions Page 43 Instruction Manual 01 2007 A5E00456472N 001 17 Application planning 4 2 Distances and arrangement of units 4 2 Distances and arrangement of units The unit should be positioned on a hard level and horizontal surface It should be positioned at least 10 cm from the side and rear of a wall or 15 cm from other equipment The display can be mounted flushly with horizontally and vertically adjacent displays SCD 21310 18 Instruction Manual 01 2007 A5E00456472N 001 Assembly 5 1 Mounting the display CE See also SCD 21310 Instruction Manual 01 2007 A5E00456472N 001 19 More information Please note the instructions in the user manual Multifunctional base for LC displays SAB 2107 RC SAB 2110 RC The SCD 21310 display is supplied in portrait orientation on an SAB 2110 RC stand For ceiling mounting VESA adapter 100 x 100 mm or installation with the stand proceed as follows e Affix the display on the VESA adapter or stand using the appropriate screws to secure it Because the screws are extremely important for securin
Download Pdf Manuals
Related Search
SOLUTIONS solutions solutions synonym solutions reach solutions architect solutions navigator solutionreach solutions by text solutions peptides solutions pest and lawn solutions health my chart solutions engineer solutions center solutions to climate change solutions bank solutions 2 go solutions for change solutions definition solutions for patient safety solutions pet products solutions north bank login solutions architect salary solutions pest and lawn website solutions north bank solutions bank login solutions staffing
Related Contents
MidiFly Manual - Parker Guitars Monarch Specialties I 3019 Instructions / Assembly 品番 ACW・62。 Operators Manual - Wolf Laboratories Ergotron LX HD Wall Mount Pivot ユニバーサルデザインの まちづくりの考え方 第 章 Propex 1800 - Club 80-90 Bedienungsanleitung User`s Manual Software de Administración de Seguros y Fianzas Copyright © All rights reserved.