Home
Corning LSE 6L Digital Water Bath
Contents
1. at set temperature by holding UP or DOWN key for more than 2 seconds 12 0 TROUBLESHOOTING Review the information in the table below to troubleshoot operating problems Problem on Display Cause Er1 Temperature probe is not in range Er2 Temperature is more than 5 C higher than the set temperature Er3 If after two hours temperature does not reach SET temperature 13 0 CLEANING AND MAINTENANCE Solution Examine temperature probe Faulty control board Examine SET temperature Examine parameters for heating Examine parameters for heater Faulty control board Examine temperature probe Examine heater Examine temperature safety fuse Faulty control board Prior to any cleaning or maintenance of the unit disconnect the power cord from the wall outlet 1 No routine maintenance is required on the unit 2 Clean the unit internally and externally with a damp cloth and mild soap Do not immerse the unit or spill liquids into or on the unit Allow the water bath to dry completely inside and out before reconnecting the power and using 14 0 TEMPERATURE RE ADJUSTMENT Water bath has a software for temperature re adjustment You must first measure temperature in the middle of the chamber or in the sample for a minimum of 2 hours which Is required to stabilize the temperature You must measure temperature with a certified reference temperature measuring instrument with precision of 0 1 C or more After 2 hours or more
2. read temperature on thermometer and compare it with temperature on LED display Difference between thermometer reading and display is the value which you will enter in the software Example 1 Reference Temperature is 37 9 C temperature on display is 37 C Difference is 37 9 37 0 9 Value is 0 9 value which you enter Example 2 Reference temperature is 36 2 C temperature on display is 37 C Difference is 37 36 2 0 8 Value is 0 8 value which you enter Procedure for Temperature Re adjustment e Press UP and DOWN key at the same time for more than 5 seconds until Cor correction appears on the display and DOT blinks e With UP or DOWN key define value when DOT is blinking Pay attention to the sign of value plus or minus After 5 of defining the temperature display will automatically show temporary temperature 15 0 PRODUCT DISPOSAL INFORMATION FOR CORNING CUSTOMERS AND RECYCLERS According to Directive 2012 19 EU of the European Parliament and Council of 4 July 2012 on waste and electronic equipment WEEE as amended the Digital Water Bath is marked with the crossed out wheeled bin and must not be disposed of with domestic waste Consequently the buyer shall follow the instructions for reuse and recycling of waste electronic and electrical equipment WEEE provided with the products and available at the following link www corning com weee For additional information visit www corning com lifescie
3. this water bath in explosive or flammable environments Do not heat or incubate flammable explosive or highly reactive materials in this water bath as serious injury may result Temperature of water bath cover and chamber edges may reach over 100 C Use protective gloves to avoid burns and injuries When opening the water bath cover beware of steam It may be very hot and can cause burns on your face and hands Do not use tap water in water bath You may only use distilled water Do not use the water bath without water Minimum level of water is at least PEPE gt 20 mm 1 Place the unit on a level surface and plug the unit into a properly grounded outlet of appropriate voltage 2 Fill the chamber with desired quantity of distilled water Normal level of water is 2 3 of chamber height Minimum level of water is at least 50 mm 3 Put the stainless steel cover on unit 4 Press the power switch to the ON position The switch should illuminate The water bath automatically detects and displays the supply frequency as F50 or F60 Water bath starts to heat 5 Set desired temperature with UP or DOWN key You must hold UP or DOWN key for more than 2 seconds DOT on DISPLAY blinks When DOT starts to blink dot blinks for 5 seconds set temperature to desired value between room temperature 5 C to 99 C Two seconds after the desired temperature is set the display will automatically show momentary temperature 6 You can always look
4. 27 The Netherlands t 31 20 655 79 28 f 31 20 659 76 73 United Kingdom t 0800 376 8660 f 0800 279 1117 All Other European Countries t 31 0 20 659 60 51 f 31 0 20 659 76 73 LATIN AMERICA grupoLA corning com Brasil t 55 11 3089 7400 f 55 11 3167 0700 Mexico t 52 81 8158 8400 f 52 81 8313 8589 AAXYGEN GOSSELIN PYREX
5. Corning LSE GOINGS 6L Digital Water Bath INSTRUCTION MANUAL Models 6783 120V US 6784 230V EU Table of Contents 1 0 2 0 3 0 4 0 5 0 6 0 7 0 8 0 9 0 10 0 11 0 12 0 13 0 14 0 15 0 Warning Safety Information Package Contents Unpacking Installation Intended Use Symbols and Conventions Product Specifications Getting Started Important Guidelines Operation Troubleshooting Cleaning and Maintenance Temperature Re adjustment Product Disposal for Corning Customers and Recyclers 1 0 WARNING Configuring and operating this equipment requires precise knowledge and observance of the instructions in this user s manual The equipment is designed only for its intended application Important safety symbols used in this manual f Warning This is a warning of risk situations and dangers Failure to observe this warning could be life threatening These warnings must be heeded i This equipment is designed for 230V 50Hz or 120V 60Hz Make sure it is connected to the correct power supply Disconnect the power cord before carrying out maintenance or cleaning work 2 0 SAFETY INFORMATION The responsibility for operating the equipment lies exclusively with the owner or user if said equipment is incorrectly serviced maintained altered by persons not employed by an authorized dealer or if the equipment is used in a manner contrary to its specified purpose The equipment
6. must be maintained and operated in accordance with this user s manual Ensure this user s manual is accessible at all times and has been read and understood by all users 3 0 Package Contents Description Qty Water Bath 6L 1 User s Manual 1 Power Cord 1 4 0 5 0 6 0 7 0 UNPACKING Carefully inspect the unit and accessories thoroughly upon receipt for any damage that may have occurred during shipping Severe damage to the packaging container may indicate damage to the unit If you suspect damage to the unit immediately file a claim with the carrier in accordance with their instructions before contacting Corning Customer Service Also verify that all accessories are included and that the unit is in good working order before discarding the shipping packaging INSTALLATION Your satisfaction and safety require a complete understanding of this unit Read the instructions thoroughly and be sure all operators are given adequate training before attempting to put the unit in use NOTE This equipment must be used only for its intended application any alterations or modifications will void your warranty and may cause injury INTENDED USE The Corning Digital Water Bath is a compact general purpose convection bath capable of uniformly heating samples up to a maximum of 99 C SYMBOLS AND CONVENTIONS The following chart is an illustrated glossary of the symbols that are used in this manual CAUTION This symbol indicates a p
7. nces or call 1 800 492 1110 Outside the United States call 978 442 2200 For Corning technical information e mail us at ScientificSupport corning com or call 1 800 492 1110 Outside the United States call 978 442 2200 For more specific information on claims visit the Certificates page at www corning com lifesciences Warranty Disclaimer Unless otherwise specified all products are for research use only Not intended for use in diagnostic or therapeutic procedures Corning Life Sciences makes no claims regarding the performance of these products for clinical or diagnostic applications Corning Incorporated Life Sciences 836 North St Building 300 Suite 3401 Tewksbury MA 01876 t 800 492 1110 t 978 442 2200 f 978 442 2476 www corning com lifesciences Worldwide Support Offices ASIA PACIFIC Australia New Zealand t 61 427286832 China t 86 21 3338 4338 f 86 21 3338 4300 India t91 124 4604000 f 91 124 4604099 CORNING For a listing of trademarks visit www corning com clstrademarks All other trademarks are the property of their respective owners Corning Incorporated All rights reserved Printed in USA 11 15 CLS EQ 111 REV1 Japan t 81 3 3586 1996 f 81 3 3586 1291 Korea t 82 2 796 9500 f 82 2 796 9300 Singapore t 65 6572 9740 f 65 6861 2913 Taiwan t 886 2 2716 0338 f 886 2 2516 7500 FALCON EUROPE France t 0800 916 882 f 0800 918 636 Germany t 0800 101 1153 f 0800 101 24
8. otential risk and alerts you to proceed with caution and warns the user to proceed with caution CAUTION This symbol indicates risks associated with hot surfaces CAUTION This symbol indicates the presence of high voltage 8 0 9 0 10 0 PRODUCT SPECIFICATIONS Temperature range Ambient 5 C to 99 C Controls PID control Chamber maximum load 20 Ibs 9 09 kg Exterior dimensions W x D x H 14 9 x 11 4 x 9 0 inches 12 9 inches with cover 38 x 29 x 23 mm 33 mm with cover Weight 19 lbs 8 6 kg Electrical 120V 50 60 Hz 4 8 amps or 230V 50 60 Hz 2 4 amps or Chamber dimensions W x D x H 12 6 x 6 7 x 6 9 inches 32 x 17 x 18 cm GETTING STARTED 9 1 Check the data plate for voltage cycle wattage and ampere requirements If matched to your power source plug the power cord into a grounded outlet Voltage should not vary more than 10 from the data plate rating 9 2 In selecting a location consider all conditions which might affect performance such as heat from radiators autoclaves etc Avoid direct sun fast moving air currents heating cooling ducts and high traffic areas Allow a minimum of 5 cm between the unit and walls or partitions which might obstruct free airflow IMPORTANT GUIDELINES Description of front overlay CORNING Wanton POWER power key shines when water bath is ON fl UP key DOWN key Display DOT is blinking when you set temperature 11 0 OPERATION Do not use