1. 3 wire ground 3 0 APPLICABLE SPECIFICATIONS 4 0 ASTM F150 Electrical Resistance of Conductive Resilient Flooring NFPA 99 Standard for Health Care Facilities ESDA STM 4 1 Worksurfaces ESDA STM 7 1 Floor Materials ESDA STM 9 1 Footwear ESDA STM 12 1 Seating UNPACKING Before using your new Model 2003 Audit Kit check the contents against the Kit Contents List in Table 4 1 Contact ETS immediately should any shortages be noted TABLE 4 1 2003 Kit Contents List RER a O O Oe Y Y VV V z 3 1 Shielded Lead Black 5 1 5m 1 1 Blue Lead 5 i Sm 6 2 JhRed amp BlheClps 7 1 Black Lead 25 7 6m Plug Plug 8 24 BlackQlip 9 2 blb 2 27kg Probe Model 850 3 Alum Plate Gnd Plane 5 0 TEST PROCEDURES The testing of flooring material generally requires the measurement of point to point Pt Pt and point to ground RTG resistance and the recording of humidity and temperature Typical test voltages for these measurements are 100 or 500 Volts DC although measurements may be made at higher or lower voltages The Model 5646A is used to measure the humidity and temperature at the time of measurement The Operating Instructions are shown in Figure 5 0 1 g Ab a D Multi purpose cio 1 PA Holder for positioning stick in cuct 0 6 oz Dia The humidity sensor is orotected Dy a cuic TA of t 1S 5 Dot om c of t e probe stem t e protective caos clo
2. and workmanship for a period of one 1 year from date of invoice The Seller at his discrection will either replace or repair without charge F O B Glenside similar equipment or a similar part to replace any equipment or part of its manufacture which within the above stated time is proved to have been defective at the time it was sold All equipment claimed defective must be returned properly identified to the Seller or presented to one of its agents for inspection This warranty only applies to equipment operated in accordance with Seller s operating instructions Sellers warranty with respect to those parts of the equipment that are purchased from other manufacturers shall be subject only to that manufacturer s warranty The Seller s liability hereunder is expressly limited to repairing or replacing any parts of the equipment manufactured by the manufacturer and found to have been defective The Seller shall not be liable for damage resulting or claimed to result from any cause whatsoever This warranty becomes null and void should the equipment or any part thereof be abused or modified by the customer or if used in any application other than that for which it was intended This warranty to replace or repair is the only warranty either expressed or implied or provided by law and is in lieu of all other warranties The Seller denies any other promise guarantee or warranty with respect to the equipment or accessories and in particular a
3. 0 Ohms Testing should be performed in accordance with applicable specifications for work surfaces Applicable specifications include ESDA STM 4 1 DOD STD 1686A DOD HDBK 263 and FED STD 209C 7 0 SEATING To measure the resistance characteristics of seating follow the procedure specified in ESDA STM 12 1 It is similar to the RTG measurement The meter probe connections are shown in Figure 7 0 1 Figure 7 0 1 Meter Probe seating connections 8 0 Other Applications 8 1 Surface and Volume Resistance Resistivity Measurements 8 1 1 Volume Resistance Resistivity Volume resistance can also be measured with the standard Model 2003 Kit by connecting one 5lb 2 27kg Probe and the 3 76mm aluminum Plate to the Model 1035 as shown in Figure 8 1 1 Figure 8 1 1 Volume resistance hook up Volume resistivity is calculated as follows Py 31 7cm tcm x Rm Ohms cm Where t is the thickness of the sample in cm and R is the measured resistance in Ohms 8 1 2 Optional ETS Model 803B Surface Volume Resistance Probe The Model 803B is designed in accordance with ANSI ASTM D257 requirements for measuring surface and volume resistivity of planar material having a minimum diameter of 2 5 64mm It is specified in ESDA STM 11 11 and 11 12 for measuring the surface and volume resistance of planar material respectively This 5 lb 2 27kg Probe is compatible with the standard specimen size used in most electrostatic material evaluati
4. FLOORING WORKSURFACE RESISTANCE AUDIT KIT Model 2003 Operating Manual 4 09 electro tech systems Inc www elec trotechsy stems com 3101 Mt Carmel Avenue Glenside PA 19038 Tel 215 887 2196 Fax 215 887 0131 1 0 INTRODUCTION The Model 2003 Flooring Worksurface Resistance Audit Kit shown in Figure 1 0 1 contains a Megohmmeter a pair of NFPA ESDA STM4 5 Ib 2 27kg 2 5 64mm diameter probes 3 76mm dia Aluminum Plate Ground plane test leads Wiring Verifier and a Humidity Temperature Dew Point indicator that is housed in a sturdy ABS carrying case The kit enables the user to make a wide range of resistance measurements Although it is designed primarily for making point to point and point to ground resistance measurements on installed flooring and work surfaces where either a 100 or 500 Volt test voltage is specified plus it may be used in any application where the 5 Ib 2 27kg probe can be used With the addition of appropriate optional probes or test fixtures the kit can also measure surface and volume resistance resistivity of material and product plus wrist and heel ground strap resistance This manual along with the Model 1035 Megohmmeter manual covers the many applications for this versatile test kit The user is urged to read both manuals to become completely familiar with its many uses and applications Figure 1 0 1 Model 2003 Flooring Worksurface Resistance Audit kit 2 0 DESCRIPT
5. ION 2 1 Model 1035 Megohmmeter Resistance Ranges 1 to 400 kOhms 3 of reading 1 ct Vg 7 9V 10k to 2 0 GigOhm 3 of reading 5 ct lt 40 MegOhm 2 ct gt 40 MegOhm 2 Ve 50V 20k to 2 0 GigOhms 2 Ve 100V 50k 20 GigOhm 2 Vg 250V 100k to 20GigOhm Vg 500V Short circuit current 3ma Voltage Ranges 0 to 600 Volts AC or DC Audible Tone Upper lower alarm thresholds switchable ON OFF Batteries oix 6 x1 5 Volt AA alkaline Battery Life 3500 tests 5 seconds each 2 2 Model 850 Surface Resistance Probe Weight 5lbs 1 oz 2 27kg each Electrode Conductive elastomer rubber carbon loaded Hardness 65410 shore A durometer Diameter 2 5 4 032 641mm Resistivity 8 Ohms cm Minimum resistance lt 100 Ohms 2 3 Model 5646A Humidity Temperature Dew Point Indicator The Model 5646A Humidity Temperature Dew Point Indicator is a lightweight portable electronic instrument that measures both temperature and relative humidity and calculates dew point on a single LCD readout The capacitive thin film type humidity sensor has fast response and accurately measures over the range of 5 98 3 RH The resistance type temperature sensor provides a temperature measurement range of 32 to 122 1 F 20 to 70 0 5 C 2 4 Model 256 Wiring Verifier The Model 256 Wiring Verifier is used to both verify the correct wiring of the AC outlet North American Standard only and to provide a standard banana jack connection to the
6. Insulation Resistance Measurement Continuity Measurement The Model 1035 Megohmmeter is equipped with a low resistance range that can be used to verify the integrity of low resistance 0 to 400 kOhm circuits Refer to the Model 1035 instruction manual for making continuity measurements Ground Strap Testing Wrist straps and heel grounders may be tested for proper resistance using the Model 1035 The Black cable with alligator clip attached is plugged into the input on the meter The lead from the wrist strap can be plugged directly into the Red jack on the meter otherwise use the Red cable with alligator clip This will be required for measuring heel grounders The 50 Volt range is recommended since there is no advantage to testing at any higher voltage level A standard wrist strap should measure between 0 75 and 1 25 megOhms Voltage Measurement The meter may be used to measure AC or DC voltages up to 400 Volts Hefer to the Model 1035 instruction manual for details 10 9 0 10 0 OPTIONAL ITEMS AVAILABLE The Model 2003 Kit may incorporate additional instruments and accessories Instruction manuals for these items are either contained within this manual as an Appendix or included separately Please contact ETS for information on any of these optional items 4 09 WARRANTY Electro Tech Systems Inc warrants its equipment accessories and parts of its manufacture to be and remain free from defects in material
7. ess than 5 1 52m from the area being tested the 5 1 5m Black Shielded Cable can be used instead of the 25 7 6m cable as shown in Figure 5 2 2 Figure 5 2 2 Alternate RTG meter probe hook up using 5 1 5m cable Place the Probe at the desired test location and remove your hand from the Probe Select the desired test range and test voltage using the rotary switch on the Megohmmeter Without touching the Probe depress the test button on the Meter and record the resistance reading After the test button is released the Probe may be moved to the next test location and the next resistance reading taken NOTE Some test methods may specify that the 2 5 64mm face of the test probe be covered with conductive foil heavy duty aluminum foil or equivalent The kit does not contain this material The Model 850 uses a conductive rubber electrode with a maximum contact resistance of less than 100 Ohms If the user desires to utilize foil this may be accomplished by adding foil to each probe face and securing the foil to the probe using heavy duty rubber bands or 3 76mm hose clamps 6 0 WORK SURFACE TESTING The Model 1035 with the Model 850 Probe may be used to measure point to point or point to ground resistance of static dissipative work surfaces To make these measurements follow the same procedures and test set up as for testing floors Most work surfaces must be grounded and exhibit resistance on the order of 10 to 1
8. igh voltage is applied to the Probes when the button is down Once the button is released the high voltage is removed and the Probes may be handled Move the two Probes to the next test location and repeat the test procedure to obtain the next resistance reading 5 2 Resistance to Ground Testing The following kit items are required Model 1035 Megohmmeter 5 1 5m Red Cable 25 7 6m Plug to Plug Cable Model 850 5lb 2 27kg Probe Black Alligator Clip if required Model 256 Wiring Verifier Used with North American outlets oe a Referring to Figure 5 2 1 connect the Model 850 Probe to the Red Source terminal using the 5 1 5m Red Cable Connect the Megohmmeter to ground using the standard 25 7 6m Black Cable Plug one end of this cable into the Black terminal on the Meter and the other end into the ground of the 3 wire ground connection on a nearby AC outlet using the AC Wiring Verifier North American Outlets only The AC Wiring Verifier provides both a convenient banana jack connection to ground and checks the correct wiring of the AC outlet If a grounded metal object is used for the ground connection then plug the supplied Black alligator clip onto the end of the 25 7 6m cable and clip it to the metal ground For non North American Std Outlets Mains attach alligator clip amp connect to grounded metal obiect Figure 5 2 1 Typical RTG meter probe hook up NOTE If the ground point is l
9. ons Refer to Figure 8 1 2a and b respectively for surface and volume resistance measurements When used with the Model 1035 surface and volume resistance measurements over the range of lt 1 to 2x10 Ohms can be made When applicable surface resistivity ps is calculated using 10Rm and volume resistivity is calculated using 7 1 t x Rm 8 e E REsisrANCE RE STANCES 4 pro m s y ets sE lt a Surface Resistance b Volume resistance Figure 8 1 2 Meter Model 803B connections 8 2 Alternate Pt Pt and Volume Resistance Measurements The Model 1035 Megohmmeter may be used to measure the end to end and volume resistance of objects including tubing sample plaques integrated circuit IC shipping tubes etc using with the ETS Model 832 or Model 846 Clamp Electrodes The Black and Red leads are used and are connected to the Electrodes as shown in Figures 8 2 1a and b Refer to the Model 832 instruction manual for more detailed testing information 8 3 8 4 8 5 8 6 a End end Resistance b Volume resistance Figure 8 2 1 Meter Clamp Electrode hook up Insulation Resistance Measurements The test kit is capable of making high voltage high resistance measurements of many types of components including cables transformers motors generators relays and other electrical items where the insulation resistance between two points is to be determined Refer to the Model 1035 instruction manual for details concerning
10. rect sunlight ID i 9 H es per hancing cancels your warranty Changing the battery d instrument is switched off 1 Open battery compartment cover ve spent l d inse TON ia 55 moa Resmo r artis 3x AAA into the battery Observe polant IL y 3 Close battery compartment cover Auto Off function Ss If no button nas bean pressed in approx 70 _ minutes the instrument switc es itself of automaticaly WX EX HE 5 y gt T dad ESCH 9 0 d ru FS al Ke Ke Mer AE ef CEA BE Figure 5 0 1 Model 5646A operating instructions 5 1 Point to Point Resistance Testing The following 2009 kit items are required 1 Model 1035 Megohmmeter 2 5 1 5m Black Shielded Cable 3 5 1 5m Red Test Cable 4 2 Model 850 5lb 2 27kg Probes Connect each Model 850 Probe to the Megohmmeter using the Black and Red Meter terminals as shown in Figure 5 1 1 The thicker Black cable plugs into the dual input jack labeled Figure 5 1 1 Pt Pt resistance meter probe hook up Place the Model 850 Probes on the surface to be measured at the spacing required by the test specification A 3 0 9m spacing is typically required select the desired voltage and resistance range using the Megohmmeter rotary selector switch Without touching either Probe depress the large yellow test button on the Megohmmeter and record the resistance reading NOTE Never touch the Probes while the Megohmmeter test button is depressed H
11. s to its or their suitability for the purposes of the buyer or its or their performance either quantitatively or qualitatively or as to the products which it may produce The buyer is expected to expressly waive rights to any warranty other than that stated herein ETS must be notified before any equipment is returned for repair ETS will issue an RMA Return Material Authorization number for return of equipment Equipment should be shipped prepaid and insured in the original packaging If the original packaging is not available the equipment must be packed in a sufficiently large box or boxes if applicable of double wall construction with substantial packing around all sides The RMA number description of the problem along with the contact name and telephone number must be included in formal paperwork and enclosed with the instrument Round trip freight and related charges are the owner s responsibility WOODEN CRATES MUST NOT BE USED PACKAGING OF DELICATE INSTRUMENTS IN WOODEN CRATES SUBSTANTIALLY INCREASES THE CONTENT S SUSCEPTIBILITY TO SHOCK DAMAGE DO NOT PLACE INSTRUMENTS OR ACCESSORIES INSIDE OTHER INSTRUMENTS OR CHAMBERS ELECTRO TECH SYSTEMS INC WILL NOT ASSUME RESPONSIBILITY FOR ADDITIONAL COST OF REPAIR DUE TO DAMAGE INCURRED DURING SHIPMENT AS A RESULT OF POOR PACKAGING 1
12. sed should only be left open for t e duration cf the measurement Swivel head Ihe d olay can always be Leid K e 2 Commissioning Remove t e protective film on the display Switching on Q Frees button m ag SS sgmenttes Cu rrent reac ng testo BOb H2 Changing the parameters Press button repeatedly TE S T mperature Dewpont Wet bulb testo 696 H1 testo 6U6 H2 Yih Humidity Switching off Keep button pressed for 3 seconds Selecting between C F o When switching on keep button pressed until th segment test is finished instrument sw o configuration menus Press button again nstrwmsnt alternates Detween SC and GE Ws at eo befors pressing the button again The selected unit ther applies Technical data Type of application Measuring range cs Short Measurement 51 595 WAH 32 to 122 F 4 to 122 F td Resolution E D 1 SORA 0 1 F Accuracy a accuracy at 25 C F 23 958H 20 9 F Parameters RH C F eme temperature vk 122 F rage temparature 4 to 180 F t ypz 3x M Battery lifetime Approx 1000 h Probe stem D 0 5 in Length Aan Warranty d years CE guideine eerta UDA TOS EWG Please read before using instrument e Avoid condensation Cossrvs measurement ranges of sensor Do not sxosed maximum sto ora ge and operating abge erit eo protect measuring instrument from di