1. e Insert new battery with the positive end down Chart Rotation Speed The unit has a switchable clock you can change the speed from a 7 day to a 24 hour rotation and from a 24 hour to a 7 day rotation Make sure the right chart is selected for your rotation Simply move the switch located on the back of the unit to the speed that fits your application Rev 09 06 a toa a zZ eQ wn o Qa I suopnazsu Buresado Charts for current pricing go to www dicksonweb com or call 1 800 323 2448 F and C Switchable 7 Day 7 Day 24 Hour 24 Hour Range F Chart C Chart F Chart C Chart 14 to 32 F 20 to 0 C C302 C310 C301 C309 4 to 50 F 15 to 20 C C304 C312 C303 C311 22 to 68 F 5 to 20 C C320 C318 C319 C317 50 to 96 F 10 to 35 C C306 C341 C305 C313 76 to 122 F 25 to 50 C C308 C316 C307 C315 Accessories for current pricing go to www dicksonweb com or call 1 800 323 2448 Description Order NIST Traceable Calibration 3 pt new unit N300 A2LA Accredited Calibration 3 pt new unit N400 Pens 6 red P222 Locking Wall Mount Case A715 Waterproof Pouch A711 gt a a a n o z wu Rev 09 06 Calibration Services New Units N100 NIST Traceable Calibration 1 Point Includes documentation to one Dickson pre selected point on new units only N300 NIST Traceable Calibration 3 Point Includes documentation of three Dickson pre selected po
2. SC3 Temperature Chart Recorder Operation Contents Product Applications and Useful Features Product Specifications Operating Instructions Getting Started Charts amp Accessories Calibrations Troubleshooting Warranty Factory Service amp Returns Order Form Rev 09 06 y A A u fe Z Product Applications The SC3 sets a new standard for compact recorders while providing the accuracy and resolution required for regulatory compliance and quality control The reusable SC3 is rugged enough to withstand the rigors of transportation and storage applications Best of all its unbeatable low price makes temperature recording affordable for virtually any application Useful Features e Rugged case ideal for transportation and storage applications e F C Switchable 24 hour 7 day Switchable e Easy user calibration Security slots for recorder security e Special ranges for refrigerators freezers incubators and ambient conditions e Optional tamper seals can be used to insure the integrity of the chart e Compact less than 4 tall wide and durable e Operates for up to 2 years on a single AA battery Direct sensing no remote probe compact enough to be placed almost anywhere Rev 09 06 c v h a v lt 7 3 suoljeriddy Product Specifications Temperature Sensor Recorder Accuracy Ambient Operating Temperature Response Time Temperature Ranges Recording Tim
3. 01 400 24 50 18 401 700 33 66 21 701 1 000 43 70 30 1 001 1 500 64 89 45 1 501 2 000 84 111 53 2 001 over Please call Dickson Customer Service All shipments UPS 2nd day unless otherwise requested w404 13p410
4. e Power Supply Resolution Mounting Chart Size Dimensions Weight Includes Rev 09 06 Precise spiral wound bi metallic strip transducer 4 F 2 5 C over full scale 22 to 122 F 30 to 50 C 11 minutes to move 63 of full scale in air moving 5 per second F or C depending on chart installed 14 to 32 F 25 to 0 C 4 to 50 F 15 to 10 C 22 to 68 F 5 to 20 C 50 to 96 F 10 to 35 C 76 to 122 F 25 to 50 C 24 hour or 7 day Single AA battery up to 2 years average life Not less than 1 per 55 F in 1 C or 2 F steps Free standing wall mounted or suspended 3 75mm DIA 3 667 x 3 667 x 2 232 93mm x 93mm x 57mm 0 5 Ibs 227 g Battery pen and instruction manual wn 3s 5 a a Y s wu Getting Started e Place chart on recorder e Set time by inserting a coin into chart hub and turning clockwise until the current hour is referenced to the time index clip Time index clip is directly to the right and below the pen tip e Remove protective pen cap The chart movement is self starting and runs continuously Allow unit to stabilize in environment for 15 minutes Pen Replacement Slide the used pen cartridge off and slide the new one on Typical pen life is one year Battery Replacement The battery compartment is located on the back of the recorder The recorder uses a standard AAA battery for up to two years of battery life
5. ervice Department at 1 800 323 2448 or 630 543 3747 in Illinois Software Return Policy IMPORTANT Read your Software License Agreement carefully before installing software Dickson will accept returns for replacement of defective disks and CDs only DICKSON 930 South Westwood Avenue Addison Illinois 60101 Phone 630 543 3747 E mail DicksonCSR dicksonweb com Rev 09 06 suinjay 8 mu gY A e lt n S A D gY g s Coa lt Fax to 1 800 676 0498 Mail to Dickson 930 S Westwood Ave Addison IL 60101 Company Address City State Zip Phone Email Step 2 Ship To if different than above Name Company Address City State Zip Phone Email Step 3 Ordering Information Order Quantity Price Unit Total leach leach leach leach Subtotal In Illinois add 7 5 sales tax gt Tax Freight All Prices in U S Dollars gt Total Check Check oney Order Credit Card Credit Card Number Expires mm yy 0000 0000 O Signature Purchase Order P O Net 15 days for established customers Customer Total Order UPS 2nd Day UPS Next Day UPS Ground 0 100 19 39 12 1
6. ints a high medium and low on new units only N400 Deluxe A2LA Accredited NIST Traceable Calibration 3 Point ISO Guide 25 A2LA Documen tation of 3 pre selected points of as found data before and after calibration for Dickson temperature and or humidity instrumentation on new units only N995 NIST User Selected Temperature Points Documentation of one customer specified point Should be selected in addition to one of the above calibration options The Importance and Benefits of Regular Calibrations Once you begin to use your precision Dickson instrumentation regular calibrations are necessary to ensure accurate readings The following Calibration Services are available N150 NIST Traceable Calibration 1 Point Includes documentation to one Dickson pre selected point after re calibration N350 NIST Traceable Calibration 3 Point Includes documentation of three Dickson pre selected points a high medium and low after re calibration N450 Deluxe A2LA Accredited NIST Traceable Calibration 3 Point ISO Guide 25 A2LA Documen tation of 3 pre selected points of as found data before and after calibration for Dickson temperature and or humidity instrumentation N995 NIST User Selected Temperature Points Documentation of one customer specified point Should be selected in addition to one of the above calibration options Why should I recalibrate my instrumentation Over time dirt dust and normal handling can throw your prec
7. ision instrumentation out of calibration Regu lar calibrations ensure that you receive the most accurate readings possible How often should recalibrate my instrumentation Depending on the environment your instrument is used in and how often it is handled you will want to recalibrate your instrument every 6 to 12 months Instruments in environments where there are extreme tem peratures wide temperature ranges humidity or pressure variations high condensation dirt dust and other debris will require calibration at least every 6 months Instruments that are frequently moved or in locations with heavy machinery that cause vibrations should also be calibrated at least every 6 months Why should I return my instrument to Dickson for calibration Dickson calibrates your instrument at the factory using proprietary production calibration software tha guarantees proper calibration Our Capabilities Dickson is the first manufacturer of humidity and temperature instrumentation to receive A2LA accreditation We are also NIST Traceable our procedures conform to MIS STD 45662A ANSI NCSL 2540 1 1994 ISO IEC Guide 25 and 1SO10012 We are experts in the manufacture and calibration of humidity and temperature instruments Fast Service Our turnaround time is 3 days or less so you receive not only expert service but fast service as well Easy We make it easy for you No phone calls for Return Authorization Numbers are required We remind you when y
8. n pen chart and battery replacement The foregoing warranty and remedy are exclusive and in lieu of all other warranties either expressed or implied Dickson shall not be liable for consequential or incidental damages resulting from failure or malfunction of its products Dickson makes no warranty for products not manufactured by it or for any products modified by buyer or subject to misuse or neglect Factory Service amp Returns Contact the factory 630 543 3747 for a Return Authorization RA Number before returning any instrument The model number serial number and a purchase order number will be requested before an RA number is issued e Carefully repack the instrument label the outside of the box with the RA and return the instrument freight pre paid to Dickson e All instruments that do not have the RA clearly marked on the outside of the box will be refused When returning instruments for credit please include all accessories in shipment e Calibration Freight charges are non refundable NOTE Dickson shall not be liable for consequential or incidental damages resulting from failure or malfunction of its products Customer Satisfaction Dickson takes pride in providing you the customer with the highest quality instrumentation We welcome the opportunity to help you in any way possible Whether it be a question or a new idea in documentation the Dickson Company would like to hear your response Please call our Customer S
9. our instrument is due for calibration You simply send in the completed Calibration Order Form with your unit for calibration with freight prepaid to Dickson Rev 09 06 A 2 S gv s wv Troubleshooting Pen Tension If the pen is not touching the paper remove red pen and push the metal pen arm down near the middle of the arm If that does not work unscrew the pen arm from the recorder and bend the pen arm down Make sure the unit is in a stable environment replace the pen arm at the same spot from where it was removed Chart Does Not Rotate If the chart is not rotating make sure the center strip of the chart is inserted properly into the black center hub of the recorder making sure the chart is not torn or ripped If the chart is properly installed and there is still no rotation the clock needs to be replaced please call customer service for a return authorization number lt M 7 o fa Z iC Rev 09 06 Warranty Dickson warrants that the products it sells will be free from defects in material and workmanship under normal use and service for a period of twelve months after delivery In the event of a claim under this warranty the product or part must be returned to the factory for repair or replacement shipping pre paid with a Return Authorization Number see Return Information above It will be repaired at Dickson s option without charge This warranty DOES NOT cover routine calibratio