1. lf a personal monitor is due for a bump test the personal monitor will emit alternating flashes left and right every 5 seconds And the test icon will remain even after a button push This alert can be cleared by either placing the personal monitor in a C Test Station or manually performing a Bump Test To manually perform a Bump Test affix the flow plate to the sensor aperture and ensure the correct level of gas is connected via a fixed flow regulator 0 5 I m An exhaust pipe should be used into a safe area where no personal are present Press the button down once the word GAS will show on the display while the TEST icon flashes The personal monitor will wait for 45 seconds for the target gas to be applied or a button press to skip the bump test If the Crowcon Clip or Crowcon Clip is bumped while showing GAS then it will record as a bump test in the event log instead of as an exposure If no gas Is applied it will return to the normal screen and will not record anything in the event log 6 Oxygen Calibration A Crowcon recommends users of the O Crowcon Clip to bump test the personal monitor before using each day The Crowcon Clip Oxygen Oz personal monitor factory default will prompt the user to calibrate the personal monitor every 30 days The user will be prompted by the screen flashing CAL please see calibration instruction below A Only perform O2 calibration in clean air 20 9 Oxygen that is free of hazardous gas
2. High 15 ppm Low 2ppm High 5 ppm Low 25 ppm High 50ppm Low 35 ppm High 50 ppm Low 30 ppm High 100 ppm Min 19 5 Max 23 5 or Min 19 Max 23 5 Note these set points can be changed using the Crowcon IR Link Please refer to section XX for further details To display the personal monitor alarm set points press the button on the front of the personal monitor A use caution when changing alarm set points Confirm these levels with your company safety officer DO NOT use IR communications when an explosive atmosphere may be present Issue 1 June 2012 Page 6 Clip amp Clip User Manual CROWCON Gas Detection You Can Trust 4 Event Log By default the personal monitor stores the last twenty five 25 alarm events The system stores events by first in first out i e the 26 event will replace the first event and so on This information can be downloaded using the Crowcon IR Link For each alarm event the personal monitor records the following The personal monitor serial number Bump Test Yes or No Life remaining on the personal monitor Number of self tests performed Number of events Alarm Condition Low High or OL Specific event date and time Peak gas concentration in ppm or 5 Bump Test Interval Using the IR Link personal monitors can be programmed to alert the user if a bump test is due This interval can be set anywhere from 1 to 365 days Note the default is to have no bump interval programmed
3. read please contact Crowcon or your local gt months days hours O registered outlet immediately O 2 someses OOOO Ps o restren CS Pa i feeen 1 2 Day to Day Usage Oo a Indicatoricon Used During Real Time Gas Reading Prior to daily use the personal monitor will prompt the eee ___ O MS 00 9 000 no _ amp e a user to perform a self test by showing the test reminder Alarm Condition Icons icon this process is a simple and effective way to ensure Instrument Life Remaining Icons safe operation of the personal monitor During the Self Test the audio visual and vibrating alarms are activated sia LOE ers gion ath ae and the sensor is tested Below is a step by step process for performing the test Screen Display Detail o ee Screen 1 uw 0 When the Test Icon appears in the upper left hand corner a self test is required we moma Press the button on the front of the personal monitor to perform the test ALARM Screen 2 CO e F After pressing the button the following screen will appear During the self test amp OOO in ensure that the following occurs 1 The personal monitor emits one audible beep all LED s light up and the vibrate function is activated 2 All LCD display elements appear Screen 3 ws T F After the full element LCD screen the low alarm and high alarm set points will be ALARM oss displayed Note these alarm set points can be adjusted using
4. the IR Link or Clip C Test configuration options pom See section 5 2 for standard set points A 32 Screen 7 see Sections 4 6 for additional screen displays oO When a self test is successful the personal monitor will turn to the original screen months and display a check mark in place of where the test icon was previously displayed and one short audible beep will sound The personal monitor will by default prompt another Self Test in 20 hours from when the button was pressed Note this value can be changed via the IR Link software anywhere from 8 to 20 hours See the IR Link Quick Reference Guide for further detail TEST Hes i i Issue 1 June 2012 Page 4 Clip amp Clip User Manual CROWCON Gas Detection You Can Trust 2 Additional Steps A self test can be performed at any time during operation Simply press the button wait for the low alarm level to show and press the button once more This will activate the self test mode allowing a user to verify the personal monitor s functionality Above we outlined the most common screen configurations However if the personal monitor has been programmed via the IR Link or has been exposed to gas additional screens may appear screen 4 If applicable If programmed with a User ID after the alarm set points are displayed a combination of numbers and or letters will scroll across the screen This will be a max of 2 screens with a maximum character limit o
5. your local Crowcon office This software will work with Windows versions XP Vista and Windows 7 10 2 Getting started Below are instructions on how to use the IR Link software Once you have successfully installed the software and click on the IR Link icon you will see the screen to the right In order to interact with your Clip amp Clip make sure to have the IR Link plugged into your computer Place the monitor face down at least 2 inches in front of the IR Reader with the top side facing the IR Reader Then click Read Device if the read was successful you will see a green bar at the bottom of the screen 10 3 Changing Settings Both the Clip amp Clip monitors allow you to change the following fields below is a brief description of each field and changes possible a High Alarm Low Alarm Set Points Adjust the high low alarm set point Note When doing so make sure to set the alarm limit not to exceed the local or site guidelines Some sites may have different set points than others a f Crowcon IR LINK ver 1 3 0 2 Hardware Version Firmware Version Serial Number Life Remaining Sensor Type High Alarm Low Alarm Calibration Interval Bump Interval User ID Self Test Interval Display SH10000000 24 Months 0 Days H2S 10 0 pA 0 VA Days 0 WA Days 20 4 hours SensorlLife Remaining X Hide bump due LED flashing Read File Read Devic 2 Write File Write Device Hib
6. CROWCON Gas Detection You Can Trust After you have made your requested changes click on the Write Device Button If a successful write has occurred you will see a green bar at the bottom see above 10 4 Firmware Upgrade Both the Clip amp Clip monitors allow for Firmware upgrades to occur during the lifetime of the unit These may be product improvements to add additional features or to enhance monitor performance Most firmware upgrades are not required unless notified by Crowcon In order to perform a firmware upgrade make sure to position your monitor in front If the IR reader as instructed in the Getting Started section Be sure to read the device before performing any upgrades to insure a connection has been established Once a successful read has occurred click on the Folder Icon located in bottom middle right Select the firmware file you would like to upgrade to and then click Upgrade Firmware This process will take approx 30 seconds to perform DO NOT move the unit during a firmware upgrade Once a successful upgrade has occurred a green bar will appear with download image ok Firmware upgrades may also be performed through the Clip Dock Docking station 10 5 Hibernating Clip Your Crowcon Clip unit allows you to place the monitor in hibernate mode when it is not being used In order to perform hibernation place your monitor in front of the IR Reader as instructed in the Getting Started Section Be sur
7. Clip amp Clip User Manual CROWCON Gas Detection You Can Trust Crowcon Clip and Crowcon Clip Fixed life single gas monitor Installation and Operation Manual Crowcon Contacts UK 2 Blacklands Way Abingdon Business Park Abingdon OXON OX14 1DY 44 0 1235 557700 sales crowcon com US 21 Kenton Lands Road Erlanger Kentucky 41018 1845 1 859 957 1039 salesusa crowcon com NL Vlambloem 129 3068JG Rotterdam 31 10421 1232 eu crowcon com SG Block 194 Pandan Loop 06 20 Pantech Industrial Complex Singapore 128383 65 6745 2936 sales crowcon com sg CN Unit 316 Area 1 Tower B Chuangxin Building 12 Hongda North Road Beijing Economic amp Technological Development Area Beijing PRC 100176 86 10 6787 0335 saleschina crowcon com www crowcon com or www crowcon us Issue 1 June 2012 Page 1 Clip amp Clip User Manual CROWCON Gas Detection You Can Trust ieee FIRST BEFORE OPERATION Crowcon Clip and Crowcon Clip personal monitors are personal safety devices accordingly please read and understand the manual ensuring you have been properly trained on the use of the equipment and appropriate actions in the event of an alarm condition A WARNING STATEMENTS A A gt gt gt bP BP BP B amp B If the personal monitor is past the Activate Before Date on the package please do not activate Do not attempt part replacement or substitution as this could impair the intrinsic safety r
8. ating and will void the warranty of the product Before daily use check the following Sensor and Audio ports are clear of any obstructions i e debris or blockage Perform the Self Test to ensure proper operation of visual audible and vibrating alarms Confirm receipt of a check mark in the upper left hand corner signalling a successful Self Test Calibrate the O personal monitor at least every 30 days Make sure to calibrate in a clean air environment See O personal monitor section for instructions on calibration Both the CO and H2S versions of the Crowcon Clip and Crowcon Clip do not require calibration for the life of the product however we do offer this capability with the C Test and recommend regular Bump Tests please see our bump test recommendations below Bump Test the personal monitor at least every 90 days by challenging the sensor with a known concentration of the target gas to be detected Recommended target gas concentrations H2S 25 15 ppm CO 200 100 ppm O2 18 Bump Tests can be performed either manually or with the C Test If a manual test is to be performed make sure to test in a clean air environment lf a personal monitor fails the Self Test or Bump Test please discontinue use of the personal monitor The personal monitor contains a lithium battery that must be disposed of by a qualified recycler and in accordance with the EU WEEE Regulations Do not substitute internal components as this may interfere wit
9. dance with the instructions in this manual The warranty on replacement consumable items supplied under warranty to replace faulty items will be limited to the unexpired warranty of the original supplied item Crowcon reserves the right to determine a reduced warranty period or decline a warranty period for any sensor supplied for use in an environment or for an application known to carry risk of degradation or damage to the sensor Our liability in respect of defective equipment shall be limited to the obligations set out in the guarantee and any extended warranty condition or statement express or implied statutory or otherwise as to the merchantable quality of our equipment or its fitness for any particular purpose is excluded except as prohibited by statute This guarantee shall not affect a customer s statutory rights Crowcon reserves the right to apply a handling and carriage charge whereby units returned as faulty are found to require only normal calibration or servicing which the customer then declines to proceed with For warranty and technical support enquiries please contact Customer Support Tel 44 0 1235 557711 Fax 44 0 1235 557722 Email customersupport crowcon com Issue 1 June 2012 Page 10 Clip amp Clip User Manual 10 IR Link Software User Guide 10 1 Software Installation CROWCON Gas Detection You Can Trust Up to date software is available from your local service agent www crowcon com or from
10. e monitor in the C Test 3 Press and hold the bump test and calibration buttons down simultaneously for approx 2 seconds Issue 1 June 2012 Page 8 Clip amp Clip User Manual CROWCON Gas Detection You Can Trust 8 Certifications Specifications Intrinsic Safety QD A355 Groups A B C D T4 ETL 4003191 C 0359 Il 1 G Ex ia IIC T4 Ga IECEx Ex ia IIC T4 Ga ITS 12ATEX2547 IECEx ETL 12 0007 35 C lt Tamb lt 50 C O 40 C lt Tamb lt 50 C H2S 40 C lt Tamb lt 50 C CO Humidity 5 to 95 non condensing relative humidity Alarms Visual vibrating audible minimum 95dB Battery Life 24 months of operation 4 minutes of alarm per day Size 85mm x 50mm x 28mm 3 3 in x 2 0 in x 1 1 HxWxD Weight 76 g 2 7 oz Event Log Storage Last 25 events Newer events replace older events 9 Warranty Clip amp Clip 9 1 Limited Warranty Crowcon warrants this product to be free from defects in material and workmanship under normal use and service for a period of one year beginning upon the date of delivery for all Crowcon C Test monitors This warranty extends only to the sale of new and unused products to the original buyer 9 2 Warranty Procedure To facilitate efficient processing of any claim contact your local Crowcon agenit distributor a Crowcon regional office or our global customer support team English working language on 44 0 1235 557711 or customersupport crowcon com to obtain a return
11. e to read the device before performing hibernation to insure a connection has been established Note before you perform hibernation it is recommended that you save the event log on the monitor To save the log click on the Save Event Log button This will allow you to select a specific folder to save the event log Once hibernation has occurred all prior event log information will be erased tbat 0K Once the log has been saved you can proceed with hibernation To hibernate the unit press the Hibernate button A screen will pop up asking if you would like to save the event log If you have already saved the log click yes Successful hibernation will give an OK green bar Issue 1 June 2012 Page 12 Clip amp Clip User Manual CROWCON Gas Detection You Can Trust 10 6 Troubleshooting Connection Problems Clip Connect Error This error occurs when the IR Link cannot communicate with the unit The screenshot to the right displays the error Read Crowcon Clip Information lf this error occurs check the following items e Make sure the unit is placed face down in front of the ee eee ee IR Link at least 2 inchs i 7 Check IR LINK and Crowcon Clip connection e Also check that the IR window on the unit is not clogged with dirt or debris e Finally make sure the area where the unit and link are is not in a high light area i e bright light bulb or bright sunlight as light can interfere with IR connectivity IR Link Con
12. ernate Save Event Log Sensor Calibration ej Crowcon Clip Firmware Upgrade Select File CROWCON Abingdon Oxf Crowcon Detection Instruments 2 Blacklands Way i i ordshire 0X14 1DY Support 44 0 1235 557711 sales crowcon com 12 06 15 11 42 20 b Bump Interval In order to help with bump test accountability users can adjust the Bump Interval If a unit has not been bump tested within the time frame the unit will emit a flash until it has been tested either manually or via the C Test Station c User ID Each individual unit can now have a unique electronic user ID You are able to program in up to 6 characters Note Some alphabetic characters do not display well d Self Test Interval By default all Clip amp Clip monitors have a default 20 hour self test Frequency can be set anywhere from 8 to 20 hours e Show Sensor Readings Once the unit starts to see gas the unit will display the real time reading with a 0 1 ppm resolution If this is unchecked the unit will always display the life remaining or flash low high alarm in an alarm condition f Hide bump due LED Flashing If you prefer to not have the units flash when due for a bump test please make sure to check this box g Hide Clock This will turn the monitor into a real time ppm display monitor By default always showing a real time ppm reading Issue 1 June 2012 Page 11 Clip amp Clip User Manual
13. es Issue 1 June 2012 Page 7 Clip amp Clip User Manual CROWCON Gas Detection You Can Trust Calibration Instructions 1 Press and hold down the operator button for four 4 seconds 2 The screen will display CAL and the O2 icon will flash in the lower left hand side 3 After a successful calibration the personal monitor will emit one 1 beep vibration and LED flash 4 After an unsuccessful calibration the personal monitor will not beep or flash and continue to display calibration If after a few failed calibration please contact Crowcon customer support or your local service centre 7 Hibernation Instructions Crowcon Clip only Crowcon Clip Hibernation via the IR Link 1 Check that you have installed the IR Link Software and the IR Link USB connections are olugged in 2 Click Read Device on the IR Link Software 3 Note when the monitor is hibernated the event log will be cleared It is highly recommended to save the event log by pressing the save event log before hibernating 5 Keepthe Clip in front of the IR Link until the Hibernate OK message is displayed at the bottom of the IR Link Software 6 Confirm the monitor screen is blank 7 lf you encounter any problems please contact Crowcon or your local service centre Crowcon Clip Hibernation via C Test 1 Check that the C Test is turned on and the USB memory stick is present 2 The Clip C Test is capable of hibernating the Crowcon Clip place th
14. h the intrinsic safety of the device and will void all warranties Do not attempt to replace the battery or sensor this product is designed to be disposable Changing these components will void the warranty If you Suspect any malfunction or have any technical problems please contact Crowcon your local Crowcon office regional sales manager or registered service centre Issue 1 June 2012 Page 2 Clip amp Clip User Manual CROWCON Gas Detection You Can Trust Prologue Thank you for purchasing the new Crowcon Clip or Crowcon Clip At Crowcon we recognise the need for reliable and robust personal monitors which are sized to be worn without intruding on primary tasks and are simple to use The Clip and Clip are personal single gas monitors designed to detect the presence of specific toxic gases Carbon monoxide CO or Hydrogen sulphide H2S or Oxygen Oz deficiency Focused on users and fleet managers alike Crowcon offers application focused fixed system and portable solutions to protect personal and plant from flammable toxic and oxygen hazards For further information on this or any other Crowcon product please contact your local regional office regional sales manager or service centre Contents te ACTIV AUOMradatsantraneiiwasauiahwasinisaceaaeahnGndndsaceaaeaNnGrdndsnsiaausahnGsinesnsicausahuGsinesniicnuiahwGsanesnMicauichwasandsts tension 4 1 1 Display DETANS i sutrcatavwsscaumswstentasenpenupswaipentmsnspsmapsdoamd
15. ils 5 consecutive times the LCD will go blank In case of a blank LCD discontinue use and contact Crowcon or your local service centre for a replacement personal monitor Issue 1 June 2012 Page 5 Clip amp Clip User Manual CROWCON Gas Detection You Can Trust 3 Alarms 3 1 Alarm Types Screen Display E Audible Alarm One 1 slow beep every second H2S H Visual Alarm One 1 slow flash every second Vibrating Alarm One 1 slow vibration every second HIGH ALARM and OVER LIMIT OL ALARM Audible Alarm Two 2 fast beeps every second Visual Alarm Two 2 fast flashes every second Vibrating Alarm Two 2 fast vibrations every second PERSONAL MONITOR LIFE COUNTDOWN ALARM Once less than One 1 month of life is remaining the display will switch to days remaining when it is less than One 1 day remaining the display will switch to hours remaining Once the Eight 8 hours are remaining it will begin to beep flash and vibrate Eight 8 slow beeps per minute Eight 8 slow flashes per minute Eight 8 slow vibrations per minute To discontinue the alerts and end the unit life press the button When the battery life remaining clock displays 0 hours the personal monitor will operate for 8 hours before deactivating 3 2 Alarm Set Points In order to comply with the 3 main regional alarm requirements Crowcon offer the following standard Default factory set points Low 5 ppm High 10 ppm Low 10 ppm
16. mcnaesmapendoandacnaesnapsdesndacearenapenntsanavantsoaeunsedae 4 te DV ODV US GE aaa nee eee eee eee eee ee eee 4 Ze Adonai SC 0 Serre cere EE E EE E EREE E EEEE EEEE EE i EE Ei ES ii EEEa SEDE 5 A a S E E Rear semm enn sem encoun atpeens 6 3 1 Alarm Types tere re nretccsnmatoranew a recin eed caansatd aes sukesnssatsvacsneuissnssees aeeneukueniseesocaueniuecisehdncausnhuecasteyd 6 Que Alar Del F ONMG 2 ssapccasecasccubacanesasteapssandeapecuetcapscusacaaezme asaj seen ceacetinedeaseceiaeanssapacapecu asaystmndcaeesescead 6 a IMO Gy see eet atari aerate cient ates cians A AE A E E 7 J BUMD TeSa soie O e A e O E E O a E A O Ea 7 O COSA UIC oa ses pee a pees ree EEEE EERE EEEE REE EEEE EEE 7 7 IDEATION INSWUCTIONS CIDE ONY cscs cecncczncacesncnsmcwpnenteansedndeanensnancesesudeceeeendeanpeindeamenceaneetaiuieeeanee 8 8 Certifications Specifications cceccccseececeeeeceeeeeceeeeceeeeeseeeeseaeesseeeessaeessaeeeseecessaeeeseueesseessaeeeaaeeesaes 9 9 Waranty GID ODF csi cccptcendecndsstecustcendsandastecapdstedsandastecaedseadsanessteccudstsdeendsstereedseasandossenentseataaeeous 9 ele TAMING Waran y cereyan E ERE E EAN E 9 Jee Wan a 816 210 se E E E E E 9 Jos Waran DISCO asese A A E inns 10 10 JR Cink S ftware User GUIde iesnasioi aE a 11 10 1 Software Installation ec ccccceececeeeeeceeeeceeeeeseeeeseeeesseeeeseecesseeeeseeeeseaeesseeeeseneetseeetseeeeas 11 10 2 Gening SIAMCO si sercccksccsnancacacnciecaceancesadeancio
17. n the User ID of 6 characters The User ID can be changed modified via the IR Link software screen 5 if applicable If the personal monitor has been exposed to gas exceeding the low alarm set point a value will appear with max next to it This represents the peak value highest that the personal monitor has seen After this screen there will be another screen displaying a value with hours days or months this represents the amount of time past since the peak reading screen 6 if applicable After the peak reading and time since screens a screen will appear with CLP this stands for clear peak If the user presses the button down at while this is displayed the peak value on the personal monitor will be reset Note while the value will be cleared on the display the value will be held in the personal monitor s log See the event log section for further details This value can be cleared on the next screen Failures FAQ s If the self test fails the personal monitor emits five short beeps and flashes before displaying Test Repeat the self test lf the self test fails three 3 consecutive times the personal monitor will enter Fail Safe mode Please contact Crowcon or your local service centre for a replacement personal monitor During normal operation the battery is continuously monitored If the battery is low for more than 3 hours the personal monitor enters Fail Safe mode lf the battery self test fa
18. nect Error This error occurs when the IR Link is either not connected to the computer or the driver did not install correctly Make sure to check the USB connection on both the computer and the IR Link itself to insure all connections are secure If all connections are secure attempt to uninstall and reinstall the software 10 7 Warranty IR Link Limited Warranty Crowcon warrants this product to be free from defects in material and workmanship under normal use and service for a period of one year beginning upon the date of delivery This warranty extends only to the sale of new and unused products to the original buyer Warranty Procedure amp Warranty Disclaimer Please see sections 9 2 amp 9 3 Issue 1 June 2012 Page 13
19. nncearnenedaanclosnaGarawcedennclesnapaancadenneeenaeaeanmadetereasanamaees 11 6S Ganong 2111 0 cee ee eee eee eee ee eee eee eee eee eee eee 11 TO IMI WAS WIDOT AOS eenia E EE E demsmeanieasneuddeneutatder E 12 TO soe ANDOU UIC Oe acyece see cnseesemas se serene A E E O E E 12 VOC TORII SIN eps accesses aces dees scenes oem eee seco 13 107 Waranty R CINK serasa EEEE EEEE 13 Issue 1 June 2012 Page 3 Clip amp Clip User Manual CROWCON Gas Detection You Can Trust 1 Activation To activate the Crowcon Clip or Clip press and hold down the button for approximately 5 seconds On activation the personal monitor will vibrate flash red LED s and sound the audible alarm A successful activation will display the life remaining in months on the screen of the personal monitor as 24 months 1 1 Display Details The personal monitor utilises a special high viewing angle LCD that is designed to enhance the screen visibility In the absence of gas it displays life remaining In those cases where gas is present the display will automatically shift to a display that shows gas concentration and a battery icon Note the displayed number can be changed in the IR Link software with the Show Sensor Readings and Hide pe E Clock user options ma arg o ALARM C A Warning Users must familiarize themselves with the icons in both non alarm and alarm states A Warning If the display is missing icons or cannot be m o a clearly
20. s form for identification and traceability purposes This form may be downloaded from our website crowconsupport com and requires the following information e Your company name contact name phone number and email address e Description and quantity of goods being returned including any accessories e Instrument serial number s e Reason for return Products will not be accepted for warranty without a Crowcon Returns Number CRN It is essential that the address label is securely attached to the outer packaging of the returned goods The guarantee will be rendered invalid if the instrument is found to have been altered incorrectly stored modified dismantled tampered with or has not used Crowcon spares for replacement parts including sensors or has been serviced or repaired by any party not authorised and certified by Crowcon to do so The warranty does not cover misuse or abuse of the unit including use outside of specified limits Issue 1 June 2012 Page 9 Clip amp Clip User Manual CROWCON Gas Detection You Can Trust 9 3 Warranty Disclaimer Crowcon accept no liability for consequential or indirect loss or damage howsoever arising including any loss or damage arising out of the use of the instrument and all liability in respect of any third party is expressly excluded This warranty does not cover the accuracy of the calibration of the unit or the cosmetic finish of the product The unit must be maintained in accor

Download Pdf Manuals

image

Related Search

Related Contents

Manual de instrucciones  Sony MZ-E75 User's Manual  Mise en page 1 - Echo d`Alsace  Six Cell Docking Station User Manual  MaiorFlip® 85 / 90 / 100 / 900 / 900R  DIVCON User Manual  Livret de l`étudiant - Université de Bretagne Occidentale  Télécharger - La Bibliothèque Départementale de l`Yonne  Imagine™ I51-2 High- Performance Mobile Audio    