User`s Manual
Contents
1. SPECIFICATIONS Dimension 41 6mm x 82 3mm x 21 4mm Power source two 2 AAA 1 5V alkaline batteries Battery life Over 3 000 determinants Display Large LCD Memory 450 measurement results with date and time Auto electrode inserting detection Auto sample loading detection Auto reaction time count down Sleeping mode Power consumption less than 20uA Auto turn off after 3 minutes without action Temperature warning Operating condition 50 F 104 F 10 C 40 C Operating humidity range 10 90 R H non condensing Storage Transportation condition 4 F 158 F 20 C 70 C 5 95 R H Measurement Units Either mg dL or mmol L Measurement Range 20 600mg dL 1 1 33 3mmol L Performance characteristics e Accuracy Whole blood glucose test results were compared with YSI 2300 glucose analyzers 100 were with 15 mg dL 0 83 mmol L at concentration lt 75 mg dL 4 2 mmol L and 100 with glucose concentration lt 75 mg dL 4 2 mmol L were with 20 of the reference values The study shows that the Prodigy glucose monitoring system compares well with a laboratory system e Precision This study shows the intermediate precision and repeatability of less than 5 e Traceability of glucose monitoring system The results of Prodigy glucose monitoring system were compared to parallel results obtained on YSI 2300 which is the manufacturer s standard measurement procedure The study shows that the Prodigy glucose monitoring system
2. 90 day averages of your blood glucose test results You can review the test results in memory with these easy steps STEP 1 Enter the Memory Mode With the meter turned off press the main button twice The 7 day average will appear indicating that you are in the memory mode If you continue to press The Main Button the 14 21 28 60 and 90 day average will appear You can then review the last 450 tests in the memory The 7 day average is calculated from the blood glucose results obtained during the last 7 days It also indicates how many blood glucose tests have been performed within this period e g 58 58 tests in the last 7 days When using the meter for the first time will appear showing that f there are no test results in memory 1 20 Prodigy The 14 day average is calculated from the blood glucose results obtained during the last 14 days It too indicates how many blood glucose tests have been performed e g 24 24 tests in the last 14 days When using the meter for the first time 14 DAYa will appear showing that there are no test results in memory It s all the same to the 21 28 60 and 90 day average STEP 2 Recall Test Results After the 90 day average the most recent test result with date and time will be shown Press the Main Button once and the next most recent test result will appear Each time you press and release the main button the meter will recall up to your last 450 te
3. number Code the meter if necessary For meter model No 51720 you will hear a voice message please apply blood into the strip Prodigy STEP 2 Apply Sample Obtain a rounded drop of blood using a Lancing Device The blood sample must be at least 0 6 microliter in vol ume Qactual size or you may get an inaccurate test result When the symbol appears on the display apply the drop of blood to the absorbent hole of the test strip Do not push your finger against the test strip or try to apply a smeared sample The blood should completely fill the confirmation win dow before the meter begins to count down If the con firmation window does not fill completely before the meter begins to count down do not add more blood to the test strip discard the test strip and retest If you have trouble filling the test strip contact the technical support line NOTE If you do not apply a blood sample within 3 minutes the meter will turn itself off You must remove the test strip and insert it back into the meter to restart the test procedure STEP 3 Accurate Results in 6 Seconds Your blood glucose test result will appear after the meter counts down Blood glucose test results are auto matically stored in the meter memory Tum the meter off by removing the test strip Discard the used test strip carefully to avoid contamination For meter model No 51720 you will hear a voice Plasma calibrated message the blood glucose is
4. 20 25 C before testing Shake the vial discard the first drop of control solution and wipe off the dispenser tip to ensure a good sample and an accurate result Use only for 90 days after first opening Record the discard date date opened plus 90 days on the control solution vial Discard after 90 days Store the control solution tightly closed at temperatures below 86 F 30 C Do not refrigerate CAUTION The control solution range printed on the test strip vial is for Prodigy control solution only It is used to test meter and test strip performance It is not the recommended range for your blood glucose level 13 Prodigy How to do a control solution test STEP 1 Insert Test Strip Insert a test strip with contact bars end first and facing up into the test port Contact bars must be inserted all the way into the meter or you may get an inaccurate test result The meter will turn on automatically CH and strip symbol will appear briefly on the display Then the ambient temperature is displayed followed by the symbol with the code number Be sure the code number on the display matches the code number on the test strip vial If the code numbers do not match code the meter correctly After the symbol appears on display press the Main Button CTL will appear on the display With the CTL sign on the display the meter will not store your test result in memory If you decide not to perform a control
5. cool dry place below 104 40 C Keep away from direct sunlight and heat Store your test strips in their original vial only After removing a test strip from the vial immediately replace the vial cap and close it tightly With clean dry hands you may touch the test strip anywhere on its sur face when removing it from the vial or inserting it into the meter Use each test strip immediately after removing it from the vial Write the discard date on the vial label when you first open it Discard remaining test strips 90 days after first opening date Apply only blood sample or control solution to the absorbent hole Applying other substances to the absorbent hole will cause inaccurate results Do not use test strips beyond the expiration date printed on the package since they may cause inaccurate results Do not bend cut or alter a test strip in any way CAUTION Keep the test strips away from children the strips can be a potential choking hazard Never chew or swallow a test strip 9 Prodigy BEFORE TESTING Checking the Display E Each time you insert a test strip into the oa meter to turn the meter on the meter jer will display CH with the strip symbol briefly This tells you that the system is performing several self checks For meter model No 51720 you will hear a voice message the code number is number Coding the Meter The Code Button is used to calibrate the test strips with the met
6. correlated very well with manufacturer s reference method The device has certified to meet the following standards 98 79 EC IEC 60601 1 IEC 61010 1 IEC 60601 1 2 IEC61326 and ISO 15197 31 Prodigy The Prodigy Blood Glucose Monitoring System is warranted to be free of defects in material and workmanship for a period of one 1 year from the date of purchase This warranty extends only to the original purchaser and is not transferable For warranty assistance please contact Diagnostic Devices Inc Technical Support 1 800 243 2636 32 Prodigy EXPECTED TEST RESULTS Time of day Before Before lunch or dinner 1 hour after meals 2 hours after meals Between 2 and 4AM Glucose range for people without Your target range diabetes mg dL mmol L mg dL mmol L 70 105 3 9 5 8 mg dL mmol L 70 110 3 9 6 1 mg dL mmol L Less than 160 8 9 mg dL mmol L ane ____ mg dL mmol L Greater than 70 3 9 mg d mmol Source Krall LP and Beaser R S Joslin Diabetes Manual Philadelphia Lea and Febiger 1989 138 33 Prodigy Symbols convey information essential for proper use Referent Do not re use Consult operating instruction Keep away from sunlight Use by Date of manufacture Batch code Catalogue number Serial number Control 34
7. message indicates that there may by Repeat the test with a new a problem with the meter test strip If the problem persists please call the Technical Support Line at 1 800 243 2636 Error massage indicates that you may Review the instructions 6 E F have removed the strip after applying and try again with a new blood to the absorbent channel test strip 7 E A Error message indicates that there is a Review the instructions problem with the meter and try again with a new test strip If the problem persists please call the Technical Support Line at 1 800 243 2636 If the meter does not display a message after inserting a test strip PROBABLE CAUSE WHAT TO DO a Batteries exhausted Replace the battenes b Batteries incorrectly installed or Check that the batteries are correctly absent installed following the diagram c Test strip inserted upside down or Insert the test strip correctly with the incompletely contact bars end first and facing up If the test does not start after applying the sample a Insufficient blood sample Repeat the test with a new test strip and larger sample b Defective test strip Repeat the test with a new test strip c Sample applied after automatic Repeat the test with a new test strip shutoff two minutes after last user apply sample only when appears action on the display d Defective meter Please call the Technical Support 30 Line at 1 800 243 2636 Prodigy
8. solution test the CTL sign will disappear when you press the Main Button again CAUTION Every time you perform a control solution test you must enter into the CTL test mode so that the test will not be stored in the meter memory 14 Prodigy STEP 2 Apply Control Solution Shake the control solution vial well Remove the cap Squeeze the vial discard the first drop and wipe off the dispenser tip to ensure an accurate result Squeeze the vial again to get another drop and place the drop on your finger tip first Then apply the drop to the absorbent hole of the test strip where it meets the narrow channel until the drop is drawn into the test strip the confirmation window is filled and the meter begins to count down CAUTION To avoid contaminating the control solution with the content of the test strip you have to place a drop of control solution on a clean surface or on your finger tip first Then touch the test strip to the drop 15 correct incorrect Prodigy Prodigy STEP 3 Result Appears in 6 Seconds The meter will count down and the control solution test result will appear Compare the result with the range printed on the test strip vial The result should fall within this range For meter model No 51720 you will hear a voice message the blood glucose is number unit Comparing control solution results If test results fall outside the range printed on the test strip vial repeat
9. OLVING GUIDE Following is a summary of some display messages and symbols These messages help to identify certain problems but do not appear in all cases when a problem has occurred Improper use may cause an inaccurate result without producing an error message or a symbol In the event of a problem refer to the information under action What it means Action The symbol appears with this 1 E b message on the display by itself This means that the batteries will not provide enough power for a test Error message could be caused by a used test strip or a temporary or permanent electronics problem occurred 2 E U The temperature of the environment 3 E t meter or test strip was below or above the system operation range You can not perform a test until the meter and test strip reach a temperature within the operation range of 50 104 F 10 40 C Replace the batteries at once Repeat the test with a new test strip If the error message appears again please call the Technical Support Line at 1 800 243 2636 Repeat the test after the meter and test strip have reached a temperature within the operation range Error message indicates that there is a 4 E problem with the meter 29 Review the instructions and try again with a new test strip If the problem persists please call the Technical Support Line at 1 800 243 2636 Prodigy Message___Whatit means Action e 5 E O Error
10. User s Manual Prodigy Blood Glucose Monitoring System Prodigy Dear Prodigy System Owner You have chosen one of the best blood glucose monitoring systems available This booklet has important information about the system Please read it carefully Your system provides you with accurate plasma calibrated test results This feature makes it easy for you and your doctor to compare your meter results with laboratory results Blood glucose monitoring plays an important role in diabetes control A long term study showed that keeping blood glucose levels close to normal can reduce the risk of diabetes complications by up to 60 The results you get with the system can help you and your healthcare professional monitor and adjust your treatment plan to gain better control of your diabetes A warranty registration card is included with your system Please complete it and mail it to us American Diabetes Association position statement on the Diabetes Control and Complications Trial 1993 Prodigy The system is intended for use outside the body in vitro diagnostic use It should be used only for testing glucose sugar and only with fresh capillary whole blood samples The system is intended for use in the home and in clinical settings It should not be used for the diagnosis of diabetes or for the testing of newborns CAUTION Before using any product to test your blood sugar blood glucose read all instructions and pra
11. Your Prodigy Blood Glucose monitoring system uses two 2 AAA 1 5 V alkaline batteries included with your system that need to be installed in the meter before use The new batteries will provide you with enough power to perform about 3 000 tests When replacing the batteries use only 2 AAA 1 5 V alkaline batteries The meter will alert you when the power is getting low by displaying two different messages 1 The symbol appears on the display when the meter is turned on and all other display messages are functional From the time the symbol first appears there is enough power left for about 50 tests The test results will be accurate but it is time to change the batteries 2 The symbol appears with the E b symbol on the displays by itself This means that the batteries will not provide enough power for a test You must change the batteries Replacing the batteries a Make sure that the meter is turned off b Tum the meter over and locate the battery compartment c Press the buckle on battery cover and lift up to remove cover d Remove the old batteries Insert two 2 AAA 1 5 V alkaline batteries making sure you follow the batteries diagram for a proper installation e Replace battery cover back into place NOTE Replacing the batteries does not affect the meter s memory previous test results stored inmemory However the time and date settings may need to be updated 28 Prodigy DISPLAY MESSAGES AND PROBLEM S
12. cantly affect the test results e This system maybe used at altitudes up to 10 742 feet 3275m without an effect on test results TABLE OF CONTENTS IMPORTANT INFORMATION GETTING STARTED The Prodigy Blood Glucose Monitoring System Prodigy Blood Glucose Meter Meter Display Segments Prodigy Test Strips BEFORE TESTING Checking the Display Coding the Meter Checking the System With Prodigy Control Solutions TESTING YOUR BLOOD Step by Step Test Procedure Special Messages Using the Meter Memory COMPARING METER AND LABORATORY RESULTS SETTING THE METER AND DELETING THE MEMORY Setting the Time Date Unit of measurement Unit of temperature and Memory deletion Downloading Test results to your PC CARING FOR YOUR METER Maintenance Power DISPLAY MESSAGES AND PROBLEM SOLVING GUIDE SPECIFICATIONS Performance Characteristics WARRANTY EXPECTED TEST RESULTS SYMBOL INFORMATION Prodigy GETTING STARTED The Prodigy Diabetes Blood Glucose Test System The Prodigy system consists of three main products the Prodigy blood glucose meter Prodigy test strips and Prodigy control solutions These products have been designed tested and proven to work together as a system to produce accurate blood glucose test results Use only Prodigy test strips and Prodigy control solutions with the blood glucose meter CAUTION Check your system to be sure the outer carton seal is not missing or broken If either of these conditions occurred please retu
13. ctice the test Do all quality control checks as directed and consult with a diabetes healthcare professional These recommendations apply to all blood glucose monitoring systems and are supported by the Diabetes Educator Section of the Canadian Diabetes Association Prodigy Important Information e Severe dehydration and excessive water loss may cause false low results If you believe you are suffering from severe dehydration consult a healthcare professional immediately e Test results below 60 mg dL 3 3 mmol L mean low blood glucose hypoglycemia Test results greater than 240 mg dL 13 3 mmol L mean high blood glucose hyperglycemia If you get results below 60 mg dL or above 240 mg dL and do not have symptoms first repeat the test If you have symptoms or continue to get results that fall below 60 mg dL or above 240 mg dL follow the treatment advice of your healthcare professional e If you are experiencing symptoms that are not consistent with your blood glucose test results and you have followed all instructions described in the Prodigy owner s booklet call your healthcare professional A red blood cell count hematocrit that is very high above 60 or very low below 20 can cause false results e Elevated blood triglyceride reducing substances such as uric acid and ascorbic acid at normal blood concentration or acetaminophen dopa methyldopa L dopa and tolbutamide at normal blood concentrations do not signifi
14. ctly It is very important that you do this simple check routinely to make sure you get accurate results Prodigy control solutions contain a known amount of glucose that reacts with test strips The Low and High level Control solutions are intended to check the monitoring system in different measurement ranges Compare your control solution test results with the expected range printed on the test strip vial label Control solution should be used to e Practice the test procedure e Make sure your meter and test strips are working together properly e Ensure you are performing the test correctly Before you use the meter to test your blood for the first time practice the procedure using control solution When you can do three tests in a row that are within the expected range you are ready to test your blood Do a control solution test e When you begin using a new vial of test strips At least once a week e Whenever you suspect that the meter or test strips are not working properly e When your blood glucose test results are not consistent with how you feel or when you think your results are not accurate e If you drop the meter After you change the meter batteries 12 Prodigy Prodigy Important Control Solution Test Information Use only Prodigy control solutions Check the expiration date on the control solution vial Do not use if expired Control solution meter and test strips should come to room temperature 68 77 F
15. er for accurate results gt Please code the meter before using it j for the first time and then every time you change to another vial of test strips Each time you test check that the code number on the meter display matches the code number on the test strip vial CAUTION If the code number displayed on the meter does not match the number printed on the vial test results may be inaccurate 10 Prodigy STEP 1 Enter the Code Mode Code Code Start with the meter turned off Press the Code Button The Code symbol and SEL L will appear on the LCD display followed by the code number with the number blinking STEP 2 Match the Code Numbers Compare the code number flashing on the LCD display with the code number on the test strip vial If the two code numbers match you may begin testing If they do not match follow Step 3 When you first use the meter you need to code the meter before testing os STEP 3 Code the Meter Press and release the Code Button to advance Gode one number When the code you need appears mM c on the display press the Main Button to set the LJ code and turn off the meter If you do not press any button in two minutes the meter will turn off automatically 11 Prodigy Checking the System with Prodigy Control Solution Prodigy control solutions are used to check that the meter and the test strips are working together as a system and that you are performing the test corre
16. minutes of each other 22 Prodigy Prodigy e Wash your hands before obtaining a blood sample e Never use your meter with blood that has been collected in a gray top test tube e Use fresh capillary blood only You may still have a variation from the result because blood glucose levels can change significantly over short periods especially if you have recently eaten exercised taken medication or experienced stress In addition if you have eaten recently the blood glucose level from a finger stick can be up to 70 mg dL 3 9 mmol L higher than blood drawn from a vein venous sample used for a lab tesf Therefore it is best to fast for eight hours before doing comparison tests Factors such as the amount of red blood cells in the blood a high or low hematocrit or the loss of body fluid severe dehydration may also cause a meter result to be different from a laboratory result References 1 Surwit R S and Feinglos M N Diabetes Forecast 1988 April 49 51 2 Sacks D B Carbohydrates Burtis C A and Ashwood E R ed Tietz Textbook of Clinical Chemistry Philadelphia W B Saunders Company 1994 959 23 Prodigy SETTING THE METER AND DELETING THE MEMORY Before you start using the meter for first time you may need to setup the time date unit of measurement and unit of temperature and language You also may need to enter the set mode and reset the meter if you need to change the time or if y
17. number unit result example The Prodigy meter provides accurate plasma calibrated test results that can be compared directly with laboratory results 18 Prodigy Special Messages Some special symbols or messages will appear depending on your test result 1 When below 20 mg dL 1 1 mmol L Lo will appear on the meter display instead of numbers 2 When between 20 to 70 mg dL 1 1 and 3 9 mmol L the low symbol will appear with your test result Lo or low symbol indicates hypoglycemia low blood glucose You should immediately treat hypoglycemia as recommended by your healthcare professional 3 When within optimal range which means 70 to 120 mg dL 3 9 to 6 6 mmol L a smile will appear with your test result 4 When between 120 to 240 mg dL 6 6 and 13 3 mmol L the high symbol will appear with your test result 5 When above 240 mg dL 13 3 mmol L KETONE symbol will appear on the meter display This indicates there is a possibility of Ketone accumulation if you are type 1 diabetic Please seek immediate medical assistance 6 When above 600 mg dL 33 3 mmol L Hi will appear on the meter display instead of number This indicates severe hyperglycemia Please seek immediate medical assistance 19 Prodigy Prodigy USING THE METER MEMORY Your Meter stores the 450 most recent blood glucose test results with date and time in its memory It also provides you with 7 14 21 28 60 and
18. ou remove the batteries Setting the Time Date Unit of Measurement Unit of Temperature and Memory Deletion To set the time you must first enter the setting mode Start with the meter off Then press the Set Button located in the batteries compartment The meter is now in the set mode STEP 1 Set the Year First all symbols on LCD will be displayed Then the year ST ws will appear with the number flashing Press and release a Lj the Main Button to advance one year To move faster hold N the Main Button down With the correct year on the dis io play press the Set Button and the date will appear on the display with the month segment flashing u cornu cow STEP 2 Set the Month Press and release the Main Button until the correct month appears To move faster hold the Main Button down With the correct I month on the display press the setting button and the day segment will start flashing 24 Prodigy STEP 3 Set the Day Press and release the Main Button until the correct day appears To move faster hold the Main Button down With the correct day on the display press the Set Button and the time will appear on the display with the hour segment flashing Before setting the time you must have first set the date STEP 4 Set the Hour Press and release the Main Button to advance one hour To move faster hold the Main Button down With the D 0 correct hour on the display press the Set Button and
19. rn your system to the place of purchase Prodigy Prodigy Blood Glucose Meter TEST SLOT The test slots is where you insert the test strip for testing The meter will turn on automatically when you insert a test Codel cr q strip SET BUTTON The Set button in the back of the a S TE meter is used to set the date time units and other meter functions DISPLAY Your test results are displayed here The display guides you through the test using symbols and simple messages MAIN BUTTON S The main button in front of the meter D f7 CODE BUTTON is used to set up the meter enter the N memory mode change the value of The code button date and time unit setting memory at the side is deleting and to turn the meter on used to CODE and off the meter to the strip Prodigy Meter Display Segments VOICE ON OFF Indicat Joti TEST STRIP SYMBOL Ony fer notat ental 64a 817203 Flashes when you turn the meter on The symbol stays on when the strip TEST RESULT AREA has been inserted Test results are displayed here A decimal point appears when BLOOD DROP SYMBOL the unit of measurement is Appears when the meter is ready for you to OOS eect add blood to strip LOW BATTERY SYMBOL g when it is ti CODE Codet I4 A Change ma betteries Displayed with the code number UNITS OF j currently in the meter MEASUREMENT Depending upon how the meter TEMPERATURE low igh se is
20. set either mmol L or mg dL SYMBOL iB cT AT will appear with the test result Appears with ambient a KETONE WARNING temperature B B M When your blood glucose test result is above 13 3mmolV L 240mg9 dL this symbol will LOW SYMBOL appear with 7 sign on the Appears when test result display is lower than normal HAPPY FACE DATE amp TIME Appears when test it mal HIGH SYMBOL eres Appears when test result MEMORY SYMBOL is higher than normal Indicates a test result store in memory CTL SYMBOL Appears when in control solution test Its result won t be stored in the memory F Prodigy Prodigy Test Strips Your system measures the amount of sugar glucose in whole blood Blood is applied to the absorbent hole of the test strip and is automatically drawn into the reaction cell where the reaction takes place The test strip consists of the following parts CONTACT BARS e Insert this end of the test strip into the meter Push it in firmly until it will go no further TEST STHIP HANDLE mem Hold this part to insert the test strip into the slot CONFIRMATION WINDOW This is where you confirm if enough blood has been applied to the absorbent channel of the strip ABSORBENT PAE atc Apply a drop of blood here the blood will be drawn automatically See pages 17 19 Testing Your Blood for complete instructions 8 Prodigy Pro Important Test Strip Information Store test strip packages in a
21. st results in order When the memory is full the oldest result is dropped as the newest is added When reaching the last set of result End will be shown on LCD display and the meter turns off automatically NOTE If no button is pressed for 2 minutes the meter will show OFF and turn off automatically STEP 3 Exit The Memory Mode Press the Main Button for three seconds to turn off the meter 21 Lu T IHE i20 e caru ow Y E U e faal 2 a com coun v L IO Prodigy Prodigy COMPARING METER AND LABORATORY RESULTS Test results from meter and laboratory are both expressed in plasma equivalent units However the result you obtain from your meter may differ somewhat from your laboratory result due to normal variation Meter results can be affected by factors and conditions that do not affect laboratory results in the same way See test strip package insert for typical accuracy and precision data and for important information on limitations To make an accurate comparison between meter and laboratory results follow the guidelines below Before you go to the lab e Perform a control solution test to make sure that the meter is working properly e Itis best to fast for at least eight hours before doing comparison tests e Take your meter with you to the lab While at the lab e Make sure that the samples for both tests the meter test and the lab test are taken and tested within 15
22. the minutes segment will start flashing STEP 5 Set the Minutes Press and release the Main Button to advance one minute To move faster hold the Main Button down With the correct minute on the display press the Set Button and the current unit of measurement will start flashing 0 NOTE Your meter displays 7 14 21 28 60 and 90 day averages which you can access from the meter memory These averages are calculated from results obtained during the 7 14 21 28 60 and 90 days preceding the current date and time settings When the date and time are changed the 7 14 21 28 60 and 90 day averages may change art Caru Your meter can display test results in milligrams per deciliter mg dL or mil limoles per litre mmol L The mg dL unit is standard in the United States The mmol L unit is standard in Canada 25 Prodigy STEP 6 Select mg dL or mmol L Press and release the Main Button until the unit of measurement you are choosing appears on the display Press the Set Button and the current unit of temperature will start flashing STEP 7 Select C or F Press and release the Main Button until the unit of temperature you are choosing appears on the display Press the Set Button and the meter will display dEL with flashing mem symbol STEP 8 Delete memory and exit the setting mode dEL and flashing M symbol appears on the display When you press the Main Button dEL will flash too Press the Main Bu
23. the test Out of range results may be caused by one or more of the following e Error in performing the test Failure to shake the control solution vial vigorously Expired or contaminated control solution Control solution that is too warm or too cold Failure to discard the first drop of control solution and wipe the dispenser tip clean e Improper coding of the meter e Test strip deterioration e Meter malfunction CAUTION If you continue to get control solution test results that fall outside of the range printed on the vial the system may not be working properly Do not use the system to test your blood if you continue to get test results that fall outside of the range If you are unable to resolve the problem call the Technical Support Line at 1 800 243 2636 16 Prodigy TESTING YOUR BLOOD Be sure to read this section and the test strip package insert found in the test strip box carefully before testing Make sure you have all items needed to test Blood Glucose Meter Test Strip Lancing Device Sterile Lancet 9Om gt Step by Step Test Procedure STEP 1 Insert the test strip Insert the test strip with contact bars end first and facing up into the test port fully The bars must be all the way into the meter to avoid an inaccurate result The meter will turn on automatically CH and strip symbol will appear briefly on this display Next the ambient temperature will appear and followed by the symbol with the code
24. tton again to delete all the meter Memory records When the cis is ail rai l meter displays the smile it means that the mem ory is deleted You can press the Set Button to turn off the meter If you do not want to delete the memory do not press the Main Button but press the Set Button again to turn off the meter 26 Prodigy DOWNLOADING TEST RESULTS TO YOUR COMPUTER Prodigy allows you to download all your test results in your computer helping you or your health care professional track changes in your blood glucose level over time To start downloading your results you need to connect the downloading cable to your computer and to the Prodigy meter Download Port and install a free software to your PC that will let you complete this process easily You can get the free copy of this software on our web site at www diagnosticdevicesinc com Downloading cable sold separately CARING FOR YOUR METER Maintenance Your meter does not require special maintenance As no blood or control solution comes in contact with the meter there is no special cleaning required Take care to avoid getting dirt dust blood control solution or water inside the meter through the test port or data port Store the meter in its carrying bag after each use A cloth dampened with water and mild detergent can be used to wipe down the outside of the meter Your meter is a precision instrument Please handle it with care 27 Prodigy Power