ELECTRO-SURGERY UNIT ANALYZER

image

Contents

1. Color coded scales in Watts and Milliamps 50 100 200 300 400 amp 500 ohms non inductive Each range s ohms load value corresponds to its full scale watts value Example 50 ohms gt 50 watts Load accuracy is 2 0 5 of full scale 0 10MHz 0 1000 mA for all ranges 500 mA maximum into 200 ohm load non inductive Digital display 0 999 ohms 1 ohm resolution 3 ohms 3 accuracy Isolated amp uncalibrated FWH 1 2 0 5 amp fast blow High impact plastic enclosure 4 125 H x 8 4 W x 9 1 D 4 pounds net weight 10 Accessories Included Accessory Red test lead Black test lead Green test lead Alligator clips Red clip insulators Black clip insulators Spare fuses User manual ESM 1 Switching Module EN ee Optional Accessories Carrying Case P MANUALS BCGroup ESU 2000A ESU 2000A_UM_RevO2 doc 11 NOTES 12
2. IS ASSUMED FOR INACCURACIES Manual ESU 2000A Copyright 2005 www bcgroupintl com Made in the USA 10 05 Rev 02 BC GROUP ESU 2000A ELECTRO SURGERY UNIT ANALYZER The Model ESU 2000A is a precision Electro Surgery Unit Analyzer ESU It tests the RF power output of Electro Surgery Units The power reading is taken from a large 4 inch analog display It offers six resistive loads for RF power testing from 50 to 500 ohms along with a 200 ohm load for RF leakage current testing The patient return electrode resistance simulation is adjustable in one ohm increments from 0 to 999 ohms The following are highlights of some of the main features LARGE 4 INCH ANALOG DISPLAY 5 OF FULL SCALE ACCURACY RETURN CONTACT MONITOR TEST RF LEAKAGE TEST PORTABLE OSCILLOSCOPE ISOLATED OUTPUT 0 10 MHz BANDWIDTH EASY OPERATION ACCESSORY KIT INCLUDED HIGH IMPLACT PLASTIC ENCLOSURE OVERVIEW This section looks at the layout of the ESU 2000A and gives descriptions of the elements that are present 4 INCH ANALOG FWH 0 5 AMP Fast Blow Fuse DISPLAY Patient Plate Return Electrode Monitor Test Te RETURN ELecraooe MONITOR TEST Oscilloscope Isolated Output ESU Power Ohms Load amp Watts Full Scale Potentiometer Ohms RF Leakage Input 304 dn pua sry Carrying Handle Switching Module ESM 1 RF Leakage Testing amp ESU Power Testing SAFETY PRECAUTIONS CA
3. Ge Biomedical A Unit of BC Group International nc Y ELECTRO SURGERY UNIT ANALYZER L RF Leakage Te Testing H CE al ESU 2000A USER MANUAL BC BIOMEDICAL ESU 2000A TABLE OF CONTENTS DESCRIPTION socorro a a RER UL etd dada eld eeciateaed 3 OVER VLE ev eV N TEET a ee 4 SAFETY PRECAUTIONS cin 5 OPERATING INSTRUCTIONS iiianoe tala on la tone tod 6 MANUAL REVISIONS a 2 er ven auegeiec Due ceneveneceereaeceds ces 9 WARRANTY zerstreut 9 SPECIFICATIONS a aa a 10 NOTES Sr rs rer nen N 12 WARNING All connections to patients must be removed before connecting the Device Under Test DUT to the Analyzer A serious hazard may occur if the patient is connected when testing with the Simulator Do not connect any leads from the patient directly to the Analyzer or DUT CAUTION Always turn the ESU under test off before making or removing connections Never touch exposed metal surfaces on test leads or other current carrying parts while ESU is activated Never open the ESU 2000A when connected to an ESU Electro Surgery units generate high frequency RF voltages at high levels NOTICE BC GROUP INTERNATIONAL INC RESERVES THE RIGHT TO MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME WITHOUT NOTICE IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE HOWEVER NO RESPONSIBILITY
4. UTION Electro Surgery Units ESUs produce high frequency RF output voltages up to 500 volts CAUTION Always turn the ESU under test Off before making or removing connections CAUTION Never touch exposed metal surfaces on test leads or other current carrying parts while the ESU is activated CAUTION Refer to ESU manufacturer s manuals and specifications for safety and testing information CAUTION Only replace the ESU 2000A fuse with the specified rating and type CAUTION Never open the ESU 2000A case when it is connected to an ESU OPERATING INSTRUCTIONS ESU RF Power Measurements 1 Consult the ESU manufacturer s manual for load specifications 2 Turn off ESU under test Remove all connections between the ESU under test and the ESU 2000A 3 Install the BC Biomedical Switching Module on the ESU 2000A front panel so that the words ESU Power Testing are right side up When installed correctly the Active Electrode connectors should be exposed and the RF Leakage connectors should be inaccessible 4 Connect the ESU s Patient Plate to the ESU 2000A s black Patient Plate connector using a black test lead 5 Connect the ESU s Active Electrode to the ESU 2000A s desired load range connector in the Active Electrode area on the left side of the front panel The Active Electrode input connectors are color coded and each corresponds to a color coded range on the analog meter The full scale watts ran
5. ge for each range equals the input resistance For example the 50 ohms input resistance range has a full scale watts range of 50 watts etc 6 Turn the controls on the ESU under test to minimum power and then activate the CUT or COAG modes Turn the ESU power to the desired levels being careful to not exceed the full scale wattage range selected on the ESU 2000A Note the reading on the ESU 2000A on the proper color coded wattage range and record it if desired If the milliamp output rather than watts is required that reading may be obtained by reading the lower 0 1000 milliamp scale 7 Turn the ESU power to minimum and deactivate the output before disconnecting from ESU 2000A or before moving the Active Electrode test lead to a different input range connector ESU RF Leakage Current Measurements 1 Turn off ESU under test Remove all connections between the ESU under test and the ESU 2000A 2 Install the BC Biomedical Switching Module on the ESU 2000A front panel so that the words RF Leakage Testing are right side up When installed correctly the Active Electrode connectors should be inaccessible and the RF Leakage connectors should be exposed 3 Connect the ESU s Active Electrode to the ESU 2000A s red RF Leakage connector using a red test lead DO NOT MEASURE ACTIVE ELECTRODE LEAKAGE ON NON ISOLATED ESUs with the ESU 2000A Damage may occur to the ESU 2000A 4 Connect the ESU s case ground to the ESU 2000A s
6. green ESU Ground connector using a green test lead 5 Turn the controls on the ESU under test to minimum power then activate the CUT or COAG modes Turn the ESU power to the desired level Note the Active Electrode RF Leakage current reading on the ESU 2000A s red 0 500 milliamps RF Leakage scale and record it if desired 6 Turn the ESU power to minimum and deactivate the output 7 Move the red test lead to the Patient Plate connection on the ESU under test Leave the ground connection in place 8 Turn the controls on the ESU under test to minimum power then activate the CUT or COAG modes Turn the ESU power to the desired level Note the Patient Plate RF leakage current reading on the ESU 2000A s red 0 500 milliamps RF Leakage scale and record it if desired 9 Turn the ESU power to minimum and deactivate the output Disconnect all test leads Patient Plate Return Electrode Monitor Test 1 Turn off ESU under test Remove all connections between the ESU under test and the ESU 2000A 2 Use an actual return electrode monitor patient plate connected to the ESU under test Cut off the plate and expose the two wires 3 Connect both wires to the two ESU 2000A Patient Plate Return Electrode Monitor Test jacks in the upper right corner via test leads of any color with alligator clips 4 Apply power to the ESU under test and verify operation of its return electrode monitor mode as specified by the ESU manufacturer using vario
7. us Ohms settings on the ESU 2000A 5 Turn off power to the ESU under test and remove all connections MANUAL REVISIONS Revision Engineering Revisions Made Rev 01 DT7375 New Manual Format Rev 02 DT7375 Accessory Fuse Updated LIMITED WARRANTY WARRANTY BC GROUP INTERNATIONAL INC WARRANTS ITS NEW PRODUCTS TO BE FREE FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY ARE INTENDED THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF SHIPMENT EXCLUSIONS THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE BC GROUP INTERNATIONAL INC IS NOT LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES NO PERSON OTHER THAN AN OFFICER IS AUTHORIZED TO GIVE ANY OTHER WARRANTY OR ASSUME ANY LIABILITY REMEDIES THE PURCHASER S SOLE AND EXCLUSIVE REMEDY SHALL BE 1 THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS WITHOUT CHARGE 2 AT THE OPTION OF BC GROUP INTERNATIONAL INC THE REFUND OF THE PURCHASE PRICE SPECIFICATIONS Display Selectable Loads amp Ranges Watts Measurement Accuracy Frequency Response ESU Output Current Range RF Leakage Test Return Contact Monitor Test Minimum resistance at zero setting Oscilloscope Output Fuse Case Size amp Weight ESU 2000A 4 1 2 analog meter Meter accuracy is 2 0 of full scale