Manual: Hydro 2000 G/S user manual issue 1.0

image

Contents

1. Hydro 2000 G S l u u Hydro 20006 5 User Manual mAnO385 Issue 10 March 2007 Malvern Instruments Ltd 2007 Malvern Instruments makes every effort to ensure that this document is correct However due to Malvern Instruments policy of continual product development we are unable to guarantee the accuracy of this or any other document after the date of publication We therefore disclaim all liability for any changes errors or omissions after the date of publication No reproduction or transmission of any part of this publication is allowed without the express written permission of Malvern Instruments Ltd Head office Malvern Instruments Ltd Enigma Business Park Grovewood Road Malvern Worcestershire WR14 1XZ United Kingdom Tel 44 0 1684 892456 Fax 44 0 1684 892789 Windows 2000 and XP are registered trademarks of the Microsoft Corporation Tygon is a registered trademark of Norton Co Perlast is a registered trademark of Precision Polymer Engineering Printed in England Table of Contents Part 1 Operator s Guide Introduction to this manual Introducir ane Sg PA a ames 1 1 Access to the instrument 0 000 ees 1 2 Assumed information LL 1 3 Where to get help 1 3 Hardware features Introduction ar sui ware SON de ta Saw nn Ok ea a 2 1 What the dispersion unit does LL 2 1 How the dispersion unit Works 2 2 How the dispersion unit is controlled 2 3 Featu
2. Dispersion mechanisms Continuously variable and independent pump stirrer and ultrasonics Modes of operation Automatic via SOPs Manual via computer operating dialogues Weight 18 7kg Dimensions Width 348mm Height 333mm Depth 365mm Power 100 120V 200 240 V AC 50 60Hz 240VA Maximum size of particles Hydro 200064 S Up to 2000 microns depending on particle shape and density Page A 1 A Appendix A Specification Hydro 2000S Dispersion type Wet Capacity 50 150ml Typical applications Solvent based suspensions Pharmaceuticals Dispersion mechanisms Continuously variable combined pump stirrer Continuously variable ultrasonics Modes of operation Automatic via SOPs Manual via computer operating dialogues Weight 16 8kg Dimensions Width 348mm Height 333mm Depth 365mm Power 100 120V 200 240 V AC 50 60Hz 240VA Maximum size of particles Up to 600 microns depending on particle shape and density Purge air specification Hydro 2000G This section only applies to the Hydro 2000G Ifa particularly aggressive dispersant is used this may cause the grease in the pump motor bearings to leach out If this may be a problem the air purge inlet port can be used to connect an air supply This will keep fumes away from the bearings If required to use the air purge inlet connector see Chapter 2 check that the air supply conforms to the following specificat
3. Use the up and down arrows to set the threshold for the dispersant then click the Apply button The values of some commonly used dispersants are listed below m Water default 64 m White spirit 36 m Propan 2 ol 64 m 2 2 4 Trimethyl pentane 36 The level sensor threshold must always be set when a different dispersant is used Level sensor thresholds are also stored in the dispersant database Rear panel This illustration identifies the main features of the rear panel ill 4917 Drain and overflow The drain and overflow port is the exit point where sample dispersant leaves the dispersion unit A warning triangle warns that the contents of the tank are drained Page 2 14 MAM 0385 Hardware features Chapter 2 from this pipe The risk depends on the hazardous nature of the dispersants and samples being measured The drain pipe should not exceed 2m in length and should have no loops or kinks Dispersant input port The dispersant input is the inlet port which is connected to a clean source of dis persant A warning triangle warns that the dispersant may be hazardous though for most applications it will be water Tank fill rate This adjuster alters an internal pressure regulator that controls the rate at which the tank is filled with dispersant This is always set by the installation engineer but ifit s found that the dispersant pressure has changed one sign is that the tank takes a long tim
4. 2 8 Maximum size of particles A 1 A 2 Measurement delay 3 9 Menu commands 1 3 Model number 1 1 Multiple accessories 2 8 Multiple dispersion units 2 15 O Operator functions 1 2 P Particle size A 1 A 2 Pipe From cell 2 8 To cell 2 8 Page i Index Pipe grommets 2 10 Power requirements A 1 A 2 Pre measurement period box 3 6 Pump and stirrer settings 3 6 Pump control 3 3 Pump error 3 3 Pump speed changing 2 3 3 2 Pump stirrer 2 5 2 13 Purge air specification A 2 Purpose of accessory 2 1 R Rear panel details 2 7 2 14 overview 2 4 2 11 Regulatory statements C 1 S Sampler selection tab 3 5 Sampler settings tab 3 6 Software features 3 1 SOP control 3 5 Specification A 1 Status indicator overview 2 3 2 11 Status indicators 3 2 Stirrer control 3 2 Stirrer error 3 4 Stirrer settings 3 6 Stirrer speed changing 2 3 3 2 Supervisor functions 1 2 Pace ii Hydro 2000G S T Tank cleaning 3 3 emptying 3 3 filling 2 5 2 13 3 3 Tank area details 2 4 2 12 overview 2 3 2 11 Tank fill rate 2 8 2 15 Tank filling options 3 7 Termination plug 2 8 Tip displacement 3 7 To cell pipe 2 15 TO CELL port 2 2 U Ultrasonic power changing 2 3 3 2 Ultrasonic probe function 2 2 2 5 2 13 safety 2 5 2 12 Ultrasound error 3 4 selection 3 6 Ultrasound control 3 3 V Valves 2 2 device indicators 3 4 Vari flow 3 8 man 0385
5. The Liquid sensed indicator illuminates if the level sensor detects dispersant i e the tank is full If this indicator illuminates but the tank is still empty the level sen sor may need cleaning see Chapter 4 of the Essentials Manual MAM 0385 Software features Chapter 3 Writing an SOP for the dispersion unit Hydro 2O00G S An SOP can be configured to control all settings for the dispersion unit automati cally When an SOP is run the software will depending on how the SOP has been set up automatically fill the tank flush the tank to clean it run a de gas routine to remove bubbles from the dispersant and then set the pump stirrer and ultrasound to predefined settings The SOP will even report ifthe wrong cell or dispersion unit is fitted The SOP Wizard is described in full in the main User Manual There are ten dialogues in all but here we just describe those which have parameters specific to the Hydro 2000G S These are the Sampler Selection and Sampler Settings dialogues Sampler Selection dialogue The Sampler Selection dialogue is shown below Sampler Selection Sample Handling Unit TN The following sample handling units are available Select the one for which this SOP is being designed Sample handling unit Hydro 20006 4 e 0 lt Back Next gt Finish Cancel Help ill 7758 Select the dispersion unit to be used from the drop down list All dispersion units available will be
6. button will close the drain valve MAN 0385 Hardware features Chapter 2 Hydro 2O00G S Flow cell The flow cells for the Hydro 2000S and Hydro 2000G are identical except for internal identification encoding and the label badge This diagram shows the cell nt ill 4915 The flow cell passes the sample dispersant through the analyser beam of the optical bench so that it can be measured The cell windows are part of the optical path of the system and must be kept clean and free from scratches at all times The Essen tials Manual gives details on cleaning maintaining the cell Caution A The flow cell is an optical device Treat it as carefully as an expensive camera Scratches to the surfaces of the cell may affect performance Each cell is encoded so that as soon as it is loaded into the optical bench the system will detect and identify it The cell features are Cell shroud The cell shroud prevents stray light entering the measurement area that may affect a measurement More importantly the cell shroud stops human access to laser radiation Never use the system if the cell shroud is damaged Page 2 9 Chapter 2 Page 2 10 Hardware features Cell windows The cell windows allow the analyser beam of the optical bench to pass through the cell and hence the particle field The cell windows can be removed to allow clean ing the cell window tool used to remove the window is located within t
7. identifies the main features of the rear panel These are the following Drain The drain port is the exit point where sample dispersant leaves the dispersion unit A warning triangle warns that the contents of the tank are drained from this pipe The risk depends on the hazardous nature of the dispersants samples being meas ured The drain pipe should not exceed 2m in length must run downwards and should have no loops or kinks Dispersant input port The dispersant input is the inlet port which is connected to a clean source of dis persant A warning triangle warns that the dispersant may be hazardous though for most applications it will be water Air purge inlet If using a particularly aggressive dispersant this may cause the grease in the pump motor bearings to leach out If this may be a problem the air purge inlet port can be used to connect an air supply This will keep fumes away from the bearings The air supply must be maintained at all times when aggressive dispersants are used Appendix A gives the purge air supply specifications Page 2 7 Chapter 2 Page 2 8 Hardware features O Tank fill rate This adjuster alters an internal pressure regulator that will control the rate at which the tank is filled with dispersant This is always set by the installation engineer but if it s found that the dispersant pressure has changed one sign is that the tank takes a long time to fill alter this adjuste
8. the blood and its unknown effects never put fingers in the tank when the ultrasonic probe is in operation The ultrasonic probe is used to help disperse samples The power of the probe is controlled through the software Level sensor The level sensor automatically stops the tank being filled above a certain level If the tank fails to fill properly then the level sensor may require cleaning or the Page 2 5 Chapter 2 Page 2 6 Hardware features threshold settings may need adjusting The Essentials Manual has more details on cleaning the level sensor The level sensor must be set to detect different types of dispersant by setting the level sensor threshold This is done by selecting Configure Accessories The Level Sensor Threshold section is shown below Level Sensor Threshold Apply Use the up and down arrows to set the threshold for the dispersant used then click the Apply button The values of commonly used dispersants are listed below Water default 64 White spirit 36 Propan 2 ol 64 2 2 4 Trimethyl pentane 36 The level sensor threshold must always be set when a different dispersant is used Level sensor thresholds are also stored in the dispersant database Tank cover The tank cover is there to protect the operator from splashes from the tank Always replace the cover after adding sample to the tank MAM 0385 Hardware features Chapter 2 Hydro 2O00G S Rear panel This illustration
9. the command Configure New SOP refers to selecting the New SOP item in the Configure menu The same rules apply for sub menus of sub menus so that Tools Options Instrument Port refers to the Instrument Port item in the Options sub menu which itself is a sub menu of the Tools menu Menu commands are always shown in bold text Where to get help Hydro 2O00G S Full details on where and how to obtain help can be found in Chapter 1 of the main User Manual The Essentials Manual gives information on the following m Site requirements m Health and Safety m Maintenance m Installation in case the system has to be moved after its initial installation by Malvern Instruments personnel Page 1 3 Chapter 1 Introduction to this manual Page 1 4 MAM 0385 Hardware features Introduction This chapter describes the physical features of the dispersion unit It covers What the dispersion unit does How the dispersion unit works How the dispersion unit is controlled The physical features of the Hydro 2000G in detail The physical features of the Hydro 20008 in detail What the dispersion unit does Hydro 200064 S The sole purpose of any sample dispersion unit is to prepare the sample and then deliver it to the optical bench so that it can be measured The Hydro 2000G and Hydro 2000S allow the Mastersizer to be used for particle in liquid particle sizing Functionally the two dispersion units are virtually identi ca
10. ant slider This tab will show a single control for the pump stirrer if the Hydro 20008 is selected in the Sampler selection dialogue The Advanced button allows the vari flow options to be set these are described below Ultrasonics selection The ultrasonics selection radio buttons enable the ultrasonics The options are m None no ultrasound m Continuous from sample addition ultrasound is activated after sample is added and will run continuously mE Pre measurement activates ultrasound for a set time before measurement Type the required duration in seconds in the Pre measurement period box Page 3 6 MAM 0385 Software features Chapter 3 Hydro 20006G S m Continuous from start ultrasound is active throughout the measure ment Ultrasound will start after the electrical background is complete The stabilising period will add a delay between the electrical background and the optical alignment During this delay ultrasound will be operating allowing bubbles to be driven from the dispersant Tip displacement slider The tip displacement sets the power of the ultrasonics Moving the slider to 100 gives full power ultrasonic dispersion Pre Measurement Delay A pre measurement delay can be added before the measurement to allow thermal equilibration to occur after ultrasound has been applied If this is not done the anal ysis may show the presence of large particles which are not in the sample This is caus
11. d above the tank holds the sample and dispersant The pump and stirrer in the tank kecp the sample in suspension and continually circulate the sample and dispersant through the cell This diagram identifies the features of the tank area ill 4914 MAM 0385 Hardware features Chapter 2 Hydro 2O00G S These are the following Tank The tank can hold up to one litre of sample and dispersant for circulation through the cell of the optical bench Warning Due to the possible risk of sonication of the blood and its unknown effects never place fingers in the tank when the ultrasonic probe is in operation Also never put fingers in the tank when the pump stirrer shaft is rotating Sample and dispersant can be poured directly into the tank If this is done do not cause the tank to overflow by filling it too quickly Clean any spillage off the covers immediately to prevent damage Y Pump stirrer The stirrer located in the base of the tank agitates the sample dispersant mixture The pump circulates the sample through the flow cell located in the sample area of the optical bench The pump and stirrer speeds are controlled independently by the software Warning Never put fingers into the tank when the pump stirrer shaft is rotating Caution Never operate the pump stirrer at more than half speed with the tank empty Ultrasonic probe Warning A Due to the possible risk of sonication of
12. e to fill alter this adjuster so that a fill tank cycle takes about six seconds O From cell pipe This is the sample return port where the sample dispersant returns to the disper sion unit from the flow cell in the optical bench To cell pipe This is the sample out port where sample dispersant leaves the dispersion unit for the flow cell Mains power supply The mains power supply is a standard IEC socket that supplies mains power to the dispersion unit It also houses the dispersion unit s fuses and its on off switch Accessory comms in connector The communications cable from the optical bench or from another dispersion unit connects here Accessory comms out connector If more than one dispersion unit is connected to the system and this is the first dis persion unit in line then a communication cable will be connected from this con nector to the accessory comms in connector of the second dispersion unit The termination plug supplied with the optical bench must be fitted here if this is the only dispersion unit connected or it is the last dispersion unit in the line O Auxiliary connector This is not used Manual drain The manual drain button can be pressed to drain the tank Users may wish to do this if for example they need to drain the dispersion unit without having to turn on the control software Pressing and holding the button will open the drain valve Releasing the button will cl
13. ed Place of issue Name Function Malvern Instruments Limited UK Alan Bragginton Production Director ill 8007 Page C 2 MAM 0385 Regulatory statements Appendix C CE Declaration of Conformity AWA2001 The CE badge on this product signifies conformance to European Commission Directives Li CE Declaration of Conformity in accordance with ISO IEC 17050 1 Form 0734F CE Declaration of Conformity Date 15 02 2007 Issue 0001 Supplier details Document NAME NUMBER Issuer s Address ADDRESS OF MANUFACTURER Enigma Business Park Grovewood Road Malvern Malvern Instruments Limited Worcestershire WR14 1XZ UK CE declaration of conformity for AWA2001 2007 doc Issuer s Name NAME OF MANUFACTURER Product details Product description name and model number s Name Hydro 2000S Model AWA2001 Description Particle Dispersion Unit Directive s and Standard s to which conformity is declared EMC directive 89 336 EEC as amended by 92 31 EEC and 93 68 EEC Low Voltage Directive 2006 95 EC BS EN 50081 2 1994 BS EN 50082 2 1995 BS EN 61000 3 2 1995 CLASSA BS EN 61000 3 3 1995 BS EN 61010 1 2001 and part 2 as required Additional Information Declaration hereby declare that the product specified conforms Signature Date of issue with the provisions of Council Directive s and sta
14. ed by the refractive index differences observed in the sample cell when the dis persant is not equilibrated Choose a delay sufficiently long enough for adequate thermal equilibration to occur To set the delay highlight either the minutes or seconds in the Delay period box then use the selection arrows to alter the time alternatively the time can be typed in A delay of between 0 and 59 59 minutes can be chosen Tank Fill options The user can fill the tank with dispersant manually or automatically For the tank to be filled automatically the dispersant must be plumbed into the dispersant inlet port on the rear panel a solenoid valve then controls the flow into the tank If however a solvent is to be used as a dispersant it is unlikely that this will be plumbed into the dispersion unit Here select Manual This switches off the sole noid valve and the tank can be filled by hand This means that if the unit is con nected to a water supply the dispersant inlet does not have to be disconnected during a manual fill The software states when to fill the tank during an SOP measurement Page 3 7 Chapter 3 Software features Advanced software features Vari flow Some materials such as metal flakes may form structures when they are circulated through the flow cell at normal pump speeds These structures cause a secondary diffraction pattern that is incorrectly interpreted as a small peak at 1000 microns Vari flow accessed by
15. eed than that used for sample addition 0 RPM 2500 Reduced pump speed Q io RPM Dispersant equilibration O seconds Measurement delay lo gt seconds Page 3 8 MAM 0385 Software features Hydro 2O00G S Chapter 3 The pump speed falls to a predetermined reduced level Reduced pump speed in the dialogue this can be zero and an additional Measurement delay allows the background level to fall before the alignment is performed and the background measured This Optical equilibration period is the same as the measurement delay The pump speed is returned to the higher rate for the sample to be added and dispersed in the normal way Before the measurement stage is entered the pump speed drops to the lower level and the system waits for the duration of the Measurement delay before measuring the sample This delay allows the relaxation of the sample in the flow cell so thata normal measurement can be obtained If repeat measurements are required the pump will return to the high level for a brief re dispersion period the Measurement delay then drop down again for the measurement process to be repeated This will continue for as long as prescribed in the SOP Page 3 9 Chapter 3 Software features Page 3 10 MAM 0385 Part e Appendices Specification Hydro 2000G Dispersion type Wet Capacity 1 litre Typical applications Minerals fillers chemicals foodstuffs emulsions
16. ety of the disper sion unit and of its operation The supervisor is responsible for the training of the operators The supervisor can perform all user maintenance routines identified in Chapter 4 of the Essentials Manual including changing the fuses Warning A The supervisor operator must never remove the covers of the instrument or dispersion unit Removal of the covers by unauthorised personnel will invalidate the warranty of the dispersion unit Operator An operator is a person trained in the use of the dispersion unit The operator can perform all user maintenance routines identified in Chapter 4 of the Essentials Manual except for changing the fuses Warning A Failure to follow this guideline could result in exposure to hazardous volt ages Page 1 2 MAM 0385 Introduction to this manual Chapter 1 Assumed information All pictures and text show the Hydro 2000G unless information is specific to the Hydro 2000S Naming convention Within this manual m The Mastersizer 2000 optical bench is referred to as the optical bench or the instrument m The sample dispersion units are referred to in full e g the Hydro 2000G or as the dispersion unit m The combination of the optical bench one or more dispersion units and the computer is referred to as the system Menu commands Menu commands from the Malvern software are referred to in the form main menu menu item As an example
17. he cell holder The cell windows are delicate and should be treated with caution refer to the Essentials Manual for details on cleaning them Cell window seals The cell window seals prevent the windows from leaking Inspect them regularly especially if using aggressive solvents as a dispersant Details on changing the cell window seals can be found in the Essentials Manual Locking handle The cell locking handle must be rotated anti clockwise to lock the cell in place on the optical bench To remove the cell rotate the handle clockwise and lift the cell clear Warning Never try to lift the optical bench by its locking handle The Health and Safety section Essentials Manual gives correct moving techniques Drip tray Any small leaks will be captured within the drip tray and drained through a drain port in the base of the optical bench If any liquid is noticed in the drip tray remove the cell and tip the contents away Locate and fix any leaks Cell in connector The sample tube from the dispersion unit is connected here It is important that the sample flows up through the cell so any bubbles in the flow path can escape Cell out connector The sample return pipe from the cell to the dispersion unit is connected here Pipe grommets The pipe grommets prevent light entering the measurement area These must be replaced if they are damaged in any way MAM 0385 Hardware features Chapter 2 Feat
18. ion m 1800ml air min at 212mbar g m 0 064 cu ft min at 3 0psi g m Filtered to remove all particles larger than 0 5 microns if the dispersion unit is to be used in unclean environments Page A 2 MAN 0385 Chemical compatibility Introduction The Hydro 2000G and Hydro 2000S are manufactured from materials that give the widest protection from chemical attack However it is important to check that any sample or dispersant used is chemically compatible with the materials that they will come into contact with within the dispersion unit This appendix lists all materials that come into contact with the sample and disper sant in normal operation Components in contact with sample dispersant Component Materials Tank and pump cavity 316 stainless steel Viton Internal pipes 316 stainless steel and F E P Level sensor P E E K Pump stirrer 316 stainless steel and P T F E Ultrasonic probe Titanium Drain valve 316 stainless steel P T F E Perlast and Viton Sample pipes to cell Tygon Drain pipe PVC Cell Glass 316 stainless steel Viton Splash guard Hydro 2000G only Acrylic Note For the Hydro 2000S Viton seals can be upgraded to Perlast to improve the chemical compatibility Contact the Malvern representative for details Hydro 2000G S5 Page B 1 Appendix B Chemical compatibility Page B 2 MAM 0385 Regulatory statements This section presents the CE Decla
19. l The speed of the stirrer is altered by moving the slider Use the up and down arrows under the slider for fine adjustment of the stirrer speed Page 3 2 MAM 0385 Software features Chapter 3 Hydro 2O00G S Note For the Hydro 2000S the Pump and Stirrer slide bars are combined into a single slider Pump control The speed of the pump is altered by moving the slider Use the up and down arrows under the slider for fine adjustment of the pump speed Ultrasound control The power of the ultrasonic probe is altered by moving the slider Use the up and down arrows under the slider for fine adjustment of the ultrasonic power Tank cleaning Pressing the Clean button enables a clean sequence If the sample dispersant in use is not normally removed with a single flush enter a number in the Number of Cycles box to cycle through the flush sequence this number of times Tank filling Pressing the Fill button opens the dispersant valve and fill the tank The dispersant valve automatically closes when the level sensor detects liquid Selecting the Degas at end of fill clean sequence check box enables the Degas facility outlined in O below after the tank has filled Tank emptying Pressing the Empty button opens the drain valve and empty the tank The Man ual Drain button on the rear panel of the dispersion unit has the same function Degassing Degassing helps to remove dissolved gasses and bubbles from the dis
20. l except that the Hydro 2000G can circulate 800ml of liquid while the Hydro 2000S has a smaller capacity of 150ml The small volume of the Hydro 2000S is ideal when using solvent dispersants or when samples and dispersants are either expensive or hazardous The materials used in the manufacture of the dispersion units maximise the range of samples which can be handled Variable pump stirrer speeds allow a wide range of particle sizes and densities to be suspended and circulated whilst a variable power ultrasonic system enables particle agglomerates to be dispersed Page 2 1 Chapter 2 Hardware features How the dispersion unit works This is the Hydro 2000G e80 00 ill 4911 The dispersion unit comprises a tank that holds one litre of sample dispersant liquid A stirrer O controlled from the software agitates the sample and stops it from settling or separating out The pump controlled from the software forces the sample from the TO CELL port of the dispersion unit to the flow cell located in the optical bench via the sample tubing The pump and stirrer can be independently controlled The sample is pumped through the flow cell and then returns to the tank via the sample tubing and the FROM CELL port An ultrasonic probe is provided to aid sample dispersion and can also be used to remove bubbles from the flow path The dispersion unit contains two fluid handling valves controlled from the sof
21. listed even those that are not attached to the system This enables an SOP to be written remotely from the system The system allows up to two of each dispersion unit to be connected to the optical bench In this situation the dispersion units will be identified as unit A and B for example there might be an entry for Hydro 2000G A and Hydro 2000G B If using just one dispersion unit select the A option Page 3 5 Chapter 3 Software features All systems the SOP is distributed to must have the dispersion unit connected Sampler Settings dialogue The Sampler Settings dialogue controls the pump stirrer and ultrasonic settings for the dispersion unit This dialogue changes depending on the dispersion unit selected in the Sampler Selection dialogue This is the dialogue for the Hydro 2000G Sampler Settings Set the pump amp stirrer speed ultrasonics amp type of tank fill to suit your sample Pump and Stirrer D 0 RPM 2500 Pump Speed i J E 850 RPM 0 RPM 1000 Stirer Speed 4 850 RPM Ultrasonics None O Pre measurement Continuous Continuous from start fi from sample e addition trai Pre Measurement Delay 14 Delay period 0 00 mm ss O Tank Fill Automatic O Manual lt Back Next gt Finish Cancel Help ill 7759 Its components are Pump and stirrer settings The speed of the pump and stirrer is altered by moving the relev
22. n unit is and explain in simple terms how it works m Identify the physical features of the dispersion unit m Describe the software and explain how to use the dispersion unit to make a measurement on the system Warning A The dispersion unit or the samples to be measured may be hazardous if misused Users must read the Health and Safety information in the Essentials Manual before operating the system We recommend that users who have never operated a Malvern particle analyser before read this manual fully before starting the first measurement Page 1 1 Chapter 1 Introduction to this manual Those who are more familiar with particle size analysers may wish to jump straight to Chapter 4 ofthe main User Manual which gives details on making measure ments However the importance of sample preparation before measurement can not be overstated so we recommend reading the chapter on sample preparation Chapter 8 of the main User Manual as a priority Access to the instrument Within this manual reference is made to the various people who will have access to the dispersion unit Below is a list of these people and their responsibilities Malvern personnel Malvern personnel service engineers representatives etc have full access to the dispersion unit and are authorised to perform all service procedures that may require the removal of the covers Supervisor The supervisor is the person responsible for the management saf
23. ndard s listed above Signed for and the behalf of Malvern Instruments Limited Place of issue Name Function Malvern Instruments Limited UK Alan Bragginton Production Director ill 8006 Hydro 2000C S Page C 3 Appendix C Regulatory statements Page C 4 MAM 0385 Index A Accessory chaining 2 15 Accessory control dialogue 3 2 Air purge inlet 2 7 Air supply purge A 2 Cc Capacities 2 1 CE Declaration C 2 Cell holder Overview 2 4 2 12 Cell in connector 2 10 Cell out connector 2 10 Cell shroud 2 9 Cell window seals 2 10 Cell windows purpose 2 10 Chaining dispersion units 2 8 Chemical compatibility B 1 Cleaning software sequence 3 3 Communications cable 2 8 2 15 Controlling the accessory manually 3 2 D Degassing dispersant 3 3 Device indicators 3 4 Dispersant input 2 15 Dispersant input port 2 7 Dispersant valve 3 4 Dispersion unit chaining 2 8 Drain 2 7 Drain and overflow 2 14 Drain valve 3 4 Drain valve error 3 3 Drip tray 2 10 E Error indicators 3 3 Hydro 2O00G S F Features overview 2 2 Flow cell details 2 9 Hydro 2000S 2 16 Fluid handling valves 2 2 From cell pipe 2 15 FROM CELL port 2 2 H Help 1 3 Hydro 2000G features 2 3 Hydro 2000S features 2 11 L Level sensor function 2 5 2 13 Locking handle 2 10 Mains power supply 2 8 2 15 Making a measurement 3 1 Manual control 3 2 Manual drain 2 15 Manual drain button
24. ose the drain valve Hydro 2000C S Page 2 15 Chapter 2 Hardware features Hydro 2000S flow cell The flow cells for the Hydro 2000S and Hydro 2000G are identical except for internal identification encoding and the label badge Refer to the Flow Cell section for the Hydro 2000G above to identify the features ofthe cell Page 2 16 MAM 0385 Software features Introduction This chapter describes the features of the software which are specific to the disper sion unit It covers m Making a measurement the basics of making a measurement using the dis persion unit m Manually controlling the dispersion unit m Writing a Standard Operating Procedure SOP for the dispersion unit m Advanced software the vari flow facility Making a measurement Making a measurement using the Hydro 2000S and Hydro 2000G is fully docu mented in Chapter 4 of the Mastersizer 2000 User Manual Refer to that man ual for details The dispersion unit can either be controlled manually or automatically by running an SOP The system automatically detects which dispersion unit and cell is con nected If more than one dispersion unit is connected the system detects all disper sion units connected but only the dispersion unit that has its cell installed on the optical bench will be active Hydro 2000G S Page 3 1 Chapter 3 Software features Manually controlling the dispersion unit To control the dispersion unit manuall
25. persant Press ing the De Gas button pulses the ultrasonic probe for the amount of time speci fied in the adjacent seconds box Error indicators The following indicators illuminate if an error condition is detected Pump Error illuminates if the pump fails to reach the requested speed As an example this may be due to the sample dispersant being particularly viscous Drain Valve Error illuminates if the drain valve has failed to drive i e it is stuck open or closed This may be caused by leaving a sample in the tank for long periods this may set and bond the drain valve Page 3 3 Chapter 3 Page 3 4 Software features Caution Never leave sample in the tank for long periods always flush the tank and leave it filled with clean dispersant Contact the Malvern representative if the Drain Valve Error indicator illuminates E Stirrer Error illuminates if the stirrer fails to reach the requested speed Again this may be due to the sample dispersant being particularly viscous Ultrasonics Error illuminates if the ultrasonics fail to reach the requested power Contact the Malvern representative for assistance Device indicators The device indicators give the status of the drain and dispersant valves Drain Valve open tank empties m Drain Valve closed tank remains full m Dispersant Valve open tank fills with dispersant m Dispersant Valve closed tank stops filling with dispersant
26. r so that a fill tank cycle takes about 20 seconds From cell pipe This is the sample return port where the sample dispersant returns to the disper sion unit from the flow cell in the optical bench To cell pipe This is the sample out port where sample dispersant leaves the dispersion unit for the flow cell Mains power supply The mains power supply is a standard IEC socket that supplies mains power to the dispersion unit It also houses the dispersion unit s fuses and its on off switch Accessory comms in connector The communications cable from the optical bench or from another dispersion unit connects here Accessory comms out connector If more than one dispersion unit is connected to the system and this is the first dis persion unit in line then a communication cable will be connected from this con nector to the accessory comms in connector of the second dispersion unit The termination plug supplied with the optical bench must be fitted here if this is the only dispersion unit connected or it is the last dispersion unit in the line Auxiliary connector Not used Manual drain The manual drain button can be pressed to drain the tank the dispersion unit must be powered up first Users may wish to do this if for example they need to drain the dispersion unit without having to turn on the control software Pressing and holding the button will open the drain valve Releasing the
27. rations of Conformity for the products Hydro 2000C S Page C 1 Appendix C Regulatory statements CE Declaration of Conformity AWA2000 The CE badge on this product signifies conformance to European Commission Directives CE Declaration of Conformity in accordance with ISO IEC 17050 1 Date 15 02 2007 Issue 0001 nee Form 0734F CE Declaration of Conformity Supplier details Document NAME NUMBER Issuer s Address ADDRESS OF MANUFACTURER CE declaration of conformity for AVVA2000 2007 doc Enigma Business Park Grovewood Road Issuer s Name NAME OF MANUFACTURER Malvern Malvern Instruments Limited Worcestershire WR14 1XZ UK Product details Product description name and model number s Name Hydro 2000G Model AWA2000 Description Particle Dispersion Unit Directive s and Standard s to which conformity is declared EMC directive 89 336 EEC as amended by 92 31 EEC and 93 68 EEC Low Voltage Directive 2006 95 EC BS EN 50081 2 1994 BS EN 50082 2 1995 BS EN 61000 3 2 1995 CLASS A BS EN 61000 3 3 1995 BS EN 61010 1 2001 and part 2 as required Additional Information Declaration hereby declare that the product specified conforms with the provisions of Council Directive s and standard s listed above Signature Date of issue Signed for and the behalf of Malvern Instruments Limit
28. res of the Hydro 2000G 1 ec ee 2 3 Features of the Hydro 20008 2 11 Software features Introductioni sv ie tiie ie wel E 3 1 Making a measurement 3 1 Manually controlling the dispersion UNIt LL 3 2 Writing an SOP for the dispersion unit LL 3 5 Advanced software features Vari flOW o ooooooommmoom o oo o 3 8 Hydro 2000C S Page i Table of Contents Hydro 2000G S5 Part 2 Appendices Specification Hydro 20006 sis ab tetra hg we ed an Bl ware A 1 Hydr6 20008S vadis dara en pe ani a A 2 Purge air specification Hydro 2000G LL A 2 Chemical compatibility Introduction wi ei aca a ann B 1 Regulatory statements CE Declaration of Conformity AWA2000 LL C 2 CE Declaration of Conformity AWA2001 saasaa aana C 3 Index Pace ii MAM 0385 Part 1 Operators Guide Introduction to this manual Introduction Hydro 200064 S This manual describes the operation of the large volume Hydro 2000G and small volume Hydro 2000S sample dispersion units Instrument Model number Hydro 2000G AWA2000 Hydro 2000S AWA2001 This manual is a supplement to the following manuals m Mastersizer 2000 User Manual termed the main User Manual m Mastersizer 2000 Essentials Manual termed the Essentials Manual This dispersion unit manual focuses on some specific issues of the dispersion unit that are not covered by the above manuals It aims to m Describe what the dispersio
29. selecting the Advanced button in the Sampler Settings dialogue allows these materials to be dispersed in the tank in the normal way The pump is then briefly turned off or its speed reduced before measurement of the sample When the pump is turned off the structures in the sample are randomised and the measurement can be made without the occurrence of the secondary dif fraction pattern When the flake material is suspended in a volatile solvent such as white spirit vari flow is even more useful since it assists the rapid thermal stabilisation of the disper sant When the sample tank is filled with fresh solvent very high backgrounds are often observed These arise from thermal gradients in the solvent causing steering ofthe laser beam onto the inner rings of the detector array As the temperature equilibrates the background falls to normal levels 80 100 and measurement can proceed Pausing the pump after the solvent has had a chance to thoroughly dis perse through the system will accelerate the fall of the background to normal levels What is the sequence of events The Vari flow cycle is as follows 1 The pump is operated at normal speed set in the SOP for an initial period to allow the dispersant to be thoroughly mixed and become thermally homoge nous This initial period the Dispersant equilibration period can be set in this dialogue Advanced Vari flow Yari flow Allows measurements to be performed at a different pump sp
30. t ware m A motor driven drain valve that allows the tank to be drained m A dispersant valve that opens to allow clean dispersant into the tank The operation of the Hydro 2000S is virtually identical except that it has a com bined pump stirrer mechanism and a smaller 150ml tank capacity Page 2 2 MAM 0385 Hardware features Chapter 2 How the dispersion unit is controlled The dispersion unit is controlled using a single software dialogue with sliders for the pump speed stirrer speed and ultrasonic power Chapter 3 has details When controlled through a Standard Operating Procedure SOP the software tells the user what value to set for cach parameter Features of the Hydro 2000G Hydro 2O00G S This illustration identifies the main features of the dispersion unit u ill 4912 These are the following Tank area The tank holds the sample and dispersant The pump and stirrer in the tank kecp the sample in suspension and continually circulate the sample and dispersant through the flow cell More details on the tank area are given below An ultrasonic probe also helps to disperse the sample All functions of the tank arca are controlled by the Malvern software Status indicator The status indicator illuminates in one of three colours when the dispersion unit is powered up MW green if the dispersion unit is functioning correctly and its cell has been loaded into the optical bench i e the dispersion uni
31. t is active Page 2 3 Chapter 2 Page 2 4 Hardware features m amber if the dispersion unit is functioning correctly but its cell has not been loaded into the optical bench i e the dispersion unit is at standby m red if the dispersion unit detects an error If an error message does not appear on the screen selecting Configure Accessory displays the accessory control dialogue which should show the error Rear panel The rear panel contains all services and communication connectors for the disper sion unit More details on the rear panel are given later in this chapter Cell holder The Mastersizer 2000 system is designed so that more than one dispersion unit can be connected at once The cell holder provides a convenient storage location for the flow cell if another dispersion unit is in use The cell holder reduces the build up of dust on the cell windows when not in use If the cell is not used for short periods up to 24 hours leave the cell and tank full of clean dispersant so that the cell windows do not dry out leaving smears or water marks on the window surface If the push on connectors are to be removed the cell may be stored wet by using the syringe supplied to fill it with clean dispersant If the cell is not to be used for longer periods remove and dry the cell windows the Essentials Manual has details The cell holder houses the window removing tool in its base Tank area As describe
32. to the tank If doing so do not overflow the tank by filling it too quickly Do not tip a sample onto the lip of the tank as fine particles may settle out and bias the measurement Y Pump stirrer The stirrer located in the base of the tank agitates the sample dispersant mixture in the tank The pump circulates the sample through the flow cell The pump and stirrer speeds are controlled independently through the software Warning Never put fingers in the tank when the pump stirrer shaft is rotating Caution Never operate the pump stirrer at more than half speed with the tank empty Ultrasonic probe The ultrasonic probe is used to help disperse samples Its power is controlled through the software Warning A Due to the possible risk of sonication of the blood and its unknown effects never put fingers in the tank when the ultrasonic probe is in operation Level sensor The level sensor automatically stops the tank being filled beyond a certain level If the tank fails to fill properly the level sensor may require cleaning or the threshold settings may need adjusting The Essentials Manual shows how to clean the level sensor The level sensor must be set to detect different types of dispersant by setting the level sensor threshold This is done by selecting Configure Accessories The Level Sensor Threshold section is shown below Page 2 13 Chapter 2 Hardware features Level Sensor Threshold Apply
33. ures of the Hydro 2000S This illustration identifies the main features of the dispersion unit ill 4916 Tank area The tank holds the sample and dispersant A combined pump stirrer within the tank keeps the sample in suspension and continually circulates the sample and dis persant through the flow cell More details on the tank area are given below An ultrasonic probe also helps to disperse the sample All functions of the tank arca are controlled by the Malvern software Status indicator The status indicator illuminates in one of three colours at power up m Green if the dispersion unit is functioning correctly and its cell has been loaded into the optical bench i e the dispersion unit is active Amber if the dispersion unit is functioning correctly but its cell has not been loaded into the optical bench i e the dispersion unit is at standby m Red if the dispersion unit has detected an error If an error message does not appear on the screen selecting Configure Accessory will display the accessory control dialogue which should show the error Rear panel The rear panel contains all services and communication connectors for the disper sion unit The rear panel is described in detail below Hydro 2000G S Page 2 1 Chapter 2 Page 2 12 Hardware features Cell holder The Mastersizer system is designed so that more than one dispersion unit can be connected at once The cell holder pro
34. vides a convenient storage location for the flow cell if another dispersion unit is in use The cell holder reduces the build up of dust on the cell windows when not in use If the cell is not used for short periods up to 24 hours leave the cell and tank full of clean dispersant so that the cell windows do not dry out leaving smears or water marks on the window surface If the push on connectors are to be removed the cell may be stored wet by using the syringe supplied to fill it with clean dispersant If the cell is not to be used for longer periods remove and dry the cell windows refer to the Essentials Manual for details The cell holder houses the window removing tool in its base Tank area As described above the tank holds the sample and dispersant The pump stirrer in the tank keeps the sample in suspension and continually circulates the sample and dispersant through the cell Warning A Due to the possible risk of sonication of the blood and its unknown effects never put fingers in the tank when the ultrasonic probe is in operation Also never put fingers into the tank when the pump stirrer shaft is rotating This diagram identifies the features of the tank area ill 4918 MAM 0385 Hardware features Chapter 2 Hydro 2O00G S Tank The tank can hold up to 150ml of sample and dispersant for circulation through the cell of the optical bench Sample and dispersant can be directly poured in
35. y either select Configure Accessory or when making a manual measurement click the Accessory button in the Meas urement Display window The accessory control dialogue shown below appears Hydro 2000G Controls Stirrer Pump Ultrasound O 1000 2500 100 500 1250 50 00 i nn 0 CJ rpm D pmp zp 2 Tank Functions D Clean _ 27 je E F Fill Empty Oo O O gt De Gas Cancel Operation 1 Number of Cycles O Degas at end of fill or clean sequence Level Sensor Threshold Apply Error Indicators Pump Error Drain Valve Error Stirrer Error Ultrasound Error Device Indicators y oe Open Closed Drain Valve E Dispersant Valve Liquid sensed Manual Operations Dispersant Valve ill 7757 The contents of this dialogue change depending on the dispersion unit connected It allows the user to alter the pump speed stirrer speed and ultrasonic power by moving the relevant Controls slider bar This dialogue also allows the user to fill and drain the tank by pressing the relevant Tank Functions buttons Status indicators give information such as whether the drain valve is open closed and whether the level sensor has been triggered the tank is full as well as reporting any errors within the dispersion unit The features change depending on which dispersion unit is connected but are Stirrer contro