Home
User`s Manual - Infiniti Medical
Contents
1. Curato ap INFINITI Curao System Users Manual D CAUTION This Curato Wound Treatment System User s Manual is not a guarant warranty It is intended only as an operational guide For additional information and questions please contact Infiniti Medicals Customer Service department at 1 650 327 5000 In order for the Curato Wound Treatment System to provide safe reliable and propor performance the folowing conditions must be met Failure to comply with these conditions will vid pertinent warranties There are no user serviceable components in the Curato All assembly operation adjustment modification maintenance and or repair must be carried out only qualified personnel authorized by Infiniti Medical The electrical installation of the room in which the device will be used must comply withthe appropriate national electrical standards The product must be used in accordance with this manual and all associated labeling and the Instructions for Use Any device that does not function as expected must be returned to Infiniti Medical Notice to Users CAUTION Federal law restricts this device to s by or on the order of a veterinarian As with any prescription medical device failure to follow product instructions or changing settings and performing therapy applications without the express direction and or supervision of a trained clinical caregiver may lead
2. Prepare area around wound to permit adhesion of the polyurethane drape NOTE I poriwound area is excessively moist ar oly a medical grade liquid adhesive may improve sealing For frage skin use skin sealant prior to drape application or frame the wound with skin barrier layer Cut the drape to size large enough to cover the foam and the barrier layer only Take measurements of the wound dimensions and note wound type Select the appropriate oam based on wound assessment Cut tho foam dressing to a size that is appropiate for the wound NOTE Do not tim the foam dressing over or around the wound site to help prevent debris from foam dressing from fling into the wound 7 Place the foam dressing the wound site taking caro to avoid contact with the per wound skin WARNING Do not pack the foam dressings into any areas of the wound Forcing dressings in a compressed manner into any wound is contrary to approved protocols Loosely fill visible and invisible dead space in the wound INFINITI MEDICAL Curato System Users Manual NOTE The foam dressing should cover the entire wound margin including tunneling and undermining However the foam dressing should not be in contact with intact skin Size and trim the polyurethane drape to cover dressing plus 3 5 em border of intact skin extra piecas of drape can be used to seal dressing leaks Remove the drapes release liner and place over he fo
3. to improper product performance and the potential for serious or fatal injury z INFI m Curao System Users Manual Introduction Indications Contancatons Precautions Astana Precautions Safety Tos Festes 2 Care amp Cleaning Equipment Disposal Cleaning the Curato Device 3 Patient Care Acting the Dressing Canister nstalaton Canister Removal Dressing Removal Disposal of Dressings and Canister Device Renewal 4 Operating Instructions Power ONOFF PowerUp Procedure Therapy Seting Adjustment Beeper Volume Adjustment Banery Operation Operation Therapy Selection LockUniock 5 Specifications Symbas Replacement Pars 7 Questions amp information INFI 10 0 2 2 15 15 6 Curato System Users Manual E Indications The Curato Wound Treatment System is indicated for the application ol suction negative pressure ta wounds to promote wound healing and for the removal af 105 including wound exudates irrigation fuids body fluids and infectious materials Contraindications The Curato Wound Treatment System is contraindicated for patients with malignancy in the wound Untreated osteomyelitis on antari and unexplored fistulas or necrotic tissue wih aschar present Do mot place the foam dressing over exposed biood vessels or organs Precautions Precautions should be taken for patients with active bleeding d
4. Conditions Temperature Range 120 to 43 C Relative Humidity Range 20 95 Non condensing Atmospheric Pressure Range 50KPa to 110 kPa Operating Conditions Temperature Range AF 22 Relative Humidity Range 20 75 Non condensing Atmospheric Pressure Range S0kPato 110 kPa Service of Curato Wound Treatment System 3 years CAUTION Federal law restricts this device to sale by or on the order of veterinarian INFINITI MEDICAL Curato System Users Manual Explanation of Symbols 4A gt 60 6 gt lt Curao System Users Manual 6 Description Part Number Therapy Unit Wound Treatment System User s Manual Power Supply A C Power Adapter AND Accessories LV Pole or cago adaptor Carrying Bag NOTE Part numbers for Canisters Dressings and Disposable Accessories may bo oblained by visting Infiniti Medical website waw infinitimedical com NOTE in order to assure the highest safety quality and effcacy of the products the Curato Wound Treatment System should only be used with Infiniti Medical s disposables and Infiniti Medical Dressings should only be used with the Curato Wound Treatment System INFIN ME Curato System User For addtional information pertaining to the Curato Wound Treat
5. I MEDICAL Curao System Users Manual 10 4 Plug the device s A C Adapter into a suitable 100 240 VAC 50 60Hz outlet Insert the power plug into Powar Jack on the side of the device rafer o Figure 5 The Curato Device should only be used withthe supplied A C Adapter Use of an incorrectly rated adapter could create shock hazard for the patient and caregiver NOTE infiniti Medical offers 12 Voit vehicle adapter that allows the device to operate on extemal power while traveling 5 Verily tha dressing application corect the tubing is connected and the suction tubing with SlekConnact clamp is open 6 Begin therapy see OPERATING INSTRUCTIONS Canister Removal 1 Press the OFF button to turn the therapy of 2 Close sucton tubing clamp 3 Remove tubing connector from top of canister 4 Press canister release button and withdraw canister from bottom of uni 5 Dispose of canister according to local state and federal regulations as wall as institutional protocols Dressing Removal Carefuty read the SAFETY TIPS in the INTRODUCTION section of this guide prior to removing the dressing NOTE Wounds must be carefuly monitorod at regular intervals In a non infected wound dressings should be changed every 48 o 72 hours but no lass than 3 times per weak with the frequency of dressing change determined by he clinician Infected wounds must be monitored continuously For infected wounds dressings
6. am dressing and per wound 9 Pat drape material down around the wound site and aver the foam dressing to ensure dressing is property sealed 10 Cut a 1 diameter hale in the top of the drape at convenient location aver the dressing soe Fe 41 Peal tho backing from one of the SlckConnect fanges and place i above the hole made in Stop 10 Using the tips of the fingers press around tho top o the SlickConnect to ensure good seal to the dressing see Fg 2 To Fluid Colection Canister Foam Dressing Polyuretnane Drape Figure Flgure2 Canister Installation 1 Ensure that canister is propery inserted in the receptacle located on the back side of the therapy unit sae Fig 3 The canister should snap into place and lock The canister release button may need a be depressed to permit canister insertion NOTE Aways use a new canister with a new patient Curao System Users Manual SckConaset Figures Figure a 2 Inspect tno SickConnect Tube fanges to onsure that they are properly connected to the dressing and that the connections are wall sealed 3 Connect the distal end of the SlickConnect Tube with the blue tapered connector to the patient port the Canister see Figure 4 Gently twist and push the connector on just enough to secure and seal it Also make sura that the clamp on the SlckConnect Tube is open see Figure 5 care INFINIT
7. ato Wound Treatment device should only be used with the supplied A C Adaptor Use of an incorrectly rated adapter could create a shock hazard for the patient or caregiver WARNING Avoid spilling liquid on any part of the therapy unit Liquids can cause corrosion when left on electronic controls which can lead to failure Component failure may cause the therapy unit to operate erratically possibly causing a potential hazard to the patient or Caregiver WARNING Particular care must be taken when handling undiluted germicide concentrate or chlorine bleach including proper shielding of eyes Always mix by adding concentrated germicide or chlorine bleach to the water NEVER intermix germicides or mix germicides with chlorine bleach Curato System Users Manual 7 Patient It is recommended that all sections of this manual be reviewed prior to using the product Carefully read the INDICATIONS CONTRAINDICATIONS PRECAUTIONS and SAFETY TIPS in the INTRODUCTION section before attempting to perform patient car for a patient with the Curato Wound Treatment System Applying the Dressing 1 Cleansa the wound according to institutional protocols or vetennarians order 2 Debride ali necrotic tissue including eschar and hardened slough Be certain the wound has achieved hemostasis 4 Visually examine and palpate wound bed to locate any blood vessels or delicate underlying structure In close proximity 5
8. e A C Adapter into the power receptacle on the side of the therapy uni Plug the A C Adapter into suitable 120 VAC 60 Hz wall oul When the device is connected to an AC power source the green power LED on the front of the device wil iluminate indicating AC power is present and the amber charging LED located just below the power LED wil iluminate when the battery is charging Once the battery is fully charged the amber LED wil extinguish indicating the charge cycle is complete When the Curato Wound Treatment System is disconnected from the AC power source he device Wil automaticaly swich over to the internal battery and continue to operate without interruption Curato System Users Manual se Alarm Operation Clearing an Alarm Condition clear an alarm condition tum the therapy unit OFF then ON Tho alarm will clear when the power is cyce Alarm Troubleshooting NOTE Pressing the ON MUTE button after an alarm wil silence the beeper for 5 minutes NOTE In the event of an emergency please contact the treating veterinarian NOTE If an Alarm Condition persists and cannot be resolved please contact Infiniti Medical for further assistance Curao System Users Manual Therapy Selection Lock Unlock The Curato Wound Treatment System is equipped with therapy locking feature designed to provenit unauthorized individuais from changing the therapeutic setting
9. e dressing integrity i verified adjust he unit for desired therapy NOTE The device must be connected to the A C Adapter while attempting to obtain an initial dressing seal 5 check dressing for vacuum leaks and repair with additonal polyurethane drape necessary The Curato Wound Treatment System should be operated at least 22 hours out of every 24 hour period Remove the dressing if therapy is terminated or is aff for more than 2 hours in 24 hour period INFIN Curao System Users Manual Therapy Setting Adjustment CAUTION Only a veterinarian can prescribe the proper settings and protocols for the therapy unit Failure to follow product instructions or adjusting settings and performing therapy application without the express direction andor supervision of a veterinarian may lead to improper product performance and the potential for serious or fatal injury Negative Pressure Level Adjustment There ae five negative pressure settings that can be selected 50 mmHg 75 mmHg 100 mmHg 125 mmHg and 150 mmHg The pressure selection buttons are lacatad on the left side ofthe control panal The bution decreases the negative pressure setting and the button increases the negative pressura setting 1 When the unit is powered up the current setting is selected automatically unless therapy setting has been locked previously by caregiver see Therapy Selection Lock Unlock Section 2 To change the
10. e volume The LED display wil indicate the volume level Battery Operation NOTE The Curato Wound Treatment System is designed to permit use of the product while the intemal battery is charging The therapy unit wil continue to operate propery while the battery is charging Battery Life The specified battery of the Curato Wound Treatment System with a ful charged battery and a wol sealed crossing is up to 24 hours The actual fe is dependent on the integrity of the dressing A leak in the dressing can reduce overall battery longevity significantly Average Time for Recharging To ensure the battery has been fully charged the device should be connected an A C supply for approximately 3 hours Alter approximately 2 hours of charging the device wil have achieved 80 of total battery capacity Low Battery Alarm While running on battery a low battery alarm wil activate when remaining capacity of the battery is lass than 20 Soo Alarm Operation Typically the unit wil continue to operate between 30 minutes and 1 hour after the low battory alarm s activated Low Battery Shutoff Ifthe battery charge falls below a functional tevel the device wil shutoff automatically and therapy will be discontinued At this point the device must be plugged into an A C power source for therapy resume Once the A C Adapter is plugged in pressing the ON button wil restart the device Recharging the Battery Plug the power cord from th
11. erence All components ofthe Curato Wound Treatment System disposable set are latex frae sure to comply with other CONTRAINDICATIONS and PRECAUTIONS for the Curato Wound Treatment System Signs of possible infection may include fever tenderness redness swelling increased warmth in the wound area purulent discharge a strong odor Vomiting diarhea high fever relractory hypotension be added signs of more serious complications of infection WARNING Do not pack the foam dressings into any areas of the wound Forcing dressings in a compressed manner into any wound is contrary to approved protocols INFINI AED z Curato System Users Manual 4 Features Easy to use One Touch Operation Therapy activation and change of pressure settings can be accomplishod with the push of a button Therapy settings can be locked by the caregiver see Therapy Selection Lock Unlock Lighted LEDs clearly indicate current therapeutic settings Renewable Device The Curato is the worlds frst renewable NPWT device The therapy unit incorporates service timer that wil indicate when to return the device for renewal Light Weightimpact Resistant The Curato device weighs only 15 cz 0 43 and can be easly and transported The polymer enclosure is impact resistant to help prevent damage from sopping Sel4imiting Pump Tha pump is designed to mechanically selt imit the amount
12. herapy the Curato Suction Pump and dressing are a closed system and are NOT vented to atmosphere During Therapy whan a canister fils with fuid it should be replaced immediately as fluids such as wound exudate wil not be remaved from the dressing once the canister is ful INFINITI MEDICAL Curao System Users Manual 2 Safety Tips The Curato Wound Treatment System is indicated for the application of suction negative pressure to wounds to promete wound healing and for the removal of uds including wound exudates irigation ids body fluids and infectious materials KEEP THERAPY ON The Curato Wound Treatment System should bo operated at least 22 hours of every 24 hour period Remove the foam dressing if therapy is terminated ris off for more than 2 hours in a 24 hour period DRESSING CHANGES Clean the wound per veterinarian order prior to dressing application Routine dressing changes shouid occur every 48 to 72 hours Dressing changes for infected wounds should be accomplished more frequently than 48 to 72 hours Aways replace with sterile disposables from unopened packages Fallow established institution protocols regarding clean versus sterile technique MONITORING THE WOUND Inspect the dressing frequentiy to ensure that the foam is collapsed and that therapy is being delivered n a consistent manner Monitor periwound tissue and exudate for signs of infection or ther complications Extra care and attent
13. ificult wound hemostasis or who are an anticoagulants When placing the foam dressing in clase proximity to blood vessels or organs take care to ensure that they are adequately protected with overlying fascia issue or other protective barriers Exposed tendon nerves or blood vessels should be protected by moving available muscle fascia over them by a layer of synthetic material Greater care should be taken with respect to weakened iradiated or sutured blood vessels or organs Bone fragments or sharp edges could puncture dressing barier vessel or organ Wounds with enteric fistula require special precautions in order to optimize therapy Additional Precautions Defibrillation Remove the dressing f defibrillation is required in the area of dressing placement Failure to remove the dressing may inhibit electrical curent transmission andor patient resuscitation Magnetic Resonance Imaging MRI The Curato Suction Pump Unit is not MRLcompatible Do not take ino the MRI area Hyperbaric Oxygen Therapy HBO NEVER allow a device whether on or off inside hyperbaric chamber The device must be disconnected from the patient prior to HBO treatment Large Canisters Use of Large Canisters 500m may increase serious risks associated with excessive fuid loss Monitor patient status continually DO NOT USE for patients with low fad volume nor for patients at nigh risk af major hemorrhage During Negative Pressure T
14. ion should be given if there are any signs of possible infection or lated complications Infection can be serious With or without the Curato Wound Treatment System Infection can lead to many adverse complications including pain discomfort fever septi shock and various other complications Wih signs of moro serious complications of infaction discontinue the use of the Curato Wound Treatment System until the serious infection is diagnosed and properly treated DISCOMFORT ADHERENCE I patient appears uncomfortable during dressing change consider pre medication such as uso of a mon adherent prior to foam placement ar irrigation of a topical anesthetic agent such as 1 Lidocaine priar to dressing removal UNSTABLE STRUCTURES Use the lowest pressure setting on the Curata Wound Treatment System over unstable body structures such as unstable chest wall or non intact fascia INFINITI MEDICAL Curato System Users Manual BODY CAVITY WOUNDS Undorlying structures must be covered by natural tissues or synthetic materia that frm a compl barrier between the underlying structures and the dressing INFINITI MEDICAL DRESSING USE The dressings are to be used exclusively with the Curato Wound Treatment System NOTE dressing components of the Curato Wound Treatment System are packaged sterile The decision to use dean versus steroasepti technique is dependent upon wound pathophysiology and inician pref
15. ith excessive fluid loss Monitor patient status continually DO NOT USE for patents with low fid volume nor for patients at high risk of major hemorrhage INFIN Curao System Users Manual 5 amp Cleaning Introduction The following instructions are Infiniti Medica recommended cleaning and infection conia proceduras for tho Cuato Wound Treatment System The veterinarian and caretakers should review tis manual in entirety before attempting to the use the product Carel read the PRECAUTIONS and SAFETY TIPS inthe INTRODUCTION section before attempting to perform ceaning procedures on the Curato Wound Treatment Unit Protective Equipment Universal Safety Precautions should be used to minimize the risk of infection and contact with contaminated blood or bodily fluids during the dressing changes or disposal itis important ta protect all exposed skin and mucous membranes The protective equipment includes Disposable gloves latex or 19 70 Protective cap and mask Disposable impervious gown Disposal Alter patient use al disposable components of the system should be reated as contaminated These include a The foam dressing components The exudate colection canister Tubing connectors and clamps Dispose of al disposable components in accordance with state and federal regulations and instituton protocols NOTE Cleaning proceduras should nat be performed when unitis c
16. may need be changed more often than 48 hours the dressing change interval should be based cnical evaluation of the wound condilon rather than a fired schedule NOTE The canister should be replaced when full ha Full Canister alarm activates or at least once every w to minimize the potential for contamination and production af odors 1 Press the OFF button to turn the therapy 2 Close suction tubing clamp 3 Disconnect SickConnact suction tube Twisting the tapered connector vill make removing the Suction tube from the canister easier 4 Slowly pul drape up and away skin while gently stretching drape NOTE Ifthe foam dressing adheres to the wound during removal refer to tho SAFETY TIPS section of this manual 5 Discard disposables in accordance with applicable rules regulations and infection control protocols and aways follow Universal Safety Precautions Curao System Users Manual Disposal of Dressings Canister and Other Disposables minimize the risk of infection and contact with contaminated blood bodily fide during the dressing changes disposal iis important to protect all exposed skin and mucous membranes Alter patient use ali disposable components ofthe system should be treated as contaminated These include The foam dressing and polyurethane drape The exudate collection canister SckConnect tubing connectors amp camps Dispose of al disposable component
17. ment System please contact your Jacal Infiniti Medical representative or INFINITI MEDICAL Advanced Technology for Vets and Pets 525 Middlefield Road Suite 150 Menlo Park CA 94025 650 327 5000 www infinitimedical com sales infinitimedical com www lofinitimedical com A trademark designated herein are o nt Madea covered ane ar mre peg pn For questions or comments regarding the content of this User s Manual lease contact Infrnti Medical at the above address Please contact Infiniti Medical Customer Service 1 650 327 5000 for issues concerning the product and its use INFIN MEDIGAL Curato System Users Manual
18. ndicators Nam Baton Condon s PowerCharsing i sane TN INFINITI MEDICAL Curato System Users Manual EI Power Up Procedure 1 Verily the dressing application is corect tubing is connected and the suction tubing with SlckConnect camp is open 2 Place the therapy unit in an upright postion as level withthe wound as possible The device can be placed on a table or attached an LV polo using the LV Pole adapter CAUTION The LV pole clamp should only be used on poles that are in excess of 0 9 22 cm diameter and are securely attached to a bed frame or suitable stand To ensure stability of the therapy unit on tho LV pole it should be clamped no higher than two times the width of the pole base The clamp should be tightened to ensure that the therapy unit cannot slide down the pole 3 Prass the ON button ll LED indicators wil sequentially iluminate during the power on sell test 4 Each time the device is tumed the front panel LED display wil indicate the remaining therapy ifo unit using the folowing display format d XX h YY For example if the display for a 30 day devica indicates d 11 h 22 Iis means 11 days and 22 hours of therapy remain on the device This information can also be interpreted as the device has been used for 18 days and 2 hours 5 Upon tuming the device the dressing should slowly collapse indicating the presence of suction Onc
19. of suction that be applied to the wound site Electronic sensors limit the maximum applied suction to 200 mmHg 109 Intermittent Mode The Curato Wound Treatment System can be set to operate intermittently Sine ON 2 minuto OFF cyclo Unit maintains pressure at 25 mmHg during the OFF state to prevent loss of dressing seal and increase patient comfort NoiseGuard Unit is virtually statin its normal operation wih wel sealed dressing increasing patient comfort and compliance PowerGuard An internal battery provides up to 24 hours af operation from a single ful chargo Battery charges while unitis operating wih the AC adapter While running on battery a low battery alarm wil sound and the fron panal LED display wil indicate a low battery alarm condition when remaining capacity of the battery is less than 20 TherapyGuard Automated alarms for leaklow pressure full canister and low battery Alarms provide both a visual and audible indication Alarms wil sell teset once a problem is corrected or can be manually raset by tuming the therapy unit OFF and Audiblo alarms can muted for five minutos by pressing the MUTE button SlickConnect Eight foot single lumen tubing set with adhesive anges facilitate connection to dressing Single Patient Use Canisters 20000 and 50000 canisters for normal and highly exudating wounds CAUTION Use of large canisters 500ml may increase serious risks associated w
20. onnected to a patient Disconnect the unit tom the patient and power source before cleaning or servicing INFINITI Curato System Users Manual Cleaning the Curato Device Perform a visual inspection of the device Check fr any sign of contamination and ensure that the device is functioning propery Ifthe device is not operating propery refer to the Alarm Troubleshooting guide in the OPERATING INSTRUCTIONS secton of this manual or contact Infiniti Medical to replace the device help reduce the risk of infection and contact with contaminated blood and bodiy fud please wear the protective equipment identifed above when cleaning the Curato Wound Treatment device NOTE Aways follow Universal Safety Precautions Follow established institution protocols regarding clean versus stria technique The folowing cleaning procedure must be performed at least once a week and must be completed between patients The Curato Wound Treatment device should be wiped with either diluted solution ofS militrs bleach in 1 tar of warm water approximately 1 teaspoon bleach in 1 quart water or mild disinfectant Usa coarse cloth and wring out any excess solution unti the cioth is damp and not dripping AC ADAPTER INSPECTION The Adapter should be inspected regularly for damage andlor unusual wear Replace damaged worn Power Supplies immediately A C Adapters are available from Infiniti Medical WARNING The Cur
21. s in accordance with state and federal regulations and instituton protocols Device Renewal When the device s service timer indicates is time to have the therapy unit renewed please return the and A C power adapter to Infiniti Medial for servicing Please ensure the canister is removed the device before sending it and never send disposable componentslaccessores such as the bag canisters dressings or tubing in the retum package INFINITI MEDICAL Curato System Users Manual 2 This section contains instructions for setting and adjusting functions of the Curato Wound Treatment System The section explains the procedure for activating therapy and explain the major functions that adjusted from the control panel Carefuty read the PRECAUTIONS and SAFETY TIPS in the INTRODUCTION section before attempting to operato and adjust the Curato Wound Treatment System WARNING The Curato Wound Treatment System should only be used with the supplied A C Adaptor Use of an incorrectly rated adapter could create a shock hazard for the patient caregiver The part number for the adapters can be found in the REPLACEMENT PARTS section of this manual Power On Off The ON and OFF buttons are located on the front top of the control panel The ON and OFF buttons control the application of power to the therapy unit Curato Por ON o Tera i
22. s inadvertently Locking To lock the unit prass and hold the power ON button for three seconds until three audible beeps aro heard Atthis point the unit is locked The therapeutic seting wil be recalled each time the unit is powered OFF and ON and the unit wil remain locked unti itis subsequently unlocked Unlocking To unlock the unit press and hold the power ON bution unti three audible beeps are heard At this point the unit is unlocked and therapy settings can be changed Aditionall when the unit is powered OFF and ON the unit wil remain unlocked INFIN Curato System Users Manual Curato Wound Treatment System Dimensions 76x43x275in 13x 10x70 cm Weight 43K Therapy Settings 50 75 100 125 150 mmHg Canister Volume Wih respect to electric shock fre and mechanical hazards conforms to UL 60601 1 60601 1 ENGOSDI 1 IEC Classification Medical Equipment Equipment not suitable for use in presence of flammable anesthetic mixtura with air oxygen or nitrous oxide Continuous Operation B Applied Part gt Class Ii Intemally Powered Equipment Battery Duration Fuly Charged upto 24 hours Electrical Power Supply Input 100 240 VAC 80 60Hz 200 mA or 12 24 VDC 850 mA Optional Power Supply Output 5VDC 1 Amps Patient amp Enclosure leakage Current lt 100 Miro amps Environmental Conditions Storage
23. setting simply press ether tha therapy selection button or therapy selection button desired therapy selection is indicated by the green LED 3 The green LED indicator wil lash indicating the selection has baen made and will continue flashing unti the desired negative pressure level has been achieved at which ime the LED wal remain urinated If the green LED indicator begins to lash during therapy it means the device is unable to maintain the therapeutic setting This event would most likely be associated with dressing leak and Wil requira clinician intervention to correct Intermittent Mode ONIOFF The Curato operate in an intermittent suction mode with a 5 minute ON and 2 minute OFF Prass the buon to tum the Intermittent Mode on and off During intermitent operation Curato will provide target therapy pressure during the ON part of the cycle and approximately 25 mmHg during the part af the cycle By maintaining ths lower pressure tho unit is OFF the dressing seal is never compromised This method of applying intermittent pressure also increases patient comfort INFINI D Curato System Users Manual 15 Beeper Volume Adjustment The volume of the beeper can be adjusted ft various care settings or patient preferences To adjust the beeper volume press and hold the ON button while simultaneously pressing the increase the volume orthe q button to decrease th
Download Pdf Manuals
Related Search
Related Contents
1/3 HP Submersible Sump Pump WTP-8B66 Series User's Manual - Wincomm MT KIT 10 - Door Entry Direct Cizalla móvil Genesis XP Manual de Usuario Ambulance and pre-hospital services in risk situations MR-9132E USER`S MANUAL I =ALPHABET OF KIROBO= Web As Corpus Toolkit User's and Hacker's Manual Thuraya SatSleeve FRG Series User Manual - Automated Laser Corporation Copyright © All rights reserved.