Home
        EDM Users Guide - Crystal Instruments India Private Limited
         Contents
1.     300 Orn    irm     300 Or    Tus  D  m     B00 Or   300 m     1 2       Figure 4  Time Domain Waveform in CoCo    A very common reading will show a spectrum peak at 100Hz with a peak value reading  1 0208 in in s  Peak        COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 7    CRYSTAL    instruments    h1     ipeckrumfi    300  0r       0 00 100 00  Hz  div  Figure 5  FFT Spectrum in CoCo  in s Peak    If somebody is interested in the RMS value of this frequency component  he can change the  spectrum type to RMS  then the display value will be changed to 0 7218     RPM 101    ch 1     300 0rn    r     J       A  li  p E        150  0rm       0 00 100 00  Hz  Jdiw  Figure 6  FFT Spectrum in CoCo  in s RMS  Similarly  the user can look at Peak to Peak  vDB SI  and vDB US  of the spectral peak   Now let s introduce the concept of dB   Most often  spectra are shown in the logarithmic unit decibels  dB   Using this unit of measure   It is easy to view wide dynamic ranges  that is  it is easy to see small signal components in  the presence of large ones  The decibel is a unit of ratio and is computed as follows     dB   10logi0  Power Pref     where Power is the measured power and Pret is the reference power        COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 8    CRYSTAL    instruments    Use the following equation to compute the ratio in decibels from amplitude values   dB   20log10  Ampl Aret   where Ampl is the measured amplitude and Ar
2.  PAGE 30    CRYSTAL    instruments          Figure 30  Connecting Channel 1 to Accelerometer    Case 2  Tri axis Vibration Measurement    You can use either one tri axis accelerometer to measure the 3D vibration  Simply connect  chi  ch2 and ch3 of CoCo to the X  Y and Z axis of the tri axis sensor  The sensor will  generate signals for three channels simultaneously                       Figure 31  Connecting Tri axis Accelerometer  Case 3  Single Channel Vibration Measurement   Tacho    Connect chi of CoCo to the analog output of tachometer  connect ch2 of CoCo to the sensor        COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 31    CRYSTAL    instruments                Figure 32  Connecting Tachometer and Accelerometer  Case 4  Tri axis Vibration Measurement   Tacho    Connect ch1 of CoCo to the analog output of the tachometer  Connect ch2  ch3 and ch4 of  CoCo to the X  Y and Z axis of the tri axis sensor                             Figure 33  Connecting Tachometer and Tri axis Accelerometer    COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 32    
3.  first is when the  whole time capture is long enough to cover the complete duration of the signals  This can  occur with short transient signals  For example in a hammer test  if the time capture is long  enough it may extend to the point where the signal decays to zero  In this case  a data  window is not needed     The second case is when a periodic signal is sampled at such a sampling rate that is  perfectly synchronized with the signal period  so that with a block of capture  an integer  number of cycles of the signal are always acquired  For example  if a sine wave has a  frequency of 1000Hz and the sampling rate is set to 8000Hz  Each sine cycle would have 8  integer points  If 1024 data points are acquired then 128 complete cycles of the signal are  captured  In this case  with no window applied you still can get a leakage free spectrum     Figure 8 shows a sine signal at 1000 Hz with no leakage resulting in a sharp spike  Figure 9  shows the spectrum of a 1010 Hz signal with significant leakage resulting in a wide peak   The spectrum has significant energy outside the narrow 1010 Hz frequency  It is said that the  energy leaks out into the surrounding frequencies     COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 10    CRYSTAL    instruments       Figure 8  Sine spectrum with no leakage        Figure 9  Sine spectrum with significant leakage     Uniform window  rectangular   w k    1 0  Uniform is the same as no window function     Hann window    21
4.  in the exponential average is  given by     yln    y n  1    1   a    x n    a    where y n  is the nth average and x n  is the nth new record     is the weighting coefficient   Usually    is defined as 1  Number of Averaging   For example in the instrument  if the  Number of Averaging is set to 3 and the averaging type is selected as exponential averaging   then a   1 3    The advantage of this averaging method is that it can be used indefinitely  That is  the  average will not converge to some value and stay there  as is the case with linear averaging   The average will dynamically respond to the influence of new records and gradually ignore  the effects of old records     Exponential averaging simulates the analog filter smoothing process  It will not reset when a  specified averaging number is reached     The drawback of the exponential averaging is that a large value may embed too much  memory into the average result  If there is a transient large value as input  it may take a long  time for y n  to decay  On the contrary  the contribution of small input value of x n  will have  little impact to the averaged output  Therefore  exponential average fits a stable signal better  than a signal with large fluctuations     Peak Hold    This method  technically speaking  does not involve averaging in the strict sense of the word   Instead  the  average  produced by the peak hold method produces a record that at any point  represents the maximum envelope among all the component 
5. 1 0 033  Time  seconds     Figure 12  A 1 kHz sine wave sampled at 8 kHz  top  and also sampled at 5 12 kHz  bottom      It is clear that the higher the sampling frequency  the closer this digitized signal is to the true  analog waveform  When the sampling rate is low  the digital integration will have significant  calculation error  For example the 5 12 kHz sampled signal is not symmetric about O volts so  the integration will drift and a double integration may grow with accumulated error very fast     In general  you should use a sampling rate at least 10 times higher than the frequency  content that is of interest in the signal when you apply numerical integration   For example  a  motor at 3600 RPM is driving a machine through a gear box which has a 3 1 reduction gear  with 36 12 gear teeth  To detect the gear mesh frequency  the motor speed of 60 Hz is  multiplied by the number of teeth to get the gear mesh frequency of 2160 Hz  To detect  problems in the gearbox it is necessary to sample at 2 16 kHz or higher   Think of trying to  draw a single sine wave using points on a graph  It will be much more clear with 10 points or  more than with only two     DC offset is the second type of digital integration error and can be more severe  It is caused  by any measurement error before integration and may result in huge amplitude errors after  the integration  The illustration below shows how a small measurement error in acceleration  will create a constant DC offset in the acce
6. CRYSTAL    instruments    Vibration Data Collector       Signal Analysis       James Zhuge  Ph D   President  Crystal Instruments Corporation  4633 Old lronsides Drive  Suite 304    Santa Clara  CA 95054  USA    WWW gO ci com     Part of VDC User s Manual   3 16 2009    COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 1    CRYSTAL    instruments    Dynamic Signal Analysis in Vibration Data Collector       The CoCo 80 90 provides two different user interfaces for Dynamic Signal Analyzer and  Vibration Data Collector  The style and settings are different to meet industrial conventions   The user has the choice to enter one of the interfaces when system is powered on  The VDC  user interface is specifically designed for fast data collection operation and ease of use  A  professional user focused on research and development can open and use the DSA functions  instead of that of VDC  This section explains in detail about how the signals are processed  when CoCo runs in the VDC mode     CoCo uses various different technologies of digital signal processing  Among them  the most  fundamental and popular technology is based on the Fast Fourier Transform  FFT   The FFT  transforms time domain signals into the frequency domain  To perform FFT based  measurements  however  it helps to understand the fundamental issues and computations  involved  This Appendix describes some of the basic signal analysis computations  discusses  anti aliasing and acquisition front end for F
7. FT based signal analysis  explains how to use  windowing functions correctly  explains some spectrum computations  and shows you how to  use FFT based functions for some typical measurements  Users should be aware of the  subtle differences between a traditional dynamic signal analyzer and a vibration data collector  even though they all employ the same signal processing theory     General Theory of Spectral Analysis  Time Domain Waveform  A typical time waveform signal in analog form from the sensor  such as an accelerometer     velocity meter or displacement probe  could take an appearance like that shown in the  following picture     COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 2    CRYSTAL    instruments       m 03 11 2009 02 51 49 Entry2   3600 RPM    MATAITAI     in s     TRY    0 461 0 554 0 646 0 739 0 834 0 924  Time  Seconds           Figure 1  Time Domain Waveform    In a digital instrument  much the same thing is seen  However  it is necessary in a digital  instrument to specify several parameters in order to accurately represent what is truly  happening in the analog world  It is important to tell the instrument what sample rate to use   and how many samples to take  In doing this  the following are specified     p Spectral Lines FBG o       Figure 2  Parameter Setup in CoCo    Measurement Quantity  This field is required to determine what measurement quantity is to  be displayed  Even if the sensor is an accelerometer  the CoCo device can int
8. The filter cutoff frequency is specified at  3dB attenuation     To remove unwanted signals at or near DC  please set up the cutoff frequency of the digital  high pass filter as high as possible as long as it won t chop off useful frequency content of  your interest     To give an example  if you are not interested in any frequency less than 20Hz  then you can  set the cutoff frequency to approximately 10Hz  With this setting  the amplitude attenuation at  20Hz will be less than 1dB   Typically  the lowest frequency of interest on rotating machinery  will be one half the running speed of the machine  If the high pass filter is set to one third the  running speed  the half order vibration will still be detectable         COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 18    CRYSTAL    instruments    Readings in a Vibration Data Collector  Readings    Readings are overall values that represent the characteristics of the measured signals  They  are either calculated from time waveform or frequency spectrum  In CoCo these readings can  be displayed individually or together with the waveform or spectrum     g  overall mns     ch 1     infs  peak     inis  true ms     ret  mi      eH  m  1i  C    RPM  100 0    ch 1        brunt    Figure 15  Onsite Measurement Display  Peak and Peak Peak    Peak and Peak Peak values are calculated from the time waveform  Peak value is the largest  signal level seen in a waveform over a period of time  For sine signals  the peak val
9. ctor AmpCorr is calculated based on the data window shape     Step 4    Apply one of the averaging techniques to the power spectrum Sxx  see below for  averaging techniques     Step 5    Finally  take the square root of the averaged power spectrum to get final spectrum  result        COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 6    CRYSTAL    instruments    Spectrum Type    Display Preference    b Acceleration Spectrum Type  b Acceleration Engineering Unit    p Velocity Engineering Unit      p Displacement Spectrum Type PRED  k Displacement Engineering Unit     Current Spectrum Type     Peak Peak     amp  Current Engineering Unit dB  SII  P Spectrum Y Axis Type       vdB  U5   k Spectrum s Axis Type         Figure 3  Display Preference Setup    Now we come to a confusing part about the spectrum of a signal  With the same time domain  signal  the spectrum can actually be displayed in different values  This is controlled by a  parameter  spectrum type  set in the Display Preference on the CoCo     The motivation of doing so is that people may want to look at different aspect of the spectrum  and give different physical interpretation to the original time signals  For example  from the  spectrum the user may wants to know the frequency component at 1X rotating speed   represented in its Peak  Peak to Peak or RMS     To give a practical example  a 100Hz sine wave with roughly 1 0 peak in s is fed into the  CoCo system  The time waveform is shown below     h1 
10. d    10V    10V user selectable  This tacho channel accepts either the tacho sensor with regular  voltage output or a tacho sensor with IEPE ICP interface     Typical tacho measurement specification using a PLT200 tachometer from Monarch  Instrument is     RPM Range  5 200 000 RPM    Accuracy   0 01  of reading       COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 29    CRYSTAL    instruments    Resolution  0 001 to 10 RPM    Operating Range  2 inches to 25 feet       Figure 29  Monarch PLT200 Tachometer  Pocket Laser Tachometer 200 Kit includes  Tachometer  Remote Contact Assembly  RCA      Carrying Case  Factory Calibration Certificate and 5 ft roll of T 5 reflective tape  PLT200 has a  TTL compatible Pulse Output that can be connected to the channel 1 of CoCo     Typical Connections of CoCo with Accelerometers and Tachometer  Several typical connections are recommended below using a four channel CoCo device  If  you are doing the route data collection  make the same parameter setup in EDM  upload the  route to the CoCo and conduct the test  This setup cannot be changed on CoCo     If you are conducting onsite measurement  set the input channels accordingly in the Input  Channel and Sensor setup on CoCo     Case 1  Single Channel Vibration Measurement    This is the simplest measurement  Connect chi of CoCo to the sensor                                                                          ss    COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED 
11. ducer and a second engineering unit after the integration  For  example  first the instrument must provide a means to set the sensitivity of the sensor   say 100mV g in acceleration  After the double integration the instrument must have  the means to set the engineering unit to a unit that is compatible with the integration  such as mm of displacement     The CoCo instrument handles these three issues effectively so you can get reliable velocity  or displacement signals from the acceleration measurement  or displacement signals from  the velocity measurement  The CoCo hardware has a unique design to provide 130dB  dynamic range in its front end measurement  The signals with high dynamic range will create  better results after digital integration     Since such built in integration is conducted in the time domain before any other data  conditioning or spectral analysis  the time streams generated after the digital integration can  be treated in the same way as other time streams  They can be analyzed or recorded                                                                   COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 15    CRYSTAL    instruments    CoCo also provides differentiation and double differentiation to calculate the acceleration or  velocity from velocity or displacement transducers  Differentiation is not as commonly used as  integration     It must be noticed that the displacement value derived after double integration of the  acceleration s
12. e performed at the analog hardware level or at the digital level     The CoCo provides a means to digitally integrate or double integrate the incoming signals   The integration module fits into the very first stage after data is digitized  as shown below                                Anag A D  Spona  Data Spectral  Signal High Pass Filter ee    Z    B Converter       m   Conditioning m  Analysis  Conditioning and Integration                                           CoCo             Figure 11  Signal Processing Sequence in CoCo  There are several issues to address in such implementation     1  The integration and double integration algorithm has to be accurate enough and it  must find a way to reduce the effects of a DC offset  A tiny initial value  offset in the  measurement or temperature drift before the integration  may result in a huge value  after single or double integration  This DC effect can be removed using a high pass  filter     2  The initial digital signal must have a high signal to noise ration and high dynamic  range  The integration process in essence will reduce the high frequency energy and  elevate the low frequency components   f the original signals do not have good  signal to noise ratio and dynamic range  the signals after integration and double  integration will have too much noise to use  The noise will corrupt the integrated  signal     3  The instrument must be able to set two different engineering units  one engineering  unit for the input trans
13. ectrum in frequency domain between  Fmin and Fmax        f max     gt  Power f      f min    TrueRms         BW   Where   BW   noise power bandwidth of window  Fmax   maximum frequency of interest  Fmin   minimum frequency of interest    Fmax and Fmin are set in the Analysis Parameters in CoCo  They control the maximum and  minimum frequency of interest  as shown below            gt    Measurement Quantity          b Fmin  Hz     Figure 17  CoCo Display  Setting Fmax    Obviously  the true RMS will be no greater than the Overall RMS     Demodulation Spectrum    A useful technique for measuring and analyzing data is a process called Demodulation  The  demodulation process is effective for detection of high frequency low amplitude repetitive  patterns that lie embedded within the time waveform  These are characteristic of certain types  of mechanical faults  particularly rolling element bearing faults such as inner or outer race  cracks and spalls that make a clicking or ringing tone as the rolling elements pass over the  fault  Demodulation is useful as an early warning device  as it detects bearing tones before  they are visible in a normal spectrum  As the fault progresses towards failure  the frequencies  will spread out and appear more as an increase in the    noise floor  of the FFT spectrum as  the amplitude increases     The process works by extracting the low amplitude  high frequency impact signals and then  tracing an    envelope    around these signals to identify t
14. ef is the reference amplitude   As shown in the preceding equations for power and amplitude  you must supply a reference  for a measurement in decibels  This reference then corresponds to the O dB level  Different    conventions are used for different types of signals     The vibration velocity level in dB is abbreviated VdB  and is defined as     p  Vra f       Va B   2  log    y  Mab   20log    10   m sec    The Systeme Internationale  or Sl  is the modern replacement for the metric system     The reference  or  0 dB  level of 10 9 meter per sec is sufficiently small that all our  measurements on machines will result in positive dB numbers  This standardized reference  level uses the Sl  or  metric   system units  but it is not recognized as a standard in the US  and other English speaking countries   The US Navy and many American industries use a  zero dB reference of 10 8 m sec  making their readings higher than SI readings by 20 dB      The VdB is a logarithmic scaling of vibration magnitude  and it allows relative measurements  to be easily made  Any increase in level of 6 dB represents a doubling of amplitude   regardless of the initial level  In like manner  any change of 20 dB represents a change in  level by a factor of ten  Thus any constant ratio of levels is seen as a certain distance on the  scale  regardless of the absolute levels of the measurements  This makes it very easy to  evaluate trended vibration spectral data  6 dB increases always indicate doubling o
15. egrate it  digitally into velocity or displacement     Fmax  This field defines the maximum frequency of interest for analysis  The sampling rate of  the analog digital  A D  digitizer will be determined based on this parameter     Fmin  This is the low frequency cut off filter that will be applied in the frequency domain for  spectral analysis     Block Size Spectral Lines  The block size is usually defined in blocks of two  binary  to the  power of 10 or more   Block size of 2   is 1024  2   is 2048  2   is 4096  etc   The block size  and Fmax will determine the total time period of each sampling block for frequency analysis  A  larger block size for the same frequency band will increase the accuracy of the measurement   Immediately after the signal is digitized  it will also go through     e Low pass filters   to eliminate any high frequencies that are not wanted       COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 3    CRYSTAL    instruments    e High pass filters   to eliminate DC and low frequency noise that are not wanted    Additionally  the integration of the signal provides velocity or displacement from an  accelerometer or a displacement from a velocity pickup  Traditional signal analyzers have a  drawback of dynamic range in the digital domain and some argue that the analog integration  is superior to that of digital  The situation is greatly improved due to the very high dynamic  range technology in the CoCo  With more than 130dB dynamic range i
16. er must pay attention to the sensitivity of the sensor when they source it  Select an  accelerometer by matching its output for expected acceleration levels  Don t  crowd  the full   scale specifications  Allow a margin for unexpectedly large accelerations  Using only the  lower 2096 of an accelerometer s response range will ensure ample margins for unpredicted       COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 28    CRYSTAL    instruments    overloads  After you select an accelerometer that can survive predicted worst case shock  limits  compute the sensor   s output voltage  At a sensitivity of 10 mV g  for example  an  accelerometer that encounters a 100 g shock will produce a 1 V peak signal  This is well  wthin the    10V range of CoCo input channels  However it must be noted that this definition  is accounted in the acceleration domain  To transform the specification from the velocity  domain  the frequency factor has to be accounted     Integral Electronics Piezoelectric  IEPE  Sensor    IEPE accelerometers operate from a low cost  constant current power source over a two wire  circuit with signal power carried over one wire and the other wire serving as ground  The  cable can be ordinary coaxial or ribbon wire  Low noise cable is not required  Constant  current to operate the accelerometer comes from a separate power unit or it may be  incorporated inside a readout instrument such as an FFI analyzer or Data Collector   Integrated electronic acceler
17. f the  magnitudes     Data Window Selection  Leakage Effect    Windowing of a simple signal  like a sine wave may cause its Fourier transform to have non   zero values  commonly called leakage  at frequencies other than the frequency of this sine   This leakage effect tends to be worst  highest  near sine frequency and least at frequencies  farthest from sine frequency  The effect of leakage can easily be depicted in the time domain  when a signal is truncated  As shown in the picture  after data windowing  truncation has  distorted the time signal significantly  hence causing a distortion in its frequency domain        COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 9    CRYSTAL    instruments    Actual Input      l                 Windowed Input  Nor Periodic    Assumed Inout  h ra    l   i l     Figure 7  Illustration of a non periodic signal resulting from sampling    If there are two sinusoids  with different frequencies  leakage can interfere with the ability to  distinguish them spectrally  If their frequencies are dissimilar  then the leakage interferes  when one sinusoid is much smaller in amplitude than the other  That is  its spectral  component can be hidden or masked by the leakage from the larger component  But when  the frequencies are near each other  the leakage can be sufficient to interfere even when the  sinusoids are equal strength  that is  they become undetectable     There are two possible scenarios in which leakage does not occur  The
18. hem as repetitions of the same fault        COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 21    CRYSTAL    instruments    The resulting spectrum  with the low frequency data removed  will now clearly show the high  frequency impact signals and harmonics     The high frequency signals that demodulation aims to extract are do not travel well through  large structures  therefore extra care must be taken to ensure the accelerometer is setup  correctly  Ensure that     e The accelerometer is mounted close to the fault source with the shortest direct path  through the structure to the accelerometer     e The accelerometer is well coupled  using either stud mounting or a very strong  magnet on bare metal  A handheld probe or stinger is not recommended     e The accelerometer mounting is consistent between visits  If not  a trend plot of overall  RMS values will be meaningless     The demodulation process can be graphically described in the following flow chart        Acceleration Signal High pass filter Envelope Low pass amp decimation FFT Analysis    Figure 18  Demodulation Process Flow Chart    Below is a depiction of an acceleration time waveform with a repetitive high frequency  component  Because of the large difference in amplitude and frequency  a very low amplitude  high frequency signal could be overlooked during routine vibration analysis           Figure 19  Acceleration Time Waveform with Fault    The high pass filter removes the low frequency componen
19. ignal is not the same as that directly measured by a proximity probe  A  proximity probe measures the relative displacement between a moving object  such as a  rotor shaft  to the fixed coordinates seated by the probe  mounted to the case   The  accelerometer and its integration value can only measure the movement of the moving object  against the gravity field     Sensor Considerations    Accelerometer signals that are non dynamic  non vibratory  static or quasi static in nature   low acceleration of an automobile or flight path of a rocket  are typically integrated in the  digital domain  downstream of the signal conditioner  Piezoelectric and IEPE accelerometers  are commonly used to measure dynamic acceleration and  therefore  dynamic velocity and  displacement  They should not be used to measure static or quasi static accelerations   velocities  or displacements because the IEPE includes analog high pass filtering in the  sensor conditioning that cuts out any low frequency signal  At low frequencies approaching O  Hz  piezoelectric and IEPE accelerometers cannot  with the accuracy required for integration   represent the low frequency accelerations of a test article     When this slight inaccuracy is integrated in order to determine velocity and displacement  it  becomes quite large  As a result  the velocity and displacement data are grossly inaccurate  A  piezoresistive or variable capacitance accelerometer is a better choice for low frequency  signals and for integrat
20. ion  These types of sensors measure acceleration accurately at  frequencies approaching O Hz  Therefore the integration calculation of velocity and position  can be used to produce accurate results     Calculation Errors in Digital Integration    Two types of calculation errors can be introduced by parameters chosen for digital  integration  low sampling rate and DC offset     The sampling rate of a signal must be high enough so that the digital signal can accurately  depict the analog signal shape  According to the Nyquist sampling theorem as long as the  sampling speed is more than twice of the frequency content of the signals before the  integration  the integration results should be acceptable  This is not true  Satisfying the  Nyquist frequency only ensures an accurate estimate of the highest detectable frequency of a  measurement  It will not provide an accurate representation of the signal shape  Integration  error can still occur of a signal is sampled at more than twice the signal frequency  The figure  below shows a 1kHz sine wave sampled at 8kHz and 5 12kHz     COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 16    CRYSTAL    instruments          8    SI60013 BLOCK ch1          0 024 0 025 0 026 0 027 0 028 0 029 0 030 0 031 0 032 0 033  Time  seconds      Signals SIG0014 BLOCK ch1   4bx    e  u    BM    b   Begin 0 0240 End 0 0330                amp     SIG0014 BLOCK ch1   1 000  0 500     0 0004   0 500        1 000    0 024 0 026 0 028 0 029 0 03
21. k    w k    0 5    0 5 cos        N 1    Flattop window                                                                           aur  COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 11    CRYSTAL    instruments             21k 4r k 67k  w k      1     1 93 COS   1 29 cosy     0 388 cosa    fork 0 N 1         0 032 Mi    cos     The term  Hanning window  is sometimes used to refer to the Hann window  but is  ambiguous as it is easily confused with Hamming window     If a measurement can be made so that no leakage effect will occur  then do not apply any  window  in the software  select Uniform    As discussed before  this only occurs when the  time capture is long enough to cover the whole transient range  or when the signal is exactly  periodic in the time frame     If the goal of the analysis is to discriminate two or multiple sine waves in the frequency  domain  spectral resolution is very critical  For such application  choose a data window with  very narrow main slope  Hanning is a good choice  In general  we recommend Hanning  window in VDC applications     When you are extremely sensitive to the accuracy of peak estimation at certain frequency   choose Flattop window  It will give you the best estimation for the frequency components  measured at a rotating machine or reciprocating machine     Averaging Techniques    Averaging is widely used in spectral measurements  It improves the measurement and  analysis of signals that are purely random or mixed ra
22. lap means 25  of the old data will be used for each spectral  processing  0  overlap means that no old data will be reused     Overlap processing can improve the accuracy of spectral estimation  This is because when a  data window is applied  some useful information is attenuated by the data window on two  ends of each block  However  it is not true that the higher the overlap ratio the higher the  spectral estimation accuracy  For Hanning window  when the overlap ratio is more than 50    the estimation accuracy of the spectra will not be improved     Another advantage to apply overlap processing is that it helps to update the display more  quickly     Built in Digital Integration And Filtering  Introduction to Digital Integration    Ideally  a measurement is made using a sensor that directly measures the desired quantity   For example an accelerometer should be used to measure acceleration  a laser velocimeter  or velocity pickup should be used to measure velocity and a linear voltage displacement  transducer  LVDT  should be used to measure position  However since position  velocity and  acceleration are related by the time derivatives it is possible to measure an acceleration  signal and then compute the velocity and position by mathematical integration  Alternatively       COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 14    CRYSTAL    instruments    you can measure position and compute velocity and acceleration by differentiating  The  integration can b
23. leration integrated to compute velocity and result  in a drift and eventually an infinitely large magnitude of displacement after double integration     COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 17    CRYSTAL    instruments    Displacement    Velocity    a              Y Acceleration    Figure 13  A small error in acceleration results in a DC offset in velocity and a huge drift in displacement        Of course  the computed velocity and displacement signals are unrealistic  They are artifacts  of the integration errors  In order to remove such a problem caused by inaccurate  measurement and digital integration  a high pass filter can be applied before or after the  integration  It should be noted that the high pass filter will distort the waveform shape to some  extent because it alters the low frequency content of the signal  However this effect must be  tolerated if numerical integration is used     Digital High Pass Filter    The most effective way to remove the DC drift effect as described above is to apply a high  pass digital filter to the continuous time streams  In CoCo  a unique algorithm is realized so  that even the data is sampled at high rate  the high pass filter can still achieve very low cutoff  frequency  The high pass filter parameter can be entered in the channel table     Input Channel  amp  Sensor Setup    che Sensor Input Mode Hi PassFltr     accel 612 mwito    J Ere    JC  Single i       Figure 14  CoCo Input Channel Setup Table  
24. n the front end  digital  integration can achieve excellent accurate results     The Fourier Transform    CoCo fully utilizes FFT frequency analysis methods and various real time digital filters to  analyze measurement signals  The Fourier Transform is used to convert quantities amplitude  vs time in the time domain  time waveform  to amplitude vs frequency in the frequency  domain  FFT spectrum   usually derived from the Fourier integral of a periodic function when  the period grows without limit  often expressed as a Fourier transform pair  In the classical  sense  a Fourier transform takes the form of     X f    f secet dt    Where   x t    continuous time waveform  f frequency variable  j complex number    X f    Fourier transform of x t     As the theory of Jean Baptiste Fourier states  All waveforms  no matter how complex  can be  expressed as the sum of sine waves of varying amplitudes  phase  and frequencies  In the  case of rotating machinery vibration  this is most certainly true  A machine s time waveform is  predominantly the sum of many sine waves of differing amplitudes and frequencies  The  challenge is to break down the complex time waveform into the components from which it is  made     Mathematically the Fourier Transform is defined for all frequencies from negative to positive  infinity  However  the spectrum is usually symmetric and it is common to only consider the  single sided spectrum which is the spectrum from zero to positive infinity  For discrete  
25. ndom and periodic  Averaged  measurements can yield either higher signal to noise ratios or improved statistical accuracy     Typically  three types of averaging methods are available in DSA products  They are   Linear Averaging  Exponential Averaging  and Peak Hold  Linear Averaging    In linear averaging  each set of data  a record  contributes equally to the average  The value  at any point in the linear average in given by the equation     Sum of Records    A d    verage N    N is the total number of the records  The advantage of this averaging method is that it is  faster to compute and the result is un biased  However  this method is suitable only for  analyzing short signal records or stationary signals  since the average tends to stabilize  The  contribution of new records eventually will cease to change the value of the average     Usually  a target average number is defined  The algorithm is made so that before the target  average number reaches  the process can be stopped and the averaged result can still be  used     COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 12    CRYSTAL    instruments    When the specified target averaging number is reached  the instrument usually will stop the  acquisition and wait for the instruction for another collection of data acquisition     Exponential Averaging    In exponential averaging  records do not contribute equally to the average  A new record is  weighted more heavily than old ones  The value at any point
26. ometers are available under several different trademark names  such as ICPe  PCB Piezotronics   Isotrone  Endevco   Delta Trone  B amp K   and Piezotrone   Kistler  to mention a few  CoCo IEPE input mode provides 4 7mA constant current for each  channel     The main advantage of low impedance operation is the capability of IEPE accelerometers to  operate continuously in adverse environments  through long  ordinary  coaxial cables  without  increase in noise or loss of resolution  Cost per channel is less  since low noise cable and  charge amplifiers are not required  The main limitation involves operation at elevated  temperatures  above 325 F     The signal conditioning circuitry in the instrument box usually has high pass and low pass  filter  When IEPE is selected in the CoCo  the high pass filter cutoff frequency is set fixed at  0 3 Hz     3dB  and 0 7 Hz     0 1dB      IEPE sensor will not be able to measure the DC or quasi constant acceleration signal  This is    usually not a problem to the acceleration measurement because in our world  no objects can  keep moving at constant acceleration     Tachometer    Tachometer is used to measure the rotating speed of the rotating machines  There are many  kinds of tachometers that can be chosen for CoCo  In general  as long as the tachometer  claims that it output analog pulse signal  it will be able to interface to the CoCo input channel     The first analog input channel can be configured as a tachometer measurement  Threshol
27. ought of as a smoothing function  This  smoothing can be represented by an effective filter shape of the window  i e   energy at a  frequency in the original data will appear at other frequencies as given by the filter shape   Since time domain windows can be represented as a filter in the frequency domain  the time  domain windowing can be accomplished directly in the frequency domain     Because creating a data window attenuates a portion of the original data  a certain amount of  correction has to be made in order to get an un biased estimation of the spectra  In linear  spectral analysis  an Amplitude Correction is applied        COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 5    CRYSTAL    instruments    Spectrum   A spectrum in CoCo in VDC mode is calculated based on a few steps including data window   FFT  amplitude scaling and averaging  You can extract the harmonic amplitude by reading  the amplitude values at those harmonic frequencies in a spectrum    To compute the spectrum  the instrument will follow these steps    Step 1    A window is applied to the time waveform     x k    w k  x k        Where   x k     is the original data and x k  is the data used for a Fourier  transform   Step 2    The FFT is applied to x k  to compute Sx     N 1    Sx   2 x k e sere    n 0    Next the    periodogram    method is used to compute the spectra with amplitude  correction using Sx     step 3  Calculate the  Power Spectrum    Sxx   Sx Sx    AmpCorr     The fa
28. records  The equation for a  peak hold is    y n    MAX   x n     JD    Peak hold is useful for maintaining a record of the highest value attained at each point  throughout the sequence of ensembles  Peak Hold is not a linear math operation therefore it  should be used carefully  It is acceptable to use Peak Hold in auto power spectrum  measurement but you would not get meaningful results for FRF or Coherence measurement  using Peak Hold     Peak hold averaging will reset after a specified averaging number is reached     COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 13    CRYSTAL    instruments    Overlap Processing    To increase the speed of spectral calculation  overlap processing can be used to reduce the  measurement time  The diagram below shows how the overlap is realized     Signal Captured in the Time Domain                2       a                Acquired Signal Data Transformed into FFT Frames  No Overlap Processing       Acquired Signal  Post Processed with Overlap FFTs    1024 Samples        1024 Samples  1024 Samples  1024 Samples    FFTs Overlap Samples  Overlap Interval Samples    1024 Samples    Figure 10  Illustration of overlap processing     As shown in this picture  when a frame of new data is acquired after passing the Acquisition  Mode control  only a portion of the new data will be used  Overlap calculation will speed up  the calculation with the same target average number  The percentage of overlap is called  overlap ratio  25  over
29. sampled signals  this can be expressed as     N 1    X k    2 x k g Jane    n 0       COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 4    CRYSTAL    instruments    Where     x k  samples of time waveform    n running sample index   N total number of samples or  frame size    k finite analysis frequency  corresponding to  FFT bin  centers     X k    discrete Fourier transform of x k     In CoCo  a Radix 2 DIF FFT algorithm is used  which requires that the total number of  samples must be a power of 2  total number of samples in FFT   2    where m is an integer      The Fourier Transform assumes that the time signal is periodic and infinite in duration  When  only a portion of a record is analyzed the record must be truncated by a data window to  preserve the frequency characteristics  A window can be expressed in either the time domain  or in the frequency domain  although the former is more common  To reduce the edge effects   which cause leakage  a window is often given a shape or weighting function  For example  a  window can be defined as     w t    g t   T 2  lt  t  lt  T 2     0 elsewhere    where g t  is the window weighting function and T is the window duration   The data analyzed  x t  are then given by    x t    w t  x t      where x t     is the original data and x t  is the data used for spectral analysis     A window in the time domain is represented by a multiplication and hence  is a convolution in  the frequency domain  A convolution can be th
30. t of the signal  below        COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 22    CRYSTAL    instruments       Figure 20  Acceleration Time Waveform after High Pass Filter    The next step in the process is enveloping which lowers the frequency of the signal to that of  the repetitive element     MII IISA A    Figure 21  Signal after Enveloping    The final step is to process the resulting time waveform signal into a frequency spectrum   Since the signal has been altered by removal of low frequencies and enveloping  it is referred  to as the Demodulated Spectrum     COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 23    CRYSTAL    instruments       liu    Figure 22  Demodulated Spectrum       A Bearing Detection Example of Demodulation    The following examples show CoCo screens in VDC mode being used to analyze a rolling  element bearing with a slight defect               RBM  100 0 200 0  ZU nm  ERE Y  0 00   100 0rn    00    m   300 Or   400   r    ormichi    Wr awe    aM  APM 100  ED   m    Em  LL    UBI  C Masc anl i    100m  0 00 700 00 Meldi    Figure 23  CH1 Time Waveform and FFT with slight bearing defect    Spectrum       The following is the same signal with the demodulation spectrum on the lower trace     COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 24    CRYSTAL    instruments    ormichij    W awe    omm    0 30 Sidiv ENT     PM  100 0 mee   edm  eu  m  16 0m  12 0m       aim   lanl Puh Loh di k Ju or TT UN  
31. the sensor and the machine surface  the more accurate your high frequency  response will be  High frequency response is based on the sensor specified as well as the  method of attachment together with a system  Stud mounted sensors are often able to utilize  the entire frequency range that the sensor specified  Conversely  a probe tip mounted sensor  has very little surface area contact with the machine surface  and offers very little high  frequency accuracy above 500Hz  30  000CPM      The picture below shows the frequency response of a typical accelerometer  It might be  Surprising to you that how inaccurate the measurement can be at different frequency range     COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 27    CRYSTAL              instruments  1 30    f  lower frequency limit  f  calibration frequency  440 f  resonance frequency  1 05  1 00  0 95  0 90  0 71    f 2f  3f  f  0 2f  0 5f  f  f  0 3f     Figure 27  Frequency Response of a Typical Accelerometer    The following chart offers a general guideline for the range of mounting techniques available   and the corresponding high frequency response expectations     Curved Surface Quick Flat Magnet  with Magnet Disconnect with Target  Adhesive  Mount    Stud Mount    E       Tip    ad  eme       H      H       NEM CN ACT M Pn   30 000 RPM   120 000 CPM   390 000 CPM   600 000 CPM  _ 500 000 CPM  Lada    Figure 28  Accelerometer Mounting vs Maximum Frequency Response    Choose the Sensitivity    The us
32. ue is  always 1 414 times the RMS value of the signal level  For non sine signals  this formula will    not apply     The Peak Peak value is the difference between the maximum and minimum signal levels  over a period of time  For a pure sine wave  the Peak Peak level is two times the peak signal  level and 2 828 times the RMS value of the signal level  For a non sine signal this formula will    not apply     COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 19    CRYSTAL    instruments                  peak                               Figure 16  Illustration of Time Domain Peak  Peak Peak    If accelerometer is used and the Peak or Peak Peak reading is displayed for velocity or    displacement  the digital integration will be applied to the time waveform continuously before  the Peak or Peak Peak detection     Overall RMS    In CoCo  the overall RMS is calculated based on the spectrum in frequency domain across all  of the effective frequency range  i e  from DC to maximum analysis frequency range        f5 0 45     gt   Power f   BW    overallrms           Where     BW   noise power bandwidth of window  Fa   analysis frequency band   Fs   sampling frequency band   0 45   the ratio of Fa     According to  Parseval s theorem   such overall RMS is equivalent to that calculated in the  time domain        COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 20    CRYSTAL    instruments    True RMS    In CoCo  the true RMS is calculated based on the sp
33. ustrial Applications    Accelerometers are widely used in the vibration data collection  By using the feature of  CoCo s very high dynamic range  130dB   the acceleration signals can also be accurately  integrated into velocity and displacement signals  There are a wide range of accelerometers  to choose in the market  Many of them are IEPE mode  In most of applications we  recommend using IEPE accelerometers     There are three types of accelerometers in the market   1   accelerometers used for cost   sensitive market such as PDAs  electronic toys  automotive airbags or laptop computers   These are MEMS based sensors that cost a few dollars each  They do not fall into our  categories   2   The accelerometers used for testing and measurement purpose  The US  manufacturers like PCB  Endevco and Dytran all focus on such applications   3   The  accelerometers used for machine vibration  or called industrial applications  US  manufacturers include Wilcoxon  CTC and so on  Most often  the vibration data collector asks  for the sensors from the last category  These sensors are relatively large in size  rugged  less  accurate and less expensive than those from the category  2      Mounting Accelerometers    Care must be taken to insure the appropriate accuracy across the whole frequency range   The accuracy of your high frequency response is directly affected by the mounting technique  that you select for the sensor  In general  the greater the mounted surface area contact  between 
34. y Peet 4 irm   25 00 Hzldiw    Figure 24  CH1 Time Waveform and Demodulation Spectrum with slight bearing defect    chij    Spectrum        As the bearing deteriorates  the defect typically becomes larger and generates a wider range  of frequencies as the rolling elements pass over it  The following is the demodulation  spectrum with slightly deteriorated bearing     120 0m  3n    m  EU  rn  30  0rn  0 00   30   m   0   r   30    m    ormich1      VW awe     30 Stdiw   fA 100 0 175m   b   m  12 5m  REY  Om    chij    Spectrum    j    Ut f og hae A  ld Jus ks Mal ova T LL lan t ul 2  A    Zo 00 Hz Ji nv       Figure 25  CH1 Time Waveform and Demodulation Spectrum with slightly deteriorated bearing    As can be seen in the screen below  the standard FFT spectrum shows the relatively high first  order amplitude but only shows an elevated noise floor in the higher frequencies                                                           v   s  COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 25    CRYSTAL    instruments        RPM  100 0    300  0rn   50   m  0 00   150 0m   300 Orr     450   m    chi    arm i i    WW awe    0 10 Sidi  SPI  100 0 P  JU Urn  50m  60 0rm  45   r  300m  ERU    Em  Nm    ipeckrum c    100 00 e FER T     Figure 26  CH1 Time Waveform and FFT Spectrum with deteriorated bearing       COPYRIGHT    2009 CRYSTAL INSTRUMENTS  ALL RIGHTS RESERVED  PAGE 26    CRYSTAL    instruments    Using Accelerometers and Tachometer       Accelerometers for Ind
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
12151 - BatteryMINDers  molbox™ Analog Circuit Calibration Software User`s Manual  Acquisition commands and parameters - Pascal-Man  GE JVM230BL Microwave Oven User Manual  HP 100B LK561LT  Interactive Setup  MP120 Eng R3  Exmark Lazer ZXP User's Manual  Wibur Curtis Company CAFE PC1 User's Manual    Copyright © All rights reserved. 
   Failed to retrieve file