domiflex 2 wash Push-and-Ready - Hermann Bock GmbH
Contents
1. When conducting the visual and functional testing pay particularly attention to cable splits abrasions kinks and pressure points porosities and exposed wires on the connec tion cables and plug connectors It is also possible for the operators of health care beds to extend the regulated deadline to at least one safety inspection every 2 years This applies when it is proven e g through documented error statistics etc and guaranteed that the error rate e g the number of errors detected during a control check does not exceed the 2 limit as set in BGV Provisions of the German Accident Prevention and Insurance Association A3 Very important If the error rate should exceed again the 2 threshold see BGV A3 the deadline must be reduced again to 1 x per year This health care bed was developed designed and manufactured for safe operation over a long period of time If properly used the expected life service life of this health care bed is up to 10 years The service life depends on operating conditions and frequency Attention Unauthorised technical changes to the product void all warranty claims 40 TSC check list according to MPBetreibV Medical Devices Operator Ordinance BGV A3 and DIN EN 62353 Issuing date 1 July 2015 LlElectric bedL Mechanic bedL Insert frame Series inventory b k number OC Visual inspection No Description o Yes No Remark General Type plate sticker present on
2. Stefan Kettelhoit Company management Company management 38 10 Regular technical safety inspections with service Regular safety inspections ensue the maintenance of the highest possible safety level and are considered an important safety precaution Medical devices must be inspected regularly in terms of safety according to the stipulated regulations of the manufacturer and the generally accepted rules of technology The safety related protection measures are subject to different requirements and demands This also applies to the potential wear and tear in the daily use To prevent such risks constant and consistent compliance with the deadlines for regular technical safety controls TSC is absolutely necessary The manufacturer has no influence on the operator s adherence with respect to the ob servance of these regulations concerning electric beds Bock facilitates the observance of the necessary precautionary measures to be taken by means of their time saving services The execution of the inspection assessment and documentation must be performed only by or under supervision of professional persons such as electricians or electro technically instructed persons who have a thorough knowledge of the relevant provi sions and are able to recognize possible impacts and hazards In case that there is no suitable person on part of the operator to perform regular func tional testing Bock s service offers to perform the functional testi
3. These standards divide the bed in various use areas Intensive care in a hospital intensive care bed 2 Short term care in a hospital or other medical facility patient bed in the hospital 3 Long term care in medical environment stationary nursing bed 4 Home care bed 5 Home care nursing service 1 1 Intended use The health care bed in elderly care or nursing homes or rehabilitation facilities improves the positioning of patients or others in need of care who are 12 years of age or older and having a height of at least 150 cm It is used to alleviate a disability and or to facili tate the lives of people who are in need of care or to make the work of their caregivers easier Furthermore the health care bed was also designed as a convenient solution for home care of frail and elderly people as well as for home care of people with disabilities according to standard EN 60601 1 11 2010 Therefore the health care beds are de signed to be used for the application area 3 to 4 Any other use is considered improper and is excluded from a possible liability claim The Trendelenburg function is a special function that is provided upon request It is not part of the standard equipment It may be used under the supervision of medical profes sionals The beds which are determined for environment type 4 are equipped with a hand control It cannot operate the Trendelenburg function The nursing bed is not suitable for use in hospitals It is als
4. V combinatorial circuit device SMPS drive system or a Dewert 24 V combinatorial circuit device SMPS no leakage current measurement must be carried out during the service life of the health care bed when the visual and functional testing electric components was passed Device leakage current measured value smaller than 0 5mA ollo o Evaluation No Description o o Yes No Remark All values inspection within the permissible range passed EJ ol In the event the inspection result did not pass LlRepair Sorting Next inspection Signature of tester 42 43 o f f bock Hermann Bock GmbH Nickelstra e 12 D 33415 Verl Germany Phone 49 0 52 46 92 05 0 Fax 49 0 52 46 92 05 25 Internet www bock net C E Mail info bock net Our SALES PARTNERS Our business partners pursue the same strategy as we do quality innovation and above average standards that are internationally recognized You can rely on our business partners as you can rely on us Please note that only our authorised personnel and our sales partners can provide training supply of spare parts repairs safety technical controls TSC and other services Otherwise all warranty claims will be void Here is a short list of our business partners Australia alphacare www alpacare com au Belgium AXAMED nv sa www axamed be Denmark Medema Danmark AS www medema dk UK Carebase www carebase net Estonia ITAK Ltd www itak e
5. and proper maintenance and cleaning To prevent damages to the cables wiring should be conducted outside of the area in which damages could be caused Furthermore avoid touching the sharp parts To pre vent injury through an electric shock avoid the possibilities of too high contact voltages These circumstances may especially be the case if the power cable is damaged if inad missible and excessive leakage currents exist or if liquid was spilled in the motor hous ing e g during improper cleaning This damage can cause malfunction of the control which could result in unwanted movements of single bed elements posing a risk of inju ry for the operator and the person in need of care 17 4 Drives 4 1 24 Volt drives H Bock GmbH equipped health care beds with the drive systems of the companies Dew ert and Limoss 4 2 Drive systems Each drive consists always of four main components Housing Motor Gear Lift pipe and clevis The housing principle of the individual drive guarantees the permanent function of all drive components Due to a detailed interior structure the construction of the housing interior creates an essential prerequisite for the precise integration of the drive technology The individual drive is characterized through a particularly easy to follow assembly disassembly design and also an easy installation of the control electronics The drive has a primary fuse in the mains plug 4 3 The external switch mode power
6. and ready handle Hang extension piece in the lower leg bracket 7 Do not slide the foot end piece further than the centre of the frame of the lying surface 8 IMPORTANT Make sure to read the labels attached on the top and bottom of the side rails so these may not be confused with each other 9 Hook the side rails into the pre assembled metal guides and connect them 10 Push the end piece into the lying surface as far as it will go and fasten the push and ready handle ae ee gt Hazard note by Bock When using accessories on the bed or medically necessary devices as infusion stands in close proximity to the bed ensure particularly that there are no risks of crushing or shearing for the caregiver when adjusting the back and leg rests 30 6 3 Assembly accessories The following standard equipment can be com bined with the bed models domiflex 2 and 1 Zlllllllll domiflex 2 wash E f Side rail extension Scope of delivery includes Fully assembled side rail extension Open the plastic cap plug in the side rail extension position it in the centre and close the cap Please make sure that the release button of the side rail extension faces outwards Important note The side rail extension from Bock are designed for the use along with all wooden side rail models from Bock Bock assumes no liability for damages arising from the use in combi nation with third party products Lift
7. bed and legible olol Operating manual available cope Is the safe capacity working load weight of patient plus mattress weight plus accessory weight as mentioned on the type plate observed Electric components Power cable connecting cables and plug without scratches dents kinks porous parts or bare wires Strain relief firmly fastened and efficient eo Housings of motors and hand control without damages Correct and secure cable leading and cable connections o Motor lift pipes without damages Chassis with scissors construction beds end pieces of actuator beds Chassis construction free of defects with no ruptured welding seams 10 Castors and bumper rollers if available without damages o ii Plastic end caps and mechanical connecting elements screws bolts etc complete and without damages Lying surface and end pieces Sprung wooden slats aluminium steel bars carrier plate or springs without damages No cracks no fractures tight fit enough pressure etc a ruptured welding seams S bolts etc complete and without damages ER m N 13 4 1 Bed accessories lifting pole stand up help triangle handle etc safely attached and without any signs of wear and tear Side rai Distance between side rail shafts not more than 12 cm o Height of side rails above the mattress at least 22 cm o fot n 00 Distance between end pieces and side rails and or d
8. described steps on the opposite end The side rail is now in a low position If the side rails should be placed in the top position to aid in fall prevention reach with the fingers in the centre of the upper side rail shaft of the gripping groove and pull the side rail upwards until you hear it click into place at both ends The side rail is now in a pulled up position Fig 1 Fig 2 10 The side rails serve first and foremost as an aid to fall prevention If very thin persons in need of care are no longer sufficiently protected by the side rails take additional protec tive measures e g attach additional push fit side rails padding accessories The dis tances of the continuous side rails must be less than 12 cm When using the continuous side rails they may not remain in a diagonal position Top tip by Bock Hazard note by Bock Use only original Bock side rails which are available as accessories for every health care bed Use only technically flawless and non damaged side rails with the allowed gap dimen sions Make sure that the side rails are engaged securely Before installation of the side rail and each new use inspect all mechanical parts on the bed frame and side rails which secure the side rails for any possible damage The operation of side rails should be done with great care Fingers can be quickly pinched between the longitudinal pieces 11 Fig 1 Continuous wood and steel side rails divided into two
9. disabling function in hand control As it is our goal to satisfy every requirement of the person in need of care Hermann Bock offers practical and independence promoting accessories The installation is done in a quick and easy manner using the fixing points on the bed that have already been prepared for this purpose It goes without saying that every element of our additional equipment range meets the special quality and safety standards of Bock In addition to the standard accessories included in the basic equipment the customer can also choose from our variety of accessories which is available for each bed model These optional accessories vary depending on the bed model They range from technical elements around mattresses up to the occasional extra bed A wide offer of wooden finishes and a variety of colours allow for the harmonious integration of each health care bed with any kind of furniture 6 1 Special dimensions Special dimensions make up an essential part of the manufacture at Hermann Bock Optimal lying comfort for persons in need of care who have a particular physique can only be achieved by means of custom built models With its customized models Her mann Bock enables customers to have their health care bed tailored to fit the individual ly physical requirements of the person in need of care For body heights up from 190 cm Hermann Bock recommends the use of a bed extension that allows an extension of the lying surface to
10. disinfec tion Use only mild and gentle disinfection methods to retain the material resistance of the plastic elements such as the drive housing decorative elements ripolux and ripolan springs Concentrated acids aromatic and chlorinated hydrocarbons as well as deter gents containing highly concentrated alcohol ether ester and ketone may damage the material and should therefore be avoided 7 3 Avoidance of hazards Please make sure to consider the following guidelines with respect to the electrical com ponent parts of your health care bed These guidelines are very important to avoid haz ards related to cleaning and disinfection Non observance of these guidelines may result in considerable damage of the electrical lines and the drive 1 Pull the mains plug and position it in such a way that contact with excessive amounts of water or detergents can be excluded 2 Check all plug connections for correct position according to the instructions 3 Check the cables and electrical component parts for damage Should you detect any damage do not perform any cleaning operations but first have the defects repaired by the manufacturer or an authorised licensed electrician 4 Before starting the operation check the mains plug for residual moisture and dry or blow out the device if necessary 5 On any suspicion of the intrusion of moisture into the electrical components disconnect the mains plug immediately and do not re establish the connec
11. inside the patient s room or to allow access to the patient gt is suitable for persons in the need of care who are at least twelve years old and 150 cm tall gt under certain circumstances can be used if necessary for medical purposes with other electric medical equipment e g suction devices ultrasonic humidifi er food systems anti bedsore systems oxygen concentrators and similar de vices In this event disable all bed functions for the duration of the application via the integrated disabling function Attention The bed has no special connection options for a potential equalisation Electrical medical devices connected to the patient s intravascular or intracardiac system must not be used The operator of the medical products has to ensure that the combination of the equipment meets the requirements of EN 60601 1 2006 20 Making model series domiflex 2 ready for use Remove all packaging from the bed and place bed on a free and flat surface Remove the side rails and the divided lying surface from the transport bracket and put aside Tilt the end pieces with the connected transport bracket on an end piece To protect the surface lay a blanket or cardboard underneath guality made in Germ ny Loosen the tool free connection Remove the upper end piece as well as the transport bracket 21 q gt i Gijs i Stressed Released eee le ws pm 7 t Pla
12. safe operation over a long period of time With proper operation and maintenance this health care bed has an expected service life of 7 to 10 years The service life depends on operating condi tions and frequency Attention Unauthorised technical changes to the product void all warranty claims This product is not approved for the North American market particularly not for the United States of America USA Distribution and use of the health care bed in these markets including through third parties is prohibited by the manufacturer 1 5 Type plate example Model xxx O Date of manufacturing xX XX XXXX a Serial NO XXXXXXXXX XXX C z Xxx V xx HZ max XA 8 ED xx x min ON xx min OFF R E Motor protection class IPX x N xk xx kg A Z Hermann Bock GmbH Nickelstr 12 D 33415 Verl Tel 49 0 1805 262500 o 1 Model designation 2 Manufacture date Day month and year 3 Serial number Order number running number 4 Mains voltage mains frequency and power input 5 Duty cycle 6 Drive protection class 7 Maximum patient weight safe capacity 8 Manufacturer 9 Symbols located on the right side IPX 4 Standard IPX 6 domiflex 2 wash C Conformity mark according to the medical device directive IPX4 Protection of electrical equipment against splashing water IPX6 Protection of electrical equipment against strong water jet no high pressure o Medical application part type B Use onl
13. supply The plug in part of the external switch mode power supply SMPS is an electronic trans former which warms up only to a minimum degree under load and it is equipped with electronic performance monitoring The result is a constant voltage up to the maximum load no loss of speed and a high level of protection against overloading The external transformer ensures safety right from the socket because it converts the voltage directly into the safety low voltage of 29 35 V This is the voltage that operates the bed There fore the health care bed has no components carrying 230 V Only when using the hand control the health care bed is subjected to a low voltage of 29 35 V It is connected via a plug coupling to the mains supply line feeder cable and can be replace separately if defective The plug in part of the external switch mode power supply complies with the future European directives for electrical household appliances In standby mode it also has a low energy consumption of maximum 0 5 Watt and can be used internationally with variable input voltages from 100 V to 264 V Electromagnetic alternating fields are not measureable on the SMPS adapter and in operation still lower than mains isolation due to the very small DC voltage The external switch mode power supply 18 A C CEC A A h la E ARA Al AACO FA r Fo IB J 5 1 Technical data domiflex 2 Technical data domiflex 2 O 1 1 max Angle of incidence to hor
14. TS eS ee eS 14 3 1 ONV av aceersnee Sarria co eae sauna ws se Cae poate E aan aeauaa eounaaS 14 S52 keveladjustiment driven E E E 14 3 3 Lockable hand control first fault protected sessneessssesennnsssserrereessssrrereess 14 34 Handcontrol symbols asss ANE 15 3 5 Hand control lock functionS eesssssesssesessrrrsesrrrrsesrreressrrreesrrrreessreressereresseereee 16 30 Caution Electric dYa a O NA 17 ETC AE E AE E EAE E E A R E AE E E AE EA oe ae 18 AL lt 2A NOMA CS neson aaa a A 18 4 2 DIEING SYSTEMS srate a T T O ees 18 4 3 The external switch mode power supply sssessssssesssssenssssrerssserersssereesssreresse 18 3 AsseMOyV NG O e O r E st EAR E E ERAR RSSRA e 19 Sil Technical dalina a a a a a 19 5 2 Mod lseries domillexX 2 icom orrara e EA TA 20 5 3 Trans DOMUSYVSLCIM eea Ea E E ETAR 24 5 4 Emergency lowering back part or foot Part esss sssseseesssreresssrersssererssserreeo 26 Bio IDISASSEMIDIY ai E ETTAN 27 5 6 Transport storage and operating CONCITIONS ccccccssecccceseeceeesecsaeeseesaaees 27 5 7 FUNCUON NOLES asra A N Momaenluianes 27 Bee DEPOSI NOE EANNA 27 Deo TOUDIESNOOLNE iannau a E N 28 D ACOS SO ROS a A a te ates 29 OAL SPECla I CIMMCN SIONS reenen naen poaasaeeteaunee 29 6 2 ASSEMIDIV edextensiohS esrin ae E 30 63 Assembly accessories nsari a A AEN ica deealscbsteancadoedat 31 6 4 Mattresses eniai r r ETE T A T O E T 33 7 Cleaning maintenance and disinf
15. a ce Installation Instruction and User Manual eo P domiflex bock domiflex 2 domiflex 2 wash Dear valued customer Thank you for purchasing a health care bed from Hermann Bock GmbH You made a good buying choice because this health care product has a long product life and provides superior functionality and the highest level of safety Our electrically operated health care beds guarantee optimal ly ing comfort and allow professional care at the same time This product was designed with a focus on the elderly whose confidence must be rein forced and whose life needs protection With this health care product we meet these requirements We urge you to prevent potential malfunctions and the risk of accidents by complying strictly with the safety and operating instructions and by carrying out the necessary maintenance Sincerely yours Wos Kuch Klaus Bock Content 1 Preface and general instructionS iida cs sccrcesd esede st teenth vadeds at calorie rasedeas adeeb eeetansseeaks 4 1 1 Ntended USE engen SS 4 12 DeMMtOnror TNE USEF eroU Deena E E 5 To Salte INSUPUCTIONS aerer Nr EAN 5 TAS Service WIE warranty sisieaied lt onsavassddestersorzucausededied songuvadsddedtae sora cauhobedued sonasusdsdvecue 7 LS Typerplate Exaile itscccsizsceivesstacviaieicdinond a aa easa airia 8 2 General description of the TUNCTIONS vi viciccndeeverates vase satvsvtusvevvedeewsvedduseexeesdenvaenduasewoneus 9 3 1 DANS Sem SP Oe S
16. a length of up to 220 cm That way a high degree of lying comfort can also be ensured for taller persons while the functionality remains the same Hazard note by Bock Do not try to fix failures on the electrical equipment itself It could be fatal Either call the customer service of Hermann Bock GmbH or an authorised licensed electrician who conducts the troubleshooting in compli ance with all relevant VDE regulations and safety regulations Top tip by Bock The representative of the service hotline of Hermann Bock are looking forward to informing you about the best retrofitting solution for your bed Hotline no 0180 5262500 14 cents min for calls from landline phones 42 cents min for calls from mobile phones A wide product range of auxiliary furniture complements the various bed models up to the complete interior design of your home This combination creates a care and living comfort resulting in perfect harmony 29 6 2 Assembly bed extensions The delivery for a lying surface extension consists of the following parts 2 adapter unit for the left and right foot part 1 extension piece for the foot part 1set side rails How to assembly in the simplest way Remove the mattress from lying surface Loosen the foot end piece and pull slightly from the lying surface Dismantle the side rails Remove the foot end piece Plug adapter units into the frame of the lying surface at the foot end and fasten with the push
17. ce the end piece with the blue marking on the motor flat on the floor Next insert the frame of lying surface head part on the mounting latches You can recognise the head part easily by the round mountings for the lifting pole Create a solid connection by fastening the push and ready handle Insert the frame of the lying surface foot part onto the already assembled frame of lying surface head part Create a solid connection by fastening the push and ready handle Plug the plug connector of the drive motors in the sockets of the control box Observe colour cod ing Then turn the mounted lying surface with the head end piece so that the head end piece is in a vertical position Push the mountings at the foot end piece into the lying surface Do not push as far as it will go allow 5 cm protrusion 22 Insert the continuous wood or steel side rails at the head end piece in the mountings Install the side rails in diagonal position on the side rail mountings as shown in the picture Note the markings of side rails as they fit only on the corre sponding position either up or down Next slide the upper side rail on the mounting of the foot end piece and connect in a longitudinal direction Install the other side rail too Push the foot end piece into the lying surface as far as it will go and fasten the push and ready handle The plug connector of the drive motors and the hand control are colour coded Inser
18. cket key B Release hand control keys C Patient function head lowering position n Trendelenburg locked D Release hand control keys Special design Hazard note by Bock All drive components must not be opened Troubleshooting or exchanging single electrical components may be performed only by special qualified personnel Hazard note by Bock Simultaneous use of electrical appliances particularly in the vicinity of the operational bed may result in small electromagnetic interactions of these electric devices e g static noise in the radio In such rare events increase the distance of the devices Do not use the same socket or temporarily switch off the interference source and or the disturbed device if the bed should be operated with electrical medical equipment contrary to its intended use the func tions of the bed must first be disabled via the integrated locking function in the hand control for the duration of the application 16 3 6 Caution Electric drive Hermann Bock calls the electrically operated nursing and therapy beds health care beds because through their multiple functions they allow the person in need of care to sup port the healing process mentally and physically and relieve pain at the same time Elec trically operated beds that are medical products need special care in regards to constant safety checks This includes safety conscious handling of the bed daily inspection of electrical equipment
19. e Finland RESPECTA OY www respecta fi Greece Wheel Rehabilitation Products www wheel gr Israel Israel Quality of Life Center www iqlc com Italy Enrico Spadoni www bock net Croatia BEZ LIMITA d o o www bezlimita hr Lebanon ALBERT MASSAAD s a r l www albertmassaad com Luxembourg Stoll www matelas lu New Zealand The Netherlands Cubro Ltd www cubro co nz Eureva B V www eureva nl Norway Medema Norge AS www medema no Austria Reha Service GesmbH www rehaservice at Poland Timago International Group Sp z 0 0 www timago pl Portugal MACHADOS www machadosmadeira com Romania Donis srl www donis ro Russia Lazerlink Switzerland Sodimed www sodimed ch Serbia Proxi Med d o o www proxi med co rs Slovenia Medimaj d o o www medimaj com Slovakia Servis Invo www servisinvo sk Bock Spain Ferran Asensio Jou www bock net es Czech Republic Ortoservis www ortoservis cz Ukraine ADS Ukraine registered trademark 890 02174 as of 09 2015 Subject to technical changes
20. e zipper and or the hook and loop fastener Tray 4 0 kg Delivery includes Tray 1 piece The tray is applied to the side rails and prevented from getting out of place using two distance pieces Universal clamp 0 6 kg Delivery includes Clamp 1 piece fastening ring 1 piece The universal clamp is a special holding appliance that provides flexible positioning of the modular functional equipment It is optionally possible to attach fixtures for urine bottles infusion systems or a lamp on this clamp Furthermore the universal clamp can be shifted along the side rails according to prefer ence or requirement The universal clamp is clamped onto the upper side of the side rails and fixed with the fastening ring Assist handle with cross bar for actuator beds 3 0 kg Delivery includes Assist handle 1 piece Support cross bar 1 piece Screws 4 mm 4 pieces Hold the support cross bar to the frame of the lying surface and Pi mark with a pen the drill holes on the frame _ e e 32 Drill holes into the previously marked spots 3 5 mm of the lying E surface frame Use the provided screws to fasten the cross bar to the lying surface frame Push the assist handle into the support cross bar adjust it to the desired position and fasten it tightly 6 4 Mattresses In general foam and latex mattresses are suitable for the Hermann Bock health care beds A volumetric weight of at least 35 kg m3 is requir
21. ection ssssssssssssssssererererererererereresesesesesesereserere 34 7 1 Cleaning and Ma iNbe aN CO vices wraivcscbasttetseeaesccestaneoosweeesdehsaeeeeweusteaesaaaaveasernnamesstes 34 7 2 EIS WATE CUO a co satiandaddauansasiedcusacdeweasannavaernadauedsadsaes oladeudarsnaceustadabanodsusstaudeuhee aa 35 Fo AVvoidanceof Naza Sevres gacisasnuceeveds a ads aeuwacenstuavsdadaunmienssdatidnts 35 7 4 Cleaning in cleaning facilities osscak ota eversnsee sented cavearscdedeasadacesanaacy soqisaneseaiestesenbadeas 35 8 Guidelines and manufacturer s declaration ccceccecceccecceceececsececeesecseeseeseeees 36 9 Declaratio mOr CONTONMICY aeara teats sae aes 38 10 Regular technical safety INSPECTIONS with service ccc ecceceeceececeeceeceeeeeceeeeceeeess 39 1 Preface and general instructions The various bed systems from Hermann Bock meet special requirements for the use in care and treatment facilities as well as for home care Reliable functionality and a long product life make each bed particularly valuable Our beds need little maintenance with proper operation and care Each bed from Hermann Bock must pass quality testing in a final inspection before it is shipped anywhere The beds are manufactured according to the current standards for medically used beds and tested accordingly Our beds comply with Standard EN 60601 2 52 2010 The electrical building components comply with safety standard EN 60601 1 2006 for medical devices
22. ed along with the dimensions of 90 x 190 cm 100 x 190 cm 90 x 200 cm and 100 x 200 cm The max height with steel aluminium spring wood lying surfaces must not exceed 15 cm Lying surfaces of spring systems may not exceed a height of 12 cm As a general rule the mattress height must not be less than 10 cm An additional detachable rail available as an accessory must be used when using higher mattresses When using foam mattresses we recommended the use of a cut foam mat tress to allow a better combination with the lying surface 33 7 Cleaning maintenance and disinfection The individual bed elements consist of high quality materials The surface of the steel tubes is covered with a durable PES powder coating All surfaces of the wooden parts are surface sealed with an ecologically coating that is low on harmful substances All bed elements are easy to clean and cared for using wipe and spray disinfection means ac cording to the applicable cleaning requirements with respect to the various areas of application Observing the following care instructions will retain the usability and visual appearance of your health care bed for a long time to come 7 1 Cleaning and maintenance Steel tubes and vanished metal parts Please use a wet wipe and a regular mild household detergent for the cleaning and care of these surfaces Wooden decorative and plastic elements All standard furniture cleaners and cleaning detergents can be
23. ergency lowering is carried out by loosening the locking pin on the motor mountings Furthermore power adjustment allows for constant speed of the functions Therefore the safety functions comply with protection class Il and the moisture barrier protection type IPX4 domiflex 2 Wash IPX6 The maximum duty cycle is specified on the type plate of the bed For example 10 2 min ON 18 min OFF means that any electronic adjustment can be done within maxi mal 2 minutes After that the unit switches off for 18 minutes protection against over heating If the maximum setting time of two minutes is exceeded by two minutes e g someone plays continuously with the hand control which could lead to overheating of the ser vomotors the thermal fuse immediately shuts off the power supply to the bed After a cooling down time of approx one hour the power will be again automatically supplied 3 2 Level adjustment drive The level adjustment of the bed takes place via two manual controls with built in indi vidual DC drives Their adjustment range is defined by an integrated control limit switch The level adjustment drive is connected to the control box via a helical cable 3 3 Lockable hand control first fault protected The base functions can be controlled safely through the eight extra large keys on the hand control The individual keys are marked with corresponding symbols The servomo tors run until as long as a corresponding key i
24. fety of the per son in need of care as well as for the relatives and the caregivers nursing staff to avoid potentially dangerous situations This requires the correct installation and operation of the bed The individual physique of the person in need of care as well as type and the extent of their disability must be taken into account by all means when operating the bed Avoid dangers accidental motor adjustments and incorrect operation by using the disa bling function When the operator e g the nursing staff caregivers or the care providing relative leaves the room the entire operating functions of the bed should be disabled This is achieved by operating the key of the hand control First lower the lying surface to the lowest position and activate the lock function with a twist of the key located in the keylock on the backside Remove the key and check the function of the hand control for safety reasons Make sure that it is indeed locked These recommendations apply particularly if the person in need of care cannot operate the hand control safely themselves due to certain disabilities if the person in need of care or the caregivers could be at risk due to those acci dental adjustments if the side rails are in a raised position and there could be danger of trapping and crushing if children are unsupervised in the room with the bed Always make sure that the hand control when not in use is securely hooked in the suppo
25. ge 150 kHz to 80 MHz field strengths should be less than V V m Recommended separation distances between portable and mobile RF communications equipment and the medizinisches Bett The medizinisches Bett is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the medizinisches Bett can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the medizinisches Bett as recommended below according to the maximum output power of the communications equipment Separation distance according to frequency of transmitter m 150 HBG 89 MHGG0 Mhz 800 MHo 2 5 6Hz She di oe 3 10 100 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts VV according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorpti on and reflection from structures objects and people 37 9 Declaration of conformity Manufacturer Product description model Hermann Bock GmbH domiflex 2 Nickels
26. h mechanism In case of power failure the back and or leg part can be lowered through disconnecting the motor mountings locking pin The chassis The height adjustment of the bed takes place through two height adjustable manual controls The surface of the tubular steel structure is stove enamelled with a PES powder coating The side rails Any health care bed can be equipped on both sides with integrated side rails at a special safety height The side rails can be lifted and lowered through a rail The sliding pieces run particularly smoothly and quietly with an impact damper and the ends are fitted with attractive looking caps The side rails can be easily operated through an ergonomi cally designed release button Depending on the model shorter or longer partitioned side rail variations are available Operation of the continuous side rails The release button for the adjustment of the continuous side rails is located on the up per side rail shaft in the guide rail Fig 1 If lowering of the side rails is desired reach with your fingers in the gripping groove of the upper side rail shaft which is designed for this purpose and lift the side rail slightly while pressing on the one sided release button on the head or foot part Fig 2 The side rail opens at the corresponding place and can be easily lowered downwards as far as it will go Fig 3 The side rail is now diagonal To lower the other side as well carry out the previously
27. hp d 5 WP 3 7 D d sja P 800 MHz to 2 5 GHz where p is the maximum output power rating of the transmitter in watts VV according to the transmitter manufacturer and dis the recommended separation distance in metres m 80 MHz to 800 MHz Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol G NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordiess telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the medizinisches Bett is used exceeds the applicable RF compliance level above the medizinisches Bett should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the medizinisches Bett b Over the frequency ran
28. ing pole with triangle handle 5 5 kg The capacity of the lifting pole is 75 kg Scope of delivery includes Lifting pole with hook up loop 1 piece Triangle 1 piece Place the lifting pole with triangle handle in the provided loop at the head part and adjust it accordingly Next hang the triangle into the hook up loop Use only mattress with a required mattress height as described by company Bock You find this information in section 6 4 ATTENTION The lifting pole with a triangle handle must not swivel outside of the lying surface When used in line with its intended purpose the service life of the triangle handle should at least last 5 years If a lifting pole with triangle handle installed to the bed it must be tested at each safety technical control Hazard note by Bock For safety reasons use only original Bock accessories when furnishing your health care bed further Those accessories must be approved by company Bock for the respective bed model You will find a detailed overview of accessories and extras for your bed in the separate datasheet Hermann Bock assumes no liability for accidents damages and hazards arising from the use of other accessories 31 Side rail padding 1 4 kg uy Delivery includes 1 piece covering material _ 1 piece padding Open zipper of the padding and pull padding from the top over the side rails Pull the foam padding from the inside of the bed into the fabric and close th
29. istance be sa tween divided side rails less than 6 cm or greater than 31 8 cm ejeje WO N 41 Address o AA Place and postcode cw Too bock Station room oo e C e a a ee Functional testing No Description o Yes No Remark Side rails ES Smooth running of the side rail in the tracks and safe locking into afo place Sufficient mounting and firm seat of the side rail shafts parts fol ofl Load stress test of the side rails without deformation ee a Lying surface 4 witaayeesecat ea retors Polo eeu an D 5 enumeran neve olo oo even under stress Is the clamping effect of the 6 eccentric clamps domiflex 2 domiflex 2 wash sufficient If not the lock nut possibly could use a bit more fastening Chassis with scissors construction beds end pieces of actuator beds Hub adjustment properly and without any obstacles olol Safe braking effect blocking and free running of wheels olol o Electric components A a aa es Lolo properly without any defects O moe n doene a e a properly and without any defects Electric measuring Description Yes Measured value Insulation resistance must be only measured on old models before manufacture year of 2002 Insulation resistance measured value larger than 7MQ fo oOo Device leakage current It must be measured with each TSC Only it is a domiflex 2 item no 271 or domiflex 2 wash item no 272 with a Limoss 24
30. izontal 190 155 back part w Lower leg part 15 39 9 7 Trendelenburg position OO o ae oe Side rail height from lying surface cm 3a f 89 Side rail options ee Continuous wood side rail Continuous steel side rail Lifter space cm Sound level dB A lt 65 Le Pe Total incl continuous Wood side rail kg Total incl continuous Steel side rail kg Lying surface back part kg Lying surface foot part kg Control actuating device including motor end piece kg Continuous wood side rail kg set Continuous steel side rail kg set Electric data ass 100 240 Input voltage V 100 240 100 240 Limoss 0 5 0 5 9 9 2 cai a F soo 2 0 1 2 All parts and data are subject to a constant further development and therefore may differ from the mentioned data 19 1 back part 2 fixed seat part 3 Upper leg rest 4 Lower leg rest 5 2 Model series domiflex 2 Domiflex model series 2 consisting of models domiflex 2 and domiflex 2 wash were specifically designed for the requirements of daily use of the product as part of home care The above mentioned models provide high comfort to frail people patients who need care and people with disabilities These models offer high lying comfort and sup port through easy operation as well as optimal care Model series domiflex 2 gt is not suitable for use in hospitals gt is not suitable for patient transport The beds must be only moved for cleaning purposes
31. led downwards Once the second person has left the danger zone the first person can lower gently the back part Hold the back part by all means until it is fully lowered Hazard note by Bock Emergency lowering may be only carried out in an emergency by people who safely master this operation 26 5 5 Disassembly Pull out the mains plug before disassembly Disassembly of domiflex 2 takes place in reverse order of the assembly see manual Change of location Ifthe bed must be moved to another location please follow these safety in structions Bring the lying surface to the lowest position Before proceeding pull out the mains plug and attach with the suspension de vice to the side rail or the end piece to secure the power cable against falling and being crushed by anything Make sure that the cable is not dragged over the floor Before inserted the mains plug again inspect the power cable visually for me chanical damage dents and kinks abrasions and bare wires Place the power cable in a way that it will not be rolled over or strained during the op eration of the bed or could be damaged when inserting the mains plug again 5 6 Transport storage and operating conditions Transport and storage Operation Temperature 0 C to 40 C 10 C to 40 C Relative humidity 20 to 80 20 to 70 Air pressure 800hPa to 1060hPA 5 7 Function notes To keep the bed in one location you must block the b
32. mestic and CISPR 11 partly those directly connected to the public voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions Class A IEC 61000 3 2 Voltage fluctuations flicker Complies emissions IEC 61000 3 3 Electromagnetic immunity The medizinisches Bett is intended for use in the electromagnetic environment specified below The customer or the user of the medizinisches Bett should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge 6 kV contact 6 kV contact Floors should be wood concrete or ceramic tile ESD If floors are covered with synthetic material the IEC 61000 4 2 8 kV air 8 kV air relative humidity should be at least 30 Electrostatic transient 2 kV for power supply lines 2 kV for power supply ines Mains power quality should be that of a typical burst commercial or hospital environment IEC 61000 4 4 1 kV for input output lines 1 kV for input output lines Surge 1 kV differential mode 1 kV differential mode Mains power quality should be that of a typical IEC 61000 4 5 commercial or hospital environment 2 kV common mode 2 kV common mode Voltage dips short lt 5 U lt 5 U Mains power quality should be that of a typical interruptions and voltage gt 95 dip in U for 0 5 cycle gt 95 dip in U for 0 5 cycle commercial or hospital envir
33. ng including check and observance of the respective inspection terms for you for a fee The company Hermann Bock GmbH stipulates that a safety technical inspection is to be executed with the first use and at least once a year and with each re use of the bed To facilitate the execution of all necessary safety inspections Hermann Bock GmbH provides you with the TSC checklist for the regular functional testing which can be found in the assembly and operation manual Please make a copy of the checklist and use it as a form for your functional testing The checklist serves as evidence report of the performed inspection and must be kept on file You can also download this TSC checklist from our website www bock net Attention If you suspect or recognise an error deterioration or malfunctions immediately dis connect the bed from the mains remove it from service and label it out of order A waiver of the annual leakage current measurement is only permitted with health care beds of the series domiflex 2 and domiflex 2 wash manufacture started on 1 July 2015 by Hermann Bock GmbH which are equipped with a 24 V drive system SMPS from Dewert or Limos All 230 volt parts of this drive system are outside of the health care bed In regards to the TSC conduct visual and functional testing If the testing was 39 passed for the electrical components the leakage current measurement can be omit ted during the service life of the health care bed
34. o not designed to transport patients The beds must be only moved inside a patient s room for cleaning purposes or to allow access to the patient This user manual contains safety instructions All persons working with the beds must be acquainted with the contents of these instructions Improper operation can result in personal injuries 1 2 Definition of the user group Operator Operator e g medical supply stores specialist shops facilities and budget holders is any natural or legal person who uses the beds or has the bed used The briefing on the use of the product shall generally be conducted by the operator User Users are persons whose training experience or briefing on the product allows them to operate the health care bed or carry out works on it The user is able to recognize possi ble hazards and or to avoid them and to assess the health condition of the patient Patient resident Persons in need of care as well as disabled and fragile people who are lying in the health care bed Qualified personnel Employees of the operator are referred to as qualified personnel They are entitled to deliver the health care bed assemble dismantle and transport it on the basis of their training or instructions These persons must be instructed according to the guidelines concerning the cleaning and disinfection of the health care bed 1 3 Safety instructions The intended use operation of all moving parts is as important for the sa
35. onment If the user variations on power supply of the medizinisches Bett requires continued input lines 40 U 40 U operation during power mains interruptions it is IEC 61000 4 11 60 dip in U for 5 cycles 60 dip in U for 5 cycles recommended that the medizinisches Bett be powered from an uninterruptible power supply 70 U 70 U or a battery 30 dip in U for 25 cycles 30 dip in U for 25 cycles lt 5 UT lt 5 U gt 95 dip in U for 5 sec gt 95 dip in U for 5 sec Power frequency 50 60 3 A m 3 A m Power frequency magnetic fields should be at Hz magnetic field levels characteristic of a typical location in a IEC 61000 4 8 typical commercial or hospital environment NOTE U is the a c mains voltage prior to application of the test level 36 Electromagnetic immunity The medizinisches Bett is intended for use in the electromagnetic environment specified below The customer or the user of the medizinisches Bett should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Conducted RF Portable and mobile RF communications equipment should be used no closer to any part of the EQUIPMENT medizinisches Bett including cables than the recommended separation distance calculated from the equation applicable to the IEC 61000 4 6 frequency of the transmitter Radiated RF IEC 61000 4 3 Recommended separation distance d
36. parts D21895 H2220 D21895 All dimensions in mm Depending on the length of the lying surface The dimension in brackets is optional Legend Distance between the head part and the side rail Height 1 of the side rail Height 2 of the side rail Width 1 of the side rail Distance between the elements within the side rail Distance between the divided side rails Distance between the lying surface and the upper edge of the side rail Height of the top edge of the side rail above the mattress without compression Thickness of the mattress for the intended use Width 2 of the side rail Smallest dimension between side rail and lying surface or panel if any Distance between the foot part and the side rail ors toe VO oO 2 D 12 Item numbers Designation item no Fig 1 Continuous wood side rail divided into two parts Wood side rail set 91907 Fig 2 Continuous wood side rail divided into three parts Wood side rail set 80410 Fig 3 Continuous steel side rail divided into two parts Steel side rail set 91908 13 3 Electric parts 3 1 Drive The drive unit consists of individual drives A switching power supply with a rectifier that converts the input voltage into low voltage power of 35 VDC direct current is part of the external control The motors and the hand control function with this non hazardous low voltage The cables are double insulated and the mains plug has a primary fuse Internal em
37. rakes on castors of the chassis To accomplish this use your foot to move the locking lever on the chassis downwards If necessary the integrated side rails must be pulled up until they lock into place With different mattress thicknesses the minimum height must not be below 22 cm taken from the upper edge of the side rail above the mattress excluding any compression if the minimum height is not observed we recommended the installation of a third de tachable rail 5 8 Disposal Each of the components made of plastic metal and wood are recyclable and can be disposed recycled in compliance with the relevant legal provisions Please note that electric adjustable health care beds or nursing beds are considered commercially used electronic scrap according to the WEEE EC directive 2012 19 EC b2b All replaced elec trical and electronic components of the electrical adjustment system must be disposed 27 properly in accordance with the requirements of the Waste Electrical and Electronic Equipment Act in short ElektroG When disposing components the operator has to make sure that none of the replaced and disposed components are infectious or con taminated 5 9 Troubleshooting This overview helps you to detect and correct malfunctions on your own and explains what kind of malfunctions require the consultation of suitably qualified service person nel Malfunction The drive units cannot be controlled via the hand control When bu
38. rt hook at the bed and cannot drop As a general rule the bed should be operated by instructed nursing staff caregivers relatives or in attendance of instructed persons When adjusting the lying surface it is particularly important to ensure that no limbs are placed in the side rails of the adjustment range If the side rails themselves are adjusted pay attention to the correct lying position of the person in need of care Prior to making any electrical adjustment it should as a general rule be made sure that no limbs are positioned in the adjustment range between the chassis and the head or foot part especially that there are no persons animals in the area between the floor and the raised lying surface The danger of being crushed is particularly high in these areas The permitted person s weight depends on the total weight of the equipment that has been mounted to the bed mattresses and other electronic medical devices For the safe capacity please refer to the type plate on the lying surface frame of the bed Attention The beds come with no special connection options for a potential equalisa tion Electrical medical devices connected to the patient s intravascular or intracardiac system must not be used The operator of the medical products has to ensure that the combination of the equipment meets the requirements of EN 60601 1 2006 1 4 Service life warranty This health care bed was developed designed and manufactured for
39. s pressed and held A coiled cable allows the necessary freedom of movement while operating With the rear mounted suspension unit the hand control can be attached to the side rail 14 3 4 Hand control symbols fe gt N X Back part upwards A Ss Back part downwards s gt Lower leg part upwards Lower leg part downwards 3 960900900090 Autocontour upwards 0 0 0 0 O 0 O O Q o Autocontour downwards 2o Lying surface upwards i Lying surface downwards n No Comfort sitting position up bock bock lt wards E Head lowering position oe 2 ste Trendelenburg Standard hand control with special functions foe 208 Reset Sun eum AIS Lightening Floor light Press keys Back part up wards and Back part down wards at the same time O Comfort sitting position just goes up All adjusted positions must be lowered separately Hazard note by Bock Do not exceed the maximum duty cycle of 2 minutes Observe a subsequent break of at least 18 minutes by all means 15 3 5 Hand control lock functions The hand control comes with an integrated disa bling function that can be activated and deac tivated with the corresponding key To disable the entire electrical function insert the key in the keylock on the backside and turn the lock function on or off with a corresponding twist of the key Cc A So
40. t them into the corresponding socket of the control unit Next tighten the strain relief cap again The blue marked plug connector must be located on the end piece of the head part Hazard note by Bock Check all screw and push and ready connections again before operating the bed Is the clamping effect of the 6 eccentric clamps sufficient If not the lock nut possibly could use a bit more fastening Hazard note by Bock Do not squeeze crush the cables Adjustment of moving parts may only be used for the intended use Hermann Bock GmbH assumes no liability for unauthorized technical changes 23 5 3 Transport system Versions Standard 3 health care beds on one pal let Transport system to be changed to the VAN or Wash version A aa xy k ANo ene 1835 mm To change the standard transport bracket to the VAN or Wash version loosen the lower horizontal push and ready handle and pull the end pieces on both sides ap prox 2 cm from the connector Remove the attached foot and head part frames and one end piece Next pull the long con nectors from the brackets to the top and insert them horizontally in the short con nectors Make sure that the welded con nections point towards the side on which the wire basked for the side rails is located Attach the previously removed end piece back to the connector and establish a solid connection with the push and ready han dles First a
41. tion Put the bed out of operation immediately attach an appropriate visible label and contact the manufacturer supplier 7 4 Cleaning in cleaning facilities Cleaning in cleaning facilities is only possible with the model domiflex 2 See the separately available brochure Item no 890 02160 Hazard note by Bock Use of abrasive cleansers and or detergents containing grinding particles cleaning pads or stainless steel cleaners for the cleaning is absolutely not recommended Neither use organic solvents such as halogenated aromatic hydrocarbons and ketones nor detergents containing acid or alkaline Never clean the bed using a water hose or high pressure cleaner as this might lead to the intrusion of fluid into the electrical component parts which causes malfunctions and hazards 35 8 Guidelines and manufacturer s declaration Electromagnetic emission The medizinisches Bett is intended for use in the electromagnetic environment specified below The customer or the user of the medizinisches Bett should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions Group 1 The medical used bed uses RF energy only for its internal function Therefore its RF emissi CISPR 11 partly ons are very low and are not likely to cause any interference in nearby electronic equipment RF emissions Class B The medizinisches Bett is suitable for use in all establishments other than do
42. tra e 12 domiflex 2 wash 33415 Verl Classification Choose conformity evaluation process Medical products class Annex VII of MDD norm 1 and 12 referring to appendix IX of MD Hereby we declare that the above specified products comply with the provisions of the directive 93 42 EWC concerning advice for medical products The entire associated documentation is kept in the premises of the manufacturer Applied standards Harmonized standards for which the proof of concordance can be supplied EN 60601 1 2006 Medical electric devices part 1 General definitions for the safety including the essential performance characteristics EN 60601 1 2 2007 Medical electric devices part 1 2 General definitions for the safety including the essential performance characteristics collateral standard Electro magnetic tolerance requirements and testing EN 60601 1 11 2010 Medical electric devices part 1 11 General definitions for the safety including the essential performance characteristics collateral standard Require ments for medical electrical devices and medical electrical systems for medical care in a home setting EN 60601 2 52 2010 Medical electric devices part 2 52 General definitions for the safety including the essential performance characteristics of medical beds EN ISO 14971 2012 Application of the risk management for medical products Verl 25 June 2015 eur XK 9 iS A tee 2 Klaus Bock Dr
43. ttach the foot part frame and then the head part frame to the transport bracket The sprung wooden slats must show in the cross direction to the end piec es and the head and foot support must be placed with the bended end upwards 24 Version VAN or Wash Depending on the available height transport the lifting pole and the side rails separately from the bed 1443 mm Version Eco 3 health care beds on one pallet NOT repositionable Can only be used for transport on pallets with appropriate strapping Hazard note by Bock Do not try to fix failures on the electrical equipment itself It could be fatal Either call the cus tomer service of Hermann Bock GmbH or an authorised licensed electrician who conducts the troubleshooting in compliance with all relevant VDE regulations and safety regulations Clean and disinfect the bed before using it again Also at the same time perform a visual inspec tion to check for any mechanical damages You will find details hereto in our Safety Guidelines 25 5 4 Emergency lowering back part or foot part In case of power or drive system failure you can lower the elevated back part manually Must be carried out always by two people One person lifts the back part slightly to take pressure off and holds it in this position As next step the second person removes the locking pin from the motor The motor is now separated from the back part and can be swivel
44. ttons are pressed the drive units stop after a short time The drives stop after a longer adjustment time Opposite functions when operating the hand control Individual drive units run in one direction only Drive units stop and bed re mains in a tilted position Potential causes Power cable is not connected No voltage in the socket Plug connector of the hand control not fixed firmly Hand control or drive unit defective Disabling function or control box in the hand control activated There is an obstruction in the adjustment range The safe capacity working load has been exceeded The adjustment time or safe capaci ty working load has been exceeded and the polyswitch in the transformer of the control unit has responded to increased heat Motor connector switched internally Hand control drive unit or controller defective Constant operation of adjustment func tions 28 Remedy Insert power cable Check the socket or the fuse box Check the plug in connection on the motor Notify the operator or Bock customer service Disabling function or control box in the hand control deac tivated Remove obstruction Reduce the load Allow the drive system to cool down sufficiently for at least one minute Notify the operator or Bock customer service Notify the operator or Bock customer service Move lying surface in bottom or top position as this will straighten it again horizontally Activate
45. used Using a wet wipe without detergent additives for the cleaning of the plastic elements should generally be sufficient For care of the plastic surfaces use a product that is specifically suitable for plastics Drive To prevent the intrusion of moisture into the motor housing we recommended using only a damp rag to clean outside housing Spring systems ripolux neo Use a damp rag without adding any detergents or if deemed necessary a detergent that is exclusively suitable for plastics and clean the spring elements made of plastics In case of heavy contamination remove the spring elements from the supporting elements and the supporting elements from the frame of the lying surface The dismounted plas tics elements can be rinsed or spray washed with hot water to get them clean For the disinfection the components should be sprayed with a detergent suitable for plastics Most of the moisture drips off the plastic surface by slightly shaking it while the rest will dry on its own within a very short time Remount the elements after they have com pletely dried If required you can also remove each of the individual lying surface ele ments completely from the frame to clean them Hazard note by Bock Clean and disinfect the bed before using it again At the same time perform a visual inspection to check for any mechanical damages 34 7 2 Disinfection All methods in accordance to the standard EN 12720 can be used for the wipe
46. y in dry rooms Protection class II double insulation insulated for protection Within the European Union this product must be disposed in a separate waste collection Product may not be disposed via the separated municipal waste Symbol for maximum patient weight Symbol of capacity Qpe XOP gt Symbol for observance of the user manual 2 General description of the functions Construction design and function The lying surface with 4 function areas The lying surface standard consists of a slatted comfort frame as an option it can be also provided with a metal lying surface or special suspension systems and is divided into four functional areas Backrest solid seat upper and lower leg rest The comprehensive lying surface frame is welded from a steel tube and stove enamelled with a PES powder coating The variable height adjustment of the lying surface is electric and stepless It is carried out by 24 V DC motors and controlled by the smooth keys of the hand control The backrest can be adjusted electrically The leg part consists of a foot support that is divided into two parts With a touch of a button on the hand control each individual position can be adjusted continuously The electronic hand control also allows an automatic 3 way function to set a stretched leg elevation and the chest and the knee bending part Availability depends on the model The motor of the foot part can be replaced if necessary with a manual catc
Download Pdf Manuals
Related Search
Related Contents
Electrónica Digital - Manteniment industrial.cat Hustler X-ONE General Service Manual Sun Ultra 20 Workstation User Guide - it Panas。nm 直管LED 施工説明童 JBL gto1201.1 II Car Amplifier User Manual Mode d`emploi Francis Online \(WebSPIRS 5\) Set Up Guide & User Manual ASSMANN Electronic AK-330109-010-S Copyright © All rights reserved.