Home

Grain Size Horiba Analysis

image

Contents

1. Blank Feed Circulate Alignment Blank Measure Please Add Sample Wait 10 Seconds Alignment Measure Measure Ultrasonic Please Add Sample Ultrasonic On Ultrasonic Wait Alignment Measure Ultrasonic On Add Sample Measure Ultrasonic Ultrasonic On Please Add Sample Ultrasonic Off Ultrasonic On Ultrasonic Wait Alignment Measure Measure then Measure Ultrasonic Please Add Sample Wait 10 seconds Alignment Measure Ultrasonic On Ultrasonic Wait Alignment Measure 16 Rinse Feed Circulate Drain Average 3 Repeat Measurements Save Average Average Data in Memory 1 2 amp 3 Move Average to Main Result Position Prompt Filename then Save Auto Initial Align Feed Circulate Initial Align Drain References Jiillavenkatesa A Dapkunas S J and Lum L H 2001 Particle Size Characterization U S Department of Commerce Chapter 6 Size characterization by laser light diffraction techniques Matthew Kirby CSFU Paleoclimatology Laboratory Detailed Lake Mud Grain Size Standard Operating Procedure 10 April 2009 Horiba Inc LA 920 Instruction Manual Horiba Inc LA 920 Users Manual Horiba Inc LA 920 Software Manual Document History Written by Laura Triplett for the LRC with contributions from Trevre Andrews Revised June 2007 by Laura Triplett Revised May 2010 by Jessica Heck 17
2. Listed below are the details for sequences available for use with the Horiba Many are standard sequences that come with the Horiba but some are custom programmed by LRC staff Additional sequences have been written for specific projects and are available for use 15 Start Up LRC Standards LRC Polydisperse Standard LRC Open Conditions Files Feed Circulate Alignment Blank Please Add Sample Ultrasonic On Ultrasonic Wait Alignment Measure Drain Feed Circulate Drain Run Feed Circulate Alignment Blank Please Add Sample Wait 10 seconds Alignment Measure Drain Feed Circulate Drain Run Ultrasonic Feed Circulate Alignment Blank Please Add Sample Ultrasonic On Ultrasonic Wait Alignment Measure Drain Feed Circulate Drain Run then Run Ultrasonic Feed Circulate Alignment Blank Please Add Sample Wait 10 seconds Alignment Measure Ultrasonic On Ultrasonic Wait Alignment Measure Drain Feed Circulate Drain This will produce two measurement files for the same sample It can be useful when trying to determine if you want to use ultrasonicator on your samples or not Look for the addition of 10 20um bubbles that may be a result of the ultrasonicator Ultrasonic On Add Sample Run Ultrasonic Feed Circulate Alignment Blank Ultrasonic On Please Add Sample Ultrasonic Off Ultrasonic On Ultrasonic Wait Alignment Measure Drain Feed Circulate Drain
3. Feed to fill the plumbing Click Align Blank to align the optics and blank the sensors for any remaining material in the machine Add your sample to the sampler introducer chamber by pouring using an eyedropper or scoop Sample can be added in either dry or wet form Wet sample Add no more than about 25 ml or the cell will overflow and skew your measurements Add the sample quickly and rinse down any excess caught in the sample area Dry sample Add only a few milligrams Add the sample quickly and rinse down any excess caught in the sample area Turn on the ultrasonicator to disperse the sample in solution Watch the two transmission percentages The red is for the laser and the blue is for the incandescent bulb They will decrease rapidly as the sample circulates past the sensor If the transmission percentages are too high or too low you can add more sample add more DI or click Rinse and start over at the beginning of this procedure If you wish to ultrasonicate turn on the ultrasonicator The controller on off switch is the yellow Y button near the top of the screen One minute of ulltrasonication at power level 6 should be enough for most samples Setting can be changed using the up and down arrows near the ultrasonicator icon Click the Align and Measure button and the optics will align and measure the sample This takes about a minute 10 Click the Rinse button 11 Repeat this
4. computer w o the Horiba the newest software is version 3 58 available for download at Horibalabs com Install Gilson Special 586 a Select COM 2 when prompted b Click yes to restart your computer 14 7 8 9 Copy the file FEEDINFO sys in to the following directory C Program Files HORIBA LA 920Wet Open the Horiba LA 920 software When prompted for the optical data click browse and locate the folder optical data select OK When prompted select yes to build a database 10 For use with the autosampler you need to re map the robot arm 11 Go to Sequence Scheduler Set Up Robot Feeder and change the XPOS and YPOS to match the following table POSITION XPOS YPOS Sample 1 80 1600 Sample 2 80 1000 Sample 3 80 300 Sample 4 750 1600 Sample 5 750 1000 Sample 6 750 300 Sample 7 1400 1600 Sample 8 1400 1000 Sample 9 1400 300 Sample 10 2050 1600 Sample 11 2050 1000 Sample 12 2050 300 Sample 13 2650 1600 Sample 14 2650 1000 Sample 15 2650 300 Sample Chamber 1920 2400 Rinse Pos 2650 2400 Home Pos 0 0 Click on OK to save changes If these values do not work you will need to use the manual control to locate the correct positions then enter them in the spread sheet 12 To use the autosampler you also need to set up a methods file following the instructions in the methods section above Sequences usedto operate the Horiba
5. the Horiba should be rinsed and cleaned with Q tips and DI water 7 For use with the autosampler only clean the rinse chamber Remove the autosampler rinse chamber next to the sample collection funnel Push the silver button to disconnect the white tubing from the chamber then lift the chamber upwards out of the tray Rinse it with DI and scrub it with Q tips or paper towels if necessary Replace the rinse chamber on the sample tray reconnect the tubing and fill the rinse chamber with DI water 8 For use with the autosampler only fill the pump line Fill the beaker on top of the autosampler with DI water Visually inspect the pump line if sediment is present it will need to be flushed out or the tubing will need to be changed Password is Horiba without the quotes a To fill the pump line go to Sequence Schedule Robot Control Manual Control b Set arm position x axis 2650 y axis 2400 z axis 2150 c Click Move Arm Arm will move so that it is positioned over the rinse chamber d In the pump control section click Delivery then On e Let it draw water for 5 6 seconds until the pump line is full and water sprays out of the autosampler into the rinse chamber f Click off Change the pump control back to Suction Exit back to the main software window TA 9 On the Measurements screen click the Start Up button This process take about 45 minutes Do not skip this step it includ
6. Execute to begin the analysis Once you click Execute the Horiba will run though all of the necessary steps outlined in the Feeding Samples Manually without the autosampler section above and then prompt the user to add the sample Add your sample to the sampler introducer chamber by pouring using an eyedropper or scoop Sample can be added in either dry or wet form Wet sample Add no more than about 25 ml or the cell will overflow and skew your measurements Add the sample quickly and rinse down any excess caught in the sample area Dry sample Add only a few milligrams Add the sample quickly and rinse down any excess caught in the sample area Click OK The Horiba will finish running the sequence to take a measurement rinse and proceed on with the next sample You will be prompted each time to add sample Measuring samples added by hand with a preset sequence for just a few samples standards 6 minutes per sample machine time 1 Click Select Program on the measurement screen From the drop down menu select the sequence you would like to run See the section on sequences below to choose the correct sequence For most samples select Run Ultrasonic Other examples Standards City of Mpls Water Works Wait while the Horiba runs the sequence When prompted add the sample to the sample introducing funnel Wet sample Add no more than about 25 ml or the cell will overflow and skew your measurement
7. data it is a good idea to move your Horiba d2w and txt files to Transfers Analytical Data New Data Grain Size Your Project Name Not only does clean up the active files in the Horiba database but it assures that your data is backed up and will get archived Shut down 1 Turn off the computer First select Shut Down in the Windows menu then when the computer says You may now turn off the computer push the button to turn it off 2 Turn off the Horiba using the switch located in the back on the lower left Horiba Settings The Horiba has several different settings the following are four that can have the largest affect on analysis results and the time it takes to analyze samples Measurement Conditions Used to analyze samples cannot be changed after measurement Display Conditions Used to display results can be changed after measurement Method Files Used by the scheduler to analyze samples cannot be changed after measurement Automatic Concentration Settings Used by the scheduler to analyze samples with the autosampler cannot be changed after measurement Measurement Conditions Used to analyze samples cannot be changed after measurement Usually only the parameters in the Settings tab need to be changed Go to Conditions Measurement Conditions to set these parameters Meas Conditions Setting Auto Functions Feeding and Rinse Control Circulation Speed W le Ultrasonic Time t vil
8. in case you change your mind about the settings you want to use Save the conditions to a file so you can use them for all of your samples and so you can designate that file when you are setting up the autosampler When you start if you do not set new conditions or open 10 a conditions file the machine will just use whichever conditions were used by the last person Usually only the parameters in the Distributions tab need to be changed but the fixed values and span tab can be edited if you will be using calculated data directly from the Horiba without reprocessing Go to Conditions Display Conditions Display Conditions Datso Fed Vises and Scan Setting on of Darton gt Sardi Shep Manal Voume Ques Length Number Sge Class Retained O verses Pasg Undersice m Type Oa le to Display conditions screen Distribution tab Form of distribution Standard R R Index 116a001i Distribution Base Volume Density Distribution Standard Axis Selection LogX LinY ISO 9276 1 Format Off Cumulative Distribution Graph On Size Class Passing Undersize Axis Type Bar or Line Notes Form of distribution Standard should always be used with sediment samples because of the wide distribution range Relative Refractive Index 116a001i may be used as a default for quartz samples Please consult LRC staff for assistance in choosing an appropriate relative refractive index This setting has a huge impact on
9. procedure for each sample Clean Up 1 2 Command the Horiba to rinse feed circulate for 3 minutes and rinse Optional feed add 1 2 drops of natural or weak alkaline detergent circulate 30 sec rinse feed circulate 30 seconds and rinse Wipe down the sample introducing funnel Wipe down the outside surface of the Horiba if necessary with damp paper towel If the autosampler was used rinse the autosampler by going to Sequence Schedule Robot Control Manual Control 1 Set arm position x axis 2650 y axis 2400 z axis 2150 2 Click Move Arm Arm will move so that it is positioned over the rinse chamber 3 In the pump control section click Delivery then on 4 Let it draw water for 10 seconds to flush any sediment residue out of the autosampler and into the rinse chamber 5 Click off 6 Change the pump control back to Suction 7 Click Exit then OK 6 If the autosampler was used remove the autosampler rinse chamber from the autosampler tray empty it into the sink rinse with DI empty and then replace on the autosampler tray 7 lf the autosampler was used remove the sample funnel on the autosampler rinse with DI and then replace on the autosampler tray 8 If the autosampler was used wash all beakers carefully using water and paper towels Make sure there is no sediment residue inside each beaker You may set them upside down on paper towels to d
10. LacCore SOP Series Horiba Grain Size Analyzer LA 920 Wet SOP Purpose To measure the particle size distribution of sediment samples using laser diffraction analysis Cautionary Notes The Horiba is not designed to contain vapors that may be hazardous All potentially dangerous chemical reactions should be complete before analysis in the Horiba and the dispersant should be safe to use out of a fume hood The use of vigorous pumping and the ultrasonic probe may increase the amount of harmful gas emitted The ultrasonic probe may be damaged by the repeated use of strong acids at high temperatures Horiba should be consulted before any strong acids are used The following are all compatible with the Horiba aqueous solutions keytones organic solvents alcohols ethers esters chlorinated solvents oxidizers oils fuels acids alkalis Pretreatment If you are considering removal of organic matter biogenic silica or carbonates details concerning pretreatment procedures can be found a separate Grain Size Pretreatment SOP Each researcher should consider pretreatment but each pretreatment has benefits and hazards depending on the research question Start Up Preparing and calibrating the Horiba may take 1 2 hours at the beginning of each day After that you may plan to measure about 10 samples per hour at maximum efficiency Clean up takes about 30 minutes at the end of the day If the Horiba asks for a password it is Horiba wit
11. T is still falling it may prematurely decide that a concentration correction is needed If the T for your samples repeatedly falls below your minimum setting by the time a real measurement is taken change this setting On the other hand it is desirable that the automatic readjustment process be done as quickly as possible to avoid any settling or sample bias in the plumbing Maintenance Cleaning the cell based on Horiba User s Manual p 9 1 2 gi oe D ee Start up the Horiba Rinse twice then drain Add isopropanol to fill the sample inducer Set the pump speed to 15 start the pump Look to see if the T increases or channel baseline decreases for the next 5 minutes Rinse twice with deionized water Repeat steps 3 6 two more times Rinse 10 times with deionized water If necessary see the LA 920 Instruction Manual p 7 and the User s Manual p 5 for instructions on how to remove the optical cell and clean it The instructions are located in the drawer below the computer monitor Instructions for installing the Horiba on a new computer You will need windows XP and two serial ports The autosampler should be connected to the PCI card and the Horiba should be connected to the onboard serial port In device manager check to make sure that the PCI card is COM 2 and the on board serial port is COM 1 Install LA 920 Wet Robot 3 46 a This is the newest wet robot version of the software For use ona
12. e when you close the software or view another sample this sample will revert to its original Display conditions and R R Index setting To change the Display Conditions for multiple samples after analysis use the Batch Edit feature 1 In File Open Sample make sure the database has all the samples that you need to change 2 If the database has some samples that you would like to exclude use the Create Index button in the data base screen and enter a parameter that is common to the files you wish to change and create a group Click OK to exit the database screen In Group Processing menu select Batch Operation In the R R Index drop down list box select the new RRI value Click Selected Database check box Make sure this has the selected number and not the total database number Mark the checkbox Update Data on Disk Press OK to convert the data Oo OL Oe N Method File Used by the scheduler to analyze samples cannot be changed after measurement Itis good to know about these setting but they probably won t need to be changed unless you need to set up a method for running samples by hand with the scheduler 12 Go to Sequence Schedule Setup Methods Click on existing file or create a new file Dispersant From Reservoir Blanking Delay 20 seconds Yes Blank Measurement Add sample From Autosampler Measurement Delay 15 sec Num times 1 Repeat delay 0 seconds D
13. es the necessary warm up time 10 On the Measurements screen click the Standard button 11 The Horiba will prompt you to add the standard when it is ready Add 1 level scoop of the Alox 240 standard click OK and the Horiba will measure the standard It is recommended that you run a standard at the start of the day and at the end of the day 12 Then the machine will prompt you to select display and user conditions for your samples Select an existing condition file from the list and click Open You will be asked if you want to apply the Display Conditions to the current open sample Choose no unless it is your sample and you want to change its Display Conditions If you do want to permanently change the current sample s Display Conditions you must then save sample in the File Menu For most samples select LRC cn2 for both condition files Different ways to measure This SOP outlines 4 different ways to run measurements on the Horiba 1 2 3 4 Measuring samples using the autosampler recommended Measuring samples added by hand with a preset sequence and scheduler for lots of samples Measuring samples added by hand with a preset sequence for just a few samples standards Manual measurements not recommended Measuring samples using the autosampler This process uses the autosampler and the scheduler Using the autosampler is recommended when possible to standardize the s
14. hout the quotes 1 Turn on the Horiba using the switch on the back of the machine on the left The autosampler is usually left on because it is a little harder to get to but it has a separate switch on the back of the machine on the left 2 The computer and monitor on buttons are on the front of those machines 3 Open LA 920 from the desktop The software has three views once opened a 3D view a Measurement view and a Display view The measurement view contains the controls for the machine the display view contains the data from the measurements you have taken and the 3D view displays multiple data plots at once 4 Make sure that there is at least half a tank of 0 5 g L Na hexametaphosphate SHMP in the large bottle to the right of the Horiba Dry SHMP is on the shelf above the computer monitor and there should be a scale in the Horiba room Mix using dry chemical and low purity deionized water 5 The Horiba has a base unit with an autosampler tray mounted on top that slides on rails Remove the small silver pin on the autosampler tray and lift the black sample collection funnel upwards out of the tray Then slide the autosampler tray to the left to reveal the inner part of the Horiba Use several Q tips and DI water to rinse out the black funnel Sample introducer Make sure it is clean of any sediment and film 6 For use with the autosampler only the black sample collection funnel on top of
15. mber This is a fast process and appears to be quite efficient This automatic feature may be disabled in the software so that you get an error message when the sample is too dilute in that case you need to add more sample manually If many of your samples are too dilute you may increase the delivery time so that the autosampler initially delivers more sample to the chamber For instructions on how to do that see 4 above Measuring samples added by hand with a preset sequence and scheduler for lots of samples This process is a combination of using the scheduler and adding sample by hand The machine is programmed to complete all the steps and prompt the user to add the sample when needed This process frees up the user to deal with the samples more and the less with the Horiba and keeps the timing between the steps more uniform than when everything is done by manually This is the preferred method when the autosampler cannot be used 1 8 Go to Sequence Schedule to open the Scheduler Enter the sample name Leave position blank for each sample In the drop down menus choose the methods and conditions files you wish to apply to your samples There is a condition file named by hand that is preset to run samples using this method It can be changed following the steps outlined in the Conditions section below The Scheduler will auto fill those selections down through your list Still in the Scheduler click
16. min Setup ina sec Ultrasonic Power U Sonic works during Meas C7 Sampling Times Laser Preferable T zo Max 25 10 No of Blanking Scans i Laser Measurement conditions screen settings tab Circulation Speed 6 Ultrasonic Time 1 minute Ultrasonic Power U Sonic works during Meas No Sampling Times Laser 25 Lamp 2 Preferable T Max 90 Min 70 No of Blanking Scans 10 Notes Circulation Speed You may use 4 8 depending on grain size max circulation speeds may lead to bubbles being measured Ultrasonic Time 30 sec to 2 min is usually sufficient for most samples Long periods of ultrasonication can lead to heat build up and increase the solubility of some particles U Sonic ON during Measurement If yes be sure to check for 10 20um air bubbles that may form and be measured Preferable T Varies Horiba technicians state that 85 transmittance is ideal for most analyses However samples with distributions that span orders of magnitude may need lower min and max settings possibly as low as 30 transmittance The Horiba does not take any actions based on this setting The only thing this setting does is change the size position of the gray bars on the transmission percent graph on the measurement screen Display Conditions These are parameters that the Horiba will apply to your sample in order to display the results These conditions can be changed after you have run your samples
17. raining and rinsing Yes Drain and check Rinse Rinse flue From Reservoir Notes Add sample Change to By Hand if you would like to set up an automated process to measure samples by feeding manually with the scheduler Automatic Concentration Settings These settings determine how the autosampler automatically adjusts the sample concentration to fit within your transmittance boundaries You may increase the range if the machine has trouble fitting both the laser and blue lamp Go to Conditions Measurement Conditions to set these parameters Go to Sequence Schedule Setup Concentration Yes Auto Adjust Sample Concentration Use T from Laser Minimum Feed 4 Maximum Feed 30 Required T 84 T Plus Minus 10 Conc Delay 60 Notes Use T from The laser usually diminishes more slowly than the blue light so using the laser to check transmittance will most quickly get you to your desired T Required T and T Plus Minus Varies Horiba technicians state that 85 transmittance is ideal for most analyses However samples with distributions that span orders of magnitude may need lower min and max settings possibly as low as 30 transmittance 13 Conc Delay This setting may be lengthened if the Horiba is too aggressive in automatically adjusting your sample concentration For instance when the Horiba delivers a sample to the sample changer The T decreases If the Horiba takes a measurement while the
18. ry Saving Data The software will automatically save a copy of your data but it is in the format of the software so you must export it as a text file To save individual files with different file names ils With the desired data set Sample open in the display window click on Export Export and export as a text file txt to a folder that you choose Save all files by sample name There are arrow tabs at the top of the Display screen to scroll through data so you can look at your past sample measurements and export them Do this for every sample Once all of your files have been exported you can save your text files on a flash drive or compress them and e mail them to yourself Large numbers of data files can be saved as text files in bulk The file name will be the Horiba ID but can be renamed later OS Click on Export Export Click on Data Convert and select the desired data files Click on Title re title file names optional make sure the files are selected Click on Open It may look like nothing happened but if you use Windows Explorer to go to the directory where the data is located C Program Files HORIBA LA 920WET data you will find your text files There is a link to Data on the desktop that will take you to the files Data management The Horiba auto saves all files to the last used save location As a default use C Program Files HORIBA LA 920WET data to save data Once you are done exporting your
19. s Add the sample quickly and rinse down any excess caught in the sample area Dry sample Add only a few milligrams Add the sample quickly and rinse down any excess caught in the sample area Turn on the ultrasonicator to disperse the sample in solution You will know when you have added enough sample by watching the T on the left side of the screen For most samples T should be between 95 70 Click OK and the Horiba will continue the sequence to measure the sample Repeat steps 2 5 until all samples have been run If you are running the Standards sequence select the conditions files you wish to use before continuing Manual measurements Care should be taken when using this method It is important to try and standardize the measurements as much as possible and this method leaves the most room for changes in timing f Remove the silver pin and the black sample funnel from the machine and slide the autosampler tray to the left You can now see the back sample introducer funnel in the machine Go to File Open Conditions and apply the conditions file you wish to use with your samples There is a pump that circulates sample within the Horiba during analysis The pump controller on off switch is the blue square button near the top of the screen A pump speed of 6 is usually adequate unless you have grains larger than 500 um as a significant gt 5 part of the sample Click the blue box to turn on the pump 4 Click
20. t coordinates Making a mistake in these categories could damage the autosampler Sample Pump Duration section all in seconds Withdrawal 15 Delivery 20 Rinse 5 Empty liquid 0 Stirrer ON 100 Drip Spin 3 Needle Rinse 40 Delivery time should always be slightly longer than withdrawal to make sure all the sediment is washed out of the autosampler You may change the Withdrawal time if you need more or less sample to be analyzed For instance if the machine always had to add more sample to adjust the concentration to be within your transmittance boundaries you could lengthen the withdraw and delivery times This is not necessary but could save you some time Ultrasonic power of 4 6 and run time of 1 minute is also standard Still in the Scheduler click Execute to begin the analysis If the sample is too opaque i e transmittance is too low The Horiba will automatically add a small amount of SHMP solution then re evaluate the transmittance This is a slow process and if it continues too long some coarse particles may settle down into the plumbing If you see that the initial transmittance measurement is far too low you can use an eyedropper to quickly add more DI into the sample chamber Be sure not to overflow the sample introducer funnel or your results will be skewed If the sample is too dilute i e transmittance is too high The autosampler will quickly get more sample and add it to the sample cha
21. the dataset especially in the lt 20 um range This setting has two components a real and an imaginary component The real component is the refractive index of the sediment relative to the dispersal solution water In the case of quartz 1 16 is the relative refractive index The imaginary component 001i where i V 1 varies with particle characteristics such as particle absorbance porosity density variation and surface roughness As long as you re 11 internally consistent with your optical model and R R I settings you can make valid comparisons between your samples However it is not clear how well you can compare your grain size results to other people s grain size results It is strongly recommended that you note the R R I setting and associated assumptions when publishing your results Distribution Base Volume should always be used unless comparing results to other methods of grain size analysis Conversion to number base should only be used if the volume base distribution is symmetrical and only for comparison to other methods such as microscopy To change the Display Conditions after analysis 1 Open the sample File Open Sample on the Display screen 2 Goto Conditions Display Conditions 3 Click on the R R Index button scroll down through the options to the setting you want and click Open then click OK 4 If you want to save the sample with this new R R Index applied go to File Save Sample Otherwis
22. ubsampling and analysis This process may not be recommended for samples where significant portions of the distribution are greater than 250 um 1 If the lid of the Horiba is open be sure to close it and replace the black sample collection funnel The samples should be dispersed in 5 g L SHMP and agitated on a wrist action shaker at least 8 hours overnight preceding analysis Pour each sample into a numbered Horiba beaker and add DI water or 5 g L SHMP so there is at least 50 75 ml in each beakers The analysis will be most efficient if all of your samples have approximately the same transmittance e g all dark brown or all light brown Place the samples on the autosampler Go to Sequence Schedule to open the Scheduler Enter the sample name and position 1 through 15 for each sample Beaker positions are identified on a sticker on the autosampler tray In the drop down menus choose the method and condition files you wish to apply to your samples The Scheduler will auto fill those selections down through your list LRC and autosampler are the common condition and methods files chosen You may also save this sample list in the Scheduler under File Save As but it is not necessary to do so While the Scheduler is open check the delivery system settings under Setup Robot Feeder To begin try these settings Under most circumstance you should not change the sample location stirrer position or needle heigh