1.       Symptom    Possible Cause    Action       No response to gas    Gas cylinder empty  Cylinder as needed    Hose blocked or kinked    Check gauge  replace    Ensure no restriction  to flow       Gasman fails gas test    Gas cylinder empty  Cylinder as needed    Gas cylinder out of date  Hose blocked or kinked  Calibration drifted    Gas flow not started  immediately    Check gauge  replace    Check date and  replace as needed    Ensure no restriction  to flow    Calibrate Gasman    Repeat test  starting  gas immediately       Gasman fails  calibration    Gas cylinder empty  Gas cylinder out of date  Hose blocked or kinked    Calibration drifted  Stabilization time too short    Check gauge replace  bottle as needed  Check date and  replace as needed    Ensure no restriction  to flow       Calibrate Gasman    Reset using PC soft   ware       Gasman passes gas  test but will not enter  calibration mode    Menu Zero not performed    Gasman not field  calibration version    Select Zero from  menu    Send for re configu   ration       Note  Remove regulator from gas cylinder when not in use over a pro     longed period     For parts list  see section XI     One button Calibration will check for small drifts of stored calibration  value  but Crowcon recommends Gasman is sent for full certified calibra   tion at six monthly intervals      gt    CROWCON    Gas Detection You Can Trust    21    Maintenance and Calibration Gasman       VII  Maintenance and Calibration       Gasman is
2.   Please see the  section on replacement parts and acces   sories        Periodically   bump test  the sensor with  gas to confirm its ability to respond to gas  by exposing the detector to a targetted gas    concentration that exceeds the high alarm  setpoint  Manually verify that the audible  and visual alarms are activated       It is recommended that the detector be cal   ibrated before first time use  and then at  least once every 90 days and only by quali   fied personnel       For use in potentially explosive atmospheres  where oxygen concentrations do not  exceed 21  v v        Do not turn off the detector during a work  shift  Turning off the detector resets the  time weighted average gas exposure values  to zero       Ensure that the sensor is not blocked       High off scale  LEL readings may indicate  an explosive concentration       Only charge the battery using a Crowcon  supplied charger       Do not use any other charger  Other charg   ers may damage the detector making it  unsafe for use in hazardous areas       Do not change or charge batteries in a haz   ardous location  Doing so may lead to fire  or explosion     Flammable Gas Units Only      Extended exposure of the flammable sensor  to certain concentrations of combustible  gases and air may stress the sensor ele   ment  which can seriously affect its per   formance  If an alarm occurs due to a high  concentration of combustible gases  recali   bration should be performed  or if needed   the sensor repla
3.  1  To view the additional display Option menu  double click the operator  button  The menu icons will appear on the screen  as shown     Note  Only  instruments for toxic  gas monitoring will  display the TWA  menu option        2  Press the operator button with a single click to scroll across the  list  When the underscore appears under your choice  double click  the operator button    If Peak or TWA is selected  Gasman will display the icon on the  operator screen      gt    CROWCON 11    Gas Detection You Can Trust    Operation Gasman    Peak test   When performing a Peak test  such as a vertical entry check   previous readings can be cleared by selecting the Peak Reset menu  option    1  Scroll 2  Select       m Peak display       Zero   To perform Autozero  select the Zero function from the menu   When Zero is completed  the instrument will return to normal Run  mode     2 4 Logging    Gasman incorporates both data and event logging which can be  accessed using the RS232 communications link available with the Single  Way Charger Plus PC Interface  using Crowcon Portables PC software   See section VIII for part numbers     Data is recorded every minute  this rate can be adjusted using the PC  software   The log can record 900 hours of data at 1 minute intervals     Gasman also records the time and date for a number of operating and  diagnostic events including     12      Switch on and switch off       Level 1  Level 2 and Time Weighted Average Alarms  TWA    alarm on  a
4.  Portables PC software  for the instrument to lock in the  event of the calibration date expiring  to prevent further operation of the instrument     3  Autozero  Press the operator button with a single click to  confirm Autozero  Flammable and toxic units will be NB  If Autozero  set to read zero  and oxygen units set to read     s  20 9   If the operator button is not pressed within displayed   10 seconds Gasman will proceed directly to normal  operation without performing a zero      gt    CROWCON 9    Gas Detection You Can Trust    Operation Gasman    Switch off  To switch off the unit  press and hold the operator button for 5  seconds  The display will count down from 5  continue to hold the  button until the unit counts down to shut off     2 2 Run mode  The Gasman unit displays the gas reading on the operator display screen   A typical display is shown below       Screen Icons b                         x Warm up  OK Flashing  icon   Gasman    running  The sensor channel will display the current value of the normally  gas being monitored and units of measurement  the  ay _4   OK  symbol will flash indicating unit is working cor  Alarms  Me      AL  2  rectly  Familiarize yourself with the gas currently being  monitored in your unit  Ensure you understand site H Battery  health and safety procedures  For information on Peak   Ii  and TWA readings  go to section 2 4     Confidence signals NT aZe  To reassure users the unit is working correctly  the Gasman unit will   emit
5.  USA  UL913   Canada  CSA22 2  152   Operation EN50270  EN50271  IEC61508  EN61779        gt    CROWCON    Gas Detection You Can Trust    27    Accessories and spare parts Gasman       XI  Accessories and spare parts       Accessory list    Crowcon    part number    Description    Single way chargers                         C01941 12 V DC input single way charger  C01942 Single way charger with 230 V UK style power supply  C01943 Single way charger with 230 V EUR style power supply  C01944 Single way charger with 110 V US style power supply  C01945 Single way charger with 90 260 V in line power supply  C01297 Vehicle lighter socket lead  C01940 Single Way Combined charger and PC interface  C01947 Single way charger interface with 230 V UK style power supply  C01948 Single way charger interface with 230 V EUR style power supply  C01949 Single way charger interface with 110 V US style power supply  C01950 Single way charger interface with 90 260 V in line power supply  Multiway charger  C01951 5 way Multicharger with 90 260 V in line power supply  i modules   5011424 0 10096 LEL methane  5011436 0 100  LEL propane  5011437 0 100  LEL pentane  5011439 0 100  LEL butane  S011440 0 100  LEL ethylene  S011460 0 100  LEL hydrogen  S011423 0 25  oxygen  S011422 0 500 ppm carbon monoxide  S011421 0 50 ppm hydrogen sulphide  S011425 0 10 ppm sulphur dioxide  S011429 0 1000 ppm hydrogen  S011426 0 10 ppm nitrogen dioxide  S011428 0 5 ppm chlorine  S011432 0 1 ppm ozone  S011430 0 25 ppm
6.  a short beep accompanied by a blue LED flash every 10 seconds   and the OK icon will flash     2 3 Display symbol guide  Battery    A full battery is represented by a battery icon showing a  full three bars  A low battery charge will show 1 bar  When  d J l zero bars are shown the battery icon flashes and the   a sounder will emit warning bleeps    If the battery becomes too low  Gasman will switch off     Full    TWA alarm    Gasman will display the TWA alarm when the 15 minute or 8 hour time  weighted average alarm threshold is passed for toxic gases   Gasman will display    LTWA    and  STWA   TWA alarms cannot be cleared     10  gt    CROWCON    Gas Detection You Can Trust    Gasman Operation    2 4 Display options    Gasman provides four additional selectable displays     Peak display   hd When Peak mode is selected the instrument shows the  highest value for flammable and toxic gases  or the lowest  value for oxygen since switch on or last Peak Reset  This is  useful for vertical entry checks where the whole instrument    can be lowered down the shaft rather than using a sampling  tube and for viewing peak exposure at the end of a shift     TWA display    amp  P   Shows the 15 minute or 8 hour time weighted average   TWA   for toxic gases  monitored since last turn on     Peak Reset  Before a Peak entry test is performed select this menu  option to clear any previously stored Peak value     Zero   lt  0 0 Performs Zero on your Gasman unit     How to display the menu   
7.  designed to operate almost maintenance free under most con   ditions  However  some small items of routine maintenance are recom   mended     General    To keep the display panel and operator button free from dirt build up  reg   ularly wipe over your Gasman unit with a damp cloth     Filter    Inspect the front filter at regular intervals for dirt or damage  Clean as nec   essary     Zero and calibration    Gasman is supplied with an Autozero function on start up  This function  can be configured to operate automatically  on user confirmation  see  quick start guide   or can be disabled  This configuration can be set with  the Crowcon Portables PC software  see section VIII  Gasman also has a  Zero function in the menu  See section 2 4     Crowcon recommends  as a minimum  a monthly gas test to confirm sen   sor operation  A test gas of known composition  needs to be applied  to  verify sensor response and alarm function  see section 6 1     Instrument calibration of sensor should be performed at 6 month regular  intervals     Calibration method    Gasman calibration can be performed using the Gas Test Kit as per 6 3  or  with Crowcon Portables PC software and calibration gas mixtures   Appropriate certified calibration gases should be used  Calibration gas is  applied using the appropriate flow cap  Refer to Crowcon Portables PC  software help file for further information     22  gt    CROWCON    Gas Detection You Can Trust    Gasman PC interface and software       VIII  P
8.  hydrogen cyanide  28       CROWCON    Gas Detection You Can Trust    Gasman Accessories and spare parts        011435 0 50 ppm ammonia   011438 0 1000 ppm ammonia  S011431 0 2 ppm phosphine  S011434 0 1 ppm fluorine   S011433 0 10 ppm hydrogen fluoride  Sampling accessories    M02340 Alligator Clip   M04851 Flow cap   C01937 Aspirator Assembly    Calibration gas contact Crowcon    required gases depend on sensor combination    Carrying and wearing              C01952 Harness Plate   C01843 Shoulder Strap   C01844 Chest Harness Straps   C01953 Hard Hat Clip   Communications    E07532 PC Interface Lead   C01832 PC Software   C02097 USB to RS232 adaptor   Spares   consumables    M04856 Rear Molding   C03329 Metal Clip   E01918 Disposable Battery   E07621 i module Elastomer   E07620 LCD Elastomer   M04682 Sensor Sealing Ring   MO3705 Clip retaining nut M3   MO3793 Case screw   MO5910 Crowcon domed label   EO1535 Power supply for charger 230 V for UK only  EO1536 Power supply for charger 230 V for EUR only  EO1537 Power supply for charger 110 V for USA only  E01552 Power Supply for charger   230 V No plug  E01553 Power Supply for charger   115 V No plug  CO1979 Power supply for charger Universal 90 260 V     gt    CROWCON 29    Gas Detection You Can Trust    Troubleshooting guide    Gasman       XII  Troubleshooting guide       Symptom   error message    Cause    Instrument won t switch on Flat battery    No confidence beep    Gas reading when no gas  present    Unstable inaccura
9.  innovative and rigorous design  technology  we have introduced several new features      gt    CROWCON 7    Gas Detection You Can Trust    Introduction Gasman    i module gas sensor    Gasman uses unique plug and play i module sensor technology  Each sen   sor unit incorporates its own intelligent processor holding sensor configu   ration and calibration data  Different sensors can be purchased  and once  inserted  are immediately ready to run  Flammable units are only available  with rechargeable batteries  Plug and play will ease maintenance time and  cost  and the intelligent modular system will remove the need to calibrate  the sensor  Gasman can be reconfigured by purchasing an additional pre   calibrated i modules from your local supplier     Reliable  anti shock mechanics and robust housing    The Gasman housing is built from resilient material  giving it strength and  flexibility to withstand the hardest of working conditions  water and dust  tight to IP65 and IP67  with a non slip grip  If the unit is dropped  there  will be no disruption of power or function  ensuring reliability and service  for years to come     Software    The internal software in Gasman has been designed and written in accor   dance with the requirement of IEC 61508 to ensure quality and integrity  of operation  Gasman has been designed to give a truly reliable personal  gas monitoring system  The internal circuitry includes an independent  watchdog  the software monitors for any malfunction wi
10.  over   range to prevent burn out  This locks out for 200 seconds after which a  button press will reconnect power to the catalytic bead  If the sensor  power is reconnected when the unit is exposed to an over range gas con   centration there is a risk of damage to the catalytic bead sensor  Restart  should be carried out in a known fresh air environment  Depleted oxygen  levels can reduce the combustible gas reading  and if oxygen levels are  below safe breathing limits it should be assumed that the combustible  reading is low           Electrochemical gas sensors  toxic gases or oxygen  contain chemicals   Extreme levels of humidity can also cause problems  The sensors are rated  for an  average  ambient of 15 9096 R H  However they are used from the  tropics to deserts to tundra without this normally being a problem     Water should not be allowed to collect on the sensor as this may impede  gas diffusion     Persistent exposure to high levels of toxic gas will shorten the life of toxic  sensor  If the high level gas is corrosive  e g  hydrogen sulphide  damage  may occur over time to metal components     Sensors may be cross sensitive to other gases  If unsure  contact Crowcon  or your local agent      gt    CROWCON 31    Gas Detection You Can Trust        A        HALMA    GROUP  COMPANY       UK Office   Crowcon Detection Instruments Ltd  2 Blacklands Way    Abingdon Business Park  Abingdon   Oxfordshire OX14 1DY   United Kingdom   Tel   44  0 1235 557700  Fax   44  0 1
11.  please see the troubleshooting  guide in the first instance or contact Crowcon     5  To abort the gas test press the button at any time whilst the test is  in progress   6 3 How to perform a one button  calibration test    To perform a one button calibration test  you must first perform a Zero  on your Gasman unit     1  Ensure you are in clean air   Double click the button and select Zero from the Options menu   Gasman will perform a Zero     To perform a one button Calibration  complete the next steps within 15  minutes of completing the Zero     2  Follow steps 1 to 3 given in 6 2  Gasman will display  CAL  and         alternately  Press the button to confirm one button  Calibration    If the button confirmation for calibration is not made within 10  seconds then the process will revert to test as in 6 2    3  Apply calibration gas following step 4 in 6 2    4  To abort the Calibration test press the button at any time whilst  the test is in progress    Your Gasman unit will adjust the value for the gas channel to match the  stored calibration gas value within the sensor i module     If your Gasman unit does not calibrate successfully  Gasman will display     FAIL     Your Gasman unit must be sent for re calibration     Gas Test    pass    and    fail     and calibration    pass        fail    and values are stored  in the Event log     20  gt    CROWCON    Gas Detection You Can Trust    Gasman    Gasman gas test accessory kit    6 4 Gas test calibration troubleshooting 
12.  the case to change the battery     To replace the battery pack  ensure you are in a safe  non hazardous area   Remove the back cover and remove the battery pack  Replace the Lithium  Cell  then reinsert the battery into the instrument and replace and secure   ly fasten the back cover     14    CROWCON    Gas Detection You Can Trust    Gasman Alarm indications       IV  Alarm indications       Gasman provides two instantaneous alarm levels  designated level 1 and  level 2  For toxic gas sensors  there are also two time weighted average  alarms  TWA   one for short term exposure  STEL   based on a 15 minute  time weighted average  and the second TWA alarm is for long term expo   sure  based on a 8 hour time weighted average     Alarm configurations are set via the Crowcon Portables PC software   The following settings can be made     Alarm thresholds for each sensor  Level 1 and HL   l  level 2 alarms can be set for the gas sensor  HL 7 B    Alarm trigger  This can be set to rising levels of gas  concentration  or as falling  Oxygen units are set to falling for defi   ciency monitoring     Alarm latching  Alarms can be set to be latched or unlatched   Latched alarms will require the operator button to be pressed in  order to clear the alarm  This is the default setting  Unlatched  alarms will clear automatically when the gas hazard has passed     Alarm mute  The sounder can be set to mute for level 1 alarm  only  pressing the operator button during an alarm condition i e   pre
13. 235 557749  Email  crowcon crowcon com  Web site  www crowcon com    Rotterdam Office   Crowcon Detection Instruments Ltd  Vlambloem 129   3068JG  Rotterdam   Netherlands   Tel   31 10 421 1232   Fax   31 10 421 0542   Email  crowcon crowcon net   Web site  www crowcon com    USA Office  Crowcon Detection Instruments Ltd  21 Kenton Lands Road   Erlanger   Kentucky 41018 1845  USA  Tel   1 800 527 6926 or  1 800 5 CROWCON   1 859 957 1039  Fax   1 859 957 1044  Email  sales crowconusa com  Web site  www crowcon com    Singapore Office   Crowcon Detection Instruments Ltd  Block 192 Pandan Loop    5 01 Pantech Industrial Complex  Singapore 128381   Tel   65 6745 2936   Fax   65 6745 0467   Email  sales crowcon com sg   Web site  www crowcon com    
14. C interface and software    Gasman can be connected to a PC using the single way charger unit with  optional PC interface  The charger unit is fitted with a D type 9 pin R 232  Socket which is located at the rear of the charger  see diagram below  The  PC requires Crowcon Portables PC software  A USB RS232 adaptor is also  available from Crowcon                       The software provides the user with access to reconfigure alarm levels   operation  run calibrations  print reports and to access data and event log  files        Set up  1  Install Portables PC software on PC and attach the RS 232 cable to  the charger and PC   2  Switch on the Gasman unit and drop it into the charger unit  ensuring it is upright and the display is facing forward   3  Open the Portables PC software and either use the Wizard or the  Engineer s Form  select Gasman and upload the configuration   For more information on using the Crowcon Portables PC software  see  installed help file      gt    CROWCON 23    Gas Detection You Can Trust    i module replacement Gasman       IX  i module replacement       1  Ensure you are in a non hazardous  safe  area   Switch off the unit   2  Remove the back cover by unscrewing the four M2 5  12 mm  Allen screws as shown the in the drawing  point            3  Lift off near corner and unclip the i module    Installing or replacing an i module  1  Unwrap the i module from any packaging  ensure the sensor is  fully seated on the module board     If replacing an i mo
15. Gasman  Personal Gas Monitor                ME jCROWCON    October 2005     Issue 3 Gas Detection You Can Trust    Safety Certification     Gasman is classified to US standards by  Underwriter Laboratories Inc as suitable for use  in Class 1 Groups A  B  C  D  Division I locations   Before use read the manual to ensure a full  understanding of how to operate the unit and as  to any limitations of use        Si  Fo DG     uNES KINGDOM  n    ONLY AS TO INTRINSIC SAFETY FOR   L USE IN HAZARDOUS LOCATIONS   UL CLASS    DIV I  GROUPS A B C D   EA TEMPERATURE CODE T4  FOR INTRINSIC SAFETY PRECAUTIONS SEE MANUAL  WARNING  UNDERSTAND MANUAL BEFORE OPERATING  ONLY CHANGE BATTERY IN A NOW HAZARDOUS AREA    Ei ia lICTA  AF  20 C Ta 149F 65 C    IIC T4 mE 20 C Ta 149F 65 C  EEFAO4ATEX0384  ea BAS 05 0039                      QXU IDY  N  I UNTED KINGDOM       Os IN HAZARDOUS LOCATIONS   UL CLASS    DIV I  GROUPS A B C D  66Y6 TEMPERATURE CODE T4   FOR INTRINSIC SAFETY PRECAUTIONS SEE MANUAL  WARNING  UNDERSTAND MANUAL BEFORE OPERATING  ONLY CHARGE BATTERY IN A NON HAZARDOUS AREA  EEx ia d IIC T4  4F  20 C Ta 149F 65 C  Ex ia d IIC T4  4F  20 C Ta 149F 65 C  BASEEFAO4ATEX0383   IECEx BAS 05 0038    CE 1180    Warning  To avoid possible personal injury                Substitution of components may impair  intrinsic safety       Do not use the detector if it is damaged or  something is missing  contact Crowcon  immediately       Use only sensor modules specifically  designed for your Gasman 
16. Run mode without performing Zero  The third  option is to disable the Autozero completely     Autozero confirm     Screen Icons N    iw   w Warm up    OK Flashing  icon   Gasman  running  normally          AL    Alarms  AL 2  gl Battery  i          Your unit is now ready to use  NT Aulorero          Familiarize yourself with the gas being monitored in your unit and  make sure you understand site health and safety procedures in the  event of alarm conditions        CROWCON    Gas Detection You Can Trust    Gasman Quickstart guide    Confidence signals   In normal Run mode  Gasman will emit a short beep accompanied by a blue LED flash every 10  seconds and the OK icon flashes to show operational health  These confidence signals can be  disabled using the PC software     2  In the event of an alarm    Alarm signals    In the event of gas concentrations exceeding the alarm thresholds for the  gas being monitored  Gasman will activate the alarm signals     Alarm signals   The red and blue alarm LEDs will  flash  the sounder will emit a loud   fast series of beeps  the internal  vibrator alarm will activate  The  operator screen will display the  alarm level and gas reading  alternately    See the figure to the left     AL   1     Alarm level one       AL   2     Alarm level two    1  When the gas level returns to normal  press the operator button   This will reset your Gasman unit to normal Run mode  If the gas  level is still in alarm  the button will have no effect     The Gasma
17. ced       Do not test the combustible sensor s  response using a butane lighter  Doing so  may damage the sensor       Prior to each days usage  sensitivity must be  tested on a known concentration of  methane equivalent to 25 50  of full scale  concentration  accuracy must be within  022096 of actual   Accuracy may be cor   rected by recalibrating the instrument       Any rapid up scale reading followed by a  declining or erratic reading may indicate a  gas concentration beyond the upper scale  limit  which may be hazardous     Caution    CAUTION  For safety reasons this equipment  must be operated and serviced by qualified  personnel only  Read and understand instruc   tion manual completely before operating or  servicing     ATTENTION  Pour des raisons de s  curit    cet    quipement doit   tre utilis    entretenu et  r  par   uniquement par un personnel qualifi       tudier le manuel d instructions en entir avant  d utiliser  d entretenir ou de r  parer  l   quipement     To avoid damage to the detector       Do not expose the detector to electrical  shock or severe mechanical shock       Do not attempt to disassemble  adjust or  service the detector unless instructions for  that procedure are contained in the manu   and or that part is listed as a replacement  part  Only use Crowcon replacement parts        v    Crowcon Detection Instruments Ltd  2 Blacklands Way  Abingdon  OX14 1DY UK    Tel   44  0 1235 557700    Fax  444  0 1235 557749  WWNW crowcon com  Email  sales 
18. crowcon com         Do not immerse the detector in liquids       The detector warranty will be voided if cus   tomer personnel or third parties damage  the detector during repair attempts  Non   Crowcon repair service attempts void this  warranty        Gasman  Personal Gas Monitor       Contents   UNPACKING 3 3 d dae eae yest gl ote tede debts 1  Quickstart guide    n    anaana illis 3  IRITLOCUGLION    o ON Oe S ees eios des 7  Operation    bed LubP RE ene EAS 9  Batteries x ioi ie Va A et EO SMS EYE GARS 13  Alarm iridications    44er renes ue E esi 15  Fixing accessories       2    17  Fow sampling i5  in ct deae tp pbi eR ttes 18  Maintenance and calibration                      22  PC interface and software                 0 005  23  i module replacement                   0000005 24  Specification steno  cadet ege tex oe ba ees RE 26  Accessories and spare parts                       28  Troubleshooting guide                   2 20005  30    Appendix  Limitations of sensors                   31    Gasman Unpacking       Gasman  Personal Gas Monitor    Thank you for purchasing the new Gasman Personal Gas Monitor   Gasman has redefined portable gas monitoring and will give you years  of unparalleled service and reliability        Please read the instructions carefully before use  Keep the manual for  future reference     Unpacking  Remove the Gasman Personal Gas Monitor from the packaging  The  Gasman accessories will be located in the bottom of the box  Check the  conten
19. dule with one of the same type  instrument specific configuration will be  retained  If replacing with a different i module its default configuration will be loaded     2  Ensure the gasket is in place on the sensor  slip the sensor onto  the sensor housing  Click the quick release fixings around the    24     CROWCON    Gas Detection You Can Trust    Gasman i module replacement    i module board  ensuring the i module is held in place firmly and  the sensor is still tightly located on the module board     3  Place the Gasman unit face down on a surface     4  Unclip the i module from its retaining support clip  Ease one side out at  a time  Ensure the elastomer which is retained with the body of the  support clip remains in place     5  Unpack the new i module and carefully clip it into the support clip   6  Replace the back of the Gasman case     7  Switch on the Gasman unit  The new i module will be automatically  recognized     8  Crowcon advise carrying out a calibration check when any new sensor  is fitted   Re assembling the Gasman unit  1  Replace the back cover and tighten the screws   2  Switch on your Gasman unit  The new sensor will be automatically  identified     Check the filter is in good condition  Replace if any items are faulty   Refer to the troubleshooting guide if necessary      gt    CROWCON 25    Gas Detection You Can Trust    Specification    Gasman       X  Specification       Dimensions    90 x 48 x 24 mm  37 x 1 9 x 1 inches        Weight    130 g 
20. ed to maintain compliance with  certification           CROWCON    Gas Detection You Can Trust    Gasman Quickstart guide       Quickstart guide       1  Getting started    Review your Gasman unit         Gas  Sensor    Sounder  outlet    Operator      Button    Operator LCD  Display Screen    CROWCON    Switching on your unit    Gasman requires little setting up  follow these simple steps to get your  unit ready for use   1  Ensure the unit is in clean air   2  Switch on  Press and hold the operator button for  approximately 3 seconds until the red LEDs  flash   The operator display screen will light up and  the unit will begin a warm up sequence         gt    CROWCON    Gas Detection You Can Trust    Quickstart guide Gasman    Gasman warm up sequence vu c paar au    a  The unit will test the alarm    Sounder    LEDs  sounder  vibration alerts  Alarm test  and the operator display ANS f  screen  The sounder may be XN  silenced by pressing the  button        2  JJ Vibrator Alarm test    b  The unit will continue through a warm up sequence as shown    e   lt        below  this will take approximately 20 seconds           Autozero  There are three Autozero options available  The default setting for  the US will run the Autozero function automatically with no input  from the user  The second option is the manual confirmation of  the Autozero with a single click of the operator button  If this  button is not pressed within the 10 second time out  Gasman will  proceed directly to 
21. flammable  105 g oxygen  90 g toxic       Housing  degree of protection    Ingress protection IP65  NEMA 4   amp  IP67       Operating temperature     20 to  55  C   4 to  131 F        Humidity    0 99  RH  non condensing for continuous  operation       Display    Custom LCD with backlight  Starburst charac   ters for number and text display  plus screen  icons for status and mode        Warm up time    1   minute maximum       Response time  typical      T90    approx 20 seconds for most toxic and  flammable sensors  10 seconds for oxygen        Audible Alarms    95 dBA Multiple alarm sounds allow selection  of distinctive tones for different alarms        Visible Alarms    Dual color red blue flashing LEDs in gas hazard        Vibrating Alarm    Internal vibrating alarm        Repeatability     2  FSD  6 months       Explosion protection    Intrinsically Safe       ATEX    Safety certificate no     Essential Health and Safety Requirement   clause 15 9    BASEEFAO4ATEX0383 Flammable Gas  BASEEFAO4ATEX0384 Oxygen or Toxic Gas       IECEx    IECExBASO5 0038 Flammable Gas  IECEXBASO5 0039 Oxygen or Toxic Gas       Approval codes          Europe  ATEX II 1G EEx ia IIC T4   Tamb    20 to  65  C   Toxic Oxygen  ATEX II 2G EEx ia d IIC T4   Tamb    20 to  65  C   Flammable   USA  Class 1 Division 1  Groups A  B  C and D    Canada  Approvals pending    26       CROWCON    Gas Detection You Can Trust       Gasman Specification  Standards   Safety  EN50014  EN50020  EN50018  94 9 EC  
22. into a mains socket  and connect to  the charger   3  To charge your Gasman unit  simply drop the Gasman into the  charger in an upright position with the screen facing outwards   see diagram below     Power  Supply        Green LED  charging  complete   12 V dc    Red LED  unit charging         gt    CROWCON 13    Gas Detection You Can Trust    Batteries Gasman    The Gasman unit would normally be left switched off for charging  and will display a battery icon on the display screen sweeping from  empty to full  During charging the red LED on the front of the  charger unit will be lit  When charging is complete  the green LED  on the charger unit will illuminate    If the unit is switched on during charging  the normal display will  show the battery icon sweeping from empty to full  On removing  Gasman from the charger unit  the display icon will update in 20  seconds to show actual charge state     The charging time will be longer if the unit is switched on during charging     Gasman is fully charged when the charging battery icon is  flashing  and the green LED on the charger unit is illuminated  In  Run mode  the battery icon will display three bars when it is full     Changing rechargeable batteries    It is recommended that rechargeable batteries should be changed by an  authorized Crowcon service centre     3 3 Non rechargeable batteries  Gasman uses a Lithium Cell battery pack which will give up to two years  of operation     Always switch the Gasman off before opening
23. larm off and the peak level during the alarm       Zero  calibration and gas test with success or failure     Flammable sensor saver on and off       The battery condition is logged every switch on and switch off  while the instrument is operating  and certain configuration  changes are also logged       The Event log can record over 4800 events        CROWCON    Gas Detection You Can Trust    Gasman Batteries       Ill  Batteries       3 1 Rechargeable batteries   Recharge time for the Li ion batteries is less then 6 hours  less  if they are  not fully discharged   Rechargeable batteries will typically last 12  hours  for flammable units     3 2 Gasman Charging Unit   There are 3 models of charging units available for Gasman  a single way  drop in charger unit  a single way charger unit with integral PC  Interface  and a multiway charger unit  The multiway charger unit can  accommodate up to 5 Gasman units  Note  the multiway charger unit  does not have a PC interface option     The charger units are powered by a nominal 12 V dc input  Power sup   plies are available with UK  EUR or US pins  Alternately  an in line supply  of 230 V or 115 V  or a universal 90 260 V power supply are also available  for any other configuration  The multiway charger includes a universal  power supply  A vehicle lighter socket lead can also be supplied  please  see section XI Parts and Accessories     Charging the batteries    1  Ensure you are in a safe area    2  Plug the charger power supply 
24. lowing the user to wear Gasman on a belt  sleeve   jacket lapel etc     Universal harness plate  Crowcon provide a universal harness plate which can be used with either  a chest harness or a shoulder strap     How to wear your Gasman unit       Chest harness C72     Shoulder neck  Use the M3 fixing on the back of yi    3 sonneetors  your Gasman unit to attach the       Ra        chest harness plate  Create a  CES O5 3 M3 nut  chest harness by attaching  e   d i   Waist  strap to the top connectors  to go ST L S    connectors    around the neck  and the other to w  link around the waist using the   side connectors  Adjust the    lengths until the Gasman unit is in  a comfortable working position  Ghost harness plate    Shoulder strap    With the universal harness plate in place on the belt clip  attach the shoul   der strap accessory onto the top connectors  Adjust to a comfortable  working position     See accessories  section XI  for full list     CROWCON 17    Gas Detection You Can Trust    Flow sampling Gasman       VI  Flow sampling    Attaching the flow cap    To perform manual sampling using Gasman  a flow cap must be fitted  onto the front of the unit  over the sensor  A flow cap is included with    your Gasman unit   AN  oO           1  To fit the flow cap  slip the flow cap over the sensor on the front  of your Gasman unit  until the cap is tightly fitted into place     2  Attach the sampling tube or flow accessory onto the gas inlet  nozzle     3  Attach the aspirat
25. n alarm is set to latch by default  The unit will still continue in alarm mode even  when the gas level returns to normal  pressing the operator button will clear the alarm     3  Switch off unit and storage    Switching off unit    1  Press and hold the button for 5 seconds until the display shows  OFF  The display will count down to switch off     P   CROWCON 5    Gas Detection You Can Trust    Quickstart guide Gasman    Storing conditions    In order to optimize sensor performance and lifetime  your Gasman unit  should be stored in a safe  non hazardous area  0 30  C  10 90 RH     4  Additional information   For battery recharging information go to section Ill   For fixing accessories go to section V    For sampling section go to section VI    For calibration information go to section VII     For troubleshooting guide go to section XII     6   CROWCON    Gas Detection You Can Trust    Gasman Introduction       I  Introduction       Thank you for purchasing the new Gasman Personal Monitor  Gasman is  a portable single gas detector  designed to be carried or worn by individ   uals working in hazardous environments such as confined spaces  It is suit   able for use in classified hazardous areas  Gasman monitors a single gas  and displays the reading on a display screen  Alarm warnings are given  through a combination of a loud audible alarm  a bright visual alarm of  blue red flashing LEDs and an internal vibrator  Gasman can be fitted with  a wide range of modular  plug and 
26. or bulb onto the gas outlet nozzle     4  To remove the flow cap  gently tease the cap from the sensor and  lift the cap away from the instrument     When using the manual aspirator kit  adopt a consistent style whilst using  the hand aspirator  Crowcon recommend squeezing once per second to  achieve a flow rate of approximately 0 5   1 litre min  At least 10 pumps  per sample are recommended     The sampling tube supplied is normally a 2m  6ft  length  Longer lengths of sampling tube can  be provided  but will increase the time taken to get a sample from the point of sampling to the  Gasman instrument  When using an extended length of tubing a response time test is  recommended  Gas of known concentration should be sampled along the full length of tubing  to be used and the time taken for the sensor reading to reach the known gas levels should be  noted  This time should be used as the minimum for sampling before readings should be taken     18   CROWCON    Gas Detection You Can Trust    Gasman Gasman gas test accessory kit    Gasman gas test accessory kit    The Gas Test Accessory is a gas testing kit designed to enable gas testing  and one button calibration of your Gasman gas detector using a special   ly formulated  high stability long life single gas mix  It can be used with  Gasman units having sensors for Flammable  Oxygen  Carbon Monoxide  and Hydrogen Sulphide gases     6 1 Gas testing   Gas testing checks the sensor is responding within set limits to an applied  ga
27. play gas sensors  The sensor carries an  intelligent processor which contains calibration and sensor information     Gasman is battery operated and is available with rechargeable batteries or  dry cell battery options  The dry cell battery option is only available for tox   ics and oxygen Gasman units  A battery charger for single and multiple  Gasman units is available for rechargeable units  see section XI for more  information     At Crowcon we recognized the need for a reliable and robust personal  monitoring system  which is both small  light and easy to use  Gasman has  a single operator button  and an intelligent user friendly display with  automatic backlight  The gas level is continuously monitored providing  normal gas readings  peak readings and time weighted averages  TWA    Gasman is available as a diffusion sampling instrument  see section XI for  sampling accessories  Configuration and data event logging is handled by  Crowcon Portables PC software  the PC communication link is provided as  a link on the charger unit        Gasman s compact shape and design makes it comfortable to wear and  as non intrusive as possible  with a non slip grip for better handling  Extra  accessories  such as pocket clip  hard hat clip  alligator clip  shoulder strap  and chest harness  can be purchased     Gasman has been designed from top to bottom to bring you a lighter   compact design with single button operation for ease of use  mainte   nance and extreme reliability  Through
28. s of known composition  This can be performed as often as desired  but  would usually be carried out each time the Gasman is issued for use  The  Gasman itself will determine Pass Fail status for the gas test     In order to perform successful gas tests ensure       The gas used has the correct gas concentration  and that it is  within the validity date specified by the supplier      The gas flow path is leak tight  It is important to check that the  flow cap is properly fitted to the Gasman unit  and the outlet  tubing is not restricted in any way  nor additional tubing length  used    The Gas Test Accessory Kit comprises a gas cylinder containing the gas  a  regulator with interconnect tubing  a magnet   used to activate Test  mode  a flow cap to attach to the Gasman  and a vent line     6 2 How to perform a gas test  1  Ensure your Gasman unit is switched on and in normal operation   2  Fit the flow cap onto the front of the sensor and attach the hose  from the regulator   Attach the outlet hose to    vent gas away      do not extend this  hose and do not restrict or allow kinks   3  Swipe the magnet passed the Crowcon label  Your Gasman unit  will activate the Gas Test and show    TEST    on the display   4  Gasman will display a progress bar  Apply the gas whilst the  progress bar is counting down   Gasman will display    PASS    or    FAIL      gt    CROWCON 19    Gas Detection You Can Trust    Gasman gas test accessory kit Gasman    In the event Gasman displays  FAIL  
29. sence of hazardous gas  will silence the sounder and stop the  vibration alarm  The alarm LEDs will continue to flash     Alarm sounder tone  Different distinctive tones can be selected  for the each alarm condition to achieve the best performance for  the monitoring conditions available     In the event of a Time Weighted Alarm  TWA     In the event the 15 minute or the 8 hour TWA is triggered   Gasman will go into alarm and display a TWA warning with  the toxic gas readings  The 8 hour TWA alarm cannot be  cleared      H  2  amp   0A    A a A a   H    P   CROWCON 15    Gas Detection You Can Trust    Alarm indications Gasman    In the event of a Flammable overange alarm    If flammable gas concentrations exceed 10096 LEL quququ  Gasman locks into alarm and displays  9999  showing an   overange condition  Gasman will temporarily cut off power to the sensor  to prevent burn out and display a progress bar for 200 seconds  When the  timeout is complete  either press the button to continue or optionally  switch your unit off and on to restart  This option is programmable with  Crowcon Portables PC software     16  gt    CROWCON    Gas Detection You Can Trust    Gasman Fixing Accessories       V  Fixing Accessories       Clip Accessories  Gasman is supplied with a simple pocket clip  other clip options are     Hard hat clip  Allows Gasman to be clipped to a hard hat  a method of wearing a  portable gas detector preferred by many users     Alligator clip  A strong alligator clip al
30. te gas  reading    Autozero failed  Cannot Autozero due to  alarm    Calibration expired    Display shows empty  battery symbol on  switch on    30    Function disabled  Zero drifted    Sensor failure    Zeroing in  contaminated  atmosphere  Zeroing in  contaminated  atmosphere    The calibration  due date has  passed    Battery depleted    Action    Recharge or replace battery   Reconfigure with PC software     Restart instrument in clean  air    Do not use  exit hazardous  area immediately  Return  instrument for recalibration  or sensor replacement   Switch off and restart in  clean air     Switch off and restart in  clean air     Send for calibration     Charge or change as  appropriate      gt    CROWCON    Gas Detection You Can Trust    Gasman Appendix  Limitations of sensors       Appendix  Limitations of sensors       Sensor limitations  The sensors used in Gasman have limitations common to all such gas sen   sors  and users should be aware of the points listed below  Crowcon can  advise on particular situations and suggest alternative sensors if the instru   ment is likely to experience extreme conditions     Gasman Flammable uses a catalytic gas sensor  which measures the com   bustibility of the gas  For this reason  readings displayed on the unit will  be unreliable over concentrations of approximately 12096 LEL  Oxygen is  necessary for catalytic sensors to operate  A  catalytic bead saver  is used  to disconnect power to the catalytic bead sensor in the event of
31. thin the unit and  will display an error warning to the user should they occur     8    CROWCON    Gas Detection You Can Trust    Gasman Operation       Il  Operation       2 1 Switch on sequence    1  Ensure the unit is in clean air     2  Switch on  Press and hold the operator button for approximately 3 seconds  until the red LEDs flash     The instrument begins with testing all the LCD segments on the  operator display screen  the red and blue alarm LEDs  sounder and  internal vibrator alert  The sounder may be silenced by pressing  the button  The unit enters a warm up mode and displays a  sequence of screens  see page 3 for more details  At the end of  warm up  the Autozero menu will be displayed        automatically without user confirmation Use this time to check there is   which is the default setting in the USA  or sufficient charge in the battery  is can be disabled altogether  pack    See section VIII PC Interface and software     The Autozero function can be set to run H I Battery check  I  id        Calibration check   During the warm up sequence  if the next calibration date is due in less than 31 days    CAL   nn  will be displayed  where nn is the number of days to the next calibration  If  the date has expired  the Gasman unit will display a warning message  CAL   The  instrument can still function  but it is strongly recommended the unit is sent for  calibration as soon as possible     Press the button to continue to Run mode     Gasman can be set  using the
32. ts are complete  you should have      Gasman unit with standard pocket clip      A configuration report detailing the sensor installed  alarm   settings and a calibration certificate     Calibration flow cap and tubing     User manual     Optional battery chargers and other accessories will be packed in a sep   arate box     Gasman units ordered with an Alligator clip will receive this clip in place  of the standard pocket clip     Battery check  The Gasman Personal Gas Monitor has two battery options  Li ion  rechargeable or non rechargeable batteries     Rechargeable units    Gasman uses a Li ion battery pack and should arrive with sufficient  charge so that the unit can be used straight out the box  However  if  this is the first time you have used the Gasman unit  you may need to  charge the battery to attain the full operating time   The actual operat   ing time will depend on the types of sensor installed   The flammable  Gasman will operate for at least 12 hours on a fully charged battery     gt    CROWCON 1    Gas Detection You Can Trust    Unpacking Gasman    Warning  rechargeable units    Do not attempt to recharge the rechargeable battery with any other  charger other than the Crowcon charger supplied with this unit  Failure  to comply could invalidate safety certification and may result in perma   nent damage to the unit     Non rechargeable units     Gasman uses a CR2 battery which can be easily field replaced  Ensure    only the correct types of battery are fitt
    