Aquamax KF Portable Manual
Contents
1. User Manual GRScientitic Aquamax KF Portable UK Distribution by IMA Ltd Parkwell House Otley Road Guiseley West Yorkshire LS20 8BH England Tel 44 0 1943 878877 Email info ima co uk Website www ima co uk Aquamax KF Portable v 13 08 Aquamax KF Portable v 13 08 CONTENTS Ji G lir ON sox pncd teed o O A Page 3 Principle of Measurement 0 cece cece cece cece nn n nn nn Page 4 Techical SPecticalionS risa a a Page 5 Linpachki e Ser Farts LIST asesinos caidas Page 6 Titration Cell Assembly Instructions Page 7 8 OBS WOE kK I 01 uu uu assasi nasus usnu Page 9 Filling amp Changing Reagents Page 10 Front amp Rear Panel Description Page 11 Serio Time de Dato scroll esco Page 12 Connection amp Start Up Procedure Page 13 Programming A Method Page 14 20 Water standa Lester biie Page 21 Example Methods amp Applications Page 22 24 How To Run A Test Examples V SG amp W w Page 25 26 A usos s sss Loading the paper Page 27 Example Print Out Page 28 Drift COMPpPCNS all OW u u u uuu aber satis esibas Page 29
2. Our dedicated service staff are on hand to help in the unlikely event that your equipment develops a fault Please contact us using the details provided below to discuss the next step For further information about any of our products pricing or details about local representatives just contact Telephone 44 0 1525 404747 Fax 44 0 1525 404848 Email info Ogrscientific com Web www grscientific com GRScientifie 34
3. 2 Weight Dilution ratio W K Use this method if you need to dissolve a sample in another solvent Suitable for samples which will not dissolve in the anode reagent or which might interfere with the KF reaction Result ppm or microgram count bl XK W where W weight of sample dissolved gm K total volume of diluent sample volume injected into Cou Lo Aquamax bl micrograms of water for same injected volume of diluent ug e g Accurately weigh approximately 2 gm of sample into a 20ml stoppered graduated flask Make upto the volume line with suitable solvent diluent Using appropriate syringe inject 100ul of diluted sample into the titration cell Calculation parameters for this example would be W 2 gm K 20ml _ 200 100ul bl microgram count of 100ul injected volume diluent GRScientitic 17 Aquamax KF Portable v 13 08 3 Volume Specific Gravity V SG The most widely used technique for liquid samples However care must be taken to ensure removal of all air bubbles from the syringe Result ppm or microgram count V x SG where V volume of sample ml SG specific gravity 4 Volume V v Result volume microgram count V x 1 0 where V volume of sample ml This is essentially the same as V SG mode except that the sample S G is fixed at 1 0 PRINTER MODES Selecting this program allows the printer to be turned ON or OFF It also enables you to print out the titration par
4. Trace moisture on cell walls If excess drift occurs switch the instrument off remove titration cell from clamp gently swirl the anode reagent around the cell walls replace cell in clamp switch instrument on and allow to precondition OVERTITRATION If the detector electrode senses too much Iodine in the titration cell usually caused by instrument being left unoperated for a long period the display will read OVERTITRATION SEE INSTRUCTION MANUAL Simply add 3 5 microlitres of water or a small amount of a known wet sample until the detector signal bars on bottom left of display are activated Then instrument will automatically carry out its precondition REAGENT LIFE There are three main factors which govern the life of reagents and therefore the frequency of cleaning and recharging the titration cell I The physical size of the titration cell allows for 50 60 mls of sample to be added For the analysis of most liquid samples e g solvents the Aquamax KF Portable normally only requires a few microlitres to be injected therefore the maximum volume of 50 60 mls should not become a limiting factor Oil samples are an exception and are dealt with elsewhere in this manual Ze The total amount of water that can be analysed before saturation of the reagents varies slightly from one reagent manufacturer to another Typically a 100 ml charge of Anode reagent Formula A will analyse upto 1 gm water whilst a 5
5. Drift Test After the initial drift has been calculated in micrograms of water per minute the Aquamax KF is ready for operation XXXXUg Ready The background drift value will usually slowly decrease with time as the titration vessel becomes drier and more stable Although it is possible for the titrator to be used at high drift values it is advisable to wait until the drift value is below 20 ug minute and stable before commencing especially for low water content samples in the ppm ranges The lower and more stable the drift the more accurate the result Aquamax KF Portable v 13 08 PROGRAM FUNCTIONS When the Aquamax KF Portable displays Ready you can enter titration parameters recall or edit existing methods enter new method files etc by using the PROGRAM ENTER CLEAR and e keys Key Functions PROGRAM press PROGRAM key to scroll through menus listed ENTER allows parameter values for the displayed menu to be entered into memory this key has two functions 1 If you have finished entering data press CLEAR and the Aquamax KF Plus will return to the Ready condition 2 If you have started to enter an incorrect value press CLEAR and the Aquamax KF Plus will display the previously stored value you can now select the correct value and press ENTER CLEAR decimal point key allows scroll through of individual parameters in each menu When the required parameter is displayed press ENTER to store it
6. O eE E tas Page 29 A a AAA ukasa sec tauaceneuaewseacto Page 29 leaning the GISSSWare_ uu u us susta sugsbayaqaswaphasquqasaqqahusquapun Page 30 MTOUDICSNOGUMS uu u oops Page 31 NC Hi ass A s uuu ns L s Page 31 AECE OU uu s A Page 32 34 GRScientific 2 Aquamax KF Portable v 13 08 G R Scientific Ltd Tel 44 1525 404747 Fax 44 1525 404848 E mail info gerscientific com Website www grscientific com Correspondence Address Delivery Address Building 69 Building 69 Wrest Park Wrest Park Silsoe Silsoe Bedfordshire Bedfordshire MK45 4HS MK45 4HS United Kingdom United Kingdom INTRODUCTION Key personnel at G R Scientific are recognised experts in coulometric Karl Fischer titration technology and electrochemisty They have designed and manufactured coulometric titrators since the early 1980 s and are widely regarded as some of the leading specialists in this technique Designed primarily for the oil and power industries the Aquamax KF Portable is the only model on the market with a carry handle attached directly to the unit It also features a protective stainless steel cage built in printer and single button operation for water standards analysis using either distilled water or water pellets Specially designed Low Drift Cell twin port glassware avoids the need for Teflon sleeves or grease on the electrodes to seal titration cell from moisture invasion Also available is a carry case for transportation in the field
7. 1 e POW INULA WNVWISOYSd gt gt gt e fpd t Aihe yeuuo4 Byw ynsey wdd soouo e e POGBWII9918 t t a gt gt gt J lt ON Gi ejdwes oa gt gt gt gt rea lt lt lt lt 15 Aquamax KF Portable v 13 08 PROGRAMMING A METHOD After the initial drift has been calculated in micrograms ug of water per minute the Aquamax KF Portable is ready for operation It Is now possible to step through the PROGRAM parameters and enter the required calculation data XXXXUg Ready The Flow Chart on the previous page may prove helpful Press the PROGRAM key until the required option is displayed then simply follow the Flow Chart to enter the specific data values Pressing PROGRAM will move you on to the next prompt pressing CLEAR will return to Ready condition It is not necessary to step through all of the programme functions for every method for example if you only require to change the Beep Mode from Off to Both simple press PROGRAM until display shows Beep Mode press the decimal point key e until display shows Both then press ENTER and CLEAR to return to Ready SAMPLE ID NUMBER Pre
8. PROGRAM or waiting until the end of titration when display will prompt TARE Weight Enter the weight then press CLEAR ANALYSIS METHOD 2 1 Confirm that Aquamax KF Portable is in Ready mode 2 Flush 1 0ml syringe several times minimum 6 times with sample 3 Fit luer needle and flush through with sample 4 Draw sample into syringe beyond the 1 0ml marking 5 Invert syringe so that any air bubbles can be ejected through the needle and adjust syringe plunger to the 1 0 ml mark 6 Wipe off excess sample from outside of needle using a clean dry tissue or paper towel 7 Place syringe on balance and tare to zero 8 Press PROGRAM until Calc Mode displayed then select W w 9 Enter the sample weight as 0 0 gram then press ENTER CLEAR 10 Pierce needle through injection septa of titration vessel 1 2 cm 11 Press START 12 Push needle into anode reagent and inject sample 13 Withdraw needle from titration vessel 14 Reweigh the syringe 15 The net weight of syringe can be entered either during the titration by pressing PROGRAM or waiting until the end of titration when display will prompt TARE Weight Enter the weight then press CLEAR GRScientific 26 Aquamax KF Portable v 13 08 LOADING THE PRINTER SHOWN WITHOUT PROTECTIVE CAGE OR CARRY HANDLE To load the paper into the printer first open the paper roll holder lid above the print head and remove the old roll and rollers Now fit rollers to new paper roll 5030
9. and functions ideal for general laboratory QC and GLP based applications Producing rapid and reliable results that are accurate and reproducible are paramount in all field and laboratory environments All products within the E chem series cover a wide range of applications in areas including food and beverage water environmental power and petrochemical industries E chem electrodes buffers standards We offer a range of combined pH glass epoxy tough and reference electrodes for every application and budget All of our electrodes are manufactured in the UK so you are guaranteed Uncompromised quality and fast delivery For accurate pH and conductivity measurements we offer certified standards traceable to NIST and are tested using ISO 17025 accredited method Our conductivity standards have an accuracy of 1 0 and their shelf life is valid regardless of when the bottle was first opened provided good laboratory practice is adhered to Our pH Buffers are manufactured to exacting specifications according to NIST with an extended shelf life Temperature dependence data is printed on the labels as are lot numbers and expiry dates GR Scientific Service We want you to maximise the return on your equipment through a service program tailored to your requirements from when you purchase our equipment which includes Support and repair Maintenance and calibration Training and installation Quality assurance and certification
10. ml charge of Cathode reagent Formula C will analyse upto 250 mg water Considering that the injected sample volume is normally small and also that the analysis 1s usually for the determination of low levels of water these water capacity values are not normally a limiting factor 3 Similar to all other Karl Fischer reagents coulometric reagents will deteriorate when exposed to sunlight and with increases in temperature Placing the instrument in direct sunlight or near to a heat duct can decrease the reagent life As reagents age the titration speed will reduce and the drift value will increase Aquamax KF Portable v 13 08 CLEANING GLASSWARE IMPORTANT The platinum mesh and wires of the electrodes are fragile and can be easily bent or broken Extreme care MUST be taken when disassembling cleaning and reassembling the titration cell The generator electrode is the most expensive part of the Aquamax KF glassware DO NOT USE BRUSH TO CLEAN INSIDE ELECTRODE AS THIS COULD DAMAGE CONNECTIONS Coulometric reagents are flammable and toxic so care should always be taken when handling them and when cleaning the glassware Whenever possible this should be done in a safe area e g inside a fume cupboard Reagents should not be recharged while the titration cell is still attached to the titrator See page 11 on charging procedure Under normal circumstances the titration cell can be used for a large number of samples before having to repleni
11. recalled method parameters or press CLEAR and the instrument will go back to the Ready condition If you do not want to recall a method press PROGRAM to view the next prompt Aquamax KF Portable v 13 08 WATER STANDARD TEST The display prompt when first switched on reads Water Standard Test By simply pressing just one button the titrator is ready for the operator to inject 10 ul of distilled water as required by the ASTM methods for these analyses After running the 10 000 ug check the titrator reverts back to ordinary user mode The Water Standard Test can also be used to check lower microgram count levels using water standard pellets and pellet sampler In Water Standard Test mode the Aquamax KF Portable will display only microgram counts and print out the final result in micrograms This is an absolute result to check that the titrator is performing correctly and within required parameters For 10 000 ug counts use the 10ul syringe with distilled water for lower counts use the pellet sampler and Water Standard Pellets as shown below When titrator is in Ready mode press Program key and display will show Water Standard Test Press Enter to activate and display will return to Ready mode Either using the 10 microlitre syringe with distilled water or using the water standard pellets and pellet sampler press Start and display will prompt Add standard Inject the 10 microlitres of water or turn the 3 way
12. storage of all results whilst in the field Pellet sampler Power pack amp Water standard pellets in car adapter 10ul 1 0ml amp 5 0ml syringes Results Manager Syringe USB flash drive tray Above the flash drive are two bottles anode reagent two cathode reagent vials and distilled water bottle GRScientific 9 Aquamax KF Portable v 13 08 FILLING amp CHANGING REAGENTS The Aquamax KF Portable has been designed to operate with all major coulometric Karl Fischer reagents however optimum performance can be obtained by using Aquamax KF Formula reagents which have been specifically formulated for use with these instruments For most routine applications 100ml of Formula A anode reagent and 5ml of Formula C cathode reagent are used When analysing Transformer oils Crude oils and other petroleum products Formula A is especially suited as 1t contains other solvents to improve sample miscibility and solubility When analysing samples of Ketones amines or others which may interfere with the reaction it is advisable to use specialised reagents which can be obtained from various suppliers Although reagents can be poured into the titration vessel whilst it is located on the titrator we recommend that the vessel and electrodes are removed from the instrument whilst this procedure is performed to avoid reagent spillage onto the instrument casing Any spillage onto the instrument casing should be wiped off imm
13. the drying tube and seal the top of electrode using the injection port septa and screw cap 3 Connect outlet adapter to wet gas flow meter 4 Flush the sample lines and allow instrument to stabilise 5 Press START and allow gas sample to flow through the cell at an approximate flow rate of 0 5 litres minute 6 Turn off gas flow and allow titration to reach end point Calculation G x 273 t x 22 4 W W moisture content V x 273 x 18 G microgram count V gas volume litres t water temperature of wet gas flow meter C Gas sample adapter Fitted to vessel GRScientific 24 Aquamax KF Portable v 13 08 HOW TO RUN A TEST Please read the section PROGRAMMING A METHOD then use the PROGRAMMING FLOW CHART before running any samples Example Transformer oil sample by V SG Program the Aquamax KF Portable with parameters for the analysis For Transformer Oil samples these parameters are usually Result Format mg kg or ppm Calculation Mode V SG Sample Volume 1 0 ml Sample SG 0 875 Other parameters such as Sample ID number Printer Mode etc are optional ANALYSIS METHOD I Confirm that Aquamax KF Portable is in Ready mode Za Flush 1 0ml syringe several times minimum 6 times with sample 3 Fit luer needle and flush through with sample 4 Draw sample into syringe beyond the 1 0ml marking 5 Invert syringe so that any air bubbles can be ejected through the needle and adjust syringe plunger to t
14. 44 1525 404848 INSTRUMENT CARE Although the instrument casing is highly resistive to most solvents care should be taken to ensure that any spillage is wiped off immediately to avoid staining or possible damage to the paint surface WARRANTY Under the conditions of our warranty all faults which are proved to be due to material construction or manufacturing defects which occur within 36 months of the delivery date will be repaired or replaced at our discretion free of charge at our premises Freight costs will be chargeable along with any inspection work which was not necessitated by manufacturing or material defects Breakages of glassware or electrodes are not covered by this guarantee In the event of instrument failure during the warranty period written authorisation must be obtained from the manufacturer prior to shipping Aquamax KF Portable v 13 08 OTHER PRODUCTS IN THE GR SCIENTIFIC RANGE Aquamax KF ECO This is an economical entry level titrator which has single injection port glassware full numeric keypad for data entry results manager software for data storage and retrieval For use with generator electrodes either with or without frit This small footprint unit offers excellent value Aquamax KF Plus This is the truly versatile model which includes a built in battery and printer to enable use in areas where the power supply is irregular Low Drift Cell twin port glassware results manager software and accessories a
15. 73 load as shown and insert the paper into the slot at the rear of the print head As soon as the printer sees the paper it will automatically load it To avoid damage to the printer mechanism it is advisable to only use the correct part number 503073 thermal paper rolls Cut the end of the paper roll at a slight angle as shown This will make it easier to feed into the printer The mechanism has optical sensors which detect the paper and will auto feed Schematic of paper feed Paper Roll Fit rollers onto both sides of paper roll Locate paper roll in to compartment GRScientific af Aquamax KF Portable v 13 08 PRINT OUTS The Aquamax KF Portable allows you to select different print modes If you choose to have the printer on you will receive a hard copy of the result obtained together with the selected parameters for that particular titration every time a titration is performed Options available Off no result print out Every result will print result after every titration first result also prints all calculation input data Full Report will print results after every titration and then print out statistical data after a selected number of titrations The Aquamax KF Portable automatically increments the next run number every time you press the start button The run counter is reset to zero when power is switched off and back on again A duplicate set of results can be printed out simply by holding down the start b
16. WVdOOtd gt gt gt t e e AVWATO ANVEYOOUd gt gt gt gt gt gt gt 3IN3 sun 66 oJdn Juud sonsnelS 1 quinu Jajua gt gt gt gt SUINIOA tt e OOCUd INI A ajdwes oney fiM lt SOtHd LN3 vonnia DOYA LN3 ldues JUBISM E tt e DOYNd IN3 ROL YABIC um joA YIIN3 SjiodSs DOYA LN3 adwes n A J u anjea J1 u y3LN3 gt gt gt gt lt yYIL1N3 oo gt lt y31N3 gt gt gt Y1N3 Td gt gt gt a Y31N3 gt gt gt Y1N3 l 4 gt gt gt o e y31N3 Dro gt gt 1 SOYdA LN3 SOYd LN gt gt gt gt SOYd LN3 SOYA LN3 SDOYd LN3 SOYd LN3 gt suojdo WYYDONd JO YEJS 0 unya 6 0Jaqunupajas lt lt pOUJAA SALES WYeDONd gt gt gt gt e AHJ3ouMA lt WM soouo gt gt gt gt gt eabenfue7 pajas asooyo gt gt gt gt gt gt e e e yog lt uoge pug tt ssad y lt soouo gt gt gt gt gt gt gt gt E e spuocoes 60 e lt lt anjea 1 u gt gt gt y gt gt gt gt a gt e sojnuiw og odn lt lt au
17. ameters being used at that time Simply scroll through the options using the decimal point key e until the required mode is displayed then press ENTER Options available Off no result print out Every result will print result after every titration first result also prints all calculation input data Full Report will print results after every titration and then print out statistical data after a selected number of titrations STATISTICS PRINT The Aquamax KF Portable can be programmed to print out the statistical data from a series of results when using Full Report print mode If no number has been selected the Aquamax will automatically clear after 99 runs If 3 or more runs have been selected the Aquamax will print out the minimum maximum mean SD amp CV values PRINT DRIFT The Aquamax KF Portable can be programmed to print out the drift value for each titration This is a useful indicator of titration cell stability If drift is above 25 ug minute print will show XX This is a warning that a high drift value was present at start of the titration Aquamax KF Portable v 13 08 DELAY TIMES The Aquamax KF Portable has been programmed to accept two different types of delay times one at the start of titration and one at the end as a minimum titration time START DELAY TIME The Aquamax KF Portables has a preset delay of five seconds which allows samples to be dispersed into the reagents before the titration star
18. ample injected Weight of sample analysed Res 0 0000 Calculated result water content Aquamax KF Portable v 13 08 DRIFT RATE COMPENSATION At the start of any titration the drift value in the top left corner of the screen in micrograms per minute 1s stored in memory This ensures that the displayed result Is corrected for any ingress of atmospheric moisture during the titration period Whilst in READY condition the Aquamax KF Portable continually compensates for any drift caused by atmospheric moisture ingress or reagent decomposition and displays this information which is updated every 10 seconds If the drift value is above 25 ug minute the display will show High Drift XX This is a warning that the value is high however the instrument can still be used by simply pressing START to continue Provided the drift rate 1s stable then it usually acceptable to continue operation The Aquamax titration cell design can provide drift values down to less than 5 ug minute but can also operate at much higher values The maximum drift value is 60 ug minute but we recommend operating upto a maximum of 25 whenever possible If the drift value is above 60 ug per minute the display will read EXCESS DRIFT SEE INSTRUCTION MANUAL Most common causes of excess drift are Titration vessel not properly sealed check septa amp fittings Reagent almost depleted clean amp recharge the cell Sample introduced before pressing START key
19. and an OPS Workstation which includes everything the user needs such as syringes standards reagents etc all in one footprint for ultimate efficiency Aquamax KF Formula reagents have been specially formulated for use with Aquamax KF titrators Formula A anode reagent is suitable for most routine applications and is especially useful for water content determination of oil samples e g transformer oils crude oils etc This anode reagent is supplied in single shot bottles of 100ml No measuring of volumes required No mixing of other solvents required Cathode reagent Formula C is supplied in single shot 5ml vials which have safety snappers pre fitted thereby reducing risk to the operator Where appropriate G R Scientific products include traceable calibration certificates and are backed up by a three year parts and labour warranty These titrators are used by many leading companies in the oil petrochemical gas pharmaceutical biotechnology automotive and power generation industries For a list of users typical applications or details of our network of international distributors please visit our website www grscientific com or contact G R Scientific Ltd on Tel 44 1525 404747 Fax 44 1525 404848 E mail infoO grscientific com To obtain the full benefits that this instrument can offer it 1s recommended that you read this user manual before assembling the unit Cou Lo and Aquamax KF are registered trade ma
20. cock to allow pellet to drop into titration vessel Remove syringe from titration vessel or close the 3 way cock of pellet sampler immediately after sample introduction After completing Water Standard Test press Program and display will prompt Water Standard Cancel If you want to repeat standard sample analysis simply press Start and introduce next sample Otherwise press Enter to cancel water standard test parameters and revert back to pre programmed methods Tear open the foil pouch containing water standard pellet Carefully pour this into the funnel of pellet sampler device avoid contact with skin to reduce contamination of sample with moisture from your fingers Press Start and turn the 3 way cock so that pellet drops into titration vessel After sample introduction close 3 way cock to avoid ingress of atmospheric moisture Aquamax KF Portable v 13 08 EXAMPLE COULOMETRIC METHODS The Aquamax KF series of Coulometric Karl Fischer titrators can be used for the following standard methods either by direct sample introduction or in conjunction with accessories such as water evaporators ASTM D 1533 D 3401 D 4928 D 5460 D 6304 D 6869 E 1064 API MPMS Ch 10 9 El IP 386 438 BS 6829 1 5 ISO TC 158 SC 10101 1 10101 3 10337 12937 IEC 60814 Insulating liquids Halogenated organic solvents Crude oils Rubber compounding materials Petroleum products lubricating oils and additives Plastics Or
21. ediately to avoid damage or staining Remove the drying tube and injection septa Using the funnel supplied charge the titration vessel upto the lower line with Formula A reagent For your convenience these reagents are supplied in single shot bottles which contain 100ml so it is not necessary to measure any volumes simply pour in the complete bottle Also using the funnel charge the inner chamber of the generator electrode with Formula C reagent which are supplied in single shot 5 ml vials which havce safety snappers pre fitted thereby reducing risk to the operator It is not necessary to clean the funnel between reagents Reconnect the drying tube and injection septa so that the titration vessel is sealed from ingress of atmospheric moisture Locate the complete titration vessel onto the titrator and connect the electrode leads onto the appropriate sockets The colours of anode and cathode reagent are only an illustration to show that there are two different liquids These are not necessarily the exact colours of the reagents being used The Aquamax KF Portable titrator is now ready to be switched on Cathode Reagent Anode Reagent Aquamax KF Portable v 13 08 DESCRIPTION Front Panel Printer Display panel Keypad Stirrer speed Rear Panel Power Jack Nominal 18v DC 45w Supply Input Fuse 3 15amp T type Time delay RS232 Standard 4 wire RS232 Input Output RS232 Por
22. ei MENU LIST Sample ID Number upto 8 digits Result Format ug ppm mg kg or water Calculation Mode V SG W W W K or V v Printer Mode off every result or full report Statistics Print upto 99 runs will print min max mean SD amp CV Print Drift on or off Start Delay Time upto 30 minutes Min Titration Time upto 30 minutes End Sensitivity 0 9 seconds Beep Mode off keypress end titration or both Language select language Generator Electrode with frit without frit Save Method store upto 10 methods See PROGRAMMING A METHOD for an explanation of each Menu List parameter Aquamax KF Portable v 13 08 PROGRAMMING FLOW CHART CECCECEECEC CCC CC tt 4 M R ee ek ee gt 1 e yVI19 MNVyS9OYd gt gt gt gt gt gt gt gt gt en on do YVATD AVYSOYA e e e e e e e e e e gt gt gt o o lt t e ee e e e gt njeA 4 lt e e 331N3 yuejg e e e e e _ gt gt gt gt 1 eee wqV31O WWVdOOHgdd gt gt gt o 1 e yYVI1D MWNV9OYAd gt gt gt C e e yV3ID MWNVISOYdA gt gt gt a gt t t e e yvV3ID9 NVI9QOYd e gt gt gt gt lt e yV3IID MWNVEQOYAd gt gt gt e e e yVW319 MWNVy9OYd gt gt gt gt e wWqV3D
23. eper inactive Keypress beep once every time a key is pressed End titration beep three times at the end of a titration Both beep after every keypress and three times at the end of a titration Aquamax KF Portable v 13 08 LANGUAGE The Aquamax KF Portable can be programmed to show all display prompts and print outs in various languages Use the decimal point key e to select language then press ENTER GENERATOR ELECTRODE The Aquamax KF Portable can be operated using generator electrodes with or without frit diaphragm Use the decimal point key e to select either with or without frit then press ENTER The titration electrolysis current settings are automatically adjusted for each style of generator and appropriate reagents SAVE METHOD If you want to save a method for future use this can be done by pressing a numerical key O 9 and ENTER All titration parameters printer mode beeper mode etc are stored automatically Display then shows Save Method Done Then press CLEAR and the instrument will revert back to the Ready condition RECALL METHOD If any methods have been saved to memory then the first prompt before Sample ID number will be Recall Method When this prompt is displayed simply press the number of the method file you wish to recall and press ENTER There are 10 method files available numbered from 0 to 9 The titration parameters etc will all be automatically recalled Press PROGRAM to view the
24. er sample analysis compared to standard potentiometers Supplied with an optimized titration vessel and micro combination electrode that regenerates to optimum efficiency in minutes and offers excellent stability for non aqueous titrations microTAN OPS Workstation Everything you need in one footprint and transportable The MicroTAN workstation is an organised workspace containing all equipment and chemicals required to run your samples MicroTAN unit chemicals and accessories are contained in a sturdy outer case with a firm foam lining in an ergonomic arrangement Optional accessories include a bottle top dispenser for dedicated dispensing of titration solvent and a mini thermal printer TitraMax Titrator Is a simple versatile yet robust potentiometer for measuring TAN TBN mercaptan sulphur and analysis of petroleum products Pre programmed methods that conform to ASTM standards and an interchangeable burette system with intelligent volume recognition are key features of the Titramax In addition each unit is supplied with bespoke UK manufactured electrodes for optimum performance and reproducible results GRScientific 33 Aquamax KF Portable v 13 08 E chem pH amp Conductivity meters The E chem range of pH conductivity benchtop meters have been developed to provide cost effective robust and simple to use solutions for your needs The E512 pH meter offers research grade specifications with a comprehensive range of features
25. ganic liquids Crude oil Crude petroleum Petroleum products Surface active agents Natural gas and gas substitutes Natural gas Natural gas Crude petroleum Petroleum products Insulating liquids Aquamax KF Portable v 13 08 EXAMPLE APPLICATIONS LIQUID SAMPLES Liquid samples are usually introduced into the titration vessel by using a syringe The greater the amount of water introduced into the titration cell will cause the titration to take longer and also consume the reagents more quickly It 1s recommended that the smallest representative amount of sample be used to ensure fastest titration speed and maximum sample throughput before reagents have to be replenished Supplied as standard with every Aquamax KF titrator is a Iml syringe and luer needle These are suitable for injection of oil samples and water standards Other volume syringes are commercially available from various suppliers Please ensure that the needle is long enough to reach into the anode reagent This guide may be of assistance when selecting injection volume of samples Expected water content Suggested sample size l WO A 1 or 2 ml 10 100 ppm 1 or 2 ml 100 500 ppm 0 5orlml 50 1000 ppm 0 Sorlml 0 001 0 01 a l or 2 ml 0 0 0 1 C 2 asasshakayaaska or 2 ml Orel O89 Tesis 0 5 or 1 ml 03 I O Vinnie 0 5 or 1 ml TsO S0 A 0 2 ml above FO as ssssrzussasa 0 1 ml o
26. he 1 0 ml mark 6 Wipe off excess sample from outside of needle using a clean dry tissue or paper towel he Pierce needle through injection septa of titration vessel 1 2 cm 8 Press START 9 Push needle into anode reagent and inject sample 10 Withdraw needle from titration vessel 11 Read result in mg kg ppm water on display and printout 12 Repeat steps 2 11 1f duplicate result required Aquamax KF Portable v 13 08 Example Crude Oil by W w Program the Aquamax KF Portable with parameters for the analysis ANALYSIS METHOD 1 1 Confirm that Aquamax KF Portable is in Ready mode 2 Flush 1 0ml syringe several times minimum 6 times with sample 3 Fit luer needle and flush through with sample 4 Draw sample into syringe beyond the 1 0ml marking 5 Invert syringe so that any air bubbles can be ejected through the needle and adjust syringe plunger to the 1 0 ml mark 6 Wipe off excess sample from outside of needle using a clean dry tissue or paper towel 7 Weigh syringe 8 Press PROGRAM until Calculation Mode displayed then select W w 9 Enter the syringe weight when prompted by display then press ENTER CLEAR 10 Pierce needle through injection septa of titration vessel 1 2 cm 11 Press START 12 Push needle into anode reagent and inject sample 13 Withdraw needle from titration vessel 14 Reweigh the syringe 15 The net weight of syringe can be entered either during the titration by pressing
27. is has been done the display will start to show the calculated result in ppm mg kg or water Aquamax KF Portable v 13 08 CALCULATION MODE The Aquamax KF Portable has four standard calculation modes in memory Each of these allow the instrument to divide the microgram count by the gram weight of sample introduced so that the calculated result can be printed out in either ppm or percentage of water Press the decimal point key e to scroll through the options available Select the calculation mode required and press ENTER The display will prompt you with the appropriate parameters Enter the numerical value for each parameter and press ENTER The four standard calculations are 1 Weight difference W w This is the most accurate calculation and is suitable for almost all types of samples powders liquids gases etc Result ppm or microgram count W w Where W total weight of sampler sample gm w tare weight of empty sampler gm After entering the total weight W the instrument will return to the Ready condition The Tare weight w can be entered at any time during the titration by pressing PROGRAM or at the end of titration For ease of operation it is possible to enter the Total weight W as 0 0000gm then place the full sampler on the balance and tare to zero Thereafter it is only necessary to enter the weight difference tare weight w for each titration This procedure reduces potential data entry errors
28. is means that upto 50 samples would fill the titration cell Provided that the other criteria governing reagent life total water titrated and reagent age have not been reached then it is possible to switch off the titrator and allow the oil to separate from the reagent The oil can then be siphoned off and the instrument switched on again After this operation it may take upto 30 minutes before the instrument settles down to a stable baseline If it takes longer than 30 minutes then it is advisable to clean the cell and recharge with fresh reagents For optimum performance use Aquamax KF Formula A reagent TURBINE LUBRICATING OILS For most Lube Oil applications a sample size of Iml is sufficient However if it is suspected that the water content is above then a sample size of 0 2 to 0 5 ml is suggested Some lubricating oils contain additives which can coat the detector electrode or block the glass frit of the generator electrode For these applications we suggest using the Aquamax KF Portable in conjunction with an oil evaporator such as the model 51000 GAS SAMPLES Gas samples can be introduced by fitting a needle onto the gas sampling bomb and selecting Calculation mode W w Alternatively gas samples can be bubbled through the titration vessel and the passed volume measured For this method we suggest using the gas analysis kit 503074 1 Fit the gas inlet and outlet adapters in place of the injection septa 2 Remove
29. n Jajua gt gt gt gt gt gt gt tt e sanu o odn lt au 19 u gt gt gt tt lt UO asooyo gt gt gt gt gt gt gt gt gt gt gt gt tft e e ynsay Ai9A3 J lt Hod glind4d soouo gt gt gt gt gt 1 NA Ll SOtid LN3 WM o AM 8POW 9eD di 9S A Et tt tt 2 lt ygi yai e 60J9qunu pajas gt gt gt o gt gt 1 gt dword 3x8u 0 uo SAQU 07 INWHODONG Sseud YVAID NVIDOYA URINE lt lt gt gt gt unKue Je AGV38t 0 wN 0 yV310 Ssd Sybip g odn suequunu Ja ua gt gt gt o o gt t gt gt gt gt gt gt gt gt gt 1 gt gt gt gt gt gt gt gt gt dd 1 e e ed Joyeseuag WVYDOYd 1 gt gt gt gt gt gt gt 1 1 gt gt gt gt 1 e t to E 1 pow desg WNV3Y9OYdA E Gs 1 1 yAnisuas puq dd gt gt gt gt gt gt gt gt gt unjuoneji UW NWYDONd 1 gt gt gt gt gt gt gt gt gt au ed HEIS WVYDOYd 1 gt gt gt gt gt gt gt gt tt e t ct e yug Jung es gt gt gt A gt gt gt 3 gt 1 e ds
30. n battery may lose some charge if stored for prolonged period We advise connecting the Aquamax KF Portable to mains power supply for 12 hours before first switching the power on After unpacking your Aquamax KF Portable check that the glassware accessories pack contains Item Description Low Drift Cell l pce Titration Vessel twin port 503053 l pce Detector electrode LDC 503054 1 pce Generator electrode LDC 503055 1 pce Desiccant tube amp cap 503058 pack Injection septa 10 503059 1 pce Iml Gas Tight syringe 503063 pce Luer needle Pair 17 amp 19 gauge 503067 1 bottle Molecular sieve 503070 1 pce Stirrer bar 503071 l pce Funnel 503072 1 pce Thermal paper roll 503073 ACCESSORIES 1 pce Mains power pack 401607 1 pce In Car Adapter 401421 GRScientific 6 Aquamax KF Portable v 13 08 AQUAMAX KF TITRATION CELL STEP BY STEP ASSEMBLY INSTRUCTIONS SHOWN WITHOUT PROTECTIVE CAGE OR CARRY HANDLE Position titration vessel on Aquamax KF titrator The glass joints do not require PTFE sleeves or grease The special screw fittings seal the vessel from ingress of moisture from the atmosphere and are easily released even after several weeks Position electrodes on titration vessel It is not necessary to overtighten these screw joints Connect leads to generator and detector electrodes then connect the leads to the titrator Generator electrode lead connects to socket marked GENERATOR detector electrode lead connects
31. operating time is 8 hours continuous use This may be reduced if repeated high water contents are being determined or if the printer is being continually used GRScientific 11 AAA Aquamax KF Portable v 13 08 SETTING THE DATE AND TIME 1 Switch on the Aquamax KF Portable and wait for it to display the following Aquamax KF Portable Press START 2 Hold down the START key until the Aquamax KF Portable displays Set Date Time DD MM YYYY HH MM SS 3 Use the decimal point key to move the cursor across the screen 4 Use the numerical keys to set alter the required number 5 When completed press the ENTER key to save the changes and exit Aquamax KF Portable v 13 08 CONNECTION amp START UP PROCEDURE Assemble glassware see Assembly Instructions Charge vessel and generator electrode with reagents see Filling amp Changing Reagents Locate assembled titration vessel onto the Aquamax KF Portable Connect electrode leads Connect mains lead power cord Switch ON using switch on rear panel and wait for display to show a Aquamax KF Portable Press START If required adjust the stirrer speed using the black toggle switch on the front of the instrument 7 Press START and allow the instrument to precondition equilibrate microgram count Aquamax KF XXXX Ug MMII 4 Precondition detector signal titration speed At the end of the precondition period the display will show XXXXUg
32. or PTFE sleeves required AC polarisation Visual display print out acoustic beep Possible lug 200 mg water Typical lug 10mg water Ippm 100 water O lug 2 24 mg per minute 400 ma Automatically controlled 10 100ug 3ug 100ug 1mg 3ug above Img 0 3 0 30 minutes user selectable 0 30 minutes user selectable Weight weight W w Weight dilution ratio W K Volume density V SG Volume volume V v ug mg kg ppm ug mg kg ppm max mean min values up to 99 runs 10 user programmable methods stored plus multi method via pc link User programmable Microprocessor controlled Multi languages user selectable Analysis time amp date print out Display amp print out indication USB and RS232 ports Flash drive memory stick 15 key touchpad 40 character alphanumeric backlit LCD 42 character high speed thermal printer Stainless steel Built in 90 264V AC 47 63 Hz 12V DC car adapter internal battery 250 x 245 x 300 mm incl cage 4 kg Aquamax KF Portable v 13 08 UNPACKING amp PARTS LIST Inside the shipping carton you will find a plastic case which contains the glassware and accessories pack and the user manual The Aquamax KF titrator is located underneath the glassware case or inside the optional carry case if ordered The mains power pack and in car adapter are inside the carry case lid compartment Remove all of these items and check that they have arrived undamaged The built i
33. r less Please note that these volumes are only a guide You may wish to experlment with sample size to determine the optimum balance between repeatability reagent depletion and speed of analysis CRUDE OILS IP 386 amp 438 ASTM D1533 D4928 D6304 API MPMS Chapter 10 9 ISO 10337 amp 12937 Most coulometric reagents contain Methanol as the main solvent and as such it is possible that waxy deposits from crude oils can drop out and contaminate the electrodes requiring more frequent and thorough cleaning of the glassware This problem can be reduced by using Aquamax KF Formula A anode reagent which has been specially formulated for this application The most common interference in crude oil samples is usually caused by mercaptans or sulphides Samples containing alkyl groups will react stoichiometrically so that 100 ppm Mercaptans would show as approximately 30 ppm water 100 ppm Sulphides would show as approximately 50 ppm water In normal practice oils containing less than 500 ppm mercaptans or sulphides are regarded as having no appreciable effect Repeatability problems are usually caused by sampling inaccuracies or improper homogenisation of the sample Aquamax KF Portable v 13 08 TRANSFORMER INSULATING OILS ASTM D 1533 BS EN 60814 IEC 60814 Due to the viscosity of these samples and the fact that typical water content levels are in the range 5 40 ppm it is usually necessary to inject sample volumes of 1 ml Th
34. re supplied as standard As well as being a stand alone instrument the Aquamax KF Plus can be combined with our Oil Evaporator or a Solids Evaporator system Suitable for a wide range of applications the Aquamax KF Plus can be used to determine water contents of liquids gases and solids Aquamax KF Oil Evaporator For combined operation with the Aquamax KF Plus titrator to determine water content of additives lube oils base oils automatic transmission fluids hydrocarbon solvents and other petroleum products Aquamax KF Solids Evaporator For combined operation with the Aquamax KF Plus titrator to determine water content of solids such as plastics nylons paper powders etc Aquamax KF Portable v 13 08 Aquamax KF Volumetric An adaptable titrator capable of titrating with superior accuracy and precision even for samples with low moisture content Dual syringes enable the operator to dispense titrant and standard automatically eliminating cell opening which may introduce moisture to the system The Aquamax KF Volumetric comes equipped with a solvent handling system to reduce cell conditioning time and can be connected directly to your laboratory balance via serial interface Choose from four pre programmed methods or develop a custom method for almost any application or sample type microTAN Titrator An innovative micro system for measuring Total Acid Number which enables the user to save up to 75 of organic solvent p
35. rks of G R Scientific Ltd ACE control system is patented technique owned by G R Scientific Ltd All Aquamax KF power supplies are double insulated units that do not need an earth ground connection These units are sealed and conform to CSA LR84459 UL listings and CE requirements Aquamax KF titrators only require a low voltage 15V input to the electronic control circuitry The moulded plastic casing of the instrument also provides additional insulation Aquamax KF Portable v 13 08 PRINCIPLE OF MEASUREMENT Karl Fischer titration is simply a means to measure water content of samples Modern instruments such as the Aquamax KF use the coulometric principle whereby the water present in the sample is coulometrically titrated to a predefined end point at which there is a minute excess of free iodine present Stoichiometrically 1 mole of water will react with 1 mole of iodine so that 1 milligram of water is equivalent to 10 71 coulombs of electricity Combining the coulometric technique with Karl Fischer titration Aquamax KF titrators determine the water content of the sample by measuring the amount of electrolysis current necessary to produce the required iodine This is an absolute technique which does not require calibration of the reagents For further information visit www grscientific com and download the PDF article titled Recent Advances 502 l2 502 3C5H5N H20 gt 2C5H5N HI C5H5 N S O SO2 p SO4CH3 CH30H g
36. rying procedure has not been properly carried out Aquamax KF Portable v 13 08 TROUBLESHOOTING OVERTITRATION Detector senses too much Iodine Usually caused by old reagents or stirrer switched off Check stirrer speed Add 3 5 microlitres of water until detector bars activate and allow to precondition EXCESS DRIFT Indicates excess moisture ingress or condensation on cell walls Switch OFF Remove titration cell from clamp and gently swirl anode reagent around the cell walls Reconnect switch ON and allow to Precondition LONG PRECONDITIONING TIME Normally caused by glassware not being sufficiently dried before assembly Re clean glassware and electrodes dry thoroughly and charge with fresh reagents PROGRESSIVELY LOWER RESULTS FOR SAME SAMPLE Moisture contamination of syringe Flush syringe and needle with sample several times before beginning new test POOR REPEATABILITY FOR SAME SAMPLE Sample size probably too small or moisture contamination of syringe Increase sample size and flush syringe and needle several times before beginning new test WARNING LETTERS ON PRINT OUT Warning letters will be printed after time of each analysis to signify the following U Uncalibrated E Calibration expired B Battery level low A ACE system active Call G R Scientific Ltd or your local dealer for suggestions on sampling technique for your application E mail info grscientific com TEL 44 1525 404747 FAX
37. sh the reagents Once the reagents have been exhausted or when the titration cell maximum volume has been reached it should simply be necessary to 1 DISASSEMBL E titration cell disconnect leads from titrator Ze EMPTY the titration cell and generator electrode 3 RINSE all parts with methanol Do not use brush on electrodes 4 DRY all parts 5 REASSEMBLE glassware 6 RECHARGE with fresh reagents If the cell 1s heavily contaminated then it may be necessary to clean it more thoroughly For oil samples cleaning with chloroform or xylene is suggested whilst for salt deposits a water wash may be required Use whichever solvent is most suited for the sample type The titration vessel can even be cleaned with hot soapy water and a bottle brush However after cleaning with suitable solvent all glassware parts MUST be rinsed inside and out with methanol They can then be dried with a warm air blower such as a domestic hair dryer placed in a low temperature oven at 40 50 C or left in a desiccator After being fully dried reassemble the titration cell and charge with fresh reagents THE MORE THOROUGHLY THAT THE TITRATION CELL IS CLEANED AND DRIED THE FASTER THE INSTRUMENT WILL STABILISE READY FOR OPERATION AFTER RECHARGING REAGENTS Under normal conditions the Aquamax KF Portable should be ready for operation within 5 10 minutes after reassembly however it could take up to 20 minutes before to completely stabilise if the d
38. ss PROGRAM Display will show Key upto 8 numbers gt Press ENTER PROGRAM Sample ID No Sample ID No ENTER 12345678 If you do not want to enter any further parameters simply press CLEAR to go back to the Ready condition Press PROGRAM if you want to continue entering other parameters RESULT FORMAT Press PROGRAM gt Display will show Result Format PROGRAM Te Results can be selected in ug ppm mg kg or water If you want to select the format already stored in memory and shown on display press PROGRAM and the instrument will go to the next prompt If you want to change format to give results in ppm mg kg or you can scroll through these options by pressing decimal point key e until the required format is displayed then press ENTER to store this in memory and PROGRAM to move onto the next prompt If you have selected a result format in ppm mg kg or the instrument will prompt you for the necessary calculation mode and data values Result Format Calculation Mode ENTER PROGRAM V SG The display shows the calculated water content according to the programmed Result Format and Calculation Mode program settings ie whatever has been chosen to be shown printed at the end of the titration The only exception to this is in the W w mode when the Tare Weight has not yet been entered In this case the display will show the ug count The PROGRAM key can be used to enter the Tare Weight at any time during the titration once th
39. t 4 15vDC 40w USB B Direct cable to PC Results Manager Port USB A Removable Flash Memory Store Port RS232 Port For updating titrator software connecting to other equipment for remote controlling software specific protocols to be provided USB B Port For connecting directly to PC to export data to Results Manager and enable results to be saved on files for future reference See Results Manager User Manual section USB A Port For exporting results data to a Removable Flash Drive memory stick when running sample tests in the field or remote locations and subsequent transfer to PC at a later date for loading onto Results Manager files To use the memory store insert Flash Drive into USB port A titrator will see it and a symbol will appear on the display The symbol will appear blacked out while the drive is being written to and then as when idle The drive should only be removed when in the idle mode To read data saved on drive insert 1t into PC USB port Start Results Manager load selected file and then save in the format you require see Results Manager user manual for full instructions There 1s a separate power ON OFF switch on the rear panel Whilst connected to mains power supply a trickle charge is continuously applied to the battery It is therefore recommended that the instrument is used from mains power supply whenever possible so that the battery is kept fully charged Maximum battery
40. t C5 H5 N S CsHsN 2 2e gt l2 Using the latest pulse current technology and our patented ACE control system Patent No GB2370641 the Aquamax KF automatically selects the appropriate titration speed dependent upon the amount of water present in the sample The titration speed is reduced as the end point is approached and when the titration is completed the instrument prints out and displays the results For a more detailed explanation of the technique and a comparison with volumetric Karl Fischer titration please visit our website www grscientific com SCHEMATIC OF TYPICAL TITRATION ELECTROLYSIS CURRENT DETECTOR SIGNAL Technical Specifications Titration Method Electrolysis Control Titration vessel End Point Detection End point indication Measuring range Moisture range Max sensitivity Max titration speed Max current Drift compensation Precision Start delay time End delay time Calculation modes user programmable Display format Print format Statistics Method storage Sample ID number Stirrer speed Languages Calendar clock Battery low indicator Data outputs Removable Data storage Data Entry Display Printer Protection cage Carry handle Power supply Dimensions Weight Aquamax KF Portable v 13 08 Aquamax KF Portable Coulometric Karl Fischer titration Patented ACE control system GB2370641 Low Drift Cell design no grease
41. to socket marked DETECTOR Place stirrer bar inside titration vessel Fit injection septa into plastic screw cap connectors and locate onto injection ports of titration vessel Fill drying tube with molecular sieve and locate on top of generator electrode The assembled titration cell is now ready to be charged with reagents prior to use Please be sure to assemble correctly every time the cell is cleaned or re filled It is recommended that the titrator is NOT switched on until after the titration vessel has been charged with reagents This will avoid damage being caused to the electrodes by the stirrer bar Avoid twisting the electrode leads otherwise they may become damaged Aquamax KF Portable v 13 08 AQUAMAX KF TITRATION CELL Generator III i k Drying Tube Electrode Cap D Ring T II 4 assembly o Screw Cap generator _ Electrode Detector Electrode Injection Port Cap Sept Titration Vessel Aquamax KF Portable v 13 08 OPS WORKSTATION The Aquamax KF Portable can also be supplied complete with an OPS Workstation A sturdy outer case secures and enables transportation of the titrator complete with spare anode reagent bottles cathode reagent vials distilled water and 10 microlitre syringe of water standard test pellet sampler and water standard pellets 1 0ml amp 5 0ml gas tight syringes and luer needles power pack in car adapter and Results Manager USB flash drive for
42. ts Some samples particularly powders and viscous samples may need additional time to fully dissolve By selecting the start delay program you can increase this time by upto 30 minutes Display shows Start Delay Time 00 mins 00 secs Select required time delay and press ENTER Whilst a delay time 1s in operation the display will show add sample and the clock counter will show the time remaining before the start of titration A start delay can be overridden by simply pressing START MINIMUM TITRATION TIME For some applications such as certain gas samples or when using the Aquamax KF Portable combined with an oil evaporator or solids evaporator system it can prove useful to delay the end point of the titration By selecting this parameter you can increase the minimum titration time by upto 30 minutes Display shows Min Titration Time 00 mins 15 secs Select required minimum time and press ENTER END SENSITIVITY The sensitivity of detector signal as the end point is approached can be set from 0 9 seconds The default value is 3 seconds and there are very few occasions when this may need to be altered Decreasing this value can prove useful if samples cause interference e g some Lube Oil samples BEEP MODE Four options are available for the audible alarm beeper and they can be scrolled through using the decimal point key e until the required one is displayed Then press ENTER The options are Off be
43. utton for 3 seconds Results memory is cleared after duplicate print out or when power is switched off PRINT OUT EXPLANATION EXAMPLE OF V SG CALCULATION Aquamax KF Portable V XXXX Serial No 71000 Calibrated DD MM YYYY Titrator model Software version Titrator serial number Date titrator calibrated Run Date Date sample run Run 1 Run Time F Run number amp time of analysis F generator electrode with frit Nett Count 0 0 ug Nett microgram count after drift correction Drift 00 Drift value at start of titration Result Format To Result format selected Calculation Mode V SG Calculation mode selected Volume 1 0000 ml Sample volume Density 1 0000 Sample density Res 0 0000 Calculated result water content Subsequent runs of the same sample and calculation parameters show run Run 2 Run Time number run time nett count Nett Count 0 0 ug Drift value and calculated result Drift 00 Res 0 0000 EXAMPLE OF W w CALCULATION Aquamax KF Portable V XXXX Serial No 71000 Calibrated DD MM YYYY Run Date Run 1 Run Time N Nett Count 0 0 ug Drift 00 Result Format Calculation Mode W w Total Weight 0 0000 gm Tare Weight 0 0000 gm Sample Weight 0 0000 gm Run number amp time of analysis N generator electrode without frit Nett microgram count after drift correction Drift value at start of titration Result format selected Calculation mode selected Weight of syringe plus sample Weight of syringe after s