Home
        Product Manual - Drive Medical
         Contents
1.           ways   Alarm level Audible prompt  High    DO DO DO      DO DO  DO DO DO      DO DO     Medium    DO DO DO     Low    DO      5 3 2 Alarm Lamp    When a physiological alarm occurs  the alarm levels are indicated in the following  different visual ways                 Alarm level Visual prompt   High Alarm lamp flashes in red with 2 Hz    Medium Alarm lamp flashes in yellow with 0 5 Hz    Low Alarm lamp lights on in yellow without flashing                 When a technical alarm occurs  the alarm levels are indicated in following different  visual ways                   Alarm level   Visual prompt       Medium   Alarm lamp flashes in blue with 0 5 Hz        Low   Alarm lamp lights on in blue without flashing     Caution      When multiple alarms of different levels occur at the same time  the  monitor will select the alarm of the highest level and give visual and    audible alarm indications     5 3    Vital Signs Monitor User s Manual       5 3 3 Alarm Message      Physiological alarm  Physiological alarm messages are displayed in the physiological alarm area  The  system uses different symbols and background colors for the alarm messages to match    the alarm level as follows                                Alarm level symbol background color  High kkk red  Medium     Low a yellow  Caution    O The monitor won t give physiological alarm under clinic mode     e Technical alarm  Technical alarm messages are displayed in the technical alarm area  The system  uses differ
2.     Medium    SpO   Low SpO  measuring value is below the low  alarm limit  SpO   Desat SpO  measuring value is too low  High  PR High PR measuring value is above the high  alarm limit        Medium    PR Low PR measuring value is below the low  alarm limit  SpO  No Pulse SpO  signal is predominantly invalid and High  therefore cannot be analyzed  Oo NIBP  Alarm messages Cause Level  NIBP Sys High NIBP Sys measuring value is above high  alarm limit  NIBP Sys Low NIBP Sys measuring value is below low  alarm limit  NIBP Dia High NIBP Dia measuring value is above high  alarm limit                Medium      NIBP Dia Low NIBP Dia measuring value is below low  alarm limit  NIBP Map High NIBP Map measuring value is above high  alarm limit  NIBP Map Low NIBP Map measuring value is below low  alarm limit                Vital Signs Monitor User s Manual                                                                                        e CO   Alarm messages Cause Level  EtCO  High EtCO  measuring value is above high alarm  limit   EtCO  Low EtCO  measuring value is below low alarm  limit   FiCO  High FiCO  measuring value is above high alarm 3  bats Medium    limit   awRR High awRR measuring value is above high alarm  limit   awRR Low awRR measuring value is above high alarm  limit   CO  Apnea No breath is detected in the set period  High  o Temp  Alarm messages Cause Level    Temp measuring value is above high alarm  Temp High E  limit         Medium      Temp measuring value is b
3.    Technical alarm indicating    lamp    1 Blue        Power indicating lamp    1 Green Orange     Green  When powered with AC  it lights green while turn on    and off the monitor     Orange  When powered with battery  it lights orange only    while turn on the monitor        Battery charging indicating    lamp          1  Orange        A 3 5 Audio indicating       Speaker    Gives audible alarm  QRS tone     Supports Pitch Tone and multi level volume     Alarm tones meet the requirement of IEC 60601 1 8              Alarm pressure    45 dB to 85 dB  Testing place is 1 meter from the tone                 V6 Vital Signs Monitor User s Manual                         A 3 6 Input device  ieem NG  PAN 5  NIBP  record  suspend silence  screen switch and power switch  Knob With  Barcode scanner Optional  Temp module  Button 1  START          A 3 7 Connectors       Center computer connector    RJ 45  10M 100M  TCP IP       Serial port    RS232 serial port       Nurse call    Nurse call connector       Equipotential grounding    point    1       USB connector    Barcode scanner or other USB devices supported by monitor       SD card connector    reserved connector       Wireless network    reserved connector       Temp module connector       Temp probe connector          A 3 8 Signal Output       Nurse call output       Drive mode    Relay       Electric specification     lt 60W   lt 2A   lt 36VDC   lt 25VAC                               Isolated voltage 1500VAC   Signal type N C   N
4.    The monitor can be fitted with rechargeable battery to ensure its continuous work  after the failure of alternating current power supply  and it needs no special  maintenance under the normal condition  While the monitor connecting with  alternating current power  no matter whether the monitor is operating or not  the  battery always can be charged  In the case of sudden power off  the monitor will  automatically get power supply from battery without interruption of monitoring work     Indicative message under the screen will display battery states     HE The icon indicates that the battery is fully charged     HE The icon indicates that the battery is four grids left     HE  The icon indicates that the battery is three grids left     QM The icon indicates that the battery is two grids left     BN  The icon indicates that the battery is one grid left     EE  The icon indicates that the battery is almost depleted and need to be    charged immediately     Warning    e Use only batteries specified in this manual     e Keep the batteries out of children   s reach     Caution    O Remove the batteries prior to shipping or if the monitor is not likely to be    used for an extended period     12 1    Vital Signs Monitor User s Manual       12 2 Installing a Battery    The battery compartment is in the bottom part of the monitor  please refer to the    following steps when installing or charging the batteries     1     2  3   4   5    Turn off power of the monitor  and disconnect 
5.   Manufacture date          de  O    ul  pi       Manufacturer                               SN Serial number  IPX1 Degree of protection against ingress of liquid  Symbol marked on a tag attached to the supply cord of the monitor to  Hospital Only warn that the supply cord should be connected to the sockets which are  Hospital Only to achieve grounding reliability   SpO  Short for    Pulse Oxygen Saturation     NIBP Short for    Non invasive Blood Pressure     CO  Short for    Carbon dioxide     Temp Short for    Temperature     K Symbol for the marking of electrical and electronics devices according to   m Directive 2002 96 EC           2 5       Vital Signs Monitor User s Manual       Chapter 3 Operations    3 1 Unpacking and Checking    1  Unpacking    Before unpacking  examine the packing case carefully for signs of damage  If any    damage is detected  contact the carrier  If the packing case is intact  open the package     2  Remove the monitor and accessories carefully     3  Keep all the packaging materials for future use in transportation or storage     4  Check the monitor and accessories    Check the monitor and its accessories one by one in accordance with the packing    list  Check to see if the parts have any mechanical damages  In case of problems     please contact us or our agent     Warning      Keep the packing materials out of children s reach  Disposal of the  packing materials should observer the applicable waste control  regulations      The monitor mig
6.   Menu   Display Setup     2  Select  Color Setup  choose a desired color for the parameter and waveform     3  Select  Default Setup  set all the parameters and waveforms to default color     4 1 3 Screen Switch    You can set the screen layout as required  The setting method is as follows   1  Select  Menu   Display Setup     2  Select  Screen Switch  choose a desired screens to display  There are  NIBP    Review     Trend Screen  and  Alarm Screen       4 1 4 Standby Mode    The monitor has standby function under the clinic mode  Please set as follows    1  Select  Menu   Display Setup   Screen Save Time      2  Set the screen save time  the monitor will enter standby mode while with no  operations during the setting time     3  Press any button  including the touch screen  to exit the standby mode     Vital Signs Monitor User s Manual       4 2 Screen Layout    The monitor has configured with a color TFT LCD to display parameters and  waves of patient s SpO   NIBP and CO   The following figures are screens of the    monitor     4 2 1 Standard Screen    Parameters displayed under the screen are SpO2  CO2  SpO   and TEMP  Waves  displayed under the screen are SpO2 and CO       12 1  11     wer   2  3  4  10  5  6  9 7  8       1  Alarm Pausing Silence Area    It is to display alarm pausing icon and time or silence icon     2  NIBP Parameter Area    It is to display NIBP parameters  Select    NIBP    to enter  NIBP Setup  menu   3  CO  Parameter Area    It is to display CO  
7.   SpO2 Desat Limit  to a desired value     6 5 6 Setting Wave Color    1  Select   Pleth    to enter  Pleth Setup  menu   2  Set  Wave Color  to a desired color     6 5 7 Masimo SpO  module setup  When using masimo SpO  module  You may need to do the following operations    inthe  SpO  Setup  menu      Setting Sensitivity Mode   You can set sensitivity mode of Masimo SpO   module according to use condition   Select  Sensitivity Mode  in the SpO  Setup menu with the options of  Max        Normal  or  APOD        Max   This mode should be used for the sickest patients  where obtaining a reading    6 5    Vital Signs Monitor User s Manual       is most difficult  The mode is recommended during procedures and when clinician  and patient contact is continuous    Normal    This mode provides the best combination of sensitivity and probe off  detection performance  The mode is recommended for the majority of patients    APOD    This mode is the least sensitive in picking up a reading on patients with  low perfusion but has the best detection for probe off conditions  The mode is useful  for patients that are at particular risk of the sensor becoming detached  pediatric     combative  etc        Setting Fast Sat mode    The Fast Sat mode is clinically applicable during procedures when detecting rapid  changes in oxygen saturation is paramount such as induction  intubation and sleep  studies    Select  Fast Sat Mode  in the SpO  parameter setting menu  you can select  On   or  Off  to
8.   to a desired value   7 8 Setting Venipuncture Press    You can use the NIBP cuff to cause sub diastolic pressure  and block the venous  blood vessel to assist venous puncture    1  Select    NIBP    to enter  NIBP Setup  menu    2  Set  Veni  Press  to a desired value    3  Select  Venipuncture      4  Puncture vein and draw blood sample    5  Select   Venipuncture   again to deflate the cuff  The cuff deflates    automatically after a set time if you do not deflate it   7 9 NIBP Resetting    Select  Reset  in  NIBP Setup  menu to restore the inflation value of blood  pressure pump to the initial value  In case the blood pressure pump doesn t work as    normal but without any prompt  the blood pressure pump can be checked by reset     7 5    Vital Signs Monitor User s Manual       thus the blood pressure pump in abnormal condition due to unexpected reason will    automatically restore   7 10 Air Leakage Testing    Air Leakage Testing is to test the status of air way s air leaking  If no error  information displays on NIBP parameter area  it indicates that the airway is in good  situation and no air leaks exist  However if the airway may have air leaks  the prompt  message appears in NIBP displaying area  NIBP Air Leakage Testing should be done  every two years or when you feel the reading is incorrect    Before testing  please get ready the following things   O One adult cuff  O One inflate hose    One Cylinder Metal Vessel  Procedure of the air leakage test   1  Enter  Pat
9.  9 5    Vital Signs Monitor User s Manual       Chapter 10 Reviewing    Review means reviewing the patients    relevant data that have been saved by the  monitor previously  Review function is available in Clinic Mode only  In Monitor  Mode  you may review the trend data  NIBP measurement data and alarm events of    the monitored patient by use of screen changing key   10 1 Entering the Reviewing Menu    You can enter the reviewing menu by the following two methods   O Select  Menu   History Review  to enter the reviewing window     O Select  Menu   Data Manage  to enter  Data Manage  window     10 2 Reviewing Details  O Through  History Review  window     1  Select  History Review  to enter  History Review  menu    2  Turn the Trim Knob to select a patient in the window  then select  LookUp 1  to view patient s  Detail Record      3  Select  Record  or press the button on the front panel of monitor to print    out the current page     O Through  Data Manage  window     1  View one patient   s ID record   a  Use a barcode scanner to input the patient   s ID number in  Lookup ID  Record   or input the ID number manually    b  Select the button  Lookup ID Record  on the right to view patient s  Detail  Record      2  View record during a time   a  Set  Lookup Jto  Within a week 1   Within a month 1   Within half a year   or TAM      b  Select  Lookup  on the right to enter  History Review  menu     10 1    Vital Signs Monitor User s Manual       Caution      No matter where the
10.  Biolight Meditech Co   Ltd   Address  No 2 Innovation First Road  Technical Innovation Coast  Hi tech Zone  Zhuhai  P R   China    Post code  519085    EC Representative   Shanghai International Trading Corp  GmbH  Eiffestrasse 80   20537 Hamburg    Germany    PN  22 038 0013    
11.  Display screen  4     NIBP  press this button to start or stop NIBP measurement   5  RECORD  press this button to start or stop recording     6  22 SPEND SILENCE     Press this button to pause or reactive the alarms       Press and hold this button for 1 second to silence the system sound     de  mm  Press this button to     Change the screen       Return to the main screen when a menu is opened     8  Trim Knob  The Trim Knob is used for   O Turn left or turn right to move the cursor   O Press down to perform an operation  such as open a menu dialog or select    one option     9  NIBP connector  10  CO  connector    11  SpO  connector    Vital Signs Monitor User s Manual       12  0 0 Power button  O Press this button to turn on the monitor after AC power is connected or the  battery is installed     O Press and hold it for 2 seconds to turn the monitor off     13  Power indicating lamp  It is a LED that lights green and orange  the status of the LED is specified as  follows   O Green  When the AC mains is connected   O Orange  When the AC mains is not connected and monitor is powered by  battery   O Off  When the AC mains is not connected     14  Battery charging indicating lamp    Light up  When the battery is being charged     Off  When the battery is fully charged or no battery in monitor     15  Handle    1 2 2 Side View    Left side         SD card slot    Recorder    Vital Signs Monitor User s Manual       1 2 3 Rear View       1  Temp connector  2  Grounding termina
12.  Meanwhile the  system clanks and enters main screen and the temperature module displays screen    information after a sound of    ba      Warning  e If the monitor is mechanically damaged  or if it is not working properly     do not use it for any monitoring procedure on a patient  Contact your    service personnel     Caution  e The monitor does not have mains switch  The monitor is switched    completely only by unplugging the power cable from the AC power    source     3 2    Vital Signs Monitor User s Manual       3 2 3 Starting Monitoring    1  Decide which parameter should be monitored or measured    2  Install required modules or sensors    3  Check whether the installation of modules or sensors is correct   4  Check whether all kinds of settings are correct     5  Start monitoring on a patient  Detailed information refers to the related chapters   3 3 Shutting off the Monitor    Please follow the below steps to shut off the monitor   Confirm that the patient monitoring is finished   Disconnect the cables and sensors form the monitor     Confirm that the monitoring data is stored or cleared     PO UN A    Press the power switch and hold it for 1s to turn off the monitor     Caution    e If the monitor can t be switched off normally  forced close the monitor by  pressing and holding the power switch for more than 5s  This may cause    some damages to the device     3 4 Operation Modes    1  Select  EJ or pa in the Icons Area after starting the monitor to enter  Mode  S
13.  O   A 3 9 Data Storage   Clinic Mode   Patient quantity 1000   Recording number 16000   Single patient clinic  1 16000           record number                   V6 Vital Signs Monitor User s Manual                      Monitor Mode   Patient quantity 1   Trend data 3 kinds resolution  1Min  5Min  10Min   1Min  can store 96 hours  5 Min  can store 480 hours  10 Min  can store 960 hours   Alarm events 1000   NIBPmeasurement 5000   record                A 4 Measurement Specifications    A41 SpO     e Digital SpO  module                      Measurement technic Digital SpO  technic   Monitoring parameters   SpO  and PR   SpO    Measurement range  0   100      Resolution 1    Accuracy   2    70   100   SpO      0   69    unspecified   PR   Measurement range  257255  bpm   Resolution 1 bpm   Accuracy  1  or   1 bpm  whichever is the greater    SpO  alarm range  0   100     high low limit can be adjusted continuously      PR alarm range  0   300  bpm  high low limit can be adjusted continuously      O Nellcor SpO2 module      Measurement technic Nellcor SpO  technic           Monitoring parameters   SpO  and PR         SpO   Measurement range  0   100            V6 Vital Signs Monitor User s Manual                      Resolution 1   Accuracy   2    70   100   SpO     adult     3    70 100   SpO     neonate     2    70 100   SpO   Clow perfusion    unspecified  0   69    PR  Measurement range  20 300  bpm  Resolution 1 bpm  Accuracy  1  or   1 bpm  whichever is the greater  SpO  al
14.  Sensor or module is not initialized   o Temp  Alarm messages Cause Level  Temp Sensorerrorl The temp sensor is error   Temp Sensorerror2 The temp sensor is error   The measurement value is higher than  Temp OverrangeUp    43  C   The measurement value is lower than  Temp OverrangeDown 34  C  Low  Temp Environment   ee  The ambient temp is higher than 40    OverrangeUp  Temp Environment      The ambient temp is lower than 10    OverrangeDown  Temp Disconnect The temp probe is disconnected   C 3 Prompt Messages  O System    Messages Cause Level    Recording   Recorder is in printing operation    No level       17    Vital Signs Monitor User s Manual                                                                         SpO   Messages Cause Level  Search pulse SpO  module is searching for pulse   E 7   No level  Motion interference Patient movement is too much      NIBP  Messages Cause Level  Software Overpress NIBP is testing Software Over Pressure   No level  Hardware Overpress NIBP is testing Hardware Over Pressure   Manometer NIBP is testing Manometer   Air Leakage Testing NIBP is testing Air Leakage  No level  Venipuncture NIBP is in venipuncture   e CO     Messages Cause Level    Zero in Progress Zeroing is in progress     7 No level    CO  Sensor Warm Up Module is warming up        Vital Signs Monitor User s Manual       Manufactured for   MEDQUIP Inc   11 Parmenter Road Bluffton  SC 29910 USA  Catalog  MQ3600  Model  V6 Vital Signs Monitor    Manufactured by  Guangdong
15.  Unit    mmHg  kPa  0 mmHg to 40 mmHg   2 mmHg  41 mmHg to 70 mmHg   5  of reading   Accuracy  760mmHg  71 mmHg to 100 mmHg   8  of reading    temperature is 25      101 mmHg to 150 mmHg   10  of reading   when RR  gt 80 rpm  all the range is  12  of reading     Gas temperature at 25  C       Total system response    time     lt 3s       Sample flow rate and    accuracy    Rate  50 ml min    Accuracy   7 5ml min    15ml min       Alarm range    0   20 0  OmmHg  152mmHg    high low limit can be adjuste  continuously        Alarm accuracy     0 1  or  1mmHg          O awRR      Measurement range 0 rpm 150 rpm         V6 Vital Signs Monitor User s Manual          Measurement accuracy    0 rpm   70 rpm   1 rpm   71 rpm  120 rpm   2 rpm   121 rpm  150 rpm   3 rpm        Alarm range and accuracy       Range  Orpm 150rpm  high low limit can be adjusted  continuously     Accuracy   1 rpm       A 4 4 Temp    Ambient temperature    Measurement range    10  C  40  C  34 0  C 43 0  C                               Measurement part Ear cavity   Unit C  F   Resolution 0 1C F   Kaanib Between 35 0  C 42 0  C   lt  0 2  C  Beyond 35 0  C 42 0  C   lt  0 3  C   clinical repeat accuracy  lt x0 3C   Measurement time  lt 0 8s   Measurement interval time   12s   Standby time 40s   Memory storage 10       Temp probe alarm    The measured value is beyond the range  Alarm occurs when the  value is lower than 34 0  C or higher than 43 0  C        The ambient temp is beyond the range  Alarm occurs when
16.  aati 6 1  62 Salety Iron ae OMe 3 NANA ANAN PLA NN tad 6 1  6 3 Monitoring Procedure Lanka RANA ANA aaa aadd 6 2  64 SpOs Display eiii idad 6 2  o PA Astute etna ag 6 4  6 6 Masimo Information mac am o Rec at a 6 7  6 7 N  llcor Infomation  nsoni la 6 7  Chapter 7 NIDO cincel 7 1  A ERA NAAN R Cal 7 1  LN A o NA 7 1  7 3 Measurement Limitations ara 7 2  7 4 Measurement Mods AA A A AALAGA NAA 7 2  ida a maan a aaa naa a ada aa 7 2  1 6 NIBP Display did 7 4  NT DA IB o as ale cdas 7 5  LG Setting Venpres Maa AA Uh 7 5  TD NIBP RESCUING ninia adas ARNAN 7 5  TAQ Air Weakage  TESUNP A ee kaaa 7 6  Chapter   CO paa a A A eee 8 1  S1 Dangal li mO ss tse R a ea doe ellen A A ale Rash  8 1  8 2  Monitoring Procedure usar nG Bank 8 1  8 3 CO DISPIAY e dl od ed 8 4  O o o ere 8 4  8 3 LAO iaa 8 5  BG Calibrating A rea 8 6  91 Remoyina BUTAS da 8 6    IV    Vital Signs Monitor User s Manual       Chapter 9 AA 9 1  OA NMG UCU OTN ii illa kak 9 1  9 2 Safety Inform atl OW ni NAGIIBA REE TANAA KHAN 9 2  9 3 Monitorine Procedure e E KAN 9 2  OP Temp Display A akala peta 9 3  OS Settins AA AA PAA 9 4  XO Temp Anna aa AKA ADA ANA 9 4  A erases 9 5   Chapter 10 Revie AAA ANGGE AA AA 10 1  10 1 Entering the Reviewing MENU    UNA NGANGA ANAN 10 1  10 2 REVIEW E Details ka AA NANANA 10 1   Chapter AAA NAIG NGA LINIBEAEDR UE 11 1  Jk NG IR  ea AA AA AA 11 1  11 2 Recordins PP AK ARALAN 11 1  11 3 Starting Stopping Recording GIKAN KARNE 11 1  LEA Settine RECOMEND E E e 11 1  11 5 Installing Record
17.  absorbed by CO  in the gas  At another side of infrared light source  a  photodetector is used to measure the remaining infrared energy and convert it to  electric signal  which will be compared with the energy of infrared light source and  adjusted so as to correctly reflect the CO  concentration in the gas sample    The method to measure carbon dioxide in the patient s airway for this monitor is  microstream  That is takeing a sample of the respiratory gas with a constant sample    flow from the patient   s airway and analyzes it with the CO2 sensor   8 2 Monitoring Procedure    1  Attaching the CO  Module Cable  Plug the cable into the CO  connector on the monitor     2  Attaching the Sampling Tube  Insert the sampling tube into the sampling tube receptacle  Shown as follows        Exhaust pa  Tube     Fal Pa Receptacle  if a IA  Sampling Tube po 2422 Module Cable H  Receptacle NS      i  i       an  Pa    Eo     Ni Sampling Tube    CO2 Module    8 1    Vital Signs Monitor User s Manual       Note    Inserting the sampling tube into the receptacle automatically starts the  sampling pump  Removal of the sampling tube turns the sample pump  off    e To remove the sampling tube from the sampling tube receptacle  press    down on the locking tab and pull the sampling tube from the receptacle     3  If the sampling pump fails to turn on  or runs intermittently  perform a    Zero     procedure    4  Ensure that the CO  module exhaust tube vents gases away from the module  envir
18.  burning  of patients  do not use damaged sensor or extension cable  Do not soak the  sensor into water or make it wet  otherwise it may be damaged    O When disposing the disposable SpO  probe or useless SpO  probe  please  observe all local  state  and federal regulations that relate to the disposal    of this products or similar products     6 3 Monitoring Procedure    1  Depending on the patient category  weight and application site  you can select  the SpO   sensor as required      Clean the application site  such as colored nail polish      Apply the sensor to the patient      Select the extension cable according to the SpO   connector     nA A UU N      Plug the SpO  sensor into the extension cable     6 4 SpO  Display  6 4 1 BLT SpO2    Oo Waveform Display    Pleth 1       1  Pleth waveform    6 2    Vital Signs Monitor User s Manual       2     9o    indicates SpO  unit while       indicates signal intensity  Details are as  the following table                               SpO  value  4  Pleth bar  Pulse Rate  obtained from SpO     Indicator of signal intensity Description   ni The signal strength is low   eae The signal strength is good    ee The signal strength is best    6 4 2 Masimo SpO2    The Masimo module is intended to monitor the SpO2  PR  PI  SpMet  SpHb  PVI   SpCO SpOC of patients  The following figure shows the SpO  display screen  the  display on your monitor may be looked slightly different          SpO  value      SpO  unit      PI  Perfusion Index  
19.  current cursor is in the Data Manage dialog box  the  successful scanning result by using a barcode scanner will fill in the  Lookup ID Record automatically  The ID will be replaced instead of    accumulating when repeated scanning occurs     10 2    Vital Signs Monitor User s Manual       Chapter 11 Recording    11 1 Recorder    This monitor uses the thermal recorder which supports various record type  It can  output the patient information  measurement data  review data and two waveforms at    most   11 2 Recording Type    The records can be divided into the following types according to trigger modes   1  Real time record of manual startup   2  The circular record of automatic startup of the recording meter in line with the  given time interval   3  The alarm record triggered by out of limit parameter and so on     4  Record started by manual operation and related to special function     11 3 Starting Stopping Recording    Through the following methods  you can start and stop recording     Press the button on the front panel of monitor to start real time recording       Press the button again to stop recording     11 4 Setting Recorder    Select to enter  Recorder Setup  menu   11 4 1 Setting Cycle Record   1  Enter  Recorder Setup  menu through the above methods    2  Set  Cycle Record  to  5 Min     10 Min     30 Min    1 H  or  2 H    11 4 2 Setting Cycle Record Time   You can set a certain time interval  and the recorder will automatically start    recording in line wi
20.  enable or disable the Fast Sat mode     O Setting Smart Tone Mode     Smart Tone Mode  is a feature that affects pulse beep and can be selected in  the SpO  parameter setting menu  When you set it to  On    it will allow the audible  pulse beep to beep when the pleth shows signs of motion  The pulse beep is    suppressed during signs of motion when Smart Tone is set to KOff       Setting Waveform Mode  You can select  Waveform Mode Jin the SpO  parameter setting menu  and select     Resp  out Jor  Resp  in   Resp  out is short for Respiration Signal Filtered Out  and  Resp  in is short for Respiration Signal Included     O Setting Alarm Delay   In the SpO2 setting menu  select  Alarm Delay    and you can select the delay  time in the alarms as required in the options    O Setting SpHb Mode   While monitoring Hemoglobin levels  there are two blood sample sources from  which Hemoglobin readings can be obtained  arterial and venous  You can select      SpHb Mode  in the SpO gt  parameter setting menu  and you can select  Arterial  or      Venous       6 6    Vital Signs Monitor User s Manual         Setting SpHb Average Time    You can select the averaging mode for SpHb value  select  SpHb Average Time    in the SpO2 parameter setting menu  and you can select  Short     Medium  or   Long         Setting SpHb Precision    You can select the SpHb precision to be displayed on the screen  select  SpHb    Precision  in the SpO2 parameter setting menu  and select a proper time as req
21.  or data displayed on your product    Conventions   e Bold Italic text is used in this manual to quote the referenced chapter or  sections   e    is used to enclose screen texts     o 1s used to indicate operational procedures     II    Vital Signs Monitor User s Manual       Contents  Chapter 1 General Introduction        oooomommss 1 1  VA Intended Use  al 1 1  LAMA Uttar 1 1  13  Work A ANA AA NALANG NA AN AA papa 1 5  Chapter 2 Sale ty A G INT TA AA 2 1  Zeal SALLY AA A IA 2 1  AAA A A 2 4  2 3 Equipment Symbol S senil diia idad 2 4  Chapter 3 Opera tions GINANGBANG KGAD EGR 3 1  3 1 Unpacking and Check cinc cnica AGA 3 1  SL GEN AA AA 3 2  3 3 Shutting OFF the Monitor miii KANAN 3 3  3 4 Operation Modes o e 3 3  Sus ng NGA NA Ba o e e edo 3 3  3 6 General SEIP sA dat ANG Gia 3 4  3 1 DELAS NAAGAPAN ANA BRIAN UA BEBANG GANG 3 5  3 09 Nurse all hie A BABALA A a AAA A ales 3 5  A AN BA KANAN PANA RA eae  3 7  A AA AA 3 7  3 11 Viewing the Machine Td ios 3 7  Chapter 4 User Interface           escceccossssesssecssecccossccocssoocsooossscccossesoesoscesosecsssscosssssessose 4 1  A a O  a  GA PS PHC a 4 1  4 2 Sereen LAVOUL ini ardid 4 2  4 3  Patient IIA Ad vee 4 7  AA NA O 4 8  4 5 Machine Malena led 4 9  Chapter 3 Adaro 5 1  3  Alarm Caesar gt ae 5 1  SD tl a AA NP 5 2    III    Vital Signs Monitor User s Manual       IA CAES A ANAN NAA GAI GG 5 2  o NTG ANN 5 4  5 5 Parameter Alai 5 5  JO Paust Alarms BAN EE E E EEE E Hi 5 5  SOSEN AA N O 5 6  A AGE 6 1  a AN AA ine ell aR
22.  possible concentration    Do not immerse part of the equipment in the liquid    Do not pour liquid onto the equipment or accessories    Do not allow liquid to enter the case    Never use abrasive materials  such as steel wool or silver polish   or erosive    cleaners  such as acetone or acetone based cleaners      Warning    O Be sure to shut down the system and disconnect all power cables from the    outlets before cleaning the equipment     Caution    e If you spill liquid onto the equipment or accessories  contact your service    personnel or us     13 2 Cleaning of Monitor    1     Common detergent and non corrosive disinfectant used in hospital can be  applied to clean monitor  however you must be aware that many kinds of  detergents must be diluted prior to utilization  and please use it according to    the instruction of detergent manufacturer       Avoid the use of alcohols  amino or acetonyl detergent     3  The enclosure and screen of monitor shall be free of dust  and they can be    wiped with lint free soft cloth or sponge soaked in detergent  While cleaning     13 1    Vital Signs Monitor User s Manual       be careful and do not spill liquid onto the instrument and keep any liquid out  of it  When wiping the side panel of monitor  you must be especially careful to    keep water out of all kinds of cable and outlet on the panel       Do not use abrasive material including wire brush or metal brightener during    cleaning because this material will damage the pan
23.  radiation    Before connecting the monitor to the power line  check that the voltage  and frequency ratings of the power line are the same as those indicated on  the monitor   s label or in this manual    Always install or carry the monitor properly to avoid damage caused by  drop  impact  strong vibration or other mechanical force    When THERMAL CUT OUT operates  the monitor will function  continually without any SAFETY HAZARD by using its rechargeable  battery  However  the battery is not being charged anymore  While the  battery is low the monitor will give both visible and audible warnings in 5  minutes    The American mains plug whose ground is in the same plug with its other  two cords is Hospital Only  The supply cord should be connected to the    sockets which are Hospital Only to achieve grounding reliability     2 1 3 Note    Note    Put the monitor in a location where you can easily see the screen and  access the operating controls    Keep this manual in the vicinity of the monitor so that it can be obtained  conveniently when needed    The software was developed in compliance with IEC 60601 1 4  The  possibility of hazards arising from software errors is minimized     This manual describes all features and options  Your monitor may not    2 3    Vital Signs Monitor User s Manual       have all of them    O The service life of this monitor is 5 years  At the end of its service life  the  monitor  as well as its accessories  must be disposed of in compliance with 
24.  range exceeds module  Range Exceeded oh Ny   specification        Excessive Motion    1 Too many retries due to interference of  motion artifact    2 Signal is too noisy during measurement   e g  patient has severe tremor     3  Irregular pulse rate  e g  arrhythmia        Overpressure Sensed    Cuff pressure exceeds the specified high  safety limit  Could be due to rapid    squeezing or bumping of cuff        Signal Saturated    Large motion artifact that saturates the BP    amplifier   s amplitude handing capability        Pneumatic Leak    Module reports Air Leakage failure while    in the Pneumatic Test mode        System Failure    Module occurs abnormal processor event                 Time Out Measurement took more than 120 seconds  in adult  90 seconds in neonate mode   Cuff Type Err Neonate cuff is used in adult mode           Low             Vital Signs Monitor User s Manual                                                                                           e CO   Alarm messages Cause Level  CO  sensor is off patient or off the  CO  sensor off    monitor   nG   Low  i CO  airway adapter is disconnected with  Check airway adapter  CO  sensor   CO  measurement over CO  measurement is over range and need  range verify zero   Zero Required Sensor or module is not initialized   Check CO  Sampling Line Sampling tube is occluded or damaged   Sampling tube is kinked or pinched  Low  Exhaust tube is blocked   CO  sensor error CO  sensor is error   Sensor no initialized
25.  the  ambient temp is lower than 10 0  C or higher than 40 0  C           Fever alarm  Alarm occurs when the measured value is higher than  37 5  and lower than 43 0  C        V6 Vital Signs Monitor User s Manual       Appendix B Factory Defaults    B 1 Patient messages                                                                              Patient messages   Factory Defaults    Type   Adult  B 2 Alarm    Alarm setup   Factory defaults    ALM Volume   2    Alarm paused time   2min  B 3 Interface Setup  Interface setup Factory defaults  Brightness 3  B 4 SpO   B 4 1 General Setups  SpO  setup Adult Pediatric Neonatal  Alarm switch On  Alarm Level Medium  Alarm Print Off  Average Time 8s  High alarm limit of SpO  100  100  95   Low alarm limit of SpO  90  90  85   Desat Limit 85   Pleth  Wave Speed 25mm s  Color Yellow                   V6 Vital Signs Monitor User s Manual       B 4 2 Special Setups  Masimo                                                           SpO  setup Factory defaults  Sensitivity Mode Normal  Fast Sat Mode off  Smart Tone Mode off  Waveform Mode Resp out  Alarm Delay off  SpHb Mode Arterial  SpHb Average Time Long  SpHb Precision 0 1  SpHb Unit g dL  B 5 NIBP  NIBP setup Adult Pediatric Neonatal  Alarm switch On  Alarm Level Medium  Alarm Print Off  High alarm limit of Sys 160 mmHg 120 mmHg 90 mmHg       Low alarm limit of Sys    90 mmHg 70 mmHg 40 mmHg       High alarm limit of Map    110 mmHg 90 mmHg 70 mmHg       Low alarm limit of Map    60 mmH
26.  the guidelines regulating the disposal of such products  If you have  questions concerning disposal of products  please contact manufacturer    or its representatives     2 2 Safe Operation Conditions          Methods of sterilization or Sterilization  not applicable   disinfection recommended by the Disinfection  Refer to Maintenance and Cleaning chapter  manufacturer   Electromagnetic interference No mobile telephone nearby       Electrosurgical interference damage   No damage          Diathermy instruments influence Displayed values and prints may be disturbed or erroneous       during diathermy       2 3 Equipment Symbols       Symbol Symbol Note       Type CF applied part  defibrillation protected     ot The unit displaying this symbol contains an F Type isolated  floating     applied part providing a high degree of protection against shock  and is    defibrillator proof        Type BF applied part  defibrillation protected    TA  The unit displaying this symbol contains an F Type isolated  floating     applied part providing a high degree of protection against shock  and is    defibrillator proof        A Attention  Consult accompanying documents  this manual                 Non ionizing radiation                   2 4       Vital Signs Monitor User s Manual                   Symbol Symbol Note  h Dangerous voltage  y Equipotential grounding  100V 240V    Alternating current input range  100V 240V       USB socket       Network connector       Nurse call connector     
27.  to dissipate before zeroing      Do not attempt to zero the module while the adapter or cannula is in the    patient   s airway     8 5    Vital Signs Monitor User s Manual       O Do not attempt zeroing if the temperature is not stable    O Zeroing with CO  in the adapter or cannula can lead to inaccurate  measurements or other error conditions  If you attempt zeroing while  CO  remains in the adapter or cannula  the time required to zero the    module may be increased     8 6 Calibrating    The monitor has been calibrated before leaving factory  You can use the monitor  directly except the following three situations  When one of the following situations  happens  please calibrate the CO  module             The CO  module has been used for half a year or a whole year           The therapeutist doubt about the reading of measurement           tThe atmospheric pressure or height above sea level has changed a lot after    last calibrating     Caution    O Please calibrate the monitor under the direction of the technician  authorized by manufacturer  Incorrect calibrating process should lead    incorrect reading     8 7 Removing Exhaust Gases    Warning    O When using the microstream CO  measurement on patients who are  receiving or have recently received anesthetics  connect the outlet to a  scavenging system  or to the anesthesia machine ventilator  to avoid    exposing medical staff to anesthetics     Use an exhaust tube to remove the sample gas to a scavenging system  Atta
28. 2000     A physician must determine the clinical significance of the NIBP measurement     7 2 Safety Information    Warning    O Check the patient category before monitoring  Incorrect settings may  result in some risk for patient safety  Higher adult setting is not suitable  for pediatric and neonatal patients      Do not measure NIBP on patients with sickle cell disease or any condition  where skin damage has occurred or is expected    e Use clinical judgement to decide whether to perform frequent Auto BP  measurements on patients with severe blood clotting disorders because of  the risk of hematoma in the limb fitted with the cuff      Use clinical judgement to decide whether to perform Auto BP  measurement on the patients of thrombasthemia    O Do not use the NIBP cuff on a limb with an intravenous infusion or  arterial catheter in place  This could cause tissue damage around the  catheter when the infusion is slowed or blocked during cuff inflation    O If you doubt the NIBP measurements  check the patient   s vital signs by    other device  and then check the monitor     Vital Signs Monitor User s Manual       7 3 Measurement Limitations    NIBP measurements are impossible with heart rate extremes of less than 40 bpm or  greater than 240 bpm  or if the patient is on a heart lung machine     The measurement may be inaccurate or impossible        with excessive and continuous patient movement such as shivering or    convulsions        if a regular arterial pressure pulse 
29. Flo   Accessories PN  LoFlo CO  sensor 16 100 0016  Airway adapter  adult  15 100 0045  CO  nasal cannula  adult  15 100 0044  CO   O  nasal cannule  adult  15 100 0046  CO  nasal cannule  pediatric  15 100 0048  CO  nasal cannule  infant  15 100 0049  CO  nasal cannule  infant  15 031 0010  14 4 Temp    Accessories PN    Earcap  Disposable  Probe Cover   038       14 5 Barcode scanner       Accessories Model    Barcode scanner FG2100             14 3    V6 Vital Signs Monitor User s Manual       Appendix A Product Specifications    A 1 Safety Specifications    A 1 1 Classification       Classification of Protection against    electric shock       Degree of protection against electric  shock    SpO    NIBP  Temp  CF  CO  BF       Degree of protection against hazards    of explosion    Not suitable       Degree of protection against ingress    of liquid    IPX1       Mode of operation          Continuous          Note     I  Class I  internally and externally powered equipment     CF  Type CF applied part  BF  Type BF applied part    Not suitable  Equipment is not suitable for use in the presence of flammable    anesthetic mixture with air or with oxygen or nitrous oxide     A 1 2 Environmental Specifications                   Operating  5 40      Temperature  Transportation and Storage   20    55    C   he Operating 15  85   noncondensing    Humidity  Transportation and Storage 10  93   noncondensing    t Operating  700 1060  hPa     Atmospheric Pressure   Transportation and 
30. Q  MEDGUIP    V6 Vital Signs Monitor  MQ3600    User   s Manual    Product Information    Manufactured for   MEDQUIP Inc   11 Parmenter Road Bluffton  SC 29910 USA  Catalog  MQ3600  Model  V6 Vital Signs Monitor    OEM Manufacture  Guangdong Biolight Meditech Co   Ltd   Add No  2 Innovation First Road  Technical Innovation Coast    Hi Tech Zone  Zhuhai  P R  China 519085    Vital Signs Monitor User s Manual       Revision History    This manual has a revision number  This revision number changes whenever the  manual is updated due to software or technical specification change  Contents of this  manual are subject to change without prior notice      Document No   J V6 0 118 001  O Revision number  V1 0    Release time  2013 12  Copyright   2013 Guangdong Biolight Meditech Co   Ltd  All rights reserved     CE mark    C     6123    EC Representative Name   Shanghai International Trading Corp  GmbH  Hamburg   EC Representative Address     Eiffestrasse 80  20537 Hamburg Germany    Statement    Manufacturer holds the copyright of this manual  and we are also entitled to deal  with this manual as confidential files  This manual is only used for operation   maintenance and service of product  someone else can not publish the manual    This manual contains exclusive information protected by copyright laws and we  reserve its copyright  Without written approval of manufacturer no parts of this  manual shall be photocopied  Xeroxed or translated into other languages    The contents cont
31. S 10  BAPOR o 10  O 11  BO COS CLGEIO  ri dad 12  BATES AS 12  Appendix C Alarm Messages PA ABA ceneeevancenatonceusacansscnetanecaeaeess 13  CA Physiological alarm Mess a  A tt 13  C 2 Technical alarm Messages nA a 14  Co Prompt  Messages idas SEESE 17    VI    Vital Signs Monitor User s Manual       Chapter 1 General Introduction    1 1 Intended Use    The monitor is intended to be used for monitoring  displaying  reviewing  storing  and alarming of multiple physiological parameters of patients  including Pulse  Oxygen Saturation  SpO2   Pulse Rate  PR   Non invasive Blood Pressure  NIBP    Carbon dioxide  CO2  and Temperature  Temp      The monitor is intended to be used in outpatient departments and emergency  treatment rooms of hospitals  community clinics  private clinics and other medical  institutions  It is not intended for helicopter transport  hospital ambulance or home    use     1 2 Main Unit    1 2 1 Front View       Vital Signs Monitor User s Manual       1  Physiological alarm indicating lamp  When a physiological alarm occurs  this lamp will light up as defined below     High level alarm  the lamp quickly flashes red     Medium level alarm  the lamp slowly flashes yellow       Low level alarm  the lamp lights yellow without flashing     2  Technical alarm indicating lamp  When a technical alarm occurs  this lamp will light up as defined below     Medium level alarm  the lamp slowly flashes blue     O Low level alarm  the lamp lights blue without flashing   3 
32. Select  Modify   Yes  to finish setting     3 4    Vital Signs Monitor User s Manual       3 6 4 Adjusting the Volume      Alarm Volume    Under the Monitor Mode   1  Select  Menu   Sound Setup       2  Select  Alarm Volume  choose a desired value       Beep Volume    1  Select  Menu   Sound Setup       2  Select  Beep Volume  choose a desired value   3 7 Default Setups    Caution      In case of power failure  after restart  the system will restore the setup    before power failure automatically     It is possible that you change some setup during operation  but these changes are  not always appropriate or correct  Therefore  you may restore some setup to factory  default setup during actual operation to ensure that various setup of the monitor is  applicable to the monitored patient    1  Select  Menu   System Setup  to open  System Setup  menu   2  Select  Default Config  to enter  Remind  menu   3  Select  Yes  to load the user config and alarm limit shown in the Remind    menu  Select  No  to exit the Remind menu   3 8 Nurse Call    Nurse Call is a function that the monitor will send signal to call nurse when the  alarm conditions destined occur    The monitor has a nurse call output connector  connect the connector to the nurse  call system of the hospital by the nurse call cable  the nurse call function can be  realized    The nurse call function is valid when the following conditions are concurrent              The nurse call function is open     3 5    Vital Signs Mo
33. Setup  menu   O Select    Pleth    to enter   Pleth Setup  menu     6 5 2 Setting Beep Volume    1  Select   SpO gt     to enter  SpO  Setup  menu   2  Set  Beep Volume  to  Off  or  1 5       6 5 3 Setting Scan Speed    O SpO  Scan Speed    1  Select   SpO gt     to enter  SpO  Setup  menu   2  Set  Scan Speed  to  6 25mm s     12 5mm s     25 mm s  or  50 mm s       O Pleth Scan Speed    6 4    Vital Signs Monitor User s Manual       1  Select    Pleth    to enter  Pleth Setup  menu   2  Set  Scan Speed  to  6 25mm s     12 5mm s     25 mm s  or  50 mm s       6 5 4 Setting Average Time    The SpO  reading shown on the monitor is the average of data collected within a  specific time  The shorter the averaging time is  the quicker the monitor responds to  the change in the patient s oxygen saturation level  Contrarily  the longer the  averaging time is  the slower the monitor responds to the change in the patient s  oxygen saturation level  but the measurement accuracy will be improved  When a  critical patient is monitored  selecting shorter averaging time will help understanding  the patient s state    Select    SpO     to enter  SpO  Setup  menu  Set  Average Time  to a desired    value   6 5 5 Setting Desat Limit    Desat Limit means that when the measuring value is lower than the limit  high  alarm will be triggered  Set the Desat Limit as follows   1  Select  Menu       System Setup       Machine Mainte      enter the password  to enter  Machine Mainte  menu     2  Set
34. Storage  500   1060  hPa    V6 Vital Signs Monitor User s Manual       A 1 3 Power Specifications    Input voltage AC  100 240  V   50 60  Hz          Input power TOVA  Fuse T1 6AL 250V  2  5x20mm       A 2 Physical Specifications               Part   Weight  kg  Size  WxHxD   mm       Mainframe About 2 5  Including a lithium battery   lt 160x130x260      Temp module About 0 16  Including the connecting wire   153x40 5x60         A 3 Hardware Specifications                   A 3 1 Display  Type Color TFT LCD  Size  diagonal  7 inch  Resolution 234x480 pixels  Anti glare screen With  LCD switch With       Temp module                            Type black and white LCD  Size 1 2 inch  A 3 2 Recorder    Type Thermal dot array      Paper width 50 mm    Recording width 40 mm  Recording speed 25 mm s  Recording waveform double tracks          V6 Vital Signs Monitor User s Manual                            A 3 3 Battery  Type Rechargeable lithium ion battery  Size 105mmx78mmx20mm  Weight s360 g  Quantity 1  Rated voltage 11 1 VDC  Capability 4000 mAh  8 hours    Operating time    Using a new and fully charged battery at 25  C ambient  temperature  connecting SpO  sensor and NIBP work on    AUTO mode for 15 minutes interval        Charge time    6h to 100   Standby                    Turn off delay 5 min  15 min after the low battery alarm first occurs   Indicator of battery i  me With  capability  A 3 4 Mainframe LED       Physiological alarm    indicating lamp    1  Yellow Red     
35. Trim  Knob to select  Step    then the options of  1Min    5Min  and  10 Min     will pop up  select one you need     4 3 Patient Information    Select    Patient Information    icon to enter  Patient Info   menu  In the menu you    can view patient s ID  Name  Type  Gender and Age   4 3 1 Continuous the same patient    When the patient to be monitored is the one displayed in  Patient Info  menu   please select  Return    The monitor will continuous the monitoring of the same    patient and add current data to the previous one     4 7    Vital Signs Monitor User s Manual       4 3 2 Admitting a new patient    When the patient to be monitored is a new one  please select  New Patient  to  enter  New Patient  menu  Input new patient s ID  Name  Type  Gender and Age  At  last  select  OK  to keep the setting of new patient     4 3 3 Patient ID    e Clinic Mode   In  Patient Info   menu  when selecting  New Patient    Patient ID  will be  created according to  ID Name  automatically  You can also input  Patient ID   through barcode scanner or input manually  Through  Patient Info  menu  you can  change the patient info  After amending the patient info  selectingl Return lto save the  changing   O Monitor Mode   In  Patient Info  menu  when selecting  New Patient    you need to input    Patient ID  through barcode scanner or input it manually  Through  Patient Info     menu  you can change the patient info  After amending the patient info  selecting      Return  to save the changin
36. age down button to browse more  trendgraph information  The left button is page up button while the right one is page  down button  As shown in the above figure    3  Time scale   The scale is to display time  You can browse the selected parameter   s trend graph  during the time between the left time and the right time on the time scale  The date    and time displayed under the time scale is the date and time the cursor now indicates     4 6    Vital Signs Monitor User s Manual       4  Scale  It is to scale the value of parameters   5  Parameter displaying area  It is to display values of all the monitored parameters at the time where the  cursor indicates now   6  The selected parameter  It is to display the name of the selected parameter and its value at the time where  the cursor is   7  Setting interval  Interval is the time between the left time and the right time on the time scale  It  can be 1  2  3  4 or 5 hours  Turn the Trim Knob to select  Setting interval    The  sign of         is to add the time value while           is to decrease it  Each time you press  the two buttons  the time will add 1 hour or decrease 1 hour   8  Cursor  It is to indicate the selected parameter s trend graph at the time where it is  Turn  the Trim Knob to select  Tcursor  button  the color of the button will turn into  green like the cursor  Then turn Trim Knob to move the cursor where you need   9  Setting step  Step is the time the cursor move on the time scale at one time  Turn the 
37. ained in this manual are subject to amendments without    notification     Vital Signs Monitor User s Manual       Manufacturer s Responsibility    Only under the following circumstances will manufacturer be responsible for the   safety  reliability and performance of the instrument      All the installation  expansion  readjustment  renovation or repairs are conducted  by the personnel certified by manufacturer    O The storage condition  operation condition and electrical status of the instrument  conform to the product specification     O The instrument is used in accordance with the user s manual     About this manual    This manual contains the instructions necessary to operate the product safely and  in accordance with its function and intended use  Observance of this manual is a  prerequisite for proper product performance and correct operation and ensures patient  and operator safety    This manual is based on the maximum configuration and therefore some contents  may not apply to your product  If you have any question  please contact us    This manual is an integral part of the product  It should always be kept close to the  equipment so that it can be obtained conveniently when needed    The manual is geared for clinical professionals who are expected to have a working  knowledge of medical procedures  practiced and terminology as required for  monitoring patients    All illustrations in this manual serve as examples only  They may not necessarily  reflect the setup
38. arm   area      The technical alarm message will still be shown in the technical alarm area    After the alarm paused time  or when a new medium level technical alarm occurs  during the alarm pausing  the monitor will automatically cancel the alarm pausing     Press again the button 244   the alarm pausing can be cancelled by manual operation       Setting Alarm Pausing time    5 5    Vital Signs Monitor User s Manual       Set the silence time only under Monitor Mode   1  Select  Menu   System Setup   Machine Mainte  enter the required  password   2  Set  Alarm suspend Time  to  1Min     2Min  or  3Min       5 7 Silence    Press and hold the button 24 on the front panel of monitor for 1 second to set the  system silent  That is  all the sound of system is shut off  And the icon Xx will be  displayed in the upper right corner of the screen  When in the silence status  the alarm  indicators are valid except audible alarm  Press the button 4  again to exit the silence    status  A new alarm will cancel the silence automatically     5 6    Vital Signs Monitor User s Manual       Chapter 6 SpO     6 1 Introduction    The measurement of oxygen saturation of arterial blood  also known as pulse  oxygen saturation  usually shortened as SpO2  adopts the principles of light spectra  and volume tracing  The LED emits lights with two specific wavelengths  which are  selectively absorbed by oxygenated hemoglobin and deoxyhemoglobin  The optical  receptor measures the changes in the light 
39. arm range  0   100    high low limit can be adjusted continuously  PR alarm range  0   300  bpm  high low limit can be adjusted continuously         Masimo SpO  module          Measurement range 0    100   Resolution 1   70    100    2   Adult Pediatric  non motion conditions    70    100    3   Neonate  non motion conditions    Accuracy    70  to 100    3   motion conditions    0  69   undefined       Average time  PR    Measurement range    2s 4s  4s 6s  8s  10s  12s  14s  16s       25 bpm 240 bpm       Accuracy    Resolution  SpCO    Measurement range     3 bpm  non motion conditions       5 bpm  motion conditions      1 bpm         0  to 100        Accuracy    SpMet    Measurement range    0  to 40    3   non motion conditions       gt 40   unspecified         0  to 100        Accuracy    PI    Measurement range    0  to 15    1   non motion conditions       gt 15   unspecified            0 05  to 20           SpHb    V6 Vital Signs Monitor User s Manual          Measurement range    Accuracy    0 g dl to 25 g dl       8 g dl to 17 g dl   1   non motion conditions       lt 8 g dl or  gt 17 g dl  unspecified         SpOC          Measurement range    A 4 2 NIBP    Measurement way       0 ml dl to 35 ml dl  Automatic oscillometry                                          Sys  30   270  mmHg    Adult Dia  10   220  mmHg     Map  20 235  mmHg   Sys  30 235  mmHg   MGA Pediatric   Dia  10 220  mmHg  Map  20225  mmHg    Sys  30   135  mmHg     Neonatal   Dia  10   110  mmHg   Ma
40. at the patient is in a life threatening situation and an  emergency treatment is necessary  This is the highest level alarm    O Medium level alarms  Indicate that the patient   s vital signs appear abnormal and  an immediate treatment is required    e Low level alarm  Indicate that the patient   s vital signs appear abnormal and an  immediate treatment may be required    The levels of some physiological alarms are predefined before the monitor leaves  the factory and can not be changed by users  While some levels of physiological  alarms can be changed by users    The monitor   s technical alarms are classified into two categories  medium level  and low level    The levels of technical alarms are predefined before the monitor leaves the factory    and can not be changed by users   5 3 Alarm Indicators    When an alarm occurs  the monitor will indicate it through the following means      Alarm tone  According to alarm level  speaker in the monitor gives alarm sound  in different tone      Alarm lamp  According to alarm level  alarm lamp on monitor flashes in different  color and speed      Alarm message  Alarm messages are displayed on the screen       Flashing numeric  The numeric of parameter in alarm flashes     Caution    O The concrete presentation of each alarm prompt is related to the alarm    level     5 2    Vital Signs Monitor User s Manual       5 3 1 Alarm Tone    The different level alarms are indicated by the system in following different audio                  
41. breathing circuit  In reverse  always remove the  airway adapter from the breathing circuit before removing the sensor    e Always disconnect the camnula  airway adapter or sampling tube from  the CO  sensor when not in use    e Do not insert the things other than sampling tube into receptacle of  sampling tube    e Only use the CO  sampling tubes specified by our company    e The sampling tubes are disposable  Please keep the sampling tube clean   and prevent the tube from clogging by dust  It is advised to replace the  sampling tube every 12h  up to 120h of use with filter tip   the sampling  tube leaks or has been damaged and contaminated     8 3    Vital Signs Monitor User s Manual       8 3 CO  Display    e Waveform Display         CO  waveform     Unit of CO      Inspired minimum CO   FiCO      Airway respiration rate  awRR     End tidal CO2 value  EtCO2     R A UU N e    8 4 Setting CO     8 4 1 Entering CO  Menu    O Select    CO     to enter   CO  Setup  menu   O Select    CO  Wave    to enter  CO  Wave Setup  menu     8 4 2 Setting Apnea Alarm    1  Select    CO     to enter  CO  Setup  menu   2  Set  Apnea Alarm  to  10 Sec    15 Sec    20Sec    30Sec    45 Sec  J or  60 Sec       8 4 3 Setting Unit    1  Select    CO     to enter  CO  Setup  menu   2  Set  Unit  to  mmHg     KPa  and  96       8 4 4 Setting Scan Speed      CO  Scan Speed    8 4    Vital Signs Monitor User s Manual       1  Select    CO     to enter   CO  Setup  menu   2  Set Scan Speed   to  6 25 
42. ch it to    the outlet connector of microstream CO  module     8 6    Vital Signs Monitor User s Manual       Chapter 9 Temp    9 1 Introduction    9 1 1 Measuring principle    The emission infrared wave length is Sum 14um while the human body  temperature is in 35  C 44   C  The science proves that human s eardrum approaches  the brain hypothalamus  and the hypothalamus is the human body temperature control  center  thus the ear temperature can reflect the body core temperature more accurately  than other parts of human  Obviously the human ear cavity temperature and the body  temperature have the relatively stable corresponding relations  According to this  principle  we can monitor the human ear cavity   s infrared launch through the infrared  temperature measurement probe  After signal transforming and microprocessor  processing  the relatively accurate human body temperature will be outputted    This temperature module has the ambient temperature automatic compensation  function  It can detect the ambient temperature change and revise its reading to reduce  the influence of difference temperature to the monitoring and avoid the shortcoming  that it is unable to monitor correctly when used in the big different temperature  environment  This temperature module uses the ear cavity mode to monitor human    temperature rapidly and obtain the temperature information     9 1 2 Introduction of Temp probe    The structure of the Temp probe is as follows     earcap    Measuring bu
43. connect the monitor from the patient and stop all monitoring and    measuring procedures     12 2    Vital Signs Monitor User s Manual       2  Place the monitor in the charger stand and connect the AC mains  Allow the  battery to be charged uninterruptedly for above 6 hours    3  Disconnect AC mains and allow the monitor to run on the battery until it  shuts off     4  The operating time of a battery reflects its performance directly   Caution    The operating time of a battery depends on the configuration and  operation of the monitor  NIBP measurement  SpO2 measurement and    using of recorder will deplete the battery faster than other parameters       measurement     12 5 Disposing Batteries    Batteries that are damaged or depleted should be replaced and discarded properly     Dispose of used batteries according to local regulations     Caution   O The service life of battery depends on the service time and frequency   This monitor battery can be charged and discharged for 300 times  generally     Warning    O Do not disassemble batteries  dispose of them in fire  or cause them to    short circuit  They may ignite  explode  or leak  causing personal injury     12 3    Vital Signs Monitor User s Manual       Chapter 13 Maintenance and Cleaning    13 1 Introduction    Keep your equipment and accessories free of dust and dirt  To avoid damage to the    equipment  follow these rules     1     O a AR NG    Always dilute according the manufacturer s instructions or use lowest 
44. d for use only by clinical professionals or under  their guidance  It must only be used by persons who have received  adequate training in its use  Anyone unauthorized or untrained must not  perform any operations on it    O The monitor can only be applied to one patient at one time    O Before using  please check the connecting cables and accessories are in  correct working order and operating condition    O Please connect the monitor to a socket with protective earth  If the socket    does not have protective earth conductor  please do not use the socket and    Vital Signs Monitor User s Manual       use battery to provide power to the monitor    O To avoid explosion hazard  do not use the monitor in the presence of  flammable anesthetics  vapors or liquids    O Do not open the monitor housings  electric shock hazard may exist  All  servicing and future upgrades must be carried out by the personnel  trained and authorized by manufacturer only    O When using the monitor with electrosurgical units  ESU   make sure the  patient is safe    e If you spill liquid onto the equipment or accessories  particularly if there  is a chance that it can get inside the tubing or the measurement device   contact your service personnel  Please insure the measurement device and  its accessories dry during using electrosurgery and defibrillator    O Do not come into contact with the patient during defibrillation  Otherwise  serious injury or death could result    O Do not rely exclusively on t
45. d value is displayed on the module screen  while it is    transmitted simultaneously to the main unit to display     Caution    O After each monitoring  the temperature module will locks for 12 seconds   Please do not carry out the next monitoring until the temperature unit  sign glints       If there is no operation in 40 seconds  the temperature module will enter    standby mode automatically and turn off the screen     9 4 Temp Display      Unit    Temp value       9 3    Vital Signs Monitor User s Manual       9 5 Setting Temp    9 5 1 Setting Unit    1  Select    TEMP    to enter the temp setup menu     2  Set  Unit  toL Jor      The new unit will become effective at the next    monitoring     9 5 2 Setting Alarm    Select  menu   Alarm Setup  to enter the alarm setup menu  Set the alarm    level and alarm limit  The setting value will become effective at the next monitoring     9 6 Temp Alarm    1  The main unit temperature alarm     If the measured value is lower than 34 0   an alarm for that the temperature is  lower than the low limit occurs    If the measured value is higher than 43 0   the alarm for that the temperature  1s higher than the high limit occurs    If the ambient temperature is lower than 10 0   the alarm for that the ambient  temperature is lower than the low limit occurs    If the ambient temperature is higher than 40 0   the alarm for that the ambient  temperature is higher than the high alarm occurs    If the measured temperature value is higher tha
46. e  to enter ID Name menu     2  Define the ID Name rule according to your usage     3 10 Patient ID    You can use a barcode scanner to input the Patient ID  please go through the  following steps before input the patient ID   1  Open the barcode scanner switch in factory configuration   2  Connect the barcode scanner to the monitor     B shown in the bottom of the  display screen indicates the barcode scanner is supported by monitor  Otherwise        g displayed indicates the barcode scanner can   t be used     Caution    O In case of interfere with other USB device  including USB keyboard   please use the barcode scanner provided or designated by the    manufacturer       Please check whether the switch of scanner is open before start the  barcode scanner      Only in the patient ID input interface can the barcode scanner work  The  monitor will not process any input operation when using the scanner in    other input interface     3 11 Viewing the Machine Info    1  Select  Menu   Machine Info       2  View the detail information about this monitor     3 7    Vital Signs Monitor User s Manual       Chapter 4 User Interface    4 1 Display Style    Display style of user interface can be set according to your need     Including        Screen brightness        Display color of wave and parameter         Sweep mode of wave   4 1 1 Adjusting the Screen Brightness    1  Select  Menu   Display Setup     2  Select  Back Light  set its value     4 1 2 Selecting the Color    1  Select
47. el and monitor screen     5  Do not submerge the monitor in liquid       While cable or plug of attachment accidentally gets wet  please rinse 1t with    distilled water or deionized water and dry it in the environment of temperature    40 to80 for at least one hour     13 3 Cleaning and Sterilizing of Accessories    13 3 1 SpO2 Sensor    The recommended disinfector include isopropyl alcohol 70   10  decolourant    solution can be used for sterilization at lower standard  Don t use undiluted    decolourant  5   5 25  sodium hypochlorite  or other non recommended    disinfector in order to avoid damage to sensor     Attention     Do not sterilize sensor by ray  steam or epoxy ethane    Do not directly submerge sensor in liquid    To avoid long time harm to sensor  it is suggested that sterilization to  theproduct be conducted only when necessary according to the regulation of    your hospital     13 3 2 NIBP Cuff      Please regularly clean the product       Remove cuff from connector and pull out airbag from sheath     3  Submerge clean and soft medical gauze pad or other soft cleaning tools into    fresh water or neutral soapy water  and wring out surplus water from the    submerged gauze then wipe airbag and pipe       Wash the cuff sheath in the clean neutral soapy water     5  After the sheath and airbag intensive drying  enclose airbag with cuff sheath    and put into operation     13 2    Vital Signs Monitor User s Manual       Attention     Excessive or frequent cleani
48. elect  menu    2  Select  Clinic Mode  or  Monitor Mode  to enter  Continue  menu    3  If the patient you are going to monitor is the one displayed in  Continue   menu  please select  Continue      4  If the patient is a new one  please select  New Patient       5  Input the new patient s information in  New Patient  menu     3 5 Using Menu    Turn the Trim Knob left or right to select  Menu  to open the following main    menu  You can finish most operations and sets through  Menu    The main menu    3 3    Vital Signs Monitor User s Manual       under Monitor Mode is different from that under Clinic Mode  The following figure is    main menu under Monitor Mode             Style of other menus is the same as that of the main menu  parts of which are as           follows    1  Menu title  Name of the menu    2  Main display area  Area to display options  buttons or prompt messages      M     means you can enter its submenu     3   Return    Press this button to exit the menu     3 6 General Setup    3 6 1 Changing the Language    1  Select  Menu   System Setup     2  Select  Language  choose a desired language     3  Select  Return  exit the current menu     3 6 2 Setting the Screen Saver    Under the Clinic Mode   1  Select  Menu   Display Setup       2  Select  Screen Save Time  choose a desired setup     3 6 3 Setting the Date and Time    1  Select  Menu   System Setup    IDatetime Setup     2  Set  Year     Month     Day    Hour     Min  and  Sec  to a desired value   3  
49. elow low alarm  Temp Low PERA  limit   C 2 Technical alarm Messages    System  Alarm messages Cause Level  Battery Failure Battery failure or no battery  Low  Battery Low Voltage of battery is too low  Medium  SD Write Protected SD memory card is write protect  Low  SD Unknowpart The inserted SD card is unrecognized  Low  SD Write Error SD card is miswriting   SD No Space SD card has no space  Medium  Flash No Space Interior flash has no space   No paper in the recorder when recording  Recorder Error or the recorder door is open or recorder is   Low             absent                 Vital Signs Monitor User s Manual       O SpO        Alarm messages Cause Level       SpO  sensor may be disconnected from the a  SpO2 Sensor off     Medium  patient or the monitor                                                                                                             SpO  sensor error SpO  sensor failure  Low  SpO  signal weak SpO  signal is weak   O Masimo SpO  module   Alarm messages Cause Level  SpO2 No cable No cable connected  SpO2 Replace cable Cable life expired    Cable is defective  SpO2 Incompatible cable Cable is incompatible  SpO2 Unrecognized cable Cable is unrecognized  SpO2 No sensor No sensor connected  SpO2 Replace sensor Sensor life expired    Sensor is defective  SpO2 Invalid sensor Sensor is incompatible    Sensor is unrecognized  SpO2 No tape No tape  SpO2 Replace tape Tape life expired   Tape is defective  SpO2 Invalid tape Tape is incompatible    Tape i
50. ent symbols and blue background color for the alarm messages to match the    alarm level as follows        Alarm level symbol background color       Medium pa        oa a blue         Prompt messages  Prompt messages are displayed in technical alarm area  Prompt messages have no    color and visual and audible alarm indication   5 4 Setting Alarm Volume    You need to set alarm volume only under monitor mode     1  Select  Menu   Sound Setup      5 4    Vital Signs Monitor User s Manual       2  Select  Alarm Volume  choose a desired value     5 5 Parameter Alarm    You need to set  Alarm Setup  only under Monitor Mode   5 5 1 Turn on off the Alarm    Take SpO2 for example   1  Select  Menu   Alarm Setup     2  Set  Class  to any option but  Off  to turn on the alarm  while set  Class    to  Off  to turn off the alarm     5 5 2 Setting Alarm Level    Take SpO  for example   1  Select  Menu   Alarm Setup     2  Set  Class  to  Medium  or  High       5 5 3 Setting Alarm Limit    Take SpO  for example   1  Select  Menu   Alarm Setup     2  Set  High Lim  and  Low Lim  to a desired value     5 6 Pausing Alarms    Press the button 48 on the front panel of monitor  you can suspend all alarm  indicators of the monitor      The visual alarm and audible alarm are all suspended      The parameters of physiological alarm stop flashing      The alarm message in the physiological alarm area will not be displayed    O The remaining time and the icon  4 will be shown in the physiological al
51. eps     1     Use the pily stick or soft cloth bedewed with water or alcohol to wipe the lens gently  Do  not pot the probe in the water or other liquid   Dry the lens for over 30 min before cover the new earcap  and avoid solarizing directly    while the temperature is beyond 10  C 40     13 3    Vital Signs Monitor User s Manual       Attention    e Do not sterilize sensor by ray  steam or epoxy ethane    e Do not directly submerge sensor in liquid   To avoid long time harm to sensor  it is suggested that sterilization to  theproduct be conducted only when necessary according to the regulation  of your hospital      Please change the earcap after each use to ensure the veracity and avoid    cross infecting     Note      The electric schematic and element list can only be offered to the eligible    service center or personnel     13 4    Vital Signs Monitor User s Manual       Chapter 14 Accessories    Warning    e Use only accessories specified in this manual  Using other accessories may  cause damage to the monitor    e Disposable accessories are designed for single patient use only  Reuse of  them may cause a risk of contamination and affect the measurement  accuracy     e Check the accessories and their packages for any sign of damage  Do not    use them if any damage is detected                                                        14 1 SpO     SpO  Sensor  Nellcor SpO   Type Model Patient category  MAX A Adult finger  patient size gt 30kg   MAX P Pediatric foot hand  
52. g     Caution    O No matter where the current cursor is in the New Patient dialog box  the  successful scanning result by using a barcode scanner will fill in the  Patient ID automatically  The ID will be replaced instead of accumulating  when repeated scanning occurs    O The history data will be deleted while admitting a new patient under  Monitor Mode     4 4 Demo    4 4 1 Turn on Demo    1  Select  Menu   System Setup  to enter the system setup menu     2  Set  Demo  to  On    a password entering window will pop up     4 8    Vital Signs Monitor User s Manual       3  Input the desired password and then select  OK  to enter the demo mode   4 4 2 Turn off Demo    1  Select  Menu   System Setup  to enter the system setup menu   2  Set  Demo  to  Off       4 5 Machine Maintenance    Select  Menu   System Setup   Machine Mainte  input the required    passwords to enter  Machine Mainte  menu       NIBP Test  Select  NIBP Test  to test NIBP      NIBP PR  Select  On  or  Off       Touch Adjust  Select  Touch Adjust  to adjust touch screen     Nurse Call  Select  Nurse Call  to set nurse call      Desat limit  Set SpO2 desat value       Alarm suspend Time  Set the value of Alarm suspend Time     TI Tm 4 A U Ne      Factory Mainte  Used for factory maintenance     4 9    Vital Signs Monitor User s Manual       Chapter 5 Alarm    Alarm refers to a prompt that is given by the monitor for medical personnel  through visual  audible and other means when a vital sign appears abnormal o
53. g 50 mmHg 25 mmHg       High alarm limit of Dia    90 mmHg 70 mmHg 60 mmHg       Low alarm limit of Dia    50 mmHg 40 mmHg 20 mmHg                      Measure Mode Manual  Unit mmHg  Interval 15 min  Color red       Venipuncture Press    60 mmHg   40 mmHg   30 mmHg       Inflation          170 mmHg   130 mmHg   100 mmHg             V6 Vital Signs Monitor User s Manual                                                                                        B 6 CO   LoFlo   CO  setup Adult Pediatric Neonatal  Alarm switch On  Alarm Level Medium  Alarm Print Off  Limit Display Off  Unit mmHg  High alarm limit of EtCO  50 mmHg 50 mmHg 45 mmHg  Low alarm limit of EtCO  20  minke 20 mma 30rmmakle  High alarm limit of FiCO  4 mmHg 4 mmHg 4 mmHg  High alarm limit of awRR 30 rpm 30 rpm 100 rpm  Low alarm limit of awRR 8 rpm 8 rpm 30 rpm  Scale 61 mmHg  Wave Speed 12 5 mm s  Color White  Apnea alarm time 20s  B 7 Temp  Temp setup Factory defaults  Alarm switch On  Alarm Level Medium  Unit    CC  High alarm limit of Temp 39  C  Low alarm limit of Temp 36  C                Vital Signs Monitor User s Manual       Appendix C Alarm Messages    C 1 Physiological alarm Messages    The third line in the cable is    Alarm level    of factory default  the ones with            mean that the level can be changed by users                                                                             O SpO   Alarm messages Cause Level  SpO gt  High SpO  measuring value is above the high  alarm limit f  
54. gns Monitor User s Manual       The screen is called    Trend Screen    under Monitor Mode while called    Clinic  Review    under Clinic Mode  Under the screen you can view the measurement of each  parameter      Storage Setup   1  Select  Menu   Storage Setup  to enter  Storage Setup  menu   2  Set  Trend StorInter  to a desired value between  1 Min     1 Min  and   10 Min       4 2 4 Alarm Screen    Akm Screen    tae   tino  O TO      4113   Aecorder Error                              Under the screen  you can view each alarm event including its data  time and the    parameters leading the alarm     4 5    Vital Signs Monitor User s Manual       4 2 5 Trend Graph    Trend  reph  Sws Die Map    00 lic   ce  5  Parameter displaying area    owkR    6  The selected parameter    4  Scale 7  Setting interval    3  Time scale 8  Cursor    2310432115 17 22    2  Page up 2  Page down  1  Selecting parameter aan a 9  Setting step       You can enter TrendGraph screen through the screen changing button as shown in  chapter 1 section 1 2 main unit  Under the screen  you can do the following  operations    1  Selecting the parameter  Turn the Trim Knob to select  Param    Press the Trim Knob   Et Fi      awRR    NIBP    Sp02    PR  and  TEMP  will pop up  Turn and  press the Trim Knob to select the parameter you need to view its trend graph  And the  selected parameter will turn to be the parameter you selected just now    2  Turning pages   You can turn pages through the page up and p
55. he audible alarm system for patient  monitoring  Adjustment of alarm volume to a low level or off may result  in hazard to the patient  Remember that alarm settings should be  customized according to different patient situations and always keeping  the patient under close surveillance is the most reliable way for safe  patient monitoring    O The physiological data and alarm messages displayed on the monitor are  for reference only and cannot be directly used for diagnostic  interpretation    O To avoid inadvertent disconnection  route all cables in a way to prevent a  stumbling hazard  Wrap and secure excess cabling to avoid risk of    entanglement or strangulation by patient or personnel     2 1 2 Caution    Caution    O To ensure patient safety  use only parts and accessories specified in this  manual     O At the end of its service life  the monitor  as well as its accessories  must    2 2    Vital Signs Monitor User s Manual       be disposed of in compliance with the guidelines regulating the disposal of  such products  If you have any questions concerning disposal of the  monitor  please contact us    Magnetic and electrical fields are capable of interfering with the proper  performance of the monitor  For this reason make sure that all external  devices operated in the vicinity of the monitor comply with the relevant  EMC requirements  Mobile phone  X ray equipment or MRI devices are  a possible source of interference as they may emit higher levels of  electromagnetic
56. he part of air pipe plug  of cuff close to socket with fingers before pulling it out     5  Tie the cuff to the upper arm or thigh of the patient    Ensure the mark         on the cuff shall lie above artery while the air pipe shall be  under the cuff  ensuring the air pipe outside the cuff does not knot and the white line  on the cuff shall be within the range      lt   gt       otherwise the cuff shall be replaced    The monitor is applicable for standard neonatal cuff  pediatric cuff and adult cuff      Including arm cuff and thigh cuff      Note      While measuring blood pressure  the patient must keep calm without  any talk    O The cuff tied on the limb shall be on the same level as the patient s heart  so as to avoid the reading error resulting from the hydrostatics effect of  the blood flow between the heart and cuff  If the cuff position is higher  than heart level  the BP reading will be lower  the measured result shall be  added 0 75mmHg  0 1kPa  for each centimeter higher  in case the cuff  position is lower than heart level  the BP reading will be higher  the  measured result shall be deducted 0 75mmHg  0 1kPa  for each    centimeter lower     7 3    Vital Signs Monitor User s Manual       7 5 2 Starting Stopping Measuring    Press the button      on the front panel of monitor to start NIBP measuring while    press the button   s again to stop NIBP measuring   7 5 3 Auto Measurement    1  Select    NIBP    to enter  NIBP Setup  menu    2  Set  Interval  to any o
57. ht be contaminated during storage and transport   Before use  please verify whether the packages  especially the package of  disposable accessories  are intact  In case of any damage  do not apply it  to the patient      Please ensure the monitor is working under specified conditions   otherwise  the technical specifications mentioned in this manual will not  be met  thus possibly leading to damage of equipment and other    unexpected results     Caution      Please put the monitor onto a horizontal and stable supporting plane   Avoid putting the monitor in the locations where it easily shakes or  wobbles  Enough space shall be left around the monitor so as to guarantee    normal ventilation     Vital Signs Monitor User s Manual       3 2 Getting Started    3 2 1 Inspecting the Monitor    1  Before you start to make measurements  carry out the following checks on the  monitor including all connected modules        Check for any mechanical damage             Check for any incorrect connection of all the external cables and accessories   2  Plug the power cord into the AC power source  If you are using battery power   ensure that the battery has sufficient power for monitoring  When you use a battery  for the first time  you must charge it  following the instructions given in Battery    chapter   3 2 2 Starting the Monitor    Press the power switch  the technical alarm lamp will light up in blue  and then the  physiology alarm lamp lights up in yellow and red  at last goes out 
58. ient Info  menu  set  Type  to  Adult     2  Connect the cuff securely with the socket for NIBP air hole     3  Connect the cuff to a suitable Cylinder Metal Vessel  shown as follows     Metal        Cylinder       Monitor       Cuff    4  Enter  NIBP Setup  menu  select  Leakage      After 20 seconds or so  the system will automatically open the deflating valve   which marks the completion of an air leakage test    6  If no error information displays on NIBP parameter area  it indicates that the  airway is in good situation and no air leaks exist  However if the prompt  Air  Leak  appears in the place  it indicates that the airway may have air leaks  In this  case  the user should check for loose connection  After confirming secure  connections  the user should re perform the air leakage test  If the failure prompt    still appears  please contact the manufacturer for repair   If the leakage still exists after all the above operations  please contact us     7 6    Vital Signs Monitor User s Manual       Chapter 8 CO     8 1 Introduction    The monitor adopts infrared absorption technology to measure the carbon dioxide   CO   concentration in the breathing airway of patient  Because CO  molecule can  absorb infrared light of special wavelength  and the amount of absorbed infrared light  directly relates to the concentration of CO  therefore while the infrared light radiated  from the infrared light source passing through the gas sample containing CO    part of  energy will be
59. ing Paper  aaa aNG NGI 11 2  M6 Elan Jam Papas ass suid Sel AA APA a i 11 3  11 7 Cleaning Recordere eining cds 11 3   Chapter 12 Battery ii NANANG UNAN RANGE 12 1  12 1  MAE TUG Ca ma A AN Taka aanak 12 1  12 211stalling a Baterias 12 2  12 3 Optimizing Battery Performance      oooonocccnnococooncccononnconncnononanonnnonononccnnnncnnno 12 2  12 4 Checking Battery Performance iss si ssccissecacsansaveasisesenccaspascdssecsencsavonadeussentens 12 2  12 5 Disposing Batteries cistitis KAU KIA Li Gama 12 3   Chapter 13 Maintenance and Cleaning       oooocooocsosssc 13 1  VES WMHS EH Lo y RON AER E Eu o BANA BABABA NLA 13 1  132 Cleaning OF Mont a ips 13 1  13 3 Cleaning and Sterilizing of Accessories   ococccnoccnocaconncnoncnonnnonn non nono ncnnnnnnnnno 13 2   Chapter 14 Accessories         eesessesseseesossesossossesceosoesossossesossossescosoesessosoesossossesosssssessoss 14 1    Vital Signs Monitor User s Manual       APO AKA BANG o 14 1  TA NIBE a aaa mah MB AA AA AA e eae 14 2  14 3 CO   LOFI0  aaa NAN NAGANA dd 14 3  NA AA ee a ae 14 3  14 3 Barcode  Scanner a di as 14 3  Appendix A Product Specifications      ooooomo 1  a AA ALTA A NG AG KAO 1  Ai PHYSIC al Spa ima SNANG TANAKA 2  A 3 Hardware Specifications  bu aNG sons e raii D E sE 2  AA Measurement Specifications ssc  dics ascicissseccsensacessdigesadtashascevsbedeeaeeaveszeeesacensentecses 5  Appendix B Factory Delas 24ANG 10  Bil Patient messages aa NAN a NAA paalaala 10  AA PA a Be ee A ee 10  a a apace sa A A E 
60. intensity after the light passes the capillary  network and estimates the ratio of oxygenated hemoglobin and the total hemoglobin     Oxygenated hemoglobin    Spo h  A Q   i Oxyhemoglobin   deoxyhemoglobin 5100     Wavelengths of the light emitted by the pulse oximeter probe are nominally  660nm for red LED and 940nm for infrared LED     6 2 Safety Information    Warning    Use only SpO2 sensors specified in this manual  Follow the SpO2 sensor s   instructions for use and adhere to all warnings and cautions      When a trend toward patient deoxygenation is indicated  blood samples  should be analyzed by a laboratory co oximeter to completely understand  the patients conditions    O Do not use the monitor and the SpO  sensor during magnetic resonance  imaging  MRI   Induced current could cause burns    O Prolonged continuous monitoring may increase the risk of unexpected  changes in skin characteristics  such as irritation  reddening  blistering or  burns  Inspect the sensor site every two hours and move the sensor if the  skin quality changes  For neonates  or patients with poor peripheral blood  circulation or sensitive skin  inspect the sensor site more frequently    O Check the SpO  sensor and its package for any sign of damage before use     Do not use the sensor if any damage is detected     Vital Signs Monitor User s Manual         Before use  the operator must ensure the compatibilityies of the monitor   SpO  sensor and extension cables  otherwise  this may lead to the
61. is hard to detect        with cardiac arrhythmias        with rapid blood pressure changes        with severe shock or hypothermia that reduces blood flow to the peripheries        on an edematous extremity     7 4 Measurement Mode    There are three modes of measuring NIBP   O Manual  measurement on demand   e Auto  continually repeated measurements in the set interval     STAT  rapid series of measurements over a five minutes period  then the    monitor returns to the previous mode  Use only on supervised patients     7 5 Monitoring Procedure    7 5 1 Preparing to Measure NIBP    1  Check the patient category  if you want to change the patient category  Select   Ci to enter  Patient Info  menu  And select the patient category as required    2  Select the appropriate cuff according to patient category    O Check the limb circumference of patient      Select the appropriate cuff  The applicable limb circumference for cuff is  marked on the cuff   The width of the cuff should be about 40  of the limb  circumference  50  for neonate  or 2 3 of the upper arm   s length  The  inflatable part of the cuff should be long enough to encircle 50  80  of  the limb     1 2    Vital Signs Monitor User s Manual       Note      The accuracy of measurement of BP depends on the suitability of the  cuff     3  Confirm the cuff has been entirely deflated    4  Plug the air pipe plug of cuff into the connector  NIBP  of monitor until the  plug and socket contact well   Attention  you shall nip t
62. l  3  AC power input connector  Connect to USB device  such as keyboard and mouse   4  Serial port  5  Wired network connector  Standard RJ45 socket  it is used for connection with the central monitoring system  provided by manufacturer   6  USB socket  7  Nurse call connector  8  Fan and horn orifice    Note            The wired network connector is used for connection with the central    monitoring system provided by manufacturer        Vital Signs Monitor User s Manual       1 2 4 Bottom View    Battery compartment       Fig 1 4    1 3 Work mode  The monitor has two working modes  Clinic Mode and Monitor Mode     1 3 1 Clinic Mode    Clinic Mode refers to that the monitor is used for monitoring of a number of  patients one by one by outpatient doctor  Each patient has his her own ID number  and  the monitoring results of patients are saved according to their ID numbers  When a  patient with the same ID number appears again  the device will automatically find out  the previous monitoring data  and will add current data to the previous one    Clinic mode may be used in one of the following occasions    1  Medical personnel are present  e g  monitoring of patients by outpatient doctor   or house call doctor    2  A number of measurements of many patients  e g  monitoring of patients by   physical checkup doctor    3  Rounds of wards  e g  doctor or nurse can monitor relevant physiological    status of patient by the monitor when checking in ward   1 3 2 Monitor Mode    Monito
63. mm s    12 5 mm s    25 mm s  or  50 mm s      e CO  Wave Scan Speed    1  Select    CO  Wave    to enter  CO  Wave Setup  menu   2  Set Scan Speed 1 to  6 25 mm s     112 5 mm s     25 mm s  or 50 mm s       8 4 5 Setting Wave Color    1  Select    CO  Wave    to enter  CO  Wave Setup  menu     2  Set  Wave Color  to a desired value   8 5 Zeroing    Zeroing allows the CO  module to adjust to the optical characteristics  in order to  obtain accurate readings  While zeroing is recommended the first time a CO  module  is connected to the monitor  it is only absolutely necessary when the message    Zero  Required    is displayed    Follow these steps    1  Ensure that the nasal cannula or airway adapter is not connected to the patient  or close to any source of CO   including the patient s  your own  exhaled  breath and ventilator exhaust valves     2  Select  Zero  in the CO  parameter setting menu  this will start zeroing  The  monitor zeroes the module and displays the message    Zero In Progress    for  about 15 20s on the screen  The message disappears upon completion of the  zeroing     Caution      Always ensure that the sampling tube is properly connected to the  microstream CO  module before zeroing    O Always ensure that the mainstream CO  sensor is properly connected to  the airway adapter before zeroing    O Do not attempt zeroing for 20s after removing the adapter or cannula  from the patient   s airway  This time allows any CO  remaining in the  adapter or cannula
64. n the high limit or is lower than    the low limit  the alarm for that the temperature is high or low occurs     2  The temperature module alarm     If the measured value is lower than 34 0   the LCD will display    Lo    and the  buzzer sounds    B B B B B       If the measured value is higher than 43 0   the LCD will display    Hi     and  the buzzer sounds    B B B B B       If the ambient temperature is lower than 10 0   the LCD will display    AL     and the buzzer sounds    B B B B B       If the ambient temperature is higher than 40 0   the LCD will display    AH     and the buzzer sounds    B B B B B       The temperature module has a fever alarm function  While the temperature is    higher than 37 5   the buzzer sounds    B B B        9 4    Vital Signs Monitor User s Manual       9 7 Calibrating    The monitor has already been calibrated before leaving factory  User can directly  apply it to measuring in normal conditions  to the exclusion of the below conditions   For temp module  please carry out temperature deviation calibration  when the    following conditions happened   More details refer to Service Manual              The temperature module has been used for half a year or one year       The accuracy of Temp reading has been doubted by clinical physician   Caution      User may only calibrate the device under the instruction of the technical  personnel authorized by manufacturer  Moreover  incorrect calibrating    procedure may result in incorrect reading    
65. ng  otherwise   the recording meter may be damaged   e Unless for paper replacement or fault remedy  do not keep the recorder    door open     11 6 Clearing Jam Paper    While the sound of recorder operation or printing of recording meter is abnormal   please first check whether there is paper jam in the recording meter   If so  please clear it as per following steps   1  Open the recorder door   2  Pull out the recording paper  and cut off the wrinkle part     3  Load recording paper once again and close the recording meter door   11 7 Cleaning Recorder    After long time service  some paper scrap and impurity will accumulate on the  printing head  and affect printing quality as well as the service life of printing head  and roll shaft  The recorder can be cleaned according to the following methods    1  Before cleaning  the measures such as wearing anti static wrist strap shall be  adopted to avoid the damage to recording meter resulting from static    2  Open the recorder door and pull out recording paper    Use a tampon with some alcohol to sweep slightly the surface of  thermo sensitive parts of printing head    4  After the alcohol entirely vaporizes  load recording paper once again and    close the recorder   s door   Caution  e Don t use any article that can damage the thermo sensitive parts of    recorder during cleaning     e Don t heavily press the printing head of recorder     11 3    Vital Signs Monitor User s Manual       Chapter 12 Battery    12 1 Introduction 
66. ng may damage airbag  so don t clean airbag  unless necessary    Do not dry airbag and sheath in high temperature    If higher sterilization level is required  please choose disposal cuff    One disposal cuff can only be used for one patient    Carefully keep water and cleaning solution out of the connecting parts ofcuff    and monitor     13 3 3 CO2 Sensor and Reusable Airway Adapter    The outside of the module or sensor may be cleaned and disinfected by wiping  with 70  isopropyl alcohol  a 10  bleach solution  or mild soap  After  cleaning  wipe with a clean  water dampened cloth to rinse  Dry before use   Reusable airway adapters may be cleaned by rinsing in a warm soapy solution   followed by soaking in a liquid disinfectant such as 70  isopropyl alcohol  a  10  bleach solution  Cidex  or System 1   refer to the disinfectant  manufacturer s instructions for use   Adapters should then be rinsed with  sterile water and dried    Reusable airway adapters may also be pasteurized or autoclaved  Autoclave at  121  250   for 20 minutes  unwrapped    Before reusing the adapter  ensure the windows are dry and residue free  and  that the adapter has not been damaged during handling or by the cleaning    process     13 3 4 Temp probe    The recommended disinfector is isopropyl alcohol 70   Don   t use undiluted    decolourant  5   5 25  sodium hypochlorite  or other non recommended    disinfector in order to avoid damage to sensor     Please clean the probe lens as the following st
67. nitor User s Manual          An alarm condition destined occurs          The monitor is not in the state of alarm paused or system silence     Warning      The nurse call function should not be used as the primary patient alarm  inform source  It is necessary for combining the auditory and visual  alarm signal and the patient clinical feature and symptom as the primary  information to medical and nursing staff about the physiological  condition of the patient     3 8 1 Turning on off Nurse Call    Select  Menu   System Setup   Machine Mainte  enter the required  password    Select  Factory Mainte  enter the required password    Select  Nurse Call    setitto  on      3 8 2 Setting Nursecall Type    Select  Menu   System Setup   Machine Mainte  enter the required  password    Select  Nurse Call  to enter the nurse call menu    Set  Nursecall Type  to  Normally Open  or  Normally Closed       3 8 3 Setting Call Time    1     Select  Menu   System Setup   Machine Mainte  enter the required  password    Select  Nursecall Setup   to enter nurse call setup menu    Set  Call Time  to  1 Sec  or  Continuous       3 8 4 Triggering Nurse Call    Select  Menu   System Setup  to enter system setup menu    Select  Nursecall Setup   to enter nurse call setup menu    Set  PhyAlarm Trigger  and  TecAlarm Trigger  to  Off    Low      Med  or  High       3 6    Vital Signs Monitor User s Manual       3 9 ID Name    You need to set ID name under the Clinic Mode   1   Menu   System Setup   ID Nam
68. onment    5  Wait for the CO  module to warm up  The monitor will display the    Sensor    Warm Up    message for approximately Iminute while the module warms up  to operating temperature  The message disappears when the module is ready  for use     6  Apply Microstream airway adapter or cannula     1  For intubated patients requiring an airway adapter  Install the airway adapter  at the proximal end of the circuit between the elbow and the ventilator Y    section  Shown as follows        2  For intubated patients with an integrated airway adapter in the breathing  circuit  Connect the male connector on the straight sample line to the female    port on the airway adapter  Shown as follows      gt      o   O    8 2    Vital Signs Monitor User s Manual       3  For non intubated patients  Place the nasal cannula onto the patient  Shown as    follows   um     ba Wa    4  For patients prone to mouth breathing use an oral nasal cannula  Trim the oral  sampling tip if necessary to fit the patient  It should extend down past the  teeth and be positioned in the mouth opening  Remove the cannula from the    patient if the tip needs to be trimmed  Shown as follows     FI    5  For nasal or oral nasal cannulas with oxygen delivery  place the cannula on  the patient as shown then attach the oxygen supply tubing to the oxygen    delivery system and set the prescribed oxygen flow     Caution    e Always connect the airway adapter to the sensor before inserting the  airway adapter into the 
69. p  20 125  mmHg   Cuff pressure range  0   300  mmHg       Resolution    1 mmHg       Pressure accurac y    Static   3 mmHg    Clinic  Average error   5 mmHg  standard deviation   lt 8 mmHg             Unit mmHg  kPa  PR range  40 240  bpm  PR Resolution lbpm       Cuff auto deflation    The cuff will deflate automatically when power is off or time of  measurement is beyond 120 seconds  90 seconds for neonate  or the  cuff pressure is beyond the overpressure protection set by software    and hardware        Measurement time    Normally  it is 20s to 45s  depending on HR and moving    interference typically        Overpressure protection  Adult  Pediatric    Neonatal        29743  mmHg   25243  mmHg   14743  mmHg             V6 Vital Signs Monitor User s Manual          Alarm range              0 300  mmHg  high low limit can be adjusted   PE continuously   ba  0 300  mmHg  high low limit can be adjusted  continuously   Map  0 300  mmHg  high low limit can be adjusted  continuously             A 4 3 CO   LoFlo     O Microstream CO  module    Measurement way    Infrared spectrum         Measurement mode       Warm up time    microstream  Capnogram displayed in less than 15 s  at an ambient temperature o    25  C  full specifications within 2 minutes        Measurement range     0 19 7     0 150 mmHg                 Resolution 0 1  or 1mmHg  Short term drift   0 8mmHg over four hours   Stability Long term drift  Accuracy specification will be maintained over a 12  hour period   
70. parameters  Select    CO     to enter  CO  Setup  menu   4  CO  Wave Area    It is to display CO  wave  Select    CO  Wave    to enter  CO  Wave Setup   menu     4 2    Vital Signs Monitor User s Manual       5  TEMP Parameter Area   It is to display temperature parameter  Select    TEMP    to enter  TEMP Setup   menu   6  SpO2 Wave Area   It is to display SpO2 wave  Select     SpO2 Wave    to enter  SpO2 Wave Setup 1  menu   7  Time and Battery Status Displaying Area   It is to display system time and battery status   8  Prompt Messages Area   It is to display the menu   s meaning which the cursor is on or prompt messages   9  Icons Area   From left to right  the icons are  Menu     Patient Information     Recorder  Setup     Network Setup  and  SD Card     10  SpO  Parameter Area   It is to display SpO  parameters  Select    SpO     to enter  SpO  Setup  menu   11  Physiological Alarm Area   It shows the physiological alarm messages  When multiple alarms of different  levels occur at the same time  the monitor will give visual and audible alarm  indications according to the alarm level   12  Technical Alarm Area  It shows technical alarm messages  When multiple alarms of different levels occur   at the same time  the monitor will give visual and audible alarm indications according    to the alarm level     4 3    Vital Signs Monitor User s Manual       4 2 2 NIBP Review       Under the screen  you can view the measurement of NIBP     4 2 3 Trend Screen       4 4    Vital Si
71. patient size 10 50kg   Disposable MAX I Infant foot hand  patient size 3 20kg   UARN Aui finger or neonatal foot hand  patient  size  gt 40 kg or  lt 3 kg   DS 100A Adult  Reusable OXI A N Adult   neonatal  OXI P I Pediatric   infant  BLT SpO   Type Patient category PN  Adult 15 100 0013  Reusable Pediatric 15 100 0014  Neonatal 15 100 0015                14 1       Vital Signs Monitor User s Manual         SpO  Extension cable       Accessories    PN          Extension cable    15 031 0016          Masimo SpO  Sensor                            Type Model   PN Patient category  DCI   2501 Adult finger  Reusable  DCIP   2502 Pediatric finger  Disposable Neo   2514 Infant foot hand  O Masimo SpO2 Extension cable  Accessories   Model   PN          Rainbow RC 12 20pin Patient cable      Rainbow RC 12   2404       14 2 NIBP      Disposable cuffs                                           Model Patient category Limb circumference  cm    Bladder width  cm   M1866A 3 1 5 7 2 5  M1868A 4 3 8 0 3 2  Neonatal  M1870A 5 8 10 9 4 3  M1872A 7 1 13 1 5 1       O Reusable cuffs                               Patient category Limb circumference  cm  Bladder width  cm  PN   Large adult 15 100 0021  Adult 25 35 14 4 15 100 0019  Small adult 20 28 11 15 100 0023  Pediatric 13 20 8 15 100 0022  Infant 10 18 5 15 100 0024  Adult  Thigh  15 100 0020  Neonatal 6 11 15 100 0025                14 2             Vital Signs Monitor User s Manual                                                 14 3 CO   Lo
72. ption but  Manual      3  Start the Auto measurement manually for the first time  and then enter the  Auto mode  The monitor will start the measurement continually repeated in  the set interval after the first measurement  If you press the button     during  the auto measurement  the measuring will be paused  You should press the    button  gt  again to start the auto measurement   Warning      Prolonged NIBP measurements in Auto mode are associated with  purport  ischemia and neuropathy in the limb wearing the cuff  When  monitoring a patient  examine the extremities of the limb frequently for  normal color  warmth and sensitivity  If any abnormality is observed     stop the NIBP measurements     7 5 4 STAT Measurement    1  Select  NIBP    to enter  NIBP Setup  menu   2  Select  STAT  to start the STAT Measurement     7 6 NIBP Display    There is no waveform displayed for NIBP measurement  the NIBP readings are  displayed in the parameter area  The following figure shows the NIBP display screen     the display on your monitor may be looked slightly different     7 4    Vital Signs Monitor User s Manual            Pressure unit     Systolic blood pressure     Mean arterial blood pressure     Pulse Rate  obtained from NIBP       Diastolic blood pressure    nA A U N e    7 7 Setting NIBP   Select    NIBP    to enter  NIBP Setup  menu   7 7 1 Setting Unit   In  NIBP Setup  menu  set  Unit  to  mmHg  or  kPa     7 7 2 Setting Initial Press   In  NIBP Setup  menu  set  Init Press
73. r    technical problem occurs     Note      The monitor generates all the audible and visual alarms through speaker   alarm lamp and screen  When the monitor powers on  the alarm lamp  will be lighted one time and the speaker will give a beep voice  which    indicates the alarm system of the monitor is working order     5 1 Alarm Category    According to character of alarm  the monitor   s alarms can be classified into three  categories  physiological alarms  technical alarms and prompt messages   e Physiological alarms   Physiological alarms are triggered by a monitored parameter value that violates   set alarm limits or an abnormal patient condition  Physiological alarm messages are  displayed in the physiological alarm area     Technical alarms   Technical alarms are triggered by a device malfunction due to improper operation  or system problems  The problems may result in system abnormal operation or  irresponsible monitoring parameters  Technical alarm messages are displayed in the  technical alarm area   O Prompt messages   As a matter of fact  prompt messages are not alarm messages  Apart from the  physiological and technical alarm messages  the monitor will show some messages to    indicate the system status     Vital Signs Monitor User s Manual       5 2 Alarm Level    According to severity of alarm  the monitor s physiological alarms are classified  into three categories  high level alarms  medium level alarms and low level alarms      High level alarms  Indicate th
74. r Mode refers to that the monitor is used to monitor the same patient in  ward for long time  In this mode  the small sized monitor is used to monitor the  patient s SpO2  NIBP and CO      Monitor mode may be used in one of the following occasions     Vital Signs Monitor User s Manual       1  Monitor of a single patient for a long time  e g  hospitalized patient   2  Ward without on duty medical personnel     3  Remote monitoring  connect to central unit through network      1 3 3 Difference Between two Modes                                              Function Mode Monitor Clinic  Silence With With  Alarm Pause With Without  Technical Alarm With With  Physiological Alarm With Without  Recording Review With With  Alarm Review With With  Parameter Storage With With  Alarm Messages Storage With Without  Nurse Call With Without  Communicating with Center   With With  Computer   Multi patient Without With  Standby Mode Without With                Vital Signs Monitor User s Manual       Chapter 2 Safety    2 1 Safety Information  Warning      Indicates a potential hazard or unsafe practice that  if not avoided  will    result in death or serious injury     Caution      Indicates a potential hazard or unsafe practice that  if not avoided  could    result in minor personal injury or product property damage     Note    O Provides application tips or other useful information to ensure that you    get the most from your product     2 1 1 Warning    Warning    O The monitor is intende
75. s unrecognized  SpO2 Sensor Calibrating Sensor is calibrating    SpO2 Sensor off Sensor may be disconnected from patient  SpO2 Pulse search Searching pulse  SpO2 Interference Interference detected  SpO2 Low perfusion Low perfusion index  SpO2 Demo Mode The monitor is at demo mode  SpO2 Check sensor Check sensor is connecting  SpO2 Low SIQ SpO2 Signal IQ is low  SpO2 Low PR SIQ PR Signal IQ is low  SpO2 Low PI SIQ PI confidence is low  SpO2 Low SpCO SIQ SpCO Signal IQ is low  SpO2 Low SpMet SIQ SpMet Signal IQ is low  SpO2 Low SpHb SIQ SpHb Signal IQ is low  SpO2 Low SpOC SIQ SpOC Signal IQ is low          Vital Signs Monitor User s Manual                SpO2 Low PVI SIQ PVI Signal IQ is low  SpO2 Board Failure SpO2 board is failure  SpO2 Failure SpO2 module is failure       SpO2 Communication Error    SpO2 communication is error                            SpO2 Enter Programming SpO2 is entering programming mode  Mode  SpO2 Weak signal SpO2 signal is weak  SpO2 PR Over 239bpm SpO2 PR is over 239bpm    NIBP  Alarm messages Cause Level       Transducer or other hardware failure           Self test Failed   1  Cuff is completely unwrapped   Loose Cuff 2  The cuff is not connected    3  Adult cuff used in neonate mode   Air Leak Air leak in pneumatics  hose  or cuff        Air Pressure Error    Unable to maintain stable cuff pressure     e g  kinked hose        Very weak patient signal due to a loosely       Weak Signal wrapped cuff  The pulse of patient is too  weak   Measurement
76. th the given time interval     11 1    Vital Signs Monitor User s Manual       1  Enter  Recorder Setup  menu through the above methods   2  Set  Cycle Record Time  to  5 Sec     10 Sec  or  30 Sec       11 4 3 Setting Alarm Record    1  Enter  Recorder Setup  menu through the above methods   2  Set  Alarm Record  to  On  or  Off       11 4 4 Setting Alarm Record Time    1  Enter  Recorder Setup  menu through the above methods   2  Set  Alarm Record Time  to  5 Sec     10 Sec  or  30 Sec       11 4 5 Setting Gridding    1  Enter  Recorder Setup  menu through the above methods   2  Set  Open Grid  to  On  or  Off       11 4 6 Setting Recorder Speed    1  Enter  Recorder Setup  menu through the above methods   2  Set  Recorder Speed   to  6 25 mm s     12 5 mm s    25 mm s  or  50  mm s       11 5 Installing Recording Paper    If the record paper runs out  please install the record paper as the following step    1  Press both sides of the recorder door with one hand and pull outwards to open  the recorder door    2  Put the recording paper into the recorder with the thermal side  which is  smoother up    3  Close the door of the recorder  and pull some recording paper outside of the    paper out port   Caution  e Must use the thermo sensitive record paper  otherwise  it will lead to  recording failure  bad quality record or damage of thermo sensitive    printing head     11 2    Vital Signs Monitor User s Manual       e Do not pull out the recording paper during recorder printi
77. the power wire and other  connected wires    Open the battery door towards the direction labeled on it    Take out the old battery    Insert the new battery towards the direction labeled     Close the battery door     12 3 Optimizing Battery Performance    A battery needs at least two optimizing cycles when it is put into use for the first    time  A battery cycle is one complete  uninterrupted charge of the battery  followed by    a complete  uninterrupted discharge of the battery  A battery should be conditioned    regularly to maintain its useful life  Condition a battery once when it is used or stored    for two months  or when its run time becomes noticeably shorter     To optimize a battery  follow this procedure     1     6     Disconnect the monitor from the patient and stop all monitoring and  measuring procedures    Place the battery in need of optimizing into the battery compartment to the  monitor    Place the monitor in the charger stand and connect the AC mains  Allow the  battery to be charged uninterruptedly for above 6 hours    Remove the AC mains and allow the monitor to run from the battery until it  shuts off    Replace the monitor in the charger stand and connect the AC mains  Allow  the battery to be charged uninterruptedly for above 6 hours     The optimizing of the battery is over     12 4 Checking Battery Performance    The performance of a battery may deteriorate over time  To check the    performance of a battery  follow this procedure     1     Dis
78. tton            mmm Connecting wire       Vital Signs Monitor User s Manual       9 2 Safety Information    Warning      Please change the earcap after each use to ensure the veracity and avoid  cross infecting      At least every half year or one year carries out a temperature measure  calibration  or carries out the calibration according to the hospital    regulation     9 3 Monitoring Procedure    1  Connect the temperature module to the main unit  press the power button of the    main unit to turn on the monitor     2  Installs earmuffs to the temperature probe    While monitoring  please use new clean earmuffs with non breakage and ensure  that the ear cavity is clean  Please install the earmuffs as follows to ensure the  monitoring accuracy    The earmuffs installing way    1  Put the earmuffs paper film into the assistant ring groove   2  Insert the probe to the end when its center is at the center of the assistant ring     3  Ensure that the earmuffs on the probe is fixed on     Refer to the following pictures        9 2    Vital Signs Monitor User s Manual       3  Pull back the ear cavity gently to make it straight  and fix the head of the patient        4  Measuring method  When the temperature module is in the off state or the unit sign  glints  please carry out the monitoring  Insert the temperature probe into the ear at  the ear hole  then press    START     please don t read the temperature information  until you hear the buzzer sound    The temperature measure
79. uired     Setting SpHb Unit    You can select  SpHb Unit  in the SpO   parameter setting menu  and select the  unit as required  The options are  g dL 1 and  mmol L       6 6 Masimo Information    UNASHO    rainbow         E Masimo Patents    This device is covered under one or more of the following U S A  patents   5 758 644  5 823 950  6 011 986  6 157 850  6 263 222  6 501 975 and other  applicable patents listed at  www masimo com patents htm    E No Implied License    Possession or purchase of this device does not convey any express or implied  license to use the device with unauthorized sensors or cables which would  alone  or  in combination with this device  fall within the scope of one or more of the patents    relating to this device     6 7 Nellcor Information       This is the trademark of Nellcor Puritan Bennett Inc     6 7    Vital Signs Monitor User s Manual       Chapter 7 NIBP    7 1 Introduction    The monitor uses the oscillometric method for measuring NIBP  It is applicable    for adult  pediatric and neonatal patients  The method of oscillometric indirectly    estimates the systolic and diastolic pressures within the blood vessels by measuring    the change of the pressure within blood pressure cuff along with the volume of the    arteries and calculates the average pressure     The NIBP measurement is suitable for use in the presence of electrosurgery and    during the discharge of a cardiac defibrillator according to IEC 60601  2 30 1999 EN  60601 2 30 
80. value and unit      SpMet  Methemoglobin Saturation  value and unit      SpHb  Total Hemoglobin  value and unit      SpOC  Oxygen Content  value and unit      SpCO  Carboxyhemoglobin Saturation  value and unit     PVI  Pleth Variability Index  value and unit      PR value    VD 0 N Mm 4 bh W Nh    Note    O When one of the above parameters    signals is weak  its label will fresh       Parameters Description    6 3    Vital Signs Monitor User s Manual       a  Perfusion index  PI   PI is a value that indicates arterial pulse signal strength as the  percentage of pulsatile signal to non pulsatile signal  The perfusion index allows  clinicians to place sensors on optimal sites    b  Methemoglobin Saturation  SpMet   SpMet is a value that represents the  percentage of methemoglobin saturation within the blood    c  Total Hemoglobin  SpHb   SpHb is a measure of the total hemoglobin  concentration in arterial blood    d  Oxygen Content  SpOC   SpOC is a measure of the total oxygen content present in  the blood    e  Carboxyhemoglobin Saturation  SpCO   SpCO is a value that represents the  percentage of carboxyhemoglobin saturation within the blood    f  Pleth Variability Index  PVD  PVI is a measure of peripheral perfusion changes  secondary to respiration  or the PI amplitude modulation over a respiration  and can be    closely related to intrathoracic pressure changes     6 5 Setting SpO     6 5 1 Opening the SpO  Menu    To open the SpO  Menu   O Select  SpO     to enter  SpO  
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
G DATA TotalProtection 2013, 1Y, 1PC, Box  Samsung 700DX-2 ユーザーマニュアル  Transition Networks 1000MCC-1 fiber optic cable  Nokia CK-300 GPS Receiver User Manual  Ressentir !  Poulan 532 43 34-32 Lawn Mower User Manual  No.14 1995  Samsung NP-R411 用户手册(Windows 7)  Samsung Galaxy S4 Active Manual de utilizare(Jellybean)  Silicon Graphics® 1600SW    Copyright © All rights reserved. 
   Failed to retrieve file