1. RH 10 90 RH 1 Dimensions and Weight 195 x 55 x 30 mm 250g 2 Electrical Safety to IEC 1010 1 e Dual Insulation e Installation Category III e Degree of Pollution 2 e Rated Voltage or Max Live Earth 560V RMS 3 CE Mark Conforms to EN 50081 1 and 50082 1 standards 5 OPERATING PROCEDURE Connect the leads to the input and place the wire grip on the circuit to be tested Connect the probe to the oscilloscope with the accessory BNC BNC lead Adjust the vertical zero adjustment of the oscilloscope if necessary Select the attenuation ratio and the vertical deviation of the oscilloscope in accordance with the conversion table below NOTE The POWER light must be ON and note the Max input voltage of each range The conversion table gives the real vertical deviation Attenuation x 200 x 100 x 50 x 20 Max Input DC 800V 400V 200V 80V Max Input AC p p 1600V 800V 400V 160V Max Input AC RMS 560V 280V 140V 56V 10 DP 200pro Instruction Manual Vertical Deviation on the Real Deviation In V div Oscilloscope in V div x200 x100 x50 x20 1V 200V 100V 50V 20V 0 5V 100V 50V 25V 10V 0 2V 40V 20V 10V 4V 0 1V 20V 10V 5V 2V 50 mV 10V 5V 2 5V 1V 20 mV 4V 2V 1V 0 4V 10mV 2V 1V 0 5V 0 2V 5 mV 1V 0 5V 0 25V 0 1V N B The real vertical deviation in V div is equal to the attenuation factor multiplied by the range of vertic
2. DIFFERENTIAL PROBE DP 200pro Super High Frequence Model 20mVp p 1 6KVp p DC 200MHz DP 150pro Super High Range Model 10mVp p 8KVp p DC 150MHz Ce 0 INSTRUCTION MANUAL TABLE OF CONTENTS DP 150pro Differential Probe 1 FEATURES 1 2 SPECIFICATIONS 1 3 PANEL DESCRIPTION 2 4 OPERATING ENVIRONMENTAL CONDITIONS 3 5 OPERATING PROCEDURE 3 6 EXT POWER SOURCE 4 7 MAINTENANCE 5 8 CLEANING 5 9 WARRANTY 5 10 REPAIR 6 DP 200pro Differential Probe 1 FEATURES 8 2 SPECIFICATIONS 8 3 PANEL DESCRIPTION 9 4 OPERATING ENVIRONMENTAL CONDITIONS 10 5 OPERATING PROCEDURE 10 6 EXT POWER SOURCE 11 7 MAINTENANCE 11 8 CLEANING 12 9 WARRANTY 12 10 REPAIR 12 DIFFERENTIAL PROBE COMPARISON CHART 13 ACCESSORIES 14 e Differential Voltage Probe Read the instructions before using the instrument 1 Must acquire a differential voltage probe amp get the best service from instrument 2 Read carefully the USER MANUAL 3 Respect the safety precautions e SAFETY PRECAUTIONS Warning Risk of Electric Shock Respect the max input voltages DP 150pro 1 Max differential voltage 8KV DC AC peak or 2 8KV RMS 2 Max voltage between each input terminal and ground 3 5KV RMS DP 200pro 1 Max differential voltage 1 6KV DC AC peak or 560V RMS 2 Max voltage between each input terminal and ground 600V RMS Do not use the probe in damp en
3. able for normal use of our instruments and does not apply to damage or destruction caused notably by error in mounting mechanical accident faulty maintenance defective use overload or exceed voltage Our responsibility being strictly limited to the pure and simple replacement of the faculty parts of our equipment the buyer expressly renounces any attempt to find us responsible for damages or losses caused directly or indirectly Our guarantee is applicable for twelve 12 months after the date at which the equipment is made available The repair modification or replacement of a part during the guarantee period will not result in this guarantee being extended 10 REPAIR Maintenance repairs under or out of guarantee Please return to product to your distributor 12 PINTEK DIFFERENTIAL PROBE COMPARISON CHART PINTEK PINTEK DP 40 650V MODEL PINTEK PINTEK DP 25 DP 30 FEATURE Economic Super High Model 1300V Voltage 20mV lmV 10mV 1300V 65V 650V 3 x20 3 x1 x2 3 x10 x50 x200 x10 x20 x100 Economic Model 6500V 4 x100 x200 x500 x1000 Attenuator Selection Range 4MQ 2MQ 4MQ Each Input R C 2MQ IMQ 2MQ Differential R C Adapter PINTEK DP 65pro 1600V Performance Model 3 x20 x50 x200 2010 5 28 PINTEK PINTEK DP 150pro DP 200pro PINTEK DP 22Kpro SuperRange Gain Super High PINTEK PINTEK PINTEK DP 100 DP 14K DP 20K Voltage Input 7TSMHz 20MH
4. al deviation selected on the oscilloscope It will be doubled in the case of use a 500 load Example With the differential probe on factor x 100 the oscilloscope on 0 5 V div the real vertical deviation is 100 x 0 5 50 V div With a 500 load on the input of the oscilloscope the deviation becomes 100 V div 6 EXT POWER SOURCE Power consumption of the probe are 150mA 1 4W thus it not suit for battery please use the accessory adapter only If there are any damage on the adaptor please contact us and use the adaptor supply by us only If the input power over 12V DC or the opposite polarity will caused to the probe hard damage 7 MAINTENANCE For maintenance only use specified spare parts 11 DP 200pro Instruction Manual The manufacturer can not be held responsible for any accident arising following a repair made other than its after sales service or approved repairers 8 CLEANING This probe does not require any particular cleaning If necessary clean the case with a cloth slightly moistened with soapy water 9 WARRANTY Unless notified to the contrary our instruments are guaranteed against any manufacturing defect or material defect They do not bear the specification known as the safety specification Our guarantee which may not under any circumstances exceed the amountof the invoiced price goes no further than the repair of our faulty equipment carriage paid to our workshops It is applic
5. ange 3 Accuracy 2 4 Voltage Input Ranges RANGE MAX DC MAX AC RMS MAX AC p p x1000 4KV 2 8KV 8KV x300 1 2KV 840V 2 4KV x100 400V 280V 800V x30 120V 84V 240V x10 40V 28V 80V 5 Permitted Max Input Differential Voltage Max differential voltage 8KV DC AC peak to peak Max voltage between each input terminal and ground 2 8KV RMS NOTE The Max operation voltage between terminal and ground of each range are the same as the Max AC RMS input DP 150pro Instruction Manual 6 Input Impedance Differential 100MQ 1 0 pF Between terminals and ground 50MQ 2 0 pF 7 Output 5V 8 Output Impedance 50 Q 9 Rise Time 3 5 ns for x10 and 2 4 ns for x30 x100 x300 x1000 10 Rejection Rate on Common Mode 60 Hz gt 80dB 100 Hz gt 60dB 1 MHz gt 50dB 11 EXT Power Supply External 9 V DC power supply 12 Consumption 150 mA max 1 4 WATT 3 PANEL DESCRIPTION Input Connector ii Over Range Indication Power ON OFF Indication Attenuation Function Switch External Power Source DP 150pro Instruction Manual 4 OPERATING ENVIRONMENTAL CONDITIONS Reference Use Storage Temperature 20 C 30 C 0 C 50 C 30 C 70 C Relative Humidity lt 70 RH 10 85 RH 10 90 RH 1 Dimensions and Weight 245 x 76 x 32 mm 350g 2 Electrical Safety to IEC 1010 1 e Dual Insulatio
6. cilloscope accuracy is 3 and DMM is 10 times SPECIFICATIONS 1 Bandwidth DC to 200MHz 3 dB for x50 x100 x200 DC to 100MHz for attenuation x 20 2 Attenuation x20 x50 x100 x200 3 Accuracy 2 4 Voltage Input Ranges RANGE MAX DC MAX AC RMS MAX AC p p x200 800V 560V 1600V x100 400V 280V 800V x50 200V 140V 400V x20 80V 56V 160V 5 Permitted Max Input Differential Voltage Max differential voltage 1 6KV DC AC peak to peak Max voltage between each input terminal and ground 560 V RMS NOTE The Max operation voltage between terminal and ground of each range are the same as the Max AC RMS input DP 200pro Instruction Manual 6 Input Impedance Differential 18MQ 1 7 pF Between terminals and ground 9MQ 3 4 pF 7 Output 5V 8 Output Impedance 500 9 Rise Time 3 5 ns for x20 and 1 7 ns for x50 x100 x200 10 Rejection Rate on Common Mode 60 Hz gt 80 dB 100 Hz gt 60 dB 1 MHz gt 50 dB 11 EXT Power Supply External 9 V DC power supply 12 Consumption 150 mA max 1 4 WATT 3 PANEL DESCRIPTION Input Connector _4 de External Power Source Output Connector gt DP 200pro Instruction Manual 4 OPERATING ENVIRONMENTAL CONDITIONS Reference Use Storage Temperature 20 C 30 C 0 C 50 C 30 C 70 C Relative Humidity lt 70 RH 10 85
7. elected on the oscilloscope It will be doubled in the case of use a 500 load Example With the differential probe on factor x 10 the oscilloscope on 0 5 V div the real vertical deviation is 10 x 0 5 5 V div With a 500 load on the input of the oscilloscope the deviation becomes 10 V div 6 EXT POWER SOURCE Power consumption of the probe are 150mA 1 4W thus it not suit for battery please use the accessory adapter only f there are any damage on the adaptor please contact us and use the adaptor supply by us only If the input power over 12V DC or the opposite polarity will caused to the probe hard damage DP 150pro Instruction Manual 7 MAINTENANCE For maintenance only use specified spare parts The manufacturer can not be held responsible for any accident arising following a repair made other than its after sales service or approved repairers 8 CLEANING This probe does not require any particular cleaning If necessary clean the case with a cloth slightly moistened with soapy water 9 WARRANTY Unless notified to the contrary our instruments are guaranteed against any manufacturing defect or material defect They do not bear the specification known as the safety specification Our guarantee which may not under any circumstances exceed the amountof the invoiced price goes no further than the repair of our faulty equipment carriage paid to our workshops It is applicable for normal use of ou
8. n e Installation Category III e Degree of Pollution 2 e Rated Voltage or Max Live Earth 2 8KV RMS 3 CE Mark Conforms to EN 50081 1 and 50082 1 standards 5 OPERATING PROCEDURE e Connectthe leads to the input and place the wire grip on the circuit to be tested e Connectthe probe to the oscilloscope with the accessory BNC BNC lead e Adjustthe vertical zero adjustment of the oscilloscope if necessary e Select the attenuation ratio and the vertical deviation of the oscilloscope in accordance with the conversion table below e NOTE The POWER light must be ON and note the Max input voltage of each range The conversion table gives the real vertical deviation Attenuation x 1000 x 300 x 100 x 30 x 10 Max Input DC 4KV 1 2KV 400V 120V 40V Max Input AC p p 8KV 2 4KV 800V 240V 80V Max Input AC RMS 2 8KV 840V 280V 84V 28V DP 150pro Instruction Manual Vertical Deviation on the Real Deviation In V div Oscilloscope in Vidiv gt 41000 x300 x100 x30 x10 1V IKV 300V 100V 30V 10V 0 5V 500V 150v 50v 15V 5V 0 2V 200V 60v 20V 6v 2V 0 1V 100v 30v tov 3v 1V 50 mV 50V 15V 5V 1 5V 0 5V 20 mV 20V 6V 2V 0 6V 0 2V 10 mV 10V_ 3V 1V 03V 01V 5 mV 5V 15V 0 5V 150mv 50mv NOTE The real vertical deviation in V div is equal to the attenuation factor multiplied by the range of vertical deviation s
9. r instruments and does not apply to damage or destruction caused notably by error in mounting mechanical accident faulty maintenance defective use overload or exceed voltage Our responsibility being strictly limited to the pure and simple replacement of the faculty parts of our equipment the buyer expressly renounces any attempt to find us responsible for damages or losses caused directly or indirectly Our guarantee is applicable for twelve 12 months after the date at which the equipment is made available The repair modification or replacement of a part during the guarantee period will not result in this guarantee being extended DP 150pro Instruction Manual 10 REPAIR Maintenance repairs under or out of guarantee Please return to product to your distributor DP 200pro Differential Probe Super High Frequence Model DP 200pro DIFFERENTIAL PROBE 1 2 FEATURES The DP 200 differential voltage probe provides a safety means of measuring floating potentials for all models of oscilloscopes incomplete safety It converts the high differential voltage 1 6KV peak into a low voltage 5V with reference to the earth for display on the oscilloscopes The BNC output is designed to operate on an input with an impedance of 1 MQ Itis 2 times of the 50Q Recommend to use Powertek PL 10 with DP 200 and expand the measuring ranges From DMM can observe more exact measured testing voltage Os
10. vironment or where there is risk of explosion Do not use the probe with its case open Disconnect the inputs and outputs of the probe before opening the case e TO ORDER Differential Voltage Probe and Accessories a An Insulated BNC BNC lead and two 44 mm length 3 inches BP 250 Supplied a Adapter preset 9 V DC 115 V or 230 V a 2 x high voltage IC clips BP 256N BP 266 2 x Banana to Banana high voltage plug BP 356N BP 366 a 2 x Alligator plug BP 276N a 2 x Test Lead BP 286 Instruction Manual Carry Box PX 501 for DP 200pro PX 502 for DP 150pro DP 150pro Differential Probe Super High Range Model DP 150pro DIFFERENTIAL PROBE 1 2 FEATURES The DP 150 differential voltage probe provides a safety means of measuring floating potentials for all models of oscilloscopes incomplete safety It converts the high differential voltage 8KV peak into a low voltage 5V with reference to the earth for display on the oscilloscopes The BNC output is designed to operate on an input with an impedance of 1 MC Itis 2 times of the 50 Q Recommend to use Powertek PL 10 with DP 150 and expand the measuring ranges From DMM can observe more exact measured testing voltage Oscilloscope accuracy is 3 and DMM is 10 times SPECIFICATIONS 1 Bandwidth DC to 150 MHz 3 dB for x30 x100 x300 x1000 DC to 100 MHz for attenuation x10 2 Attenuation x10 x30 x100 x300 x1000 5 r
11. z 150MHz 200MHz 120MHz 200mV 300mV 10mV 20mV 30mV 14KV 20KV 8KV 1600V 22KV 4 x300 5 7x10 x30 4 x20 x50 5 7x30 x100 x600 1500 x100 x300 x100 x200 X300 x1000 x3000 x1000 X3000 100MQ 100MQ 100MQ 18MQ 100MQ 50MQ 50M Q 50M Q IMQ 50M Q CARRI BOX o o o 0 BP 250 BP 250 BP 256N BP 276N BP 286 BP 356N BP 256N BP 266 BP 356N BP 276N BP 366 e BP 250 BP 256N BP 276N BP 286 BP 356N BP 256N BP 276N BP 286 BP 356N BP 276N BP 286 BP 366 BP 276N BP 286 BP 366 BP 276N BP 286 BP 366 BP 276N BP 286 BP 366 BP 276N BP 286 BP 356N X Specifications and appearance design subject to change without notice 13 ACCESSORIES BP 256N IC Clip 1000Vrms 5A BP 266 HV IC Clip 7KVrms Alligator Clip 1KVrms 10A BP 356N Banana Plug to Banana Plug 6KVDC 10A 60cm BP 366 HV Banana Plug to Banana Plug 20KVDC 10A 60cm Test Leader 1KVrms 10A BP 250 BNC Plug to BNC Plug 50 Q cable 100cm 1GHz 3A PX 501 PX 502 Carry Box Adapter ADP 110 ADP 220 DC 9V 300mA DP 200pro BP 256N BP 356N BP 250 BP 276N BP 286 PX 501 ADP 110 ADP 220 DP 150pro BP 266 BP 366 BP 250 BP 276N BP 286 PX 502 ADP 110 ADP 220 Option PL 10 to Connect With Digital DMM Oscillossope DP 150pro DP 200pro Differeritial Probe