1. or other competent individual is close at hand when the patient is in tilt Ensure that all electrical cables are positioned away from the moving parts of the table A Press to raise table Tilt Press Y to lower table Tilt AKRON 8601B 3 Operation Head Section CAUTION Ensure Head Extension Frame is fully located in desired position before use To adjust the length of the Head Extension A Locate locking pin on underside of top frame B Pull the pin out C Adjust Head Extension to desired position D The pin will spring into the holes on the Head Extension to secure Foot Board WARNING For reasons of safety always check the Foot Board is secure before use To adjust the Foot Board into position A Unscrew locking knob at the base of the board B Slide Foot Board to desired position to suit patient C Tighten locking knob to secure Foot Board in new position Pommel To adjust the Pommel A Unscrew locking knob below the support B Slide Pommel to desired position C Tighten locking knob to secure Pommel in new position AKRON 8601B 3 Operation Hand Grips To fit Hand Grips A Remove Work Table if fitted B Slide hand grips into fixation brackets on both sides of table C Tighten locking knobs on fixation brackets to secure Hand Grips To adjust height of Hand Grips A Loosen locking knobs on sliding fixation brackets
2. contamination is known or suspected the surface of the straps should be sprayed liberally using a hypochlorite solution of 1 000 ppm or a 1 000 ppm solution of sodium dichloroisocyanurate NaDCC after removing excessive organic material The straps should then be air dried at ambient temperature and inspected as above prior to patient use 5 Troubleshooting Electric Actuator Symptom Possible cause Action Actuator will not work Power disconnected Ensure mains supply is connected Fuse blown Replace fuse in socket Hand footswitch Check that the hand foot switch is fully disconnected inserted into the actuator If the problem persists consult your local Akron Supplier 10 AKRON 8601B 6 Maintenance Check table visually before use Carry out routine servicing every six months or more frequently if in heavy use Routine servicing or repair must only be performed by suitably qualified person WARNING Before commencing any maintenance activity disconnect the table from the electrical supply Plug acts as disconnect device Examine the table for any signs of damage Make sure that all nuts bolts and other fasteners are tight Do not over tighten as this may cause binding Upholstery fasteners should be checked tightened or replaced if necessary Check section hinge pins to ensure that no pin protrudes Check gas strut controls throughout the full range of movement for locking and assistance Check hydra
3. the UK by Huntleigh Akron As part of the ongoing development programme this company reserves the right to modify specifications and materials of this couch without notice 8 Revision Status PAGES ISSUE WRITTEN DATE ALL PAGES UM 01 04 06 98 ALL PAGES UM 02 18 12 98 ALL PAGES UM 03 25 01 99 PAGE 12 amp 13 UM 04 15 11 99 PAGE 2 UM 05 DCN958 30 10 00 13 AKRON 8601B 9 Technical Specification 8601B Width cm 33x15 Overall Length cm 17123 Height Range cm 8341 5 Max Load Weight 50 Kg Tilting Control Electrical Data 230v Voltage 220 240v Actuator Linak 32K 300 34 005 Power Supply CB 1A 60 24 Charger Input Output Batteries Main Plug Fuse Rt _ _ EEE Linak 32K 300 34 005 Power Supply CB 1A 60 24 CD08 110 110v 60 Hz 14 Va 24v 220mA 5 3 Va Re order Code 5F3600Z IEC 601 1 Classification nl Double Insulated Type B equipment A Attention consult this manual 14 AKRON 8601B 15 AKRON 8601B 16 Jr Huntleigh AKRON Huntleigh Akron 1 Farthing Road Ipswich Suffolk IP1 5AP England Tel 44 0 1473 461042 Fax 44 0 1473 462924 792 209 05
4. to bottom using a damp cloth and clean water C Wipe couch carefully dry with clean paper towels from top to bottom If DECONTAMINATING then repeat steps a to c using a dilute Hypochlorite warm water mixture 10 000ppm instead of warm water and soap Safely discard cleaning decontaminating materials Wash hands thoroughly Put couch back into service AKRON 8601B 4 Cleaning and Decontamination Decontamination of Straps It is not recommended that the straps are laundered on a regular or routine basis as this will eventually compromise the integrity of the material and associated Tac Flex However under certain conditions ie when the straps become heavily soiled or when they are suspected or known to have been in contact with an infectious organism they can be laundered or decontaminated using one of the following methods General Wash The straps should be washed separately as the Tac Flex will attract debris from other material which will ultimately reduce its effectiveness The Tac Flex should be joined prior to washing The straps may then be laundered in the normal fashion at a temperature of approximately 71 C and allowed to dry in ambient conditions After such wash the straps should be thoroughly inspected prior to patient use If deterioration is present or the Tac Flex fails to maintain integrity the straps must be replaced Decontamination Eye protection and suitable clothing should be worn Where
5. Mi Huntleigh A KR ON AKRON CHILDS TILT TABLE Model 8601B USER MANUAL AKRON 8601B Contents 1 Introduction Warnings Cautions General Warnings Installation Operation Wheel Locking Tilting Control Head Section Foot Board Pommel Hand Grips Pelvic Support Work Table Fixation Belts amp Harnesses Cleaning and Decontamination Troubleshooting Maintenance Warranty and Service Revision Status Technical Specification CE The C mark on this product demonstrates conformity with the EC Directive 93 42 EEC NN 9 1O0 ta I 4 d oo 11 12 13 13 14 AKRON 8601B 1 Introduction Always keep this Operating Manual available for reference Warnings amp Cautions WARNINGS Warnings identify possible hazards in procedures or conditions which if not correctly followed could result in death injury or other serious adverse reactions CAUTIONS Cautions identify conditions or procedures which if not correctly followed could result in equipment failure or damage General Warnings WARNINGS Use only accessories that have been designed or approved for use with this table Do not use electric couches in the presence of flammable gasses such as an anaesthetic agent Maximum Load 50 Kg 110 Ibs distributed evenly Do not concentrate weight on either side of the table Plug acts as a disconnect device Do not allow children to play with the table at any tim
6. e Adequately supervise children in the proximity of the table AKRON 8601B 2 Installation Battery Charging CAUTION Do NOT use table when batteries are being charged To Charge Batteries A Plug charger lead adaptor into handswitch as shown B Although batteries are charged prior to packing charge for a minimum of 12 hours before use C Charge batteries at the end of each day after use Ensure mains supply cable is not stretched when connected Ensure during table use no obstacles are in its path Ensure handswitch tubes or leads are not trapped by moving parts of the table CAUTION Ensure that the electricty supply is compatible with the table The Voltage required is marked on the Serial No Label Serial No Label AKRON 8601B 3 Operation Wheel Locking WARNING Lock table wheels castors prior to patient transfer or treatment The retractable castors are not designed for transportation of patients Disconnect and coil electrical cord away from moving parts prior to repositioning the table Castors Castors Unlocked Castors Locked Tilting Controls WARNING Ensure that the patient is safely positioned prior to tilting the table Always secure the patient using the straps provided prior to tilting the table Under no circumstances should the table be moved with the patent in situ and in tilt mode In the interest of patient safety always ensure that a member of staff
7. ly Always ensure that the straps and harnesses are in the correct position and that the patient is fully secure before operating the tilt facility Only use Huntleigh Akron straps and harnesses The Table is supplied with the following Fixation Harnesses and Belts TT3 x Wide Belt TT2 2 x Wrap Straps 1H0402A The Wide Belt is used across the patient s knees and fastened using Velcro strips which wrap around the chrome rails One Narrow Belt is fitted across the chest with the other across the waist The Wrap Straps are used to hold the patients feet in place The straps will pass through the slots in the Foot Board to secure the feet 1f required Secure in place using the Velcro part of the straps AKRON 8601B 4 Cleaning and Decontamination WARNING Do not use Alkalis or other organic solvents for routine cleaning Replace torn or damaged vinyl immediately Always cover vinyl with fresh paper or sheet between patients If contamination is known or suspected always decontaminate couch before use Frequency of routine couch cleaning will depend on exact application CLEANING DECONTAMINATION Wear suitable protective clothing Disconnect mains supply Locate couch in a suitable cleaning area with castors disengaged or wheels locked A Using a clean damp cloth warm water and soap carefully wipe couch from top to bottom taking care around electrical components B Wipe couch carefully from top
8. ng 3 Attach the decontamination certificate or other written statement declaring that the product has been cleaned to the outside of the package 4 Mark the package Service Department Manufactured in the UK by Huntleigh Akron As part of the ongoing development programme this company reserves the right to modify specifications and materials of this couch without notice 12 AKRON 8601B 7 Warranty and Service UK only Huntleigh Akron s standard terms and conditions apply to this couch A copy is available upon request These contain full details of warranty terms and do not limit the statutory rights of the consumer For service maintenance and any questions regarding this or any other Huntleigh Akron Product please contact Huntleigh Akron Farthing Road Ipswich Suffolk IP1 SAP England Telephone 44 0 1473 461042 Fax 44 0 1473 462924 Huntleigh Akron T A a Division of Huntleigh Nesbit Evans Limited A Member of the Huntleigh Technology PLC Group of Companies Huntleigh Technology PLC 2001 If for any reason your couch is being returned please inform Huntleigh Akron prior to sending and 1 3 4 Clean the product as described in the cleaning section Pack it in suitable packing Attach the decontamination certificate or other written statement declaring that the product has been cleaned to the outside of the package Mark the package Service Department Manufactured in
9. on both sides of table B Slide hand grips along chrome rails to desired height C Re tighten locking knobs to secure hand grips Pelvic Support CAUTION The Pelvic Supports cannot be fitted at the same time as the Accessory Worktable To fit and adjust Pelvic Support A Slide support into fixation bracket on chrome rail B Secure pelvic support to desired height by tightening locking knob on fixation bracket C Adjust support to width required by tightening the second locking knob on the fixation bracket AKRON 8601B 3 Operation Accessory Work Table WARNING The Accessory Work Table has a safe working load of 20 Kg To fit the Work Table A Insert Work Table Frame into sliding fixation brackets B Tighten locking knobs on fixation brackets to secure Work Table Work Table To adjust the height of the Work Table A Loosen locking knobs on sliding fixation brackets on both sides of table B Slide Worktable Assembly along chrome rails to desired height C Re tighten locking knobs to secure Work Table To alter the angle of the Work Table A Unscrew hand wheel on the underside of the Work Table Hand wheel B Adjust Work Table to the angle required C Tighten hand wheel to secure Work Table in the new position AKRON 8601B 3 Operation Fixation Belts and Harnesses WARNING Check condition of straps and harnesses If worn out or damaged replace immediate
10. ulic tables for smooth operation and signs of leakage WARNING All electrical checks should be performed by a suitably qualified and licensed electrician Neither BLUE or BROWN wires are to be connected to the earth terminal of a three pin plug Examine the flexible cables for cuts abrasions kinking or other deterioration and replace if necessary Ensure that the power supply connector plug is securely attached Check mains plug connections are tight and fuse is correctly rated 5A 11 AKRON 8601B 7 Warranty and Service UK only Huntleigh Akron s standard terms and conditions apply to this couch A copy is available upon request These contain full details of warranty terms and do not limit the statutory rights of the consumer For service or maintenance please contact Service Department Huntleigh Healthcare 310 312 Dallow Road Luton Beds Telephone 01582 745700 LUI 1TD Fax 01582 745869 For information on this product or any other Huntleigh Akron product please contact Huntleigh Akron Farthing Road Ipswich Telephone 01473 461042 IP1 SAP Fax 01473 462924 Huntleigh Akron T A a Division of Huntleigh Nesbit Evans Limited A Member of the Huntleigh Technology PLC Group of Companies Huntleigh Technology PLC 2001 If for any reason your couch is being returned please inform Huntleigh Akron prior to sending and 1 Clean the product as described in the cleaning section Pack it in suitable packi