Home
Installation/User Guide - Vidar Systems Corporation
Contents
1. 25 Unpacking n AD a Oa a dE 26 To Look for damage nii eroi et i ce Ie teet n enia 26 2 Unpack the shipping Carton 27 3 Unpack IG accessory boxu u uu rte E rd teta ates 28 LB cC 30 5 Identify the GomponentS u u ecce te dnce beue idee 32 6 If anythinig is missing eee icr e tn d i pe ear etes 33 7 Activate your Droduct waan 33 SO 34 Identify important features 35 Assemble the GigitiZer u u a S a sau akay kiwa nnne nnn nnns 36 Connect the power con 40 Where Io ger from here teu ete eec meteo Qata Ee de aat 41 Interfacing ET TR A3 VIDAR Drivers and Toolkit Installation CD and Driver Information Tab 4 VIDAR device driver summary sese enemies 45 About the STI drivers u ana Susu haha hay saa nnne nes 45 E Uu EE 47 If necessary set the digitizer s SCSI ID 48 If necessary terminate the SCSI bus 49 Connect the SCSI cable to the digitizer eene nenr tne nennnerenrne 50 Installing SCSI bhardware u sua nnne nenas nsnm s innen 51 1 Install the SCSI card ironiei io adeno en adani eaea enne nnne nns 53 2 Configure the SCSI BIOS sse enne 55 Installing SCSI IT 59 UE Ile RTR u u I ui Au aR aa 67 USB interface requirements A 67 Updating a previous VIDAR STI driver installation 68 Installing the USB drivers esses n
2. Cleaning the Auto Feed Roller and Suction Cups CADPRO Advantage only After extensive use dust or lint particles may build up on the Auto Feed Roller and Suction Cups and they must be cleaned to maintain reliable film feeding VIDAR recommends cleaning the Auto Feed Roller and Suction Cups at least twice per year During cleaning inspect the Suction Cups for wear 1 Turn off the digitizer using the switch on the rear panel 2 Remove the Film Guide from the feeder by loosening four screws Loosen each screw a few turns then move to the next screw Repeat until the Film Guide is freed from the feeder Note The screws are captive but will come entirely out if unscrewed too far and could then fall into the digitizer base Be careful not to unscrew them from their captive locations Removal of screws or other items from the inside of the digitizer is not covered under your warranty Auto Feed Roller Captive screws Captive screws d Film Guide B putri VIDAR Advantage Series Installation and Operation Guide 100 3 Remove residue from the Auto Feed Roller Loop a piece of light adhesive tape such as Scotch Magic Tape around your fingers with the adhesive side out then gently pat the roller Pat and turn the roller until its entire surface is clean 4 Gently pull the Suction Cups forward 5 Examine the Suction Cups If you see cracks or irregularities you should replace the Suction Cu
3. Note m On Windows 2000 computers Service Pack 2 or higher must be installed for SCSI installations Computers without Service Pack 2 or higher will not be supported m On Windows XP Professional Edition Service Pack 1 or higher must be installed for SCSI installations VIDAR s STI device driver provides compatibility between VIDAR film digitizers scanners and Windows 2000 and Windows XP only The STI driver takes advantage of Microsoft s Still Image Architecture STD for digital imaging devices You will need m VIDAR Drivers and Toolkit Installation CD Note These instructions have been successfully tested on a wide range of Windows 2000 and Windows XP systems In rare instances you may have difficulty installing the STI driver If that happens try again after logging on as Administrator or as a user with Administrator rights 1 Be sure the digitizer is turned OFF and is NOT connected to the PC Wait until you are prompted to connect the digitizer 2 Turn on the PC 3 Insert the VIDAR Drivers and Toolkit Installation CD into the CD ROM The installer should launch automatically VIDAR Advantage Series Installation and Operation Guide 60 If the installer does not launch automatically double click the appropriate CD drive icon under My Computer in Windows Explorer 4 In the first screen click Install Software INSTALL SOFTWARE CD CONTENTS VIEW OUR WEBSITE CONTACT US Ins
4. The digitizers do not include application specific software Picture Archiving and Communications PAC system Teleradiology Oncology Systems or Computer Aided Detection CAD software The manufacturer of the application software will determine specific indications for use These third party software packages or complete systems are approved separately from a regulatory perspective The digitizers are marketed as a component to application software development companies who will incorporate the digitizer into their respective PACS or Teleradiology CAD system s The software developer is ultimately responsible for detailing the Contradictions for the PACS System or Teleradiology software package or Oncology Systems as a whole including the digitizer Caution No operator serviceable parts inside Refer servicing to qualified personnel Achtung Geh use nicht ffnen Wartung uno reparatur nur durch eletrofachkr fte Attention Aucune piece ne peut etre remplacee par l utilisateur Toute operation de maintenance doit etre effectuee par une personne qualiee Atencion Acceso interno solo autorizado a personal tecnico cualificado Attenzione Non appire Rivolgersi a personale qualificado 365 Herndon Parkway Sales 1 800 471 SCAN or 1 800 471 7226 D A R Herndon VA 20170 U S A Email sales vidar com systems corporation Phone 1 703 471 7070 Technical Support Fax 1 703 471 7665 1 703 471 7070 or 1 800 471 7226 Medical I
5. 2000 or Windows XP L Update to STI drivers See Computers with Windows 2000 and Windows XP in the Installing device drivers chapter L Check all SCSI cables and connectors m Problem Digitizer is not detected on SCSI bus L Set up SCSI card See the Installing SCSI hardware chapter L Check all SCSI cables and connectors VIDAR Advantage Series Installation and Operation Guide 119 m Problem SCSI card driver not loading or hanging during Windows 2000 installation This applies to all AdaptecTM PCT SCSI adapters cards running the Windows 2000 operating system Note Use these troubleshooting steps in sequence Stop when the problem is corrected and do not execute the next step 1 Update the motherboard BIOS to make it A C P I compliant ACPI Advanced Configuration amp Power Interface 2 Disable P M and A C P I in motherboard CMOS PM Power Management 3 Set the option for PnP OS to No in motherboard CMOS even though Windows 2000 is a PnP OS 4 Press the F6 key at Windows 2000 CD boot It allows selecting another SCSI card if necessary Unfortunately only the four top drives are displayed and there is no way to scroll a Microsoft limitation 5 Press the F6 key at Windows 2000 CD boot Selecting Standard PC installs Windows 2000 without ACPI support VIDAR Advantage Series Installation and Operation Guide 120 Packing the film digitize
6. Connect the other end of the power cord to a grounded power outlet Note The film digitizer has an auto sensing power supply When you apply power the digitizer will detect the voltage and set itself for proper operation VIDAR CADPRO Advantage Installation Guide 41 Where to go from here How you proceed depends on how you will be connecting the digitizer to your computer m If the digitizer will be connected via a SCSI interface continue the installation using the following sequence of sections in this manual L VIDAR Drivers and Toolkit CD and Driver Information L SCSI Settings L Installing SCSI hardware L Installing SCSI Drivers L Updating the VIDAR Toolkit L Operating CADPRO Advantage m If the digitizer will be connected via a USB interface continue the installation using the following sequence of sections in this manual L1 VIDAR Drivers and Toolkit CD and Driver Information L Installing USB Drivers L Updating the VIDAR Toolkit L Operating CADPRO Advantage VIDAR CADPRO Advantage Installation Guide 42 VIDAR Advantage Series Installation and Operation Guide 43 Interfacing Guide IDAR V systems corporation Medical Imaging VIDAR Advantage Series Installation and Operation Guide 44 VIDAR Advantage Series Installation and Operation Guide 45 VIDAR device driver summary Operating system Device EZ SCSI
7. Film starts and stops during scanning 1 Increase memory allocation for scanning software Especially if scanning at high resolutions memory allocation must be sufficient to accept data stream from digitizer If scanning to disk ensure sufficient space is available on disk drive Ensure PC has enough memory available to support scanning application Close applications not needed for image acquisition from digitizer If problem persists get qualified technical help Digitizer stops scanning and PC 1 locks up Reset entire system Remember that SCSI protocol requires that digitizer be powered before PC a Turn digitizer off b Shut down PC c Turn digitizer on d After digitizer all three digitizer LEDs turn solid green turn on PC Launch scanning software on PC then try scanning again If problem persists get qualified technical help Check for correct device driver installation continued VIDAR Advantage Series Installation and Operation Guide 113 Symptom Digitizer was working properly but after installing or reinstalling the scanning application there are Toolkit errors or the digitizer is not detected on the SCSI bus Corrective action 1 If you installed a new scanning application or if you reinstalled the existing application the older Toolkit vscsi32 dll may have been installed Run the VIDAR Drivers and Toolkit Installation CD see appropriate
8. HS re Emissions Pupply lines SUDRIWdi itizers ara amp yai amp Bleceimeraiebnreaesplita ments lass A IEC 61000 3 2 other than domestic and those directly connected to the public low voltage power supply network that supplies Voltage Fluctuations buildings used for domestic purposes flicker emissions Complies IEC 61000 3 3 VIDAR Advantage Series Installation and Operation Guide 134 IEC 61000 4 4 1 kV for 1 kV for environment input output lines input output lines Surge ES US SC ditioni Mains power quality should be that of IEC 61000 4 5 a typical commercial or hospital 2 kV common mode 2 kV common mode environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 lt 5 UT gt 95 dip in Ur for 0 5 cycle 40 Ur 60 96 dip in Ur for 5 cycles 70 Ur 30 dip in Ur for 25 cycles lt 5 UT gt 95 dip in Ur for 0 5 cycle 40 Ur 60 dip in Ur for 5 cycles 70 Ur 30 dip in Ur for 25 cycles Mains power quality should be that of a typical commercial or hospital environment H the user of the Digitizer requires continued operation during power mains interruptions it is recommended that the Digitizer be powered from an uninterruptible power supply or a battery lt 5 Ur lt 5 Ur gt 95 dip in Ur gt 95 dip in Ur for 5 sec for 5 sec Power frequen
9. 100 sheet feeder can hold up PD Uc oan ji to 100 sheets at a time pees 6 Using your scanning software execute the Scan command 7 Monitor the exit tray and remove films periodically to prevent it from overfilling The exit tray can hold up to 100 films If it becomes overfilled films may be damaged or jam in the digitizer VIDAR Advantage Series Installation and Operation Guide Ss Advantage Series Other Information IDAR systems corporation Medical Imaging VIDAR Advantage Series Installation and Operation Guide 98 Preventive maintenance Cleaning the Auto Feed Roller DiaGNosTICPRO Advantage and DosiMETRYPRO Advantage only After extensive use dust or lint particles may build up on the Auto Feed Roller VIDAR recommends cleaning the roller at least twice per year Auto Feed Roller Loop a piece of light adhesive tape such as Scotch Magic Tape around your fingers with I the adhesive side out then gently pat the roller Pat and turn the roller until its entire surface is clean CAUTION m Do not use any type of cleaning solvent on the feed roller and idler wheels as this could damage these components and cause improper operation m Do not use tape with a strong adhesive such as packing tape or strapping tape m Do not use tape requiring the adhesive to be wetted such as brown paper packing tape VIDAR Advantage Series Installation and Operation Guide 99
10. O Single film feeder O Multi film feeder Not important Performance Style Reliability Size Maintenance record Image quality Packaging Documentation Software TWAIN Multi sheet capability Service and support F3 Eh L3 L3 Eh Fri rA rA Fra Fr Fm rmm N N N NI FA NY FA NM NY NY NY FA N CO Ww WwW Ww CO Ww CU CO CO CO CO CO CO PSP A A A A A A A A A A DA Important 5 a Mm oT Ui G oO oOo Ln nn aAa Un End user information End user administrative contact Name Title Company name Telephone no E mail address Fax no End user technical contact Name Title Company name Telephone no E mail address Fax no Part Number 16667 004 REV A April 2005 Printed in the USA Copyright 2005 by VIDAR Systems Corporation All rights reserved No part of this publication may be reproduced stored in a retrieval system or transmitted in any form or by any means electronic mechanical photocopying recording or otherwise without the prior written permission of VIDAR Systems Corporation Terms denoted by the trademark symbol TM or 9 used in this publication are trademarks or service marks registered in the United States or other countries NOTE The VIDAR digitizers are used to digitize radiographs X ray film or other transparent targets When the digitizer is used to digitize radiographic films the digital image is intended for use in primary secondary and over reading applications
11. OF THIS AGREEMENT INCLUDING THE SOFTWARE LICENSE AND DISCLAIMER OF SOFTWARE WARRANTY BELOW PLEASE READ THIS DOCUMENT CAREFULLY BEFORE USING THE SOFTWARE IF YOU DO NOT AGREE WITH THE TERMS AND CONDITIONS OF THIS AGREEMENT YOU SHOULD PROMPTLY RETURN THE UNUSED SOFTWARE TO THE PLACE FROM WHICH YOU OBTAINED IT AND YOU ILL BE GIVEN A FULL REFUND OF ANY LICENSE FEE THAT YOU PAID FOR SUCH SOFTWARE 1 GRANT OF LICENSE SOFTWARE USE RESTRICTIONS fidar provides this program and licenses its use You assume responsibility for the selection of the program to achieve your intended results and for the installation use and results obtained EI Click Install to begin th tion If yo nt to re i change any of your installation settings click Back Cli ancel to exit the InstallShield VIDAR Advantage Series Installation and Operation Guide 63 10 The Setup Status window will appear Wait while the files are installed iver InstallShield Wizard VIDAR Medical Film Digitizer Driver is config i new software installation Installing 12 InstallShield will indicate that installation is complete However drivers are not yet installed In the Installation Complete window activate No I will restart my computer later then click Finish 13 Shut down the computer Leave the VIDAR CD in the CD drive 14 Earlier you connected VIDAR SCSI cable part number 4270 to the digitizer Now attach the 50 pin connector
12. Update Toolkit CDs needed driver Software X vscsi32 dll Windows XP Vidar STI No Yes VIDAR Drivers and Toolkit Professional Edition Installation CD Service Pack 1 or higher Windows 2000 with Vidar STI No Yes VIDAR Drivers and Toolkit Service Pack 2 or higher Service Pack 4 or higher is required for USB connectivity About the STI Installation CD WARNING VIDAR has validated the STI driver and replacement of vscsi32 dll with all supported VIDAR digitizers on both Windows 2000 and Windows XP systems using several third party software applications No bugs or problems were identified during validation testing However you must check with your system integrator or scanning software vendor to verify that they support the replacement of VIDAR s vscsi32 dlll drivers Previously most system integrators built their scanning applications for VIDAR digitizers based on either the VIDAR SCSI Toolkit or the ActiveX control based on the SCSI Toolkit VIDAR s SCSI Toolkit was based on Adaptec s ASPI drivers However Adaptec is not supporting its ASPI drivers for Microsoft Windows 2000 or XP Therefore VIDAR developed a toolkit that takes advantage of Microsoft s Still Image Architecture STI for digital imaging devices The new VIDAR Toolkit and VIDAR STI driver support several operating system enhancements VIDAR Advantage Series Installation and Operation Guide 46 The VIDAR STI driver was d
13. as a controller or host adapter is required to enable the VIDAR film digitizer to communicate with your computer VIDAR digitizers have been validated with the Adaptec 2930CU SCSI adapter card Do not use other SCSI adapter cards Ultra wide SCSI adapters are not compatible with VIDAR film digitizers A computer equipped with a SCSI bus can communicate with multiple SCSI devices for example a film digitizer a scanner and a disk drive Each device must have a unique SCSI ID number so the computer can distinguish it from other SCSI devices Valid SCSI ID numbers range from 1 to 7 but ID 7 is reserved for the SCSI adapter Instructions for setting the film digitizer s SCSI ID number are provided in the previous chapter VIDAR Advantage Series Installation and Operation Guide 52 Warning Before you begin the installation procedure turn off all power to the computer and peripherals Connecting the SCSI cable with the power on can cause serious damage to the unit or your computer Pr caution Avant de commencer la connexion assurez vous que votre ordinateur soit bien teint Toute connexion du cable SCSI et courant pourrait gravement endommager votre num riseur ou ordinateur Advertencia Antes de continuar con la instalaci n favor de apagar su computadora y perif rico Conectando el cable SCSI con la computadora prendida puede causar dario al equipo o a su computadora Warnung Vor dem Installieren den Computer u
14. conforms to standards EN60601 1 1998 with Amendments 1 and 2 UL60601 1 2003 CAN CSA C22 2 NO 601 1 M90 with Amendments 1 and 2 This device is classified as a Group 1 Class A device for Electro Magnetic Compatibility per EN55011 1998 This device complies with standards EN55011 EN60601 1 2 2001 EN61000 3 2 1995 with Amendments A1 2 and 14 EN61000 3 3 1995 EN61000 4 2 1995 with Amendments 1 and 2 EN61000 4 3 1997 with Amendment 1 EN61000 4 4 1995 with Amendments 1 and 2 EN61000 4 5 1995 with Amendment 1 EN61000 4 6 1996 with Amendment 1 EN61000 4 8 1993 EN61000 4 11 with Amendment 1 FDA 501 k Certification CADPRO Advantage 993598 DIAGNOSTICPRO Advantage and DOSIMETRYPRO Advantage 993599 Acceptable shipping conditions Operating conditions m Temperature 25 to 65 C 13 to 150 F m Temperature 10 C to 35 C 50 F to 100 F m Relative humidity 10 to 95 non condensing m Relative humidity 20 to 80 non condensing Atmos pressure 500 to 1060hPa 18 000 to 1 200ft m Atmos pressure 697 to 1060hPa 10 000 to 1 200ft Electrical supply Voltage 100 to 240 VAC Current 1 5 to 0 75 A Frequency 47 to 63 Hz S C SS 5RA9 le o US Safety and compliance information MEDICAL EQUIPMENT WITH RESPECT TO ELECTRIC SHOCK FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL60601 1 IEC60601 1 AND CAN CSA C22 2 No 601 1 DIAGNOSTICPRO Advantage DOSIMETRYPRO Ad
15. filmdigitizer com Note that m The digitizer will report as VXR 12 for backward compatibility m You may not see the same firmware and hardware version as shown in the window above m The serial number will not report if the digitizer was just powered on Your digitizer s serial number will be different from the one shown in the example above NEXT Goto Updating the VIDAR Toolkit gt VIDAR Advantage Series Installation and Operation Guide 79 Updating the VIDAR Toolkit IMPORTANT The scanning application must be installed before performing the Toolkit DLL Update The Update Toolkit DLL feature on the VIDAR Drivers and Toolkit CD updates the scanning application with the current VIDAR Toolkit Dynamic Link Library DLL Check with the scanning application vendor to ensure that the scanning application has tested vscsi32 dll version 4 1 81 WARNING VIDAR has validated the STI driver and replacement of vscsi32 dll with all supported VIDAR digitizers on both Windows 2000 and Windows XP systems using several third party software applications No bugs or problems were identified during validation testing However you must check with your system integrator or scanning software vendor to verify that they support the replacement of VIDAR s vscsi32 dll 1 Insert the VIDAR Drivers and Toolkit Installation CD into the CD ROM The installer should launch automatically If the installer does not
16. for EMI radiated emissions VIDAR Advantage Series Installation and Operation Guide 68 Updating a previous VIDAR STI driver installation If older VIDAR STI drivers are installed on this computer you must first remove them then perform a clean driver installation To remove older drivers 1 For Windows XP Professional Start gt Control Panel gt Printers and Other Hardware Scanners and Cameras or On a Windows 2000 system Start gt Settings gt Control Panel gt Scanners and Cameras 2 VIDAR Film Digitizer should report in the Devices window Click the Remove button 3 Continue in Installing the USB drivers next VIDAR Advantage Series Installation and Operation Guide 69 Installing the USB drivers You will need m VIDAR Drivers and Toolkit Installation CD Note These instructions have been successfully tested on a wide range of Windows 2000 and Windows XP systems In rare instances you may have difficulty installing the STI driver If that happens try again after logging on as Administrator or as a user with Administrator rights 1 Be sure the digitizer is turned OFF and is NOT connected to the PC Wait until you are prompted to connect the digitizer Note If the USB cable is connected to the digitizer before installing driver software and the hardware wizard is allowed to install the drivers from the CD you will be prompted to install the CD each time the digitiz
17. ire tite eet e tac cet agere e EE EE wa qusha 111 Weights and dimensions l U U U ul ul J J J 114 SCSI information and troubleshooting guide 115 lA aKo eE eE A E 115 Digitizer and SCSI controller compatibility 115 ER 116 Troubl shootirig ioi e e ec teu tee rel e Re ce Le Ee cS Ere dette 117 Packing the film digitizer for shipment J 120 T Turn ege ET 120 2 Disconnect the power cord and SCSI or USB cable 121 3 Remove the exit tray from the digitizer base 122 4 Remove the feeder from the digitizer base 123 5 Remove the digitizer base from the stand DIAGNOSTICPRO Advantage and DOSIMETRYPRO Advantage only sss 124 6 Prepare the feeder CADPRO Advantage only 125 7 Wrap certain items arridet eeii Ea AE aein ae Ai OPAN NE iS e AA ERNE S 126 8 Pack the accessory Dos 127 9 Pack the shipping Carton 131 Electromagnetic guidance l l u J J J T T 137 i DIAGNOSTICPRO Advantage and DOSIMETRYPRO Advantage Installation Guide IDAR systems corporation Medical Imaging Unpacking 1 Look for d
18. is positioning the film Slow flashing green BUSY digitizing film None This is normal operation The LED will flash slowly while a film is being digitized Slow flashing green BUSY ejecting film None This is normal operation The LED will flash slowly while the scanning application is ejecting the film Off ERROR 1 Make certain power is supplied to the digitizer and that the digitizer power switch is on 2 If problem persists get qualified technical help see In case of difficulty later in this chapter VIDAR Advantage Series Installation and Operation Guide 110 Bottom LED power status LED state Digitizer condition Corrective action Solid green Ready None Fast flashing green ERROR DSP not Get qualified technical help see In case of communicating difficulty later in this chapter Fast flashing green ERROR unknown Get qualified technical help see In case of ballast difficulty later in this chapter Off ERROR 1 Make certain power is supplied to the digitizer and that the digitizer power switch is on 2 If problem persists get qualified technical help see In case of difficulty later in this chapter VIDAR Advantage Series Installation and Operation Guide 111 In case of difficulty How to use this section 1 Look through the left columns to find a description of the problem you are having 2 Follow the instructions in order i
19. launch automatically double click the appropriate CD drive icon under My Computer in Windows Explorer VIDAR Advantage Series Installation and Operation Guide 80 2 In the first screen click Install Software INSTALL SOFTWARE BROWSE CD CONTENTS VIEW OUR Y CONTACT US Install Software INSTALL DIGITIZER DRIVER re to install or Tone Saree nue MAIN MENU VIDAR Advantage Series Installation and Operation Guide 81 4 ADOS window will open Wait while the files are updated 3 JE scsi32Update ysdllupd exe Welcome to the VIDAR Systems Film Digitizer x Programming Library File Update Program w Searching all local drives for copies of csi32 d11 please wait Documents and Settings Administrator My Documents Save for Dan Uscsi S8cantest colortarget WUSCSI32 d11 Scantest deltaepi21i Uscsi32 d11 Scantest deltagWscsi32 d11 NScantestNp119 Uscsi WINNT twain_32 uxr Usesi32 d11 i32 d11 files found 12 C C C C C G G G G G G 5 When you see the message Exiting press any key to quit press any key or close the window by clicking the X in the upper right corner 6 Remove the VIDAR Drivers and Toolkit Installation CD IMPORTANT You will need to update the tookit DLL if you m Uninstall or reinstall the drivers m Upgrade the digitizer firmware m Replace the digitizer with another one having different firmware Reinstall the scanning application Instal
20. long their tops ends are near the top of the feeder Also note the connector on the bottom of the feeder It mates to the connector on top of the digitizer base Top ends of long screws U U Feeder rear view i i U L U qe i CH 1 a wee ibs d v o amp Long screw Short screw Long screw Feeder bottom view o 4 Connector VIDAR CADPRO Advantage Installation Guide b Place the feeder on top of the digitizer and slide the feeder connector into the digitizer connector c On the rear of the feeder tighten the three screws to secure the feeder to the digitizer base Screws m You may need to move the feeder slightly to get the screws started in the holes m Tighten each screw a few turns then repeat until all are finger tight VIDAR CADPRO Advantage Installation Guide 20 4 Install the exit tray by hooking it over the notched plastic edge just below the rollers When the exit tray is properly positioned it will be parallel to the roller shaft Roller shaft Exit tray Properly installed exit tray VIDAR CADPRO Advantage Installation Guide 21 5 Insert the film support bar legs into the two holes on the film feeder s top edge The film support bar s curved center portion must face the front of the film feeder Holes in film feeder Film support ba
21. must correct this behavior in the SCSI BIOS 1 Turn on or restart your computer When the computer boots up it will detect the SCSI Adapter When this message appears on the screen immediately press CTRL A Press Ctrl A for SCSISelect Utility The message appears for just a few seconds Note In the next steps use keyboard ARROW keys to navigate within a screen When you have selected the desired option press ENTER 2 In the Options menu see below select the Configure View Host Adapter Settings option then press ENTER AHA 2930CU at Bus 00h Device DD Would you like to configure the host adapter or run the SCSI disk utilities Select the option and press Enter Press F5 to switch between color and monochrome modes Options SCSI Disk Utilities VIDAR Advantage Series Installation and Operation Guide 57 3 Under Additional Options select SCSI Device Configuration see below Press ENTER AHA 2930U at Bus 00h Device DDb Configuration SCSI Bus Interface Definitions Host Adapter SCSI ID SCSI Parity Checking Enabled Host Adapter SCSI Termination Automatic Additional Options Boot Device Options Press lt Enter gt SCSI Device Configuration Advanced Configuration Options Press lt Enter gt lt F6 gt Reset to Host Adapter Defaults BIOS Information Interrupt IRQ Channel I O Port Address 4 You will see SCSI Device ID s 0 to 7 Initiate Sync Negotiation will be
22. presorting Light box loading head up normal reading left justified Width 8 to 14 20 3cm x 35 6cm Length 8 to 17 20 3cm x 43 2cm Thickness 0 006 to 0 010 0 15mm to 0 51mm VIDAR Advantage Series Installation and Operation Guide 85 DOSIMETRYPRO Advantage specifications Nominal Pixels Spot size DPI Line pairs Digitizing resolution 14 x17 film um per mm speed 1k x 1 25k 997 x 1211 356 71 25 1 4 12 2 sec 2k x 2 5k 1995x2422 178 1425 28 12 2 sec 4k x 5k 3990x4845 89 285 5 6 24 4 sec ACR Standard for Teleradiology Revision 35 1998 recommends 2 5 line pairs mm minimum Scan modes 32 bit mapped to 12 bit and 8 bit grayscale outputs Film sizes Width 8 to 14 20 3cm to 35 6cm manual feed Auto film feeder Length 8 to 51 t 20 3cm to 129 5cm Thickness 0 006 to 0 010 0 15mm to 0 51mm TLonger films require user support during feeding and a scanning application that handles long films Standard 25 film capacity mixed sizes no presorting Light box loading head up normal reading left justified Width 8 to 14 20 3cm x 35 6cm Length 8 to 17 20 3cm x 43 2cm Thickness 0 006 to 0 010 0 15mm to 0 51mm VIDAR Advantage Series Installation and Operation Guide 86 About films The film digitizer handles standard radiograph films up to 14 x 17 35 5cm x 43cm You can feed multiple films up to 25 and you can mix film sizes ranging from 8 x 10 20 3cm
23. section in the Installing device drivers chapter 2 If problem persists contact medtech vidar com Digitizer is not listed in Windows 1 Update to STI drivers See Computers with Control Panel under Scanners and Windows 2000 and Windows XP in the Cameras Installing device drivers chapter 2 Check all SCSI cables and connectors Digitizer is not detected on SCSI 1 Set up SCSI card See the Installing SCSI bus hardware chapter 2 Check all SCSI cables and connectors Auto Feeder does not reliably feed 1 Clean the lift roller on the Feeder films 2 CadPro only Clean the Suction Cups inside the feeder see Cleaning the Auto Feed Roller and Suction Cups in the Preventive maintenance chapter Be sure Suction Cups are not loose Computer does not recognize the digitizer when it is connected through a USB cable VIDAR firmware does not support USB 1 1 The digitizer will not be detected by the host computer if it is connected to a USB 1 1 port The digitizer must be connected to a USB 2 0 port Check the operating system to ensure the latest service pack is installed Windows XP must have Service Pack 1 or higher Windows 2000 must have Service Pack 4 or higher Microsoft does not support USB 2 0 in older service packs If the computer does not have a USB 2 0 port install a USB 2 0 adapter Adaptec www adaptec com offers several USB 2 0 adapters VIDAR has tested Adaptec s USB2Con
24. touch the diffuser material with your fingers 4 Place the lamp cartridge on a flat surface with the diffuser facing up 5 Clean the diffuser using a lint free wipe VIDAR part number 15194 Wipe in one direction only Don t wipe back and forth E Wipe in one direction only d i VIDAR Advantage Series Installation and Operation Guide 104 6 Insert the lamp cartridge into the HIS GIIGHGE basa TR TP li digitizer base mg m The diffuser must face the rear of the digitizer base z O Bulb m The end of the lamp cartridge with Cartridge the white connector goes in first Slot in lamp cartridge m Ensure the slots along the top and bottom of the cartridge align with the slides in the digitizer see T E Slide in digitizer base illustration at right m When the cartridge is nearly all the way in push it firmly until you feel the connector snap into place Slide cartridge in then push to seat connector VIDAR Advantage Series Installation and Operation Guide 105 7 Close the door on the side of the digitizer base Be sure the door closes fully 8 Turn on the digitizer The digitizer will recalibrate to the cleaned diffuser VIDAR Advantage Series Installation and Operation Guide 106 Replacing the lamp cartridge all Advantage models Follow the instructions in Cleaning the diffuser earlier in this chapter but work with the new lamp cartridge in st
25. tray SS Accessory box VIDAR Advantage Series Installation and Operation Guide 130 b Place these items on the cardboard retainer flap SCSI cable or USB cable Power cord EE Accessory box c Close the accessory box and seal it with tape VIDAR Advantage Series Installation and Operation Guide 181 9 Pack the shipping carton a Place the items shown in the shipping carton as shown here Orient the items as shown Upper foam support Note For clarity this illustration does not show the plastic bags surrounding the digitizer and feeder E Shipping carton VIDAR Advantage Series Installation and Operation Guide 132 b Close the shipping carton and seal it with tape Note If you are shipping the digitizer to VIDAR for repair or any other reason 1 Obtain a Returned Materials Authorization RMA number by sending an e mail request to medtech vidar com Include the digitizer s serial number and the reason for return 2 Write the RMA number on the box or shipping label If the shipment does not have an RMA processing will be delayed when it reaches VIDAR VIDAR Advantage Series Installation and Operation Guide 133 Electromagnetic guidance Caution Medical electrical equipment EMC Electro Magnetic Compatibility must be considered before any medical electrical equipment is installed or put into service Follow the information in the accompanying docume
26. x 24 5cm to 14 x 17 35 5cm x 43cm in one stack as long as you follow the rules below IMPORTANT Remove stickers tape staples paper clips etc from films before scanning These may cause serious film feeding problems and may damage the digitizer s internal mechanisms Failure to remove extraneous items from films will void your warranty Load films just as you would view them on a light box with thess qualifications m Films less than 8 wide will not be detected by the digitizer s film sensor and therefore will not be staged or digitized CAUTION Films smaller than 8 x 8 can fall into the internal optics box and cause damage not covered by your warranty m Do not load films more than 17 43 2cm long in the auto feeder See Digitizing a single film later in this chapter when loading films longer than 17 VIDAR Advantage Series Installation and Operation Guide 87 Basic operating instructions 1 Turn on the film digitizer and wait for all three LEDs to turn green Note Always turn on the film digitizer before turning on the computer This enables the computer to recognize the digitizer Note Several minutes are required for the digitizer to carry out its internal diagnostics and establish communications with the scanning application These activities must be completed before you can load film into the digitizer 2 Turn on the computer 3 Launch the scanning application IMPORTANT R
27. DAR CADPRO Advantage Installation Guide 24 Where to go from here How you proceed depends on how you will be connecting the digitizer to your computer m If the digitizer will be connected via a SCSI interface continue the installation using the following sequence of sections in this manual L VIDAR Drivers and Toolkit CD and Driver Information L SCSI Settings L Installing SCSI hardware L Installing SCSI Drivers L Updating the VIDAR Toolkit L Operating DIAGNOSTICPRO Advantage and DOSIMETRYPRO Advantage m If the digitizer will be connected via a USB interface continue the installation using the following sequence of sections in this manual L1 VIDAR Drivers and Toolkit CD and Driver Information L Installing USB Drivers L Updating the VIDAR Toolkit L Operating DIAGNOSTICPRO Advantage and DOSIMETRYPRO Advantage VIDAR CADPRO Advantage Installation Guide WWW W Ns aN mm SS ESA HE 7 Y 8 CADPRO Advantage Installation Guide IDAR systems corporation Medical Imaging VIDAR CADPRO Advantage Installation Guide 26 Unpacking 1 Look for damage Before unpacking the film digitizer examine the shipping carton for damage If the carton is damaged notify the shipper immediately IMPORTANT Save the carton and all packing materials If you need to ship the digitizer later you must repack it using th
28. Go to Setting up gt Setting up The digitizer must be placed on a counter or tabletop WNWIKeW Wou VIBE VIDAR Advantage Series Installation and Operation Guide 16 Identify important features Look over the digitizer base and locate the features shown in this section You will need to know where these features are when you assemble and operate the digitizer in later chapters Film feeder connector n MEE LL J y Film feeder screw holes Digitizer base top view from rear SCSI port ON OFF switch a rocker switch SCSI ID switch SCSI termination switch Press to turn power ON Press 0 to turn power OFF USB port Power entry module lt Digitizer base rear view Power entry module VIDAR CADPRO Advantage Installation Guide Assemble the digitizer 1 Place the stand on a solid flat surface such as a table 2 Place the digitizer base on the stand as shown below Tighten the thumbscrews on the bottom of the stand to secure it to the digitizer Front view Side view Thumbscrews 2 Longer portion toward front Vent in bottom of base must be over cutout in stand VIDAR CADPRO Advantage Installation Guide 3 Attach the feeder to the digitizer a Note the three screws protruding from the bottom of the feeder These mate to holes in the digitizer base Two of the screws are very
29. Installation and operation guide for DIAGNOSTICPRO Advantage DOSIMETRYPRO Advantage CADPRO Advantage x j with 50 sheet SmartFeeder V and 100 sheet SmartFeeder Serial Numbers 300000 to 339999 Warranty information for VIDAR film digitizers Instructions 1 Cut along dotted line at left 2 Fill in 3 Make a copy for your records 4 E mail orfax to VIDAR System integrator information System integrator administrative contact Name Company name E mail address System integrator technical contact Name Company name E mail address Shipping information Company name Street address City State Country Telephone no systems corporation Medical Imaging 365 Herndon Parkway Herndon VA 20170 USA Phone 1 703 471 7070 e Fax 1 703 471 7665 E mail medtech vidar com Title Telephone no Fax no Title Telephone no Fax no Postal code Fax no Product information What product are you registering Select only one L1 VXR 12 Plus L1 DiagnosticPRO L1 DiagnosticPRO plus What is the product s serial number Whatis the SCSI ID Where is this film digitizer installed Or where will it be installed L1 Hospital O Office L1 Clinic Why did you purchase this product Select all that apply and rate importance Feature nuuuuuuuuuuiui Price L1 Other L1 TeleradPRO L1 SIERRA O Single film feeder L1 Multi film feeder L1 SIERRA plus
30. Place the film support bar in a padded bag Seal the bag with tape VIDAR Advantage Series Installation and Operation Guide 127 8 Pack the accessory box DIAGNOSTICPRO Advantage and DosIMETRYPRO Advantage only a Place these items in the accessory box in the order shown here Orient the items as shown When the cardboard retainer is in place the two runners on the exit tray should protrude through the two slots in the retainer Cardboard retainer When cardboard retainer a is in place exit tray runners must protrude through slots in retainer a Exit tray Note For clarity this illustration does not show the protective plastic bag surrounding the exit tray suene Accessory box VIDAR Advantage Series Installation and Operation Guide 128 b Place these items on the cardboard retainer Film support bar in padded bag bag not shown SCSI cable or USB cable e Power cord Exit tray runners mista BW protrude through ES cardboard retainer slots UE Accessory box as specified at time of order c Close the accessory box and seal it with tape VIDAR Advantage Series Installation and Operation Guide 129 CADPRO Advantage only a Place these items in the accessory box in the order shown here Orient the items as shown Cardboard retainer Exit tray Note For clarity this illustration does not show the protective plastic bag surrounding the exit
31. THAT YOU PAID FOR SUCH SOFTWARE 1 GRANT OF LICENSE SOFTWARE USE RESTRICTIONS idar provides this program and licenses its use You assume responsibilty for the selection of the program to achieve your intended results and for the installation use and results obtained EI VIDAR Advantage Series Installation and Operation Guide 73 10 In the Ready to Install the Program window click Install VIDAR Medical Film Digitizer Driver InstallShield Wizard Ready to Install the Program installation InstallShield 11 The Setup Status window will appear Wait while the files are installed VIDAR Medical Film Digitizer Driver InstallShield Wizard Setup Status InstallShield VIDAR Advantage Series Installation and Operation Guide 74 13 InstallShield will indicate that installation is complete However drivers are not yet installed In the Installation Complete window not shown activate No I will restart my computer later then click Finish 14 Connect the digitizer to the computer s USB 2 0 port with a USB 2 0 cable 15 Shortly after connecting the digitizer to the computer the New Hardware Found Wizard window will appear m Ifthe computer is running Windows XP activate the Install the software automatically option then click Next m Ifthe computer is running Windows 2000 activate the Install from a list or specific location option then click Next I
32. amage Before unpacking the film digitizer examine the shipping carton for damage If the carton is damaged notify the shipper immediately IMPORTANT Save the carton and all packing materials If you need to ship the digitizer later you must repack it using the original wire ties plastic bags foam supports and cartons Instructions for repacking are provided in Appendix Packing the film digitizer for shipment 2 Unpack the shipping carton Remove items from the shipping carton in the order shown below Upper foam support Note For clarity this illustration does not show a the plastic bags surrounding the digitizer and feeder Leave in box Accessory box lower foam support Kee Shipping carton 3 Unpack the accessory box a Remove these items from the accessory box t Film support bar in padded bag O bag not shown ch s c Power cord c 2 t SCSI cable e Documentation in or USB cable padded bag bag not shown EES Accessory box SCSI kit as specified at time of order 11 Note For clarity this illustration does not show the plastic bag surrounding the exit tray b Then remove these items from the accessory box Caraboard retainer a Exit tray PERE SAH Accessory box 4 Unwrap a Unwrap all items b Remove the foam block from the rear of the digitizer base IMPORTANT Save the carto
33. ar before packing the feeder otherwise the feeder will not fit into the foam support Screws b Loosen the three thumbscrews securing the feeder to the digitizer base c Lift the feeder up and off the digitizer base VIDAR Advantage Series Installation and Operation Guide 124 5 Remove the digitizer base from the stand DiaGNosTICPRO Advantage and DosiMETRYPRO Advantage only Loosen the thumbscrews on the bottom of the stand then lift the digitizer base from the stand Thumbscrews 2 VIDAR Advantage Series Installation and Operation Guide 125 6 Prepare the feeder CADPRO Advantage only a Secure the film feeder Suction Cup Arm using the foam wedge as shown below Front of feeder Wedge Then slide wedge into opening and under bar with other hand DD Gently lift and hold suction cup arm with one hand Bottom of feeder Slide slot in wedge over metal edge on bottom of feeder Feeder with wedge inserted should look like this b Place the film feeder in a plastic bag Seal the bag with tape VIDAR Advantage Series Installation and Operation Guide 126 7 Wrap certain items a Insert the foam block in the rear of the digitizer base b Place the digitizer base in a plastic bag Seal the bag with tape c Place the film feeder in a plastic bag Seal the bag with tape d Place the exit tray in a plastic bag Seal the bag with tape e
34. ase wait while the wizard installs the software A VIDAR Film Digitizer SCH Vidarcpl cpl To CSwIND WS system32 19 When the wizard is finished installing the driver click Finish Found New Hardware Wizard Completing the Found New Hardware Wizard The wizard has finished installing the software for e VIDAR Film Digitizer 39 Click Finish to close the wizard VIDAR Advantage Series Installation and Operation Guide 77 20 Verify that the digitizer drivers are fully installed by checking the device properties Use the path for your operating system m For Windows M XP Professional Start gt Control Panel gt Printers and Other Hardware Scanners and Cameras m For Windows 2000 Professional Start gt Settings gt Control Panel Scanners and Cameras 21 In the device window right click Vidar Film Digitizer then click Properties in the pop up menu VIDAR Advantage Series Installation and Operation Guide 78 22 In the VIDAR Film Digitizer Properties window click the VIDAR Properties tab VIDAR Film Digitizer Properties Color Management Film Digitizer Vendor Identification VIDAR Model Name VXR 12 Serial Number 300143 Firmware Version 45 9 Hardware Version 112 Manufacturer VIDAR Systems Corporation 365 Herndon Parkway Herndon Virginia 20170 USA phone 1 703 471 7070 toll free 800 471 7226 fax 1 703 471 7665 info filmdigitizer com http www
35. bances are controlled The customer or the user of the Digitizer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the Digitizer as recommended below according to the maximum output power of the communications equipment Rated maximum output Separation distance according to frequency of transmitter m power of transmitter W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz d 1 2VP d 1 24P d 2 3VP 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 2 1 2 2 3 10 3 7 3 7 7 4 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people VIDAR Advantage Series Installation and Operation Guide 137 Disposal of Waste Equipment by Users in the European Union This symbol on the product indicates that this product must not be disposed of as unsorted municipal waste Instea
36. ch to 0 7 8 9 A B C Dor F CAUTION Do not force switch rotation Do not use a large screwdriver to rotate the switch VIDAR Advantage Series Installation and Operation Guide 49 If necessary terminate the SCSI bus The film digitizer has a single SCSI port connector and must be connected directly to the computer or connected at the end of a SCSI chain In this configuration the digitizer must provide termination for the SCSI bus The digitizer was set for SCSI bus termination at the factory VIDAR strongly recommends that you not change the factory SCSI termination setting Note If you cannot install the digitizer at the end of a SCSI chain please contact VIDAR Technical Support for instructions 1 703 471 7070 or medtech vidar com Note When the film digitizer is at the end of the SCSI chain the SCSI termination switch must be set for SCSI termination Consult your SCSI controller manual regarding proper use of termination If you determine that the film digitizer s SCSI termination is not set for internal termination left picture below here is how to change the setting to deactivate internal termination 1 Locate the SCSI termination switches on the back of the digitizer 2 Push both switches as shown below Switches Switches pushed pushed this way this way Factory recommended Not recommended termination switch setting VIDAR Advantage Series Installation and Operation Guid
37. cy Power frequency magnetic fields 50 60 Hz should be at levels characteristic of a magnetic field P SAM typical location in a typical commercial IEC 61000 4 8 or hospital environment NOTE Ur is the a c mains voltage prior to application of the test l vel VIDAR Advantage Series Installation and Operation Guide 135 Guidance and manufacturer s declaration electromagnetic immunity Table 204 The Digitizer is intended for use in the electromagnetic environment specified below The customer or end user of the Digitizer should assure that it is used in such an environment IEC 61000 4 3 80 MHz to 2 5 GHz Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Portable and mobile RF communication equipment should be used no closer to any part of the Digitizer including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance Conducted RF 3 Vrms 3V d 1 2NP IEC 61000 4 6 150 kHz to 80 MHz Radiated RF 3V m 3 V m d 1 24 P 80MHzto 800 MHz d 2 34 P 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and dis the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey s
38. d it is your responsibility to dispose of your waste equipment by handing it over to a designated collection point for recycling of waste electrical and electronic equipment The separate collection and recycling of your waste equipment will help to conserve natural resources and ensure that it is recycled in a manner that protects human health and the environment For more information about where you can dispose of your waste equipment for recycling please contact your local city office your household waste disposal service or the shop where you purchased the product Note on WEEE 6799D237 VIDAR Advantage Series Installation and Operation Guide 138 VIDAR Advantage Series Installation and Operation Guide VIDAR systems corporation Medical Imaging 365 Herndon Parkway Herndon VA 20170 USA Phone 1 703 471 7070 Main fax 1 703 471 1165 Web www filmdigitizer com Technical Support Phone 1 703 471 7070 1 800 471 7226 Fax 1 703 471 7665 medtech vidar com
39. e 50 Connect the SCSI cable to the digitizer Use only a shielded 1 8 meter 6 ft VIDAR SCSI cable part number 4270 This cable has a 68 pin high density connector that attaches to the digitizer and a 50 pin connector that attaches to the SCSI adapter in the computer to be installed later Use of SCSI cables from other sources may violate safety and emissions compliances Note Attach or remove the SCSI cable carefully to avoid bending the connector pins Insert the connector straight into the SCSI port Inserting the connector at an angle may damage the pins If the pins or cable are damaged improper SCSI operation may result Note Before connecting the SCSI cable be certain that the SCSI cable isn t connected to anything else and the digitizer power supply is unplugged 1 Carefully attach the 68 pin high density SCSI connector to the film digitizer s SCSI port located on the rear of the digitizer Apply even pressure to avoid bending SCSI connector pins Do not force the connectors together Do not connect the other end of the cable at this time WARNING If the SCSI connector has bent pins it will damage the SCSI port in the digitizer Do not attempt to straighten bent pins Discard any SCSI cable with bent pins and replace it with a new cable NEXT Goto Installing SCSI hardware gt VIDAR Advantage Series Installation and Operation Guide 51 Installing SCSI hardware A SCSI card also known
40. e original wire ties plastic bags foam supports and cartons Instructions for repacking are provided in Appendix Packing the film digitizer for shipment VIDAR CADPRO Advantage Installation Guide 27 2 Unpack the shipping carton Remove items from the shipping carton in the order shown below E Upper foam support Note For clarity this illustration does not show the plastic bags surrounding the digitizer and feeder Leave in box Accessory box lower foam support Re Shipping carton VIDAR CADPRO Advantage Installation Guide 28 3 Unpack the accessory box a Remove these items from the accessory box E S amp Power cord A 9 A t SCSI cable PA J Documentation or USB cable package EEN Accessory box SCSI kit optional VIDAR CADPRO Advantage Installation Guide 29 Note For clarity this illustration does not show the plastic bag surrounding the exit tray b Then remove these items from the accessory box f Cardboard retainer a Exit tray EEN Accessory box VIDAR CADPRO Advantage Installation Guide 30 4 Unwrap a Unwrap all items b Remove the foam block from the rear of the digitizer base VIDAR CADPRO Advantage Installation Guide 31 d Remove the wedge from the feeder as shown below Front of feeder with other hand DD Gently lift and hold suction cup arm with one
41. eder These mate to holes in the digitizer base Two of the screws are very long their tops ends are near the top of the feeder Also note the connector on the bottom of the feeder It mates to the connector on top of the digitizer base Top ends of long screws U U Feeder rear view i i U L U qe i e i wee tbs d v amp Long screw Short screw Long screw Feeder bottom view o 4 Connector VIDAR CADPRO Advantage Installation Guide 38 b Place the feeder on top of the digitizer and slide the feeder connector into the digitizer connector c On the rear of the feeder tighten the three screws to secure the feeder to the digitizer base Screws m You may need to move the feeder slightly to get the screws started in the holes m Tighten each screw a few turns then repeat until all are finger tight VIDAR CADPRO Advantage Installation Guide 39 3 Install the exit tray by hooking it over the notched plastic edge just below the rollers When the exit tray is properly positioned it will be parallel to the roller shaft Roller shaft Exit tray Properly installed exit tray VIDAR CADPRO Advantage Installation Guide 40 Connect the power cord 1 Connect the power cord to the digitizer s Power Entry Module PEM located on the rear of the digitizer 2
42. emove feeder 3 Inspect electrical connector and screws on bottom of feeder Fast flashing green ERROR unknown feeder attached Check feeder 1 Turn off digitizer 2 Remove feeder 3 Inspect electrical connector and screws on bottom of feeder Slow flashing green BUSY film is staging None This is normal operation Fast flashing green ERROR film staging problem Eject film by clicking the eject button in the scanning application VIDAR Advantage Series Installation and Operation Guide 109 Center LED digitizer status LED state Solid green Digitizer condition Ready Corrective action None The digitizer is ready to accept commands from the scanning application Slow flashing green BUSY normalizing None This is normal operation The LED will flash slowly during initial calibration after power is applied and when the scanning application initiates a calibration Slow flashing green BUSY tracking ADC None This is normal operation The LED will flash periodically when the digitizer is idle Automatic Digitizer Calibration ADC is independent of the scanning application Slow flashing green BUSY staging film None This is normal operation The LED will flash slowly while the digitizer is staging the film Slow flashing green BUSY Positioning film None This is normal operation The LED will flash slowly while the scanning application
43. emove stickers tape staples paper clips etc from films before scanning These may cause serious film feeding problems Failure to remove extraneous items from films will void your warranty 4 If you will be digitizing films 17 43 2cm or longer raise the film support bar 5 Load the film How you load the film depends on whether you are digitizing one film or multiple films See the next pages for details CAUTION X ray images displayed on a computer monitor are representative only A variety of factors influence image density and size including monitor luminance and calibration window and level settings zoom level and dimensional and grayscale inaccuracies resulting from build up of tolerances in the digitizer display board and computer For this reason special precautions must be exercised e g calibrating the software ruler and computer monitor when taking measurements from digitized images Please refer to the user s manual for your digitizing software for more information VIDAR Advantage Series Installation and Operation Guide 88 Digitizing a single film Options for digitizing single films There are two ways to scan a single film m By manually inserting the film behind the blue bar as described in this section or m Bytreating the single film as a batch of one if the scanning application supports this approach for instructions see Scanning multiple films next sectio
44. eps 4 5 and 6 VIDAR recommends replacing the lamp cartridge once per year as part of the scheduled preventive maintenance on the film digitizer Preventive maintenance is not covered under warranty Only failures during the published warranty period are covered by warranty replacement Failures must be confirmed by running VIDAR s ScanInfo 5 2 diagnostic software The software generates a report showing light values indicative of lamp performance You can request a copy of the ScanInfo program from VIDAR Technical Support Contact VIDAR Technical Support if you have any questions about installing or using your VIDAR film digitizer Phone 1 800 471 SCAN 1 800 471 7226 1 703 471 7070 outside the U S E mail medtech vidar com VIDAR Advantage Series Installation and Operation Guide 107 Troubleshooting Reading the LEDs Three LEDs on the front of the digitizer provide operating information LEDs Location Function Normal state Top LED Feeder status Solid green Ga Solid green but slow flashing e Center LED Digitizer status when digitizer is busy Bottom LED Power status Solid green Tables on the next three pages explain all LED states VIDAR Advantage Series Installation and Operation Guide 108 Top LED feeder status LED state Digitizer condition Corrective action Solid green Ready None The feeder is ready to operate Off ERROR no feeder 1 Turn off digitizer attached 2 R
45. er is moved to another USB port To correct this the driver software will need to be installed twice The first time will be done as detailed in steps 1 through 19 the second time will follow the same method with one exception the installation wizard will recognize that the software has been installed and will give two options prior to installation Remove or Repair Select Repair and continue with the process After completing this process you will no longer need the CD if the digitizer is moved to another USB port VIDAR Advantage Series Installation and Operation Guide 70 Turn on the PC Insert the VIDAR Drivers and Toolkit Installation CD into the CD ROM The installer should launch automatically If the installer does not launch automatically double click the appropriate CD drive icon under My Computer in Windows Explorer 4 In the first screen click Install Software gt CONTENTS Ww OUR WEBSITE VIDAR Advantage Series Installation and Operation Guide 71 5 In the next screen click Install Digitizer Driver Install Software STALL DIGITIZER DRIVER n e to install or AB UPDATE TOOLKIT DLL MAIN MENU 6 You will see a message that the InstallShield Wizard is starting 7 In the InstallShield Welcome screen click Next VIDAR Medical Film Digitizer Driver InstallShield Wizard VIDAR Medical Film Digitizer Driver Installation Program Welcome to the VIDAR Medical Fi
46. er mm speed 6720 x 5376 44 570 11 50 sec Scan modes 32 bit mapped to 16 bit and 12 bit grayscale outputs Film sizes Fixed width 24cm Length 18cm or 30cm Thickness 0 006 to 0 010 0 15mm to 0 51mm Auto film feeder 50 or 100 film capacity mixed sizes Light box loading head up normal reading VIDAR Advantage Series Installation and Operation Guide 94 About films The film digitizer handles standard 18cm x 24cm and 24cm x 30cm mammography films You can feed multiple films up to 50 or 100 depending on the feeder and you can mix film sizes IMPORTANT Remove stickers tape staples paper clips etc from films before scanning These may cause serious film feeding problems and may damage the digitizer s internal mechanisms Failure to remove extraneous items from films will void your warranty To prepare films for digitizing m Stack each group of films according to the sequence specified by the CAD system manufacturer A typical sequence is shown below m Place a separator in each group Use separators approved by the CAD system manufacturer Separator Sheet Separator Sheet 18cm long x 24cm wide film stack 30cm long x 24cm wide film stack VIDAR Advantage Series Installation and Operation Guide 95 Operating instructions 1 Turn on the film digitizer and wait for all three LEDs to turn green Note Always turn on the film digitizer before turning on the computer T
47. eries Installation and Operation Guide 67 Installing USB drivers Note VIDAR digitizers with serial numbers 310000 and higher have USB 2 0 functionality Some older models have USB ports but require factory hardware upgrades to operate as USB devices Contact VIDAR for details Note Do not connect the digitizer to the computer until you are prompted during the driver installation process USB interface requirements m Thecomputer must have a USB 2 0 port If you are not certain that the computer has a USB 2 0 port contact its manufacturer Be aware that the manufacturer may tell you that a USB 2 0 peripheral can be used with a USB 1 1 port That combination may be possible for other devices but the operating system will not recognize the VIDAR digitizer if it is connected to a USB 1 1 port m Ifthe computer does not have a USB 2 0 port you may install a USB 2 0 adapter Adaptec www adaptec com offers several USB 2 0 adapters however VIDAR has only tested Adaptec Model USB2Connect part number AUA 2000 m The computer s operating system must be upgraded to the following level e Windows M XP Service Pack 1 or higher e Windows 2000 Service Pack 4 or higher Microsoft does not support USB 2 0 on earlier service packs m The digitizer must be connected to the computer with a USB 2 0 cable VIDAR provides an approved USB 2 0 cable part number 4618 with all USB digitizer orders this cable has been tested and approved
48. esigned to provide compatibility between VIDAR medical film digitizers and Windows 2000 and XP Features of this driver include m Installation setup wizard for easy and consistent installation m Scanners and Cameras Control Panel which provides a common interface for still image devices m No need to rely on unsupported ASPI layer drivers from Adaptec for SCSI adapters The VIDAR STI driver works with the new Toolkit vscsi32 dlI The Toolkit was developed specifically to support the STI architecture m tisadirect replacement for previous vscsi32 dlls m Norecompiles are necessary m Existing applications are easily configured to work with the Toolkit NEXT f connecting digitizer via SCSI go to SCSI settings gt or If connecting digitizer via USB go to Installing USB drivers gt VIDAR Advantage Series Installation and Operation Guide 47 SCSI settings Warning Before you begin the installation procedure turn off all power to the computer and peripherals Connecting the SCSI cable with the power on can cause serious damage to the unit or your computer Pr caution Avant de commencer la connexion assurez vous que votre ordinateur soit bien teint Toute connexion du cable SCSI et courant pourrait gravement endommager votre num riseur ou ordinateur Advertencia Antes de continuar con la instalaci n favor de apagar su computadora y perif rico Conectando el cable SCSI con la compu
49. hand Bottom of feeder Gently lower suction cup arm IMPORTANT Save the carton and all packing materials If you need to ship the digitizer later you must repack it using the original wire ties plastic bags foam supports and cartons Instructions for repacking are provided in Appendix Packing the film digitizer for shipment VIDAR CADPRO Advantage Installation Guide 32 5 Identify the components Item Digitizer base Exit tray 50 sheet film feeder or 100 sheet film feeder Power cord 115V p n 4043 001 or 230V p n 2104 006 Important Safety Information card on film feeder DID DmDmDrpnrs Quick setup card Digitizer base Exit tray D or tm EN 50 sheet feeder or 100 sheet feeder Y Other items see footnote O Adaptec 2930 SCSI kit p n 3053 and SCSI cable 6 foot MMD 50 to SCSI III 68 pin male p n 4270 Ll USB cable 6 foot p n 4618 You will receive either a SCSI kit and SCSI cable or a USB cable as specified at time of order VIDAR CADPRO Advantage Installation Guide 33 6 If anything is missing Immediately contact your VIDAR supplier 7 Activate your product warranty Complete the warranty information customer survey form at the back of this manual and mail it to VIDAR Systems Corp If your digitizer needs service this information should be on file at VIDAR Contact VIDAR Technical Support if you have any questio
50. his enables the computer to recognize the digitizer Note Several minutes are required for the digitizer to carry out its internal diagnostics and establish communications with the scanning application These activities must be completed before you can load film into the digitizer 2 Turn on the computer 3 Launch the scanning application IMPORTANT Remove stickers tape staples paper clips etc from films before scanning These may cause serious film feeding problems Failure to remove extraneous items from films will void your warranty CAUTION X ray images displayed on a computer monitor are representative only A variety of factors influence image density and size including monitor luminance and calibration window and level settings zoom level and dimensional and grayscale inaccuracies resulting from build up of tolerances in the digitizer display board and computer For this reason special precautions must be exercised e g calibrating the software ruler and computer monitor when taking measurements from digitized images Please refer to the user s manual for your digitizing software for more information continued VIDAR Advantage Series Installation and Operation Guide 96 4 Hold the stack in front of you as you would view it on a light box 5 Slide the stack into the feeder between the guides Do not overfill the feeder m The 50 sheet feeder can hold up to 50 sheets at a time m The
51. hould be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol 2 people Notel At80MHz and 800MHz the higher frequency range applies Note2 These guidelines may not be applicable in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and reorienting or relocating the Digitizer a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the Digitizer is used exceeds the applicable RF compliance level above the Digitizer should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as b Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m VIDAR Advantage Series Installation and Operation Guide 136 Recommended separation distances between portable and mobile RF communications equipment and the Digitizer Table 206 The Digitizer is intended for use in an electromagnetic environment in which radiated RF distur
52. ies in the pop up menu 2 In the VIDAR Film Digitizer Properties window click the VIDAR Properties tab VIDAR Advantage Series Installation and Operation Guide 65 VIDAR Film Digitizer Properties General i Color Management Film Digitizer Vendor Identification VIDAR Model Name VXR 12 Serial Number 300143 Firmware Version 45 9 Hardware Version 112 Manufacturer VIDAR Systems Corporation 365 Herndon Parkway Herndon Virginia 20170 USA phone 1 703 471 7070 toll free 800 471 7226 fax 1 703 471 7665 info filmdigitizer com http www filmdigitizer com Note that m The digitizer will report as VXR 12 for backward compatibility m You may not see the same firmware and hardware version as shown in the window above m The serial number will not report if the digitizer was just powered on Your digitizer s serial number will be different from the one shown in the example above Note The Windows New Hardware Wizard will force you to reinstall the drivers if any of the following occur after driver installation Q The digitizer is moved to another SCSI card Q The digitizer firmware is changed The digitizer is switched to another SCSI ID The driver files are placed on the local hard disk during digitizer installation so that the driver CD should not be needed VIDAR Advantage Series Installation and Operation Guide 66 NEXT Goto Updating the VIDAR Toolkit gt VIDAR Advantage S
53. information and offers troubleshooting guidance for common problems Digitizer and SCSI controller compatibility VIDAR film digitizers have been validated with the Adaptec 2930CU SCSI adapter card Do not use other SCSI adapter cards Ultra wide SCSI adapters are not compatible with VIDAR film digitizers VIDAR Advantage Series Installation and Operation Guide 116 SCSI cable Use VIDAR part number 4270 only Connecting the cable Note Attach or remove the SCSI cable carefully to avoid bending the connector pins Insert the connector straight into the SCSI port Inserting the connector at an angle may damage the pins If the pins or cable are damaged improper SCSI operation may result Note Before connecting the SCSI cable be certain that the PC and digitizer are OFF 1 Carefully attach the SCSI connector to the digitizer s SCSI port Apply even pressure to avoid bending SCSI connector pins Do not force the connectors together 2 Connect the other end of the cable to the SCSI port on the computer WARNING If the SCSI connector has bent pins it will damage the SCSI port in the digitizer Do not attempt to straighten bent pins Discard any SCSI cable with bent pins and replace it with a new cable VIDAR Advantage Series Installation and Operation Guide 117 Troubleshooting m Problem A BIOS banner does not appear for an adapter that has a BIOS chip Or A BIOS banner appears but there is a
54. l the digitizer and scanning application on another computer NEXT Go to the operating chapter for your digitizer gt VIDAR Advantage Series Installation and Operation Guide 82 VIDAR Advantage Series Installation and Operation Guide Advantage Series Operating Guide IDAR V systems corporation Medical Imaging VIDAR Advantage Series Installation and Operation Guide 84 Operating DIAGNOSTICPRO Advantage and DosiMETRYPRO Advantage About digitizers and film sizes This section describes important differences between the DIAGNOSTICPRO Advantage and DOSIMETRYPRO Advantage DIAGNOSTICPRO Advantage specifications Nominal Pixels Spot size DPI Line pairs Digitizing resolution 14 x17 film um per mm speed 1kx1 25k 1008x1124 340 75 1 12 2 sec 2k x 2 5k 2002 x 2431 170 150 3 12 2 sec 4k x 5k 3990 x 4845 85 300 6 24 4 sec Mammography film 18cm x 24cm 4k x 5k 4104 x 5472 44 570 11 27 4 sec ACR Standard for Teleradiology Guidelines Revision 35 1998 recommends 2 5 line pairs mm minimum Scan modes Film sizes manual feed Auto film feeder 32 bit mapped to 12 bit and 8 bit grayscale outputs Width 8 to 14 20 3cm to 35 6cm Length 8 to 51 t 20 3cm to 129 5cm Thickness 0 006 to 0 010 0 15mm to 0 51mm TLonger films require user support during feeding and a scanning application that handles long films Standard 25 film capacity mixed sizes no
55. lm Digitizer Driver for Wind D and program nstallation NOT CONNECT THE DIGITIZER TO YOUR COMPUTER UNTIL AFTER THE ALLATION IS COMPLETE PLEASE FOLLOW THE ON SCREEN INSTRUCTIONS Click Met to continue VIDAR Advantage Series Installation and Operation Guide 72 8 If you are updating from older VIDAR SCSI drivers you will see the message below Click Yes CINE E K i If you proceed with this installation existing VIDAR drivers will be removed Do you wish to continue 9 In the License Agreement window activate the I accept the terms of the license agreement then click Next END USER SOFTWARE LICENSE AGREEMENT FOR VIDAR SYSTEMS CORPORATION a SOFTWARE THIS IS LEGALLY BINDING AGREEMENT BETWEEN YOU EITHER AN INDIVIDUAL OR A SINGLE ENTITY AND VIDAR SYSTEMS CORPORATION VIDAR FOR THE VIDAR SOFTWARE AND ASSOCIATED MEDIA AND PRINTED MATERIALS AND MAY INCLUDE ONLINE OR ELECTRONIC DOCUMENTATION SOFTWARE PRODUCT OR SOFTWARE BY INSTALLING COPYING OR OTHERWISE USING VIDAR S SDFTWARE PRODUCT DU ARE AGREEING TO BE BOUND BY THE TERMS AND CONDITIONS OF THIS AGREEMENT INCLUDING THE SOFTWARE LICENSE AND DISCLAIMER OF SOFTWARE WARRANTY BELOW PLEASE READ THIS DOCUMENT CAREFULLY BEFORE USING THE SOFTWARE IF YOU DO NOT AGREE WITH THE TERMS AND CONDITIONS DF THIS AGREEMENT YOU SHOULD PROMPTLY RETURN THE UNUSED SOFTWARE TO THE PLACE FROM WHICH YOU OBTAINED IT AND YOU ILL BE GIVEN A FULL REFUND OF ANY LICENSE FEE
56. maging Internet www filmdigitizer com E mail medtech vidar com Authorized representative Emergo Europe Phone 31 1114 11515 in Europe P O Box 149 Fax 31 1114 10104 4300 AC Zierikzee Email info emergogroup com The Netherlands VIDAR Advantage Series Installation and Operation Guide Radio Frequency Emissions This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area can cause harmful interference in which case the user will be required to correct the interference at his or her own expense Product compliance testing was conducted using VIDAR shielded cables Modifications to the digitizer or the VIDAR shielded cables or the use of cables other than those available from VIDAR could void the user s authority to operate the equipment CE Declarations VIDAR Systems Corporation declares the product is classified as a Class I medical device per Annex IX Rule 10 and are in conformity with the essential requirements and provisions of Council Directive 93 42 EEC and
57. n and all packing materials If you need to ship the digitizer later you must repack it using the original wire ties plastic bags foam supports and cartons Instructions for repacking are provided in Appendix Packing the film digitizer for shipment 5 Identify the components Y Item L Digitizer base Ll Exittray Ll Film feeder Ll Film support bar Ll Stand Ll Power cord 115V p n 4043 001 or 230V p n 2104 006 Ll Important Safety Information card on film feeder Ll Quick setup card DEI YS q ns c X L d Digitizer base Film feeder WK Film support bar Stand Exit tray Y Other items see footnote O Adaptec 2930 SCSI kit p n 3053 and SCSI cable 6 foot MMD 50 to SCSI III 68 pin male p n 4270 Ll USB cable 6 foot p n 4618 You will receive either a SCSI kit and SCSI cable or a USB cable as specified at time of order 6 If anything is missing Immediately contact your VIDAR supplier 7 Activate your product warranty Complete the warranty information customer survey form at the back of this manual and mail it to VIDAR Systems Corp If your digitizer needs service this information should be on file at VIDAR Contact VIDAR Technical Support if you have any questions about installing or using your VIDAR film digitizer Phone 1 800 471 SCAN 1 800 471 7226 1 703 471 7070 outside the U S E mail medtech vidar com NEXT
58. n error message Host Adapter Configuration Error This error at system boot indicates the PCI interrupt has not been configured correctly Do not continue with installation until this problem has been resolved Check these items in the system CMOS BIOS they will usually be found under PCI or PnP configuration L PCI slot should be enabled L Bus mastering should be enabled L1 IRQ triggering should be set to Level L There should be enough IRQs available for PCI PnP Also try installing the adapter in another PCI slot m Problem Driver is installed but Windows fails to run when you reboot There is probably a resource conflict m Problem System locks up with memory errors or no video There could be a BIOS address conflict Contact the BIOS manufacturer and update the system BIOS to the most recent version VIDAR Advantage Series Installation and Operation Guide 118 m Problem System locks up or freezes when the driver loads O Check for IRQ conflicts L Check the system CMOS for any PCI slot or IRQ options that enable you to reserve an IRQ for PCI PnP L Try the adapter in another PCI slot primary or first slot preferred m Problem You installed a PnP adapter in Windows and the PnP does not load the driver Disable the PnP on the adapter Check adapter documentation for instructions m Problem Digitizer is not listed in Windows Control Panel gt Scanners and Cameras Windows
59. n in this manual The digitizer cannot feed films smaller than 8 wide x 8 long 20 3cm wide x 20 3cm long 1 Hold the film in front of you as you would view it on a light box 2 Align the left edge of the film with the film guide on the left of the feeder 3 Place the film into the slot BEHIND the blue bar Yos The blue bar has a lip and a groove on the top edge of the digitizer where the feeder is attached Place w a single film directly behind the blue bar he THAT A When the digitizer senses the film it automatically stages the film in preparation for digitizing The film will first move down then up This is normal operation CAUTION If the film is not staged properly do not manually pull the film out of the digitizer Instead use the scanning application s EJECT command to remove the film continued VIDAR Advantage Series Installation and Operation Guide 89 Note When a film has been staged it is in the digitizer s light path The ADC Automatic Digitizer Calibration feature requires that the light path be clear of the film for proper background calibration of the digitizer The digitizer will automatically adjust the film s position to properly proceed with ADC Depending on the location of the film s leading or trailing edge and the length of time the film has been blocking the light path the film will be 8 pushed up or b pushed down 4 Using your scanning software e
60. n the Corrective action column If the one corrective action doesn t solve the problem then carry out the next corrective action 3 When the instruction is Get qualified technical help then a Contact your system integration specialist the company or person that installed your VIDAR film digitizer b If your system integration specialist isn t available then contact VIDAR Customer Support medtech vidar com Please record system information and digitizer serial number before calling and have it available when calling Tip Check www filmdigitizer com for current troubleshooting information tools and software updates Symptom Corrective action Can t turn digitizer on or off Can t The on off switch is located on the rear of the digitizer find power switch body Look at the digitizer from the rear to see the switch Streaks in image Clean lamp diffuser see Cleaning the diffuser in the Cleaning and maintaining chapter earlier in this manual continued VIDAR Advantage Series Installation and Operation Guide 112 Symptom Bottom LED power status is off Corrective action Ensure wall outlet is providing AC power obtain another electrical device known to be working and plug it into that AC wall outlet If other device doesn t work AC power is not available at that wall outlet Get help from building services If other device does work AC power is available from that outlet
61. n the next screen not shown select the CD drive Found New Hardware Wizard Welcome to the Found New Hardware Wizard This wizard helps you install software for VIDAR Film Digitizer If your hardware came with an installation CD lt gt or floppy disk insert it now What do you want the wizard to do Click Next to continue Nest Cancel VIDAR Advantage Series Installation and Operation Guide 75 16 The wizard will search for the USB driver files Found New Hardware Wizard Please wait while the wizard searches a NO VIDAR Film Digitizer Cancel 17 You may see a warning about Windows Logo Testing Click Continue Anyway Hardware Installation The software you are installing for this hardware VIDAR Film Digitizer has not passed Windows Logo testing to verify its compatibility with Windows XP Tell me why this testing is important Continuing your installation of this software may impair or destabilize the correct operation of your system either immediately or in the future Microsoft strongly recommends that you stop this installation now and contact the hardware vendor for software that has passed Windows Logo testing Continue Anyway STOP Installation VIDAR Advantage Series Installation and Operation Guide 76 18 The wizard will indicate that files are being installed This may take a minute or so to complete Found New Hardware Wizard Ple
62. nd angeschlossene Ger te ausschalten Durch AnschluB des SCSI Kabels an angeschaltete Ger te k nnen ernsthafte Sch den entstehen VIDAR Advantage Series Installation and Operation Guide 53 1 Install the SCSI card This section provides general instructions for installing the SCSI card Refer to your computer s manual for specific instructions a Turn off power to the computer and disconnect the computer s power cord from the electrical outlet b Remove the cover from the computer Refer to the computer s manual for instructions on removing the cover Tip If the computer is a tower model it s easier to install the SCSI card when the tower is laid on its side c Locate an unused PCI expansion slot PCI connectors are typically white or ivory Remove the expansion slot cover Save the slot cover screw for use in step g PCI expansion slots d Before handling the SCSI card discharge your static electricity by touching a metal part on the computer chassis e Remove the SCSI Card from its anti static packaging Handle the card only by its edges continued VIDAR Advantage Series Installation and Operation Guide 54 f Insert the SCSI Card into the PCI expansion slot Press down firmly until the SCSI card clicks into place These slightly tilted cards are incorrectly seated in the slot When properly seated the gold fingers along the bottom of the card should hardl
63. nect p n AUA 2000 Windows XP does not recognize the USB connected film digitizer when another USB scanner is connected to the system Disconnect the other scanner from the system Windows XP SP1 and SP2 supports only one scanner per system VIDAR Advantage Series Installation and Operation Guide 114 Weights and dimensions Shipping weight and dimensions Shipping dimensions 24 wide x 29 deep x 24 high 610mm x 737mm x 610mm Shipping weight 60 Ib 27 kg Digitizer weight and dimensions DIAGNOSTICPRO Advantage and DosiMETRYPRO Advantage Overall dimensions 19 wide x 23 deep x 29 25 high 483mm x 584mm x 743mm Footprint 19 wide x 23 deep 483mm x 584mm Weight 40 Ib 18 kg Digitizer weight and dimensions CADPRo Advantage Overall dimensions 19 wide x 21 25 deep x 25 5 high 483mm x 540mm x 648mm Footprint 19 wide x 21 25 deep 483mm x 540mm Weight 40 Ib 18 kg VIDAR Advantage Series Installation and Operation Guide 115 SCSI information and troubleshooting guide Introduction VIDAR digitizer installation instructions provide basic guidance for installing SCSI hardware and software That information enables most customers to successfully install and connect their digitizers to recommended SCSI controllers Some customers however may encounter situations in which the basic SCSI installation information is insufficient This appendix provides additional
64. nns 69 Updating the VIDAR Toolkit l U U u uu ul J J 79 Advantage Series Operating Guide J J 83 Operating DIAGNOSTICPRO Advantage and DOSIMETRYPRO Advantage oeconomia a d aid 84 About digitizers and film sizes 84 leie TEE 86 Basic operating Imstruchons u 87 Digitizing a single Tim 88 Digitizing multiple lm 90 Operating CADPRO Advantage 93 Film sizes and scanning specifications seen 93 ee TEE 94 Operating Instr ctions ciere ecce aede e a ER eit au neces aed 95 Advantage Series Other Information 97 Preventive maintenancoe u u uu u u u 98 Cleaning the Auto Feed Roller DIAGNOSTICPRO Advantage and DOSIMETRY PRO Advantage onby tun didt reote 98 Cleaning the Auto Feed Roller and Suction Cups CADPRO Advantage ejje T M a Duan 99 Cleaning the diffuser all Advantage models 102 Replacing the lamp cartridge all Advantage models 106 Troubleshooting 2 2 11er q ceceeadee Zeeche deischter e destecueedeseesene 107 Reading the L EDS s irte edi tt d sette 107 Micase Of difficulty siinse Er
65. ns about installing or using your VIDAR film digitizer Phone 1 800 471 SCAN 1 800 471 7226 1 703 471 7070 outside the U S E mail medtech vidar com NEXT Go to Setting up gt VIDAR CADPRO Advantage Installation Guide 34 Setting up The digitizer must be placed on a cart counter or tabletop SAFETY WARNING Never place the digitizer on the floor WOU WEE qem um Em e Minim em 20i im VIDAR Advantage Series Installation and Operation Guide 35 Identify important features Look over the digitizer base and locate the features shown in this section You will need to know where these features are when you assemble and operate the digitizer in later chapters Film feeder connector n MEE LL J y Film feeder screw holes Digitizer base top view from rear SCSI port ON OFF switch a rocker switch SCSI ID switch SCSI termination switch Press I to turn power ON Press 0 to turn power OFF USB port Power entry module lt Digitizer base rear view Power entry module VIDAR CADPRO Advantage Installation Guide 36 Assemble the digitizer 1 Place the digitizer base on a solid flat surface such as a table VIDAR CADPRO Advantage Installation Guide 37 2 Attach the feeder to the digitizer a Note the three screws protruding from the bottom of the fe
66. ntation when installing and operating the Digitizer Caution Portable or mobile RF communication equipment can effect Medical Electrical equipment Caution Using the Digitizer adjacent to or stacked with other equipment may cause interference between the equipment Before utilizing stacked or adjacent equipment verify proper functionality of all equipment in the actual configuration in which it will operate Caution Connecting the Digitizer to equipment that is not rated CISPR 11 class A or class B may alter the electromagnetic characteristics Caution In a domestic environment this product may cause radio interference in which case the user may be required to take adequate measures Guidance andnialautze ture risanerfieratieri s ddet ratiungnetic immunity l ti l xz 105 20d Tive RIGONOS e eet ENE GArAeNeskaae aah CIR VE nn ded fpigigedr dheuatesemegnatig gnydegmeni amp panitiegesimMent he customer or end user should assure that they are usedHn such an environment test Esau mihsteelt Campllance CoheglianeRisy bt Q aI met guidance Lee The Digi Wels USE BUTE SREICOBb doit Anger Fane F tas L SEN 6 KYheRbEact Therefore ek BF lemiesionearevveBal Qum GS PENUD and are not likelyctgygseserameimierfatenaeri n Rarity 042 FS KV al 8 amp alfonic Geer be at least 30 Rrebmissians fed Sent eSlagst r power 2kVforpower Mains power quality should be that of
67. on the other end of that cable to the external port on the SCSI adapter 15 Turn on the digitizer 16 Turn on the computer 17 After Windows loads the Found New Hardware window will appear indicating that the digitizer is connected to the computer 18 Next the Digital Signature Not Found window will appear Click Yes VIDAR Advantage Series Installation and Operation Guide 64 Digital Signature Not Found E gt x The Microsoft digital signature affirms that software has been tested with Windows and that the software has not been altered since it was tested The software you are about to install does not contain Microsoft digital signature Therefore there is no guarantee that this software works correctly with Windows VIDAR Film Digitizer If you want to search for Microsoft digitally signed software visit the Windows Update Web site at http windowsupdate microsoft com to see if one is available Do you want to continue the installation i No More Info 19 Verify that the digitizer drivers are fully installed by checking the device properties Use the path for your operating system m For Windows M XP Professional Start gt Control Panel gt Printers and Other Hardware gt Scanners and Cameras m For Windows 2000 Professional Start gt Settings gt Control Panel gt Scanners and Cameras 20 In the device window right click Vidar Film Digitizer then click Propert
68. ps to prevent film feeding problems Replacement Suction Cups p n 4533 001 can be ordered from VIDAR 6 Clean the Suction Cups using an isopropyl wipe VIDAR part number 4100 CAUTION m Do not use any type of cleaning solvent on the feed roller as this could damage it and cause improper operation m Do not use tape with a strong adhesive such as packing tape or strapping tape m Do not use tape requiring the adhesive to be wetted such as brown paper packing tape 7 Gently push the Suction Cups back in VIDAR Advantage Series Installation and Operation Guide 101 8 Secure the faceplate to the feeder with the four captive screws Tighten the screws a little at a time in the order shown here VIDAR Advantage Series Installation and Operation Guide 102 Cleaning the diffuser all Advantage models If you notice streaks vertical artifacts in images you should clean the diffuser Turn off the digitizer Allow 10 to 15 minutes for the bulb cartridge to cool to a comfortable temperature before removing it 2 Open the door on the side of the digitizer base Note The door has a door open sensor When the door is open all processes including calibration and scanning will stop VIDAR Advantage Series Installation and Operation Guide 103 3 Grasp the top and bottom of the lamp cartridge then pull it out Do not touch diffuser CAUTION Do not
69. r Front of film feeder Curved portion of film support bar Front of film feeder IMPORTANT Do not lift or carry the digitizer using the film support bar VIDAR CADPRO Advantage Installation Guide 22 Warning Before you begin the installation procedure turn off all power to the computer and peripherals Connecting the SCSI cable with the power on can cause serious damage to the unit or your computer Pr caution Avant de commencer la connexion assurez vous que votre ordinateur soit bien teint Toute connexion du cable SCSI et courant pourrait gravement endommager votre num riseur ou ordinateur Advertencia Antes de continuar con la instalaci n favor de apagar su computadora y perif rico Conectando el cable SCSI con la computadora prendida puede causar dario al equipo o a su computadora Warnung Vor dem Installieren den Computer und angeschlossene Ger te ausschalten Durch AnschluB des SCSI Kabels an angeschaltete Ger te k nnen ernsthafte Sch den entstehen VIDAR CADPRO Advantage Installation Guide 23 Connect the power cord 1 Connect the power cord to the digitizer s Power Entry Module PEM located on the rear of the digitizer 2 Connect the other end of the power cord to a grounded power outlet Note The film digitizer has an auto sensing power supply When you apply power the digitizer will detect the voltage and set itself for proper operation VI
70. r for shipment IMPORTANT When shipping the film digitizer you must use the original packing materials including wire ties plastic bags plastic wrap foam supports and cartons Improper packaging may allow damage to the digitizer which will incur additional charges for repair 1 Turn off the digitizer Press 0 on the power switch VIDAR Advantage Series Installation and Operation Guide 121 2 Disconnect the power cord and SCSI or USB cable a Disconnect the power cord from the Power Entry Module on the rear of the digitizer b Disconnect the SCSI or USB cable from the rear of the digitizer SCSI cable USB cable h not s own ayy Av Power cable D 2 a VIDAR Advantage Series Installation and Operation Guide 122 3 Remove the exit tray from the digitizer base Lift the exit tray up to disengage it from the plastic edge just below the rollers then pull it gently toward you Lift exit tray up and off plastic edge Removing DIAGNOSTICPRO Advantage and DOSIMETRYPRO Advantage exit tray Lift exit tray up and off plastic edge Removing CADPRO Advantage exit tray VIDAR Advantage Series Installation and Operation Guide 123 4 Remove the feeder from the digitizer base a DIAGNOSTICPRO Advantage and DOSIMETRYPRO Advantage only Remove the film support bar from the top of the feeder Note You must remove the film support b
71. r Calibration feature requires that the light path be clear of the film for proper background calibration of the digitizer The digitizer will automatically adjust the film s position to properly proceed with ADC Depending on the location of the film s leading or trailing edge and the length of time the film has been blocking the light path the film will be 8 pushed up or b pushed down 4 Using your scanning software execute the Scan command Note During extended batch scans more than 5 films the digitizer may pause periodically to recalibrate This is normal operation Note Proper orientation of the film during loading depends on the scanning application program you are using Some scanning programs rotate images 180 for display If images appear upside down you may be able to set the default orientation in the scanning application If this option isn t available then insert films into the digitizer upside down as shown here IS WA VIDAR Advantage Series Installation and Operation Guide 92 VIDAR Advantage Series Installation and Operation Guide 93 Operating CADPRO Advantage Film sizes and scanning specifications 18cm long x 24cm wide mammography film landscape orientation Spot size Line pairs Digitizing Pixels um DPI per mm speed 4032 x 5376 44 570 11 30 sec 30cm long x 24cm wide mammography film portrait orientation Spot size Line pairs Digitizing Pixels um DPI p
72. set to Yes or No Set them all to No Set Maximum Sync Transfer Rate to 10 0 for all Press ESC to exit this screen SCSI Device Configuration SCSI Device ID 0 1 2 3 4 5 6 7 Initiate Sync Negotiation no no no no no no no Maximum Syne Transfer Rate MB Sec 10 0 10 0 10 0 10 0 10 0 10 0 10 0 10 0 Enable Disconnection yes yes yes yes yes yes yes yes Options Listed Below Have NO EFFECT if the BIOS is Disabled Send Start Unit Command no no no no no no no no BIOS Multiple LUN Support no no no no no no no no include in BIOS Ben ui 9n ew yes yes yes yes yes yes yes yes VIDAR Advantage Series Installation and Operation Guide 58 5 Select Advanced Configuration Options see below and press ENTER AHA 2930U at Bus 00h Device 0Dh Configuration SCSI Bus Interface Definitions Host Adapter SCSI ID SCSI Parity Checking Enabled Host Adapter SCSI Termination Automatic Additional Options Boot Device Options Press Enter SCSI Device Configuration Press Enter Advanced Configuration Options Press Enter lt F6 gt Reset to Host Adapter Defaults BIOS Information Interrupt IRQ Channel I O Port Address 6 Select Host Adapter Bios Set to Disabled 7 Press ESC then SAVE SETTINGS then REBOOT NEXT Goto Installing SCSI drivers D gt VIDAR Advantage Series Installation and Operation Guide 59 Installing SCSI drivers
73. source of the interference This product is intended to be turned on and left on Operation is continuous Correct and safe operation of the digitizer requires familiarity with information that is not marked on the product The following symbol indicates the operator should consult the manual for additional information Turn off the scanner before mounting or removing the feeder Mount the feeder to the scanner by mating the connector found on the bottom of the feeder to the connector found on the top of the scanner Secure the feeder to the scanner with the three thumbscrews To remove the feeder reverse the mounting process Contents DIAGNOSTICPROTM Advantage ANd DosimetRYPRO Advantage Installation G Ild e ORCE D NN 7 Upnpechking aen eect eae Su ee Ee E 8 T Look for damage teet ceat ep a aber EL ERR E42 8 2 Unpack the shipping carton eeeeeeesseeseeseseeeeeee trennen trenes 9 3 Unpack the accessory box 10 LM Iro C 12 5 Identify the components L enne 13 6 If anything is missing enne enne 14 7 Activate your product warranty eene nennen 14 IO iA HC DE 15 Identify important features 16 Assemble the doittzer essen 17 Connect the power con 23 Where to go from bere 24 CADPRO Advantage Installation Guide
74. tadora prendida puede causar dario al equipo o a su computadora Warnung Vor dem Installieren den Computer und angeschlossene Ger te ausschalten Durch AnschluB des SCSI Kabels an angeschaltete Ger te k nnen ernsthafte Sch den entstehen VIDAR Advantage Series Installation and Operation Guide 48 If necessary set the digitizer s SCSI ID A computer equipped with a SCSI bus can communicate with multiple SCSI devices for example a film digitizer a scanner and a disk drive Each device must have a unique SCSI ID number so the computer can distinguish it from other SCSI devices Valid SCSI ID numbers range from 1 to 6 When selecting SCSI ID numbers m The SCSI Card 2930U is preset to SCSI ID 7 and should not be changed m Set the film digitizer to any SCSI ID between 1 and 6 as long as that number is not used by another SCSI device attached to the computer Note The digitizer is shipped from the factory with the SCSI ID preset to 3 If the system is configured to boot from a SCSI disk drive it s best to set the disk s SCSI ID to 0 or 1 Most SCSI disks are preset to SCSI ID 0 at the factory CAUTION The digitizer and computer MUST be turned OFF before changing the SCSI ID To set the digitizer s SCSI ID 1 Locate the SCSI ID switch on the back of the digitizer below the SCSI port connector 2 Change the SCSI ID switch using a 1 8 or 2 flat blade screwdriver Note Do not set the swit
75. tall Software INSTALL DIGITIZER DRIVER ose the software to install or UPDATE TOOLKIT DLL MAIN MENU 6 You will see a message that the InstallShield Wizard is starting VIDAR Advantage Series Installation and Operation Guide 61 7 In the InstallShield Welcome screen click Next VIDAR Medical Film Digitizer Driver InstallShield Wizard VIDAR Medical Film Digitizer Driver Installation Program to the VIDAR Medical Film Digitizer Dri DD and XP installation OT CONNECT THE DIGITI ALLA M InstallShield 8 If you are updating from older VIDAR SCSI drivers you will see the message below Click Yes pun UU xi K i If you proceed with this installation existing VIDAR drivers will be removed Do you wish to continue WE No VIDAR Advantage Series Installation and Operation Guide 62 9 In the License Agreement window activate the I accept the terms of the license agreement then click Next License Agreement Ple END USER SOFTWARE LICENSE AGREEMENT FOR VIDAR SYSTEMS CORPORATION a SOFTWARE THIS IS A LEGALLY BINDING AGREEMENT BETWEEN YOU EITHER AN INDIVIDUAL OR A SINGLE ENTITY AND VIDAR SYSTEMS CORPORATION VIDAR FOR THE VIDAR SOFTWARE AND ASSOCIATED MEDIA AND PRINTED MATERIALS AND MAY INCLUDE ONLINE OR ELECTRONIC DOCUMENTATION SOFTWARE PRODUCT OR SOFTWARE BY INSTALLING COPYING OR OTHERWISE USING VIDAR S SOFTWARE PRODUCT OU ARE AGREEING TO BE BOUND BY THE TERMS AND CONDITIONS
76. vantage and CADPRO Advantage carry the CE mark issued by BSI This product s function and intended use is as an X ray Film Digitizer To maintain Medical Equipment Certification the digitizer must be connected to a host computer that has been configured in accordance with IEC 60601 1 1 This product is in the Ordinary Equipment Class It provides no protection against the ingress of water This product is not suitable for use in the presence of flammable anesthetic mixtures with air or with oxygen or with nitrous oxide Class I Medical Device No Applied Parts This product provides Class I medical device protection against electrical shock Power cords used with this device in North America must be rated by Underwriters Laboratories for hospital use Power cords used with this device in Europe must meet the requirements of IEC 227 Designation 53 or IEC 245 Designation 53 The use of portable or mobile communications equipment and or the presence of strong electromagnetic and or x ray fields may interfere with proper operation of this product This product should not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary verify normal operation in the configuration in which it will be used Should such interference occur the user is required to provide adequate isolation between the digitizer and the source of the interference Isolation is typically achieved by moving the digitizer away from the
77. xecute the Scan command Note Proper orientation of the film during loading depends on the scanning application program you are using Some scanning programs rotate images 180 for display If images appear upside down you may be able to set the default orientation in the scanning application If this option isn t available then insert films into the digitizer upside down as shown here VIDAR Advantage Series Installation and Operation Guide 90 Digitizing multiple films CAUTION Do not load more than 25 films at one time 1 Hold the film stack in front of you as you would view it on a light box Top Films at the back of the stack are digitized first Bottom Note The digitizer cannot feed films smaller than 8 wide x 8 long 20 3cm wide x 20 3cm long When loading small films the vertical dimension must be at least 8 20 3cm mm 2 Align the left edge of the film stack with the film guide on the left edge of the feeder 3 Place the film stack IN the groove in the blue bar The blue bar has a lip and a groove on the top edge of the digitizer where the feeder is attached Place a single film directly behind the blue bar Note If you are digitizing a single film place the film BEHIND the blue bar continued VIDAR Advantage Series Installation and Operation Guide 91 Note When a film has been staged it is in the digitizer s light path The ADC Automatic Digitize
78. y show g Secure the SCSI card with the expansion slot screw you removed in step c h Replace the cover on the computer VIDAR Advantage Series Installation and Operation Guide 55 2 Configure the SCSI BIOS Note This step is required for SCSI installations A Turn on the digitizer Ensure the PC is turned off Note Always apply power to the digitizer before turning on the computer This enables the computer to recognize the digitizer 2 Turn on the digitizer 3 Watch the LEDs on the front of the digitizer Wait until all LEDs are solid green center LED may flash slowly Then proceed to step 2 next in this chapter LE a DIAGNOSTICPRO Y Advantage Note When you turn on the digitizer several minutes are required for the digitizer to carry out its internal diagnostics LEDs x Location Function Normal state Top LED Feeder status Solid green Solid green but slow flashing Center LED Digitizer status when digitizer is busy Bottom LED Power status Solid green e eve Note For more information about LED states see Appendix troubleshooting later in this manual VIDAR Advantage Series Installation and Operation Guide 56 B Edit the SCSI adapter BIOS settings In most cases the digitizer is not properly detected after you install the SCSI adapter This is because the computer tries to boot from the adapter but no SCSI hard disk is connected You