The PowerMAX Ultrasonic Scaler (PM 25&30) User Manual

image

Contents

1. TE Do not turn water control knob anti clockwise repeatedly or knob will come off 3 POWER Indicating the output power intensity control knob When turning clockwise the output power intensity will be increased When turning counter clockwise the output power intensity will be decreased This power intensity control knob gives the flexibility to adjust the output power intensity based on the difficulty of the cleaning job e g toughness of the calculus Label 1 means the minimum power output and labels 3 means maximum power output 4 Indicator This indicator represents whether the unit s master power is on or off Light will illuminate when the power is on The LED will be off when the units main power is off NOTE Follow power setting guidelines from insert manufacturer for different type of tips ll INSTALLATION A Safety Instructions Grounding Before any connection to the output connectors is made the unit must be connected to a protective earth conductor via the three core main cable The main plug must be inserted only into a socket outlet that has a protective earth contact Main voltage range and fuse Before plugging the main plug into the outlet make sure that the instrument is suitable for local main voltage The fuse holder is located on the rear panel in the main input socket To replace a new fuse proceed as follow Disconnect the unit from the power supply outlet Remove the cover of th
2. The PowerMAX Ultrasonic Scaler PM 25 amp 30 User Manual Powe rVMAX Scaling System Power you want where you want it TPC Advanced Technology Inc Tel 800 560 8222 www tocadvanceadtechnology com REV 1 1 13 TPC ADVANCED TECHNOLOGY INC The Power Company TPC ADVANCED TECHNOLOGY IN owns all rights to this publication The publication should be used solely for the purpose of reference operation maintenance or repair TPC equipment No part of this document may be reproduced or disseminated in any form or by any means electronically or mechanically for other purposes In the event of inadvertent or deliberate publication TPC intends to enforce its rights to this manual under copyright law No one may copy the materials presented in this manual without the express written permission of TPC ADVANCED TECHNOLOGY INC Product specifications as well as information presented in this manual are subject to change without prior notice PROPERTY OF TPC ADVANCED TECHNOLOGY INC ALL RIGHTS RESERVED TABLE OF CONTENTS Attention Description of Components I Description of Control Panel II Installation JII Operation Procedures IV Maintenance V Cleaning and Sterilization VI Specifications Vil ATTENTION Thank you for choosing TPC PowerMAX Scaler PM 25 30KHz Before using the unit please carefully read this User Manual since it contains important information including installation procedures operation tips a
3. e Before cleaning remove the Insert from the Handpiece Let the Handpiece run for a couple of seconds to drain out the water and any possible contamination left inside the Handpiece The outer surface of the handpiece should be cleaned with an antiseptic soap or solution Rinse off with water and wipe or spray with a chemical disinfectant that is compatible with the handpiece material such as a 1 6 dilute solution of Sporicidin brand disinfectant A sterilized Insert or Tip is then reassembled to the Handpiece in preparation of the next patient Warning Do not put the Handpiece and the extension wire directly into any sterilization fluid Any such fluid left inside the machine will interfere with the normal operations of the system At the end of the day with the Insert removed the Handpiece and cable should be scrubbed with an antiseptic soap or solution and rinsed off with water Warning The chemical disinfectant should not be allowed to remain on the surface longer than the recommended time or damage may result e Insert After each usage there may be saliva and or blood or other debris left on the Insert consequently it is necessary to clean the Insert with a supersonic cleaner This can be done manually by scrubbing with a brush or by use of an ultrasonic cleaner with a solution of detergent and water After scrubbing the Insert should be rinsed thoroughly with water to remove all detergent and dried Finally the Inser
4. e fuse holder with a small screwdriver Fit a new fuse of the correct rating and refit the cover of the fuse holder WARNING The unit should be disconnected from the power supply outlet when replacing a fuse B Installation Instructions Water connection Connect the scaler s water to a water supply with the male quick disconnect that is at the end of the water line NOTE Water supply should be of drinking quality Purified water or distilled water is recommended Electrical Connection Plug the foot control and AC power cord into the back of the main unit Then plug the AC power line to a grounded power outlet Inserts Choose the correct INSERT for your scaler TPC 25k long type inserts for your 25HKz scaler or TPC 30k short type inserts for your 30HKz scaler The handpiece will also accept Densply and Hu Friedy inserts NOTE When pushing the insert into the handpiece do not put the insert all the way into the handpiece at once First push the insert into the handpiece half way step on the foot control to turn on the scaler and allow some water to come out of the handpiece then push the insert all the way into the handpiece The above procedure will help to reduce air inside the handpiece This will prevent the handpiece from generating too much heat due to extra air and thus extending the product life of the handpiece IV OPERATION PROCEDURES e Preparation Step 1 Push the POWER SWITCH to light ON indica
5. iece is not enough to perform treatment it is very possible that the Insert is worn out If the Insert is worn out replace it with a new one otherwise have your authorized agent provide maintainservice to the system V MAINTENANCE The PowerMAX Scaler does not need special maintenance routine The following daily start up Shut off routine and tips are useful to extend the product life of the Scaler a Daily start up Turn on the water supply Push the MAIN POWER SWITCH to light the on indicator LED NOTE If no water comes out please check the water supply NOTE if the power LED does not light please contact your local authorized agent m Daily Shut off Tips Place the Main Unit on an open area where air can flow freely If Main Unit need to be moved handle with care Before leaving the Clinic Room check that the AC power is turned off and the water faucet is tightly closed VI CLEANING AND STERILIZATION In this section we describe the procedures to clean and sterilize the scaler It is important to follow these procedures before using the machine on human bodies otherwise patients and or personnel may have the possibility of infection It is mandatory that operators wear sterilized gloves during these procedures at all times to avoid any possibility of incomplete sterilization and or infection Below we detail the procedures for the Handpiece the Insert and the Main Unit respectively e Handpiec
6. itch Cable Specification 110 125V 60Hz 0 85A 24 6W Scaling mode 8 9W Scaling mode 24 5KHz for 25KHz Insert Model 30 KHz for 30KHz Insert Model 25 60 pounds inch2 5 50ml min 8 W x 8 D x 3 H 4Kg include handpiece Cable 250 cm 250 cm Environment Temperature Operation 10 C 35 C Storage 0 C 60 C e Humidity Operation 20 80 non condensing Storage 10 90 non condensing e Pressure Atmospheric pressure 860 1060 Hpa Classification e Class II with transformer e Type B IP40
7. nd warnings Warnings Important Notes The equipment is only to be used by a qualified doctor or dentist Patients with pacemakers cannot be treated with the equipment A shielded AC power cord must be used with this equipment PowerMAX should be powered from separate wall outlet with a grounding point I DESCRIPTION OF COMPONENTS Description Part Number 1 Main unit in 25KHz or 30 KHz with attached handpiece PM25 PM30 2 Foot control FC 3 Water line with male quick connect coupler WLWCCM 4 AC power line ACPC 5 Scaler Insert 1025 S1030 ll DESCRIPTION OF CONTROL PANEL The function of each button or Knob on the control panel is detailed below ON dh gt TP C Power Max 25 Ultrasonic Scaler T lt gt OFF di e s MAX INDICATOR POWER Gs 1 ON OFF Indicating the main power switch When pushed the AC power is sent to the Main unit and the power indicator LED will be lit up If pushed again the power is then turned off and the power indicator will be extinguished 2 WATER Indicating the water volume control knob When turning clockwise the water volume coming out of the handpiece will be decreased Turning counter clockwise will increase the water volume When operating a continuous flow of water is required to keep the handpiece and insert cool Adjustment of the water flow rate controls the water temperature the greater the water flow the lower temperature and vice versa NO
8. t is put into a sterilization pouch and placed into a medical equipment autoclave Sterilize the Insert at 260 F 127 C for 30 minutes or as recommended by the manufacturer of the particular sterilizer used Warning High room temperature conditions improper dilutions or excessive immersion time in a chemical sterilant can result in damage to the plastic and elastomeric materials of the Insert Caution The use of a dry heat over incompatible chemical vapor style sterilizers quaternary ammonium compounds or alcohol based compounds must be avoided as damage can result to the plastic and elastomeric materials e Main Unit The Main Unit should be cleaned as follows Carefully swab the Main Unit with isopropyl alcohol and keep it away from dust VII SPECIFICATION PowerMAX Scaler is designed and manufacturers to meet the most demanding environment Its specification are list below 1 Standard List Emission EN55011 Class B EN60555 2 Class A EN60555 3 Immunity IEC 1000 4 2 IEC 1000 4 4 IEC 1000 4 5 IEC 801 3 ENV50141 Safety IEC 60601 1 EN1640 2 Function List Scaler power control Water flow control Footswitch control Power Turbo mode Power perio mode 25 KHz Or 30KHz Insert Selectable Using different Handpiece 3 Specifications ltem Power supply Max Output Power Min Output Power Basic Vibration Frequency Water e Input pressure e Output volume Dimension Weight Handpiece Footsw
9. tor LED Step 2 Check the water Step 3 Select the needed INSERT TIP wet the O ring then place the INSERT into the HANDPIECE with a clockwise motion until it is fully seated CAUTION Insert should be sterilized before each use Step 4 Set the POWER KNOB to a suitable level for the insert Step 5 Hold the HANDPIECE with the insert end pointing up over a suitable drain Step on the foot control and allow water to run from the handpiece for a few seconds until it flows without spurting CAUTION Make sure that the water spray is at the desired temperate and is reaching the working edge of the tip Step 6 Place insert tip into the patient mouth and use the foot control to activate the handpiece and water follow e Operation Tips During treatment keep the angle between the patient s tooth surface and the Handpiece Insert as close to 15 degree as possible If the patient is not comfortable during the treatment try the following steps A When treating areas where patient is very sensitive try to increase the Handpiece movement speed on the surface B Leave these sensitive areas for later treatment and do other areas first C If the problem persists reduce the output power intensity of the Handpiece D When operating a continues flow of water is required to keep the Handpiece cool The greater the water flow the lower the water temperature and vice versa E After six months or if you find the output power of the Handp