Gilibrator User Manual - Enviro
Contents
1. Figure 6 E Connect the Sampler 1 Connect the air sampler to be calibrated to the Upper Outlet Boss 26 of the Flow Cell Assembly with 1 4 ID tubing NOTE An auxillary liquid trap between sampler and flow cell is recommended to prevent moisture carry over into the sampler during continuous calibration periods 2 _ Operation A Conditioning the Flow Tube 1 Turn on the sampler Depress the Bubble Initiate Button 22 several times to wet the inner walls of the flow tube 24 You will not be able to initiate a timing bubble without first Priming the flow tube The operator will develop a feel for bubble generat ion with practice B Power Up 2 After the Flow Tube walls have been primed turn on the Power Switch 12 of the Cilibrator Control Unit base and the Printer Module if one is being used Wait approximately 10 seconds while the system runs through it s check sequence The Run LED 20 will light at this time as well as a Lo Battery indication and a series of five dashes displayed on the LCD Readout 19 Do not Operate the Cilibrator until the Run LED signal extinguishes Ready operation is indicated by a series of 4 dashes C Bubble Generation 1 For optimum bubble generation depress the Bubble Initiate Button 22 and hold to initiate 1 bubble up the flow tube Re lease the button to initiate a second bubble up the flow tube This will be the standard procedure to making clean consistent bubbles at
2. H20 No leakage should be observed c Cleaning To clean the exterior of the Bubble Generator use a mild detergent and warm water NEVER USE ALCOHOL ACETONE OR ANY OTHER HARSH CLEANERS TO CLEAN THE BUBBLE GENERATOR d Transportation When transporting the Gilibrator especially by air it is important that one side of the seal tube which connects the inlet and outlet boss be removed thereby allowing f o r equalizing internal pressure within the generator Do Not transport unit with soap solution or storage tubing in place CAUTION Do Not Pressurize the Flow Cell Excessive pressure may cause cell to rupture resulting in personal injury Section 6 Printer Module Operation Part C 800274 1 Introduct ion The Cilibrator Printer provides a hardcopy record of all calibration data identical to data calculated and displayed on the LCD readout of the Gil ibrator Control Unit base Handling Precautions Do not use where the temperature is extremely hot or cold Do not leave in direct sunlight or where it is dusty Do not operate near liquids or beverage Do not operate without the heat sensitive roll paper loaded Use only specified Cilian Replacement Roll Paper A 400681 Do not attempt to disassemble the unit bd ECH ET NOTE The printer should only be used with the Cilian Interconnect Cable providedwhich supplies the power for the printer Use of this printer with any other cable may cause permanent damage to the print
3. High and Medium flow ranges 12 2 As the bubble rises up the Flow Tube 24 it will initiate a timing sequence when it passes the lower sensor Run LED will light and culminate the timing sequence upon passing the upper sensors Run LED will extinguish The timing information is then transmitted to the control unit which will perform all the necessary calculation A flow reading will instantaneously appear on the LCD display 19 However if a bubble breaks before the time sequence is completed timing will continue until another bubble is generated to trip the second sensors This will cause an erroneous reading and should be subtracted from the average by hitting the Delete Button 16 If a Printer Module is used be sure the printer has comple ted it s printing sequence before pressing the Delete Button When the Delete Button is activated a negative symbol will be display ed on the LCD of the Control Unit and the printer will initiate a line showing this subtraction D Flow Readout With Printer Option The printer will print in sequence the flow average and sample number of each successive bubble reading Without Printer Option The Control Unit will display the actual flow for each sample and will accumulate and average each succes sive reading 1 Average 15 In order to display average and number of samples depress and hold the Average Button Releasing the button will display the last flow reading Repres
4. source as the control unit 3 Theory of Operation The Printer Module is powered and interfaced to the Control Unit of the Cilibrator by means of an Interconnect Cable During a test series the Control Unit transfers the sum of calculations of a given calibration bubble and provides the flow rate the sample number and an average of the successive flow rates determined in that calibration series The buttons on the Control Unit are used in conjunction with the Printer module during calculation Delete Auto Average and Reset buttons are provided to subtract false readings and to initiate new calibration sequences Calculations are automatically provided by the Control Unit and transferred to the printer for a hardcopy readout This is printed onto a thermal sensitive roll paper which advances a single line after printing out each successive reading These hard copy calculations are retained for calibration records 4 Operation Procedures a Initial set up 1 Connect ing the Printer a While the Gilibrator Control Unit is in the OFF mode connect the large end of the Interconnect Cable 2 to the Printer Modules Parrallel Interface Connector and snap holding clips in place Connect the smaller end to the Control Unit s Printer connect port Power is supplied to the printer from the Control Unit s battery system 2 Turn ON the Control Unit and the Printer Module a 3 Installing the Remove the heat sensitive roll pap
5. the Damper Diaphragm assembly If the diaphragm becomes ruptured or worn please use the following procedure for its replacement a Removing and Replacing Damper Plate Diaphragm First remove flexible Safety Tape 27 from around the lip of the Damper Plate assembly 1 Using a small flat blade screwdriver remove the Damper Plate 2 from the Upper Cell Chamber using the notch provided Remove the large O ring 2b and the Pulsation Damper diaphragm 2a See fig 10 Damper Plate Assembly x Figure 11 Insert a Flathead Screwdriver between the Damper Plate Assembly and the tip of the Upper Flow Cell Chamber Gently pry apart the two by twisting the screwdriver K Diaphragm Damper Plate Bubble Generator Assembly 2 To replace center new diaphragm over Damper Plate aperture and roll O ring over diaphragm and into the O ring groove If wrinkles occur repeat the procedure to achieve a smooth place ment See fig 11 16 Reposition Damper Plate Assembly into Upper Flow Cell Chamber and press firmly into place Damper Plate Asse Upper Flow Cell Ch Figure 12 Bubble Generator Assembly 3 Wet QO ring of Damper Plate amp press into the Upper Cell Chamber firmly see Fig 12 4 Replace Safety Tape around lip of plate and the upper flow cell chamber b Leakage Check The system should be leak checked at 6 H20 by connecting a manometer to the outlet boss and evacuate the inlet to 6
6. Breaker Plate 4 This gives completeaccess to the interior of the Flow Cell Tube Continue to run clear water through the cell until water runs clear Rock cell to empty out all excess water Replace the Bubble Breaker Plate 4 and center the Air Outlet Boss 26 with the plates largest hole Next insert the spacer To replace the Damper Plate assembly moisten the O ring 2b with soap solution and then press the Damper Plate into the top of the Upper Cell Chamber Use fingers and firmly sqeeze plate into upper flow cell chamber Rock the Flow Cell Assembly in a see saw fashion Figure 9 Bubble Generator Assembly Damper Plate Assembly Flush with Water Sensor Block Insert a Flathead Screwdriver between the Damper Plate Assembly and the lip of the Upper Flow Cell Da Chamber Gently pry apart the two Sensor Block Holding Screws by twisting the screwdriver Do Not exert your body weight on the Bubble Generator to press plate into place Exces Lower Flow Cell Chamber sive weight on the flow cell may result in breakage Replace the Seal Tubing between the Upper and Lower cell bosses for storage l Reposit ion Damper Plate Assembly into Upper Flow Cell Chamber Damper Plate and press firmly into place Upper Flow Cell hamber Bubble Generator Assembly 15 2 Maintenance The Cilibrator is designed so that little maintenance is required The area which may need replacement is
7. D 800275 D 800265 D 800266 D 800267 B 800331 D 800268 C 800274 A 400681 B 400674 A 400450 A 400667 B 800269 D 800273 23 24 Section 9 Warranty The Seller warrants to the Purchaser that any equipment manufactured by it and bearing its nameplate to be free from defects in material or workman ship under proper and normal use and service as follows If at any time within 90 days from the date of sale the Purchaser notifies the Seller that in his opinion the equipment is defective and returns the equipment to the Seller s originating factory prepaid and the Seller s inspection finds the equipment to be defective in material or workmanship the Seller will promptly correct it by either at its option repairing any defective part or material or replacing it free of charge and returned shipped lowest cost transportation prepaid if Purchaser requests premium transportation Purchaser will be billed for difference in transportation costs If inspection by the Seller does not disclose any defect in material or workmanship the Seller s regular charges will apply This warranty shall be effective only if installation and maintenance is in accordance with our instructions and written notice of a defect is given to the seller within such period This warranty is exclusive and is in lieu of any other Warranties written oral or implied specifically without limitation there is no warranty of merchantability or fitness for any purpo
8. Instruction Manual for the Gilian The Gilibrator Primary Standard Airflow Calibrator with Interchangeable Flow Cell Assemblies Flow Cell Assembly Ranges Low 1 to 250 cc min Standard 20 cc min to 6 LPM High 2 to 30 LPM On Site Instruments 1 800 7 On Site 1 800 766 7483 689 North James Road Columbus Ohio 43219 Prologue This Instruction Manual describes the basic principles installation Operation controls and maintenance for the Cilibrator Primary Standard Airflow Calibrator manufactured by Cilian Instrument Corp Table of Contents The Gil tobrator Section Page Introduct ion 2 2 General Description 2 6 3 Theory of Operation 7 4 Operating Procedures 8 12 Initial Set up amp Operation 5 Storage amp Maintenance 13 17 6 The Printer Module 18 Introduction amp General Description 18 19 Theory of Operation 19 Operation Procedures 19 21 Storage and Maintenance 21 7 Spec if ications The Gil ibrator 22 The Printer Module 23 8 Cilibrator Parts List 23 d Warranty amp Service Policy 24 Section 1 Introduction The Gilibrator is a high accuracy electronic bubble flowmeter that provides instantaneous air flow readings and a cumulative averaging of multiple samples The Gilibrator system provides a large dynamic range through the use of 3 interchangeable flow cell assemblies A delete function on the Control Unit subtracts eroneous readings to insure accurate data The Control unit a
9. e interfacing code to define the cell volume as well as sensitivity adjustments for the optical sensor systems b As the bubble traverses between the sansors first one and then the second sensors are tripped thereby providing the time for the bubble traverse This timing information is sent to the micro processor of the control base which in turn provides the crystal control time base for the system The timing information along with the volume information are then sent to the micro processor which in turn does the necessary mathematical calculations which allow the flow to be displayed directly on the LCD readout In order to insure the highest accuracy possible a Delete and Average function are provided on the Control Unit The Delete allows for subtracting out an obvious malformed bubble and the average allows the user to obtain average information without pencil or paper A printer interface allows connection of a Printer Module so that hard copy can be produced Section 4 Operating Procedures 1 Initial Set up This covers all steps necessary to bring the Gilibrator into operating status This includes charging cell mounting installing soap solution connecting the printer optional and connecting the sampling source A Charging the Gilibrator 1 Prior to operation plug the 120V charger into the wall and connect to the Charging Jack 13 on the right side of the Control Unit The unit s Charging LED 11 will light indicating
10. er and to the equipment from which the cable is connected The use of a non shielded interface cable with the printer is prohibit ed 2 General Description See fig 13 The Printer is comprised of the following basic components The Printer Module 1 Interconnect Cable 2 and the Thermal Roll Paper A brief description is provided of their use and operation The Gil ibrator System The Printer Module Assembly Nomenclature 1 Printer Module 2 Printer Interconnect Cable 3 Thermal Paper Cover 4 Paper Cutter 5 Paper Feed 6 Power On Off 7 Parallel Interface Connect Figure 13 a Printer Module 1 Power 6 This switch activates operation of the printer from the Control Unit s power source The printer will initially print a heading line upon activation 2 Paper Feed 5 This switch is used to feed the Thermal Paper continuously Press it lightly Note Use the Paper Feed switch to feed out the paper If paper is pulled out by hand always feed it one line with the Paper Feed Switch before starting printing 3 Paper Cutter 4 This is used to tear of the paper Tear off by pulling it in the arrow direction only b Thermal Paper w A specially treated paper which is activated by the printer head Cl Interconnect Cable 2 The interconnect cable provides electrical connection between the printer 1 and the Gilibrator Control Unit 14 The printer operates from the same power
11. er cover erma aper Se fig 14 a S ide back and remove the Thermal Paper Cover 3 from the printer section see Fig 14 b Use the special Gilian heat sensitive roll paper A 400681 c Cut the end of the paper in a half circle or triangular shape 2 Cut the end of the heat sensitive paper a shown below d While inserting the end of the paper press the Paper Feed Switch 5 until the paper is fed out Replace protective paper cover see Fig 15 Figure 14 D 3 While inserting the end of the heat sensitive roll paper in the direction of the arrow press and hold the paper feed switch until the paper is fed out Since the outside of the heat sensi tive roll paper is coated with a heat sensitive agent pay careful attention to the front and back of the paper Figure 15 20 b Operation 1 Start Sequence a Turn the Control Unit ON The Printer Module will print out an identification heading for the current calibration test The program version is indicated after the identifica t ion line for future reference See sample below b Initiate a bubble up the Flow Tube and observe the reading on the Control unit and also on the Printer Module The Printer will print in sequence the flow average and sample number after each successive reading NOTE If you do not need a hard copy printout from the Printer Module turn it OFF This will preserve the battery life of the Gilibrator as the printe
12. g Jack Section 3 Theory of Operation 1 Primary Airflow Standard To be a primary standard all values must be absolute and measured as absolute A primary standard airflow measurement is a volume divided by a time interval as performed by the Control Unit of the Gilibrator The volume V is measured volume of space between two infrared sensors The time is that interval needed for a soap film bubble to traverse between the two sensors which bound the volume Therefore V t the volume per unit of time becomes the airflow and is prime because all measurements are basic volume and time In today s technology time is measured by an electronic clock whose accuracy exceeds that of volume measurements by orders of magnitude hence the control accuracy volume resides solely with volume measurements 2 Bubble Generation and measurement a The Gilibrator consists of two elements the Flow Cell Assembly and the Control Unit base The function of the Flow Cell Assembly is to generate a clean consistent bubble which traverses up the flow tube Meas urement of the traverse time is done by infrared sensor pairs which are mounted at the bottom and the top of the Sensor Block The volume bound by these sensors is specifically adjusted to a volume standard by allowing the upper sensor blocks to move in unison so as to enable this calibration to be set accurately to a primary volume standard A second function of the sensor block provides th
13. generation Ring Bubble Generator The Ring Bubble Generator lifts an even soap film from the soap reservoir and provides consistent bubble generation Bubble Breaker The Bubble Breaker design allows the soap film to be drawn away form the end of the flow tube before bursting This improves flow measurement accuracy by minimizing film residue Within the flow tube Programmable Update The electronics design allows programmable updates to assure a long useful service life Accuracy Better than 0 5 Section 7 cont d Gilibrator Printer Module Printing Thermal System dot matrix Character 2 dots Spacing Character Set 159 Alphabetic upper and lower case numeric Kana symbols Printing Direction Left right Number of Columns 40 columns line Section 8 Parts List High Flow Kit 2 30 LPM Standard Flow Kit 20cc 6LPM Low Flow Kit 1 250 ec min Deluxe High Flow Cell Assembly only Standard Flow Cell Assembly only Low Flow Cell only Safety Tape 10 ft Control Unit base Printer Module Printer paper 3 Battery Charger Flow Cell Soap Soap bottle dispenser Tubing Carrying Case rolls 120v Specifications Character Font 7x5 dot matrix Character Size 2 4 vertical x 1 1 horizontal mm Paper Thermal Paper width 80 Imm roll diameter 40mm reorder Gilian Thermal Paper 3 rolls A 400681 Printing 0 6 Speed line second Part D 800270 D 800271 D 800272
14. lso supports a hard copy print out through the use of a printer accessory Section 2 General Description See fig la The Gilibrator is comprised of the following basic components Flow Cell Assembly Control Unit base Battery Charger and Soap Solution Dif ferent sized inter changeable Flow Cell Assemblies are available for use as follows Low Flow 1 to 250 cc min Standard Flow 20 cc min to 6 LPM High Flow 2 to 30 LPM In addition to the basic components an optional Printer Module is available The printer provides a hard copy record of calibration data however identical data is displayed on the LCD of the Cilibrator Control Unit during calibration 1 Flow Cell Assembly See fig Ib The Flow Cell Assembly consists of a Bubble Generator and Sensor Block Each Bubble Generator is sized to produce a bubble film stretched across the diameter of the flow cell tube which is carried by airflow from the bottom to top of the tube As the bubble traverses past two infrared sensors each sensor transmits a signal to the Control Unit base indica ting the passage of the film The Flow Cell Assembly incorporates a manual bubble initiation push button which starts the film on its travel up the tube A Bubble Generator 1 Flow Ranges Low Flow 1 to 250 cc min Standard Flow 20 cc minto 6 LPM High Flow 2 to 30 LPM 2 Pulsation Damper 2 a built in damper smoothes out any pulsation within the airflow and reduces
15. occilation of the bubble film assuring maximum accuracy 3 Bubble Initiate Button 22 This pushbutton lowers the Bubble Generator Ring 21 into the soap solution reservoir Upon releasing the button the ring lifts out of the soap solution and a film bubble is generated across the opening of the flow tube 24 4 Bubble Breaker 4 The Bubble Breaker is a secondary chamber in the upper chamber which provides the soap film a rapid expan sion path which is instrumental in breaking the bubble This prevents excessive wall wetting by the soap film and allows it to flow back into the cell 5 Storage Tubing 25 This anti spill tubing connects upper and lower cell chambers and prevents the soap solution from evapora ting which may cause the solution concentration to change CAUTION If transporting by plane be sure to disconnect this hosing from upper or lower Flow Cell Chamber Bosses 23 amp 26 to prevent pressurization and possible rupture within the Bubble Generator D Sensor Block 8 Surrounding the flow tube 24 and secured between the upper and lower chamber of the bubble generator is the Sensor Block 8 The blockincorporates lower and upper sensors for time start and time stop Thesensors consist of an infrared emitter and detector pairs whose sensitivity and accuracy is controlled by a columna ting slot This block is secured to the Bubble Generator Assembly 1 by means of two Locking Screws 7 and all
16. ome available 3 Battery Charger Standard wall operated 120V charger to charge Gilibrator Control Unit for 14 hours prior to operation The Charging LED on the Control Unit will be illuminated while charging is in progress 4 Soap Solution This specially compounded low residue soap is specifically designed to provide high film strength and compatibility with the materials used within the Flow Cell Assembly The Gilibrator System The Gilibrator N Nomenclature Bubble Generator Assembly Damper Plate 2a Pulsation Damper 2b On Ring Spacer Plate Bubble Breaker Cable Assembly Sensor Block Connecting Jack Sensor Block Locking Screw Sensor Block Base Plate Assembly Mounting Base Charging Indicator On Off Charging Jack Control Unit Average amp Sample Switch Delete Switch Reset Switch Printer Jack LCD Display Run Indicator Bubble Generator Ring Bubble Initiate Button Air Intet Boss Flow Tube Storage Tubing Air Outlet Boss AW C w The Gilibrator System The Flow Cell Assembly N Nomenclature Bubble Generator Assembly Damper Plate 2a Pulsation Damper 2b O Ring Spacer Plate Bubble Breaker Sensor Block Locking Screw Sensor Block Base Plate Assembly 9a O Ring Bubble Generator Ring Bubble Initiate Button Air Inlet Boss Flow Tube Storage Tubing Air Outlet Boss Safety Tape Figure Ib Sensor Block back Connectin
17. ontrol Unit Connecting the Control Unit s Interface Plug Interface Plug to the Sensor Block s Connecting Jack C Adding the Gilibrator Soap Solution 1 Remove the Storage Tubing 25 from the upper Outlet Boss 26 of the upper cell Fill the dispenser bottle provided with Cilibrator soap solution Using the rubber Storage Tubing as a funnel slowly add soap solution from the dispenser 2 The amount of soap needed is determined by depressing the Bubble Initiate Button 22 and holding it down in the lower position Continue to add enough soap solution until the angled edge at the bottom of the Bubble Generator Ring 21 is immersed in the solution Do Not Overfill See fig 6 3 After filling is completed the rubber Storage Tubing 25 may be removed completely Recap the soap dispenser bottle for later use NOTE If the Flow Cell Assembly is not going to be used for a prolonged period of time reinstall the rubber Storage Tubing between the inlet and outlet bosses 23 26 This will prevent evaporation from occurring which may cause the solution s concen trations to alter D Printer Connection if applicable 1 Connect printer cable to Printer Jack connector 18 on the left side of the Control Unit Be sure to properly match up connectors before engaging Soap Solution Dispenser Flow Tube Bubble Initiate inlet Boss Bubble Gellarat b Solution AEE Solution Rese pty
18. ows easy removal to facilitate cleaning 1 Electrical Interface The Electrical Interface provides power to the sensing system as well as transmits timing information to the Control Unit 2 The Control Unit base See fig la The Control Unit 14 contains a crystal controlled timing system a micro processor control system and an LCD readout for displaying flow and messages The Control Unit also contains switches for Reset 17 Delete 16 and Auto Averaging 15 functions as well as an Printer Jack 18 interface port for direct connection to a hard copy printer LED indicators are provided to note Charging 11 and Run calibration 20 operation a Power 12 switch turns the Control Units power on and off b Charge Indicator LED 11 lights when the charger is plugged into the Charging Jack c Printer Jack 18 provides interface for auxillary printer d Reset 17 sw push button deletes all current information for the micro processor in order to initiate a new sequence e Delete 16 push button automatically deletes false informa tion from the average and will reset the average and sample number to the previous reading When a printer is in line this will indicate a minus symbol and the average will return to the previous value f Average 15 push button when pushed and held will display the average of the previous sequence of readings When released will display the last actual reading and when re
19. pushed and held will show the number of samples in the sequence with display information S sample Releasing the button will automatical ly bring the display back to the last reading g Sequence Run Indicator LED 20 indicates bubble sequence by lighting as the bubble passes between the two sensors The LCD 19 will be blank The Run signal will also light when turning on the Control Unit and will extinguish after unit has finished it s initial sequence check h Low Battery sw will indicate on the LCD display 19 if insuffi cient battery voltage is available to operate the unit proper ly Since power for all Control Unit functions is derived from the rechargeable NiCad battery the batteries must be fully charged for proper operation A Low Battery indication will also appear initially when turning the Control Unit ON as a sequence check of the unit s electronics i Cable Assembly 5 mates with the Connecting Jack 6 in the back of the Sensor Block It provides power for the sensing system information regarding cell size and control of the timing information to the micro processor j Timing System The quartz controlled timing system controls infra red sensor activation to assure maximum calculation accura cy k Micro Processor controls the timing and mathematical data processing to provide optimum flow measurement characteristics This programmable micro processor can be upgraded as new programs bec
20. r is powered by same 2 The Average Reset and Delete buttons on the Control Unit are used in conjunction with the printer during calibration They are as follows a Average In order to display average and number of samples depress and hold average button Releasing the button will display the last reading Repressing will display the number of samples accumulated for that averaging sequence Releasing the button will once again display the last flow reading Additional pressing and holding will repeat this sequence b Reset To reinitiate the sequence for additional pumps hit the Reset Button This will zero out sample and average regis ters within the Control Unit and will cause the printer to index one line and reprint headings This denotes the start of a new sequence The Reset Button may also be used if a malformed bubble is generated and has not been been subtracted from the average by use of the Delete Function c Delete To delete obvious false readings press the Delete Button and subtract the false reading from the average This resets the average and sample number back to the previous reading 5 Storage amp Maintenance a Storage Daily Take Down 1 Turn the Power of the Printer Module OFF All data stored in the printer cannot be reviewed once this is done 2 Turn off the Power of the Control Unit 3 Turn off the Power of the Sampler being calibrated b Storage Long Term If the Printer Module is not to be u
21. sconnect the Cable Assembly from the back of the Sensor Block on the Flow Cell Assembly 2 Remove the Flow Cell Assembly from the Control Unit base Remove the Cell from the base in the reverse order in which it was mounted 3 Pour soap solution out of the Bubble Generator through the Lower Inlet Boss 23 By holding Bubble Generator horizontally and with the inlet boss facing down tilt at a 45 degree angle Con tinue until all of soap solution has poured out See fig 7 Figure 7 13 4 Flushing the Bubble Generator Clean There are 2 methods in which to cleanse the Bubble Generator a The unit may be flushed clean by connecting Storage Tubing to the Upper Outlet Boss and continuously running water through the generator until water runs clear Maintain a horizontal position with cell bosses facing down and flush for 15 30 seconds Then remove tubing and rock the cell in a see saw fashion to empty out all excess water Replace Seal Tubing between the Upper and Lower Cell bosses See fig 8 S Inlet Bos Storage Tubing Figure 8 Assembl Ki Cell or see Figure 9 b Remove the Sensor Block Assembly by loosening the 2 holding screwsand sliding the block out from between the Upper amp lower cell chambers Remove Safety Tape Using a small flat blade screw driver lift off the Damper Plate 2 using the notch between the upper chamber and the lid Remove the Spacer 3 and then the Bubble
22. se The liability of the Seller shall be limited to the repair or replacement of materials or parts as above set forth Limitation of Liability The Seller shall not be liable for any claim for consequential or special loss or damage arising or alleged to have arisen from any delay in delivery or malfunction or failure of the equipment The Seller s liability for any other loss or damage arising out of or connected with the manufac ture sale or use of the equipment sold including damage due to negligence shall not in any event exceed the price of the equipment supplied by us Service Policy For a minimum fee of 95 00 Gilian Instrument Corp will overhaul repair and or replace minor components and recalibrate one _ Cilibrator Gilian reserves the right to proceed with additional repairs up to a maximum cost of 130 00 per Gilibrator without notifying the customer If major components must be replaced Gilian will notify the customer before proceed ing with repairs When the instrument s is returned please include a purchase order marked Repair Cost not to Exceed 100 Without Customer Authorization Also include company name return shipping address contact name and phone number serial number s of calibrators date of purchase and description of problem Return to Gilian Instrument Corp 35 Fairfield Place W Caldwell NJ 07006 6206 Att Gilibrator Repairs Dept
23. sed for long periods of time the following steps should be taken to keep the unit in proper working order 1 Disconnect the interconnect cable from the Control Unit and from the Printer Module 2 Store in original packaging or carrying case c Cleaning Use a damp cloth with warm water Never use ALCOHOL ACETONE or any other harsh cleaners 22 Section 7 Specifications for the Gilibrator System Flow Cell Assembly Ranges Available High Flow Cell 2 30 LPM Standard Flow Cell 20 cc min 6 LPM Low Flow Cell 250 ce min Operational Features Direct Flow Readings The easy to read LCD read out on the Control Unit base instantly displays flow readings Auto Averaging A switch on the Control Unit provides Auto Averaging of the successive flow readings Flow Delete Function A Flow Delete switch allows the user to delete erroneous flow readings from the Auto Averaging function For example if a double bubble or other obvious malfunction is observed that reading can be subtracted from the auto average and will automatically reset the average to it s previous value Programmable Update The micro processor electronic design allows program mable updates to assure a long useful service life Mechanical Features Pulsation Damper The pulsation damper removes pulsation from the flow source for higher accuracy sampling as well as eliminating pulsations back into the flow source during bubble
24. sing the button will display the number of samples accumulated for that averaging sequence and releasing will once again display the last flow reading Additional pressing and holding will repeat this sequence 2 Delete 16 To delete obvious false readings push the Delete Button which will automatically delete the false information from the average and reset the average and sample number back to the previous reading 3 Reset 17 To reinitiate the sequence for additional pumps hit the Reset Button This will zero out all sample and average registers within the Control Unit and will cause the printer to index oneline and reprint sequence headings This denotes the start of a new sequence The Reset Button is also used if a malformed bubble is generated and has not been been subtracted from the average byuse of the Delete Function 1 Section 5 Storage Maintenance Storage A Take Down Turn OFF the Control Unit the sampler and the Printer Module if applicable B For Daily Use H Gilibrator is to be used daily it is recommended that the air sampler be removed and the Storage Tubing be replaced between the upper and lower cell chambers Plug in the Charger and connect into the Control Unit Charging Jack 13 Recharge overnight for next day usage C Long Term Storage If the Cilibrator is not to be used for long periods of time the following steps should be taken to keep the unit in proper working order 1 Di
25. that the unit is charging properly Allow the battery system to charge for 14 hours prior to operation B Mounting the Flow Cell Assembly See fig 2 1 Select the Flow Cell Assembly to cover the flow range required 2 The bottom of the Flow Cell Assembly employs a quick mount feature The base of the Flow Cell Assembly is positioned onto the mounting plate 10 of the Control Unit 14 Figure 2 Mounting late Mounting Pins Cell Assembly bottom Control Unit 3 Engage the pin of the cell assembly base into the mounting plate of the Control Unit When the Flow Cell Assembly is properly engaged the base of the cell will be flush to the mounting plate and the cell assembly will face towards either the right or left side See fig 3 Figure 3 E Flow Cell Assembly _ ema Mounting Pins Mounting Base ae Control Unit Cell Assembly Mounted 4 Grasp the bottom cell chamber and rotate the cell until it clicks in CAUTION Always engage disengage the cell by grasping and rotating only the bottom cell chamber The cell assembly will now face forward see Fig 4 Ce gt n Figure 4 Flow Cell Assembly Firmly twist cell into the locking position iz Control Unit 5 Insert the Control Units Cable Assembly 5 into the Sensor Block Connecting Jack 6 located on the back of the Sensor Block 8 See fig 5 Figure 5 A i M Sensor Block Connecting Jack C