Home

User Manual - Chauvin Arnoux Group

image

Contents

1. Voltage Range Adj Voltage 1 40 to 5100V Adj Voltage 2 in steps of 10V from 40 to 1000V Adj Voltage 3 in steps of 100V from 1000 to 5100V There are three adjustable settings possible To change the value for Adjustable Voltage 1 2 or 3 use the W button To select the Adjustable Voltage parameter to program press the button to highlight the voltage to be used for that selection then use the A and V buttons to increment or decrement the value When the desired voltage is on screen press the lt button to move the cursor back to the parameter selection position TIMED RUN h m This function lets you set a time from 1 minute to 49 hours and 59 minutes to run an insulation resistance test The Model 5070 will automatically end the test at the end of the timed run Range 00 to 49 01 to 59 To change the length of time that a test will run press the button to move the blinking cursor to the hour in TIMED RUN Use the A and V buttons to increment or decrement the hours Next press the to select the minute value Use the A and V buttons to increment or decrement the minutes When finished press the lt button to move the cursor back to the parameter selection position Megohmmeter Model 5070 21 SAMPLE TIME m s Data from a timed run test can be stored in the Model 5070 at an interval you select This storage interval can be as fast as once every 5 seconds to as s
2. Case 1 Switch Positions 500V 2TO Input voltage A 230V AC 1000V 4TQ Frequency 50 0 Hz 2500V 10TQ Input current 24 6nA 5000V 10TO Date 09 28 2003 Time 22 39 Megohmmeter Model 5070 55 Case2 You select an insulation measurement with one of the selectable test voltages Switch Position 50 to 5000V You can choose from one of the three selected voltages predefined in SET UP using the A and V buttons selecting adjustable voltage 1 2 or 3 or define another voltage by selecting it with the gt button to highlight the test voltage and adjusting it with the A and V but tons The new selected voltage will not be saved when the instrument is shut off or the rotary switch is moved to another position Case 3 You select an insulation measurement with a test voltage that varies in steps this is the Step Function mode Switch Position Step Voltage You can choose from the three stored step voltage schemes using the A and Y buttons m ADJUSTABLE VOLTAGE 2 2300 V Input voltage 20V AC Frequency 50 0 Hz Input current 24 6nA Date 09 28 2003 Time 22 39 i STEP FUNCTION 1 Min 2300V F Max 3900V Test Run Time 08 38 30 Input voltage 1V AC Frequency 50 0 Hz Input current 24 6nA Date 09 28 2003 Time 22 39 Pressing the START STOP button immediately begins the measurement process If the beeper is selected as ON in the SET UP mode an audible beep is emit ted every 10 se
3. MAXIMUM OUTPUT VOLTAGE The Model 5070 provides the ability to limit the maximum test voltage to a value you specify from 40 to 5100 volts When programmed the instrument will not gen erate a higher voltage to conduct the test even if the switch position indicates a higher voltage For example if you set the maximum voltage to 1250 volts and you place the rotary switch in the 5000 volt position and start a test the Model 5070 will only output 1250 volts Range 40 to 5100V 24 Megohmmeter Model 5070 SET DEFAULT PARAMETER This function allows you to reset all the user programmable functions back to the factory defaults CLEAR MEMORY The clear memory function allows you to selectively erase individual tests or all tests from the internal memory of the instrument A warning will appear first to ensure that you don t do this by accident V DISTURBANCE V OUTPUT see Appendix C This ratio defines the maximum allowable disturbance voltage depending on the selected measurement voltage If the V Disturbance exceeds the value of the ratio the instrument stops the test in process If the voltage exceeding the ratio is pres ent at the device under test before the test is started a test will not be allowed to start Three programmable values are 3 10 and 20 of the measurement voltage The default is 396 Example If a test is to be performed at 1000V and the V Disturbance V Output is set at 10 the presence of
4. o oo coat rnnt rrt etu patet dpa ener 5 1 4 Ordering Informalion itt tren e hec rie eai 6 1 4 1 Accessories and Replacement Parts ssususss 6 1 5 Accessory Information essem 6 1 5 4 DataView Software 6 2 PRODUCT FEATURES eeeeeeeeeeeen ener nnne nnne nnn 7 2 1 Control Feat res ere iie erede itte a Sa a ad EEE 8 2 2 Display Features esent eiii tede dee it een ect 10 3 SPECIFICATIONS 1eeeeeeeeee eren 11 3 1 Reference Conditions iere eterne 11 3 2 Woltage eO asiste xe ocu Re ciiin suede Ecol 11 39 Ins lation Resistance ini tein cire e eed rt tea 11 34 Powet SUpply iet te te teet i eene 16 3 5 Environmental Specifications saiisine 17 3 6 Mechanical Specifications sssssssssssse eene 17 37 gt Safety SpecifICatiOnis riui roit nice aiae 17 3 8 Variations in Operating Range ssse 18 4 SPECIAL FUNCTIONS eeeenneeeeeere eene nnne nnn 19 44 SET UP FUNCION dras cic eoa cibos cec eere 19 4 1 1 Default Configurations ees 19 4 1 2 Instrument Configuration Parameters sess 20 44 3 SET UP Menu 5er E EEE 27 4 2 MODE PRINT Bu tton inea cereis recen tn eua 28 4 2 1 Primary Function MODE seem 28 4 2 2 Secondary Function PRINT seseeeeee 32 Mego
5. Line two indicates the Model 5070 s Instrument Number and Software Version The selection cursor will be blinking and positioned to the left of the Display Contrast parameter To adjust the Display Contrast parameter press the gt button to move the blink ing cursor to display contrast value The default value is 80 Use the A and V buttons to lighten or darken the display The higher the number the lighter the display will be When finished press the 4 button to move the cursor back to the parameter selection position The arrow buttons can be used to modify any parameter Megohmmeter Model 5070 27 4 2 MODE PRINT Button 4 2 1 Primary Function MODE The primary function of this button is used before the measurement takes place to define the measurement conditions Z NOTE This button is inactive in the Step Voltage and SET UP positions Press the MODE button once to access the list of possible measurement modes Select the mode using the A W and 4 gt buttons To validate the mode selected press the MODE button again The measurement mode choices are as follows MANUAL STOP This is the conventional quantitative insulation measurement mode The measurement is started by press ing START STOP button and stopped by pressing START STOP button again In this mode the user determines the duration of the test which will be indi cated by the elapsed time indicator MANUAL STOP DD The meas
6. Capacitance 320nF i ADJUSTABLE VOLTAGE 2 2300 V Test Run Time 00 10 00 Input voltage 0 1V AC Frequency 0 0 Hz Input current 20 6pA Date 10 18 2003 Time 09 49 750 Megohmmeter Model 5070 Information Displayed Measured resistance DC test voltage Leakage current Measurement duration DAR PI Capacitance values After 1st Press on DISPLAY Test type DC test voltage Programmed duration of the test AC DC input voltage Frequency Leakage current Date time After 2nd Press on Display AC DC input voltage Frequency Leakage current Voltage bargraph 39 TIMED RUN DD mode BEFORE the measurement G DD i ADJUSTABLE VOLTAGE 2 2300 V Test Run Time 00 10 00 0 1V AC 0 0Hz 20 6pA Time 09 39 Input voltage Frequency Input current Date 10 18 2003 Information Displayed Test type and DC test voltage Programmed duration of the test AC DC input voltage Frequency Residual current Date time DURING the measurement DD N 234 5MO 2307V 24 6pA Elapsed time 00 09 43 kQ MQ GQ TQ 10 100 1 10 100 1 10 100 1 10 Information Displayed Measured resistance DC test voltage Leakage current Remaining measurement time Insulation resistance bargraph 40 500 750 Information after pressing DISPLAY AC DC input voltage Frequency Residual current Voltage bargraph pp N 234 5MO 2307V 24 6pA Elapsed time 00 09 43 DAR 30 60 PI 1 10 Capacitance
7. PI Settings 1st PI time 1 00 mm ss 00 30 59 59 Buzzer 2nd PI time 10 00 mm ss 00 30 59 59 7 C OFF Test Voltage Alarm Set Points MQ 500 lt v 5000k 30k2TQ Auto Power OFF MM DD YYYY DD MM YYYY Read from Inst Save to Disk Load from Disk ON E Defaults C ON M 2 1000V lt v 1000k 100k 4T Q o OFF M Q 2500V v 2500M 300k 10T Q m Disturbance voltage MQ 5000V lt 5 000M 300k10TQ limitation e 3x C 10 IV Enable all test voltages C 20 Figure 7 5 Set Clock Clear Memory Download Variable Voltages for programming the three selectable voltages and their alarm set points Megohmmeter 5070 Setup General Settings Variable Voltages Ramp Voltages Temperature Variable test voltage Alarm Set Points M Q Variable voltage V1 50 40 5100 V lt v 50 00k 30k 10T 2 M Q Variable voltage V2 100 40 5100 V lt v 1000k 30k10T Q M Q Variable voltage V3 250 40 5100 V lt v 2500k 30k 10T Q Figure 7 6 76 Write to Inst Read from Inst Save to Disk Load from Disk Defaults SetClock Download Megohmmeter Model 5070 Ramp Voltages for programming the three ramp profiles Megohmmeter 5070 Setup General Settings Variable Voltages Ramp Voltages Temperature Ramp 1 Ramp 2 Ramp 3 o Miete Voltage Duration Voltage Duration Voltage Duration Read from Inst 40 5100 V h m 40 5100 V h m 40 5100 V h m Step1 5
8. and trend summary frames are updated to display the values for the selected test 7 7 2 Title Page An example of the second worksheet is shown below This is the title page for the report It will contain all of the information you entered in the session properties box when saving the data Test Report 4 mas t Box x x 6 28 02 Tex ox Operator Test Site AEMC Instruments Test Company John Smith Test Person 200 Foxborough Blvd 101 Test Street Foxborough MA 02035 Test City Test State 12345 508 598 2115 Test Phone Comenta Test performed on 280V 3 Phase motor Session cummar Database C Program Files YDataView Samples MegohmSample dvb Database 9 AEMC INSTRUM EATS Figure 7 14 This worksheet contains the following frames Title Box containing the text Trend Summary Report Operator Summary containing operator information Site Summary containing test site information Comments Summary containing typed in operator comments Database containing the name of the database used to generate the report Each of these frames is customized by using the associated properties dialog box The properties dialog box for a frame is displayed by selecting the frame right clicking and selecting Properties from the list displayed i In addition to the pre designed report templates DataView allows you to totally configure reports to
9. of Reading 10 00 to 99 99pA 10nA 100 0 to 999 9pA 100nA 1000 to 30001A 1pA 10 of Reading Calculation of Terms DAR and PI Resolution Accuracy 0 02 to 50 00 0 01 5 of Reading ict Calculation of Term DD Resolution Accuracy 0 02 to 50 00 0 01 10 of Reading 1ct 3 4 Power Supply Rechargeable battery NiMh 8 x 1 2V 3 5Ah Recharge 85 to 256V 50 60Hz Minimum Battery Charge Life per NF EN 61557 2 TestVottage Nominal Charge with 25 pause between each measurement 500V 500kO 6500 1000V 1MO 5500 2500V 2 5MO 4000 5000V 5MQ 1500 Average Battery Life The operating time will be 15 days or 3 weeks based upon a 10 minute long PI measurement Recharge Time 6 hours for 100 capacity 10 hours if the battery is completely drained 0 5 hours for 1096 capacity charge life 2 days approximately 16 Megohmmeter Model 5070 AN NOTE It is possible to recharge the batteries while performing insulation measurements provided that the values measured are higher than 20M In this case the recharging time is higher than 6 hours and depends on the frequency of the measurements 3 5 Environmental Specifications Operating Range 14 to 104 F 10 to 40 C during recharging of batteries 14 to 131 F 10 to 55 C during measurement 10 to 80 RH Storage 40 to 158 F 40 to 70 C 10 to 90 RH Altitude lt 2000m 3 6 Mechanical Specifications Case Dimen
10. where Vpeak is the external voltage peak or DC at the and terminals of the Model 5070 diSt is the coefficient that can be selected in SET UP at 3 default value 10 or 20 VN is the test voltage selected for the insulation measurement Example B If the conditions defined in Example A were programmed and the test was started a disturbance voltage change of voltage of 2700V would abort the test Vpeak 0 03 1 05 2500V 2700V Note The dlSt factor is selected to optimize the measurement build up time If there is no leakage voltage dlSt can be adjusted to its minimum value to obtain the shortest possible measurement build up time If there is a large leakage voltage dlSt can be increased so that the measurement will not be interrupted as soon as a negative alternation occurs during the genera tion of the test voltage this helps optimize the measurement build up time in the presence of a leakage voltage Pressing the START STOP button again stops the measurement If the programmed time test mode Timed Run or Timed Run DD was selected as the test mode the measurement is stopped without pressing the START STOP button at the end of the test time Megohmmeter Model 5070 57 Similarly if the DAR or PI mode is selected as measurement mode before the test the measurement is stopped automatically after the time programmed to calculate them has occurred time defined in SET UP A number of speci
11. 0 Instr Nr 9600004 SW Version 1 2 3 Display contrast 80 Alarm Settings Adjustable Voltage 1 50V Adjustable Voltage 2 100V Adjustable Voltage 3 250V Timed Run h m 0 10 Sample Time m s 0 10 DAR s s 30 60 PI m m 1 010 Set Step Function 1 Set Step Function 2 Set Step Function 3 Temperature Unit Celsius Default probe temperature 23 C Rc reference temperature 40 C AT for R 2 10 C Calculate AT from Memory Maximum Output Voltage 5100V Set Default Parameter Clear Memory V Disturbance V Output 3 Buzzer Power Down Baud Rate 9600 RS232 Units Europe Date d m y 04 02 2004 Time h m 15 47 Megohmmeter Model 5070 19 4 1 2 Instrument Configuration Parameters DISPLAY CONTRAST This function allows you to change the lightness and darkness of the display to best suit the environment Range 0 to 255 The display is no longer legible above 130 To adjust the contrast press the button to move the blinking cursor to the dis play contrast value The default value is 80 Use the A and V buttons to lighten or darken the display The higher the number the lighter the display will be When finished press the lt button to move the cursor back to the parameter selection position ALARM SETTINGS This function allows you to select a low insulation resistance value that will cause an alarm symbol to appear on the display and will cause the buzzer to emit a con tinuous tone when the measured insulati
12. 100 volts 1000V 10 100 before the test starts will inhibit the test Range 3 10 or 20 BUZZER The Model 5070 is equipped with a buzzer that will emit an audible tone when a key is pressed at regular intervals during a timed test or will be on continuously during an alarm trip This function lets you toggle the buzzer on or off Range ON or OFF POWER DOWN The Model 5070 has a power save feature which turns the unit s display off after 5 minutes of no activity if a timed test is not in progress This function lets you toggle this feature on or off Range ON or OFF Megohmmeter Model 5070 25 BAUD RATE This function lets you program in the communication speed between the Model 5070 and your computer It also lets you select a parallel communication mode for direct printing Press the down arrow YW button until the blinking cursor is to the left of Baud Rate To adjust the baud rate press the right arrow 9 button to highlight the current setting Next press the up A or down V arrow buttons to select the desired baud rate Your choices are 300 600 1200 2400 4800 9600 or parallel Press the right arrow 9 button after you have selected the baud rate to complete the process Range 300 to 9600 RS 232 or Parallel UNITS The Model 5070 can display the date in either US m d y or European d m y fashion This choice can be toggled here Press t
13. 100GQ 8s 90s 5MQ 4s 16s 5000V 100GQ 8s 120s Typical discharge time for a capacitive component to reach 25Vpc Initial Voltage 500V 1000V 2500V 5000V Discharge time C in pF Cx3s Cx4s Cx4s Cx7s Megohmmeter Model 5070 13 Typical changes in test voltages as a function of the load Output voltage V Output voltage V 14 600 500V Range 500 400 300 200 100 0 0 01 1200 1000 0 1 Resistance MQ 1000V Range 800 600 400 200 0 01 Resistance MQ Megohmmeter Model 5070 2500V Range Megohmmeter Model 5070 3000 2500 o 2000 D S 9 41500 5 Q 5 1000 O 500 0 0 1 1 10 Resistance MO 5000V Range 6000 5000 v 4000 D S 9 3000 5 Q 5 2000 O 1000 0 0 1 1 10 Resistance MO 15 Capacity Measurement after discharge of tested component Resolution Accuracy 0 005 to 9 999uF 1nF 10 of Reading 1ct 10 00 to 49 99uF 10nF Leakage Current Measurement Resolution Accuracy 0 000 to 0 250nA 10A 15 of Reading 10cts 0 251 to 9 999nA 3 10 of Reading 10 00 to 99 99nA 10pA 100 0 to 999 9nA 100pA 1 000 to 9 999pA 1nA 5
14. Duration Voltage N h mm def actual 1 0 10 1000 V 1020 V 2 627 GOhm 2 0 10 2000 V 2043 V 2 411 GOhm 3 0 10 3000 V 3060 V 2 347 GOhm 4 0 10 4000 V 3755 V 2 182 GOhm 5 0 10 5000 V 3237 V 2 023 GOhm AR 604 GOhm AV 4000 V AR R AV ppm V 57 ppm Capacitance 110 nF OBJECT 01 TEST 03 EQ sonos tetenteni met Date of next test Pow P REMARKS seien d adc te nme y Ee cm Operator Gg aeda i ode eec aes a Signatures e Ree URS Rma Ue 70 Resistance Megohmmeter Model 5070 CHAPTER 7 DATAVIEW SOFTWARE 7 1 Installing DataView INSTALL DATAVIEW BEFORE INSTALLING THE USB DRIVERS AND BEFORE CONNECTING THE INSTRUMENT TO THE COMPUTER NOTE When installing the user must have Administrative access rights during the installation The users access rights can be changed after the installation is complete DataView must be reinstalled for each user in a multi user system USB Flash Drive Install 1 Insert the USB stick into an available USB port wait for driver to be installed 2 If Autorun is enabled then an AutoPlay window should appear as shown J Install Ji PDFXChange f CEES oee Javorov f E us J use FT232 DATAVIEW LI J Usb2 A Ji User Manuals 8 AUTORUNINF DocsEN ini DocsFR ini 8 DocsSP ini General options A Open folder to view files using Windows Explorer x BA Setup exe Bp or Satupini 2 g Windows ReadyBoos Upgrades ini 2 2 View more AutoPlay
15. SIOTdge s oe eite En ien e aetates Dee dure oes 86 APPENDIX A UTILIZING THE GUARD TERMINAL 87 APPENDIX B SERIAL CABLE REQUIREMENTS 89 APPENDIX C V DISTURBANCE V OUTPUT FEATURE 90 Repair and Calibration cuc iuo rete ortos tete te coenae cet teeta 91 Technical and Sales Assistance sssss eee 91 limited Waranty erri recette eret e s eR retiro de eas 92 Warranty Repairs 5 ce iei eerie cod b recette Eee Re eo ER Re EEEE SUR ER ERES 92 Megohmmeter Model 5070 3 CHAPTER 1 INTRODUCTION WARNING These safety warnings are provided to ensure the safety of personnel and proper operation of the instrument Do not attempt to perform any tests with these instruments until you have read the instruction manual Safety is the responsibility of the operator Tests are to be carried out only on non energized circuits Check for live circuits before making resistance measurements Safety Check High voltage is present as is the sample connected to it Anyone performing or assisting in testing must follow all safety precautions to prevent electrical shock to themselves and to others AEMC considers the use of rubber gloves to be an excellent safety practice even if the equipment is properly operated and correctly grounded When testing samples with a capacitive component make sure they have been properly disc
16. Timed Run test the list of recorded samples To stop printing move the position of the selector switch to any other position 64 Megohmmeter Model 5070 Depending on the function used the following formats are obtained Any measurement except step function measurements AEMC Instruments Model 5070 000 001 Instrument number OBJI ral Q r3 01 TEST 01 INSULATION RESISTANC Date d EST 01 31 2003 14h55 00 15 30 23 C Relative humidity TE a 1000 V 385 GOhm Starting time Running time Temperature n Test voltage Insulation resistance Rc calculated resistance 118 5 GOhm at reference temperatur 40 C with AT for R 2 10 C DAR 1 30 1 234 PI 10 1 2 345 DD ul i Capacitance 110 nF Elapsed time Utest Resistance 00 00 10 1020 V 35 94 GOhm 00 00 30 1020 V 42 0 GOhm 00 00 50 1020 V 43 5 Gohm GE gue uod so uu me eoe M SORIA E iu sev ui ener do Megohmmeter Model 5070 printed only in MR mode after timed run test 65 Date of next test iced cac ere seria Operatory auct ga annaa deere RD LET Case Yrs SCTOnatu ei 2x uev ge Y e ques Step function measurement AEMC Instruments Model 5070 Instrument number 000 001 OBJECT 01 TEST 01 STEP FUNCTION TEST Date 01 31 2003 Starting time 14h55 Running time 00 00 50 Temperature 23 C Relative humidity PEE printed only in MR mode Step
17. now select another option to download e g Adobe Reader or close the window 17 Restart your computer then connect the instrument to the USB port on the computer 18 Once connected the Found New Hardware dialog box will appear Windows will complete the driver installation process automatically Shortcuts for DataView and each instrument control panel selected during the installation process have been added to your desktop NOTE If you connected your instrument to the computer before installing the software and drivers you may need to use the Add Remove Hard ware utility to remove the instrument driver before repeating the process 74 Megohmmeter Model 5070 7 2 Connecting the Model 5070 to your Computer The Model 5070 is supplied with a serial interface cable necessary for connecting the instrument to the computer This cable Cat 2119 45 is equipped with a 9 pin connector on each end see diagram in Appendix B To connect the megohmmeter to your computer Connect the 9 pin connector from one end of the cable to the serial port on the front panel of the Model 5070 megohmmeter Connect the 9 pin connector from the other end of the cable to an avail able serial port on the back of your computer 7 3 Establishing Communication to the Instrument Double click the Megohmmeter Icon that was created during software instal lation located on the desktop The Connection dialog box will appear Make sure that the
18. options in Control Panel UpgradesES ini UpgradesFR ini Megohmmeter Model 5070 71 NOTE If Autorun is disabled it will be necessary to open Windows Explorer then locate and open the USB stick drive labeled DataView to view the files on the drive 3 Inthe AutoPlay window select Open Folder to view Files 4 Double click on Setup exe from the opened folder view to launch the Data View setup program i NOTE If installing onto a Vista based computer the User Account Control dialog box will be displayed Select the Allow option to proceed 5 ASet up window similar to the one below will appear E Francais Espafiol Exit Select an option from the list below Options Description PEEN Adobe Reader Dataview Update Firmware Upgrades Ground Tester Model 6470 B 6471 and 6472 User Manuals Megohmmeter Model 1060 5060 5070 6550 and 6555 Micro Ohmmeter Model 6240 and 6250 Power Energy Logger Model PEL102 and PEL103 Power Meter Model 8220 Power Quality Logger Model PQL100 and PQL120 PowerPad Model 3945 and 3945 B PowerPad II Model 8335 PowerPad Jr Model 8230 Power Clamp Model 407 and 607 Sentinel Data Logger Simple Logger II DTR Model 8510 Installs the English version of DataView onto this computer This version of DataView supports the following instruments Figure 7 1 There are several different options to choose from Some options require an interne
19. provides the ability to display and store temperature corrected Insula tion Resistance readings Therefore tests compared to historical data can all be referenced to a common temperature for better qualitative analysis NOTE The TEMP function can be activated only after a measurement has been completed either before or after it is stored e If the result of your measurement is outside of the instrument s range the display shows the or symbol next to the reading the Tempera ture Correction function cannot be applied Temperature correction is not available on Step Voltage tests 46 Megohmmeter Model 5070 Procedure NOTE You have made a measurement and not yet stored it Check that the result is within the Instrument s measurement range Enter the TEMP mode by pressing the 2nd TEMP buttons Enter the estimated temperature Probe Temperature at which you made the measurement by default the instrument offers the value set in SET UP With the blinking cursor next to TEMPERATURE Probe Temperature press the gt button to activate probe tempera D Probe Temperature 23 7 C ture Use the A or V buttons to iban x iod pie increase or decrease the tempera Cnererence 1emperature i o P AT for R 2 23 0 C R measured 273 7MQ Press the lt or gt buttons to move Rc at 28 5 C 328 5MO the cursor back to the selection position Next make sure Resistance Correction is turned on to pe
20. tem perature defined in SET UP see 4 10 6 2 Step Function Mode This function is based upon the principle that ideal insulation produces the same resistance no matter what the test voltage applied Any negative variation ofthis resistance means thatthe insulation may be defective The resistance of defective insulation will decrease as the test voltage increases This phenomenon is rarely observed with low test voltages therefore at least 2500V must be applied The usual test condition is a voltage increasing in 5 steps each lasting for 1 minute minimum Assessment of the result A deviation of the resistance f test voltage curve that exceeds 500ppm V generally indicates the presence of mold or other deterioration Alarger deviation or a sudden drop indicates the presence of localized physical damage such as arcing or perforation of the insulation etc 60 Megohmmeter Model 5070 Procedure n the SET UP menu select Set Step SET UP Function 1 2 or 3 by scrolling down Ramp 3 definition to set Step Function 1 2 or 3 using Step Voltage Duration h m the V button Press the gt button to 7 M4 enter the Set up mode for the pro B 1000V 01 00 fi 2000V 01 00 ile selected The screen on the right 3000V 01 00 shows the default profile for Step Function Number 3 4000V 01 00 unction Number 3 5000V 01 00 Define the Step Function and the Total duration h m 05 00 desired number of
21. the 2nd TEMP buttons at the same time as the measurements without applying the correction Resistance Correction OFF In the SET UP mode select Calculate AT from Memory and press the gt button The display shows all values recorded with a temperature Select at least 3 measurements using the A V 4 gt buttons AT is calculated and recorded automatically once 3 stored measure ments have been selected and will be updated if more measurements are selected The larger the number of measurements the more accurate the calcula tion of AT AMS NOTE This calculation is available only for resistance values lt 20060 SET UP Instr Nr 9600004 SW Version 1 2 3 Calculate AT from Memory Maxumum Output Voltage 5100V Set Default Parameter Clear Memory V Disturbance V Output 10 Buzzer on Power Down on Baud Rate 9600 RS232 Megohmmeter Model 5070 AT Calculation for R 2 Obj Test 47 99 13_59 13 58 IDA 02 02 02 01 SET UP Res 228 5MQ 208 5MQ 178 5MQ 328 5MQ 328 5MQ 328 5MQ Volt 5078V 5078V 5078V 5078V 5078V 5078V 23 7 C Temp 23 C 30 C 37 C 23 C 23 C 23 C 53 4 11 Maximum Output Voltage In the SET UP menu select Maximum Output Voltage pressing the Y button to highlight it and the gt button to choose it Adjust the maximum output voltage using SET UP the A Y buttons Instr Nr 9600004 SW Version 1 2 This function prohibits the u
22. to 49 Next press the button to highlight minutes Use the A or V buttons to select the desired minutes from 0 to 59 The minimum selectable duration is 1 minute and the maximum is 49 hours and 59 minutes e Press the gt button until the minute value in the sample rate is high lighted then use the A or V buttons to adjust the minute value between 0 and 5 Next press the button until the seconds value is highlighted Use the A or V buttons to select the desired seconds from 1 to 59 The shortest possible sample rate is 5 seconds and the longest sample rate is 5 00 minutes When finished with Duration and Sample selections press the lt or gt buttons again until the blinking cursor is at Timed Run You are now ready to begin a Timed Run Test for the selected Test Voltage Press the Yellow START STOP button to begin the test The display will briefly show OK followed by the active test screen When the measurement is started the timer counts down showing the time remaining in the measurement When the Remaining Time is zero the measurement is stopped During the timed run test the intermediate samples are automatically stored They are used to plot a curve of insulation resistance vs time This curve can be dis played after the measurement is completed by pressing the Yellow 2 button and the GRAPH button as long as no new measurement has been started The samples and the curve are automaticall
23. your needs Refer to the DataView HELP file on Templates to learn more about templates 84 Megohmmeter Model 5070 CHAPTER 8 MAINTENANCE Use only factory specified replacement parts AEMC will not be held responsible for any accident incident or malfunction following a repair done other than by its service center or by an approved repair center 8 1 Recharging the Battery If the symbol is flashing the battery needs to be recharged Connect the instrument to the 120Vac power cord via the connector charging starts automatically even if the instrument is turned off If the instrument is charging in the OFF position the battery symbol is displayed and the 3 bars flash throughout the charging Charging bat tery is also indicated When the battery is full the symbol and its 3 bars are lit steadily and Charging Full is indicated Ifthe instrument is charging in a measurement position the battery symbol flashes There is no full charge indication The Charging Full indication is displayed only when the instrument is returned to the OFF position If the instrument is turned on and the battery voltage is gt 8V then the normal use of the device is permitted i The battery should be replaced by an authorized repair facility recognized by AEMC Instruments WARNING Changing the battery causes data to be lost from the memory Press the MEM MR button OFF is displayed Proceed with a total clearin
24. 0 oo roo 1000 1 Save to Disk Load from Disk Step2 100 0 01 300 2000 __toed tom Dik Step3 150 0 01 500 Step4 200 0 01 700 4000 250 0 01 300 01 5000 Rit sample mm ss 0 10 n 0 0 0 Step5 01 ot Defaults 01 3000 01 10 Set Clock Clear Memory Download Figure 7 7 Temperature for programming the temperature correction variable Megohmmeter 5070 Setup General Settings Variable Voltages Ramp Voltages Temperature d White to Inst Temperature unit Read from Inst a Save to Disk Default probe temperature 23 15 C 75 C Load from Disk Defaults Ric reference temperature fo 15 C 75 C AT for R 2 10 Cisse pic Set Clock Clear Memory Download Figure 7 8 These settings configure every aspect of the megohmmeter Each field is identical to the programmable features available from the instrument front panel itself Refer to Chapter 4 for a complete description of each configurable function Megohmmeter Model 5070 77 Several of the functions are configured by typing the appropriate value in the field provided such as test run time which may be programmed from a value of one minute to 49 hours 59 minutes Others are configured by clicking on the appropriate radio button such as activat ing or deactivating the buzzer by either selecting ON or OFF The right hand side of this setup dialog box contains eight command buttons Write to Inst Programs the meg
25. 00 4000 5000 Variable Voltage V Resolution and Accuracy Chart see curves pages 14 and 15 Test 500V 1000V Voltage 500V 2500V 500V 1000V 2500V 5000V Banos 30 to 100 to 300 to 1MQ to 4 00 to 40 0 to 9 99kQ 299kQ 999kQ 3 999MQ 39 99MO 399 9MQ Resolution 1kQ 10kQ 100kQ Accuracy 5 of Reading 3cts Test 1000V 2500V 2500V Voltage 500V 1000V 2500V 5000V 5000V 5000V 400MQ to 4 00 to 40 0to 400GQ to 4 00 to Range 39996O 39 9962 399960 1 999ro 00010 399970 450000 Resolution 1MO 10MQ 100MQ 1GQ 10GQ Accuracy 5 of Reading 3cts 15 of Reading 10cts 12 Megohmmeter Model 5070 DC Voltage Measurement during insulation test Range Resolution Accuracy 40 0 to 99 9V 0 1V 100 to 1500V 1V 1 of Reading 1ct 1501 to 5100V 2V DC Voltage Measurement after insulation test Range Resolution Accuracy 25 to 5100V 0 2 Vn 5 of Reading 3cts Typical build up time for the measurement according to components tested Vdist 0 03Vn These values include the influences caused by the charge of the capacitive com ponent by the automatic range system and to the test voltage control Non Capacitive With capacitance Test Voltage Load unsmoothed of 1yuF smoothed measurement measurement 1MO 3s 4s 500V 100GQ 8s 40s 1MO 3s 4s 1000V 100GQ 8s 80s 3MO 3s 4s 2500V
26. 000 30 100 200 2500 75 250 500 5000 150 500 1000 If a sudden change in voltage occurs during the test the formula for calculating V Disturbance V Output is slightly different It is V Output 1 05 dist Using 500 volts as the test voltage and 3 as dlSt the inhibit voltage that must occur after a test is started is 500 1 05 03 540 volts The table below shows the inhibit voltages for the 4 fixed test voltage positions and the three V Disturbance V Output percentages that must occur after a test starts During a Test 90 Test V Disturbance V Output Voltage 395 10 20 500 540 575 625 1000 1080 1150 1250 2500 2700 2875 3125 5000 5400 5750 6250 Megohmmeter Model 5070 Repair and Calibration To ensure that your instrument meets factory specifications we recommend that it be scheduled back to our factory Service Center at one year intervals for recalibration or as required by other standards or internal procedures For instrument repair and calibration You must contact our Service Center for a Customer Service Authorization Number CSA This will ensure that when your instrument arrives it will be tracked and processed promptly Please write the CSA on the outside of the shipping container If the instrument is returned for calibration we need to know if you want a standard calibration or a calibration traceable to N I S T Includes calibration certificate plus recorde
27. 1V DC Frequency 0 0 Hz Input current 20 6pA Date 10 18 2003 Time 10 03 0 400 250 500 Megohmmeter Model 5070 Information Displayed Measured resistance DC test voltage Leakage current Measurement duration DAR PI Capacitance After 1st Press on DISPLAY Test type DC test voltage Programmed duration of the test AC DC input voltage Frequency Leakage current Date and time After 2nd Press on Display AC DC input voltage Frequency Leakage current Voltage bargraph 45 4 3 2 Secondary Function GRAPH At the end of a programmed Time Test Timed Run or Timed Run DD press ing the 2nd GRAPH buttons will bring up a display of the Insulation Resistance versus time curve The resolution of this presentation is GRAPH dependent on the sample time selected before the test or during set up 5078V 328 5MQ 00 02 30 The lt and gt buttons can be used to move the vertical cursor along the time axis to display the Insulation Resistance and elapsed test time at the cursor posi tion Pressing the DISPLAY button returns to Ww 0 0 30 1 00 1 30 2 00 2 30 3 00 the Digital Presentation of Results 4 4 4 TEMP Button 4 4 1 Primary Function lt Selects the parameter to be modified to the left or moves the cursor on a graph screen to the left 4 4 2 Secondary Function TEMP At the completion of a test press the 2nd TEMP buttons This brings up the Temperature Correction display This function
28. 2nd Press on Display AC DC input voltage Frequency Leakage current Voltage bargraph NOTE Because the test will stop after the DAR calculation PI will not be calculated if the time values for this test are longer than those set for DAR Megohmmeter Model 5070 43 PI mode BEFORE the measurement il FIXED VOLTAGE 5000 V Test Run Time 00 01 00 0 1V DC 0 0Hz 20 6pA Time 20 02 Input voltage Frequency Input current Date 10 18 2003 Information Displayed Test type DC test voltage Programmed duration of the test AC DC input voltage Frequency Residual current Date and Time DURING the measurement A 234 MO 5007V Elapsed time ue kQ MQ GQ TQ 10 100 l 10 100 1 10 100 i 10 Information Displayed Measured resistance DC test voltage Leakage current Remaining measurement time Insulation bargraph 44 0 499 250 500 750 Information after pressing DISPLAY AC DC input voltage Frequency Residual current Voltage bargraph A 234 5MO 5007V Elapsed time 05045 DAR 30 60 PI 1 10 Capacitance Information after pressing DISPLAY Measured resistance DC test voltage Leakage current Remaining measurement time DAR PI Capacitance in process Megohmmeter Model 5070 AFTER the measurement i 234 5MO 5007V 20 6pA Elapsed time 00 10 00 DAR 30 60 RSS PI 1 10 2 74 Capacitance 320nF m FIXED VOLTAGE 5000 V Test Run Time 00 10 00 Input voltage 0
29. 8 5 4 Residual Current Measurement sseem 58 6 OPERATION 1 eeeeeeeeeeee eene rn nnne nnn nnn nnn nnn 59 6 1 Performing Measurements ssssse emn 59 6 2 Step Function Mode uii neci n as He cet neta 60 6 3 Operation Examples etin ceder n eal adie 62 2 Megohmmeter Model 5070 6 4 Printing Measured Values MODE PRINT Button 64 6 5 Instantaneous Printing of Measurements Print Result 64 6 6 Printing Data in Memory Print Memory eeeeee 67 7 DATAVIEW SOFTWRBE i5 inkxx kimi uii oi cia nnn 71 744 Installing BataView iiec nete Een ket ge eue datu 71 7 2 Connecting the Model 5070 to your Computer sssss 75 7 3 Establishing Communication to the Instrument 75 7 4 Configuring the Instrument ssssee e 76 1 5 Runningithe Test eit terret eet evenit 79 7 6 Exporting the Report to a PDF File essssseseeeese 82 7 7 DataView Megohmmeter Templates see 83 7 7 1 Megohmmeter Summary Report Template 83 1 52 Mile Pageant eed 84 8 MAINTENANGE 11eeeeeeeee eren nnne nnnm nnn nnnm nnn nnnm nnn 85 8 1 Recharging the Battery eemper eret 85 82 FUSE Replacement rented emet ripe i neso tieu Enea tie 86 EMMEGI I 86 9 4
30. 94 8E 6 R Ohms 8 2005 6 08 47 084 PH 994 1E 6 R Ohms 3 8 2005 5 08 52 S88 7E 6 R Ohms 3 8 2005 6 08 57 254 PH 989 9E 6 R Ohms 3 8 2005 6 09 02 338 PH DBS 9E 6 R Ohms 8 2005 5 09 07 423 PH 990 2E 6 R Ohms 3 8 2005 6 09 12 508 PHM 990 7E 6 R Ohms 3 8 2005 6 09 17 593 PH 989 BE 6 R Ohms s 8 2005 s 991 7E 6 R Ohms 3 10 col a on en co eo i 8 2005 6 09 27 762 PH m 593 0E 6 R Ohms eee e eee es en Figure 7 13 This worksheet contains the following frames Channel a Summary Trend gee Summary List Trend Graph This frame provides a graphical look into the database it is attached to The database is attached to the Trend Graph Frame when it is created Channel Summary This frame displays a textual summary of the chan nel record trace selected in its parent frame Channel List This frame displays a textual summary of the channel records contained in the database attached to its parent frame Trend Summary List This frame displays a textual list of the sample points for a given test Each summary line contains user specified param eters of a group of sample points Megohmmeter Model 5070 83 The results for each test is displayed by selecting the desired test from the Chan nel List When a test is selected the graph channel summary
31. DD current DD Information after pressing DISPLAY Measured resistance DC test voltage Leakage current Remaining measurement time DAR PI Capacitance Residual current calculation of DD DD Megohmmeter Model 5070 AFTER the measurement DD a 234 5MO 2307V 24 6pA Elapsed time 00 10 00 DAR 30 60 2 24 PI 1 10 1 55 Capacitance 320nF DD current 11 55pA DD aos Information Displayed Measured resistance DC test voltage Leakage current Measurement duration DAR PI Capacitance Residual current calculation of DD DD 500 750 After 2nd Press on Display AC DC input voltage Frequency Leakage current Voltage bargraph Megohmmeter Model 5070 G DD i ADJUSTABLE VOLTAGE 2 2300 V 0 1V AC 0 0 Hz 24 6pA Time 10 05 Input voltage Frequency Input current Date 10 18 2003 After 1st Press on DISPLAY Test type DC test voltage AC DC input voltage Frequency Leakage current Date time G DD i 234 5MO 2307V 24 6pA Elapsed time 00 10 00 DAR 30 60 2 24 PI 1 10 1 55 Capacitance 320nF DD current 11 55pA DD 2 55 After 1 Minute Measured resistance DC test voltage Leakage current Total test time DAR PI Capacitance DD test current DD 41 DAR mode BEFORE the measurement il FIXED VOLTAGE 5000 V Test Run Time 00 01 00 0 1V DC 0 0 Hz 20 6pA Time 20 02 Input voltage Frequency Input current Date 10 18 2003 Information Displayed Te
32. DED eene Cat 2119 86 Lead Set of 3 Color coded 45 ft 5kV Safety with Integral Hippo Clips JUMPER LEAD NOT INCLUDED eene Cat 2119 87 Extra Large Classic Tool Bag es Cat 2133 73 Inverter 12Vpc to 120VAc 200 Watt for Vehicle use Cat 2135 43 Battery Rechargeable 9 6V ences Cat 2960 21 US 115V Power Cord ecrit tede a ach Cat 5000 14 Adapter RS 232 to USB 2 0 sse Cat 5000 60 1 5 Accessory Information 1 5 1 DataView Software DataView software makes it possible to Retrieve data from memory and plot graphs of the changes in insulation as a function of the time over which the test voltage is applied R t Print out protocols of personalized tests depending on the user s needs Create text files for use on spreadsheets Setup and control the instrument entirely by the RS 232 6 Megohmmeter Model 5070 CHAPTER 2 PRODUCT FEATURES The Model 5070 megohmmeter is a top of the line measuring instrument portable in a rugged housing with a graphic display It is capable of operating from internal battery or line power Main Functions Automatic detection and measurement of voltage frequency input cur rent Quantitative and qualitative insulation measurement Measurement at 500 1000 2500 5000Vpc or other test voltage between 40 and 5100Vopc adjustable voltage Me
33. Duration Voltage Resistance No h mm def actual 1 0 10 1000 V 1020V 2 627 GOhm 2 0 10 2000 V 2043 V 2 411 GOhm 3 0 10 3000 V 3060 V 2 347 GOhm 4 0 10 4000 V 3755 V 2 182 GOhm 5 0 10 5000 V 3237 V 2 023 GOhm AR 604 GOhm AV 4000 V AR R AV ppm V 57 ppm 66 Megohmmeter Model 5070 Capacitance 110 nF Elapsed time Utest Resistance 00 00 10 1020 V 2 627 GOhm 00 00 30 1020 V 2 627 GOhm 00 00 50 1020 V 2 627 Gohm SG Cs T Date of next test ee aatas Remarks iun aedi ovs ene eS ELE 6 ner ae Pane Operatori asa d y on 3 cR gu 6 mr a i SUOMACUR SS 3qwxeuweeg wey mra T Ee rye ei 6 6 Printing Data in Memory Print Memory When this printing mode is selected the content of the memory is displayed Stored measurements to be printed are PRINT selected using the A V lt gt but 12 15 2003 07 04 625V SESS 12 07 2003 Here the measurements selected for sii printing are RB 11 29 2003 13 59 Wms 11 24 2003 15 04 2150V sre OMA 10 29 2003 2145 975V 02 0 09 302003 0243 5000V 09 02 2003 1507 x O Once they have been selected do one of the following e To start printing press the PRINT key again e To exit without printing change the rotary switch position e To stop printing change the rotary switch position The printing of each group of data is reduced to the main results Megohmmeter Model 5070 67 Depending on the meas
34. E MEGOHMMETER 5 O 7 O EJ AEMC INSTRUMENTS User Manual VES EaS CxS SSS RSS SS oS SS SS SSS A ave E AVC INL Statement of Compliance Chauvin Arnoux Inc d b a AEMC Instruments certifies that this instrument has been calibrated using standards and instruments traceable to international standards We guarantee that at the time of shipping your instrument has met its published specifications An NIST traceable certificate may be requested at the time of purchase or obtained by returning the instrument to our repair and calibration facility for a nominal charge The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer For recalibration please use our calibration services Refer to our repair and calibration section at www aemc com Serial Catalog 2130 30 Model 5070 Please fill in the appropriate date as indicated Date Received Date Calibration Due Chauvin Arnoux Inc D d b a AEMC Instruments INSTRUMENTS www aemc com ERES IEEE S SS nS BS SoS IEEE SER ES ERS Table of Contents 1 INTRODUCTION eeeeeeeeeeneeereenennnn nennen nnn nnn nnns 4 1 1 International Electrical Symbols sn 5 1 2 Definition of Measurement Categories eese 5 1 3 Receiving Your Shiprmielit
35. ET UP mode NOTE In this mode the DAR ratio will also be calculated automatically if the times needed to calculate it are less than the second time needed to calculate the PI ratio 30 MODE Total Run Time Manual Stop Manual Stop DD Duration Sample him mis 02 30 01 40 02 31 00 Timed Run E Timed Run DD DAR s s 30 60 PI m m 1 10 MODE Total Run Time Manual Stop Manual Stop DD Duration Sample him m s 02 30 01 40 00 01 00 Timed Run Timed Run DD n ps 30 60 PI m m 1 10 MODE Total Run Time Manual Stop Manual Stop DD Duration Sample h m mis 02 30 01 40 00 10 00 Timed Run Timed Run DD DAR s s D El mm 30 60 1 0 10 Megohmmeter Model 5070 NOTE 1 Whatis the DD Dielectric Discharge index In the case of multilayer insulation if one of the layers is defective but the resis tance of all the others is high neither the quantitative insulation measurement nor the calculation of the PI and DAR quality ratios will reveal the problem This makes it important to perform a dielectric discharge test from which the DD Ratio can be calculated This test measures the dielectric absorption of heteroge neous or multilayer insulation and disregards parallel surface leakage currents It involves applying a test voltage for long enough to electrically charge the insu lation to be measured typically a voltage of 500V is applied f
36. Null Modem Cable Cat 2119 45 9 pin female connector 9 pin female connector to Model 5070 to serial printer Megohmmeter Model 5070 89 APPENDIX C V DISTURBANCE V OUTPUT FEATURE The V Disturbance V Output feature referred to in the text of this manual as dlSt serves two purposes First it acts as a safety device inhibiting the instrument from generating a test voltage when live signals are present In the set up mode you can choose a percentage of the test voltage that constitutes a condition that requires the test to be inhibited before it starts If the Model 5070 mea sures a voltage higher than the V Disturbance V Output setting allows no test voltage will be generated when the Start button is pressed The calculation for this condition is simple In the set up mode you have a choice to set the ratio for V Disturbance V Output stated as dISt to 3 10 or 20 The inhibit voltage is then calculated by multiplying this percentage times the test voltage Example Test voltage is set to 500 and the V Disturbance V Output is set to 3 there fore the voltage present at the input terminals before the start of a test that will inhibit testing is 500 0 03 15 volts The table below shows the inhibit voltages for the 4 fixed test voltage positions and the three V Disturbance V Output percentages Before Testing Test V Disturbance V Output Voltage 395 10 20 500 15 50 100 1
37. RM Button 4 6 1 Primary Function A Increases the value of the flashing parameter displayed or selects the next param eter up 4 6 2 Secondary Function ALARM To activate the ALARM function before or during a test press the 2nd ALARM buttons The word ALARM will appear at the top right side of the display If the insulation resistance drops below the set value at any time during the test the ALARM symbol will flash and the buzzer will be continuously on if activated as long as the alarm condition exists The alarm function can be deactivated at any time by pressing the 2nd ALARM buttons again 4 7 Button 4 7 4 Primary Function gt Selects the parameter to be modified to the right or moves the cursor on a graph screen to the right NM 4 7 2 Secondary Function 7e To activate the display backlight press the 2nd e buttons To deactivate press the 2nd le buttons again Megohmmeter Model 5070 49 4 8 MEM MR Button 4 8 1 Primary Function MEM Save Results can be recorded in the instruments memory and stored along with addresses defined by an object number OBJ and a test number TEST An object represents a box into which you can place 99 tests An object can rep resent a single piece of equipment such as a motor or pump or a department or an entire facility Tests will represent the insulation resistance measurements carried out on the object Example If the object is a 3 phas
38. State e Phone Fax Default Delete Figure 7 11 The Session Properties dialog box allows you to specify Operator Site and Custom parameters that are to be saved with recorded data These parameters are used when generating reports The Operator and Site tabs allow you to maintain lists of operators and sites saving you time when specifying parameters for reports On the left of the Operator and Site tabs is the list of previously defined Operators and Sites On the right of the Operator and Site tabs is the individual parameters that will be saved in an associated database Only a single set of operator and site fields are saved in the recording database The Custom tab allows you to set up user defined parameters Click the New button to create a user defined parameter Double click on the name or click the Edit button to assign a name and value A checkbox appears to the left of the value Items that are checked will be added to an associated database and used in displaying a report Only a single set of Custom parameters can be maintained unlike the Operator and Site lists 5 Once you have completed the Session Properties click OK Type a descrip tive name for the object and test just performed 6 Press the OK button The information is now saved to memory 7 You will then be asked if you want to open the database just downloaded Click the Yes button which will open the Create Report from Template dialog
39. adings Surface leakage is fundamentally a low resistance path in parallel with the most important resistance path through the body of the insulation The guard termi nal can be used to stop the effects of the surface leakage from influencing the measurement of resistance through the insulation In the example of a cable shown in the diagram below connecting the guard terminal of the Model 5070 to the surface of the insulation will redirect surface leakage currents away from the true measured value of leakage current from the conductor through the insula tion The use of the guard terminal is most important when there is a significant exposed surface in a cable When testing the insulation at the end of a cable it is necessary to eliminate the error from surface leakage which occurs par ticularly at high resistance values The guard terminal provides a third terminal within the path of the surface leakage Connect the instrument as shown in Figure A 1 Outer jacket The guard terminal Exposed Surface Conductor Insulation Shield to to to Line Guard Earth Terminal Terminal Terminal is useful when measuring very high resistance values Figure A 1 Shielded Cable Connect the black test lead to the conductor at A Connect the red test lead to the shield layer of the cable at C Connect the blue Guard G lead to the insulation layer of the cable at B If there is no shield use a copper wire wound
40. adually fall off they affect the measurement at the start for a length of time that depends on whether the insulation is in good condition or degraded These ratios complete the absolute insulation value and reliably reflect whether the insulation layers are in good or poor condition Megohmmeter Model 5070 31 In addition changes in these ratios over time can be observed and used for pre ventative maintenance e g to monitor the aging of the insulation of a population of rotating machines The DAR and PI ratios are calculated as follows PIl R 10 min R 1 min 2 values to be noted during a 10 minute measurement are at the end of 1 minute and 10 minutes DAR R 1 min R 30 sec 2 values to be noted during a 1 minute measurement are at the end of 30 seconds and at the end of 1 minute Note that the times of 1 and 10 minutes for the calculation of PI and 30 amp 60 sec onds for the calculation of DAR are those currently considered standards and pro grammed as defaults in the instrument They can be modified in SET UP mode to adapt to a possible change in a standard or to the needs of a specific application Interpretation of the results DAR PI Condition of Insulation 1 125 lt Poor or lt 2 even Dangerous 1 6 4 Good 1 6 24 Excellent 4 2 2 Secondary Function PRINT The secondary function of this button gives access to the print menu below There are two or three choices in
41. age AC DC input voltage Frequency Leakage current Date time i 234 5MO 507V 24 6pA Elapsed time 01 22 43 DAR 30 60 2 24 PI 1 10 1 55 Capacitance 320nF DD current 24 6pA DD 2 55 After 1 Minute Measured resistance DC test voltage Leakage current Total test time DAR PI Capacitance DD test current DD 37 TIMED RUN mode BEFORE the measurement E ADJUSTABLE VOLTAGE 2 2300 V Test Run Time 00 10 00 0 1V AC 0 0 Hz 20 6pA Time 09 39 Input voltage Frequency Input current Date 10 18 2003 Information Displayed Test type DC test voltage Programmed duration of the test AC DC input voltage Frequency Residual current Date time DURING the measurement i A 234 5MO 2307V 24 6pA Remaining time 00 09 43 kQ MQ GQ TQ 10 100 1 10 100 i 10 100 i 10 Information Displayed Measured resistance DC test voltage Leakage current Remaining measurement time Insulation resistance bargraph 38 Information after pressing DISPLAY AC DC input voltage Frequency Residual current Voltage bargraph i A 234 5MO 2307V 24 6pA Remaining time 00 09 43 DAR 30 60 PI 1 10 Capacitance Information after pressing DISPLAY Measured resistance DC test voltage Leakage current Remaining measurement time DAR PI Capacitance Megohmmeter Model 5070 AFTER the measurement i 234 5MO 2307V 20 6pA Elapsed time 00 10 00 DAR 30 60 2 64 PI 1 10 1 05
42. al functions can be used during the measurement see 4 5 3 Capacitance Measurement The capacitance measurement is performed automatically during the insulation measurement and is displayed after the measurement stops and the circuit has been discharged 5 4 Residual Current Measurement The residual current circulating in the installation is measured and displayed auto matically upon connection to the installation it is all measured during and after the insulation measurement 58 Megohmmeter Model 5070 CHAPTER 6 OPERATION NOTE Charge the instrument fully before use 6 1 Performing Measurements To begin an insulation resistance test first insert the cables as necessary then turn the rotary switch to the appropriate test voltage position The instrument can measure insulation values from 10kO to 10TO depending on the test voltage selected from 40V to 5100Vpc The screen displays i The battery symbol and battery charge FIXED VOLTAGE condition 2500 V The test voltage selected The voltage frequency and residual current at the input terminals Jie A pus The date and time Input current 24 6nA Date 09 28 2003 Time 22 39 Connect the Red and the Black cables to the test object Connect the Blue Guard jumper and cable if required for the test see Appendix A When to use the Guard Terminal Next unless the step function mode is selected select the measurement mode to be used Ma
43. asurement in voltage step mode the applied voltage increases in steps Automatic calculation of DAR PI and DD dielectric discharge index quality ratios Automatic calculation of measurement result referred to a reference temperature Automatic capacitance measurement Automatic measurement of residual current This megohmmeter helps ensure the safety of electrical installations and equip ment The operation is controlled by a microprocessor for data acquisition pro cessing and display of measurements and the storage and printing of results Advantages Digital filtering of insulation measurements Automatic voltage measurement Automatic detection of external AC or DC voltages on the terminals before or during insulation measurements disabling or stopping the mea surements when measurement accuracy is no longer guaranteed Threshold programming to trigger audible alarms Timer for measurement time checks Fuse protection of the instrument with detection and display of defective fuses Operator safety by automatic discharge of residual high voltage on the equipment tested Megohmmeter Model 5070 7 Automatic shutdown of the instrument to save battery power Battery charge indication Large backlit graphic display that is very easy to read Memory 128KB real time clock and serial interface Control and programming of the instrument from a PC using DataView software Direct printing with a serial pri
44. box Megohmmeter Model 5070 81 The Megohmmeter Instrument Group and Megohmmeter 5070 Summary Report template will be selected automatically If you wish to open the data base using another template select it Create Report from Template Step 1 Specify Template Select Instrument Select Template Continuity Test Report Template Megohmmeter 1060 Summary Report Templates Groups list Megohmmeter 5060 Summary Report list box P Megohmmeter 5070 Summary Report box Megohmmeter Summary Report Resistance Test Report v or Template cs E Megohmmeter 5070 Summary Report dvt Template Name field Browse Step 2 Specify Databases button 1 C Megohmmeter Object dvb 4 Database Eee list box button Cancel Help Figure 7 12 9 Inthe Step 2 Specify Database s window the file you just saved and named 10 11 7 6 82 should be visible If this is the file you wish to generate a report from click the OK button if it is not click on the file you wish to work with and then click OK The report will now be available on the screen including graph individual data points and all statistical data To print this report either click on the Print icon or choose File gt Print To save or export to a spreadsheet or PDF file keep the report open and read the next section 7 7 Exporting the Report to a PDF File With a report open go to File gt Print In the Print dialog
45. box that appears make sure that PDF XChange 3 0 is selected from the drop down menu then click OK When the PDF is complete the Save As dialog box will appear Choose a location to save it to and click Save Megohmmeter Model 5070 7 7 DataView Megohmmeter Templates 7 1 1 Megohmmeter Summary Report Template This template generates a view with two worksheets Each worksheet is selected via the Worksheet tabs located just above the status bar Trend Graph Bm Trend Graph nc Channel Summary a Channel List in 6 3 8 2005 08 42 000 PM 5 0 000 M S 1 Min Div 3 8 2005 6 13 42 000 PM 1000846 800 0E 6 R ohms 600 0E 6 Rc 0hns 400 OE 6 200 0E 6 0 000 3 8 2005 6 08 42 000 PI 5 0 000 M S 1 Min Div 3 8 2005 6 13 42 000 PM Test name Function ate Time Test run time Voltage Resistance R t ref Object 1 Test 1 R Ohms Megohm Variable Voltage test 3 8 2005 6 08 42 000 PM 5 0 000 M S 997 6E 6 Ohm 307 0E amp 6 Ohm D R 30s R value 60s R value PI First R value Second R value DD Capacitance Residual Current 1 013 987 4E 6 1 000E49 Ohm Ohm Undefined 996 6E 6 Ohm Ohm Undefined Undefined 104 9E 9F A Undefined Name Object i Test i R Ohms Object 1 Test 1 Rc Ohms ndex Date Time Val Units Samples El I 8 2005 s 08 a2 000 PN 9
46. communication port matches the port that the serial cable is plugged into Connection Communications Port COM1 X Communication Rate 9600 Cancel Help Figure 7 4 The communication rate baud rate will default to 9600 and should be acceptable to any computer This is also the default rate for the Model 5070 You may select a different baud rate by clicking on the Communication Rate drop down menu and selecting from the available rates see page 26 for selecting the baud rate DataView will automatically try each rate should the selected baud rate that was initially selected not match that which is programmed into the Model 5070 Once the proper communication parameters have been specified click OK If the instrument is turned ON at the time the Control Panel is opened a commu nication link will happen automatically If the instrument is turned OFF at the time the Control Panel is opened establish a communication link by going to Instrument Connect in the main menu Megohmmeter Model 5070 75 7 4 Configuring the Instrument Click the Configure button in the Control Panel The following settings appear General Settings for programming the 5070 s main functions Megohmmeter 5070 Setup General Settings Variable Voltages Ramp Voltages Temperature Test Run Settings Date Format Test run time o 10 hh mm 00 01 49 59 USA Sample interval R t 0 10 mm ss 00 10 10 00 C E uropean
47. conds to indicate that a test is in process A number of special functions can be performed during the test Chapter 4 describes these functions i WARNING Insulation measurements cannot be started if there is an excessively high external voltage on the and terminals The Model 5070 will automatically inhibit testing 56 Megohmmeter Model 5070 When the START button is pressed if the external voltage of the terminals of the Model 5070 is greater that the value V peak defined in the set up mode and described next the insulation test is not started and an audible alarm is emitted the instrument then returns to automatic voltage measurement mode V peak gt dlSt x VN where Vpeak is the external voltage peak or DC at the and terminals of the Model 5070 diSt is the coefficient that can be selected in SET UP at 3 default value 10 or 20 VNis the test voltage selected for the insulation measurement Example A The test is to be conducted at 2500V dlSt is set to 3 Therefore before the test starts a voltage present at the and terminals of 75V will cause the Model 5070 to inhibit testing Vpeak 0 03 x 2500V 75V Similarly if during the insulation test an external voltage greater than the V peak as defined below is detected the measurement is stopped and the symbol A appears next to the value of the external voltage measured V peak gt dISt 1 05 x VN
48. d calibration data Ship To Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 800 945 2362 Ext 360 603 749 6434 Ext 360 Fax 603 742 2346 or 603 749 6309 E mail repair aemc com Or contact your authorized distributor Costs for repair standard calibration and calibration traceable to N I S T are available NOTE You must obtain a CSA before returning any instrument Technical and Sales Assistance If you are experiencing any technical problems or require any assistance with the proper operation or application of your instrument please call mail fax or e mail our technical support team Chauvin Arnoux Inc d b a AEMC Instruments 200 Foxborough Boulevard Foxborough MA 02035 USA Phone 800 343 1391 Ext 351 508 698 2115 Ext 351 Fax 508 698 2118 E mail techsupport aemc com www aemc com NOTE Do not ship Instruments to our Foxborough MA address Megohmmeter Model 5070 91 Limited Warranty The Megohmmeter Model 5070 is warranted to the owner for a period of one year from the date of original purchase against defects in manufacture This limited warranty is given by AEMC Instruments not by the distributor from whom it was purchased This warranty is void if the unit has been tampered with abused or if the defect is related to service not performed by AEMC Instruments Full warranty coverage and product registration is available o
49. e Screen and the graph on the right side of the screen e Save to Disk Saves the current data to a file in the computer Configure Returns to the megohmmeter Setup dialog box for configur ing the instrument Download Retrieves data from the memory of the megohmmeter for storage in the computer Megohmmeter Control Panel File View Instrument Help Test Operation 500v C 1000V Clear Display C 2500V Save to Disk C 5000V Configure C Vay Download C Var Mur Calculate DD Calculate R t refl 10 00 m s D s 5075 V 5084n 996 4 M Ohm 1 000 362 4 M Ohm 962 4 M Ohm 1 027 d 3 970 5 M Ohm 10 00 R value 996 7 M Ohm 9 50 m s l Min Div Figure 7 9 To run a timed test perform the following steps 1 Select the appropriate voltage for the test by clicking on one of the six radio buttons next to the desired selection If Var V or Var Ramp is selected you will also need to select profile 1 2 or 3 using the drop down arrow next to that selection 2 Press the Start button This will run the test at the selected voltage As the test is running all other buttons are grayed out and the Start button becomes a Stop button Once the test has started a graph will begin to appear and build on the right side of the Control Panel representing a standard test and a step voltage test respec tively Megohmmeter Model 5070 79 Megohmmeter Control Panel File View Instrument H
50. e first reading Range 0 5 to 30 1 to 90 22 Megohmmeter Model 5070 SET STEP FUNCTION 1 2 3 Step voltage testing can be a valuable tool in determining defective insulation or used to dry out cables that have been penetrated by moisture With this function you can program three different profiles each containing up to 5 voltage steps and dwell times from 1 minute to 9 hours and 59 minutes per step Step Default Value Range Function Voltage Duration h m Voltage Duration h m Step Function 1 h m h m Step 1 50V 00 01 00 to 09 01 to 59 Step 2 100V 00 01 40 to 5100V 00 to 09 01 to 59 Step 3 150V 00 01 in 10V then 00 to 09 01 to 59 Step 4 200V 00 01 100V steps 00 to 09 01 to 59 Step 5 250V 00 01 00 to 09 01 to 59 sample time 00 01 m s 00 to 59 00 to 59 Step Function 2 h m h m Step 1 100V 00 01 00 to 09 01 to 59 Step 2 300V 00 01 40 to 5100V 00 to 09 01 to 59 Step 3 500V 00 01 in 10V then 00 to 09 01 to 59 Step 4 700V 00 01 100V steps 00 to 09 01 to 59 Step 5 900V 00 01 00 to 09 01 to 59 sample time 00 01 m s 00 to 59 00 to 59 Step Function 3 h m h m Step 1 1000V 00 01 00 to 09 01 to 59 Step 2 2000V 00 01 40 to 5100V 00 to 09 01 to 59 Step 3 3000V 00 01 in 10V then 00 to 09 01 to 59 Step 4 4000V 00 01 100V steps 00 to 09 01 to 59 Step 5 5000V 00 01 00 to 09 01 to 59 sample time 00 01 m s 00 to 59 00 to 59 The minimum sample
51. e motor then each test number will represent a specific insulation resistance measurement e g Test 1 Phase A to housing Test 2 Phase B to housing Test 3 Phase C to housing and so on When the MEM button is pressed the following screen is displayed The flashing cursor identifies the first free Obj Test location Here 02 59 the Obj number is that Store MEMORY of the last measurement stored but mm 3 Obj Test Date Time Fet 302 59 12 07 2003 22 39 3800V 02 58 11 29 2003 1547 50VO the Test number is incremented by 1 Itis always possible to modify Obj Test using the A V 4 buttons 02 03 11 24 2003 15 04 2150V If a new Obj is selected the test 02 02 10 29 2003 21 45 975V number is set to 01 automatically 02 01 09 30 2003 02 43 5000V If you select a memory address that 01 02 09 02 2003 1507 L O is already occupied the screen to the right is displayed and prompts you to confirm or cancel deletion of the con tent of that address Store MEMORY e To confirm overwriting press the gt WARNING button Old data set will To cancel press the WV button to be overwritten highlight CANCEL then press the gD O K button to confirm r n CANCEL When the MEM key is pressed again the measurement results are recorded at the selected memory address whether occupied or not All information from the measurement will be stored at a single locatio
52. ed Measured resistance DC test voltage Residual current Measurement duration Insulation resistance bargraph 34 0 400 250 500 750 Information after pressing DISPLAY AC DC input voltage Frequency Residual input current Voltage bargraph N 234 5MO 507V Elapsed time 000043 DAR 30 60 PI 1 10 Capacitance Information after pressing DISPLAY Measured resistance DC test voltage Residual current Measurement duration DAR PI Capacitance Megohmmeter Model 5070 AFTER the measurement rj 234 5MO 507V 24 6pA Elapsed time 01 02 43 DAR 30 60 2 64 PI 1 10 1 05 Capacitance 320nF m FIXED VOLTAGE 500 V Input voltage 0 1V AC Frequency 0 0 Hz Input current 24 6pA Date 09 28 2003 Time 22 49 0 0Hz 24 6pA AC 0 199 250 500 750 1000 L L Megohmmeter Model 5070 Information Displayed Measured resistance DC test voltage Leakage current Measurement duration DAR PI Capacitance values After 1st Press on DISPLAY Test type and test voltage AC DC input voltage Frequency Leakage current Date time After 2nd Press on Display AC DC input voltage Frequency Leakage current Voltage bargraph 35 MANUAL STOP DD mode BEFORE the measurement i FIXED VOLTAGE 500 V 0 1V AC 0 0 Hz 24 6pA Time 09 39 Input voltage Frequency Input current Date 10 08 2003 Information Displayed Test Type DC test voltage AC DC in
53. elp Test peration C 500V C 1000V Clear Display C 2500V Save to Disk C 5000V Configure C Var Download Var Ramp fr Help I Calculate Rit refl 5 09 m s 249 1 V 248 3 n 1 003 G Ohm 308 7 M Ohm 4 50 m s 30 Sec Div Figure 7 10 The statistical data for the time of the test DAR PI resistance value and other information will appear as they are completed on the lower left side of the Control Panel The test will stop automatically at the end of the programmed run time The test can be stopped at any time by pressing the Stop button In either case when the test is completed the Model 5070 will automatically calculate dielectric discharge ratio if the calculate DD box is checked This test takes 60 seconds i NOTE On small motors and machinery it is possible that the capaci tance values are too low for this test to be completed accurately In these cases simply press the Stop button to stop the dielectric discharge test or uncheck this box before starting the test 3 Once the test has been completed press the Save to Disk button to save the results of the test just completed to a disk drive in your computer 4 Name the file in the dialog box that opens then click Save 80 Megohmmeter Model 5070 This will open the Session Properties dialog box Session Properties Operator site Custom Operators list Operator Company Address City
54. g of memory in SET UP see 84 8 so the MEM MR functions can be used again Megohmmeter Model 5070 85 8 2 Fuse Replacement If GUARD FUSE appears on the display the fuse must be replaced taking all the necessary precautions when opening up the instrument WARNING Make sure that none of the terminals are connected and that the selector switch is set to the OFF position before opening up the case Only use the type of fuse shown on label inside the unit s cover 0 1A Fast Acting 380V 5x20mm 10kA Cat 2119 84 8 3 Cleaning WARNING Disconnect the instrument from any source of electricity Use a soft cloth lightly dampened with soapy water Rinse with a damp cloth and then dry with a dry cloth Do not get water inside the case This may lead to electrical shock or damage to the instrument Donotuse alcohol solvents or hydrocarbons 8 4 Storage If the instrument is not used for an extended time period longer than two months it is recommended that a complete charge and discharge of the battery is per formed three times before re using the instrument Complete battery discharge can be performed By removing the battery from the instrument and applying a 3 Amp load to it e On the 5000V position with the backlight on consumes the most power 86 Megohmmeter Model 5070 APPENDIX A UTILIZING THE GUARD TERMINAL Guard terminals are useful when measuring high resistance values and for sta bilizing re
55. ge and the initial resistance with the lowest test voltage The difference AV between the final and initial test voltages The slope of the curve in ppm V The capacitance Pressing the GRAPH button displays the resistance versus applied test voltage curve Using the lt and gt buttons you can scroll through the various samples recorded and for each see The insulation resistance value The applied test voltage The time of measurement 6 3 Operation Examples m 234 5MO 5078V 24 6pA Elapsed time 00 05 00 AR 8 5MQ AV 4000V AR R AV ppm V 0 28 Capacitance 320nF Resistance 328 5MO Output Voltage 5078V DC Elapsed Time 00 0 1 00 2 00 3 00 400 5 00 Connection diagram for measurement of low resistance insulation e g motor BLACK RED 62 Pines Or MODEL 5070 Megohmmeter Model 5070 When measuring high levels of insulation gt 1G it is advisable to use the G guard terminal to eliminate the influence of surface leakage currents The guard is connected between the two measurement contact points and the surface suscep tible to surface currents e g dusty damp cable or transformer insulation In this case alligator clips are preferable to hand held test probes As soon as the insulation measurement is stopped the test circuit is automatically discharged using the instrument s internal discharge feature Connection diagram for meas
56. ge of the parameter ALARM Activates or deactivates the alarms programmed in SET UP V Selects a function or decrements the parameter on which the cursor is positioned Pressing and holding the key increases the rate of change of the parameter SMOOTH Stops starts smoothing of displayed values during testing lt q Selects a parameter to be modified to the left TEMP activates temperature correction of the measurement to the reference temperature programmed in SET UP P Selects a parameter to be modified to the right 3 Turns the display backlight ON or OFF MEM Saves measured values MR Recalls saved data MODE Before starting a test the mode function allows selection of run time sample rate DAR PI and ratio times PRINT direct printing of test result s on a serial or parallel printer DISPLAY Browses through the various screens accessible before during and after the measurement GRAPH After a timed run the Graph mode displays insulation resis tance versus time in a graph format Megohmmeter Model 5070 9 2 2 Display Features REMOTE COM 2nd DAR PI DD SMOOTH ALARM A A 10 Indicates that the instrument is controlled remotely via the RS 232 interface In this mode all of the keys and the rotary switch are inac tive except for the OFF position Flashes when data is transmitted to the serial interface On continu ously if there is a problem in transmissio
57. harged and are safe to touch Dielectric insulation samples should be short circuited for at least five times the amount of time they were energized Megohmmeters should never be used in an explosive environ ment Only use the leads that are supplied with the megohmmeter If they are defective or worn replace before testing This instrument can be used on installations rated for 1000V CAT IIl or 2500V CAT I 4 Megohmmeter Model 5070 1 1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced insulation This symbol on the instrument indicates a WARNING and that the operator must refer to the user manual for instructions before operating the instrument In this manual the symbol preceding instructions indicates that if the instructions are not followed bodily injury installation sample and product damage may result Risk of electric shock The voltage at the parts marked with this symbol may be dangerous Ex B BI In conformity with WEEE 2002 96 EC 1 2 Definition of Measurement Categories CAT II For measurements performed on circuits directly connected to the electrical distribution system Examples are measurements on household appliances or portable tools CAT Ili Formeasurements performed in the building installation at the distribution level such as on hardwired equipment in fixed installation and circuit breakers CAT IV For measurements perfo
58. he down arrow VW button until the blinking cursor is to the left of Units To select the desired format press the right arrow 9 button to highlight the current setting Next press the up A or down V arrow buttons to select either USA or Europe Press the right arrow button after you have selected the desired units to complete the process Range Europe or USA DATE The day month and year value can be set with this function Press the down arrow V button until the blinking cursor is to the left of Date To adjust the Date press the right arrow button to highlight the current setting The cursor will move to the first variable day for European format or month for US format Next press the up A or down V arrow buttons to select the desired value Press the right arrow gt button to highlight the next date field month for European format or day for US format Next press the up A or down V arrow buttons to select the desired value Press the right arrow button once more to highlight the last date field which is year Press the up A or down V arrow buttons to select the desired year Finally press the right arrow gt button after you have selected the date set up to complete the process Europe USA dd mm yyyy mm dd yyyy 26 Megohmmeter Model 5070 TIME h m Time can be set here A 24 hour clock is used in the Model 5070 Therefore 3 30pm would be programmed in as 15 30 Press the down ar
59. hmmeter Model 5070 1 4 3 DISPLAY GRAPH Button sesenta 33 4 3 1 Primary Function DISPLAY 33 4 3 2 Secondary Function GRAPH esee 46 4 4 Left TEMP BUOD airiai aaa ae d in 46 4 4 1 Primary F nction Left iie eee 46 4 4 2 Secondary Function TEMP 46 4 5 Down SMOOTH Button iae see cepe erbe rg DER dos 48 4 5 1 Primary Function Down eseeeen 48 4 5 2 Secondary Function SMOOTH see 48 4 6 Up ALARM Buttons sas eere ooi cede tete oe prede Red 49 4 6 1 Primary Function Up eene nenne 49 4 6 2 Secondary Function ALARM esse 49 4 7 Right Backlight BUON cosirer sirina 49 4 7 1 Primary Function Right e 49 4 7 2 Secondary Function Backlight un 49 4 8 MEM 7 MR BULOFR tero t d ee tate e base bete he 50 4 8 1 Primary Function MEM Save 50 4 8 2 Second Function MR Recall esee 51 4 9 Clearing the Memory eene eren kunt eher Led hne ee 52 4 10 Calculation of AT from Stored Data 53 4 11 Maximum Output Voltage sssssssessesee enne 54 4 12 List Of Emor CodeS rniii ene tere eR E Ren 54 5 MEASUREMENT FUNCTIONS eee 55 5 4 AC DC Voltage en tte aee ei reed ares eld 55 5 2 Insulation Measurement seeeeee eene an 55 5 3 Capacitance Measurement ssseeeeen 5
60. j and Test number Here 47 99 Use the A V 4 gt buttons to select the desired Obj and Test number Recall MEMORY Obj Test Date Time After selecting the Obj and Test press the gt button to access the first item of 347 99 12 15 2003 07 04 information relative to this measure 13 59 12 07 2003 22 39 ment 13 58 11 29 2003 1547 02 03 11 24 2003 15 04 To access more data press DISPLAY 02 02 10292003 2145 repeatedly or GRAPH if the mode selected before the start of the measurement allows 02 01 09 30 2003 02 43 01 02 09 02 2003 15 07 To exit the MR function press the MR button once again or turn the selector switch to another position Megohmmeter Model 5070 51 4 9 Clearing the Memory In SET UP select Clear memory by using the Y button to highlight this function Press the gt button to enter this mode To erase the content of one or more specific OBJ TEST numbers Select Select Data Sets to Clear by pressing the button Then select each memory to be erased using the A V buttons to choose it and the gt button to select it The lt button will deselect it Validate by pressing DISPLAY The operation is confirmed by pressing the gt button with the OK option highlighted or cancelled by pressing the gt button with the CANCEL option highlighted SET UP Instr Nr 9600004 SW Version 1 2 Rc reference temperature 40 C AT for R 2 10 C Calculate AT from Memory Maxu
61. low as once every 10 minutes Range 00 to 59 05 to 59 DAR s s The Dielectric Absorption Ratio is a ratio of insulation resistance measured at two predetermined times The reading at the first time mark is then divided into the reading at the second time mark to calculate the ratio The typical times used for this test are 30 seconds and 60 seconds These are the factory default settings You can adjust the times here in the set up mode from 10 to 90 seconds for the first reading and from 15 to 180 seconds for the second reading both in 5 second increments NOTE The Model 5070 will not allow the time for the second reading to be set lower that the set time for the first reading Range 10 to 90 15 to 180 PI m m The Polarization Index is a ratio of insulation resistance measured at two rede termined times The reading at the first time mark is then divided into the reading at the second time mark to calculate the ratio The typical times used for this test are 1 minute and 10 minutes These are the factory default settings The insula tion materials used today require less time to determine this ratio Therefore you can adjust the times here in the set up mode to suit your needs from 0 5 minutes to 30 minutes for the first reading and from 1 minute to 90 minutes for the second reading NOTE The Model 5070 will not allow the time for the second reading to be set lower that the set time for th
62. m
63. measurement sam R t sample m s 00 20 ples R t sample by scrolling to the desired Step using the A or V buttons followed by the button to access the selected steps voltage and time variables Adjust the voltage for the Step by pressing the A or V buttons Then press the button to move the selection cursor to the duration setting Adjust the hours and minutes for the step in the same fashion used to select the voltage When finished press the button until the cursor is again in the Step Column Repeat this process for each of the five steps i If a Step is not used select as the duration Next adjust the sample rate by the same procedure using the A V and lt P gt buttons to select and adjust the sample rate Once the Step Function is defined turn the selector switch to Step Voltage the desired profile 1 2 or 3 using the A V buttons Then start the measurement by pressing the START STOP button During the measurement the following screens can be accessed by pressing the DISPLAY button Q9 HW oi A 234 5MQ A 234 5MO 4078V 24 6pA 4078V 24 6pA Remaining time 00 01 43 Remaining time 00 01 43 AR MQ AV 3000V kQ MO GO TO 10100 1 10100 1 10100 1 10 AR R AV ppm V Doro gr dod Capacitance Megohmmeter Model 5070 61 At the end of the measurement the following are indicated The difference AR in insulation resis tance between the final resistance with the highest test volta
64. mum Output Voltage 5100V Set Default Parameter 3 Clear Memory V Disturbance V Output 3096 Buzzer on SET UP Clear memory Obj Test Date Time Fct 47 99 15 12 2003 07 04 625V IKE 07 12 2003 22 39 3800V OEE 29 112003 1547 50VO 02 03 24 11 2003 15 04 2150V To erase the entire memory SET UP Clear memory O Select Data Sets to Clear Clear All SET UP WARNING All selected data sets will be cleared Do K CANCEL e Select Clear All by pressing the Y button to highlight it and the gt button to choose it e The operation is confirmed or cancelled by pressing the button when the appropriate choice is highlighted as described above SET UP Clear memory Select Data Sets to Clear O Clear All 52 SET UP WARNING All data sets will be cleared BDo K CANCEL Megohmmeter Model 5070 4 10 Calculation of AT from Stored Data The coefficient AT is used to calculate the insulation resistance at a temperature other than the measurement temperature It is the temperature difference at which the insulation resistance concerned is divided by 2 This coefficient is variable and depends on the type of insulation When it is not known the instrument can calculate it from three or more stored measurements NOTE These 3 measurements must have been made on identical insu lation but at 3 different temperatures and the temperatures must have been recorded using
65. n Indicates that the secondary function of a key will be used Indicates that the programmed time test mode was selected before the measurement was started Indicates that the automatic calculation of Dielectric Absorption Ratio mode was selected before the measurement was started Indicates that the automatic calculation of Polarization Index mode was selected before the measurement was started Indicates that the automatic calculation of Dielectric Discharge Index mode was selected before the measurement was started Smooths the insulation measurement readings on the display for better visibility Indicates that the alarm is activated An audible alarm will sound if the value measured is above the limit defined in the SET UP mode Indicates the battery charge condition Voltage generated dangerous V gt 120Vpc External voltage present this symbol is activated when the Start button is pressed if V 25VRMs Megohmmeter Model 5070 CHAPTER 3 SPECIFICATIONS 3 1 Reference Conditions Influence Quantity Reference Values Temperature 23 C 3 K Relative Humidity 45 to 55 Supply Voltage 9 to 12V Frequency Range DC and 15 3 to 65Hz Capacitance in parallel with the input OuF resistance Electric Field nil Magnetic Field lt 40A m 3 2 Voltage Measurement Range 1 0 to 99 9V 100 to 999V 1000 to 2500V 2502 to 4000V Freque
66. n in memory date time test mode and voltage insulation resistance capacitance residual cur 50 Megohmmeter Model 5070 rent and possibly DAR PI DD if available temperature corrected resistance and the R t graph i WARNING If a button other than MEM is pressed or the selector switch is moved before pressing MEM a second time the instrument exits from MEM mode without saving the results Memory Capacity Total memory space 128kb Data management 8kb Free memory space 120kb An insulation measurement requires 80 bytes Therefore approximately 1500 insulation measurements can be stored Free Memory Space This function is automatically activated when a result is saved Press the MEM button once to bring up the first available Object and Test storage location A bargraph is displayed corresponding to the available memory location If the entire memory is free all of the bargraph segments appear e If the entire memory is full the arrow to the left of the bargraph flashes As soon as the storage is complete the bargraph disappears Each segment of the bargraph equals approximately 50 recordings 4 8 2 Second Function MR Recall The MR function allows you to recall any of the stored data from memory regard less of the position of the rotary selector switch except OFF or SETUP When the MR button is activated the following screen is displayed The flashing cursor identifies the last occupied Ob
67. n our website at www aemc com warranty html Please print the online Warranty Coverage Information for your records What AEMC Instruments will do If a malfunction occurs within the one year period you may return the instrument to us for repair provided we have your warranty registration information on file or a proof of purchase AEMC Instruments will at its option repair or replace the faulty material REGISTER ONLINE AT wwW aemc com Warranty Repairs What you must do to return an Instrument for Warranty Repair First request a Customer Service Authorization Number CSA by phone or by fax from our Service Department see address below then return the instrument along with the signed CSA Form Please write the CSA on the outside of the shipping container Return the instrument postage or shipment pre paid to Ship To Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 800 945 2362 Ext 360 603 749 6434 Ext 360 Fax 603 742 2346 or 603 749 6309 E mail repair aemc com Caution To protect yourself against in transit loss we recommend you insure your returned material NOTE You must obtain a CSA before returning any instrument 92 Megohmmeter Model 5070 EJ AEMC INSTRUMENTS 10 15 99 MAN 100283 v23 Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 603 749 6434 Fax 603 742 2346 www aemc co
68. ncy Range DC and 15Hz to 500Hz DC Only Resolution 0 1V 1V 2V 2V Accuracy 1 of Reading 5cts 1 of Reading 3ct Input Impedance 750kQ at 3MQ depending on measure voltage Over 500Hz the small display indicates and the main display gives only an assess ment of the peak value of the measured voltage Measurement Category 1000V CAT III or 2500V CAT I transients lt 2 5kV 3 3 Insulation Resistance Method Voltage current method according to EN 61557 2 ed 02 97 Nominal Output Voltage 500 1000 2500 5000 VDC or selectable from 40V to 5100V Adjustments Available in Variable Mode 10V from 40V to 1000V 100V from 1000V to 5100V Open circuit Voltage x 1 1 x Vn x2V Vn 2 in variable mode Max Overload of Voltage Vn 1 05 dISt Vn 50V with dlSt 3 10 or 20 NOTE Vn Test Voltage Megohmmeter Model 5070 11 Nominal Current gt 1mApc Short circuit Current lt 1 6mA 5 Load Current 3mApc approx when starting measurement Maximum Acceptable AC Voltage Vpeak 1 05 dISt Vn with dISt 3 10 or 20 NOTE dlSt is the ratio of V Disturbance V Output and is selectable in SET UP mode Measurement Ranges 500V 30kQ to 1 999TQ 2500V 100kO to 9 99TQ 1000V 100kQ to 3 999TO 5000V 300kQ to 9 99TQ Variable 40V to 5100V see graph below Resistance Range in Voltage Mode Rmax TQ Rmin kQ 12 300 10 250 8 200 6 150 4 100 2 50 0 0 0 1000 2000 30
69. nter optional accessory Cat 2140 21 2 1 Control Features a PF o NM SET UP mo 50 5000 Mo 5000V FF OA 880V 10kA 5x20mm ALARM SMOOTH START STOP MEGOHMMETER MODEL 5070 Safety terminals G and Protective fuse access AC power plug direct operation on AC and battery recharge Back lit liquid crystal display RS 232 serial interface male plug 9 pin For connection to a PC or printer Rotary selector switch with 8 positions OFF Instrument is off Megohmmeter Model 5070 MQ 500V Insulation measurement to 2TQ MQ 1000V Insulation measurement to 4TQ MQ 2500V Insulation measurement to 10TQ MQ 5000V Insulation measurement to 10TQ MQ 50V 5000V Insulation measurement with selectable test voltage from 40V to 5100V 10V steps from 40V to 1000V and 100V steps 1000V to 5000V Step Voltage Insulation measurement with voltage step function up to 5 steps can be configured SET UP User programming of the megohmmeter configuration 7 START STOP button starts and stops measurements 8 Eight buttons each with a primary and secondary function The secondary functions are highlighted in yellow italics below each button 2 yellow button Selects the second function on each button A Selects a function or increments the parameter on which the cursor is positioned Pressing and holding the key increases the rate of chan
70. nual Stop Manual Stop DD Timed Run Timed Run DD DAR or Pl by pressing the MODE button see 4 2 1 Press the START STOP button to begin the measurement test i If the voltage present is greater than the maximum allowed value the measurement will be prohibited The DISPLAY button can be used to scroll through the information available during the test This information depends on the measurement mode selected see 4 2 1 If the insulation values displayed are very unstable a digital filter can be activated by pressing SMOOTH to smooth them see 4 5 2 Megohmmeter Model 5070 59 The alarm mode can be activated by pressing the ALARM button An audible beep will continuously sound if the measurement result is below the value defined in SET UP see page 20 Press the START STOP button again to stop the test The last result remains displayed until the next measurement is made or the switch is turned When the insulation measurement has stopped the device under test will be auto matically discharged by the Model 5070 The DISPLAY button can be used to view the test results available after the test This information depends on the measurement mode selected see 4 2 1 If the measurement mode was either Timed Run or Timed Run DD pressing the GRAPH button displays the insulation measurement versus time curve see 4 3 2 Press the TEMP button to correct the measurement result to the reference
71. ohmmeter using the current settings Read from Inst Reads the current configuration of the megohmmeter attached via the serial cable e Save to Disk Saves the current configuration This file will reside on the computer s disk drive Saving different configuration setups can be useful for future functions and tests Load from Disk Retrieves a saved file from the computer s disk drive to be used in programming the megohmmeter Defaults Resets all fields in the Setup dialog box back to the original factory settings e Set Clock Programs the computer s time and date into the configuration of the megohmmeter Clear Memory Erases all contents in all memory locations in the megohmmeter Download Retrieves the stored data from all memory locations within the megohmmeter for naming and storing on the computer s disk drive memory Once all fields in each of the 4 tabs of the Setup dialog box have been configured it is recommended that you perform the following steps 1 Click on the Save to Disk button 2 Name and save the file 3 Click on the Write to Inst button to configure the megohmmeter with this new configuration 4 Click on the Close button to exit Setup and return to the Control Panel 78 Megohmmeter Model 5070 7 5 Running the Test There are five command buttons in the Control Panel Start This button begins a timed test Clear Display Clears the statistical data on the lower left side of th
72. on resistance falls below this value Voltage Range 500V 30kO to 2TQ 1000V 100kQ to 4TQ 2500V 300kQ to 10TQ 5000V Adj Voltage 1 Adj Voltage 2 10kQ to 10TQ Adj Voltage 3 To adjust the alarm settings press the W button until the blinking cursor is to the left of alarm settings To adjust the alarm value press the 9 button From the screen displayed you can adjust low limit resistance values for 500V 1000V 2500V 5000V or the three adjustable voltage positions To change the low limit for any of these voltage positions use the A and V but tons to select the voltage then press the gt button to highlight the resistance value for that voltage Next use the A and V buttons to increase or decrease the value at the blinking cursor position Once the desired resistance value has been programmed press the lt button to move the cursor again to the voltage selection part of the screen You may now select the different voltage and adjust its value as just described To exit the alarm setting function press the DISPLAY button This will bring you back to the top of the SET UP menu 20 Megohmmeter Model 5070 ADJUSTABLE VOLTAGE 1 2 3 This function defines a specific voltage that will be used for Insulation Resistance Measurement when the rotary switch is in the adjustable voltage position The values for this setting can be anywhere from 40 to 5100 Volts adjustable in 10 or 100 Volt increments
73. or the features you do not want to install Megohmmeter support files Ground Tester Power Quality Logger PowerPad PowerPad II Model 8335 PowerPad Jr Sentinel Data Logger Simple Logger II 407 and 607 Wattmeter 8220 I PDF XChange Figure 7 2 12 In the Ready to Install the Program window click on Install 13 If the instrument selected for installation requires the use of a USB port a warning box will appear similar to Figure 7 3 Click OK Megohmmeter Model 5070 DataView InstallShield Wizard Do not connect the AEMC USB Instrument or cable until after the LP installation of the drivers and the DataView software has finished If the AEMC USB Instrument or cable is connected to the computer now then disconnect it from the computer before proceeding Figure 7 3 NOTE The installation of the drivers may take a few moments Windows may even indicate that it is not responding however it is running Please wait for it to finish 14 When the drivers are finished installing the Installation Successful dialog box will appear Click on OK 15 Next the Installation Wizard Complete window will appear Click on Finish 16 A Question dialog box appears next Click Yes to read the procedure for con necting the instrument to the USB port on the computer NOTE The Set up window remains open You may
74. or 30 minutes At the end of the measurement the instrument causes a rapid discharge during which the capacitance of the insulation is measured 1 minute later the residual current circulating in the insulation is measured The DD Ratio is then calculated as follows DD current measured after 1 minute mA test voltage V x measured capacitance F The insulation quality rating as a function of the value found is as follows Value of DD Quality of Insulation DD 7 Very Poor DD between 7 and 4 Poor DD between 4 and 2 Doubtful DD 2 Good The dielectric discharge test is especially well suited to insulation measurements on rotating machines and in general to insulation measurements on heteroge neous or multilayer insulations containing organic materials 2 Whatare the DAR Dielectric Absorption Ratio and the PI Polarization Indexes It is beneficial to calculate insulation quality ratios in addition to the quantitative insulation resistance value because they can be used to eliminate the influence of certain parameters likely to invalidate the absolute insulation measurement The most important of these parameters are Temperature and relative humidity with which insulation resistance varies to a quasi exponential law e The leakage currents capacitive charging current dielectric absorption cur rent are created by the application of the test voltage Even though they gr
75. put voltage Frequency Residual current Date time DURING the measurement DD fal A 234 5MO 507V 24 6pA Elapsed time 00 00 43 kQ MQ GQ TQ 10 100 1 10 100 i 10 100 i 10 Information Displayed Measured resistance DC test voltage Leakage current Measurement duration Insulation resistance bargraph 36 0 400 250 500 750 Information after pressing DISPLAY AC DC input voltage Frequency Leakage current Voltage bargraph DD E A 234 5MO 507V 24 6pA Elapsed time 00 00 43 DAR 30 60 PI 1 10 Capacitance DD current DD Information after pressing DISPLAY Measured resistance DC test voltage Leakage current Measurement duration DAR PI Capacitance Residual current calculation of DD DD Megohmmeter Model 5070 AFTER the measurement DD il A 234 5MO 507V 24 6pA Elapsed time 00 22 43 DAR 30 60 2 24 PI 1 10 1 55 Capacitance 220nF DD current 11 55pA DD esc Information Displayed Measured resistance DC test voltage Leakage current Measurement duration DAR PI Capacitance values Residual current calculation of DD DD 0 1V AC V 0 499 250 500 750 1000 24 6pA After 2nd Press on Display AC DC input voltage Frequency Leakage current Voltage bargraph Megohmmeter Model 5070 i FIXED VOLTAGE 500 V 0 1V AC 0 0Hz 24 6pA Time 10 09 Input voltage Frequency Input current Date 10 08 2003 After 1st Press on DISPLAY Test Type DC test volt
76. rence Temperature RT Measured Insulation Resistance at Probe Temperature KT Coefficient T defined as follows KT 1 2 Rc Temperature Reference T AT T Probe Temperature AT Temperature change for which the Insulation Resistance is divided by 2 Rc Temperature Reference Temperature to which the Insulation Resistance measurement is adjusted Example A motor winding is tested at 1000V in a 68 F environment At the end of the timed test the Insulation Resistance Value is 991 megohms All the historical data is at 104 F The corrected Insulation Resistance value corrected to 104 F for this motor is calculated by the Model 5070 to be 284 4 megohms using the formula above when temperature correction is activated 4 5 V SMOOTH Button 4 5 1 Primary Function V Decreases the value of the flashing parameter displayed or selects the next para meter down 4 5 2 Secondary Function SMOOTH The SMOOTH function activated by pressing the 2nd SMOOTH buttons acti vates deactivates an insulation measurement digital filter Only the display is affected which is smoothed not the measurements This function is useful if the insulation values displayed are very unstable The filter is calculated as follows RSMOOTH RSMOOTH R RSMOOTH N Since N is set to 20 the time constant of this filter is approximately 20 seconds Therefore displayed results are presented more smoothly 48 Megohmmeter Model 5070 4 6 A ALA
77. rform the calculation the default value in TEMPERATURE SET UP is off Probe Temperature 23 7 C O Resistance Correction EJ RcReference Temperature 28 5 C AT for R 2 23 0 C Press the V button to select Resis tance Correction Press the gt button to highlight off then press the A or V buttons to select on R measured 273 7MQ Rc at 28 5 C 328 5MQ The calculation is performed imme diately and the result is displayed as Rc along with the Reference Temperature This indicates what the measurement result would have been at the refer ence temperature The Reference Temperature Rc Reference Temperature and the coef ficient AT indicated and used for the calculation are those defined in SET UP Note To store this calculation press the 2nd TEMP buttons again OK is then displayed before storing everything During the procedure pressing the DISPLAY button or turning the rotary switch cancels the calculation in progress If the coefficient AT used for the calculation is not known the instrument can calculate it in advance using at least 3 stored measurements made at different temperatures see 4 10 Megohmmeter Model 5070 47 The calculation is performed as follows The insulation resistance varies with the measurement temperature This depen dence can be approximated by the expression Rc KT RT where Rc Corrected Insulation Resistance to the reference temperature Rc Refe
78. rmed at the primary electrical supply 1000V such as on primary overcurrent protection devices ripple control units or meters 1 3 Receiving Your Shipment Match the contents with the ordering information Notify your distributor of any missing items If the equipment appears to be damaged file a claim immediately with the carrier and notify your distributor at once i NOTE Charge the instrument fully before use Megohmmeter Model 5070 5 1 4 Ordering Information Megohmmeter Model 5070 sssseeeeee Cat 2130 30 Includes extra large tool bag set of three 10 ft 5kV leads red black blue with integral hippo clips 5000V one guard terminal jumper lead RS 232 DB9 F F 6 ft null modem cable RS 232 to USB adapter US 115V power cord rechargeable battery pack and a USB stick with DataView software and user manual 1 4 1 Accessories and Replacement Parts Cable PC RS 232 DB9 F F 6 ft Null Modem Cable Cat 2119 45 Set of 3 color coded 10 ft safety leads 5000V 3 color coded alligator clips 600V CAT IV Cat 2119 76 Lead Replacement 1 ft jumper lead iriiri Cat 2119 78 Fuse set of 3 0 1A 380V 5 x 20 10kA iiris Cat 2119 84 Replacement Lead Set of 3 Color coded 10ft 5kV Safety with Integral Hippo Clips JUMPER LEAD INCLUDED Cat 2119 85 Lead Set of 3 Color coded 25 ft 5kV Safety with Integral Hippo Clips JUMPER LEAD NOT INCLU
79. rn the instrument Simply use the SET UP mode to restore the default parameters Set Default Parameter Data Storage Error When it is impossible to store data the entire memory must be erased using the Clear Memory function in the SET UP mode see 84 9 54 Megohmmeter Model 5070 CHAPTER 5 MEASUREMENT FUNCTIONS 5 1 AC DC Voltage Turn the rotary selector switch to any insulation position other than OFF or SET UP The instrument is now automatically in AC DC voltage measurement mode The voltage between the input terminals is measured at all times and indicated on the display next to the input voltage heading Also as soon as the switch is turned the frequency and the residual DC current at the terminals of the instrument are measured The residual current is mea sured in order to evaluate its effect on the insulation measurement about to be performed NOTE Measurement is prohibited if an external voltage above a preset value is present at the terminals before pressing START STOP Similarly if an interference voltage of a preset value is detected during measurement the measurement is stopped and that voltage is indicated see V Disturbance V Output on page 25 5 2 Insulation Measurement When the switch is turned to an insulation resistance measurement position one of following appears i FIXED VOLTAGE You select an insulation measurement with a fixed standard test voltage in 2500 V manual mode
80. row VW button until the blinking cursor is to the left of Time To adjust the Time press the right arrow gt button to highlight the current setting The cursor will move to the hour selection Next press the up A or down V arrow but tons to select the desired hour from 0 to 24 Press the right arrow gt button again to highlight minutes Next press the up A or down V arrow buttons to select the desired minutes from 0 to 59 Finally press the right arrow gt button after you have selected the time set up to complete the process 4 1 3 SET UP Menu Turning the rotary switch to SET UP gives you access to the menu of all program mable functions Select the function to be modified using the A and V buttons Move the blinking cursor to the value to be modified using the lt and gt buttons Adjust the value at the blinking cursor using the A and V buttons Move back to the select function position by pressing the lt and gt buttons until the blinking cursor is again to the left of the function Pressing the DISPLAY button while modifying a parameter will bring you back to the top of the SET UP menu When you first enter the SET UP mode a display similar to this will appear SETUP Instr Nr 9600004 SW Version 1 2 O Display contrast 80 Alarm Settings Adjustable Voltage 1 Adjustable Voltage 2 Adjustable Voltage 3 Timed Run h m Sample Time m s DAR s s The top line on this screen indicates that you are in SET UP mode
81. se of test Calculate AT from Memory voltages higher than the selected maxi I Maximum Output Voltage 5100V mum output voltage for the insulation Set Default Parameter measurement Clear Memory V Disturbance V Output 10 The instrument can then be used by Buzzer on less experienced personnel for specific Power Down on applications where it is important not to Baud Rate 9600 RS232 exceed a maximum test voltage Example if the maximum output voltage is set to 750V and the measurement is made in the 5000V switch position only 750V will be generated 4 12 List of Error Codes If an inconsistency is detected when the instrument is started up or in operation the display unit indicates an error code The format of this error code is a 1 or 2 digit number This number identifies the problem and the action to be taken Here are some examples of possible errors i Codes from 0 to 9 identify fatal errors in the hardware The instrument must be returned to AEMC Instruments for corrective action Codes from 20 to 25 identify semi fatal errors except for codes 21 and 25 The instrument must be returned to the factory for corrective action Error 20 Communication failed Error 21 Check of options failed Error 22 Check of contents failed Error 23 Check of calibration values failed Error 24 Check of instrument identification number failed Error 25 Check of print file failed For non fatal errors 21 and 25 it is not necessary to retu
82. several times around the exposed Megohmmeter Model 5070 87 surface Note If a shield is not available and you do not make up a shield around B and connect to the guard terminal the measurement will be erroneous and lead to confusion as to the cable s condition If the guard terminal is not connected at B the instrument measures the current i flowing through the insulation and a surface leakage current i1 See Figure A 2 No Connection to Guard Terminal B To EARTH To LINE Terminal Terminal A C Figure A 2 No Guard Terminal Connected With the guard terminal connected the surface leakage i1 is removed and has no effect on the reading See Figure A 3 To Guard Terminal T B To EARTH To LINE Terminal Terminal A C Figure A 3 Guard Terminal Connected 88 Megohmmeter Model 5070 APPENDIX B SERIAL CABLE REQUIREMENTS RS 232 Specifications The Baud Rate can be set to 300 600 1200 2400 4800 9600 or Parallel for printing results on parallel printers via the optional serial to parallel adapter in the SET UP mode see 4 1 The Data format is 8 data bits 1 stop bit no parity Xon Xoff protocol Serial Printer Connection Cable PC RS 232 DB9 F F 6 ft for serial printer Cat 2119 46 9 pin female connector 9 pin female connector to Model 5070 to serial printer Connection to a PC or to a Parallel Printer Cable PC RS 232 DB9 F F 6 ft
83. sions LxWxH 10 63 x 9 84 x 7 09 270 x 250 x 180mm Weight 9 5 Ibs 4 3kg approx 3 7 Safety Specifications Electrical safety according to EN 61010 1 Ed 2 for 2001 EN 61557 ed 97 CE Double insulation E 1000V CAT III 2500V CAT I Pollution Degree 2 Electromagnetic Compatibility NF EN 61326 1 Ed 97 A1 industrial environment category Mechanical Protection IP 53 per NF EN 60529 Ed 92 IK 04 per NF EN 50102 Ed 95 Megohmmeter Model 5070 17 3 8 Variations in Operating Range Influential Range of Quantity Influence Quantity Influence Influenced Typical Max Battery V ict 2cts Voltage SERE MO ict 3cts a V 0 15 R 10 C 0 3 R 1ct Temperature 10 to 55 C MO 0 20 R 1 R ict V 0 2 R 1 R 2cts Humidity 10 to 80 RH MQ 10kQ to 40GO 0 2 R 1 R 5cts MQ 40GQ to 10TQ 3 R 15 R 5cts Frequency 15 to 500Hz V 0 3 R 0 5 R 1ct AC voltage superimposed on 0 to 20 Vn MO 0 1 R Vn 0 1 R 9o Vn 5cts test voltage The terms DAR PI DD and the capacity and current leak measurements are included in the quantity MQ 18 Megohmmeter Model 5070 CHAPTER 4 SPECIAL FUNCTIONS 4 1 SET UP Function This function located at the top rotary switch position Blue SET UP position is used to change the configuration of the instrument 4 1 1 Default Configurations The default configurations are as follows D 0 ERUP
84. st type DC test voltage Programmed duration of the test AC DC input voltage Frequency Residual current Date and Time DURING the measurement DAR o A 234 5MO 5007V 20 6pA Elapsed time 00 00 46 kQ MQ GQ TQ 10 100 1 10 100 10 100 1 10 Information Displayed Measured resistance DC test voltage Leakage current Remaining measurement time Insulation bargraph 42 0 1V DC V 0 400 250 500 750 1000 20 6pA Information after pressing DISPLAY AC DC input voltage Frequency Residual current Voltage bargraph DAR o A 234 5MO 5007V 20 6pA Elapsed time 00 00 46 DAR 30 60 PI 1 10 Capacitance Information after pressing DISPLAY Measured resistance DC test voltage Leakage current Remaining measurement time DAR PI Capacitance in process Megohmmeter Model 5070 AFTER the measurement DAR i 5007V 20 6pA Elapsed time 00 01 00 DAR 30 60 2 64 PI 1 10 Capacitance 320nF DAR i FIXED VOLTAGE 5000 V Test Run Time 00 01 00 Input voltage 0 1V DC Frequency 0 0 Hz Input current 20 6pA Date 10 18 2003 Time 20 03 DAR i 0 0 Hz 20 6pA DC V O 400 250 500 750 1000 Information Displayed Measured resistance DC test voltage Leakage current Measurement duration DAR PI Capacitance After 1st Press on DISPLAY Test type DC test voltage Programmed duration of the test AC DC input voltage Frequency Leakage current Date and time After
85. t connection DataView Version x xx xxxx Installs DataView onto the PC Adobe Reader Links to the Adobe website to download the most recent version of Adobe Reader to the computer Adobe Reader is required for viewing PDF documents supplied with DataView DataView Updates Links to the online DataView software updates to check for new software version releases Firmware Upgrades Links to the online firmware updates to check for new firmware version releases 72 Megohmmeter Model 5070 Documents Shows a list of instrument related documents that you can view Adobe Reader is required for viewing PDF documents supplied with DataView 6 DataView Version x xx xxxx option should be selected by default Select the desired language and then click on Install 7 The Installation Wizard window will appear Click Next 8 To proceed accept the terms of the license agreement and click Next 9 Inthe Customer Information window enter a Name and Company then click Next 10 In the Setup Type window that appears select the Complete radio button option then click Next 11 In the Select Features window that appears select the instrument s control panel that you want to install then click Next NOTE The PDF XChange option must be selected to be able to generate PDF reports from within DataView i pa DataView InstallShield Wizard InstallShield Vi Ground Integrity Monit
86. the print mode depending on when it is accessed f you access the print mode right after the completion of a test you will see Print result as the first choice e If you access the print mode without running a test Print memory will be your first choice Print result PRINT Immediate printing of a measurement O Print result following a measurement or after access Print memory to the MR Memory Recall mode Baud rate Port 9600 RS232 Print memory Printing of stored data Baud rate Port Display of baud rate selected in the SET UP menu 32 Megohmmeter Model 5070 4 3 DISPLAY GRAPH Button 4 3 1 Primary Function DISPLAY The primary function of this button is used to browse through the various screens of information available before during or after the measurement The screens vary depending on the mode selected before the measurement is started This section starting on the following page shows typical screens that can be displayed for each test mode Megohmmeter Model 5070 33 MANUAL STOP mode BEFORE the measurement il FIXED VOLTAGE 500 V Input voltage 0 1V AC Frequency 0 0 Hz Input current 24 6pA Date 09 28 2003 Time 22 39 Information Displayed Test type DC test voltage AC DC input voltage Frequency Residual input current Date and Time DURING the measurement A 234 5MO 507V Elapsed time 000043 kQ MQ GQ TQ 10 100 1 10 100 i 10 100 i 10 Information Display
87. time is related to the total duration of the test Total Run Time It is equal to Sample Time seconds h 1 5 where h total run time in hours TEMPERATURE UNIT This function toggles the display between Fahrenheit and Celsius scales for tem perature di splay Range C or F Megohmmeter Model 5070 23 DEFAULT PROBE TEMPERATURE With this function you can program a default value for the temperature of the device under test This will be used when temperature correction is applied if no value is programmed right after the test is conducted Range 15 C to 75 C Rc REFERENCE TEMPERATURE Reference temperature to which the measurement result must be referred Range 15 C to 75 C AT for R 2 Insulation resistance changes with temperature The typical rule of thumb is that for every 10 C increase in temperature the leakage current doubles and the resis tance halves Some materials have a different rate of change This feature lets you program the temperature change AT at which the resistance halves for the equipment you will be testing This value will be used when correcting results to a reference temperature Range 15 C to 75 C CALCULATE AT FROM MEMORY The Model 5070 has the ability to calculate AT from three previously stored testsre sults at different temperatures from one material in the event that no AT is selected for the present test
88. urement is started by press ing START STOP button and stopped by pressing START STOP again 1 minute after the end of the measure ment the instrument calculates and dis plays the Dielectric Discharge DD ratio The time remaining during this minute is displayed TIMED RUN This mode is used to perform a measure ment for a duration defined in advance with a predetermined number of mea surement samples The measurement is started by pressing the START STOP button and stops automatically after the time programmed by the user has occurred 28 MODE Total Run Time n Manual Stop DD Duration Sample him mis Timed Run 02 30 01 40 Timed Run DD DAR s s 30 60 PI m m 1 10 MODE Total Run Time Manual Stop n Duration Sample him mis Timed Run 02 30 01 40 Timed Run DD DAR s s 30 60 PI m m 10 MODE Total Run Time Manual Stop Manual Stop DD Duration Sample him mis BENE 02 30 0140 Timed Run DD DAR s s 30 60 PI m m 1 10 02 30 00 Megohmmeter Model 5070 This duration Duration and the time interval between samples Sample must be specified when the Timed Run mode is selected To set the Duration and Sample rate from the MODE screen proceed as follows Use the A or V buttons to highlight Timed Run e Press the button to move the cursor to Duration the hour value will be highlighted Use the A or V buttons to select the desired hours from 0
89. urement of high resistance insulation a Example of a motor reduction of capacitive effects External insulator Braid Megohmmeter Model 5070 63 6 4 Printing Measured Values MODE PRINT Button To use a serial printer first choose the appropriate communication baud rate in the SET UP menu between 300 9600 baud Then program the printer to the 5070 format There are two print modes Instantaneous printing of measurement PRINT Printing of recorded data PRINT memory Pressing the PRINT button gives access to the following menu Print result Immediate printing of a measurement following that mea surement or after accessing values in the MR mode Print memory Printing of stored data Baud rate Port Baud rate adjusted in the SET UP menu The choices are 300 600 1200 2400 4800 9600 or parallel After selection of the print mode f the transmission of data to the printer is successful the COM symbol flashes at the top left of the display unit e Ifa problem occurs the COM symbol remains lit steadily at the top left of the display unit 6 5 Instantaneous Printing of Measurements Print Result When this print mode is selected the following are printed in order General information concerning the measurement The measurement result Ifthe TEMP function was activated the measurement result referred to the reference temperature Fora
90. urement performed the following formats are obtained Any measurement except step function measurements AEMC Instruments Model 5070 Instrument number 000 001 OBJECT 01 TEST 01 INSULATION RESISTANCE TEST Date 01 31 2003 Starting time 14h55 Running time 00 15 30 Temperature 23 C Relative humidity TP S Test voltage 1000 V Insulation resistance 385 GOhm Re calculated resist 118 5 GOhm at reference temperatur 40 C with AT for R 2 i0 c DAR 1 30 1 234 PI 10 1 2 345 DD Eu Capacitance 110 nF OBJECT 01 TEST 02 INSULATION RESISTANCE TEST Date 01 31 2003 Starting time 17h55 Running time 00 17 30 68 Megohmmeter Model 5070 a Temperature 23 C Relative humidity Siei A Test voltage 1000 V Insulation resistance 385 GOhm Rc calculated resist 118 5 GOhm at reference temperatur 40 C with AT for R 2 10 C DAR 1 30 1234 PI 10 1 2 345 DD ege Capacitance 110 nF Ho Date of next test 7200 DR RemarksSi esdews ae eee eee Operatori aro aye 33 xen us ere car 1e hare Signatures quede aUe Deren aos rares Step function measurement AEMC Instruments Model 5070 Instrument number 000 001 Megohmmeter Model 5070 OBJECT 01 TEST 01 STEP FUNCTION TEST Date 01 31 2003 Starting time 14h55 Running time 00 00 50 Temperature 23 C Relative humidity E Step
91. y stored with the final value of the resistance if it is stored i During the measurement if the position of the rotary switch is changed or the START STOP button is pressed the measurement is stopped Megohmmeter Model 5070 29 TIMED RUN DD This mode is identical to the Timed Run except that 1 minute after the end of the measurement the instrument calculates and displays the Dielectric Discharge DD The measurement duration is there fore the duration of timed run 1 minute The insulation resistance vs time curve can be displayed after the measurement by pressing the Yellow 2 button and the GRAPH button as long as no new mea surement has been started see 4 3 2 for a typical graph DAR The DAR measurement is started by pressing the START STOP button and stops automatically when the DAR ratio has been calculated e g after 1 minute the time taken to measure the second insulation resistance value needed for the calculation the ratio times can be modified in the SET UP mode The default is 30 60 meaning the first reading will be taken in 30 seconds and the second reading will be taken in 60 seconds from the Start PI The PI measurement is started by press ing the START STOP button and stops automatically when the PI ratio has been calculated e g after 10 minutes the time taken to measure the second insulation resistance value needed for the calcula tion the ratio times can be programmed in the S

Download Pdf Manuals

image

Related Search

Related Contents

Traffic Simulation Toolbox  Martha Stewart Living 1607200400 Instructions / Assembly  Samsung S760 User Manual  Elizabeth Fireplace Screen Model # 885A  LA SIGNALÉTlQUE EN BIBLIOTHEQUE  Energy Tablet a4 Dark Iron 8GB Energy Sistem  

Copyright © All rights reserved.
Failed to retrieve file