Home
Clifton NE 5 latest model
Contents
1. Class 1 thermal resetting cut out unit protection 100 C Alarm auto set Set actual temp C 0 Press twice Press either button to change set temperature Press to confirm set temperature Press to return to Normal operation screen The Clifton Range 3 Year Warranty 9 High Quality Products at Affordable Prices Alarm events and status If the actual temperature is 4 above or below set temperature the temperature value is replaced with the word HIGH or LOW and screen changes from GREEN to RED If OPEN or ERROR appears seek advice from our Service department do not use Trend view Trend view graphs actual water temperature data with set temperature on screen Press twice To review data pressing either down or up moves the cursor through the last 240 temperature readings Trend view graphs actual water temperature data solid line with set temperature dotted line every 30 seconds Trend scale adjusts automatically to actual water temperature data Data is not retained when unit is turned OFF or if sample interval is changed Press to return to Normal operation screen The Clifton Range 3 Year Warranty 10 High Quality Products at Affordable Prices Program setup The Shaker bath can be programmed by choosing program options Delay start Ramp timer and Dwell at set temprature Program delay start ramp timer and dwell at set
2. BX0843 www nickel electro co uk
3. clips racks etc and glassware then adjust the water level to suit Optimal temperature performance liquid level needs to cover the top of the trolley 75 to 120mm Always ensure the product is disconnected from the electrical supply when emp tying and filling The filling level depends on the size and number or vessels being placed into the bath Suitable Liquids Operating temperatures of ambient to 99 C for general use we recommend Distilled water Heat transfer liquid The LB range is formulated for temperatures from 15 C to 90 C and provides complete protection from freezing and algae growth and safe guards against corrosion See accessories for the full range available VIRKON dissolved in distilled water has proven efficacy against bacteria including mycobacteria viruses spores and fungi in a variety of independent tests using different protocols Above 60 C or below room temperature it is recommended that to achieve optimum performance the bath should be covered with SL1 lid or polypropylene spheres The Clifton Range 3 Year Warranty High Quality Products at Affordable Prices NE5 10D Front Shaker Drive Arm connects to Trolley Recessed Tray for flasks to stand on Display interface temperature settings Clifton temperature profile temeprature log Indicators heating alarm tune and program Keyboard menu navigation up and down function acknowledge buttons Speed cont
4. 1787 223232 or http www day impex co uk The ultimate high level surface disinfectant dissolve VIRKON in water providing a safe working solution with a faint lemon odor It has proven efficacy against bacteria including mycobacteria viruses spores and fungi in a variety of inde pendent tests using different protocols Presents no serious long term health risks to staff obviating the need for costly ventilation equipment and health monitor ing Also provides high level disinfection of laboratory equipment and instruments where autoclaving is neither practical nor necessary For more detailed informa tion relating to how Virkon should be used with access to test reports www relyon dupont com Is Virkon solution corrosive Virkon solution requires only 10 minutes contact time to be effective so long term exposure is not necessary and therefore will not corrode most materials Care should be taken with Stainless steel water bath tanks these surfaces should not be affected however it is important that generally you do not leave Virkon solution in contactwith metal surfaces FOR LONGER THAN IS NECESSARY Virkon is Registered in accordance with the requirements of the Medical Devices Directive 93 42 EEC as a Medical Device Disinfectant Sterilant We recommend PeraSafe a powder product for the safe and rapid chemical sterilant of equipment in a wide variety of situations available from your distributor or contact Day Impex Ltd
5. for more details Telephone 44 0 1787 223232 or http www day impex co uk PeraSafe has a proven safety profile for end users with none of the undesirable properties of skin sensitisation toxic fumes or unpleasant odours that are associated with aldehyde solutions Leading UK and USA microbologists have proven PeraSafe to be active against viruses mycobacteria and fungi It is microbiologically superior to glutaraldehyde destroying sporing bacteria in one minute It has also been independently proven that PeraSafe sterilises in just 10 minutes For more detailed information relating to how PeraSafe should be used with access to test reports www relyon dupont com The Clifton Range 3 Year Warranty 16 High Quality Products at Affordable Prices 3 Year Warranty Our service engineers are fully trained in the assembly calibration and servicing of all Clifton instru mentation Products can be returned to our comprehensively equipped service centre where a fast and efficient turnaround is guaranteed Service Department Nickel Electro Limited Oldmixon Crescent Weston super Mare North Somerset BS24 9BL UK Tel 44 0 1934 626691 Fax 44 0 1934 630300 Out of Warranty Our Service Department has comprehensive stock of charegeable spare parts maintaining working life of equipment or units can be returned for quotation before repairs are undertaken End of Life This symbol indicates that this product should not be disposed
6. of with your waste Instead dispose waste electrical equipment by handing it over to a designated col lection point for the recycling of waste electrical and electronic equipment The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protects human health and the environment For more information about where a you can drop off your waste equipment for recycling in UK please contact Service Department rest Europe contact your Distributor Health amp Safety unless in receipt of a Decontamination Notice or Report the unit cannot be returned or accepted for disposal Clifton electrical and electronic equipment has been designed for recycling and takes into account the dismantling and recovery its components and materials Clifton products are easily recycled with majority of the product constructed from stain less or mild steels which can readily be re used or recycled In excess of 78 of this product range can be easily re cycled economically Portable Appliance Testing When conducting testing ensure it is conducted by a qualified person DO NOT PAT TEST THE BATH UNLESS IT CONTAINS WATER THIS EQUIPMENT MUST NOT BE FLASH TESTED The Clifton Range 3 Year Warranty 17 High Quality Products at Affordable Prices Service Diagram for 5 28 Service Diagram for NE5 10D and NE5 28D The Cl
7. Anti bacterial active Ag ions tHE Clifton rance Shaking Baths NE5 Series IMPORTANT ALWAYS FILL the water bath before connecting to the power supply Contents About this Manual General Notes Amendments Symbols Location Unpacking Safety Assembly Power Supply Lead and Connection to Electrical Supply Liquid Level Suitable Liquids Front Top and Rear Views NE5 10D Front and Rear Views NE5 28 and NE5 28D Operating Instructions NE5 28 Operating Instructions NE5 10 and NE5 28D Setting Shaking Speed Stroke adjustment General Cleaning Cleaning external painted surfaces with Anti bacterial finish Cleaning the stainless steel tank Descaling the stainless steel tank Decontamination of Equipment Virucidal Disinfectant Disinfectant Sterilant 3 Year Warranty Out of Warranty End of Life Portable Appliance Testing Declaration of Conformity Final Inspection and Electrical Safety Test Report The Clifton Range 3 Year Warranty CONOUA AA A A A AW WW W Enclosed Rear High Quality Products at Affordable Prices About this Manual This user Manual contains instructions which must be followed in order that the product is operated correctly General Notes Please observe the following safety precautions 1 Always follow good laboratory practice by ensuring substances being heated offer no risk of a hazard explosion implosion or release of toxic or flammable gases o
8. and Clip trays see accessories can then be placed on the trolley 10 Litre Bath Place the trolley into bath locating the drive arm into the bush and tighten the locking knob on the drive arm by hand Fit the eyebolt handles used for liting the perforated tray out of the bath one on each side using one of the pair of outer holes Safety Do not touch any electrical contacts or open any closure panels RISK OF ELECTRICAL SHOCK NE5 10D and NE5 28D are Class 1 IEC519 Part 2 reference to over temperature condition provides product over temperature protection NE5 28 is a Class 2 IEC519 Part 2 reference to over temperature condition provides product over temperature protection and liquid over temperature protection suitable for heating materials which could present a safety hazard at the operating temperature of the over temperature system Power Lead and Connection to Electrical Supply Before connecting the product to the electrical supply check the information on the rating label is compatible IF INDOUBT CONSULT AN ELECTRICIAN THE PRODUCT MUST BE EARTHED Where the mains supply or plug connection differs refer to local regulations or qualified electrician Liquid Level Minimum working liquid level must cover the top of the trolley by 40mm Maximum flask immersion 120mm from top of trolley Recommend fill to minimum liquid level first with trolley in position place the perforated shelf with requried
9. and switch the bath back on The Clifton Range 3 Year Warranty 14 High Quality Products at Affordable Prices Cleaning General Important please follow these instructions to avoid possible damage to the unit otherwise affecting its performance and warranty Always disconnect the product from the electrical supply before cleaning Cleaning External Painted Surfaces featuring Anti bacterial Paint Finish The water bath should be cleaned at regular intervals wiping external surfaces with a cloth or sponge soaked in warm soapy water with a mild detergent All surfaces should be cleaned using a soft cloth or sponge Do not under any circumstances use strong solvents or solutions containi ng Chlorinated Hydocarbons Esters Ketones or abrasive cleaners or polish on the paint finish otherwise it damages the built in anti bacterial properties All painted surfaces on Clifton range products features an Anti bacterial paint finish identified with this authenticating logo on the unit This Anti bacterial paint finish inhibits the growth of bacteria It has been Pr tection a ie tested by independent specialist test houses such as the Law Laboratories in active Ag ions the UK using internationally recognized test methods and proven to be f effective versus a wide range of bacteria species including Escherichia coli and Staphylococcus aureus MRSA We recognise hygienic coatings are
10. d high concentrations of halogens especially chloride With such a high quality and resistant tank it may show symptoms of these halogens as rust which are deposits from external sources in the water supply We recommend always empty the bath of liquid after use and wipe out the internal faces of the tank with a non abrasive cloth and allow to dry Any deposits can be removed with nitric acid 10 ona cloth WEAR PROTECTIVE EQUIPMENT It is also recommended to use an accessory lid to prevent contaminates landing in bath liquids The Clifton Range 3 Year Warranty High Quality Products at Affordable Prices Descaling the Stainless Steel Tank Descale the stainless steel tank regularly to maintain it in as new condition ensuring the corrosion resistance and normal operating conditions are maintained throughout its working life Descale by adding 1 litre of vinegar to water and gently heating to 50 C for an hour empty and brush the lime away Rinse thoroughly afterwards Decontamination of Equipment Clifton laboratory equipment can be decontaminated after spillage or contact with potentially HIV and Hepatitis infected blood samples during analysis using following recommended rapid disinfect ants Virucidal Disinfectant We recommend Virkon tablets for the safe and rapid disinfection of equipment in a wide variety of situations available from your distributor or contact Ltd for more details Telephone 44 0
11. electrical supply details IEC connector for power supply lead and plug Fuse draw fitted underneath The Clifton Range 3 Year Warranty 6 High Quality Products at Affordable Prices Operating Instructions N The Clifton Range 3 Year Warranty E 5 28 Analogue Shaking Bath 1 Ensure the bath has the required liquid level then switch the unit ON I using the mains switch located on the front of the unit 2 Adjust the temperature control to the desired setting using the blue knob Temperature can be be set between ambient 5 C to 99 C Heater neon will illuminate indicating heater activity 3 Allow the water temperature to stabilise taking an actual reading in the middle of the bath 4 Adjust the shaker speed by rotating the dial Speed can be set between 0 380 strokes per minute 5 Optional Setting Adjustable Overtemperature Safety Device a Ensure the adjustable overtemperature safety device dial is turned fully clockwise b Set the temperature on the controller at which you require the device to operate at using previous instructions for set point 1 Allow the liquid to reach this temperature and settle c Slowly turn the dial back anticlockwise until it clicks d Now adjust the temperature on the controller to the actual required operating temperature Allow the bath to cool down to this temperature and it will control at this set point and the adjustable overtemperature safety dev
12. evel Issue 11 November 2010 2011 models Pro4 Issue 12 March 2011 UpdatesJMD Symbols HOT SURFACES Paragraphs marked by this symbol indicate that a potential hazard to your personal safety exists from heated surfaces or other appendages on the outside or inside of the equipment CAUTION This icon accompanies text and or other international symbols dealing with poten tial damage to equipment When present it indicates that there is a potential danger of equipment damage may occur if information stated within the CAUTION paragraph is not adhered to or procedures are executed incorrectly PROTECTIVE EARTH OR GROUND TERMINAL Protective earth conductor terminal The Clifton Range 3 Year Warranty High Quality Products at Affordable Prices Location The product must be placed on a smooth level and sturdy work surface preferably near a sink or drain for emptying NE5 28 versions are fitted with a drain outlet Use in a ventilated room Suitable for use in ambient temperatures 5 C to 40 C with a maximum humidity 80 temperature 31 C decreasing to 50 temperature 40 C The product is designed for laboratory use Unpacking Remove the product from its packaging If there is any damage please notify your dealer immediately Retain the packaging over the warranty period Assembly 28 Litre Baths Place the trolley into bath locating the drive arm into white bush and tighten the locking knob on the drive arm by h
13. ice is now set During a fault condition where the safety device operates heating is cut off 6 The stroke length can be adjusted by opening the access door and rotating the adjustment screw Stroke length is variable between 0 40mm a Open the access door on the side A warning neon will illuminate indicating it is safe to adjust the stroke Switch the bath off b Turn the drive block by hand until a knurled adjuster screw is visible c Rotate this screw using a flat screwdriver clockwise to reduce stroke length or anticlockwise to increase stroke length On the drive arm there are a number of graduation lines indicating stroke length the first line indicates 10mm further lines are then positioned at 10mm intervals The factory setting is 20mm d Increasing stroke length decreases quoted speed and likewise decreasing stroke length increases quoted speed e Close the door and switch the bath back on If the access door warning neon DOES NOT illuminate when the door is opened switch the bath OFF at the mains switch and disconnect from main electrical supply The Shaker bath must be serviced by a qualified electrician High Quality Products at Affordable Prices Operating Instructions NE5 10D and NE5 28D Switching ON and OFF Switching ON the unit may be turned ON 1 at the mains switch located at the rear When ON 1 the switch is illuminated and unit performs a self test Clifton logo appears and then actual and se
14. ifton Range 3 Year Warranty High Quality Products at Affordable Prices EC Declaration of Conformity We herewith confirm the following product NE5 Shaking Waterbath Range Conforms with the requirements outlined by following European Directives Low Voltage Directive 2006 95 EC EMC Directive 89 336 EEC We confirm the declaration NICKEL ELECTRO LIMITED Manufacturers of laboratory medical and clinical equipment Oldmixon Crescent Weston super Mare North Somerset BS24 9BL United Kingdom Tel 44 0 1934 626691 Fax 44 0 1934 630300 Email clifton nickel electro co uk www nickel electro co uk Nickel Electro Ltd is also registered ISO9001 reference No Q09820 Conforms with the requirements of following Standards BS EN 61010 1 BS EN 61010 2 010 Safety requirements for electrical equipment for measurement control and laboratory use BS EN 61326 Electrical equipment for measurement control and laboratory use EMC requirements Nickel Electro Ltd is also registered 1509001 reference No Q09820 The Clifton Range 3 Year Warranty 20 High Quality Products at Affordable Prices Final Inspection and Electrical Safety Test Report NICKEL ELECTRO Ltd Manufacturers of laboratory medical and clinical equipment idmgon Crescent VWeeston super Mare North Somerset B534 SBL United Kingdom 44 0 1934 630300 Email cliftong nickol oloctro co uk www nickel electro co uk
15. part of a controlled approach to a cleaner working environment Within its formulation an active ingredient with proven anti bacterial properties is bound into the paint finish The efficacy of the paint finish applied to the Clifton range is maintained over its lifetime as the anti bacterial agent is integral within the paint In a laboratory environment it makes this one less source of contamination contributing to essen tial clean working practices A benefit of such a paint finish can lead to a reduction in cleaning schedules because surfaces are more protected and improves protection between cleaning Unlike detergents Anti bacterial paint finish does not offer an instantaneous action but is intended for long term general protection against bacterial growth Moisture on the painted surface is necessary for the bacterium to absorb the agent and be affected by it The coating is therefore less active in very dry conditions but dependent upon relative humidity moisture in the atmosphere maintains activity Areas where moisture is trapped are also areas that normally are difficult to clean and where bacteria proliferate but these areas are most active for the anti bacterial coating improving the defence against bacterial growth Cleaning the Stainless Steel Tank The stainless steel crevice free tank with smooth corners should provide years of valuable service and is resistant to chloride containing solutions it is however important to avoi
16. r that these have been addressed When heating substances where liberation of gases occurs suitable extraction should be used 2 If this product is not used in accordance with these instructions then basic safety protection afforded by the water bath may be affected NE5 28 is classified as Class 2 IEC519 with regards to over temperature protection 3b The NE5 10D and NE5 28D is classified as Class IEC519 with regards to overtemperature protection 4 Check the operation of the over temperature device regularly 5 The mains supply cord fitted to this products is a heat resistant type and should be replaced by an equivalent type It is recommended the units are connected to the mains supply with RCD protection 7 Before using any cleaning or decontamination method except those recommended check with your distributor that the proposed method will not damage the equipment On Amendments Issue 1 January 1996 Initial issue instruction book Issue 2 May 1996 Address change Issue 3 August 1996 Multilingal version fof first issue book Issue 4 January 2001 Update to product range Issue 5 September 2001 Layout changes Issue 6 November 2004 Revised product and instruction manual design Issue 7 March 2006 Service diag update OT amended NE5 28 WEEE Issue 8 July 2006 NE5 10D trolley update NE5 28D adjustment stroke length Issue 9 December 2006 Anti bacterial paint finish and WEE Issue 10 September 2008 Liquid L
17. rol dial Top View Eyebolt handles for easy removal of perforated shelf Perforated shelf for accessory clips trays test tube racks can be fitted Rear View Mains power switch I ON O OFF Accessory RS232 communication No longer available Chart logging featured on Display interface Serial number and electrical supply details IEC connector for power supply lead and plug Fuse draw fitted underneath The Clifton Range 3 Year Warranty High Quality Products at Affordable Prices NE5 28D Shaker Drive Arm connects to Trolley Front View Access door side view stroke adjust ment Recessed Tray for flasks to stand on Display interface temperature settings temperature profile temeprature log Indicators heating alarm tune and program Keyboard menu navigation up and down function acknowledge buttons Access door open warning indicator Speed control dial Drain outlet for accessory tap NE5 28 Shaker Drive Arm connects to Trolley Front View Access door side view stroke adjustment Recessed Tray for flasks to stand on Indicators Temperature control dial Adjustable Over temperature safety device Speed control dial NE5 28D and NE5 28 Drain outlet for accessory tap Rear View Mains power switch I ON O OFF Accessory RS232 communication No longer available Chart logging featured on Display interface Serial number and
18. t water temperature is displayed ready to use Switching OFF the unit may be turned OFF QO at the mains switch located at the rear All temperature and time values remain in memory Heating Control Modes The Clifton NE5D range feature an advanced PID temperature controller that can be used in either of the following ways temperature control process control with delay ramp and soak at set temperature Heating indicator Alarm indicator Tune indicator Program indicator Controls LED indicator labels Actual water temperature Set temperature Heater power bar graph 0 to 100 Shaking speed dial Access door NE5 28D Keyboard Description MOVE BACK IN MENU Accept value Or use to move back through screens NEXT ITEM DECREASE Used to decrease a value Or next item in menu NEXT ITEM INCREASE Used to increase a value Or next item in menu MOVE FORWARD IN MENU ACKNOWLEDGE Accept value Or use to move forward through screens MOVE DOWN ONE MENU LEVEL Press and hold simultaneously Move down one menu level The Clifton Range 3 Year Warranty High Quality Products at Affordable Prices Normal operation screen Following power up self test and Clifton logo screen the next screen below will appear Most screens will revert back to this screen after couple of minutes without activity Temperature Control only Temperature control Temperature range gt
19. temperature with timer And showing alarm and messaging Temperature range 7 Class 1 thermal resetting cut our unit protection Unstable sensor error message Unstable condition 10 triggers alarm Alarm auto set 3PC Set actual temp orc Delay 13 Dwell Timer 12 start setting setting hh mm ss hh mm ss Press 5 times to access program setup Press to move cursor between each time value hh mm ss for delay ramp and dwell Also navigates to set temperature for dwell Press either button to change values Delay time Ramp time Dwell time hh mm ss hh mm ss hh mm ss Delay timer upto 99h 99m 99s Setting of 00h 00m 00s provides no delay program starts Ramp timer controls rate and rise in temperature with time upto 99h 99m 99s A setting of 00 00 00 provides immediate rise to set temperature Dwell timer set temperature held upto 99h 99m 99s Setting of 00 00 00 produces an infinate time The Clifton Range 3 Year Warranty 11 High Quality Products at Affordable Prices Keep pressing until Program Control screen shown below appears Press to navigate to highlight start Press Then press to highlight yes Press to start program To re run the program press to navigate to Program screen Navigate to highlight start The Clifton Range 3 Year Warranty High Quality Products at Affordable Prices Main Menu Press both b
20. uttons together Main menu screen will appear Automatic Tuning Select this option enter code 10 using the increase decrese buttons and then acknowledge Select YES and autotune will start Useful function allows the bath to learn about your applications its environment and temperatures Configuration Menu Product information Our service engineers may guide you through these menus if required Service information Provides contact details for product servicing and support The Clifton Range 3 Year Warranty 13 High Quality Products at Affordable Prices Setting Speed Control Adjust the shaker speed by rotating the dial Speed is variable between 0 380 strokes per minute Stroke adjustment NE5 28 and NE5 28D Stroke length is variable between 0 40mm 1 Open the access door on the side 2 Power is isolated from controls and drive system a warning indicator will illuminate safe to adjust the stroke 3 Turn the drive block by hand until a knurled adjuster screw is visible 4 Rotate this screw using a flat screwdriver clockwise to reduce stroke length or anticlockwise to increase stroke length On the drive arm there are a number of graduation lines indicating stroke length the first line indicates 10mm further lines are then positioned at 10mm intervals The factory setting is 20mm Increasing stroke length decreases quoted speed and likewise decreasing stroke length increases quoted speed Close the door
Download Pdf Manuals
Related Search
Related Contents
Transferarbeit Urs Zihlmann · HSLU iPMT RK 80 B INSTRUCCIONES DE INSTALACIÓN DE COCINAS DE GAS DE 30" API 550 User Manual Frigidaire FBD2400KB Energy Guide Electro-Voice Loudspeaker System XiA-Series User's Manual Panasonic S-07MD1U6 Installation Manual User's Guide FPAR, LAI (ESDT: MOD15A2) 8 電磁流量計 変換器 Samsung SM-N900VZKEVZW User's Manual Copyright © All rights reserved.