Home
EDP3-Plus - Mettler Toledo
Contents
1. 2 5 0 025 1 2 0 012 range 0 5 10 pL 5 1 5 0 075 0 6 0 03 10 1 0 0 10 0 4 0 04 E8 20 E12 20 2 0 02 7 5 0 15 2 0 0 04 E2X8 20 E24 20 10 1 5 0 15 05 0 05 range 2 20 pL 20 1 0 0 20 0 3 0 06 E8 50 E12 50 5 0 05 35 0 18 1 5 0 075 range 5 50 pL 25 1 2 0 30 04 0 1 50 08 0 40 0 2 0 1 E2X8 100 10 0 1 3 5 0 35 10 0 1 E24 100 50 08 04 0 24 0 12 range 10 100 pL 100 08 08 0 15 0 15 E8 200 E12 200 20 0 2 25 0 5 1 0 0 2 range 20 200 pL 100 08 08 0 25 0 25 200 0 8 1 6 0 15 0 3 E8 300 E12 300 30 0 2 2 5 0 75 1 0 0 3 range 20 300 pL 150 0 8 1 2 0 25 0 375 300 08 24 0 15 0 45 E8 1200 100 1 0 3 6 36 0 60 0 60 E12 1200 600 08 48 0 20 1 20 range 100 1200 pL 1200 08 96 0 15 1 80 Specifications subject to change without notice Service If you experience any problems with your multichannel EDP3 pipette contact RAININ Technical Support at 800 543 4030 Ordering Information Cat No Description Pipettes 8 channel E8 10 8 channel 0 5 to 10 uL E8 20 8 channel 2 uL to 20 uL E8 50 8 channel 5 uL to 50 uL E8 200 8 channel 20 uL to 200 uL E8 300 8 channel 20 uL to 300 uL E8 1200 8 channel 100 uL to 1200 uL 12 channel E12 10 12 channel 0 5 to 10 uL E12 20 12 channel 2 uL to 20 uL E12 50 12 channel 5 uL to 50 uL E12 200 12 channel 20 uL to 200 uL E12 300 12 channel 20 uL to 300 pL E12 1200 12 channel 100 uL to 1200 uL 2x8 16 channel E2X8 20 16 channel in 2x8 configuration 2 uL
2. E Miguel Hernandez 69 71 08908 L Hospitalet de Llobregat Barcelona ES Ventas 0034 93 223 7666 mtemkt mt com Servicio T cnico 0034 93 223 7666 mtemkt mt com Mettler Toledo A S Naverland 8 2600 Glostrup DK Salg 43 27 08 25 info mtdk mt com Service 43 27 08 25 info mtdk mt com Rainin Instrument LLC 7500 Edgewater Drive Box 2160 Oakland CA 94621 0060 a METTLER TOLEDO Company Prices and specifications are subject to change without notice Copyright 2007 Rainin Instrument LLC 9920 256 Rev L
3. Liability Statement Please complete and return the Warranty Registration Card on receipt of your pipette RAININ pipettes are calibrated with RAININ tips To assure excellent reproducibility and performance use only RAININ tips as recom mended in this manual Specified performance is guaranteed only when RAININ tips are used Contacting RAININ Technical Information T 800 543 4030 F 510 564 1617 tech support rainin com Pipette Service T 800 662 7027 F 781 935 7631 service rainin com Direct Order Line T 800 472 4646 F 510 564 1617 pipets rainin com RAININ website www rainin com From outside North America T 1 510 564 1600 global rainin com from outside the U S METTLER TOLEDO Offices METTLER TOLEDO website www mt com rainin Mettler Toledo Schweiz GmbH Im Langacher 8606 Greifensee CH Verkauf 044 944 45 45 salesRN ch mt com Service 044 944 47 47 serviceRN ch mt com Mettler Toledo GmbH Ockerweg 3 35396 Giessen DE Verkauf 49 641 507 222 info mtd mt com Service 49 641 507 307 info mtd mt com Mettler Toledo SAS 18 20 avenue de la P pini re 78220 Viroflay FR Ventes 013097 17 17 marcom fr mt com Laboratoire d Etalonnage Accr dit 264 rue Abraham Lincoln Service Pipette 62400 Bethune FR Service 03 21 64 54 66 marcom fr mt com N V Mettler Toledo S A Leuvensesteenweg 384 1932 Zaventem BE Sales 32 2334 0211 general mtb mt com Service 32 2334 0211 general mtb mt com Mettler Toledo S A
4. to 20 uL E2X8 100 16 channel in 2x8 configuration 10 uL to 100 uL 24 channel E24 20 24 channel in 2x12 configuration 2 uL to 20 uL E24 100 24 channel in 2x12 configuration 10 uL to 100 uL Note LTS multichannel pipettes require LTS tips Accessories HU M3 Package of 3 Magnetic Hang Ups Includes adhesive disk for attachment to a non ferrous vertical surface HU S3 Shelf clamp with 3 Hang Ups attached CR 7 Free standing Pipette Carousel for up to 7 pipettes HU A3 Adapter for Multichannels for HU M3 HU S3 amp CR 7 6107 040 Battery 6107 249 Battery for E12 1200 Note One of the following charging devices must be used to charge EDP3 Plus battery E3 WPS Wall Power Supply Recharges EDP Plus in about an hour Allows the pipette fo be used while being charged E3 RCS Rapid Charge Stand Recharge up to three EDP3 Plus pipettes Includes E3 WPS Bench or wall mounting To order pipettes or accessories call 800 472 4646 or visit www rainin com on the web RAININ is a registered trademark and EDP3 and LTS are trademarks of Rainin Instrument LLC EDP3 Pipettes are manufactured under U S Patents 4 671 123 4 905 526 5 187 990 5 614 153 6 299 841 and 6 254 832 LTS LiteTouch Tip Ejection System is protected under U S patents 6 168 761 B1 and 6 171 553 B1 Other U S and national patents pending Prices and specifications subject to change without notice Limited Warranty See the Limited Warranty and Limitations of
5. IN LTS tips E8 10 E8 20 E12 10 E12 20 E2X8 20 amp E24 20 20uL LTS tips E8 50 E12 50 E8 200 E12 200 E2X8 100 amp E24 100 250 uL LTS tips E8 300 and E12 300 300 uL LTS tips E8 1200 and E12 1200 use 1200 uL LTS tips recommended OR 1000 pL LTS tips picks up 1200 pL except aerosol resistant LTS tips have a cylindrical cross section with a well defined seal ring thin wall and positive stop They seal properly on the shaft and cannot be jammed or forced too far onto the shaft nozzles Mounting LTS Tips 8 and 12 channel Mounting racked LTS tips onto E8 and E12 pipettes is simple First align the shaft nozzles into the row of tips holding the pipette at a slight angle Figure 2 1 Position the pipette upright and press the nozzles into the tips until the positive stop is reached Figure 2 2 The tips are now mounted with proper sealing FIGURE 2 MOUNTING RACKED LTS TIPS 8 AND 12 CHANNEL You do not need hand tighten use heavy pressure or rock the nozzles onto the tips to obtain a good seal Mounting LTS Tips 2x8 and 24 channel 1 Align the first row of shaft nozzles with the first row of tips holding the pipette almost vertically Move the pipette as shown by the arrow to align the second row of tips You may need to rock or rotate the pipette to get all tips aligned E2X8 20 E24 20 E2X8 100 E24 100 TIPRACK FIGURE 3 MOUNTING RACKED LTS TIPS DOUBLE RO
6. RAININ Pipetting 360 EDP3 Plus Multichannel Electronic Pipette 8 12 16 2x8 24 channel models EDP3 Plus LTS Electronic Multichannel Pipette EDP3 Plus multichannel pipettes are based on EDP3 Plus single channel models which use the patented LTS LiteTouch Tip Ejection System Four versions of the multichannel EDP3 Plus are available 8 channel E8 12 channel E12 16 channel in two rows E2X8 and 24 channel in two rows E24 8 and 12 channel models are available in 6 volume ranges 0 5 10 uL 2 20 uL 5 50 uL 20 200 pL 20 300 pL and 100 1200 uL The 2x8 channel and 24 channel models are available in 2 volume ranges 2 20 uL and 10 100 pL This is an addendum to the EDP3 Plus manual 9920 235 which describes pipette operation modes and options Packing List In addition to this manual the pipette box should contain EDP3 instruction manual 9920 235 Rainin Test Report Performance Assurance Brochure sample tips and labels Speed Setting The maximum speed setting for E8 and E12 multichannel pipettes is 10 for the 10 300 uL models and 9 for the 1200 pL model The max speed setting for E2X8 and E24 pipettes is 10 DISPLAY 4 KEY PAD o ood DUAL TRIGGERS me DuaL TRIGGERS sE TIP EJECTOR BUTTON N SHAFT COUPLING LIQUID END MANIFOLD n y LTS SHAFT NOZZLES LTS DISPOSABLE TIPS FIGURE 1 E8 200 PIPETTE Tip Selection EDP3 multichannel pipettes must be used with RAIN
7. W MODELS 2 When you are satisfied that all tip nozzles are aligned with the tips lightly press both rows of shaft nozzles into the tips until you feel the positive stops in the tips 3 While maintaining slight pressure rock the pipette back and forth and in a circular direction to ensure both rows of tips are attached Do not press very hard a little practice will help you find the optimal pressure and amount of rocking The tips are now mounted with proper sealing Tip Immersion Depth The recommended depth for tip insertion into the sample for each EDP3 Plus multichannel model is shown in the table below EDP3 Plus Model Immersion Depth E8 10 E12 10 2 3mm E8 20 E12 20 E8 50 E12 50 E2X8 20 E24 20 2 3mm E8 200 E12 200 E2X8 100 E24 100 3 6mm E8 300 E12 300 3 6mm E8 1200 E12 1200 6 10 mm Tip immersion depth is important If exceeded the volume mea sured may be inaccurate possibly out of specification Tip angle Operate the pipette within 20 degrees of vertical Tip Ejection Simply press on the tip ejector button All tips are ejected clean ly with minimal pressure on the thumb Positioning the Liquid End Manifold The liquid end manifold can be rotated to any angle for conve nience You do not need to loosen the coupling nut AutoPace AUTO AutoPace allows multiple aliquots of the same volume to be dis pensed repeatedly at equal time intervals after the dispense trig ger is pressed
8. e blowout sequence and obtain more time for touch off by holding in the dispense trigger for the dispense stroke instead of pressing and releasing the trigger Hold the trigger only long enough to touch off both rows of tips then release the trigger You can also touch off both rows of tips at the same time touch off all tips together at the bottom of the wells Practice both methods to develop your technique Power requirements for E12 1200 E12 1200 has different power requirements than other RAININ electronic multichannel pipettes It can be operated a with the battery b with the battery as well as connected to the option al wall power supply However the high forces in E12 1200 preclude operation with the wall power supply alone Note Unique tips for 1200 pL models The RT L1200 and RT L1200S tips are specifically designed for use with 1200 uL models in MULTIDISPENSE mode The unique design of these tips prevents drop formation and eliminates the need for touch off between dispenses in MULTIDISPENSE mode The RT L1200F aerosol resistant tip allows a full 1200 uL to be aspirated under the filter However fluid dynamics dictate that touch off may be required when multidispensing Specifications These manufacturer s specifications should be used as guide lines when establishing your own performance specification Volume Increment Accuracy Precision Model uL uL UP ul s E8 10 E12 10 1 0 01
9. once For example on the 300 pL volume pipette all channels can dispense ten 30 uL aliquots at 2 second intervals with one touch of the trigger In MULTI mode press and hold the mode key to reach the options menu Then click the mode key until auto appears on the dis play The first time used the display will flash off Use the arrow keys to set the time interval between dispenses The range is 0 2 seconds fo 25 seconds in 0 1 second increments MULTIDISPENSE AUTO f Autodispensing set in MULTI mode Al Note AUTO symbol flashes when paused con and is steady when active SX 20404 pl FIGURE 4 AUTOPACE When the time interval is set press either trigger to pick up the set volume Then press the trigger again to begin dispensing automatically at the set time interval Pressing either trigger pauses autodispensing until either trigger is pressed again when autodispensing restarts At the end of the last dispensed aliquot press RESET to begin the next pickup Note for 2x8 and 24 channel models omai 109 4 FIGURE 5 E2X8 20 L AND E24 100 R On the 2x8 channel and 24 channel models both rows of tips need to be touched off Normally dispense and blow out are performed automatically by EDP3 Plus as you press and release the trigger for dispensing This may not provide enough time to touch off the second row on double row models if you touch off each row separately You can delay the dispens