to download. - Pulsatom Healthcare

image

Contents

1. 88 88 2 B8 B B8 Low mg dL fil Aa mmol L LILI cope ipat EDEN ERR HOUR MIN DAY MONTH YEAR 10 01 11 02 09 COS o HOUR MIN DAY MONTH YEAR 10 01 n 02 09 If no sample is applied within 60 seconds the monitor switches OFF automatically Press the POWER Button for 2 seconds to manually switch OFF the monitor 6 Date and Time Settings Your GM3 Blood Glucose Meter is preset with IST date and time Even ifthe date and time are incorrect test results are not affected You can reset the date and time on your GM3 Glucose Meter When you press the POWER Button the monitor switches ON and displays the initial screen in which all the segment s of the display are shown Press the POWER button once again on your monitor when the initial display as shown in the picture is shown to enter the time and date setting mode Your monitor now displays the current time and date set on your monitor and the hour is blinking Press NEXT or POWER Button to increase or decrease the hour setting till it displaysthe desired hour value 10 nour MIN AVG pay MONTH YEAR 88 88 MEN 88 18 88 i 8 I Amen II cone ch ir ERR DAY MONTH YEAR 01 81 01 HOUR MIN 22 01 DAY MONTH YEAR i Bi Wait for 2 seconds to save the hour setting and to go to the minutes set mode The minutes now start to blink Press the NEXT or POWER Button to increase or decrease the minutes setting till it displays the desired
2. arrow For an even finer adjustment the arrow may be placed between any two numbers 17 A Next hold the tip in one hand and pull on the sliding barrel with the other hand The device will be cocked when a click is felt as the ends are pulled apart Release the sleeve It will automatically move back to its original position near the trigger Button Warm your fingers slightly to improve blood flowto your fingertips Hold the loaded device against the side of your finger Next press the trigger and lift up the device Set aside the Auto Lancet Wait for a few seconds till a droplet of blood is formed Do not press hard or milk your fingers to draw the blood You may press your finger gently Li For safety reasons and to prevent cross contamination discard the used lancet into an appropriate scrap or bio hazard container Never allow another person to use a lancet that has already been used 18 11 Checking your Blood Glucose 1 Remove atest strip from its pack 2 Insert the strip into the GM3 Blood Glucose Meter with the two electrodes pointing towards themonitor and facing upwards 3 Press the POWER Button to turn on the monitor It displays initial message followed by CODE currently set on the monitor A For best results the test strip must be used immediately afterremoving from its pack A Ensure that the calibration code displayed matches the calibration code of strips being used If not change
3. 11 02 80 mg dL 3 Continue to press NEXT or POWER Button Tea continuously to scroll through all memory 24 13 Important Do s amp Don t s 1 Use strips before the expiry date marked on the test strip pack 2 Storethe stripsin a cool and dry place below 30 C 3 DO NOT use strips which have incorrectly stored or have been left open for a long time Remove the strip at the time of testing only 4 Ensure that Code on the meter matches the code on the strips pack 5 Ensure that your hands are clean at the time of obtaining a sample 6 Ifusing spirit or alcohol swab before pricking your finger wait for 3 4 minutes for the finger to dry completely before pricking 7 Use only freshly obtained capillary or venous blood for testing DO NOT use stored blood fortesting 8 Avery small drop of blood is reguired for the test Do not flood the test strip with excessive blood Gently touch the drop of blood to the testarea and the blood will fill up the test area by itself 9 Care should be taken to avoid extremes in temperature variations humidity shock dust and exposure to direct sunlight A Your GM3 Monitor is an invaluable tool to monitor your Diabetes Ensure thatyou handle the meterwithcare 25 14 Trouble Shooting 1 Displays ERROR Remove and reinserttest strip Ifthe problem persists repeat test with a new test strip Ensure sample is applied after apply sample symbol is displayed 2 In ca
4. GM3 can store the last 75 test results in its memory To maintain a permanentlog record the glucose values in your health log DAY MONTH YEAR ona La ing Enda HI mg dl A DAY MONTH YEAR ne mg dL 11 Remove the test strip from the meter and discard it safely Unscrew the Auto Lancet tip Carefully stick the protective disk cover back on the used lancet Grip the holder firmly and pull the lancet safely out of the holder Dispose the lancet safely 21 A If ERROR is displayed during testing refer to the Trouble Shooting section for more information A If the readings obtained on your GM3 blood glucose meter does not confirm with your symptoms or the way you feel review proper testing procedure and repeat test with a new strip new pack of strips If the reading that appears still does not confirm with your symptoms seek your doctors advice When the Blood Glucose test is over 1 Ifyou wish to do another test insert a new strip into the monitor and press the NEXT Button The monitor restarts for you to do anothertest 2 The monitor automatically switches OFF after 1 minute after the test is over or after 10 seconds after the used strip is removed 3 To manually switch OFF your monitor press the POWER Button for 2 seconds A The results obtained are saved in memory and are not lost when the monitor is switched off automatically 22 12 Memory Function GM3 Glucose Meter has a memory ca
5. minutes value Wait for 2 seconds to save the minutes setting and to go to the set day mode The day now starts to blink Press the NEXT or POWER Button to increase or decrease the day setting till it displays the desired date value Wait for 2 seconds to save the day setting and to go to the set month mode The month now starts to blink Press the NEXT or POWER Button to increase or decrease the month setting till it displays the desired month value 11 HOUR MIN 22 46 HOUR MIN 22 46 HOUR MIN 22 46 DAY MONTH YEAR 01 01 01 DAY MONTH YEAR 23 01 01 DAY MONTH YEAR 23 09 01 10 Wait for 2 seconds to save the month setting and to go to the set year mode The year now a starts to blink DAY MONTH YEAR 22 46 23 09 08 11 Press the NEXT or POWER Button to increase or decrease the year setting till it displays the desired year value 12 Wait for 2 seconds to save the year setting and return to using your GM3 Glucose Meter A The displayed Date amp Time is saved and your monitor automatically exits the time setting mode 12 7 Coding your Meter Before you use your GM3 Glucose Meter for the first time and thereafter every time you open a new box of test strips you need to set the GM3 Glucose Meter to match the strip being used This is called Coding ofthe monitor A Incorrectly coded monitors will give false results Coding needs to be done every time you use a different or new
6. pack of strips 1 Code of the GM3 Gm3 Blood Glucose Test strips can be found printed on the vial of strips as shown in the figure For example let us code your monitor to use a packet of strips with Code C11 2 When your press the POWER Button the meter switches ON and displays initial screen in which moua mn DAY MONTH YEAR 10 01 11 82 89 all the segment of the display are shown Then DI co the code number currently set on your GM3 monitor is displayed identified by the CODE indicator on the display 13 3 A Press both the buttons on your meter simultaneously to change the calibration code r 3 The CODE symbol begins to blink E CODE Now press the NEXT or POWER Button to increase or decrease the code setting till it displays the desired code For example C11 to mE a uo aa use a pack of strips with code C11 ae Widest C I i CODE DO NOT press any key for 2 seconds and the code will be saved and the monitor will return to the operational mode If no Button is pressed for 2 seconds the displayed code is saved and your monitor automatically exits the code setting mode 14 8 Supplies for Blood Glucose Testing Gather all your testing supplies before you begin to test You will need the following items 1 GM3 Blood Glucose Monitor 2 GM3 ACE Blood Glucose Test Strips 3 Comfylets Lancets Super Fine Lancets to ensure easy flow of blood with maximum comfort 4 Safe
7. replace both batteries at the same time and both should be of the same brand Use good quality alkaline batteries only Inferior Cheap batteries may damage your monitor If the batteries show signs of leakage please replace immediately Leaking batteries will damage your monitor 7 The life of the batteries depends on the extent of use and the brand of battery used 1 Press lightly on the battery compartment cover lock and lift the cover off the back of monitor 2 Insert two 1 5 Volts AAA alkaline batteries type LRO3 one at a time into the battery compartment 3 While inserting the batteries into the compartment ensure that the positive amp negative terminals are aligned correctly as marked in the compartment 4 Place the battery cover back on the GM3 and snap it in place to lock the battery compartment lid 5 Be sure that the positive side of the each battery is facing in the correct direction If batteries are incorrectly installed the monitor may be permanently damaged 5 Switching ON 1 When you press the POWER Button the monitor switches ON and displays the initial screen in which all the segment s of the display are shown It next displays the current Code set on your GM3 Blood Glucose Meter This is indicated by the tag CODE displayed on the monitor It then displays the blood drop symbol indicating it is ready to accept a sample for a test HOUR min AVG DAY MONTH YEAR
8. GME Blood Glucose Monitoring System Revised 2014 Operating Instructions NOTICE A The essential features of the type described and illustrations herein remaining unaltered Pulsatom Healthcare Pvt Ltd reserves the right to carry out modifications at any moment on the monitor strips or any other accessories as may be convenient and necessary without being obliged to bring this booklet up to date The descriptions and illustrations in this booklet are not to be taken binding on the manufacturers A Any text that appears in bold italics and with these symbols is important information that should be read carefully and CONTENTS 1 Introduction 2 GM3 Blood Glucose Meter 3 GM3 Blood Glucose Test Strips 4 Batteries 5 Switching ON 6 Date and Time Settings 7 Coding your Meter before use 8 Supplies for Blood Glucose Testing 9 Preparingthe Puncture Site 10 Obtaining a Blood sample for Glucose Testing 11 Checking your Blood Glucose 12 Memory Function 13 Important Do s amp Don t s 14 Trouble Shooting 15 Specifications 16 Warranty Information 17 Service Instructions on oO A 13 15 16 17 19 23 25 26 27 28 31 1 Introduction GM3 Blood Glucose Meter is the latest Glucometer coming from Pulsatom Healthcare Pvt Ltd manufacturers of India s first indigenous Blood Glucose Meter Please read this users manual thoroughly before you begin using the GM3 Blood Glucose Meter as it provid
9. care Lancing Device To help you prick your fingerfor a blood sample with ease 5 Sterile Cotton Swab A Items packed with your Kit is marked on your GM3 Blood Glucose Meter Carton Box Package contents vary according to the package purchased 15 9 1 2 A Preparing the Puncture Site Wash your hands with soap and water Wipe dry If you use an alcohol wipe or medicated spirit to clean the puncture site ensure that you let your finger dry completely before testing AI Wait for at least for 3 to 4 minutes before SA testing a NOW Wipe the test area with sterile dry cotton Take your Auto Lancet Device and Comfylets lancet and follow instructions given in Section 10 to draw a sample of blood If you are using any other lancet device or other method of obtaining blood follow manufacturers instruction for the same Ensure puncture site is cleaned as indicated above Failure to clean puncture site as indicated above may result in wrong results 16 10 Obtaining a Blood sample for Glucose Testing 1 Unscrew the Auto Lancettip 2 Insert a new lancet firmly into the carrier and pinch the lancet to twist off the cover e Replace the threaded tip 4 The comfort tip offers five different levels of skin penetration Select best depth as follows 1 2 for soft or thin skin 3 for average skin thickness 4 5 for thick or calloused skin A To select a desired depth move the number to match the
10. ed as the period as long as the product is in production and 1 year afterthe production ofthe product has ceased If during this period the monitor does not work properly because ofa defect in material or workmanship Pulsatom Healthcare Pvt Ltd will repair the monitor or replace the monitor with another monitor or an equivalent product free of charge The Warranty of the repaired or replacement monitor will expire on the date of expiration of the original Warranty or ninety 90 days after shipment of a replacement system whichever period is longer This Warranty applies only to the original purchaser of the monitor who has registered with Pulsatom Healthcare Pvt Ltd as a user of this monitor by sending the customers registration card duly filled in Pulsatom Healthcare Pvt Ltd undertakes to repair or replace the instrument at its discretion subject to the following conditions 28 1 The purchase Memo Bill Invoice must be preserved and produced along with the defective unit 2 In an event of any complaint the unit must be returned to the nearestservice center or to us freight paid for examination 3 This Warranty is not valid for damages resulting from accident mishandling negligence improper repair tampering loss of components or accessories or if the instrument is not used according to the manual 4 The purchaser s exclusive remedy with respect to the GM3 Blood Glucose meter shall be repair or replacement at the op
11. es all the information you need to use it correctly It consists of GM3 Blood Glucose Meter GM3 Test Strips Lancing Device Lancets Carry case User Manual This is a comprehensive list of items that form a part of the GM3 Blood Glucose Meter Kit Actual contents of the package purchased by you may vary according to the package 2 Know your GM3 Blood Glucose Meter Display Power Button Increase Button Strip Slot Front View of your GM3 Battery Compartmet Serial Number Back View of your GM3 5 Date Units mg dl Memory Symbol HOUR MIN DAY MONTH YEAR D 01 DEN 1 02 09 Pel Code mg dL Code Symbol Display CODE Low Battery ERR Error Symbol Symbo Apply Sample Symbol Display Panel GM3 Blood Glucose Meter 3 GM3 Blood Glucose Test Strips LEN E68 IMTS y O Strips for ACE and GM3 Blood Glucose meter are compatible for each other 1 Contact Area Indicated by the arrow mark Insert this end of the test strip into the GM3 Blood Glucose Meter with the arrow facing upwards and towards the monitor 2 Test Area Blood Sample is applied to this area of this test strip for testing 4 Installing Batteries in your GM3 Your monitor requires two external 1 5 Volts AAA alkaline batteries type LR 03 that need to be installed in the monitor before the monitor can be used O When replacing external batteries always
12. nline http www pulsatom com E mail info pulsatom com Tel 91 80 2357 9166 91 990 003 1600 Mail to Pulsatom Healthcare Pvt Ltd 86D2 A Industrial Suburb Yeshwanthpur Bengaluru 560 022 Karnataka INDIA 31 lala Healthcare Pvt Ltd 86D2 A Industrial Suburb Yeshwanthpur Bengaluru 560 022 INDIA Tel 91 80 23579166 Fax 91 80 23579170 Email info pulsatom com Web www pulsatom com
13. pacity to automatically store your last 75 Blood Glucose test results along with smart tags You can review these readings in order from the most recent reading to the oldest reading or vice versa The Smart tag Date and Time of the test result is also displayed enabling you to know when those readings were taken GM3 also displays the average of the all Random values saved followed by the Before Meal Average and After Meal average of the values saved in its memory While calculating the average it ignores all results under 20 mg dl and above 600mg dl A Ifthe memory is full with 75 glucose results and a new result is added then the oldest result is automatically deleted A Ifthere is no Button input for 4 seconds the program automatically exits the memory mode and returns to test mode 23 Reviewing Average of Stored Values AVG 1 When in the test mode press the POWER Button once till AVG symbol is displayed to review the average memory values 2 The monitor shows your average readings with Ar Smart tags CUE ee 3 Press the NEXT or POWER Button to review other averages sequentially AFTER MEAL Reviewing Stored Values 1 In the test mode press the POWER Button till the MEM symbol is displayed on the monitor to begin using the memory function 2 The meter shows your latest reading with Smart tag Press the NEXT or POWER Button to HOUR MIN DAY MONTH YEAR review stored values sequentially 58 5 BEN
14. se of Incorrect Blood Glucose Values Check Was the correct code set on the monitor Was the strip used before its expiry date Was the strip stored under 30 C in a cool and dry place Was the puncture site cleaned amp dried well before the test Was alcohol or spirit used to clean the puncture site If yes did you wait for 3 4 minutes for it to dry Did you apply too less or excess quantity of blood sample A Forany other assistance please contact customer care 26 15 Product Application Size Weight Specification GM3 Blood Glucose Monitor To determine blood glucose in whole blood by amperometricmethod 120 mm x 70mm x 24 mm 125 grams including batteries Operating Conditions Temperature 100C to 400C Power Supply Battery Life Display Measuringrange Measuringtime Test strip Storage Memory Capacity Humidity lt 98 Two 1 5V AAA batteries Type LR 03 Approx 2 000tests 1 test day Large Display with 12 mm charactersfor results 20 mg dlto 600 mg dl 20 seconds Between 100C and 300C Last Test Value with Time and Date or 75 Blood Glucose Test Results with Time Date and Smart Tag for Random or Before or After Meal It also displays average for each tag 27 16 Warranty Information This monitor is covered by a PRODUCT LIFE TIME WARRANTY against any defect due to defective materials or workmanship for a period ofthree years from the date of purchase LIFE TIME is defin
15. the calibration code of your monitor as explained earlier 4 The monitor now displays the Blood Drop BB HEEE E Symbol which indicates that it is ready to acceptthe blood sample 19 5 If instead of the test symbol ERROR is displayed as shown here refer to the trouble shooting section for more information When the meter displays the blood drop symbol press the NEXT key to select the smart tag BEFORE MEAL or AFTER MEAL or none to tag your reading with your testing status With the strip inserted in the GM3 Glucose Meter apply the drop of blood to the test area You may lift the monitor in your hand and move it to come in contact with the blood drop The blood droplet automatically fills up the test area Remove your finger as soon as you see blood flowing into the test area A very small droplet of blood approx 2ul is required for the test 20 HOUR MIN 10 01 DAY MONTH YEAR 11 02 0939 7 As soon as the sample is applied the machine begins a countdown of 20 seconds and displays the time remaining for test to complete as 20 19 18 12 9 At the end of 20 seconds your blood glucose value is displayed in mg dl units on the GM3 monitor This is also saved in the memory along with the time and date of the test 10 If the blood glucose result is higher than 600mg dl Hi is displayed and if the blood glucose result is lesser than 20mg dl Lo is displayed A
16. tion of Pulsatom Healthcare Pvt Ltd 5 In no event shall Pulsatom Healthcare Pvt Ltd be liable to the purchaser or any other person for any direct incidental consequential special or punitive damages arising from or in any way connected with the purchaser or operation of the monitor or its 29 6 7 8 9 lt parts even if Pulsatom Healthcare Pvt Ltd has been advised of the possibility of such damages No Warranty of merchantability or fitness for a particular purpose if any is implied from the sale of the monitor shall extend for longer duration than product lifetime as defined earlier While the company will make every effort to repair or replace the unit under this Warranty as soon as possible it is expressly made clear that the company shall not be liable to do so within any specified time or period All warranties covering GM3 blood glucose monitoring systems are valid only when GM3 is used with GM3 GM3 Blood Glucose Strips All warranties covering the GM3 meter are voided by use of any other strips other than GM3 GM3 Strips Claims ifany to this Warranty shall be made only before the courts having jurisdiction in Bengaluru 30 17 Service Instructions Customers experiencing difficulties should review the troubleshooting guide of their user manual For your service requirements and any other enquires please contact us You can contact us by any ofthe following means O