Home
Molift Air
Contents
1. Wall Battery Mounted indicator Battery Charger Charger Light Hand Control Mains cable disconnecting device Battery Charger The lifter is charged through the Hand control The battery charger shall be mounted to a vertical wall surface close to a power outlet minimum 120 cm above the floor The charger has to be used outside the patient environment The Charger must be placed or installed in a way that makes it easy to disconnect mains cable plug Take care not to damage the cable The charger can be connected to the power outlet at all times and the hand control should be placed in the charger console when the lifter is not in use Charger for handcontrol Nomad AIR Art no 1240100 Charger Light Description of charger light Charger light Yellow Ready for use no bat tery connected Yellow Initialization Red Fast Charge Green Yellow Top off charge Trickle charge Red Green Error The lifter is charged through a built in charger in lift motor for continuous charging through the entire rail system length Molift Air IRC has a built in charger connected to the battery It must be installed with IRC Trolley in a rail system with a conductive tape The IRC trol ley is connected to the battery charger inside the Molift Air IRC The IRC trolley is always in contact with the the conductive tape inside the rail The IRC power supply is connected to the conductive tape through IRC connection at
2. adjusted to the correct length height by cutting tube and cord Use this checklist to verify that the hoist is properly installed and can operate correctly an d safely before use O End stops on rail are installed after mounting of trolley 1 Make sure that the hoist is properly fastened to the rail and that the lift does not have any loose parts Perform one lift with normal load 60 80 kg Make sure battery is fully charged OO How to use Molift Air General Safety Precautions Safe Working Load Only use accessories and slings that are adjusted If maximum load SWL differs between hoist suspension and body to fit the user type of disability size weight and type of transfer support unit then the lowest maxi mum load shall always be used Working pause ratio Duty Cycle Molift Air should not be run constantly for more than 2 minutes with maximum load and rest for minimum 18 minutes Duty Cycle 10 Intermittence according to standard ISO EN 10535 Molift lifters shall only be used to lift persons Never use the lifter to lift or move objects of any kind Inspection to be performed daily or before use O Make sure lifting strap does not have visible damage or frays O Inspect lifter and rail system has no visible O Test run lifters operation functions and make damage defects or deformations sure lifter does not make any abnormal 1 Make sure suspension connection and all sounds detachable parts a
3. no 1840000 Can be combined with all Molift Air compatible suspensions Scale is approved Class Ill The scale can also calculate BMI Body Mass Index with a single touch Read the manual that comes with the weight scale prior to use Molift supplies a wide selection of slings for dif ferent types of transfers The Molift RgoSling sling series is developed to be combined with a 4 point suspension but the sling also works with a 2 point suspension See the combination list in the slings user manual for the correct sling and suspension combination The Molift RgoSling sling series is available in sizes XXS XXL in polyester and polyester mesh Rgosling Mediumback Padded XS XXL Rgosling Highback Padded XS XXL Rgosling Mediumback Net XXS XXL Rgosling Highback Net XXS XXL Rgosling Toilet Lowback XXS XXL Rgosling Toilet Highback XXS XXL Rgosling Ampu Mediumback XXS XXL Rgosling Ampu Highback XXS XXL Rgosling Ambulating Vest Rgosling Comfort Highback S L RgoSling Fabric Stretcher Accessories Rgosling Extension Loops Art no 1721600 Rgosling Ambulating Vest Groin strap XXS XL Molift Air www etac com Maintenance Cleaning and disinfection Clean on a regular basis Clean surfaces with a damp cloth using an appropriate pH neutral detergent Do not use solvents or strong liquids this may damage surfaces on the lifter For disinfection when needed use isopropyl alcohol Avoid abrasive cleaning pro
4. sent to Etac molift etac com In the event of danger to safety the hoist shall immediately be taken out of service and marked clearly with out of order and shall not be used until the hoist is repaired When periodic inspection is completed the inspector shall mark the hoist with a sticker on the control label showing the date when periodic inspection is performed The control label can be found on the inside of the Side cover and this will then indicate when next service should be performed Mark label with month and year of inspection Molift Air www etac com Troubleshooting Symptom Possible Cause Action The lifter does not respond to Hand control action Emergency Stop is activated Deactivate by push ing button back in The lifters electronics is overheated Wait for it to cool down Hand control is not plugged in properly Open cover and plug in Hand control Hand control or plug or cord can be broken an should be replaced The lifter does not respond to Hand control up Lifting strap angle sensor is activated Adjust button lifting strap or move hoist to reduce lifting strap angle 14 Molift Air www etac com Etac AS Etac Supply Gjavik Hadelandsveien 2 2816 Gjavik Norway Tel 47 4000 1004 molift etac com www etac com a part of Etac
5. the end of the rail and supplies power to the battery charger inside Molift Air IRC Conductive IRC connection RH62 connected to IRC power supply IRC fits both Single rail and Traverse systems Conductive Traverse trolley IRC Connected to traverse trolley IRC connection RH62 connected to IRC power supply Conductive Tape IRC Power Supply Art no 2510119 Charger 2515 DC DV Part of Molift Air IRC Art no 2510147 Transfer Plan the lifting operation in advance to ensure that it is as safe and smooth as possible Remember to work ergonomically Assess the risks and take notes The assistant is responsible for the safety of the user Molift Air can be used with both 2 point and 4 point suspension for different types of transfers sitting recumbent or ambulating The suspensions hook design prevents sling to be inadvertently detached when mounted correct It is important that the sling has been tested with the individual user and for the intended lifting situation a Read User Manual for the sling prior to use D tt d d r bad 1 Check that the sling is correctly fitted around ER the user and that the strap loops are correctly worn slings fitted to the suspension hooks 2 Stretch the sling straps without lifting the user Ensure that all four loops of the sling are Slings made by other manufacturers securely fastened to avoid the user slipping or We recommend only using Molif
6. weeks The lifter has two indicator lights AO Ca O Wrench symbol Service light Battery symbol Battery light The electrical system has a power save function which will turn off the electrical system after ten minutes without activity All lights will turn off The system is activated when pushing one of the operating buttons Battery Light The battery indicator light will illuminate and make a sound when battery level is low and the lifter requires charging When this occurs the lifter will have sufficient power available for one full lift ing cycle with max load When battery is critically low it is only possible to lower the lifting belt Service Light The lifters electronics record the loads lifted and number of lifts After a certain period of operation a signal is given to indicate that service is required Service light No light Power saving Stand by Ready for use Yellow Order service Red Perform service Red sound Perform service immediately Service involves replacing the lifting belt and inspection replacement of worn parts This must be carried out by authorized personnel Service is needed when the service light is red calculated 5 000 lifts in weight class 2 or service is required after 5 years if the service counter not has displayed red yet Molift Air can be charged in two ways Through hand control with a wall mounted battery charger or through the rail with In Rail Charging
7. Molift Air EN User manual BM16101 Rev E 2015 01 15 CE English Manual Molift Air 205 300 pene nen eee ene Te 2 Hoist COMPONENTS ds unies 2 About Molift Air mens den ma stut 2 G n ral SE re 3 Declaration of conformity 3 Conditions for USE ts semester 3 Warranty ier acetates Sav RER EEE SEES 3 Product identification sans 3 Technical Data eira fs EE e 4 Installation ca Set Sith adhe 5 Trolley installation essere 5 Lifter installation iesss ti snsasss 5 Checklist after installation 1 1 1 1 5 How to use Molift AP ies 6 Before use Daily check 6 Hand control save re 6 Mounting Suspension 7 Emergency Stop Emergency lowering 7 EOS aaret 8 Transport and Operating Conditions 8 Bele norr eee 8 Battery and Service indicator ooooo0 8 Service SCOP Greens added 8 Ch rie 9 Charging through Hand Control 9 IRC In Rail Charging 9 ASTM At dt 10 Using SINOS se 10 Lifting and lowering 10 ACCESSONES sd Sn telnet na 11 SUSPENSIONS En ananas 11 SINGS ae ec AT 11 Slate EE Severe eer ere ete tetas EE ES 11 Main tenan es uses insu ss 12 Cleaning and disinfeciton 2 0 12 Reconditioning seins ne 12 REC ClO sas a ar me a 12 Spare Pafta sains 12 Periodic INSpection aent 12 Pericodic Inspection Scope 0 12 TOUDI S RGO Nr h
8. c Expected Lifetime The hoist has an expected lifetime of 30 000 cycles or 10 years Dimensions Lx W xH Length Width Height without trolley 36 x 19 x 19 5 cm 14 2 x 7 5 x 7 7 inches A Hoisting Range 300 cm 118 Inches B Minimum distance from ceiling to Suspension coupling point 32 7 cm 12 7 8 Inches C29 5mm 72 5mm H195mm E10mm Min 300mm Installation Assembly and installation The hoist is marked with Safe Working Load SWL this should not exceed rail systems max load capacity The rail system can only be installed by certified personnel in accordance with applicable installa tion instructions Do not start using hoist before completing control according to checklist after installation Mounting trolley in rail system Remove End stop and insert Trolley into rail Mount end stop and make sure it is securely fastened Make sure end stops are mounted in all ends of the Rail system before using the hoist Mounting of Lifter on trolley No tools required The lifter has two connection points Connect one point first Push button all the way in and place lifter in connection point on trolley Release button and make sure it is clearly show ing green before connecting the next point Make sure both connection point buttons completely return and clearly show green after installation Emergency lowering cord and tube can be
9. ducts Check emergency stop and emergency lowering after cleaning The lift should not be exposed to run ning water Make sure not to damage or remove labels when cleaning A Reconditioning Follow cleaning and installation procedure com plete periodic inspection and use checklist after installation to recondition the lifter Recycling Refer to Recycling instructions for how to properly dispose of product This can be found on www etac com Spare Parts A list of spare parts is available on request Periodic Inspection Periodic Inspection Scope Periodic inspection is a visual examination par ticularly of the hoist s load bearing structure and lifting mechanism with attachments brakes con trols safety devices and person support devices according to Periodic Inspection Report for Molift Air This can be found on www etac com Periodic Inspection shall be performed at least once a year or more frequently if required by local requirements The inspection must be performed by service personnel authorized by Etac Contact Etac for training and authorization or recommen dation of an approved service partner When performing a periodic inspection the inspector shall fill out the inspection report for Molift Air The reports should be retained by the person s responsible for servicing the hoist If the inspection reveals defects and damages the owner shall be notified and a copy of the report should be
10. een tested and approved by a third party according to standards IEC 60601 1 IEC 60601 1 2 and NS EN ISO 10535 2006 Conditions for Use Lift and transfer of a person will always pose a certain risk and only trained personnel are allowed to use the equipment and accessories covered by this user manual The rail system must be installed by certified personnel in accordance with applicable installation instructions Modifications and use of components made by other manufacturers We recommend only using Molift components and spare parts Declaration of conformity is not valid and Etac is not responsible for warranty if any modifications are made to the product Etac shall not be liable for faults or accidents that can occur when using components made by other manufacturers A Only certified personnel are allowed to open hoist or accessories to perform service or repair Risk of injury from rotating parts and electric shock The lifter is not intended to be operated by the person being lifted If a hoist is to be used by a disabled person living on their own then some form of communication device shall be installed in the area of use of the hoist so that in the event of an emergency the disabled person is able to summon assistance This may for example be the fitting of an alarm system or the supply of a conveniently placed telephone etc Warranty 2 year warranty against defects in workmanship and materials of our products For Terms a
11. ill come out and hoist will stop Lifter must be sent to service after performing manual emergency lowering Push with finger or use tube on cord to push but ton back in to reset emergency stop Electronics Transport and Storage For long time storage it is recommended that the emergency stop button is activated pulled out The lifter can be stored and transported under temperatures between 25 70 C Operating The lifter is designed for use at standard room temperatures 5 to 40 C Air Pressure 70 106 kPa Relative Humidity 15 93 Following storage or transport at other tempera tures leave the lifter in a room with a suitable temperature until it reaches a safe operating temperature Medical electrical equipment requires special precautions regard ing electromagnetic compatibility EMC Portable or mobile radio communication equipment may affect the medical electrical equip ment and should be kept minimum 25 cm 10 inches from the lifters electronics Molift Air has a 26 4 V NiMH battery pack The battery has a life expectancy of approx 500 charge cycles Used batteries must be disposed of as special waste according to local rules and regulations Do not dispose in general waste Batteries must be fully charged and depleted completely a couple of times before it achieves full capac ity Also follow this procedure if the lifter has not been in use for a long period 3 4
12. nd conditions see www etac com Product label The Product labels barcode contain article number serial number and production date ODEL ace me Refer to user manual RO Indoor Use only pattie 200090 me 255 E Max user weight Duty cycle 10 g i LR molift Do not dispose in general waste erial no www molift com Emergency stop Warning labels and Symbols Emergency lowering Symbols used on the product explained in more detail E Class I double insulated CE marked Service Light Battery Light Hex tool symbol for Manual Emergency lowering O lox Technical Data Safe Working Load SWL Molift AIR 205 205 kg 450 Ibs Molift AIR 300 300kg 660 Ibs Weight of unit 8 46 kg 18 65 lbs Battery Art 2510023 26 4 V NiMH 2 2 Ah Battery charger Charger for handcontrol Mascot type 2215 Art no 1240100 100 240 V AC 50 60 Hz max 0 9A Output 0 9A IRC Power Supply Mascot type 9920 Art no 2510119 Input 100 240 VAC 50 60 Hz max 0 9A Output 24VDC 1 6A 40W IRC Charger Mascot type 2515 Art no 2510147 In 20 30VDC 1 3A Output 0 5A Lifting speed 60mm second 2 36 inches second with 75kg 165 35 lbs load Propulsion motor speed 0 2m second 7 87 inches second Protection class Hand control IP24 Lift motor IP24 Maximum A weighted sound power level Lyn 63 dB Operating forces button Buttons on handset 3 4 N Material Aluminum Plasti
13. onte out run 13 This User Manual contains important safety instructions and information regarding the use of the lifter and accessories In this manual the user is the person being lifted The assistant is the person operating the lifter Warning This symbol indicates important information related to safety Follow these instructions carefully Read User Manual before usel It is important to fully understand the content of the user manual before attempting to use the equipment Visit www etac com for download of documenta tion to ensure you have the latest version Molift Air 205 300 Hoist components Trolley Hoist Side covers Emergency stop Emergency lowering Lifting belt Quick release pin Hand control Suspension BRWN gt ON O UT Molift AIR is a strong and smooth ceiling lift that enables patients and residences to be transferred in a comfortable and safe way Suitable for sitting and lateral transfers as well as standing and gait training situations together with the comprehen sive accessory program from Molift Molift AIR is a very light motor which in combina tion with the quick release coupling system gives an easy handling when mounting and servicing General Declaration of conformity The Molift Air and related accessories described in this operator manual are CE marked in accordance with EU Council Directive 93 42 EEC concerning medical devices class 1 and has b
14. re properly connected and secured If there are any faults or defects the lift needs to 1 Test emergency stop button and emergency be taken out of operation and marked out of lowering order 1 Make sure battery indicator or service light is not illuminated The Hand control has 2 buttons for lifting and lowering or 4 buttons if the hoist is equipped with propulsion The Hand control has an indica tor light that will illuminate when battery level is low and the lifter requires charging Do not pull Hand control to move the hoist along the rail mot Electrical Emergency Lowering Pull and hold to start lowering 1 Align suspension in connection point 2 Push button on locking pin and insert all the Manual Emergency lowering way through Allen key or bit for power drill can be found inside 3 Make sure locking pin is properly fastened the lifter Remove cover to locate 4 Push button and push bolt from opposite side while holding in button to remove locking pin Emergency Stop Activation and reset of emergency stop Push down and press cover outwards with a flat screwdriver in the two slots to open cover Locate sticker over the hole for manual emergency lowering Punch hole or remove sticker and insert tool Rotate clockwise to lower lifting belt Using a power drill is recommended Pull to activate emergency stop The button w
15. t slings Etac shall falling not be liable for faults or accidents that can occur 3 Lift user and perform transfer when using slings made by other manufacturers Angle sensor The hoist has a directional safety feature that prevents lifting if angle on lifting strap is to steep When moving the user stand to the side of the person you are lifting Make sure that arms and legs do not obstruct the seat bed etc Keep eye contact with the user to help them feel safe 1 If the hoist doesn t respond to Hand control up button it might be because lifting strap angle is to steep or lifting belt is twisted and sensor is activated The 4 point suspension must always be positioned across the user to prevent the user from uninten tionally sliding out of the sling Accessories Recommended optional equipment and acces sories for Molift Air Molift Air can be used with 2 or 4 point suspension Aluminum suspension 2 point Small aluminum 340 mm Art no 1830003 2 point Medium aluminum 440 mm Art no 1830002 2 point Large aluminum 540 mm Art no 1830001 4 point Small aluminum 340 mm Art no 1830012 4 point Medium aluminum 440 mm Art no 1830011 4 point Large aluminum 540 mm Art no 1830010 4 point X Large aluminum 640 mm Art no 1530006 i 8 point suspension for fabric stretcher steel Art no 2140003 Molift Scale Set without Suspension Art