Home

ART-P6 Piezoelectric Ultrasonic Scaler

image

Contents

1. Do not reuse scaling tips piezo tips that are damaged bent or reshaped Discard immediately Use the long axis of the scaler tip piezo tip to wipe accretions from the tooth Do not gouge the tooth with the tip 3 C Precautions for Hand piece Do Not Twist And Pull Cable Do Not Bend Do Not Twist Do Not Bundle Pull To Disconnect ee Section IV Infection Control 4 A General Infection Control Recommendations As with all dental procedures use standard protective equipment such as face masks eyewear face shields gloves and protective gowns To ensure safety to the operator and patient carefully follow the Infection Control Procedures detailed in Section IV 4 B Water Supply Recommendations All dental water supply systems should conform to applicable Centers for Disease Control and Prevention CDC and American Dental Association ADA standards These standards should apply to flushing chemical flushing and general infection control procedures 4 C Cleaning and Sterilization All sterilization procedures must be followed in accordance with the EN ISO17664 standards Autoclave sterilization conditions In order to avoid bacterial or viral infections always clean and sterilize the following components after each treatment 1 Hand piece 2 Tips 3 Wrenches The components above are composed of materials that can withstand a maximum temperature of 135 C or 275 F for a maximum of 3
2. s head gently so that the oral cavity can be easily reached without causing discomfort to the patient see figure below 4 During treatment try to keep the angle between the surface of the tooth and scaling insert at 15 degrees If the patient experiences any discomfort during treatment follow the suggestions below e Increase the hand piece movement speed on the surface when treating sensitive areas e Treat less sensitive area first and then return to sensitive areas e Ifthe problem persists reduce the power output intensity of the hand piece sjan 7 C Applications of the TIP A Tangential application BS1 TIP Do not apply the BS1 tip directly to the tooth Doing so C lt ss may damage the enamel Control the hand piece with a x slow and steady motion ES B Frontal application BS2 TIP Apply the BS2 tip directly against the tartar but not CEZARY directly against the tooth Use a slight amount of pressure N C Tangential application BS3 TIP Do not apply pressure when applying the BS3 tip to the enamel Section VIII System Maintenance 8 A Hand piece maintenance 1 After each use the hand piece and tip should be rinsed with clean water for about 20 to 30 seconds 2 Inspect the hand piece cable daily to ensure it is in good condition 3 For sterilization of the hand piece please refer to Section IV 8 B TIP maintenance Worn out tips can adversely affect performa
3. 6 unit e After the above installation requirement is fulfilled thoroughly flush the water prior to connecting to the scaler Please use the manual shut off valve on the dental water system supply line to shut off when the office is unoccupied 5 C Electrical Requirements Refer to Section X Specifications 5 D Unpacking the Unit Carefully unpack your ART P6 unit and verify that all components and accessories are included Figure 1 All components and accessories Item Quantity Item Quantity 1 Main Unit l 2 Aluminum Sterilization Box Optional 3 Plastic Wrench 1 4 Metal Torque Limited Wrench Optional 5 Hand piece 1 6 PU Water Tubing 1 7 User Manual amp Literature 1 8 AC Power Core 1 Packet 9 Foot Switch 1 10 ART BS1 Universal Pointed Tip 2 ART BS2 Chisel Head Flat Tip ART BS3 Round Flat Tip 1 NOTE Other piezo tips are available and can be purchased separately through distributors Included items are subject to change depending on location NOTE Upon receipt of your ART P6 check your ART P6 unit for any damages If any damages are found please contact your dealer immediately Enter the unit serial number on your warranty card and mail it within 14 days from the date of purchase 5 E Power Cord Power Connection Always make sure that the power switch is set to the OFF position before performing the following tasks e Plug the detachable AC cord i
4. ART P6 Piezoelectric Ultrasonic Scaler Directions for Use pec BONART CO LTD Jorgen Kruuse A S 4F 11 No 3 Wuquan 1st Rd Havretoften 4 DK 5550 Langeskov Xinzhuang City Taipei County Denmark Taiwan R O C Tel 886 2 22983980 C Fax 886 2 22983981 0434 Email sales bonartmed com COPYRIGHT 2011 BONART CO LTD BONART Co Ltd owns all rights to this publication This publication shall be used solely as areference for operation maintenance and repair of BONART equipment No part of this document may be reproduced or distributed in any form for any other purpose In the event of inadvertent or deliberate publication BONART Co Ltd intends to enforce its rights to this manual under the copyright law Reproduction of the materials presented in this manual without the express written permission of BONART Co Ltd is prohibited The contents of this manual are subject to change without prior notice PROPERTY OF BONART CO LTD ALL RIGHTS RESERVED ART is a registered trademark of BONART Co Ltd amp its affiliates Document No P61001 Rev 1 3 06 11 Printed in Taiwan R O C Operator Safety Please read this manual before operating unit The ART P6 piezoelectric ultrasonic scaler should be operated maintained and repaired by qualified and properly trained personnel NOTE CAUTION AND WARNING STATEMENTS NOTE Provides tips and advice CAUTION Provides correct operating or maintenance procedures WAR
5. NING Alerts user of danger that may cause severe injury when proper procedures are not followed e SYMBOLS C Compliance with MDD 93 42 EEC 0434 Type BF equipment Grounding terminal Attention Please read instructions Preface About the ART P6 Piezoelectric Ultrasonic Scaler The ART P6 Piezoelectric Ultrasonic Scaler is manufactured by Bonart Co Ltd an ISO 13485 2003 company The ART P6 is also CE 0434 certified in compliance with the applicable requirements of the Council Directive 93 42 EEC The reviewing council also classifies the ART P6 in accordance to EN60601 1 and EN60601 1 2 safety requirements How it works The ART P6 Piezoelectric Ultrasonic Scaler generates ultrasonic waves in the hand piece to vibrate the tip which allows the ART P6 to remove calculus and tartar easily These ultrasonic waves are created by 4 ceramic piezoelectric plates subjected to high frequency alternating currents The piezoelectric transducer operates at a frequency of 29khz with minimal noise and heat and efficiently reduces the amount of cooling water required during treatment This electric mechanical feature provides greater efficiency than traditional magnetostrictive systems Technical Support For technical support and repairs in the U S call 1 888 5 BONART Monday through Friday 9 00 AM to 5 00 PM Pacific Time For other areas contact your local Bonart representative Supplies amp Replacement Parts To order s
6. ice center The tip does not thread Use a metal torque wrench to properly secure the tip in place No atomization a tip is deformed or worn Replace the tip Water flow is inadequate Check the water flow setting The tip does not thread properly Use a metal torque wrench to secure the tip in place Molas ound patoi The hand piece is broken Replace the hand piece The cord is damaged Replace the cord The tip does not thread Ultrasound is too properly Use a metal torque wrench to secure the tip in place strong or too weak The tip is deformed or worn out Replace the tip sje 9 B Technical Support and Repairs For technical support and assistance call 1 888 5 BONART 1 888 526 6278 or 909 594 8066 Monday through Friday 9 00 AM to 5 00 PM Pacific Time For other areas contact your local Bonart representative Section X Specifications Power supply 115V 5 50 60Hz 28V A 230V 5 50 60Hz 28VA Working frequency 26KHz 32KHz Water supply 25 60 PSIG 172 414KPa Dimension 19 5cm L x 16 2cm W x 8 3cm H Weight 1 6 Kg include hand piece Hand piece Cable 250 cm Footswitch Cable 250 cm Operation environment Temperature 0 60 Relative Humidity 10 90 non condensing Transport and storage conditions e Temperature 0 60 e Relative Humidity 10 90 non condensing e Atmospheric pressure 860 1060 hPa Classifications Protection against electric shoc
7. k Class I Protection against electric shock degree Type BF Mode of operation Continuous Medical device directive lla _18 Complied Directives and Standards ISO13485 Medical devices Quality Management Systems Requirements for regulatory purposes 93 42 EEC MDD CE directive EN60601 1 Medical Electrical Equipment General Requirement for Safety EN60601 1 2 Medical Electrical Equipment Electromagnetic Compatibility Requirements and Tests Section XI Disposal Please follow county and state regulations for disposal of the ART P6 19 20
8. minutes After completing the sterilization cycle air dry it for about 20 30 minutes before using it WARNING When autoclaving the hand piece tips and wrenches flash sterilize by using an unwrapped sterilization cycle in order to extend the life of the hand piece NOTE The Sterilization Assurance Level SAL of steam autoclave should be 10 According to ISO 13683 Sterilization of health Care products NOTE Prior to cleaning and sterilizing your P6 unit always remember to turn off the device by using the power switch and to disconnect the power plug from the outlet Preparations for sterilizing the tip s 1 Remove the tip from the hand piece 2 Clean and disinfect the surface of the outer casing cords and connectors with a cloth and mild detergent or disinfectant solution with a neutral pH pH7 3 Allow the disinfectant solution to air dry before sterilizing 4 Seal the tip inside a sterilization pouch If you are autoclaving more than 1 tip place each tip in an individual bag You may also use the tip holder that comes with the ART P6 in place of the sterilization pouch 5 Steam autoclave the tips for sterilization WARNING Before starting the sterilization cycle make sure that the inside of the tip is completely dry by blowing air through the internal hole with the syringe This will prevent stains and patches from appearing on the surface of the tip WARNING Do not use any other method of sterilization dry hea
9. nce resulting in insufficient power and vibration Check tips regularly for wear and tear and replace as necessary _15 8 C Main Unit Maintenance The housing of the ART P6 as well as the hand piece cord should be cleaned and disinfected using alcohol or soap and warm water on a daily basis sis Section IX Troubleshooting Although service and repair of the ART P6 system should be performed by Bonart personnel the following are some basic trouble shooting procedures that will help avoid unnecessary service calls Generally check all lines and connections to and from the system a loose plug or connection will often create problems 9 A Troubleshooting Guide PROBLEM POSSIBLE CAUSE SOLUTION f Check if the power outlet is The device does not The power cord is faulty working properly turn on when the switch is positioned to ON Fuses are damaged Contact BONART or an authorized service center for fuse replacement The device is turned on but does not The footswitch has gone out work Replace the footswitch Contact BONART or an authorized service center The water supply is not connected to the unit Check water supply No liquid comes out The filter or may be clogged of the tip during Clean or replace the filter operation The tip socket is clogged Clean the tip socket The internal tubing of the Contact BONART or an hand piece is clogged authorized serv
10. nto the back of the unit 10 e Plug the 3 prong plug into a grounded outlet N Safety Instructions A Grounding Prior to connecting accessories to the unit check that the main unit is plugged into a grounded wall outlet B Main voltage range and fuse Prior to plugging the AC adapter into the power outlet check that the voltage is supported 5 F Foot Control Cable Assmbly Connection Align the pins of the foot control plug with the receptacle on the back of the device and push in until the plug is firmly seated 5 G Water Supply Line Connection Push the blue water tube hose into the stainless steel receptacle until the hose cannot be pushed any further Then tighten the screw Connect the quick connect to the water supply line The ART P6 should be prepackaged with a male quick connect If a female quick connect is necessary please contact a Bonart distributor to purchase the part Inspect all connections for leaks To remove the water line from the ART P6 scaler first turn off the water supply or disconnect the water supply line Then loosen the screw on the water tube from the receptacle of the unit and gently pull the water hose out ait sd 5 H Hand piece connection and tip compatibility The ART P6 hand piece is not only compatible with Bonart s tips but also with tips designed for Satelec Acteon piezoelectric scalers The hand piece is also detachable from the hand piece cable During use the o
11. perating frequency of the tip is measured to be approximately 29khz 5 I Assembling the Scaler Unit 1 Ensure that the power switch is set to off If plugged in the power indicator light should be unlit 2 Plug the power cord of the main unit into a grounded AC power outlet 3 Insert the tip into the hand piece and lock it into place by using the enclosed plastic wrench Be careful not to over turn the tip as it may cause the tip and hand piece to break 4 Ensure that the water tube connection to the water supply source and main unit is secure Figure 2 System Installation Diagram Bay SectionVI General Information and Parts Description Figure 3 Outlook diagram 1 Main Unit The main unit generates power and produces a signal that is passed to the hand piece The hand piece is then powered up and vibrates the installed tip Hand piece Footswitch e Power cord Water In connector An gt WN e Power Panel The power panel allows you to adjust the power level within 1 20 Power is indicated by a lit LED 7 Main power switch with indicator 8 Water Adjust knob 9 Power Adjust knob Section VII Techniques 7 A Performing Ultrasonic Scaling Procedures 1 Use purified or distilled water to prevent infection when patients experience tissue laceration during treatment 2 Keep the power cord tidy to avoid tripping and other accidents 3 Position
12. rative dental procedures Do not use the ART P6 if the patient or operator is wearing a pacemaker Do not immerse the ART P6 in water or liquid If the ART P6 has water damage return the machine to Bonart for servicing Do not modify the ART P6 Modifications will invalidate the warranty on the machine as well as invalidate safety codes and endanger the patient and operator Section III Precautions 3 A Precautions for all Ultrasonic Scaler Units and Systems Ensure sufficient water flow to the scaler tip during use to cool the hand piece and tips Take precaution as all ultrasonic scalers produce aserosols that may transmit contagious diseases Use tips manufactured by Bonart for optimal performance and result Keep the unit away from intense heat Excessive heat may damage the electronic components Switch the water valve off when the ART P6 is not in use Avoid treating patients with pacemakers due to the magnetic filed produced by the ART P6 The magnetic field may cause monitoring devices to malfunction and provide inaccurate readings 3 B Precautions for Ultrasonic Prophylaxis Procedures Scaler tips piezo tips will wear with use Tips with 2mm of wear will lose approximately 50 of their scaling efficiency In general it is recommended that tips be discarded and replaced after 2mm of wear to maintain optimal efficiency and avoid breakage Avoid direct contact of tips with the patient s lips cheeks and tongue
13. t radiation ethylene oxide gas low temperature plasma etc Preparations for Autoclave sterilization of the hand piece 1 Take special care not to break the threading pin of the hand piece The threading pin is where the tips are installed 2 Disinfect the hand piece with a cloth and mild disinfectant containing a neutral pH 3 Dry the electric contacts by blowing air onto them with the syringe 4 Seal the hand piece in a sterilization pouch without any tips If you are short on sterilization pouches you may use the surgical tray in place of the sterilization pouch 5 Follow the instructions of your autoclave machine to sterilize the hand piece in the autoclave machine Preparations for Autoclave sterilization of the wrench 1 Clean the wrench 2 Disinfect the wrench with a cloth and mild disinfectant containing a neutral pH 3 Seal the wrench in a sterilization pouch You may also an autoclavable surgical tray in place of a sterilization pouch 4 Autoclave the wrench Section V Installation 5 A General Information If the installation of your ART P6 system is performed by non Bonart distributor personnel check to see that requirements below are followed 5 B Water Line Requirements Incoming water supply line pressure to the Scaler must be 25 psi 172 kPa minimum to 60 psi 414Kpa maximum If your dental water supply line pressure is above 60 psi install a water regulator on the water supply line to your ART P
14. the footswitch in an easily accessible spot for the user Keep the footswitch cord tidy to avoid tripping and other accidents ites 4 Lock the tip in the hand piece securely using the enclosed wrench CAUTION DO NOT FORCE OR OVERTURN THE TIP INTO PLACE Doing so will damage the tip hand piece and wrench NOTE The scaler will not operate if the tip is either removed from the hand piece or improperly installed 6 Hold the hand piece over a sink or drain with a tip installed Verify that water is reaching the tip 7 Check your Bonart s piezo tip for wear and replace as needed o0 Use the foot control to regulate water flow to the hand piece Hold down the foot control to allow water to flow through Release the foot control to stop 9 Adjust the water flow to the desired level using the water adjustment knob Increase the water flow by turning the water adjust knob clockwise Decrease the water flow by turning the water adjust knob counter clockwise NOTE Increased water flow results in a cooler hand piece temperature and vice versa CAUTION During operation a continuous flow of water is required to keep the hand piece cool 7 B Patient Comfort Considerations 1 Position the patient comfortably on the patient chair Adjust the chair angle and position to access the patient s oral cavity with ease and comfort 2 Remove excess saliva and debris in the patient s mouth with a saliva ejector 3 Rotate the patient
15. upplies or replacement parts in the U S contact your Bonart distributor or dealer A list of Bonart distributors can be found on our website at www bonartmed com Table of Contents Copytishts se deen Sudo Geena healed Operator Safety Preface gg Aaaa Oa comes AE Contents aS acta hee Na nternbiatea g ates Section I Indications for Use r a a R D Section II Contraindications and Warnings Section II Precautions j zaiaagubcaeesmendedantumumauseene Section IV Infection Control Heated iisewy see siaeeateeiiaes Section V Installation g Section VI General Descriptions and Information of Parts Section VII Techniques 9 ar LJ l L L Lee Section VIII System Maintenance _ Section IX Troubleshooting Section X Specifications an eeceecceeceeceeeeeeeeneeeeeneeees Section XI Disposal ad cecccecc cece uu D u e as Section I Indications for Use Ultrasonic procedures Removal of calculus and plaque during dental prophylaxis General supra and sub gingival scaling applications Periodontal debridement for all types of periodontal diseases Endodontic procedures Section II Contraindications and Warnings Do not use the ART P6 for resto