Contents
1. Bump Test IEC 68 2 29 Vibration Test BS EN 61010 clause 8 3 The E3640 has been designed with safety in mind However no design can Impact Test BS EN 61010 clause 8 2 completely protect against incorrect use Electrical circuits can be dangerous and even lethal through lack of caution or poor safety practices Drop Test BS EN 61010 clause 8 4 1 1 Read the Instruction Manual carefully and completely so that you fully understand the instructions before using the tester 5 3 ENVIRONMENTAL Follow the instructions in the Manual for every test Take all the recommended precautions Never exceed the limits of the tester Operating Temperature 15 C to 55 C 1 2 The device under test DUT must be de energised and isolated Storage Temperature 25 C to 65 C before connections are made for any tests Operating Humidity 80 RH at 40 C 1 3 Check that the Power selector on the tester is set correctly Cold Temperature TEC 68 2 1 Dry Heat IEC 68 2 2 1 4 Always set the voltage control to RESET or set the power switch to OFF before connecting disconnecting the earth test lead and Damp Heat IEC 68 2 3 the HV gun probe to from the DUT Installation Category I BS EN 61010 Annex J Pollution Degree 2 BS EN 61010 clause 3 7 3 2 1 5 The earth test lead and the HV gun probe must be clean and in good working order with no broken or cracked insulation 1 6 The HV gun probe must always be held correctly with fingers 6 SP ARES AND AC CE
2. Martindale Electric Metrohm House Penfold Trading Estate Imperial Way Watford WD24 4YY T 01923 441717 F 01923 446900 Email sales martindale electric co uk web www martindale electric co uk LITE3640 ISSUE 7 07 07 Metrohm E3640 FLASH TESTER INSTRUCTION MANUAL Martindale Electric Co Ltd Martindale Electric Co Ltd reserves the right to discontinue models at any time or to change specifications price or design without notice and without incurring any obligation LIMITED WARRANTY Martindale Electric Co Ltd warrants instruments and test equipment manufactured by them to be free from defective material or factory workmanship and agree to repair or replace such products which under normal use and service reveal the defect to be the fault of our manufacturing with no charge for parts and service If we are unable to repair or replace the product we will make a refund of the purchase price Consult the Instruction Manual for instructions regarding the proper use and servicing of instruments and test equipment Our obligation under this warranty is limited to repairing replacing or making refund on any instrument or test equipment which proves to be defective within 24 Months from the date of original purchase Proof of purchase may be required This warranty does not apply to any of our products which have been repaired or altered by unauthorised persons in any way so as in our sole judgement to degrade their stabil
3. SSORIES behind the guard Never try to defeat the safety trigger mechanism 1 7 Do not touch the DUT during testing with any part of the body Mains Power Supply Lead MARTL205 Earth Lead METDFH 0366 1 8 Always use the recommended replacement fuses HV Gun Probe METDFH 0369 1 9 Do not use the tester in wet conditions 2 0 Please ensure the test lead is pushed fully into the test socket and locked in position prior to carrying out any tests NOTE This tester should only be used by a competent suitably trained person 2 At Trip Level Range 0 3 to 3mA 50 60Hz Setting Accuracy 5 of full Trip Response Leakage Will trip for fault duration gt 5mS but not for lt 100pS Breakdown Will trip for fast transients of 30uS and above Safety LVD Meets BS EN 61010 1 EMC Meets BS EN 50081 1 BS EN 50082 2 Output Terminal Characteristics E3640 OUTPUT CHARACTERISTICS OUTPUT VOLTAGE kV 2kY POSITION A VOLTAGE CONTROL SETTO GIVE 42mA ON SHORT CIRCUIT LOAD 24V POSITION VOLTAGE CONTROL SET TO GIVE 2kV ON OPEN CIRCUIT LOAD o o 1 2 3 5 POSITION POSITION OUTPUT CURRENT mA 5 2 MECHANICAL Case Dimensions 330 x 263 x 144mm Case Material ABS Weight 4kg 10 2 DESCRIPTION 2 1 GENERAL The E3640 is an AC flash tester for leakage and breakdown testing It is designed for voltage testing of electrical apparatus components and insulating materials The tester is housed in a
4. ating that the high voltage circuit is energised 3 2 4 Touch the HV gun probe onto the test point of the DUT and slowly rotate the voltage control clockwise until the value indicated on the display reaches the recommended test voltage for the DUT 3 2 5 Maintain this voltage for the desired length of time If during this test period the voltage is steady and the failure indicators do not energise fail lamp and buzzer then the DUT has passed its test The recommended test time is 1 minute according to BSEN 61010 3 2 6 A failure of the DUT will result in the fail lamp illuminating and the buzzer sounding The kV ON lamp will extinguish but the voltage and leakage current at the instant of failure will by held on the displays Measurements can then be recorded accurately for each item tested Je
5. ction item combines an IEC inlet with a double pole switch fuse drawer and 115 230 volt mains power selector 2 2 6 INDICATORS kV ON A red lamp which indicates that the high voltage output circuit is energised Fail A red lamp which indicates that the DUT has failed and the trip has triggered A buzzer sounds in unison with the fail lamp to give an audible indication that the DUT has failed 2 2 7 DISPLAYS Output Voltage kV The voltage applied to the DUT is displayed and the value at failure is held until the tester is reset Leakage Current mA Leakage current through the DUT is displayed and the value at failure is held until the tester is reset 5 2 2 8 HIGH VOLTAGE SOCKET This unique safety socket has an interlock which prevents accidental disconnection of the HV gun probe lead 3 INSTRUCTIONS FOR USE 3 1 PRELIMINARY PROCEDURE NOTE OBSERVE ALL SAFETY RULES AT ALL TIMES 3 1 1 The tester can be powered from 115 or 230V 15 50 60Hz Ensure that the power source available is compatible with the voltage setting on the tester To change the setting remove the fuse drawer using a small screwdriver and pull out the fuse holder Fit a 5 x 20mm 0 25A type HBC fuse for 230 volt operation or a 0 5A of the same type for 115 volt operation Insert the fuse holder so that the desired voltage shows in the drawer window then replace the drawer NOTE If the voltage selector is not set correctly t
6. e IEE Wiring regulations BS7671 Testing is mainly carried out on factory assemblies but if an on site repair is carried out or a factory assembled item is fitted on site then the same tests are carried out Please ensure the test lead is pushed fully into the test socket and locked in position prior to carrying out any tests 5 TECHNICAL SPECIFICATION 5 1 ELECTRICAL 230 Volts 15 50 60Hz 40VA selectable 115 Volts 15 50 60Hz 40VA selectable Power Requirement Fuses 230V 5x20mm 0 25A HBC Type F 115V 5x20mm 0 5A HBC Type F Output Voltage Range 0 2 2 2 0 2 4 2kV 50 60Hz Output Voltage Accuracy 2 of reading digit Leakage Current Range 0 5 mA nominal 50 60Hz in 10uA steps Leakage Current Accuracy 2 of reading digit 29 To clear the displays the output voltage control is returned to reset NOTE a b c d e In the burn mode the trip circuit is inoperable and the tester will continue to drive current into a fault The fault current is limited to a nominal 5mA and the output voltage will decrease with increase in current In the breakdown mode if excessive current is drawn by the DUT the desired output voltage may not be achieved The specification for the DUT would then have to be checked to decide on the validity of the test The leakage current is not displayed in this mode It is advisable to return the output voltage control to reset before sw
7. e voltage control to reset will activate the HV circuits Maximum output voltages of 2 2kV and 4 2kV are set such that if these 3 limits are exceeded the output will shut down until the voltage control is turned to reset Illuminated displays show that the tester is switched on and the kV ON lamp indicates that high voltage may be present at the outlet The HV gun probe has been designed with operator safety as the prime consideration The trigger for the retractable high voltage probe sleeve is situated behind the finger guard and the integral lead is terminated in a custom designed high voltage safety connector 2 2 CONTROLS Fig 1 Refers 2 2 1 VOLTAGE This control adjusts the output of the electronic AC regulator between reset and maximum 2kV or 4kV which ever is selected The reset position is used to re energise the high voltage output circuit after 2 13 ZA a CMetrohm E3640 Flash Tester exceeding the maximum level or after power switch on Reset is also used to reset the trip circuit after a failure of the DUT 2 2 2 VOLTAGE SELECT A two way switch allows 0 2kV or 0 4kV to be selected 4 APPLICATIONS 4 1 PRELIMINARY CHECKS The E3640 flash tester is ideal for flash and applied AC voltage testing of all kinds of electrical and electronic equipment and components Requirements for testing are contained in many British and International standards and in the sixteenth edition of th
8. hen the tester may be damaged Connect the mains lead to the power inlet and plug it into a suitable power outlet If the neon to the right of the inlet socket lights reverse the plug in the socket 3 1 2 Select the mode of operation either breakdown trip or burn Breakdown When this mode is selected the trip circuit responds to high frequency currents produced by arcing faults only Note If there is a short on the DUT there will be no indication on the displays and the trip mode should be selected NOTE Leakage currents are not displayed when the break down mode is selected Trip In the trip mode mains frequency currents associated with leakage and short circuits and high frequency arcing currents that are produced by breakdown are detected Set the trip level to the required current by adjusting the control potentiometer Burn In this mode the tester will not trip regardless of any failure mode in the DUT Current will continue to flow allowing faults to be traced by observing the resulting arcing 3 2 PERFORMING A TEST 3 2 1 Attach the spade connector of the earth lead to the earth terminal 2 on the tester and the crocodile clip to the earth reference point of the DUT 3 2 2 Insert the HV gun probe into the HV socket and select 2kV or 4kV output voltage as required 3 2 3 Set the voltage control fully anti clockwise to reset and switch on the tester The kV ON lamp will illuminate indic
9. itching off the tester During applied voltage tests the current which flows into and across the surface of the DUT is the leakage current and is the resultant of two components resistive and reactive Resistive current is that which flows as a result of losses in the insulation of the DUT Reactive current is that which flows as a result of capacitance in the DUT and will increase with increase in capacitance It must be taken into account when carrying out an applied AC voltage test If the voltage select switch is operated when the voltage control is not at the Reset position the HV circuit will de energise and it will be necessary to return the voltage control to Reset to reactivate the HV circuit 2 2 3 MODE SELECT Three modes of operation are selectable Breakdown The tester will trip should breakdown occur in the DUT high frequency leakage currents Trip This position switches in a level detector whose input is variable over the range 0 to 3mA The tester will trip if the leakage current in the DUT exceeds the level that is set The tester will also trip for breakdown Burn The tester will not trip in this mode regardless of breakdown or leakage currents but the burn current is limited to a nominal 5mA 2 2 4 TRIP LEVEL This control sets the level on the trip detector that must be exceeded before a failure of the DUT is indicated 2 2 5 POWER INLET This multi fun
10. ity or reliability or which have been subject to misuse abuse misapplication negligence or accident or which have had the serial numbers altered defaced or removed Accessories not of our manufacture used with this product are not covered by this warranty To register a claim under the provisions of this warranty return the instrument or test equipment to Martindale Electric Co Ltd Metrohm House Penfold Trading Estate Imperial Way Watford WD24 4YY UK Upon our inspection of the product we will advise you of the outcome of your claim ALL WARRANTIES IMPLIED BY LAW ARE HEREBY LIMITED TO A PERIOD OF 24 MONTHS AND THE PROVISIONS OF THE WARRANTY ARE EXPRESSLY IN LIEU OF ANY OTHER WARRANTIES EXPRESSED OR IMPLIED The purchaser agrees to assume all liability for any damages and bodily injury which may result from the use or misuse of the product by the purchaser his employees or others and the remedies provided for in this warranty are expressly in lieu of any other liability that Martindale Electric Co Ltd may have including incidental or consequential damages TABLE OF CONTENTS 1 SAFETY RULES 2 DESCRIPTION 2 1 GENERAL 2 2 CONTROLS 3 INSTRUCTIONS FOR USE 3 1 PRELIMINARY CHECKS 3 2 PERFORMING A TEST 4 APPLICATIONS 4 1 PRELIMINARY CHECKS 5 TECHNICAL SPECIFICATION 5 1 ELECTRICAL 5 2 MECHANICAL 5 3 ENVIRONMENTAL 6 SPARES AND ACCESSORIES 7 LIMITED WARRANTY PAGE 8 9 9 10 10 10 11 1 SAFETY RULES
11. rugged moulded plastic case with integral carrying handle The hinged detachable lid is fitted with a pouch which contains all accessories and the instruction manual All operating functions are mounted on a panel underneath the protective lid A two position power selector adjusts the tester to operate from 115 or 230V 15 50 60Hz The tester has an output AC test voltage which may be set in the range 0 to 4kV by adjusting the voltage control The mode select switch allows the selection of either breakdown trip or burn Leakage current passing through the DUT is detected In the breakdown mode the high frequency current will cause the tester to trip In the trip mode the tester will trip if the current exceeds the trip level setting 0 to 3mA of the trip level control or if breakdown occurs However when the burn mode is selected the trip does not occur and the output voltage is maintained but it reduces as current increases When the tester trips the fail lamp will illuminate a buzzer will sound and the KVON lamp will extinguish The output voltage will drop to zero but the output voltage and the leakage current at the instant of trip will be held on their respective displays To reset the tester the voltage control is rotated anti clockwise to the reset position If the tester is switched on when the voltage control is not in the reset position the HV output is inhibited and the kV ON lamp will not illuminate Turning th
Download Pdf Manuals
Related Search
Related Contents
Toshiba Satellite C855D-S5340 Samsung Samsung GALAXY U4.71.00 - Code_Aster Plantronics 46186-01 headphone pillow KOALA+AAOmega Documentation - Anglo Corsair HX1000W, UK CMS Submission User Manual - Falcon Healthcare Systems Untitled - G 身の回りの標準サイズ A4のものがタテでもヨコでもOK! Plaquette Rythmes scolaires Copyright © All rights reserved.