        Dyn-Loc IV Quick Start Guide
         Contents
1.  a safe speed   Be sure that stable speed control of the DUT is achieved     After the DUT and dynamometer are rotating and a speed value is displayed  on the Dyn Loc IV  proceed to Section 9  Enabling the Dyn Loc IV  Controller        Dyn Loc IV Quick Start Guide V1 0 Fed   O Copyright Dyne Systems  Inc i    9 ENABLING THE DYN LOC IV CONTROLLER       Before enabling the controller for the first time  set the over speed trip by using the  thumb wheel switches on the back of the Dyn Loc IV  Based on DUT performance  specifications  select an appropriate over speed value  Note that the over speed  setting is in x100 rpm increments     IF STARTING IN TORQUE MODE     1     Press the TORQUE push button  this will put the Dyn Loc IV in Torque  Control Mode     Set the desired Linear Acceleration  LAC  also known as ramp rate  into the  active lever wheels     Press the LAC SET button to set the Linear Acceleration  display units per  second  value     NOTE  LAC should be set to match application requirements   Set both front lever wheels to different desired torque setpoint values   Press the DYNE ON button     Observe  the Dyn Loc IV should control to the active torque setpoint  Ensure  accurate and stable control response     Press the non illuminated ACTIVE button     Observe  the Dyn Loc IV should control to the newly active torque setpoint   Ensure accurate and stable control response     When done  press the DYNE OFF button  next to the lever wheel switch  to  unload the devi
2. DSUB Connector Male for 4QUAD AMP 205206 9  2 25 Pin  DSUB Connector Male for DTC AMP 5205208 1  3 37 Pin  DSUB Connector Male for REMOTE AMP 205210 8  4 Male Crimp Pin for DSUB Connector AMP 205089 1  5 Female Crimp Pin for DSUB Connector AMP 205090 1  6 Metalized Plastic Shell for 15 Pin DSUB Connector 3M 3357 6515 1C  7 Metalized Plastic Shell for 25 Pin DSUB Connector 3M 3357 6525 1C  8 Metalized Plastic Shell for 37 Pin DSUB Connector 3M 3357 6537 1C  9 7 Pin Male RPM AMPH MS3106A 16S 1P  10 7 Pin Strain Relief AMPH 97 3057 1008 1  11 6 Pin Male LOAD CELL AMPH MS3106A 14S 06P  12 6 Pin Strain Relief AMPH 97 3057 1007 1  13 9 Pin DSUB Connector Female for Serial Port AMP 205203 1   14 Metalized Plastic Shell for 9 Pin DSUB Connector 3M 3357 6509 1C                   Dyn Loc IV Quick Start Guide  O Copyright Dyne Systems  Inc    
3. Dyn Loc IV Quick Start Guide V1 0    o Single Ended Encoder Wiring     If the system uses an Encoder  refer  to the table below  the Dyn Loc IV User Manual and supplied drawings  for wiring details  Complete the sensor wiring as indicated     Encoder     Single Ended Output                   Belden 9534  Item 2 in Cable table   Pin Wire Function  G Red Encoder Excitation   12 VDC   Black Encoder Excitation Common     Shield Not connected on the sensor end  D Green Direction  Optional   Green Channel B  Optional   White Channel A                   NOTE  Never apply power to the Dyn Loc IV during sensor wiring     The cable shield must be connected on the Dyn Loc IV end only   leave shields unconnected at the transducer end to avoid ground  loop signal problems     Single ended encoders require additional hardware adjustments     3  Using the DIP switches listed below select the type of speed feedback input  source you are using  The following DIP switches are located on the DS501  circuit board  bottom center board inside the Dyn Loc IV               Setting  OFF ON OFF OFF OFF Two wire magnetic pickup   ON OFF OFF OFF OFF Encoder pickup non direction  sensitive   ON OFF ON OFF OFF Encoder pickup with direction    bit output        Encoder pickup with direction  ON OFF OFF ON OFF derived from B channel  encoder signal        Encoder pickup with direction  derived from two channel  encoder signal but inverts the  direction     ON OFF OFF OFF ON                            Energi
4. Fr Inc     Midwest  amp  Dynamatic  Dynamometers    TORQUE    fon DyneSystemsCo     e e e e  DYN LOC IV x el  MASTER TER    RPM TORQUE a Lock    0000   co Za  ar 2       Dyn Loc IV  QUICK START GUIDE    www dynesystems com Phone  1 800 657 0726    Dyn Loc IV Quick Start Guide  For Eddy Current Dynamometers        This Quick Start Guide helps to get the Dyn Loc IV dynamometer controller  configured and running quickly  Other Quick Start Guides are available for specific  AC dynamometer drive combinations  While this Guide helps to get started quickly  it  is not a substitute for The Dyn Loc IV User Manual  Refer to the manual for details  not specifically covered in this guide  The manual can be found on our web site   www dynesystems com       NOTE  This quick Start Guide applies to units with serial number greater than  748 only   Assure that the following required material is on hand before beginning          Dyn Loc IV User Manual      Various types of multi conductor shielded cable for transducer connections   see table at the end of this document       Mating MS Series connectors   see table at the end of this document     Use of this guide requires that the dynamometer be connected to a Device Under  Test  DUT   The DUT must be capable of rotating under the control of a speed  control or throttle position control     NOTE  If the DUT is not operational  energizing the eddy current dynamometer is  NOT recommended  The DUT must be fully operational to carry out the Quick 
5. Start  Guides    process in its entirety  However  connecting and testing of the speed and  torque transducers may be carried out without a functional DUT  reference sections 3  and 4 of this quick start guide     This guide outlines steps needed to bring the Dyn Loc IV to operational status as  quickly as possible  Execute these steps exactly in the order presented     Safety Review   Dyn Loc IV Power Connections   Speed Transducer Connection  amp  Configuration   Torque Transducer Connection  amp  Configuration  Dyn Loc IV Eddy Current Power Amplifier Connections  Water Safeties Interlock   Field Current Adjustments   Enabling the DUT   Enabling the Dyn Loc IV Controller    oO On oar onN      ae Inc      Midwest  amp  Dynamatic    Dynamometers    Controls  Dynamometers  System Integration and Test Cell Automation  Dyne Systems  Inc    W209 N17391 Industrial Drive Jackson  Wisconsin 53037  phone  800 657 0726 _website www dynesystems com    Dyn Loc IV Quick Start Guide V1 0 1  O Copyright Dyne Systems  Inc i       1 SAFETY REVIEW       Installing and operating this equipment involves hazards relating to rotating  equipment and dangerous electrical voltages  To avoid the possibility of serious or  fatal injury  only qualified persons familiar with industrial electrical power systems  should install power wiring  Review all safety considerations in the manual before  executing any steps in this guide or making any adjustments to the Dyn  Loc IV  It is  recommended that a licen
6. ce under test     IF STARTING IN SPEED MODE     1     Dyn Loc IV Quick Start Guide  O Copyright Dyne Systems  Inc    Press the RPM push button  this will put the Dyn Loc IV in Speed Control  Mode     Set the desired Linear Acceleration  LAC  also known as ramp rate  into the  active lever wheels     Press the LAC SET button to set the Linear Acceleration  display units per  second  value     NOTE  LAC should be set to match application requirements     Set both front lever wheels to different desired speed setpoint values  The  active setpoint value must match idle base speed of the DUT  The inactive  setpoint value must be just under the idle base speed of the DUT     Press the DYNE ON button     Observe  the Dyn Loc IV should control to the active speed setpoint  This  should result in little speed variation and minimal torque application     Press the non illuminated ACTIVE button     Observe  the Dyn Loc IV should control to the newly active speed setpoint   Ensure accurate and stable control response     When done  press the DYNE OFF button  next to the lever wheel switch  to  unload the device under test     V1 0    REFERENCE       CONNECTORS AND RELATED PARTS                                                          CABLE    Item Description Manufacturer   P N  1 Multi conductor cable  1 twisted pair  22 AWG  shielded Belden 8761  2 Multi conductor cable  4 conductors  24 AWG  shielded Belden 9534   CONNECTOR COMPONENTS    Item Description Manufacturer   P N  1 15 Pin  
7. d in  Section 5  if the dynamometer nameplate calls for more than 16 Amps field current   or the field transformer is greater than 277VAC  an external ECPAU must be used   Exceeding these limitations will damage the Dyn Loc IV     Ensure Dyn Loc IV is powered down   2  Determine the appropriate current measurement location for the application   a  Clamp on a current meter  capable of measuring DC current  around  one of the dynamometer field wires     b  Voltage measurement from test points TP7 to TP8 on the DS256  board  In standard applications  0 01 volts across the resistor is  equivalent to 1 amp of field current  TP7 and TP8 is equivalent to the  voltage across the gold power resistor found on the heat sink    3  Turn the CFB potentiometer fully clockwise  1 2 Amps minimum field current  will result     4  Determine the specified field current rating  on nameplate  for the  dynamometer  Where applicable  use the    hot amps     lesser  value    Apply power to the Dyn Loc IV   Assure the dynamometer is at a standstill   Press the TORQUE push button    Set the active lever wheel switches to 9000    Press the DYNE ON push button    NOTE  If the H20 interlocks are made  the control should latch DYNE ON  If  the DYNE ON lamp drops out  check the H20 Interlock  If the under speed  trip has not been disabled  the Dyn Loc should trip after about one second   This will cause the OS US RESET push button to light and a high braking    condition  current limit  on the field  You wi
8. er strip and are labeled as    H20 and I L      Safety devices  such as pressure  flow   and temperature dry contacts  must be connected in series  When adequate coolant  pressure and flow is supplied to the dynamometer all these contacts should be  closed  Without closed water safeties the Dyn Loc IV will not maintain a DYNE ON  condition     NOTE  Only use dry contacts  which are contacts not connected to any other circuits  or power sources     Dyn Loc IV Quick Start Guide V1 0     Copyright Dyne Systems  Inc      7 FIELD CURRENT ADJUSTMENTS       DYNAMOMETER CURRENT SETTING    Dyn Loc IV Quick Start Guide  O Copyright Dyne Systems  Inc    The current limit adjustment is made using the CFB  Current Feedback   potentiometer on the DS256 circuit board  location  Dyn Loc IV left side panel   This  adjustment allows adaptation to a wide range of dynamometer field specifications   provides protection for the dynamometer coil  and can influence stability of the  dynamometer   s control     NOTE  The fuses located on the rear panel of the Dyn Loc IV must be sized to  protect the dynamometer coil  The minimum fuse size allowed is 2 0A  The fuse  chosen should provide adequate over current and short circuit protection and be  sized based on the field coil current rating on the dynamometer s nameplate  The  voltage rating of the fuse must exceed the voltage used on the secondary of the field  transformer     Use the following procedure for eddy current dynamometers only  As mentione
9. imal place must be used to match the displayed decimal   point location     5  Ensure that the dynamometer and sensor mounting are free of any external  forces that could interfere with torque readings     Energize both the control and field transformers     Calibrate the torque transducer as follows        Dyn Loc IV Quick Start Guide     Copyright Dyne Systems  Inc    a     Attach the calibration arm to the dynamometer and attach the weight  hanger     Press and release the AUTO ZERO button on the Dyn Loc IV  This  will zero the torque display     Hang the calibration weights  Calculate the calibration torque reading  by multiplying the check arm length  from the dynamometer  nameplate  with the amount of weight applied    Torque ft     lbs    Check Arm  ft  x Weight lbs    Configure the active lever wheel for the calculated calibration torque  and press the    Auto Span    button to calibrate    Remove the calibration weights    Ensure the torque display returns to zero  If yes proceed to step    g      if no  return to step    b       Remove weight hanger  and calibration arm    Press and release the AUTO ZERO button on the Dyn Loc IV     Ensure the torque display value returns to zero  if not  return to step     R        Refer to the HELP number descriptions in the Dyn Loc IV User  Manual  Chapter 9  for HELP XXXX messages appearing on the  display during the calibration process    When the eddy current dynamometer is absorbing and the system is  rotating with positive s
10. ll have one minute to adjust the  CFB potentiometer  If it times out repress the DYNE ON button     o On on    10  Turn CFB potentiometer counter clockwise until you achieve the  dynamometer rated field current  Do not exceed the field rating     11  Push DYNE OFF push button  and OS US RESET push button  simultaneously  Leave the meter connected for the next test     V1 0    ADJUSTING THE FAULT CURRENT    The fault current is the current that will be applied to the field if an E stop  over  speed  or under speed occurs  The fault current needs to be adjusted for the  application  adjust the fault current as follows     1  Press the RPM push button to place the controller in RPM mode   2  Press E Stop button     NOTE  You will have one minute to adjust the Fault Brake pot found on the  DS501 board  This is the center board found in the bottom of the Dyn Loc   The DYNE ON times out after one minute  If adjustments are incomplete   press the DYNE ON button again     Press the E Stop reset button    Press DYNE OFF push button and OS US RESET push button  simultaneously    8 ENABLING THE DEVIGE UNDER TEST  DUT        An eddy current dynamometer is an energy absorbing device that is not capable of  driving a system  The eddy current dynamometer is driven by the DUT  typically  either an engine or an electric motor  Follow these steps in the order presented     NOTE  Speed control of the DUT is required for the below process     Enable the DUT speed control and rotate the system to
11. peed values  the torque readings should also  be positive  If not  a wiring error may exist   If so  swap wires on pins  B and C of the 6 pin M S  torque connector     8  Refer to the calibration section of the Dyn Loc IV User Manual for additional  calibration information     NOTE  In line torque sensors can be calibrated either with calibration arms  and weights or by performing a shunt calibration   Refer to Chapter 4 of the  Dyne Loc IV User Manual     9 EDDY CURRENT POWER AMPLIFIER CONNECTIONS       The internal Eddy Current Power Amplifier  ECPAU  must be connected to the  dynamometer coil  F1  F2  and the external field transformer  L1  L2   See Section 2  Dyn Loc IV Power Connections   The water safeties of the dynamometer must also  be wired to the H20 and IL terminals on the barrier strip     NOTE  The ECPAU internal to the Dyn Loc IV is rated for 277VAC maximum and  16AMPS maximum  If dynamometer coil specifications exceed these values an  external ECPAU is required  contact Dyne Systems for additional information        6 WATER SAFETY INTERLOCK       Water safety interlocks on the Dyn Loc IV are critical in detecting a dynamometer  coolant loss condition  Loss of coolant while the dynamometer is being loaded will  result in overheating and likely cause severe damage to the dynamometer     The water safeties interlock can be found on the barrier strip on the back of the Dyn   Loc IV  The water safeties interlock connections are located on the left end of the  barri
12. sed electrician perform all power connections exceeding 50    Volts     2 DYN LOG IV POWER CONNECTIONS       Follow the below steps to connect and supply power to the Dyn Loc IV     Dyn Loc IV Quick Start Guide  O Copyright Dyne Systems  Inc    Ensure all incoming power is de energized prior to proceeding     Place the Dyn Loc IV controller in a rack or on a suitable surface     NOTE  The Dyn Loc IV is a rack mountable unit  Rack mounting is  recommended due to exposed connections on the Dyn Loc IV barrier strip     Connect the control transformer secondary  typically 120 VAC  power wires  to terminals L3 and L4 on the barrier strip found on the back of the controller     NOTE  If the control transformer secondary supplies 230 VAC to L3 and L4  then the 115V 230V switch inside the Dyn Loc IV must be set for 230VAC   This switch is located in the back left corner of the bottom circuit board     Connect the field transformer secondary  typically 240 VAC  277VAC  maximum  power wires to terminals L1 and L2 on the barrier strip found on  the back of the controller     NOTE  Incoming power to both the control and field transformer primary  windings must be fed from the same phase pair  Damage may result from  improper phasing        Energize both the control and field transformers  The RPM  MASTER  EM  STOP RESET push buttons  digital LED display  and ACTIVE button on the  front of the controller will illuminate     De energize both the control and field transformers     If the pu
13. sh buttons and digital LED display illuminate as expected  proceed to  the next step  If not  reference the Dyn Loc IV user manual or contact Dyne  Systems before proceeding     V1 0    3 SPEED TRANSDUCER CONNECTION  amp  CONFIGURATION       Refer to the manual  Chapter 8  Section 3  for details on wiring and  Chapter 4   Sections 3 and 4 and Chapter 5  Section 11 4  for configuring a speed transducer  All  speed transducers are terminated at the    RPM F B     connector located on the back of  the Dyn Loc IV  A mating 7 pin MS connector is required and can be constructed  using items 9  and 10 listed in the Connector Components table at the end of this  guide  Follow these steps in the order presented     1  Ensure all incoming power is de energized prior to proceeding    2  Determine the type of speed transducer to be used  Two types of speed  transducers are supported  Magnetic Pickup Units  MAGPU   s   and Single   Ended encoders  A complete list of pin assignments on the    RPM F B      connector is shown in the Dyn Loc IV User Manual  Chapter 8      Magnetic Pickup Unit Wiring     If the system uses a Magnetic  Pickup  refer to the table below  the Dyn Loc IV User Manual and  supplied drawings for wiring details  Complete the sensor wiring as  indicated     Magnetic Pickup    O          Belden 8761  Item 1 in Cable table   Pin Wire Function  A Clear MAGPU      B Black MAGPU      C Shield Not connected on the sensor end                O Copyright Dyne Systems  Inc       
14. the Dyne Loc IV User manual and supplied  drawings for wiring details  Complete the sensor wiring as appropriate                          Belden 9534  Item 2 in Cable table   Pin Wire Function  A Red Excitation   10 VDC   Black Excitation Common     Shield Excitation Common  Green Input     Signal  White Input     Signal  DS105 PC board  solder  jumpers  JP2 installed  JP1 not installed             In Line Torque Sensor    Belden 8761  Item 1 in Cable table           Pin Wire Function  B Clear Torque Sensor Input  Both C and D Black Torque Sensor Input Common  D Shield                DS105 PC board  solder  jumpers  JP1 installed  JP2 not installed          Never apply power to the Dyn Loc IV during sensor wiring     The cable shield must be connected on the Dyn Loc IV end only  leave  shields unconnected at the transducer end to avoid ground loop signal  problems     Additional hardware may be required for In Line torque sensor applications     4  The torque display decimal point location is set from the Dyn Loc IV front  panel  Access the DIP switches by removing the torque display bezel  Set  switches  4 and  5 as shown on the table below     Torque Display Decimal Place    TQ display Switch  5 TQ display Switch  4 TQ Display                   OFF OFF 00000  OFF ON 0000 0  ON OFF 000 00  ON ON 00 000                NOTE     The power display decimal point location matches the torque  display decimal point    When configuring lever wheel set points  torque   an implied  dec
15. ze both the control and field transformers     Test speed transducer by rotating the system  by hand if possible   A speed  reading should appear on the RPM display     NOTE  If a magnetic pickup is used  a minimum speed  as a few hundred  rpm  may be as high required before a RPM reading is displayed     Dyn Loc IV Quick Start Guide V1 0     Copyright Dyne Systems  Inc      4 TORQUE TRANSDUCER CONNECTION  amp  CONFIGURATION       Refer to the Dyn Loc IV User Manual  Chapter 8  Section 2  for details on wiring and   Chapter 4  Sections 5 9  for configuring a torque transducer  All torque transducers   are terminated at the    LOAD CELL    connector located on the back of the Dyn Loc IV   A mating 6 pin MS connector is required and can be constructed using items 11 and    12 liste    d in the Connector Components table at the end of this document  Follow    these steps in the order presented        Dyn Loc IV Quick Start Guide  O Copyright Dyne Systems  Inc    Ensure all incoming power is de energized prior to proceeding     Determine the type of torque transducer to be used  Two types of torque  transducers are supported  Load Cell and Inline Sensor  A complete list of  pin assignments on the  LOAD CELL    connector is shown in the Dyn Loc IV  User Manual  Chapter 8  Section 2     NOTE  Some sensors labeled    Inline Torque Sensors    should be wired and  configured as a Load Cell  Contact Dyne Systems for assistance with this  configuration    Refer to the following tables  
    