Home

ITC-155/320/650 - Cameron Instruments Inc.

image

Contents

1. Place the probe in the calibrator and follow the instruc tions on the display Third party labs and calibration facili ties will be able to perform this function if a certificate from an independent source is necessary Of course AMETEK can provide you with a traceable calibration cer tificate from our labs when you require a higher level of confidence Calibration of up to 24 sensors with JOFRA ASM Using the JOFRA ITC series together with the ASM Advanced Signal Multi scanner offers a great time saving automatic solution to calibrate multiple temperature sen sors at the same time The ASM series is an eight channel scanner controlled by JOFRACAL software on a PC Up to 3 ASM units can be stacked to calibrate up to 24 sensors at the same time It can handle signals from 2 3 and 4 wire RTD s TC s transmitters thermisters temperature switches and volt age Please also see more in specification sheet SS CP 2360 which can be found at www jofra com www jofra com JOF RA JOFRACAL CALIBRATION SOFTWARE JOFRACAL calibration software ensures easy calibration of RTD s thermocouples transmitters thermoswithes pressure gauges and pressure switches JOFRACAL can be used with JOFRA DPC 500 APC CPC and IPI pres sure calibrators all JOFRA temperature calibrators as well as JOFRA AMC900 ASC300 multi signal calibrator and ASM 800 signal multi scanner JOFRACAL calibration software may also be used for manu
2. UNDRILLED INSERTS FOR ITC SERIES ITC 155 A ITC 320 650 A 4 mm Reference sensor 5 pack undrilled inserts 5 pack undrilled inserts with a 4 mm hole for the reference probe 5 pack undrilled inserts with a 1 4 in hole for the reference probe One undrilled insert Undrilled insulation plug 1 4 in Reference sensor 9 www jofra com Industrial Temperature Calibrator e ITC MULTI HOLE INSERTS FOR ITC SERIES METRIC MM Spare part no for multi hole inserts metric mm Insert code ITC 155 A ITC 320 650 A M x57 n N we 5s N Note All multi hole inserts metric and inches for ITC 155 are supplied with a matching insulation plug Note Remember to use matching insulation plugs Note 1 Use the insert code when ordered as the standard insert together with a new calibrator A A A 4 mm Reference 4 mm Reference Sensor ee Sensor Multi hole M01 ITC 155 A Multi hole M02 ITC 155 A 1 4 in Reference 1 4 in Reference 49 Sensor Multi hole M06 Multi hole M07 ITC 155 A ITC 155 A 4 mm Reference sensor sensor 1 4 in Reference ee Multi hole MO1 ITC 320 650 A 4 mm Reference sensor Multi hole M03 ITC 155 A 1 4 in Reference sensor amp Ts Multi hole M08 ITC 155 A MULTI HOLE INSERTS FOR ITC SERIES IMPERIAL INCH Spare part no for multi hole inserts imperial inch NA M A 122760 125380 Note All multi hole
3. temperature JOFHA Wide temperature range ITC 155 23 to 155 C 9 to 311 F ITC 320 33 to 320 C 91 to 608 F ITC 650 33 to 650 C 91 to 1202 F Improved temperature homogeneity The unique dual zone heating block ensures good temperature homogeneity in the critical calibration zone of the heating block Enhanced accuracy and stability MVI circuitry ensures temperature stability despite mains supply variations Timesaving features Fast one key one function access to the auto matic switch test and the step function High accuracy and long term stability Specified drift over a one year period of time Improves the reliability of the JOFRA ITC series Documentation made easy RS232 communication and JOFRACAL cali bration software are included in the standard delivery ISO 9001 Manufacturer Specification Sheet SS ITC Industrial Temperature Calibrator ITC 155 320 650 P Lu ui e p The JOFRA ITC series is the mid range dry block calibrator model offered by AMETEK The design basis for the ITC series is portability and ease of use supplied at a reasonable cost without sacrificing accuracy performance and features The ITC series incorporates the features of the high end ATC series with the functionality of the standard CTC series dry block calibrators The ITC series employs the slim and rugged design of the CTC series This series also features the intuitive user interface the clear LCD dis
4. 0 18 C 0 32 F Ors 4 0 0 25 C 0 45 F ITC 650 2 0 45 C 0 81 F 12 month period Specification by use of the internal reference Radial homogeneity difference between holes NEW 1 REC A M 0 02 C 0 04 F ITO 320 Fy c 0 04 C 0 07 F ITC650 A eee ee ee eee 0 1 C 0 18 F Immersion depth TG 155 A M M 160 mm 6 30 in ITC 320 A ITC 650 A nnnm 150 mm 5 91 in Well diameter FISH 20 mm 0 79 in ITC 320 ITC 650 A ern 30 mm 1 18 in Heating time ITC 155 A 20 to 23 C 4 to 73 F 4 minutes 23 10 155 C 73 to ONT F aeceecessess 14 minutes 20 to 155 C 4 to 311 F 18 minutes ITC 320 A 50 to 320 C 122 to 608 F 7 minutes ITC 650 A 50 to 650 C 122 to 1202 F 25 minutes 1 Improved specifications from January 1 2008 BEOR Industrial Temperature Calibrator JOFRA Cooling time ITC 155 A 155 to 100 C 311 to 212 F 4 minutes 155 to 23 C 311 to 73 F 14 minutes 23 to 20 C 73 to 4 F 23 minutes 155 to 20 C 311 to 4 F 3 minutes ITC 320 A 320 to 100 C 608 to 212 F 30 minutes 320 to 50 C 608 to 122 F 60 minutes ITC 650 A 650 to
5. This both improves efficiency and extends the useful life of the heat ing cooling block Peltier effect ITC 155 In 1834 Jean Peltier a French physicist found that an opposite thermocouple effect could be observed when an electric current was connected to a thermocouple Heat would be absorbed at one of the junctions and discharged at the other junction This effect is called the PELTIER EFFECT A A pA B B 5 4 7 The practical Peltier element electronic heating pump consists of many elements of semiconductor material con nected electrically in series and thermally in parallel These thermoelectric elements and their electrical interconnec tions are mounted between two ceramic plates The plates serve to mechanically hold the overall structure together and to electrically insulate the individual elements from one another WWw jofra com Maximum temperature From the setup menu the user can select the maximum temperature limit for the calibrator This function prevents damage to the sensor under test caused by the applica tion of excessive temperatures The feature also aids in reducing drift resulting from extended periods of expo sures to high temperatures This feature can be locked with an access code MVI Improved temperature stability MVI stands for Mains power Variance Immunity Unstable mains power supplies are a major contributor to on site calibration inaccuracies Traditional temperature calibrat
6. and advanced documenting functions with JOFRACAL calibration software JOFRA Pressure Instruments Convenient electronic systems ranging from 25 mbar to 1000 bar 0 4 to 15 000 psi multiple choices of pressure ranges pumps and accuracies fully temperature compensated for problem free and accurate field use JOFRA Signal Instruments Process signal measurement and simulation for easy control loop calibration and measurement tasks from handheld field instruments to laboratory reference level bench top instruments JOFRA JF Marine Instruments A complete range of calibration equipment for temperature pressure and signal approved for marine use FP Temperature Sensors A complete range of temperature sensors for industrial and marine use M amp G Pressure Testers Pneumatic floating ball or hydraulic piston dead weight testers with accuracies to 0 015 of reading M amp G Pumps Pressure generators from small pneumatic bicycle style pumps to hydraulic pumps generating up to 1 000 bar 15 000 psi because calibration is a matter of confidence AMETEK Mansfield amp Green North America T 1 800 527 9999 cal info ametek com AMETEK Singapore Pte Ltd Singapore T 65 6 484 2388 aspl ametek com sg AMETEK Inc Beijing Rep Office China T 86 10 8526 2111 jofra ametek com cn AMETEK GmbH Germany T 49 2159 9136 510 info mct de ametek de AMETEK Calibration Instruments UK T 44 0
7. 100 C 1202 to 212 F 56 minutes 650 to 50 C 1202 to 122 F 95 minutes Switch input dry contact Test voltage ee Maximum 5 VDC Test OBE Fel xuucuosssprcp cese nt dervoieruuiexduscouses Maximum 2 5 mA PHYSICAL SPECIFICATIONS Instrument dimensions LXWXH 241 x 139 x 375 mm 9 5 x 5 5 x 14 8 in Instrument weight Ges p eE P 7 6 kg 16 8 Ib G21 8 P o E 6 5 kg 14 3 Ib psc eer oe 8 5 kg 18 7 Ib Insert dimensions ITC 155 A Outer diameter 19 9 mm 0 78 in Inner diameter 16 9 mm 0 67 in Length NEN IET TTE UT 150 mm 5 91 in ITC 320 650 A Outer diameter 29 7 mm 1 17 in Inner diameter multi hole 25 9 mm 1 02 in Inner diameter single hole 22 0 mm 0 87 in Length 160 mm 6 30 in Weight of non drilled insert approximate REED X9 130 g 4 6 oz 15 320 650 A cista ia duda tias Dict cu Edi dduE 940 g 33 2 oz Shipping including optional carrying case Weight ITC 155 A eeeeeeseeees 14 0 kg 30 9 Ib Weight ITC 320 A esee 13 7 kg 30 2 Ib Weight ITC 650 A eese 15 7 kg 34 6 Ib Size L x W x H 490 x 220 x 405 mm 19 3 x 8 7 x 15 9 in Shipping without carrying case Weight ITC 155 A 11 0 kg 24 3 Ib Weight ITC 320 A ccc
8. 1243 833 302 jofra ametek co uk www jofra com Pub code SS ITC Issue 0912
9. inserts metric and inches for ITC 155 are supplied with a matching insulation plug Note Remember to use matching insulation plugs Note 1 Use the insert code when ordered as the standard insert together with a new calibrator Moz N A moa 128297 N A mos 123298 122758 Moo A 4 mm Reference sensor sensor Multi hole M04 ITC 155 A ITC Industrial Temperature Calibrator 4 mm Reference Multi hole M05 ITC 155 A 4 mm Reference sensor Multi hole M05 ITC 320 650 A 1 4 in Reference sensor JOFRA 4 mm Reference 1 4 in Reference sensor sensor Multi hole M02 ITC 320 650 A 1 4 in Reference a sensor 4 mm Reference sensor Multi hole M03 ITC 320 650 A 1 4 in Reference sensor 4 mm Reference sensor Multi hole M04 ITC 320 650 A 4 mm Reference sensor Multi hole M06 ITC 320 650 A Multi hole M09 ITC 155 A www jofra com ORDERING INFORMATION Order number ITC155A ITC320A ITC650A 115 230 TQomnmuoou C X ITC320A115BMOGCGXX Description Base model number ITC 155 series 23 to 155 C 9 to 311 F ITC 320 series 50 to 320 C 122 to 608 F ITC 650 series 50 to 650 C 122 to 1202 F Power supply US deliveries 60 Hz only 115VAC 230VAC Mains power cable type European 230V USA CANADA 115V UK 240V South Africa 220V Italy 220V Australia 240V Denmark 230V Switzerland 2
10. play and the functionality that is used in the successful CTC series However the ITC is designed with the state of the art dual zone heating block and MVI circuitry that has been adopted from the ATC series The MVI circuitry ensures stable temperatures even when the mains supply is unstable The ITC series is designed for both on site and maintenance shop use The applications are generally critical process control but can vary based on calibration and testing requirements The ITC series dry block calibrators are available in 3 different tempera ture ranges and all models are equipped with RS232 serial communica tion capabilities The standard delivery also includes the JOFRACAL calibration PC software AMETEK CALIBRATION INSTRUMENTS JOFABA ITC 320 amp ITC 650 dual zone heating block The specialized block design increases the tempera ture homogeneity in the critical calibration zone It also minimizes the need to insulate the sensors under test and makes it possible to calibrate liquid filled and other mechanical sensors The main or lower zone ensures optimum heat dissipa tion throughout the entire block The secondary or upper zone compensates for the heat loss from the top of the block and from the sensor under test el E ue m s i A E Internal reference ITC 155 heating cooling block The model ITC 155 features improved Peltier elements that employ a Multi Stage Technology
11. 0 124702 123300 2 105642 inserts 105644 Set of insulation plugs 123374 105646 Note All inserts metric and inches are supplied with a hole for the 105648 4 mm OD reference probe Note All inserts metric and inches for ITC 155 are supplied with a Package of the above a 124699 124701 matching insulation plug Note 1 Use the insert code when ordered as the standard insert Set of insulation plugs 123374 together with a new calibrator Note 2 ITC 155 12 mm 13 mm 7 16 in and 1 2 in inserts are delivered without the 4 mm reference hole but supplied with a matching insulation plug PREDRILLED INSERTS FOR ITC SERIES 1 4 IN 6 35 mm REFERENCE HOLE Probe diameter Insert code ITC 155 A ITC 320 650 A Probe diameter ITC 155 A ITC 320 650 A 808 809 A 7 16im 906 123301 125307 Package of the above E 125390 125391 Set of insulation plugs A A 125283 Note All inserts metric and inches are supplied with a hole for 125285 the 1 4 in OD reference probe Note All inserts metric and inches for ITC 155 are supplied with Package of the above a matching insulation plug j 125387 125388 inserts 125988 Note 1 Use the insert code when ordered as the standard insert Set of insulation plugs 125510 together with a new calibrator Note 2 ITC 155 12 mm 13 mm 7 16 in and 1 2 in inserts are delivered without the 1 4 in reference hole but supplied with a matching insulation plug 125510 N A
12. 20V Israel 230V Insert type and size 1 x Insert for dry block configuration please see the previous insert pages for the right insert codes Calibration certificate Traceable calibration certificate standard for Europe Asia Australia and Africa NIST traceable calibration certificate standard for Western Hemisphere Accredited calibration certificate Options Carrying case No option used Sample order number JOFRA ITC 320 A dry block calibrator 115VAC power with US power cord and predrilled multi hole insert type 6 including carrying case and NIST traceable certificate AMETEK CALIBRATION INSTRUMENTS AMETEK Denmark A S Gydevang 32 34 3450 Aller d Denmark T 45 4816 8000 ametek ametek dk Sales amp Service Europe Asia Africa Middle East and South America Information in this document is subject to change without notice 2009 by AMETEK Inc www ametek com All rights reserved www jofra com AMETEK Calibration Instruments is one of the world s leading manufacturers and developers of calibration instruments for temperature pressure and process signals as well as for temperature sensors both from a commercial and a technological point of view JOFRA Temperature Instruments Portable precision thermometers Dry block and liquid bath calibrators 5 series with more than 25 models and temperature ranges from 90 to 1205 C 130 to 2200 F All featuring speed portability accuracy
13. al calibrations as it can be set up to accept manual entry of calibration data together with other liquid baths ice points or dry block heat sources The calibration data collected may be stored on a PC for later recall or analysis The ITC calibrator stores the cali bration procedure and may be taken out to the process site without using a personal computer Once all calibrations are completed the data may be uploaded to the JOFRACAL calibration software for post processing and printing of certificates The calibration data collected may be stored on the personal computer for later recall or analysis The JOFRACAL temperature calibration software may be donwloaded from our web page www jofra com Please also see more about JOFRACAL calibration software in specification sheet SS CP 2510 which can be found at www jofra com JOFRACAL software Minimum hardware requirements for JOFRACAL calibra tion software e INTEL 486 processor e PENTIUM 800 MHz recommended e 32 MB RAM 64 MB recommended e 80 MB free disk space on hard disk prior to installation e Standard VGA 800 x 600 16 colors compatible screen e 1024 x 786 256 colors recommended e CD ROM drive for installation of the program e 1 free RS232 serial port www jofra com STANDARD DELIVERY ITC dry block calibrator user specified Mains power cable user specified Traceable certificate temperature performance Insert user specified 3 pc
14. cscecsecsssessessesseseeseeseee 10 7 kg 23 6 Ib Weight ITC 650 A 12 7 kg 28 0 Ib Size L x W x H 460 x 216 x 405 mm 18 1 x 8 5 x 15 9 in Shipping carrying case only Lar e gy capuc cm E 5 0 kg 11 Ib Size L x W x H 490 x 220 x 405 mm 19 3 x 8 7 x 15 9 in Miscellaneous Serial data interface RS232 9 pin Male Operating temperature 0 to 40 C 32 to 104 F Storage temperature 20 to 60 C 4 to 140 F Humidity e 0 to 9096 RH Protection class e E IP 10 www jofra com JOF RA PREDRILLED INSERTS FOR ITC SERIES 4 MM REFERENCE HOLE JOFRA dry block insert compatibility and materials All specifications on hole sizes are referring to the outer ATC 320 ATC 650 ITC 320 ITC 650 made of brass diameter of the sensor under test ATC 155 ATC 156 made of aluminum The correct clearance size is applied in all predrilled ATC 157 ITC 155 made of aluminum inserts ATC 140 ATC 250 made of aluminum Probe diameter ITC 155 A ITC 320 650 A 003 128270 105620 amm o 128271 105624 5mm o 128272 105020 7 mm 007 008 Probe diameter ITC 155 A ITC 320 650 A mwn s wwe 1596 250 12928 105680 312 129282 1056082 U2in 500 123302 105688 nein 562 NA j 105690 sin 65 NA j 10692 pm Package of the above 12470
15. or under test in their position while calibrat ing Includes 2 sensors grips and 2 fixtures for sensor gribs Industrial Temperature Calibrator e ITC FUNCTIONAL SPECIFICATIONS Mains specifications Voltage ITC 155 320 A 115V 90 127 230V 180 254 Voltage ITC 650 A 115V 100 127 230V 200 254 Frequency non US deliveries 50 Hz 5 60 Hz 5 Frequency US deliveries 60 Hz 5 Power consumption max ITC 155 A 150 VA Power consumption max ITC 320 650 A 1150 VA Temperature range ITC 155 A MaxiMUT iicet ate tse Ria 159 C 311 F Minimum ambient temp 0 C 32 F 39 C 38 F Minimum ambient temp 23 C 13 F 23 C 9 F Minimum ambient temp 40 C 104 F 10 C 14 F 145320 Assesse 33 to 320 C 91 to 608 F OZON rs 33 to 650 C 91 to 1202 F Resolution user selectable All temperatures eeeeeeeeeeeennene nnn 1 or 0 1 Stability NEW ITG 155 A c 0 01 C 0 02 F N20 F cd 0 02 C 0 04 F TC 650 p scm cent 0 04 C 0 07 F Measured after the stability indicator has been on for 10 minutes Measuring time is 30 minutes Time to stability approximate All models eren 10 minutes Accuracy NEW 1 aie
16. ors often become unstable in production environ ments where large electrical motors heating elements and other devices are periodically cycled on or off The cycling of supply power can cause the temperature regu lator to perform inconsistently leading to both inaccurate readings and unstable temperatures The ITC series employ the MVI thus avoiding such stabil ity problems The MVI circuitry continuously monitors the supply voltage and ensures a constant energy flow to the heating elements Easy to use intuitive operation All instrument controls may be performed from the front panel The heat source is positioned away from the panel This design helps to protect the operator The main functions on the ITC series are designed with one key one function logic This means that there are no sub menus or difficult to remember multiple keystrokes necessary to access primary functions The easy to read backlit display features dedicated icons which help in identifying instrument conditions and opera tional steps AMETEK ki peu a pom em UC eed TR Industrial Temperature Calibrator e ITC Set temperature The Up and Down arrow keys allow the user to set the exact temperature desired with a resolution of 0 1 Instrument setups The ITC series stores the complete instrument setup including engineering units stability criteria resolution display contrast slope ramp rate auto step settings and maximum tempera
17. s insulation plugs for 5 8 11 mm sensors ITC 155 only Tool for insertion tubes RS232 cable JOFRACAL calibration software User manual Reference manual English Test cables 1 x red 1 x black ACCESSORIES 122832 Cleaning Brushes 4 mm Package of 3 pcs 60F174 Cleaning Brushes 6 mm Package of 3 pcs 122822 Cleaning Brushes 8 mm Package of 3 pcs 123374 Set of 3 pcs of insulation plugs 4 mm ref hole 5 8 and 11 mm 02 0 31 and 0 43 in TC 155 A only 125510 Set of 3 pcs of insulation plugs 1 4 in ref hole 5 8 and 11 mm 0 2 0 31 and 0 43 in ITC 155 A only 125068 Support rod set for sensors 2 gribs 2 fixtures Support rod set can be mounted on all JOFRA dry blocks 125066 Extra fixture for sensor grib 125067 Extra sensor grib 125002 Edge port Converter with 4 pcs of RS232 ports 123396 Carrying Case for ITC Series Carrying case Optional 123396 The optional protective carrying case ensures safe transportation and storage of the instrument and all associated equipment Heat shield Optional 104216 An external heat shield is avail able and may be placed on top of the calibrator to reduce the hot air stream around the sensor under test This is especially important for testing thermocouples having head mounted transmitters with cold junction compensation Support rod set Optional 125068 Support rod for sensors to be mounted on all JOFRA dry block calibrators Holds the sens
18. ture Stability indicator A bold checkmark on the display indicates that the cali brator has reached the desired set temperature and is stable The operator may change the stability criteria and establish a greater sense of security in the calibration results A convenient countdown timer is activated five minutes before the unit reaches stability Automatic switch test Operators can save a lot of time using the automatic ther moswitch test function to find values for the Open and Close temperatures Additionally this feature displays the hysteresis deadband between the two points The feature ensures a very high repeatability when testing thermoswitches Simply press the SWITCH TEST key to activate the function AMETEK CALIBRATION INSTRUMENTS Auto stepping This feature saves manpower The operator may stay in the control room or another remote location monitor ing the output from the sensor under test while the ITC series calibrator is placed in the process and automatically changes the temperature using a programmed step value and rate Up to 9 different temperature steps may be pro grammed including the hold time for each step BEOR Industrial Temperature Calibrator JOFRA Re calibration adjustments The ITC series has a very easy and straightforward proce dure for re calibration adjustment There is no need for a screwdriver or PC software The only thing you need is a reliable reference thermometer