1.      6 2 CONTROL MODE    Few seconds after the warming up process of the lamp  the Turbi Quick will automatically enter in the  Control Mode and on the display a Control sample will be required  the instrument now is ready to  analyze one or more Control     PT  Control n  1  insert cuvette    and press button    When this indication  Stand by mode  appears on the display the Operator has to press rapidly the  button  On the display will appear     PT  Control n  1  when T 0 add  PT Rea  T 5     A 5 seconds countdown will start  when the T   zero seconds  on the display will appear the following  message   PT  Control n  1  add PT Rea       Now the operator has 3 seconds to inject the starter reagent   During these three seconds  on the display will appear     PT  Control n  1  mixing       And immediately after     PT  Control n  1  Time  XX    Once the analysis is finished  the clot is formed   the timer stops  and on the display appears   Control n  1  PT  26 3 INR 2 43    PT  222  Press the button       Pag 12   24    Vital Diagnostics    TURBI QUICK   USER S MANUAL       If the a printer is connected  the same information will be also printed out   Once the analysis is finished  the clot is formed   the timer stops  and on the display appears     Control n  1    PT  26 3    INR 2  43    PT  222  Press the button       If the a printer is connected  the same information will be also printed out     Now the operator can     a  Press the button rapidly  less than 2  seconds
2.    Turbi Quick    USER   S MANUAL    Manual code MAN 037     Revision 5  Revision date  05 26  2008    TURBI QUICK     USER S MANUAL Vital Diagnostics       Pag 2   24    Vital Diagnostics TURBI QUICK     USER S MANUAL       INSTRUMENT NAME  Turbi Quick      INSTRUMENT MANUFACTURER  Vital Diagnostics S r l   Via Balzella 41 g 4  47100 Forl    ITALY    INSTRUMENT DISTRIBUTOR     Please read this manual completely before using the instrument    Pag 3   24    TURBI QUICK   USER S MANUAL Vital Diagnostics       1  INTRODUCTION E NEE E E T E E a cubase tasse E sonia 5  Eli  MAAND D US E mura es er ti tm reti tnt ich rode iu aim hr tute 5  1 2  IMPROPER US E iii EN Ia E E E E einen ten 5   2  INSTRUMENT DESCRIPTION             cccsssssssscccccccscsssssccccccccccssssccecccccecsssscccecceecessscccceccceeecssscsscccceseesscees 5  2L SCUVECIESHOEDER  SEA a M o ec dea e den aio a dee bn e oo i Miao pei 5  2 2  SMART CARD    na ANNE A A A USNJEN VR EE 6  2 3 RESET BUTTON                 eee 7  24   FILTERSSEEECTOR   5 2 did  7   3  INE NS aka zaka AE eve okove N 8  Sed   TURBEOUICK POSTITONINGE lalla AA aje A A teak Re Sees ates A ae 8  3 2  POWER O NEDENARNE ESS K SN NE UN NE eti  8  3 3 WARM UP AND CALIBRATION kk 9  3 4    LAMP WARMING UP PROCESS dinas enee AR Up ob ad 9   4  OPERATING RECOMMENDATION G         cccccccssssssssscccccccssssssccccccccccssssscccccccecesssccccecceecccssscccccceeeeessscees 10   5  SAFE IA O 10  5 1 USER PRECAUTIONS                 ceri eee 10  5 2 ELECTRICAL 
3.   In this case  the analyzer will enter in  the stand by mode  and is ready for a new  control     b  Press the button more than 2  seconds If the operator wants to start to  analyze patients sample  he can press more  than 2 second the button  the instrument will    PT enter to Sample mode and on the display will  Control n  2 appear   insert cuvette   PT    and press button Sample n  1    insert cuvette    and press button    Now  follow instructions described at point 6 3   6 3 SAMPLE MODE  PRESS THE BUTTON MORE THAN 2 SECONDS   In case the operator does not want to perform any control  he can start immediately to analyze patient   s  samples  To pass directly to Sample mode  press the button for more than 2 seconds  then on the  display will appear   PT  Sample n  1    insert cuvette    and press button    When this indication  Stand by mode  appears on the display the Operator must insert the cuvette into  the reading channel and then press rapidly  less than 2 seconds  the button  In this way a 5 seconds  countdown will start  when the T   zero seconds  on the display will appear the following message     PT  Sample n  1    add PT Rea       Now the operator has 3 seconds to inject the starter reagent   During these three seconds  on the display will appear     PT  Sample n  1    mixing     And immediately after     PT  Sample n  1    Time  XX    Pag 13   24    TURBI QUICK   USER S MANUAL Vital Diagnostics       Once the analysis is finished  the clot is formed   the tim
4.  6  Inject the Starter reagent  straight on the light bundle  Do not insert the whole pipette tip into the  cuvette  but just the end of the tip     7  When insert the cuvette into the Reading channel  push it down strongly  until the cuvette  reaches the bottom of the channel  This operation is very important  because an incorrect cuvette  positioning will cause an erroneous sample   s reading     8  Each sample and or control must be prepared following the instruction sheet inside each Turbi  Kit  Once the sample is ready  transfer the cuvette into channels 1     4 for the pre incubation  Maintain  the cuvette into the pre incubation channel for the time shown on the instruction sheet  Once the pre   incubation time is finished  transfer the cuvette into the reading channel and follow all the indication  previously described to perform the test     5  SAFETY MEASURES  5 1 USER PRECAUTIONS    Before using the analyzer  the operator must know the rules for handling potentially infectious materials  and electro mechanical systems     5 2 ELECTRICAL EQUIPMENT    As with all electrical equipment  the power supply is a potential source of danger  To prevent the risk of  electrical shock to the user and or damage the instrument  the operator should not open any part   electrical  of the instrument  Only authorized personnel  such as service technicians  may open the  instrument to perform maintenance or repair     5 3 SAMPLES    All biological fluids must be considered by the oper
5.  Diagnostics S r l  and supplied together  with the instrument     1 2 IMPROPER USE    Following uses are considered improper   1  The use of the device by not qualified personnel  The use of cuvettes different from those specified on the manual and on the inserts  The cuvette re use  The use of reagents different from those specified on the the manual and on the inserts  Every attempt to analyze samples different from human serum or plasma  Every attempt to obtain different parameters form those specified on the inserts  The exposure of the device and of the optical part to liquids  which can soak into the reading  system   8  The insertion of devices different form cuvettes into the reading channel  9  Disregard instructions given on this manual  The above mentioned uses and every attempt to use the Turbi Quick with a purpose different from the  intended use  must be considered improper     tO Or A    2  INSTRUMENT DESCRIPTION  2 1 CUVETTES HOLDER    In the upper part of the Turbi Quick reader  there is the cuvettes holder  that is automatically  thermostatized at 37  C  It has with 5 channels  Channel from 1 to 4 are pre incubation channels and can  be used to reach the sample   s temperature of 37  C    The channel marked with    V    is instead the reading channel where cuvettes must be transferred after the  pre incubation phase     The reading channel is a photometer with two different filters  A and B  available     Pag 5   24    TURBI QUICK     USER S MANUAL Vital Di
6.  In case the injection time has been expired   and the reagent has not been injected  on the  display will appear     PT  Control n  1  PT Rea  not detected  repeat sample       In this case  press the button rapidly  less than 2  sec    on the display will appear     PT  Control n  1  insert cuvette    and press button    Now the instrument is ready again to perform  the test  press again the button and insert the  starter reagent when the display requires the  injection     In case the reagent has been injected  due to  the small reagent volume and or the small  reagent absorbance  the analyzer can t  recognize the injection  the operator has to  press the button within 2 seconds and on the  display will appear consequently     CRP Latex  Control n  1  mixing       CRP Latex  Control n  1  reading       CRP Latex  Control n  1  Time  119     Once the reading time has expired  on the  display will appear     CRP Latex  Control n  1  conc  12 3 mg L  press the button       If the a printer is connected  the same  information will be also printed out    Now the operator can run another sample as  previously described    Pag 17   24    TURBI QUICK   USER S MANUAL    Vital Diagnostics       8 2 SMART CARD CHANGE    It is possible to change the smart card  also without switching off the instrument  This special feature is  available only during the analyzer stand by mode  The stand by mode can be displayed both during    samples analyses and during controls analyses     A  during
7.  control mode  The stand by mode is displayed as follows     PT  Control n  1  insert cuvette    and press button    Under this condition  to change the smart card   it is necessary firstly to pass from control to  sample mode  pressing the button for more than  2 seconds  the analyzer will enter in the sample  mode     PT  Sample n  1  insert cuvette    and press button    Now it is possible to remove the smart card and  insert a new one and pass from  e g   PT  analysis to APTT analysis  To do that  follow the  instructions described under point B      Pag 18   24    B  during sample mode  The stand by mode is displayed as follows     PT  Sample n  1  insert cuvette    and press button    Under this condition  it is possible to remove the  smart card and insert a new one and pass from   e g   PT analysis to APTT analysis    To do that  remove the old smart card and insert  the new one  Then press more than 2 seconds  the button  so the instrument will start again with  the calibration procedure       Turbi Quick      Reader V X X                             APTT Filter  A  Lot  9999 16 05 05    test counter  9999      Warmup to 37 0 C     Temperature  34 6       Filter calibration      please wait         APTT  Control n  1  insert cuvette    and press button    Now the instrument is in Control Mode  ready to  analyze the first APTT control     Vital Diagnostics TURBI QUICK     USER S MANUAL       8 3 HIGH LOW SAMPLE   S CONCENTRATION    For those analytes expressed in conc
8.  etici 23   13    DISPOSAL  AND  RECYCLING iesssdssccssecessessscccectdecsseessssecosssedascedsncecesssocssocsestesesssecssncteotedeessondcustestesesseses 24    Pag 4   24    Vital Diagnostics TURBI QUICK     USER S MANUAL       da INTRODUCTION    Dear User  we hope you will be satisfied with your Turbi Quick   analyzer    The Turbi Quick    analyzer is a reader especially developed for coagulation and immunochemistry  analyses    The coagulation tests are performed using the optical detection of the clot  once the clot is formed  there  is an absorbance shift which allows the detection of the end of the reaction    For what concerns instead Immunochemistry  a complete panel of quantitative immunoturbidimetric  analyses is available for the Turbi Quick   reader  all the tests have been studied by Vital Diagnostics and  optimized for the Turbi Quick   reader  All the tests are based on the Turbidimetry principle  the  attenuation of the intensity of the light through a solution containing Antigen Antibody aggregates  is  proportional to the Antigen concentration     This manual describes the Turbi Quick   way of working  Please read it completely to understand the  capabilities of your instrument     1 1 INTENDED USE    The Turbi Quick analyzer is a semi automated photometer developed for the analysis of coagulation and  immunochemistry parameters  performed on human serum or plasma  The complete list of tests  available on the instrument  is supplied and updated only by Vital
9. 20   24    Vital Diagnostics TURBI QUICK     USER S MANUAL          Picture 4 Picture 5    Unscrew the screw blocks the lamp as shown in picture 5 and then remove the lamp   Insert a new lamp  block it whit the screw  connect the cable and mount the back panel of the  photometer     10  PERFORMANCE CRITERIA AND LIMITATIONS    10 1 PERFORMANCE CRITERIA    A  Optical detection precision   0 001 O D   Photometric Measurement    B  Analysis Reproducibility   C V    lt  6  test depending    C  Cuvette Holder block thermostatized 37  C  Automatic  room temperature between 15 32   C   D  Photometric zeroing   Automatic   E  measuring points   2   F  Measuring range   0 1 50 D     10 2 LIMITATIONS    A  Strongly lipemic or haemolytic samples may alter instrument   s reading capability     B  Samples with a concentration  gt  or  lt  than linear range     Pag 21   24    TURBI QUICK   USER S MANUAL    Vital Diagnostics       11  TECHNICAL SPECIFICATIONS    Area of application    Instrument size    Weight    Voltage    Reading Channel  Pre incubation Channels  Thermostatization  Analysis time   Loading pattern    Results    Measuring Principle    Filters    Reading resolution  Display interface  Interface   Software    CE Marked    Pag 22   24    Immunochemistry and coagulation analyses  Height 145 mm   Width 210 mm   Depth 120 mm   1100 gr     external power supply 240 V  110 V  A C  50  60  Hz  output  12 V dc 3 5 A    1  4   37   C   kit s dependent  One test a time    Shown on the d
10. 24    
11. EOUIPMENT      e iii 10  SN A NA 10  5 4        NOTES   ON SAFETY  MEASURES ii I e rea 11  5 5 RESIDUALRISKS             eee 11   6  COAGULATION OPERATING PROCEDURE              ccccccccssssssssscccccccccsssssccccccccecssssssceccceeecssccccccceseeees 12  Gal    FOREWORDS tt eet I ce ieee wean Tdi 12  6 2 CONTROL MODE                    rire 12  6 3 SAMPLE MODE  PRESS THE BUTTON MORE THAN 2 SECONDS     13  6 4 READING ERROR MESSAGE kakec 14   7  IMMUNOTURBIDIMETRIC OPERATING PROCEDURE             sccssscccccccssssssssscccccceccsssscccccceeeecssscees 14  TEL KONRAD da 14  7 2 SAMPLE MODE  PRESS THE BUTTON MORE THAN 2 SECONDS     aaa 15   8  TROUBLESHOOTING oreesa ieat ROTAIA RATA TIRA ici 17  8 1 THE INSTRUMENT DOESN   T RECOGNIZE THE STARTER INJECTION         cscccccceseccccessceccueccecuueceseueecsseueceeeaners 17  622     SMART CARD CHANGE Sci 18  8 3 HIGH LOW SAMPLE    S CONCENTRATION                        erre 19  0 47  READING ERRORS  tit is ti ld a E eta ba Baia eS 19   9  MAINTENANCE AA O 20  9 1     GLEANING INSTRUGTIONS  sni ne nee aa tai 20  9 2 LAMPREPLACEMENT                    erre rie eee 20   10  PERFORMANCE CRITERIA AND LIMITATIONS              cccccsssssssssccccccccscsssscsccccccecssssccscccceeeccssscsceeees 21  10 1 PERFORMANCE CRITERIA 2 50 en a ente 21  10 2 MEAN NS a dit i de di o 21   11  TECHNICAL SPECIFICATIONS iisscccscccssssssescuscsccsscessssccosssscsscossetesessscsssonscesssssescsssacsseesveeesoaccosseseesessenes 22   12    EGCDEGEARATION    cnc RR
12. agnostic use only    A t i 4 ELECTROSTATIC DISCHARGE SENSITIVE DEVICE  ESDS      The device could be damaged by electrostatic potentials    5 5 RESIDUAL RISKS    Despite of the measures taken in the designing of the machine to guarantee a safe use of it  there might  happen reasonably predictable occurrences  whose risk was possible to reduce  but not to eliminate    completely        RESIDUAL RISKS    PROTECTION MEASURES       Biological contamination    The operator must wear always gloves and  protection glasses  as prescribed by laboratory  regulations  Do not ever open tubes       Cuvettes breaking and or overthrow    In case of cuvettes accidental broken and or  overthrow  clean the instrument with a sodium  hypochlorite solution        Cuvettes re use          Cuvettes are disposable  so the re use of cuvettes  can false the analysis result       Pag 11   24       TURBI QUICK   USER S MANUAL Vital Diagnostics       6  COAGULATION OPERATING PROCEDURE  When a Coagulation smart card is inserted in the Turbi Quick  follow these instructions   6 1 FOREWORDS    The cuvettes required to perform coagulation test have a volume of 1 5 mL and a particular shape  studied to reduce the volume of reagent needed  code  ACC 16 047   Only this kind of cuvette is suitable  to perform coagulation tests  All the cuvettes different from ACC16 047 must be discarded    Coagulation tests are very sensible to the temperature  Strictly respect pre incubation time to obtain  reproducible results
13. agnostics               Pre incubation channels  from 1 to 4     Reading Channel       2 2 SMART CARD    The Turbi Quick   reader works only in association with Turbi kits    Turbi kits   are the reagents needed  to perform analyses on the instrument and inside each Turbi kit box  there is a Smart Card  The smart  card must be inserted in the Smart Card Slot  located in the rear panel of the instrument  Without the  smart card  the instrument doesn   t work        m         ss    Smart Card  In the Smart Card s chip in fact  the following data are stored       Calibration Curve  in each card the calibration curve of the test is stored  in this way the Operator  doesn t need to perform the calibration curve  but he can start immediately with patient s samples      Test Settings  reading time  mixer behavior and kind of filter are already stored in the card in  order to avoid incorrect readings due to wrong filter  mixing or reading time setting errors      Expiry Date  the Expiry Date of the kit is stored on the card in order to avoid the use of expired  kits      Number of test  the number of tests performable with the kit  is stored in the card  this counter  will decrease each time the user performs one test     Pag 6   24    Vital Diagnostics TURBI QUICK     USER S MANUAL       Smart Card Slot    Power ON OFF       Smart Card Slot and Power ON OFF switch  2 3 RESET BUTTON    On the upper part of the cuvettes holder  between channel 2 and 3  is located the Reset Button whic
14. ample T 5     Pag 14   24    Vital Diagnostics TURBI QUICK   USER S MANUAL       A 5 seconds countdown will start  when the T   zero seconds  on the display will appear the following  message   CRP Latex  Control n  1  add sample       Now the operator has 3 seconds to inject the starter reagent  Once the sample has been injected  on the  display will appear     CRP Latex  Control n  1  mixing     And immediately after   CRP Latex  Control n  1  reading       CRP Latex  Control n  1  Time  119     Once the reading time has expired  on the display will appear     CRP Latex  Control n  1  conc  12 3 mg L  press the button       If the a printer is connected  the same information will be also printed out   Now the operator can     a  press the button more than 2 b  press the button rapidly  less than 2  seconds seconds   If the operator wants to start to analyze patient s In this case  the analyzer will enter in the  sample  and not controls   he can press more stand by mode  and is ready for a new control   than 2 second the button  the instrument will  enter to Sample mode and on the display will CRP Latex  appear  Control n  2  insert cuvette    CRP Latex and press button  Sample n  1  insert cuvette   Now the operator can run another control as    previously described  Now follow instructions described at point 7 2     7 2 SAMPLE MODE  PRESS THE BUTTON MORE THAN 2 SECONDS     In case the operator does not want to perform any control  he can start immediately to analyze patie
15. ator as potentially infectious  For this reason  the  operator must adopt the national and international standards of precautions to avoid the biological  danger    Specimens  patient samples and controls  and liquid waste should be considered potentially infectious  and capable of transmitting human immunodeficiency virus  HIV   hepatitis B virus  HBV  and other    Pag 10   24    Vital Diagnostics    TURBI QUICK     USER S MANUAL       blood borne pathogens     The handling of these substances must be performed in accordance with established laboratory safety  regulations in order to minimize risk for laboratory staff    This includes wearing of gloves  splash protection  etc  Contact with skin and mucous membranes must  be avoided  This also applies to all components of the instrument that are exposed to these substances   If any specimen is spilled on the instrument  wipe it up immediately and clean the contaminated surface  with a disinfectant of 0 5  sodium hypochlorite solution    Compliance with local regulations pertaining to the disposal of waste is the responsibility of the operator   Refer to local sources for additional information on correct biohazardous waste disposal    Qualified technical operators must apply the same warning procedures for instrument maintenance     5 4 NOTES ON SAFETY MEASURES    On the instrument  to assure a correct use of the instrument  may be placed the following symbols     Attention  read use instruction    A    IVD    For in vitro di
16. ee the Operating Procedures of this manual     Pag 9   24    TURBI QUICK     USER S MANUAL Vital Diagnostics       4  OPERATING RECOMMENDATIONS    These recommendations are essential to work correctly with Turbi Quick   reader and with Turbi Kits   Please  before performing analyses  read carefully these recommendations     1  Once the cuvette has been inserted into the Reading Channel  don   t touch the cuvette until the  reading time is finished  the cuvette   s movement could cause an erroneous sample   s reading     2  Once the cuvette has been inserted into the Reading Channel  don   t remove the cuvette until the  reading time is finished  the cuvette   s removing causes an erroneous reading process     3  To obtain reproducible results with Turbi Quick   reader  it is necessary to use calibrated  pipettes  an erroneous volume of reagents will cause an incorrect ratio between reagents and  consequently results won t be reproducible  Vital Diagnostics recommends to calibrate pipettes at least  every six months     4  The Smart Card can be removed from the Turbi Quick   only when the reader is switched off or  is in stand by mode  Removing the Smart Card while the instrument is working can damage Turbi   Quick   and cause an erroneous sample   s reading     5  During the injection of the Starter reagent  avoid bubble formation  Bubbles can create  interference with the light source and cause an erroneous reading  In case of bubbles  we suggest to  prepare a new sample    
17. entration  e g  Fibrinogen   if a sample has a concentration lower  than the minimum value of the linear range  on the display will appear     Fibrinogen  Sample n  1  conc  lt 10 0 mg L  press the button       While if a sample has concentration higher than the maximum value of the linear range  on the display    will appear     Fibrinogen  Sample n  1  conc  gt 300 mg L  press the button       In this case  the operator can repeat the test after sample   s dilution  follow insert instructions     8 4 READING ERRORS    In case of reading errors  the following messages can appear        Reading error       Check r  channel     Check filter pos     land press button      In this case the operator can   a     check if the filter selected  is the right one  if not  move  the filter lever on the right  position  and press the  button  The instrument will  check the reading system  again    b     check if the reading  channel is empty  In case the  reading channel is not empty   remove the cuvette and  press the button   If these 2 operations  won t  solve the problem  call the  technical assistance         White error    t   Check r  channel    land press button      In this case the operator can   a     check if the filter selected  is the right one  if not  move  the filter lever on the right  position  and press the  button  The instrument will  check the reading system  again    b     change the lamp as  described on this manual    If these 2 operations  won t  solve the prob
18. er stops  and on the display appears     Sample n  1  PT  26 3 INR 2  43  PT  222  Press the button       If the a printer is connected  the same information will be also printed out   Now the operator should press the button rapidly  less than 2 seconds   in this way the analyzer will enter  in the stand by mode  and is ready for a new sample     PT  Sample n  2  insert cuvette    and press button    6 4 READING ERROR MESSAGE    In some particular tests like PT or APTT  can happen that the clot doesn t form  so the instrument can t  fix the end of the reaction  When this happens  the following message will appear on the display     PT  Sample n 1  Reading error  Press the button    In this case  remove the cuvette from the reading channel and press the button rapidly  less than 2 sec    in this way the instrument will return in the stand by mode     PT  Sample n  2  insert cuvette    and press button    7  IMMUNOTURBIDIMETRIC OPERATING PROCEDURE  When an Immunoturbidimetric smart card is inserted in the Turbi Quick  follow these instructions   7 1 CONTROL MODE    Few seconds after the warming up process of the lamp  the Turbi Quick6 will automatically enter in the  Control Mode and on the display a Control sample will be required  the instrument now is ready to  analyze one or more Control     CRP Latex  Control n  1  insert cuvette    and press button    The operator now has to press rapidly the button  on the display will appear   CRP Latex  Control n  1    when T 0 add  s
19. h  can be used to reset the instrument  The complete functioning of this button is explained in this manual   Please read it carefully before use the Turbi Quick       Reset Button       2 4 FILTERS SELECTOR    On the front side of the cuvettes holder  just in front of the reading channel  is located the Filters  Selector  This selector allows the user selecting between 2 different filters  filter A and filter B    Each test  requires a specific filter  the filter needed  appears on the Turbi Quick   s display as soon as the  instrument  with a Smart Card inserted into the slot  is switched on    To change the filter  move the lever to the left for filter A or to the right for filter B    The filter should be selected before switching on the instrument  The filter required is also written in the  internal insert of each kit     Pag 7   24    TURBI QUICK   USER S MANUAL Vital Diagnostics       3  INSTALLATION  3 1 TURBI QUICK POSITIONING    The Turbi Quick   shouldn  be placed near centrifuges  oscillating agitators or other vibrating  instruments that might cause movement of the bench   It should be placed far from windows or direct light sources to avoid Turbi Quick   overheating     3 2 POWER ON    Connect power supply outlet to the instrument    Insert the power supply plug in a socket with an earth connection    If the instrument has an optional printer  it should be connected to the Turbi Quick   with the appropriate  cable and plugged in    Insert the Smart Card in the 
20. he new  Smart Card  Now you can switch on again the Turbi Quick       3 3 WARM UP AND CALIBRATION    Once the instrument is switched on with a correct Smart Card  it will start the warm up procedure  during  this time  the Turbi Quick   will heat the cuvette holder until the temperature of 37  C is reached   On the display will appear      Warmup to 37 0 C      Temperature  34 6      Once 37  C are reached  the Turbi Quick   will perform the auto calibration and the filter checking   On the display will appear      Filter calibration      please wait         If some calibration problems occur  on the display will appear        Reading error       Check r  channel     Check filter pos     land press button      This message can be displayed when    The filter selected is wrong  in this case  move the filter s selector to the other filter and press the Reset  Button  The Turbi Quick   will repeat again the calibration with the right filter    The reading channel is not empty  in this case  remove the cuvette from the reading channel and then  press the Reset Button  The Turbi Quick6  will repeat the calibration     3 4 LAMP WARMING UP PROCESS    Once the calibration is finished  the Turbi Quick   reader starts the Lamp warming up process and on the  display will appear       Lamp warmup      please wait         During this phase  the lamp reaches the correct operating temperature  Once this process has been  completed  the instrument is ready to work   To know how to proceed  s
21. isplay and printed out  if a printer is connected   Expressed in concentration    Photometric    Filter A  near UV   Filter B  UV         0 001 O D    LCD 20 x 4 with back light  RS 232 for printer  Integrated    yes    Vital Diagnostics TURBI QUICK   USER S MANUAL       12  EC DECLARATION       DICHIARAZIONE DI CONFORMIT   CE  EC DECLARATION OF CONFORMITY    conforme all Allegato III della Direttiva 98 79 CE Dispositivi Medico Diagnostici In Vitro  conforme all Allegato Il della Direttiva 2006 42 CE Direttiva Macchine  according to Annex III of the Directive 98 79 CE In Vitro Diagnostic Medical Devices  according to Annex II of the Directive 2006 42 CE                         labbricante Vital Diagnostics S r l   manufacturer  Via Balzella 41 G 4  indirizzo    address 47100 FORLI  ITALIA  telefono fax  phone 0039 0543 721220 fax 0039 0543 796001  Lettore per analisi immunochimiche effettuate con metodo  Identificazione dei prodotti turbidimetrico  Product identification Reader for immunochemistry analyses performed by turbidimetric  method  Nome commerciale TURBI QUICK  Brand name  Numero i di catalogo PRD IMP XXX  Part number s  classificazione dei prodotti dispositivi diversi da quelli elencati nell Allegato Il della Direttiva 98 79 CE  product identification devices other then those mentioned in Annex Il of the Directive 98 79 EC       Si dichiara sotto la propria responsabilit   che  i dispositivi sopraelencati rispettano le disposizioni applicabili delle seguenti dirett
22. ive     Hereby we declare under our sole responsibility that  the above mentioned devices meet the applicable provisions of the following Directives     Direttiva 98 79 CE  Direttiva 2006 42 CE  Direttiva 2004 108 CE  Compatibilit   Elettromagnetica   Direttiva 2006 95 CE  Bassa Tensione   Direttiva 2002 96 CE e 2003 108 CE  RAEE   Direttiva 2002 95 CE  RoHs     Tutta la documentazione tecnica comprovante il rispetto dei requisiti applicabili delle Direttive elencate      conservata a cura del Fabbricante    All the technical documents required to demonstrate the conformity to the listed Directives   are kept by the Manufacturer       anno di immissione in commercio  year of introduction on the market    luogo e data    place and date Forl    10 03 2008    2005             firma Mar  yA yan    signature    timbro della Societa       Pag 23   24       TURBI QUICK     USER S MANUAL Vital Diagnostics       13  DISPOSAL AND RECYCLING    Herewith we declare that this instrument is subject to the European Directive 2002 96 EC  RAEE  Directive   Therefore the instrument must be disposed separately  not as urban waste and delivered to  the specific collection center in according to the Directive 2002 96 EC    The user can ask to the dealer the collection of the instrument if a new instrument is ordered to replace  the old one  On the instrument there is a label with the symbol shown in this page  The symbol means  that the instrument cannot be disposed as urban waste        Pag 24   
23. lem  call the  technical assistance         Black error    t   Check r  channel    land press button      In this case the operator can   a     check if there is a direct  source of light on the  instrument  Remove the  direct source of light from the  instrument   If this operation  won t solve  the problem  call the  technical assistance     Pag 19   24    TURBI QUICK     USER S MANUAL Vital Diagnostics       9  MAINTENANCE    Its simplicity and component parts mean that the Turbi Quick does not require special maintenance  The  most sensitive part is the lamp which can be easily replaced in case of damage  see further   Care and  attention are required above all regarding the cleanliness of the upper part  reading part  which must be  kept dry and clean  The soaking of the upper part  can damage the reading system  Do not clean the  upper plate with liquids or damp cloths     9 1 CLEANING INSTRUCTIONS    Dust can be removed using an ordinary vacuum cleaner  In case of cuvettes accidental broken and or  overthrow  clean the instrument with a sodium hypochlorite solution     9 2 LAMP REPLACEMENT    In each Turbi Quick package  is inserted a lamp as spare part   In case of blown lamp  follow these instructions to replace the lamp        Picture 1 Picture 2    Unscrew the 3 screw as shown in pictures 1 and  2  then remove the back panel of the photometer  as shown in picture 3    Now disconnect the cable of the lamp  white  cable  as shown in picture 4        Picture 3    Pag 
24. nt   s  samples  To pass directly to Sample mode  press the button for more than 2 seconds while the  instrument is in stand by mode  then on the display will appear     CRP Latex  Sample n  1  insert cuvette    and press button    Pag 15   24    TURBI QUICK     USER S MANUAL Vital Diagnostics       When this indication  Stand by mode  appears on the display the Operator must insert the cuvette into  the reading channel and then press rapidly  less than 2 seconds  the button  In this way a 5 seconds  countdown will start  when the T   zero seconds  on the display will appear the following message     CRP Latex  Sample n  1    add sample       Now the operator has 3 seconds to inject the starter reagent  During these three seconds  on the display  will appear     CRP Latex  Sample n  1  mixing       And immediately after   CRP Latex  Sample n  1  reading       CRP Latex  Sample n  1  Time  119     Once the reading time has expired  on the display will appear     CRP Latex  Sample n  1  conc  12 3 mg L  press the button       If the a printer is connected  the same information will be also printed out  Now the operator should press  the button rapidly  less than 2 seconds   in this way the analyzer will enter in the stand by mode  and is  ready for a new sample    CRP Latex  Sample n  2  insert cuvette    and press button    Pag 16   24    Vital Diagnostics    TURBI QUICK   USER S MANUAL       8  TROUBLESHOOTING    8 1 THE INSTRUMENT DOESN   T RECOGNIZE THE STARTER INJECTION   
25. slot  Select the Filter  Switch on the instrument    Once the Turbi Quick   is switched on  on the display you can see       Turbi Quick      Reader V X X      If the printer is connected to the instrument  the same information will be also printed out   If the Smart Card is not recognized by the instrument  or is not inserted in the slot  on the display will  appear       Invalid card       turn off and      replace the card      In this case  the instrument must be switched off  the Smart Card changed or inserted into the slot  and  then the instrument can be switched on again   Now  on the display will appear this message     XX Filter  A    Lot  9999 16 05 05  test counter  9999    and  if the printer is connected  the following information will be also printed out     Analysis   XX  Filter A  Lot nr    9999  Exp date   16 05 05  T counter  9999    Where  First Line  Name of the test   Second Line  Filter reguired   Third Line  Kit s lot number   Fourth Line   Kit s expiry date   Fifth Line  Test counter  This counter will decrease each time the user performs one test     Pag 8   24    Vital Diagnostics TURBI QUICK     USER S MANUAL       If the test counter is zero  on the display will appear      test counter   0     turn off     land replace card        This message will appear when the maximum number of test per kit have been achieved and kits  reagents are finished    Once will appear this message  switch off the instrument  open a new kit and insert into the slot  t
    