Home

MAX-5000 User`s Manual - Arizona Instrument LLC

image

Contents

1. 8 4 INSTRUMENT DESCRIPTION 10 A a td 11 sp Amt 5 g Wea eae nN PPR Pe OR EMER STON ee ENTE a SRN ESOT S R ECR IEE RCP PR RE nek aE Sere eR oe een 11 a O A rae ccc eater isto caste ET cet alunite sian titania tend odode eat 12 5 INSTRUMENT SETUP mm 13 Ole AMENA OL VEST CANDOR coc 14 e APO O Ne Goes 15 A nn hecuaistasseueanosdegeeeasesustetiaddssaniaeuanns 17 Dit COMECUINTa Pte A E o 17 Dd CONMECUN Gia COMPUERTA 20 5 6 Calibrate the Balance 50 gram weight ooooo ooccccccccccconononnnnnnnananonononononoconononanannn nos 21 A A saacaauasaauuasusactudsesiteddos E TE E ESA 23 6 MAIN MENU rm 24 6 1 Memon Stans Menu sersan o 24 Edatng Test PM Sn 23 A A ER 25 Cit II ON 26 ENCINA id 27 Sample Size OU OMS oi 30 Taro OPUOAS iii oia a 31 Result Display ODUONS ai ld 33 Lill SOMPEAS AMO Mid da ld 34 Linked TESPODUOINS 1d di id 35 CUT MEMON tai lis 37 COPY Memos duda ida 37 INSER T Memory tatiana dada 37 PRINEMEMON Stalin dada 38 6 2 Stored Data Menu Limited on 21 CFR Part 11 compliant units 39 Delete Stored Data daa dai 39 Analyze Stored Dalarnas 39 SEI CAD ada 40 VEW Ord Dala eE A daa aia 41 PACS d Dalarnas 41 6 3 CADAN Men U sosa T A T 42 Balance Calma Maicao 42 calibrate Balan CE aisla 42 View Calibration Report oooo
2. AZI Customer Service 800 528 7411 or 602 470 1414 e Answer appropriately and the system will DID THE SAMPLE BURN DEGRADE ask you to start the next test DURING THE TEST IF YOU ANSWER YES THE TEST TEMPERATURE WILL BE LOWERED A SERIES TESTS AT DIFFERENT e In this example the answer was YES In TEMPERATURES WILL BE RUN this case the parameterization will be done TO DETERMINE THE HIGHEST at a lower temperature Press start to begin ACCEPTABLE TEMPERATURE TO PRODUCE THE SHORTEST TEST TIME the next test TARGET TEMP 150 C PRESS START TO BEGIN TEST The display will recommend opening the lid if the oven temperature is too high Open the lid and press OK A SERIES TESTS AT DIFFERENT e Once the temperature in the chamber is TEMPERATURES WILL BE RUN correct press Start The system will lead Me High oven temperature you through the next test te Open lid for faster cooling We TARGET TEMP 150 C PRESS START TO BEGIN TEST Note the test temperature is lowered from TESTING WONTON 150 to 134 deg C CURRENT 0 0031 Moisture TIME 0 11 RATE 0 0031 Minute 29 49 AZI Customer Service 800 528 7411 or 602 470 1414 Mete aa SAMPLE WEIGHT 10 3707 q DID THE SAMPLE BURN DEGRADE Again the system prompts to see if the DURING THE TEST temperature is too high indicated by a burned or degraded sample In this case the IF YOU ANSWER YES THE TEST answer is NO TEMPERATURE WILL BE LOWERED eT
3. The oven has been at a stable temperature for at least the time period set CHAMBER STABILIZING LOW TEMP Oven is below temperature setpoint TEMPERATURES TEST TEMPERATURE 21608 TEST HISTART TEMP 25 C IDLE TEMPERATURE 25 C AZ Customer Service 800 528 7411 or 602 470 1414 Page 26 of 92 CHAMBER STABILIZING HIGH TEMP Oven is above temperature setpoint CHAMBER STABILIZING xx xx Oven temperature is within the correct range and the time remaining until specified time is displayed gt Ash Rate This is used for high test temperatures that will ignite a sample YOU MUST USE THIS TO PREVENT HEATER DAMAGE The Ash Rate value is a rate of moisture loss used to limit the rate of increase in oven temperature to slow ignition and prevent flaming The normal temperature setpoint is limited to keep the moisture loss at less than the set ash rate For example if an ash rate of 7 Minute is used the temperature increase is controlled to maintain a lt 7 Minute moisture loss during the test gt Start Temperature Control Prevents a test from starting if the chamber temperature is more than two degrees hotter or colder than the IDLE TEMPERATURE This feature can improve the standard deviation of results of some temperature sensitive materials The criteria used by the instrument to automatically end a test This feature makes it unnecessary for the operator to monitor the analyzer while it is performing a test
4. 37 5097 15 7233 0 0458 15 7306 15 7386 15 7461 15 532 5 2279 SEL ALL VIEW PRINT The screen will immediately display STORED DATA the number of samples mean standard deviation and relative standard deviation of the selected SAMPLES data set To the right is an example 5 of a calculated report This report can MEAN AVERAGE 15 7388 be printed by pressing PRINT Press the ESC key to return to the previous menu SD 0 0123 RSD 0 078 h All stored data can be selected to be deleted analyzed viewed or printed To select all stored data press SEL ALL AZI Customer Service 800 528 7411 or 602 470 1414 Page 40 of 92 View Stored Data This function is used to view a list of results for a selected test Move the selector bar to a particular test and LOT NUMBER press VIEW to display the data There PRODUCT ID DATE 27 Apr 2005 TIME OF DAY 09 43 FINAL RESULT 37 4863 5 are three 3 pages of data RESULT TYPE Moisture TEST MODE Rate LAST ACTUAL 37 4663 h START WEIGHT 8 0020 q END WEIGHT 5 0023 g TEST DURATION 1 30 A O O EAS STORED DATA WEIGHT STORED DATA WEIGHT START TEMP 28 C ASHING RATE 0 0 Minute TEST TEMP 94 C LINKED TO NEXT TEST No HI START TEMP 25 C LINK RESULT TYPE Individual PAN TARE Standard LINK USE FOR TOTAL Ho SAMPLE TARE Start After Us FOIL ID END RATE 0 0000 Minute FOIL WEIGHT 0 0000 q RATE TARGET 1 0000 Minute END RELIAB
5. All Cal Points verification of twelve 12 preset temperatures from 50 C to 575 C CUSTOM TEMP VERIFICATION POINTS CUSTOM POINT 1 100 C Four 4 custom temperatures may be CUSTOM POINT 2 140 C entered to verify only those CUSTOM POINT 3 200 C temperatures used in your processes CUSTOM POINT 4 250 C Entering a number above 575 C will produce a warning message on the display and set in the default highest temperature of 575 C Press OK to remove the warning from the display Calibration Module Setup This menu allows entry of the CALIBRATION MODULE SETUP Temperature Calibration Interface TCI data Serial numbers of the interface and RTD used for the calibration and the certification date of the TCI is entered here and saved for printing in the various reports MODULE 35555 PROBE bb CERT DATE 22 Feb 4005 AZI Customer Service 800 528 7411 or 602 470 1414 Page 44 of 92 Perform Calibration TEMPERATURE CALIBRATION This selection starts the automatic process of calibration and verification if selected of the MAX 5000 when the Temperature Calibration Interface TEMPERATURE CALIBRATION TAKES TCI is connected to the unit A SEVERAL HOURS TO COMPLETE detailed procedure begins on page 48 The calibration uses an NIST traceable MODE TO THE SERIAL PORT HENOVE THE device to measure the temperature at PAN AND PAN SUPPORT AND PLACE THE the sample level and compare that to CALIBR ATION PLATE IN
6. START TEMPERATURE CONTROL Off oe ee el ee l DEL Sample Name This menu allows an alpha numeric identification for the test to be established If the name is not blank when selected the entry may be deleted and new alpha numeric characters entered Note that the editing box will hold more characters than can be displayed on the Main Test Screen In addition the box will hold more narrow letters like I than wide letters like M Edit the value to any desired name or number to identify the product under test AZI Customer Service 800 528 7411 or 602 470 1414 Page 25 of 92 Temperatures This menu gives access to three temperature parameters an oven stable indication an ashing moisture rate control parameter and an enable switch used during a test gt Test Temperature The current temperature setpoint is displayed to the right of TEST TEMPERATURE Press ENT and the cursor will be positioned at the left Pg pe de Fl re 5 a CHAMBER STABILIZATION O Seconds E a ASH RATE 0 0 Minute desired test temperature EROS When finished press ENT to stop editing and save the new figure or ESC to A CA A A A E gt Test HiStart Temperature When active the HiStart temperature is used at the start of the test and the test continues at that temperature until the rate of moisture loss falls to half of its peak value The heater then changes to the Test temperature for the remainder of the test This shortens
7. e The keys at the bottom of the display allow the memory starts to be cut copied inserted and printed See page 37 and 38 for more information on CUT COPY INSERT 7 INSERT and PRINT memory starts AZ Customer Service 800 528 7411 or 602 470 1414 Page 24 of 92 The memory start parameters may be MEMORY START Memory Start edited by highlighting a memory start and pressing the ENT key From the eight 8 primary selections there are many other selections available to allow changing the data stored and altering the test characteristics e The keys at the bottom of the display aid the editing process gt DEL deletes the character to the left of the cursor gt CLEAR erases all characters in the edit field gt DEL deletes the character to the right of the cursor gt Displayed only when applicable changes the polarity pos neg of the number e fa number key is pressed without moving the cursor the current content will be erased and the new key press entered Press ENT to save the changes or ESC to abort the change and restore the original content 1 A 3 4 5 b f 1 CHAMBER STABILIZATION SAMPLE NAME TEMPERATURES ENDING CRITERIA SAMPLE SIZE OPTIONS TARE OPTIONS RESULT DISPLAY OPTIONS LIFT COMPENSATION LINKED TEST OPTIONS TEMPERATURES TEST TEMPERATURE 21608 TEST HISTART TEMP 245 C IDLE TEMPERATURE 25 C 0 Seconds ASH RATE 0 0 Minute
8. 0500 MINUTE 6 DRIFT 200 80 25 500 1 0 100 NO TIME 20 MINUTES 128 SAMPLE 128 160 025 025 80 1 0 100 NO RATE 0 1000 MINUTE 129 SAMPLE 129 160 025 025 80 1 0 100 NO RATE 0 1000 MINUTE 250 SAMPLE 250 160 025 025 80 1 0 100 NO PRED Page 0X NOTE If all locations have been entered a full printout of the Memory Starts will take five pages to print all 250 sets otherwise only those Memory Starts that have been entered will be printed The column headings are abbreviations for the following po Identification Name of the Memory Start Set an in the list indicates the set is linked to the next Memory Set C Sample Window Programmed Lift Test selection using Total or Individual test result Test calculation based on total weight Ending Criteria MODE Details for selected ending criteria Page 38 of 92 This function is used to work with data stored in the instrument from past tests Up to 1 000 test results and 100 graphs WEIGHT 31 Mar 2005 37 509 can be stored before the oldest is pushed _ WEIGHT 31 Mar 2005 37 5097 out overwritten To access this function TARTRATE 31 Mar 2005 15 7233 go to the Main Menu and select the 4 DRIFT el cris STORED DATA MENU On 21 CFR Part TARTRATE A deans 11 compliant units the DELETE is not TARTRATE 31 Mar 2005 15 7481 available TARTRATE 31 Mar 2005 15 7532 LACTOSE MONOH 31 Mar 2005 5 2279 LACTOSE MONOH 31 Mar 2005 5 1965 DELETE ANALY E SEL ALL VIEW PR
9. 528 7411 or 602 470 1414 65 Sample Foils Selecting this setup option will begin the process of weighing and storing the weights of 50 foils The foils are used in paint processes where the paint is sprayed onto the foil as the end product is being painted When placed into the MAX 5000 the weight of the foil will be subtracted from both the start weight and ending weight so that only the spray product will be considered in the moisture determination Pan tare will first tare the empty sample pan see pg 31 e From Main Menu select Sample SAMPLE FOILS Foils Select ADD or DELETE to 1 7 add or delete a foil 08 Jun 2005 0 2447 17 May 7005 0 1494 17 May 7005 0 1574 12 May 7005 0 1456 17 May 7005 0 1465 17 May 7005 0 1440 17 May 7005 0 1532 FEATURE LIST FULL TEMP RANGE PARAMETER EXPERT WEB SERVER A A ee A AZI Customer Service 800 528 7411 or 602 470 1414 Page 57 of 92 ge i ac Follow the prompts to open the lid place a foil on the sample pan and close the lid OPEN LID LOAD SAMPLE FOIL 0 00294 A A A O L ABORT e After the foil is weighed and the lid is opened the weight of the foil is given and a box appears so that a unique identifying name may be given to the foil NAME e Save the weight and ID so they will be available during sample test e Press SAVE to save the information or CANCEL to return to the SAMPLE FOIL menu WEIGHT 0 2447 CANCEL AZI Customer Service 800 528 74
10. 602 470 1414 for assistance AZ Customer Service 800 528 7411 or 602 470 1414 Page 85 of 92 No heat cont Door sensor failure Check display for LID OPEN message when open during a test call AZI at 800 528 7411 or 602 470 1414 for assistance Temperature setpoint too low Raise temperature above room ambient or standby temperature see pages 26 amp 70 No power to Check switch position Instrument confirm power cord seating remove and test fuse see page 15 test the outlet with a No weight reading Balance not initialized Turn power off wait 5 seconds then turn power on wait 15 seconds Call AZI at 800 528 7411 or 602 470 1414 for assistance Erratic results Poor AC ground Ensure instrument is grounded from case ground to earth ground Call AZI at 800 528 7411 or 602 470 1414 Dirt in chamber Clean unit see pages 9 to 83 Dirt in pan support See page 79 or call AZI at 800 528 7411 or 602 470 1414 for assistance Display erratic Unit cold near freezing Allow to warm up UNDERLOAD when Sample pan or pan support not Install pan and or support START pressed in place see page 14 Quit test and press START again Check tare options pan support 79 to 83 and following Instrument won t tare Test chamber temperature fluctuates changes AZI Customer Service 800 528 7411 or 602 470 1414 Page 86 of 92 CONSUMABLES 800 0037 800 0067 800 0068 990 0003 990 0008 990 001
11. 9 rU ent LLCO USER MANUAL COMPUTRAC MAX 5000 MOISTURE ANALYZER OPERATION MANUAL Firmware Rev 2 21 April 2006 ARIZONA INSTRUMENT LLC 3375 N Delaware Street Chandler AZ 85225 USA 800 528 7411 602 470 1414 f 602 281 1745 WWW azic com Email General azi azic com International inth azic com Customer Service support azic com ARIZONA INSTRUMENT LLC 3375 N Delaware St Chandler AZ 85225 800 528 7411 sales azic com www azic com MADE IN THE USA AZI P N 700 0031 G2 Last update April 2006 Computrac MAX 5000 PROPRIETARY RIGHTS NOTICE The material in this manual contains valuable information developed by Arizona Instrument LLC for use with the Computrac line of moisture analyzers No part of this manual can be reproduced or transmitted in any form or by any means electronic mechanical or otherwise This includes photocopying and recording or in connection with any information storage or retrieval system without the express written permission of Arizona Instrument LLC ALL RIGHTS RESERVED Copyright 2006 Arizona Instrument LLC AZI Customer Service 800 528 7411 or 602 470 1414 Page 2 of 92 TABLE OF CONTENTS 1 WARNINGS AND SAFETY INFORMATION 5 2 CHANGES 2 02 nn nn nn en nn nnn nn nnn n nnn n nnn n nen n anne nneneennens 7 3 PACKING AND SHIPPING INSTRUCTIONS
12. bar to AUTO FORM FEED Pressing ENT toggles between YES and NO Selection will depend on the printer used Try Auto Form Feed YES for most printers If the printer seems to waste paper switch it to NO e Press ESC to return to the REPORT SETUP Menu Select REPORT CONTROL OPTIONS press ENT e Press ENT key to toggle between YES and NO Select YES if you want to send test results and reports to a printer or REPORT START Test Start computer and NO to prevent sending test results as outputs REPORT INTERVAL 10 Seconds OUTPUT PORT RS 232 REPORT ENABLED Pres INCLUDE GRAPH Ho AZI Customer Service 800 528 7411 or 602 470 1414 Page 18 of 92 Move the selection box down to REPORT CONTROL OPTIONS REPORT START and press ere Select between Test Start begins to send data at the start of the test gt First Prediction begins to REPORT START Test Start send data as soon as the instrument makes the first REPORT INTERWAL Test End prediction or gt Test End waits to send data INCLUDE GRAPH No until after the test is over and then sends only the final results of the test Press ENT to select your choice Move the selection box down to REPORT INTERVAL and press ENT Select an interval that will provide the test progress report you need and press ENT again to REPORT ENABLED Yes select your choice Move the selection box down to INCLUDE GRAPH Pressing ENT will toggle between YES and N
13. code permits access to memory start editing and setup menus If the master key code is forgotten call AZI Customer Service at 800 528 7411 or 602 470 1414 Calibration Key A calibration key code permits access to calibration menus This menu provides the necessary controls to interface with a local area ETHERNET SETUP network A Web Server option can be purchased for the MAX 5000 The NAME MAX 5000 MAX 5000 can be accessed from the PASSWORD intranet by a PC Generally the MAX OBTAIN IP ADDRESS Manual 5000 cannot be viewed on the Internet because of company firewalls but it is IP ADDRESS 192 168 002 254 possible and requires a registered IP SUBNET MASK 255 235 295 000 address GATEWAY 000 000 000 000 See the Web server optional feature MAC ADDRESS 00 08 50 00 00 27 section on page 65 for details on the STATUS Initialized Web Server Call AZI customer service for more information and purchasing questions AZ Customer Service 800 528 7411 or 602 470 1414 Page 55 of 92 restrict access to the instrument over the ER intranet When using IE enter NAME for the EE User name and enter PASSWORD for the py a Password to gain access to the instrument The NAME and PASSWORD can be used to APT 103 calrpeyiew htm User name E e Remember my password This can be set to Automatic Manual or Ethernet Off e Automatic Sets everything automatically A DHCP server is required to obtain
14. displayed as a AETA h sae percentage of weight loss lo Ash Solids Test results are w LOI displayed as a percentage of weight remaining The instrument is optimized to run in Solids for samples gt 99 5 volatiles Dry Weight Test results are displayed as a dry weight basis Dry Weight Winiria Weinac WeimaL 100 Note that the dry weight value ratio can exceed 100 and often does in applications such as soil and forest fuels analysis Ash Test results are displayed as a percentage of weight remaining LOI Test results are displayed as a percentage of weight loss ASS FAIL Toggles ON or OFF ein eerie acceptable limit for the test A final test value below this figure will cause a FAIL indication to appear on the DISPLAY RESULT AS Moisture display and if selected in setup on printed reports PASS FAIL On High Limit sets the highest acceptable limit for the moisture test A final test value above this figure will cause the FAIL indication to appear on the display and if selected in setup on printed reports 4 4 4 v U LOW LIMIT 0 0000 HIGH LIMIT 0 0000 AZI Customer Service 800 528 7411 or 602 470 1414 Page 33 of 92 Lift Compensation ii r ea is a feature found only in Computrac Instruments Itis included because of heated air currents in the oven chamber that LIFT COMPENSA TION 100 apply lift to the sample pan This causes a
15. false apparent weight loss If not compensated for this error leads to erroneous moisture results too high The amount of lift generated by the hot sample varies from sample to sample and increases with the test temperature gt n some cases the lift is approximately 5 milligrams and is small enough that less sensitive instruments do not provide compensation The lift must be compensated for when testing low moisture samples For example some plastic resins are controlled to a moisture content of 0 01 so for a 30 gram sample the weight change is only 3 milligrams Similarly some lyophilized drugs are so valuable that the sample weighs only 150 milligrams At 2 moisture the weight change is only 3 milligrams gt The MAX 5000 has an adjustable lift compensation based on the Memory Start Each instrument is characterized for lift produced under standard conditions and for the way that lift changes for various test temperatures The lift compensation can be adjusted by setting the Memory Start lift to a value other than 100 Use 100 for all samples unless it is found by experiment that another value is more appropriate Most users will not need to adjust this value Contact your AZI technical sales representative at 800 528 7411 for help on determining the proper value for difficult applications gt Low Moisture CF For extremely low moisture products a compensation factor can be used to align MAX 5000 results to the reference st
16. for assistance VVVV WV AZI Customer Service 800 528 7411 or 602 470 1414 Page 77 of 92 10 5 Retained Samples The weight test and the sodium tartrate test are usually sufficient to satisfy industry guidelines for quality control in moisture measurement Some additional checks can be performed using replicate testing of retained samples Retained samples when stored under airtight conditions should maintain moisture levels for long periods Routine testing of these samples should produce the same results day after day Retained samples further verify the unit s operation on the customer s own product If hygroscopic materials are used the values can change with time and repeated opening of the container e Let the system warm up for 30 minutes e Set the MAX 5000 to the optimum test temperature for the retained sample or use the ID number of a test that has been stored in memory e Place a clean aluminum sample pan onto the pan support in the test chamber e Close the lid and press START e When the OPEN LID display appears GENTLY add the optimum amount of retained sample evenly across the sample pan e Close the lid and allow the test to continue e Record the result in the quality control log e The results for the retained sample should fall within the established acceptable range of results for that sample If results are out of range repeat the test to verify results and or perform standard reference testing to cer
17. is available to remind you when scheduled calibrations are due The default factory settings are ENABLE No TIME 00 00 INTERVAL 0 Weeks To turn the prompt on off or adjust the interval go to the Main Test Screen and select in order MENU CALIBRATION MENU BALANCE CALIBRATION and CALIBRATION SETUP REMINDER gt Enable Toggle to YES to enable the prompt or NO to prevent the prompt from appearing gt Time The time the prompt is to appear usually the start of the shift using a 24 hour format INTERVAL 1 Weeks gt Interval The interval between calibrations in weeks If the interval is set to 00 the reminder will occur daily ENABLE Yes TIME OF 30 AZI Customer Service 800 528 7411 or 602 470 1414 Page 43 of 92 Temperature Calibration The MAX 5000 temperature calibration routines will calibrate and verify the accuracy of the oven temperature throughout the usable range The oven VERIFICATION SETUP is calibrated at twelve 12 preset _ CALIBRATION MODULE SETUP increments and verified at PERFORM CALIBRATION temperatures that are user selectable 4 PERFORM VERIFICA TION VIEW CALIBRATION REPORT CALIBRA MON REMINDER SETUP Verification Setup ee a allows the verification of the temperature calibration to be enabled or disabled and selection of which AUTO VERIFY AFTER CAL Yes temperatures are to be verified Selecting ALL CAL POINTS will cause POINTS TO VERIFY
18. minutes before beginning the calibration e When the verification is complete the screen will display gt The verification temperatures and gt The pass or fail condition at each calibration temperature To print out the calibration report highlight VIEW CAL REPORT and press ENT Press PRINT Calibration is complete Open the lid and wait for it to cool Turn off the instrument disconnect the TCI and remove the temperature calibration disk fixture Carefully reinstall the ceramic balance cover and pan support AZ Customer Service 800 528 7411 or 602 470 1414 Page 49 of 92 This selection is the start of an oven cleaning process that burns off most residue build up in the oven Use a fume hood to remove possible toxic emissions from the work place Similar to the self cleaning mode in a home oven the process turns the heating elements on to reach an oven temperature of 550 C in less than 22 minutes The oven temperature will be maintained for a total oven REMOWE THE PAN AND PAN SUPPORT cleaning time of 45 minutes from FROM THE OWEN AND CLOSE THE LID start to finish The heat generated makes the top of the instrument hotter than normal Extreme caution should be used to prevent injury until the instrument has cooled down at the conclusion of the cleaning cycle AO A AO CANCEL A During the cleaning cycle there may be smoke odor and possible toxic emissions produced by the instrument This is normal as the resi
19. results between 50 C and MAXS5000 TCI DIFFERENCE 575 C degrees See page 44 for 49 8 404 0 7 Pass information on how set up the 100 0 99 2 0 8 Pass temperature verification 150 0 149 0 1 0 Pass 200 0 200 0 0 0 Pass 250 0 250 2 0 2 Pass 300 0 299 9 0 1 Pass 330 1 390 0 0 1 Pass 400 0 399 9 0 1 Pass 450 0 450 0 0 0 Pass 499 6 499 8 0 0 Pass MAXS5000 DIFFERENCE 549 7 0 4 Pass 574 5 0 6 Pass AZI Customer Service 800 528 7411 or 602 470 1414 Page 46 of 92 Calibration Reminder Setup A TEMPERATURE CALIBRATION REMINDER is available to remind you when scheduled calibrations are due The default factory settings are ENABLE No TIME 00 00 INTERVAL 0 Weeks e To turn the prompt on off or adjust the interval go to the Main Test Screen and select in order MENU CALIBRATION MENU TEMPERATURE CALIBRATION and CALIBRATION REMINDER SETUP gt Enable Toggle to YES to enable the prompt or NO to CALIBRATION REMINDER SETUP prevent the prompt from appearing gt Time The time the prompt is to appear usually the start of the shift using a 24 hour format gt Interval The interval between calibrations in weeks If the interval is set to 00 the reminder will occur daily AZI Customer Service 800 528 7411 or 602 470 1414 INTERVAL 4 Weeks Page 47 of 92 e Ensure the instrument is turned ON gt A Temperature Calibration Kit AZI P N Y990 0193 with 9 VDC 100 m
20. serial port The content of each message can be selected from the REPORT SETUP menu which is described on page 52 e The MAX 5000 is equipped with RS 232 communications available at a 9 pin tl standard serial communications socket at A oe the rear of the instrument The standard communications parameters used are 115 2 kbits s no parity eight data bits and one stop bit Refer to your computer communications program manual for instructions on configuring the computer to read this data format e The MAX 5000 does not test to see if the computer is actually connected or ready to receive data lt does not respond to a request from the computer for status information e The MAX 5000 can send data to both the RS 232 port and the parallel printer port at the same time However the graph that can be printed by the printer is not in a format that can be received by the computer Therefore the graph will not be sent to the RS 232 port LINE If you have any questions about connecting a printer or computer to your MAX 5000 call AZI Customer Service at 800 528 7411 or 602 470 1414 AZI Customer Service 800 528 7411 or 602 470 1414 Page 20 of 92 NOTE The MAX 5000 must be powered on for 30 minutes to allow the balance to stabilize before beginning the balance calibration The unit should not be moved jarred or otherwise disturbed in any way during the calibration period A vibration free table will aid the calibration process If vibr
21. size and test parameters Results from these repeated tests are tabulated to calculate the mean or average and the standard deviation S D Analysis of the mean and standard deviation and their comparison with reference method results determines if the proper test temperature has been used If mean values do not compare well or if the standard deviation is too large test at another temperature along the plateau AZI Customer Service 800 528 7411 or 602 470 1414 Page 70 of 92 8 3 Selection of Ending Criteria The decision of which ending criteria to use is application dependent Best results are typically given with PREDICTION or RATE rather than TIME Experiment with your particular sample to determine which criterion is best for your product See Ending Criteria on page 27 For PREDICTION the MAX 5000 uses a sample dependent test to automatically end a test The test called the reliability criterion is equal to the actual moisture loss at a given moment divided by the end point predicted by the microprocessor times 100 The ending ratio can be set to a value from 00 to 99 depending on the user s desire to balance speed and accuracy The smaller the number the sooner each test will end A value of 99 means the actual moisture loss must exceed 99 of the predicted end point For RATE the MAX 5000 ends the test when the sample weight loss rate has fallen below a preset percentage value This is useful for samples that do not conf
22. tartrate test in Section 10 3 multiple times to determine the instruments repeatability This is a more critical verification of the instrument s performance Mean Standard Deviation and Relative Standard Deviation are calculated using the following formulas Mean total of results divided by number of results 2 Standard Deviation E n Relative Standard Deviation expressed as percentage Standard Deviation divided by Mean multiplied by 100 e Perform the steps in the sodium tartrate test a total of five 5 times gt After all five tests are complete press MENU and select the STORED DATA MENU gt Press ENT to select tests gt Arrow up or down to select the five tartrate tests and press ANALYZE gt The mean of the results should be within 0 1 of the value on the bottle label e lf the result is out of tolerance the most common causes of such errors are bent or dirty sample pans or debris that may have fallen into the balance shaft Switch the unit OFF Remove the pan support and the balance cover Carefully clean the instrument See page 79 Clean the oven and check the temperature sensor See page 81 If a temperature calibration kit AZI P N Y990 0193 is available calibrate the system See page 44 e Repeat the tartrate test again using a clean sample pan for each of the five samples and analyze the results e lf the mean is still out of tolerance call AZI customer service at 800 528 7411 or 602 470 1414
23. technicians only Any unauthorized inspection and or maintenance of the MAX 5000 will make the warranty null and void Lithium Coin Cell Battery The MAX 5000 contains a lithium battery to keep power on the clock when the unit is turned off CAUTION The battery may explode if mistreated Do not recharge disassemble or dispose of in fire Prevent excess heat build up around the analyzer When setting up the analyzer leave enough space to prevent heat from building up and to keep your analyzer from overheating Leave 20 cm about 8 in around the analyzer and 1 m about 3 ft above the unit Do not put any flammable substances on under or near the analyzer because the area around the unit will heat up especially during the self cleaning cycle Use extreme care when removing the sample from the chamber Although the ceramic heating chamber cools rapidly the sample itself the sample pan and the heating element can still be extremely hot Page 5 of 92 HAZARDS FOR PERSONS OR EQUIPMENT POSED BY USING SPECIFIC SAMPLES Flammables explosives substances that contain flammables explosives solvents and or substances that release flammable or explosive gases or vapors during the drying process e The user shall be liable and responsible for any damage that arises in connection with the MAX 5000 and potentially harmful substances e Substances containing toxic caustic or corrosive substances may be analyzed with the
24. 0 990 0196 Sodium Tartrate 500 gram bottle Ultrapure Water 1L Clear Polypropylene Filter Paper 100 PCS Waffle Bottom Sample Pans 100 PCS Flat Bottom Sample Pans 100 PCS Thick Sample Pan 5 PCS SPARE PARTS PARE PARTS 170 0014 190 1001 200 0002 250 0001 300 0470 300 0472 690 0003 690 0004 690 0025 690 0012 700 0031 Y990 0158 OPTIONS Y990 0193 6000 1044 990 0093 Y990 0098 9600 0019 980 0011 Y990 0197 Y990 0198 Fan 8 Amp Fuse Power Cord 110V Insulator Seal Pan Support Assembly Screen Vent Cover 3g Weight 5g Weight 50g Weight Tweezers MAX 5000 User s Manual MAX 5000 Accessory Kit Temperature Calibration Interface Kit with Adaptor 9 VDC output 100ma 3 5mm male plug positive tip input is 110V Null MODEM Cable Carrying Case Printer Kit Parallel 110 V 220V to 110V Power Converter Vibration Table Parameter Expert Web Server AZ Customer Service 800 528 7411 or 602 470 1414 Page 87 of 92 14 MAX 5000 SPECIFICATIONS Specification Size 18 5 length x 12 75 width x 9 5 height 470 mm length x 323 mm width x 240 mm height Height varies slightly with adjustable feet for leveling purposes Weight Casework Fused on off switch is part of an EMI power entry module Operating 0 40 C 50 humidity Environment 0 31 C 80 humidity Non condensing non explosive Heater Calibration Menu Driven NIST Traceable using an accessory Temperature Calibration In
25. 0 unit call AZI Customer Service or your technical sales representative at 800 528 7411 or 602 470 1414 AZ Customer Service 800 528 7411 or 602 470 1414 Page 91 of 92 TRADEMARK AND COPYRIGHT PROTECTION Computrac Arizona Instrument and the stylized AZI are all registered trademarks of Arizona Instrument LLC MAX 5000 is a trademark of Arizona Instrument LLC Instrument firmware is copyright protected The MAX 5000 uses the Graphics Interchange Format as part of its Ethernet interface The Graphics Interchange Format is the Copyright property of CompuServe Incorporated GIF sm is a Service Mark property of CompuServe Incorporated Copyright 2003 2006 Arizona Instrument LLC All Rights Reserved Arizona Instrument LLC Computrac MAX 5000 Moisture Analyzer Part Number 700 0031 Revision G2 April 2006 If you have any questions regarding the operation of this instrument please call or e mail for assistance Arizona Instrument LLC 33 75 N Delaware St Chandler AZ 85225 800 528 7411 602 470 1414 Fax 480 804 0656 http www azic com email azi azic com General inth azic com International support azic com Customer Support AZ Customer Service 800 528 7411 or 602 470 1414 Page 92 of 92
26. 00 graphs gt Sent directly to a personal computer and gt Printed on many graphics capable printers with a printed graphical representation of the drying process Calculation of mean standard deviation and relative standard deviation for up to 1 000 stored test results gt Calculations and data storage is not available on 21 CFR Part 11 compliant instruments Use a printer to capture data from those instruments AZI Customer Service 800 528 7411 or 602 470 1414 Page 10 of 92 4 1 Hardware The MAX 5000 consists of A keypad used for most selection and entry functions An input power module set to 100 120 VAC A ceramic convection air oven A precision balance weighing mechanism with a resolution of 0 0001 gram and a full scale range of 100 grams A backlit color graphic liquid crystal display e A microprocessor that controls the test process data handling communications and performs all necessary calculations e A small fan in the lower chassis to maintain the electronics at a stable low temperature and e Connectors on the back panel for an Ethernet connection parallel printer and serial communications devices Other connections are provided that will be implemented in later revisions to the instrument The MAX 5000 is designed to remain powered up at all times The balance requires a lengthy warm up period The first test of the day can be made sooner and with more accuracy if the unit remains powered over nig
27. 05 lt 0 003 Standard Deviation Repeatability Ashing Contact AZI for specific sample repeatability Moisture Level gt 10 lt 0 5 Standard Deviation gt 0 1 lt 5 Relative Standard Deviation gt 0 01 lt 0 005 Standard Deviation gt 0 005 lt 0 003 Standard Deviation Contact AZI for specific sample repeatability Self diagnostics Built in Hardware and Software diagnostics software Options Temperature calibration interface kit Printer kit Parameter Expert Web Server Calibration Weights Factory calibration weights Class 1 Field calibration and test weights Class F Warranty Domestic U S 2 years factory parts and labor Inter unit Correlation AZI Customer Service 800 528 7411 or 602 470 1414 Page 90 of 92 15 WARRANTY Arizona Instrument LLC warrants the Computrac MAX 5000 to be free from defects in materials or workmanship for two 2 years from the date of purchase for those units sold in the USA All other sales normally carry a one 1 year warranty AZI will repair or replace at its option products that AZI determines to be defective during the warranty period All defective parts replaced by AZI become the property of AZI Replacement parts are warranted for the remaining portion of the effective warranty period This warranty does not apply to expendable or maintenance items such as pans and pan supports The above warranty does not extend to any product which has been subjected to misuse abuse
28. 11 or 602 470 1414 Page 58 of 92 166 Parameter Expert Optional Feature A Parameter Expert option can be PARAMETER EXPERT MENU purchased for the MAX 5000 The Parameter Expert PE assists a user through the development of a memory 1 DEVELOP FROM SCRATCH start PE queries the user about his 2 NUDGE TEMP SPEED REPEAT ABILITY reference method sets up a memory start and walks the user through a series of tests to determine the highest acceptable test temperature To speed development tests are stopped if it looks like they will not end in a reasonable time a higher test temperature will be tried instead In the end PE programs the memory start which produces a fast and accurate test PE works on most samples up to 300 C For ashing PE sets up a memory start and runs a single test to determine the proper ending rate if possible Ashing may require a more complex parameter development Typical PE development time is lt 2 hours for products lt 25 moisture PE can be exited and reentered and PE remembers the previous tests so development will continue were it was previously stopped Develop From Scratch The Parameter Expert is informative and guides the user through the steps mata memory start name REF METHOD TEMP Enter the temperature used by the reference method If temp gt 300 then PE NAME NAME assumes ashing mode and sets up a memory start and runs a single REF METHOD TEMP 125 C test to dete
29. 1414 Page 64 of 92 7 WEB SERVER OPTIONAL FEATURE A Web Server can be purchased for the MAX 5000 MS Internet Explorer V6 is used to view a calibration report view the LCD screen save load memory starts to from disc and save test results to disc To access the instrument with IE type the IP address into the IE address field 192 168 2 254 for example The IP address is the one that is assigned in the Ethernet Setup section on page 55 Memory starts can be retrieved from one instrument and saved to another Test data is saved in a comma delimited format which can be read by MS Excel NOTE Results data cannot be modified or deleted through the Web Server Data can only be deleted at the instrument See the Ethernet Setup section on page 55 to setup the Web Server 7 1 Calibration Report The calibration report is viewed by selecting CALIBRATION REPORT in hui MAX 5000 471 Preferred Customer SN 500039 CALIBRATION REPORT h f r Thi r i LCD VIEW MAX 5000 CALIBRATION REPORT ad tne ett ba eport a MEMORY STARTS COMPANY MAME AZI Preferred Customer combination of the balance TEST RESULTS MAX SOOO de 500039 and temperature calibration FACTORY CAL DUE 01 Jan 2006 report The report can be downloaded to disc as an HTML file BALANCE CALIBRATION 18 Jul 2005 TEMP CALIBRATION 00 13 0000 TEMP VERIFICATION 00 1an 0000 CAL MODULE 0 CAL PROBE o MODULE CERT DATE OO Jan 0000 VERIFY RESU
30. 28 7411 or 602 470 1414 Page 53 of 92 This menu brings up the two DATA FORMAT SETUP categories of data one display oriented and the other report oriented EC DISPLAY Off LOT NUMBER Off EC Display This selection controls the displaying of the word RELIABILITY on the Main Test screen for PREDICT ENDING CRITERIA Displaying this value is useful during parameter development to aid in selecting the best ending criteria It may also help the operator to determine the remaining test time if the ending criteria target is Known Lot Number iaa uri serbia starting a material test brings up a screen labeled ENTER LOT NUMBER AND PRODUCT ID These alpha numeric entries may be used to MUA e insert the lot designation and unique product identification into reports Not all industries or processes will use them so they may be enabled only when needed LOT NUMBER Alpha numeric code to identify the product lot number PRODUCT ID Alpha numeric code DEL ee CLEAR DEL used by some companies to identify specific products PRODUCT ID 1ADX AZI Customer Service 800 528 7411 or 602 470 1414 Page 54 of 92 ACCESS KEY SETUP This menu provides a means of preventing unauthorized access to memory start editing calibration and setup menus Once the correct key has been entered the menus remain ACTIVATE KEYS No unlocked until the instrument is powered off MASTER KEY CAL KEY Master Key The master key
31. 6 OB TAINMP ADDRESS cuidaba 56 VACADDRE SO nto obran 56 AS Sane rea REC me ene A RN EE me eee ne cee 56 Special Feat Syria IR REC en A TAAI 57 Oos SAMPE FOS AAA EE E EE NS 57 RAP PP A Ne eR aS 57 6 6 Parameter Expert Optional Feature oo occcccnnnooononcncconononancnonccononannnnnnoncconannnn cons 59 Devclo pP From OCIO praia 59 Optimize Temp Speed Repeatability ooooocccccccccnnnnnnnnnnnananannnnonocncccononananann nos 63 7 WEB SERVER OPTIONAL FEATURE 65 C SAMDEAUOR ATSC DOI aria EAA 65 aza ECO VICW e a e Lo 66 Toe Meno tati dida 66 4A A A na Stance itactt etocaa AE 67 8 PERFORMANCE ENHANCEMENT 68 Sala AINTFOdUCION lodos 68 8 2 Determining Proper Test Temperature coccnnnncucuuoooonnccnccococononnnannnnnnnnnnnnnononocononananannnons 70 8 3 Selection of Ending Criteria occccccnnnnnnononananananonononononananann ono nonnnnononocononnnannnnn ns 71 9 SAMPLE PREPARATION TESTING AND ANALYSIS 71 9 1 Preparing Granular and Powdery Samples occccccccnnnnnnnonanananananononococononanananonos 71 9 2 Preparing and Testing Liquid Samples oocnnnnooooouoncncnocococonononnnnanananononcnnnncconononanannnnos T2 9 3 Selecting the Optimum Sample Size occnnnnnnnucouananancnon
32. 70 1414 to order Occasionally some samples cannot be absorbed into the filter paper At the end of the test examination of the sample pan shows the sample sitting high above the filter paper with a thick crust The crust may be singed from the heat or even show bubbling where vapor has been trapped beneath the surface SOLIDS results will be high or MOISTURE results will be low Samples such as these are treated differently After the sample has been added to the sample pan remove the pan from the test chamber and place it onto a flat surface Grasp part of the filter paper tweezers are suggested and flip it over so that the sample lies between the pan and the paper Squash the sample with a flat bottom glass or jar to a uniform layer Return the sample pan to the test chamber and close the lid The upper filter paper surface absorbs moisture from the sample and minimizes bubbling and spattering to produce more accurate reproducible results AZI Customer Service 800 528 7411 or 602 470 1414 Page 72 of 92 9 3 Selecting the Optimum Sample Size Each particular sample will have its own optimum sample size Best results will be obtained if the sample consists of a single layer of particles as long as this is sufficient to develop a weight loss of 10 milligrams or more A rule of thumb is the lower the moisture the larger the sample Low moisture samples less than 0 10 such as chemicals or plastics require larger amounts the instrum
33. A 3 5 mm male plug positive tip adaptor must be used to perform the calibration gt A graphics capable printer may be connected to the MAX 5000 if a printed report is required e Gently remove the pan support and ceramic balance cover e Install the temperature calibration fixture so the RTD mount and wires are as shown and gently close the lid e Plug the disk interface cable into the matching connector on the TCI box Plug the RS 232 interface cable into the matching connector on the TCI box and the serial port on the back of the unit Plug the 9VDC adaptor into the TCI box and an appropriate 110V AC outlet a j e From the Main Test Screen select in order MENU CALIBRATION MENU TEMPERATURE CALIBRATION e Highlight ENTER FIXTURE DATA and press ENT Enter the appropriate data for the TCI serial number disk serial number and calibration certification data gt The interface serial number and certification date are located on the TCI box label AZI Customer Service 800 528 7411 or 602 470 1414 Page 48 of 92 gt The disk number is located on the RTD fixture and should match the probe number on the TCI box If the RTD number does not match the serial number on the TCI box call customer service at 800 528 7411 or 602 470 1414 for assistance e Return to the TEMPERATURE CALIBRATION MENU highlight VERIFICATION SETUP and press the ENT key gt Select AUTO VERIFY AFTER CAL YES to automatically run the verifi
34. BALANCE Press GO Follow the prompts to complete the calibration The first prompt is OPEN LID and PLACE CLEAN PAN ON TRAY SUPPORT After placing a clean pan on the pan support and closing the LID the screen will display CALIBRATING PLEASE WAIT As soon as the tare is finished the next prompt OPEN LID and LOAD 50G WEIGHT AND WAIT will appear As the lid is opened the screen prompts LOAD 50G WEIGHT AND WAIT Avoid excess pressure on the scale Use the tweezers included with the MAX 5000 to gently place the weight on the pan AZI Customer Service 800 528 7411 or 602 470 1414 Page 42 of 92 As soon as the instrument detects weight on the balance the prompt CLOSE LID appears With the lid closed the prompt CALIBRATING PLEASE WAIT appears At the conclusion of the calibration the OPEN LID REMOVE WEIGHT prompt appears with a PASS FAIL indication of the calibration success If the result is PASS open the lid remove the weight and press PROCEED to continue If the result is FAIL the calibration was unsuccessful Rerun the calibration If the calibration fails again contact AZI Customer Service at 800 528 7411 or 602 470 1414 Press the ESC key three 3 times to return to the Main Test Manu View Calibration Report When VIEW CALIBRATION REPORT is selected the last balance calibration date will be displayed Calibration Reminder Setup A prompt
35. ENT key Press PAGE 2 or PAGE 1 to switch between pages REPORT ITEMS TO PRINT REPORT ITEMS TO PRINT SAMPLE NAME lt START WT TARGET RELIABILITY TARGET LOT NUMBER J END WEIGHT J LIFT PRODUCT ID lt TEST DURATION LOW MOISTURE CF DATE lt START TEMP J ASHING RATE J TIME OF DAY J TEST TEMP J LINK OPTIONS FINAL RESULT J HI START TEMP FOIL ID Y RESULT TYPE Y PAN TARE FOIL WEIGHT Y PASS FAIL J SAMPLE TARE PURITY lt TEST MODE END RATE Y LAST ACTUAL lt RATE TARGET Y START WEIGHT END RELIABILITY A A G ee eee EAS Add Edit Company Name This selection provides a means of customizing reports to show the name of the company test location or other alpha numeric identifiers This selection is used to enter the correct date and time 24 hour format into the instrument s clock system The internal battery saves the clock value when the power is Off Use this function to change the time at initial setup or when changing for daylight savings time The clock in the MAX 5000 uses a small battery to maintain correct time and date If the clock resets whenever the power is off the internal backup battery may need to be replaced If this happens call AZI Customer Service at 800 528 7411 or 602 470 1414 A battery that fails does not affect the memory of the MAX 5000 since the memory does not depend on a battery backup during power off conditions TIME 08 59 DATE 27 May 2005 AZ Customer Service 800 5
36. ILITY 0 0 RELIABILITY TARGET 390 h LIFT 100 i A A O EE CEN A O O GRAPH PAGE 1 Print Stored Data Stored data can be printed to a parallel printer attached to the MAX 5000 See page 17 That same data can also be routed to a computer See page 20 To print stored data press PRINT on the STORED DATA MENU Only report items that have been selected in the REPORT ITEMS TO PRINT menu will print See page 53 AZI Customer Service 800 528 7411 or 602 470 1414 Page 41 of 92 6 3 Calibration Menu Balance calibration is explained and performed as part of instrument setup see page 21 The sequence is repeated here for continuity The balance should be calibrated Thirty 30 minutes or longer after turning the instrument on after it has been turned off for several hours or stored After the instrument is moved Whenever a problem is suspected and or As often as required by your quality policy NOTE Best results are obtained if the MAX 5000 has been powered on for 30 minutes to allow the balance to stabilize The unit should not be moved jarred or otherwise disturbed in any way during the calibration period A vibration free table will aid the calibration process lf vibration is excessive the instrument will stop the calibration process and indicate the reason Calibrate Balance From the Main Test Screen press Menu Select the CALIBRATION MENU Select BALANCE CALIBRATION Select CALIBRATE
37. INT This function is used to delete individual results and is non reversible Normally results are deleted to make management of the data easier If the test results need to be saved they should first be printed or stored in a computer file Moving the selector bar to a particular test and pressing DELETE will erase the record so be careful If the list is empty the screen will show NO MENU ITEMS Mean Standard Deviation and Relative Standard Deviation are calculated using the following formulas available on many hand calculators Mean total of results divided by number of results 7 2 Standard Deviation E n Relative Standard Deviation expressed as percentage Standard Deviation divided by Mean multiplied by 100 AZ Customer Service 800 528 7411 or 602 470 1414 Page 39 of 92 To analyze a set of data Use the arrow keys to highlight the data items to be included in the analysis and press the ENT key EGET col Notice that each selected item is _ WEIGHT 31 Mar 7005 marked with a triangular symbol at the TARTRATE 31 Mar 2005 left of the data item If one has been 4 DRIFT 31 Mar 2005 designated in error move the selector i TARTRATE E e ar 4005 bar to it and press the ENT key to _ TARTRATE 44 Mar 2005 deselect and unmark it When all TARTRATE 31 Mar 2005 desired items are selected press LACTOSE MONOH 31 Mar 2005 ANALYZE DELETE ANALY E STORED DATA 16 107 37 3038 of 3082
38. LTS wt Copyright 2005 Arizona Instrument LLC AZI Customer Service 800 528 7411 or 602 470 1414 Page 65 of 92 7 2 LCD View The LCD screen is viewed by selecting LCD VIEW in the left Ai MAX 5000 AZI Preferred Customer SN 500039 bar This may help to remotely CALIBRATION REPORT monitor the instrument LCD VIEW MEMORY STARTS TEST RESULTS 08 Sep 2005 13 07 37 TESTED Memory Start RESULT 40 3460 Moisture TIME 1 50 TO TEST NAME PRESS START MEMORY SELECTS PRESS MEMORY h NEW PARAMETERS PRESS PARAMS 100 0 0045g CONTRAST MEMORY MENU GRAPH PARAMS Copyright 2005 Arizona Instrument LLC 7 3 Memory Starts Memory starts can be viewed by selecting MEMORY STARTS in the left bar AZI Preferred Customer SN 500039 YY max 5000 CALIBRATION REPORT LCD VIEW MEMORY STARTS TEST RESULTS Factory WEIGHT Factory LACTOSE MONOHYDRATE POLYPROPYLENE CLEAR Factory ULTRAPURE WATER Factory MILK MOISTURE Factory MILK 45H Factory DRIFT Memory Start y Memory Start 6 Select memory starts to download Use Shift key to select range Lise CS file extension when saving files Copyright 2005 Arizona Instrument LLC Memory starts can be downloaded from the instrument to disc or can be uploaded to the instrument from disc Thus memory starts can be retrieved from one instrument and saved to another Typically setup one MAX 5000 with the desired memor
39. O e Select NO if you do not need a printed graph of the test e Select YES if you have only the parallel port selected and want to print the graphic plot of weight and moisture ash results for a test Press ESC to return to the REPORT SETUP MENU Select REPORT ITEMS TO PRINT See page 53 By default all items are selected to print If you do not want or need items use the arrow keys to select the item s and press ENT to toggle the check mark off If you later need the item position the cursor in front of the item and press ENT to toggle the check mark back on Press PAGE 2 to see additional selections Press ESC until the Main Test Screen appears REPORT ENABLED Yes REPORT START Test End REPORT INTERVAL Second INCLUDE GRAPH The output to the printer will be tested when a weight test is performed later If you are connecting a printer to an instrument that has been in operation it is necessary to Turn the MAX 5000 OFF Power ON the printer Turn the MAX 5000 back ON to establish the communications Run a weight test and monitor the output to the printer AZ Customer Service 800 528 7411 or 602 470 1414 Page 19 of 92 5 5 Connecting a Computer If you wish to monitor the test with a computer connect a null modem cable to the serial connector at the rear of the instrument Use any serial port communications program not provided on your computer to read the reports transmitted from the MAX 5000
40. PERFORMANCE VERIFICATION Periodic checks of the system are recommended to verify proper system operation These periodic checks will satisfy some quality assurance requirements as specified in your quality assurance control policy A log is recommended either hand written or computer managed which will record the weight test results sodium tartrate test results and retained samples results Shifts in results that are out of range can be quickly detected and the appropriate corrections made 10 1 The Weight Test The weight test as described in the section beginning on page 23 simulates moisture loss with the removal of one of two known weights This test may be used as part of your control verification program e Atthe interval established by your quality assurance control program e Any time the unit is moved the power switched off the instrument is jarred or bumped and or e lf there is any concern about the accuracy of the instrument AZI Customer Service 800 528 7411 or 602 470 1414 Page 74 of 92 The precision weight test consists of performing the weight test in Section 10 1 multiple times to determine the instruments repeatability This is a more critical verification of the instrument s performance Mean Standard Deviation and Relative Standard Deviation are calculated using the following formulas Mean total of results divided by number of results 7 2 Standard Deviation a n Relative Standard Deviatio
41. Selecting this END ON Prediction option will display the ending criteria options Temp Predict gt Prediction The MAX 5000 Temp gt Rate uses an advanced version of Temp gt Time or Rate the AZI moisture prediction algorithm to automatically end each test early to speed test results The prediction method relies on the fact that most samples exhibit a weight loss END ON Prediction trend such that the rate of volatilization is proportional to PREDICTION TYPE Auto the amount of volatile material remaining in the sample When the prediction is sufficiently reliable the test ends and the predicted result is displayed stored in the instrument and printed if an accessory printer is enabled and attached Proper selection of the ending criteria can greatly improve the test times and repeatability AZI Customer Service 800 528 7411 or 602 470 1414 Page 27 of 92 Auto The instrument monitors the progress of moisture determination and automatically selects ENDING CRITERIA the ending criteria most END ON Prediction appropriate to the test under way PREDIC TION TYPE Manual e Mantel AN A TEW END TEST WHEN ACTUAL IS WITHIN instances best 90 performance will be OF PREDICTED VALUE achieved by adjusting the ending criteria This ending criterion is called the Reliability and is activated when the predicted moisture value and the actual moisture loss are sufficiently close together The number displayed is a p
42. THE OVEN the temperature measured by the RTD temperature sensor in the heater PRESS GO WHEN READY Any difference between the two readings is then supplied to the MAX 5000 to correct the RTD reading The A Se O 1 2 verification process monitors the corrected RTD temperature and the TCI temperature and reports as an error any difference between the two The systematic process is explained later in this section Perform Verification TEMPERATURE VERIFICATION The Perform Verification selection allows a temperature verification to be performed at any time as long as the TCI is connected to the unit TEMPERATURE VERIFICATION TAKES SEVERAL HOURS TO COMPLETE ATTACH THE TEMPERATURE CALIBRATION MODULE TO THE SERIAL PORT REMOVE THE PAN AND PAN SUPPORT AND PLACE THE CALIBRATION PLATE IN THE OVEN PRESS GO WHEN READY CANCEL AZI Customer Service 800 528 7411 or 602 470 1414 Page 45 of 92 View Cal Report The VIEW CAL REPORT selection provides information about the last calibration date and performance A record of the calibration date LAST CALIBRATION 70 Jan 2005 verification date and calibration LAST VERIFICATION 17 Jan 2005 devices used is shown The data ma e cada i ay MODULE 55555 e printed from this screen by pressing PROBE y 66 PRINT MODULE CERT DATE 22 Feb 2005 MAX VERIFY ERROR 1 0 C AT 150 0 C Pressing DETAIL gives a listing of the calibration temperatures and their verification
43. a long test and may be useful for tests gt 15 minutes When used try a HiStart temperature 25 C above the Test temperature To deactivate the HiStart feature set the HiStart temperature to 25 C gt Idle Temperature The temperature that is maintained by the instrument between tests This speeds testing by decreasing the time required to heat the sample chamber at the beginning of the test It can save 10 to 120 seconds of test time if properly used A common idle temperature value is 50 C or 80 C for most products For plastics a 100 C idle is desirable For ashing a 100 C idle may be desirable to reduce time between tests Any value between room temperature and 300 C can be used Experiment to see what value is best for your application Too high a value will cause low results because the sample is evaporating while the basis weight is being measured at the beginning of the test gt Chamber Stabilization When a number other than zero is entered a chamber stability warning will appear at the bottom of the Main Test Screen Although this function will not prevent tests from being run the results from extremely low high moisture samples will be improved if the test is not started until the display indicates CHAMBER STABLE For ultra low moisture samples 5 minute stabilization will help repeatability For ultra low ash samples 10 minute stabilization will help repeatability gt The possible displays are CHAMBER STABLE
44. actory for annual calibration NOTE Shipment of the instrument with the pan support in place will damage the ceramic oven Damage caused by improper shipment is not covered by the warranty e Open the lid of the MAX 5000 by lifting the handle just above the display e Install the pan support gently but firmly into the instrument The stem of the pan support fits snugly into the balance e Check the clearance between the pan support arms and the heat shield by observing the space between the arms and the heat shield The clearance should be between 1 5mm to 4 5mm 1 16 to 3 16 and equal for all five arms of the pan support If the clearance is considerably different for one arm or the pan support appears to be crooked call AZI Customer Service at 800 528 7411 or 602 470 1414 CAUTION Avoid excess pressure from the top or sides of the pan support that can damage the force balance AZI Customer Service 800 528 7411 or 602 470 1414 Page 14 of 92 5 2 AC Power Connection Power Requirements 100 130 VAC 50 60 Hz 8 Amps fuse is F8A T250V Fused on off switch is part of an EMI power entry module Ensure the AC power receptacle is set for the correct voltage of 110 VAC The MAX 5000 will draw up to eight 8 amps Ensure the instrument is provided with its own DEDICATED and GROUNDED electrical power outlet Do not put it on a circuit with motors blenders heaters coolers grinders or other high current electrical d
45. aging i e 3s 5s 8s or 20s gt Special Tare There are TARE OPTIONS three special tare options Two of these set up operations that are not required for most types of PAN TARE Standard testing Normal Tare selects normal operation of tests SPECIAL TARE Normal Syringe Tare Used to Syringe obtain an accurate Foil sample tare for materials that rapidly evaporate during the sample loading process An external balance with 0 0001 gram 0 1 mg resolution is used to weigh the syringe before and after loading which are entered into the MAX 5000 The difference between the loaded and empty syringe weights is start weight Foil Tare is used when a sample of sprayed material such as paint is to be tested Up to 50 pieces of foil can be pre weighed before actual testing begins The alpha numeric identity of each foil is stored in memory until it is replaced by other foil tare entries During a test the ID of the foil is entered from a menu of stored foils The weight of the foil is subtracted from the measured weight giving the true weight of the sprayed sample See section 6 5 for more information SAMPLE TARE Start After s AZI Customer Service 800 528 7411 or 602 470 1414 Page 32 of 92 Result Display Options RESULT DISPLAY OPTIONS gt DISPLAY RESULT AS Selects the final display value of moisture solids or dry weight ash or LOI Moisture Test results DISPLAY RESULT AS i Moisture are
46. alculated moisture solids dry weight ash and loss on ignition Current and programmed chamber temperature Elapsed test time Sample name Rate of weight loss Electronic Interface Ethernet direct connection to internal intranets via 10Base T connection Serial Port 9 pin Parallel Port 25 pin USB and AUX will be supported in a later version User Interface User programmable through keypad Ethernet connection to internal intranets via 10Base T connection RS 232 Serial Port Parallel Printer Port Computer Internet communication using a separate communications Communication package Web Server or simple monitor with Internet Explorer Applications Software Menu driven through keyboard and display Memory Storage for 250 parameter sets 1 000 test results and 100 data curves Statistical Calculation Mean standard deviation relative standard deviation Access Code Programmable through front panel Two codes available for different levels of instrument access Results Display Moisture Solids Dry Weight Ash LOI Linked test results individual and total results are displayed after a linked series Programmable Test parameters Features Report content Controlled ashing rate AZI Customer Service 800 528 7411 or 602 470 1414 Page 89 of 92 Moisture Level gt 10 lt 0 5 Standard Deviation gt 0 1 lt 5 Relative Standard Deviation gt 0 01 lt 0 005 Standard Deviation gt 0 0
47. an 2006 LACTOSE MONOHYDRATE 28 Dec 2005 LACTOSE MONOHYDRATE 27 Dec 2005 LACTOSE MONOHYDRATE 27 Dec 2005 LACTOSE MONOHYDRATE 27 Dec 2005 LACTOSE MONOHYDRATE 27 Dec 2005 LACTOSE MONOHYDRATE 27 Dec 2005 TARTRATE 27 Dec 2005 TARTRATE 27 Dec 2005 TARTRATE 27 Dec 2005 TARTRATE 27 Dec 2005 TARTRATE 27 Dec 2005 TEMP RUN 22 Dec 2005 TEMP RUN 22 Dec 2005 TEMP RUN 22 Dec 2005 TEMP RUN 22 Dec 2005 Select test results to download Use Shift key to select range Use CSV file extension when saving files Download Tests Download Test amp Graph Data Copyright 2005 Arizona Instrument LLC E Done Pop a Internet A For more information regarding the Web Server option please contact Arizona Instrument Sales or customer service representative 800 528 7411 or 602 470 1414 AZI Customer Service 800 528 7411 or 602 470 1414 Page 67 of 92 8 PERFORMANCE ENHANCEMENT 8 1 Introduction Please note The information presented here is general in nature The Computrac MAX Series loss on drying moisture analysis system takes advantage of computer technology to reduce test times and increase precision over standard vacuum oven or convection oven drying procedures The sample is added to an aluminum sample pan sitting atop an electronic force balance within a test chamber As the sample heats and loses weight due to moisture evaporation the balance records that weight loss and transmits the data to the microprocessor The microprocessor i
48. andard The percentage difference between the stabilized MAX 5000 mean values and the reference standard is entered as the LOW MOISTURE CF This feature is not typically used CURRENT LIFT 0 0046 q LOW MOISTURE CF 0 0000 Moisture AZI Customer Service 800 528 7411 or 602 470 1414 Page 34 of 92 Linked Test Options Linking tests enables multiple adjoining memory starts to be run together as one test Increasing temperatures rates times etc may be selected to monitor the progress of moisture loss for research and or control purposes In some materials such as gypsum powder moisture occurs in more than one molecular state usually free water and one or two bound water states A third condition exists in the ashing process when other components of the sample are driven off at very high temperatures and only the ash remains The Linked Test Option will also prevent rehydration of the sample between tests As the test for free moisture is completed the next memory start in sequence will be initiated This subsequent memory start must contain the parameters for the bound water test As this test ends the next memory start may be initiated and so on for as many links as needed for the process The sample integrity will not be compromised between tests since the instrument s lid is not opened e Free water evaporates from the sample at relatively low temperatures typically 40 to 80 C e Bound water however requires higher temper
49. ation is excessive the instrument will stop the calibration process and indicate the reason The balance should be calibrated e Immediately after initial setup and CALIBRATE BALANCE warm up e Whenever the instrument is moved BALANCE CALIBRATION REQUIRES THAT THE e Whenever a problem is suspected MAI ES or LOCATED IN A VIBRATION FREE AREA e As your quality policy dictates A 50g CALIBRATION WEIGHT IS REQUIRED e From the Main Test Screen press Menu PRESS GO WHEN READY gt Select the CALIBRATION MENU gt Select BALANCE CALIBRATION ee oo A gt Select CALIBRATE BALANCE The first screen reminds the operator that the instrument must be free of vibration during the calibration and that a 50 gram calibration weight must be available e Press GO If the chamber temperature is too hot the screen will prompt the operator to OPEN LID CHAMBER TEMPERATURE IS OPEN LID TOO HOT WAIT FOR COOL DOWN If the chamber temperature is cool Henne Eee TARE AS et p WAI FOR COOLDOWA enough to begin the calibration one of two prompts will appear e lf the lid is closed the OPEN LID PLACE CLEAN PAN ON TRAY OPEN LID SUPPORT prompt will appear e Ifthe lid is open the PLACE CLEAN PAN ON TRAY SUPPORT CLOSE LID prompt will appear PLACE CLEAN PAN ON TRAY SUPPORT AZ Customer Service 800 528 7411 or 602 470 1414 Page 21 of 92 After placing a clean pan on the CALIBRATING BALANCE ba
50. ator plate e Replace the worn seal AZI P N 250 0001 e Reassemble AZ Customer Service 800 528 7411 or 602 470 1414 Page 80 of 92 This selection is the start of an oven cleaning process that burns off most residue build up in the oven Similar to the self cleaning mode in a home oven the process turns the heating elements on to reach an oven temperature of 550 C in less than 22 minutes The oven cleaning cycle lasts a total of 45 minutes The heat generated makes the top of the instrument hotter than normal and extreme caution should be used to prevent injury until the instrument has cooled down after the cleaning cycle oir daci n e CALIBRATION MENU and OVEN SELF CLEANING e Press ENT The screen display OVEN SELF CLEANING TAKES will provide instructions as 45 MINUTES TO COMPLETE Sa EA REMOVE THE PAN AND PAN SUPPORT e When the pan support has been FROM THE OVEN AND CLOSE THE LID gently removed see page 83 close the lid and press GO PRESS GO WHEN READY A A O L CANCEL e The temperature and time left in OWEN SELF CLEANING the cleaning process will displayed e Atthe conclusion of the cleaning EPEORNING CLEAN cycle the screen will display OVEN SELF CLEANING IS DONE TIME LEFT 44 58 Press OK or ESC to return to MAIN MENU A A O E ee Cleaning frequency will vary depending on the application and frequency of use We recommend that you establish a periodic cleaning schedule to ensure unit t
51. atures some around 135 C and others around 240 C to break the molecular bonds e The MAX 5000 also provides temperatures to 600 C for ashing purposes The linked tests may also be helpful in determining the temperature at which moisture is freed in unknown sample material Setting the temperature a few degrees higher in each succeeding test will eventually reach a point of weight loss due to evolved moisture Gatun the meee test or to make this the last segment of linked series toggle this parameter to NO LINKED TO NEXT TEST Ho e Result Type The two selections for RESULT TYPE Individual result type offer an Individual determination for this memory start or a Total determination of two or more USE THIS TEST WEIGHT FOR TOTAL CALCULATIONS linked memory starts DISPLAY PURITY gt Individual Result This selection PURITY FACTOR provides the result of this test when two or more tests are linked ALLOW MOISTURE GAIN gt Total Result This selection provides the result using the start weight of a preceding memory start and the final weight of this memory start AZ Customer Service 800 528 7411 or 602 470 1414 Page 35 of 92 Use This Test Weight For Total Calculations E eee if this test s starting weight is to be used in Total Result calculations for this or later LINKED TO NEXT TEST No linked tests The starting RESULT TYPE Total weight for this test will be used for total calculations until anoth
52. cation after the calibration is complete and the instrument has cooled down gt If NO is selected a verification may be performed manually later if desired gt To verify the twelve 12 factory points between 50 C to 575 C ensure that POINTS TO VERIFY is set to ALL CAL POINTS gt To verify up to four 4 custom points set POINTS TO VERIFY to 1 2 3 or 4 Set CUSTOM TEMPERATURE VERIFICATION POINTS to the desired temperatures up to 575 C Factory defaults are 100 150 200 and 250 The entries are saved if there are no highlights on the display NOTE The remaining instructions assume that verification is being run automatically e Return to the TEMPERATURE CALIBRATION MENU highlight PERFORM CALIBRATION and press the ENT key gt A screen prompt appears to ensure that you have the temperature calibration device connected If the device is hooked up press GO NOTE Temperature calibration and verification is a lengthy process that requires about 6 hours to complete consider performing the calibration at night Once begun do not open the lid during the calibration or verification process Although the cool down process may be speeded up by opening the lid between calibration and verification disturbing the oven may cause verification errors One way to save time is to start the calibration when the instrument s oven is below 40 C The instrument must adjust the temperature to and be stable at 40 C for several
53. ccccccccnnnnnananannanannnnnnnccconononananana nono nonnnnononocononnnnnnnnn ns 43 Calibration Reminder Setup ccccnnnncnonononanananonocononononananananononnnnncnnnonocoononannnnn nro nonononnss 43 Temperature CAMDM ATION ses tias 44 VENINCATION SUD ad aii A 44 Calibration Module Setup ccccccnnnononononanananononoconononannnnanan ono nnnonononocoononnnnnnnanononononoss 44 Adi nO aiy A A AA A aensetnne 45 Perform Ve nica lOs aiii 45 View CIl FRED OM assi 46 Calibration Reminder Setup ccccnnnnnononononanananonononononnnnnnananonnnnnnnnononoconononnnnnnnn nono nononnos 47 Temperature Calibration Instructions ooococccccccnnnnnnnnanananananonononococonononanananono 48 Oven CILA isis dilataci n 50 0 Setup Melero nad oia lala 51 o A CR O US 0 ON NON 51 REPO SELL Drill aora 52 Ropo CON TO DUO ae 52 Report Items TORIN bic ancndasndsanaisharncceuaeate E E 53 PO GIEGIE COMPANY IN AN sr o ise 53 AZI Customer Service 800 528 7411 or 602 470 1414 Page 3 of 92 Date TIMMS SE LUD tri ia 53 Data Formato ni ni a 54 EC IS AY ana aa laa aiaa 54 EOUNUMDO iaa Ot io 54 Access Key OEI Diiia tnd sonumbnse connie E ensue detainbeseagameiteuaecboageseies 55 Master Vico ria een aetan aceneteene 55 C MONON AA E PO O ecu saeaad enteaceurrpenaed een cue taaen toesiace teases 55 Ethernet Setup Not Functional on Version 2 0 or Earlier oocccccccnnnnnonanananononos 55 NAME ana PASSWORD aaa aaa 5
54. d more repeatable results Solid granular sample material should be shaken onto the pan from a spoon shaking sample directly from the collection jar will usually deposit only the larger granules not a representative mixture Liquid samples or slurries can be spooned or AZI Customer Service 800 528 7411 or 602 470 1414 Page 71 of 92 poured onto the sample pan however a pipette or sampling straw will collect a uniform mixture of the sample All samples should be evenly distributed over the sample pan 9 2 Preparing and Testing Liquid Samples Some liquid samples will dry slowly even at high temperatures Drops or puddles may form on the pan and dry slowly as the sample concentrates A skin may form which limits evaporation from the interior Long test times greater than 25 minutes and low percent moisture or high percent solids values suggest incomplete drying Pre dried glass fiber filter paper and flat bottomed pans are recommended to successfully test samples such as these The glass fiber filter paper AZI P N 990 0003 serves as a support for the liquid sample The sample is absorbed through the paper for more even and rapid evaporation to reduce test times Waffle bottomed sample pans are not recommended for use with filter paper The liquid sample may soak through the filter paper and collect in puddles in the pan s depressions Flat bottom pans are available from AZI as P N 990 0010 Call AZI parts sales at 800 528 7411 or 602 4
55. due is burned off the oven lining Use a fume hood to remove the fumes and smoke from the work place Consult the MSDS on your sample materials to determine the content of the emissions OWEN SELF CLEANING TAKES 45 MINUTES TO COMPLETE PRESS GO WHEN READY For a complete description of the oven cleaning procedure see section beginning on page 81 AZI Customer Service 800 528 7411 or 602 470 1414 Page 50 of 92 64 Setup Menu The SETUP MENU is used to program the MAX 5000 with information relevant to all SETUP MENU operations and tests From the SETUP MENU it is possible to setup the printer items to be printed time and date certain items on the display PRINTER SETUP access codes network addressing and ne ae DATE TIME SETUP access and prepare foils for testing 4 DATA FORMAT SETUP ACCESS KEY SETUP ETHERNET SETUP SPECIAL FEATURES Printer Setup This menu provides a means of selecting PRINTER SETUP the type of printer that is attached to the MAX 5000 which port is to be used and if an automatic form feed is desired If the port option of Both is selected the data eee Re 932 will be sent to the computer serial port and PRINTER TYPE Both baster the parallel printer port Graphics will not be sent to the RS 232 port See AUTO FORM FEED Yes Connecting a Printer on page 17 and Connecting a Computer on page 20 PRINTER SETUP OUTPUT PORT RS 237 AUTO FORM FEED Seik
56. e during transit WARNING SYMBOLS Protective Ground Identifies the connection on the chassis and power inlet module for connection of the safety ground green yellow wire Warning Information or procedure that must be observed Electrical Shock Hazard Observe all steps of the procedure to prevent electrical shock Warning Hot surface or area of possible severe burns Use listed precautions when opening the oven lid and handling samples to prevent injury S A AN Page 6 of 92 2 CHANGES Firmware Rev 1 11 Manual Rev D 1 Temperature calibration changed from 10 points 50 to 500 to 12 points 50 to 575 2 Temperature Calibration Interface Kit pictures added 3 Total Percent Moisture calculation added to linked tests 4 New Test Ending Criteria to reach temperature before checking for rate or prediction added Firmware Rev 1 12 Manual Rev E 1 Updated accessory kit list on page 12 2 Updated menu selection criteria throughout the manual to change ON OFF to YES NO toggle operation and added menu items for additional linked test options Firmware Rev 2 0 Manual Rev F 1 Updated menu selection criteria throughout manual to update to new firmware configurations 2 Edited Parts and Consumables listings to include latest offerings 3 Updated maintenance procedures to include insulator seal oven vent and fan shroud maintenance Firmware Rev 2 1 Manual Rev F 1 Revised balance calibration for
57. e the balance readings to drift during taring and are reflected as an error in the final value Because the conditions that cause the temperature fluctuations are not always constant the error often appears as a higher standard deviation for a series of tests on a particular sample The Pan Tare option has four 4 user selections to adjust the relationship between tare time and tare error to improve overall operation for a given product or purpose The following table summarizes the menu choices and the parameters associated with each choice MENU MINIMUM TARE CHOICE TIME PAN TARE Standard ULTRA LOW MOISTURE The ULTRA LOW MOISTURE option should be used for samples with moisture solids ash content between 0 005 and 0 2 percent This option may be used to improve accuracy for samples with moisture content in excess of 0 2 percent but will increase the overall testing time AZI Customer Service 800 528 7411 or 602 470 1414 Page 31 of 92 gt Sample Tare Records the average Initial sample weight during a programmable period of time After lid closure there is a 7 second delay before the PAN TARE Standard sample tare time begins There are five 5 possible choices under this menu The factory default is Start After 8s Start When Stable Taring continues until the weight is found stable lt 0 2mg deviation over 4s of time Start After selectable Seconds Test starts after selectable seconds of aver
58. ements However improper use or handling of the instrument can result in damage to the equipment and or injury to personnel The MAX 5000 is Equipment Class I Pollution Degree 2 and Installation Category Il Read these operating instructions thoroughly before using your moisture analyzer to prevent damage to the equipment Keep these instructions in a safe place An up to date electronic copy of most AZI manuals can be found on the AZI web site at http www azic com products aspx The following instructions will help to ensure safe and trouble free operation of your MAX 5000 e Use the analyzer only for performing moisture solids ashing analysis of samples Any improper use of the analyzer can endanger persons and may result in damage to the analyzer or other material assets e Place analyzer in an appropriate fume hood for ashing and or burning a sample e When ashing and or burning a sample use the ash rate to slow ignition and prevent flaming If the sample flames it will damage the heater and produce poor test results e Do not use this analyzer in a hazardous area location operate it only under the ambient conditions specified in these instructions e Although the analyzer is simple to operate and is very user friendly it should only be operated by qualified persons who are familiar with the properties of the sample being analyzed e Ensure before getting started that the voltage rating printed on the manufactur
59. ent is optimized to run low moisture samples with a 30g load The larger sample size is necessary to provide sufficient weight loss information to the microprocessor A larger weight change between the initial and final sample weights will give more accurate repeatable results A smaller sample size will give faster tests Generally it is best to use a thin layer that just covers the surface of the pan Very light or very expensive test materials may necessitate weighing as little as 150 milligrams Be aware that the standard error will be larger for samples this small unless the moisture content is sufficient to cause a weight change of 10 milligrams or more for each test If the first result is too low increase sample size or raise the temperature If it is too high decrease sample size or lower the temperature The sample should be added evenly in a thin layer across the entire surface of the pan Once the desired amount has been added the pan can be removed from the test chamber to evenly distribute the sample The sample should be distributed over the surface by gentle shaking tapping or for very viscous samples spread with a spoon or knife Do not spread granular products with a spoon as the particles compress even under gentle pressure affecting test times and results An even layer of sample across the pan will reduce test times and increase precision of results AZI Customer Service 800 528 7411 or 602 470 1414 Page 73 of 92 10
60. er YES is encountered Note the starting weight of the first test in a linked series is used for total calculations until a YES is USE THIS TEST WEIGHT FOR TOTAL CALCULATIONS Yes encountered e Purity Factor Toggle this parameter to YES if a value DISPLAY PURITY No provided by the user is to be PURITY FACTOR 4 778 entered as a parameter for the purity calculation The default ALLOW MOISTURE GAIN No number is 4 778 which is the purity factor for gypsum The purity result is Purity Factor times the Moisture loss e Allow Moisture Gain When testing some materials the linked test option is used The first test is the free moisture test which drives off any free moisture However some materials if already dried might actually collect moisture during this test With the Allow Moisture Gain enabled the collected moisture will be reported Ending Criteria for this test is usually Rate or Time If Rate is used the test would never end if the sample gained moisture AZI Customer Service 800 528 7411 or 602 470 1414 Page 36 of 92 eri oil start which can be optionally inserted 9 97 at another location Use this Factory WEIGHT Factory TARTRATE Factory LACTOSE MONOHYDRATE Factory POLYPROPYLENE CLEAR Factory ULTRAPURE WATER carefully The Factory memory starts cannot be deleted with this option Factory MILK MOISTURE Factory MILK ASH Factory DRIFT Memory Star
61. er s label is identical to your local line voltage see the section on AC Power Connection beginning on page 15 e The MAX 5000 is provided with a power cord that has a protective grounding conductor Do not operate the unit without the supplied three wire power cord or an identical recognized equivalent that meets all applicable standards e AZI does not recommend the use of an extension cord However if one must be used use an extension cord that meets all applicable AZI Customer Service 800 528 7411 or 602 470 1414 standards and has a protective grounding conductor e Do not remove the protective grounding wire from the power supply circuit e With the power switch off power is still applied to the power entry module Remove the power cord from the MAX 5000 to ensure complete power removal from the instrument e Position the power cable so that it cannot touch any hot areas of the analyzer e Use only AZI supplied accessories and options with the MAX 5000 or check with AZI before using any third party accessories such as printers and analytical balances to verify compatibility e If there is visible damage to the MAX 5000 or its power cord unplug and isolate the unit first and then call Customer Service at 800 528 7411 or 602 470 1414 e Do not open the housing of the MAX 5000 There are no user serviceable parts or adjustments inside the unit Any maintenance inside the unit is to be performed by factory trained
62. ercentage equal to the actual moisture loss divided by the predicted moisture times 100 For example if 90 is entered the test will end when the actual moisture loss is within 90 of the predicted moisture gt Rate This feature ends the test when the weight loss rate falls below a selected value This value is a percent per minute For example entering 0 100 END ON Rate means that the test will end when the rate of change of the actual RATE 0 1000 Minute weight loss is less than 0 100 of the original sample weight per minute The instrument s program assumes that the sample contains volatile material that will completely evaporate after some time It further assumes that at first the rate will increase to a maximum point and then gradually decrease to zero Selecting some threshold rate value will cause the test to end before the sample is completely dry Rate is useful to profile a sample s weight loss Some applications are faster in the rate ending criteria Others are faster in the prediction ending criteria AZI Customer Service 800 528 7411 or 602 470 1414 Page 28 of 92 Some samples contain more than one volatile material for example nylon resins with large amounts of unpolymerized material Caprolactam In these cases the rate will fall to some fixed value when all the moisture is evaporated but while the monomer is still evaporating The graphical display available from the Main Test Screen
63. evices The MAX 5000 with a single 100 120 VAC ceramic heater can be used only with a 100 120 VAC input power source Check the setting on the fuse holder located at the rear of the instrument for the correct voltage 100 120 volts For 220 240 VAC operation a separate 220 to 110 step down transformer must be used Such a transformer is available from Arizona Instrument AZI P N 5600 0019 and is shipped with instruments destined for countries using 220 240 VAC as their primary voltage To ensure the entry module is set for 110 VAC operation e Pull the fuse access cover open and down Mm e Pry the fuse holder from the power entry module e Ensure only 8 amp timed fuses are installed e Install the fuse holder into the power entry module with 115V showing as in the picture e Ensure the power switch located next to the fuse holder is in the OFF position e Remove the power cord from the packing material and insert the rectangular end into the power receptacle located on the other side of the fuse holder 110 120 V F8A T250V The power source outlet used by the MAX 5000 MUST BE GROUNDED If you are not sure if the socket is grounded check with your plant electrician before proceeding The MAX 5000 will appear to function correctly even if the socket is not grounded but there is a danger of possible electrocution If grounded outlets are not available consider using a ground fault interrupter to protect person
64. he Optimize feature allows the user to raise or lower the test temperature to improve test speed and repeatability Typically a higher temperature results in a faster and less repeatable test A lower temperature results in a slower and more repeatable test Raise the temperature to speed up the test Press GO Have the sample prepared and ready to start the optimization test Press Start AZI Customer Service 800 528 7411 or 602 470 1414 THIS FEATURE ALLOWS THE USER TO SLIGHTLY CHANGE THE CURRENT MEMORY START TEMP TYPICALLY A HIGHER TEMP RESULTS IN A FASTER AND LESS REPEATABLE TEST A LOWER TEMP RESULTS IN A SLOWER AND MORE REPEATABLE TEST ONLY CHANGE THE TEMP A FEW DEGREES AT A TIME PRESS GO WHEN READY NAME WONTON TARGET RESULT 1 0000 RESULT TYPE 5 Moisture CURRENT TEMP 150 C NEW TENMP 160 C A SINGLE TEST WILL BE RUN AT THE SPECIHED TEMPERATURE IF POSSIBLE AN ENDING RATE WILL BE DETERMINED PRESS START TO BEGIN TEST Page 63 of 92 PARAMETER EXPERT Optimization is complete and the memory start has been modified Press OK and exit Parameter Expert to use the optimized THE MEMORY START IS READY FOR USE memory start THE ENDING RATE HAS BEEN DETERMINED For more information regarding the Parameter Expert option please contact Arizona Instrument Sales or a customer service representative at 800 528 7411 or 602 470 1414 AZI Customer Service 800 528 7411 or 602 470
65. he balance calibration and Hi PRESS PARAMS weight test If the result is still out of tolerance call AZI customer service at 800 528 7411 or 602 470 1414 for assistance CONTRAST MEMORY MENU GRAPH PARAMS 0 003454 AZ Customer Service 800 528 7411 or 602 470 1414 Page 23 of 92 6 MAIN MENU Press MENU to display the MAIN MAIN ME MENU MEMORY START MENU STORED DATA MENU CALIBRATION MENU 4 SETUP MENU SAMPLE FOILS The MEMORY START MENU allows access to the stored test parameter sets This is where the contents of the stored parameters may be printed out or changed MEMORY STARTS are stored test i MEMORY START EDIT parameter sets They are sequentially numbered from 1 to the total of Memory q 22 Starts entered into the instrument The gt Fue TARTRATE above figure shows memory start 9 of 22 3 Factory LACTOSE MONOHYDRATE selected 9 22 in the upper right hand E Factory TRADE WATR i a actor cone Mesa INSERT TO 6 Factor MILK MOISTURE INSERT will insert a memory start 7 Factory MILK ASH above or below the highlighted position 8 Factory DRIFT respectively All other memory starts will 9 Memory Start 7 be moved down unchanged 10 Memory Start 8 e The Factory presets are for ENTER TO EDIT OR START TO BEGIN TEST standard tests to check the performance and or calibrate the instrument e The Up and Down Arrow Keys move the highlight one line at a time
66. he system will then lead you to the optimal A SERIES TESTS AT DIFFERENT temperature for the fastest test time Press in DEE eee Start to begin the next test ACCEPTABLE TEMPERATURE TO PRODUCE THE SHORTEST TEST TIME 134 C lt TARGET TEMP 150 C PRESS START TO BEGIN TEST The system will continue through the iterative optimization process until the THE HIGHEST ACCEPTABLE TEMPERATURE optimal temperature Is achieved in the AND ENDING RATE HAS BEEN DETERMINED shortest amount of test time without WHICH SHOULD PRODUCE THE SHORTEST degrading or burning the sample A memory TEST TIME ds THE MEMORY START IS READY FOR USE elf the iterative process goes through 5 times without completing it would be best to restart the Parameter Expert as the initial temperature was significantly incorrect This may also occur for materials that tend to burn or degrade easily Once the memory start has been developed exit the Parameter Expert screen and run 5 tests to determine if the accuracy speed and repeatability are acceptable Use the Optimize feature to make minor improvements AZ Customer Service 800 528 7411 or 602 470 1414 Page 62 of 92 Optimize TemprspeediRepestabiteyy The Parameter Expert Optimize feature determines the proper ending rate criteria for a user specified test temperature for the currently active memory start The Optimize feature is for an already developed memory start that needs minor improvement T
67. ht When not in use select a test with start temperature of 25 C and keep the lid down The oven will remain cool and the Liquid Crystal Display LCD back light will turn off after a period of inactivity to conserve energy Press any key to restore illumination to the LCD If the test process is interrupted by outside interference such as static discharge surges or drops in line voltage or otherwise freezes its operation and or display reset the instrument by recycling power By turning the instrument off and back on the instrument should return to normal operation If the instrument does not return to normal operation contact AZI Customer Service at 800 528 7411 602 470 1414 or by e mail at support azic com The ceramic sample oven maintains accurate temperature control to prevent sample degradation as it removes volatile components as quickly as possible A Resistance Temperature Device RTD mounted near the heater measures oven temperature Temperature is held within one degree of the set value after equilibrium is reached The oven usually reaches test temperature in 90 to 180 seconds depending on the difference between the starting and setpoint temperatures The oven is automatically turned off when the test is completed Due to the nature of the oven lining the oven is cool almost immediately after opening the lid The sample will remain hot much longer than the oven itself so caution is required when removing used sam
68. ht produce force on the balance e Adjust the feet under the instrument to center the bubble in the level located under the lid P q y JUL e Keep all cables as short as possible less than 2 meters 6 feet in length if possible e Avoid placing the analyzer in close proximity to a heater or otherwise exposing it to heat or direct sunlight e Avoid exposing the analyzer to extreme temperature fluctuations Protect the analyzer from drafts that come from opening windows doors and air conditioning systems Keep the analyzer protected from dust and liquids as much as possible Protect the analyzer from caustic vapors Do not expose the analyzer to saturated humidity environments Moisture in the air can condense on the surfaces of a cold analyzer whenever it is brought into a substantially warmer place If you move the MAX 5000 to a warmer area make sure to condition it at room temperature disconnected from AC power for at least 2 hours or until moisture condensation no longer occurs For those familiar with moisture analysis instruments the MAX 5000 should be simple to operate However please read all of the sections in A this manual before attempting your first analysis AZ Customer Service 800 528 7411 or 602 470 1414 Page 13 of 92 e Remove the sample pan support from its separate box and packing foam e Save the box and foam so that the support can be packaged when the instrument is returned to the f
69. is of great help in visualizing these cases The rate ending criteria if set above the rate of monomer volatilization may give the best results gt Time This feature places a time limit on a test The test will end when a selected amount of time has passed Fixed time tests are END ON Time seldom superior to the prediction algorithm but can be useful when conducting application characterization tests or when investigating long term stability of the instrument gt Jima ce Rate hie fete is used during tests when a change in moisture may not occur When the moisture characteristics of a sample END ON Time or Rate are not known linked tests with slightly higher temperatures for each successive test may be used With the Time or Rate selection and linked tests enabled testing will continue for the set amount of time and then start the next test in sequence Ifa moisture loss occurs the test will end on the rate criteria and the next test will begin ENDING CRITERIA TIME 5 Minutes RATE 0 1000 Minute TIME 5 Minutes AZI Customer Service 800 528 7411 or 602 470 1414 Page 29 of 92 gt Reliability This feature is similar to Rate but is based on a percent of accuracy With reliability set to 99 END ON Reliability the test would end when the actual moisture loss was RELIABILITY 90 within 99 of the predicted result A value may be set that would give exceptional results when the ac
70. lance close the lid The screen will display CALIBRATING ZERO LOAD PLEASE WAIT As soon as the tare is finished the CALIBRATING 7ERO LOAD PLEASE WAIT next prompt appears e Open the lid and gently place a 50 gram weight in the center of the sample pan or press ABORT to return to the menu OPEN LID Avoid excess pressure on the scale LOAD 50g WEIGHT AND WAIT A Use the tweezers included with the MAX 5000 to gently place the weight on the pan As soon as the instrument detects weight on the balance the prompt CLOSE LID appears With the lid closed the prompt CALIBRATING PLEASE WAIT appears At the conclusion of the calibration one of two prompts will appear indicating the PASS FAIL condition of the calibration Ifthe result is PASS remove the weights and the instrument is ready for operation e Press PROCEED to return to the OPEN LID BALANCE CALIBRATION menu e Press ESC three 3 times to return to the Main Test Screen e Ifthe result is FAIL the calibration was unsuccessful Rerun the calibration CALIBRATION PASSED REMOVE 50g WEIGHT OPEN LID e lf the calibration fails again contact AZI Customer Service at CALIBRATION FAILED 800 528 7411 or 602 470 1414 UNSTABLE ENVIRONMENT REMOVE 50g WEIGHT AZ Customer Service 800 528 7411 or 602 470 1414 Page 22 of 92 5 7 Weight Test To ensure proper calibration and become familiar with the instrume
71. n expressed as percentage Standard Deviation divided by Mean multiplied by 100 e Perform the steps in the weight test a total of five 5 times O ga fle oh press MENU and select the STORED DATA MENU WEIGHT 10 Jun 7005 374995 gt PRESS ENT to select tests WEIGHT 10 Jun 2005 37 5076 gt Arrow up or down to select the WEIGHT 10 Jun 2005 37 5177 five weight tests and press 4 came OS ae a a OU z ANALYZE WEIGHT 10 Jun 2005 37 5081 gt The mean ofthe results should WEIGHT 10 Jun 2005 374998 be 37 500 020 for a test WEIGHT 08 Jun 2005 37 4999 using the five and three gram E 08 Jun 2005 weights WEIGHT 31 Mar 2005 37 5175 e Ifthe results obtained from these tests are outside the specified STORED DATA range call AZI Customer Service at 800 528 7411 or 602 470 1414 for assistance SAMPLES 5 MEAN AVERAGE 37 5118 SD 0 0060 RSD 0 071 5 AZI Customer Service 800 528 7411 or 602 470 1414 Page 75 of 92 10 3 Sodium Tartrate Test The sodium tartrate test verifies the heater and force balance operation Sodium tartrate dihydrate is a chemical with a known moisture concentration that releases moisture at temperatures greater than 120 C Because of its consistent moisture concentration it is used as a standard in moisture measurement using oven or titration methods To obtain consistently reliable results use a high quality or reagent grade ACS certified granular sodium tartrate Sift
72. neglect accident improper application modifications or service performed by persons other than AZI s own service representatives power surges or spikes negligence in use maintenance storage transportation or handling or acts of God If a MAX product is defective in workmanship or materials the owner s sole remedy shall be repair or replacement of the defective part or parts as provided above Under no circumstances shall AZI be liable in any way to the owner or any user for any damage including but not limited to any loss of business or profits or any other direct indirect special incidental or consequential damages whether or not foreseeable and whether or not based on breach of warranty contract or negligence in connection with the sale of such products Some states do not allow the exclusion or limitation of incidental or consequential damages so the above limitations or exclusions may not apply to you No other warranty is expressed or implied including the warranties of merchantability or fitness for a particular purpose In no event shall AZI be liable for consequential and or incidental damages The effective warranty begins on the date of purchase by or lease to the first end user owner Keep the dated bill of sale or invoice for evidence of the effective warranty date when warranty service is requested In the event that any questions or problems should arise in the use or application of your Computrac MAX 500
73. nel against electrical shock AZ Customer Service 800 528 7411 or 602 470 1414 Page 15 of 92 All accessory equipment printer computer etc should be connected before applying power to the instrument See the following sections 5 4 and 5 5 to connect a printer and or computer Plug the line cord into a dedicated electrical power outlet Line noise on the power lines may affect accuracy A dedicated AC power line or an isolation transformer for the MAX 5000 may be beneficial in eliminating this noise Turn the instrument on by pressing the switch on the back of the analyzer next to the power receptacle After displaying the opening screen for about five seconds the main test screen shown below will appear 26 May 2005 13 39 02 TESTED L2 RESULT 0 0000 Moisture TIME 1 00 TO TEST WEIGHT PRESS START MEMORY SELECTS PRESS MEMORY d HEW PARAMETERS PRESS PARAMS 3 27589 CONTRAST MEMORY MENU GRAPH PARAMS e The keypad functions are gt The Left and Right Arrow Keys move the cursor one space left or right in edit mode or to the next field on some input screens They move the highlight bar one page at a time when selecting stored memory starts or stored data test results Stored data is not available on all instruments gt The Up and Down Arrow Keys move the cursor up and down through lists in edit mode They move the highlight bar one line at a time when selecting stored memory starts stored data o
74. ning a unit e For all shipments boxes and packing materials are available from AZI e Pack the MAX 5000 in a Computrac MAX 5000 shipping container only AZI WILL NOT BE RESPONSIBLE FOR SHIPPING DAMAGE IF YOU RETURN THE INSTRUMENT IMPROPERLY PACKAGED OR SHIPPED YOU SHOULD INSURE IT FOR FULL VALUE AZI Customer Service 800 528 7411 or 602 470 1414 Page 8 of 92 A Please save the shipping box and all A packaging materials for future use AZI Customer Service 800 528 7411 or 602 470 1414 Page 9 of 92 4 INSTRUMENT DESCRIPTION The Computrac MAX 5000 is a high performance programmable loss on drying analyzer Typical Moisture Level Repeatability gt 10 lt 0 5 Standard Deviation gt 0 1 lt 5 Relative Standard Deviation gt 0 01 lt 0 005 Standard Deviation gt 0 005 lt 0 003 Standard Deviation Ashing Contact AZI for specific sample repeatability Periodic balance and temperature calibration maintains accuracy Test temperatures from 25 C to 600 C set at increments of one degree Programming is through the front panel keyboard and menu system Programming features can be placed under access code control preventing unwanted changes to operating parameters Stored test parameters for 250 different materials to include 242 user programmable sets and 8 factory set test parameters For 21 CFR Part 11 non compliant instruments test data can be gt Stored in the instrument the last 1 000 tests and 1
75. not clean the oven call AZI Customer Service at 800 528 7411 or 602 470 1414 for recommendations e Every precaution has been taken to prevent contaminants from entering the unit However if liquids or powders do enter the case call AZI Customer Service at 800 528 7411 or 602 470 1414 for recommendations The user shall be liable and responsible for any damage that arises in connection with this analyzer AZI Customer Service 800 528 7411 or 602 470 1414 ENVIRONMENTAL CONDITIONS Storage and Shipping Temperature should be between 0 C to 40 C 32 F to 104 F Relative Humidity should be between 10 and 80 OPERATIONAL CONDITIONS INDOOR USE Optimum results will be achieved when the unit is set on a smooth level non vibrating surface in a non condensing non explosive environment of 0 40 C at 50 relative humidity and 0 31 C at 80 relative humidity Do not expose the moisture analyzer unnecessarily to extreme temperatures moisture shocks blows or vibration UNPACKING THE MOISTURE ANALYZER After unpacking the moisture analyzer check it immediately for any visible damage resulting from rough handling during shipment Save the box and all parts of the packaging to use when returning your moisture analyzer for calibration Only the original packaging provides the AZl recommended protection for shipment Before packing your moisture analyzer for shipment unplug all connected cables to prevent damag
76. nt run a weight test This test simulates moisture loss by the removal of one of two known weights Both a three and a five gram weight are required Ensure that a clean dry empty sample pan is placed on the pan support 1 22 Factory WEIGHT ici pl a Test Screen press Factory TARTRATE 1 l Factory LACTOSE MONOHYDRATE If the highlight bar is not already Factory POLYPROPYLENE CLEAR there highlight Factory WEIGHT Factory ULTRAPURE WATER O f Factory MILK ASH ne of severa prompts may Factory DRIFT appear depending on the last measured weight Follow the Memory Start f instructions given until the OPEN ENTER TO EDIT OR START TO BEGIN TEST Memory Start 8 LID ADD SAMPER prompt COPY INSERTA INSERTH PRINT appears on the screen Open the lid and carefully place both the 5 gram and the 3 gram weights near the center of the sample pan Close the lid After the Main Test Screen is displayed gently open the lid carefully remove the 3 gram weight and gently close the lid Wait for the end of the test If this is the first test of the day discard the results of this test for TESTED WEIGHT Stability reasons and perform the test again RESULT 37 4999 Moisture Record the final test result The EA fw a value should fall between TIME 2 11 Sf AOU aNd 3132040 TO TEST WEIGHT PRESS START If the results are out of tolerance MEMORY SELECTS PRESS MEMORY repeat t
77. nterprets the information and compares the sample s weight loss to a standard drying curve The final moisture concentration is extrapolated from the curve and results are available within minutes A typical drying curve is shown at the right Section A B represents the sample s weight loss as it heats from room to testing temperature Section B C is linear representing moisture loss from the interior of each particle of the tested sample Section C D is exponential in nature Typical Loss On Drying Curve and represents the final evolution of moisture from within the particle The point at which no additional weight loss occurs represents the complete evolution of water from the original sample E Section D E is the longest and most time consuming portion of the entire testing process The time for a sample to reach zero residual moisture in an oven can take from one to possibly 24 hours depending upon the sample sample preparation and testing temperature The MAX 5000 loss on drying system calculates the sample s moisture concentration through mathematical extrapolation from the exponential C D portion of the curve Weight Reading and Moisture Content An aluminum pan sits on a pan support coupled to a sensitive electronic force balance The force balance registers the initial weight of the sample before testing and relays the sample s decreasing weight to the microprocessor for evaluation Balance readings are averaged t
78. o DPU Canon AZI Customer Service 800 528 7411 or 602 470 1414 Page 51 of 92 This menu allows the report outputs to be enabled and the printout to be customized to your needs Report Control Options REPORT ENABLED A YES sends a progress report optional see below to the printer during a test and a final report to REPORT CONTROL OPTIONS the printer after a test REPORT START These three options determine if a progress report begins at REPORT ENABLED Yes the start of the test when the first prediction of final result is made or just a REPORT START Test Start final report at the end of the test REPORT INTERVAL This selection determines how often the progress report INCLUDE GRAPH No data is sent to the printer or computer Times of 1 5 10 15 and 30 seconds or one 1 minute are available The short durations are good for trials and research and the long intervals are good for normal use INCLUDE GRAPH When YES is set a graph of the results will be sent to the printer after the final report Graphic data is not sent to the computer REPORT INTERVAL 10 Seconds AZI Customer Service 800 528 7411 or 602 470 1414 Page 52 of 92 Report Items to Print Test results are available to be printed sent to a computer or used for statistical analysis mean standard deviation and relative standard deviation To be printed the data fields must be selected Highlight the data to be printed and press the
79. o eliminate signal interference or erroneous data AZI Customer Service 800 528 7411 or 602 470 1414 Page 68 of 92 Within each memory start test temperatures can be programmed between normal room temperature 25 C and 600 C in either a one step or multi step temperature contour A heater is located on the underside of the test chamber s lid An RTD monitors the temperature of the test chamber and sends that information back to the microprocessor Under the microprocessor s control the heater element is cycled on and off to bring the test chamber to the programmed level During the entire testing process the microprocessor monitors the sample s weight loss The sample s decreasing weight is compared to the initial sample weight and the calculated moisture concentration will appear on the display Simultaneously the microprocessor is predicting a final moisture concentration based upon the sample s rate of weight loss compared to the exponential portion of a standard drying curve Testing will continue until the predicted moisture concentration agrees within a certain percentage of the actual moisture concentration appearing on the display The percent agreement between the predicted and the actual moisture concentrations will vary depending upon the system s programming and the moisture level of the sample The MAX 5000 is an extremely flexible instrument This flexibility means that some time is required to determine the best set of paramete
80. o unit correlation and performance Call AZI Customer Service at 800 528 7411 or 602 470 1414 if you have any questions AZ Customer Service 800 528 7411 or 602 470 1414 Page 81 of 92 11 5 Oven Vent Cleaning The oven vent should be inspected regularly for debris build up To clean debris from the vent e Switch the main power switch OFF e Remove the three 3 screws securing the vent cover e Remove the vent cover e Mechanically remove debris from the wire screen with a stiff bristled brush e Inspect the vent to ensure that there is no excessive corrosion Corrosion may cause debris falling back into the instrument possibly compromising the readings and performance of the instrument e Replace vent cover if needed AZI P N 300 0472 e Reinstall vent cover with the three 3 screws 11 6 Fan Shroud Cleaning The fan shroud vent should be inspected regularly for debris build up To clean debris from the fan shroud e Switch the main power switch OFF e Remove shroud at tabs on top and bottom See picture at right e Inspect foam filter for debris build up Clea with compressed air if necessary e Reinstall shroud by pushing back into place at tabs with flat side of shroud against the foam filter AZI Customer Service 800 528 7411 or 602 470 1414 Page 82 of 92 11 7 Sample Pan Support The sample pan support may need to be cleaned if high volatile material condenses on it Turn the unit OFF Use
81. ococonononanananananononnonococononanannnnnns 73 10 PERFORMANCE VERIFICATION 74 A A A cad a a icuetlpedateene 74 10 2 The Precision Weight Test ini A lt x 75 10233 Sodium Tarate Test T E T 76 10 4 Precision Sodium Tartrate Testa cad it 77 103 RETAINS SANDS S tas ita 78 11 ROUTINE MAINTENANCE 2 0 22 0022 0022 00022 0 2 79 Mad echamber IMENO aaa dada 79 11 2 SMTOUA GaSK aaa das 80 MES INSUlator Cala 80 wA Oven Chamber clean utili 81 LL OWEN vent Clean distal 82 110 Fan Shroud Cleaning onara a r 82 Ned Way Sample Pan SUP 1 eaa nn S 83 11 8 Lithium Battery Replacement ooooooooooccccccccccccccnononananananananononononononononnnnnann nono nnnnnnnnnccononons 83 12 TROUBLE DIAGNOSIS 2 22 022 0002220 022 0 2222 22 84 13 ACCESSORIES 8 MAINTENANCE PARTS 87 14 MAX 5000 SPECIFICATIONS 88 15 WARRANTY ms 91 AZI Customer Service 800 528 7411 or 602 470 1414 Page 4 of 92 1 WARNINGS AND SAFETY INFORMATION The MAX 5000 moisture analyzer complies with the Underwriter Laboratories Inc standards and European Council Directives for Electrical Equipment for Laboratory Use electromagnetic compatibility and the stipulated safety requir
82. of 92 11 2 Shroud Gasket The shroud gasket creates a seal between the heater shroud and ceramic plate Over time this gasket will require replacement Check the shroud gasket AZI P N 300 0446 regularly for cracks or deterioration Perform a Kimwipe test to verify the seal s integrity See Kimwipe test below If Kimwipe pulls out easily then either the shroud gasket or insulator seal will need replacement j See Insulator Seal Section 11 3 below for more information If replacement is needed contact AZI customer service at 800 528 7411 602 470 1414 or by e mail at support azic com for further details and information Kimwipe Test e Place a Kimwipe single thickness on the ceramic insulator plate at the 3 o clock position e Close lid e Attempt to pull Kimwipe out from between the gasket and plate being careful not to rip the Kimwipe e Repeat at the 6 o clock and 9 o clock positions o If it removes easily then either the shroud gasket or insulator seal or both need to be replaced o If it does not pull out then the shroud gasket and insulator seal are o k 11 3 Insulator Seal Beneath the ceramic insulator plate is an insulator seal that should be checked for compression The above Kimwipe test should be performed If the Kimwipe is removed easily then the insulator seal needs to be replaced To replace e Switch the main power switch OFF e Remove the pan support stem and ceramic insul
83. of printing for each report can be custom configured from the menu system These same reports can be directed to a computer through the RS 232 serial port at the back of the instrument e Unpack the printer and cable Load the paper according to the instructions in the printer manual Plug the large end cable connector into its mating socket on the printer and the small end into its mating socket on the T back of the MAX 5000 Plug the power cord into the printer and into a GROUNDED power receptacle e Turn the printer power on On some older printers it is also necessary to place the printer On Line SAMPLE TEXT A e e 2 AZI Customer Service 800 528 7411 or 602 470 1414 Page 17 of 92 To set up the MAX 5000 begin at the Main Test Screen and press Menu center key below display e Select SET UP MENU press ENT e Select PRINTER SETUP press ENT e Highlight Parallel for printer only RS 232C for computer only and AUTO FORM FEED Seiko DPU Both for the printer and computer at canon the same time When the correct port selection is highlighted press ENT again e Move the select box to PRINTERTYPE and press ENT e Select the type printer you have and press ENT e Epson ESC P2 is for older style Epson printers e Epson Raster is for Epson C84 and newer e Seiko DPU is for the small thermal printer e Canon is for the older Canon Bubble Jet line of printers e Move the select
84. orm to the standard exponential drying model but will show a consistent moisture rate loss at some point in the process The parameter may be set for values from 0 0010 to 9 9999 loss rate For TIME the MAX 5000 ends the test after a preset time of up to 120 minutes has past This is sometimes useful for diagnostic and research purposes Use TEMP THEN to prevent a test from ending before the test temperature has been obtained This can be helpful for ashing which takes a while to reach the final temperature and for testing ultra low moisture samples which don t readily release moisture at the beginning of the test The MAX 5000 can be used to test a wide range of products from finely ground low moisture powders to high moisture slurries For best results the sample should be of a uniform particle size distributed evenly over the sample pan lf sample material is collected into a jar or sampling bag shake the jar or bag to produce a uniform mixture Liquid samples should also be mixed or shaken to ensure a more homogeneous mixture Large chunk samples such as charcoal or cookies should be crushed or ground to smaller pieces approximately 1 4 or less in diameter For best results take care not to over grind the sample causing it to heat up and causing moisture loss Some samples such as dried vegetables may also need to be ground The smaller pieces will easily release moisture during testing leading to shorter test times an
85. ple pans from the oven AZ Customer Service 800 528 7411 or 602 470 1414 Page 11 of 92 4 3 Inventory e Carefully unpack the instrument and locate the following items Part Name AZI Part Number MAX 5000 MAX 5000 Accessory Kit Y990 0158 Line Cord 110 VAC 200 0002 Test Weights 690 0003 690 0004 690 0025 Optional Accessories RS 232 Cable null modem for 6000 1044 computer interface Temperature Calibration Kit Y990 0193 Sample Pans Flat 100 ea 990 0010 Filter Paper 990 0003 See page 87 for a complete listing of available accessories AZI Customer Service 800 528 7411 or 602 470 1414 Page 12 of 92 5 INSTRUMENT SETUP Contact Arizona Instrument Customer Service at 800 528 7411 or 602 470 1414 or by e mail at support azic com if you have any questions about the following instructions The MAX 5000 is designed to provide reliable results under normal ambient conditions in the laboratory and on the production floor Compliance with the following guidelines will help ensure optimum accuracy in minimum time e Setup the MAX 5000 on a stable even surface that is not exposed to vibrations Select a location for the instrument that is Firm and near level gt Free from vibration that will affect the balance measurements gt Close to an isolated power outlet that will provide dedicated electrical power gt v and Away from the direct flow of a fan or heating air conditioning outlet that mig
86. r lines on input screens gt The Alpha numeric Keys toggle through the listed characters plus a blank space gt The ENT Key selects most functions Other functions are selected with keys below the screen As operating modes change the function of the keys below the display change The function performed by the key is displayed and changed with operating mode changes gt The ESC Key is used to exit menus Many functions that are Yes No or On Off type selections will be saved when the ESC Key is depressed If the field is editable with numbers or letters the ESC key allows an abort without changing contents AZ Customer Service 800 528 7411 or 602 470 1414 Page 16 of 92 5 3 Contrast Adjustment If necessary adjust the display contrast controls for better viewing of the display The left most soft key below the display has the word CONTRAST above it Press this key and the display changes to instructions USE on setting contrast Use the up and down TO E TR Ast arrows on the key pad until the display is ESC TO SAVE AND EXIT acceptable and press the ESC key to save the setting and exit back to the Main Test screen The LCD will darken after a few minutes of inactivity but will brighten with any key press 5 4 Connecting a Printer An optional graphic capable printer is available from AZI pictured is printer example only It connects to the parallel printer port at the rear of the instrument The content and time
87. rmine the proper ending rate if possible APPLICATION Enter the application type i e food plastics etc Voc output TARGET RESULT Enter the reference method result FILTER PAPER Liquids and creams may require filter paper to produce repeatable results Because filter paper is expensive try developing with No filter paper first APPLICA TION Food RESULT TYPE Moisture TARGET RESULT 1 0000 FILTER PAPER Ho AZI Customer Service 800 528 7411 or 602 470 1414 Page 59 of 92 USE A WAFFLE PAN THE TEST MATERIAL MUST BE HOMOGENOUS GRIND SHRED MATERIAL THAT 15 gt 1mm IN DIAMETER OR HAS A SKIN e This is an interim screen Ensure You PRESS G0 WHEN READY have the material properly prepared eThis is another interim screen You will A SERIES TESTS AT DIFFERENT see these throughout the Parameter TEMPERA TURES WILL BE RUN Expert testing Continue to press start to go TO DETERMINE THE HIGHEST l ACCEPTABLE TEMPERATURE TO to next step in the parameter development PRODUCE THE SHORTEST TEST TIME process PRESS START TO BEGIN TEST Add sample to pan until white bar is within green area Instrument will beep indicating CLOSE LID the proper weight Close lid M SAMPLE WEIGHT 10 37g TESTING WONTON os 7 e The test will complete and additional SERRE ES TUBES a MIOS tre instructions will come to the screen TIME 0 29 RATE 0 0073 Minute TIME LEFT 29 31 SAMPLE WEIGHT 10 3735 q
88. rs for a particular application The determination of the proper parameters requires a series of tests and good laboratory procedure Call your AZI technical sales representative at 800 528 7411 or 602 470 1414 for assistance AZI Customer Service 800 528 7411 or 602 470 1414 Page 69 of 92 If the moisture of the sample material 40 is known as determined by a 35 reference method the MAX 5000 s 30 temperature can be adjusted until test 950 results correlate with the established 20 moisture concentration If the 15 moisture is not Known prepare a 10 moisture versus temperature curve to 5 decide the proper test temperature for 80 90 100 110 120 130 140 150 a particular sample Moisture Vs Temperature Curve An appropriate curve can be prepared by testing sample material at increasing temperatures of five degree increments The starting temperature is usually the standard or reference method temperature for that particular sample Results for each increasing temperature will usually rise to a plateau where several temperature increases do not change the results significantly Beyond this plateau results can rise dramatically Examination of a sample tested at these increased temperatures may show burning or charring Results along the plateau usually correlate well with the standard or reference method results Repeated testing should be done at temperatures in this range Several tests are performed using the same sample
89. t f Memory Start 8 ENTER TO EDIT OR START TO BEGIN TEST COPY INSERTt INSERT4 PRINT gt So So oo On e l e To quickly create a new set similar to an old one use COPY from one location move the highlight bar to the desired location and INSERT 7 or INSERT into a new location e Highlight the memory start set you wish to copy e Press COPY e Using the Up and Down Arrow Keys highlight the location where you want the new memory start to be and press INSERT or INSERT The contents of the locations are safe Memory starts above the insert will not be affected and memory starts below the insert will be moved down the list Pressing INSERT 7 or INSERT will insert a new memory start either above or below the cursor location and move the following memory starts down AZI Customer Service 800 528 7411 or 602 470 1414 Page 37 of 92 AZI Customer Service 800 528 7411 or 602 470 1414 This command prints the stored parameters All of the numbered Memory Start Sets are printed giving the values of the parameters shown below in the print out example COMPUTRAC MAX5000 MEMORY START REPORT ID TST IDL HST SSZ SWN LFT T A TOT EC MODE 1 WEIGHT 105 025 025 80 1 0 100 NO RATE 1 000 MINUTE 2 TARTRATE 150 80 025 80 1 0 100 NO RATE 0 0250 MINUTE 3 LMH 135 80 25 40 05 100 NO RATE 0 1000 MINUTE 4 MILK MOIST 210 50 25 50 0 5 100 NO RATE 0 1000 MINUTE 5 MILK ASH 575 25 25 50 05 100 T NO RATE 0
90. terface module Calibration tolerance 1 5 Cup to 250 C 2 C from 251 C to 575 C Twelve calibration points from 50 C to 575 C Twelve preset verification points 50 C to 575 C or up to four user programmable verification points Programmable calibration reminder 2 C at temperatures between 250 C and 600 C Test Temperatures User programmable through front panel keypad to control idle temperature test temperature and high start temperature for all tests Multiple test temperatures are programmable for linked test sequences Heater Chamber Composite ceramic fiber lined heater chamber Isolated from electronics to minimize damage from corrosive sample materials Oven self cleaning cycle 550 C for 45 minutes Balance Integrated balance with automatic operation Balance capacity to 100 grams Resistant to nominal vibration with options available for isolation feet Resistant to air currents from internal exhaust fan Isolated from extreme temperatures in the test chamber Isolated from extreme changes in humidity caused by evaporation in the test chamber AZ Customer Service 800 528 7411 or 602 470 1414 Page 88 of 92 Balance Calibration Menu driven calibration and verification in the field Annual factory servicing recommended Programmable calibration prompt 200 milligrams to 100 grams 0 0001 moisture Y VGA 320 x 240 pixel color with adjustable contrast via contrast screen Test Display Current and c
91. the IP address Because the IP address is not known until the DHCP server assigns it intranet users will have to be given the IP address whenever the instrument powers up which is undesirable Manual mode is typically used because the IP address is fixed Automatic mode is useful to see if the Web Server works and then the user should switch to manual mode e Manual Requires the manual entry of IP Address Subnet Mask and Gateway Consult with your IS department for the correct settings e Ethernet Off Disables the Ethernet port The MAC address is factory set and Is unique to every Internet device It may be helpful to your Information Services IS department This displays the Ethernet status If everything is setup correctly then the Status should turn to Initialized shortly after power up Otherwise an error will be reported AZI Customer Service 800 528 7411 or 602 470 1414 Page 56 of 92 Special Features SPECIAL FEATURES This menu provides optional features for the instrument A Web Server and Parameter development tools are examples of additional features that can be purchased If you have ENTER CODE purchased one of these features you have been given an access code to enable disable the feature Entering the code will toggle the feature on off For more information regarding the special features available on your MAX 5000 please call your technical sales representative or AZI Customer Service at 800
92. the sodium tartrate through a 20 sieve to obtain a more uniform size that will achieve better results Lumps in sodium tartrate usually indicate moisture absorption that will yield inconsistent results Discard the larger chunks or pestle through sieve and return them to the bottle Sodium tartrate is available from AZI as P N 800 0037 Call AZI part sales at 800 528 7411 or 602 470 1414 for information or to order The sodium tartrate test uses approximately 8 grams of sample and is run at 150 C Results should be within a range of 0 1 of the assayed value on the label of the chemical Sodium tartrate assayed moistures typically range from 15 61 15 71 From the Main Test Screen press MEMORY Highlight Factory TARTRATE Press START When the prompt appears on the screen open the lid and carefully place a new clean sample pan on the pan support Close the lid and allow the unit to tare e When the prompt appears on the screen again open the lid and gently and evenly shake sodium tartrate over the surface of the sample pan until the display shows that 7 9 to 8 1 grams of sample are loaded Close the lid and the test will begin Wait for the end of test signal If this is the first test of the day discard the results of this test for stability reasons Results should be within a range of 0 1 of the assayed value on the bottle label AZI Customer Service 800 528 7411 or 602 470 1414 Page 76 of 92 Perform the sodium
93. the supplied tweezers to carefully remove the pan support up out of the test chamber Clean the support post and pan support arms by wiping with a soft cloth or paper towel Carefully reinstall the pan support into the balance stem Ensure the arms of the support are all level and about 1 16 to 3 16 from the heat shield See page 14 for inspection instructions 11 8 Lithium Battery Replacement If the clock fails to keep accurate time when the instrument is turned off for long periods the battery on the main control board may be depleted and need to be changed The battery is not considered to be field replaceable Contact AZI customer service at 800 528 7411 602 470 1414 or by e mail at support azic com for further details and information AZI Customer Service 800 528 7411 or 602 470 1414 Page 83 of 92 12 TROUBLE DIAGNOSIS SYMPTOM DIAGNOSIS REMEDY Final result is always Temperature too low Zero or near Zero see pages 268 70 table Sample too small Increase sample size Sel PS Lift compensation too high Reduce lift comp pct Final results vary too Test temperature incorrect Determine much Ending criteria incorrect temperature ending criteria to use or call AZI at 800 528 7411 or 602 470 1414 for assistance Sample size too small Increase sample size A PS connections see page 15 Microprocessor failure Turn unit OFF and back ON Call AZI at 800 528 7411 or 602 470 1414 for assistance Backlight connec
94. tify the percent moisture value AZI Customer Service 800 528 7411 or 602 470 1414 Page 78 of 92 11 ROUTINE MAINTENANCE The MAX 5000 requires routine cleaning to keep the test chamber free of excess sample material Failure to keep the instrument interior clean reduces its accuracy and repeatability Sample material collecting in the test chamber prevents free movement of the pan support In addition sample material collecting on the rim of the heater shroud prevents the lid from closing completely causing erratic results Material may fall into the system s interior causing balance problems Routine cleaning of the test chamber reduces the possibility of problems and repairs extending the life of the instrument 11 1 Chamber Interior The chamber interior should be cleaned when needed to eliminate the buildup of sample material To clean debris from the chamber s R interior e Switch the main power switch OFF e Use a vacuum line portable vacuum cleaner or a small paintbrush to remove debris from the area around the balance cover and pan support stem A compressed air line or canned air is not recommended since debris migrates into the system s interior and settles in the electronic components and balance stem Do not clean the shroud or heater element with a brush vacuum cleaner liquid or any object Use the Oven Chamber Cleaning procedure AZ Customer Service 800 528 7411 or 602 470 1414 Page 79
95. tor Call AZI at 800 528 7411 or disconnected 602 470 1414 for assistance Heat will not turn off Circuit board failure Call AZI at 800 528 7411 or 602 470 1414 for assistance AZ Customer Service 800 528 7411 or 602 470 1414 Page 84 of 92 Moistures too high or Temperature set too high Lower temperature 602 470 1414 for assistance Lift compensation too low Increase lift comp pct DN Po Dirty heater assembly Clean heater assembly M e o Ending criteria incorrect Determine new ending criteria see page 27 Call AZI at 800 528 7411 or 602 470 1414 for assistance Moistures too low or Temperature too low Increase temperature Sample size too small or too Adjust sample size Sample needs spreading Spread sample evenly eee seme Lift compensation too high Decrease lift comp pct see page 34 Ending criteria incorrect Determine new ending criteria see page 27 Call AZI at 800 528 7411 or 602 470 1414 for assistance Moisture never stops Other volatiles Try lower temperature Mere asIina see pages 30 amp 73 Balance drifting Call AZI at 800 528 7411 or 602 470 1414 for assistance No graphical display Graphics not selected Press GRAPH key at Main Test Screen No heat Voltage selection wrong Check power module at rear of instrument set to correct voltage See page 15 Wiring failure in heater circuit Call AZI at 800 528 7411 or 602 470 1414 for assistance Heater failure Call AZI at 800 528 7411 or
96. tual result is within 90 to 99 percent of the predicted result This ending criterion is less dependent on actual moisture loss than Rate gt Temp Then These features are used when rate or predict criteria alone would end the test after the release of free or bound moisture but before the full temperature is reached The MAX 5000 first checks to see that the temperature is within two 2 degrees of the test temperature before checking for the other criteria as previously described This can be helpful for ashing which takes a while to reach the final temperature and testing ultra low moisture samples which don t readily release moisture at the beginning of the test Sample Size Options SAMPLE SIZE OPTIONS This feature enters a target sample size and the minimum and maximum sample range Sample size ranges can be programmed from 0 1 gram to 99 9 grams The total difference in values SAMPLE SIZE 8 0 g between size and window must be between 0 grams minimum and 100 grams maximum SAMPLE WINDOW 1 0g INSIDE LIMITS Two Beeps gt Sample Size The sample size should be consistent for all tests OUTSIDE LIMITS No Beeps on a particular product in order to get the best standard deviation possible High volatile materials such as coatings typically require smaller sample sizes one to three grams Low moisture materials such as plastic resins typically require larger sample sizes The instrument is optimized for lo
97. unit in an appropriate fume hood only The fume hood must keep the work area below the lower toxic limit established by appropriate standards e Substances that release caustic vapors such as acids should be analyzed in the smallest sample size possible to still achieve satisfactory results If vapors condense on the MAX 5000 housing parts stop the testing wipe the surface clean with a suitable neutralizer and place the unit in a fume hood capable of removing the vapors e The user has the responsibility for carrying out appropriate decontamination if hazardous material is spilled on or inside the MAX 5000 CLEANING the MAX 5000 e Clean the MAX 5000 according to the cleaning instructions only Use of strong detergents and oven cleaners will damage the case and heater housing of the MAX 5000 e Before using any cleaning or decontamination method except those recommended by Arizona Instrument check with AZI Customer Service at 800 528 7411 or 602 470 1414 to ensure that the proposed method will not damage the equipment e The outside housing of the MAX 5000 should be cleaned with a mild household detergent such as Formula 409 and a soft lint free cloth Paper towels should never be used as they will scratch the instrument s finish and the clear lens over the display Since the MAX 5000 has a built in cleaning mode of operation no cleaning product should be necessary for the oven If the 550 C self cleaning cycle does
98. use in high vibration environment No User changes resulted Firmware Rev 2 2 Manual Rev G 1 Updated menu selection criteria throughout manual to update to new firmware configurations Firmware Rev 2 21 Manual Rev G1 1 Added ability to lockout special features on trials and loaners AZI Customer Service 800 528 7411 or 602 470 1414 Page 7 of 92 PACKING AND SHIPPING INSTRUCTIONS IMPORTANT Movement handling and packaging of the instrument must be done with EXTREME CARE to avoid permanent expensive internal damage The MAX 5000 uses a delicate and very sensitive electronic force balance to measure small weights and weight losses The balance is capable of measuring weight to a resolution of 0 1 milligram Careful handling is necessary to prevent damage to the balance With proper care the instrument will provide years of dependable service UNPACKING Unpack the instrument carefully and set it down gently Select a place where it will be safe from bumping jarring excessive air currents and vibration Retain all packaging materials for any future shipment of the instrument If the instrument is returned to AZI for any reason place it in the original packaging materials Only the original packaging materials have been tested and proven effective during shipment REPACKAGING FOR SHIPMENT e Call AZI Customer Service at 800 528 7411 or 602 470 1414 for Return Material Authorization RMA information prior to retur
99. w moisture testing with a 30g load Ashing typically requires a smaller sample size to reduce the residue that is emitted during testing The instrument is optimized for ashing with a 4g load Contact your technical sales representative for appropriate sample sizes gt Sample Window The number entered sets the allowable tolerance of the sample to be loaded AZ Customer Service 800 528 7411 or 602 470 1414 Page 30 of 92 gt Inside Limits and Outside Limits Selections of no sound one beep two beeps or three beeps may be programmed to have the instrument indicate when the weight of the sample is inside or outside the sample weight tolerance Tare Options Two tare options define how stable the balance must be before a test will start The third tare option is labeled Special Options and is used to tare a syringe foil tray or other lightweight device used to carry or hold the SAMPLE TARE Start After 8s sample on the pan gt Pan Tare Records the SPECIAL TARE Normal average empty pan weight after ensuring that its weight is sufficiently stable over a programmable period of time Taring continues until the weight is found stable lt 0 2mg deviation over the programmable period of time Temperature fluctuations occur in the test chamber dependent on a number of different criteria such as previous test temperature the amount of cool down time between tests and the current idle temperature The temperature fluctuations caus
100. y starts transfer the memory starts to disc and transfer the memory starts from disc to other instruments Memory start data is saved in a comma delimited format which can be read by MS Excel Although memory starts can be edited in Excel it is not recommended AZ Customer Service 800 528 7411 or 602 470 1414 Page 66 of 92 7 4 Test Results Test results can be viewed by selecting TEST RESULTS in the left bar Test results can be downloaded from the instrument to disc Test data is saved in a comma delimited format which can be read by MS Excel Graph data contains only the data points that are used by the instrument to generate the graph so the actual visual graph is not downloaded A http 192 168 2 149 m5k htm Microsoft Internet Explorer y lr File Edit View Favorites Tools Help Q back x S x B TN JO Search 52 Favorites O Ew bead Fan rel er 2o Address http 192 168 2 149 m5k htm Eco u Google molecular weight al C Search amp y E 2 blocked ALE Check AutoFill fea Options 8 a molecular l weight y A Autolink YY max 5000 AZI Preferred Customer SN 500002 CALIBRATION REPORT LCD VIEW PE WONTON 23 Feb 2006 MEMORY STARTS Pe WONTON 23 Feb 2006 Pe WONTON 23 Feb 2006 TEST RESULTS PE WONTON 23 Feb 2006 PE WONTON 23 Feb 2006 PE WONTON 23 Feb 2006 Pe WONTON 23 Feb 2006 Pe WONTON 23 Feb 2006 Pe WONTON 23 Feb 2006 TARTRATE 03 J

Download Pdf Manuals

image

Related Search

Related Contents

GR5 - Oerlikon Servicios > Welding Assistance  USER`S MANUAL - Moore Industries International  取扱説明書  HP Officejet K7100 User's Manual  Rinnai FS35ETRBN/US User's Manual    Manual de usuario de Webmail  User`s Guide Guide de l`utilisateur Guía del usuario  第13回外部評価委員会 配付資料01~09 (PDF:1423KB)  電気メス用付属品(未滅菌)(大人ポリヘッシブⅡ 対極板 コード無)  

Copyright © All rights reserved.
Failed to retrieve file