Home
        TO-Clean Manual
         Contents
1.     To fill the humidifier  release the clips on the metal strap around the bottle and  remove the metal strap  While supporting the humidifier  angle it away from the  TO Clean and unscrew the PTFE lid  Fill the humidifier about a quarter full with  HPLC grade water only  Do not fill more than half full  If contamination from room    d  10  WW WASSON ECE    INSTRUMENTATION    Hardware Description and Assembly    air is a concern  make sure to thoroughly purge the humidifier each time the  water is changed before cleaning cans     Vacuum Valve Assembly    The vacuum valve assembly is located at the back of the TO Clean and consists  of the vacuum valve and tee fitting  shown in Figure 6  Attach the vacuum valve  hose assembly from the dewar U tube to the bottom of the vacuum valve  If not  already connected  attach the 1 8    vacuum valve actuation air line to the 1 8     fitting on the vacuum valve  Attach the 1 8    nitrogen charge line from the  bulkhead labeled    To Tee Fitting     located on the back of the controller box  to    the tee on the vacuum valve assembly     11    NW  WASSON ECE    INSTRUMENTATION    Hardware Description and Assembly    Figure 6   Vacuum Valve Assembly                  Nitrogen charge gas      to bulkhead  labeled    To Tee  Fitting           Valve Actuation Air      to bulkhead labeled     Air or N2     60 psig   Max       Vacuum Hose  to Dewar  U tube    Controller Box    The controller box  located at the upper right side of the TO Clean  
2.     Vacuum Pump Default Settings      Description  Default Setpoint    A higher gas ballast allows the pump to handle  Gas Low Flow  SW condensable vapors such as water but also  Ballast  Position I     increases the pump ultimate pressure     Mode High Vacuum Selects high vacuum mode over high    Selector  small droplet  throughput mode        The pump performs most efficiently when the oil is warm  If cleaning cycles are  performed frequently the pump may be left on     For further information on the Edwards pump see the manual and CD included in    the shipment     Figure 1   Vacuum Pump       Vacuum Hose    to Dewar U   Tube    NAN 6  TI WASSON ECE    INSTRUMENTATION    Hardware Description and Assembly    Dewar    The dewar houses the U tube where contaminants are trapped  This aids in  cleaning the nitrogen entering the system  protects the pump from contaminants  in the canisters  and prevents any back diffusion from the pump to the system   Connect the dewar brackets to the right side of the TO Clean as shown in Figure  2  Slide the dewar into the brackets and place both the vacuum and nitrogen U   tubes into the dewar along with the provided insulating cap     Figure 2   Dewar Bracket    Dewar  Brackets                Attach the vacuum hose from the pump to the dewar U tube using the metal  clamp  as shown in Figure 3  Attach the vacuum hose  which goes to the  assembly in Figure 6  to the other side of the dewar U tube using the metal  clamp     NW  WASSON ECE    IN
3.     lines  and valves before operating the TO Clean  Always turn off  the flammable gas at its source when the instrument is not being  used  or before working on the instrument  Insure the TO Clean   vent and vacuum vent are plumbed to a safe location  free of any    ignition sources     The TO Clean is not designed to be used with toxic samples  without a secondary ventilation system  If the user must runa    toxic gas in the TO Clean  it should be housed in a ventilated       secondary enclosure designed for that purpose  such as a  laboratory fume hood      It is the responsibility of the customer to inform Wasson ECE if the  TO Clean has been used with toxic chemicals prior to any    instrument service being performed     When handling any chemicals for use in the TO Clean  all  applicable local and national laboratory safety practices must be  followed  This would include  but not be limited to  correct use of  Personal Protective Equipment  PPE   correct use of storage  containers  and correct handling of chemicals as defined in the  laboratory   s internal safety analysis and standard operating  procedures  Failure to adhere to laboratory safety practices could  lead to injury or death  Any disposal of chemicals must be in  compliance with applicable local and national laws     NW  WASSON ECE    INSTRUMENTATION    Table of Contents    Table of Contents    Instrument Description wisscacssdicsenssisccanesdidsancndecuacasdncsandndncesessiudsanieandvandeus 1  Product Feat
4.    N A WASSON ECE Save Method       INSTRUMENTATION       Method Name                              1 2 3ja4 sje 7 sel e    alw  el RITI Y  ul iol P    A  sjeler  G  H  J   Kit       zix c vip n m del  Save Cancel             Canister Cleaning    In the Start Menu  Figure 8  select Clean  The Clean Menu  Figure 17  displays  the Method that is currently loaded  To start cleaning using the loaded method    MAN SS  WU WASSON ECE    INSTRUMENTATION    Operation    parameters  select Clean  If no method is listed as loaded or a different method  is desired  select Load Method     Figure 17   Clean Menu          IN DEE Clean Menu  Method  Default Load Method  Clean Cancel                In the Load Method screen  Figure 18  scroll to the desired method using the  arrows on the side of the screen and choose Load     Figure 18   Load Method Screen       NW  WASSON ECE Load Method    INSTRUMENTATION       Fast Clean Method A VI   Test Method   Default Load  Cancel  Delete                   MAN e  TI WASSON ECE    INSTRUMENTATION    Operation    Once the method has been loaded the user is returned to the Clean Menu and  the loaded method is displayed on the screen  The method has now been   loaded and is ready to clean  Select Clean  The system will start to run the  method and the Cleaning status screen  Figure 19  will be displayed  Cleaning  status will display the status of the system including number of cycles remaining   cycle status  pressure and time remaining for the current c
5.   e 40 mTorr vacuum for 12  6 liter canisters   e Automated system leak test   e Electronic valve control   e Custom built trays for different canister sizes   e User defined methods   e Save up to ten methods   e Isothermal oven  controllable to 110  C    e Edwards RV 8 vacuum pump    NW  WASSON ECE    INSTRUMENTATION    Instrument Description  Operating Specifications    The TO Clean was designed to be used according to the following specifications   Any use outside of the intended purpose and specifications of this instrument  may result in damage to the system  bodily injury and possible loss of warranty        Feature    Specifications       TO Clean Model Number    TO0108       TO Clean Serial Number       General    TO Clean is designed to purge vacuum  sample canisters for cleaning per EPA  Method TO 14 TO 15        Ambient operation temperature    5 to 40  C  41 to 104 F        Canister Sample Type    Specially passivated stainless steel  canisters  Silco Can or  equivalent  Ambient Air Samples       Required Gases    High purity nitrogen and actuation air       Nitrogen supply input pressure  range    10   35 psig max  see page 3 for details        Actuation supply input pressure  range    55     60 psig max  see page 3 for details        Required Cryogen    Liquid argon for dewar       Required Water    HPLC grade water for humidifer       Venting Requirements    Purge vent and vacuum vent should be  vented to a safe location       TO Clean Power    120 VAC  50 60 
6.  the user is presented with the Start Menu  Figure 8  which allows    18  NW  WASSON ECE    INSTRUMENTATION    Operation    the user to select one of three options  Configure Methods  Diagnostics  or                         Clean   Figure 8   Start Menu  VEE Start Menu  Configure Methods Diagnostics  Clean  Diagnostics    The control software allows the user to diagnose the system for potential  problems  By selecting Diagnostics from the Start Menu the user can leak check  the system and actuate valves in the system to diagnose valve problems     The Diagnostics screen  Figure 9  allows the user to select Valve Control or Leak    Test the system     19    NW  WASSON ECE    INSTRUMENTATION    Operation    Figure 9   Diagnostics        IT Diagnostics  Valve Control Leak Test    Cancel             Valve Testing    All system valves can be tested using the Valve Control menu  Figure 10   This  menu allows the user to actuate the Pressure Valve  Vent Valve  and Vacuum  Valve  When the valve is actuated the user will hear the valve switch  If the  valve switch is not audible then contact the Wasson ECE service department for  support at  970  221 9179     Figure 10   Valve Control             e Valve Control  Pressure Valve  Off  Vent Valve  Off  Vacuum Valve  Off  Done             N A 20  TI WASSON ECE    INSTRUMENTATION    Operation    Leak Testing    Before running cleaning cycles  the system should be leak checked  If the  system is not leak tight  achieving lower vacuum will 
7. Hz  20A or 220 240V   50 60Hz  9 6A       Vacuum Pump Power    120 VAC  50 60 Hz  7 8A or 220 240V   50 60Hz  4 0A             Operation Continuous  Weight 525 lbs  238 kg   Dimensions 44 7     H x 49   W x 27    D       Oven Temperature Range    Room temperature   110  C       Soak Pressure Range    5 psig     30 psig       Vacuum Pressure Range    40 mTorr     1000 mTorr          Cycle Pressure Time       0 minutes     999 minutes       INSTRUMENTATION    NW  WASSON ECE          Initial Setup    Initial Setup  Gas Supply    Connect all gas lines to the proper fittings  as described in the Hardware  Description and Assembly section  Gas bottle regulators should be high quality  two stage regulators  Set the gas bottle pressures leading to the TO Clean to the  following gas pressures     pes Minimum Quality Maximum Set Pressure  Nitrogen or Zero Air Ultra High Purity 35 psig max       Note   Nitrogen or zero air pressures greater than 35 psig can damage the vacuum transducer     Valve actuation requires approximately 55 psig dry instrument air or nitrogen     Industrial compressed    house air    is usually inadequate for valve actuation  without additional treatment  Water and particulates can damage the actuators   Connect gases as described in the Hardware Description and Assembly section     Wasson ECE recommends leaving the nitrogen source connected and  pressurized unless you are changing bottles  When you need to change bottles   follow the procedures outlined in the D
8. RESTEK  di WASSON   ECE    INSTRUMENTATION          TO Clean Manual     Version 4 3        101 Rome Ct    Fort Collins  CO 80526  Ph   970  221 9176   F   970  221 9364      Copyright 2012 by Wasson ECE Instrumentation  Inc   MG 10 12   All rights reserved worldwide  Printed in the United States of America    This publication is protected by federal copyright law  No part of this publication  may be copied or distributed  transmitted  transcribed  stored in a retrieval  system  or transmitted into any human or computer language  in any form or by  any means electronic  mechanical  magnetic  manual or otherwise  or disclosed  to third parties without the express written permission of Wasson ECE  Instrumentation  Inc   101 Rome Court  Fort Collins  CO 80524     Disclaimer of Warranty and Limited Warranty    This publication is provided    as is    without warranty of any kind  Wasson ECE  Instrumentation  Inc  does not warrant  guarantee  or make any representation  regarding the use  or the results of the use  of this publication in terms of  correctness  accuracy  reliability  currentness  or otherwise  Further  Wasson   ECE Instrumentation  Inc  reserves the right to review this publication and to  make changes from time to time in the content herein without obligation of  Wasson ECE Instrumentation  Inc   to notify any person or organization of such    revision or changes     NW  WASSON ECE    INSTRUMENTATION    Note to the End User    This manual should be read in its enti
9. STRUMENTATION    Hardware Description and Assembly    Figure 3   Dewar Vacuum Hose Connections                 Vacuum FANN Vacuum Hose  Hose to GEM to Valve  Pump l Assembly              Connect the 1 8    line from the compressed nitrogen regulator to the 1 8    moisture  trap provided  Connect the other end of the 1 8    moisture trap to the 1 8    dewar  U tube fitting as shown in Figure 4  Connect the other side of the 1 8    dewar U   tube line to the bottom of the humidification chamber  described in the next  section  The dewar can be filled with liquid argon once the system has been  assembled and leak checked per the procedures in the Leak Testing section on  page 21  The 1 8    fittings on the charge gas line need to be disconnected to  remove the dewar lid  The user should be able to lift the insulating lid with the    vacuum hose connections intact     If the dewar needs to be moved to the liqud argon source  remove one of the  screws holding the top strap around the dewar  For safety  it is preferable if a    NAN 8  TI WASSON ECE    INSTRUMENTATION    Hardware Description and Assembly    liquid argon source can be plumbed to the system  Fill the dewar with liquid  argon to three inches below the top  exerting extreme caution not to expose skin  to the liquid argon  This should be enough cryogen for a 12 hour period  Place  the dewar back on the shelf and re screw the strap  if necessary  Carefully  replace the dewar lid and connect the fittings on the charge gas 
10. at any time by pressing    Cancel        Figure 12   Leak Test Status       II EH Leak Test  Leak Test Status    Pressure  12 23 PSIG  Pressure Change  0 23 PSIG  Time  10 00 min    Cancel             NW  WASSON ECE    INSTRUMENTATION    22    Operation  Method Configuration    From the Start Menu select Configure Methods  In the Configuration screen   Figure 13  the user is able to configure the method including Cycle Setup   Pressure Setup  and Canister End State     Figure 13   Configuration Screen       NW WASSON ECE Configuration    INSTRUMENTATION    Cycle Setup   Method Name  Default      Canister End State  Press  Setup                                                                Vacuum Pressurized        Load   Save   Accept    To configure a method  start by selecting Cycle Setup  In the Cycle Setup                      screen  Figure 14   the user can select the number of cycles desired by touching  the up and down arrows  Notice that for each cycle  both the Low and High  Pressure Cycle Times can be adjusted  Each cycle can have different low and  high pressure cycle times  Once the desired settings have been achieved select  Accept to return to the Configuration screen     MAN  WU WASSON ECE    INSTRUMENTATION    Operation    Figure 14   Cycle Setup       Oe Cycle Times    Number of Cycles  3 A    y  Cycle Number  1 A v  V    Low Pressure Cycle Time  1 00 min                High Pressure Cycle Time  1 00 min A V       Accept Cancel                From the Confi
11. be difficult if not impossible     First  the charge gas line should be tested for leaks using a pressure decay test   This is done by adjusting the regulator to pressurize the line to 35 psig  Once the  pressure gauge reading has stabilized  close the regulator  If the gauge pressure  remains constant after ten minutes  the charge gas line is leak tight  If the gauge  pressure does not remain constant  check the fittings  between the regulator and  V3 on the plumbing diagram  and perform this test again  Once the charge gas  line is tested the rest of the system can be tested for leaks using the Leak Test  Menu  Figure 11      From the Diagnostics menu select Leak Test  Using the arrows choose a  desired charge pressure and a soak time  If the system holds pressure to     0 4  psi the system is considered leak tight and will pass the test  If the system fails  the leak test secure all fittings and run the leak test again  The status of the leak  test is displayed in the Leak Test menu     The leak test feature can also be used to ensure the canisters have been  attached properly and that none of the canister fittings are loose before cleaning     21  NW  WASSON ECE    INSTRUMENTATION    Operation    Figure 11   Leak Test Menu    NW  WASSON ECE Leak Test       INSTRUMENTATION    Charge Pressure  30 PSIG A vy  Test Time  1 00 min A v    Start Cancel             As the leak test is performed  the Leak Test Status screen  Figure 12  is    displayed  The leak test can be stopped 
12. enough clearance in front of the TO Clean to open the door     38     and also to access the back of the controller box     Note   The TO Clean weighs 525 Ibs  238 1 kg   Do not attempt to lift this system by yourself  Use a    mechanical lifting device and secure edges to ensure the system does not tip over  Follow standard safety    guidelines when moving this system        Vacuum Pump    The TO Clean comes with an Edwards RV 8 vacuum pump  Unpack the pump  and place it on the floor next to the canister cleaner  Attach the vacuum hose to  the top of the pump  using the metal clamp  as shown in Figure 1     The oil mist filter and oil ballast have been pre assembled at the factory and are  connected by a short piece of tubing  Unscrew the gray oil ballast cap and screw  the oil ballast into the threaded hole until tight  Remove the clear cap from the  black mist filter fitting on the pump  Place the oil mist filter on top of the black    fitting and secure using the metal clamp     Unscrew the two gray oil caps on the pump and pour the oil that came with the  vacuum pump in either of the inlets  There is an oil level window on the back of  the vacuum pump  The oil reservoir should be 1 3     1 2 full  You should not   need the entire bottle for the first fill  Screw the oil caps back on before turning    the pump on     NW  WASSON ECE    INSTRUMENTATION       Hardware Description and Assembly    The pump settings have been set at factory to the values in Table 1     Table 1 
13. ently cooled  close all the canister valves   Close the valve on the nitrogen bottle    On the Start Menu  select Diagnostics    On the Diagnostics menu  select Valve Control     oa fF wh    On the Valve Control menu  click the Vent Valve  On  This will bring the  manifold and lines between V3 and V1 to atmospheric pressure   7  Click the Pressure Valve    On     You should hear nitrogen flowing out of  the purge vent   Once the regulator pressure reads    0     close the regulator   On the Valve Control screen click the Vent Valve and Pressure Valve     Off      10  Replace the nitrogen bottle   11  Open the nitrogen bottle valve and set the regulator to desired pressure   35 psig maximum     12  The system is now ready to run cleaning cycles     MAN e  WU WASSON ECE    INSTRUMENTATION    Operation    Powering Down the System    W    After the last cleaning cycle is finished  reduce the oven temperature  setpoint to 25  C  Open the door for faster cooling  Do not touch oven  components until they are sufficiently cooled    Once the oven is sufficiently cooled  close all the canister valves    If you are planning on disconnecting the nitrogen source follow the  instructions in the previous section    Shut off power to the TO Clean using the switch on the back of the  controller box  shown in Figure 7    Shut off the vacuum pump using the power switch located on the back of  the pump     29  WASSON ECE  INSTRUMENTATION    Methods    Methods  Default Method    The TO Clean is d
14. esigned around EPA Method TO 15 and comes with a Default  method based on these parameters  The table below shows these settings as    well as the EPA recommended method for non heated cleaning cycles     Customers may adjust these parameters to create methods for their specific    needs  as described in the Method Configuration section on page 23     Method    Canister  End  State    Low  Pressure  Cycle  Time   min     Table 2     Method Parameters    High  Pressure  Cycle  Time   min     Pressure   psig     Vacuum  Pressure   mTorr        Default    Vacuum    60    5    30       TO 15    Recommended    Procedure    without Heating       Vacuum    NW  WASSON ECE    INSTRUMENTATION    60    Not  specified    30       60    Not  specified    30                60       Not  specified       30          30    Maintenance    Maintenance    This section can help you solve operating problems that may arise while  operating the TO Clean     If you cannot solve the problem  call the Wasson ECE Instrumentation Service  Department at  970  221 9179     Troubleshooting    NW  WASSON ECE    INSTRUMENTATION    31    Maintenance    Wen   Symptoms Recommended Action    Setpoint is too low  Although the software can set the vacuum  pressure to 10 mtorr  the vacuum pump is not strong enough to  achieve this setpoint  Change vacuum settings to 40 mtorr     There is a system leak  Run the Leak Test method  If there is a  leak  check canister connections and fitting in controller box  If  Leak T
15. ess Steel U Tube  pressure trap    A2552 002       Edwards RV 8 Vacuum pump    70218       Edwards Oil  1 quart    70487       Edwards Oil Mist Filter    NW  WASSON ECE    INSTRUMENTATION          70219    36    Warranty Statement    Warranty Statement    Wasson ECE provides quality analytical systems that are designed  tested and  warranted to perform in a reasonable manner consistent with the performance  criteria as detailed in the Quotation from which the system was purchased  The  warranty includes all materials provided as part of the system such materials  originating with Wasson ECE  Agilent Technologies  or any other    third party     components incorporated into the solution and identified on the Quotation     Wasson ECE analytical systems are warranted to perform to design  specifications for a period of one year after shipment from the factory in  Colorado  Both materials and labor are included in the warranty in North  America where the sale is directly from Wasson ECE to the customer who is the  end user  Outside of North America  only materials are warranted  labor is the  responsibility of local distributors or other resale agents     This warranty excludes the end user   s operation of the systems in a manner  outside of their design specifications  Examples of this misuse include  but are  not limited to     e The application of inappropriate samples   e Over temperature of ovens  columns  or valves   e Low quality carrier gas  or out of carrier gas   e Par
16. est does not indicate a leak  check the connections of the  PVC tubing     System is not  reaching There is water in the U tube  Remove U tube from dewar and  vacuum thaw in a properly vented location   setpoint   The pump oil needs to be replaced  Follow procedures in the  vacuum pump manual     The vacuum valve is not actuating  Use diagnostics screen to  turn on the vacuum valve  There should be a loud click when it    actuates  If there is no sound  check the pressure of the  actuation air  H the valve is still not working call Wasson ECE for  a replacement     dee The U tube or check valve is frozen  Remove U tube or check  ret valve from the dewar and thaw in a properly vented location   not reaching  the pressure The pressure transducer is not reading correctly  Call Wasson   setpoint     ECE for assistance     Vacuum Pump       The charge pressure is not high enough or the pressure setpoint  The valves are   is too high  The system turns the pressure valve on and off as it  clicking but gets closer to the pressure setpoint  The charge pressure must  won t reach be set at at least 5 10 psig above the pressure setpoint or the  pressure set   valve will just keep clicking on and off  Please note that the  point  setpoint should not be greater than 30 psig as the system can  only handle a maximum of 35 psig   Pressure    A periodic maintenance and inspection schedule of the vacuum pump can lead to  increased uptime  reliable results  and longer instrument lifetime  Refer 
17. fittings on either side of the Pressure U tube    2  After the last cleaning cycle is finished  reduce oven temperature setpoint  to 25  C   Once the oven is sufficiently cooled  close all the canister valves   Remove the vacuum hoses on either side of the Vacuum U tube by  unscrewing metal clamps   Carefully remove U tubes from dewar cap   Allow the traps to come to room temperature in a properly vented  location   Pour out any moisture in the U tubes   Allow to dry completely  They may also be placed in a heater to facilitate  removal of organic components     9  Replace dewar cap and attach U tubes to respective fittings     This should be done on a daily basis  or more frequently  if needed      NAT 35  WW WASSON ECE    INSTRUMENTATION    Replacement Parts    Replacement Parts    Refer to the plumbing diagram on page 39 for the valve that needs to be    replaced     Wasson ECE Part Number  Valve    Item    Wasson ECE Part Number       O Ring Seals for Vacuum Hose and Tee  Fitting    25566       Jo Nuts with caps    12910 001       Jo Vespel ferrules    1499       Tee Fitting  NW25    25572       Metal Clamps for Vacuum and Tee    25570       Pressure Transducer  0 50 PSIG    25560       Pressure Transducer  15 mTorr    25562       PVC Vacuum Tubing     6 ft  3 ft    25574 003       Humidifier    25568 019       Humidifier cap    25568 020       Dewar  Pyrex Brand Glass  4300 mL    25568 022          2    Stainless Steel U tube  vacuum trap    A2552 004       1 8    Stainl
18. guration screen  Figure 13  select Pressure Setup  In the  Pressure Setup screen  Figure 15  cycle set the charge pressure and the    vacuum pressure for each cycle     Figure 15  Pressure Setup       II EE Pressure    Cycle Number  1 A v  Charge to Pressure  30 PSI A v    Vacuum Pressure  200 mTorr A v    Accept Cancel             Select Accept to return to the Configuration screen  Figure 13   Before the  method is saved choose a Canister End State  The canister end state allows the    user to select if the canisters will be under vacuum or under pressure at the end    24  NW    WASSON ECE    INSTRUMENTATION    Operation    of the cycle  If the canisters will need to be tested for certification after cleaning   select Pressure  The pressure of the canister will be left at the Charge Pressure  of the last cycle  If the canisters will need to be readied for sampling then select  Vacuum  The canisters will be left at the Vacuum Pressure of the last cycle   This pressure is set in the Pressure Setup screen  Figure 15      After the method has been configured the user can save and name the method  using the keyboard provided in the Save Method screen  Figure 16  by choosing  Save from the Configuration  Figure 13  screen  Type the name of the new  method and choose Save  After saving the method the software returns to the  Configuration screen  From the Configuration screen choose Accept  This will  return the user to the Start Menu  Figure 8      Figure 16   Save Method Screen 
19. houses the  electrical components  including the heater controller and the touch screen  controller  Operation of these components is described in the Operation section    of this manual     There are five 1 8    bulkheads located on the back of the controller box  shown in  Figure 7  If not already connected  attach the tubing shown in Figure 6  Valve  Actuation Air    and  Nitrogen Charge Gas    to the bulkheads labeled  Vacuum    12  NW    WASSON ECE    INSTRUMENTATION    Hardware Description and Assembly    Valve    and    To Tee Fitting     respectively and connect tubing from the humidifier  shown in Figure 5 to the bulkhead labeled    UHP Nitrogen        You will need to connect actuator gas to the bulkhead labeled    Air or Nitrogen    60 PSIG Max    and the bulkhead labeled    Purge Vent    to a proper vent location     Troubleshooting note     The vacuum valve requires pressures between 55 60 psig to actuate  Gas pressures outside of this range  will either fail to actuate the valve or cause damage to hardware components        Figure 7   Controller Box     Back       Plug the power cable to the back of the controller box and to the appropriate    electrical outlet  The power switch is located on the back of the control box  Set  the power switch to the    On    position  The fan on the back of the controller box  should start running and the heater and control screen on the front of the    controller box will turn on     13  NW    WASSON ECE    INSTRUMENTATION    O
20. isconnecting the Nitrogen Source section  on page 28     Wasson ECE does not recommend using the same nitrogen source for the valve  actuation and the charge pressure lines as improperly disconnecting the nitrogen  charge source could allow water to flow into the actuation valve line and cause    damage to the electrical components  If the same nitrogen source is used  two    N A 3  VW WASSON ECE    INSTRUMENTATION    Initial Setup    regulators will be required to control the pressures within the correct operating  range of each line     Cryogenic Liquid    The system requires liquid cryogen to fill the 4 3L dewar provided  Wasson ECE  recommends the use of liquid argon from a low pressure dewar or a bulk    laboratory source     Caution   Exert extreme caution when working with cryogen to prevent injuries from frostbite  asphyxiation  and    explosions  Always wear proper personal protection equipment  PPE  including gloves and eye protection    when working with cryogenic liquids        Power Requirements    The TO Clean system and vacuum pump are configured for 120VAC  50 60 Hz  or 220 240VAC  50 60 Hz  See the Operating Specifications on page 2 for    amperage requirements        NW  WASSON ECE    INSTRUMENTATION    Hardware Description and Assembly    Hardware Description and Assembly    Place the TO Clean on a bench top or on the optional system cart in an area with  appropriate electrical outlets and gas supply as described in the previous section   There needs to be 
21. line     Figure 4   Dewar with Nitrogen Charge Gas Connection    z y                         Nitrogen or  charge gas  through Dewar       Nitrogen or  charge gas  through Dewar                Humidification Chamber    The humidification chamber is filled with water and humidifies the nitrogen  entering the canisters  This aids in removing volatile organic compounds from the  canisters  If not already connected  attach the 1 8    nitrogen charge gas tubing  from the dewar to the top of the tee on the humidifier cap  Do not remove the  check valve between the dewar and humidifier  This prevents water from backing    WN WASSON ECE    INSTRUMENTATION    Hardware Description and Assembly    into the dewar and freezing the nitrogen charge gas line  Attach the 1 8    tubing  from the side fitting on the tee on the humidifier cap to the 1 4    bulkhead labeled       UHP Nitrogen    on the back of the controller box  shown in Figure 7     Note     The maximum pressure of the humidifier is 60 psig  If the nitrogen line exceeds this pressure  the    humidification chamber could explode  The user should take care to set the system pressure at the  maximum pressure of 35 psig        Figure 5   Humidification Chamber                     To    UHP N2     To Dewar Nitrogen or     bulkhead on back  Charge gas 1 8    1 of controller box  fitting e        Humidification          Humidification Chamber  Chamber Strap and E   Release Clip Z a         b      9   K  E    hi e  vi                   
22. lower display  The  upper display should show  110   use the  W  and   A  keys to adjust  the number if it is not set to  110      Press the tool key until  Loop  is in the upper display    Press the Advance   0    key until  t tUn  is in the lower display  The    upper display should show  yES   use the  W  and   A  keys to adjust          the parameter if it is not set to  yES       15  NW  WASSON ECE    INSTRUMENTATION    Operation    14     15     16     17     18     19     20     21     22     23     24   25     26     27     28     29     30     31     Press the     key until  Loop  is in the upper display  Use the  V   and   A  keys to scroll to  Ai  in the upper display    Press the Advance  0   key until  Ai  is in the lower display  The   upper display should be set to  2     Press the Advance  0  key to scroll to  SEn  in the lower display   Use the  V   and           nu    keys to scroll to  r0 1H  in the upper display              Press the Advance   0    key to scroll to  rt L  in the lower display   Use the  VW  and   A  keys to scroll to  2  in the upper display   Press the     key to get back to  Ai  in the upper display  Use the   V  and  A  keys to scroll to  dio  in the upper display    Press the Advance   0    key to scroll to  dir  in the lower display     Use the  V  and  A  keys to scroll to  in  in the upper display        Use the Advance  0  key to scroll to  LEv  in the lower display  Use  the  V  and  A  keys to scroll to  Low  in the upper displa
23. o Dewar U tube from the pump inlet port   Figure 1      MAN 33  WU WASSON ECE    INSTRUMENTATION    Maintenance    Remove the filter and O ring and inspect for cleanliness   If needed  remove the O ring from the filter and wash the filter with a  proper cleaning solution and allow it to dry  Do not let the O ring come  into contact with the cleaning solution    4  Re assemble the filter  O ring  and Vacuum Hose to Dewar U tube     Inspect and Clean the Gas Ballast Control    1  Remove the oil ballast  Figure 1  from the top of the pump   2  Wipe with a clean dry cloth and check that the air hole is not blocked   3  Screw the oil ballast back into place     Clean Oil Level Sight Glass           Drain the oil as described above     2  Undo the two screws at the top and bottom of the oil level window and  remove the bezel  window  and O ring    3  Clean the screws  bezel and window with a proper cleaning solution    4  Wipe the O ring with a dry cloth    5  Re assemble the oil level window    6  Fill the pump with oil and check for any leaks through the window     Clean the Motor Fan Cover and Enclosure    1  Turn the pump OFF and disconnect from the electrical supply   2  Use a dry cloth and soft brush to remove dirt and deposits from the fan     cover and enclosure     Dewar U Tube Traps    34  NW  WASSON ECE    INSTRUMENTATION    Maintenance    The Vacuum U tube and Pressure U tube housed in the Dewar will periodically  need cleaning  To do this     1  Remove the 1 8    tube 
24. ol in the upper display    35  Press the     key until the Main display appears again  showing the  temperatures in the upper and lower displays    36  To enter the Oper Menus press and hold both  VW  and  A  until the       lower display says  OPEr   Use the  VW  and  A  keys to scroll to   ALM  in the upper display    37  Press the Advance  0  key until the lower display shows  A hi   Use  the  V  and  A  keys to change the upper display to  300     38  Press the     key until the Main display appears again  showing the    temperatures in the upper and lower displays     Autotuning Controllers    You may occasionally need to perform an autotune to ensure the Watlow  controller is working optimally     Before autotuning  make sure the controller hardware is installed correctly  and  the sensor type and function  along with the scale setting  are set in the  configuration parameters  First  enter the desired set point or one that is in the  middle of the expectant range of what needs to be tuned  Next  turn on the  autotuning by following the steps below   1  From the Main display  press the Advance  0   key until the lower  display shows  AUt1   Use the  V  and  A  keys to change the  upper display to  yES   When autotuning is working  the display will       flash back and forth between the current display and the upper display    17  NW  WASSON ECE    INSTRUMENTATION    Operation    showing  tUn1  and the lower display showing  Attn   Once  autotuning is finished  the contr
25. oller will start controlling at the normal set  point  using the new parameters    2  Press the      key until the Main display appears again  showing the    temperatures in the upper and lower displays     Connecting the Canisters    Place the 6 liter canisters upside down in the slots in the oven shelves and  connect them to the 1   4    Swagelok fittings provided with the steel flex tubing  It is  easiest to start loading cans on the top shelf in the back and work forward and  down  It is helpful to make sure the valves are facing forward when connecting  the canisters  Do not open the valves until all positions are connected so the  canister contents can be properly directed to the vent  If there are any positions    that are unused ensure that position is plugged with the provided 1   4    Swagelok    plug   Connecting 1 Liter Canisters    If the optional upgrade to 24   1 Liter Canister capacity was purchased  the  system will include 12 sheet metal inserts  These are removable and are  fastened to the standard rack with four screws each  The 1 liter canisters are  placed upside down over the opening in the insert  The steel flex tubing will have  a removable tee with two connections to handle double the canisters  To convert  back to 6 liter canisters  remove the metal inserts by unscrewing the 4 screws   and remove the tees from the flex tubing lines     Software Control    When the controller is powered on using the power switch located on the back of  the controller
26. peration    Operation    To start using the system  the TO Clean and the vacuum pump must be powered  on  The power switches are located on the back of the controller box and the  back of the vacuum pump  Once these have been switched on  the following  sections can be used to set the oven temperature and configure cleaning  methods     Setting the Oven Temperature     Watlow EZ Zone PM  Controller    The oven temperature is controlled by the Watlow controller located on the front  of the control panel  Turn on the TO Clean to power up the controller  Make  sure the oven door is closed to operate the controller  as a safety latch is  installed to shut off heaters when the door is open  Set the controller  temperature to the temperature at which you wish to clean the canisters  To  change the operating temperature  press the   V  or  A  arrow keys on the  Watlow controller  The lower display will show the set temperature  the upper  display will show the actual temperature  The oven   s maximum temperature is  110  C     Note     If you are using Restek canisters  the blue Parker valves can withstand heating to 250  C  but the gauges    on the 3 way valves can only be heated to 80  C  Please verify what type of valve gauge is on the canister    before setting the oven temperature        The Watlow parameters have been factory set  If you need to re set these  parameters  follow the instructions described below  Complete information about  the Watlow controller and all default 
27. rety  prior to use of the instrument  Once    read  please follow the instructions in the    Initial Setup    section of the manual in    setting gas pressures and supplying appropriate power to the TO Clean     A       This symbol alerts the user to a potentially hazardous situation   The user should refer to this section of the manual before    operating the equipment     Important Safety Warnings    A  A  A       Parts of the TO Clean operate at temperatures high enough to  cause serious burns  You should always cool the TO Clean to  room temperature before working on the system  If you must  perform maintenance on hot parts  use a tool and thermally  protective gloves     If the TO Clean is connected to a power source  even if the power  switch is off  potentially dangerous voltages exist  The hazardous  parts are shielded by covers  With the covers in place  it should  be difficult to accidently make contact with dangerous voltages   Before performing any procedure that requires removal of a cover   disconnect the gas chromatograph from the power source     If the power cord insulation is frayed or worn  the cord must be  replaced  Contact Wasson ECE for a suitable replacement part     The TO Clean is not designed for use with flammable samples  If  using a flammable sample with the TO Clean  be aware that a  leak in the system may cause a fire or explosion hazard  In any  application using flammable gases  leak check all connections     NW  WASSON ECE    INSTRUMENTATION
28. telemeemeanaeeiere 34  Replacement E e CN 36  UE E E E 37  CE Certification     Declaration of Conformity         rrrrrnnnnnnnnnvvvvnnnnnnnnnnnn 38  Plumbing Diagram  vececeiciecasesssesewecicerennsaneresecsnecrdsssnersdensnerenciiveranniineswees 39    NW  WASSON ECE    INSTRUMENTATION    Instrument Description    Instrument Description    TO Clean is a revolutionary canister cleaning system developed by Wasson ECE  Instrumentation  The TO Clean canister cleaner allows for fully automated  canister cleaning  The TO 14 15 compliant TO Clean system was designed to  take the guess work out of canister cleaning  The system has a capacity of  twelve  6 L canisters  increasing your lab   s efficiency  The system is fully  automated  allowing the user to start a cleaning cycle and walk away  Using the  touch screen controller  up to ten custom methods can be defined and loaded as  needed  The system also comes with an automated leak check method to  ensure system performance  This provides a high performance system that is  easy to use and consistent     An on board touch screen embedded controller makes system operation as easy  as pressing start  The user can define the parameters for each method  including  number of cycles  pressure  and soak times  Once the method has  been defined  it is saved and is accessible for later use     Product Features  e EPA Method TO 14 15 compliant    e 12 canister capacity  expandable to 24 one liter canisters  e Embedded touch screen controller 
29. ticles  precipitates  or heavy liquids introduced   e Power fluctuations such as over voltage or off frequency  e Method loss  corruption  or modifications not authorized  e Fire or mechanical damage from the environment    e Damage from transportation outside of contracted scope    Revised August 1  2007    MAN S    TI WASSON ECE    INSTRUMENTATION    CE Certification     Declaration of Conformity    CE Certification     Declaration of  Conformity    NW  WASSON ECE    INSTRUMENTATION    38    Plumbing Diagram    Plumbing Diagram    NW  WASSON ECE    INSTRUMENTATION    39    
30. to Table  3 for the recommended maintenance schedule for the vacuum pump   Maintenance instructions are described below  For further information refer to the  Edwards Instruction Manual CD which is included with the system     MAN 32  WU WASSON ECE    INSTRUMENTATION    Maintenance    Table 3   Recommended Maintenance Schedule for Vacuum Pump    Maintenance Frequency  Check oil level Weekly  Replace the oil Every 3000 hours of operation             Inspect and clean inlet filter Yearly  Inspect and clean gas ballast control Yearly  Clean oil level window Yearly   Clean motor fan cover and enclosure Yearly                   Check Oil Level    The user must ensure that there is sufficient oil in the vacuum pump  This can be  quickly monitored using the oil level window on the back of the pump  If you need  to add more oil  follow the steps in the Hardware Description and Assembly  section of this manual on page 5     Replace the Oil    1  Operate the pump for ten minutes to warm the oil   Turn the pump    Off      Place a block under the front of the pump to tilt it toward the drain in the  back   Unscrew one of the oil caps on the top of the pump   Place a vessel under the drain and remove the drain plug   Replace drain plug and fill the pump with new oil as described in the  Hardware Description and Assembly section of the manual on page 5   7  Replace oil caps and remove block before operating the pump     Inspect and Clean the Inlet Filter    1  Disconnect the Vacuum Hose t
31. ures ERE rnrn tesenn tEEnn Enese n nnen n n 1  Op    rating Ee Le E 2   Initial Set  p EN 3  G   SUN seede E 3  Cryogenic LVL 4  Power e TE 4   Hardware Description and Assembly       nnnnnvnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnen 5  VEC UU Pumpueaaesseamsmsiansesannuiakkeinaitensnuneasheinetkvinnskein 5  Ee 7  Humidification Chamber E 9  Vacuum Valve Assembly L  rum  sne 11  C  ntroll  r  gt  REE EE EE 12   el  EE nen aeknnnsinibden 14  Setting the Oven Temperature     Watlow EZ Zone PM Controller                   14   Setting your Watlow Parameter  15  Autotuning Controllers            ccccccceeeseeeeeeeeceeeeeceaeeeeneeceeeeesaaeeesaaeseeeeeseaeeesaeeseaaeseeeeesaas 17  Connecting the Canisters LL Lansoreusamaiednmsmmaridreimsmmddisemmmsidn 18  Connecting 1 Liter Canisters             eccceceeceeeeeeeeeeeeeeeeeeceaeeesaaeeeeaeeseeeeeseaeessaeeeeneeenaees 18  EHNEN eege ee eege eege 18  Nell te 19    NW  WASSON ECE    INSTRUMENTATION    Table of Contents    Method Configuration           ccccccccceeeeeeceeeeceeeeecaeeeeaeeseeeeesaeeesaaeseeeeeseeeetaeeseaeesenneess 23  Canister GICANING DEE 25   d ERR 27  Safe Shutdown Geen te 28  Disconnecting the Nitrogen Gource nere renrnrresnnrnnennnnnrnnenenn 28  Powering Down the System       msrnnannnnrnrrnnnnrrnnnnvnnenvnnrvrrennnrnnennnnnennenrsrreennrnssnnnnnennenenn 29  Methods EE 30  Default Method WEE 30  Ee EEN EA  Tren 31  Vacuum PP eee 32  Dewar U Tube Traps i isscssscssinentcierenasessiceneebenasanttesuenntbenmeeseeanacnne
32. values can be found in the Watlow Series  EZ Zone PM PID Controller User   s Manual on the CD included with the shipment     14  NW  WASSON ECE    INSTRUMENTATION    Operation    Setting your Watlow Parameters     gt A    o OD NOWA    11     12   13       To enter the Setup Menus  press and hold both   V   and   A  keys until       the lower display says  SEt   The upper display should give you several  different options  Use the   V   and   A  keys to scroll through these  options    Scroll to  gLbL  and hit the Advance key  0    C_F  should show up  in the lower display  The upper display should be  c   Use the       and   A  keys to scroll through the options for the upper display    Hit the    to return to the  gLbL  display    Use the  VW  and   A  keys to scroll to  Ai  in the upper display and hit  the Advance   0    key     Keep pressing the Advance   0  key until  SEn  is in the lower display        Hit an arrow key to scroll to  tc  in the upper display    Hit the Advance       key and scroll to  Lin  in the lower display    Use the  VW  and   A  keys to scroll to  H  in the upper display    Hit the     key two times  The upper display should show  Loop   use    the  V  and  A  keys to adjust the parameter if it is not set to  Loop         Press the Advance  0   key until  L SP  is in the lower display  The    upper display should show  0   use the  VW  and  A  keys to adjust the  number if it is not set to  0      Press the Advance       key until  h SP  is in the 
33. y    Use the Advance  OI key to scroll to  Fn  in the lower display  Use  the  V  and  A  keys to scroll to  oFF  in the upper display    Press the     key to get back to  dio  in the upper display    Use the Advance  0   key to scroll to  dir  in the lower display  Use  the  V  and  A   keys to scroll to  Ot Pt  in the upper display    Use the Advance  0   key to scroll to  Fn  in the lower display     Use the  VW  and   A  keys to scroll to  ALM  in the upper display              Use the Advance       key to scroll to  Fi  in the lower display  Use the   V  and  A  keys to scroll to  1  in the upper display    Use the     key to get back to  dio  in the upper display  Use the    V  and  A  keys to scroll to  ALM  in the upper display    Press the Advance  0   key to scroll to  A  ty  in the lower display   Use the  V   and   A  keys to select  Pr AL  in the upper display   Press the Advance   0    key to scroll to  Sr A  in the lower display     Use the  VW  and   A  keys to scroll to  Ai  in the upper display     INSTRUMENTATION    16  NW  WASSON ECE    Operation    32  Press the Advance  0  key to scroll to  iSA  in the lower display  Use  the  V  and  A   keys to scroll to  2  in the upper display    33  Press the Advance  0  key to scroll to  A hy  in the lower display   Use the  VW  and         nu    keys to get to  1  in the upper display              34  Press the Advance  0   key to scroll to  A  L9  in the lower display   Use the  V  and   A  keys to scroll to  AL 
34. ycle  Using the Cancel  button the cleaning can be cancelled at any time     Figure 19   Cleaning Status       N A WASSON ECE Cleaning    INSTRUMENTATION       Cleaning Status  Cycles Left  1   Cycle Status  Venting  Pressure  12 23 PSIG  Time  10 00 min    Cancel                      Initial Use    The TO Clean system components have been tested to be free of contamination   However  contamination can occur during transport  from exposure to laboratory  air  or from impure water in the humidifier  Before using the system for the first  time  it is recommended that the user purges the system extensively with  humidified air or nitrogen before cleaning canisters  This can be done by setting  up a method with multiple cycles and running the method with the manifold  capped     27    WN WASSON ECE    INSTRUMENTATION    Operation   This procedure is also recommended anytime the water in the humidifier is  changed or the system has been exposed to room air for an extended period   Safe Shutdown Procedures    Disconnecting the Nitrogen Source    To ensure uninterrupted operation of the TO Clean  the following procedures  must be followed when reducing nitrogen supply pressure or changing nitrogen       bottles  Failure to follow these procedures may result in water from the humidifier  backing into the dewar U tube and freezing     1  After the last cleaning cycle is finished  reduce oven temperature set   point to 25  C  Open the door for faster cooling    Once the oven is suffici
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Samsung SC4530 Instrukcja obsługi    計測データワイヤレス通信システム U-WAVE-T 改版6  Conceptronic 2.5"/3.5" HDD Station USB 2.0  Instrumentos multifunción para la verificación y análisis de  User Guide - TP-Link  Minox 35 ML Owner's Manual    Fujitsu 9368987046 User's Manual  LX7 II - ヒビノプロオーディオセールス Div.    Copyright © All rights reserved. 
   Failed to retrieve file