Home

Pioneer User Manual

image

Contents

1. END UNIT 1 END LOCK __ END CAL 7 GBL RST po Yn sad ke ad N _InCal models only Not available in InCal models Pioneer EN 6 The mode menu has the following sub menus pd A Ero SET OFF l ENA ER DYN gt A F o l A l SET OF SET OFF l DENSITY SET ON SOLID WATER T20 0C On ga Ser Oe SET OFF OTHER 01 00000g cc l Vol l LIQUID INKER 010 000g cm END MOD Note Use No key to change the temperature value Enter Menu When the balance is on press and hold Unit Menu until PEM appears Release the button and the Calibrate CAL menu will display Menu Navigation Select menus menu items and settings through use of the Yes and No buttons Solid arrows point to the content displayed when Yes is pressed Dashed lines when No is pressed Changing Settings To select the displayed Setting press Yes To move to the next Setting press No Exit Menu When Efid is displayed press Yes to exit the menu function or press No to return to the Cal menu Note Press and hold No at any time to exit quickly 4 2 Calibration Menu CAL InCal or Span calibration should be performed daily and when the room temperature changes InCal inf AL calibrates the balance using an internal mass Span Calibration SF AM uses two weight values zero and a weight between 50 and 100 of the capacity of the balance Linearity calibration t 7 uses three weight values
2. SET ON SET OFF Dynamic Mode d n SET ON OFF Auto SET ON OFF Density Mode dEn5 4 SET ON OFF Solid Liquid Water Other T 20 0C 01 00000 Sinker 010 000 ml When T 20 0C 01 00000 010 000 occur use No to change the number and Yes to accept the change 4 5 Unit Menu tin t The Unit menu is used to enable or disable a specific unit SET ON SET OFF The unit is indicated by a small character next to Unit in the display g grams The default setting is Grams SET ON and all other units SET OFF T Units When unit t displays press Yes to show the T Unit settings SET OFF SET TT Taiwan Tael SET TH Hong Kong Tael SET TS Singapore Tael SET TO Tola or SET TI Tical M Units When unit m is displayed press Yes to show the M Unit settings SET OFF SET MO Momme or SET ME Mesghal Custom Unit Custom Unit C is used to create a unit of measure not provided with the balance The Custom Unit is defined by a factor a multiplier E and a least significant digit LSD The balance will use this to convert grams to a custom unit of measure Example 1 gram 0 257206 Avoirdupois Dram using a 4100g x 0 01g balance To create a custom unit press Yes when unit c is displayed Factor The Factor F is a value from 0 1000000 to 1 999999 When the Factor is displayed the first digit is flashing Press Yes to accept its value and activate the next digit
3. 2 OHAUS Pioneer Balances Instruction Manual EN 1 1 INTRODUCTION 1 1 Safety Precautions Please follow these safety precautions M Pioneer e Verify that the AC Adapter input voltage matches the local AC power supply e Use the balance only in dry locations e Do not operate the balance in hostile environments Do not drop loads on the platform e Do not place the balance upside down on the platform or platform mounting cone e Service should be performed only by authorized personnel 1 2 Controls Button Functions O T On Short Press Off Long Press Yes Short press Menu Unit Short Press Menu Long Press No Short press Menu Long press Menu Print Short Press Cal Long Press 2 INSTALLATION 2 1 Package Contents 0 1g and 0 01g Models Instruction Manual Power Adapter Balance Pan Pan Support Wind Ring InCal models only Warranty Card Turns balance on sets display to zero Turns balance off Selects or accepts setting Steps through active units and modes Enters Menu Steps through available settings Exit menu or abort out of menu item Sends data Initiates Span Calibration 0 001g and 0 0001g Models Instruction Manual Power Adapter Balance Pan Glass Doors and Panels Warranty Card Pioneer EN 2 2 2 Install Components 0 1g and 0 01g Models 1 Install Wind Ring InCal only 2 Install Pan Support 3 Install Pan 0 001g and 0 0001g Models Instal
4. EN 5 Pioneer Press the right O T button to accept the weight as liquid weight The density is calculated and result and SE are shown alternately Press the right O T button to stop the blinking and go back to ready status Begin Density Liquid mode The display shows A and LE SHt alternately Place the sinker on the on density determination kit the display shows the weight value Press the right O T button to accept the weight as air weight The display shows t Zu and LE SHE alternately Place the sinker on the density determination kit in the liquid please make sure that the entire sinker is submerged in the liquid the display shows the weight value Press the right O T button to accept the weight as liquid weight The density is calculated and result and Sf are shown alternately Press the right O T button to stop the blinking and go back to ready status 3 6 Weigh Below Feature n am i Remove Weigh Below Cover Attach wire or string to Hook Suspend sample 4 SETTINGS 4 1 Menu Navigation gt OUT STA f 1 i i i SPAN FILTER i i i 1 LINEAR ATARE i DN KILOGRAM CONTENT HAND SH MODE i 1 7 i i I I I I 1 f I I LI fi CAL ADJ GRAD _ END SET EEN EE NO gt NOT ALL UNITS SHOWN YES
5. is displayed blinking Press the right 0 T button to stop the blinking and go back to ready status Begin Dynamic Auto mode lf there is no weight less than 10d the display shows Fut and LE 54t alternately Place object s on weighing pan more than 10d the dynamic weighing will be started automatically The count down is shown on the display eg t 5 t 4 t 1 Once the averaging is finished the result is displayed blinking Remove the object s or press the right O T button to clear the result and go back to ready status 3 5 Density Mode Use this mode to determine an object s density A density determination kit is needed to use this mode For liquid density determination one more sinker is needed To access the Density mode press Unit until the display shows AE NS 4 After accessing this mode the two O T buttons will have different functions as shown in below picture On gt 0 T lt Off gt 0 T lt a Yes Zero Tare Accept Begin Density Solid mode The display shows A and LE 3Ht alternately Place object s on the density determination kit in the air the display shows the weight value Press the right 0 T button to accept the weight as air weight The display shows 3u d and LE 3Ht alternately Place object s on the density determination kit in the liquid please make sure that the entire object is submerged in the liquid the display shows the weight value
6. or No to edit When editing press No until the desired value appears then press Yes to accept Repeat until all digits have been accepted When the Factor flashes on the display press Yes to accept or No to re edit Example F 0 257206 Multiplier E The settings are E amp Fx1 E Fx10 E 2 Fx100 E 3 Fx1000 E 3 F 1000 E 2 F 100 and E d F 10 Press No to display the next setting Yes to accept Example E 0 Note The multiplier selections are limited when the capacity in grams is exceeded LSD The Least Significant Digit LSD is the number of displayed divisions d by which the weight is incremented The values are 1d 2d 5d 10d 100d or 0 5d Press No to go to the next setting press Yes to accept Example LSD 1d Note LSD options are limited if the readability in grams is exceeded The example custom unit will display 5 25 C when 1g is placed on the pan 4 6 Print Menu Pr ant Output Stable Hut St Ab Data will only be sent when the Stable indicator is on This setting works with manual button pressing or continuous and interval Auto Print SET ON OFF Auto Print A Pr at Data will be continuously sent when Cant mu is set Interval Er sends data every 1 to 3600 seconds When Stable St ALLE will send data when the balance detects a new stable reading This can be a weight value only t GA or it can also include a stable zero t 4ZEr al GFF disables automatic
7. ref wt x in grams Set ref wt x in grams 6 3 2 RS232 DB9 Pin Connections Pin 2 Balance transmit line TxD Pin 3 Balance receive line RxD Pin 5 Ground signal GND Pin 7 Clear to send hardware handshake CTS Pin 8 Request to send hardware handshake RTS EN 13 Pioneer 6 4 Compliance Compliance to the following standards is indicated by the corresponding mark on the product This product conforms to the EMC Directive 2004 108 EC the Low Voltage Directive 2006 95 EC and the Non Automatic Weighing Directive 2009 23 EC The complete Declaration of Conformity is available from Ohaus Corporation AS NZS4251 1 AS NZS4252 1 P CAN CSA C22 2 No 1010 1 92 UL Std No 3101 1 C US FCC Note This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense Industry Canada Note This Class A digital apparatus complies with
8. Canadian ICES 003 ISO 9001 Registration ISO 9001 Registration In 1994 Ohaus Corporation USA was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International BVOI confirming that the Ohaus quality management system is compliant with the ISO 9001 standard s requirements On June 21 2012 Ohaus Corporation USA was re registered to the ISO 9001 2008 standard LIMITED WARRANTY Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period During the warranty period Ohaus will repair or at its option replace any component s that proves to be defective at no charge provided that the product is returned freight prepaid to Ohaus This warranty does not apply if the product has been damaged by accident or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other than Ohaus In lieu of a properly returned warranty registration card the warranty period shall begin on the date of shipment to the authorized dealer No other express or implied warranty is given by Ohaus Corporation Ohaus Corporation shall not be liable for any consequential damages As warranty legislation differs from state to state and country to country please contact Ohaus or your local Ohaus dealer for further details
9. Reference Weight Press Yes to establish a new Reference Weight Put the reference sample on the pan and press Yes to accept or No to abort Percent Place the object s to be compared to the reference weight on the pan 3 4 Dynamic Mode Use this mode to weigh an unstable load such as a moving animal To access the Dynamic mode press Unit until the display shows dr Two different start reset modes can be selected Manual start and stop via key press Automatic start and stop automatically The default Averaging Time is 5s After accessing this mode the two O T buttons will have different functions as shown in below picture On gt 0 T lt Off gt 0 T lt Yes Zero Tare Accept Set up Averaging Time The default Averaging Time is 5s The initial display shows t 5 The time t5 is blinking Press No to scroll the time selection 5 10 15 20 and 30 The time unit is s second Press the right 0 T button to select the desired option After time is set dynamic weighing will begin Begin Dynamic Manual mode lf there is no weight less than 10d the display shows Put and bJE SHE alternately Place object s on weighing pan more than 10d the display shows the current weight value Press the left O T button to zero or tare if needed Press the right O T button to start the dynamic weighing The count down is shown on the display eg t 5 t 4 t 1 Once the averaging is finished the result
10. he following ambient conditions Ambient temperature 10 C to 30 C Relative humidity 15 to 80 at 31 C non condensing decreasing linearly to 50 at 40 C Height above sea level Up to 2000 m Operability is assured at ambient temperatures between 5 C and 40 C Power AC adapter Balance power input 8 14 5 VAC 50 60Hz 4VA or 8 20 VDC 4W Protection Protected against dust and water Pollution degree 2 Installation category Class II 6 1 Drawings 0 01g and 0 1g Models 19 6cm 9 2 cm 32cm 0 0001g and 0 001g Models 19 6 cm EN 11 Pioneer 6 2 Specifications PA84 PA163 PA323 3 54 in 9 cm diameter 4 72 in 12 cm diameter Net Weight 10 1 Ib 4 6kg 11 2lb 5 2kg for Incal Model ceder Paton Pama paar pam ok omen Repeatability Tare Range To capacity by subtraction Stabilization Time 3 seconds 7 1 in 18 cm diameter Net Weight 10 Ib 4 5 kg 11 1 Ib 5 1 kg for Incal Model Pioneer EN 12 6 3 Communication 6 3 1 Commands The RS232 Interface allows a computer to control the balance as well as receiving data such as the displayed weight The balance will return ES for invalid commands Command Function SSSss OSO IP Immediate Print of displayed weight a ae Print displayed weight uses Stable ON OFF menu settings Interval Print x Print Interval 1 3600 sec OFF PSN Version Print product name software revision and LFT ON if LFT is set ON Set PC
11. l Side Doors Insert fully into Top Frame then down over retainer 1 2 Install Panels Insert bottom edge in groove then press until locked k me PE nd 3 Install Top Door 4 Install Pan EN 3 Pioneer 2 3 Level Balance Level the balance on a firm steady surface Avoid locations with excessive air current vibrations heat sources or rapid temperature changes Ill a 2 4 Connect Power g mee A For use with CSA Certified or equivalent approved power source which must have a limited circuit output 2 5 Initial Calibration Without InCal Power on the balance by pressing O T Press and hold Print Cal until ZAL is displayed The display flashes the calibration mass needed To select the alternate calibration weight press No Put the calibration mass on the pan The display flashes b15 then TLEAr FFn Remove the mass When calibration is complete d0nE is displayed InCal Press and hold Print Cal until TAL is displayed The display flashes H154 when calibration is complete 42E is displayed Note Calibrations should be performed after a warm up time of 60 minutes 3 OPERATION Count APW Optimization Percentage Dynamic Density or specific units of measure must be activated in the MODE or UNIT menus if they are not initially available 3 1 Weighing Mode Repeatedly press Unit until the desired unit icon is displayed Press 0 T to zero the balance and then p
12. lace objects to be weighed on the pan 3 2 Count Mode Use the Count mode to count parts of uniform weight To access Count Mode press Unit until the display shows Count Establish an Average Piece Weight APW Each time a new type of part is to be counted the nominal weight of one piece APW must be established using a small quantity of pieces With EL APLu displayed press No to use the previously saved APW or press Yes to establish a new APW The display indicates the number of pieces to be used to establish the new APW If a different sample size is preferred press No until the desired sample size is displayed 5 10 20 50 or 100 Put the specified number of pieces on the pan Press Yes to accept new APW or No to abort Count Place the quantity to be counted on the pan Pioneer EN 4 APW Optimization Since the weight of each piece varies slightly APW Optimization may be used to increase the accuracy of the count The balance automatically recalculates the Average Piece Weight when the number of parts on the pan is less than three times the original sample size The display shows FFLU ZP 7 each time the APW is optimized 3 3 Percent Mode Use this mode to measure the weight of a sample as a percentage of a reference weight To access the Percent Mode press Unit until the display shows PErCEnt Establish a new Reference Weight With Clear reference Et r r EF displayed Press No to use the previously saved
13. libration Lac SEt Setup L or ad Mode t ac tin amp Unit Lac Pr t Print Lac die RS232 4 9 Sealing Access to the Balance Settings The Menu Lock switch prevents changes to the Lock Menu The switch can be secured using paper seals wire seals or plastic ties Un Locked Locked with Plastic Tie EN 9 Pioneer 5 0 MAINTENANCE 5 1 Troubleshooting Symptom Possible Cause Remedy Unstable environment Move balance to suitable location Incorrect calibration masses Use correct calibration masses E70 Time EEPROM Checksum error Contact the authorized dealer REF Err Reference Weight is too small Add additional samples LOWREF Reference Weight is too low for accurate Add additional samples or continue to weigh parts counting or percent weighing with less accurate results 5 2 Service Information If the troubleshooting section does not resolve or describe your problem contact your authorized Ohaus service agent Please visit our web site www ohaus com to locate the Ohaus office nearest you 5 3 Accessories Security device 80850000 Auxiliary Display 80251396 Density Determination Kit 80253884 0 1mg and 1mg models only Printer Thermal Contact Ohaus Printer Impact Contact Ohaus Cable Kit Thermal Printer Contact Ohaus Cable Kit Impact Printer Contact Ohaus Software Winwedge 80850080 Pioneer EN 10 6 TECHNICAL DATA Ambient conditions The technical data is valid under t
14. printing Content COA EN The content in the data transmission can be modified Each of the following settings can be set on or off Number Only hu aE will only send the numeric result Balance ID bAL td will add the Balance serial number for traceability purposes Reference EFEr will add reference information relevant to the current mode GLP L F will send additional items to allow proper documentation of laboratory results Pioneer EN 8 Line Format t mE Fa Single line format 5 nt E will put all the data in one line separating each output with a comma Multi line format tut t will put each data output on a new line Multi 4 7 41 F will add 4 line spaces between each output AAAAAAAAAAAAAAAAAAAAA Multi Line Format with 4 line feeds Line Feed 2 Line Feed 3 Line Feed 4 User ID sei ER GLP ON Bal ID 123456789 Balance ID ON PrOD eten GLP ON MGE vernemen GLP ON Date EEn ieee GLP ON 120 01 g Result AAAAAAAAAAAAAAAAAAAAA 4 7 RS232 Menu r 5232 Baud Rd The RS232 baud rate can be set to 600 1200 2400 4800 9600 and 19200 Parity PAr t 4 Parity can be sent to 7 bits even parity 7 EGER 7bits odd parity 7 add 7bits no parity 7 Fa or 8bit no parity B Hel Handshake HAnd5h Handshake can be set to off BFF X on X off an aFF or hardware HAr duur 4 8 Lock Menu t 5E When a Lock Menu item is SET ON the indicated menu cannot be changed Lae CAL Ca
15. zero 50 of capacity and full capacity Generally this calibration is not required unless testing shows that the linearity error exceeds the Linearity tolerance in the Specification table Not available in InCal models Calibration Adjust TAL Adu allows adjustment to the result of the internal calibration by 99 divisions InCal models only 4 3 Setup Menu 5 amp t Automatic Zero Setting A257 Environmental changes can cause the display to drift The Automatic Zero Setting Mechanism AZSM is designed to keep the balance set at zero despite these slight changes OFF SET 5d SET 1d SET 2d SET 5d Filter F LeEr Use the low setting SET LO when environmental disturbances are not present Use the medium setting SET MED for normal environments Use the high setting SET HI when vibrations or air currents are present Auto Tare R t AE The initial item placed on the balance is assumed to be a container so it is zeroed out The next item is then weighed When the pan is cleared the balance resets waiting for a container SET OFF SET ON uF EN 7 Pioneer Grad Lr Ad Select the readability displayed Reducing the readability may be needed for approval SET 1d SET 1 d SET 10d Global Reset Lat r5t Resets all settings to factory default values RESET 4 4 Mode Menu ad Count Mode faunt SET ON or OFF Average Piece Weight APW Optimization RPu GP 7 SET ON SET OFF Percent Mode PErLEnt