1. 0 025 150 FS Full scale 1 Accuracy defined as non linearity hysteresis and repeatability 2 Add 0 001 FS oC from 10 C to 10 C and 30 C to 50 C 14 F to 50 F and 86 F to 122 F 3 The system is protected from over pressure with an internal pressure relief valve Pressure media Most gases compatible with aluminium brass stainless steel nitrile and polyurethane seals PTFE acetal nylon Pressure connection Tool less quick fit connection Supplied with G1 8 female and 1 8 NPT female adaptors Other adaptors available see accessories External pressure modules IDOS intelligent digital output sensor pressure modules can be connected via a USB converter P N 0620 USB IDOS to extend the measuring range of the instrument Please refer to the IDOS UPM datasheet Electrical measurement and source Total uncertainty Additional error Resolution 10 C to 30 C 10 C to 10 C 50 to 86 F amp 30 C to 50 C for one year 14 F to 50 F and 86 F to 122 F Rdg FS FS C Measure mode DC Voltage 200 mV 0 018 0 005 0 001 0 001 2000 mV 0 018 0 005 0 001 0 01 20 V 0 018 0 005 0 001 0 00001 30 V 0 018 0 005 0 001 0 0001 Current 20 mA 0 015 0 006 0 001 0 0001 55 mA 0 018 0 006 0 001 0 0001 Source mode DC Voltage 10V Fixed 25mA max omi 0 0 001 24V Fixed 25mA max 0 1 0 0 0 001 Current Oto 24 mA 0 018 0 006 0 001 0 001 O to 24
2. 3lbs including batteries Power supply 8 x AA alkaline batteries Optional mains adaptor P N I0620 PSU 100 260V 10 50 60Hz AC Output DC V 5A 1 6A Battery life 18 to 26 hours depending on functions Connectivity USB type A USB type mini B Ordering Information Please use the following part numbers when ordering DPI611 05G for 1 to 1 bar 14 5 to 15 psi gauge range DPI611 07G for 1 to 2 bar 14 5 to 30 psi gauge range DPI611 10G for 1 to 7 bar 14 5 to 100 psi gauge range DPI611 11G for 1 to 10 bar 14 5 to 150 psi gauge range DPI611 13G for 1 to 20 bar 14 5 to 300 psi gauge range Please order accessories by part number as separate line items Each DPI 611 is supplied with a set of alkaline batteries hand strap test leads G1 8 female and 1 8 NPT female adaptors calibration certificate quick user guide and an electronic copy of the user manual stored in memory on the instrument Accessories Carrying case P N TBA A tailored fabric carry case with carrying strap allows the DPI 611 to be used without removing it from the case Rechargeable battery pack P N TBA Use in place of AA cells The battery pack can be charged within the instrument or externally P N 10620 PSU mains adaptor is also required for charging Mains adaptor P N 10620 PSU A universal input mains adaptor Input voltage 100 to 240 VAC 50 60 Hz Mains socket adaptors are provided USB cable P N 10
3. soft over moulding provide you with a firm grip for hand held use and prevent the instrument from sliding when on a test bench The proven mechanical design provides a simple to use dependable system without the pitfalls of electromechanical devices namely poor reliability regular servicing dependence on battery condition and long pressure cycle times Pressure accuracy Using advanced silicon technology the digitally corrected Druck pressure sensor achieves 0 0185 FS accuracy compared with 0 025 FS for the DPI 610 615 Taking stability and temperature errors period the DPI 611 is more than twice as accurate as its predecessor This is expressed as a total uncertainty to give you complete confidence in the measurement accuracy between annual calibrations Electrical capability The DPI 611 retains the comprehensive electrical measurement and sourcing capability of the DPI 610 series but has better accuracy and simplified connections Ergonomic design a hand strap you can position on the left into account over the one year calibration P IDOS mA V mV 10 24V Switch Vdc Measure Y Y Y Y Y Y MMeasureMeasur Source Y Y Y Y P2 IDOS is an optional external pressure sensor e The DPI 611 mA measurement accuracy includes one year stability temperature errors and calibration uncertainty making it three times more accurate than the DPI 610 series e For ease of use the electrical connectors have be
4. 620 USB PC Connects the DPI 611 to a PC IDOS to USB converter P N 10620 IDOS USB Allows connection of an IDOS universal pressure module to the DPI 611 P N 10620 USB PC is also required to connect the converter to the DPI 611 USB port USB to RS 232 cable P N I0620 USB RS232 Connects the DPI 611 to an RS 232 interface Dirt moisture trap P N 10620 IDT621 Prevents contamination of the DPI 611 pneumatic dol system and cross contamination from one device under test to another The trap connects directly to el the pressure port and replicates the DPI 611 quick fit connection for compatibility with the standard adaptors adaptor kits and hoses Pneumatic hose A high pressure pneumatic hose rated to 400 bar 5800 psi The hose connects directly to the DPI 611 pressure port and replicates the quick fit connection for compatibility with the standard adaptors CA supplied and the adaptor kits Hf Mi P N 10620 HOSE P1 1m 3 28ft pneumatic hose kit P N 10620 HOSE P2 2m 3 28ft pneumatic hose kit Pressure adaptor set A set of test point adaptors dci a to connect the tool less as T IB quick fit DPI 611 pressure n e port or the extension hoses to the device under test P N 10620 BSP G1 8 male and G1 4 male G1 4 female G3 8 female and G1 2 female P N 10620 NPT 1 8 male and 1 4 male 1 4 female 3 8 female and 1 2 female P N 10620 MET 14 mm female and 20 mm female Comparator
5. GE Measurement amp Control DPI 611 Hand held Pressure Calibrator This fully self contained pressure test and calibration system combines pressure generation signal measurement and loop power to provide all the convenience of the Druck DPI 610 615 and yet it s half the size twice as accurate and easier to use DPI 611 hand held pressure calibrator The DPI 611 is the fourth generation in the DPI 600 family which was first introduced in 1984 The DPI 600 family revolutionized test and calibration by providing all the tools for pressure generation and signal measurement in self contained portable packages The DPI 600 soon became the industry workhorse and today it is simply known as the Druck Building on the technical legacy and more than three decades of experience in pressure measurement and calibration the DPI 611 provides all the convenience and reliability of a true Druck yet offers twice the performance in a product half the size e 50 smaller and 33 lighter than the DPI 610 e Generates 0 to 20 bar 300 psi in less than 30 seconds e Creates 95 vacuum e Pressure measurement Is twice as accurate e Three times better electrical accuracy e Simplified touch screen interface with application DASHBOARD quick TASK selection and FAVOURITES storage e Fast three touch set up for any application e Calculates PASS FAIL errors documents results and interfaces with calibration software Precision en
6. adaptor P N I0620 COMP For greater efficiency two test devices can be connected at the same time The adaptor connects to the pressure port of the DPI 611 and provides two outlet ports Compatible with the standard adaptors supplied and the adaptor kits Related Products For information on the wide range of pressure temperature and electrical test and calibration equipment please visit our web site at www ge mcs com en pressure and level 920 650A 2014 General Electric Company All Rights Reserved Specifications are subject to change without notice GE is a registered trademark of General Electric Company Other company or product names mentioned in this document may be trademarks or registered trademarks of their respective companies which are not affiliated with GE
7. ble settle and run times The start and end pressures are reported along with the pressure change and leak rate a En Pressure Gauge 2549 58 20000 00 mbar Start 2579 16 End 2558 05 Change E Rate per minute 63 32 Max min Captures maximum and minimum values and calculates the mean Relief valve Captures the venting pressure of a relief valve Scaling Scales the measured value into a process value For example mA expressed as a Flow correction is available for scaling differential flow transmitter outputs Resolution Adjustable from four to seven digits this matches the displayed value to that of the test device for easy comparison Help The DPI 611 is provided with a multi lingual quick user guide to get you up and running without delay and for convenience the full manual is stored digitally in the instrument Using a tool in the help application the manual can be transferred to a PC for viewing or printing In the help application you can also view wiring connection diagrams 15 23 CONNECTIONS Specifications Pressure gauge ranges referenced to atmosphere Pressure Accuracy Total uncertainty Over Range 10 to 30 C pressure 50 to 86 F for one year bar psi FS FS FS 1tol 14 5 to 15 0 0185 0 025 150 1to 2 14 5 to 30 0 0185 0 025 150 1to 7 14 5 to 100 0 0185 0 025 150 1 to 10 14 5 to 150 0 0185 0 025 150 1to 20 14 5 to 300 0 0185
8. en rationalised to four 4mm sockets Truly hand held The DPI 611 is truly a hand held pressure calibrator being 50 smaller and 33 lighter than the DPI 610 e Redesign of the pressure assembly has resulted in a more efficient and higher performance system that is both smaller and lighter e State of the art microelectronics has reduced the size of the PCA printed circuit assembly and lowered the power consumption which in turn means smaller batteries Even though the instrument is much smaller the touch display is almost twice as large for clear viewing and more information Quick to Nt pressure connections Making a leak tight pressure connection in the field is inevitably frustrating The DPI 611 comes with a quick to fit adaptor system that has a number of advantages over conventional methods e All adaptors hoses and accessories including the dirt moisture trap are quick and simple to fit No tools or sealing are required and connections are leak free e Damaged adaptors are very simply replaced and there s no repair downtime e Making leak tight connections wastes time and for several joints it takes longer than a calibration The DPI 611 system is proven to significantly reduce set up time ss Simplified touch screen DASHBOARD The DPI 611 uses the same interface design as the unique and award winning DPI 620 Genii Measures magazine innovative product of 2014 e The DASHBOARD allows quick applicati
9. g The DPI 611 can record data from three channels simultaneously by manually touching a record button or automatically at a user set interval Data can be reviewed on screen or the data file can be transferred to a PC for further analysis Connectivity with leading calibration and maintenance software The DPI 611 integrates with leading calibration and maintenance software including 4Sight from GE Typically such applications provide an automated and paperless solution to calibration and realise significant benefits including reduced operating costs regulatory compliance and improved process efficiency 4Sight calibration and maintenance software ya AN 4Sight WG a RS 4Sight is the new state of the art integrated web based Software as a Service SaaS calibration manager e Maintains compliance with industry standards e Provides a full time and date stamped audit trail e Significantly reduces your operating costs e Provides automated paperless solutions Ensures that you are always ready for an audit e Optional web hosting means no IT overhead 4Sight calibration and maintenance software gives you total control for all your calibration and maintenance tasks e Software e Mobile solutions e Workshop solutions e Global service 4Sight calibration management software will help you comply with regulations reduce running costs and improve process efficiency As your calibration manager i
10. gineering Performance is a function of precision engineering The innovative design of the DPI 611 pressure system can only achieve efficient generation and precise control through the use of carefully selected materials with high tolerance machining and perfect finishes The choice of case material and precision moulding ensures that the DPI 611 is both rugged and weatherproof State of the art analogue and digital microelectronics contribute both accuracy and processing power to provide a class leading pressure and electrical capability with a unique simple to use interface TE I E IDT621 Instrument Cee sni lt Jan Druck Ltd LEG OFH UK BE To za aa e gt gt 4 COD IPI ESA H 5 im IM Bw Pressure generation With the DPI 611 you can generate from 95 vacuum to 20 bar 300 psi pneumatic pressure A simple selector lets you convert from vacuum to pressure and with a few stokes ofthe pump you will generate the required pressure Fine adjustment is made with the built in volume adjuster and falling calibration points are achieved with the precision vent valve The redesigned mechanical system provides significantly improved performance allowing you to generate 20 bar 300 psi while holding the instrument in one hand It s also quicker taking just 30 seconds to generate maximum pressure even with a one metre hose connecting the device or right and a
11. mA internal 0 018 0 006 0 001 0 001 loop power FS Full scale Rdg reading Multiple parameter display capability The display can be configured to show a maximum of three simultaneous reading windows as follows pressure electrical measurement or electrical source IDOS external pressure module General specifications Display Size 110 mm 4 3in diagonal 480 x 272 pixels LCD colour display with touch screen Internal memory 8Gb nominal user memory for automated procedures calibration data and data log files Languages English default Chinese Dutch French German Italian Japanese Korean Portuguese Russian Spanish Operating 10 to 50 C 14 to 122 F 0 to 40 C 32 to 104 F when temperature powered from optional mains adaptor I0620 PSU Storage 20 to 70 C 4 to 158 F temperature Ingress protection IP 54 Protected against dust and splashing water from any direction Humidity O to 90 RH none condensing To Def Stan 66 31 8 6 cat III Shock vibration BS EN 61010 1 2010 MIL PRF 28800F CLASS 2 Altitude Up to 2000m EMC BS EN 61326 1 2013 Electrical safety BEN DIO 7 2070 Pressure safety Pressure Equipment Directive Class Sound Engineering Practice SEP Enclosure materials Polycarbonate polyamide polypropylene acrylic cotton Approved CE marked Size L W H 27O x 10x 120mm 06 5 15 4 7 in Weight 1 96Kg 4
12. on selection without menus or special keys just tap the app e The TASK menu provides a library of popular configurations From the calibrator screen three simple gestures completely reconfigure the DPI 611 for the next job DOCUMENTING e From the FAVOURITES menu it s even quicker to access regularly used and customised TASKS e The DPI 611 touch screen only shows function keys when they re O O required making it quicker and simpler to use than complex SETTINGS HELP keypads with special function keys and key combinations Easy as 1 2 3 1 Swipe from right to left 2 Touch to make a selection 3 Touch to select the TASK gt l Current 20 0000 55 0000 mA TASK MENU CALIBRATOR FAVOURITES m CALIBRATOR 5 a Pressure CUSTOM TASK Gauge 20 000 DPI 611 external features Quick to fit pressure adaptor system no Pressure vent tools required or seals valve precision needle valve for fine control 10V DC regulated Electrical connections for supply and 24V loop measuring MA mV V power and detecting switch contacts Hand strap attachment io E cover Full colour touch screen display er USB master and slave sockets On Off Sleep Hand strap attachment Padded leather hand strap attaches to left or right Pressure vacuum selection Volume adjuster for precise pressure setting Anti slip feet front and rear Pump Fully documenting The DPI 611 is a simple to u
13. se everyday tool for maintaining and calibrating pressure instruments It also has the advanced features of the DPI 620 Genii for automating calibration procedures calculating errors and interfacing with PCs and calibration and maintenance systems e Automated calibration procedures e PASS FAIL error analysis e Multi channel data logging e 8Gb nominal user memory e Connectivity with leading calibration and maintenance software including 4Sight from GE Automated calibration procedures Procedures generated by calibration management software can be downloaded to the DPI 611 These procedures are presented as a list of work orders and when selected each one will configure the DPI 611 to calibrate a specific device The procedures run automatically and all you have to do is set the pressure The data is recorded digitally ready to be uploaded to the management software Using the DPI 611 with automated procedures significantly reduces the time taken to calibrate a device from typically 40 minutes to less than 10 minutes including the time to set up Further time is saved when assessing the data and creating calibration reports because these operations are automated within the software PASS FAIL error analysis Error analysis calculates the error of the device being tested and reports a pass or fail The error is displayed live allowing zero and span adjustments to be assessed as they are made Multi channel data loggin
14. ts automated workflow robust data and complete traceability will significantly reduce calibration and maintenance costs Please visit http www ge mcs com 4sight for more information Advanced features Step and ramp mA output Simply configured for simulating transmitter outputs into control loops testing valve positioners and checking safety systems The function has programmable end points manual or automatic sequencing and the following options for quick set up e step The step size is defined as a percentage For example 25 provides five test points of 4 8 12 16 and 20 mA Defined step The step size is defined as a value in mA Span check Toggles between two end points for example 4 and 20 mA for checking zero and FS e Ramp A linear ramp between two end points with programmable travel and dwell times is perfect for dynamically testing switches a e Current Fan m ee Current Ti AN 4 000 gt 15 208 mA 25 step manual advance RAMP automatic cycle Nudge Simply used to make a small incremental change to a mA output using up down keys This is great for determining trip values Switch test Automates the capture of pressure switch actuation and de actuation values and calculates the hysteresis Gouge 2 616 0 000 bar Opened at 2 615 Closed at 2 790 Hysteresis 0 175 Pressure leak test This automated procedure for detecting leaks and determining leak rates has programma