Home
SIARETRON 4000 ICU
Contents
1. Use destination High performance Intensive care ventilator for adults children and newborns with colour TFT at 12 Operation principle e Time cycled at constant volume e Pressure cycled e Microprocessor controlled flow e Spontaneous breath with integrated valve Ventilation modalities VC VAC VC VAC BABY APCV BILEVEL ST APCV TV P SIMV PS V SIMV PS SPONT PSV BILEVEL PSV TV Auto Weaning APRV CPAP SIGH NEB Apnoea BACK UP NIV MANUAL Breathing rate From 5 to 150 bpm Inspiratory Time Expiratory Time maximum minimum e Timin 0 036sec minimum inspiratory time e Ti max 9 6sec maximum inspiratory time e Temin 0 08sec minimum expiratory time e Te max 10 9sec maximum expiratory time SIMV Breathing rate From 0 to Rate 1 Rate 1 SIMV Inspiratory time From 0 2 to 4 0 sec Tidal volume from 2 to 3000 ml from 2 to 100 ml in VC VAC BABY mode 1 E ratio From 1 10 to 4 1 Inspiratory pause From 0 to 60 of the inspiratory time Inspiratory pressure limit From 2 to 80 cmH20 PEEP From OFF to 50 cmH20 O2 concentration Adjustable from 21 to 100 with electronic integrated mixer Trigger e Pressure trigger adjustable from OFF 1 to 20 cmH20 under PEEP level e Flow trigger adjustable from OFF 0 3 to 15 l min Trigger E From 5 to 90 of the inspiratory flow peak Trigger detective method Through sensor
2. Gas feeding O2 Gas feeding Air CAN BUS error Maintenance Battery at 50 Battery at 25 10 Minutes Low lt 2 7 bar Low lt 2 7 bar Electronic boards CAN connection wrong 1000 hours Gas supply EXP INSP Flow sensor Airways pressure sensor Patient circuit Battery Siaretron_4000ICU_960401_te_rev5 Verification of the presence of Air and O2 supply pressure Verification of EXP flow sensor operation Verification of pressure sensor operation through control of PAW reading Verification of patient circuit Checking on battery power 6 8 Oxygen cell Cell condition Acoustic alarm Verification by the user of acoustic signal emission the confirmation of the test is made by silencing of that alarm Electric power supply 100 230 Vac 50 60Hz 12 Vdc with external auxiliary battery Power 120 Watt Internal power supply 12 Vdc 4 5 Ah Internal battery operation 1 5 hours max Optional external battery 8 hours max Re charging time About 8 hours External connections RS232 for CO2 module or PC transfer patient data events and trends optional Capnometry connection Available CO2 Measuring mode Mainstream or Sidestream General specifications See on relative Multigas Analyzer Modules for technical data sheet ISO 5369 EN 1281 1 IEC 601 1 IEC 601 1 2 Directive 93 42 EEC EN 4135 IEC 601 1 4 NF S 90 118 EN 794 1 UNI CEI ISO 14971 UNI EN 475 UNI EN ISO 9703 3 Class and type acc
3. pressure or flow Max inspiratory flow 240 l min Flow by 2 l min Flow Trigger Siaretron_4000ICU_960401_te_rev5 2 8 Leakage compensation In NIV mode SIGH Interval 40 500 bpm step 1 Amplitude OFF 10 100 of set Tidal Volume step 10 APRV Time 1 and Time 2 from 10 to 200 sec Level 1 and Level 2 from 3 to 30 cmH20 CPAP From 3 to 30 cmH20 NEB Drug nebulizer selectable to 6 l min with automatic compensation on forced ventilation modes and dedicated output Supply pressure Oo Aria pressure included between 280 kPa and 600 kPa 2 8 6 bar Max flow requested from ventilator 120 l min Patient circuit Double hose non rebreathing Other controls e INSP Block and EXP Block max 20 seconds e Button 100 O2 x 5 min e MAN control manual ventilation e Loops e Curves selection Other features External alarm Nurse call Expandability Software upgradeable for future modalities Dead space compensation Automatic compensation of mechanical and patient circuit dead space 12 TFT colour display The display allows e Setting of Operative Mode e Visualization of alarm messages and signals e Setting and monitoring of physiological breathing parameters e Visualization of additional graphs and breathing parameters e The function MENU for setting operation parameters e Activation of special functions e Visualization of operative mode clock date and tim
4. SIARETRON 4000 ICU A E Intensive care ventilator pe code 960401 gas driven l e A Adults Paediatric Newborns A y rev 5 01 09 2013 Main characteristics The Siaretron 4000 electronic lung ventilator is equipped with a TFT 12 colour monitor displaying the curves of pressure flow volume the loops of breathing parameters the trends and the ventilatory parameters The ventilator is suitable for ventilation of adult paediatric and neonatal patients It is equipped with a flow and pressure trigger it provides the most advanced volume controlled ventilation modalities VC VAC VC VAC BABY pressure controlled ventilation modalities APCV APCV TV SIMV by Volume and by Pressure Pressure supported modalities PSV PSV TV CPAP BILEVEL S ST SIGH Non Invasive Ventilation NIV Drug Nebulizer and Manual Ventilation MAN Siaretron 4000 is supplied with back up long lasting batteries and its software can be updated for new modes and last generation ventilatory strategies Optional Medical air compressor for trolley mounting Dimensions Ventilator unit 518 x 535 x 269 mm W x H x D Ventilator unit and trolley 555 x 1207 x 554 mm W x H x D Weight Ventilator unit 14 Kg Ventilator unit and trolley 25 Kg Relative Humidity use 30 95 RH Working temperature From 10 to 40 C Automatic compensation of Present atmospheric pressure on measured pressure Siaretron_4000ICU_960401_te_rev5 1 8
5. e functions e Visualization of software version Siaretron_4000ICU_960401_te_rev5 3 8 Display keyboard Lateral keyboard for rapid access of functions Encoder knob for selection set up and confirmation of physiological breathing parameters selection and direct activation of function Set physiological breathing FLOW L min FR bpm FRsimv bpm 1 E Pausa PEEP cmH20 parameters SET PLIM cmH20 PMax cmH20 PS cmH20 SIMV RATE bpm Ti s Ti max s Ti min s Trig E Trig L min Vte ml Vti ml O2 SIGH bpm CPAP cmH20 APRV sec cmH20 BACK UP parameters MENU function settings e e SETUP adjustments Alarms Trends Events Patient data Default parameters SETUP function settings e Language Graphic Volume Energy saving Brightness Apnoea time Gas sensor N20 unit of measurement Password TCP setting Technical contact Test on demand Gas sensor Colour selection Siaretron_4000ICU_960401_te_rev5 4 8 Measured parameters e PAW peak mean plateau PEEP range 20 80 cmH20 e _Tinsp Texp Tpause range 0 036 10 9sec e I E ratio range 1 99 99 1 e Static and dynamic compliance range 10 150 ml cmH20 e Resistance range 0 400 cmH20 I s e of FiO range 0 99 e Rate range 0 150 bpm e Tidal Volume Vte Vti range 0 3000 ml e Minute Volume range 0 40 l min e Inspirat
6. ording to IEC 601 1 Class 1 Type B Class according to 93 42 EEC Directive Class IIb Siaretron_4000ICU_960401_te_rev5 7 8 Supplied Accessories e User s Manual e O92 supply hose e AIR supply hose e Nebulizer set e Silicone patient circuit e Antibacterial filter e Power cable SHUKO VDE Oo cell Optional Accessories See on Export Price List SIARE applies the UNI EN ISO 13485 2004 Quality System and the 93 42 EEC SIARE ENGINEERING INTERNATIONAL GROUP s r l Via Pastore 18 40056 Crespellano Bologna ITALY Tel 39 051 969802 Fax 39 051 969366 Email mail siare it http www siare it Siaretron_4000ICU_960401_te_rev5 8 8
7. ory Peak Flow range 1 200 l min e Expiratory Peak Flow range 1 150 l min e EtCOo with optional CO2 module range 0 10 Additional measured MAP cmH20 Pplateau cmH20 Fi L min Fe L min Ti sec Te sec parameters Tpause sec Ri c mH2O L sec Cs ml cmH20 Displayed graphics e CURVES Pressure Flow Volume CO2 optional e LOOPS Pressure Volume Flow Volume Pressure Flow e Auto range Trends Storage capacity 72 h of all measured parameters Events Memory storage up to 100 machine events including the alarms Flow sensor e Magnetic perturbance patented multi usage e Automatic flow sensor calibration started by the operator e Steam or chemical disinfection Oxymeter Electronic with automatic calibration CO2 module optional CO2 module main stream or side stream Siaretron_4000ICU_960401_te_rev5 5 8 Alarm types With limits set by the operator By default the operator cannot set them up Alarm priority High Mean Standby Airways pressure Breathing rate Expired Minute Volume Expired Tidal Volume FiO2 concentration PEEP Electric power supply Apnoea EtCO2 High Low High Low High Low High Low High Low High Low Alarm occurs in case of failure of external power supply Low Rate function of Apnoea BACK UP High Low with optional CO2 module Level charge Level charge Battery Level low
Download Pdf Manuals
Related Search
Related Contents
Philips GoGear MP3 player SA3MXX04R Integral IN2T4GFWNEX4K2 memory module Denver TR-61 TECHCONNECT TC2-MATRIXBRIDGE OWNERS MANUAL Instalación del Navegador VW MFD en un Passat 3BG con hueco Ednet 31809 mobile device charger CILINDROS HIDRÁULICOS LG VS450PP Data Sheet 日常のお手入れのお願い Hand Care Systems HCS FRE 0412 5M Copyright © All rights reserved.