Keysight InfiniiVision 3000 X-Series Oscilloscopes User`s Guide
Contents
1. of WS LEFT CHANNEL i RIGHT CHANNEL ws Lo Ri 2 sex TV X LN g SDATA Ii o 1 2 AE ol1l 2 n2 MSB LSB MSB LSB 12 Press the WS Low softkey then turn the Entry knob to select whether WS Low indicates Left or Right channel data The on screen diagram changes with your selection WS Low Left Channel Left channel data corresponds to WS low right channel data corresponds to WS high WS Low Left is the oscilloscope s default WS setting Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 389 26 12S Triggering and Serial Decode LEFT CHANNEL RIGHT CHANNEL ws gi U UL sork il Ul tl ad gt ONE SCLK CYCLE SDATA o l2 n2 n1 ol1 2 p2 MSB LSB MSB LSB WS Low Right Channel Right channel data corresponds to WS low left channel data corresponds to WS high LEFT CHANNEL RIGHT CHANNEL sou qu Win Ir ONE SCLK CYCLE SDATA of 1 2 l n2 n1 7 o 1 2 n 2 n 1 MSB LSB MSB LSB
2. 13 Press the SCLK Slope softkey then turn the Entry knob to select the SCLK edge on which data is clocked in your device under test either rising or falling The on screen diagram changes with your selection I2S Triggering To set up the oscilloscope to capture 2S signals see Setup for I2S Signals on page 387 After you have set up the oscilloscope to capture S signals you can then trigger on a data value 1 Press Trigger 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select the serial slot Serial 1 or Serial 2 on which the 12S signals are being decoded Trigger Menu Trigger Trigger Setup S PS 390 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide I2S Triggering and Serial Decode 26 3 Press the Trigger Setup softkey to open the 2S Trigger Setup Menu IS Trigger Setup Menu Audio Trigger Base Left Decimal 4 Press the Audio softkey then turn the Entry knob to choose to trigger on Left channel events Right channel events or events that occur on Either channel 5 Press the Trigger softkey and choose a qualifier Equal triggers on the specified audio channel s data word when it equals the specified word Not equal triggers on any word other than the specified word Less than triggers when the channel s data word is less than the specified value Greater than triggers when the channel s data word is greater tha
3. Width on page 225 Yes Width on page 225 Yes X at Max Y on page 229 Yes The resultant units are in Hertz X at Min Y on page 228 Yes The resultant units are in Hertz Use the cursors to make other measurements on FFT Note that additional Power App measurements are available when the DSOX3PWR power measurement and analysis license is installed and the Power Application is enabled For more information see the DSOX3PWR Power Measurement Application User s Guide at www keysight com find 3000X Series manual or on the Documentation CD Note that additional measurements are available with the N2820A high sensitivity current probe when both the Primary and Secondary probe cables are used Zoom In waveform data below the probe s clamp level is joined with Zoom Out waveform data above the probe s clamp level to create the waveform on which the measurement is made These measurements are only valid for analog input channels Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 213 14 Measurements Dual Channel N2820A Probe Notes Measurement Amplitude See Amplitude on page 216 Charge Charge in Amp hours is the measured area under the waveform See Area on page 230 Average See Average on page 219 Base See Base on page 217 Peak Peak See Peak Peak on page 216 DC RMS See DC RMS on page 220 AC RMS See AC RMS on page 220
4. k 10 ns Within a range of time values gt lt k 10 ns gt Trigger __ Tigger For example for a positive pulse if you set t gt 10 ns and t lt 15 ns k 10ns 4 H 15 ns gt k 12ns __ Trigger m L l 7 Select the qualifier time set softkey lt or gt then rotate the Entry knob to set the pulse width qualifier time The qualifiers can be set as follows 2 ns to 10 s for gt or lt qualifier 5 ns to 10 s for 350 MHz bandwidth models 10 nsto 10 s for gt lt qualifier with minimum difference of 5 ns between upper and lower settings Pulse width When the less than lt qualifier is selected the Entry knob sets the oscilloscope trigger lt qualifier to trigger on a pulse width less than the time value displayed on the softkey time set softkey time range value Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide When the time range gt lt qualifier is selected the Entry knob sets the upper 147 10 Triggers Pulse width trigger gt qualifier time set softkey When the greater than gt qualifier is selected the Entry knob sets the oscilloscope to trigger on a pulse width greater than the time value displayed on the softkey When the time range gt lt qualifier is selected the Entry knob sets the lower time range value Pattern Trigger 148 The Pattern trigger identifies a trigger condition by looki
5. 3867 KEYSIGHT TECHNO IES Acquisition Channels Acquire Menu Acq Mode Segmented Peak Detect i Figure 37 Sine With Glitch Peak Detect Mode Using Peak Detect Mode to Find a Glitch 190 Connect a signal to the oscilloscope and obtain a stable display To find the glitch press the Acquire key then press the Acq Mode softkey until Peak Detect is selected Press the Display key then press the oo Persistence infinite persistence softkey Infinite persistence updates the display with new acquisitions but does not erase previous acquisitions New sample points are shown at normal intensity while previous acquisitions are displayed at reduced intensity Waveform persistence is not kept beyond the display area boundary Press the Clear Display softkey to erase previously acquired points The display will accumulate points until co Persistence is turned off Characterize the glitch with Zoom mode a Press the zoom key or press the Horiz key and then the Zoom softkey b To obtain a better resolution of the glitch expand the time base Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Acquisition Control 12 Use the horizontal position knob 4 to pan through the waveform to set the expanded portion of the normal window around the glitch Averaging Acquisition Mode The Averaging mode lets you average multiple acquisitions together to reduce noise and increase vertical resolution
6. Header Size A 32 bit integer that is the number of bytes in the header Waveform Type A 32 bit integer that is the type of waveform stored in the file 0 Unknown 1 Normal 2 Peak Detect 3 Average 4 Not used in InfiniiVision oscilloscopes 5 Not used in InfiniiVision oscilloscopes 6 Logic Number of A 32 bit integer that is the number of waveform buffers required to read the Waveform data Buffers Points A 32 bit integer that is the number of waveform points in the data Count A 32 bit integer that is the number of hits at each time bucket in the waveform record when the waveform was created using an acquisition mode like averaging For example when averaging a count of four would mean every waveform data point in the waveform record has been averaged at least four times The default value is 0 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 335 22 Reference 336 X Display Range A 32 bit float that is the X axis duration of the waveform that is displayed For time domain waveforms it is the duration of time across the display If the value is zero then no data has been acquired X Display Origin A 64 bit double that is the X axis value at the left edge of the display For time domain waveforms it is the time at the start of the display This value is treated as a double precision 64 bit floating point number If the value is zero then no data has been acquired
7. High Pass and Low Pass Filter on page 92 High Pass and Low Pass Filter The high pass or low pass filter functions available with the DSOX38ADVMATH advanced math measurements license apply the filter to the selected source waveform and display the resut in the math waveform The high pass filter is a single pole high pass filter The low pass filter is a 4th order Bessel Thompson filter 92 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Math Waveforms 4 Use the Band width softkey to select the filter s 3 dB cutoff frequency The ratio of the input signal s Nyquist frequency and the selected 3 dB cutoff frequency affects how many points are available in the output and under some circumstances there are no points in the output waveform 1 25 KEYSIGHT TECHNO IES Acquisition 4 Channels s Measurements Freq 1 PKPRI Freq M 500kHz m Pk Pk M tit LPF Ch1 Scale 1 00V Function Operator Source 1 f tit Low Pass Filter 5 0MHz Figure 15 Example of Low Pass Filter Math Visualizations With the DSOX3ADVMATH advanced math measurements license you can apply visualization math functions that give you different ways of viewing captured data and measurement values Magnify on page 94 Measurement Trend on page 94 Chart Logic Bus Timing on page 96 Chart Logic Bus State on page 97 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 93 4 Math
8. Overload Frame The oscilloscope will trigger on CAN overload frames If you select a condition that lets you trigger on ID or data values use the Bits softkey and the CAN Bits Menu to specify those values For details about using the CAN Bits Menu softkeys press and hold the softkey in question to display the built in help You can use the Zoom mode for easier navigation of the decoded data If the setup does not produce a stable trigger the CAN signal may be slow enough that the oscilloscope is AutoTriggering Press the Mode Coupling key then press the Mode softkey to set the trigger mode from Auto to Normal To display CAN serial decode see CAN Serial Decode on page 347 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide CAN LIN Triggering and Serial Decode 23 CAN Serial Decode See Also To set up the oscilloscope to capture CAN signals see Setup for CAN Signals on page 343 For CAN triggering set up see CAN Triggering on page 345 To set up CAN serial decode 1 Press Serial to display the Serial Decode Menu OVLD 000000000 0 0 ERA 000001572 1 9 UTIL 24 0 Mode Signals Reset CAN a Counters 2 Ifthe decode line does not appear on the display press the Serial key to turn it on 3 If the oscilloscope is stopped press the Run Stop key to acquire and decode data If the setup does not produce a stable trigger the CAN signal may be slow enough that the oscillos
9. Utility key Press this key to access the Utility Menu which lets you configure the oscilloscope s 1 0 settings use the file explorer set preferences access the service menu and choose other options See Chapter 20 Utility Settings starting on page 287 Quick Action key Press this key to perform the selected quick action measure all snapshot print save recall freeze display and more See Configuring the Quick Action Key on page 302 Analyze key Press this key to access analysis features like trigger level setting measurement threshold setting Video trigger automatic set up and display mask testing see Chapter 15 Mask Testing starting on page 237 or the DSOX3PWR power measurement and analysis application Wave Gen key Press this key to access waveform generator functions See Chapter 17 Waveform Generator starting on page 253 6 Trigger controls These controls determine how the oscilloscope triggers to capture data See Chapter 10 Triggers starting on page 139 and Chapter 11 Trigger Mode Coupling starting on page 173 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 35 1 36 Getting Started Horizontal controls The Horizontal controls consist of Horizontal scale knob Turn the knob in the Horizontal section that is marked N to adjust the time div sweep speed setting The symbols under t
10. 1 000 10 00 Channels Measurements Top 1 Measurement Menu Source y d Settings Clear Meas A Figure 46 Isolating area for Top measurement Base The Base of a waveform is the mode most common value of the lower part of the waveform or if the mode is not well defined the base is the same as Minimum The Y cursor shows the value being measured Overshoot Overshoot is distortion that follows a major edge transition expressed as a percentage of Amplitude The X cursors show which edge is being measured edge closest to the trigger reference point Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 217 14 Measurements 218 local Maximum D Top 4 Amplitude i Rising edge overshoot Base D local Minimum x 100 Falling edge overshoot Amplitude Overshoot local Minimum Overshoot Channels x Measurements Measurement Menu Source Add Settings Clear Meas Measurement a ad gt Figure 47 Automatic Overshoot measurement Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Measurements 14 Preshoot Preshoot is distortion that precedes a major edge transition expressed as a percentage of Amplitude The X cursors show which edge is being measured edge closest to the trigger reference point local Maximum D Top Rising edge preshoot Amplitude x 100 _ Base D local Minimum Falling edge preshoot Amplitude x 100 s Preshoot local Maximu
11. 2 Softkeys The functions of these keys change based upon the menus shown on the display directly above the keys The S Back Up key moves up in the softkey menu hierarchy At the top of the hierarchy the 8 Back Up key turns the menus off and oscilloscope information is shown instead 3 Intensity key Press the key to illuminate it When illuminated turn the Entry knob to adjust waveform intensity You can vary the intensity control to bring out signal detail much like an analog oscilloscope Digital channel waveform intensity is not adjustable More details about using the Intensity control to view signal detail are on To adjust waveform intensity on page 127 34 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Getting Started 1 4 Entry knob The Entry knob is used to select items from menus and to change values The function of the Entry knob changes based upon the current menu and softkey selections Note that the curved arrow symbol a above the entry knob illuminates whenever the entry knob can be used to select a value Also note that when the Entry knob a symbol appears on a softkey you can use the Entry knob to select values Often rotating the Entry knob is enough to make a selection Sometimes you can push the Entry knob to enable or disable a selection Pushing the Entry knob also makes popup menus disappear 5 Tools keys The Tools keys consist of
12. 232 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Measurements 14 6 Press the Upper softkey then turn the Entry knob to set the upper measurement threshold value Decreasing the upper value below the set middle value will automatically decrease the middle value to be less than the upper value The default upper threshold is 90 or 1 50 V If threshold Type is set to the upper threshold value can be set from 7 to 95 Measurement Window with Zoom Display When the zoomed time base is displayed you can choose whether measurements are made in the Main window portion of the display or the Zoom window portion of the display 1 Press the Meas key 2 In the Measurement Menu press the Settings softkey 3 Inthe Measurement Settings Menu press the Meas Window softkey then turn the Entry knob to select from Auto Select The measurement is attempted in the lower Zoom window if it cannot be made there the upper Main window is used Main The measurement window is the upper Main window Zoom The measurement window is the lower Zoom window Measurement Statistics Press the Meas key to enter the Measurement Menu By default statistics are displayed and frequency and voltage are measured on channel 1 Select the measurements you desire for the channels you are using see Measurements Summary on page 211 From the Measurement Menu press the Statistics softkey to enter the Statist
13. 268 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data Saving Setups Screen Images or Data 269 Recalling Setups Masks or Data 277 Recalling Default Setups 279 Performing a Secure Erase 280 Oscilloscope setups reference waveforms and mask files can be saved to internal oscilloscope memory or to a USB storage device and recalled later You can also recall default or factory default setups Oscilloscope screen images can be saved to a USB storage device in BMP or PNG formats Acquired waveform data can be saved to a USB storage device in comma separated value CSV ASCII XY and binary BIN formats There is also a command to securely erase all the oscilloscope s non volatile internal memory Saving Setups Screen Images or Data 1 Press the Save Recall key 2 Inthe Save Recall Menu press Save 3 In the Save Trace and Setup Menu press Format then turn the Entry knob to select the type of file you want to save KEYSIGHT TECHNOLOGIES 269 18 Save Recall Setups Screens Data 2 0 Setup scp The oscilloscope s horizontal timebase vertical sensitivity trigger mode trigger level measurements cursors and math function settings that tell the oscilloscope how to make a particular measurement See To save setup files on page 271 8 bit Bitmap image bmp The c
14. If infinite persistence display mode see To set or clear persistence on page 129 is on when mask test is enabled it stays on If infinite persistence is off when mask test is enabled it is switched on when mask test is switched on then infinite persistence is switched off when mask test is switched off Troubleshooting If you press Create Mask and the mask appears to cover the entire screen check the Mask Setup Y and X settings in the Automask Menu If these are set to zero the resulting mask will be extremely tight around the waveform If you press Create Mask and it appears that no mask was created check the Y and X settings They may be set so large that the mask is not visible Mask Test Setup Options From the Mask Test Menu press the Setup softkey to enter the Mask Setup Menu Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 239 15 Mask Testing 240 Run Until The Run Until softkey lets you specify a condition on which to terminate testing Forever The oscilloscope runs continuously However if an error occurs the action specified using the On Error softkey will occur Minimum of Tests Choose this option and then use the of Tests softkey to select the number of times the oscilloscope will trigger display the waveform s and compare them to the mask The oscilloscope will stop after the specified number of tests have been completed The specified minimum number
15. X Increment A 64 bit double that is the duration between data points on the X axis For time domain waveforms this is the time between points If the value is zero then no data has been acquired X Origin A 64 bit double that is the X axis value of the first data point in the data record For time domain waveforms it is the time of the first point This value is treated as a double precision 64 bit floating point number If the value is zero then no data has been acquired X Units A 32 bit integer that identifies the unit of measure for X values in the acquired data 0 Unknown 1 Volts 2 Seconds 3 Constant 4 Amps 5 dB 6 Hz Y Units A 32 bit integer that identifies the unit of measure for Y values in the acquired data The possible values are listed above under X Units Date A 16 byte character array left blank in InfiniiVision oscilloscopes Time A 16 byte character array left blank in the InfiniiVision oscilloscopes Frame A 24 byte character array that is the model number and serial number of the oscilloscope in the format of MODEL SERIAL Waveform Label A 16 byte character array that contains the label assigned to the waveform Time Tags Segment Index A 64 bit double only used when saving multiple segments requires segmented memory option This is the time in seconds since the first trigger A 32 bit unsigned integer This
16. Channels Annotation Menu Annotation Text Color Background Inverted See Also To save BMP or PNG image files on page 271 To print the oscilloscope s display on page 281 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 305 20 Utility Settings 306 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface Accessing the Web Interface 308 Browser Web Control 309 Save Recall 315 Get Image 317 Identification Function 318 Instrument Utilities 318 Setting a Password 320 When the Keysight InfiniiVision X Series oscilloscopes have the DSOXLAN LAN VGA option module installed you can access the oscilloscope s built in web server using a Java enabled web browser The oscilloscope s web interface lets you View information about the oscilloscope like its model number serial number host name IP address and VISA address connect string Control the oscilloscope using the Remote Front Panel Send SCPI Standard Commands for Programmable Instrumentation remote programming commands via the SCPI Commands applet window Save setups screen images waveform data and mask files Recall setup files reference waveform data files or mask files Get screen images and save or print them from the browser Activate the Identification function to identify a particular instrument by causing a message to b
17. DE Set all Digits E Digital channel values at cursor See Pattern Trigger on page 148 for more information on Pattern triggering Digital channel signal fidelity Probe impedance and grounding When using the mixed signal oscilloscope you may encounter problems that are related to probing These problems fall into two categories probe loading and probe grounding Probe loading problems generally affect the device under test while probe grounding problems affect the accuracy of the data to the measurement instrument The design of the probes minimizes the first problem while the second is easily addressed by good probing practices Input Impedance The logic probes are passive probes which offer high input impedance and high bandwidths They usually provide some attenuation of the signal to the oscilloscope typically 20 dB 114 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Digital Channels 6 Passive probe input impedance is generally specified in terms of a parallel capacitance and resistance The resistance is the sum of the tip resistor value and the input resistance of the test instrument see the following figure The capacitance is the series combination of the tip compensating capacitor and the cable plus instrument capacitance in parallel with the stray tip capacitance to ground While this results in an input impedance specification that is an accurate model for DC and low frequencies the hi
18. Defaulting labels without erasing the default library Pressing Default Setup sets all channel labels back to the default labels but does not erase the list of user defined labels in the library 138 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers Adjusting the Trigger Level 140 Forcing a Trigger 141 Edge Trigger 142 Edge then Edge Trigger 144 Pulse Width Trigger 145 Pattern Trigger 148 OR Trigger 151 Rise Fall Time Trigger 152 Nth Edge Burst Trigger 153 Runt Trigger 155 Setup and Hold Trigger 157 Video Trigger 158 USB Trigger 170 Serial Trigger 172 A trigger setup tells the oscilloscope when to acquire and display data For example you can set up to trigger on the rising edge of the analog channel 1 input signal You can adjust the vertical level used for analog channel edge detection by turning the Trigger Level knob In addition to the edge trigger type you can also set up triggers on rise fall times Nth edge bursts patterns pulse widths runt pulses setup and hold violations TV signals USB signals and serial signals if option licenses are installed You can use any input channel or the External Trigger Input on page 178 BNC as the source for most trigger types KEYSIGHT TECHNOLOGIES 139 10 Triggers Changes to the trigger setup are applied immediately If the oscillosc
19. Figure11 Example of Ax B Use the Gain A softkey to specify the gain Use the Offset B softkey to specify the offset The Ax B function differs from the Magnify math visualization function in that the output is likey different than the input See Also Magnify on page 94 Square The square function available with the DSOX3ADVMATH advanced math measurements license calculates the square of the selected source point by point and displays the result Press the Source softkey to select the signal source See Also Square Root on page 87 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 89 4 Math Waveforms Absolute Value The absolute value function available with the DSOX3ADVMATH advanced math measurements license changes negative values in the input to positive values and displays the resulting waveform KEYSIGHT TECHNO IES Acquisition Channels x Measurements Pk Pk 3 519mV Pk Pk M 257mV fset 100 16m Figure 12 Example of Absolute Value See Also Square on page 89 Common Logarithm The Common Logarithm log function available with the DSOX3ADVMATH advanced math measurements license performs a transform of the input source Where the transform is undefined for a particular input holes zero values appear in the function output See Also Natural Logarithm on page 91 90 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide
20. In the decoded data Right channel data values appear in green along with the R characters Left channel data values appear in white along with the L characters Decoded text is truncated at the end of the associated frame when there is insufficient space within frame boundaries Pink vertical bars indicate you need to expand the horizontal scale and run again to see decode Red dots in the decode line indicate that more data can be displayed Scroll or expand the horizontal scale to view the data Aliased bus values undersampled or indeterminate are drawn in pink Unknown bus values undefined or error conditions are drawn in red When the receiver word size is greater than the transmit word size the decoder fills the least significant bits with zeros and the decoded value does not match the trigger value 394 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide I2S Triggering and Serial Decode 26 Interpreting 12S Lister Data 1 00 2 00 37 475 10 0057 Stop 25 1 12 Time Left Channel Right Channel Errors x KEYSIGHT TECHNOLOGIES Acquisition Channels 20 53us Zoom to Options Selection Zoom w In addition to the standard Time column the 12S Lister contains these columns Left Channel displays the left channel data Right Channel displays the right channel data Errors highlighted in red and marked with an X Aliased data is highlighted in
21. Math Waveforms 4 Natural Logarithm The Natural Logarithm In function available with the DSOX8ADVMATH advanced math measurements license performs a transform of the input source Where the transform is undefined for a particular input holes zero values appear in the function output Acquisition Channels x Measurements Pk Pk 3 519mV Pk Pk M 4 16 tit In Ch3 Scale 1 00 Function Operator tit In Figure 13 Example of Natural Logarithm See Also Common Logarithm on page 90 Exponential The Exponential e x function available with the DSOX3ADVMATH advanced math measurements license performs a transform of the input source See Also Base 10 Exponential on page 91 Base 10 Exponential The Base 10 Exponential 10 x function available with the DSOX3ADVMATH advanced math measurements license performs a transform of the input source Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 91 4 Math Waveforms 200 05 o O 0V KEYSIGHT TECHNO IES Acquisition 1 Channels x Measurements Pk Pk 3 519mV Pk Pk M 1 244 Operator Source 1 Figure 14 Example of Base 10 Exponential See Also Exponential on page 91 Math Filters With the DSOX3ADVMATH advanced math measurements license you can use math filters to create a waveform that is the result of a high or low pass filter on an analog input channel or on the result of an arithmetic operation
22. S safety warning 29 sample rate 3 sample rate and memory depth 187 sample rate current rate displayed 48 sample rate oscilloscope 184 185 sampling theory 183 sampling overview 182 save 303 save file 291 save files via web interface 315 save segment 273 save setup files 271 save times data 274 Save to 276 save Quick Save 303 save recall from web interface 315 Save Recall key 38 saver screen 294 saving data 269 SCL 12C trigger 370 SCLK 12S trigger 388 SCPI Commands window 313 screen image via web interface 317 screen saver 294 SDA 369 SDA I2C trigger 370 Search key 36 secure erase 280 segmented memory 193 segmented memory re arm time 196 segmented memory saving segments 273 segmented memory statistical data 196 select digital channels 111 select knob 111 Selected 292 selecting values 35 self test front panel 300 self test hardware 300 serial clock I2C trigger 370 serial clock I2S trigger 388 serial data 369 serial data 12C trigger 370 serial decode controls 37 Serial key 37 serial number 300 307 service functions 297 Set all Digits softkey 150 setup and hold triggering 157 setup files saving 271 setup automatic 107 setup default 29 setups recall 278 SGM 193 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide SGM license 332 shipping damage 23 shipping precautions 302
23. The XY time mode converts the oscilloscope from a volts versus time display to a volts versus volts display using two input channels Channel 1 is the X axis input channel 2 is the Y axis input You can use various transducers so the display could show strain versus displacement flow versus pressure volts versus Current or voltage versus frequency Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 51 2 52 Horizontal Controls Example This exercise shows a common use of the XY display mode by measuring the phase difference between two signals of the same frequency with the Lissajous method 1 Connect a sine wave signal to channel 1 and a sine wave signal of the same frequency but out of phase to channel 2 2 Press the AutoScale key press the Horiz key then press Time Mode and select XY 3 Center the signal on the display with the channel 1 and 2 position knobs Use the channel 1 and 2 volts div knobs and the channel 1 and 2 Fine softkeys to expand the signal for convenient viewing The phase difference angle 8 can be calculated using the following formula assuming the amplitude is the same on both channels sind Sor Signal must 4a be centered in Y A Measuring Signals 90 degrees Signals phase difference out of phase in phase Figure 3 XY time mode signals centered on display 4 Press the Cursors key 5 S
24. 2 bits Label Data Label 8 bits Data 23 bits 4 Ifthe decode line does not appear on the display press the Serial key to turn it on 5 Ifthe oscilloscope is stopped press the Run Stop key to acquire and decode data If the setup does not produce a stable trigger the ARINC 429 signal may be slow enough that the oscilloscope is AutoTriggering Press the Mode Coupling key then press the Mode softkey to set the trigger mode from Auto to Normal You can use the horizontal Zoom window for easier navigation of the decoded data See Also Interpreting ARINC 429 Decode on page 409 ARINC 429 Totalizer on page 410 Interpreting ARINC 429 Lister Data on page 411 Searching for ARINC 429 Data in the Lister on page 411 408 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide MIL STD 1553 ARINC 429 Triggering and Serial Decode 27 Interpreting ARINC 429 Decode 100 057 Channels 000A6 67 ERRORS 0000400034 Signals Settings Word Format Lister gt L SDI _ Depending on the selected word decode format the ARINC 429 decode display is color coded as follows When the decode format is Label SDI Data SSM Label yellow 8 bits displayed in octal SDI blue 2 bits displayed in binary Data white red if parity error 19 bits displayed in the selected Base SSM green 2 bits displayed in binary When the decode format is Label Data SSM Label yellow 8 bits displa
25. 6 EXT TRIG IN connector External trigger input BNC connector See External Trigger Input on page 178 for an explanation of this feature 7 USB Host port 8 USB Device port This port functions identically to the USB host port on the front panel USB Host Port is used for saving data from the oscilloscope and loading software updates See also USB Host port see page 39 This port is for connecting the oscilloscope to a host PC You can issue remote commands from a host PC to the oscilloscope via the USB device port See Remote Programming with Keysight IO Libraries on page 314 Learn the Oscilloscope Display The oscilloscope display contains acquired waveforms setup information measurement results and the softkey definitions Analog channel Trigger point sensitivity _ time reference Status line Trigger level Analog channels and ground levels aA Digital channels S A Measurement statistics Menu line Softkeys Delay Time Run Stop Trigger Trigger time div status type _ source N fo a Channels Trigger level or digital threshold _ Information area Slice Cursors defining trea Tee measurement bed Pk Pki1 500mv N j Ampl 2 Measurements 3 5V Settings Clear Meas Statistics gt al Figure 1 Interpreting the oscilloscope display Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 43 1 Getti
26. Accessing the Web Interface on page 308 2 When the oscilloscope s web interface is displayed select the Save Recall tab from the left side of the Welcome screen 3 Click the Recall link 316 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Web Interface 21 4 On the Recall page a Click Browse b Inthe Choose file dialog select the file you want to recall then click Open c When recalling reference waveform data files select the To Reference Waveform location Support KEYSIGHT i Another wq pea x Mala C Temp web_interface scope h5 Cee aoe R1 v R1 d Click Recall Get Image To save or print the oscilloscope s display from the web interface 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 308 2 When the oscilloscope s web interface is displayed select the Get Image tab from the left side of the Welcome screen After a delay of several seconds the oscilloscope s screen image will be displayed 3 Right click on the image and select Save Picture As or Print Picture 4 Select a storage location for the image file and click Save Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 317 21 Web Interface Identification Function The Identification web interface feature is useful when trying to locate a specific instrument in a rack of equipment 1 Access the oscilloscope s web interface see Accessing
27. Channel 2 ground ___ level Channel 1 Menu Coupling Imped BW Limit DC 1M Ohm The ground level of the signal for each displayed analog channel is identified by the position of the icon at the far left side of the display To turn waveforms on or off channel or math 1 Press an analog channel key turn the channel on or off and to display the channel s menu When a channel is on its key is illuminated Turning channels off You must be viewing the menu for a channel before you can turn it off For example if channel 1 and channel 2 are turned on and the menu for channel 2 is being displayed to turn channel 1 off press 1 to display the channel 1 menu then press 1 again to turn channel 1 off 62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Vertical Controls 3 To adjust the vertical scale 1 Turn the large knob above the channel key marked A N to set the vertical scale volts division for the channel The vertical scale knob changes the analog channel scale in a 1 2 5 step sequence with a 1 1 probe attached unless fine adjustment is enabled see To change the vertical scale knob s coarse fine adjustment setting on page 65 The analog channel Volts Div value is displayed in the status line The default mode for expanding the signal when you turn the volts division knob is vertical expansion about the ground level of the channel however you can change this to expand about the c
28. Channels Fast Debug Channels Acq Mode All Normal Press and hold front panel key or softkey or right click softkey when using web browser remote front panel Quick Help remains on the screen until another key is pressed or a knob is turned To select the user To select the user interface and Quick Help language interface and 4 press Help then press the Language softkey Quick Help language 2 Repeatedly press and release the Language softkey or rotate the Entry knob until the desired language is selected The following languages are available English French German Italian Japanese Korean Portuguese Russian Simplified Chinese Spanish and Traditional Chinese Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 45 1 46 Getting Started Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls To adjust the horizontal time div scale 48 To adjust the horizontal delay position 49 Panning and Zooming Single or Stopped Acquisitions 50 To change the horizontal time mode Normal XY or Roll 50 To display the zoomed time base 54 To change the horizontal scale knob s coarse fine adjustment setting 55 To position the time reference left center right 56 Searching for Events 56 Navigating the Time Base 58 The horizontal controls include The horizontal scale and position
29. Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 95 4 Math Waveforms Chart Logic Bus Timing The Chart Logic Bus Timing function available with the DSOX3ADVMATH advanced math measurements license displays bus data values as an analog waveform like a D A conversion When the bus value is transitioning the function output is the bus s last stable state 100 02 Acquisition Channels Cursors tit Timing Chart Bus 1 Scale 100m Htset 200 32m Function Operator Source 1 Units Code O Offset Units tit Bus Timing Bus1 1 0mY 128mY Volts Figure 18 Example of Chart Logic Bus Timing Use the Units Code softkey to specify the analog value equivalent of each increment in the bus data value Use the 0 Offset softkey to specify the analog value equivalent of a bus data value of zero Use the Units softkey to specify the type of values the bus data represents volts amps etc See Also Chart Logic Bus State on page 97 96 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Math Waveforms 4 Chart Logic Bus State The Chart Logic Bus State function available with the DSOX3ADVMATH advanced math measurements license displays bus data values sampled on a clock signal s edge as an analog waveform like a D A conversion 12 59 KEYSIGHT IES TECHNO Acquisition 1 Channels Cursors AX 7 950000us YINI 64 00mv 2 M AY 126 95mi Source 1 More Chart Bus
30. qualifier pulse width 147 Quick Action key 35 302 Quick Clear Display 303 Quick Freeze Display 303 Quick Help 44 Quick Help language 45 Quick Measure All 302 Quick Print 302 Quick Recall 303 Quick Save 303 Quick Trigger Mode 303 R ramp waveform generator output 255 random noise 173 ratio measurement 222 ratio measurement trend 95 ratio X cursor units 202 428 ratio Y cursor units 202 raw acquisition record 273 Real Scope Remote Front Panel 310 rear panel connectors 41 re arm time 196 recall 303 recall files via web interface 316 recall mask files 278 recall setups 278 recall Quick Recall 303 Rectangular FFT window 82 Ref key 37 99 reference point waveform 293 reference waveforms 99 remote control 287 Remote Front Panel 313 remote programming Keysight lO Libraries 314 remote programming web interface 313 replacement parts 118 required oscilloscope bandwidth 186 reset network password 322 restart condition 12C trigger 371 return instrument for service 302 Rise time measurement 226 rise time measurement trend 95 rise time oscilloscope 185 rise time signal 186 rise fall time triggering 152 rising edge count measurements 230 RMS AC measurement trend 95 roll mode 51 RS232 trigger 415 Run Control keys 36 runt pulses 223 runt triggering 155 Russian front panel overlays 41 Russian user interface and Quick Help 45
31. 401 Interpreting MIL STD 1553 Lister Data 402 Searching for MIL STD 1553 Data in the Lister 403 Setup for ARINC 429 Signals 404 ARINC 429 Triggering 405 ARINC 429 Serial Decode 407 Interpreting ARINC 429 Decode 409 ARINC 429 Totalizer 410 Interpreting ARINC 429 Lister Data 411 Searching for ARINC 429 Data in the Lister 411 28 UART RS232 Triggering and Serial Decode Setup for UART RS232 Signals 413 UART RS232 Triggering 415 UART RS232 Serial Decode 417 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 22 Index Interpreting UART RS232 Decode 418 UART RS232 Totalizer 419 Interpreting UART RS232 Lister Data 420 Searching for UART RS232 Data in the Lister 420 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started Inspect the Package Contents 23 Tilt the Oscilloscope for Easy Viewing 26 Power On the Oscilloscope 27 Connect Probes to the Oscilloscope 28 Input a Waveform 29 Recall the Default Oscilloscope Setup 29 Use Auto Scale 30 Compensate Passive Probes 32 Learn the Front Panel Controls and Connectors 33 Learn the Rear Panel Connectors 41 Learn the Oscilloscope Display 43 Access the Built In Quick Help 44 This chapter describes the steps you take when using the oscilloscope for the first time Inspect the Package Contents Inspect the shipping
32. AL 17 Square Use the Frequency Frequency Fine Period Period 100 mHz to 20 mVpp to 2 50 V Fine Amplitude High Level Offset Low Level and 10 MHz 5 Vpp Duty Cycle softkeys to set the square wave signal parameters The duty cycle can be adjusted from 20 to 80 Ramp Use the Frequency Frequency Fine Period Period 100 mHz to 20 mVpp to 2 50 V Fine Amplitude High Level Offset Low Level and 200 kHz 5 Vpp Symmetry softkeys to set the ramp signal parameters Symmetry represents the amount of time per cycle that the ramp waveform is rising and can be adjusted from 0 to 100 Pulse Use the Frequency Frequency Fine Period Period 100 mHz to 20 mVpp to 2 50 V Fine Amplitude High Level Offset Low Level and 10 MHz 5 Vpp Wid th Wid th Fine softkeys to set the pulse signal parameters The pulse width can be adjusted from 20 ns to the period minus 20 ns DC Use the Offset softkey to set the DC level n a n a 2 50 V Noise Use the Amplitude High Level and n a 20 mVpp to 2 50 V Offset Low Level to set the noise signal parameters 5 Vpp Sine Cardinal Use the Frequency Frequency Fine Period Period 100 mHz to 20 mVpp to 1 25 V Fine Amplitude and Offset softkeys to set the sinc 1 MHz 5 Vpp signal parameters Exponential Use the Frequency Frequency Fine Period Period 100 mHz to 20 mVpp to 2 50 V Rise Fine Amplitude High Level and Offset Low Level 5 MHz 5 Vpp softkeys to set the exponential rise signal parameters
33. Analog Channel Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 339 22 Reference Fie Meader Number of Waveforms N Number of Segments 12 bytes Number of Waveform Buffers 1 Index 1 Time Tag 0 0 Waveform Header 1 140 bytes Waveform Data Header 1 12 bytes Voltage Data 1 buffer size Waveform Header 2 140 bytes Buffer Type 1 floating point Bytes per Point 4 Number of Waveform Buffers 1 Index 2 Time Tag time between segment 1 and 2 Waveform Date Buffer Type 1 floating point Bytes per Point 4 Header 2 12 bytes Voltage Data 2 buffer size Number of Waveform Buffers 1 Index N Time Tag time between segment 1 and N Waveform Header N 140 bytes Waveform Data Buffer Type 1 floating point Bytes per Point 4 Header N 12 bytes Voltage Data N buffer size CSV and ASCII XY files CSV and ASCII XY file structure on page 341 Minimum and Maximum Values in CSV Files on page 341 340 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Reference 22 CSV and ASCII XY file structure In CSV or ASCII XY format the Length control selects the number of points per segment All segments are contained in the CSV file or in each ASCII XY data file For example If the Length control is set to 1000 points there will be 1000 points rows in the spreadsheet per segment When saving all segments there are three header ro
34. Serial 1 or Serial 2 on which the C signals are being decoded Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide I2C SPI Triggering and Serial Decode 25 Trigger Menu Trigger Trigger 5 FC EEPROM 4 Press the Trigger softkey then turn the Entry knob to select the trigger condition Start Condition The oscilloscope triggers when SDA data transitions from high to low while the SCL clock is high For triggering purposes including frame triggers a restart is treated as a start condition Stop Condition The oscilloscope triggers when data SDA transitions from low to high while the clock SCL is high soa NAT TT NIDO nA 1 i I 1 scL 1 7 8 _ 9 1 7 _ 8 9 f71 Sy i t bam ii Jol j ees Start Address R Ack Data Ack Stop Condition Condition Missing Acknowledge The oscilloscope triggers when SDA data is high during any Ack SCL clock bit Address with no Ack The oscilloscope triggers when the acknowledge of the selected address field is false The R W bit is ignored Restart The oscilloscope triggers when another start condition occurs before a stop condition EEPROM Data Read The trigger looks for EEPROM control byte value 101 Oxxx on the SDA line followed by a Read bit and an Ack bit It then looks for the data value and qualifier set by the Data softkey and the Datais softkey When this event occurs the oscilloscope will trigger on the clock edge for the Ack bit af
35. all statistics and begins recording statistical data again Each time a new measurement for example frequency period or amplitude is added the statistics are reset and accumulation of statistical data begins again When the Single key is pressed statistics are reset and a single measurement is done count 1 Successive Single acquisitions accumulate statistical data and the count is incremented Press the Transparent softkey to disable the Transparent mode This shows statistics with a gray background Press the Transparent softkey again to enable nsparent mode This writes measurement values statistics and cursor values e screen without a background The Transparent setting affects measurement tics reference waveform information and the optional Mask test feature s PE tics display Ora Zg GeF nn ch ct a An Relative o When enabled the standard deviation shown in measurement stati tatistics becomes a relative standard deviation that is standard deviation mean n Max Count This softkey specifies the number of values used when calculating measurement statistics The Increment Statistics softkey only appears when the acquisition is stopped and the optional segmented memory feature is off Press the Single or Run Stop key to stop the acquisition You can use the horizontal position control in the Horizontal control section of the front panel to pan through the waveform Act
36. label list loading from text file 136 labels 133 labels auto increment 136 labels default library 137 LAN connection 289 LAN interface remote control 287 LAN port 42 LAN Settings softkey 289 290 LAN VGA module 26 42 LAN VGA module installation 26 landscape mode 284 language user interface and Quick Help 45 length control 273 length softkey 273 level trigger 140 LF Reject 176 library labels 135 licenses 331 333 LIN serial decode 354 LIN trigger 353 line voltage 27 Lister 122 load file 291 Load from 276 localized front panel overlay 40 Location 276 292 logic presets waveform generator 262 logic threshold 110 low pass filter math function 92 low frequency noise rejection 176 M magnify math function 94 mask files recall 278 MASK license 332 mask test trigger output 241 297 mask testing 237 mask TRIG OUT signal 297 math filters 92 Math key 37 math operators 74 math transforms 76 math visualizations 93 math 1 2 75 math 1 2 75 math addition 74 math differentiate 77 math divide 75 math FFT 81 math functions 71 math integrate 78 math multiply 75 math offset 73 math scale 73 math subtract 74 math transforms or filters on arithmetic operations 72 math units 73 math using waveform math 71 MATLAB binary data 334 Maximum measurement 216 maximum sample rate 187 Meas key 37 209
37. math and display settings is also saved in a separate file with a TXT extension Length sets the number of data points that will be output to the file For more information see Length Control on page 273 Save Seg when data is acquired to segmented memory you can specify whether the currently displayed segment is saved or all acquired segments are saved See also Saving Data from Segmented Memory on page 196 4 Finally press the Press to Save softkey A message indicating whether the save was successful is displayed See Also Binary Data bin Format on page 333 CSV and ASCII XY files on page 340 Minimum and Maximum Values in CSV Files on page 341 Length Control The Length control is available when saving data to CSV ASCII XY or BIN format files It sets the number of data points that will be output to the file Only displayed data points are saved The maximum number of data points depends on these things Whether acquisitions are running When stopped data comes from the raw acquisition record When running data comes from the smaller measurement record Whether the oscilloscope was stopped using Stop or Single Running acquisitions split memory to provide fast waveform update rates Single acquisitions use full memory Whether only one channel of a pair is turned on Channels 1 and 2 are one pair channels 3 and 4 are the other Acquisition memory is divided among the channels in
38. measurements 231 threshold digital channels 110 thumb drive 39 tilt for viewing 26 time measurements 222 time reference indicator 56 time re arm 196 timebase 50 times for saving data 274 timing logic bus chart 96 Tools keys 35 Top measurement 216 totalizer ARINC 429 410 totalizer CAN 349 totalizer FlexRay 365 totalizer UART rs232 419 tracking cursors 200 trademarks 2 Traditional Chinese front panel overlays 41 Traditional Chinese user interface and Quick Help 45 transforms math 76 transient withstand capability 325 Transparent backgrounds 294 TRIG OUT connector 42 296 Trig d trigger indicator 175 Trig d trigger indicator 175 Trigger controls 35 trigger coupling 175 trigger indicator Auto 175 trigger indicator Trig d 175 trigger indicator Trig d 175 trigger level 140 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Index trigger mode auto or normal 174 trigger mode Quick Trigger Mode 303 trigger output 296 trigger output mask test 241 297 trigger type ARINC 429 405 trigger type CAN 345 trigger type edge 142 trigger type edge then edge 144 trigger type FlexRay 360 trigger type glitch 145 trigger type hex bus 150 trigger type 12C 370 trigger type 12S 390 trigger type LIN 353 trigger type MIL STD 1553 399 trigger type Nth edge burst 153 trigger type OR 151 trigger type pattern 148 t
39. see UART RS232 Serial Decode on page 417 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide UART RS232 Triggering and Serial Decode 28 UART RS23 Serial Decode To set up the oscilloscope to capture UART RS232 signals see Setup for UART RS232 Signals on page 413 For UART RS232 triggering setup see UART RS232 Triggering on page 415 To set up UART RS232 serial decode 1 Press Serial to display the Serial Decode Menu Serial Decode Menu AX FRAMES 0000276334 TX FRAMES 0000165780 ERR 000003 82 0 9 Serial Mode Signals Bus Contig Settings Serial 1 UART AS232 y T y 2 Press Settings 3 In the UART RS232 Settings Menu press the Base softkey to select the base hex binary or ASCII in which decoded words are displayed Framing off When displaying words in ASCII the 7 bit ASCII format is used Valid ASCII characters are between 0x00 and 0x7F To display in ASCII you must select at least 7 bits in the Bus Configuration If ASCII is selected and the data exceeds 0x7F the data is displayed in hex When Bits is set to 9 in the UART RS232 Bus Configuration Menu the 9th alert bit is displayed directly to the left of the ASCII value which is derived from the lower 8 bits 4 Optional Press the Framing softkey and select a value In the decode display the chosen value will be displayed in light blue However if a parity error occurs the data will be displayed in red 5 Ifthe decode
40. select is highlighted in the OR line directly above the softkeys and in the upper right corner of the display next to the OR gate symbol Adjust the trigger level for the selected analog channel by turning the Trigger Level knob Press the Digital key and select Thresholds to set the threshold level for digital channels The value of the trigger level or digital threshold is displayed in the upper right corner of the display 5 For each channel you select press the Slope softkey and select Rising Falling Either or X Don t Care The selected slope is displayed above the softkeys Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 151 10 Triggers 100 057 Trig d D TTL KEYSIGHT TECHN S Acquisition Channels CE rt aT 0 AMC LCT WL WL Hl JI If all channels in the OR trigger are set to don t care the oscilloscope will not trigger 6 Tosetall analog and digital channels to the edge selected by the Slope softkey press the Set all Edges softkey Rise Fall Time Trigger The Rise Fall Time trigger looks for a rising or falling transition from one level to another level in greater than or less than a certain amount of time High level l Low level Rising edge time Falling edge time 1 Press the Trigger key 152 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Triggers 10 2 Inthe Trigger Menu press the Trigger softkey
41. the Quick Action Key on page 302 To select the grid type When the Video trigger type is selected see Video Trigger on page 158 and the vertical scaling of at least one displayed channel is 140 mV div the Grid softkey lets you select from these grid types Full the normal oscilloscope grid mV shows vertical grids labeled on the left from 0 3 V to 0 8 V IRE Institute of Radio Engineers shows vertical grids in IRE units labeled on the left from 40 to 100 IRE The 0 35 V and 0 7 V levels from the mV grid are also shown and labeled at the right When the IRE grid is selected cursor values are also shown in IRE units Cursor values via the remote interface are not in IRE units The mV and IRE grid values are accurate and match Y cursor values when the vertical scaling is 140 mV division and the vertical offset is 245 mV To select the grid type 130 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Display Settings 8 1 Press Display 2 Press the Grid softkey then turn the Entry knob a to select the grid type To adjust the grid intensity To adjust the display grid graticule intensity 1 Press Display 2 Press the Intensity softkey then turn the Entry knob a to change the intensity of the displayed grid The intensity level is shown in the Intensity softkey and is adjustable from O to 100 Each major vertical division in the grid corresponds to the vertical sensiti
42. the vertical amplitude of the waveform is tracked and measured The time and voltage positions are shown for the markers The vertical Y and horizontal X differences between the markers are shown as AX and AY values Binary Logic levels of displayed waveforms at the current X1 and X2 cursor positions are displayed above the softkeys in binary The display is color coded to match the color of the related channel s waveform Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Cursors 13 Cursors Menu Mode Source Cursors Binary X1 Hex Logic levels of displayed waveforms at the current X1 and X2 cursor positions are displayed above the softkeys in hexadecimal Cursors Menu Mode Source Cursors Units Hex X1 a Manual and Track Waveform modes can be used on waveforms that are displayed on the analog input channels including math functions Binary and Hex modes apply to digital signals of MSO oscilloscope models n Hex and Binary modes a level can be displayed as 1 higher than trigger evel O lower than trigger level indeterminate state 1 or X don t care n Binary mode X is displayed if the channel is turned off n Hex mode the channel is interpreted as 0 if turned off 4 Press Source or X1 Source X2 Source in the Track Waveform mode then select the input source for cursor values 5 Select the cursor s to be adjusted Push the Cursors knob then turn the Cursors
43. 00ms rm Points Menu Point Voltage 3 4 950 Omv Insert Point Remove Point Transparent To specify the voltage values of points Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 259 17 Waveform Generator 1 Press Point then use the Entry knob to select the point whose voltage value you wish to set 2 Press Voltage then use the Entry knob to set the point s voltage value To insert a point 1 Press Point then use the Entry knob to se point will be inserted 2 Press Insert Point All points are adjusted to maintain uniform To remove a point 1 Press Point then use the Entry knob to se 2 Press Remove Point All points are adjusted to maintain uniform lect the point after which the new time spacing between points ect the point you want to remove time spacing between points The Transparent softkey enables or disables transparent backgrounds When enabled you can view underlying waveforms When disabled the background is shaded Capturing Other Waveforms to the Arbitrary Waveform The Edit Waveform Menu is opened by pressin Waveform Generator Menu rm Menu Create New Edit Existing Interpolate g Edit Waveform in the main w To capture another waveform to the arbitrary waveform 1 Press Source then use the Entry knob to se reference location whose waveform you wis 2 Press Store Source to Arb lect the analog channel math or h to ca
44. 1553 ARINC 429 Triggering and Serial Decode 27 Interpreting MIL STD 1553 Lister Data on page 402 Searching for MIL STD 1553 Data in the Lister on page 403 Interpreting MIL STD 1553 Decode To display serial decode information you must press Run or Single after switching on serial decode 20 005 10 005 Channels D 0000 Mode Signals MIL 1553 wy Lister Auto Setup The MIL STD 1553 decode display is color coded as follows Command and Status decoded data is colored green with the Remote Terminal Address 5 bits of data being displayed first then the text C S followed by the value of the remaining 11 bits of aCommand Status word Data word decoded data is colored white preceded by the text D Command Status or Data words with a Parity error have the decode text displayed in red instead of green or white SYNC errors are displayed with the word SYNC within red angle brackets Manchester encoding errors are displayed with the word MANCH within blue angle brackets blue instead of red because a valid Sync pulse started the word Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 401 27 MIL STD 1553 ARINC 429 Triggering and Serial Decode Interpreting MIL STD 1553 Lister Data 7 3965 Stop 553 250 RTA Word Type Data Errors KEYSIGHT TECHNOLOGIES Acquisition 1 Channels Zoom to Options c pla S MIL 1553 Selection In add
45. 220 AC RMS N Cycles measurement 220 accessories 23 326 330 acquire 181 191 Acquire key 38 acquisition memory 140 acquisition memory saving 273 acquisition mode 187 acquisition mode averaging 191 acquisition mode high resolution 193 acquisition mode normal 188 acquisition mode peak detect 188 acquisition mode preserve during AutoScale 295 acquisition modes 181 active serial bus 348 356 375 384 394 418 activity indicator 109 actual sample rate 187 adding digital channels license 333 addition math function 74 address with no ack condition 12C trigger 371 Addresses softkey 290 ADVMATH license 331 AERO license 331 aliasing 183 aliasing FFT 85 alternating edge trigger 143 AM amplitude modulation waveform generator output 264 Amplitude measurement 216 amplitude modulation AM waveform generator output 264 analog channel inputs 39 analog channel probe attenuation 67 analog channel setup 61 analog filters adjusting 81 Analyze key 35 Analyze Segments 194 196 235 annotation adding 303 arbitrary generated waveforms editing 257 arbitrary waveform generator output 254 arbitrary waveforms copying from other sources 260 arbitrary waveforms creating new 258 arbitrary waveforms editing existing 259 Area Full Screen measurement 231 Area N Cycles measurement 231 ARINC 429 serial decode 407 ARINC 429 totalizer 410 ARINC 429 trigger 405 AR
46. 284 5 Press the Palette softkey to select the print palette See To specify the palette option on page 284 6 Press the Press to Print softkey You can stop printing by pressing the Cancel Print softkey The oscilloscope will print the last menu visited before you pressed the Print key Therefore if you have measurements Amplitude Frequency etc showing on the display before you press Print the measurements will be shown on the printout To print the display showing the Print Configuration Menu at the bottom press the Print key twice then press the Press to Print softkey You can also configure the Quick Action key to print the display See Configuring the Quick Action Key on page 302 See Also Adding an Annotation on page 303 To set up network printer connections When the DSOXLAN LAN VGA module is installed you can set up network printer connections A network printer is a printer attached to a computer or print server on the network 1 Press the Print key on the front panel 282 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Print Screens 19 2 In the Print Configuration Menu press the Print to softkey then turn the Entry knob to select the network printer you want to configure either 0 or 1 3 Press the Network Setup softkey 4 In the Network Printer Setup Menu press the Modify softkey then turn the Entry knob to select the network parameter you want
47. 429 Data in the Lister on page 411 Searching for CAN Data in the Lister on page 351 Searching for FlexRay Data in the Lister on page 366 Searching for I2C Data in the Lister on page 376 Searching for 12S Data in the Lister on page 395 Searching for LIN Data in the Lister on page 358 Searching for MIL STD 1553 Data in the Lister on page 403 Searching for SPI Data in the Lister on page 385 Searching for UART RS232 Data in the Lister on page 420 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 125 7 Serial Decode 126 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Display Settings To adjust waveform intensity 127 To set or clear persistence 129 To clear the display 130 To select the grid type 130 To adjust the grid intensity 131 To freeze the display 131 To adjust waveform intensity You can adjust the intensity of displayed waveforms to account for various signal characteristics such as fast time div settings and low trigger rates Increasing the intensity lets you see the maximum amount of noise and infrequently occurring events Reducing the intensity can expose more detail in complex signals as shown in the following figures 1 Press the Intensity key to illuminate it This key is located just below the Entry knob 2 Turn the Entry knob to adjust the waveform instensity Wavefor
48. 50 ohm High level high Z expected output load expected output load TTL OV 2 5 V TTL compatible 5 V CMOS 5 0V OV Not Available 5V CMOS 3 3V OV 2 5 V CMOS 3 3 V compatible CMOS 2 5V OV 2 5V 2 5 V ECL 1 7 V 0 8 V ECL compatible 0 9 V To add noise to the waveform generator output 1 If the Waveform Generator Menu is not currently displayed on the oscilloscope s softkeys press the Wave Gen key 2 Inthe Waveform Generator Menu press the Settings softkey 262 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Waveform Generator 17 3 In the Waveform Generator Settings Menu press the Add Noise softkey and turn the Entry knob to select the amount of white noise to add to the waveform generator output Note that adding noise affects edge triggering on the waveform generator source see Edge Trigger on page 142 as well as the waveform generator sync pulse output signal which can be sent to TRIG OUT see Setting the Rear Panel TRIG OUT Source on page 296 This is because the trigger comparator is located after the noise source To add modulation to the waveform generator output Modulation is where an original carrier signal is modified according to the amplitude of a second modulating signal The modulation type AM FM or FSK specifies how the carrier signal is modified To enable and set up modulation for the waveform generator output 1 Ifthe Waveform Gene
49. ARINC 429 signal is being decoded 3 Press Search then select from these options Label finds the specified label value Label values are always displayed in octal Label Bits finds the label and the other word fields as specified Parity Error finds words with a parity error Word Error finds an intra word coding error Gap Error finds an inter word gap error Word or Gap Error finds either a Word or Gap Error All Errors finds any of the above errors For more information on searching data see Searching Lister Data on page 124 For more information on using the Navigate key and controls see Navigating the Time Base on page 58 412 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 28 UART RS232 Triggering and Serial Decode Setup for UART RS232 Signals 413 UART RS232 Triggering 415 UART RS232 Serial Decode 417 UART RS232 triggering and serial decode requires Option COMP or the DSOX3COMP upgrade Setup for UART RS232 Signals To set up the oscilloscope to capture UART RS232 signals 1 Press Label to turn on labels 2 Press Serial 3 Press the Serial softkey turn the Entry knob to select the desired slot Serial 1 or Serial 2 and press the softkey again to enable decode 4 Press the Mode softkey then select UART RS232 trigger type 5 Press the Signals softkey to open the UART R
50. Digital Channels starting on page 103 Chapter 7 Serial Decode starting on page 121 Chapter 8 Display Settings starting on page 127 Chapter 9 Labels starting on page 133 When setting up triggers or changing how data is acquired see Chapter 10 Triggers starting on page 139 Chapter 11 Trigger Mode Coupling starting on page 173 Chapter 12 Acquisition Control starting on page 181 Making measurements and analyzing data Chapter 13 Cursors starting on page 199 Chapter 14 Measurements starting on page 209 Chapter 15 Mask Testing starting on page 237 Chapter 16 Digital Voltmeter starting on page 249 When using the built in license enabled waveform generator see Chapter 17 Waveform Generator starting on page 253 When saving recalling or printing see Chapter 18 Save Recall Setups Screens Data starting on page 269 Chapter 19 Print Screens starting on page 281 When using the oscilloscope s utility functions or web interface see Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Chapter 20 Utility Settings starting on page 287 Chapter 21 Web Interface starting on page 307 For reference information see Chapter 22 Reference starting on page 323 When using licensed serial bus triggering and decode features see Chapter 23 CAN
51. Exponential Use the Frequency Frequency Fine Period Period 100 mHz to 20 mVpp to 2 50 V Fall Fine Amplitude High Level and Offset Low Level 5 MHz 5 Vpp softkeys to set the exponential fall signal parameters Cardiac Use the Frequency Frequency Fine Period Period 100 mHz to 20 mVpp to 1 25 V Fine Amplitude and Offset softkeys to set the 200 kHz 5 Vpp cardiac signal parameters Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 255 17 Waveform Generator Waveform Characteristics Frequency WEVA Offset Type Range Amplitude High Z High Z Gaussian Pulse Use the Frequency Frequency Fine Period Period 100 mHz to 20 mVpp to 1 25 V Fine Amplitude and Offset softkeys to set the 5 MHz 4 Vpp gaussian pulse signal parameters 1When the output load is 50 Q these values are halved Pressing a signal parameter softkey can open a menu for selecting the type of adjustment For example you can choose to enter amplitude and offset values or you can choose to enter high level and low level values Or you can choose to enter frequency values or period values Keep pressing the softkey to select the type of adjustment Turn the Entry knob to adjust the value Notice that you can select between coarse and fine adjustments for frequency period and width Also pushing the Entry knob is a quick way to toggle between coarse and fine adjustments The Settings softkey opens the Waveform Generato
52. Interface To send remote programming commands to the oscilloscope via the SCPI Commands applet window 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 308 2 When the oscilloscope s web interface is displayed select Browser Web Control then select Remote Programming The SCPI Commands applet appears within the browser web page Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 313 21 Web Interface KEYSIGHT TECHNOLOGIES pegmatite te Remote Programming You can send remote programming commands directly to your MSO X 2024A oscilloscope using the tool below commands see the Programming Commands Quick Reference mi Browser Web Control Clear J Commands Options aaa IDN lf Successfully connected to Agilent MSOX2024A InfiniiVision US50210029 gt IDN lt AGILENT TECHNOLOGIES MSO X 2024A US50210029 02 38 2014102201 Remote Programming with Keysight IO Libraries While the SCPI Commands applet window lets you enter and remote programming commands remote programming for automated test and data acquisition is typically done using the Keysight IO Libraries which are separate from the instrument s web interface The Keysight IO Libraries let a controller PC communicate with Keysight InfiniiVision oscilloscopes via their USB LAN when the LAN VGA option module is installed or GPIB when
53. MIL STD 1553 ARINC 429 Triggering and Serial Decode 27 Use the Set all Bits softkey to set all bit values to O 1 or X 6 If you select the Label Range condition use the Lable Min and Label Max softkeys to specify the ends of the range Again label values are always displayed in octal You can use the Zoom mode for easier navigation of the decoded data To display ARINC 429 serial decode see ARINC 429 Serial Decode on page 407 ARINC 429 Serial Decode To set up the oscilloscope to capture ARINC 429 signals see Setup for ARINC 429 Signals on page 404 For ARINC 429 triggering set up see ARINC 429 Triggering on page 405 To set up ARINC 429 serial decode 1 Press Serial to display the Serial Decode Menu Serial De e 000707471 ERRORS QOQI Signals 2 Inthe submenu accessed by the Settings softkey you can use the Base softkey to select between hexadecimal and binary display of the decoded data The base setting is used for data display in both the decode line and in the Lister Label values are always displayed in octal and SSM and SDI values are always displayed in binary 3 Press the Word Format softkey and specify the word decode format Label SDI Data SSM Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 407 27 MIL STD 1553 ARINC 429 Triggering and Serial Decode Label 8 bits SDI 2 bits Data 19 bits SSM 2 bits Label Data SSM Label 8 bits Data 21 bits SSM
54. Measure controls 37 measure Quick Measure All 302 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide measurement category definitions 324 measurement definitions 211 measurement record 273 measurement statistics 233 measurement thresholds 231 measurement trend math function 94 measurement window with zoom display 233 measurements 211 measurements automatic 209 measurements delay 212 measurements overshoot 212 measurements phase 213 measurements preshoot 213 measurements time 222 measurements voltage 215 MegaZoom IV 4 mem4M 332 memory depth and sample rate 187 memory segmented 193 menu line 44 MIL STD 1553 serial decode 400 MIL STD 1553 trigger 399 Minimum measurement 216 missing acknowledge condition 12C trigger 371 Mode Coupling key trigger 173 model number 300 307 Modify softkey 290 modulation waveform generator output 263 module installed 300 module slot 42 MSO 4 MSO feature upgrade 333 MSO license 332 Multicast DNS 289 multiplexed position knob 37 multiplexed scale knob 37 Multiply math function 75 N N2820A high sensitivity current probe 213 N8900A InfiniiView oscilloscope analysis software 270 natural logarithm math function 91 navigate files 291 Navigate key 36 navigating the time base 58 negative pulse width measurement trend 95 network configuration parameters 307 network printer connec
55. Oscilloscopes User s Guide 337 22 Reference pla eedat Number of Waveforms N 12 bytes Waveform Header 1 Number of Waveform Buffers 1 140 bytes Waveform Date Buffer Type 1 floating point Bytes per Point 4 Header 1 12 bytes Waveform Data Header 2 12 bytes Waveform Data Header N 12 bytes Voltage Data N buffer size Voltage Data 1 buffer size Waveform Header 2 Number of Waveform Buffers 1 140 bytes Buffer Type 1 floating point Bytes per Point 4 Voltage Data 2 buffer size Waveform Header N Number of Waveform Buffers 1 140 bytes Buffer Type 1 floating point Bytes per Point 4 Single Acquisition The following picture shows a binary file of a single acquisition with all pods for the All Pods Logic logic channels saved Channels 338 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Reference 22 pla eedat Number of Waveforms 2 12 bytes Waveform Header 1 Number of Waveform Buffers 1 140 bytes Waveform Data Buffer Type 6 unsigned char Bytes per Point 1 Header 1 12 bytes Waveform Data Header 2 12 bytes Pod 2 Timing Data buffer size Pod 1 Timing Data buffer size Waveform Header 2 Number of Waveform Buffers 1 140 bytes Buffer Type 6 unsigned char Bytes per Point 1 Segmented The following picture shows a binary file of a segmented memory acquisition on Memory one analog channel Acquisition on One
56. Press the Signals softkey to open the MIL STD 1553 Signals Menu 3 Signals Menu High Threshold Low Threshold 250mV 250mV Press the Source softkey to select the channel connected to the MIL STD 1553 signal line The label for the MIL STD 1553 source channel is automatically set Press the O Back Up key to return to the Serial Decode Menu 8 Press the Auto Setup softkey which performs these actions Sets the probe attenuation factor of the input Source channel to 10 1 Sets upper and lower thresholds to a voltage value equal to 1 3 division based on the current V div setting Turns off trigger noise reject Turns on Serial Decode Sets the trigger type to MIL 1553 If the upper and lower thresholds are not set correctly by Auto Setup press the Signals softkey to return to the MIL STD 1553 Signals Menu Then Press the High Threshold softkey then turn the Entry knob to set the high trigger threshold voltage level Press the Low Threshold softkey then turn the Entry knob to set the low trigger threshold voltage level The threshold voltage levels are used in decoding and will become the trigger levels when the trigger type is set to the selected serial decode slot Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide MIL STD 1553 ARINC 429 Triggering and Serial Decode 27 MIL STD 1553 Triggering To set up the oscilloscope to capture a MIL STD 1553 signal see Setup for MIL STD 1553 Signa
57. Preview User 180min Spell A 3 Press the Wait softkey then turn the Entry knob to select the number of minutes to wait before the selected screen saver activates When you turn the Entry knob the number of minutes is displayed on the Wait softkey The default time is 180 minutes 3 hours 4 Press the Preview softkey to preview the screen saver you have selected with the Saver softkey 5 To view the normal display after the screen saver has started press any key or turn any knob To set AutoScale preferences 1 Press Utility gt Options gt Preferences gt AutoScale 2 Inthe AutoScale Preferences Menu you can Press the Fast Debug softkey enable disable this type of autoscale When fast debug is enabled autoscale lets you make quick visual comparisons to determine whether the signal being probed is a DC voltage ground or an active AC signal Channel coupling is maintained to allow easy viewing of oscillating signals Press the Channels softkey and turn the Entry knob to specify the channels to be autoscaled All Channels The next time you press AutoScale all channels that meet the requirements of AutoScale will be displayed Only Displayed Channels The next time you press AutoScale only the channels that are turned on will be examined for signal activity This is useful if you only want to view specific active channels after pressing AutoScale Press the Acq Mode softkey and turn the Entry knob
58. See To navigate storage locations on page 276 3 Press the Press to Recall softkey A message indicating whether the recall was successful is displayed 4 If you would like to clear the display press Clear Display See Also To edit arbitrary waveforms on page 257 Recalling Default Setups 1 Press the Save Recall key 2 Inthe Save Recall Menu press Default Erase 3 In the Default Menu press one of these softkeys Keysight InfiniiVision 3000 X Series Oscil loscopes User s Guide 279 18 Save Recall Setups Screens Data Default Setup recalls the oscilloscope s default setup This is the same as pressing the front panel Default Setup key See Recall the Default Oscilloscope Setup on page 29 Some user settings are not changed when recalling the default setup Factory Default recalls the oscilloscope s factory default settings You must confirm the recall because there are no user settings that are left unchanged Performing a Secure Erase 1 Press the Save Recall key 2 Inthe Save Recall Menu press Default Erase 3 In the Default menu press Secure Erase This performs a secure erase of all non volatile memory in compliance with National Industrial Security Program Operation Manual NISPOM Chapter 8 requirements You must confirm the secure erase and the oscilloscope will reboot when finished 280 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Keysight Infinii
59. Sigma minimum 240 Simple Remote Front Panel 311 Simplified Chinese front panel overlays 41 Simplified Chinese user interface and Quick Help 45 sinc waveform generator output 255 sine waveform generator output 254 single acquisition 36 Single key 181 single ended active probes 327 single shot acquisitions 175 single shot events 181 size 109 skew analog channel 68 slope trigger 142 Snapshot All measurements 214 snapshot all quick action 302 softkey labels 44 softkeys 6 34 software updates 333 software version 300 Span FFT 81 Spanish front panel overlays 41 Spanish user interface and Quick Help 45 specifications 323 spectral leakage FFT 87 SPI serial decode 383 SPI trigger 381 square math function 89 square root 87 square waveform generator output 255 square waves 184 stand alone connection 290 start acquisition 36 start condition 12C 371 state logic bus chart 97 statistics increment 235 statistics mask test 242 statistics measurement 233 statistics using segmented memory 196 status line 44 status User Cal 301 Std Deviation measurement 220 stop acquisition 36 stop condition 12C 371 storage locations navigate 276 subnet mask 289 subtract math function 74 sync pulse waveform generator 261 T template front panel 40 test mask 237 Thai front panel overlays 41 theory sampling 183 threshold analog channel
60. To start acquiring data press Run Stop To capture and display a single acquisition whether the oscilloscope is running or stopped press Single The Single key is yellow until the oscilloscope triggers For more information see Running Stopping and Making Single Acquisitions Run Control on page 181 Default Setup key Press this key to restore the oscilloscope s default settings details on Recall the Default Oscilloscope Setup on page 29 10 Auto Scale key When you press the AutoScale key the oscilloscope will quickly determine which channels have activity and it will turn these channels on and scale them to display the input signals See Use Auto Scale on page 30 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Getting Started 1 11 Additional waveform controls The additional waveform controls consist of Math key provides access to math add subtract etc waveform functions See Chapter 4 Math Waveforms starting on page 71 Ref key provides access to reference waveform functions Reference waveforms are saved waveforms that can be displayed and compared against other analog channel or math waveforms See Chapter 5 Reference Waveforms starting on page 99 Digital key Press this key to turn the digital channels on or off the arrow to the left will illuminate When the arrow to the left of the Digita
61. Triggering With the built in MegaZoom technology you can simply AutoScale the waveforms then stop the oscilloscope to capture a waveform You can then pan and zoom through the data using the Horizontal and Vertical knobs to find a stable trigger point AutoScale often produces a triggered display Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 143 10 Triggers Edge then Edge Trigger The Edge then Edge trigger mode triggers when the Nth edge occurs after an arming edge and a delay period The arm and trigger edges can be specified as 4 Rising or Falling edges on analog or digital channels 1 Press the Trigger key 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select Edge then Edge Channels Edge then Edge D7 100us Chi t 2 Trigger Sources Delay Nth Edge B Edge then Edge se 100us 2 3 Press the Sources softkey 4 In the Edge then edge Sources Menu 144 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Triggers 10 Edge then Edge D 100us Chi t 2 Arm amp Slope A Trigger B a Press the Arm A softkey and turn the Entry knob to select the channel on which the arming edge will occur b Press the Slope A softkey to specify which edge of the Arm A signal will arm the oscilloscope c Press the Trigger B softkey and turn the Entry knob to select the channel on which the trigger edge will occur d Press the Slope B soft
62. Triggering on an infrequent signal from a serial bus for example 12C SPI CAN LIN etc or another signal that arrives in bursts The Normal trigger mode lets you stabilize the display by preventing the oscilloscope from auto triggering Making single shot acquisitions with the Single key Often with single shot acquisitions you must initiate some action in the device under test and you don t want the oscilloscope to auto trigger before that happens Before initiating the action in the circuit wait for the trigger condition indicator Trig d to flash this tells you the pre trigger buffer is filled See Also Forcing a Trigger on page 141 To set the trigger holdoff on page 177 To position the time reference left center right on page 56 To select the trigger coupling 1 Press the Mode Coupling key Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 175 11 Trigger Mode Coupling 2 Inthe Trigger Mode and Coupling Menu press the Coupling softkey then turn the Entry knob to select DC coupling allows DC and AC signals into the trigger path AC coupling places a 10 Hz high pass filter in the trigger path removing any DC offset voltage from the trigger waveform The high pass filter in the External Trigger input path is 50 Hz for all models Use AC coupling to get a stable edge trigger when your waveform has a large DC offset LF low frequency Reject coupling adds a high pas
63. User s Guide Cursors 13 1 00 KEYSIGHT TECHNO Acquisition Channels Cursors AY 1 Mode Source Cursors Manual X1X2 Figure 44 Moving the cursors together to check pulse width variations Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 207 13 Cursors 208 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 14 Measurements To make automatic measurements 210 Measurements Summary 211 Voltage Measurements 215 Time Measurements 222 Count Measurements 229 Mixed Measurements 230 Measurement Thresholds 231 Measurement Window with Zoom Display 233 Measurement Statistics 233 The Meas key lets you make automatic measurements on waveforms Some measurements can only be made on analog input channels The results of the last four selected measurements are displayed in the easurements information area on the right hand side of the screen Cursors are turned on to show the portion of the waveform being measured for the ost recently selected measurement bottom most on the right side measurement area 3 Post Acquisition Processing In addition to changing display parameters after the acquisition you can perform all of the measurements and math functions after the acquisition Measurements and math functions will be recalculated as you pan and zoom and turn channels on and off As you zoom in an
64. User s Guide 157 10 Triggers 50 002 Sal 1 76V EEE Acquisition MEMAN gger lt Setup Setup and Hold 16 0ns 9 Press the lt Hold softkey and turn the Entry knob to select the hold time Video Trigger 158 Video triggering can be used to capture the complicated waveforms of most standard analog video signals The trigger circuitry detects the vertical and horizontal interval of the waveform and produces triggers based on the video trigger settings you have selected The oscilloscope s MegaZoom IV technology gives you bright easily viewed displays of any part of the video waveform Analysis of video waveforms is simplified by the oscilloscope s ability to trigger on any selected line of the video signal It is important when using a 10 1 passive probe that the probe is correctly compensated The oscilloscope is sensitive to this and will not trigger if the probe is not properly compensated especially for progressive formats 1 Press the Trigger key Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Triggers 10 2 In the Trigger Menu press the Trigger softkey then turn the Entry knob to select Video 200 08 Vid Field1 KEYSIGHT TECHNOLOGIES Acquisition Channels Settings gt 3 Press the Source softkey and select any analog channel as the video trigger source The selected trigger source is displayed in the upper right corner of the display Turning the Trigger Level knob does not ch
65. Waveforms Magnity The magnify math function available with the DSOX3ADVMATH advanced math measurements license lets you display an existing input source at different vertical settings to provide more vertical detail 200 08 Channels x Measurements Pk Pk 3 ae Pk Pk M TF Scale 100m V Operator Source 1 Magnify Figure 16 Example of Magnify See Also Ax B on page 88 Measurement Trend The measurement trend math function available with the DSOX3ADVMATH advanced math measurements license shows measurement values for a waveform based on measurement threshold settings as the waveform progresses across the screen For every cycle a measurement is made and the value is displayed on the screen for the cycle 94 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Math Waveforms 4 47 405 Channels Cursors 989 94kHz 850 78kHz Operator Source 1 Thresholds Meas Trend Ema Figure 17 Example of Measurement Trend Use the Type softkey to select the measurement whose trend you want to look at You can display trend values for these measurements Average RMS AC Ratio Period Frequency Width Width Duty Cycle Rise Time Fall Time Use the Thresholds softkey to access the Measurement Threshold Menu See Measurement Thresholds on page 231 If a measurement cannot be made for part of a waveform the trend function output is a hole that is no value until a measurement can be made
66. When using the N2820A probe to make measurements on a battery powered floating device always connect the supplied ground lead between ground on your device and the probe s ground connector as shown in the following figure Simply snap the end of the ground lead onto the probe s connector Without the ground connection the probe s common mode input amplifier cannot properly display waveforms Ground Connectior on Probe Battery Powered Device Figure 45 Measurements on Battery Powered Devices Using the N2820A Probe Snapshot All The Snapshot All measurement type displays a popup containing a snapshot of all the single waveform measurements 214 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Measurements 14 1 00 1 00 500 05 KEYSIGHT TECH Acquisition Channels x Measurements PKPRI easurement Menu Source Type Settings Clear Meas c n hot Me gt w You can also configure the Quick Action key to display the Snapshot All popup See Configuring the Quick Action Key on page 302 Voltage Measurements The following figure shows the voltage measurement points Maximum Measurement units for each input channel can be set to Volts or Amps using the channel Probe Units softkey See To specify the channel units on page 67 The units of math waveforms are described in Units for Math Waveforms on page 73 Peak Peak on page 216 Maximum on page 216 Keysight Infinii
67. a pair Whether reference waveforms are on Displayed reference waveforms consume acquisition memory Whether digital channels are on Displayed digital channels consume acquisition memory Whether segmented memory is on Acquisition memory is divided by the number of segments Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 273 18 Save Recall Setups Screens Data The horizontal time div sweep speed setting At faster settings fewer data points appear on the display When saving to a CSV format file the maximum number of data points is 64K When necessary the Length control performs a 1 of n decimation of the data For example if the Length is set to 1000 and you are displaying a record that is 5000 data points in length four of each five data points will be decimated creating an output file 1000 data points in length When saving waveform data the save times depend on the chosen format Data File Format Save Times BIN fastest ASCII XY medium CSV slowest See Also Binary Data bin Format on page 333 CSV and ASCII XY files on page 340 Minimum and Maximum Values in CSV Files on page 341 To save Lister data files Lister data files can be saved to an external USB storage device 1 Press Save Recall gt Save gt Format then turn the Entry knob to select Lister data file 2 Press the softkey in the second position and use the Entry knob to navigate
68. and Serial Decode Interpreting UART RS232 Lister Data 1 0003 Stop 1 12V KEYSIGHT TECHNOLOGIES Acquisition 210 8ms Zoom to Options Selection Zoom w In addition to the standard Time column the UART RS232 Lister contains these columns Rx receive data Tx transmit data Errors highlighted in red Parity Error or Unknown Error Aliased data is highlighted in pink When this happens decrease the horizontal time div setting and run again Searching for UART RS232 Data in the Lister The oscilloscope s search capability lets you search for and mark certain types of UART RS232 data in the Lister You can use the Navigate key and controls to navigate through the marked rows 1 With UART RS232 selected as the serial decode mode press Search 2 Inthe Search Menu press the Search softkey then turn the Entry knob to select the serial slot Serial 1 or Serial 2 on which the UART RS232 signals are being decoded 3 In the Search Menu press Search then select from these options Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide UART RS232 Triggering and Serial Decode 28 Rx Data Finds a data byte that you specify For use when the DUT data words are from 5 to 8 bits in length no 9th alert bit Rx 1 Data For use when the DUT data words are 9 bits in length including the alert bit the 9th bit Finds only when the 9th alert bit is 1 The spec
69. be available to make additional measurement settings on some measurements 5 Press the Add Measurement softkey or push the Entry knob to display the measurement Cursors are turned on to show the portion of the waveform being measured for the most recently added measurement bottom most on the display To view the cursors for a previously added measurement but not the last one add the ce measurement again By default measurement statistics are displayed See Measurement Statistics on page 233 6 To turn off measurements press the Meas key again he Measurements are erased from the display 7 To stop making one or more measurements press the Clear Meas softkey and choose the measurement to clear or press Clear All ments Menu s1 Clear Meas 2 Freq 1 Pk Pk 1 After all measurements have been cleared when Meas is pressed again the default measurements will be Frequency and Peak Peak Measurements Summary The automatic measurements provided by the oscilloscope are listed in the following table All measurements are available for analog channel waveforms All measurements except Counter are available for reference waveforms and math waveforms other than FFT A limited set of measurements is available for math FFT waveforms and for digital channel waveforms as described in the following table Measurement Valid for Notes Digital Channels Snapshot All on page 214 Keysight Infini
70. container for damage If your shipping container appears to be damaged keep the shipping container or cushioning material until you have inspected the contents of the shipment for completeness and have checked the oscilloscope mechanically and electrically Verify that you received the following items and any optional accessories you may have ordered KEYSIGHT TECHNOLOGIES 23 1 Getting Started InfiniiVision 3000 X Series oscilloscope Power cord country of origin determines specific type Oscilloscope probes Two probes for 2 channel models Four probes for 4 channel models Documentation CD ROM 24 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Getting Started 1 6 O mm 4 N2862B N2863B or N2890A probes Qty 2 or 4 Documentation CD Digital Probe Kit MSO models only Power cord Based on country of origin N6450 60001 Digital Probe Kit contains 54620 61601 16 channel cable qyt 1 01650 82103 2 inch probe ground leads qyt 5 5090 4832 Grabber qyt 20 Digital probe replacement parts are listed in the Digital Channels chapter See Also Accessories Available on page 330 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 25 1 Getting Started Install the Optional LAN VGA or GPIB Module If you need to install a DSOXLAN LAN VGA module or a DSOXGPIB GPIB module perform this
71. device The softkey in the second position can have these labels Press to go when you can push the Entry knob to navigate to a new folder or storage location Location when you have navigated to the current folder location and are not saving files Save to when you can save to the selected location Load from when you can recall from the selected file When saving files The proposed file name is shown in the Save to file line above the softkeys To overwrite an existing file browse to that file and select it To create a new file name see To enter file names on page 276 To enter file names To create new file names when saving files to a USB storage device 1 In the Save Menu press the File Name softkey You must have a USB storage device connected to the oscilloscope for this softkey to be active 2 Inthe File Name Menu use the Spell Enter and Delete Character softkeys to enter the file name 276 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Save Recall Setups Screens Data 18 Spell press this softkey and turn the Entry knob to select the character at the current position Enter press this softkey to enter characters and move the cursor to the next character position Pushing the Entry knob is the same as pressing the Enter softkey Delete Character press this softkey to delete the character at the current position You can use a connected USB keyboard i
72. displayed To enable mask test order Option LMT at time of oscilloscope purchase or order DSOX3MASK as a stand alone item after oscilloscope purchase To create a mask from a golden waveform Automask A golden waveform meets all chosen parameters and it is the waveform to which all others will be compared 1 Configure the oscilloscope to display the golden waveform 2 Press the Analyze key 3 Press Features then select Mask Test 4 Press Features again to enable mask testing KEYSIGHT TECHNOLOGIES 237 15 Mask Testing 238 Failures Channel 1 O Features Automask w 5 Press Automask In the Automask Menu press the Source softkey and ensure the desired analog channel is selected Automask Menu Source Units Create Divisions 0 10 0 10 Mask Adjust the mask s horizontal tolerance Y and vertical tolerance X These are adjustable in graticule divisions or in absolute units volts or seconds selectable using the Units softkey Press the Create Mask softkey The mask is created and testing begins Whenever the Create Mask softkey is pressed the old mask is erased and a new mask is created Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Mask Testing 15 77505 2 0005 Channels Curr Si ate Create Divisions 0 10 ERS 9 Toclear the mask and switch off mask testing press the Back Up key to return to the Mask Test Menu then press the Clear Mask softkey
73. environment Pollution Degree 2 Normally only dry non conductive pollution occurs Occasionally a temporary conductivity caused by condensation may occur Example General indoor environment Pollution Degree 3 Conductive pollution occurs or dry non conductive pollution occurs which becomes conductive due to condensation which is expected Example Sheltered outdoor environment Probes and Accessories 326 AutoProbe Interface See Also This setion lists the probes and accessories that are compatible with the 3000 X Series oscilloscopes Passive Probes on page 327 Single Ended Active Probes on page 327 Differential Probes on page 328 Current Probes on page 329 Accessories Available on page 330 Most Keysight single ended active differential and current probes are compatible with the AutoProbe interface Active probes that do not have their own external power supply require substantial power from the AutoProbe interface In the following tables for AutoProbe interface compatible probes Quantity Supported indicates the maximum number of each type of active probe that can be connected to the oscilloscope If too much current is drawn from the AutoProbe interface an error message will be displayed indicating that you must momentarily disconnect all probes to reset the AutoProbe interface then connect only the supported quantity of active probes For more information on probes and acce
74. f t Displayed 3 Use the Operator softkey to select an operator or transform For more information on the operators see Math Operators on page 74 Math Transforms on page 76 Math Filters on page 92 Math Visualizations on page 93 4 Use the Source 1 softkey to select the analog channel on which to perform math You can rotate the Entry knob or repetitively press the Source 1 softkey to make your selection If you choose a transform function differentiate integrate FFT or square root the result is displayed 5 If you select an arithmetic operator use the Source 2 softkey to select the second source for the arithmetic operation The result is displayed oe 6 To re size and re position the math waveform see To adjust the math waveform scale and offset on page 73 TIP Math Operating Hints If the analog channel or math function is clipped not fully displayed on screen the resulting displayed math function will also be clipped Once the function is displayed the analog channel s may be turned off for better viewing of the math waveform The vertical scaling and offset of each math function can be adjusted for ease of viewing and measurement considerations The math function waveform can be measured using Cursors and or Meas To perform transforms or filters on an arithmetic operation To perform a transform function see Math Transforms on page 76 or filter see Math Filters on
75. filters on an arithmetic operation on page 72 Units for Math Waveforms on page 73 Integrate J dt integrate calculates the integral of the selected source You can use integrate to calculate the energy of a pulse in volt seconds or measure the area under a waveform J dt plots the integral of the source using the Trapezoidal Rule The equation is dy Se Sy i 0 Where integrated waveform At point to point time difference y channel 1 2 3 or 4 or g t internal arithmetic operation 78 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Math Waveforms 4 co arbitrary constant i data point index The integrate operator provides an Offset softkey that lets you enter a DC offset correction factor for the input signal Small DC offset in the integrate function input or even small oscilloscope calibration errors can cause the integrate function output to ramp up or down This DC offset correction lets you level the integrate waveform Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 79 4 Math Waveforms 0 V level Integrate without DC offset correction f t s ChT dt Scale 20 OuVs Function Operator Source 1 f t fdt Integrate with DC offset correction _ Scale 5 00uV s Operator Source 1 Jat 1 Figure 8 Integrate and Signal Offset See Also on page 73 F 1 2 12V AV KEYSIGHT TECHN ES Acquisition i al Channels 10 0 1 1 00 1 Of
76. installation before you power on the oscilloscope 1 If you need to remove a module before installing a different module pinch the module s spring tabs and gently remove the module rom the slot 2 To install a module slide the module into the slot on the back until it is fully seated The module s spring tabs will latch into the slot keepi a i LANIVGA N Module zil m Device Host TRIG OUT EXT Tha IN ee R te Module Slot GPIB Module ng the module in place The LAN VGA or GPIB module must be installed before powering on the oscilloscope Tilt the Oscilloscope for Easy Viewing 26 There are tabs under the oscilloscope s front feet that ca oscilloscope Keysight InfiniiVision 3000 X Seri n be flipped out to tilt the es Oscilloscopes User s Guide Getting Started 1 Flip Out Tabs Power On the Oscilloscope Power Line voltage frequency and power Requirements Line 100 120 Vac 50 60 400 Hz 100 240 Vac 50 60 Hz 100 W max Ventilation The air intake and exhaust areas must be free from obstructions Unrestricted air Requirements flow is required for proper cooling Always ensure that the air intake and exhaust areas are free from obstructions The fan draws air in from the left side and bottom of the oscilloscope and pushes it out behind the oscilloscop
77. interval in the specified field 1 or 2 1 Press the Trigger key 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select Video 3 Press the Settings softkey then press the Standard softkey to select the appropriate TV standard 4 Press the Mode softkey and select Field1 or Field2 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 165 10 Triggers 200 05 o Vid Field KEYSIGHT TECHNOLOGIES Acquisition Channels Mode Field Holdoff Field1 O Oflds Figure 29 Triggering on Field 1 To trigger on all fields of the video signal To quickly and easily view transitions between fields or to find the amplitude differences between the fields use the All Fields trigger mode 1 Press the Trigger key 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select Video 3 Press the Settings softkey then press the Standard softkey to select the appropriate TV standard 4 Press the Mode softkey and select All Fields 166 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide All Fields Figure 30 Triggering on All Fields To trigger on odd or even fields Triggers 10 200 08 o Vid All Fld KEYSIGHT TECHNOLOGIES Acquisition Channels Field Holdoff O Oflds To check the envelope of your video signals or to measure worst case distortion t trigger on he odd or even fields When Field 1 is selected the oscilloscope tr
78. is being measured The Y cursor shows the middle threshold point Frequency Frequency is defined as 1 Period Period is defined as the time between the middle threshold crossings of two consecutive like polarity edges A middle threshold crossing must also travel through the lower and upper threshold levels which eliminates runt pulses The X cursors show what portion of the waveform is being measured The Y cursor shows the middle threshold point See Also To isolate an event for frequency measurement on page 223 To isolate an event for frequency measurement The following figure shows how to use Zoom mode to isolate an event for a frequency measurement You may need to change the measurement window setting so that the measurement is made in the lower Zoom window See Measurement Window with Zoom Display on page 233 If the waveform is clipped it may not be possible to make the measurement Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 223 14 Measurements 224 Counter 50 005 Measurement Menu Source Settings Clear Meas Measurement T Ea Figure 48 Isolating an event for Frequency measurement The InfiniiVision 3000 X Series oscilloscopes have an integrated hardware frequency counter which counts the number of cycles that occur within a period of time known as the gate time to measure the frequency of a signal The gate time for the Counter measurement is automatically adjusted to be 1
79. is the segment number Only used when saving multiple segments Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Reference 22 Waveform Data A waveform may have more than one data set Each waveform data set will have a Header waveform data header The waveform data header consists of information about the waveform data set This header is stored immediately before the data set Waveform Data A 32 bit integer that is the size of the waveform data header Header Size Buffer Type A 16 bit short that is the type of waveform data stored in the file 0Q Unknown data 1 Normal 32 bit float data 2 Maximum float data 3 Minimun float data 4 Not used in InfiniiVision oscilloscopes 5 Not used in InfiniiVision oscilloscopes 6 Digital unsigned 8 bit char data for digital channels Bytes Per Point A 16 bit short that is the number of bytes per data point Buffer Size A 32 bit integer that is the size of the buffer required to hold the data points Example Program for Reading Binary Data To find an example program for reading binary data direct your web browser to www keysight com find 3000X Series examples and select Example Program for Reading Binary Data Examples of Binary Files Single Acquisition The following picture shows a binary file of a single acquisition with multiple Multiple Analog analog channels Channels Keysight InfiniiVision 3000 X Series
80. knob To finalize your selection either push the Cursors knob again or wait about five seconds for the popup menu to disappear Or Press the Cursors softkey then turn the Entry knob The X1 X2 linked and Y1 Y2 linked selections let you adjust both cursors at the same time while the delta value remains the same This can be useful for example for checking pulse width variations in a pulse train The currently selected cursor s display brighter than the other cursors 6 Tochange the cursor units press the Units softkey In the Cursor Units Menu Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 201 202 13 Cursors Cursor Units Menu X Units Phase Use X Cursors Y Units As 360 ase You can press the X Units softkey to select Seconds s Hz 1 s Phase wh location as 0 Ratio en selected use the Use X Cursors softkey to set the current X1 degrees and the current X2 location as 360 degrees when selected use the Use X Cursors softkey to set the current X1 location as 0 and the current X2 location as 100 You can press the Y Units softkey to select Base the Ratio same units used for the source waveform wh en selected use the Use Y Cursors softkey to set the current Y1 location as 0 and the current Y2 location as 100 For phase or ratio units once the O and 360 degree or 0 and 100 locations are set adjusting cursors wi
81. light blue Decoded text is truncated at the end of the associated frame when there is insufficient space within frame boundaries Pink vertical bars indicate you need to expand the horizontal scale and run again to see decode When the horizontal scale setting does not permit the display of all available decoded data red dots will appear in the decoded bus to mark the location of hidden data Expand the horizontal scale to allow the data to display An unknown undefined bus is shown in red A parity error will cause the associated data word to be shown in red which includes the 5 8 data bits and the optional 9th bit UART RS232 Totalizer The UART RS232 totalizer consists of counters that provide a direct measure of bus quality and efficiency The totalizer appears on screen whenever UART RS232 Decode is ON in the Serial Decode Menu The totalizer is running counting frames and calculating the percentage of error frames even when the oscilloscope is stopped not acquiring data The ERR error counter is a count of Rx and Tx frames with parity errors The TX FRAMES and RX FRAMES counts include both normal frames and frames with parity errors When an overflow condition occurs the counter displays OVERFLOW The counters can be reset to zero by pressing the Reset UART Counters softkey in the UART RS232 Settings Menu Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 419 28 420 UART RS232 Triggering
82. off 109 To switch all digital channels on or off 109 To switch groups of channels on or off 110 To change the logic threshold for digital channels 110 To reposition a digital channel 111 To display digital channels asa bus 111 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Digital channel signal fidelity Probe impedance and grounding 114 Input Impedance 114 Probe Grounding 116 Best Probing Practices 118 To replace digital probe leads 118 7 Serial Decode Serial Decode Options 121 Lister 122 Searching Lister Data 124 8 Display Settings To adjust waveform intensity 127 To set or clear persistence 129 To clear the display 130 To select the grid type 130 To adjust the grid intensity 131 To freeze the display 131 9 Labels To turn the label display on or off 133 To assign a predefined label to a channel 134 To define anew label 135 To load a list of labels from a text file you create 136 To reset the label library to the factory default 137 10 Triggers Adjusting the Trigger Level 140 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Forcing a Trigger 141 Edge Trigger 142 Edge then Edge Trigger 144 Pulse Width Trigger 145 Pattern Trigger 148 Hex Bus Pattern Trigger 150 OR Trigger 151 Rise Fall Time Trigger 152 Nth Edge Burst Trigger 153 Runt Trigger 155 Setup and Hold Trigger 157 Video Trigger 158 T
83. oscilloscope triggers a rising edge occurs on TRIG OUT The rising edge is delayed 30 ns from the oscilloscope s trigger point The output level is 0 5 V into an open circuit and 0 2 5 V into 50 Q See Chapter 10 Triggers starting on page 139 296 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Utility Settings 20 Mask The pass fail status is evaluated periodically When the evaluation of the testing period results in a failure the trigger output pulses high 5 V Otherwise the trigger output remains at low 0 V See Chapter 15 Mask Testing starting on page 237 Waveform Generator Sync Pulse All of the waveform generator output functions except DC Noise and Cardiac have an associated Sync signal The Sync signal is a TTL positive pulse that occurs when the waveform rises above zero volts or the DC offset value See Chapter 17 Waveform Generator starting on page 253 The TRIG OUT connector also provides the User Cal signal See To perform user calibration on page 297 Performing Service Tasks The Service Menu under Utility gt Service lets you perform service related tasks Service Menu rt Hardware Front Panel User Cal Self Test Self Test To perform user calibration on page 297 To perform hard ware self test on page 300 To perform front panel self test on page 300 To display oscilloscope information on page 300 To display the user calibrati
84. page 92 on the add subtract or multiply arithmetic operations 72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Math Waveforms 4 1 Press the Function softkey and select g t Internal 2 Use the Operator Source 1 and Source 2 softkeys to set up an arithmetic operation 3 Press the Function softkey and select f t Displayed 4 Use the Operator softkey to select a transform function or filter 5 Press the Source 1 softkey and select g t as the source Note that g t is only available when you select a transform function in the previous step To adjust the math waveform scale and offset 1 Make sure the multiplexed scale and position knobs to the right of the Math key are selected for the math waveform lf the arrow to the left of the Math key is not illuminated press the key 2 Use the multiplexed scale and position knobs just to the right of the Math key to re size and re position the math waveform 3 Math Scale and Offset are Set Automatically Any time the currently displayed math function definition is changed the function is automatically scaled for optimum vertical scale and offset If you manually set scale and offset for a function select a new function then select the original function the original function will be automatically rescaled See Also Units for Math Waveforms on page 73 Units for Math Waveforms Units for each input channel can be set to Volts or Amps using the
85. pink When this happens decrease the horizontal time div setting and run again Searching for 12S Data in the Lister The oscilloscope s search capability lets you search for and mark certain types of 12S data in the Lister You can use the Navigate key and controls to navigate through the marked rows 1 With 12S selected as the serial decode mode press Search 2 Inthe Search Menu press the Search softkey then turn the Entry knob to select the serial slot Serial 1 or Serial 2 on which the 12S signals are being decoded 3 In the Search Menu press Search then select from these options Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 395 26 396 12S Triggering and Serial Decode Equal finds the specified audio channel s data word when it equals the specified word Not Equal finds any word other than the specified word lt Less than finds when the channel s data word is less than the specified value gt Greater than finds when the channel s data word is greater than the specified value gt lt In Range enter upper and lower values to specify the range to find lt gt Out of Range enter upper and lower values to specify range not to find Errors finds all errors For more information on searching data see Searching Lister Data on page 124 For more information on using the Navigate key and controls see Navigating the Time Base on
86. portion of EOP Press Default Setup 2 Press the Label key to switch labels on 3 Turn on any analog or digital channels that you will be using for the USB signals 4 Press the Trigger key Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Triggers 10 5 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select USB 15 005 Channels Trigger Speed SOP Full Speed 6 Press the Trigger softkey to select where the USB trigger will occur SOP Start of Packet triggers at the Sync bit at the beginning of the packet EOP End of Packet triggers at the end of the SEO portion of the EOP RC Reset Complete triggers when SEO is gt 10 ms Suspend Enter Suspend triggers when the bus is idle gt 3 ms Exit Sus Exit Suspend triggers when exiting an idle state gt 10 ms This is used to see the suspend resume transition 7 Press the Speed softkey to select the speed of the transaction being probed You may select Low Speed 1 5 Mb s or Full Speed 12 Mb s 8 Press the D and D softkeys to select the channel connected to the USB signal D and D lines The D and D labels for the source channels are automatically set As you press the D or D softkey or rotate the Entry knob the D and D label for the source channel is automatically set and the channel you select is shown in the upper right corner of the display next to USB Keysight Infini
87. scale and run again to see decode Red dots in the decode line indicate that there is data that is not being displayed Scroll or expand the horizontal scale to view the information Unknown bus values undefined or error conditions are drawn in red If there is an error in the synch field SYNC will appear in red If the header exceeds the length specified in the standard THM will appear red Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide CAN LIN Triggering and Serial Decode 23 If the total frame count exceeds the length specified in the standard TFM will appear red LIN 1 3 only For LIN 1 3 a wakeup signal is indicated by WAKE in blue If the wakeup signal is not followed by a valid wakeup delimiter a wakeup error is detected and displayed as WUP in red Interpreting LIN Lister Data 30 572 370 487 Stop 1 11 a Checksum Errors KEYSIGHT 13 31ms 4D 53 4F EE a TECHNOLOGIES 2 759ms 00 F Acquistion 7 798ms 27 AA 90 AO A3 24 18 35ms AQ A er 0 0 39 46ms 13 55 48 DO 51 12 50 02ms D4 53 B a Channels 60 57ms 71 13ms 89 2A 24 E8 28 89 Data 00 80 Lister Menu Display Zoom to S LIN Selection In addition to the standard Time column the LIN Lister contains these columns ID frame ID Data LIN 1 3 only data bytes Checksum LIN 1 3 only Data and Checksum LIN 2 0 only Errors highlighted in red Aliased data is highlighted in p
88. see decode Red dots in the decode line indicate that there is data that is not being displayed Scroll or expand the horizontal scale to view the information Aliased bus values undersampled or indeterminate are drawn in pink Unknown bus values undefined or error conditions are drawn in red Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide I2C SPI Triggering and Serial Decode 25 Interpreting SPI Lister Data KEYSIGHT TECHNOLOGIES 7 477ms_ FF 02 05 FF Aequisition 8 362ms FF 02 06 41 47 10 33ms_ FF 05 FF FF 05 FF gt FE FF FF FF FF FF FF 02 08 49 4C 45 4E 54 16 39ms_ FF FF 4D CF 49 4C 45 4E 54 03 06 00 00 00 00 00 00 00 06 MOS ESR Reh FE be abe ble In addition to the standard Time column the SPI Lister contains these columns Data data bytes MOSI and MISO Aliased data is highlighted in pink When this happens decrease the horizontal time div setting and run again Searching for SPI Data in the Lister The oscilloscope s search capability lets you search for and mark certain types of SPI data in the Lister You can use the Navigate key and controls to navigate through the marked rows 1 With SPI selected as the serial decode mode press Search 2 Inthe Search Menu press the Search softkey then turn the Entry knob to select the serial slot Serial 1 or Serial 2 on which the SPI signals are being decoded 3 Press Search then select from th
89. serial decode requires Option FLEX or the DSOX3FLEX upgrade Setup for FlexRay Signals FlexRay signal setup consists of first connecting the oscilloscope to a differential FlexRay signal using a differential active probe the Keysight N2792A is recommended specifying the signal source threshold voltage trigger level baud rate and bus type To set up the oscilloscope to capture FlexRay signals 1 Press Label to turn on labels 2 Press Serial 3 Press the Serial softkey turn the Entry knob to select the desired serial bus Serial 1 or Serial 2 and press the softkey again to enable decode 4 Press the Mode softkey then select FlexRay mode 5 Press the Signals softkey to open the FlexRay Signals Menu KEYSIGHT TECHNOLOGIES 359 24 FlexRay Triggering and Serial Decode Auto Setup 6 Press Source and select the analog channel that is probing the FlexRay signal 7 Press Threshold then turn the Entry knob to set the threshold voltage level The threshold level should be set below the idle level The threshold voltage level is used in decoding and will become the trigger level when the trigger type is set to the selected serial decode bus 8 Press Baud and select the baud rate of the FlexRay signal being probed 9 Press Bus and select the bus type of the FlexRay signal being probed It is important to specify the correct bus because this setting affects CRC error detection 10 Press Auto Setup to perform t
90. skew 68 To calibrate a probe 68 4 Math Waveforms To display math waveforms 71 To perform transforms or filters on an arithmetic operation 72 To adjust the math waveform scale and offset 73 Units for Math Waveforms 73 Math Operators 74 Add or Subtract 74 Multiply or Divide 75 Math Transforms 76 Differentiate 77 Integrate 78 FFT Measurement 81 Square Root 87 Ax B 88 Square 89 Absolute Value 90 Common Logarithm 90 Natural Logarithm 91 Exponential 91 Base 10 Exponential 91 Math Filters 92 High Pass and Low Pass Filter 92 Math Visualizations 93 Magnify 94 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Measurement Trend 94 Chart Logic Bus Timing 96 Chart Logic Bus State 97 5 Reference Waveforms To save a waveform to a reference waveform location 99 To display a reference waveform 100 To scale and position reference waveforms 101 To adjust reference waveform skew 101 To display reference waveform information 102 To save recall reference waveform files to from a USB storage device 102 6 Digital Channels To connect the digital probes to the device under test 103 Probe cable for digital channels 104 Acquiring waveforms using the digital channels 107 To display digital channels using AutoScale 107 Interpreting the digital waveform display 108 To change the displayed size of the digital channels 109 To switch a single channel on or
91. the oscilloscope in a known operating condition The major default settings are Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 29 1 Getting Started Table2 Default Configuration Settings Horizontal Normal mode 100 us div scale 0 s delay center time reference Vertical Analog Channel 1 on 5 V div scale DC coupling 0 V position 1 MQ impedance Trigger Edge trigger Auto trigger mode 0 V level channel 1 source DC coupling rising edge slope 40 ns holdoff time Display Persistence off 20 grid intensity Other Acquire mode normal Run Stop to Run cursors and measurements off Labels All custom labels that you have created in the Label Library are preserved not erased but all channel labels will be set to their original names In the Save Recall Menu there are also options for restoring the complete factory settings see Recalling Default Setups on page 279 or performing a secure erase see Performing a Secure Erase on page 280 Use Auto Scale Use Auto Scale to automatically configure the oscilloscope to best display the input signals 1 Press Auto Scale You should see a waveform on the oscilloscope s display similar to this 30 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Getting Started 1 200 08 1 34V IGHT EYSIGH TECHNOLOGIES Acquisition N Channels Autoscale Menu Undo Fast Debug Channels Acq Mode Autosc
92. the GPIB option module is installed interfaces The Keysight IO Libraries Suite connectivity software to enables communication over these interfaces You can download the Keysight IO Libraries Suite from www keysight com find iolib 314 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Save Recall Web Interface 21 Information about controlling the oscilloscope through remote commands is contained in the Programmer s Guide which is included on the documentation CD supplied with this oscilloscope You can also access this document on the Keysight web site For more information about connecting to the oscilloscope refer to the Keysight Technologies USB LAN GPIB Interfaces Connectivity Guide For a printable electronic copy of the Connectivity Guide direct your Web browser to www keysight com and search for Connectivity Guide You can save setup files screen images waveform data files or mask files to your PC via the oscilloscope s web interface see Saving Files via the Web Interface on page 315 You can recall setup files reference waveform data files or mask files from your PC via the oscilloscope s web interface see Recalling Files via the Web Interface on page 316 Saving Files via the Web Interface To save setup files screen images waveform data Lister data or mask files to your PC via the oscilloscope s web interface 1 Access the oscilloscope s web interface see Accessing
93. the event occurs while the pre trigger buffer is being filled This may be more likely for example when the horizontal scale knob is set to a slow time div setting such as 500 ms div The trigger indicator at the top right of the display shows whether triggers are occurring In the Auto trigger mode the trigger indicator can show Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Trigger Mode Coupling 11 Auto flashing the trigger condition is not found after the pre trigger buffer has filled and forced triggers and acquisitions are occurring Auto not flashing the trigger condition is found or the pre trigger buffer is being filled In the Normal trigger mode the trigger indicator can show Trig d flashing the trigger condition is not found after the pre trigger buffer has filled and no acquisitions are occurring Trig d not flashing trigger condition is found or pre trigger buffer is being filled When the oscilloscope is not running the trigger indicator area shows Stop When to Use Auto The Auto trigger mode is appropriate when Trigger Mode Checking DC signals or signals with unknown levels or activity When trigger conditions occur often enough that forced triggers are unnecessary When to Use The Normal trigger mode is appropriate when Normal Trigger Mod You only want to acquire specific events specified by the trigger settings ode
94. then turn the Entry knob to select Rise Fall Time 200 08 Auto J KEYSIGHT TECHNOLOGIES Acquisition Channels Level Select High 3 Press the Source softkey and turn the Entry knob to select the input channel source 4 Press the Rising Edge or Falling Edge softkey to toggle between edge types 5 Press the Level Select softkey to select High then turn the Trigger Level knob to adjust the high level 6 Press the Level Select softkey to select Low then turn the Trigger Level knob to adjust the low level You can also push the Trigger Level knob to toggle between High and Low selection 7 Press the Qualifier softkey to toggle between greater than or less than 8 Press the Time softkey and turn the Entry knob to select the time Nth Edge Burst Trigger The Nth Edge Burst trigger lets you trigger on the Nth edge of a burst that occurs after an idle time Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 153 10 Triggers 154 I lt Idle Time gt 1 Tigger OLII ULUY Nth Edge Burst trigger set up consists of selecting the source the slope of the edge the idle time and the number of the edge 1 Press the Trigger key 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select Nth Edge Burst Channels Trigger Menu Trigger Source Nth Edge 3 Press the Source softkey and turn the Entry knob to select the in
95. to caution you about rs inaccura To use the digital volt te RMS measurement results meter 1 Press the Analyze key Press Features then select Digital Voltmeter 2 3 Press Features again to enable the DVM measurements 4 Press the Source softkey and turn the Entry knob to select the analog channel on which digital vo tmeter DVM measurements are made The selected channel does not have to be on displaying a waveform in order for DVM measurem ents to be made 5 Press the Mode softkey and turn the Entry knob to select the digital voltmeter DVM mode AC RMS displays the root mean square value of the acquired data with the DC component removed DC displays th DC RMS displa Frequency disp e DC value of the acquired data ys the root mean square value of the acquired data lays the frequency counter measurement Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Digital Voltmeter 16 6 Press Transparent to toggle between a transparent and shaded background for the DVM display 7 Ifthe selected source channel is not used in oscilloscope triggering press Auto Range to disable or enable automatic adjustment of the DVM channel s vertical scale vertical ground level position and trigger threshold voltage level used for the counter frequency measurement When enabled Auto Range overrides attempted adjustments of the channel s vertical scale and position
96. uniquely reconstructed without aliasing fmax fo 2 Nyquist frequency fy folding frequency Aliasing Aliasing occurs when signals are under sampled fg lt 2f yay Aliasing is the signal distortion caused by low frequencies falsely reconstructed from an insufficient number of sample points Figure 33 Aliasing Oscilloscope Bandwidth and Sample Rate An oscilloscope s bandwidth is typically described as the lowest frequency at which input signal sine waves are attenuated by 3 dB 30 amplitude error Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 183 12 Acquisition Control At the oscilloscope bandwidth sampling theory says the required sample rate is fs 2fgw However the theory assumes there are no frequency components above fmax fay in this case and it requires a system with an ideal brick wall frequency response OdB 3dB c 2 w gt c oO E lt fn fs FE QUENCY ma Figure 34 Theoretical Brick Wall Frequency Response However digital signals have frequency components above the fundamental frequency square waves are made up of sine waves at the fundamental frequency and an infinite number of odd harmonics and typically for 500 MHz bandwidths and below oscilloscopes have a Gaussian frequency response 184 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Acquisition Control 12 OdB 3dB 2 w c o lt i Aliased frequency components i fg4
97. 0 1 500 02 50 1 500 Region Number 4 Number of vertices 5 10 00 1 000 12 50 0 500 15 00 0 500 15 00 1 500 12 50 1 500 Region Number 5 Number of vertices 5 10 00 1 000 12 50 0 500 15 00 0 500 15 00 1 500 12 50 1 500 Region Number 6 Number of vertices 6 05 00 1 000 02 50 0 500 02 50 0 500 05 00 1 000 02 50 1 500 02 50 1 500 Region Number 7 Number of vertices 5 10 00 1 000 12 50 0 500 15 00 0 500 15 00 1 500 12 50 1 500 Region Number 8 Number of vertices 4 12 50 MIN 10 00 1 750 10 00 1 750 12 50 MIN setup MTES ENAB 1 CHAN1 RANG 4 00E 00 OFFS 0 0E 00 COUP DC IMP ONEM DISP 1 BWL 0 INV 0 CHAN1 LAB 1 UNIT VOLT PROB 1 0E 00 PROB SKEW 0 0E 00 STYP SING CHAN2 RANG 16 0E 00 OFFS 1 62400E 00 COUP DC IMP FIFT DISP 0 BWL 0 INV 0 CHAN2 LAB 2 UNIT VOLT PROB 1 0E 00 PROB SKEW 0 0E 00 STYP SING Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 247 15 Mask Testing CHAN3 RANG 40 0E 00 OFFS 0 0E 00 COUP DC IMP ONEM DISP 0 BWL 0 INV 0 CHAN3 LAB 3 UNIT VOLT PROB 1 0E 00 PROB SKEW 0 0E 00 STYP SING CHAN4 RANG 40 0E 00 OFFS 0 0E 00 COUP DC IMP ONEM DISP 0 BWL 0 INV 0 CHAN4 LAB 4 UNIT VOLT PROB 1 0E 00 PROB SKEW 0 0E 00 STYP SING EXT BWL 0 IMP ONEM RANG 5E 00 UNIT VOLT PROB 1 0E 00 PROB STYP SING TIM MODE MAIN REF CENT MAIN RANG 50 00E 09 PO
98. 00 ms or twice the current time window whichever is longer up to 1 second The Counter can measure frequencies up to the bandwidth of the oscilloscope The minimum frequency supported is 1 2 X gate time The hardware counter uses the trigger comparator output Therefore the counted channel s trigger level or threshold for digital channels must be set correctly The Y cursor shows the threshold level used in the measurement Analog and digital channels can be selected as the source Only one Counter measurement can be displayed at a time Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Measurements 14 Width Width is the time from the middle threshold of the rising edge to the middle threshold of the next falling edge The X cursors show the pulse being measured The Y cursor shows the middle threshold point Width Width is the time from the middle threshold of the falling edge to the middle threshold of the next rising edge The X cursors show the pulse being measured The Y cursor shows the middle threshold point Burst Width The Burst Width measurement is the time from the first edge to the last edge on screen l l l l I4 gt Burst width Duty Cycle The duty cycle of a repetitive pulse train is the ratio of the positive pulse width to the period expressed as a percentage The X cursors show the time period being measured The Y cursor shows the middle threshold point Duty cycle awh
99. 000 X Series Oscilloscopes User s Guide 83 4 84 Math Waveforms FFT DC Value on page 85 FFT Aliasing on page 85 FFT Spectral Leakage on page 87 Units for Math Waveforms on page 73 FFT Measurement Hints The number of points acquired for the FFT record can be up to 65 536 and when frequency span is at maximum all points are displayed Once the FFT spectrum is displayed the frequency span and center frequency controls are used much like the controls of a spectrum analyzer to examine the frequency of interest in greater detail Place the desired part of the waveform at the center of the screen and decrease frequency span to increase the display resolution As frequency span is decreased the number of points shown is reduced and the display is magnified While the FFT spectrum is displayed use the Math and Cursors keys to switch between measurement functions and frequency domain controls in FFT Menu FFT Resolution The FFT resolution is the quotient of the sampling rate and the number of FFT points fs N With a fixed number of FFT points up to 65 536 the lower the sampling rate the better the resolution Decreasing the effective sampling rate by selecting a greater time div setting will increase the low frequency resolution of the FFT display and also increase the chance that an alias will be displayed The resolution of the FFT is the effective sample rate divided by the number of points in the FFT
100. 09 digital display interpreting 108 Digital key 37 digital probes 103 114 digital probes impedance 114 digital voltmeter DVM 249 Display key 38 display multiple acquisitions 182 display area 44 display interpreting 43 display persistence 129 display signal detail 127 display softkey labels 44 display status line 44 displayed channels AutoScale 295 distortion problems 81 Divide math function 75 DNS IP 289 dual channel N2820A probe measurements 213 Duty cycle measurement 225 duty cycle measurement trend 95 DVM digital voltmeter 249 DVM license 332 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Dynamic DNS 289 E ECL threshold 110 edge speeds 186 edge then edge triggering 144 edge triggering 142 EDK license 332 EEPROM data read I2C trigger 371 either edge trigger 143 EMBD license 332 energy of a pulse 78 English user interface and Quick Help 45 Entry knob 35 Entry knob push to select 35 erase secure 280 event table 122 expand about 63 293 expand about center 294 expand about ground 293 exponential fall waveform generator output 255 exponential math function 91 exponential rise waveform generator output 255 exporting waveform 269 EXT TRIG IN as Z axis input 53 EXT TRIG IN connector 43 external memory device 39 external trigger 178 external trigger input impedance 179 external trigger pro
101. 0H2 protect button 3 GPIB 4 TRIG OUT connector 3 Module slot 2 Kensington lock hole oy 1 Power cord connector 1 Power cord connector Attach the power cord here 2 Kensington lock hole This is where you can attach a Kensington lock for securing the instrument 3 Module slot A DSOXLAN LAN VGA module may be ordered and installed separately LAN port lets you communicate with the oscilloscope and use the Remote Front Panel feature using the LAN port See Chapter 21 Web Interface starting on page 307 and Accessing the Web Interface on page 308 VGA video output lets you connect an external monitor or projector to provide a larger display or to provide a display at a viewing position away from the oscilloscope The oscilloscope s built in display remains on even when an external display is connected The video output connector is always active For optimal video quality and performance we recommend you use a shielded video cable with ferrite cores A DSOXGPIB GPIB module may be ordered and installed separately 4 TRIG OUT connector Trigger output BNC connector See Setting the Rear Panel TRIG OUT Source on page 296 5 Calibration protect button See To perform user calibration on page 297 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Getting Started 1
102. 1 D ai Figure19 Example of Chart Logic Bus State Use the Clock softkey to select the clock signal Use the Slope softkey to select the edge of the clock signal to be used Use the More Chart softkey to open a submenu for specifying the analog value equivalent of each bus value increment the analog equivalent of a zero bus value and the type of values the charted bus data represents volts amps etc tit State Chart Bus 1 Seale 50 0mV Units Code D Offset Units 1 0mY B4mv Volts Use the Units Code softkey to specify the analog value equivalent of each increment in the bus data value Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 97 4 Math Waveforms Use the 0 Offset softkey to specify the analog value equivalent of a bus data value of zero Use the Units softkey to specify the type of values the bus data represents volts amps etc See Also Chart Logic Bus Timing on page 96 98 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 5 Reference Waveforms To save a waveform to a reference waveform location 99 To display a reference waveform 100 To scale and position reference waveforms 101 To adjust reference waveform skew 101 To display reference waveform information 102 To save recall reference waveform files to from a USB storage device 102 Analog channel or math waveforms can be sa
103. 1 0 or X don t care Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 399 27 MIL STD 1553 ARINC 429 Triggering and Serial Decode Parity Error triggers if the odd parity bit is incorrect for the data in the word Sync Error triggers if an invalid Sync pulse is found Manchester Error triggers if a Manchester encoding error is detected For MIL STD 1553 decode information see MIL STD 1553 Serial Decode on page 400 MIL STD 1553 Serial Decode 400 See Also To set up the oscilloscope to capture MIL STD 1553 signals see Setup for MIL STD 1553 Signals on page 397 For MIL STD 1553 triggering setup see MIL STD 1553 Triggering on page 399 To set up MIL STD 1553 serial decode 1 Press Serial to display the Serial Decode Menu Signals MIL 1553 w Base Hex Auto Setup 2 Use the Base softkey to select between hexadecimal and binary display of the decoded data The base setting is used for the display of the remote terminal address and the data in both the decode line and in the Lister 3 If the decode line does not appear on the display press the Serial key to turn it on 4 Ifthe oscilloscope is stopped press the Run Stop key to acquire and decode data You can use the horizontal Zoom window for easier navigation of the decoded data Interpreting MIL STD 1553 Decode on page 401 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide MIL STD
104. 1 Press the Digital key to toggle the display of digital channels The Digital Channel Menu is displayed above the softkeys If you want to switch the digital channels off and the Digital Channel Menu is not already displayed you must push the Digital key twice to switch the digital channels off The first push displays the Digital Channel Menu and the second push switches the channels off Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 109 6 To switch groups of channels on or off 1 Press the Digital key on the front panel if the Digital Channel Menu is not already displayed 2 Press the Turn off or Turn on softkey for the D15 D8 group or the D7 DO group Each time you press the softkey the softkey s mode toggles between Turn on and Turn off To change the logic threshold for digital channels 110 1 Press the Digital key so that the Digital Channel Menu is displayed 2 Press the Thresholds softkey 3 Press the D15 D8 or D7 DO softkey then select a logic family preset or select User to define your own threshold Logic family Threshold Voltage TTL 1 4 V CMOS 2 5 V ECL 1 3 V User Variable from 8 V to 8 V The threshold you set applies to all channels within the selected D15 D8 or D7 DO group Each of the two channel groups can be set to a different threshold if desired Values greater than the set threshold are high 1 and values less than the set thresh
105. 186 An oscilloscope s rise time is not the fastest edge speed that the oscilloscope can accurately measure It is the fastest edge speed the oscilloscope can possibly produce Oscilloscope Bandwidth Required See Also The oscilloscope bandwidth required to accurately measure a signal is primarily determined by the signal s rise time not the signal s frequency You can use these steps to calculate the oscilloscope bandwidth required 1 Determine the fastest edge speeds You can usually obtain rise time information from published specifications for devices used in your designs 2 Compute the maximum practical frequency component From Dr Howard W Johnson s book High Speed Digital Design A Handbook of Black Magic all fast edges have an infinite spectrum of frequency components However there is an inflection or knee in the frequency spectrum of fast edges where frequency components higher than fknee are insignificant in determining the shape of the signal fknee 0 5 signal rise time based on 10 90 thresholds fknee 0 4 signal rise time based on 20 80 thresholds 3 Use a multiplication factor for the required accuracy to determine the oscilloscope bandwidth required Required accuracy Oscilloscope band width required 20 few 1 0 X fknee 10 few 1 3 X fknee 3 few 1 9 X fknee Choosing an Oscilloscope with the Right Bandwidth for your Application Keysight Application
106. 24A DSO X 3034A DSO X 3054A DSO X 3104A The Keysight In Keysight InfiniiVision 3000 X Series Osci 2 and 4 ch finiiVision 3000 X Series oscilloscopes deliver these features 100 MHz 200 MHz 350 MHz 500 MHz and 1 GHz bandwidth models loscopes User s Guide annel digital storage oscilloscope DSO models 2 16 channel and 4 16 channel mixed signal oscilloscope MSO models An MSO lets you debug your mixed signal designs using analog signals and tightly correlated digital signals simultaneously The 16 digital channels have a 1 GSa s sample rate 1 25 GSa s for the 1 GHz models with a 50 MHz toggle rate 8 5 inch WVGA display Interleaved 2 Mpts or non interleaved 1 Mpts MegaZoom IV memory for the fastest waveform update rates uncompromised Upgradeable to 4 Mpts 2 Mpts All knobs are pushable for making quick selections Trigger types edge edge then edge pulse width pattern OR rise fall time Nth edge burst runt setup amp hold video and USB Serial decode trigger options for CAN LIN FlexRay 2C SPI S UART RS232 and MIL STD 1553 ARINC 429 Lister for serial decode Math waveforms add subtract multiply FFT d dt integrate and square root With the DSOX3ADVMATH option you get these additional math waveforms Ax B square absolute value common logarithm natural logarithm exponential base 10 exponential low pass filter high pass filter magnify measurement tr
107. 6000 and 7000 Series InfiniiVision Oscilloscope Probes and Accessories Data Sheet 5968 8153EN Loading Licenses and Displaying License Information License files are loaded from a USB storage device using the File Explorer see File Explorer on page 291 License information is displayed with other oscilloscope information see To display oscilloscope information on page 300 For more information about the licenses and other oscilloscope options available see Licensed Options Available on page 331 Other Options Available on page 333 Upgrading to an MSO on page 333 Licensed Options Available The following licensed options can be easily installed without returning the oscilloscope to a Service Center See data sheets for details Table 10 Licensed Options Available License Description After purchase model number notes ADVMATH Advanced Math Measurements Order DSOX3ADVMATH AERO MIL STD 1553 and ARINC 429 Serial Triggering and Analysis Order DSOX3AERO AUDIO Audio Serial Triggering and Analysis 12S Order DSOX3AUDIO AUTO Automotive Serial Triggering and Analysis CAN LIN Order DSOX3AUTO COMP Computer Serial Triggering and Analysis Order DSOX3COMP RS232 422 485 UART Provides trigger and decode capability for many UART Universal Asynchronous Receiver Transmitter protocols including RS232 Recommended Standard 232 Keysight InfiniiVision 3000 X Series Oscilloscopes Us
108. 65536 points Still once the of acquired points gt 65536 some form of decimation is required The decimator used to produce a CSV record is configured to provide a best estimate of all the samples that each point in the record represents Therefore the min and max values may not appear in the CSV file Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 341 22 Reference Acknowledgements RealVNC HDF5 CUPS mDNSResponder 342 RealVNC is licensed under the GNU General Public License Copyright C 2002 2005 RealVNC Ltd All Rights Reserved This is free software you can redistribute it and or modify it under the terms of the GNU General Public License as published by the Free Software Foundation either version 2 of the License or at your option any later version This software is distributed in the hope that it will be useful but WITHOUT ANY WARRANTY without even the implied warranty of MERCHANTABILITY or FITNESS FOR A PARTICULAR PURPOSE See the GNU General Public License for more details This license is located on the Keysight InfiniiVision Oscilloscopes Documentation CD ROM RealVNC source code can be obtained from RealVNC or by contacting Keysight Keysight will charge for the cost of physically performing the source distribution Reference Waveform files use HDF5 HDF5 was developed by The HDF Group and by the National Center for Supercomputing Applications at the University of Illinois a
109. 7 With the DSOX38ADVMATH advanced math measurements license these additional transforms are available Ax B on page 88 Square on page 89 76 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Math Waveforms 4 Absolute Value on page 90 Common Logarithm on page 90 Natural Logarithm on page 91 Exponential on page 91 Base 10 Exponential on page 91 Differentiate d dt differentiate calculates the discrete time derivative of the selected source You can use differentiate to measure the instantaneous slope of a waveform For example the slew rate of an operational amplifier may be measured using the differentiate function Because differentiation is very sensitive to noise it is helpful to set acquisition mode to Averaging see Selecting the Acquisition Mode on page 187 d dt plots the derivative of the selected source using the average slope estimate at 4 points formula The equation is _ Vita Ying 2Y Yi di SAT Where d differential waveform y channel 1 2 3 or 4 or g t internal arithmetic operation data points i data point index At point to point time difference Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 4 Math Waveforms KEY TECHN Acquisition Channels orm Math Menu f t d dt Ch1 Function Operator Source 1 tit didt Figure 7 Example of Differentiate Function See Also To perform transforms or
110. A indicates No Ack high Decoded text is truncated at the end of the associated frame when there is insufficient space within frame boundaries Pink vertical bars indicate you need to expand the horizontal scale and run again to see decode Red dots in the decode line indicate that more data can be displayed Scroll or expand the horizontal scale to view the data Aliased bus values undersampled or indeterminate are drawn in pink Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 375 25 2C SPI Triggering and Serial Decode Unknown bus values undefined or error conditions are drawn in red Interpreting I2C Lister Data 101 257 Stop 2c 2 254 Missing Ack KEYSIGHT 896 1us a TECHNOLOGIES 735 1us 574 1us 06 gt 41 47 49 4C 45 4E 54 10 4D 53 4F 3 443ms 3 740ms 3 901ms Data 41 47 49 4C 45 4E 54 In addition to the standard Time column the I2C Lister contains these columns Restart indicated with an X Address colored blue for writes yellow for reads Data data bytes Missing Ack indicated by an X highlighted in red if an error Aliased data is highlighted in pink When this happens decrease the horizontal time div setting and run again Searching for I2C Data in the Lister The oscilloscope s search capability lets you search for and mark certain types of 2C data in the Lister You can use the Navigate key and controls
111. CSV files minimum and maximum values 341 current probes 329 cursor measurements 199 cursor units 201 Cursors key 37 Cursors knob 37 cursors binary 200 cursors hex 201 cursors manual 200 cursors track waveform 200 D D 37 111 d dt math function 77 damage shipping 23 data sheet 323 DC channel coupling 64 DC offset correction for integrate waveform 79 DC RMS Full Screen measurement 220 DC RMS N Cycles measurement 220 DC signals checking 175 DC waveform generator output 255 dead time re arm 196 decibels FFT vertical units 82 decimating samples 187 decimation for measurement record 341 decimation for screen 341 default configuration 29 default label library 137 default setup 29 280 Default Setup key 36 defaults waveform generator 267 delay knob 49 Delay measurement 212 226 delay time indicator 56 delayed sweep 54 delete character 277 delete file 291 Demo 1 terminal 39 Demo 2 terminal 39 depth AM modulation 264 deviation FM modulation 266 DHCP 289 290 differential probes 328 differentiate math function 77 Digit softkey 150 digital bus mode 111 digital channel controls 37 digital channel inputs 39 digital channel menu 109 digital channels 109 digital channels AutoScale 107 digital channels enabling 333 digital channels logic threshold 110 digital channels probing 114 digital channels size 1
112. Cursor Examples 203 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 14 Measurements To make automatic measurements 210 Measurements Summary 211 Snapshot All 214 Voltage Measurements 215 Peak Peak 216 Maximum 216 Minimum 216 Amplitude 216 Top 216 Base 217 Overshoot 217 Preshoot 219 Average 219 DCRMS 220 ACRMS 220 Ratio 222 Time Measurements 222 Period 223 Frequency 223 Counter 224 Width 225 Width 225 Burst Width 225 Duty Cycle 225 Rise Time 226 FallTime 226 Delay 226 Phase 227 XatMinY 228 XatMaxY 229 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Count Measurements 229 Positive Pulse Count 229 Negative Pulse Count 230 Rising Edge Count 230 Falling Edges Count 230 Mixed Measurements 230 Area 230 Measurement Thresholds 231 Measurement Window with Zoom Display 233 Measurement Statistics 233 15 Mask Testing To create a mask from a golden waveform Automask 237 Mask Test Setup Options 239 Mask Statistics 242 To manually modify a mask file 243 Building a Mask File 246 How is mask testing done 248 16 Digital Voltmeter 17 Waveform Generator To select generated waveform types and settings 253 To edit arbitrary waveforms 257 Creating New Arbitrary Waveforms 258 Editing Existing Arbitrary Waveforms 259 Capturing Other Waveforms to the Arbitrary Waveform 260
113. DTV 1080p 60 Progressive Tri level HDTV 1080i 50 Interlaced Tri level HDTV 1080i 60 Interlaced Tri level 160 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Triggers 10 The Generic selection lets you trigger on custom bi level and tri level sync video standards See To set up Generic video triggers on page 162 N Press the Auto Setup softkey to automatically set up the oscilloscope for the selected Source and Standard Source channel vertical scaling is set to 140 mV div Source channel offset is set to 245 mV Source channel is turned on Trigger type is set to Video Video trigger mode is set to All Lines but left unchanged if Standard is Generic Display Grid type is set to IRE when Standard is NTSC or mV see To select the grid type on page 130 Horizontal time division is set to 10 us div for NTSC PAL SECAM standards or 4 us div for EDTV or HDTV standards unchanged for Generic Horizontal delay is set so that trigger is at first horizontal division from the left unchanged for Generic You can also press Analyze gt Features and then select Video to quickly access the video triggering automatic set up and display options 8 Press the Mode softkey to select the portion of the video signal that you would like to trigger on The Video trigger modes available are Field1 and Field2 Trigger on the rising edge of the first serration pulse of field 1 or field 2 interlace
114. FM broadcast stations or noise from fast system clocks from the trigger path 1 Press the Mode Coupling key 2 Inthe Trigger Mode and Coupling Menu press the HF Reject softkey to enable or disable To set the trigger holdoff Trigger holdoff sets the amount of time the oscilloscope waits after a trigger before re arming the trigger circuitry Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 177 11 Trigger Mode Coupling Trigger Holdoff Operating Hints Use the holdoff to trigger on repetitive waveforms that have multiple edges or other events between waveform repetitions You can also use holdoff to trigger on the first edge of a burst when you know the minimum time between bursts For example to get a stable trigger on the repetitive pulse burst shown below set the holdoff time to be gt 200 ns but lt 600 ns Holdoff Oscilloscope triggers here H gt gt gt gt gt a gt gt 200 ns 600 ns To set the trigger holdoff 1 Press the Mode Coupling key 2 Inthe Trigger Mode and Coupling Menu press the Holdoff softkey then turn the Entry knob to increase or decrease the trigger holdoff time The correct holdoff setting is typically slightly less than one repetition of the waveform Set the holdoff to this time to generate a unique trigger point for a repetitive waveform Changing the time base settings does not affect the trigger holdoff t
115. INC 429 words errors counter 410 ARINC 429 decode signal speed 404 ARINC 429 decode signal type 404 ARINC 429 decode word format 407 ASCII file format 270 attenuation probe 67 attenuation probe external trigger 179 attenuators 68 AUDIO license 331 Auto Increment 277 AUTO Option 331 Auto Scale key 36 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Auto Setup FFT 82 Auto trigger mode 174 Auto trigger indicator 175 AutolP 289 290 automatic measurements 209 211 automatic setup 107 AutoProbe interface 39 64 AutoScale preferences 295 AutoScale digital channels 107 AutoScale undo 31 Average Full Screen measurement 219 Average N Cycles measurement 219 average measurement trend 95 averaging acquire mode 187 191 Ax B math function 88 B Back Up key 34 bandwidth 300 bandwidth limit 65 bandwidth required oscilloscope 186 bandwidth oscilloscope 183 base 10 exponential math function 91 Base measurement 217 BIN file format 270 binary data bin 333 binary data file examples 337 binary data in MATLAB 334 binary data example program for reading 337 bits SPI trigger 382 Blackman Harris FFT window 82 blanking 53 BMP file format 270 brick wall frequency response 184 brightness of waveforms 34 Browser Web Control 309 310 311 S12 313 Browser Based Remote Front Panel 312 423 Index built in help 44 burst width measurement
116. Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide KEYSIGHT TECHNOLOGIES Notices Keysight Technologies Inc 2005 2013 No part of this manual may be reproduced in any form or by any means including elec tronic storage and retrieval or translation into a foreign language without prior agreement and written consent from Keysight Technolo gies Inc as governed by United States and international copyright laws Manual Part Number 75019 97073 Edition Eighth edition April 2013 Printed in Malaysia Published by Keysight Technologies Inc 1900 Garden of the Gods Road Colorado Springs CO 80907 USA Print History 75019 97000 January 2011 75019 97013 February 2011 75019 97014 June 2011 75019 97027 October 2011 75019 97040 February 2012 75019 97051 March 2012 75019 97062 July 2012 75019 97073 April 2013 Trademarks Java is a U S trademark of Sun Microsys tems Inc Sun Sun Microsystems and the Sun Logo are trademarks or registered trademarks of Sun Microsystems Inc in the U S and other countries Warranty The material contained in this docu ment is provided as is and is subject to being changed without notice in future editions Further to the maxi mum extent permitted by applicable law Keysight disclaims all warranties either express or implied with regard to this manual and any information con tained herein including but not limited to the impli
117. Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter The Digital Voltmeter DVM analysis feature provides 3 digit voltage and 5 digit frequency measurements using any analog channel DVM measurements are asynchronous from the oscilloscope s acquisition system and are always acquiring To enable the digital voltmeter analysis feature order Option DVM at time of oscilloscope purchase or order DSOXDVM as a stand alone item after oscilloscope purchase The DVM display is a seven segment readout like you would see on a digital voltmeter It shows the selected mode as well as the units Units are selected using the Units softkey in the channel s Probe Menu After pressing the Analyze key the DVM display also appears in the graticule along with a scale and the frequency counter value The DVM scale is determined by the channel s vertical scale and reference level The scale s blue triangle pointer shows the most recent measurement The white bar above that shows the measurement extrema over the last 3 seconds KEYSIGHT TECHNOLOGIES 249 16 Digital Voltmeter 250 Source DYM The DVM makes accu 100 057 Auto 6259 TECHNOLOGIES DVM Channels Frequency Mode Transparent Auto Range AC RMS rate RMS measurements when the signal frequency is between 20 Hz and 100 kHz When the signal frequency is outside this range lt BW Limit or gt BW Limit appears in the DVM display
118. Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 65 3 Vertical Controls 2 Inthe Channel Menu press the Invert softkey to invert the selected channel When Invert is selected the voltage values of the displayed waveform are inverted Invert affects how a channel is displayed However when using basic triggers the oscilloscope attempts to maintain the same trigger point by changing trigger settings Inverting a channel also changes the result of any math function selected in the Waveform Math Menu or any measurement Setting Analog Channel Probe Options 66 See Also 1 Press the probe s associated channel key 2 Inthe Channel Menu press the Probe softkey to display the Channel Probe Menu This menu lets you select additional probe parameters such as attenuation factor and units of measurement for the connected probe Channel 1 Probe Menu 1 00 1 U Probe 1 00 1 The Channel Probe Menu changes depending on the type of probe connected For passive probes such as the N2862A B N2863A B N2889A N2890A 10073C 10074C or 1165A probes the Probe Check softkey appears it guides you through the process of compensating probes For some active probes such as InfiniiMax probes the oscilloscope can accurately calibrate its analog channels for the probe When you connect a probe that can be calibrated the Calibrate Probe softkey appears and the probe attenuation softkey may change See To calibrate
119. LIN Triggering and Serial Decode starting on page 343 Chapter 24 FlexRay Triggering and Serial Decode starting on page 359 Chapter 25 I2C SPI Triggering and Serial Decode starting on page 369 Chapter 26 I2S Triggering and Serial Decode starting on page 387 Chapter 27 MIL STD 1553 ARINC 429 Triggering and Serial Decode starting on page 397 Chapter 28 UART RS232 Triggering and Serial Decode starting on page 413 Abbreviated instructions for pressing a series of keys and softkeys Instructions for pressing a series of keys are written in an abbreviated manner Instructions for pressing Key1 then pressing Softkey2 then pressing Softkey3 are abbreviated as follows Press Key1 gt Softkey2 gt Softkey3 The keys may be a front panel Key or a Softkey Softkeys are the six keys located directly below the oscilloscope display Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide InfiniiVision 3000 X Series Oscilloscopes At a Glance 3 In This Guide 5 Inspect the Package Contents 23 Install the Optional LAN VGA or GPIB Module 26 Tilt the Oscilloscope for Easy Viewing 26 Power On the Oscilloscope 27 Connect Probes to the Oscilloscope 28 A Maximum input voltage at analog inputs 28 Do not float the oscilloscope chassis 29 Input a Waveform 29 Recall the Default Oscilloscope Setup 29 Use Auto Scale 30 Compensate Passive Probes 32 Lea
120. Lister Data on page 124 For more information on using the Navigate key and controls see Navigating the Time Base on page 58 Setup for SPI Signals Serial Peripheral Interface SPI signals setup consists of connecting the oscilloscope to a clock MOSI data MISO data and framing signal then setting the threshold voltage level for each input channel and finally specifying any other signal parameters To set up the oscilloscope to capture SPI signals use the Signals softkey which appears in the Serial Decode Menu 1 Press Label to turn on labels 2 Press Serial 3 Press the Serial softkey turn the Entry knob to select the desired slot Serial 1 or Serial 2 and press the softkey again to enable decode 4 Press the Mode softkey then select SPI trigger type Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 377 25 2C SPI Triggering and Serial Decode 5 Press the Signals softkey to open the SPI Signals Menu SPI Signals Menu Clock MOSI w w SPI Clock Menu Display Info Clock Threshold Display Info TTL 1 4V In the SPI Clock Menu a Press the Clock softkey then turn the Entry knob to select the channel connected to the S PI serial clock line The CLK label for the source channel is automatically set b Press the Threshold signal threshold vo softkey then turn the Entry knob to select the clock tage level The threshold voltage level is used in decoding and it wi
121. Modulation AM In the Waveform Generator Modulation Menu under Wave Gen gt Settings gt Modulation 1 Press the Type softkey and turn the Entry knob to select Amplitude Modulation AM 2 Press the Waveform softkey and turn the Entry knob to select the shape of the modulating signal Sine Square Ramp Sine Card inal Exponential Rise Exponential Fall When the Ramp shape is selected a Symmetry softkey appears so that you can specify the amount of time per cycle that the ramp waveform is rising 3 Press the AM Freq softkey and turn the Entry knob to specify the frequency of the modulating signal 4 Press the AM Depth softkey and turn the Entry knob to specify the amount of amplitude modulation AM Depth refers to the portion of the amplitude range that will be used by the modulation For example a depth setting of 80 causes the output amplitude to vary from 10 to 90 90 10 80 of the original amplitude as the modulating signal goes from its minimum to maximum amplitude The following screen shows an AM modulation of a 100 kHz sine wave carrier signal Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Waveform Generator 17 4 28V KEYSIGHT TECHNOLOGIE Acquisition Channels or Modulation Menu Type Waveform AM Depth Symmetry AM v Ramp 50 100 To set up Frequency Modulation FM In the Waveform Generator Modulation Menu under Wave Gen gt Settings gt Modula
122. NTSC video standard To trigger on a specific line of video on page 163 To trigger on all sync pulses on page 164 To trigger on a specific field of the video signal on page 165 To trigger on all fields of the video signal on page 166 To trigger on odd or even fields on page 167 To set up Generic video triggers When Generic available with the DSOX3VID extended video triggering license is selected as the Video trigger Standard you can trigger on custom bi level and tri level sync video standards The Video Trigger Menu changes as follows 162 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Triggers 10 Video Trigger Menu andard Edge Horiz Syne Horiz Syne Field Holdoff 1 126 0us 0 Oflds 1 Press the Time gt softkey then turn the Entry knob to set the time to greater than the sync pulse width so that the oscilloscope synchronizes to the vertical sync 2 Press the Edge softkey then turn the Entry knob to select the Nth edge after after vertical sync to trigger on 3 To enable or disable the horizontal sync control press the first Horiz Sync softkey For interleaved video enabling the Horiz Sync control and setting the Horiz Syne adjustment to the sync time of the probed video signal allows the Edge function to count only lines and not double count during equalization Additionally the Field Holdoff can be adjusted so that the oscilloscope triggers once per frame Similarl
123. No trigger Trigger value Armed value g 4h bree Decreasing value similar to the description above except the trigger occurs on a descreasing data word value and the Armed value is the value to which the data must rise in order to re arm the trigger 6 Press the Base softkey and select a number base for entering data values Binary 2 s complement When Binary is selected the Bits softkey appears This softkey opens the 12S Bits Menu for entering data values When the trigger qualifier requires a pair of values as with In Range Out of Range Increasing value or Decreasing value the first softkey in the 12S Bits Menu lets you select which value of the pair In the I2S Bits Menu press the Bit softkey and rotate the Entry knob to select each bit then use the 01 X softkey to set each bit value to zero one or don t care You can use the Set all Bits softkey to set all bits to the value chosen on the 0 1 X softkey Signed decimal When Decimal is selected the softkey s to the right let you enter decimal values with the Entry knob These softkeys can be Data lt gt or Threshold depending on the selected trigger qualifier If the setup does not produce a stable trigger the I2S signal may be slow enough that the oscilloscope is AutoTriggering Press the Mode Coupling key then press the Mode softkey to set the trigger mode from Auto to Normal To display I2S serial decode see I2S Serial Decod
124. Note 1588 http literature cdn keysight com litweb pdf 5989 5733EN pdf Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Acquisition Control 12 Memory Depth and Sample Rate The number of points of oscilloscope memory is fixed and there is a maximum sample rate associated with oscilloscope s analog to digital converter however the actual sample rate is determined by the time of the acquisition which is set according to the oscilloscope s horizontal time div scale sample rate number of samples time of acquisition For example when storing 50 us of data in 50 000 points of memory the actual sample rate is 1 GSa s Likewise when storing 50 ms of data in 50 000 points of memory the actual sample rate is 1 MSa s The actual sample rate is displayed in the right side information area The oscilloscope achieves the actual sample rate by throwing away decimating unneeded samples Selecting the Acquisition Mode When selecting the oscilloscope acquisition mode keep in mind that samples are normally decimated at slower time div settings At slower time div settings the effective sample rate drops and the effective sample period increases because the acquisition time increases and the oscilloscope s digitizer is sampling faster than is required to fill memory For example suppose an oscilloscope s digitizer has a sample period of 1 ns maximum sample rate of 1 GSa s anda 1 M m
125. Passive probe 10 1 1 1 350 MHz 1 2 m N2890A Passive probe 10 1 500 MHz 1 2 m Single Ended Active Probes The following single ended active probes can be used with the InfiniiVision 3000 X Series oscilloscopes Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 327 22 Reference Table 6 Active Probes Model Description Quantity Supported 1130A 1 5 GHz InfiniiMax amplifier requires one or more 2 InfiniiMax probe heads E2675A E2668A E2669A 1131A InfiniiMax 3 5 GHz probe 2 1132A InfiniiMax 5 GHz probe 2 1134A InfiniiMax 7 GHz probe 2 1156A Active probe 1 5 GHz 4 1157A Active probe 2 5 GHz 4 1158A Active probe 4 GHz 4 N2744A T2A probe interface adapter Unknown depends on probes connected N2795A Active probe 1 GHz with AutoProbe interface 2 N2796A Active probe 2 GHz with AutoProbe interface 2 1See AutoProbe Interface on page 326 Differential Probes The following differential probes can be used with the InfiniiVision 3000 X Series oscilloscopes Table 7 Differential Probes Model Description Quantity Supported 1141A Active differential probe 200 MHz 200 VDC peak AC max requires 1142A power supply 1144A Active probe 800 MHz requires 1142A power supply 1145A Active probe 750 MHz 2 ch requires 1142A power supply 328 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Reference 22
126. Pressing the FFT Auto Setup softkey will automatically rescale the waveform and span and center will again automatically track the horizontal scale setting 82 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Math Waveforms 4 3 To make cursor measurements press the Cursors key and set the Source softkey to Math f t Use the X1 and X2 cursors to measure frequency values and difference between two frequency values AX Use the Y1 and Y2 cursors to measure amplitude in dB and difference in amplitude AY 4 To make other measurements press the Meas key and set the Source softkey to Math f t ia You can make peak to peak maximum minimum and average dB measurements on the FFT waveform You can also find the frequency value at the first occurrence of the waveform maximum by using the X at Max Y measurement The following FFT spectrum was obtained by connecting a 4 V 75 kHz square wave to channel 1 Set the horizontal scale to 50 us div vertical sensitivity to 1 V div Units div to 20 dBV Offset to 60 0 dBV Center frequency to 250 kHz frequency Span to 500 kHz and window to Hanning Channels UI PEPR tit FFT Ch1 Scale 20dB ffset 60 0004BV FFT Resolution 763Hz Window Vertical Units Hanning Decibels Auto Setup See Also To perform transforms or filters on an arithmetic operation on page 72 FFT Measurement Hints on page 84 FFT Units on page 85 Keysight InfiniiVision 3
127. S 0 0E 00 TRIG MODE EDGE SWE AUTO NREJ 0 HFR 0 HOLD 60E 09 TRIG EDGE SOUR CHAN1 LEV 75 00E 03 SLOP POS REJ OFF COUP DC ACQ MODE RTIM TYPE NORM COMP 100 COUNT 8 SEGM COUN 2 DISP LAB 0 CONN 1 PERS MIN SOUR PMEM1 HARD APR AREA SCR FACT 0 FFE 0 INKS 1 PAL NONE LAY PORT SAVE FIL mask_0 SAVE IMAG AREA GRAT FACT 0 FORM NONE INKS 0 PAL COL SAVE WAV FORM NONE MTES SOUR CHAN1 ENAB 1 LOCK 1 MTES AMAS SOUR CHAN1 UNIT DIV XDEL 3 00000000E 001 YDEL 2 00000000E 00 1 MTES SCAL BIND 0 X1 0 0E 00 XDEL 1 0000E 09 Y1 0 0E 00 Y2 1 00000E 0 0 MTES RMOD FOR RMOD TIME 1E 00 WAV 1000 SIGM 6 0E 00 MTES RMOD FACT STOP 0 PRIN 0 SAVE 0 end_setup How is mask testing done 248 InfiniiVision oscilloscopes start mask testing by creating a database that is 200 x 640 for the waveform viewing area Each location in the array is designated as either a violation or a pass area Each time a data point from a waveform occurs in a violation area a failure is logged If Test All was selected every active analog channel is tested against the mask database for each acquisition Over 2 billion failures can be logged per channel The number of acquisitions tested is also logged and displayed as of Tests The mask file allows greater resolution than the 200 X 640 database Some quantization of data occurs to reduce the mask file data for display on screen Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide
128. S232 Signals Menu Threshold x Threshold 12V 1 92 KEYSIGHT TECHNOLOGIES 413 28 UART RS232 Triggering and Serial Decode 6 For both the Rx and Tx signals a Connect an oscilloscope channel to the signal in the device under test b Press the Rx or Tx softkey then turn the Entry knob to select the channel for the signal c Press the corresponding Threshold softkey then turn the Entry knob to select the signal threshold voltage level The threshold voltage level is used in decoding and it will become the trigger level when the trigger type is set to the selected serial decode slot The RX and TX labels for the source channels are automatically set 7 Press the O Back Up key to return to the Serial Decode Menu 8 Press the Bus Config softkey to open the UART RS232 Bus Configuration Menu 3us Configuration Menu Parity Baud Rate Polarity Bit Order LSB Odd E S Idle low Set the following parameters a Bits Set the number of bits in the UART RS232 words to match your device under test selectable from 5 9 bits b Parity Choose odd even or none based on your device under test c Baud Press the Baud Rate softkey then press the Baud softkey and select a baud rate to match the signal in your device under test If the desired baud rate is not listed select User Defined on the Baud softkey then select the desired baud rate using the User Baud softkey You can set the UART baud rate from 1 2 kb s
129. Screens Data starting on page 269 Setting Oscilloscope Preferences The User Preferences Menu under Utility gt Options gt Preferences lets you specify oscilloscope preferences To choose expand about center or ground on page 293 To disable enable transparent backgrounds on page 294 To load the default label library on page 294 To set up the screen saver on page 294 To set AutoScale preferences on page 295 To choose expand about center or ground When you change a channel s volts division setting the waveform display can be set to expand or compress about the signal ground level or the center of the display To set the waveform expansion reference point 1 Press Utility gt Options gt Preferences gt Expand and select Ground The displayed waveform will expand about the position of the channel s ground This is the default setting The ground level of the signal is identified by the position of the ground level icon at the far left side of the display The ground level will not move when you adjust the vertical sensitivity volts division control Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 293 20 Utility Settings 294 If the ground level is off screen the waveform will expand about the top or bottom edge of the screen based on where the ground is off screen Center The displayed waveform will expand about the center of the display To disable ena
130. Series Oscilloscopes User s Guide 4 Enter your desired password and click Apply Changes 321 21 Web Interface Support Pro Ww KEYSIGHT PRT another web TECHNOLOGIES _ from Keysight Modify Network Configuration Undo Changes LANReset Apply Changes Parameter Configured Value Edit Configuration IP Settings may be configured using the following Automatic ON OFF ON IP Settings to use in non automatic mode 49 aa5 IP Address 141 121 237 192 Subnet Mask 255 255 248 0 Default Gateway 141 121 232 1 Name service settings a mx2024a 10029 gt Help with ad this Page Hostname a mx2024a 10029 Requires reboot to take effect DNS Server 156 140 24 15 156 140 24 15 Multicast DNS ON OFF ON Other settings Agilent MSOX2024A InfiniiVision r aiaia aN Description US50210029 Agilent MSOX2024A InfiniiVision L Enter password GPIB Address 7 7 Set to blank for factory default value When accessing the password protected oscilloscope the user name is the IP address of the oscilloscope Toresetthe Do one of these things to reset the password password Using the keys on the front panel of the oscilloscope press Utility gt 1 0 gt LAN Reset Using the web browser select the Configure Network tab select Modify Configuration erase the Password and select Apply Changes 322 Keysig
131. Table 7 Differential Probes continued Model Description Quantity Supported N2772A Active differential probe 20 MHz 1 2 kVDC peak AC max requires N2773A power supply N2790A High voltage differential probe 50 1 or 500 1 switchable 100 MHz with 4 AutoProbe interface N2791A High voltage differential probe 25 MHz 700 V 1 MOhm termination 10 1 or 100 1 switchable N2792A Differential probe 200 MHz 10 1 50 Ohm termination N2793A Differential probe 800 MHz 10 1 15 V 50 Ohm termination N2891A 70 MHz 7 kV high voltage differential probe 1See AutoProbe Interface on page 326 Current Probes The following current probes can be used with the InfiniiVision 3000 X Series oscilloscopes Table 8 Current Probes Model Description Quantity Supported 1146A Current probe 100 kHz 100 A AC DC 1147A Current probe 50 MHz 15 A AC DC with AutoProbe interface 2 N2774A Obsolete replaced by N2782A with N2775A power supply N2780A Current probe 2 MHz 500 A AC DC use with N2779A power supply N2781A Current probe 10 MHz 150 A AC DC use with N2779A power supply N2782A Current probe 50 MHz 30 A AC DC use with N2779A power supply N2783A Current probe 100 MHz 30 A AC DC use with N2779A power supply N2893A Current probe 100 MHz 15 A AC DC with AutoProbe interface 2 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 329 22 Refe
132. The actual resolution of the display will not be this fine as the shape of the window will be the actual limiting factor in the FFTs ability to resolve two closely space frequencies A good way to test the ability of the FFT to resolve two closely spaced frequencies is to examine the sidebands of an amplitude modulated sine wave For the best vertical accuracy on peak measurements Make sure the probe attenuation is set correctly The probe attenuation is set from the Channel Menu if the operand is a channel Set the source sensitivity so that the input signal is near full screen but not clipped Use the Flat Top window Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Math Waveforms 4 Set the FFT sensitivity to a sensitive range such as 2 dB division For best frequency accuracy on peaks Use the Hanning window Use Cursors to place an X cursor on the frequency of interest Adjust frequency span for better cursor placement Return to the Cursors Menu to fine tune the X cursor For more information on the use of FFTs please refer to Keysight Application Note 243 The Fundamentals of Signal Analysis at http literature cdn keysight com litweb pdf 5952 8898E pdf Additional information can be obtained from Chapter 4 of the book Spectrum and Network Measurements by Robert A Witte FFT Units O dBV is the amplitude of a 1 Vrms sinusoid When the FFT source is channel 1 or channel 2 or channel 3 o
133. This turns on labels for the displayed analog and digital channels Labels are displayed at the left edge of the displayed traces The figure below shows an example of displayed labels KEYSIGHT TECHNOLOGIES 133 9 Labels 2 05V KEYSIGHT TECHNOLOGIES Acquisition Channels ew labe l Channel Delete Apply New Library c Character Label ACK 2 To turn the labels off press the Label key again To assign a predefined label to a channel 1 Press the Label key 2 Press the Channel softkey then turn the Entry knob or successively press the Channel softkey to select a channel for label assignment 134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Labels 9 2 05 KEYSIGHT TECHNOLOGIES Acquisition Channels Delete Apply New Library Character Label ACK The figure above shows the list of channels and their default labels The channel does not have to be turned on to have a label assigned to it 3 Press the Library softkey then turn the Entry knob or successively press the Library softkey to select a predefined label from the library 4 Press the Apply New Label softkey to assign the label to your selected channel 5 Repeat the above procedure for each predefined label you want to assign to a channel To define a new label 1 Press the Label key 2 Press the Channel softkey then turn the Entry knob or successively press the softkey to select a channel for label assi
134. To output the waveform generator sync pulse 261 To specify the expected output load 261 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 To use waveform generator logic presets 262 To add noise to the waveform generator output 262 To add modulation to the waveform generator output 263 To set up Amplitude Modulation AM 264 To set up Frequency Modulation FM 265 To set up Frequency Shift Keying Modulation FSK 266 To restore waveform generator defaults 267 18 Save Recall Setups Screens Data Saving Setups Screen Images or Data 269 To save setup files 271 To save BMP or PNG image files 271 To save CSV ASCII XY or BIN data files 272 Length Control 273 To save Lister datafiles 274 To save reference waveform files to a USB storage device 275 To save masks 275 To save arbitrary waveforms 275 To navigate storage locations 276 To enter filenames 276 Recalling Setups Masks or Data 277 To recall setup files 278 To recall mask files 278 To recall reference waveform files from a USB storage device 278 To recall arbitrary waveforms 279 Recalling Default Setups 279 Performing aSecure Erase 280 19 Print Screens To print the oscilloscope s display 281 16 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide To set up network printer connections 282 To specify the print options 284 To specify the palette option 284 20 Utility S
135. Units softkey in the channel s Probe Menu Units for math function waveforms are add or subtract VorA Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 73 4 Math Waveforms Math function Units multiply V A2 or W Volt Amp d dt V s or A s V second or A second fdt Vs or As V seconds or A seconds FFT dB decibels See also FFT Units on page 85 V square root V2 A12 or W12 Volt Amp When the FFT source is channel 1 2 3 or 4 FFT units will be displayed in dBV when channel units is set to Volts and channel impedance is set to 1 MQ FFT units will be displayed in dBm when channel units is set to Volts and channel impedance is set to 50Q FFT units will be displayed as dB for all other FFT sources or when a source channel s units has been set to Amps A scale unit of U undefined will be displayed for math functions when two source channels are used and they are set to dissimilar units and the combination of units cannot be resolved Math Operators Math operators perform arithmetic operations like add subtract or multiply on analog input channels Add or Subtract on page 74 Multiply or Divide on page 75 Add or Subtract When you select add or subtract the Source 1 and Source 2 values are added or subtracted point by point and the result is displayed You can use subtract to make a differential measurement or to compare two waveforms If your wavefo
136. Vision 3000 X Series Oscilloscopes User s Guide 19 Print Screens To print the oscilloscope s display 281 To set up network printer connections 282 To specify the print options 284 To specify the palette option 284 You can print the complete display including the status line and softkeys to a USB printer or a network printer when the DSOXLAN LAN VGA module is installed The Print Configuration Menu is displayed when you press the Print key The print option softkeys and the Press to Print softkey are ghosted not available until a printer is connected To print the oscilloscope s display 1 Connect a printer You can Connect a USB printer to either the USB port on the front panel or the rectangular USB host port on the rear panel For the most up to date listing of printers that are compatible with the InfiniiVision oscilloscopes please visit www keysight com find InfiniiVision printers Set up a network printer connection See To set up network printer connections on page 282 2 Press the Print key on the front panel 3 In the Print Configuration Menu press the Print to softkey then turn the Entry knob to select the desired printer KEYSIGHT TECHNOLOGIES 281 19 Print Screens 4 Press the Options softkey to select the print options O la Print Configuration Metra Print to Options Network Setup Press to NETPRTO Invert Grat ae Print See To specify the print options on page
137. Vision 3000 X Series Oscilloscopes User s Guide 215 14 Measurements 216 Minimum on page 216 Amplitude on page 216 Top on page 216 Base on page 217 Overshoot on page 217 Preshoot on page 219 Average on page 219 DC RMS on page 220 AC RMS on page 220 Ratio on page 222 Peak Peak The peak to peak value is the difference between Maximum and Minimum values The Y cursors show the values being measured Maximum Maximum is the highest value in the waveform display The Y cursor shows the value being measured Minimum Minimum is the lowest value in the waveform display The Y cursor shows the value being measured Amplitude The Amplitude of a waveform is the difference between its Top and Base values The Y cursors show the values being measured Top The Top of a waveform is the mode most common value of the upper part of the waveform or if the mode is not well defined the top is the same as Maximum The Y cursor shows the value being measured See Also To isolate a pulse for Top measurement on page 217 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Measurements 14 To isolate a pulse for Top measurement The following figure shows how to use Zoom mode to isolate a pulse for a Top measurement You may need to change the measurement window setting so that the measurement is made in the lower Zoom window See Measurement Window with Zoom Display on page 233
138. When triggering on a frame ID and data To select the frame ID value press the Frame ID softkey and use the Entry knob Note that you can enter a don t care value for the frame ID and trigger on data values only To set up the number of data bytes and enter their values in hexadecimal or binary press the Bits softkey to open the LIN Bits Menu 01 00 LIN Bits Menu Define Nibble Set all Nibs Data Hex 1 1 1 For details about using the LIN Bits Menu softkeys press and hold the softkey in question to display the built in help For LIN decode information see LIN Serial Decode on page 354 LIN Serial Decode To set up the oscilloscope to capture LIN signals see Setup for LIN Signals on page 351 354 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide CAN LIN Triggering and Serial Decode 23 For LIN triggering setup see LIN Triggering on page 353 To set up LIN serial decode 1 Press Serial to display the Serial Decode Menu Serial Decode Me Serial Mode Signals Show Parity Lister Serial 1 LIN gt 2 Choose whether to include the parity bits in the identifier field a lf you want to mask the upper two parity bits ensure that the box under the Show Parity softkey is not selected b If you want to include the parity bits in the identifier field ensure that the box under the Show Parity softkey is selected 3 If the decode line does not appear on the display press t
139. Z Z axis blanking 53 zoom and pan 48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide
140. a in the form of time and voltage pairs This file is much smaller than the ASCII XY data file See To save CSV ASCII XY or BIN data files on page 272 Lister data csv This is a CSV format file containing serial decode row information with commas separating the columns See To save Lister data files on page 274 Mask msk This creates a mask file in a Keysight proprietary format that can be read by Keysight InfiniiVision oscilloscopes A mask data file includes some oscilloscope setup information but not all setup information To save all setup information including the mask data file choose Setup scp format instead See To save masks on page 275 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Save Recall Setups Screens Data 18 Arbitrary Waveform data csv This creates a file of comma separated values for the arbitrary waveform points time and voltage values See To save arbitrary waveforms on page 275 Power Harmonics data csv When the DSOX3PWR power analysis allication is licensed this creates a file of comma separated values for the current harmonics power analysis results See the DSOX3PWR Power Measurement Application User s Guide for more information You can also configure the Quick Action key to save setups screen images or data See Configuring the Quick Action Key on page 302 To save setup files Setup files can be saved to one of 10 int
141. a probe on page 68 To specify the channel units on page 67 To specify the probe attenuation on page 67 To specify the probe skew on page 68 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Vertical Controls 3 To specify the channel units 1 Press the probe s associated channel key 2 Inthe Channel Menu press Probe 3 In the Channel Probe Menu press Units then select Volts for a voltage probe Amps for a current probe Channel sensitivity trigger level measurement results and math functions will reflect the measurement units you have selected To specify the probe attenuation This is set automatically if the oscilloscope can identify the connected probe See Analog channel inputs see page 39 The probe attenuation factor must be set properly for accurate measurement results If you connect a probe that is not automatically identified by the oscilloscope you can manually set the attenuation factor as follows 1 Press the channel key 2 Press the Probe softkey until you have selected how you want to specify the attenuation factor choosing either Ratio or Decibels 3 Turn the Entry knob a to set the attenuation factor for the connected probe When measuring voltage values the attenuation factor can be set from 0 1 1 to 1000 1 in a 1 2 5 sequence When measuring current values with a current probe the attenuation factor can be set from 10 V A to 0 001 V A When spec
142. acquisitions are made so that signal activity is displayed on the oscilloscope In the Normal trigger mode triggers and acquisitions only occur when the specified trigger conditions are found To select the trigger mode 1 Press the Mode Coupling key 2 Inthe Trigger Mode and Coupling Menu press the Mode softkey then select either Auto or Normal See the following When to Use Auto Trigger Mode on page 175 and When to Use Normal Trigger Mode on page 175 descriptions You can also configure the Quick Action key to toggle between the Auto and ormal trigger modes See Configuring the Quick Action Key on page 302 After the oscilloscope starts running after pressing Run or Single or changing the trigger condition the oscilloscope first fills the pre trigger buffer Then after the pre trigger buffer is filled the oscilloscope starts searching for a trigger and sampled data continues to flow data through the pre trigger buffer in a first in first out FIFO manner When a trigger is found the pre trigger buffer contains the events that occurred just before the trigger Then the oscilloscope fills the post trigger buffer and displays the acquisition memory If the acquisition was initiated by Run Stop the process repeats If the acquisition was initiated by pressing Single the acquisition stops and you can Pan and Zoom the waveform In either Auto or Normal trigger mode a trigger may be missed if
143. ad of the black backgound that appears on screen Aptos EEA Graticule Not Inverted Graticule Inverted Palette lets you choose between Color or Grayscale images 4 Finally press the Press to Save softkey A message indicating whether the save was successful is displayed When saving screen images the oscilloscope uses the last menu visited before pressing the Save Recall key This lets you save any relevant information within the softkey menu area To save a screen image showing the Save Recall Menu at the bottom press the Save Recall key twice before saving the image You can also save the oscilloscope s display image using a web browser See Get Image on page 317 for details See Also Adding an Annotation on page 303 To save CSV ASCII XY or BIN data files Data files can be saved to an external USB storage device 1 Press Save Recall gt Save gt Format then turn the Entry knob to select CSV data csv ASCII XY data csv or Binary data bin 2 Press the softkey in the second position and use the Entry knob to navigate to the save location See To navigate storage locations on page 276 3 Press the Settings softkey In the File Settings Menu you have these softkeys and options 272 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Save Recall Setups Screens Data 18 Setup Info when enabled setup information vertical horizontal trigger acquisition
144. adjust vertically and can be used to measure Volts or Amps dependent on the channel Probe Units setting or they can measure ratios When math functions are used as a source the measurement units correspond to that math function KEYSIGHT TECHNOLOGIES 199 13 Cursors The Y1 cursor is the short dashed horizontal line and the Y2 cursor is the long dashed horizontal line The Y cursors adjust vertically and typically indicate values relative to the waveform s ground point except math FFT where the values are relative to 0 dB In XY horizontal mode the Y cursors display channel 2 values Volts or Amps When active the Y1 and Y2 cursor values for the selected waveform source are displayed in the softkey menu area The difference between Y1 and Y2 AY is displayed in the Cursors box in the right side information area To make cursor measurements 200 1 Connect a signal to the oscilloscope and obtain a stable display 2 Press the Cursors key The Cursors box in the right side information area appears indicating that cursors are on Press the Cursors key again when you want to turn cursors off 3 In the Cursors Menu press Mode then select the desired mode Manual AX 1 AX and AY values are displayed AX is the difference between the X1 and X2 cursors and AY is the difference between the Y1 and Y2 cursors Cursors Menu Mode Source Manual Track Waveform As you move a marker horizontally
145. age location Location when pointing to a directory that is currently selected Selected when pointing to a file that can be loaded or deleted When this label appears you can press the Load File or Delete File softkeys to take the action Pushing the Entry knob is the same as pressing the Load File softkey A file that has been deleted from a USB storage device cannot be recovered by the oscilloscope Use your PC to create directories on a USB storage device Most USB mass storage devices are compatible with the oscilloscope However certain devices may be incompatible and may not be able to be read or written to When the USB mass storage device is connected to the oscilloscope s front or rear USB host port a small four color circle icon may be displayed briefly as the USB device is read Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Utility Settings 20 You do not need to eject the USB mass storage device before removing it Simply ensure that any file operation you ve initiated is done and remove the USB drive from the oscilloscope s host port Do not connect USB devices that identify themselves as hardware type CD because these devices are not compatible with the InfiniiVision X Series oscilloscopes If two USB mass storage devices are connected to the oscilloscope the first one is designated usb and the second one is designated usb2 See Also Chapter 18 Save Recall Setups
146. al bars indicate you need to expand the horizontal scale and run again to see decode Red dots in the decode line indicate that there is data that is not being displayed Scroll or expand the horizontal scale to view the information 348 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide CAN LIN Triggering and Serial Decode 23 Aliased bus values undersampled or indeterminate are drawn in pink Unknown bus values undefined or error conditions are drawn in red with a label Flagged error frames are drawn in red with an ERR label CAN Totalizer The CAN totalizer provides a direct measure of bus quality and efficiency The CAN totalizer measures total CAN frames flagged error frames overload frames and bus utilization UTIL 24 1 al Serial 1 The totalizer is always running counting frames and calculating percentages and is displayed whenever CAN decode is displayed The totalizer counts even when the oscilloscope is stopped not acquiring data Pressing the Run Stop key does not affect the totalizer When an overflow condition occurs the counter displays OVERFLOW The counters can be reset to zero by pressing the Reset CAN Counters softkey Types of Frames Active error frames are CAN frames in which a CAN node recognizes an error condition during a data or remote frame and issues an active error flag A partial frame occurs when the oscilloscope detects any error condition during a fram
147. al mode at slower time div settings extra samples are decimated in other words some are thrown away This mode yields the best display for most waveforms Peak Detect Acquisition Mode In Peak Detect mode at slower time div settings minimum and maximum samples are kept in order to capture infrequent and narrow events at the expense of exaggerating any noise This mode displays all pulses that are at least as wide as the sample period For InfiniiVision 3000 X Series oscilloscopes which have a maximum sample rate of 4 GSa s a sample is taken every 250 ps sample period See Also Glitch or Narrow Pulse Capture on page 189 Using Peak Detect Mode to Find a Glitch on page 190 188 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Acquisition Control 12 Glitch or Narrow Pulse Capture A glitch is a rapid change in the waveform that is usually narrow as compared to the waveform Peak detect mode can be used to more easily view glitches or narrow pulses In peak detect mode narrow glitches and sharp edges are displayed more brightly than when in Normal acquire mode making them easier to see To characterize the glitch use the cursors or the automatic measurement capabilities of the oscilloscope KEYSIGHT TECHNOLO Acquisition Channels Segmented Normal gt Figure 36 Sine With Glitch Normal Mode Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 189 12 Acquisition Control
148. ale All Normal 2 Ifyou want to return to the oscilloscope settings that existed before press Undo AutoScale 3 If you want to enable fast debug autoscaling change the channels autoscaled or preserve the acquisition mode during autoscale press Fast Debug Channels or Acq Mode These are the same softkeys that appear in the AutoScale Preferences Menu See To set AutoScale preferences on page 295 f you see the waveform but the square wave is not shaped correctly as shown above perform the procedure Compensate Passive Probes on page 32 f you do not see the waveform make sure the probe is connected securely to the front panel channel input BNC and to the left side Demo 2 Probe Comp terminal How AutoScale Auto Scale analyzes any waveforms present at each channel and at the external Works trigger input This includes the digital channels if connected Auto Scale finds turns on and scales any channel with a repetitive waveform that has a frequency of at least 25 Hz a duty cycle greater than 0 5 and an amplitude of at least 10 mV peak to peak Any channels that do not meet these requirements are turned off Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 31 1 Getting Started Compensate 32 The trigger source is selected by looking for the first valid waveform starting with external trigger then continuing with the lowest number analog channel up to the highest number analog channel a
149. and press the softkey again to enable decode 4 Press the Mode softkey then select CAN trigger type KEYSIGHT TECHNOLOGIES 343 23 CAN LIN Triggering and Serial Decode 5 Press the Signals softkey to open the CAN Signals Menu Threshold Sample Point Signal 1 00V 75 0 Diff L H 6 Press Source then select the channel for the CAN signal The label for the CAN source channel is automatically set 7 Press the Threshold softkey then turn the Entry knob to select the CAN signal threshold voltage level The threshold voltage level is used in decoding and it will become the trigger level when the trigger type is set to the selected serial decode slot 8 Press the Baud softkey then turn the Entry knob to select the baud rate that matches match your CAN bus signal The CAN baud rate can be set to predefined baud rates from 10 kb s up to 5 Mb s or a user defined baud rate from 10 0 kb s to 4 Mb s in increments of 100 b s Fractional user defined baud rates between 4 Mb s and 5 Mb s are not allowed The default baud rate is 125 kb s If none of the pre defined selections match your CAN bus signal select User Defined then press the User Baud softkey and turn the Entry knob to enter the baud rate 9 Press the Sample Point softkey then turn the Entry knob to select the point between phase segments 1 and 2 where the state of the bus is measured This controls the point within the bit s time where the bit value is cap
150. ange the trigger level because the trigger level is automatically set to the sync pulse Trigger coupling is automatically set to TV in the Trigger Mode and Coupling Menu Provide Correct Matching Many video signals are produced from 75 Q sources To provide correct matching to these sources a 75 Q terminator such as a Keysight 11094B should be connected to the oscilloscope input 4 Press the sync polarity softkey to set the Video trigger to either positive JL or negative LI sync polarity 5 Press the Settings softkey Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 159 10 Triggers Auto Setup 6 Inthe Video Trigger Menu press the Standard softkey to set the video standard The oscilloscope supports triggering on the following television TV and video standards Standard Type Sync Pulse NTSC Interlaced Bi level PAL Interlaced Bi level PAL M Interlaced Bi level SECAM Interlaced Bi level With the DSOX3VID extended video triggering license the oscilloscope additionally supports these standards SEU ETC Type Sync Pulse Generic Interlaced Progressive Bi level Tri level EDTV 480p 60 Progressive Bi level EDTV 567p 50 Progressive Bi level HDTV 720p 50 Progressive Tri level HDTV 720p 60 Progressive Tri level HDTV 1080p 24 Progressive Tri level HDTV 1080p 25 Progressive Tri level HDTV 1080p 30 Progressive Tri level HDTV 1080p 50 Progressive Tri level H
151. ardware self test Pressing Utility gt Service gt Hard ware Self Test performs a series of internal procedures to verify that the oscilloscope is operating properly It is recommended you run Hardware Self Test After experiencing abnormal operation For additional information to better describe an oscilloscope failure To verify proper operation after the oscilloscope has been repaired Successfully passing Hardware Self Test does not guarantee 100 of the oscilloscope s functionality Hardware Self Test is designed to provide an 80 confidence level that the oscilloscope is operating properly To perform front panel self test Pressing Utility gt Service gt Front Panel Self Test lets you test the front panel keys and knobs as well as the oscilloscope display Follow the on screen instructions To display oscilloscope information Press Help gt About Oscilloscope to display information about your oscilloscope Model number Serial number Bandwidth Module installed Software version Installed licenses See also Loading Licenses and Displaying License Information on page 331 300 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Utility Settings 20 To display the user calibration status Pressing Utility gt Service gt User Cal Status displays the summary results of the previous User Cal and the status of probe calibrations for probes that can be calibrated Note that passive probes d
152. area The Full Screen measurement interval variation measures the value on all displayed data points The N Cycles measurement interval variation measures the value on an integral number of periods of the displayed signal If less than three edges are present the measurement shows No edges The X cursors show what interval of the waveform is being measured Measurement Thresholds Setting measurement thresholds defines the vertical levels where measurements will be taken on an analog channel or math waveform Changing default thresholds may change measurement results The default lower middle and upper threshold values are 10 50 and 90 of the value between Top and Base Changing these threshold definitions from the default values may change the returned measurement results for Average Delay Duty Cycle Fall Time Frequency Overshoot Period Phase Preshoot Rise Time Width and Width 1 From the Measurement Menu press the Settings softkey then press the Thresholds softkey to set analog channel measurement thresholds You can also open the Measurement Threshold Menu by pressing Analyze gt Features and then selecting Measurement Thresholds 2 Press the Source softkey to select the analog channel or math waveform source for which you want to change measurement thresholds Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 231 14 Measurements Each analog channel and the math waveform can be assigne
153. ars on the screen SPI Triggering To set up the oscilloscope to capture SPI signals see Setup for SPI Signals on page 377 After the oscilloscope has been set up to capture SPI signals you can then trigger on a data pattern that occurs at the start of a frame The serial data string can be specified to be from 4 to 32 bits long When you select the SPI trigger type and Display Info is enabled a graphic is displayed showing the current state of the frame signal clock slope number of data bits and data bit values 1 Press Trigger 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select the serial slot Serial 1 or Serial 2 on which the SPI signals are being decoded Trigger Menu Trigger Trigger Setup Display Info 3 Press the Trigger Setup softkey to open the SPI Trigger Setup Menu Set all Bits Display Info Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 381 25 2C SPI Triggering and Serial Decode 4 Press the Trigger softkey then turn the Entry knob to select the trigger condition Master Out Slave In MOSI Data for triggering on the MOSI data signal Master In Slave Out MISO Data for triggering on the MISO data signal 5 Press the Bits softkey and turn the Entry knob to set the number of bits Bits in the serial data string The number of bits in the string can be set anywhere from 4 bits to 64 bits The data values for the serial st
154. asurement Category Definitions Measurement category is for measurements performed on circuits not directly connected to MAINS Examples are measurements on circuits not derived from MAINS and specially protected internal MAINS derived circuits In the latter case transient stresses are variable for that reason the transient withstand capability of the equipment is made known to the user Measurement category Il is for measurements performed on circuits directly connected to the low voltage installation Examples are measurements on household appliances portable tools and similar equipment easurement category Ill is for measurements performed in the building installation Examples are measurements on distribution boards circuit breakers wiring including cables bus bars junction boxes switches socket outlets in the fixed installation and equipment for industrial use and some other equipment for example stationary motors with permanent connection to the fixed installation easurement category IV is for measurements performed at the source of the low voltage installation Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units 324 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Reference 22 Transient Withstand Capability CAUTION A Maximum input voltage at analog inputs CAT 300 Vrms 400 Vpk transient overvoltage 1 6 kVpk 50Q input 5 Vrm
155. at all time div settings Averaging requires a stable trigger The number of averages can be set from 2 to 65536 in power of 2 increments A higher number of averages reduces noise more and increases vertical resolution Avgs Bits of resolution 2 8 4 9 16 10 64 11 gt 256 12 The higher the number of averages the slower the displayed waveform responds to waveform changes You must compromise between how quickly the waveform responds to changes and how much you want to reduce the displayed noise on the signal To use the Averaging mode 1 Press the Acquire key then press the Acq Mode softkey until the Averaging mode is selected 2 Press the Avgs softkey and turn the Entry knob to set the number of averages that best eliminates the noise from the displayed waveform The number of acquisitions being averaged is displayed in the Avgs softkey Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 191 12 Acquisition Control Acquisition ormal Channels 10 0 1 10 0 1 4A nu 4D Acq Mode Segmented Normal Figure 38 Random noise on the displayed waveform KEYSIGHT TECHNO Channels Segmented gt Figure 39 128 Averages used to reduce random noise 192 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Acquisition Control 12 See Also Chapter 11 Trigger Mode Coupling starting on page 173 High Resolution Acquisition Mode In High Re
156. ation The amount of delay range pre trigger and post trigger information available depends on the time div selected and memory depth The horizontal position knob works in the Normal time mode while acquisitions are running or when they are stopped When running adjusting the horizontal scale knob changes the sample rate When stopped adjusting the horizontal scale knob lets you zoom into acquired data See Panning and Zooming Single or Stopped Acquisitions on page 50 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 49 2 Horizontal Controls Note that the horizontal position knob has a different purpose in the Zoom display See To display the zoomed time base on page 54 Panning and Zooming Single or Stopped Acquisitions When the oscilloscope is stopped use the horizontal scale and position knobs to pan and zoom your waveform The stopped display may contain several acquisitions worth of information but only the last acquisition is available for pan and zoom The ability to pan move horizontally and scale expand or compress horizontally an acquired waveform is important because of the additional insight it can reveal about the captured waveform This additional insight is often gained from seeing the waveform at different levels of abstraction You may want to view both the big picture and the specific little picture details The ability to examine waveform detail after the waveform has been acquire
157. ay then turn the Entry knob to select the serial slot Serial 1 or Serial 2 on which the serial bus signals are being decoded If you select All the decode information for different buses is interleaved in time Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Serial Decode 7 1 12V KEYSIGHT TECHNOLOGIES Lister Menu Display Zoom to Options S UART AS232 Selection v Before you can select a row or navigate through the Lister data oscilloscope acquisitions must be stopped 4 Press the Single key in the Run Control group on the front panel to stop the acquisition Pressing Single instead of Stop fills the maximum memory depth When zoomed out and viewing a large number of packets the Lister may not be able to display information for all packets However when you press the Single key the Lister will contain all on screen serial decode information 5 Press the Scroll Lister softkey and use the Entry knob to scroll through the data Time tags in the Time column indicate the event time relative to the trigger point The time tags of events that are shown in the waveform display area are displayed with a dark background 6 Press the Zoom to Selection softkey or push the Entry knob to center the waveform display at the time associated with the selected Lister row and automatically set the horizontal scale setting 7 Press the Undo Zoom softkey to return to t
158. be attenuation 179 external trigger probe units 179 F f t 73 factory default settings 280 Fall time measurement 226 fall time measurement trend 95 falling edges count measurements 230 fast debug AutoScale 295 FFT aliasing 85 FFT DC value 85 FFT measurement hints 84 FFT measurements 81 FFT resolution 84 FFT spectral leakage 87 FFT units 85 FFT vertical units 82 FFT window 82 file explorer 291 file format ASCII 270 file format BIN 270 file format BMP 270 file format CSV 270 file format PNG 270 File keys 38 file name new 276 file save recall load 291 filter math function high pass and low pass 92 filters math 92 fine adjustment horizontal scale 55 firmware updates 333 firmware upgrade files 319 firmware version information 307 firmware versions 318 flash drive 39 flat frequency response 185 Flat top FFT window 82 FLEX license 332 FlexRay frame counter 365 FlexRay serial decode 363 FlexRay totalizer 365 FlexRay trigger 360 FM frequency modulation waveform generator output 265 folding frequency 183 forcing a trigger 141 frame trigger I2C 372 freeze display 303 freeze display Quick Freeze Display 303 French front panel overlays 41 French user interface and Quick Help 45 frequency deviation FM modulation 266 Frequency measurement 223 frequency measurement trend 95 frequency modulation FM wavefor
159. ble transparent backgrounds There is a preference setting for whether measurements statistics reference waveform information and other text displays have transparent or solid backgrounds 1 Press Utility gt Options gt Preferences 2 Press Transparent to toggle between transparent and solid text display backgrounds To load the default label library See To reset the label library to the factory default on page 137 To set up the screen saver The oscilloscope can be configured to turn on a display screen saver when the oscilloscope has been idle for a specified length of time 1 Press Utility gt Options gt Preferences gt Screen Saver to display the Screen Saver Menu Wait 180min Preview 2 Press the Saver softkey to select the screen saver type The screen saver can be set to Off to display any of the images shown in the list or can display a user defined text string If User is selected press the Spell softkey to select the first character of the text string Use the Entry knob to choose a character Then press the Enter softkey to advance to the next character and repeat the process You can use a connected USB keyboard instead of using the Spell and other character editing softkeys Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Utility Settings 20 The resultant string is displayed in the Text line above the softkeys Text AGILENT TECHNOLOGIES Saver Wait a
160. can be any analog or digital channel available on your oscilloscope Adjust the trigger level For analog channels turn the Trigger Level knob For digital channels press the Digital key and select Thresholds to set the threshold level The value of the trigger level or digital threshold is displayed in the upper right corner of the display Press the pulse polarity softkey to select positive JL or negative LJ polarity for the pulse width you want to capture The selected pulse polarity is displayed in the upper right corner of the display A positive pulse is higher than the current trigger level or threshold and a negative pulse is lower than the current trigger level or threshold Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Triggers 10 When triggering on a positive pulse the trigger will occur on the high to low transition of the pulse if the qualifying condition is true When triggering on a negative pulse the trigger will occur on the low to high transition of the pulse if the qualifying condition is true 6 Press the qualifier softkey lt gt gt lt to select the time qualifier The Qualifier softkey can set the oscilloscope to trigger on a pulse width that is Less than a time value lt For example for a positive pulse if you set t lt 10 ns K 10 ns oe Greater than a time value gt For example for a positive pulse if you set t gt 10 ns K 10 ns
161. cannot select 50Q mode This is done to prevent damage to the oscilloscope AC coupling places a 10 Hz high pass filter in series with the input waveform that removes any DC offset voltage from the waveform Note that Channel Coupling is independent of Trigger Coupling To change trigger coupling see To select the trigger coupling on page 175 To specify channel input impedance When you connect an AutoProbe self sensing probe or a compatible InfiniiMax probe the oscilloscope automatically configures the analog input channels to the correct impedance 1 Press the desired channel key 2 Inthe Channel Menu press Imped impedance then select either 50 Ohm matches 50 ohm cables commonly used in making high frequency measurements and 50 ohm active probes When 50 Ohm input impedance is selected it is displayed with the channel information on screen When AC coupling is selected see To specify channel coupling on page 63 or excessive voltage is applied to the input the oscilloscope automatically switches to 1M Ohm mode to prevent possible damage 1M Ohm is for use with many passive probes and for general purpose measurements The higher impedance minimizes the loading effect of the oscilloscope on the device under test This impedance matching gives you the most accurate measurements because reflections are minimized along the signal path 64 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Ve
162. channel source 3 Press the Type softkey then turn the Entry knob to select Delay 4 Press the Settings softkey to select the second analog channel source and slope for the delay measurement The default Delay settings measure from the rising edge of channel 1 to the rising edge of channel 2 5 Press the Back Up key to return to the Measurement Menu Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Measurements 14 6 Press the Add Measurement softkey to make the measurement The example below shows a delay measurement between the rising edge of channel 1 and the rising edge of channel 2 1 00 1 00 500 05 Channels s Measurements Freq 1 Pk Pk 1 Delay 14 24 191 0us Source Add Settings Clear Meas Measurement lt Phase Phase is the calculated phase shift from source 1 to source 2 expressed in degrees Negative phase shift values indicate that the rising edge of source 1 occurred after the rising edge of source 2 Delay Phase Source Period 360 i lt Period gt Source 1 eee ee ee fee ae Delay Source 2 NO A No O an 1 Press the Meas key to display the Measurement Menu Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 227 14 Measurements 2 Press the Source softkey then turn the Entry knob to select the first analog channel source 3 Press the Type softkey then turn the Entry knob to select Delay 4 Press the Settings softkey to se
163. ck softkey the mask is redrawn to match the source whenever you move the waveform For example if you change the horizontal timebase or the vertical gain the mask is redrawn with the new settings When you turn off Source Lock the mask is not redrawn when horizontal or vertical settings are changed Source If you change the Source channel the mask is not erased It is re scaled to the vertical gain and offset settings of the channel to which it is assigned To create a new mask for the selected source channel go back up in the menu hierarchy then press Automask and press Create Mask The Source softkey in the Mask Setup Menu is the same as the Source softkey in the Automask Menu Test All When enabled all displayed analog channels are included in the mask test When disabled just the selected source channel is included in the test Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 241 15 Mask Testing Mask Statistics From the Mask Test Menu press the Statistics softkey to enter the Mask Statistics Menu 7 750 Channels Transparent Clear Display Show Stats When you enable Show Statistics the following information is displayed Current mask name of mask Channel number date and time of Tests total number of mask tests executed Status Passing Failing or Untested Accumulated test time in hours minutes seconds and tenths of seconds And for each analog chann
164. com find contactus 1 Write the following information on a tag and attach it to the oscilloscope Name and address of owner Model number Serial number Description of service required or failure indication 2 Remove accessories from the oscilloscope Only return accessories to Keysight Technologies if they are associated with the failure symptoms 3 Package the oscilloscope You can use the original shipping container or provide your own materials sufficient to protect the instrument during shipping 4 Seal the shipping container securely and mark it FRAGILE Configuring the Quick Action Key The Quick Action key lets you perform common repetitive actions by pressing a single key To configure the Quick Action key 1 Press Utility gt Quick Action gt Action then select the action that should be performed Off disables the Quick Action key Quick Measure All displays a popup containing a snapshot of all the single waveform measurements The Source softkey lets you select the waveform source which also becomes the source selection in the Measurement Menu See Chapter 14 Measurements starting on page 209 Quick Print prints the current screen image Press Settings to set up the printing options See Chapter 19 Print Screens starting on page 281 302 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Utility Settings 20 Quick Save saves the current image wa
165. conditions are met the oscilloscope triggers and displays the acquisition For digital channels each time the oscilloscope takes a sample it compares the input voltage to the logic threshold If the voltage is above the threshold the oscilloscope stores a 1 in sample memory otherwise it stores a 0 To display digital channels using AutoScale When signals are connected to the digital channels be sure to connect the ground leads AutoScale quickly configures and displays the digital channels To configure the instrument quickly press the AutoScale key 100 08 Auto 2 08V KEYSIGHT TECHNOLOGIES Channels et h Undo Fast Debug Channels Acq Mode Autoscale All Normal Figure 20 Example AutoScale of digital channels MSO models only Any digital channel with an active signal will be displayed Any digital channels without active signals will be turned off Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 107 6 Digital Channels To undo the effects of AutoScale press the Undo AutoScale softkey before pressing any other key This is useful if you have unintentionally pressed the AutoScale key or do not like the settings AutoScale has selected This will return the oscilloscope to its previous settings See also How AutoScale Works on page 31 To set the instrument to the factory default configuration press the Default Setup key Interpreting the digital waveform display The foll
166. connectors The cable is a 2 m logic analyzer probe cable and the termination adapter provides the proper RC networks in a very convenient package These parts as well as the 1251 8106 20 pin low profile straight board connector can be ordered from Keysight Technologies To replace digital probe leads If you need to remove a probe lead from the cable insert a paper clip or other small pointed object into the side of the cable assembly and push to release the latch while pulling out the probe lead 118 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Digital Channels Table 3 Digital Probe Replacement Parts Part Number Description N6450 60001 Digital probe kit contains 54620 61601 16 channel cable 01650 82103 2 inch probe grounds qty 5 and 5090 4832 grabbers qty 20 6 54620 61601 16 channel cable with 16 probe leads and 2 pod ground leads qty 1 5959 9333 Replacement probe leads qty 5 also contains 01650 94309 probe labels 5959 9334 Replacement 2 inch probe grounds qty 5 5959 9335 Replacement pod ground leads qty 5 5090 4833 Grabbers qty 20 01650 94309 Package of probe labels For other replacement parts consult the nfiniiVision 2000 3000 X Series Oscilloscopes Service Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 119 6 Digital Channels 120 Keysight InfinilVision 3000 X Series Oscilloscopes U
167. cope is AutoTriggering Press the Mode Coupling key then press the Mode softkey to set the trigger mode from Auto to Normal You can use the horizontal Zoom window for easier navigation of the decoded data Interpreting CAN Decode on page 348 CAN Totalizer on page 349 Interpreting CAN Lister Data on page 350 Searching for CAN Data in the Lister on page 351 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 347 23 CAN LIN Triggering and Serial Decode Interpreting CAN Decode 100 05 Active Error Frame red _ KEYSIGHIT Ac Missing Acknowledge red CRC blue 2 Channels Data white _ Bac Data Length Code blue _ S Frame ID yellow _ UTIL 24 1 Lister gt Frame ID appears in hex digits in yellow Frames of 11 or 29 bits are eS automatically detected Remote frame RMT appears in green Data length code DLC appears in blue for data frames and green for remote frames Data bytes appear in hex digits in white for data frames Cyclic redundancy check CRC appears in hex digits in blue when valid or in red to indicate that the oscilloscope s hardware decode calculated a CRC that is different from the incoming CRC data stream Angled waveforms show an active bus inside a packet frame id level blue lines show an idle bus Decoded text is truncated at the end of the associated frame when there Is nsufficient space within frame boundaries Pink vertic
168. d Perfectly compensated Over compensated Under compensated L 7 Connect probes to all other oscilloscope channels channel 2 of a 2 channel oscilloscope or channels 2 3 and 4 of a 4 channel oscilloscope 8 Repeat the procedure for each channel Learn the Front Panel Controls and Connectors On the front panel key refers to any key button you can press Softkey specifically refers to the six keys that are directly below the display The legend for these keys is directly above them on the display Their functions change as you navigate through the oscilloscope s menus For the following figure refer to the numbered descriptions in the table that follows Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 33 1 Getting Started 5 Tools keys 6 Trigger controls 7 Horizontal controls 8 Run Control keys EI 9 Default Setup key 10 Auto Scale key 4 11 Additional waveform controls 4 Entry knob 3 Intensity key 12 Measure controls 2 Softkeys 13 Waveform keys 14 File keys 1 Power switch qn re 15 Help key 21 Waveform 20 Digital 19 USB 18 Demo 2 Ground 17 Analog 16 Vertical controls generator channel Host and Demo 1 channel output inputs port terminals inputs 1 Power switch Press once to switch power on press again to switch power off See Power On the Oscilloscope on page 27
169. d is a benefit generally associated with digital oscilloscopes Often this is simply the ability to freeze the display for the purpose of measuring with cursors or printing the screen Some digital oscilloscopes go one step further by including the ability to further examine the signal details after acquiring them by panning through the waveform and changing the horizontal scale There is no limit imposed on the scaling ratio between the time div used to acquire the data and the time div used to view the data There is however a useful limit This useful limit is somewhat a function of the signal you are analyzing Zooming into stopped acquisitions The screen will still contain a relatively good display if you zoom in horizontally by a factor of 1000 and zoom in vertically by a factor of 10 to display the information from where it was acquired Remember that you can only make automatic measurements on displayed data To change the horizontal time mode Normal XY or Roll 1 Press Horiz 2 Inthe Horizontal Menu press Time Mode then select 50 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Horizontal Controls 2 Normal the normal viewing mode for the oscilloscope In the Normal time mode signal events occurring before the trigger are plotted to the left of the trigger point W and signal events after the trigger plotted to the right of the trigger point XY XY mode changes the display from a volts v
170. d out on a signal using the horizontal scale knob and vertical volts division knob you affect the resolution of the display Because measurements and math functions are performed on displayed data you affect the resolution of functions and measurements KEYSIGHT TECHNOLOGIES 209 14 Measurements To make automatic measurements 1 Press the Meas key to display the Measurement Menu Source Settings Clear Meas Statistics gt w 2 Press the Source softkey to select the channel running math function or reference waveform to be measured Only channels math functions or reference waveforms that are displayed are available for measurements If a portion of the waveform required for a measurement is not displayed or does not display enough resolution to make the measurement the result will display No Edges Clipped Low Signal lt value or gt value or a similar message to indicate that the measurement may not be reliable 3 Press the Type softkey then rotate the Entry knob to select a measurement to be made 500 05 Channels Deviation Measurements giii Period gt TFL Frequency Measurement Menu Source Type Add Settings Clear Meas Statistics Freq Measurement Ess ae E a For more information on the types of measurements see Measurements Summary on page 211 210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Measurements 14 4 The Settings softkey will
171. d standards only All Fields Trigger on the rising edge of the first pulse in the vertical sync interval All Lines Trigger on all horizontal sync pulses Line Trigger on the selected line number EDTV and HDTV standards only Line Field1 and Line Field2 Trigger on the selected line in field 1 or field 2 interlaced standards only Line Alternate Alternately trigger on the selected line in field 1 and field 2 NTSC PAL PAL M and SECAM only 9 Ifyou select a line mode press the Line softkey then rotate the Entry knob to select the line number on which you want to trigger Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 161 10 Triggers The following table lists the line or count numbers per field for each video standard Video standard Field 1 Field 2 Alt Field NTSC 1 to 263 1 to 262 1 to 262 PAL 1 to 313 314 to 625 1 to 312 PAL M 1 to 263 264 to 525 1 to 262 SECAM 1 to 313 314 to 625 1 to 312 The following table lists the line numbers for each EDTV HDTV video standard available with the DSOX8VID extended video triggering license Video Triggering EDTV 480p 60 1 to 525 EDTV 567p 50 1 to 625 HDTV 720p 50 720p 60 1 to 750 HDTV 1080p 24 1080p 25 1080 30 1 to 1125 1080p 50 1080p 60 HDTV 1080i 50 1080i 60 1 to 1125 The following are exercises to familiarize you with video triggering These exercises use the
172. d still be damaged depending on the time constant of the signal The 50 Q input protection only functions when the oscilloscope is powered on With 10073C 10 1 probe CAT 500 Vpk With N2862A or N2863A 10 1 probe 300 Vrms 28 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide CAUTION Getting Started 1 A Do not float the oscilloscope chassis Defeating the ground connection and floating the oscilloscope chassis will probably result in inaccurate measurements and may also cause equipment damage The probe ground lead is connected to the oscilloscope chassis and the ground wire in the power cord If you need to measure between two live points use a differential probe with sufficient dynamic range Do not negate the protective action of the ground connection to the oscilloscope The oscilloscope must remain grounded through its power cord Defeating the ground creates an electric shock hazard Input a Waveform The first signal to input to the oscilloscope is the Demo 2 Probe Comp signal This signal is used for compensating probes 1 Connect an oscilloscope probe from channel 1 to the Demo 2 Probe Comp terminal on the front panel 2 Connect the probe s ground lead to the ground terminal next to the Demo 2 terminal Recall the Default Oscilloscope Setup To recall the default oscilloscope setup 1 Press Default Setup The default setup restores the oscilloscope s default settings This places
173. d unique threshold values 3 Press the Type softkey to set the measurement threshold to percentage of Top and Base value or to Absolute absolute value Percentage thresholds can be set from 5 to 95 The units for absolute threshold for each channel is set in the channel probe menu When the Source is set to Math f t the threshold Type can only be set to Percent TIP Absolute threshold hints Absolute thresholds are dependent on channel scaling probe attenuation and probe units Always set these values first before setting absolute thresholds The minimum and maximum threshold values are limited to on screen values If any of the absolute threshold values are above or below the minimum or maximum waveform values the measurement may not be valid 4 Press the Lower softkey then turn the Entry knob to set the lower measurement threshold value Increasing the lower value beyond the set middle value will automatically increase the middle value to be more than the lower value The default lower threshold is 10 or 800 mV If threshold Type is set to the lower threshold value can be set from 5 to 93 5 Press the Middle softkey then turn the Entry knob to set the middle measurement threshold value The middle value is bounded by the values set for lower and upper thresholds The default middle threshold is 50 or 1 20 V If threshold Type is set to the middle threshold value can be set from 6 to 94
174. d voltage level to the middle of the LIN signal The threshold voltage level is used in decoding and it will become the trigger level when the trigger type is set to the selected serial decode slot 8 Press the Baud Rate softkey to open the LIN Baud Rate Menu 9 Press the Baud softkey then turn the Entry knob to select the baud rate that matches match your LIN bus signal The default baud rate is 19 2 kb s If none of the pre defined selections match your LIN bus signal select User Defined then press the User Baud softkey and turn the Entry knob to enter the baud rate You can set the LIN baud rate from 2 4 kb s to 625 kb s in increments of 100 b s 10 Press the S Back Up key to return to the LIN Signals Menu 11 Press the Sample Point softkey then turn the Entry knob to select the sample point at which the oscilloscope will sample the bit value 60 70 80 Sample Point 12 Press the Standard softkey then turn the Entry knob to select the LIN standard you are measuring LIN 1 3 or LIN 2 0 352 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide CAN LIN Triggering and Serial Decode 23 For LIN 1 2 signals use the LIN 1 3 setting The LIN 1 3 setting assumes the signal follows the Table of Valid ID Values as shown in section A 2 of the LIN Specification dated December 12 2002 If your signal does not comply with the table use the LIN 2 0 setting 13 Press the Sync Break softkey and se
175. defined in FAR 52 227 14 June 1987 or DFAR 252 227 7015 b 2 November 1995 as applicable in any tech nical data Safety Notices CAUTION A CAUTION notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly performed or adhered to could result in damage to the product or loss of important data Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met A WARNING notice denotes a haz ard It calls attention to an operat ing procedure practice or the like that if not correctly performed or adhered to could result in personal injury or death Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide InfiniiVision 3000 X Series Oscilloscopes At a Glance a E a a go is a a a Table 1 3000 X Series Model Numbers Band widths Sample Rates Band width 100 MHz 200 MHz 350 MHz 500 MHz 1 GHz Sample Rate interleaved 4 GSa s 4 GSa s 4 GSa s 4 GSa s 5 GSa s non interleaved 2 GSa s 2 GSa s 2 GSa s 2 GSa s 2 5 GSa s 2 Channel 16 Logic MSO X 3012A MSO X 3032A MSO X 3052A MSO X 3102A Channels MSO 4 Channel 16 Logic MSO X 3014A MSO X 3024A MSO X 3034A MSO X 3054A MSO X 3104A Channels MSO 2 Channel DSO DSO X 3012A DSO X 3032A DSO X 3052A DSO X 3102A 4 Channel DSO DSO X 3014A DSO X 30
176. device under test from 4 to 32 bits 10 Press the Receiver softkey Turn the Entry knob to match the receiver word size of the device under test from 4 to 32 bits 11 Press the Alignment softkey then turn the Entry knob to select the desired alignment of the data signal SDATA The on screen diagram changes with your selection Standard Alignment MSB of data for each sample is sent first LSB is sent last The MSB appears on the SDATA line one bit clock after the edge of the WS transition Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide I2S Triggering and Serial Decode 26 LEFT CHANNEL RIGHT CHANNEL ws l i souk il Ul g ad gt ONE SCLK CYCLE 1 SDATA o 1 2 l n2 n1 o 1 2 h21 7 MSB eenen LSB MSB zse eeecesse ntectec LSB Left Justified Data transmission MSB first begins at the edge of the WS transition without the one bit delay that Standard format employs LEFT CHANNEL RIGHT CHANNEL 1 1 s sae T p l sci ll Ul tl UU soara oa m21 OEE n 2 n 1 MSB Sees E E E Lo 2 AOE Seo ETE LSB Right Justified Data transmission MSB first is right justified to the transition
177. ds the start of Command Status word at the end of a valid C S Sync pulse Remote Terminal Address finds the Command Status word whose RTA matches the specified value The value is specified in hex When this option is selected the RTA softkey becomes available and lets you select the hex Remote Terminal Address value to find Remote Terminal Address 11 Bits finds the RTA and the remaining 11 bits that match the specified criteria When this option is selected these softkeys become available The RTA softkey lets you select the hex Remote Terminal Address value The Bit Time softkey lets you select the bit time position The 01 X softkey lets you set the bit time position value as a 1 0 or X don t care pa Parity Error finds odd parity bits that are incorrect for the data in the word Sync Error finds invalid Sync pulses Manchester Error finds Manchester encoding errors For more information on searching data see Searching Lister Data on page 124 For more information on using the Navigate key and controls see Navigating the Time Base on page 58 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 403 27 MIL STD 1553 ARINC 429 Triggering and Serial Decode Setup for ARINC 429 Signals 404 Setup consists of first connecting the oscilloscope to a ARINC 429 signal using a differential active probe the Keysight N2791A is recommended then using the Signals Menu to sp
178. e Press the Mode softkey then select 12C trigger type I2C Triggering 370 Press the Signals softkey to open the fc Signals Menu PC Signals Menu SCL Threshold Threshold TTL 1 4V TTL 1 4 For both the SCL serial clock and SDA serial data signals a b Connect an oscilloscope channel to the signal in the device under test Press the SCL or SDA softkey then turn the Entry knob to select the channel for the signal Press the corresponding Threshold softkey then turn the Entry knob to select the signal threshold voltage level The threshold voltage level is used in decoding and it will become the trigger level when the trigger type is set to the selected serial decode slot Data must be stable during the whole high clock cycle or it will be interpreted as a start or stop condition data transitioning while the clock is high The SCL and SDA labels for the source channels are automatically set To set up the oscilloscope to capture 12C signals see Setup for I2C Signals on page 369 After the oscilloscope has been set up to capture I2C signals you can trigger ona stop start condition a restart a missing acknowledge an EEPROM data read or on a read write frame with a specific device address and data value 1 2 3 Press Trigger then select the I2C trigger type Press Trigger In the Trigger Menu press the Trigger softkey then turn the Entry knob to select the serial slot
179. e When using the oscilloscope in a bench top setting provide at least 2 clearance at the sides and 4 100 mm clearance above and behind the oscilloscope for proper cooling Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 27 1 Getting Started To power on the oscilloscope 1 Connect the power cord to the rear of the oscilloscope then to a suitable AC voltage source Route the power cord so the oscilloscope s feet and legs do not pinch the cord 2 The oscilloscope automatically adjusts for input line voltages in the range 100 to 240 VAC The line cord provided is matched to the country of origin WARNING Always use a grounded power cord Do not defeat the power cord ground 3 Press the power switch The power switch is located on the lower left corner of the front panel The oscilloscope will perform a self test and will be operational in a few seconds Connect Probes to the Oscilloscope CAUTION 1 Connect the oscilloscope probe to an oscilloscope channel BNC connector 2 Connect the probe s retractable hook tip to the point of interest on the circuit or device under test Be sure to connect the probe ground lead to a ground point on the circuit A Maximum input voltage at analog inputs CAT 300 Vrms 400 Vpk transient overvoltage 1 6 kVpk 50Q input 5 Vrms Input protection is enabled in 50 Q mode and the 50 Q load will disconnect if greater than 5 Vrms is detected However the inputs coul
180. e Searching Lister Data on page 124 366 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide FlexRay Triggering and Serial Decode 24 For more information on using the Navigate key and controls see Navigating the Time Base on page 58 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 367 24 FlexRay Triggering and Serial Decode 368 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 25 I2C SPI Triggering and serial Decode Setup for 12C Signals 369 12C Triggering 370 12C Serial Decode 374 Setup for SPI Signals 377 SPI Triggering 381 SPI Serial Decode 383 I2C SPI triggering and serial decode requires Option EMBD or the DSOX3EMBD upgrade NOTE Only one SPI serial bus can be decoded at a time Setup for I2C Signals I2C Inter IC bus signals setup consists of connecting the oscilloscope to the serial data SDA line and the serial clock SCL line and then specifying the input signal threshold voltage levels To set up the oscilloscope to capture C signals use the Signals softkey which appears in the Serial Decode Menu 1 Press Label to turn on labels 2 Press Serial KEYSIGHT TECHNOLOGIES 369 25 I2C SPI Triggering and Serial Decode 3 Press the Serial softkey turn the Entry knob to select the desired slot Serial 1 or Serial 2 and press the softkey again to enable decod
181. e RMT remote frame or Data DLC data length code Data data bytes CRC cyclic redundancy check Errors highlighted in red Errors can be Acknowledge Ack A Form Fo or Frame Fr Different kinds of errors can be combined like Fo Fr in the above example Aliased data is highlighted in pink When this happens decrease the horizontal time div setting and run again 350 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide CAN LIN Triggering and Serial Decode 23 Searching for CAN Data in the Lister The oscilloscope s search capability lets you search for and mark certain types of CAN data in the Lister You can use the Navigate key and controls to navigate through the marked rows 1 With CAN selected as the serial decode mode press Search 2 Inthe Search Menu press the Search softkey then turn the Entry knob to select the serial slot Serial 1 or Serial 2 on which the CAN signal is being decoded 3 Press Search then select from these options Remote Frame ID RTR Finds remote frames with the specified ID Press the Bits softkey to enter the ID Data Frame ID RTR Finds data frames matching the specified ID Press the Bits softkey to enter the ID Remote or Data Frame ID Finds remote or data frames matching the specified ID Press the Bits softkey to select the ID Data Frame ID and Data Finds data frames matching the specified ID and data Press the Bit
182. e on page 393 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide I2S Triggering and Serial Decode 26 I2S Serial Decode To set up the oscilloscope to capture 2S signals see Setup for 12S Signals on page 387 For I2S triggering setup see 12S Triggering on page 390 To set up 12S serial decode 1 Press Serial to display the Serial Decode Menu Serial Decode Menu Mode Bus Contig PS 2 Press the Base softkey to select the number base in which to display decoded data 3 Ifthe decode line does not appear on the display press the Serial key to turn it on 4 Ifthe oscilloscope is stopped press the Run Stop key to acquire and decode data If the setup does not produce a stable trigger the I2S signal may be slow enough that the oscilloscope is AutoTriggering Press the Mode Coupling key then press the Mode softkey to set the trigger mode from Auto to Normal You can use the horizontal Zoom window for easier navigation of the acquired data See Also Interpreting 12S Decode on page 394 Interpreting 12S Lister Data on page 395 Searching for I2S Data in the Lister on page 395 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 393 26 12S Triggering and Serial Decode Interpreting 12S Decode 5 0008 Channels Signals Bus Config Base Lister Sd Decimal ke Angled waveforms show an active bus inside a packet frame Mid level blue lines show an idle bus
183. e Trigger Buffer XX J y Acquisition Memory Adjusting the Trigger Level You can adjust the trigger level for a selected analog channel by turning the Trigger Level knob 140 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Triggers 10 You can push the Trigger Level knob to set the level to the waveform s 50 value If AC coupling is used pushing the Trigger Level knob sets the trigger level to about 0 V The position of the trigger level for the analog channel is indicated by the trigger evel icon Tp if the analog channel is on at the far left side of the display The value of the analog channel trigger level is displayed in the upper right corner of the display The trigger level for a selected digital channel is set using the threshold menu in the Digital Channel Menu Press the Digital key on the front panel then press the Thresholds softkey to set the threshold level TTL CMOS ECL or user defined for the selected digital channel group The threshold value is displayed in the upper right corner of the display The line trigger level is not adjustable This trigger is synchronized with the power line supplied to the oscilloscope You can also change the trigger level of all channels by pressing Analyze gt Features and then selecting Trigger Levels Forcing a Trigger The Force Trigger key causes a trigger on anything and displays the acquisition This key is usefu
184. e displayed or a front panel light to blink View installed options view firmware versions and install firmware upgrade files and view calibration status via the Instrument Utilities page View and modify the oscilloscope s network configuration KEYSIGHT TECHNOLOGIES 307 21 Web Interface The web interface for InfiniiVision X Series oscilloscopes also provides help for each of its pages Microsoft Internet Explorer is the recommended Web browser for communication and control of the oscilloscope Other Web browsers may work but are not guaranteed to work with the oscilloscope The Web browser must be Java enabled with Sun Microsystems Java Plug in Before you can use the web interface you must place the oscilloscope on the network and set up its LAN connection Accessing the Web Interface To access the oscilloscope s web interface 1 Connect the oscilloscope to your LAN see To establish a LAN connection on page 289 or establish a point to point connection see Stand alone Point to Point Connection to a PC on page 290 It is possible to use a point to point connection but using a normal LAN connection is the preferred method 2 Type the oscilloscope s hostname or IP address in the web browser The oscilloscope s web interface Welcome Page is displayed 308 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Web Interface 21 Support Products Keysight Site KEYSIGHT Oscillosc
185. e or disable trigger HF Reject 177 To set the trigger holdoff 177 External Trigger Input 178 To access the Trigger Mode and Coupling Menu In the Trigger section of the front panel press the Mode Coupling key id Coupling Menu Coupling Noise Rej HF Reject Holdoff External DC 40 000ns ma Noisy Signals If the signal you are probing is noisy you can set up the oscilloscope to reduce the noise in the trigger path and on the displayed waveform First stabilize the displayed waveform by removing the noise from the trigger path Second reduce the noise on the displayed waveform 1 Connect a signal to the oscilloscope and obtain a stable display 2 Remove the noise from the trigger path by turning on high frequency rejection To enable or disable trigger HF Reject on page 177 low frequency rejection To select the trigger coupling on page 175 or To enable or disable trigger noise rejection on page 177 3 Use Averaging Acquisition Mode on page 191 to reduce noise on the displayed waveform KEYSIGHT TECHNOLOGIES 173 11 Trigger Mode Coupling To select the Auto or Normal trigger mode Triggering and the Pre and Post Trigger Buffers Trigger Indicator 174 When the oscilloscope is running the trigger mode tells the oscilloscope what to do when triggers are not occurring In the Auto trigger mode the default setting if the specified trigger conditions are not found triggers are forced and
186. e same as an existing label in the library the new label will not be put in the library For example if label AO is in the library and you make a new label called A12345 the new label is not added to the library When you save a new user defined label the new label will replace the oldest label in the list Oldest is defined as the longest time since the label was last assigned to a channel Any time you assign any label to a channel that label will move to the newest in the list Thus after you use the label list for a while your labels will predominate making it easier to customize the instrument display for your needs When you reset the label library list see next topic all of your custom labels will be deleted and the label list will be returned to its factory configuration To reset the label library to the factory default Pressing the Default Library softkey will remove all user defined labels from the library and set the labels back to the factory default Once deleted these user defined labels cannot be recovered 1 Press Utility gt Options gt Preferences 2 Press the Default Library softkey This will delete all user defined labels from the library and set the labels in the library back to the factory default However this does not default the labels currently assigned to the channels those labels that appear in the waveform area Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 137 9 Labels
187. e that is not followed by an active error flag Partial frames are not counted Counters The FRAMES counter gives the total number of completed remote data overload and active error frames The OVLD counter gives the total number of completed overload frames and their percentage of the total number of frames The ERR counter gives the total number of completed active error frames and their percentage of the total number of frames The UTIL bus load indicator measures the percentage of time the bus is active The calculation is done on 330 ms periods approximately every 400 ms Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 349 23 CAN LIN Triggering and Serial Decode Example If a data frame contains an active error flag both the FRAMES counter and the ERR counter will be incremented If a data frame contains an error that is not an active error it is considered a partial frame and no counters are incremented Interpreting CAN Lister Data ID 0296A95D DLC Data Data 5 at 00 01 02 03 04 9639 KEYSIGHT TECHNOLOGIES O7F RMT O7F Data 22 Acquisition RMT Data RMT Data 42 9A AG 9E Channels 0296A95D RMT O0296A95D Lister Menu Data 00 02 04 06 08 Zoom to Selection Options w In addition to the standard Time column the CAN Lister contains these columns ID frame ID Type frame typ
188. e waveform display area The Horizontal scale sweep speed knob controls the size of the box The Horizontal position delay time knob sets the left to right position of the zoom window The delay value which is the time displayed relative to the trigger point is momentarily displayed in the upper right portion of the display when the delay time knob is turned Negative delay values indicate you re looking at a portion of the waveform before the trigger event and positive values indicate you re looking at the waveform after the trigger event To change the time div of the normal window turn off Zoom then turn the horizontal scale sweep speed knob For information about using zoom mode for measurements refer to To isolate a pulse for Top measurement on page 217 and To isolate an event for frequency measurement on page 223 To change the horizontal scale knob s coarse fine adjustment setting 1 Push the horizontal scale knob or press Horiz gt Fine to toggle between fine and coarse adjustment of the horizontal scale When Fine is enabled turning the horizontal scale knob changes the time div displayed in the status line at the top of the display in smaller increments The time div remains fully calibrated when Fine is on When Fine is turned off the Horizontal scale knob changes the time div setting in a 1 2 5 step sequence Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 55 2 Horiz
189. ecific error symbol in red Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide FlexRay Triggering and Serial Decode 24 FlexRay Totalizer The FlexRay totalizer consists of counters that provide a direct measure of bus quality and efficiency The totalizer appears on screen whenever FlexRay Decode is ON in the Serial Decode Menu Serial Deco nu FRAMES 010 NULL OOOO Serial Signals Reset FlexRay Serial 1 al Counters The FRAMES counter gives a real time count of all captured frames The NULL counter gives the number and percentage of null frames The SYNC counter gives the number and percentage of sync frames The totalizer runs counting frames and calculating percentages even when the oscilloscope is stopped not acquiring data When an overflow condition occurs the counter displays OVERFLOW The counters can be reset to zero by pressing the Reset FlexRay Counters softkey Interpreting FlexRay Lister Data len HCAC CYC Data 106 6us 9D7F3A 91 56us 815ECD 56 58us 49C7F6 41 56us 55E601 9D7F3A 8 440us 815ECD 43 42us 49C7F6 Channels 58 44us 55E601 93 45us 9D7F3A Data 01 01 Undo Options Zoom Paes Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 365 24 FlexRay Triggering and Serial Decode In addition to the standard Time column the FlexRay Lister contains these columns FID frame ID Len payload len
190. ecify the signal source the high and low trigger threshold voltage levels the signal speed and the signal type To set up the oscilloscope to capture ARINC 429 signals 1 Press Label to turn on labels 2 Press Serial 3 Press the Serial softkey turn the Entry knob to select the desired slot Serial 1 or Serial 2 and press the softkey again to enable decode 4 Press the Mode softkey then select ARINC 429 decode mode 5 Press the Signals softkey to open the ARINC 429 Signals Menu ARINC 429 Signals Menu Source High Threshold Low Threshold Speed Signal Type W 1 00 Es Diff 4 B Auto Setup 6 Press Source then select the channel for the ARINC 429 signal The label for the ARINC 429 source channel is automatically set 7 Press the Speed softkey to open the ARINC 429 Speed Menu User Baud 100kb s 8 Press the Speed softkey and specify the speed of the ARINC 429 signal High 100 kb s Low 12 5 kb s User Defined if selected press the User Baud softkey and turn the Entry knob to specify the speed 9 Press the Back Up key to return to the ARINC 429 Signals Menu 10 Press the Signal Type softkey and specify the signal type of the ARINC 429 signal Line A non inverted Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide MIL STD 1553 ARINC 429 Triggering and Serial Decode 27 Line B inverted Differential A B 11 Press the Auto Setup softkey to automatically set these op
191. ecoded 3 Press the Trigger softkey then turn the Entry knob to select Error Trigger Menu Trigger Errors All 4 Press the Errors softkey then select the error type All Errors Header CRC Error cyclic redundancy check error in the header Frame CRC Error cyclic redundancy check error in the frame Because FlexRay errors occur infrequently it may be helpful to set the oscilloscope to press the Mode Coupling key then press the Mode softkey to set the trigger mode from Auto to Normal This prevents the oscilloscope from Auto triggering while waiting for an error to occur You may need to adjust trigger holdoff to see a particular error when multiple errors exist Triggering on FlexRay Events 1 Press Trigger 362 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide FlexRay Triggering and Serial Decode 24 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select the serial bus Serial 1 or Serial 2 on which the FlexRay signals are being decoded 3 Press the Trigger softkey then turn the Entry knob to select Event Frame ID Auto Setup All for Event 4 Press the Event softkey then select the event type Wake up TSS Transmission Start Sequence BSS ByteStart Sequence FES DTS Frame End or Dynamic Trailing Sequence 5 Press Auto Setup for Event This automatically configures oscilloscope settings as shown on the display for the selected event trigg
192. ed warranties of merchant ability and fitness for a particular pur pose Keysight shall not be liable for errors or for incidental or consequential damages in connection with the fur nishing use or performance of this document or of any information con tained herein Should Keysight and the user have a separate written agreement with warranty terms covering the mate rial in this document that conflict with these terms the warranty terms in the separate agreement shall control Technology Licenses The hardware and or software described in this document are furnished under a license and may be used or copied only in accor dance with the terms of such license Restricted Rights Legend If software is for use in the performance of a U S Government prime contract or subcon tract Software is delivered and licensed as Commercial computer software as defined in DFAR 252 227 7014 June 1995 or as a commercial item as defined in FAR 2 101 a or as Restricted computer software as defined in FAR 52 227 19 June 1987 or any equivalent agency regulation or contract clause Use duplication or disclosure of Software is subject to Keysight Technologies standard commercial license terms and non DOD Departments and Agencies of the U S Government will receive no greater than Restricted Rights as defined in FAR 52 227 19 c 1 2 June 1987 U S Govern ment users will receive no greater than Lim ited Rights as
193. el Number of failures acquisitions in which the signal excursion went beyond the mask Failure rate percentage of failures Sigma the ratio of process sigma to maximum achievable sigma based on number of waveforms tested 242 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Mask Testing 15 Reset Statistics Note that statistics are also reset when Mask Test is switched on after being switched off Clear Mask softkey is pressed An Automask is created Additionally the accumulated time counter is reset whenever the oscilloscope is run after the acquisition was stopped Transparent Enable Transparent mode to write measurement values and statistics on the screen without a background Disable the Transparent mode to show them with a gray background The Transparent setting affects mask test statistics measurement statistics and reference waveform information display Clear Display Clears acquisition data from the oscilloscope display To manually modify a mask file You can manually modify a mask file that you created using the Automask function 1 Follow the steps 1 7 in To create a mask from a golden waveform Automask on page 237 Do not clear the mask after creating it ao oO FF OON Attach a USB mass storage device to the oscilloscope Press the Save Recall key Press the Save softkey Press the Format softkey and select Mask P
194. emory depth At that rate memory is filled in 1 ms If the acquisition time is 100 ms 10 ms div only 1 of every 100 samples is needed to fill memory To select the acquisition mode 1 Press the Acquire key on the front panel 2 Inthe Acquire Menu press the Acq Mode softkey then turn the Entry knob to select the acquisition mode The InfiniiVision oscilloscopes have the following acquisition modes Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 187 12 Acquisition Control Normal at slower time div settings normal decimation occurs and there is no averaging Use this mode for most waveforms See Normal Acquisition Mode on page 188 Peak Detect at slower time div settings the maximum and minimum samples in the effective sample period are stored Use this mode for displaying narrow pulses that occur infrequently See Peak Detect Acquisition Mode on page 188 Averaging at all time div settings the specified number of triggers are averaged together Use this mode for reducing noise and increasing resolution of periodic signals without bandwidth or rise time degradation See Averaging Acquisition Mode on page 191 High Resolution at slower time div settings all samples in the effective sample period are averaged and the average value is stored Use this mode for reducing random noise See High Resolution Acquisition Mode on page 193 Normal Acquisition Mode In Norm
195. en the acquisition is stopped and the segmented memory feature is on or when the serial Lister is enabled You can also turn on infinite persistence in the Display Menu and press the Analyze Segments softkey to create an infinite persistence display Segmented Memory Re Arm Time After each segment fills the oscilloscope re arms and is ready to trigger in about 1 us Remember though for example if the horizontal time per division control is set to 5 us div and the Time Reference is set to Center it will take at least 50 us to fill all ten divisions and re arm That is 25 us to capture pre trigger data and 25 us to capture post trigger data 2 0005 500 08 Channels Segment Sof 25 4 00058ms Segment 1 Time 2010 10 22 16 06 23 Segmented Current Seg of Segs 5 25 Saving Data from Segmented Memory You can save either the currently displayed segment Save Segment Current or all segments Save Segment All in the following data formats CSV ASCII XY and BIN 196 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Acquisition Control 12 Be sure to set the Length control to capture enough points to accurately represent the captured data When the oscilloscope is busy saving multiple segments progress is displayed in the upper right area of the display For more information see To save CSV ASCII XY or BIN data files on page 272 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Gu
196. end chart logic bus timing and chart logic bus state Reference waveform locations 2 for comparing with other channel or math waveforms Many built in measurements and a measurement statistics display Built in license enabled waveform generator with arbitrary sine Square ramp pulse DC noise sine cardinal exponential rise exponential fall cardiac and gaussian pulse USB ports make printing saving and sharing data easy Optional LAN VGA module for connecting to a network and displaying the screen on a different monitor Optional GPIB module A Quick Help system is built into the oscilloscope Press and hold any key to display Quick Help Complete instructions for using the quick help system are given in Access the Built In Quick Help on page 44 For more information about InfiniiVision oscilloscopes see www keysight com find scope Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide In This Guide This guide shows how to use the InfiniiVision 3000 X Series oscilloscopes When unpacking and using the oscilloscope for the first time see Chapter 1 Getting Started starting on page 23 When displaying waveforms and acquired data see Chapter 2 Horizontal Controls starting on page 47 Chapter 3 Vertical Controls starting on page 61 Chapter 4 Math Waveforms starting on page 71 Chapter 5 Reference Waveforms starting on page 99 Chapter 6
197. enter of the display See To choose expand about center or ground on page 293 To adjust the vertical position 1 Turn the small vertical position knob to move the channel s waveform up or down on the display The voltage value momentarily displayed in the upper right portion of the display represents the voltage difference between the vertical center of the display and the ground level icon It also represents the voltage at the vertical center of the display if vertical expansion is set to expand about ground see To choose expand about center or ground on page 293 To specify channel coupling Coupling changes the channel s input coupling to either AC alternating current or DC direct current TIP If the channel is DC coupled you can quickly measure the DC component of the signal by simply noting its distance from the ground symbol If the channel is AC coupled the DC component of the signal is removed allowing you to use greater sensitivity to display the AC component of the signal Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 63 3 Vertical Controls 1 Press the desired channel key 2 Inthe Channel Menu press the Coupling softkey to select the input channel coupling DC DC coupling is useful for viewing waveforms as low as 0 Hz that do not have large DC offsets AC AC coupling is useful for viewing waveforms with large DC offsets When AC coupling is chosen you
198. er FlexRay Serial Decode To set up the oscilloscope to capture FlexRay signals see Setup for FlexRay Signals on page 359 For FlexRay triggering setup see FlexRay Triggering on page 360 To set up FlexRay serial decode 1 Press Serial to display the Serial Decode Menu FRAMES 0105232 YNC 0105232199 100 0 Mode nals Reset FlexRay Counters 2 Ifthe decode line does not appear on the display press the Serial key to turn it on Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 363 24 FlexRay Triggering and Serial Decode 364 See Also 3 Ifthe oscilloscope is stopped press the Run Stop key to acquire and decode data You can use the horizontal Zoom window for easier navigation of the acquired data Interpreting FlexRay Decode on page 364 FlexRay Totalizer on page 365 Interpreting FlexRay Lister Data on page 365 Searching for FlexRay Data in the Lister on page 366 Interpreting FlexRay Decode Channels 01 01 Serial Decode Menu NC 010576 SEE SE ESSA Serial 1 exfay w Counters w Frame type NORM SYNC SUP NULL in blue Frame ID decimal digits in yellow Payload length decimal number of words in green Header CRC hex digits in blue plus red HCRC error message if invalid Cycle number decimal digits in yellow Data bytes HEX digits in white Frame CRC hex digits in blue plus red FCRC error message if invalid Frame coding errors sp
199. er s Guide 331 22 Reference Table 10 Licensed Options Available continued License Description After purchase model number notes DVM Digital Voltmeter Order DSOXDVM Provides provides 3 digit voltage and 5 digit frequency measurements using any analog channel EDK Educator s Kit Order DSOXEDK Provides training signals on the oscilloscope s Demo terminals and a lab guide tutorial for education environments EMBD Embedded Serial Triggering and Analysis 12C SPI Order DSOX3EMBD FLEX FlexRay Triggering and Analysis Order DSOX3FLEX MASK Mask Limit Test Order DSOX3MASK Lets you create a mask and test waveforms to determine whether they comply to the mask mem4M Memory Upgrade Order DSOX3MEMUP It shows the total memory depth 4 Mpts interleaved MSO Mixed Signal Oscilloscope MSO Upgrade a DSO to an MSO Order DSOX3MSO for 500 MHz and lower Adds 16 digital channels You do not have to install any bandwidth models or DSOXPERFMSO for hard ware 1 GHz bandwidth models The digital probe cable kit is supplied with the MSO license PWR Power Measurement and Analysis Order DSOX3PWR You can find the DSOX3PWR Power Measurement Application User s Guide at www keysight com find 3000X Serie s manual or on the Documentation CD SGM Segmented Memory Order DSOX3SGM Lets you capture infrequent or burst signals with great resolution by eliminating the capture of your signal s inactivi
200. erence Waveform Menu press the Ref softkey and turn the Entry knob waveform location to select the desired reference waveform location 3 Press the Clear R1 R2 softkey to clear the reference waveform location Reference waveforms are also cleared by a Factory Default or Secure Erase see Chapter 18 Save Recall Setups Screens Data starting on page 269 To display a reference waveform 1 Press the Ref key to turn on reference waveforms 2 Inthe Reference Waveform Menu press the Ref softkey and turn the Entry knob to select the desired reference waveform location 3 Then press the Ref softkey again to enable disable the reference waveform display 100 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Reference Waveforms 5 1 7605 500 08 1 14 KEYSIGHT TECHNOLOGIE Acquisition Channels Options gt Source One reference waveform can be displayed at a time See Also To display reference waveform information on page 102 To scale and position reference waveforms 1 Make sure the multiplexed scale and position knobs to the right of the Ref key are selected for the reference waveform lf the arrow to the left of the Ref key is not illuminated press the key 2 Turn the upper multiplexed knob to adjust the reference waveform scale 3 Turn the lower multiplexed knob to adjust the reference waveform position To adjust reference waveform skew Once reference waveforms are dis
201. erial Decode on page 374 SPI Serial Decode on page 383 With the DSOX3AUDIO license you can decode 12S Inter IC Sound or Integrated Interchip Sound serial buses See 12S Serial Decode on page 393 With the DSOX3COMP license you can decode many UART Universal Asynchronous Receiver Transmitter protocols including RS232 Recommended Standard 232 See UART RS232 Serial Decode on page 417 With the DSOX3AERO license you can decode MIL STD 1553 and ARINC 429 serial buses See MIL STD 1553 Serial Decode on page 400 ARINC 429 Serial Decode on page 407 To determine whether these licenses are installed on your oscilloscope see To display oscilloscope information on page 300 To order serial decode licenses go to www keysight com and search for the product number for example DSOX3AUTO or contact your local Keysight Technologies representative see www keysight com find contactus Lister is a powerful tool for investigating protocol failures You can use Lister to view large amounts of packet level serial data in a tabular format including time tags and specific decoded values After pressing the Single key you can press the Scroll Lister softkey and then rotate the Entry knob to select an event and press the Zoom to Selection softkey to jump to the event To use the Lister 1 Set up trigger and decode on the serial data signals to be analyzed 2 Press Serial gt Lister 3 Press Displ
202. ernal Keysight Flash locations or to an external USB storage device 1 Press Save Recall gt Save gt Format then turn the Entry knob to select Setup scp 2 Press the softkey in the second position and use the Entry knob to navigate to the save location See To navigate storage locations on page 276 3 Finally press the Press to Save softkey A message indicating whether the save was successful is displayed Setup files have the extension SCP These extensions appear when using the File Explorer see File Explorer on page 291 but they do not appear when using the Recall Menu To save BMP or PNG image files Image files can be saved to an external USB storage device 1 Press Save Recall gt Save gt Format then turn the Entry knob to select 8 bit Bitmap image bmp 24 bit Bitmap image bmp or 24 bit image png 2 Press the softkey in the second position and use the Entry knob to navigate to the save location See To navigate storage locations on page 276 3 Press the Settings softkey In the File Settings Menu you have these softkeys and options Setup Info setup information vertical horizontal trigger acquisition math and display settings is also saved in a separate file with a TXT extension Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 271 18 Save Recall Setups Screens Data Invert Grat the graticule in the image file has a white background inste
203. ersus time display to a volts versus volts display The time base is turned off Channel 1 amplitude is plotted on the X axis and Channel 2 amplitude is plotted on the Y axis You can use XY mode to compare frequency and phase relationships between two signals XY mode can also be used with transducers to display strain versus displacement flow versus pressure volts versus current or voltage versus frequency Use the cursors to make measurements on XY mode waveforms For more information about using XY mode for measurements refer to XY Time Mode on page 51 Roll causes the waveform to move slowly across the screen from right to left It only operates on time base settings of 50 ms div and slower If the current time base setting is faster than the 50 ms div limit it will be set to 50 ms div when Roll mode is entered In Roll mode there is no trigger The fixed reference point on the screen is the right edge of the screen and refers to the current moment in time Events that have occurred are scrolled to the left of the reference point Since there is no trigger no pre trigger information is available If you would like to pause the display in Roll mode press the Single key To clear the display and restart an acquisition in Roll mode press the Single key again Use Roll mode on low frequency waveforms to yield a display much like a strip chart recorder It allows the waveform to roll across the display XY Time Mode
204. ese options Master Out Slave In MOSI Data for searching MOSI data Master In Slave Out MISO Data for searching MISO data 4 Press the Bits softkey to open the SPI Bits Search Menu Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 385 25 2C SPI Triggering and Serial Decode 5 In the SPI Bits Search Menu use the Words softkey to specify the number of words in the data value then use the remaining softkeys to enter the hex digit values For more information on searching data see Searching Lister Data on page 124 For more information on using the Navigate key and controls see Navigating the Time Base on page 58 386 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 26 2S Triggering and Serial Decode Setup for 12S Signals 387 12S Triggering 390 12S Serial Decode 393 I2S triggering and serial decode requires Option AUDIO or the DSOX3AUDIO upgrade Only one 12S serial bus can be decoded at a time l Setup for 12S Signals IS Inter IC Sound or Integrated Interchip Sound signals setup consists of connecting the oscilloscope to the serial clock word select and serial data lines and then specifying the input signal threshold voltage levels To set up the oscilloscope to capture 12S signals 1 Press Label to turn on labels 2 Press Serial 3 Press the Serial softkey turn the Entry k
205. et the Y2 cursor to the top of the signal and set Y1 to the bottom of the signal Note the AY value at the bottom of the display In this example we are using the Y cursors but you could have used the X cursors instead Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Horizontal Controls 2 6 Move the Y1 and Y2 cursors to the intersection of the signal and the Y axis Again note the AY value KEYSIGHT TECHN Channels Cursors 0 0 1 73 Cursors XT ov TE 1 21250V 2 X2 Figure 4 Phase difference measurements automatic and using cursors 7 Calculate the phase difference using the formula below For example if the first AY value is 1 688 and the second AY value is 1 031 second AY _ 1 031 sind first AY 1 688 0 37 65 degrees of phase shift Z Axis Input in XY Display Mode Blanking When you select the XY display mode the time base is turned off Channel 1 is the X axis input channel 2 is the Y axis input and the rear panel EXT TRIG IN is the Z axis input If you only want to see portions of the Y versus X display use the Z axis input Z axis turns the trace on and off analog oscilloscopes called this Z axis blanking because it turned the beam on and off When Z is low lt 1 4 V Y versus X is displayed when Z is high gt 1 4 V the trace is turned off Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 53 2 Horizontal Controls To display the zoomed time ba
206. ettings I O Interface Settings 287 Setting up the Oscilloscope s LAN Connection 288 To establish a LAN connection 289 Stand alone Point to Point Connection toa PC 290 File Explorer 291 Setting Oscilloscope Preferences 293 To choose expand about center or ground 293 To disable enable transparent backgrounds 294 To load the default label library 294 To set up the screensaver 294 To set AutoScale preferences 295 Setting the Oscilloscope s Clock 296 Setting the Rear Panel TRIG OUT Source 296 Performing Service Tasks 297 To perform user calibration 297 To perform hardware self test 300 To perform front panel self test 300 To display oscilloscope information 300 To display the user calibration status 301 To clean the oscilloscope 301 To check warranty and extended services status 301 To contact Keysight 301 To return the instrument 302 Configuring the Quick Action Key 302 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 17 Adding an Annotation 303 21 Web Interface Accessing the Web Interface 308 Browser Web Control 309 Real Scope Remote Front Panel 310 Simple Remote Front Panel 311 Browser Based Remote Front Panel 312 Remote Programming via the Web Interface 313 Remote Programming with Keysight I0 Libraries 314 Save Recall 315 Saving Files via the Web Interface 315 Recalling Files via the Web Interface 316 Get Image 317 Iden
207. ettings 20 File Explorer The File Explorer lets you navigate the oscilloscope s internal file system and the file systems of connected USB storage devices From the internal file system you can load oscilloscope setup files or mask files From a connected USB storage device you can load setup files mask files license files firmware update cab files label files etc Also you can delete files on a connected USB storage device The USB port on the front panel and the USB port on the rear panel labeled HOST are USB Series A receptacles These are the receptacles to which you can connect USB mass storage devices and printers The square receptacle on the rear panel labeled DEVICE is provided for controlling the oscilloscope over USB See the Programmer s Guide for more information The oscilloscope s internal file system under Keysight Flash consists of 10 locations for oscilloscope setup files and four locations for mask files To use the File Explorer 1 Press Utility gt File Explorer 2 Inthe File Explorer Menu press the softkey in the first position and use the Entry knob to navigate Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 291 20 Utility Settings 292 USB Storage Devices 500 05 TECHN Acquisition Channels SMS02000 The softkey in the first position can have these labels Press to go when you can push the Entry knob to navigate to a new folder or stor
208. fN fs FrequenCy me Limiting oscilloscope bandwidth faw to 1 4 the sample rate fs 4 reduces frequency components above the Nyquist frequency fn Figure 35 Sample Rate and Oscilloscope Band width So in practice an oscilloscope s sample rate should be four or more times its bandwidth fs 4fgw This way there is less aliasing and aliased frequency components have a greater amount of attenuation ote that 1 GHz bandwidth 3000 X Series oscilloscope models have more of a brick wall type frequency response also known a flat response than the Gaussian response of lower bandwidth 3000 X Series oscilloscope models To understand the characteristics of each type of oscilloscope frequency response see Understanding Oscilloscope Frequency Response and Its Effect on Rise Time Accuracy Keysight Application Note 1420 http literature cdn keysight com litweb pdf 5988 8008EN pdf SeeAlso Evaluating Oscilloscope Sample Rates vs Sampling Fidelity How to Make the Most Accurate Digital Measurements Keysight Application Note 1587 http literature cdn keysight com litweb pdf 5989 5732EN pd f Oscilloscope Rise Time Closely related to an oscilloscope s bandwidth specification is its rise time specification Oscilloscopes with a Gaussian type frequency response have an approximate rise time of 0 35 fgy based on a 10 to 90 criterion Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 185 12 Acquisition Control
209. for a voltage probe Amps for a current probe Measurement results channel sensitivity and the trigger level will reflect the measurement units you have selected 4 Press the Probe softkey then turn the entry knob to specify the probe attenuation The attenuation factor can be set from 0 1 1 to 1000 1 in a 1 2 5 sequence The probe attenuation factor must be set properly for measurements to be made correctly Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 179 11 Trigger Mode Coupling 180 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control Running Stopping and Making Single Acquisitions Run Control 181 Overview of Sampling 182 Selecting the Acquisition Mode 187 Acquiring to Segmented Memory 193 This chapter shows how to use the oscilloscope s acquisition and run controls Running Stopping and Making Single Acquisitions Run Control There are two front panel keys for starting and stopping the oscilloscope s acquisition system Run Stop and Single KEYSIGHT TECHNOLOGIES When the Run Stop key is green the oscilloscope is running that is acquiring data when trigger conditions are met To stop acquiring data press Run Stop When stopped the last acquired waveform is displayed When the Run Stop key is red data acquisition is stopped Stop is displayed next to t
210. fore performing this procedure 2 Press the rear panel CAL button to disable calibration protection 3 Connect short 12 inch maximum equal length cables to each analog channel s BNC connector on the front of the oscilloscope You will need two equal length cables for a 2 channel oscilloscope or four equal length cables for a 4 channel oscilloscope Use 50W RG58AU or equivalent BNC cables when performing User Cal For a 2 channel oscilloscope connect a BNC tee to the equal length cables Then connect a BNC f to BNC f also called a barrel connector to the tee as shown below Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Utility Settings 20 Long cable to TRIG OUT To Channel 1 AF To Channel 2 Figure 49 User Calibration cable for 2 channel oscilloscope For a 4 channel oscilloscope connect BNC tees to the equal length cables as shown below Then connect a BNC f to BNC f barrel connector to the tee as shown below To Channel 1 Ba Figs Maes Ny To Channel 4 Longer cable to TRIG OUT Figure 50 User Calibration cable for 4 channel oscilloscope Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 299 20 Utility Settings 4 Connect a BNC cable 40 inches maximum from the TRIG OUT connector on the rear panel to the BNC barrel connector 5 Press the Utility key then press the Service softkey 6 Begin the Self Cal by pressing the Start User Cal softkey To perform h
211. from All or 1 to 2047 6 Press the Frame Type softkey to select the frame type All Frames Startup Frames NULL Frames Sync Frames Normal Frames NOT Startup Frames NOT NULL Frames NOT Sync Frames 7 Press the Cyc Ct Rep softkey and use the Entry knob to select the cycle count repetition factor 2 4 8 16 32 or 64 or All 8 Press the Cyc Ct Bas softkey and use the Entry knob to select the cycle count base factor from O through the Cyc Ct Rep factor minus 1 For example with a base factor of 1 and a repetition factor of 16 the oscilloscope triggers on cycles 1 17 33 49 and 65 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 361 24 FlexRay Triggering and Serial Decode To trigger on a particular cycle set the Cycle Repetition factor to 64 and use the cycle base factor to choose a cycle To trigger all any cycles set the Cycle Repetition factor to All The scope will trigger on any and all cycles Because specific FlexRay frames may occur infrequently it may be helpful to press the Mode Coupling key then press the Mode softkey to set the trigger mode from Auto to Normal This prevents the oscilloscope from Auto triggering while waiting for a particular frame and cycle combination Triggering on FlexRay Errors 1 Press Trigger 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select the serial bus Serial 1 or Serial 2 on which the FlexRay signals are being d
212. fset 0 0Vs lt d Offset 0 0v KEYSIGHT TEC Acquisition i al Channels 10 0 1 1 00 1 Offset 7 020uVs aD Offset ral on page 72 80 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Math Waveforms 4 FFT Measurement FFT is used to compute the fast Fourier transform using analog input channels or an arithmetic operation g t FFT takes the digitized time record of the specified source and transforms it to the frequency domain When the FFT function is selected the FFT spectrum is plotted on the oscilloscope display as magnitude in dBV versus frequency The readout for the horizontal axis changes from time to frequency Hertz and the vertical readout changes from volts to dB Use the FFT function to find crosstalk problems to find distortion problems in analog waveforms caused by amplifier non linearity or for adjusting analog filters To display a FFT waveform 1 Press the Math key press the Function softkey and select f t press the Operator softkey and select FFT tit FFT Ch1 FFT Resolution 63H2z Function Operator Source 1 Center tit FFT 500kHz 250kH2 Source 1 selects the source for the FFT See To perform transforms or filters on an arithmetic operation on page 72 for information about using g t as the source Span sets the overall width of the FFT spectrum that you see on the display left to right Divide span by 10 to calculate the number of Hertz per divi
213. fter the start address read write acknowledge and data events occur If don t care is selected OxXX or OxXXX for the address the address will be ignored The trigger will always occur on the 17th clock for 7 bit addressing or 26th clock for 10 bit addressing b Press the Data value softkey and turn the Entry knob to select the 8 bit data pattern on which to trigger You can select a data value in the range of 0x00 to OxFF hexadecimal The oscilloscope will trigger after the start address read write acknowledge and data events occur If don t care OxXX is selected for data the data will be ignored The trigger will always occur on the 17th clock for 7 bit addressing or 26th clock for 10 bit addressing c lf you have selected a three byte trigger press the Data2 value softkey and turn the Entry knob to select the 8 bit data pattern on which to trigger To display I2C serial decode see 12C Serial Decode on page 374 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 373 25 2C SPI Triggering and Serial Decode I2C Serial Decode See Also 374 To set up the oscilloscope to capture 12C signals see Setup for I2C Signals on page 369 For I2C triggering setup see I2C Triggering on page 370 To set up 12C serial decode 1 Press Serial to display the Serial Decode Menu Serial Decode Menu Serial Mode Signals Addr Size re 7B Serial 1 Bit w 2 Choose 7 bit or 8 bit address si
214. ge 57 is similar to setting up triggers In fact except for Serial events you can copy search setups to trigger setups and vice versa see To copy search setups on page 57 Searches are different than triggers in that they use the measurement threshold settings instead of trigger levels Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Horizontal Controls 2 Found search events are marked with white triangles at the top of the graticule and the number of events found is displayed in the menu line just above the sofkey labels To set up searches 1 Press Search 2 Inthe Search Menu press Search then turn the Entry knob to select the search type 3 Press Settings and use the Search Settings Menu to set up the selected search type Setting up searches is similar to setting up triggers For setting up Edge searches see Edge Trigger on page 142 For setting up Pulse Width searches see Pulse Width Trigger on page 145 For setting up Rise Fall Time searches see Rise Fall Time Trigger on page 152 For setting up Runt searches see Runt Trigger on page 155 For setting up Serial searches see Chapter 10 Triggers starting on page 139 and Searching Lister Data on page 124 Remember that searches use the measurement threshold settings instead of trigger levels Use the Thresholds softkey in the Search Menu to access the Measurement Threshold Menu See Measurement Thre
215. gh frequency model of the probe input is more useful see the following figure This high frequency model takes into account pure tip capacitance to ground as well as series tip resistance and the cable s characteristic impedance Z 8 5 pF 100 kQ Figure 21 DC and Low Frequency Probe Equivalent Circuit 250 Q 7 5 pF 1 pF 100 kQ 150 Q Figure 22 High Frequency Probe Equivalent Circuit Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 115 6 116 Digital Channels The impedance plots for the two models are shown in these figures By comparing the two plots you can see that both the series tip resistor and the cable s characteristic impedance extend the input impedance significantly The stray tip capacitance which is generally small 1 pF sets the final break point on the impedance chart 100 k 10k High Frequency Model oO Z 1k Typical 100 Model 10 1 10kHz 100kHz 1MHz 10MHz 100 MHz 1 GHz Frequency Figure 23 Impedance versus Frequency for Both Probe Circuit Models The logic probes are represented by the high frequency circuit model shown above They are designed to provide as much series tip resistance as possible Stray tip capacitance to ground is minimized by the proper mechanical design of the probe tip assembly This provides the maximum input impedance at high frequencies Probe Grounding A probe ground is the low impedance path for current to return t
216. gnment The channel does not have to be turned on to have a label assigned to it If the channel is turned on its current label will be highlighted 3 Press the Spell softkey then turn the Entry knob to select the first character in the new label Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 135 9 Labels Label Assignment Auto Increment Turning the Entry knob selects a character to enter into the highlighted position shown in the New label line above the softkeys and in the Spell softkey Labels can be up to ten characters in length 4 Press the Enter softkey to enter the selected character and to go to the next character position 5 You may position the highlight on any character in the label name by successively pressing the Enter softkey 6 To delete a character from the label press the Enter softkey until the letter you want to delete is highlighted then press the Delete Character softkey You can use a connected USB keyboard instead of using the Spell and other character editing softkeys 7 When you are done entering characters for the label press the Apply New Label softkey to assign the label to the selected channel When you define a new label it is added to the nonvolatile label list When you assign a label ending in a digit such as ADDRO or DATAO the oscilloscope automatically increments the digit and displays the modified label in the New label field after you press the App
217. gth HCRC header CRC CYC cycle n Data umber FCRC frame CRC Frames with e Searching for FlexRay The oscilloscope rrors are highlighted in red Data in the Lister s search capability lets you search for and mark certain types of FlexRay data in the Lister You can use the Navigate key and controls to navigate through the marked rows 1 With FlexRay selected as the serial decode mode press Search 2 Inthe Search Menu press the Search softkey then turn the Entry knob to select the serial bus decoded 3 Inthe Search Serial 1 or Serial 2 on which the FlexRay signals are being enu press Search for then select from these options Frame ID Finds frames with the specified ID Press the Frame ID softkey to select the ID Cycle number Frame ID Finds frames with the specified cycle number and ID Press th e Frame ID softkey to select the ID Press the Cycle softkey to select the number Data Frame ID Cycle number Finds frames with the specified data cycle number and frame ID Press the Frame ID softkey to select the ID Press the Cycle softkey to select the number Press the Data sofkey to open a menu where you can enter the data value Header CRC Error Finds cyclic redundancy check errors in headers Frame CRC Error Finds cyclic redundancy check errors in frames Errors finds all errors For more information on searching data se
218. he Serial key to turn it on 4 Ifthe oscilloscope is stopped press the Run Stop key to acquire and decode data If the setup does not produce a stable trigger the LIN signal may be slow enough that the oscilloscope is AutoTriggering Press the Mode Coupling key then press the Mode softkey to set the trigger mode from Auto to Normal You can use the horizontal Zoom window for easier navigation of the decoded data See Also Interpreting LIN Decode on page 356 Interpreting LIN Lister Data on page 357 Searching for LIN Data in the Lister on page 358 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 355 23 CAN LIN Triggering and Serial Decode 356 Interpreting LIN Decode 1 0008 Trig d 1 41 KEYSIGHT TECHNOLOGIES Channels 01 oo Signals Show Parity Lister v Angled waveforms show an active bus inside a packet frame Mid level blue lines show an idle bus LIN 1 3 only The hexadecimal ID and parity bits if enabled appear in yellow If a parity error is detected the hexadecimal ID and parity bits if enabled appear in red Decoded hexadecimal data values appear in white For LIN 1 3 the checksum appears in blue if correct or red if incorrect The checksum always appears in white for LIN 2 0 Decoded text is truncated at the end of the associated frame when there is insufficient space within frame boundaries Pink vertical bars indicate you need to expand the horizontal
219. he following actions Sets the selected source channel s impedance to 50 Ohms assuming a differential active probe that requires a 50 ohm termination is being used Sets the selected source channel s probe attenuation to 10 1 Sets the trigger level on the selected source channel to 300 mV Turns on trigger Noise Reject Turns on Serial Decode Sets the trigger type to FlexRay FlexRay Triggering 360 To set up the oscilloscope to capture a FlexRay signal see Setup for FlexRay Signals on page 359 After you have set up the oscilloscope to capture a FlexRay signal you can then set up triggers on frames see page 361 errors see page 362 or events see page 362 To display FlexRay serial decode see FlexRay Serial Decode on page 363 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide FlexRay Triggering and Serial Decode 24 Triggering on FlexRay Frames _ Press Trigger 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select the serial bus Serial 1 or Serial 2 on which the FlexRay signals are being decoded Trigger Menu Trigger Trigger Frames S FlexRay Frame ae 3 Press the Trigger softkey then turn the Entry knob to select Frame 4 Press the Frames softkey to open the FlexRay Frame Trigger Menu Cyc Ct Bas Frame Type Frame ID 1 Syne 1 ol Press the Frame ID softkey and use the Entry knob to select the frame ID value
220. he horizontal scale and delay settings before the last Zoom to Selection 8 Press the Options softkey to open the Lister Options Menu In this menu you can Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 123 7 Serial Decode Enable or disable the Track Time option When enabled as you select different Lister rows using the Entry knob while acquisitions are stopped the horizontal delay changes to the Time of the selected row Also changing the horizontal delay will scroll the Lister Press the Scroll Lister softkey and use the Entry knob to scroll though data rows in the Lister display Press the Time Ref softkey and use the Entry knob to select whether the Time column in the Lister display shows times relative to the trigger or relative to the previous packet row Searching Lister Data When serial decode is enabled you can use the Search key to find and place marks on rows in the Lister The Search softkey lets you specify events to find It is similar to specifying protocol triggers Events found are marked in orange in the far left Lister column The total number of events found is displayed above the softkeys 548 0 1 0003 Errors Channels Search for Tx Data 124 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Serial Decode 7 Each serial decode option lets you find protocol specific headers data errors etc See Searching for ARINC
221. he knob indicate that this control has the effect of spreading out or zooming in on the waveform using the horizontal scale Horizontal position knob Turn the knob marked lt to pan through the waveform data horizontally You can see the captured waveform before the trigger turn the knob clockwise or after the trigger turn the knob counterclockwise If you pan through the waveform when the oscilloscope is stopped not in Run mode then you are looking at the waveform data from the last acquisition taken Horiz key Press this key to open the Horizontal Menu where you can select XY and Roll modes enable or disable Zoom enable or disable horizontal time division fine adjustment and select the trigger time reference point Zoom key Press the zoom key to split the oscilloscope display into Normal and Zoom sections without opening the Horizontal Menu Search key Lets you search for events in the acquired data Navigate keys Press this key to navigate through captured data Time search events or segmented memory acquisitions See Navigating the Time Base on page 58 For more information see Chapter 2 Horizontal Controls starting on page 47 Run Control keys When the Run Stop key is green the oscilloscope is running that is acquiring data when trigger conditions are met To stop acquiring data press Run Stop When the Run Stop key is red data acquisition is stopped
222. he oscilloscope stops If the signal you are measuring has more than about 1 s of inactivity consider selecting the Normal trigger mode to prevent AutoTriggering See To select the Auto or Normal trigger mode on page 174 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Acquisition Control 12 2 000 vent 1 Channels Progress indicator Sample rate Segment lof 11 0 0s Segmented Current Seg of Segs 1 1000 See Also Navigating Segments on page 195 Measurements Statistics and Infinite Persistence with Segmented Memory on page 195 Segmented Memory Re Arm Time on page 196 Saving Data from Segmented Memory on page 196 Navigating Segments 1 Press the Current Seg softkey and turn the Entry knob to display the desired segment along with a time tag indicating the time from the first trigger event You can also navigate segments using the Navigate key and controls See To navigate segments on page 59 Measurements Statistics and Infinite Persistence with Segmented Memory To perform measurements and view statistical information press Meas and set up your desired measurements see Chapter 14 Measurements starting on page 209 Then press Analyze Segments Statistical data will be accumulated for the measurements you have chosen Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 195 12 Acquisition Control The Analyze Segments softkey appears wh
223. he trigger type in the status line at the top of the display To start acquiring data press Run Stop To capture and display a single acquisition whether the oscilloscope is running or stopped press Single The Single run control lets you view single shot events without subsequent waveform data overwriting the display Use Single when you want maximum memory depth for pan and zoom 181 12 Acquisition Control When you press Single the display is cleared the trigger mode is temporarily set to Normal to keep the oscilloscope from auto triggering immediately the trigger circuitry is armed the Single key is illuminated and the oscilloscope waits until a trigger condition occurs before it displays a waveform When the oscilloscope triggers the single acquisition is displayed and the oscilloscope is stopped the Run Stop key is illuminated in red Press Single again to acquire another waveform If the oscilloscope doesn t trigger you can press the Force Trigger key to trigger on anything and make a single acquisition To display the results of multiple acquisitions use persistence See To set or clear persistence on page 129 Single vs Running The maximum data record length is greater for a single acquisition than when the and Record Length oscilloscope is running or when the oscilloscope is stopped after running Single Single acquisitions always use the maximum memory available at
224. ht InfinilVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference Specifications and Characteristics 323 Measurement Category 323 Environmental Conditions 325 Probes and Accessories 326 Loading Licenses and Displaying License Information 331 Software and Firmware Updates 333 Binary Data bin Format 333 CSV and ASCII XY files 340 Acknowledgements 342 Specifications and Characteristics Please see the InfiniiVision oscilloscope data sheets for complete up to date specifications and characteristics To download a data sheet please visit www keysight com find 3000X Series Then select the Library tab followed by Specifications Or go to the Keysight home page at www keysight com and search for 3000 X Series oscilloscopes data sheet To order a data sheet by phone please contact your local Keysight office The complete list is available at www keysight com find contactus Measurement Category Oscilloscope Measurement Category on page 324 Measurement Category Definitions on page 324 Transient Withstand Capability on page 325 KEYSIGHT TECHNOLOGIES 323 22 Reference Oscilloscope Measurement Category The InfiniiVision oscilloscopes are intended to be used for measurements in Measurement Category WARNING Use this instrument only for measurements within its specified measurement category Me
225. iVision 3000 X Series Oscilloscopes User s Guide 171 10 Triggers Serial Trigger If you have connected the oscilloscope s analog source channels to the D and D signals Adjust the trigger level for each connected analog channel to the middle of the waveform by pressing the D or D softkey then turning the Trigger Level knob If you have connected the oscilloscope s digital source channels to the D and D signals this applies to MSO model oscilloscopes only Press the Digital key and select Thresholds to set an appropriate threshold level for the digital channels The value of the trigger level or digital threshold is displayed in the upper right corner of the display With serial decode option licenses see Serial Decode Options on page 121 you can enable serial trigger types To set up these triggers see 172 ARINC 429 Triggering on page 405 CAN Triggering on page 345 FlexRay Triggering on page 360 I2C Triggering on page 370 I2S Triggering on page 390 LIN Triggering on page 353 MIL STD 1553 Triggering on page 399 SPI Triggering on page 381 UART RS232 Triggering on page 415 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling To select the Auto or Normal trigger mode 174 To select the trigger coupling 175 To enable or disable trigger noise rejection 177 To enabl
226. iVision 3000 X Series Oscilloscopes User s Guide 211 14 Measurements Measurement Valid for Notes Digital Channels Amplitude on page 216 Area on page 230 Average on page 219 Yes Full Screen Base on page 217 Burst Width on page 225 Counter on page 224 Yes Not valid for math waveforms Delay on page 226 Measures between two sources Press Settings to specify the second source Duty Cycle on page 225 Yes Fall Time on page 226 Frequency on page 223 Yes Maximum on page 216 Yes Minimum on page 216 Yes Rising Edge Count on page 230 Falling Edges Count on page 230 Positive Pulse Count on page 229 Negative Pulse Count on page 230 Overshoot on page 217 Peak Peak on page 216 Yes Period on page 223 Yes 212 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Power App Measurements Dual Channel N2820A Probe Measurements Measurements 14 Measurement Valid for Notes Digital Channels Phase on page 227 Measures between two sources Press Settings to specify the second source Preshoot on page 219 Ratio on page 222 Measures between two sources Press Settings to specify the second source Rise Time on page 226 DC RMS on page 220 AC RMS on page 220 Top on page 216
227. ics Menu Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 233 14 Measurements 234 ment Display On lenu Reset Transparent Settings Clear Meas Acquisition Channels Measurements Freq 1 49 Pk Pk 1 Max Count oo Relative o The following statistics are shown Name of the measurement current measured value mean minimum measured value maximum measured value standard deviation and the number of times the measurement has been made count Statistics are based on the total number of captured waveforms count The standard deviation shown in Statistics is calculated using the same formula used for calculating the standard deviation measurement The formula is shown in the section titled AC RMS on page 220 The source channel of the measurement is shown in parenthesis after the measurement name For example Freq 1 indicates a frequency measurement on channel 1 You can switch the statistics Display On or Display Off Statistics continue to accumulate even when the statistics display is off When you navigate away from the Measurements Menu the statistics display will no longer be displayed but statistical data collection will continue Return to the Measurements Menu to see the data again Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Measurements 14 To reset the statistics measurements press the Reset Statistics softkey This resets
228. ide 197 12 Acquisition Control 198 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors To make cursor measurements 200 Cursor Examples 203 Cursors are horizontal and vertical markers that indicate X axis values and Y axis values on a selected waveform source You can use cursors to make custom voltage time phase or ratio measurements on oscilloscope signals Cursor information is displayed in the right side information area Cursors are not always limited to the visible display If you set a cursor then pan and zoom the waveform until the cursor is off screen its value will not be changed It will still be there when you return to its original location X Cursors X cursors are vertical dashed lines that adjust horizontally and can be used to measure time s frequency 1 s phase and ratio The X1 cursor is the short dashed vertical line and the X2 cursor is the long dashed vertical line When used with the FFT math function as a source the X cursors indicate frequency In XY horizontal mode the X cursors display channel 1 values Volts or Amps The X1 and X2 cursor values for the selected waveform source are displayed in the softkey menu area The difference between X1 and X2 AX and 1 AX are displayed in the Cursors box in the right side information area Y Cursors Y cursors are horizontal dashed lines that
229. iew Quick Help right click on a softkey Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 311 21 Web Interface i VNC Agilent MSOX2024A InfiniiVision US50210029 Remote Front Panel Main Menu _ z gt Function Keys Hint appears lt when you roll over with mouse Softkeys left click to select right click for Quick Help Scrollingand When using a monitor resolution of 800 x 600 or less on the remote computer you Monitor need to scroll to access the full remote front panel To display the remote front Resolution panel without scroll bars use a monitor resolution greater than 800 x 600 on your computer s display Browser Based Remote Front Panel To operate the oscilloscope using the web interface s Browser Based Remote Front Panel 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 308 2 When the oscilloscope s web interface is displayed select Browser Web Control then select Browser Based Remote Front Panel After a few seconds the Remote Front Panel appears 3 Click keys or knobs that you would normally press on the oscilloscope s front panel Buttons have been added for turning knobs 312 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Web Interface 21 Ms0 x MA MSO X 2024A Fun Control YE Agilent Technologies InfiniiVision MSO X 2024A noe 0 L J Single Remote Programming via the Web
230. ified data byte applies to the least significant 8 bits excludes the 9th alert bit Rx 0 Data For use when the DUT data words are 9 bits in length including the alert bit the 9th bit Finds only when the 9th alert bit is 0 The specified data byte applies to the least significant 8 bits excludes the 9th alert bit Rx X Data For use when the DUT data words are 9 bits in length including the alert bit the 9th bit Finds a data byte that you specify regardless of the value of the 9th alert bit The specified data byte applies to the least significant 8 bits excludes the 9th alert bit Similar choices are available for Tx Rx or Tx Parity Error Finds a parity error based on the parity you have set in the Bus Configuration Menu Rx or Tx Any Error Finds any error For more information on searching data see Searching Lister Data on page 124 For more information on using the Navigate key and controls see Navigating the Time Base on page 58 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 42 28 UART RS232 Triggering and Serial Decode 422 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Index Symbols Width measurement 225 Width measurement 225 Numerics 1 M ohm input impedance 64 50 ohm input impedance 64 A About Oscilloscope 300 absolute value math function 90 AC channel coupling 64 AC RMS Full Screen measurement
231. ifying the attenuation factor in decibels you can select values from 20 dB to 60 dB If Amps is chosen as the units and a manual attenuation factor is chosen then the units as well as the attenuation factor are displayed above the Probe softkey Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 67 3 68 Vertical Controls Channel 1 Probe Menu 0 005 4 200 1 Units Probe Amps 0 005V74 To specify the probe skew When measuring time intervals in the nanoseconds ns range small differences in cable length can affect the measurement Use Skew to remove cable delay errors between any two channels 1 Probe the same point with both probes 2 Press one of the probes associated channel key 3 Inthe Channel Menu press Probe 4 Inthe Channel Probe Menu press Skew then select the desired skew value Each analog channel can be adjusted 100 ns in 10 ps increments for a total of 200 ns difference The skew setting is not affected by pressing Default Setup or Auto Scale To calibrate a probe The Calibrate Probe softkey guides you through the process of calibrating probes For certain active probes such as InfiniiMax probes the oscilloscope can accurately calibrate its analog channels for the probe When you connect a probe that can be calibrated the Calibrate Probe softkey in the Channel Probe Menu becomes active To calibrate one of these probes 1 First plug your probe into one of the oscillo
232. iggers on color fields 1 or 3 When Field 2 is selected the oscilloscope triggers on color fields 2 or 4 1 Press the Trigger key 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select Video 3 Press the Settings softkey then press the Standard softkey to select the appropriate TV standard 4 Press the Mode softkey and select Field1 or Field2 The trigger circuits look for the position of the start of Vertical Sync to determine the field But this definition of field does not take into consideration the phase of the reference subcarrier When Field 1 is selected the trigger system will find any field where the vertical sync starts on Line 4 In the case of NTSC video the Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 167 10 Triggers oscilloscope will trigger on color field 1 alternating with color field 3 see the following figure This setup can be used to measure the envelope of the reference burst 424 0 500 08 Channels Video Trigger Menu Standard Mode Field Holdoff NTSC Field1 O 0flds Figure 31 Triggering on Color Field 1 Alternating with Color Field 3 If a more detailed analysis is required then only one color field should be selected to be the trigger You can do this by using the Field Holdoff softkey in the Video Trigger Menu Press the Field Holdoff softkey and use the Entry knob to adjust the holdoff in half field increments until the osc
233. illoscope register its hostname and use the DNS server for name resolution 5 You can enable the Multicast DNS option to let the oscilloscope use Multicast DNS for name resolution on small networks without a conventional DNS server 6 Connect the oscilloscope to the local area network LAN by inserting the LAN cable into the LAN port on the rear panel of the oscilloscope In a few moments the oscilloscope will connect to the network automatically If the oscilloscope does not automatically connect to the network press Utility gt 1 0 gt LAN Reset In a few moments the oscilloscope will connect to the network Manual 1 Get the oscilloscope s network parameters host name IP address subnet Configuration mask gateway IP DNS IP etc from your network administrator 2 Press Utility gt 1 0 3 Press the LAN Settings softkey 4 Press the Config softkey then turn the Entry knob to select Automatic and press the softkey again to disable it If Automatic is not enabled the oscilloscope s LAN configuration must be set up manually using the Addresses and Host name softkeys Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 289 20 Utility Settings 290 5 Configure the oscilloscope s LAN interface a Press the Addresses softkey b Use the Modify softkey and the other softkeys and the Entry knob to enter the IP Address Subnet Mask Gateway IP and DNS IP values When you are done go back up in
234. illoscope triggers on only one phase of the color burst A quick way to synchronize to the other phase is to briefly disconnect the signal and then reconnect it Repeat until the correct phase is displayed When holdoff is adjusted using the Field Holdoff softkey and the Entry knob the corresponding holdoff time will be displayed in the Trigger Mode and Coupling Menu 168 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Triggers 10 Table 4 Half field holdoff time Standard Time NTSC 8 35 ms PAL 10 ms PAL M 10 ms SECAM 10 ms Generic 8 35 ms EDTV 480p 60 8 35 ms EDTV 567p 50 10 ms HDTV 720p 50 10 ms HDTV 720p 60 8 35 ms HDTV 1080p 24 20 835 ms HDTV 1080p 25 20 ms HDTV 1080p 30 20 ms HDTV 1080p 50 16 67 ms HDTV 1080p 60 8 36 ms HDTV 1080i 50 10 ms HDTV 1080i 60 8 35 ms Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 169 10 Triggers USB Trigger 170 Channels Video Trigger Menu Mode Field Holdoff Field1 3 0000flds Figure 32 Using Field Holdoff to Synchronize to Color Field 1 or 3 Field 1 mode USB trigger will trigger on a Start of Packet SOP End of Packet EOP signal Reset Complete RC Enter Suspend Suspend or Exit Suspend Exit Sus on the differential USB data lines D and D USB Low Speed and Full Speed are supported by this trigger End of packet trigger on Buse OOK IOC IY Lt SOP SEO
235. ime With Keysight s MegaZoom technology you can press Stop then pan and zoom through the data to find where the waveform repeats Measure this time using cursors then set the holdoff External Trigger Input 178 The external trigger input can be used as a source in several of the trigger types The external trigger BNC input is on the rear panel and is labeled EXT TRIG IN Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Trigger Mode Coupling 11 CAUTION A Maximum voltage at oscilloscope external trigger input CAT 300 Vrms 400 Vpk transient overvoltage 1 6 kVpk 1 M ohm input For steady state sinusoidal waveforms derate at 20 dB decade above 57 kHz to a minimum of 5 Vpk With N2863A 10 1 probe CAT 600 V DC peak AC With 10073C or 10074C 10 1 probe CAT 500 Vpk The external trigger input impedance is 1M Ohm This lets you use passive probes for general purpose measurements The higher impedance minimizes the loading effect of the oscilloscope on the device under test To set the EXT TRIG IN units and probe attenuation 1 Press the Mode Coupling key in the Trigger section of the front panel Trigger Mode and Coupling Menu Mode Coupling Noise Rej HF Reject Holdoff External DE 40 000ns Era Auto 2 Inthe Trigger Mode and Coupling Menu press the External softkey External Trigger Menu 1 00 1 3 In the External Trigger Menu press the Units softkey to select between Volts
236. inary format and provides data headers that describe that data Because the data is in binary format the size of the file is approximately 5 times smaller than the ASCII XY format Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 333 22 Reference BIN data using 334 segmented memory If more than one source is on all displayed sources will be saved except math functions When using segmented memory each segment is treated as a separate waveform All segments for a channel are saved then all segments of the next higher numbered channel are saved This continues until all displayed channels are saved When the oscilloscope is in the Peak Detect acquisition mode the minimum and maximum value waveform data points are saved to the file in separate waveform buffers The minimum value data points are saved first then the maximum value data points are saved When saving all segments each segment has its own waveform header see Binary Header Format on page 335 In BIN file format data are presented as follows Channel 1 data all segments Channel 2 data all segments Channel 3 data all segments Channel 4 data all segments Digital channel data all segments Math waveform data all segments When not saving all segments the number of waveforms is equivalent to the number of active channels including math and digital channels with up to seven waveforms for each digital pod When sav
237. ing all segments the number of waveforms is equal to the number of active channels multiplied by the number of segments acquired Binary Data in MATLAB Binary data from an InfiniiVision oscilloscope can be imported to The MathWorks MATLAB You can download the appropriate MATLAB functions from the Keysight Technologies web site at www keysight com find 3000X Series examples Keysight provides the m files which need to be copied into the work directory for MATLAB The default work directory is C MATLAB work Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Binary Header File Header Waveform Header Format Reference 22 There is only one file header in a binary file The file header consists of the following information Cookie Two byte characters AG that indicate the file is in the Keysight Binary Data file format Version Two bytes that represent the file version File Size A 32 bit integer that is the number of bytes that are in the file Number of A 32 bit integer that is the number of waveforms that are stored in the file Waveforms It is possible to store more than one waveform in the file and each waveform stored will have a waveform header When using segmented memory each segment is treated as a separate waveform The waveform header contains information about the type of waveform data that is stored following the waveform data header
238. iniiVision 3000 X Series Oscilloscopes User s Guide 117 6 Digital Channels digital channels display reconstructed waveforms they do not show ringing and perturbations You will not find ground problems through examination of the waveform display In fact it is likely you will discover the problem through random glitches or inconsistent data measurements Use the analog channels to view ringing and perturbations Best Probing Practices Because of the variables L di and dt you may be unsure how much margin is available in your measurement setup The following are guidelines for good probing practices The ground lead from each digital channel group D15 D8 and D7 D0 should be attached to the ground of the device under test if any channel within the group is being used for data capture When capturing data in a noisy environment every third digital channel probe s ground should be used in addition to the channel group s ground High speed timing measurements rise time lt 3 ns should make use of each digital channel probe s own ground When designing a high speed digital system you should consider designing dedicated test ports that interface directly to the instrument s probe system This will ease measurement setup and ensure a repeatable method for obtaining test data The 01650 61607 16 channel logic probe cable and the 01650 63203 termination adapter are designed to make it easy to connect to industry standard 20 pin board
239. ink When this happens decrease the horizontal time div setting and run again Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 357 23 CAN LIN Triggering and Serial Decode 358 Searching for LIN Data in the Lister The oscilloscope s search capability lets you search for and mark certain types of LIN data in the Lister You can use the Navigate key and controls to navigate through the marked rows 1 With LIN selected as the serial decode mode press Search 2 Inthe Search Menu press the Search softkey then turn the Entry knob to select the serial slot Serial 1 or Serial 2 on which the LIN signal is being decoded 3 Press Search then select from these options ID Finds frames with the specified ID Press the Frame ID softkey to select the ID ID amp Data Finds frames with the specified ID and data Press the Frame ID softkey to select the ID Press the Bits softkey to enter the data value Errors Finds all errors For more information on searching data see Searching Lister Data on page 124 For more information on using the Navigate key and controls see Navigating the Time Base on page 58 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and serial Decode Setup for FlexRay Signals 359 FlexRay Triggering 360 FlexRay Serial Decode 363 FlexRay triggering and
240. ion see To establish a LAN connection on page 289 It is also possible to set up a point to point LAN connection between a PC and the oscilloscope see Stand alone Point to Point Connection to a PC on page 290 Once the oscilloscope is set up on the network you can use the oscilloscope s web page to view or change its network configuration and access additional settings like the network password See Chapter 21 Web Interface starting on page 307 When you connect the oscilloscope to a LAN it is a good practice to limit access to the oscilloscope by setting a password By default the oscilloscope is not password protected See Setting a Password on page 320 to set a password 288 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Utility Settings 20 Any time you modify the oscilloscope s hostname it breaks the connection between the oscilloscope and the LAN You need to re establish communication to the oscilloscope using the new hostname To establish a LAN connection Automatic 1 Press Utility gt 1 0 Configuration 2 Press the LAN Settings softkey 3 Press the Config softkey then turn the Entry knob to select Automatic and press the softkey again to enable it f your network supports DHCP or AutolP enabling Automatic lets the oscilloscope use those services to get its LAN configuration settings 4 If your network provides Dynamic DNS you can enable the Dynamic DNS option to let the osc
241. it Annotation Menu Spell Delete Character Use the Spell and Delete Character softkeys to enter the annotation text Spell press this softkey and turn the Entry knob to select the character at the current position press this softkey to enter characters and move the cursor to the next character position gt press this softkey to enter characters and move the cursor to the previous character position Delete Character press the or softkeys until the desired character is highlighted then press this softkey to delete the character You can use a connected USB keyboard instead of using the Spell and other character editing softkeys Use the Clear softkey to delete all annotation characters Press OK to save the annotation edits 304 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Utility Settings 20 6 Press the Text Color softkey and turn the Entry knob to select the annotation color You can choose white red or colors that match analog channels digital channels math waveforms reference waveforms or markers 7 Press the Background softkey and turn the Entry knob to select the annotation background Opaque the annotation has a solid background Inverted the annotation s foreground and background colors are switched Transparent the annotation has a transparent background 1 5005 500 08 Trig d 1 70 KEYSIGHT TECHNOLO S Acquisition
242. ition the cursors where you want to read the bus values Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Digital Channels 6 20 005 Auto gt a 2 06V KEYSIGHT TECHNOLOGIES X1 cursor Acquisition X2 cursor v Bus values Al Ad Set cursors Select Bus1 Bus values mode to or Bus2 source at cursors Binary or Hex shown here Bus valuesare The bus values are also displayed when using the Pattern trigger function Press displayed when the Pattern key on the front panel to display the Pattern Trigger Menu and the bus using Pattern values will be displayed on the right above the softkeys tri oe The dollar sign will be displayed in the bus value when the bus value cannot be displayed as a hex value This occurs when one or more don t cares X are combined with low 0 and high 1 logic levels in the pattern specification or when a transition indicator rising edge 4 or falling edge t are included in the pattern specification A byte that consists of all don t cares X will be displayed in the bus as a don t care X Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 113 6 Digital Channels Pattern Trigger Pattern Trigger pattern definition 1110 0110 D Qualifier Channel Entered a ZA Analog Bus values displayed channel values at cursor KEYSIGHT TECHNOLOGIE Acquisition 4 MEMAN 1 4 B 1 4 8
243. ition to the standard Time column the MIL STD 1553 Lister contains these columns RTA displays the Remote Terminal Address for Command Status words nothing for Data words Word Type Cmd Status for Command Status words Data for Data words For Command Status words the background color is green to match the decode text color Data the 11 bits after the RTA for Command Status words or the 16 bits of a Data word Errors Sync Parity or Manchester errors as appropriate The background color is red to indicate an error Aliased data is highlighted in pink When this happens decrease the horizontal time div setting and run again 402 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide MIL STD 1553 ARINC 429 Triggering and Serial Decode 27 Searching for MIL STD 1553 Data in the Lister The oscilloscope s search capability lets you search for and mark certain types of MIL STD 1553 data in the Lister You can use the Navigate key and controls to navigate through the marked rows 1 With MIL STD 1553 selected as the serial decode mode press Search 2 Inthe Search Menu press the Search softkey then turn the Entry knob to select the serial slot Serial 1 or Serial 2 on which the MIL STD 1553 signal is being decoded 3 Press Search then select from these options Data Word Start finds the start of a Data word at the end of a valid Data Sync pulse Command Status Word Start fin
244. its in length including the alert bit the 9th bit Triggers on a data byte that you specify regardless of the value of the 9th alert bit The specified data byte applies to the least significant 8 bits excludes the 9th alert bit Similar choices are available for Tx Rx or Tx Parity Error Triggers on a parity error based on the parity you have set in the Bus Configuration Menu lf you choose a trigger condition that includes Data in its description for example Rx Data then press the Data is softkey and choose an equality qualifier You can choose equal to not equal to less than or greater than a specific data value Use the Data softkey to choose the data value for your trigger comparison This works in conjunction with the Data is softkey Optional The Burst softkey lets you trigger on the Nth frame 1 4096 after an idle time you select All trigger conditions must be met for the trigger to occur lf Burst is selected an idle time 1 us to 10 s can be specified so that the oscilloscope will look for a trigger condition only after the idle time has past Press the Idle softkey and rotate the Entry knob to set an idle time If the setup does not produce a stable trigger the UART RS232 signals may be slow enough that the oscilloscope is AutoTriggering Press the Mode Coupling key then press the Mode softkey to set the trigger mode from Auto to Normal To display UART RS232 serial decode
245. ive measurements will stay on screen allowing you to measure various aspects of the captured waveforms Press Increment Statistics to add the currently measured waveform to the collected statistical data The Analyze Segments softkey only appears when the acquisition is stopped and the optional segmented memory feature is on After an acquisition has completed and the oscilloscope is stopped you can press the Analyze Segments softkey to accumulate measurement statistics for the acquired segments You can also turn on infinite persistence in the Display Menu and press the Analyze Segments softkey to create an infinite persistence display Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 235 14 Measurements 236 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing To create a mask from a golden waveform Automask 237 Mask Test Setup Options 239 Mask Statistics 242 To manually modify a mask file 243 Building a Mask File 246 One way to verify a waveform s compliance to a particular set of parameters is to use mask testing A mask defines a region of the oscilloscope s display in which the waveform must remain in order to comply with chosen parameters Compliance to the mask is verified point by point across the display Mask test operates on displayed analog channels it does not operate on channels that are not
246. key to specify which edge of the Trigger B signal will trigger the oscilloscope Adjust the trigger level for the selected analog channel by turning the Trigger Level knob Press the Digital key and select Thresholds to set the threshold level for digital channels The value of the trigger level or digital threshold is displayed in the upper right corner of the display 5 Press the Back Up key to return to the Trigger Menu 6 Press the Delay softkey then turn the Entry knob to enter the delay time between the Arm A edge and the Trigger B edge 7 Press the Nth Edge B softkey then turn the Entry knob to select the Nth edge of the Trigger B signal to trigger on Pulse Width Trigger Pulse Width glitch triggering sets the oscilloscope to trigger on a positive or negative pulse of a specified width If you want to trigger on a specific timeout value use Pattern trigger in the Trigger Menu see Pattern Trigger on page 148 1 Press the Trigger key 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select Pulse Width Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 145 10 Triggers ol 146 KEYSIGHT TECHN S Acquisition Channels Source 5 00us Press the Source softkey then rotate the Entry knob to select a channel source for the trigger The channel you select is shown in the upper right corner of the display next to the polarity symbol The source
247. knobs When disabled you can use the channel s vertical scale and position knobs normally Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 251 16 Digital Voltmeter 252 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 17 Wavetorm Generator To select generated waveform types and settings 253 To edit arbitrary waveforms 257 To output the waveform generator sync pulse 261 To specify the expected output load 261 To use waveform generator logic presets 262 To add noise to the waveform generator output 262 To add modulation to the waveform generator output 263 To restore waveform generator defaults 267 A waveform generator is built into the oscilloscope It is enabled by Option WGN or the DSOX3WAVEGEN upgrade The waveform generator gives you an easy way to provide input signals when testing circuitry with the oscilloscope Waveform generator settings can be saved and recalled with oscilloscope setups See Chapter 18 Save Recall Setups Screens Data starting on page 269 To select generated waveform types and settings 1 To access the Waveform Generator Menu and enable or disable the waveform generator output on the front panel Gen Out BNC press the Wave Gen key When waveform generator output is enabled the Wave Gen key is illuminated When waveform generator output is disabled the Wave Gen key is
248. knobs The Horiz key for accessing the Horizontal Menu The zoom key for quickly enabling disabling the split screen zoom display The Search key for finding events on analog channels or in serial decode The Navigate keys for navigating time search events or segmented memory acquisitions The following figure shows the Horizontal Menu which appears after pressing the Horiz key KEYSIGHT TECHNOLOGIES 47 2 Horizontal Controls Trigger Time Delay Time Trigger Trigger level point reference time div source or threshold N 100 05 Sample rate Channels XY or Roll ral il mode x Time Mod TLIUN SSG KiB Time Ref Normal Center Fd J J Normal Zoomed Fine Time time mode time base control reference Figure 2 Horizontal Menu The Horizontal Menu lets you select the time mode Normal XY or Roll enable Zoom set the time base fine control vernier and specify the time reference The current sample rate is displayed in the right side information area To adjust the horizontal time div scale 1 Turn the large horizontal scale sweep speed knob marked N Ay to change the horizontal time div setting Notice how the time div information in the status line changes 48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Horizontal Controls 2 The V symbol at the top of the display indicates the time reference point The horizontal scale knob works i
249. l key is illuminated the upper multiplexed knob selects and highlights in red individual digital channels and the lower multiplexed knob positions the selected digital channel If a trace is repositioned over an existing trace the indicator at the left edge of the trace will change from Dnn designation where nn is a one or two digit channel number from 0 to 15 to D The indicates that two channels are overlaid You can rotate the upper knob to select an overlaid channel then rotate the lower knob to position it just as you would any other channel For more information on digital channels see Chapter 6 Digital Channels starting on page 103 Serial key This key is used to enable serial decode The multiplexed scale and position knobs are not used with serial decode For more information on serial decode see Chapter 7 Serial Decode starting on page 121 Multiplexed scale knob This scale knob is used with Math Ref or Digital waveforms whichever has the illuminated arrow to the left For math and reference waveforms the scale knob acts like an analog channel vertical scale knob Multiplexed position knob This position knob is used with Math Ref or Digital waveforms whichever has the illuminated arrow to the left For math and reference waveforms the position knob acts like an analog channel vertical position knob 12 Measure controls The measure controls consist of Cursors k
250. l data line If the channel you selected is off switch it on The MISO label for the source channel is automatically set b Press the Threshold softkey then turn the Entry knob to select the MISO signal threshold voltage level The threshold voltage level is used in decoding and it will become the trigger level when the trigger type is set to the selected serial decode slot 9 Press the CS softkey to open the SPI Chip Select Menu Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 379 25 380 I2C SPI Triggering and Serial Decode CS Trigger Point al clk AEN e nT tae 2 Daa X io X Bi X Bini X BiN X FF FF 4D CF 49 4C 45 X FF FF 02 EOP FE Pr Ee 03 06 OO 00 00 OO 00 06 05 FF 02 10 4D 53 4F Frame by Threshold Display Info CS TTL 1 4 In the SPI Chip Select Menu a Press the Frame by softkey to select a framing signal that the oscilloscope will use for determining which clock edge is the first clock edge in the serial stream You can set the oscilloscope to trigger during a high chip select CS a low chip select CS or after a Timeout period during which the clock signal has been idle If the framing signal is set to CS or CS the first clock edge as defined rising or falling seen after the CS or CS signal transitions from low to high or high to low is the first clock in the serial stream Chip Select Press the CS or CS softkey then turn the Entry kn
251. l decode and trigger Automotive serial decode and trigger UART RS232 serial decode and trigger Segmented Memory Mask limit testing 200MHz 100MHz Bandwidth Education kitticense WaveGen license Remote Log Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 319 21 Web Interface Setting a Password 320 Whenever you connect the oscilloscope to a LAN it is good practice to set a password The password prevents someone from remotely accessing the oscilloscope via a Web browser and changing parameters Remote users can still view the Welcome screen view network status etc but they can t operate the instrument or change its setup without the password To set a password 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 308 2 When the oscilloscope s web interface is displayed select the Configure Network tab from the instrument s Welcome page 3 Click the Modify Configuration button Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Web Interface 21 KEYSIGHT TECHNOLOGIES ical vy Current Network Configuration Modify Configuration Parameter Currently in use toma on on Configure on Network Agilent MSOX2024A InfiniiVision US50210029 tab E ra 156 140 24 15 156 140 24 47 aaa of 7 on on Keysight InfiniiVision 3000 X
252. l for testing commands or entering a few commands interactively When creating automated programs for controlling the oscilloscope you will typically use the Keysight IO Libraries from within a programming environment like Microsoft Visual Studio see Remote Programming with Keysight IO Libraries on page 314 Real Scope Remote Front Panel To operate the oscilloscope using the web interface s Real Scope Remote Front Panel 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 308 2 When the oscilloscope s web interface is displayed select Browser Web Control then select Real Scope Remote Front Panel After a few seconds the Remote Front Panel appears 3 Click keys or knobs that you would normally press on the oscilloscope s front panel Drag on edges of knobs to turn them Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Web Interface 21 AW KEYSIGHT _ Infiniivision MSO X 2024A wus Stage Ondtoncepe zoome 2054 8 mena Zoom Simple Remote Front Panel To operate the oscilloscope using the web interface s Simple Remote Front Panel 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 308 2 When the oscilloscope s web interface is displayed select Browser Web Control then select Simple Remote Front Panel After a few seconds the Remote Front Panel appears 3 Use the Main Menu and the Function Keys to control the oscilloscope To v
253. l in the Normal trigger mode where acquisitions are made only when the trigger condition is met In this mode if no triggers are occurring that is the Trig d indicator is displayed you can press Force Trigger to force a trigger and see what the input signals look like n the Auto trigger mode when the trigger condition is not met triggers are forced and the Auto indicator is displayed Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 141 10 Triggers Edge Trigger 142 The Edge trigger type identifies a trigger by looking for a specified edge slope and voltage level on a waveform You can define the trigger source and slope in this menu The trigger type source and level are displayed in the upper right corner of the display 1 On the front panel in the Trigger section press the Trigger key 2 Inthe Trigger Menu press the Trigger softkey and use the Entry knob to select Edge 3 Select the trigger source Analog channel 1 to the number of channels Digital channel on mixed signal oscilloscopes DO to the number of digital channels minus one External triggers on the rear panel EXT TRIG IN signal Line triggers at the 50 level of the rising or falling edge of the AC power source signal WaveGen triggers at the 50 level of the rising edge of the waveform generator output signal Not available when the DC Noise or Cardiac waveforms are selected You can choose a chan
254. l s BNC connector to transfer information between the oscilloscope and the probe When you connect a compatible probe to the oscilloscope the AutoProbe interface determines the type of probe and sets the oscilloscope s parameters units offset attenuation coupling and impedance accordingly 18 Demo 2 Ground and Demo 1 terminals Demo 2 terminal This terminal outputs the Probe Comp signal which helps you match a probe s input capacitance to the oscilloscope channel to which it is connected See Compensate Passive Probes on page 32 With certain licensed features the oscilloscope can also output demo or training signals on this terminal Ground terminal Use the ground terminal for oscilloscope probes connected to the Demo 1 or Demo 2 terminals Demo 1 terminal With certain licensed features the oscilloscope can output demo or training signals on this terminal 19 USB Host port This port is for connecting USB mass storage devices or printers to the oscilloscope Connect a USB compliant mass storage device flash drive disk drive etc to save or recall oscilloscope setup files and reference waveforms or to save data and screen images See Chapter 18 Save Recall Setups Screens Data starting on page 269 To print connect a USB compliant printer For more information about printing see Chapter 19 Print Screens starting on page 281 You can also use the USB port to update the
255. least twice as much memory as acquisitions captured when running and the oscilloscope stores at least twice as many samples At slower time div settings because there is more memory available for a single acquisition the acquisition has a higher effective sample rate Running When running versus taking a single acquisition the memory is divided in half This lets the acquisition system acquire one record while processing the previous acquisition dramatically improving the number of waveforms per second processed by the oscilloscope When running a high waveform update rate provides the best representation of your input signal To acquire data with the longest possible record length press the Single key For more information on settings that affect record length see Length Control on page 273 Overview of Sampling To understand the oscilloscope s sampling and acquisition modes it is helpful to understand sampling theory aliasing oscilloscope bandwidth and sample rate oscilloscope rise time oscilloscope bandwidth required and how memory depth affects sample rate 182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Acquisition Control 12 Sampling Theory The Nyquist sampling theorem states that for a limited bandwidth band limited signal with maximum frequency fmax the equally spaced sampling frequency fs must be greater than twice the maximum frequency fmax in order to have the signal be
256. lect the minimum number of clocks that define a sync break in your LIN signal LIN Triggering To set up the oscilloscope to capture a LIN signal see Setup for LIN Signals on page 351 LIN triggering can trigger on the rising edge at the Sync Break exit of the LIN single wire bus signal that marks the beginning of the message frame the Frame ID or the Frame ID and Data A LIN signal message frame is shown below Sync Sync _ Identifier Data Checksum Break Field Break Fields Field 4 Sync Break Exit Press Trigger 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select the serial slot Serial 1 or Serial 2 on which the CAN signal is being decoded Trigger Menu Trigger Trigger Frame ID 5 LIN ID amp Data DXX 3 Press the Trigger softkey then turn the Entry knob to select the trigger condition Sync Sync Break The oscilloscope triggers on the rising edge at the Sync Break exit of the LIN single wire bus signal that marks the beginning the message frame Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 353 23 CAN LIN Triggering and Serial Decode ID Frame ID The oscilloscope triggers when a frame with an ID equal to the selected value is detected Use the Entry knob to select the value for the Frame ID ID amp Data Frame ID and Data The oscilloscope triggers when a frame with an ID and data equal to the selected values is detected
257. lect the second analog channel source for the phase measurement The default Phase settings measure from channel 1 to channel 2 5 Press the O Back Up key to return to the Measurement Menu 6 Press the Add Measurement softkey to make the measurement The example below shows a phase measurement between the channel 1 and the math d dt function on channel 1 KEYSIGHT TECHN Acquisition Channels x Measurements Freq 1 499 15Hz Pk Pk 1 Delay 14 gt Measurement Menu Source Type Add Settings Clear Meas Phase Measurement a al gt X at Min Y 228 X at Min Y is the X axis value usually time at the first displayed occurrence of the waveform Minimum starting from the left side of the display For periodic signals the position of the minimum may vary throughout the waveform The X cursor shows where the current X at Min Y value is being measured Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Measurements 14 X at Max Y X at Max Y is the X axis value usually time at the first displayed occurrence of the waveform Maximum starting from the left side of the display For periodic signals the position of the maximum may vary throughout the waveform The X cursor shows where the current X at Max Y value is being measured See Also To measure the peak of an FFT on page 229 To measure the peak of an FFT 1 Select FFT as the Operator in the Waveform Math Menu 2 Choose Math f t as the sou
258. line does not appear on the display press the Serial key to turn it on 6 If the oscilloscope is stopped press the Run Stop key to acquire and decode data Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 417 28 UART RS232 Triggering and Serial Decode If the setup does not produce a stable trigger the UART RS232 signals may be slow enough that the oscilloscope is AutoTriggering Press the Mode Coupling key then press the Mode softkey to set the trigger mode from Auto to Normal You can use the horizontal Zoom window for easier navigation of the acquired data See Also Interpreting UART RS232 Decode on page 418 UART RS232 Totalizer on page 419 Interpreting UART RS232 Lister Data on page 420 Searching for UART RS232 Data in the Lister on page 420 Interpreting UART RS232 Decode Channels Settings Lister w w Angled waveforms show an active bus inside a packet frame Mid level blue lines show an idle bus When using 5 8 bit formats the decoded data is displayed in white in binary hex or ASCII When using the 9 bit format all data words are displayed in green including the 9th bit The 9th bit is displayed on the left 418 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide UART RS232 Triggering and Serial Decode 28 When a data word value is selected for framing it is displayed in light blue When using 9 bit data words the 9th bit will also be displayed in
259. ll become the trigger level when the trigger type is set to the selected serial decode slot c Press the slope softkey to select rising edge or falling edge for the selected Clock sou rce This determines which clock edge the oscilloscope will use to latch the serial data When Display state of the clock si Info is enabled the graphic changes to show the current gnal 7 Press the MOSI softkey to open the SPI Master Out Slave In Menu SPI Master Out Slave In Menu MOSI Data Threshold Display Info TTL 1 4V 378 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide I2C SPI Triggering and Serial Decode 25 In the SPI Master Out Slave In Menu a Press the MOSI Data softkey then turn the Entry knob to select the channel that is connected to a SPI serial data line If the channel you selected is off switch it on The MOSI label for the source channel is automatically set b Press the Threshold softkey then turn the Entry knob to select the MOSI signal threshold voltage level The threshold voltage level is used in decoding and it will become the trigger level when the trigger type is set to the selected serial decode slot 8 Optional Press the MISO softkey to open the SPI Master In Slave Out Menu Display Info In the SPI Master In Slave Out Menu a Press the MISO Data softkey then turn the Entry knob to select the channel that is connected to a second SPI seria
260. ll display measurements relative to the set locations 7 Adjust the selected cursor s by turning the Cursors knob Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Cursors 13 Cursor Examples Channels Cursors Cursors Menu Mode Cursors Manual wil Figure 40 Cursors used to measure pulse widths other than middle threshold points Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 203 13 Cursors 100 08 KEYSIGHT TECHNO Acquisition Channels Cursors Mode Source Cursors Manual X2 Figure 41 Cursors measure frequency of pulse ringing Expand the display with Zoom mode then characterize the event of interest with the cursors 204 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 20 008 Source Cursors Units Manual Y2 D ai Figure 42 Cursors track Zoom window Cursors 13 1 00 KEYSIGHT TECH Acquisition Channels Cursors 64 400ns 1 15 5 A T Put the X1 cursor on one side of a pulse and the X2 cursor on the other side of the pulse Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 205 13 Cursors 1 00 KEYSIGHT TECH Acquisition Channels Cursors Mode Source Cursors Manual X2 Figure 43 Measuring pulse width with cursors Press the X1 X2 linked softkey and move the cursors together to check for pulse width variations in a pulse train 206 Keysight InfiniiVision 3000 X Series Oscilloscopes
261. locations or from an external USB storage device 1 Press Save Recall gt Recall gt Recall then turn the Entry knob to select Setup scp Press the softkey in the second position and use the Entry knob to navigate to the file to recall See To navigate storage locations on page 276 Press the Press to Recall softkey A message indicating whether the recall was successful is displayed If you would like to clear the display press Clear Display To recall mask files Mask files can be recalled from one of four internal Keysight Flash locations or from an external USB storage device 1 Press Save Recall gt Recall gt Recall then turn the Entry knob to select Mask msk Press the softkey in the second position and use the Entry knob to navigate to the file to recall See To navigate storage locations on page 276 Press the Press to Recall softkey A message indicating whether the recall was successful is displayed If you would like to clear the display or clear the recalled mask press Clear Display or Clear Mask To recall reference waveform files from a USB storage device Press the Save Recall key 2 Inthe Save Recall Menu press the Recall softkey 3 In the Recall Menu press the Recall softkey and turn the Entry knob to select Reference Waveform data h5 Press the To Ref softkey and turn the Entry knob to select the desired reference waveform location Press the softkey i
262. loscopes User s Guide 19 LIN Triggering 353 LIN Serial Decode 354 Interpreting LIN Decode 356 Interpreting LIN Lister Data 357 Searching for LIN Data in the Lister 358 24 FlexRay Triggering and Serial Decode Setup for FlexRay Signals 359 FlexRay Triggering 360 Triggering on FlexRay Frames 361 Triggering on FlexRay Errors 362 Triggering on FlexRay Events 362 FlexRay Serial Decode 363 Interpreting FlexRay Decode 364 FlexRay Totalizer 365 Interpreting FlexRay Lister Data 365 Searching for FlexRay Data inthe Lister 366 25 12C SPI Triggering and Serial Decode Setup for 12C Signals 369 12C Triggering 370 I2C Serial Decode 374 Interpreting I2C Decode 375 Interpreting I2C Lister Data 376 Searching for 12C Data in the Lister 376 Setup for SPI Signals 377 SPI Triggering 381 SPI Serial Decode 383 Interpreting SPI Decode 384 Interpreting SPI Lister Data 385 20 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Searching for SPI Data in the Lister 385 26 12S Triggering and Serial Decode Setup for 12S Signals 387 12S Triggering 390 12S Serial Decode 393 Interpreting I2S Decode 394 Interpreting I2S Lister Data 395 Searching for 12S Data in the Lister 395 27 MIL STD 1553 ARINC 429 Triggering and Serial Decode Setup for MIL STD 1553 Signals 397 MIL STD 1553 Triggering 399 MIL STD 1553 Serial Decode 400 Interpreting MIL STD 1553 Decode
263. ls on page 397 To set up a MIL STD 1553 trigger 1 Press Trigger 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select the serial slot Serial 1 or Serial 2 on which the MIL STD 1553 signal is being decoded Bit Times 11110 00100 Trigger Trigger Bit Time 5 MIL 1553 RTA Bits g 3 Press the Trigger softkey then turn the Entry knob to select the trigger condition Data Word Start triggers on the start of a Data word at the end of a valid Data Sync pulse Data Word Stop triggers on the end of a Data word Command Status Word Start triggers on the start of Command Status word at the end of a valid C S Sync pulse Command Status Word Stop triggers on the end of a Command Status word Remote Terminal Address triggers if the RTA of the Command Status word matches the specified value When this option is selected the RTA softkey becomes available and lets you select the hex Remote Terminal Address value to trigger on If you select 0xXX don t cares the oscilloscope will trigger on any RTA Remote Terminal Address 11 Bits triggers if the RTA and the remaining 11 bits match the specified criteria When this option is selected these softkeys become available The RTA softkey lets you select the hex Remote Terminal Address value The Bit Time softkey lets you select the bit time position The 01 X softkey lets you set the bit time position value as a
264. ly New Label softkey Therefore you only need to select a new channel and press the Apply New Label softkey again to assign the label to the channel Only the original label is saved in the label list This feature makes it easier to assign successive labels to numbered control lines and data bus lines To load a list of labels from a text file you create 136 It may be convenient to create a list of labels using a text editor then load the label list into the oscilloscope This lets you type on a keyboard rather than edit the label list using the oscilloscope s controls You can create a list of up to 75 labels and load it into the oscilloscope Labels are added to the beginning of the list If more than 75 labels are loaded only the first 75 are stored To load labels from a text file into the oscilloscope Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Labels 9 1 Use a text editor to create each label Each label can be up to ten characters in length Separate each label with a line feed 2 Name the file labellist txt and save it on a USB mass storage device such as a thumb drive 3 Load the list into the oscilloscope using the File Explorer press Utility gt File Explorer Label List Management When you press the Library softkey you will see a list of the last 75 labels used The list does not save duplicate labels Labels can end in any number of trailing digits As long as the base string is th
265. ly spaced over the time period 2 Use the Frequency Frequency Fine Period Period Fine softkey to set the time period bounding parameter repetition frequency of the arbitrary waveform 3 Use the Amplitude High Level and Offset Low Level softkeys to set the voltage bounding parameter of the arbitrary waveform 4 When you are ready to create the new arbitrary waveform press Apply amp Edit Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide CAUTION Waveform Generator 17 When you create a new arbitrary waveform the existing arbitrary waveform definition is overwritten Note that you can use the Save Recall key and menu to save arbitrary waveforms to one of four internal storage locations or to a USB storage device and you can recall them later See To save arbitrary waveforms on page 275 and To recall arbitrary waveforms on page 279 The new waveform is created and the Edit Waveform Points menu is opened See Editing Existing Arbitrary Waveforms on page 259 Note that you can also create a new arbitrary waveform by capturing another waveform See Capturing Other Waveforms to the Arbitrary Waveform on page 260 Editing Existing Arbitrary Waveforms The Edit Waveform Points Menu is opened by pressing Edit Existing in the Edit Waveform Menu or by pressing Apply amp Edit when creating a new arbitrary waveform 100 05 O 0V KEYSIGHT mA TECHNOLOGIES 1 00ms Acquisition Channels 250mV 1
266. m generator output 265 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Index frequency requirements power source 27 frequency Nyquist 183 frequency shift keying modulation FSK waveform generator output 266 front panel controls and connectors 33 front panel self test 300 front panel browser based remote 312 front panel language overlay 40 front panel real scope remote 310 front panel simple remote 311 FSK frequency shift keying modulation waveform generator output 266 G g t 73 gateway IP 289 Gaussian frequency response 184 gaussian pulse waveform generator output 256 Generic video trigger 162 German front panel overlays 41 German user interface and Quick Help 45 glitch capture 189 glitch trigger 145 golden waveform test 237 GPIB address 288 GPIB interface remote control 287 GPIB module 26 42 GPIB module installation 26 grabber 105 106 graphical user interface language 45 graticule intensity 131 graticule type 130 grid intensity 131 grid type 130 ground level 62 Ground terminal 39 H Hanning FFT window 82 hardware self test 300 Help key 38 425 Index help built in 44 hex bus trigger 150 Hex softkey 150 HF Reject 177 high pass filter math function 92 high frequency noise rejection 177 high resolution mode 187 193 holdoff 177 hop frequency FSK modulation 267 Horiz key 36 47 52 54 190 H
267. m N Preshoo local Minimum Average Average is the sum of the levels of the waveform samples divided by the number of samples Xi Average j Where x value at ith point being measured n number of points in measurement interval The Full Screen measurement interval variation measures the value on all displayed data points The N Cycles measurement interval variation measures the value on an integral number of periods of the displayed signal If less than three edges are present the measurement shows No edges The X cursors show what interval of the waveform is being measured Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 219 14 Measurements 220 DC RMS AC RMS DC RMS is the root mean square value of the waveform over one or more full periods n 2 ae RMS dc 7i Where x value at ith point being measured n number of points in measurement interval The Full Screen measurement interval variation measures the value on all displayed data points The N Cycles measurement interval variation measures the value on an integral number of periods of the displayed signal If less than three edges are present the measurement shows No edges The X cursors show the interval of the waveform being measured AC RMS is the root mean square value of the waveform with the DC component removed It is useful for example for measuring power supply noise The N Cycles measurement i
268. m intensity adjustment affects analog channel waveforms only not math waveforms referencewaveforms digital waveforms etc KEYSIGHT TECHNOLOGIES 127 8 Display Settings 2 14V KEYSIGHT TECHNO Acquisition Channels 10 0 1 nu Trigger Source Edge 2 Figure 25 Amplitude Modulation Shown at 100 Intensity Channels 10 0 1 10 0 1 Trigger gt Source Edge 2 Figure 26 Amplitude Modulation Shown at 40 Intensity 128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Display Settings 8 To set or clear persistence With persistence the oscilloscope updates the display with new acquisitions but does not immediately erase the results of previous acquisitions All previous acquisitions are displayed with reduced intensity New acquisitions are shown in their normal color with normal intensity Waveform persistence is kept only for the current display area you cannot pan and zoom the persistence display To use persistence 1 Press the Display key 2 Press Persistence then turn the Entry knob to select between Off turns off persistence When persistence is off you can press the Capture Waveforms softkey to perform a single shot infinite persistence A single acquisition s data is displayed with reduced intensity and it remains on the display until you clear persistence or clear the display oo Persistence infinite presistence Results of previous acquisitions are neve
269. multiple infrequent trigger events it is advantageous to divide the oscilloscope s memory into segments This lets you capture signal activity without capturing long periods of signal inactivity Each segment is complete with all analog channel digital channel on MSO models and serial decode data When using segmented memory use the Analyze Segments feature see Measurements Statistics and Infinite Persistence with Segmented Memory on page 195 to show infinite persistence across all acquired segments See also To set or clear persistence on page 129 for details 1 Set up a trigger condition See Chapter 10 Triggers starting on page 139 for details 2 Press the Acquire key in the Waveform section of the front panel 3 Press the Segmented softkey 4 In the Segmented Memory Menu press the Segmented softkey to enable segmented memory acquisitions 5 Press the of Segs softkey and turn the Entry knob to select the number of segments into which you would like to divide the oscilloscope s memory Memory can be divided into as few as two segments and as many as 1000 segments depending on the oscilloscope model 6 Press the Run or Single key The oscilloscope runs and fills a memory segment for each trigger event When the oscilloscope is busy acquiring multiple segments the progress is displayed in the upper right area of the display The oscilloscope continues to trigger until memory is filled then t
270. n For more information on InfiniiMax probes and accessories see the probe s User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 69 3 Vertical Controls 70 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 4 Math Waveforms To display math waveforms 71 To perform transforms or filters on an arithmetic operation 72 To adjust the math waveform scale and offset 73 Units for Math Waveforms 73 Math Operators 74 Math Transforms 76 Math Filters 92 Math Visualizations 93 Math functions can be performed on analog channels The resulting math waveform is displayed in light purple You can use a math function on a channel even if you choose not to display the channel on screen You can Perform an arithmetic operation like add subtract or multiply on analog input channels Perform a transform function like differentiate integrate FFT or square root on an analog input channel Perform a transform function on the result of an arithmetic operation To display math waveforms 1 Press the Math key on the front panel to display the Waveform Math Menu KEYSIGHT TECHNOLOGIES 71 4 Math Waveforms Waveform Math Menu f t Ch1 Ch2 Function Operator Source 1 Source 2 tit iii 2 If f t is not already shown on the Function softkey press the Function sofkey and select
271. n 3000 X Series Oscilloscopes User s Guide MIL STD 1553 ARINC 429 Triggering and Serial Decode 27 Interpreting ARINC 429 Lister Data 146 1 3 1 00 Label SDI Data SSM Errors KEYSIGHT 1 745ms TECHNOLOGIES 1 381ms 1 018ms gt 55 44us 418 6us Channels 781 8us 00041 1 129ms Options isplay 5 ARINC 429 In addition to the standard Time column the ARINC 429 Lister contains these columns Label the 5 bit label value in octal format SDI the bit values if included in the word decode format Data the data value in binary or hex depending on the base setting SSM the bit values if included in the word decode format Errors highlighted in red Errors can be Parity Word or Gap Aliased data is highlighted in pink When this happens decrease the horizontal time div setting and run again Searching for ARINC 429 Data in the Lister The oscilloscope s search capability lets you search for and mark certain types of ARINC 429 data in the Lister You can use the Navigate key and controls to navigate through the marked rows 1 With ARINC 429 selected as the serial decode mode press Search Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 411 27 MIL STD 1553 ARINC 429 Triggering and Serial Decode 2 Inthe Search Menu press the Search softkey then turn the Entry knob to select the serial slot Serial 1 or Serial 2 on which the
272. n the specified value In Range enter upper and lower values to specify the range in which to trigger Out of Range enter upper and lower values to specify range in which trigger will not occur Increasing value triggers when the data value is increasing over time and the specified value is met or exceeded Set Trigger gt to the data value that must be reached Set Armed lt to the value to which the data must fall before the trigger circuit is re armed ready to trigger again These settings are made in the current menu when Base is Decimal or in the Bits submenu when the Base is Binary The Armed control reduces triggers due to noise This trigger condition is best understood when the digital data transferred over the 12S bus is considered in terms of representing an analog waveform The figure below shows a plot of sample data transmitted over an 12S bus for one channel In this example the oscilloscope will trigger at the 2 points shown since there are two instances in which the data increases from a value below or at the Armed value to a value greater than or equal to the specified Trigger value If you select an Armed value that is equal to or greater than the Trigger value the Trigger value will be increased so that it is always greater than the Armed value Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 391 26 392 12S Triggering and Serial Decode Trigger Trigger
273. n the Normal time mode while acquisitions are running or when they are stopped When running adjusting the horizontal scale knob changes the sample rate When stopped adjusting the horizontal scale knob lets you zoom into acquired data See Panning and Zooming Single or Stopped Acquisitions on page 50 Note that the horizontal scale knob has a different purpose in the Zoom display See To display the zoomed time base on page 54 To adjust the horizontal delay position 1 Turn the horizontal delay position knob 4 The trigger point moves horizontally pausing at 0 00 s mimicking a mechanical detent and the delay value is displayed in the status line Changing the delay time moves the trigger point solid inverted triangle horizontally and indicates how far it is from the time reference point hollow inverted triangle V These reference points are indicated along the top of the display grid Figure 2 shows the trigger point with the delay time set to 200 us The delay time number tells you how far the time reference point is located from the trigger point When delay time is set to zero the delay time indicator overlays the time reference indicator All events displayed left of the trigger point happened before the trigger occurred These events are called pre trigger information and they show events that led up to the trigger point Everything to the right of the trigger point is called post trigger inform
274. n the second position and use the Entry knob to navigate to the file to recall See To navigate storage locations on page 276 Press the Press to Recall softkey Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Save Recall Setups Screens Data 18 A message indicating whether the recall was successful is displayed 7 If you would like to clear the display of everything except the reference waveform press Clear Display To recall arbitrary waveforms Arbitrary waveform files can be recalled from one of four internal Keysight Flash locations or from an external USB storage device When recalling were not saved If the file con If the file con column to arbitrary waveforms from an external USB storage device that from the oscilloscope be aware that tains two columns the second column is automatically chosen tains more than two columns you are prompted to select which oad Up to five columns are parsed by the oscilloscope columns above the fifth are ignored The oscilloscope uses a maximum of 8192 points for an arbitrary waveform For more efficien less To recall an arb t recalls make sure your arbitrary waveforms are 8192 points or itrary waveform 1 Press Save Recall gt Recall gt Recall then turn the Entry knob to select Arbitrary Waveform data csv 2 Press the softkey in the second position and use the Entry knob to navigate to the file to recall
275. ncy is also called the folding frequency because aliased frequency components fold back from that frequency when viewing the frequency domain Aliasing happens when there are frequency components in the signal higher than half the sample rate Because the FFT spectrum is limited by this frequency any higher components are displayed at a lower aliased frequency The following figure illustrates aliasing This is the spectrum of a 990 Hz square wave which has many harmonics The sample rate is set to 100 kSa s and the oscilloscope displays the spectrum The displayed waveform shows the components of the input signal above the Nyquist frequency to be mirrored aliased on the display and reflected off the right edge 50 002 O 0V KEYSIGHT TECHNOLOGIES Acquisition Channels Cursors 120 500dB f t FFT Ch3 Scale FFT Resolution 1Hz Function Operator Source 1 Center More FFT tit FFT 25 0kHz d Figure9 Aliasing Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Math Waveforms 4 Because the frequency span goes from 0 to the Nyquist frequency the best way to prevent aliasing is to make sure that the frequency span is greater than the frequencies of significant energy present in the input signal FFT Spectral Leakage The FFT operation assumes that the time record repeats Unless there is an integral number of cycles of the sampled waveform in the record a discontinuity is created at the end of the reco
276. nd N A Cardiac To specify the expected output load 1 If the Waveform Generator Menu is not currently displayed on the oscilloscope s softkeys press the Wave Gen key 2 In the Waveform Generator Menu press the Settings softkey 3 In the Waveform Generator Settings Menu press the Out Load softkey and turn the Entry knob to select 502 High Z Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 261 17 Waveform Generator The output impedance of the Gen Out BNC is fixed at 50 ohms However the output load selection lets the waveform generator display the correct amplitude and offset levels for the expected output load If the actual load impedance is different than the selected value the displayed amplitude and offset levels will be incorrect To use waveform generator logic presets With logic level presets you can easily set the output voltage to TTL CMOS 5 0V CMOS 3 3V CMOS 2 5V or ECL compatible Low and High levels 1 If the Waveform Generator Menu is not currently displayed on the oscilloscope s softkeys press the Wave Gen key 2 In the Waveform Generator Menu press the Settings softkey In the Waveform Generator Settings Menu press the Logic Presets softkey 4 Inthe Waveform Generator Logic Level Presets Menu press one of the softkeys to set the generated signal s Low and High voltages to logic compatible levels Softkey logic levels Low level High level
277. nd finally if digital probes are connected the highest number digital channel During Auto Scale the delay is set to 0 0 seconds the horizontal time div sweep speed setting is a function of the input signal about 2 periods of the triggered signal on the screen and the triggering mode is set to Edge Passive Probes Each oscilloscope passive probe must be compensated to match the input characteristics of the oscilloscope channel to which it is connected A poorly compensated probe can introduce significant measurement errors 1 Input the Probe Comp signal see Input a Waveform on page 29 2 Press Default Setup to recall the default oscilloscope setup see Recall the Default Oscilloscope Setup on page 29 3 Press Auto Scale to automatically configure the oscilloscope for the Probe Comp signal see Use Auto Scale on page 30 4 Press the channel key to which the probe is connected 1 2 etc n the Channel Menu press Probe n the Channel Probe Menu press Probe Check then follow the instructions on screen f necessary use a nonmetallic tool supplied with the probe to adjust the trimmer capacitor on the probe for the flattest pulse possible On the N2862 63 90 probes the trimmer capacitor is the yellow adjustment on the probe tip On other probes the trimmer capacitor is located on the probe BNC connector Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Getting Starte
278. nel that is turned off not displayed as the source for the edge trigger The selected trigger source is indicated in the upper right corner of the display next to the slope symbol 1 through 4 analog channels DO through Dn digital channels E External trigger input L Line trigger W Waveform generator 4 Press the Slope softkey and select rising edge falling edge alternating edges or either edge depending on the selected source The selected slope is displayed in the upper right corner of the display Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Using AutoScale to Set Up Edge Triggers Triggers 10 200 05 2 21V KEYSIGHT TECHN S Acquisition Channels Source Alternating edge mode is useful when you want to trigger on both edges of a clock for example DDR signals Either edge mode is useful when you want to trigger on any activity of a selected source All modes operate up to the bandwidth of the oscilloscope except Either edge mode which has a limitation Either edge mode will trigger on Constant Wave signals up to 100 MHz but can trigger on isolated pulses down to 1 2 oscilloscope s bandwidth The easiest way to set up an Edge trigger on a waveform is to use AutoScale Simply press the AutoScale key and the oscilloscope will attempt to trigger on the waveform using a simple Edge trigger type See Use Auto Scale on page 30 MegaZoom Technology Simplifies
279. ng Started Access the Built In Quic To view Quick Help 44 Status line The top line of the display contains vertical horizontal and trigger setup information Display area The display area contains the waveform acquisitions channel identifiers and analog trigger and ground level indicators Each analog channel s information appears in a different color Signal detail is displayed using 256 levels of intensity For more information about viewing signal detail see To adjust waveform intensity on page 127 For more information about display modes see Chapter 8 Display Settings starting on page 127 Information area The information area normally contains acquisition analog channel automatic measurement and cursor results Menu line This line normally contains menu name or other information associated with the selected menu Softkey labels These labels describe softkey functions Typically softkeys let you set up additional parameters for the selected mode or menu Pressing the Back Up key at the top of the menu hierarchy turns off softkey labels and displays additional status information describing channel offset and other configuration parameters k Help 1 Press and hold the key or softkey for which you would like to view help Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Getting Started 1 KEYSIGH TEC Acquisition 12 Quick Help message
280. ng for a specified pattern This pattern is a logical AND combination of the channels Each channel can have a value of 0 low 1 high or don t care X A rising or falling edge can be specified for one channel included in the pattern You can also trigger on a hex bus value as described on Hex Bus Pattern Trigger on page 150 1 Press the Trigger key 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select Pattern 3 Press the Qualifier softkey then turn the Entry knob to select from the pattern duration qualifier options Entered when the pattern is entered lt Less Than when the pattern is present for less than a time value gt Greater Than when the pattern is present for greater than a time value The trigger occurs when the pattern exits not when the gt softkey time value is exceeded Timeout when the pattern is present for greater than a time value In this case the trigger occurs when the gt softkey time value is exceeded not when the pattern exits gt lt In Range when the pattern is present for a time within a range of values lt gt Out of Range when the pattern is present for a time outside of range of values Pattern durations are evaluated using a timer The timer starts on the last edge that makes the pattern logical AND true Except when the Timeout qualifier is selected the trigger occurs on the first edge that makes the pattern false pro
281. ng the Spell and other character editing softkeys Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 283 19 Print Screens 6 Press the Apply softkey to make the printer connection A message appears to tell you whether the connection was successful To specify the print options In the Print Configuration Menu press the Options softkey to change the following options Setup Information Select this to print oscilloscope setup information on your printout including vertical horizontal trigger acquisition math and display settings Invert Graticule Colors Select this to reduce the amount of black ink it takes to print oscilloscope images by changing the black background to white Invert Graticule Colors is the default mode Graticule Not Inverted Graticule Inverted Form Feed Se waveform is pri Feed off if you wan the waveform T selected Also setting Landscape Se portrait mode ect this to send a form feed command to the printer after the nted and before the setup information is printed Switch Form t setup information printed on the same sheet of paper with his option only has an effect when the Setup Information option is if the amount of setup information will not fit on the same page with the waveform it will be printed on a new page regardless of the Form Feed ec t this to print horizontally on the page instead of vertically To specify
282. nob Push this knob select cursors from a popup menu Then after the popup menu closes either by timeout or by pushing the knob again rotate the knob to adjust the selected cursor position Cursors key Press this key to open a menu that lets you select the cursors mode and source Meas key Press this key to access a set of predefined measurements See Chapter 14 Measurements starting on page 209 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 37 1 38 Getting Started 13 Waveform keys The Acquire key lets you select Normal Peak Detect Averaging or High Resolution acquisition modes see Selecting the Acquisition Mode on page 187 and use segmented memory see Acquiring to Segmented Memory on page 193 The Display key lets you access the menu where you can enable persistence see To set or clear persistence on page 129 clear the display and adjust the display grid graticule intensity see To adjust the grid intensity on page 131 14 File keys Press the Save Recall key to save or recall a waveform or setup See Chapter 18 Save Recall Setups Screens Data starting on page 269 The Print key opens the Print Configuration Menu so you can print the displayed waveforms See Chapter 19 Print Screens starting on page 281 15 Help key Opens the Help Menu where you can display overview help topics and select
283. nob to select the desired slot Serial 1 or Serial 2 and press the softkey again to enable decode 4 Press the Mode softkey then select 12S trigger type KEYSIGHT TECHNOLOGIES 387 26 388 12S Triggering and Serial Decode 5 Press the Signals softkey to open the I7S Signals Menu SCLK Threshold Threshold SDATA Threshold TTL 1 4v TTL 1 4 1 12V 6 For the SCLK serial clock WS word select and SDATA serial data signals a Connect an oscilloscope channel to the signal in the device under test b Press the SCLK WS or SDATA softkey then turn the Entry knob to select the channel for the signal c Press the corresponding Threshold softkey then turn the Entry knob to select the signal threshold voltage level Set the threshold levels for the SCLK WS and SDATA signals to the middle of the signals The threshold voltage level is used in decoding and it will become the trigger level when the trigger type is set to the selected serial decode slot The SCLK WS and SDATA labels for the source channels are automatically set 7 Press the O Back Up key to return to the Serial Decode Menu 8 Press the Bus Config softkey to open the I S Bus Configuration Menu and display a diagram showing WS SCLK and SDATA signals for the currently specified bus configuration Receiver Alignment WS Low SCLK Slope 8 S Left 9 Press the Word Size softkey Turn the Entry knob to match the transmitter word size of the
284. nob to toggle between High and Low selection 7 Press the Qualifier softkey to toggle between less than greater than or None This lets you specify that a runt pulse be less than or greater than a certain width 8 Ifyou selected the less than or greater than Qualifier press the Time softkey then turn the Entry knob to select the time 156 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Triggers 10 Setup and Hold Trigger The Setup and Hold trigger looks for setup and hold violations Clock rising edge Setup time Hold time One oscilloscope channel probes the clock signal and another channel probes the data signal To trigger on setup amp hold violations 1 2 Press the Trigger key In the Trigger Menu press the Trigger softkey then turn the Entry knob to select Setup and Hold Press the Clock softkey then turn the Entry knob to select the input channel with the clock signal Set the appropriate trigger tevel for the clock signal using the Trigger Level knob 5 Press the Rising Edge or Falling Edge softkey to specify the clock edge being used 6 Press the Data softkey then turn the Entry knob to select the input channel with the data signal Set the appropriate trigger tevel for the data signal using the Trigger Level knob Press the lt Setup softkey and turn the Entry knob to select the setup time Keysight InfiniiVision 3000 X Series Oscilloscopes
285. nstead of using the Spell and other character editing softkeys When available the Increment softkey can be used to enable or disable automatically incremented file names Auto increment adds a numeric suffix to your file name and increments the number with each successive save It will truncate characters as necessary when the file name length is at maximum and more digits are required for the numeric portion of the file name Recalling Setups Masks or Data 1 Press the Save Recall key 2 Inthe Save Recall Menu press Recall 3 In the Recall Menu press Recall then turn the Entry knob to select the type of file you want to recall Setup scp See To recall setup files on page 278 Mask msk See To recall mask files on page 278 Reference Waveform data h5 See To recall reference waveform files from a USB storage device on page 278 Arbitrary Waveform data csv See To recall arbitrary waveforms on page 279 You can also recall setups and mask files by loading them using the File Explorer See File Explorer on page 291 You can also configure the Quick Action key to recall setups masks or reference waveforms See Configuring the Quick Action Key on page 302 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 277 18 Save Recall Setups Screens Data 278 To recall setup files Setup files can be recalled from one of 10 internal Keysight Flash
286. nterval measures the value on an integral number of periods of the displayed signal If less than three edges are present the measurement shows No edges The X cursors show the interval of the waveform being measured The Full Screen Std Deviation measurement interval variation is an RMS measurement across the full screen with the DC component removed It shows the standard deviation of the displayed voltage values The standard deviation of a measurement is the amount that a measurement varies from the mean value The Mean value of a measurement is the statistical average of the measurement Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Measurements 14 The following figure graphically shows the mean and standard deviation Standard deviation is represented by the Greek letter sigma o For a Gaussian distribution two sigma 10 from the mean is where 68 3 percent of the measurement results reside Six sigma 30 from is where 99 7 percent of the measurement results reside mean 30 20 10 0 10 20 30 68 3 95 4 99 7 The mean is calculated as follows DE x N where x the mean N the number of measurements taken x the ith measurement result The standard deviation is calculated as follows N 2 XiT x g N where o the standard deviation N the number of measurements taken Keysight InfiniiVision 3000 X Series Oscillosco
287. nu Type Waveform FM Freq Symmetry FM N Ramp 10kHz 100 To set up Frequency Shift Keying Modulation FSK In the Waveform Generator Modulation Menu under Wave Gen gt Settings gt Modulation 266 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Waveform Generator 17 1 Press the Type softkey and turn the Entry knob to select Frequency Shift Keying Modulation FSK 2 Press the Hop Freq softkey and turn the Entry knob to specify the hop frequency The output frequency shifts between the original carrier frequency and this hop frequency 3 Press the FSK Rate softkey and turn the Entry knob to specify the rate at which the output frequency shifts The FSK rate specifies a digital square wave modulating signal The following screen shows an FSK modulation of a 100 kHz sine wave carrier signal 50 005 Channels Naveform Generator Modulation Menu Modulation FSK Rate S 10kHz To restore waveform generator defaults 1 Ifthe Waveform Generator Menu is not currently displayed on the oscilloscope s softkeys press the Wave Gen key 2 In the Waveform Generator Menu press the Settings softkey Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 267 17 Waveform Generator 3 In the Waveform Generator Settings Menu press the Default Wave Gen softkey The waveform generator factory default settings 1 kHz sine wave 500 mVpp O V offset High Z output load are restored
288. o not need to be calibrated but InfiniiMax probes can be calibrated For more information about calibrating probes see To calibrate a probe on page 68 Results User Cal date Change in temperature since last User Cal Failure Comments Probe Cal Status To clean the oscilloscope 1 Remove power from the instrument 2 Clean the external surfaces of the oscilloscope with a soft cloth dampened with a mixture of mild detergent and water 3 Make sure that the instrument is completely dry before reconnecting it to a power source To check warranty and extended services status To learn the warranty status of your oscilloscope 1 Point your web browser to www keysight com find warrantystatus 2 Enter your product s model number and serial number The system will search for the warranty status of your product and display the results If the system cannot find your product s warranty status select Contact Us and speak with a Keysight Technologies representative To contact Keysight Information on contacting Keysight Technologies can be found at www keysight com find contactus Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 301 20 Utility Settings To return the instrument Before shipping the oscilloscope to Keysight Technologies contact your nearest Keysight Technologies sales or service office for additional details Information on contacting Keysight Technologies can be found at www keysight
289. o set up Generic video triggers 162 To trigger on a specific line of video 163 To trigger on all sync pulses 164 To trigger on a specific field of the video signal 165 To trigger on all fields of the video signal 166 To trigger on odd or even fields 167 USB Trigger 170 Serial Trigger 172 11 Trigger Mode Coupling To select the Auto or Normal trigger mode 174 To select the trigger coupling 175 To enable or disable trigger noise rejection 177 To enable or disable trigger HF Reject 177 To set the trigger holdoff 177 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide External Trigger Input 178 Maximum voltage at oscilloscope external trigger input 179 12 Acquisition Control Running Stopping and Making Single Acquisitions Run Control 181 Overview of Sampling 182 Sampling Theory 183 Aliasing 183 Oscilloscope Bandwidth and Sample Rate 183 Oscilloscope Rise Time 185 Oscilloscope Bandwidth Required 186 Memory Depth and Sample Rate 187 Selecting the Acquisition Mode 187 Normal Acquisition Mode 188 Peak Detect Acquisition Mode 188 Averaging Acquisition Mode 191 High Resolution Acquisition Mode 193 Acquiring to Segmented Memory 193 Navigating Segments 195 Measurements Statistics and Infinite Persistence with Segmented Memory 195 Segmented Memory Re ArmTime 196 Saving Data from Segmented Memory 196 13 Cursors To make cursor measurements 200
290. o the source from the probe Increased length in this path will at high frequencies create large common mode voltages at the probe input The voltage generated behaves as if this path were an inductor according to the equation Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Digital Channels 6 di vai Increasing the ground inductance L increasing the current di or decreasing the transition time dt will all result in increasing the voltage V When this voltage exceeds the threshold voltage defined in the oscilloscope a false data measurement will occur Sharing one probe ground with many probes forces all the current that flows into each probe to return through the same common ground inductance of the probe whose ground return is used The result is increased current di in the above equation and depending on the transition time dt the common mode voltage may increase to a level that causes false data generation Probe 1 lt 4 Zin iy L GND Probe Vn Common Mode Ground 4 co hitlari Probe 2 lt 4 Zin Probe N lt Zin O Figure 24 Common Mode Input Voltage Model In addition to the common mode voltage longer ground returns also degrade the pulse fidelity of the probe system Rise time is increased and ringing due to the undamped LC circuit at the input of the probe is also increased Because the Keysight Inf
291. ob to select the channel that is connected to the SPI frame line The label CS or CS for the source channel is automatically set The data pattern and the clock transition must occur during the time when the framing signal is valid The framing signal must be valid for the entire data pattern f the framing signal is set to Timeout the oscilloscope generates it s own internal framing signal after it sees inactivity on the serial clock line Clock Timeout Select Clock Timeout in the Frame by softkey then select the Timeout softkey and turn the Entry knob to set the minimum time that the Clock signal must be idle not transitioning before the oscilloscope will search for the Data pattern on which to trigger The Timeout value can be set anywhere from 100 ns to 10 s Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide I2C SPI Triggering and Serial Decode 25 When you press the Frame by softkey the Display Info graphic changes to show timeout selection or the current state of the chip select signal b Press the Threshold softkey then turn the Entry knob to select the chip select signal threshold voltage level The threshold voltage level is used in decoding and it will become the trigger level when the trigger type is set to the selected serial decode slot When Display Info is enabled information about the selected signal sources and their threshold voltage levels as well as a waveform diagram appe
292. ob to navigate to the save location See To navigate storage locations on page 276 3 Finally press the Press to Save softkey A message indicating whether the save was successful is displayed Mask files have the extension MSK Masks are also saved as part of setup files See To save setup files on page 271 See Also Chapter 15 Mask Testing starting on page 237 To save arbitrary waveforms Arbitrary waveform files can be saved to one of four internal Keysight Flash locations or to an external USB storage device Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 275 18 Save Recall Setups Screens Data 1 Press Save Recall gt Save gt Format then turn the Entry knob to select Arbitrary Waveform data csv 2 Press the softkey in the second position and use the Entry knob to navigate to the save location See To navigate storage locations on page 276 3 Finally press the Press to Save softkey A message indicating whether the save was successful is displayed See Also To edit arbitrary waveforms on page 257 To navigate storage locations When saving or recalling files the softkey in the second position of the Save Menu or Recall Menu along with the Entry knob are used to navigate to storage locations The storage locations can be internal oscilloscope storage locations for setup files or mask files or they can be external storage locations on a connected USB storage
293. obe cable is keyed so you can connect it only one way You do not need to power off the oscilloscope CAUTION A Probe cable for digital channels Use only the Keysight logic probe and accessory kit supplied with the mixed signal oscilloscope see Accessories Available on page 330 3 Connect the ground lead on each set of channels each pod using a probe grabber The ground lead improves signal fidelity to the oscilloscope ensuring accurate measurements 104 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Digital Channels 6 Channel Pod Ground Circuit Ground 4 Connect a grabber to one of the probe leads Other probe leads are omitted from the figure for clarity Grabber 5 Connect the grabber to a node in the circuit you want to test Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 105 6 Digital Channels 6 For high speed signals connect a ground lead to the probe lead connect a grabber to the ground lead and attach the grabber to ground in the device under test Ground Grabber 7 Repeat these steps until you have connected all points of interest Signals Ground 106 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Digital Channels 6 Acquiring waveforms using the digital channels When you press Run Stop or Single to run the oscilloscope the oscilloscope examines the input voltage at each input probe When the trigger
294. ode waveform and in the Lister 4 Ifthe decode line does not appear on the display press the Serial key to turn it on 5 Ifthe oscilloscope is stopped press the Run Stop key to acquire and decode data If the setup does not produce a stable trigger the SPI signal may be slow enough that the oscilloscope is AutoTriggering Press the Mode Coupling key then press the Mode softkey to set the trigger mode from Auto to Normal You can use the horizontal Zoom window for easier navigation of the acquired data Interpreting SPI Decode on page 384 Interpreting SPI Lister Data on page 385 Searching for SPI Data in the Lister on page 385 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 383 25 384 I2C SPI Triggering and Serial Decode Interpreting SP Decode 1 0003 KEYSIGHT TECHNOLOGIES Channels 100 LUCL ULULO LOV FF FF FF FF FF FF PFIED 73 FF FF 02 FF FF FF 02 10 4D 53 4F 03 10 00 00 K 06 05 FFXO2 20 21 Signals Word Size Bit Order Lister 8 MSB Angled waveforms show an active bus inside a packet frame Mid level blue lines show an idle bus The number of clocks in a frame appears in light blue above the frame to the right Decoded hexadecimal data values appear in white Decoded text is truncated at the end of the associated frame when there is insufficient space within frame boundaries Pink vertical bars indicate you need to expand the horizontal scale and run again to
295. of tests may be exceeded If an error occurs the action specified using the On Error softkey will occur The actual number of tests completed is displayed above the softkeys Minimum Time Choose this option and then use the Test Time softkey to select how long the oscilloscope will run When the selected time has passed the oscilloscope will stop The specified time may be exceeded If an error occurs the action specified using the On Error softkey will occur The actual test time is displayed above the softkeys Minimum Sigma Choose this option and then use the Sigma softkey to select a minimum sigma The mask test runs until enough waveforms are tested to achieve a minimum test sigma If an error occurs the oscilloscope will perform the action specified by the On Error softkey Note that this is a test sigma the max achievable process sigma assuming no defects for a certain number of tested waveforms as opposed to a process sigma which is tied to the amount of failures per test The sigma value may exceed the selected value when a small sigma value is chosen The actual sigma is displayed Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Mask Testing 15 On Error The On Error setting specifies the action s to take when the input waveform does not conform to the mask This setting supersedes the Run Until setting Stop The oscilloscope will stop when the first error is detected on the first
296. off The waveform generator output is always disabled when the instrument is first turned on The waveform generator output is automatically disabled if excessive voltage is applied to the Gen Out BNC KEYSIGHT TECHNOLOGIES 253 17 Waveform Generator 2 Inthe Waveform Generator Menu press the Waveform softkey and turn the Entry knob to select the waveform type Waveform Arbitrary Sine Ramp Pulse Frequency QUT Amplitude 1 000kHz 500mYpp 500 05 24 09 KEYSIGHT TEC IES EE Settings atl 3 Depending on the selected waveform type use the remaining softkeys and the Entry knob to set the waveform s characteristics Waveform Characteristics Frequency Max Offset Type Range Amplitude High Z High Z Arbitrary Use the Frequency Frequency Fine Period Period 100 mHz to 20 mVpp to 2 50 V Fine Amplitude High Level and Offset Low Level 12 MHz 5 Vpp softkeys to set the arbitrary waveform signal parameters Use the Edit Waveform softkey to define the arbitrary waveform shape See To edit arbitrary waveforms on page 257 Sine Use the Frequency Frequency Fine Period Period 100 mHz to 20 mVpp to 2 50 V Fine Amplitude High Level and Offset Low Level 20 MHz 5 Vpp softkeys to set the sine signal parameters 204 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Characteristics Frequency Range Waveform Generator WEVA Amplitude High Z Offset
297. old are low 0 If the Thresholds softkey is set to User press the User softkey for the channel group then turn the Entry knob to set the logic threshold There is one User softkey for each group of channels Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Digital Channels 6 To reposition a digital channel 1 Make sure the multiplexed scale and position knobs to the right of the key are selected for digital channels f the arrow to the left of the Digital key is not illuminated press the key 2 Use the multiplexed Select knob to select the channel The selected waveform is highlighted in red 3 Use the multiplexed Position knob to move the selected channel waveform f a channel waveform is repositioned over another channel waveform the indicator at the left edge of the trace will change from Dnn designation where nn is a one or two digit channel number to D The indicates that two channels are overlaid To display digital channels as a bus Digital channels may be grouped and displayed as a bus with each bus value displayed at the bottom of the display in hex or binary You can create up to two buses To configure and display each bus press the Digital key on the front panel Then press the Bus softkey Digital Channel Menu Ds ttt t2ttD e i Sener fal Channel Turn on Turn off Thresholds Bus v D15 D8 D7 DO E s E S Next select a bus Rotate the Entry knob then press the Entry knob
298. omplete screen image in a reduced color 8 bit bitmap format See To save BMP or PNG image files on page 271 24 bit Bitmap image bmp The complete screen image in a 24 bit color bitmap format See To save BMP or PNG image files on page 271 24 bit image png The complete screen image in a 24 bit color PNG format that uses lossless compression Files are much smaller than the BMP format See To save BMP or PNG image files on page 271 CSV data csv This creates a file of comma separated values of all displayed channels and math waveforms This format is suitable for spreadsheet analysis See To save CSV ASCII XY or BIN data files on page 272 ASCII XY data csv This creates separate files of comma separated values for each displayed channel This format is also suitable for spreadsheets See To save CSV ASCII XY or BIN data files on page 272 Reference Waveform data h5 Saves waveform data in a format that can be recalled to one of the oscilloscope s reference waveform locations See To save reference waveform files to a USB storage device on page 275 Multi Channel Waveform data h5 Saves multiple channels of waveform data in a format that can be opened by the N8900A InfiniiView oscilloscope analysis software You can recall the first Analog or Math channel from a multi channel waveform data file Binary data bin This creates a binary file with a header and dat
299. on status on page 301 For other information related to oscilloscope maintenance and service see To clean the oscilloscope on page 301 To check warranty and extended services status on page 301 To contact Keysight on page 301 To return the instrument on page 302 To perform user calibration Perform user calibration Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 297 20 Utility Settings 298 Every two years or after 4000 hours of operation If the ambient temperature is gt 10 C from the calibration temperature If you want to maximize the measurement accuracy The amount of use environmental conditions and experience with other instruments help determine if you need shorter User Cal intervals User Cal performs an internal self alignment routine to optimize the signal path in the oscilloscope The routine uses internally generated signals to optimize circuits that affect channel sensitivity offset and trigger parameters Performing User Cal will invalidate your Certificate of Calibration If NIST National nstitute of Standards and Technology traceability is required perform the Performance Verification procedure in the Keysight InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide using traceable sources To perform user calibration 1 Disconnect all inputs from the front and rear panels including the digital channels cable on an MSO and allow the oscilloscope to warm up be
300. ontal Controls To position the time reference left center right Time reference is the reference point on the display for delay time horizontal position 1 Press Horiz 2 Inthe Horizontal Menu press Time Ref then select Left the time reference is set to one major division from the left edge of the display Center the time reference is set to the center of the display Right the time reference is set to one major division from the right edge of the display A small hollow triangle V at the top of the display grid marks the position of the time reference When delay time is set to zero the trigger point indicator W overlays the time reference indicator The time reference position sets the initial position of the trigger event within acquisition memory and on the display with delay set to 0 Turning the Horizontal scale sweep speed knob expands or contracts the waveform about the time reference point V See To adjust the horizontal time div scale on page 48 Turning the Horizontal position knob in Normal mode not Zoom moves the trigger point indicator W to the left or right of the time reference point V See To adjust the horizontal delay position on page 49 Searching for Events 56 You can use the Search key and menu to search for Edge Pulse Width Rise Fall Time Runt and Serial events on the analog channels Setting up searches see To set up searches on pa
301. ope Another web enabled instrument TECHNOLOGIES from Keysight Technologies MSO X 2024A Information about this Web Enabled Instrument Instrument MSO X 2024A Oscilloscope Serial Number US50210029 Description Agilent ee R DNS Hostname 141 121 237 192 NetBIOS Name a mx2024a 10029 Multicast DNS Hostname a mx2024a 10029 local IP Address 141 121 237 192 VISA TCP IP Connect String TCPIPO 141 121 237 192 INSTR id Advanced information Identification off on Use the navigation bar on the left to access your oscilloscope and related information Keysight Technologies Inc 2006 2014 Browser Web Control The web interface s Browser Web Control page gives you access to The Real Scope Remote Front Panel see Real Scope Remote Front Panel on page 310 The Simple Remote Front Panel see Simple Remote Front Panel on page 311 The Browser Based Remote Front Panel see Browser Based Remote Front Panel on page 312 The SCPI Command window applet for Remote Programming see Remote Programming via the Web Interface on page 313 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 309 21 310 Web Interface If Java is not installed on your PC you will be prompted to install the Sun Microsystems Java Plug in This plug in must be installed on the controlling PC for the web interface s Remote Front Panel or Remote Programming operations The SCPI Command window is usefu
302. ope is stopped when you change a trigger setup the oscilloscope uses the new specification when you press Run Stop or Single If the oscilloscope is running when you change a trigger setup it uses the new trigger definition when it starts r the next acquisition You can use the Force Trigger key to acquire and display data when triggers are not occurring You can use the Mode Coupling key to set options that affect all trigger types see Chapter 11 Trigger Mode Coupling starting on page 173 You can save trigger setups along with the oscilloscope setup see Chapter 18 Save Recall Setups Screens Data starting on page 269 Triggers General A triggered waveform is one in which the oscilloscope begins tracing displaying Information the waveform from the left side of the display to the right each time a particular trigger condition is met This provides stable display of periodic signals such as sine waves and square waves as well as nonperiodic signals such as serial data streams The figure below shows the conceptual representation of acquisition memory You can think of the trigger event as dividing acquisition memory into a pre trigger and post trigger buffer The position of the trigger event in acquisition memory is defined by the time reference point and the delay horizontal position setting see To adjust the horizontal delay position on page 49 o Event Post Trigger Buffer Pr
303. or the Bus1 Bus2 softkey to switch it on Use the Channel softkey and the Entry knob to select individual channels to be included in the bus You can rotate the Entry knob and push it or push the softkey to select channels You can also press the Select Deselect D15 D8 and Select Deselect D7 D0 softkeys to include or exclude groups of eight channels in each bus Digital Bus Menu Ds ttt tttt D Bus Channel Select Deselect Base Bus 1 D15 D8 D7 DO Hex Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 111 Digital Channels Using cursors to read bus values 112 If the bus display is blank completely white or if the display includes you need to expand the horizontal scale to allow space for the data to be shown or use the cursors to display the values see Using cursors to read bus values on page 112 The Base softkey lets you choose to display the bus values in hex or binary The buses are shown at the bottom of the display 2 06V KE TECHNOLOGIES Acquisition Channels z 434 2420 Channel Select Deselect Base D15 D8 D7 DO Binary Bus values can be shown in hex or binary To read the digital bus value at any point using the cursors 1 Turn on Cursors by pressing the Cursors key on the front panel 2 Press the cursor Mode softkey and change the mode to Hex or Binary 3 Press the Source softkey and select Bus1 or Bus2 4 Use the Entry knob and the X1 and X2 softkeys to pos
304. orizontal controls 36 50 horizontal Navigate key 36 horizontal position control 36 horizontal position knob 49 horizontal scale fine adjustment 55 horizontal Search key 36 horizontal sweep speed control 36 horizontal time div control 36 horizontal Zoom key 36 host name 289 307 Host name softkey 290 1 0 interface settings 287 12C serial decode 374 12C trigger 370 12S serial decode 393 I2S trigger 390 identification function web interface 318 idle serial bus 348 356 375 384 394 418 Imped softkey 64 impedance digital probes 114 increment statistics 235 indeterminate state 201 infinite persistence 129 182 189 information area 44 input impedance analog channel input 64 installed licenses 300 installed options 318 instantaneous slope of a waveform 77 Instrument Utilities web page 318 Integrate math function 78 intensity control 127 Intensity key 34 426 interpolate arbitrary waveform option 258 invert graticule colors 272 invert waveform 65 IP address 289 307 Italian front panel overlays 41 Italian user interface and Quick Help 45 J Japanese front panel overlays 41 Japanese user interface and Quick Help 45 K keyboard USB 136 277 283 294 304 keys front panel 33 Keysight IO Libraries Suite 314 knobs front panel 33 Korean front panel overlays 41 Korean user interface and Quick Help 45 L Label key 38 label list 137
305. oscilloscope s system software when updates are available You do not need to take special precautions before removing the USB mass storage device from the oscilloscope you do not need to eject it Simply unplug the USB mass storage device from the oscilloscope when the file operation is complete CAUTION N Do not connect a host computer to the oscilloscope s USB host port Use the device port A host computer sees the oscilloscope as a device so connect the host computer to the oscilloscope s device port on the rear panel See I O Interface Settings on page 287 There is a second USB host port on the back panel 20 Digital channel inputs Connect the digital probe cable to this connector MSO models only See Chapter 6 Digital Channels starting on page 103 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 39 1 40 Getting Started 21 Waveform generator output Outputs sine square ramp pulse DC or noise on the Gen Out BNC Press the Wave Gen key to set up the waveform generator See Chapter 17 Waveform Generator starting on page 253 Front Panel Overlays for Different Languages Front panel overlays which have translations for the English front panel keys and label text are available in 10 languages The appropriate overlay is included when the localization option is chosen at time of purchase To install a front panel overla
306. owing figure shows a typical display with digital channels Delay Time Trigger Trigger time div mode or type and N run status source D Ato a T ZUBY Threshold level Channels J Digital L _ Activity channel indicators identifiers 5 33d 4 72D Channel Turn on Turn off Thresholds Bus D165 D8 D7 DO v w a Waveform Turn Turn Threshold size individual groups of menu key channels channels on off on off 108 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Digital Channels 6 Activity indicator When any digital channels are turned on an activity indicator is displayed in the status line at the bottom of the display A digital channel can be always high always low w or actively toggling logic states f Any channel that is turned off will be grayed out in the activity indicator To change the displayed size of the digital channels 1 Press the Digital key 2 Press the size n M JL softkey to select how the digital channels are displayed The sizing control lets you spread out or compress the digital traces vertically on the display for more convenient viewing To switch a single channel on or off 1 With the Digital Channel Menu displayed rotate the Entry knob to select the desired channel from the popup menu 2 Push the Entry knob or press the softkey that is directly below the popup menu to switch the selected channel on or off To switch all digital channels on or off
307. page 58 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 MIL STD 1553 ARINC 429 Triggering and Serial Decode Setup for MIL STD 1553 Signals 397 MIL STD 1553 Triggering 399 MIL STD 1553 Serial Decode 400 Setup for ARINC 429 Signals 404 ARINC 429 Triggering 405 ARINC 429 Serial Decode 407 MIL STD 1553 ARINC 429 triggering and serial decode requires Option AERO or the DSOX3AERO upgrade The MIL STD 1553 triggering and decode solution supports bi phase MIL STD 1553 signaling by using dual threshold triggering The solution supports the standard 1553 Manchester II encoding data rate of 1 Mb s and word length of 20 bits Setup for MIL STD 1553 Signals MIL STD 1553 signal setup consists of first connecting the oscilloscope to a serial MIL STD 1553 signal using a differential active probe the Keysight N2791A is recommended specifying the signal source and specifying the high and low trigger threshold voltage levels To set up the oscilloscope to capture MIL STD 1553 signals KEYSIGHT TECHNOLOGIES 397 27 MIL STD 1553 ARINC 429 Triggering and Serial Decode 398 Press Label to turn on labels 2 Press Serial 3 Press the Serial softkey turn the Entry knob to select the desired slot Serial 1 or Serial 2 and press the softkey again to enable decode 4 Press the Mode softkey then select MIL STD 1553 decode mode 5
308. pes User s Guide 221 14 Measurements x the ith measurement result x the mean Ratio The Ratio measurement displays the ratio of the AC RMS voltages of two sources expressed in dB Press the Settings softkey to select the source channels for the measurement Time Measurements 222 The following figure shows time measurement points Rise Time Fall Time gt k j k Thresholds L i Lower Width Width he Period gt i The default lower middle and upper measurement thresholds are 10 50 and 90 between Top and Base values See Measurement Thresholds on page 231 for other percentage threshold and absolute value threshold settings Period on page 223 Frequency on page 223 Counter on page 224 Width on page 225 Width on page 225 Burst Width on page 225 Duty Cycle on page 225 Rise Time on page 226 Fall Time on page 226 Delay on page 226 Phase on page 227 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Measurements 14 X at Min Y on page 228 X at Max Y on page 229 Period Period is the time period of the complete waveform cycle The time is measured between the middle threshold points of two consecutive like polarity edges A middle threshold crossing must also travel through the lower and upper threshold levels which eliminates runt pulses The X cursors show what portion of the waveform
309. pes User s Guide 257 17 Waveform Generator Softkey Description Interpolate Specifies how lines are drawn between arbitrary waveform points When enabled lines are drawn between points in the waveform editor Voltage levels change linearly between one point and the next When disabled all line segments in the waveform editor are horizontal The voltage level of one point remains until the next point Source Selects the analog channel or reference waveform to be captured and stored to the arbitrary waveform See Capturing Other Waveforms to the Arbitrary Waveform on page 260 Store Source to Arb Captures the selected waveform source and copy it to the arbitrary waveform See Capturing Other Waveforms to the Arbitrary Waveform on page 260 You can use the Save Recall key and menu to save arbitrary waveforms to one of four internal storage locations or to a USB storage device and you can recall them later See To save arbitrary waveforms on page 275 and To recall arbitrary waveforms on page 279 Creating New Arbitrary Waveforms 258 The New Waveform Menu is opened by pressing Create New in the Edit Waveform Menu Offset To create a new arbitrary waveform 1 In the New Waveform Menu press Initial Pts then use the Entry knob to select the initial number of points in the new waveform The new waveform will be a square wave with the number of points you specify The points are even
310. played you can adjust their skew 1 Display the desired reference waveform see To display a reference waveform on page 100 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 101 5 Reference Waveforms 2 Press the Skew softkey and turn the Entry knob to adjust the reference waveform skew To display reference waveform information Press the Ref key to turn on reference waveforms 2 Inthe Reference Waveform Menu press the Options softkey 3 In the Reference Waveform Options Menu press the Display Info softkey to enable or disable reference waveform information on the oscilloscope display 4 Press the Transparent softkey to enable or disable transparent information backgrounds This setting is also used for other oscilloscope information on the display like mask test statistics etc To save recall reference waveform files to from a USB storage device Analog channel math or reference waveforms can be saved to a reference waveform file on a USB storage device See To save reference waveform files to a USB storage device on page 275 You can recall a reference waveform file from a USB storage device into one of the reference waveform locations See To recall reference waveform files from a USB storage device on page 278 102 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels To connect the digi
311. pture CAUTION When you create a new arbitrary waveform the existing arbitrary waveform definition is overwritten Note that you can use the Save Recall key and menu to save arbitrary waveforms to one of four internal storage locations or to a USB storage device and you can recall them later See To save arbitrary waveforms on page 275 and To recall arbitrary waveforms on page 279 260 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Waveform Generator 17 The source waveform is decimated into 8192 maximum or fewer arbitrary waveform points If the source waveform frequency and or voltage exceed the capabilities of the waveform generator the arbitrary waveform will be limited to the capabilities of the waveform generator For example a 20 MHz waveform captured as the arbitrary waveform becomes a 12 MHz waveform To output the waveform generator sync pulse 1 If the Waveform Generator Menu is not currently displayed on the oscilloscope s softkeys press the Wave Gen key 2 In the Waveform Generator Menu press the Settings softkey 3 In the Waveform Generator Settings Menu press the Trig Out softkey and turn the Entry knob to select Waveform Generator Sync Pulse Waveform Type Sync Signal Characteristics All waveforms except The Sync signal is a TTL positive pulse that occurs when the waveform DC Noise and rises above zero volts or the DC offset value Cardiac DC Noise a
312. put channel source 4 Press the Slope softkey to specify the slope of the edge 5 Press the Idle softkey then turn the Entry knob to specify the idle time 6 Press the Edge softkey then turn the Entry knob to which edge number to trigger on Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Triggers 10 Runt Trigger The Runt trigger looks for pulses that cross one threshold but not another High level Low level Positive runt pulse Negative runt pulse A positive runt pulse crosses through a lower threshold but not an upper threshold A negative runt pulse crosses through an upper threshold but not a lower threshold To trigger on runt pulses 1 Press the Trigger key 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select Runt Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 155 10 Triggers 200 08 319 IGHT EYSIGH TECHNOLOGIES Acquisition Channels Source Level Select Qualifier Low gt 3 Press the Source softkey and turn the Entry knob to select the input channel source 4 Press the Positive Negative or Either Runt Pulse softkey to toggle between pulse types 5 Press the Level Select softkey to select High then turn the Trigger Level knob to adjust the high level 6 Press the Level Select softkey to select Low then turn the Trigger Level knob to adjust the low level You can also push the Trigger Level k
313. r condition Word Start triggers on the start of a word Word Stop triggers at the end of a word Label triggers on the specified label value Label Bits triggers on the label and the other word fields as specified Label Range triggers on a label following in a min max range Parity Error triggers on words with a parity error Word Error triggers on an intra word coding error Gap Error triggers on an inter word gap error Word or Gap Error triggers on either a Word or Gap Error All Errors triggers on any of the above errors All Bits Eye triggers on any bit which will therefore form an eye diagram All 0 Bits triggers on any bit with a value of zero All 1 Bits triggers on any bit with a value of one 4 Ifyou select the Label or Label Bits condition use the Label softkey to specify the label value Label values are always displayed in octal If you select the Label Bits condition use the Bits softkey and submenu to specify the bit values SDI XX Data 000 0000 0000 0000 OOO Define Bit 0 1 4 Set all Bits Data 12 1 Use the Define softkey to select SDI Data or SSM The SDI or SSM selections may not be available depending on word format selection in the Serial Decode Menu Use the Bit softkey to select the bit to change Use the 01 X softkey to set the bit value Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide
314. r 4 on 4 channel models FFT units will be displayed in dBV when channel units is set to Volts and channel impedance is set to 1 MQ FFT units will be displayed in dBm when channel units is set to Volts and channel impedance is set to 50Q FFT units will be displayed as dB for all other FFT sources or when a source channel s units has been set to Amps FFT DC Value The FFT computation produces a DC value that is incorrect It does not take the offset at center screen into account The DC value is not corrected in order to accurately represent frequency components near DC FFT Aliasing When using FFTs it is important to be aware of frequency aliasing This requires that the operator have some knowledge as to what the frequency domain should contain and also consider the sampling rate frequency span and oscilloscope vertical bandwidth when making FFT measurements The FFT resolution the quotient of the sampling rate and the number of FFT points is displayed directly above the softkeys when the FFT Menu is displayed Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 85 4 86 Math Waveforms Nyquist Frequency and Aliasing in the Frequency Domain The Nyquist frequency is the highest frequency that any real time digitizing oscilloscope can acquire without aliasing This frequency is half of the sample rate Frequencies above the Nyquist frequency will be under sampled which causes aliasing The Nyquist freque
315. r Settings Menu which lets you make other settings related to the waveform generator See To output the waveform generator sync pulse on page 261 To specify the expected output load on page 261 To use waveform generator logic presets on page 262 To add noise to the waveform generator output on page 262 To restore waveform generator defaults on page 267 256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Waveform Generator 17 To edit arbitrary waveforms 1 When Arbitrary is selected as the generated waveform type see To select generated waveform types and settings on page 253 press the Edit Waveform softkey to open the Edit Waveform Menu 100 08 Auto o 0V KEYSIGHT TECHN S 1 00ms Channels 250mV 1 00ms Edit Existing Interpolate Source When you open the Edit Waveform Menu you see the existing arbitrary waveform definition The voltage and time period you see in the diagram are the bounding parameters they come from the frequency and amplitude settings in the main Waveform Generator Menu 2 Use the softkeys in the Edit Waveform Menu to define the shape of the arbitrary waveform Softkey Description Create New Opens the New Waveform Menu See Creating New Arbitrary Waveforms on page 258 Edit Existing Opens the Edit Waveform Points Menu See Editing Existing Arbitrary Waveforms on page 259 Keysight InfiniiVision 3000 X Series Oscillosco
316. r erased Use infinite persistence to measure noise and jitter to see the worst case extremes of varying waveforms to look for timing violations or to capture events that occur infrequently Variable Persistence Results of previous acquisitions are erased after a certain amount of time Variable persistence gives you a view of acquired data that is similar to analog oscilloscopes When variable persistence is selected press the Time softkey and use the Entry knob to specify the amount of time that previous acquisitions are to be displayed The display will begin accumulating multiple acquisitions Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 129 8 Display Settings 3 To erase the results of previous acquisitions from the display press the Clear Persistence softkey The oscilloscope will start to accumulate acquisitions again 4 To return the oscilloscope to the normal display mode turn off persistence then press the Clear Persistence softkey Turning off persistence does not clear the display The display is cleared if you press the Clear Display softkey or if you press the AutoScale key which also turns off persistence For another method of seeing worst case extremes of varying waveforms see Glitch or Narrow Pulse Capture on page 189 To clear the display 1 Press Display gt Clear Display You can also configure the Quick Action key to clear the display See Configuring
317. rator Menu is not currently displayed on the oscilloscope s softkeys press the Wave Gen key 2 In the Waveform Generator Menu press the Settings softkey 3 In the Waveform Generator Settings Menu press the Modulation softkey 4 In the Waveform Generator Modulation Menu Waveform Generator Modulation Menu Modulation Waveform AM Freq AM Depth Symmetry Ramp 10kHz 50 1 Press the Modulation softkey to enable or disable modulated waveform generator output You can enable modulation for all waveform generator function types except pulse DC and noise Press the Type softkey and turn the Entry knob to select the modulation type Amplitude Modulation AM the amplitude of the original carrier signal is modified according to the amplitude of the modulating signal See To set up Amplitude Modulation AM on page 264 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 263 17 Waveform Generator 264 Frequency Modulation FM the frequency of the original carrier signal is modified according to the amplitude of the modulating signal See To set up Frequency Modulation FM on page 265 Frequency Shift Keying Modulation FSK the output frequency shifts between the original carrier frequency and a hop frequency at the specified FSK rate The FSK rate specifies a digital square wave modulating signal See To set up Frequency Shift Keying Modulation FSK on page 266 To set up Amplitude
318. rce in the Measurement Menu 3 Choose Maximum and X at Max Y measurements Maximum units are in dB and X at Max Y units are in Hertz for FFT Count Measurements Positive Pulse Count on page 229 Negative Pulse Count on page 230 Rising Edge Count on page 230 Falling Edges Count on page 230 Positive Pulse Count The Positive Pulse Count measurement is a pulse count for the selected waveform source Positive pulse count ee This measurement is available for analog channels Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 229 14 Measurements Negative Pulse Count The Negative Pulse Count measurement is a pulse count for the selected waveform source WUE tt ttt Negative pulse count This measurement is available for analog channels Rising Edge Count The Rising Edge Count measurement is an edge count for the selected waveform source This measurement is available for analog channels Falling Edges Count The Falling Edges Count measurement is an edge count for the selected waveform source This measurement is available for analog channels Mixed Measurements Area on page 230 Area Area measures the area between the waveform and the ground level Area below the ground level is subtracted from area above the ground level 230 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Measurements 14 Positive area I lt 4 gt i N cycles Negative
319. rd This is referred to as leakage In order to minimize spectral leakage windows that approach zero smoothly at the beginning and end of the signal are employed as filters to the FFT The FFT Menu provides four windows Hanning Flat Top Rectangular and Blackman Harris For more information on leakage see Keysight Application Note 243 The Fundamentals of Signal Analysis at http literature cdn keysight com litweb pdf 5952 8898E pdf Square Root Square root V calculates the square root of the selected source Where the transform is undefined for a particular input holes zero values appear in the function output Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 87 4 Math Waveforms 100 08 KEY TECHN Acquisition Channels easurements f MHM HIH LUIM WH EEE D orm Math Menu tit Ch1 Function Operator Source 1 tit a Figure 10 Example of V Square Root See Also To perform transforms or filters on an arithmetic operation on page 72 Units for Math Waveforms on page 73 Ax B The Ax B function available with the DSOX3ADVMATH advanced math measurements license lets you apply a gain and offset to an existing input source 88 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Math Waveforms 4 200 05 2 49V KEYSIGHT TECHNO IES Acquisition 1 Channels x Measurements Pk Pk 3 520mv Pk Pk M Source 1 Gai Offset B 250
320. region is indicated by y values of MIN for the first and last points The top region must be the lowest numbered region in the file The bottom region must be the highest numbered region in the file Region number 1 is the top mask region The vertices in Region 1 describe points along a line that line is the bottom edge of the top portion of the mask Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Mask Testing 15 Similarly the vertices in Region 2 describe the line that forms the top of the bottom part of the mask The vertices in a mask file are normalized There are four parameters that define how values are normalized X1 AX Y1 Y2 These four parameters are defined in the Oscilloscope Setup portion of the mask file The Y values normally voltage are normalized in the file using the following equation Yhorm Y Y1 AY where AY Y2 Y1 To convert the normalized Y values in the mask file to voltage Y Ynorm AYY where AY Y2 Y1 The X values normally time are normalized in the file using the following equation Xnorm X X1 AX To convert the normalized X values to time X Xnorm AX X1 Oscilloscope The keywords setup and end_setup appearing alone on a line define the Setup Information beginning and end of the oscilloscope setup region of the mask file The oscilloscope setup information contains remote programming language commands that the oscilloscope executes
321. rence Table 8 Current Probes continued Description Quantity Supported 1See AutoProbe Interface on page 326 Accessories Available In addition to passive proves Passive Probes on page 327 single ended active probes Single Ended Active Probes on page 327 differential probes Differential Probes on page 328 and current probes Current Probes on page 329 the following accessories are available for the InfiniiVision 3000 X Series oscilloscopes Table9 Accessories Available for InfiniiVision 3000 X Series Oscilloscopes Model Part Description DSOXLAN LAN VGA connection module DSOXGPIB GPIB connection module N6456A Rack mount kit N6457A Soft carrying case and front panel cover N2786A 2 leg probe positioner N2787A 3D probe positioner 1180CZ Testmobile N6459A Hardcopy user s guide various Front panel overlays see Front Panel Overlays for Different Languages on page 40 N6450 60001 16 channel logic probe and accessory kit standard with MSO models and with MSO upgrade 01650 61607 Logic cable and terminator 40 pin to 40 pin MSO cable You can find these items at www keysight com or at www parts keysight com For information on more probes and accessories see www keysight com for 330 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Reference 22 Probes and Accessories Selection Guide 5989 6162EN 5000
322. resolving two frequencies that are close together Flat Top window for making accurate amplitude measurements of frequency peaks Rectangular good frequency resolution and amplitude accuracy but use only where there will be no leakage effects Use on self windowing waveforms such as pseudo random noise impulses sine bursts and decaying sinusoids Blackman Harris window reduces time resolution compared to a rectangular window but improves the capacity to detect smaller impulses due to lower secondary lobes Vertical Units lets you select Decibels or V RMS as the units for the FFT vertical scale Auto Setup sets the frequency Span and Center to values that will cause the entire available spectrum to be displayed The maximum available frequency is half the FFT sample rate which is a function of the time per division setting The FFT resolution is the quotient of the sampling rate and the number of FFT points fs N The current FFT Resolution is displayed above the softkeys Scale and offset considerations If you do not manually change the FFT scale or offset settings when you turn the horizontal scale knob the span and center frequency settings will automatically change to allow optimum viewing of the full spectrum If you do manually set scale or offset turning the horizontal scale knob will not change the span or center frequency settings allowing you see better detail around a specific frequency
323. ress the second softkey and select a destination folder on your USB mass storage device 7 Press the Press to Save softkey This creates an ASCII text file that describes the mask 8 Remove the USB mass storage device and connect it to a PC 9 Open the msk file your created using a text editor such as Wordpad 10 Edit save and close the file The mask file contains the following sections Mask File Identifier Mask Title Mask Violation Regions Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 243 15 Mask Testing Mask File Identifier 244 Mask Title Mask Violation Regions Oscilloscope Setup Information The Mask File Identifier is MASK_FILE_548XxX The Mask Title is a string of ASCII characters Example autoMask CH1 OCT 03 09 40 26 2008 When a mask file contains the keyword autoMask in the title the edge of the mask is passing by definition Otherwise the edge of the mask is defined as a failure 50 008 Trig d 2 00V KEYSIGHT TEC IES Region 1 Region 2 Up to 8 regions can be defined for a mask They can be numbered 1 8 They can appear in any order in the msk file The numbering of the regions must go from top to bottom left to right An Automask file contains two special regions the region glued to the top of the display and the region that is glued to the bottom The top region is indicated by y values of MAX for the first and last points The bottom
324. rigger Setup Menu Note that the setting of this softkey does not affect the selected base of the decode display 5 Press the Trigger softkey and set up the desired trigger condition Rx Start Bit The oscilloscope triggers when a start bit occurs on Rx Rx Stop Bit Triggers when a stop bit occurs on Rx The trigger will occur on the first stop bit This is done automatically whether the device under test uses 1 1 5 or 2 stop bits You do not need to specify the number of stop bits used by the device Under test Rx Data Triggers on a data byte that you specify For use when the device under test data words are from 5 to 8 bits in length no 9th alert bit Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 415 28 UART RS232 Triggering and Serial Decode 416 Rx 1 Data For use when length including the alert bit is 1 The specified data byte applies to the least significant 8 bi excludes the 9th alert bit Rx 0 Data For use when the device under test da the device under test da ta words are 9 bits in bit the 9th bit Triggers only when the 9th alert oo US ta words are 9 bits in length including the alert bit the 9th bit Triggers only when the 9th alert bit is O The specified dat excludes the 9th alert bit Rx X Data For use when the device under test da a byte applies to the least significant 8 bi a US ta words are 9 b
325. rigger type pulse width 145 trigger type rise fall time 152 trigger type RS232 415 trigger type runt 155 trigger type setup and hold 157 trigger type slope 142 trigger type SPI 381 trigger type UART 415 trigger type USB 170 trigger type video 158 trigger types 139 trigger definition 140 trigger external 178 trigger forcing a 141 trigger general information 140 trigger holdoff 177 trigger mode coupling 173 trigger source 142 triggers TRIG OUT signal 296 TTL threshold 110 turn channel on 38 U UART totalizer 419 UART trigger 415 UART RS232 frame counter 419 UART RS232 license 331 UART RS232 serial decode 417 under sampled signals 183 units cursor 201 429 Index units external trigger probe 179 units math 73 units probe 67 updating software and firmware 333 upgrade files 319 upgrade options 331 upgrading the oscilloscope 333 upload new firmware 307 usb 293 USB device port 43 USB device port remote control 287 USB host port 43 281 USB host ports 39 USB printer 281 USB printers supported 281 USB storage device 39 USB CD device 293 USB eject device 39 USB storage device numbering 293 USB trigger type 170 usb2 293 user cal 297 user calibration 297 user interface language 45 User defined threshold 110 utilities 287 Utility key 35 V V RMS FFT vertical units 82 values choosing 35 variable per
326. ring 130 persistence infinite 182 Phase measurement 213 227 phase X cursor units 202 PNG file format 270 point to point connection 290 Polish front panel overlays 41 pollution degree 325 pollution degree definitions 326 Portuguese front panel overlays 41 Portuguese user interface and Quick Help 45 position digital channels 111 position knob 111 position analog 63 positive pulse width measurement trend 95 post processing 209 post trigger information 49 Power App measurements 213 power consumption 27 power cord connector 42 power requirements 27 power supply 42 power switch 28 34 power on 27 predefined labels 134 Preshoot measurement 213 219 Press to go 276 292 pre trigger information 49 print 302 Print key 38 427 Index print options 284 print screen 281 print landscape 284 print Quick Print 302 printer USB 39 281 printing the display 281 probe attenuation 67 probe attenuation external trigger 179 probe compensation 39 probe head 68 probe units 67 probe AutoProbe interface 39 probe calibrate 68 probes 326 330 probes connecting to oscilloscope 28 probes current 329 probes differential 328 probes digital 103 probes passive 327 probes passive compensating 32 probes single ended active 327 programmer s guide 315 pulse polarity 146 pulse waveform generator output 255 pulse width trigger 145 PWR license 332 Q
327. ring are displayed in the MOSI MISO Data string in the waveform area 6 For each bit in the MOSI MISO Data string a Press the Bit softkey then turn the Entry knob to select the bit location As you rotate the Entry knob the bit is highlighted in the Data string shown in the waveform area b Press the 01 X softkey to set the bit selected in the Bit softkey to 0 low 1 high or X don t care Channels Trigger Paint al TED E 00 0011 0 ta 00 10110 FF FF FF O02 4FF FF FF 03 10 00 001 06 05 FFXO2 20 21 Set all Bits Display Info 1 The Set all Bits softkey will set all bits in the data string to the value of the 01 X softkey For SPI decode information see SP Serial Decode on page 383 382 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide I2C SPI Triggering and Serial Decode 25 SPI Serial Decode See Also To set up the oscilloscope to capture SPI signals see Setup for SPI Signals on page 377 For SPI triggering setup see SPI Triggering on page 381 To set up SPI serial decode 1 Press Serial to display the Serial Decode Menu Serial Decode Menu Mode Word Size Bit Order SPI MSB 2 Press the Word Size softkey then turn the Entry knob to select the number of bits in a word 3 Press the Bit Order softkey then turn the Entry knob to select the bit order most significant bit first MSB or least significant bit first LSB used when displaying data in the serial dec
328. rms DC offsets are larger than the dynamic range of the oscilloscope s input channels you will need to use a differential probe instead 74 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Math Waveforms 4 500 05 2 08V KEYSIGHT TECHNO IES Acquisition Channels m Math Menu f t Ch1 Ch2 i Operator Source 1 Source 2 Figure 5 Example of Subtract Channel 2 from Channel 1 See Also Units for Math Waveforms on page 73 Multiply or Divide When you select the multiply or divide math function the Source 1 and Source 2 values are multiplied or divided point by point and the result is displayed The divide by zero case places holes that is zero values in the output waveform Multiply is useful for seeing power relationships when one of the channels is proportional to the current Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 75 4 Math Waveforms 500 05 o 2 08V KEYSIGHT TECHNO IES Acquisition Channels m Math Menu f t Ch1 Ch2 Operator Source 1 Source 2 Figure 6 Example of Multiply Channel 1 by Channel 2 See Also Units for Math Waveforms on page 73 Math Transforms Math transforms perform a transform function like differentiate integrate FFT or square root on an analog input channel or on the result of an arithmetic operation Differentiate on page 77 Integrate on page 78 FFT Measurement on page 81 Square Root on page 8
329. rn the Front Panel Controls and Connectors 33 Front Panel Overlays for Different Languages 40 Learn the Rear Panel Connectors 41 Learn the Oscilloscope Display 43 Access the Built In Quick Help 44 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls To adjust the horizontal time div scale 48 To adjust the horizontal delay position 49 Panning and Zooming Single or Stopped Acquisitions 50 To change the horizontal time mode Normal XY or Roll 50 XY Time Mode 51 To display the zoomed time base 54 To change the horizontal scale knob s coarse fine adjustment setting 55 To position the time reference left center right 56 Searching for Events 56 To set up searches 57 To copy search setups 57 Navigating the Time Base 58 To navigate time 58 To navigate search events 58 To navigate segments 59 3 Vertical Controls To turn waveforms on or off channel or math 62 To adjust the vertical scale 63 To adjust the vertical position 63 To specify channel coupling 63 To specify channel input impedance 64 To specify bandwidth limiting 65 To change the vertical scale knob s coarse fine adjustment setting 65 To invert a waveform 65 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Setting Analog Channel Probe Options 66 To specify the channel units 67 To specify the probe attenuation 67 To specify the probe
330. rtical Controls 3 See Also For more information on probing visit www keysight com find scope_probes Information about selecting a probe can be found in document number Keysight Oscilloscope Probes and Accessories Selection Guide part number 5989 6162EN available at www keysight com To specify bandwidth limiting 1 Press the desired channel key 2 Inthe Channel Menu press the BW Limit softkey to enable or disable bandwidth limiting When bandwidth limit is on the maximum bandwidth for the channel is approximately 20 MHz For waveforms with frequencies below this turning bandwidth limit on removes unwanted high frequency noise from the waveform The bandwidth limit also limits the trigger signal path of any channel that has BW Limit turned on To change the vertical scale knob s coarse fine adjustment setting 1 Push the channel s vertical scale knob or press the channel key and then the Fine softkey in the Channel Menu to toggle between fine and coarse adjustment of the vertical scale When Fine adjustment is selected you can change the channel s vertical sensitivity in smaller increments The channel sensitivity remains fully calibrated when Fine is on The vertical scale value is displayed in the status line at the top of the display When Fine is turned off turning the volts division knob changes the channel sensitivity in a 1 2 5 step sequence To invert a waveform 1 Press the desired channel key
331. s 225 burst capture signal bursts 193 bus display mode 111 buttons keys front panel 33 BW Limit in DVM display 250 C calibrate probe 68 calibration 297 calibration protect button 41 42 calibration status 319 CAN decode source channels 344 CAN frame counter 349 CAN serial decode 347 CAN totalizer 349 CAN trigger 345 capture signal bursts 193 cardiac waveform generator output 255 Center FFT 81 channel labels 133 channel analog 61 channel bandwidth limit 65 channel coupling 63 channel invert 65 channel on off keys 38 channel position 63 channel probe units 67 channel skew 68 channel vernier 65 channel vertical sensitivity 63 characteristics 323 chart logic bus state math function 97 chart logic bus timing math function 96 choosing values 35 cleaning 301 clear display 190 clear display Quick Clear Display 303 clear persistence 130 clock 296 CMOS threshold 110 common logarithm math function 90 COMP license 331 compensate passive probes 32 39 Config softkey 289 290 connect probes digital 103 424 connection to a PC 290 connectors rear panel 41 control remote 287 controls front panel 33 copyright 2 Counter measurement 224 counter ARINC 429 words errors 410 counter CAN frame 349 counter FlexRay frame 365 counter UART RS232 frame 419 coupling channel 63 coupling trigger 175 crosstalk problems 81 CSV file format 270
332. s Input protection is enabled in 50 Q mode and the 50 Q load will disconnect if greater than 5 Vrms is detected However the inputs could still be damaged depending on the time constant of the signal The 50 Q input protection only functions when the oscilloscope is powered on With 10073C 10 1 probe CAT I 500 Vpk With N2862A or N2863A 10 1 probe 300 Vrms CAUTION A Maximum input voltage at digital channels 40 V peak CAT transient overvoltage 800 Vpk Environmental Conditions Environment Indoor use only Ambient temperature Operating 0 C to 55 C non operating 40 C to 71 C Humidity Operating Up to 80 RH at or below 40 C Up to 45 RH up to 50 C Non operating Up to 95 RH up to 40 C Up to 45 RH up to 50 C Altitude Operating and non operating to 4 000 m 13 123 ft Overvoltage Category This product is intended to be powered by MAINS that comply to Overvoltage Category Il which is typical of cord and plug connected equipment Pollution Degree The InfiniiVision 2000 3000 X Series oscilloscopes may be operated in environments of Pollution Degree 2 or Pollution Degree 1 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 325 22 Reference Pollution Degree Pollution Degree 1 No pollution or only dry non conductive pollution occurs Definitions The pollution has no influence Example A clean room or climate controlled office
333. s filter with the 3 dB point at 50 kHz in series with the trigger waveform 0 dB 3 dB down point Pass Band DC 50 kHz Low frequency reject removes any unwanted low frequency components from a trigger waveform such as power line frequencies etc that can interfere with proper triggering Use LF Reject coupling to get a stable edge trigger when your waveform has low frequency noise TV coupling is normally grayed out but is automatically selected when TV trigger is enabled in the Trigger Menu Note that Trigger Coupling is independent of Channel Coupling see To specify channel coupling on page 63 176 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Trigger Mode Coupling 11 To enable or disable trigger noise rejection Noise Rej adds additional hysteresis to the trigger circuitry By increasing the trigger hysteresis band you reduce the possibility of triggering on noise However this also decreases the trigger sensitivity so that a slightly larger signal is required to trigger the oscilloscope Press the Mode Coupling key 2 Inthe Trigger Mode and Coupling Menu press the Noise Rej softkey to enable or disable To enable or disable trigger HF Reject HF Reject adds a 50 kHz low pass filter in the trigger path to remove high frequency components from the trigger waveform 3 dB DC 50 kHz You can use HF Reject to remove high frequency noise such as AM or
334. s greater than the channel s trigger level or threshold level X sets the pa don t care is ttern to don t care on the selected channel Any channel set to ignored and is not used as part of the pattern However if all channels in the pattern are set to don t care the oscilloscope will not trigger Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 149 10 Triggers 150 Hex Bus The rising edge or falling edge t softkey sets the pattern to an edge on the selected channel Only one rising or falling edge can be specified in the pattern When an edge is specified the oscilloscope will trigger at the edge specified if the pattern set for the other channels is true If no edge is specified the oscilloscope will trigger on the last edge that makes the pattern true Specifying an Edge in a Pattern You are allowed to specify only one rising or falling edge term in the pattern If you define an edge term then select a different channel in the pattern and define another edge term the previous edge definition is changed to a don t care Pattern Trigger You can specify a bus value on which to trigger To do this first define the bus See To display digital channels as a bus on page 111 for details You can trigger on a bus value whether you are displaying the bus or not To trigger on a bus value 1 Press the Pattern key on the front panel 2 Press the Channel softkey and rotate the Entry knob
335. s softkey to set the ID length ID value number of data bytes and data value Error Frame Finds CAN active error frames All Errors Finds any form error or active error Overload Frame Finds CAN overload frames For more information on searching data see Searching Lister Data on page 124 For more information on using the Navigate key and controls see Navigating the Time Base on page 58 Setup for LIN Signals LIN Local Interconnect Network signal setup consists of connecting the oscilloscope to a serial LIN signal specifying the signal source threshold voltage level baud rate sample point and other LIN signal parameters To set up the oscilloscope to capture LIN signals 1 Press Label to turn on labels 2 Press Serial Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 351 23 CAN LIN Triggering and Serial Decode 3 Press the Serial softkey turn the Entry knob to select the desired slot Serial 1 or Serial 2 and press the softkey again to enable decode 4 Press the Mode softkey then select LIN trigger type 5 Press the Signals softkey to open the LIN Signals Menu Threshold Baud Rate Sample Po Standard Syne Break 1 12V 60 0 LIN 1 3 ia 6 Press the Source softkey to select the channel connected to the LIN signal line The label for the LIN source channel is automatically set 7 Press the Threshold softkey then turn the Entry knob to set the LIN signal threshol
336. scope channels This could be for example an InfiniiMax probe amplifier probe head with attenuators attached 2 Connect the probe to the left side Demo 2 Probe Comp terminal and the probe ground to the ground terminal When calibrating a differential probe connect the positive lead to the Probe Comp terminal and the negative lead to the ground terminal You may need to connect an alligator clip to the ground lug to allow a differential probe to span between the Probe Comp test point and ground A good ground connection ensures the most accurate probe calibration Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Vertical Controls 3 3 Press the Channel on off key to turn the channel on if the channel is off 4 Inthe Channel Menu press the Probe softkey 5 Inthe Channel Probe Menu the second softkey from the left is for specifying your probe head and attenuation Repeatedly press this softkey until the probe head selection matches the attenuator you are using The choices are 10 1 single ended browser no attenuator 1 differential browser no attenuator 6 dB Atten single ended browser 6 dB Atten differential browser 12 dB Atten single ended browser 1 12 dB Atten differential browser 20 dB Atten single ended browser 10 1 20 dB Atten differential browser 6 Press the Calibrate Probe softkey and follow the instructions on the display a a a 4 CDi SO E N e ED SOR A oR o
337. se 54 Normal lt window Signal anomaly expanded in zoom window Zoom L window Zoom formerly called Delayed sweep mode is a horizontally expanded version of the normal display When Zoom is selected the display divides in half The top half of the display shows the normal time div window and the bottom half displays a faster Zoom time div window The Zoom window is a magnified portion of the normal time div window You can use Zoom to locate and horizontally expand part of the normal window for a more detailed higher resolution analysis of signals To turn on or off Zoom 1 Press the zoom key or press the Horiz key and then the Zoom softkey These markers show the Time div Time div Delay time beginning and end of the for zoomed for normal momentarily displays Zoom window window window when the Horizontal position knob is turned 2 21V KEYSIGHT TECHNOLOGIES Acquisition h Channels Time Ref Normal Center Select Zoom Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Horizontal Controls 2 The area of the normal display that is expanded is outlined with a box and the rest of the normal display is ghosted The box shows the portion of the normal sweep that is expanded in the lower half To change the time div for the Zoom window turn the horizontal scale sweep speed knob As you turn the knob the zoomed window time div is highlighted in the status line above th
338. select Time 3 Press the 1010 navigation keys to play backward stop or play forward in time You can press the or keys multiple times to speed up the playback There are three speed levels To navigate search events When acquisitions are stopped you can use the navigation controls to go to found search events set using the Search key and menu see Searching for Events on page 56 1 Press Navigate 58 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Horizontal Controls 2 2 Inthe Navigate Menu press Navigate then select Search 3 Press the OFS back and forward keys to go to the previous or next search event When searching Serial decode You can press the stop key to set or clear a mark The Auto zoom softkey specifies whether the waveform display is automatically zoomed to fit the marked row as you navigate Pressing the Scroll Lister softkey lets you use the Entry knob to scroll through data rows in the Lister display To navigate segments When the segmented memory acquisition is enabled and acquisitions are stopped you can use the navigation controls to play through the acquired segments 1 Press Navigate 2 Inthe Navigate Menu press Navigate then select Segments 3 Press Play Mode then select Manual to play through segments manually In the Manual play mode Press the OIC back and forward keys to go to the previous or next segment Press the Id sof
339. select the number of the line you want to examine Alternate Triggering If Line Alternate is selected the oscilloscope will alternately trigger on the selected line number in Field 1 and Field 2 This is a quick way to compare the Field 1 VITS and Field 2 VITS or to check for the correct insertion of the half line at the end of Field 1 Line F1 Channels Field Holdotf Line Fld1 O Oflds Figure 27 Example Triggering on Line 136 To trigger on all sync pulses 164 To quickly find maximum video levels you could trigger on all sync pulses When All Lines is selected as the Video trigger mode the oscilloscope will trigger on all horizontal sync pulses 1 Press the Trigger key 2 In the Trigger Menu press the Trigger softkey then turn the Entry knob to select Video Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Triggers 10 3 Press the Settings softkey then press the Standard softkey to select the appropriate TV standard 4 Press the Mode softkey and select All Lines 20 005 Vid Lines KEYSIGHT TECHNOLOGIES Channels Field Holdoff All Lines O Oflds Figure 28 Triggering on All Lines To trigger on a specific field of the video signal To examine the components of a video signal trigger on either Field 1 or Field 2 available for interleaved standards When a specific field is selected the oscilloscope triggers on the rising edge of the first serration pulse in the vertical sync
340. ser s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode Serial Decode Options 121 Lister 122 Searching Lister Data 124 Triggeringon In some cases such as when triggering on a slow serial signal for example 12C Serial Data SPI CAN LIN etc it may be necessary to switch from the Auto trigger mode to the Normal trigger mode to prevent the oscilloscope from Auto triggering and stabilize the display You can select the trigger mode by pressing the Mode Coupling key then the Mode softkey Also the threshold voltage level must be set appropriately for each source channel The threshold level for each serial signal can be set in the Signals Menu Press the Serial key then the Signals softkey Serial Decode Options Keysight s hardware accelerated serial decode options can be installed when the oscilloscope is manufactured or added later The following serial decode licenses are available With the DSOX3AUTO license you can decode CAN Controller Area Network and LIN Local Interconnect Network serial buses See CAN Serial Decode on page 347 LIN Serial Decode on page 354 With the DSOX3FLEX license you can decode FlexRay serial buses See FlexRay Serial Decode on page 363 KEYSIGHT TECHNOLOGIES 121 7 Serial Decode Lister 122 With the DSOX3EMBD license you can decode 12C Inter IC and SPI Serial Peripheral Interface serial buses See I2C S
341. sholds on page 231 To copy search setups Except for Serial event search setups you can copy search setups to trigger setups and vice versa 1 Press Search 2 Inthe Search Menu press Search then turn the Entry knob to select the search type 3 Press Copy 4 Inthe Search Copy Menu Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 57 2 Horizontal Controls Press Copy to Trigger to copy the setup for the selected search type to the same trigger type For example if the current search type is Pulse Width pressing Copy to Trigger copies the search settings to the Pulse Width trigger settings and selects the Pulse Width trigger Press Copy from Trigger to copy the trigger setup for the selected search type to the search setup To undo a copy press Undo Copy The softkeys in the Search Copy Menu may not be available when one of the settings cannot be copied or there is no trigger type that corresponds to the search type Navigating the Time Base You can use the Navigate key and controls to navigate through Captured data see To navigate time on page 58 Search events see To navigate search events on page 58 Segments when segmented memory acquisitions are turned on see To navigate segments on page 59 To navigate time When acquisitions are stopped you can use the navigation controls to play through the captured data 1 Press Navigate 2 Inthe Navigate Menu press Navigate then
342. sion It is possible to set Span above the maximum available frequency in which case the displayed spectrum will not take up the whole screen Press the Span softkey then turn the Entry knob to set the desired frequency span of the display Center sets the FFT spectrum frequency represented at the center vertical grid line of the display It is possible to set the Center to values below half the span or above the maximum available frequency in which case the displayed spectrum will not take up the whole screen Press the Center softkey then turn the Entry knob to set the desired center frequency of the display Scale lets you set your own vertical scale factors for FFT expressed in dB div decibels division See To adjust the math waveform scale and offset on page 73 Offset lets you set your own offset for the FFT The offset value is in dB and is represented by the center horizontal grid line of the display See To adjust the math waveform scale and offset on page 73 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 81 4 Math Waveforms More FFT displays the More FFT Settings Menu 2 Press the More FFT softkey to display additional FFT settings tit FFT Ch1 Scale 2UdB ffset 60 O V FFT Resolution 763Hz Window Vertical Units Hanning Auto Setup Window selects a window to apply to your FFT input signal Hanning window for making accurate frequency measurements or for
343. sistence 129 ventilation requirements 27 vernier channel 65 Vertical controls 38 vertical expansion 63 vertical position 63 vertical position knobs 38 vertical scale knobs 38 vertical sensitivity 38 63 Vertical Units FFT 82 VGA video output 42 VID license 332 video trigger 158 video trigger custom Generic 162 viewing tilt the instrument 26 VISA connect string 307 visualizations math 93 430 voltage measurements 215 zoom display measurement window with 233 W Zoom key 36 warranted specifications 323 warranty 301 Wave Gen key 35 40 waveform generator 253 waveform generator defaults restoring 267 waveform generator expected output load 261 waveform generator logic presets 262 waveform generator sync pulse 261 waveform generator sync pulse TRIG OUT signal 297 waveform generator arbitrary waveforms 257 waveform generator waveform type 253 Waveform keys 38 waveform type waveform generator 253 waveform cursor tracking 200 waveform intensity 127 waveform printing 281 waveform reference point 293 waveform saving exporting 269 WAVEGEN license 332 web interface 307 web interface accessing 308 white noise adding to waveform generator output 262 Width measurement 225 Width measurement 225 Window FFT 82 X X at Max Y measurement 229 X at Max Y on FFT 213 X at Min Y measurement 228 X at Min Y on FFT 213 XY mode 51
344. solution mode at slower time div settings extra samples are averaged in order to reduce random noise produce a smoother trace on the screen and effectively increase vertical resolution High Resolution mode averages sequential sample points within the same acquisition An extra bit of vertical resolution is produced for every factor of 2 averages Random noise is reduced by 1 2 for every factor of 4 averages The number of extra bits of vertical resolution is dependent on the time per division setting sweep speed of the oscilloscope The slower the time div setting the greater the number of samples that are averaged together for each display point High Resolution mode can be used on both single shot and repetitive signals and it does not slow waveform update because the computation is done in the MegaZoom custom ASIC High Resolution mode limits the oscilloscope s real time bandwidth because it effectively acts like a low pass filter Sweep speed Bits of resolution lt 1 ps div 8 2 us div 9 5 us div 10 10 us div 11 2 20 us div 12 Acquiring to Segmented Memory You can purchase the oscilloscope with the segmented memory option factory installed Option SGM or you can install a license to enable it order model number DSOX3SGM Segmented Memory Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 193 12 Acquisition Control To acquire to segmented memory 194 When capturing
345. ssories see www keysight com for Probes and Accessories Selection Guide 5989 6162EN 5000 6000 and 7000 Series InfiniiVision Oscilloscope Probes and Accessories Data Sheet 5968 8153EN Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Reference 22 Passive Probes All InfiniiVision oscilloscopes recognize passive probes such as the N2862A B N2863A B N2889A N2890A 10073C 10074C and 1165A These probes have a pin on their connector that connects to the ring around the oscilloscope s BNC connector Therefore the oscilloscope will automatically set the attenuation factor for recognized Keysight passive probes Passive probes that do not have a pin that connects to the ring around the BNC connector will not be recognized by the oscilloscope and you must set the probe attenuation factor manually See To specify the probe attenuation on page 67 The following passive probes can be used with the InfiniiVision 3000 X Series oscilloscopes Any combination of passive probes can be used Table 5 Passive Probes Model Description 1165A Passive probe 10 1 600 MHz 1 5 m 10070C D Passive probe 1 1 20 MHz 1 5 m 10073C Passive probe 10 1 500 MHz 1 5 m 10074C Passive probe 10 1 150 MHz 1 5 m 10076A B Passive probe 100 1 4 kV 250 MHz N2771A B Passive probe 1000 1 30 kV 50 MHz N2862A B Passive probe 10 1 150 MHz 1 2 m N2863A B Passive probe 10 1 300 MHz 1 2 m N2889A
346. t Urbana Champaign Network printing uses the CUPS Common Unix Printing System library The CUPS and CUPS Imaging libraries are developed by Apple Inc and licensed under the GNU Library General Public License LGPL Version 2 This license is located on the Keysight InfiniiVision Oscilloscopes Documentation CD ROM CUPS network printing uses the mDNSResponder library The mDNSResponder library is developed by Apple Inc and licensed under the Apache License Version 2 0 This license is located on the Keysight InfiniiVision Oscilloscopes Documentation CD ROM Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 23 CAN LIN Triggering and serial Decode Setup for CAN Signals 343 CAN Triggering 345 CAN Serial Decode 347 Setup for LIN Signals 351 LIN Triggering 353 LIN Serial Decode 354 CAN LIN triggering and serial decode requires Option AUTO or the DSOX3AUTO upgrade Setup for CAN Signals Setup consists of connecting the oscilloscope to a CAN signal using the Signals Menu to specify the signal source threshold voltage level baud rate and sample point To set up the oscilloscope to capture CAN signals use the Signals softkey which appears in the Serial Decode Menu 1 Press Label to turn on labels 2 Press Serial 3 Press the Serial softkey turn the Entry knob to select the desired slot Serial 1 or Serial 2
347. tal probes to the device under test 103 Acquiring waveforms using the digital channels 107 To display digital channels using AutoScale 107 Interpreting the digital waveform display 108 To switch all digital channels on or off 109 To switch groups of channels on or off 110 To switch a single channel on or off 109 To change the displayed size of the digital channels 109 To reposition a digital channel 111 To change the logic threshold for digital channels 110 To display digital channels as a bus 111 Digital channel signal fidelity Probe impedance and grounding 114 To replace digital probe leads 118 This chapter describes how to use the digital channels of a Mixed Signal Oscilloscope MSO The digital channels are enabled on MSOX3000 X Series models and DSOX3000 X Series models that have the MSO upgrade license installed To connect the digital probes to the device under test 1 If necessary turn off the power supply to the device under test KEYSIGHT TECHNOLOGIES 103 6 Digital Channels Turning off power to the device under test would only prevent damage that might occur if you accidentally short two lines together while connecting probes You can leave the oscilloscope powered on because no voltage appears at the probes ov 2 Connect the digital probe cable to the DIGITAL Dn DO connector on the front panel of the mixed signal oscilloscope The digital pr
348. ter the data byte This data byte does not need to occur directly after the control byte Read a ai SDA N 1010xxx JN X scl T IAN 1 7 11 8 ao ioe L J L J L ft J J Startor Control R Ack Data Ack Restart byte Trigger point Condition Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 371 25 372 I2C SPI Triggering and Serial Decode Frame Start Addr7 Read Ack Data or Frame Start Addr7 Write Ack Data The oscilloscope triggers on a read or write frame in 7 bit addressing mode on the 17th clock edge if all bits in the pattern match For triggering purposes a restart is treated as a start condition neac Write _ SDA N m a aera scL 1 7 8 8 B SINS Start or Address R Bok Data Rok Stop Pies Trigger point Condition 17th clock edge Frame Start Addr7 Read Ack Data Ack Data2 or Frame Start Addr7 Write Ack Data Ack Data2 The oscilloscope triggers on a read or write frame in 7 bit addressing mode on the 26th clock edge if all bits in the pattern match For triggering purposes a restart is treated as a start condition Read En Write SDA UA eee Sue Oe aa ar set _ 1 7 _ e _ 9 _ 1 8 _ 9 1 7 t SG Start Address R W Ack Data Ack Data2 Ack Stop Condition Trigger point Condition 26th clock edge 10 bit Write The oscilloscope triggers on a 10 bit write frame on the 26th clock edge if all bits in the pat
349. tern match The frame is in the format Frame Start Address byte 1 Write Address byte 2 Ack Data For triggering purposes a restart is treated as a start condition Write MTA EN sa TA AAAA ATA AAA U Start or Address R Aoki Address Ack2 Data Ack Stop Restart 1st byte 2nd byte Trigger point Condition Condition 26th clock edge Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide I2C SPI Triggering and Serial Decode 25 5 Ifyou have set the oscilloscope to trigger on an EEPROM Data Read condition Press the Data is softkey to set the oscilloscope to trigger when data is equal to not equal to lt less than or gt greater than the data value set in the Data softkey The oscilloscope will trigger on the clock edge for the Ack bit after the trigger event is found This data byte does not need to occur directly after the control byte The oscilloscope will trigger on any data byte that meets the criteria defined by the Data is and Data softkeys during a current address read or a random read or a sequential read cycle 6 Ifyou have set the oscilloscope to trigger on a 7 bit address read or write frame condition or a 10 bit write frame condition a Press the Address softkey and turn the Entry knob to select the 7 bit or 10 bit device address You can select from an address range of 0x00 to Ox7F 7 bit or Ox3FF 10 bit hexadecimal When triggering on a read write frame the oscilloscope will trigger a
350. the Language See also Access the Built In Quick Help on page 44 16 Vertical controls The Vertical controls consist of Analog channel on off keys Use these keys to switch a channel on or off or to access a channel s menu in the softkeys There is one channel on off key for each analog channel Vertical scale knob There are knobs marked V Ny for each channel Use these knobs to change the vertical sensitivity gain of each analog channel Vertical position knobs Use these knobs to change a channel s vertical position on the display There is one Vertical Position control for each analog channel Label key Press this key to access the Label Menu which lets you enter labels to identify each trace on the oscilloscope display See Chapter 9 Labels starting on page 133 For more information see Chapter 3 Vertical Controls starting on page 61 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Getting Started 1 17 Analog channel inputs Attach oscilloscope probes or BNC cables to these BNC connectors With the InfiniiVision 3000 X Series oscilloscopes you can set the input impedance of the analog channels to either 50 Q or 1 MQ See To specify channel input impedance on page 64 The InfiniiVision 3000 X Series oscilloscopes also provide the AutoProbe interface The AutoProbe interface uses a series of contacts directly below the channe
351. the Web Interface on page 308 2 When the oscilloscope s web interface Welcome Page is displayed select the Identification on radio button An Identify message is displayed on the oscilloscope you can either select Identification off or press the OK softkey on the oscilloscope to continue Support Products Keysight Site TECHNOLOGIES Welcome to your LK 1 Web Enabled Oscilloscope MSO X 2024A AW KEYSIGHT PETES SS Information about this Web Enabled Instrument Serial Number US50210029 Agilent agate cage sion 141 121 237 192 a mx2024a 10029 vate 141 121 237 192 TCPIPO 141 121 237 192 INSTR id Advanced information Identification off on Identification option Use the navigation bar on the leftto access your oscilloscope and related PI Slices Instrument Utilities The Instrument Utilities page of the web interface lets you View installed options View firmware versions 318 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Install firmware upgrade files View calibration status You can select these cababilities via a drop down menu TECHNOLOGIES KEYSIGHT Oscilloscope Instrument Utilities Installed Options v Installed Options Firmware Version Calibration Status talled Options Web Interface 21 MSO Instrument ue Utilities EMBD AUTO COMP SGM MASK BW20 BW10 EDK S SCPIPS MSO Acq Memory Max Embedded seria
352. the Web Interface on page 308 2 When the oscilloscope s web interface is displayed select the Save Recall tab from the left side of the Welcome screen 3 Click the Save link 4 On the Save page a Enter the name of the file you are saving to b Select the format Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 315 21 Web Interface KEYSIGHT TECHNOLOGIES Another web from Keysig Oscilloscope EH Contral ekai 8 bit Bitmap image bmp Save Recall 24 bit Bitmap image bmp SAR PNG 24 bit image png e CSV data csv ASCII XY data csv et mage Reference Waveform data n5 Multi Channel Waveform data n5 Binary data bin Instrument Lister data csv RS etra Mask msk OOO You can click Preview to view the oscilloscope s current screen image When previewing you can use the New Acquisition check box to force a new acquisition before the preview With some formats you can click Save Setup Info to save setup information to an ASCII txt format file c Click Save The current acquisition is saved d Inthe File Download dialog click Save e Inthe Save As dialog navigate to the folder where you want to save the file then click Save Recalling Files via the Web Interface To recall setup files reference waveform data files or mask files from your PC via the oscilloscope s web interface 1 Access the oscilloscope s web interface see
353. the menu hierarchy c Press the Host name softkey Use the softkeys and the Entry knob to enter the Host name When you are done go back up in the menu hierarchy d Press the Apply softkey Connect the oscilloscope to the local area network LAN by inserting the LAN cable into the LAN port on the rear panel of the oscilloscope Stand alone Point to Point Connection to a PC The following procedure describes how to establish a point to point stand alone connection to the oscilloscope This is useful if you want to control the oscilloscope using a laptop computer or a stand alone computer 1 2 3 Press Utility gt 1 0 Press the LAN Settings softkey Press the Config softkey then turn the Entry knob to select Automatic and press the softkey again to enable it If your network supports DHCP or AutolP enabling Automatic lets the oscilloscope use those services to get its LAN configuration settings Connect your PC to the oscilloscope using a cross over LAN cable such as Keysight part number 5061 0701 available on the web at www parts keysight com Cycle power on the oscilloscope Wait until the LAN connection is configured Press Utility gt 1 0 and wait until the LAN status shows configured This may take a few minutes Now the instrument is connected and the instrument s web interface or remote control via LAN may be used Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Utility S
354. the oscilloscope to capture a CAN signal 1 Press Trigger 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select the serial slot Serial 1 or Serial 2 on which the CAN signal is being decoded Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 345 23 CAN LIN Triggering and Serial Decode 346 Trigger Menu Trigger Trigger S CAN All Errors 3 Press the Trigger softkey then turn the Entry knob to select the trigger condition SOF Start of Frame The oscilloscope triggers at the start of a frame Remote Frame ID RTR The oscilloscope triggers on remote frames with the specified ID Press the Bits softkey to select the ID Data Frame ID RTR The oscilloscope will trigger on data frames matching the specified ID Press the Bits softkey to select the ID Remote or Data Frame ID The oscilloscope will trigger on remote or data frames matching the specified ID Press the Bits softkey to select the ID Data Frame ID and Data The oscilloscope will trigger on data frames matching the specified ID and data Press the Bits softkey to select the ID and set up the number of data bytes and values Error Frame The oscilloscope will trigger on CAN active error frames All Errors The oscilloscope will trigger when any form error or active error is encountered Acknowledge Error The oscilloscope will trigger when the acknowledge bit is recessive high
355. the palette option In the Print Configuration Menu press the Palette softkey to change the following options 284 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Print Screens 19 Color Select this to print the screen in color The oscilloscope s print driver cannot print color images to color laser printers so the Color option is not available when connected to laser printers Grayscale Select this to print the screen in shades of gray rather than in color Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 285 19 Print Screens 286 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 20 Utility Settings 1 0 Interface Settings 287 Setting up the Oscilloscope s LAN Connection 288 File Explorer 291 Setting Oscilloscope Preferences 293 Setting the Oscilloscope s Clock 296 Setting the Rear Panel TRIG OUT Source 296 Performing Service Tasks 297 Configuring the Quick Action Key 302 Adding an Annotation 303 This chapter explains oscilloscope utility functions I O Interface Settings The oscilloscope can be accessed and or controlled remotely via these I O interfaces USB device port on the rear panel square shaped USB port LAN interface when a LAN VGA module is installed in the rear panel module slot GPIB interface when a GPIB module is installed in the rear panel mod
356. tification Function 318 Instrument Utilities 318 Setting a Password 320 22 Reference Specifications and Characteristics 323 Measurement Category 323 Oscilloscope Measurement Category 324 Measurement Category Definitions 324 Transient Withstand Capability 325 A Maximum input voltage at analog inputs 325 A Maximum input voltage at digital channels 325 Environmental Conditions 325 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Probes and Accessories 326 Passive Probes 327 Single Ended Active Probes 327 Differential Probes 328 Current Probes 329 Accessories Available 330 Loading Licenses and Displaying License Information 331 Licensed Options Available 331 Other Options Available 333 Upgrading toan MSO 333 Software and Firmware Updates 333 Binary Data bin Format 333 Binary Datain MATLAB 334 Binary Header Format 335 Example Program for Reading Binary Data 337 Examples of Binary Files 337 CSV and ASCII XY files 340 CSV and ASCII XY file structure 341 Minimum and Maximum Values in CSV Files 341 Acknowledgements 342 23 CAN LIN Triggering and Serial Decode Setup for CAN Signals 343 CAN Triggering 345 CAN Serial Decode 347 Interpreting CAN Decode 348 CAN Totalizer 349 Interpreting CAN Lister Data 350 Searching for CAN Data in the Lister 351 Setup for LIN Signals 351 Keysight InfiniiVision 3000 X Series Oscil
357. tion 1 Press the Type softkey and turn the Entry knob to select Frequency Modulation FM 2 Press the Waveform softkey and turn the Entry knob to select the shape of the modulating signal Sine Square Ramp Sine Card inal Exponential Rise Exponential Fall When the Ramp shape is selected a Symmetry softkey appears so that you can specify the amount of time per cycle that the ramp waveform is rising 3 Press the FM Freq softkey and turn the Entry knob to specify the frequency of the modulating signal Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 265 17 Waveform Generator 4 Press the FM Dev softkey and turn the Entry knob to specify the frequency deviation from the original carrier signal frequency When the modulating signal is at its maximum amplitude the output frequency is the carrier signal frequency plus the deviation amount and when the modulating signal is at its minimum amplitude the output frequency is the carrier signal frequency minus the deviation amount The frequency deviation cannot be greater than the original carrier signal frequency Also the sum of the original carrier signal frequency and the frequency deviation must be less than or equal to the maximum frequency for the selected waveform generator function plus 100 kHz The following screen shows an FM modulation of a 100 kHz sine wave carrier signal 50 005 Stop 4 28V KEYSIGHT TECHNOLOGIES Channels Aodulation Me
358. tion 282 network connecting to 289 new label 135 noise rejection 177 noise waveform generator output 255 noise adding to waveform generator output 262 noise high frequency 177 noise low frequency 176 noisy signals 173 non volatile memory secure erase 280 normal acquire mode 188 normal mode 187 188 Normal trigger mode 174 notices 2 Nth edge burst trigger 153 Nth edge burst triggering 153 number of negative pulses measurements 230 number of positive pulses measurements 229 Nyquist frequency 86 Nyquist sampling theory 183 0 offset DC correction for integrate waveform 79 operators math 74 options print 284 OR trigger 151 oscilloscope bandwidth 183 oscilloscope bandwidth required 186 oscilloscope rise time 185 oscilloscope sample rate 185 output load expected waveform generator 261 output trigger 296 overlay localized 40 Overshoot measurement 212 217 overvoltage category 325 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Index P palette 272 pan and zoom 48 parts replacement 118 passive probes 327 passive probes compensating 32 password network reset 322 password network setting 320 pattern trigger 148 pattern SPI trigger 382 PC connection 290 peak detect mode 187 188 189 Peak peak measurement 216 Period measurement 223 period measurement trend 95 persistence 129 persistence clea
359. tions for decoding and triggering on ARINC 429 signals High Trigger Threshold 3 0 V Low Trigger Threshold 3 0 V Noise Reject Off Probe Attenuation 10 0 Vertical Scale 4 V div Serial Decode On Base Hex Word Format Label SDI Data SSM Trigger currently active serial bus Trigger Mode Word Start 12 If the high and low thresholds are not set correctly by Auto Setup Press the High Threshold softkey then turn the Entry knob to set the high trigger threshold voltage level Press the Low Threshold softkey then turn the Entry knob to set the low trigger threshold voltage level The threshold voltage levels are used in decoding and will become the trigger levels when the trigger type is set to the selected serial decode slot ARINC 429 Triggering To set up the oscilloscope to capture a ARINC 429 signal see Setup for ARINC 429 Signals on page 404 After setting up the oscilloscope to capture a ARINC 429 signal 1 Press Trigger 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select the serial slot Serial 1 or Serial 2 on which the ARINC 429 signal is being decoded Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 405 27 MIL STD 1553 ARINC 429 Triggering and Serial Decode 406 SDI XX Data 000 0000 0000 0000 0010 Trigger Trigger Label 5 ARINC 429 Label Bits 115 3 Press the Trigger softkey then turn the Entry knob to select the trigge
360. tkey to go to the first segment Press the Pl softkey to go to the last segment Auto to play through segments in an automated fashion In the Auto play mode Press the 1010 navigation keys to play backward stop or play forward in time You can press the or keys multiple times to speed up the playback There are three speed levels Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 59 2 Horizontal Controls 60 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls To turn waveforms on or off channel or math 62 To adjust the vertical scale 63 To adjust the vertical position 63 To specify channel coupling 63 To specify channel input impedance 64 To specify bandwidth limiting 65 To change the vertical scale knob s coarse fine adjustment setting 65 To invert a waveform 65 Setting Analog Channel Probe Options 66 The vertical controls include The vertical scale and position knobs for each analog channel The channel keys for turning a channel on or off and accessing the channel s softkey menu The following figure shows the Channel 1 Menu that appears after pressing the 1 channel key KEYSIGHT TECHNOLOGIES 61 3 Vertical Controls Channel Trigger Trigger level Nolts div source __ or threshold 100 05 Channels Channel 1 ground level
361. to the save location See To navigate storage locations on page 276 3 Press the Settings softkey In the File Settings Menu you have these softkeys and options Setup Info when enabled setup information vertical horizontal trigger acquisition math and display settings is also saved in a separate file with a TXT extension 4 Finally press the Press to Save softkey A message indicating whether the save was successful is displayed 274 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Save Recall Setups Screens Data 18 To save reference waveform files to a USB storage device Press the Save Recall key 2 Inthe Save Recall Menu press the Save softkey 3 In the Save Menu press the Format softkey and turn the Entry knob to select Reference Waveform data h5 4 Press the Source softkey and turn the Entry knob to select the source waveform 5 Press the softkey in the second position and use the Entry knob to navigate to the save location See To navigate storage locations on page 276 6 Finally press the Press to Save softkey A message indicating whether the save was successful is displayed To save masks Mask files can be saved to one of four internal Keysight Flash locations or to an external USB storage device 1 Press Save Recall gt Save gt Format then turn the Entry knob to select Mask msk 2 Press the softkey in the second position and use the Entry kn
362. to 8 0000 Mb s in increments of 100 b s d Polarity Select idle low or idle high to match your device under test s state when at idle For RS232 select idle low e Bit Order Select whether the most significant bit MSB or the least significant bit LSB is presented after the start bit in the signal from your device under test For RS232 select LSB In the serial decode display the most significant bit is always displayed on the left regardless of how Bit Order is set 414 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide UART RS232 Triggering and Serial Decode 28 UART RS232 Triggering To set up the oscilloscope to capture UART RS 232 signals see Setup for UART RS232 Signals on page 413 To trigger on a UART Universal Asynchronous Receiver Transmitter signal connect the oscilloscope to the Rx and Tx lines and set up a trigger condition RS232 Recommended Standard 232 is one example of a UART protocol 1 Press Trigger 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select the serial slot Serial 1 or Serial 2 on which the UART RS232 signals are being decoded Trigger Menu Trigger Trigger Setup w S UART RS232 3 Press the Trigger Setup softkey to open the UART RS232 Trigger Setup Menu 232 Trigger Setup Menu Trigger Data is Tx Data 4 Press the Base softkey to select Hex or ASCII as the base displayed on the Data softkey in the UART RS232 T
363. to enter The setting that must be entered is Printer Address this is the printer following formats or print server s address in one of the IP address of a network enabled printer for example 192 168 1 100 or 192 168 1 100 650 Optionally specified following a colon a non standard port number can be IP address of a print server followed by the path to the printer for example 192 168 1 100 printe 192 168 1 100 650 printers pri Path to a Windows network pri rs printer name or nter name nter share for example server share When the Printer Address is a Windows network printer share the Modify softkey also lets you enter these sett ings Network Domain this is the Windows network domain name Username this is your login name for the Windows network domain Password this is your login password for the Windows network domain To clear an entered password press the Clear Password softkey 5 Use the Spell Enter and Delete Character softkeys to enter the network printer settings Spell press this softkey and turn the entry knob to select the character at the current position Enter press this softkey to enter characters and move the cursor to the next character position Delete Character press the Enter softkey until the desired character is highlighted then press this softkey to delete the character You can use a connected USB keyboard instead of usi
364. to navigate through the marked rows 1 With 12C selected as the serial decode mode press Search 2 Inthe Search Menu press the Search softkey then turn the Entry knob to select the serial slot Serial 1 or Serial 2 on which the 12C signal is being decoded 376 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide I2C SPI Triggering and Serial Decode 25 3 Press Search then select from these options Missing Acknowledge finds SDA data is high during any Ack SCL clock bit Address with no Ack finds when the acknowledge of the selected address field is false The R W bit is ignored Restart finds when another start condition occurs before a stop condition EEPROM Data Read finds EEPROM control byte value 1010xxx on the SDA line followed by a Read bit and an Ack bit It then looks for the data value and qualifier set by the Data is softkey and the Data softkeys Frame Start Ad dress7 Read Ack Data finds a read frame on the 17th clock edge if all bits in the pattern match Frame Start Address7 Write Ack Data finds a write frame on the 17th clock edge if all bits in the pattern match Frame Start Ad dress7 Read Ack Data Ack Data2 finds a read frame on the 26th clock edge if all bits in the pattern match Frame Start Ad dress7 Write Ack Data Ack Data2 finds a write frame on the 26th clock edge if all bits in the pattern match For more information on searching data see Searching
365. to select Bus1 or Bus2 3 Press the Digit softkey and rotate the Entry knob to select a digit of the selected bus 4 Press the Hex softkey and rotate the Entry knob to select a value for the digit If a digit is made up of less than four bits then the value of the digit will be limited to the value that can be created by the selected bits 5 You can use the Set all Digits softkey to set all digits to a particular value When a hex bus digit contains one or more don t care X bits and one or more bit with a value or O or 1 the sign will be displayed for the digit For information regarding digital bus display when Pattern triggering see Bus values are displayed when using Pattern trigger on page 113 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Triggers 10 OR Trigger The OR trigger mode triggers when any one or more of the specified edges on analog or digital channels is found 1 On the front panel in the Trigger section press the Trigger key 2 Inthe Trigger Menu press the Trigger softkey and use the Entry knob to select OR 3 Press the Slope softkey and select rising edge falling edge either edge or don t care The selected slope is displayed in the upper right corner of the display 4 For each analog or digital channel you want to include in the OR trigger press the Channel softkey to select the channel As you press the Channel softkey or rotate the Entry knob the channel you
366. to select whether the acquisition mode should be preserved during autoscale Normal to make the oscilloscope switch to Normal acquisition mode whenever the AutoScale key is pressed This is the default mode Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 295 20 Utility Settings Preserve to make the oscilloscope remain in the acquisition mode you have chosen when the AutoScale key is pressed Setting the Oscilloscope s Clock The Clock Menu lets you set the current date and time of day 24 hour format This time date stamp will appear on hardcopy prints and directory information on the USB mass storage device To set the date and time or to view the current date and time 1 Press Utility gt Options gt Clock Clock Menu Year Hour Minute at i 2010 0 t3 52 2 Press the Year Month Day Hour or Minute softkey then rotate the Entry knob to set to the desired number The hours are shown in the 24 hour format So 1 00 PM is hour 13 The real time clock only allows selection of valid dates If a day is selected and the month or year is changed so the day is invalid the day is automatically adjusted Setting the Rear Panel TRIG OUT Source You can choose the source of the TRIG OUT connector on the rear panel of the oscilloscope 1 Press Utility gt Options gt Rear Panel 2 Inthe Rear Panel Menu press Trig Out then turn the Entry knob to select from Triggers Each time the
367. tured 60 Sample Point 80 10 Press the Signal softkey and select the type and polarity of the CAN signal This also automatically sets the channel label for the source channel CAN_H The actual CAN_H differential bus 344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide CAN LIN Triggering and Serial Decode 23 Differential H L The CAN differential bus signals connected to an analog source channel using a differential probe Connect the probe s positive lead to the dominant high CAN signal CAN_H and connect the negative lead to the dominant low CAN signal CAN_L Dominant low signals Rx The Receive signal from the CAN bus transceiver Tx The Transmit signal from the CAN bus transceiver CAN_L The actual CAN_L differential bus signal Differential L H The CAN differential bus signals connected to an analog source channel using a differential probe Connect the probe s positive lead to the dominant low CAN signal CAN_L and connect the negative lead to the dominant high CAN signal CAN_H CAN Triggering To set up the oscilloscope to capture a CAN signal see Setup for CAN Signals on page 343 The Controller Area Network CAN trigger allows triggering on CAN version 2 0A and 2 0B signals A CAN message frame in CAN_L signal type is shown below Bus Arbitration Control Data CRC ACK Intermission Idle Field Field Field Field Field SOF edge After setting up
368. ty VID Extended Video Triggering and Analysis Order DSOX83VID WAVEGEN Waveform Generator Order DSOX3WAVEGEN 332 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Reference 22 Other Options Available Table 11 Calibration Option Option Order A6J ANSI Z540 Compliant Calibration Upgrading to an MSO A license can be installed to activate the digital channels of an oscilloscope that was not originally ordered as a mixed signal oscilloscope MSO A mixed signal oscilloscope has analog channels plus 16 time correlated digital timing channels For information about upgrading your oscilloscope through licensing contact your local Keysight Technologies representative or see www keysight com find 3000X Series Software and Firmware Updates From time to time Keysight Technologies releases software and firmware updates for its products To search for firmware updates for your oscilloscope direct your web browser to www keysight com find 3000X Series sw To view the currently installed software and firmware press Help gt About Oscilloscope Once you have downloaded a firmware update file you can place it on a USB storage device and load the file using File Explorer see File Explorer on page 291 or you can use the Instrument Utilities page of the oscilloscope s web interface see Instrument Utilities on page 318 Binary Data bin Format The binary data format stores waveform data in b
369. ule slot To configure the I O interfaces 1 On the oscilloscope s front panel press Utility 2 Inthe Utility Menu press 1 0 3 In the I O Menu press Configure KEYSIGHT TECHNOLOGIES 287 20 Utility Settings LAN When a DSOXLAN LAN VGA module is installed you can use the LAN Settings and LAN Reset softkeys to configure the LAN interface See Setting up the Oscilloscope s LAN Connection on page 288 GPIB When a DSOXGPIB GPIB module is installed you can use the Address softkey to configure the GPIB address There are no configuration settings for the USB interface When an I O interface is installed remote control over that interface is always enabled Also the oscilloscope can be controlled via multiple 1 0 interfaces for example USB and LAN at the same time See Also Chapter 21 Web Interface starting on page 307 when the oscilloscope is connected to a LAN Remote Programming via the Web Interface on page 313 The oscilloscope s Programmer s Guide Remote Programming with Keysight IO Libraries on page 314 Setting up the Oscilloscope s LAN Connection When the DSOXLAN LAN VGA module is installed you can place the oscilloscope on the network and set up its LAN connection Once that is done you can use the oscilloscope s web interface or remotely control the oscilloscope via the LAN interface The oscilloscope supports methods for automated LAN configuration or manual LAN configurat
370. ved to one of two reference waveform locations in the oscilloscope Then a reference waveform can be displayed and compared against other waveforms One reference waveform can be displayed at a time When the multiplexed knobs are assigned to reference waveforms this happens when you press the Ref key and the LED to the left of it is lit you can use the knobs to scale and position reference waveforms There is also a skew adjustment for reference waveforms Reference waveform scale offset and skew information can optionally be included on the oscilloscope display Analog channel math or reference waveforms can be saved to a reference waveform file on a USB storage device You can recall a reference waveform file from a USB storage device into one of the reference waveform locations To save a waveform to a reference waveform location 1 Press the Ref key to turn on reference waveforms 2 Inthe Reference Waveform Menu press the Ref softkey and turn the Entry knob to select the desired reference waveform location 3 Press the Source softkey and turn the Entry knob to select the source waveform KEYSIGHT TECHNOLOGIES 99 5 Reference Waveforms 4 Press the Save to R1 R2 softkey to save the waveform to the reference waveform location Reference waveforms are non volatile they remain after power cycling or performing a default setup Tocleara 1 Press the Ref key to turn on reference waveforms reference 2 Inthe Ref
371. veform data or setup Press Settings to set up the save options See Chapter 18 Save Recall Setups Screens Data starting on page 269 Quick Recall recalls a setup mask or reference waveform Press Settings to set up the recall options See Chapter 18 Save Recall Setups Screens Data starting on page 269 Quick Freeze Display freezes the display without stopping running acquisitions or un freezes the display if currently frozen For more information see To freeze the display on page 131 Quick Trigger Mode toggles the trigger mode between Auto and Normal see To select the Auto or Normal trigger mode on page 174 Quick Clear Display clears the display see To clear the display on page 130 Once the Quick Action key is configured you simply press it to perform the selected action Adding an Annotation You can add an annotation to the upper left corner of the oscilloscope s display The annotation is useful for documentation purposes to add notes before capturing screens To add an annotation 1 2 3 4 5 On the oscilloscope s front panel press Utility In the Utility Menu press Annotation In the Annotation Menu press Annotation to enable the annotation Press Edit In the Edit Annotation Menu Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 303 20 Utility Settings 500 08 TECHN Acquisition with infrequent glitch Channels Ed
372. vided the time qualifier criteria has been met The time values for the selected qualifier are set using the qualifier time set softkeys lt and gt and the Entry knob Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Triggers 10 4 For each analog or digital channel you want to include in the desired pattern press the Channel softkey to select the channel This is the channel source for the O 1 X or edge condition As you press the Channel softkey or rotate the Entry knob the channel you select is highlighted in the Pattern line directly above the softkeys and in the upper right corner of the display next to Pat Adjust the trigger level for the selected analog channel by turning the Trigger Level knob Press the Digital key and select Thresholds to set the threshold level for digital channels The value of the trigger level or digital threshold is displayed in the upper right corner of the display 5 For each channel you select press the Pattern softkey then turn the Entry knob Pattern 1 1X Trigger Pattern to set the condition for that channel in the pattern 100 047 Channels Pattern vo Qualifier Channel Pattern Entered 0 0 sets the pattern level that is less than the channel s trigger level or threshold level 1 sets the pa ttern to zero low on the selected channel A low is a voltage to 1 high on the selected channel A high is a voltage level that i
373. vity shown in the status line at the top of the display Each major horizontal division in the grid corresponds to the time div shown in the status line at the top of the display To freeze the display To freeze the display without stopping running acquisitions you must configure the Quick Action key See Configuring the Quick Action Key on page 302 1 Once the Quick Action key has been configured press it to freeze the display 2 To un freeze the display press Quick Action again Manual cursors can be used on the frozen display Many activities such as adjusting the trigger level adjusting vertical or horizontal settings or saving data will un freeze the display Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 131 8 Display Settings 132 Keysight InfinilVision 3000 X Series Oscilloscopes User s Guide Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels To turn the label display on or off 133 To assign a predefined label to a channel 134 To define a new label 135 To load a list of labels from a text file you create 136 To reset the label library to the factory default 137 You can define labels and assign them to each analog input channel or you can turn labels off to increase the waveform display area Labels can also be applied to digital channels on MSO models To turn the label display on or off 1 Press the Label key on the front panel
374. waveform that does not conform to the mask This setting supersedes the Minimum of Tests and Minimum Time settings Save The oscilloscope saves the screen image when an error is detected In the Save Menu press Save Recall gt Save select an image format bmp or png destination on a USB storage device and file name which can be auto incrementing If errors occur too frequently and the oscilloscope spends all its time saving images press the Stop key to stop acquisitions Print The oscilloscope prints the screen image when an error is detected This option is only available when a printer is connected as described in To print the oscilloscope s display on page 281 Measure Measurements and measurement statistics if your oscilloscope supports them run only on waveforms that contain a mask violation Measurements are not affected by passing waveforms This mode is not available when the acquisition mode is set to Averaging Note that you can choose to Print or Save but you cannot select both at the same time All other actions may be selected at the same time For example you can select both Stop and Measure to cause the oscilloscope to measure and stop on the first error You can also output a signal on the rear panel TRIG OUT BNC connector when there is a mask test failure See Setting the Rear Panel TRIG OUT Source on page 296 Source Lock When you turn on Source Lock using the Source Lo
375. when the mask file is loaded Any legal remote programming command can be entered in this section The mask scaling controls how the normalized vectors are interpreted This in turn controls how the mask is drawn on the display The remote programming commands that control mask scaling are Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 245 15 Mask Testing MTES MTES MTES MTES MTES SCAI SCAI SCAI SCAI SCAI Bet Ete Building a Mask File 246 BIND 0 X1 400 000E 06 XDEL 800 000E 06 Y 1 359 000E 03 Y2 4 2 35900E 00 The following mask uses all eight mask regions The most difficult part of creating a mask file is normalizing the X and Y values from the time and voltage values This example shows an easy way to convert voltage and time to normalized X and Y values in the mask file The following mask file produced the mask shown above BW Limit MASK FILE 548XX All Regions Region Number 1 Number of vertices 4 12 50 10 00 10 00 12 50 MAX 1 750 1 750 MAX Region Number 2 Number of vertices 5 75 09 KEYSIGHT TECHNOLOGIES Acquisition Channels Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Mask Testing 15 10 00 1 000 12 50 0 500 15 00 0 500 15 00 1 500 12 50 1 500 Region Number 3 Number of vertices 6 05 00 1 000 02 50 0 500 02 50 0 500 05 00 1 000 02 5
376. ws so the data for the first segment starts at row 4 The second segment s data starts at row 1004 The time column shows the time since the trigger on the first segment The top row shows the selected number of points per segment BIN files are a more efficient data transfer format than CSV or ASCII XY Use this file format for fastest data transfer Minimum and Maximum Values in CSV Files If you are running a Minimum or Maximum measurement the minimum and maximum values shown in the measurement display may not appear in the CSV file Explanation When the oscilloscope s sample rate is 4 GSa s a sample will be taken every 250 ps If the horizontal scale is set to 10 us div there will be 100 us of data displayed because there are ten divisions across the screen To find the total number of samples the oscilloscope will take 100 us x 4 GSa s 400K samples The oscilloscope is required to display those 400K samples using 640 pixel columns The oscilloscope will decimate the 400K samples to 640 pixel columns and this decimation keeps track of the min and max values of all the points that are represented by any given column Those min and max values will be displayed in that screen column A similar process is used to reduce the acquired data to produce a record usable for various analysis needs such as measurements and CSV data This analysis record or measurement record is much larger than 640 and may in fact contain up to
377. x 100 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 225 14 Measurements 226 Rise Time Fall Time Delay The rise time of a signal is the time difference between the crossing of the lower threshold and the crossing of the upper threshold for a positive going edge The X cursor shows the edge being measured For maximum measurement accuracy set the horizontal time div as fast as possible while leaving the complete rising edge of the waveform on the display The Y cursors show the lower and upper threshold points The fall time of a signal is the time difference between the crossing of the upper threshold and the crossing of the lower threshold for a negative going edge The X cursor shows the edge being measured For maximum measurement accuracy set the horizontal time div as fast as possible while leaving the complete falling edge of the waveform on the display The Y cursors show the lower and upper threshold points Delay measures the time difference from the selected edge on source 1 and the selected edge on source 2 closest to the trigger reference point at the middle threshold points on the waveforms Negative delay values indicate that the selected edge of source 1 occurred after the selected edge of source 2 Source 1 COO AOO NO Z ae Delay Source 2 NOON KX Press the Meas key to display the Measurement Menu 2 Press the Source softkey then turn the Entry knob to select the first analog
378. y 1 Gently pull on the front panel knobs to remove them 2 nsert the overlay s side tabs into the slots on the front panel 3 Reinstall the front panel knobs Front panel overlays may be ordered from www parts keysight com using the following part numbers Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide Getting Started 1 Language 2 Channel Overlay 4 Channel Overlay French 75019 94324 75019 94316 German 75019 94326 75019 94318 Italian 75019 94323 75019 94331 Japanese 75019 94311 75019 94312 Korean 75019 94329 75019 94321 Polish 75019 94335 75019 94334 Portuguese 75019 94327 75019 94319 Russian 75019 94322 75019 94315 Simplified Chinese 75019 94328 75019 94320 Spanish 75019 94325 75019 94317 Thai 75019 94333 75019 94332 Traditional Chinese 75019 94330 75019 94310 Learn the Rear Panel Connectors For the following figure refer to the numbered descriptions in the table that follows Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 4 1 42 Getting Started 8 USB Device port 3 LAN VGA 7 USB Host port option module ji a use il DEVICE oat Ie TRG OUT EXT TRIG caT TS va 6 EXT TRIG IN connector 2 5 Calibration Nl WARNING MAIN TAIN GROUND TO AVOID ELECTRIC SHOCK 100 120V S0 60 400H2 100 240 50 6
379. y for progressive video with a tri level sync enabling the Horiz Sync control and setting the Horiz Sync adjustment to the sync time of the probed video signal allows the Edge function to count only lines and not double count during vertical sync When the horizontal sync control is enabled press the second Horiz Sync softkey then turn the Entry knob to set the minimum time the horizontal sync pulse must be present to be considered valid To trigger on a specific line of video Video triggering requires greater than 1 2 division of sync amplitude with any analog channel as the trigger source Turning the trigger Level knob in Video trigger does not change the trigger level because the trigger level is automatically set to the sync pulse tips One example of triggering on a specific line of video is looking at the vertical interval test signals VITS which are typically in line 18 Another example is closed captioning which is typically in line 21 1 Press the Trigger key 2 Inthe Trigger Menu press the Trigger softkey then turn the Entry knob to select Video 3 Press the Settings softkey then press the Standard softkey to select the appropriate TV standard NTSC Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 163 10 Triggers 4 Press the Mode softkey and select the TV field of the line you want to trigger on You can choose Line Field1 Line Field2 or Line Alternate 5 Press the Line softkey and
380. yed in octal Data white red if parity error 21 bits displayed in the selected Base SSM green 2 bits displayed in binary When the decode format is Label Data Label yellow 8 bits displayed in octal Data white red if parity error 23 bits displayed in the selected Base Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 409 27 MIL STD 1553 ARINC 429 Triggering and Serial Decode 410 The Label bits are displayed in the same order as they are received on the wire For the Data SSM and SDI bits the fields are displayed in the order received however the bits within those fields are displayed in reverse order In other words the non Label fields are displayed in the ARINC 429 Word Format while the bits for those fields have the opposite transfer order on the wire ARINC 429 Totalizer The ARINC 429 totalizer measures the total ARINC 429 words and errors Serial Decode Me WORDS 0000707471 ERRORS 00 Mi Signals ARINC 429 ll The totalizer is always running counting words and errors and is displayed whenever ARINC 429 decode is displayed The totalizer counts even when the oscilloscope is stopped not acquiring data Pressing the Run Stop key does not affect the totalizer When an overflow condition occurs the counter displays OVERFLOW The counters can be reset to zero by pressing the Reset ARINC 429 Counters softkey in the decode Settings menu Keysight InfinilVisio
381. ze Use 8 bit address size to include the R W bit as part of the address value or choose 7 bit address size to exclude the R W bit from the address value 3 Ifthe decode line does not appear on the display press the Serial key to turn it on 4 Ifthe oscilloscope is stopped press the Run Stop key to acquire and decode data If the setup does not produce a stable trigger the I2C signals may be slow enough that the oscilloscope is AutoTriggering Press the Mode Coupling key then press the Mode softkey to set the trigger mode from Auto to Normal You can use the horizontal Zoom window for easier navigation of the acquired data Interpreting 12C Decode on page 375 Interpreting 12C Lister Data on page 376 Searching for I2C Data in the Lister on page 376 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide I2C SPI Triggering and Serial Decode 25 Interpreting 12C Decode 200 08 GH TTL KEYSIGHT TECHNOLOGIES Channels I 41A 47A 49A 4CA 454 Addr Size Lister gt 7 Bit oe Angled waveforms show an active bus inside a packet frame Mid level blue lines show an idle bus In the decoded hexadecimal data Address values appear at the start of a frame Write addresses appear in light blue along with the W character Read addresses appear in yellow along with the R character Restart addresses appear in green along with the S character Data values appear in white A indicates Ack low