Operating Manual for Model 370 AC Resistance Bridge

image

Contents

1. 6 4 6 1 5 IEEE Interface Example Programs eee aaa aaa aaa aaa aaa aaa aaa aaa aaa aaa cacaea 6 5 6 1 5 1 IEEE 488 Interface Board Installation for Visual Basic Program eseea aaa aaa aaa 6 5 6 1 5 2 Visual Basic IEEE 488 Interface Program Setup eeeee eee aaa aaa aaa aaa aaa 6 7 6 1 5 3 Prograny Operation zi RAA R EA RENE E LE its 6 10 6 1 6 Tr ublesho0tiNg 1 seswssa izy daw gee teat dada 6 10 6 2 SERIAL INTERFACE OVERVIEW cx ecscofcssceaecsencenev sien bestend en i p e cues d a i 6 11 6 2 1 Changing Baud Rate cocinada 6 11 6 2 2 Physical Connections miii tii 6 11 6 2 3 Hardware Supporti z e t 6 12 6 2 4 Character Format iii A A OO ZO AAA 6 12 6 2 5 Message StringS iii A i A eee a 6 12 6 2 6 Message Flow Control uc daa 6 13 6 2 7 Serial Interface Example Programs oooooonccinococcnononcncnononcnononn nono aaa aaa aaa rrrnnrrrrnnrrrrrnn 6 13 6 2 7 1 Visual Basic Serial Interface Program Setup eeeu eee aaa 6 14 6 2 7 2 Program Opera dies 6 17 6 2 8 Troubleshooting it a acia 6 17 6 3 COMMAND SUMMARY civic iia 6 18 6 3 1 Interface Commands iio dai 6 20 Y OPTIONS AND ACCESSORIES cio gia 7 1 7 0 GENERAL culantro dao SANTA 7 1 7 1 MODELS ee zen BEA EN SZ EA EEC EEC 7 1 7 2 AO A O AAA 7 2 7 3 ACCESSORIES aneti odei e ada Se eee cee gt cee cee oe adsense da eee 7 2 Bio SERVICE coz biien 8 1 8 0 GENERAL dose doo cede teeter deal E E ida 8 1 8 1 C
2. RESPONSE 273 150E 00 term ENTER COMMAND HTRRNG 0 Heater range command Instrument will turn off the heater No response will be sent ENTER COMMAND HTRRNG Heater range query Instrument will return a string with the present heater range setting RESPONSE O term ENTER COMMAND HTRRNG 1 HTRRNG Heater range command followed by a query Instrument will change to heater Low setting then return a string RESPONSE 1 term with the present setting The following are additional notes on using either IEEE 488 Interface program Ifyou enter a correctly spelled query without a nothing will be returned Incorrectly spelled commands and queries are ignored Commands and queries should have a space separating the command and associated parameters e Leading zeros and zeros following a decimal point are not needed in a command string but are sent in response to a query A leading is not required but a leading is required 6 2 8 Troubleshooting New Installation 1 Check instrument Baud rate 2 Make sure transmit TD signal line from the instrument is routed to receive RD on the computer and vice versa Use a null modem adapter if not Always send terminators 3 4 Send entire message string at one time including terminators Many terminal emulation programs do not 5 Send only one simple command at a time until communication is established 6 Be sure to spell commands correctly a
3. oooonncnnnnccnnnnaniconocacanononncnanannnrnorn cnn nn aan 3 7 3 5 3 Scanner Resistor Input CabliNQ oooooonnccccnnnocccononananononcncnanonnnnnono nono aaa aan nn 3 7 ii Table of Contents Lake Shore Model 370 AC Resistance Bridge User s Manual TABLE OF CONTENTS Continued Chapter Paragraph Title Page 3 5 4 Scanner Output Connector Definition u e eeee eee aaa aaa aaa aaa aaa aaa 3 8 3 5 5 Scanner Output Cable 3 9 3 5 6 Scanner Control Connector Definition eeeeu eee ea aaa aaa aaa aaa aaa rr rn 3 10 3 5 7 Scanner Control Cabling essien ckcsncencesten twin aw LA a 3 10 3 5 8 Using the Scan Control Signals Without A Scanner eeeue eee eee aaa aaa aaa 3 10 3 5 9 Scanner Mounting Bracket lt a e sd an 3 11 3 6 MONITOR AND REFERENCE OUTPUTS eee eee eee aaa aaa aaa aaa aaa rr nan rra rnnnnns 3 12 3 6 1 Monitor Voltage Output caia ata 3 12 3 6 2 Reference Voltage Output ziii A AL Ween a ete SSS 3 12 3 7 CONTROL HEATER OUTPUT jinaga a eo cocine 3 12 3 7 1 Heater Output Isolation civic rro EPE OO apes coeensvec shot cehehavsbOvene tnepeauhateesca 3 12 3 7 2 Heater Output Connection senasis pinanigan erias spinn aina tp aaa aaa aaa nnii evia 3 13 3 7 3 Heater Output Cabling rie iiciin eteni cennau zeczy reed ea 3 13 3 7 4 Calculating Control Heater POWeT ooocononccnnncoccnnnoncncnononcnn nono non nn aaa aaa aaa rr 3 13 3 7 5 Heater Output
4. Voltage Gain Coefficient Actual Resistance RDGR Service 8 21 Lake Shore Model 370 AC Resistance Bridge User s Manual Model 370 Resistance Input Calibration Work Sheet for Paragraph 8 12 2 6 Continued Step Range Setting Coefficient Resistor Voltage Filter Actual RDGR Coefficient 6a 316nA Gain CALG 1 12 200 kQ 63 2 mV 10s 6b 100 nA Gain CALG 1 11 200 kQ 20mV 10s 6c 31 6nA Gain CALG 1 10 200 kQ 6 32mV 10s 6d 10nA Gain CALG 1 9 200 kQ 2mV 10s 6e 3 16nA Gain CALG 1 8 200 kQ 632uV 10s 6f 1 nA Gain CALG 1 7 200 kQ 200uV 10s 6g 316 pA Gain CALG 1 6 200 kQ 63 2 UV 50s 6h 100 pA Gain CALG 1 5 200 kQ 20uV 50s 61 31 6pA Gain CALG 1 4 200 kQ 6 32 uV 100s 6j 10 pA Gain CALG 1 3 200 kQ 2uV 200s Current Gain Coefficient Actual Resistance RDGR Step Range Setting Coefficient Resistor Voltage Filter Actual RDGR Coefficient Ta 3 16 pA Gain CALG 1 2 600 kQ 2uV 200s Tb 1 pA Gain CALG 1 1 2MQ 2uV 200s Voltage Gain Coefficient Actual Resistance RDGR 8 12 3 Analog Voltage Output Calibration The Model 370 has 2 analog voltage outputs with a range of 10V These outputs can be calibrated independently of each other and their calibration is independent of the measurement input A DC voltmeter with accuracy of 1 mV ona 10 V range and cable
5. 2 9 3 7 Lead Soldering When additional wire is soldered to short sensor leads care must be taken not to overheat the sensor A heat sink such as a metal wire clamp or alligator clip will heat sink the leads and protect the sensor Lead wires should be tinned before bonding to reduce the time that heat is applied to the sensor lead Solder flux should be cleaned after soldering to prevent corrosion 2 9 3 8 Heat Sinking Leads Sensor leads can be a significant source of error if they are not properly heat sinked Heat will transfer down even small leads and alter the sensor reading The goal of heat sinking is to cool the leads to a temperature as close to the sensor as possible This can be accomplished by putting a significant length of lead wire in thermal contact with every cooled surface between room temperature and the sensor Lead wires can be adhered to cold surfaces with varnish over a thin electrical insulator like cigarette paper They can also be wound onto a bobbin that is firmly attached to the cold surface Some sensor packages include a heat sink bobbin and wrapped lead wires to simplify heat sinking 2 9 3 9 Thermal Radiation Thermal black body radiation is one of the ways heat is transferred Warm surfaces radiate heat to cold surfaces even through a vacuum The difference in temperature between the surfaces is one thing that determines how much heat is transferred Thermal radiation causes thermal gradients and reduces measure
6. SHIELD Input Connectors bmp Pin Voltage Connector Current Connector 1 V Voltage L Current 2 VG Voltage Guard IG Current Guard 3 SHIELD Shield SHIELD Shield 4 VG Voltage Guard IG Current Guard 5 V Voltage I Current 6 None Not Used None Not Used Figure 8 6 Resistor Sensor INPUT Connectors MONITOR REFERENCE MonitorRefOutputs bmp Pin Description Center Signal Shell Ground Figure 8 7 REFERENCE and MONITOR Connector Details HEATER OUTPUT Heater Output bmp Pin Description Center Signal Shell Ground Figure 8 8 HEATER OUTPUT Connector Details ANALOG ANALOG OUT 2 STILL OUT 1 Analog Outputs bmp Pin Description Center Signal Shell Ground Figure 8 9 ANALOG OUT 1 and 2 STILL Connector Details 8 12 Service Lake Shore Model 370 AC Resistance Bridge User s Manual Slides into ALARM RELAYS slot in rear of Model 370 Use screwdriver to ALARM RELAYS DZ _ lock or unlock wires 30V 5A LNO LCOM LNC HNO HCOM HNC reer wire 2 into slot Alarm Relays bmp Temrinal Block eps Low Alarm Relay Normally Open Low Alarm Relay Common Low Alarm Relay Normally Closed High Alarm Relay Normally Open High Alarm Relay Common High Alarm Relay Normally Closed Serial Connector bmp Model 370 AC Resistance Bridge Typical
7. 1 Create an input short by connecting all four voltage and current measurement leads together V V I 5 Lead wires should be kept as short as possible and soldered together Spring tension lead clips will not make appropriate connections 2 Attach the input short to the instrument and allow it to warm up for 1 hour 3 Select the highest voltage range 632 mV and the highest resistance lowest current range available for that voltage range 63 2 MQ 10 nA for the highest range 4 Use the CALZ and CALG queries to read and record the coefficients for the selected voltage range A safety step to revert to the original calibration if necessary 5 Set the CALZ and CALG coefficients for the selected Voltage range to default 0 and 1 The current source CALG coefficient does not have to be set to default 1 for this operation 6 Read the short circuit voltage for the range using the CALVIN query Sufficient settling time and filtering must be used to obtain a stable reading Refer to Paragraph 8 12 2 6 for practical filter time constants The CALVIN reading is not filtered with the instrument filter feature The voltage read should be 0 volts 0 02 of the voltage range 7 Send the value read with the CALVIN query as the CALZ coefficient Example for the highest voltage range CALZ 2 12 12 3456E 6 8 Send the CALSAVE command and wait 10 seconds for the calibration to be saved in non volatile memory Alternately this s
8. 2 Verify that the following files have been installed to the Windows System folder a gpib 32 dll b gpib dll c gpib32ft dll Files b and c support any 16 bit Windows GPIB applications being used 3 Locate the following files and make note of their location These files will be used during the development process of a Visual Basic program a Niglobal bas b Vbib 32 bas NOTE If the files in Steps 2 and 3 are not installed on your computer they may be copied from your National Instruments setup disks or they may be downloaded from www natinst com 4 Configure the GPIB by selecting the System icon in the Windows 98 95 Control Panel located under Settings on the Start Menu Configure the GPIB Settings as shown in Figure 6 1 Configure the DEV12 Device Template as shown in Figure 6 2 Be sure to check the Readdress box Remote Operation 6 5 Lake Shore Model 370 AC Resistance Bridge User s Manual System Properties 2 x General Device Manager Hardware Profiles Perform MN Gr Eua a a L2 1x y A General GPIB Settings Resources View devices by type View devices by col A Computer y AT GPIB TNT Plug and Play CDROM Disk drives 7154 PnP Serial Number OO4D7FA0 Display adapters 4 2 Floppy disk controllers Interface Name r Termination Methods 1 3 Hard disk controllers GPB0 y M Send EOI at end of Write 8 Keyboard z g Monitor GPIB Add
9. 24 SH1 AH1 T5 L4 SR1 RL1 PPO DC1 DTO CO El IEEE 488 INTERFACE 12 1 13 IEEE Connector bmp DESCRIPTION PIN SYMBOL 1 DIO1 2 DIO2 3 DIO3 4 DIO4 5 EOI 6 DAV 7 NRFD 8 NDAC 9 IFC 10 SRQ 11 ATN 12 SHIELD 13 DIOS 14 DIO 6 15 DIO7 16 DIO8 17 REN 18 GND6 19 GND7 20 GND8 21 GND9 22 GND10 23 GND11 24 GND Data Input Output Line 1 Data Input Output Line 2 Data Input Output Line 3 Data Input Output Line 4 End Or Identify Data Valid Not Ready For Data Not Data Accepted Interface Clear Service Request Attention Cable Shield Data Input Output Line 5 Data Input Output Line 6 Data Input Output Line 7 Data Input Output Line 8 Remote Enable Ground Wire Twisted pair with DAV Ground Wire Twisted pair with NRFD Ground Wire Twisted pair with NDAC Ground Wire Twisted pair with IFC Ground Wire Twisted pair with SRQ Ground Wire Twisted pair with ATN Logic Ground Figure 8 12 IEEE 488 Rear Panel Connector Details Service 8 15 Lake Shore Model 370 AC Resistance Bridge User s Manual 8 12 MODEL 370 CALIBRATION The following paragraphs provide calibration information for the Model 370 AC Resistance Bridge Calibration information for the Model 3716 Scanner follows in Paragraph 8 13 There are three features of the Model 370 that can be calibrated the resistance measurement analog voltage outputs and heater output
10. gSend True End Sub Routine to handle Send button press Set Flag to True Private Sub Form Load Dim strReturn As String Dim term As String Dim strCommand As String Dim intDevice As Integer frmIEEE Show term Chr 13 amp Chr 10 strReturn Call ibdev 0 12 0 T10s 1 H140A intDevice Call ibconfig intDevice ibcREADDR 1 Main code section Used to return response Terminators Data string sent to instrument Device number used with IEEE Show main window Terminators are lt CR gt lt LF gt Clear return string Initialize the IEEE device Setup Repeat Addressing Do Do Wait loop DoEvents Give up processor to other events Loop Until gSend True Loop until Send button pressed gSend False Set Flag as False strCommand frmlIEEE txtCommand Text Get Command strReturn Clear response display strCommand UCase strCommand Set all characters to upper case If strCommand EXIT Then Get out on EXIT End End If Call ibwrt intDevice strCommand term Send command to instrument If ibsta And EERR Then Check for IEEE errors do error handling if needed Handle errors here End If If InStr strCommand lt gt 0 Then Check to see if query strReturn Space 100 Build empty return buffer Call ibrd intDevice strReturn Read back response If ibsta And EERR Then Check for IEEE errors do error handling if needed Handle errors here End If If strReturn lt gt Then
11. 3 Remove the instrument power cord 4 With a small screwdriver release the drawer holding the line voltage selector and fuse 5 Slide out the removable plastic fuse holder from the drawer 6 Rotate the fuse holder until the proper voltage indicator shows through the window 7 Verify the proper fuse value 8 Re assemble the line input assembly in the reverse order 9 Verify the voltage indicator in the window of the line input assembly 10 Connect the instrument power cord 11 Turn the line power switch On l Line Cord Power Switch Fuse Input O Off I On Drawer 1001202207240 V 10 6 Voltage 100 120V 0 5AT 250V 5x20mm 50 60 Hz 50 VA MAX 220 240V 0 25A T 250V 5x20mm Power Switch bmp Figure 8 2 Power Fuse Access 8 2 Service Lake Shore Model 370 AC Resistance Bridge User s Manual 8 5 FUSE REPLACEMENT Use the following procedure to remove and replace the line fuses WARNING To avoid potentially lethal shocks turn off controller and disconnect it from AC power before performing these procedures CAUTION For continued protection against fire hazard replace only with the same fuse type and rating specified for the line for the line voltage selected NOTE Test fuse with an ohmmeter Do not rely on visual inspection of fuse Locate line input assembly on the instrument rear panel See Figure 8 2 Turn power switch Off O Remove instrument power cord With a small screwdriver release the drawer holding the line v
12. Roman numerals Letters employed in the ancient Roman system of numeration as follows I 1 VI 6 L 50 II 2 VII 7 c 100 III 3 VIII 8 D 500 IV 4 IX 9 M 1000 V 5 X 10 root mean square RMS The square root of the time average of the square of a quantity for a periodic quantity the average is taken over one complete cycle Also known as effective value RS 232C Bi directional computer serial interface standard defined by the Electronic Industries Association EIA The interface is single ended and non addressable Seebeck effect The development of a voltage due to differences in temperature between two junctions of dissimilar metals in the same circuit self heating Heating of a device due to dissipation of power resulting from the excitation applied to the device The output signal from a sensor increases with excitation level but so does the self heating and the associated temperature measurement error sensitivity The ratio of the response or change induced in the output to a stimulus or change in the input Temperature sensitivity of a resistance temperature detector is expressed as S dR dT setpoint The value selected to be maintained by an automatic controller serial interface A computer interface where information is transferred one bit at a time rather than one byte character at a time as in a parallel interface RS 232C is the most common serial interface SI Systeme International d Unit s See International System of
13. User s Manual Model 370 AC Resistance Bridge 6 EOS im m im m Beep Tana r jm w jm Includes Coverage For Model 370S L U AC Resistance Bridge with Model 3716 L 16 Channel Scanners and 3708 8 Channel Preamp Model 370N AC Resistance Bridge Only ad LakeShore Lake Shore Cryotronics Inc 575 McCorkle Blvd Westerville Ohio 43082 8888 USA E mail Addresses sales lakeshore com service lakeshore com Visit Our Website At www lakeshore com Fax 614 891 1392 Telephone 614 891 2243 Methods and apparatus disclosed and described herein have been developed solely on company funds of Lake Shore Cryotronics Inc No government or other contractual support or relationship whatsoever has existed which in any way affects or mitigates proprietary rights of Lake Shore Cryotronics Inc in these developments Methods and apparatus disclosed herein may be subject to U S Patents existing or applied for Lake Shore Cryotronics Inc reserves the right to add improve modify or withdraw functions design modifications or products at any time without notice Lake Shore shall not be liable for errors contained herein or for incidental or consequential damages in connection with furnishing performance or use of this material Revision 1 14 P N 119 033 14 May 2009 Lake Shore Model 370 AC Resistance Bridge User s Manual LIMITED WARRANTY STATEMENT WARRANTY PERIOD ONE 1 YEAR 1 Lake Shore warrants tha
14. signs are to aid in proper lead arrangement and not indicate signal polarity 2 Active guards can be turned on or off by the user In the off state the guard pins default to signal common but do not substitute for shields Figure 3 3 Resistance Input Connectors 3 4 2 Shielding and Twisting The minimum cable configuration for the Model 370 is shown below Voltage leads and current leads are twisted and run in shielded cable The cable shield is tied to the shield pins on the input connectors Paragraph 3 4 4 explains how the shield can be extended to the experimental Dewar without creating ground loops Pin Name 5 1 6 Shield Shielded Twisted eps Figure 3 4 Twisted Shielded Cables Typical for Current and Voltage Leads 3 4 Installation Lake Shore Model 370 AC Resistance Bridge User s Manual 3 4 3 Driven Guards A configuration of guarded and shielded cables is shown below and is recommended for long cables or when measuring large resistances Variations may be necessary depending on the configuration of available tri axial cable Guards are tied only to the guard pins on the input connector and should never be tied together tied to shields or terminated in any other way They should always be left unattached at the resistor end of the cable The cable shield is tied to the shield pins on the input connectors Paragraph 3 4 4 explains how the shield can be extended to the experimental Dewar without creating ground loops Th
15. 15 35 C at rated accuracy 5 40 C at reduced accuracy Calibration Schedule 1 year Power Requirement 100 120 220 240 VAC 6 10 50 or 60 Hz 50 VA Size 435 mm W x 90 mm H x 317 mm D full rack Weight Model 370 5 9 kg 12 9 Ibs Model 3716 1 kg 2 1 Ibs Approval CE mark consult factory for availability 3716 L and 3708 Scanners Size 135 mm W x 66 mm H x 157 mm D 5 2 in x 2 6 in x 6 2 in Weight 1 kg 2 1 Ib Ordering Information Model 370S AC resistance bridge with 3716 scanner 370L AC resistance bridge with 3716L scanner 370U AC resistance bridge with 3708 scanner 370N AC resistance bridge only 3716 16 channel scanner for Model 370 3716L Low resistance 16 channel scanner for Model 370 3708 Ultra low resistance 8 channel preamp scanner for Model 370 3708 ARW Ultra low resistance 8 channel preamp scanner for Model 370 includes 370 upgrade for instruments with main version 9 27 2005 and input version 1 3 or earlier firmware Select a power configuration VAC 100 Instrument configured for 100 VAC with U S power cord VAC 120 Instrument configured for 120 VAC with U S power cord VAC 220 Instrument configured for 220 VAC with universal Euro line cord VAC 240 Instrument configured for 240 VAC with universal Euro line cord VAC 120 ALL Instrument configured for 120 VAC with U S power cord and universal Euro line cord and fuses for 220 240 Setting extra ch
16. American Standard Code for Information Exchange ASCII A standard code used in data transmission in which 128 numerals letters symbols and special control codes are represented by a 7 bit binary number as follows WOON An A w NO w BO O N ajna wjn o N lt x jz lt je ufo gt jajo y N o quin lt x lt eH v wlo v u Gi a ola jojo o lolololololololo alafojojojo olololo 7 ojo o o ojo ojoj 7 o 3 3 x o alo jajo jojo j o o z z F x H EJa nimo o w gt A mal yu fale a o mol m American Wire Gage AWG Wiring sizes are defined as diameters in inches and millimeters as follows AWG Dia In Dia mm AWG Dia In Dia mm AWG Dia In Dia mm AWG Dia In Dia mm 1 0 2893 7 348 11 0 0907 2 304 21 0 0285 0 7230 31 0 0089 0 2268 2 0 2576 6 544 12 0 0808 2 053 22 0 0253 0 6438 32 0 0080 0 2019 3 0 2294 5 827 13 0 0720 1 829 23 0 0226 0 5733 33 0 00708 0 178 4 0 2043 5 189 14 0 0641 1 628 24 0 0207 0 5106 34 0 00630 0 152 5 0 1819 4 621 15 0 0571 1 450 25 0 0179 0 4547 35 0 00561 0 138 6 0 1620 4 115 16 0 0508 1 291 26 0 0159 0 4049 36 0 00500 0 127 7 0 1443 3 665 17 0 0453 1 150 27 0 0142 0 3606 37 0 00445 0 1131 8 0 1285 3 264 18 0 0403 1 024 28 0 0126 0 3211 38 0 00397 0 1007 9 0 1144 2 906 19 0 0359 0 9116 29 0 0113 0 2859 39 0 00353 0 08969 10 0 1019 2 588
17. Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display Operation 4 31 Lake Shore Model 370 AC Resistance Bridge User s Manual Analog Outputs Continued The next analog output setup screen appears as a prompt for the high endpoint Use the data entry keys to enter the endpoint value between 1x 10 and 100x 10 depending on selected units Enter the numeric value first Press Enter to accept the new value and advance to the units multiplier Use the A or Y key to select a units multiplier of m blank k or M Press Enter to accept the new selection and return to the normal display Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display Figure 4 3 illustrates the operation of an analog output configured to follow an input channel in Kelvin with a low endpoint of 0 mK and a high endpoint of 100 mK Both unipolar and bipolar operation is represented Input OmK 50 mK 100 mK Hz Unipolar Output OV 5V 10 V Input 0 mK 50 mK 100 mK co y 4 SY Bipolar Output 10 V OV 10 V Figure 4 3 Unipolar Versus Bipolar Analog Output 4 17 STILL OUTPUT The Model 370 can provide the small amount of power required to drive the still heater of a dilution refrigerator with analog v
18. UNDER message 8 6 Service Lake Shore Model 370 AC Resistance Bridge User s Manual Noise Signals Continued The VMIX monitor point is the most revealing under normal circumstances and should be examined first The Model 370 uses an analog passband filter 3db 1 40 Hz to reduce response to 50 60 Hz but cannot eliminate all of it Therefore it is useful to observe the signal at measurement levels to discern if wiring changes are resulting in improvements or not Bear in mind that any signal observed at this point is differential in nature and impressed across the measured resistor possibly heating it VMIX is an excellent point to monitor for grounding change effects At extremely low levels 10 uV and below residual electronic noise shows as random fluctuations without a repeating character A level exceeding 10 V peak at this point triggers a VMIX OVL message The VDIF monitor point is taken ahead of the gain and passband sections This point allows broadband monitoring for all differential sources of noise that are impressed across the measured resistor including higher frequencies that the passband filter attenuated These frequencies might be computer generated power harmonic related or even radio frequencies lt 1 MHz As with VMIX signals observed at this point are differential in nature and impressed across the measured resistor possibly heating it A level exceeding 10 V peak at this point triggers a VDIF O
19. Unused leads are connected to measurement common to reduce noise pickup a persistent problem when measuring large resistances Performance of the Model 3716 scanner is so nearly identical to the Model 370 that they share specifications for resistance range accuracy and resolution noise The Model 3716L Low Resistance Scanner The Model 3716L is optimized for low input noise figure 4 nV VHz and can achieve measurement resolution to 20 nQ The scanner s lower noise figure actually improves the measurement resolution of a standalone Model 370 over much of its range It is the best choice for general measurement applications that do not require excitation current below 100 pA and resistance above 100 kQ The 3716L also enables some measurements such as low resistance Hall effect measurements that would not be possible with the 3716 Unused leads are left open allowing the scanner to perform Hall effect measurement sequencing Please note that the specifications for resistance range accuracy and resolution noise are different than the standalone Model 370 The Model 3708 Ultra Low Resistance Preamp Scanner For ultra low AC resistance measurement applications that demand the very best in low noise performance the Model 3708 8 channel preamp scanner is the best choice At just 2 nVRMS VHz the Model 3708 offers the lowest input voltage noise of the three available scanners and can achieve measurement resolution to 10 nQ The Model 3708 improves
20. enabled features Display annunciators are not visible when their associated feature is disabled In addition to the six display annunciators there are four LED annunciators to the right of the display Display Annunciators 5 Indicated that active common mode reduction is on On steady when the alarm feature is on blinks when any alarm is active Indicates scanner is present followed by active scan channel On steady when ramp feature is on blinks when actively ramping setpoint Indicates that zone tuning mode is enabled for closed loop temperature control Indicates that the instrument is under remote control of the computer interface LED Annunciators Autorange On Off On when autorange is on for resistance range Voltage Mode On when in voltage excitation mode Current Mode On when in current excitation mode Autoscan On Off On when autoscanning scanner channels 4 2 Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 4 2 5 Temperature Control Display The lower left corner of the display provides operator feedback when temperature control is enabled The top line shows the control channel if a Model 3716 Scanner is attached The second line shows the control setpoint in resistance or temperature if a temperature response curve is loaded for the control channel The third line shows the heater output and range The heater output can be displayed as either percent of current for the heater output current source range o
21. value Press Escape again to cancel the sequence and return to the normal display 4 11 2 5 Control Filter During normal operation the PID control algorithm uses the unfiltered input readings from the control channel as its feedback This method offers the best control stability and some noise immunity because of the natural averaging that takes place in the integrator I part of the control equation The control algorithm can be configured to use filtered readings instead of unfiltered readings This configuration can improve noise rejection but it is sometimes at the expense of more difficult tuning or even control instability The filter creates time lag that destabilizes the control loop Control uses the same firmware filter parameters set for the control input channel Any control application should be tuned first with the control filter off Filter action should be added gradually starting with a very short settling time and adjustments to P and I made be required along the way most often P will decrease and I will increase The filter time constant should always remain shorter than the I setting in seconds The next control setup screen appears as a prompt for control filter On Off Use the A or Y key to select control filter on or off Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display Operation 4 17 Lake Shore M
22. 1 term with the present setting The following are additional notes on using either IEEE 488 Interface program Ifyou enter a correctly spelled query without a nothing will be returned Incorrectly spelled commands and queries are ignored Commands and queries should have a space separating the command and associated parameters Leading zeros and zeros following a decimal point are not needed in a command string but are sent in response to a query A leading is not required but a leading is required 6 1 6 Troubleshooting New Installation Check instrument address Always send terminators Send entire message string at one time including terminators Send only one simple command at a time until communication is established Be sure to spell commands correctly and use proper syntax INN E TO Attempt both Talk and Listen functions If one works but not the other the hardware connection is working so look at syntax terminators and command format 7 Ifonly one message is received after resetting the interface check the repeat addressing setting It should be enabled Old Installation No Longer Working 8 Power instrument off then on again to see if it is a soft failure 9 Power computer off then on again to see if the IEEE card is locked up 10 Verify that the address has not been changed on the instrument during a memory reset 11 Check all cable connections Intermitten
23. 10 seconds and a 2 of full scale window Sample Query Format FILTER Filter Parameter Query Input FILTER lt channel gt term Format nn lt channel gt Specifies which channel to query 1 16 Returned lt off on gt lt settle time gt lt window gt term Format n nnn nn Refer to command for description Key Begins common interface command Required to identify queries aa String of alpha numeric characters mn String of number characters that may include a decimal point nnn nnnE nn Number represented in scientific notation format term Terminator characters lt gt Indicated a parameter field many are command specific lt state gt Parameter field with only On Off states NOTE Any number being represented in scientific notation may also be entered as a string of number characters If the number is only represented as a string of number characters it cannot be entered in scientific notation The following example shows two different ways of sending the same command Refer to the individual command descriptions for further details SETP 2 0E 03 term Command will set the temperature control setpoint value to 2000 SETP 2000 term Command will set the temperature control setpoint value to 2000 NOTE If there is no scanner attached all lt channel gt parameter values must be entered as 1 6 18 Remote Operation Lake Shore Model 370 AC Resistance Bridge User s Manual Table 6 6 Command Su
24. 2 General Keypad OperatiO vicios Addie 4 4 4 4 DISPLAY SETUP Guiana pain 4 4 4 4 1 Display Setup and Units SelectiON oooonoonnccccnnonicononacanononcncnnnannnonnnnn cnn nar aaa aaa aaa 4 4 4 4 2 Display ReSolUfiONi vicio ao 4 6 4 4 3 Display Brighine S 3 4 r ens eee tan eee i R 4 6 4 5 EXCITATION MODES RESISTANCE RANGE AND AUTORANGE cocccccocccccoconcncnnnnncnnnnnnnnnnannnncnnns 4 6 4 5 1 Voltage Excitation MOJE colo o ara consi iatera ta iauna inisi 4 7 4 5 2 Current Excitation MOJE u seo Z tO OLE a 4 7 4 5 3 iTumi g ExcitationOfi x asza ni a E Ea SEL 4 8 4 6 INPUT SETUP 00 ia 4 8 4 6 1 Channel Selection emitan 4 9 4 6 2 Scanner Dwell TIME piridin a edo 4 9 4 6 3 Change Pause TIME siii aaa 4 9 4 6 4 Selecting a Temperature Response Curve cocccccococccccononccononcncnnnonnnnn nn nc nano nn nr aaa aaa niania 4 10 4 6 5 Temperature Coeficiente diia 4 10 4 6 6 Tuming A Channel On ii a eet A A Ao 4 10 4 7 INSTRUMENTI SETUP a ooo ts 4 11 4 7 1 Frequency Select tatiana e st ia WPP iaa inicie agenda 4 11 4 7 2 Active Common Mode Reduction eee eee eee aaa aaa aaa aaa aaa inniia 4 11 4 7 3 Guard DIVE zt Yad at Ni NA R Nd Ad igi SI a 4 11 4 7 4 Monitor Output Selection cocinan tdi 4 12 Table of Contents iii Lake Shore Model 370 AC Resistance Bridge User s Manual TABLE OF CONTENTS Continued Chapter Paragraph Title Page 4 7 5 Range and Excitation Keys for One or All Channels oocoonoccccnnocccnnoncncnonon
25. 20 0 0338 0 8118 30 0 0100 0 2546 40 0 00314 0 07987 ambient temperature The temperature of the surrounding medium such as gas or liquid which comes into contact with the 1 apparatus ampere The constant current that if maintained in two straight parallel conductors of infinite length of negligible circular cross section and placed one meter apart in a vacuum would produce between these conductors a force equal to 2 x 107 newton per meter of length This is one of the base units of the SI ampere turn A MKS unit of magnetomotive force equal to the magnetomotive force around a path linking one turn of a conducting loop carrying a current of one ampere or 1 26 gilberts ampere meter A m The SI unit for magnetic field strength H 1 ampere meter 41 1000 oersted 0 01257 oersted analog controller A feedback control system where there is an unbroken path of analog processing between the feedback device sensor and control actuator heater analog data Data represented in a continuous form as contrasted with digital data having discrete values analog output A voltage output from an instrument that is proportional to its input For example from a digital voltmeter the output voltage is generated by a digital to analog converter so it has a discrete number of voltage levels asphyxiant gas A gas which has little or no positive toxic effect but which can bring about unconsciousness and death by displacing air and thus depr
26. 200 632 mQ 200mQ 63 2 MQ 20 mQ 2 0 mQ 10mA 600uQ 2004Q 60402 40pQ 13uQ 4pQ 13uQ 600nQ 600nQ 600nQ 600 nQ 320 4W 100pW 324W 10uW 32UW 1uW 320nW 100nW 32nw 10nW 1 0 nW 2 0 kQ 632 Q 200 Q 63 2 Q 200 6 32 Q 200 632 mQ 200mQ 63 2mQ 6 32 mQ 316pA 2mQ 60040 200pQ 130uQ 40uQ 13pQ 404Q 2040 2040 20pQ 2 0 yQ 100 pW 324W 10uW 32UW 10uW 320nW 100nW 32nw 10nw 3 2nW 320 pW 6 32 kQ 2 0 KQ 632 Q 200 Q 63 2 Q 200 6 32 Q 200 632 ma 200 ma 20 mQ 100uA 6 0mQ 20mQ 600uQ 40040 1304Q 40uQ 13pQ 60pQ 6040 6010 6 0 yQ 32 UW 10 uw 3 2 UW 1 0 uW 320 nW 100 nW 32 nW 10 nW 3 2 nW 1 0 nW 100 pW 20 KQ 6 32 kQ 2 0 kQ 632 Q 200 Q 63 2 Q 200 6 32 Q 200 632 mQ 63 2 mQ 31 6pA 20mQ 60m2 20mQ 13mQ 4004Q 130uQ 40pQ 20402 2040 2010 20 yQ 10 uw 3 2 UW 1 uw 320 nW 100 nw 32 nW 10 nW 3 2 nW 1 0 nW 320 pW 32 pW 63 2 KQ 20 kQ 6 32 kQ 2 0kQ 632 Q 200 Q 63 2 Q 200 6 32 Q 200 200 mQ 10 HA 60 mQ 20 MQ 6 0 mQ 4 0 MQ 1 3mQ 400 uQ 130 Q 60 yQ 60 Q 60 Q 60 Q 3 2 UW 1 0 pW 320 nW 100nW 32 nW 10nW 3 2nW 1 0 nW 320 pW 100 pW 10 pW 200 kQ 63 2 kQ 20 kQ 6 32 KQ 2 0kQ 632 Q 200 Q 63 2 Q 20 Q 6 32 Q 632 MQ 3 16 pA 200 ma 60 mQ 20 MQ 13 mQ 4 0 MQ 1 3 MQ 400 yQ 200 Q 200 uQ 200 Q 200 Q 1 0 uw 320 nW 100 nW 32 nW 10 nW 3 2 nW 1 0 nW 320 pW 100 pW 32 pW 3 2 pW 200 kQ 63 2 KQ 20 kQ 6 32 KQ 2 0kQ 632 Q 200 Q 63 2 Q 200 200 1 0 HA 200 ma 60 mQ 40 mQ 13 mQ 4mQ 1 3 MQ 600 Q 600 Q 600 Q 600 Q 100 nW 32 nW 10nW 3
27. 2nW 1 0 nW 320 pW 100 pw 32 pw 10 pw 1 0 pW 2 0 MQ 200 kQ 63 2 KQ 20 kQ 6 32 kQ 2 0 kQ 632 Q 200 Q 63 2 Q 6 32 Q 316 nA hs 200 ma 130 mQ 40 MQ 13mQ 40mQ 2 0 mQ 2 0 mQ 2 0 mQ 2 0 MQ 100nW 10nW 3 2 nW 1 0 nW 320 pW 100 pw 32 pw 10 pw 3 2 pW 320 fW 6 32 MQ 2 0 MQ 200 KQ 63 2 KQ 20 KQ 6 32 kQ 2 0 kQ 632 Q 200 Q 200 100 nA ES sa 400 mQ 130 mQ 40 MQ 13 mQ 10mQ 6 0 mQ 6 0 mQ 6 0 MQ 32 nW 10nW 1 0 nW 320 pW 100 pW 32 pW 10 pW 3 2 pW 1 0 pw 100 fW E 6 32 MQ 2 0 MQ 200 KQ 63 2 kQ 20kQ 6 32 KQ 2 0 kQ 632 Q 63 2 Q 31 6 nA w kk e 400 MQ 130 mQ 60 mQ 30mQ 30 ma 20 MQ 20 MQ i 3 2nw 10nW 100pW 32pW 10pW 32pw 10pw 320W 32 1W gt 6 32 MQ 2 0 MQ 200 KQ 63 2 kQ 20 KQ 6 32 KQ 2 0kQ 200 Q 10 nA gt il ka ke 600 ma 200 ma 200 ma 130 mQ 100 mQ 60 mQ 320pW 100pW 10pw 32pw 10pw 320fW 100W 10 W id ij 6 32 MQ 2 0 MQ 200 kQ 63 2 kQ 20 kQ 6 32 kQ 632 Q 3 16nA il 8 M j e 200 630 ma 600 ma 320 ma 200 ma i ll si 32 pW 10 pW 1 0 pW 320 fW 100 fW 32 W 3 2 fW gt a ki bi 6 32 MQ 2 0 MQ 200 kQ 63 2 KQ 20 kQ 2 0 ka 1 0nA hi s ida bi 6 0 Q 3 20 2 0 Q 100 ki gt 3 2 pW 1 0pW 100 fW 32 fW 10 W 1 0 W 7 i 6 32 MQ 2 0 MQ 200 kQ 63 2 kQ 6 32 KQ 316 pA 7 id gt hi ki 38 200 100 320 j i ki ii i 320 fW 100 fW 10 W 3 2 W 320 aW E i i i bi 6 32 MQ 2 0 MQ 632 kQ 200 kQ 63 2 kQ 20 kQ 100 pA i a gt lt bi kj 600 600 30 Q 200 bl y il bi ki kd 32 fW 10 W 3 2 fW 1 0 W 320 aW 100 aW d id i hi u 6 32 MQ 2 0 MQ 632 kQ 200 kQ 63 2 KQ 31 6
28. 6 mA Zero CALZ 3 7 60 Q 2 31 6 mA Gain CALG 3 7 60 Q 3 10 mA Zero CALZ 3 6 200 Q 3 10 mA Gain CALG 3 6 200 Q 4 3 16 mA Zero CALZ 3 5 600 Q 4 3 16 mA Gain CALG 3 5 600 Q 5 1 mA Zero CALZ 3 4 2 kQ 5 1 mA Gain CALG 3 4 2 kQ 6 316 uA Zero CALZ 3 3 6 KQ 6 316 UA Gain CALG 3 3 6 kQ 7 100 UA Zero CALZ 3 2 20 kQ 7 100 uA Gain CALG 3 2 20 kQ 8 31 6u4A Zero CALZ 3 1 60 KQ 8 31 6 uA Gain CALG 3 1 60 kQ Heater Output Zero Offset Coefficient Voltage 0 Actual Resistance Nominal Current x 1 07 x 100 Heater Output Gain Coefficient Nominal Current Voltage 100 Actual Resistance 8 12 5 Calibration Commands The following interface commands are only used during calibration and are in addition to those listed in Chapter 6 CALG Gain Calibration Constant Command Input CALG lt input gt lt range gt lt value gt term Format n nn nnn nnnE nn lt input gt Specifies the input or output to assign the gain calibration constant 1 Current Excitation range 2 Voltage Excitation range 3 Heater range 4 Analog Output 5 Frequency 6 Scanner lt range gt Specifies the calibration range 1 22 for Current Excitation range calibration Reference RDGRNG 1 12 for Voltage Excitation range calibration Reference RDGRNG 1 8 for Heater range calibration Reference HTRRNG 1 2 for Analog Output calibration 1 3 for Frequency compensation Reference FREQ 1 for Scanne
29. 7 This bit is set when a ramp is completed Service Request SRQ Bit 6 Determines whether the Model 370 is to report via the SRQ line If bits 2 3 4 5 and or 7 are set then the corresponding bit in the Status Byte Register will be set The Model 370 will produce a service request only if bit 6 of the Service Request Enable Register is set If disabled the Status Byte Register can still be read by the BUS CONTROLLER by means of a serial poll SPE to examine the status reports but the BUS CONTROLLER will not be interrupted by the Service Request The STB common command will read the Status Byte Register but will not clear the bits Standard Event Status ESB Bit 5 When bit 5 is set it indicates if one of the bits from the Standard Event Status Register has been set Refer to Paragraph 6 1 4 2 Error Bit 4 This bit is set when there is an instrument error not related to the bus The following errors can trigger this bit to be set CS OVL VCM OVL VDIF OVL VMIX OVL R OVER R UNDER T OVER T UNDER Refer to Paragraph 8 6 for a description of error messages Alarm Bit 3 This bit is set when there is an alarm condition Valid Read Bit 2 This bit is set when each valid reading is taken 6 1 4 2 Standard Event Status Register and Standard Event Status Enable Register The Standard Event Status Register reports IEEE bus status of the Model 370 STANDARD EVENT STATUS REGISTER FORMAT Bit ICO
30. 8 6 Instrument Hardware are related to the instrument hardware and may require help from Lake Shore service The messages listed in Paragraph 8 6 Measurement of Compliance Limit do not indicate a problem with the instrument and will disappear when input setup is complete The messages listed in Paragraph 8 6 Illegal Operation indicate user intervention corresponding to the description of the message in the table The Model 370 should be allowed to warm up for a minimum of 30 minutes to achieve rated accuracy 4 2 DISPLAY DEFINITION The Model 370 has an 8 line by 40 character vacuum fluorescent VF display capable of showing both text and graphic images This paragraph describes features of the display that appear during normal operation including resistance measurement temperature measurement and temperature control Other display configurations appear during parameter setting and data entry operations These displays are illustrated in their individual operation paragraphs Alarm On Off Only Present if 3716 Ramp On Off Zone On or Blinking Scanner Attached or Blinking or Off Remote On Para 4 14 Para 4 8 Para 4 11 7 Para 5 3 or Off CMR On or Off Para 6 1 2 Para 4 7 2 A Corresponds to Resistance Range or Autorange Buttons Para 4 5 Control Channel Para 4 11 2 2 Changing Setpoint Para 4 11 3 Corresponds to Setpoint Units Voltage Excitation Para 4 11 2 3 Heater Output Display Display Setup Calculated Power
31. 9 Heater Polarity When the heater output is used in a traditional manner to drive a resistive load the heater output needs to be unipolar since reversing the current will still heat the load There may be some rare circumstances when the heater output is not connected to a resistive load and bipolar control is desired The heater output can be set to control the current to plus and minus the full scale range by setting the heater polarity to bipolar mode For the majority of applications and any application where the heater output is connected to a resistive load the unipolar mode should be chosen The next control setup screen appears as a prompt for the heater polarity Use the A or Y key to select the heater polarity of unipolar or bipolar Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel and return to the normal display 4 11 3 Changing Setpoint The control setpoint indicates the desired load temperature It can be expressed in either ohms or Kelvin If there is not a temperature response curve assigned to the control channel the setpoint units will default to ohms To change setpoint units refer to Paragraph 4 11 2 3 The instrument allows a large setpoint range to accommodate a variety of sensor types and applications In resistance the setpoint can range from micro ohms to mega ohms In temperature the setpoint can range from milliKelvin to over 1000 Kelvin A setpoin
32. All of these calibrations are done through computer interface and the calibration constants are stored in non volatile memory There are no trim pots inside the Model 370 and the cover does not have to be removed to calibrate the instrument In contrast the Model 3716 scanner is calibrated with a single trim pot that is accessible from the scanner front panel 8 12 1 Calibration Interface Computer interface commands are included in the Model 370 specifically for calibration These commands work with either the IEEE 488 or RS 232C interface A complete description of each command is included in Paragraph 8 12 5 Communication must be established between the instrument and a computer before calibration can be attempted Some time should be spent becoming familiar with the following commands before beginning a calibration The RDGR and RDGRNG commands are also important to calibration CALZ CALZ Send or query the zero offset calibration coefficient for a measured input or calibrated output The measurement current source does not require zero offset coefficients The zero offset is in the same units as the input or output being calibrated and the default value is 0 CALG CALG Send or query the gain coefficient for a measured input or calibrated output The gain coefficient is a ratio of actual and ideal input or output and the default value is 1 CALVIN Query the voltage measured at the instrument input for calibration of voltage measurement
33. Audible Alarm Beeper Query 23 Display Brightness Command eee 24 Display Brightness Query 211112242122121 412111 1 24 Input Change All or One Cmd 63 24 Input Change All or One Query 1 1 1 1 24 Temperature Control Mode Cmd 24 Temperature Control Mode Query 24 Common Mode Reduction Cmd 24 Common Mode Reduction Query 24 Temp Control Polarity Cmd Temp Control Polarity Query curve Delete Emi cian Curve Header Cmd Cuye Header Quo Curve Data Point Cmd cooooccccnccccconcnnnnnnnnnnnancnnncncnonos Curve Data Point Query Temperature Control Setup Parameter Cmd Temperature Control Setup Parameter Query 26 Factory Defaults Cmd Display Number Cmd Display Number Query c ccccceseseseeeeeeseeeeeeeeneees Displayed Location Parameter Cmd Displayed Location Parameter Query Digital Output Cindy A Digital Output Query 021 Filter Parameter Cid coococcccncnccccnnnoncnininononononocnnncnos 27 Filter Parameter Query coooccicnincnncinnoninnnnnncnncnnnncnnnos 28 Command FREQ FREQ GUARD GUARD HTR HTRRNG HTRRNG HTRST IEEE IEEE INSET INSET KEYST LDAT LINEAR LINEAR LOCK LOCK MDAT MNMX MNMX MNMXRST MODE MODE MONITOR MONITOR MOUT MOUT PID PID RAMP RAMP RAMPST RDGK RDGPWR RDGR RDGRNG RDGRNG RDGST RELAY RELAY RELAYST
34. C Cable capacitance in farads choose largest of one times lead to lead or lead to shield Lead inductance can create similar effects but typical installation techniques will not produce large enough inductance to cause measurable attenuation Inductance should be considered if inductive noise reduction filters are placed in series with leads 2 6 7 Instrument Operation There are several features of the Model 370 intended to improve accuracy and repeatability that can have the opposite effect when used improperly This paragraph provides a brief summary of operating features that may contribute to measurement uncertainty Use of the Model 370 with the Model 3716 Scanner can improve or degrade measurement performance depending on application The scanner can be calibrated in the field to eliminate gain error but the addition of a large number of lead wires is always unpredictable More information on the Model 3716 Scanner is given in Paragraph 2 8 Measurement uncertainty increases significantly if the instrument hardware and firmware filters are not allowed to settle completely Information on settling time is given in Paragraph 2 4 6 The common mode reduction feature can significantly reduce common mode voltage on the measurement leads when used as described in Paragraph 2 4 5 5 Be sure that all four measurement leads remain isolated from measurement common shield when using the feature If one lead is grounded the instrument is able to
35. D Manual Output Heater Low High Analog Out1 Analog Out2 Range Relay Relay 0 100 0 100 P 0 1 1k 0 1 1k 0 200 0 100 On Off On Off Zone 02 H_ Seetpo int D Manual Output Heater Low High Analog Out1 Analog Out2 Range Relay Relay 0 100 0 100 P l 0 1 1k 0 1 1k 0 200 0 100 On Off On Off Zone 01 a S etpoint D Manual Output Heater Low High Analog Out1 Analog Out2 Range Relay Relay 0 100 0 100 P l 0 1 1k 0 1 1k 0 200 0 100 On Off On Off Figure 5 1 Record of Zone Settings Zone Worksheet eps Advanced Operation Lake Shore Model 370 AC Resistance Bridge User s Manual This Page Intentionally Left Blank 5 12 Advanced Operation Lake Shore Model 370 AC Resistance Bridge User s Manual CHAPTER 6 COMPUTER INTERFACE OPERATION 6 0 GENERAL This chapter provides operational instructions for the computer interface for the Lake Shore Model 370 AC Resistance Bridge Either of the two computer interfaces provided with the Model 370 permit remote operation The first is the IEEE 488 Interface described in Paragraph 6 1 The second is the Serial Interface described in Paragraph 6 2 The two interfaces share a common set of commands detailed in Paragraph 6 3 Only one of the interfaces can be used at a time 6 1 IEEE 488 INTERFACE The IEEE 488 Interface is an instrumentation bus with hardware and programming standards that simplify instru
36. Excitation range 3 Heater range 4 Analog Output lt range gt Specifies the calibration range 1 12 for Voltage Excitation range calibration Reference RDGRNG 1 8 for Heater range calibration Reference HTRRNG 1 2 for Analog Output calibration lt value gt Zero offset calibration constant value Set value to 0 for default setting Zero offset calibration constant is always a number near zero in the same units as lt input gt Use CALSAVE to permanently save the zero offset calibration constants Calibrated Reading Uncalibrated Reading CALZ x CALG Zero Offset Calibration Constant Query CALZ lt input gt lt range gt term n nn lt input gt Specifies input or output to query 1 4 lt range gt Dependent on lt input gt lt value gt term nnn nnnE nn Refer to command for description Service 8 25 Lake Shore Model 370 AC Resistance Bridge User s Manual 8 13 MODELS 3716 and 3716 L SCANNER CALIBRATION The Model 3716 and 3716 L scanners have one trim pot for gain calibration of the pre amplifier contained in the scanner The pre amplifier is used for voltage ranges under 20 mV only Higher voltage ranges do not require pre amplification Scanner calibration is independent of the Model 370 being used to calibrate with The Model 370 has a separate calibration related to scanner operation described in Paragraph 8 12 2 5 When properly calibrated scanners and bridges can be interchange
37. Output Para 4 5 1 Para 4 11 2 6 Para 4 4 Is Always Displayed or Current Excitation Para 4 2 3 Para 4 5 2 Front Display eps Figure 4 1 Model 370 Display Operation 4 1 Lake Shore Model 370 AC Resistance Bridge User s Manual 4 2 1 Reading Display The measurement reading area includes the second through fifth lines of the display The user can configure this area to display between 1 and 8 reading locations One to four readings are displayed with larger 11 x 15 block characters Five to eight readings are displayed with typical 5 x 7 block characters Each reading location can be configured with a channel if a Model 3716 Scanner is installed data source including resistance measurement temperature measurement maximum minimum or the results of the linear equation Display setup is described in Paragraph 4 4 The following symbols are used in the reading area fit m BH or Display update indicator Channel indicators for Model 3716 Scanner channels Minus sign indicates improper lead arrangement for resistance measurement Resistance in ohms Temperature in Kelvin Result of maximum capture Result of minimum capture Result of the linear equation 4 2 2 Display Update Indicator The display update indicator blinks with 3 distinct patterns to indicate the phases of a measurement cycle The indicator remains on continuously to indicate hardware settling after a range change Blinking with on time equal to
38. Phillips screwdriver to install 4 flat head screws centered front amp rear on the top amp bottom of the enclosure Use 5 64 hex key to install 4 screws attaching top panel to unit Use 5 64 hex key to tighten 4 screws attaching bottom panel to unit If required reattach 19 inch rack mounting brackets Connect power cord to rear of unit and set power switch to On I 8 8 Service Lake Shore Model 370 AC Resistance Bridge User s Manual 8 9 ELECTROSTATIC DISCHARGE Electrostatic Discharge ESD may damage electronic parts assemblies and equipment ESD is a transfer of electrostatic charge between bodies at different electrostatic potentials caused by direct contact or induced by an electrostatic field The low energy source that most commonly destroys Electrostatic Discharge Sensitive ESDS devices is the human body which generates and retains static electricity Simply walking across a carpet in low humidity may generate up to 35 000 volts of static electricity Current technology trends toward greater complexity increased packaging density and thinner dielectrics between active elements which results in electronic devices with even more ESD sensitivity Some electronic parts are more ESDS than others ESD levels of only a few hundred volts may damage electronic components such as semiconductors thick and thin film resistors and piezoelectric crystals during testing handling repair or assembly Discharge voltages below 4000 v
39. Protection ts 3 13 3 7 6 Boosting Heater Power Above 1W eee eee ea eee aaa aaa aaa aaa cnn nn aranz 3 14 3 8 ANALOG OUTPUT AND STILL HEATER OUTPUT eee e eeee eee aaa aaa wana nano nnnnnnnnnnnnns 3 14 3 8 1 Analog Output Connections ueeeee e ee eect E Ee aa aa ae E aea E ea Erea ae a a a ikia 3 14 3 8 2 Calculating Still Heater Power eee aaa aaa aaa aaa nr cnn nano nn rr aaa aan 3 14 3 8 3 Calculating Still Heater Current 1 4 13 cess yy o esses AA A AA Y 3 14 3 9 RELAY CONNECTION sos oesie aaa ii Adina ETS 3 15 3 10 RACK MOUNTING v riiintean nn ane AE aa EE E EE a TAT EAEE EE EEA E E R r 3 15 OPERATION aeara a aaa araar a ar eaaa aera a a ia 4 1 4 0 GENERAL rg RSA ena AE ENAR AK o a as te et ea AA 4 1 4 1 TURNING POWER ON cuco cc id taa 4 1 4 2 DISPLAY DEFINITION kenara e a tc 4 1 4 2 1 Reading Display O A A ie Aa il ee 4 2 4 2 2 Display Update Indicator a aaa nana aaa aaa aaa aaa ne nn cnn cnn aaa T Aai naa 4 2 4 2 3 Range Excitation and Power u uu eee eee aaa aaa aaa h ara a aaa aaa hiaai ae aaeenanaa 4 2 4 2 4 Display and LED Annunciators e eee ee aaa aaa aaa aaa aaa aaa aaa aaa aaa aaa cana aaa 4 2 4 2 5 Temperature Control Display eeeeee eee aaa aaa aaa aaa aaa rr rn 4 3 4 3 KEPAD DEFINITION see owa W A O OW dO sites Ea DAt WESOLO SL 4 3 4 3 1 Key Descriptions o ck A GEE O id 4 3 4 3
40. Refer to Paragraph 8 2 Contact information may change periodically but current contact information can always be found on the Lake Shore web site www lakeshore com Lake Shore Cryotronics Inc Instrument Service Department Mailing Address 595 McCorkle Blvd Westerville OH USA 43082 8888 E mail Address myicefdlaksshowcom _ Insert Service Telephone 614 891 2244 Sales 614 891 2243 ext 131 Instrument Service Fae 614 818 1600 Sales 614 818 1609 Instrument Service When contacting Lake Shore please provide your name and complete contact information including e mail address if possible It is often helpful to include the instrument model number and serial number located on the rear panel of the instrument as well as the firmware revision information as described in Paragraph 4 19 8 2 RETURNING PRODUCTS TO LAKE SHORE If it is necessary to return the Model 370 or accessories for recalibration repair or replacement a Return Goods Authorization RGA number must be obtained from a factory representative or from the Lake Shore web site Do not return a product to Lake Shore without an RGA number The following information must be provided to Lake Shore in order to obtain an RGA number Instrument model and serial number User name company address phone number and e mail address Malfunction symptoms R A A Description of the system in which the product is used If possible the orig
41. Register To enable an event flag bit send the command ESE with the sum of the bit weighting for each desired bit Refer to Paragraph 6 1 4 2 for a list of event flags Example To enable event flags 0 4 and 7 send the command ESE 145 term 145 is the sum of the bit weighting for each bit Bit Bit Weighting Event Name 0 1 OPC 4 16 EXE 7 128 PON 145 xESE Event Status Enable Register Query Input ESE term Returned lt bit weighting gt term Format nnn Refer to Paragraph 6 1 4 2 for a list of event flags xESR Standard Event Status Register Query Input ESR term Returned lt bit weighting gt term Format nnn Remarks The integer returned represents the sum of the bit weighting of the event flag bits in the Standard Event Status Register Refer to Paragraph 6 1 4 1 for a list of event flags IDN Identification Query Input IDN term Returned lt manufacturer gt lt model gt lt serial gt lt date gt term Format aaaa aaaaaaaa aaaaaa mmddyyyy lt manufacture gt Manufacturer ID lt model gt Instrument model number lt serial gt Serial number lt date gt Instrument firmware revision date Example LSCLMODEL370 123456 02032001 term OPC Operation Complete Command Input OPC term Remarks Generates an Operation Complete event in the Event Status Register upon completion of all pending selected device operations Send it as the last command in a command string 6 20 Remote Operation OPC Input Returned
42. SR1 RL1 PPO DC1 DTO CO E1 Software Support LabView driver for IEEE 488 interface consult factory for availability Introduction 1 9 Lake Shore Model 370 AC Resistance Bridge User s Manual Specifications Continued Interface Continued Serial Interface Capability RS 232C DE 9 Connector 9600 BAUD Alarms Number Up to 32 high and low for each channel Settings Source High Setpoint Low Setpoint Deadband Latching Non latching Audible On Off Actuators Display annunciator beeper relays Relays Number 2 Contacts Normally Open Normally Closed and Common Contact Rating 30 VDC at 5A Operation Relays may be activated on high and low alarm setpoints or manually Connector Detachable terminal block Analog Voltage Outputs Number 2 Type Variable DC voltage source Scale User specified Range 10 V Resolution 0 3 mV 0 003 of full scale Accuracy 2 5 mV Maximum Current 100 mA Maximum Power 1 W Minimum Load Resistance 100 short circuit protected Ground Reference Chassis Operation Tracks reading error DR using linear equation or use as still heater Connector BNC Monitor Output Operation User selects one of several analog voltage diagnostic points must remain isolated Connector BNC Frequency Reference Signal Type Phase sensitive detector reference must remain isolated Amplitude 0 to 5 V nominal Waveform Square wave Connector BNC General Ambient Temperature
43. See Kelvin Scale Celsius Scale and ITS 90 Proper metric usage requires that only Kelvin and degrees Celsius be used However since degrees Fahrenheit is in such common use all three scales are delineated as follows Boiling point of water 373 15 K 100 C 212 F Triple point of water 273 16 K Freezing point of water 273 15 K 0 C 32 F Absolute zero OK 273 15 C 459 67 F kelvin Celsius Fahrenheit To convert Kelvin to Celsius subtract 273 15 To convert Celsius to Fahrenheit multiply C by 1 8 then add 32 or F 1 8 x C 32 To convert Fahrenheit to Celsius subtract 32 from F then divide by 1 8 or C F 32 1 8 temperature coefficient measurement The measurement accuracy of an instrument is affected by changes in ambient temperature The error is specified as an amount of change usually in percent for every one degree change in ambient temperature tesla T The SI unit for magnetic flux density B 1 tesla 104 gauss Glossary of Terminology A 7 Lake Shore Model 370 AC Resistance Bridge User s Manual thermal emf An electromotive force arising from a difference in temperature at two points along a circuit as in the Seebeck effect tolerance The range between allowable maximum and minimum values torr Unit of pressure torr 1 mm of mercury 1 atmosphere 760 torr transistor transistor logic TTL A logic circuit containing two transistors for driving large output capacitances at high
44. Some users prefer voltage excitation mode because it matches the operation of other bridges and permits convenient limiting of excitation power Resistance range selection 1s different in these two modes as described below Autorange is available for both modes Voltage Excitation Mode The user specifies an input voltage range that is a not to exceed setting for the instrument The instrument selects the appropriate current setting that keeps the voltage across the resistor as high as possible without exceeding the voltage range setting Current is changed when the resistance range changes to maintain the voltage limit This mode is preferred when measuring negative temperature coefficient materials common in cryogenic temperature sensors With voltage limited excitation power in the resistor reduces with lowering temperature P V R limiting the effect of self heating Current Excitation Mode The user specifies a fixed excitation current The instrument selects the appropriate voltage gain for the selected resistance range The instrument changes gain when the resistance range changes In this mode it is easier to calculate the actual power in the measured resistor P I x R because the true current is displayed In voltage excitation mode the voltage range displayed is the maximum not the actual voltage 2 4 4 Resistance Ranges The Model 370 has full scale resistance ranges from 2 mQ to 2 MQ Many ranges are available with several d
45. Units stability The ability of an instrument or sensor to maintain a constant output given a constant input still A device used to evaporate liquids heat is applied to the liquid and the resulting vapor is condensed to a liquid state strain relief A predetermined amount of slack to relieve tension in component or lead wires Also called stress relief superconducting magnet An electromagnet whose coils are made of a type II superconductor with a high transition temperature and extremely high critical field such as niobium tin Nb3Sn it is capable of generating magnetic fields of 100 000 oersteds and more with no steady power dissipation See electromagnet susceptance In electrical terms susceptance is defined as the reciprocal of reactance and the imaginary part of the complex representation of admittance suscept ibility conduct ance susceptibility x Parameter giving an indication of the response of a material to an applied magnetic field The susceptibility is the ratio of the magnetization M to the applied field H x M H In both SI units and cgs units the volume susceptibility is a dimensionless parameter Multiply the cgs susceptibility by 47 to yield the SI susceptibility See also Initial Susceptibility and Differential Susceptibility As in the case of magnetization the susceptibility is often seen expressed as a mass susceptibility or a molar susceptibility depending upon how M is expressed temperature scales
46. a serial poll but does not reset the register to all zeros The integer returned represents the sum of the bit weighting of the status flag bits that are set in the Status Byte Register Refer to Paragraph 6 1 4 1 for a list of status flags Self Test Query TST term lt status gt term n lt status gt 0 no errors found 1 errors found The Model 370 reports status based on test done at power up Wait to Continue Command WAI term This command is not supported in the Model 370 Remote Operation 6 21 Lake Shore Model 370 AC Resistance Bridge User s Manual ALARM Alarm Parameter Command Input ALARM lt channel gt lt off on gt lt source gt lt high value gt lt low value gt lt deadband gt lt latch gt term Format nn n n tnnn nnnE nn nnn nnnE nn nnn nnnE nn n lt channel gt Specifies which channel to configure 1 16 0 all channels lt off on gt Specifies alarm checking on or off 0 Off 1 On lt data source gt Specifies channel data to check 1 Kelvin 2 Ohms 3 Linear Data lt high value gt Sets the value the source is checked against to activate the high alarm lt low value gt Sets the value the source is checked against to activate low alarm lt deadband gt Sets the value that the source must change outside of an alarm condition to deactivate an unlatched alarm lt latch gt Specifies a latched alarm remains active after alarm condition correction 0 Non Latc
47. aaa aaa cacca 2 21 2 10 3 1 Setting Heater Range Albi 2 21 2 10 3 2 FUNING Proportional ae ic ie tai 2 21 2 10 3 3 Tuning Integral ion ida 2 22 2 10 3 4 Tuning DeriVatiVe ss esse tiara 2 22 3 INSTALLATION 00 ita 3 1 3 0 GENERAL ed esa ia ts 3 1 3 1 INSPECTION AND UNPACKING uuu eee ae eee aaa aaa aaa E rr narrar rre 3 1 3 2 REAR PANEL DEFINITION 004480 onenen Gee AOL O biti vues reese ieee 3 2 3 3 LINE INPUT ASSEMBLY gt arenaen aa ea annan sue neednedeneess stun EIU poka ATE YE WYDZ WWE 3 3 3 3 1 LING Voltage iia aa iae aa eatarra eaaa aa ein ectet adria rta 3 3 3 3 2 Line Fuse and Fuse Holder es wow yty OD A sites sebbbi lt tr 3 3 3 3 3 POW6Pr GONG OA 3 3 3 3 4 Power Swit Chics ib 3 3 3 4 RESISTOR INPUT CONNECTION AND CABLING ueuee eee eee aaa aaa aaa aaa aan aaa naar ncnnnnnnss 3 4 3 4 1 Resistor Input Connector Definition eue ee eee aaa aaa aaa aaa aaa aaa 3 4 3 4 2 Shielding and T WISting 28 assess AA A AEO doba esa deoc aa 3 4 3 4 3 Driven Guards wat i ai cde A EE A LA SARE elza 3 5 3 4 4 iIsolation and GrOundINGj ht zk a toz deo R Epia R ka 3 5 3 4 5 Dressing Cables nania Ri O AA AA a aiii 3 6 3 4 6 Temperature SENSO ii di ajos 3 7 3 5 MODEL 3716 SCANNER INSTALLATION eee eee eeee ee aaa eee aaa aaa nr nan nn rn nn nr rr nar rr rn 3 7 3 5 1 Grounding the SCAMMER terciarios tocan EE tetra fede di 3 7 3 5 2 Scanner Resistor Input Connector Definition
48. amplifier gain and greater sensitivity than DC techniques The AC technique is also naturally bipolar allowing any remaining thermal EMF or offset voltages to be averaged out of the resistance measurement A relatively low frequency of excitation 13 7 Hz was chosen to reduce the effects of lead capacitance on the measurement In a typical cryostat lead heat sink requirements and limited space make it difficult to reduce capacitance enough to operate at significantly higher frequency Environmental noise can limit the resolution of low excitation measurement Well designed filtering is necessary to reclaim the desired signal from the noise The Model 370 takes advantage of the AC filtering technique called phase sensitive detection that is used in Lock In Amplifiers This technique has no DC equivalent Even with phase sensitive detection it is still necessary to shield these leads from environmental noise The Model 370 facilitates appropriate shielding techniques in addition to offering more active noise reduction circuits Theory of Operation 2 5 Lake Shore Model 370 AC Resistance Bridge User s Manual 2 4 3 Excitation Modes The Model 370 uses current source excitation for resistance measurements Current excitation mode is the most fundamental measurement mode because operation is tied directly to hardware function The instrument firmware is also able to simulate voltage excitation of the measured resistor to create a second operating mode
49. an appropriate resistance range The instrument will increase the resistance range decrease excitation current if the measurement reaches full scale Normally no over range error is created because of the approximate 20 over range capability of the input Over range errors are possible when resistance changes very quickly or when resistance increases beyond the limit of the voltage excitation selection The autorange algorithm will not change the excitation setting to accommodate larger resistances The instrument will decrease the resistance range increase excitation current if the measurement drops below 20 of full scale Autorange 1s not recommended when resistance changes are very fast measurements are very noisy or when data acquisition cannot tolerate a range change delay in the middle of a data set To turn autorange on or off press the Autorange key The LED annunciator above the key lights to indicate autorange is on 4 5 2 Current Excitation Mode In current excitation mode one of the instrument current source selections is chosen directly as excitation Gain of the measurement input is changed to accommodate different resistance ranges To use current excitation mode press the Current Mode key The LED annunciator above that key will light to indicate the excitation mode The excitation display line in the lower right corner of the display will also change to indicate the current excitation in amps Voltage excitation arrow keys are lo
50. an open circuit when actuator is in the free unenergized position O D Outer diameter Glossary of Terminology A 5 Lake Shore Model 370 AC Resistance Bridge User s Manual oersted Oe The cgs unit for the magnetic field strength H 1 oersted 10 ampere meter 79 58 ampere meter ohm Q The SI unit of resistance and of impedance The ohm is the resistance of a conductor such that a constant current of one ampere in it produces a voltage of one volt between its ends open loop A control system in which the system outputs are controlled by system inputs only and no account is taken of actual system output pascal Pa The SI unit of pressure equal to 1 N m Equal to 1 45 x 10 psi 1 0197 x 10 kgs cm 7 5 x 10 torr 4 191 x 10 inches of water or 1 x 107 bar permeability Material parameter which is the ratio of the magnetic induction B to the magnetic field strength H u B H Also see Initial Permeability and Differential Permeability platinum Pt A common temperature sensing material fabricated from pure platinum to make the Lake Shore PT family of resistance temperature sensor elements polynomial fit A mathematical equation used to fit calibration data Polynomials are constructed of finite sums of terms of the form aixi where a is the i fit coefficient and x is some function of the dependent variable positive temperature coefficient PTC Refers to the sign of the temperature sensitivity F
51. answer is noise or more appropriately signal to noise Noise from a variety of sources affects the small signals used to make bridge measurements Lowering excitation makes the signals even smaller but unfortunately doesn t decrease the noise The noise sources described in Paragraph 2 5 are a fact of life Even with careful installation their effects cannot be reduced below a few microvolts That is why the Model 370 does not have voltage excitation below 2 u The goal of most experiments is to choose an excitation that is the best compromise between self heating and signal to noise The monitor output Paragraph 3 6 can be used to view the amplified signal coming into the input of the Model 370 Seeing the amplitude and frequency of noise at this point can help identify 1ts sources as an aid in troubleshooting 2 7 3 Signal to Noise vs Measurement Bandwidth The noise sources described Paragraph 2 5 combine to reduce the resolution of small signal measurements There are some common themes in the way noise interacts with measurements noise affects worsen when signal size decreases when resistance increases and when bandwidth increases Excitation is limited by self heating and cannot always be increased to create larger signals Resistance can seldom be changed because it is the quantity being measured That leaves bandwidth to overcome signal to noise and improve resolution If noise cannot be reduced at its source it has to be averaged out Reducin
52. applications required the complexity of a bridge measurement Some applications benefit so much from the bridge configuration that they may never be replaced One example is strain gauge measurements that integrate both measurement and temperature compensation into the bridge Another application often seen in physics laboratories is low noise resistance measurement where the reference resistors and unknown are all cooled to reduce thermal noise The Model 370 does not use a bridge configuration for resistance measurement Its measurement technique is described in the next paragraph Side by side testing demonstrated that the precision circuits in the Model 370 would not benefit from a bridge configuration unless the reference resistors were cooled This path was not chosen because installation is more complicated than most applications require The name Bridge was retained to reflect the instrument s heritage of precision resistance measurements 2 4 2 Resistance Measurement The Model 370 uses a four lead AC measurement technique to achieve the best possible accuracy with the lowest possible excitation Similar circuits are used extensively in small signal measurement because they are flexible and can extract small signal from relatively noisy environments The Model 370 uses current excitation because controlled current is very stable and predictable Current can also be scaled easily which is necessary to achieve low excitation The Mode
53. arrangement protects low level measurement signals from noise All inactive input leads are shorted to common to minimize noise pickup Special care must be taken during installation to maintain low noise operation Refer to Paragraph 3 5 for scanner installation The Model 370 can read any of the 16 scanner channels one at a time Channel change can be done manually or with the autoscan feature The instrument stores input setup information for each channel and will use the appropriate setup for each channel when selected Refer to Paragraph 4 8 for scanner operation Theory of Operation 2 13 Lake Shore Model 370 AC Resistance Bridge User s Manual 2 8 1 Scanner Performance Up to 16 resistors with up to 64 leads can be attached to the Model 3716 L Scanner Up to 8 resistors with up to 32 leads can be attached to the Model 3708 Scanner The large number and potential length of lead wires provide an excellent path for noise to couple into the measurement system Proper installation combined with the noise reduction features built into both the Model 370 and Model 3716 L and 3708 can minimize the effects but a full loaded scanner may never be as quiet as a single guarded and shielded resistance input A measurable increase in noise is most likely when measuring large resistance of more than 100 kQ Large resistances pose an additional problem because the guards available for a single resistance input end at the scanner enclosure Scanner and lead
54. both sensors and instruments on a regular basis If the pieces are put together properly accuracy of a few millikelvin is possible at low temperatures 2 9 1 5 Packaging Sensor packaging is important for two main reasons First the package creates sensor contact with its environment playing a key role in the overall thermal resistance Second the package protects the resistive element from stress that is normally caused by thermal contraction as the load cools When choosing a pre packaged sensor watch for a specification of thermal resistance or a recommended maximum excitation to make sure it is compatible with the experiment and measuring instruments Also make sure the package is strain free or makes some accommodation to reduce stress on the resistive element during thermal cycling Many sensors are also offered without packaging for applications that require smaller size or lower mass than packaged sensors provide These must be installed with the greatest care and all manufacture recommendations should be followed 2 9 1 6 Environment AII cryogenic temperature sensors have to contend with high vacuum and drastic temperature changes but some experiments create an even more hostile environment by adding high magnetic fields ionizing radiation and more These additional requirements make sensor selection more difficult because they can limit the useful range of many sensors The Lake Shore Temperature Measurement and Control catalog incl
55. coefficient should be 0 997 2 10 Send the 9 8 Hz coefficient to the instrument with the following command CALG 5 1 Coefficient 9 8 Hz 11 Use CALG 5 3 query to read and record the existing 16 7Hz frequency compensation coefficient A safety step to revert to the original calibration if necessary 12 Set the 16 2 Hz frequency compensation coefficient to its default of 1 with the following command CALG 5 3 1 13 Change frequency to 16 2 Hz Paragraph 4 7 1 and allow the reading to stabilize 60 seconds 14 Record the resistance reading 15 Calculate the 16 2 Hz frequency compensation coefficient as Coefficient 16 2 Hz Resistance 13 7 Hz Resistance 16 2 Hz The coefficient should be 1 005 2 16 Send the 16 2 Hz coefficient to the instrument with the following command CALG 5 3 Coefficient 16 2 Hz 17 Send the CALSAVE command and wait 10 seconds for the calibration to be saved in non volatile memory Alternatively this step can be done after all ranges are calibrated 8 12 2 5 Scanner Compensation The Model 370 will operate either with or without the Model 3716 Scanner When a scanner is used its pre amplifier replaces the pre amplifier on the Model 370 A compensation factor must be entered into the Model 370 so it can accurately subtract the gain of its pre amplifier when a scanner is used This calibration is separate from the gain calibration done to the scanner itself When both the instrument and scanne
56. command and query strings through user programs the instrument issues responses Two or more command strings can be chained together in one communication but they must be separated by a semi colon Only one query is permitted per communication but it can be chained to the end of a command The total communication string must not exceed 64 characters in length A command string is issued by the computer and instructs the instrument to perform a function or change a parameter setting The format is lt command mnemonic gt lt space gt lt parameter data gt lt terminators gt Command mnemonics and parameter data necessary for each one is described in Paragraph 6 3 Terminators must be sent with every message string 6 12 Remote Operation Lake Shore Model 370 AC Resistance Bridge User s Manual Message Strings Continued A query string is issued by the computer and instructs the instrument to send a response The query format is lt query mnemonic gt lt gt lt space gt lt parameter data gt lt terminators gt Query mnemonics are often the same as commands with the addition of a question mark Parameter data is often unnecessary when sending queries Query mnemonics and parameter data if necessary is described in Paragraph 6 3 Terminators must be sent with every message string The computer should expect a response very soon after a query is sent A response string is the instruments response or answer to a query string T
57. data Polynomial interpolation has the potential to calculate a result that is well off the curve if data points are not smoothed properly and there is a small discontinuity between breakpoints 2 9 3 Sensor Installation This section highlights some of the important elements of proper temperature sensor installation Information on installing temperature sensors applies equally well to other resistance measurements For more detailed information Lake Shore sensors are shipped with installation instructions covering that specific sensor type and package The Lake Shore Temperature Measurement and Control Catalog includes an installation section as well Lake Shore also offers a line of Cryogenic Accessories Many of the materials discussed here are available through Lake Shore and can be ordered with sensors or instruments 2 9 3 1 Mounting Materials Choosing appropriate mounting materials is very important in a cryogenic environment The high vacuum used to insulate cryostats is one source of problems Materials used in these applications should have a low vapor pressure so they do not evaporate or out gas and spoil the vacuum insulation Metals and ceramics do not have this problem but greases and varnishes must be checked Another source of problems is the wide extremes in temperature most sensors are exposed to The linear expansion coefficient of a materials becomes important when temperature changes are so large Never try to permanently bond ma
58. during closed loop control and must be set to zero if not wanted The Model 370 offers two tuning modes or ways to set the necessary P I and D parameters for closed loop control In closed loop PID mode the control parameter values are set manually as described in Paragraph 2 10 Once appropriate parameters are found they can be stored in the instrument and recalled with the zone feature Zone Tuning Optimal control parameters values are often different at varied temperatures within a system Once values have been chosen for each temperature range or zone the zone feature can automatically select the correct values each time the setpoint is changed This mode does not help choose control parameter values it helps use chosen values more efficiently Refer to Paragraph 5 3 for details on how to program parameter values into the zone table Off When control is turned off the control output is set to zero If a key related to temperature control is pressed the display will show a message indicating that control is off Use the Control Setup key to enable control by choosing one of the three control modes Operation 4 15 Lake Shore Model 370 AC Resistance Bridge User s Manual 4 11 2 Control Setup Several control parameters are grouped under the Control Setup key These are considered setup parameters because they are normally set when configuring the control loop and not used often during operation Control parameters that change of
59. entry keys to enter a manual output value between 0 and full scale The units multiplier is determined by the heater range and can not be changed Press Enter to accept the new selection and return to the normal display Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display NOTE The heater output value shown on the normal display may be truncated to fit in its display space but full setting resolution is retained in the instrument 4 11 6 Heater Output Range and Heater Output Off The temperature control heater output is a low noise variable DC current source It can provide up to 1 watt of continuous power into a resistive heater on the cooled load Lower heater ranges are provided to allow good control resolution at lower temperatures where cooling power diminishes Higher current ranges can be locked out under control setup operation if they can potentially overheat and damage the load Heater range is an integral part of the closed loop control system so re tuning may be required when heater range is changed The heater output range follows the heater output display units of current or power An appropriate heater resistance must be entered under control setup for the power calculation to be meaningful The maximum current or power is used on the display to indicate the selected range Some examples of heater power versus heater resistance are given in Paragr
60. from digital circuitry and the instrument chassis Optical isolation minimizes the effect of digital noise on the measurement and breaks ground loops For applications where lead length is greater than 10 ft or resistance is greater than 100 kQ the Model 370 also includes four separate driven guards that follow the voltage on each lead to provide the lowest possible voltage difference between the lead and its shield Driven guards reduce the effect of cable capacitance and provide the best possible shielding Driven guards are not available for scanned inputs To accommodate conversion of changing output from phase sensitive detection to a stable signal the Model 370 operates with a 200 ms minimum filter time constant While this is adequate for measurement of small resistance values with large excitation the Model 370 software provides additional filtering to ensure good resolution as resistance increases and excitation decreases Linear filtering or averaging offers the best possible settling time with selections from 1 s to 200 s Excitation Modes The Model 370 provides full scale resistance ranges from 2 mQ to 2 MQ The selected resistance range is continuously displayed excitation power dissipated in the resistor is also continuously calculated and displayed The Model 370 includes both a current excitation mode and a voltage mode for resistor excitation Current excitation is the instruments primary mode of operation In current mode the Mod
61. gt 0 Unlocked 1 Locked lt code gt Specifies lock out code 000 999 Remarks Locks out all front panel entries except pressing the Alarm key to silence alarms or selecting to turn the heater off Example LOCK 1 123 term Enables keypad lock and sets the code to 123 LOCK Keypad Lock Query Input LOCK term Returned lt state gt lt code gt term Format n nnn Refer to command for description MDAT Minimum Maximum Data Query Input MDAT lt channel gt term Format nn lt channel gt Specifies which channel to query 1 16 Returned lt min value gt lt max value gt term Format nnn nnnE nn nnn nnnE nn Remarks Returns the minimum and maximum channel data Also see the RDGST command 6 30 Remote Operation MNMX Input Format Example MNMX Input Format Returned Format MNMXRST Input Remarks MODE Input Format Example MODE Input Returned Format MONITOR Input Format Remarks MONITOR Input Returned Format MOUT Input Format Example Lake Shore Model 370 AC Resistance Bridge User s Manual Minimum and Maximum Parameter Command MNMX lt channel gt lt source gt term nn n lt channel gt Specifies channel to configure 1 16 0 all channels lt source gt Specifies channel data to process through max min 1 Kelvin 2 Ohms 3 Linear Data MNMX 8 2 term Channel 8 min max function is on and processes data from the ohm readi
62. gt term Format n lt error code gt Heater error code 0 no error 1 heater open error IEEE IEEE 488 Interface Parameter Command Input IEEE lt terminator gt lt EOI enable gt lt address gt term Format n n nn lt terminator gt Specifies the terminator 0 lt CR gt lt LF gt 1 lt LF gt lt CR gt 2 lt LF gt 3 no terminator must have EOI enabled lt EOI enable gt Sets EOI mode 0 enabled 1 disabled lt address gt Specifies the IEEE address 1 30 Address 0 and 31 are reserved Example IEEE 0 0 4 term After receipt of the current terminator the instrument uses lt CR gt lt LF gt as the new terminator uses EOI mode and responds to address 4 IEEE IEEE 488 Interface Parameter Query Input IEEE term Returned lt terminator gt lt EOI enable gt lt address gt term Format n n nn Refer to command for description INSET Input Channel Parameter Command Input INSET lt channel gt lt off on gt lt dwell gt lt pause gt lt curve number gt lt tempco gt term Format nn n nnn nnn n lt channel gt Specifies which channel to configure 1 16 0 all channels lt off on gt Specifies whether the channel is on or off 0 Off 1 On lt dwell gt Specifies a value for the autoscanning dwell time 1 200 seconds lt pause gt Specifies a value for the change pause time 3 200 seconds lt curve number gt Specifies which curve the channel uses 0 no curve 1 20 user cur
63. identifies the Model 3716 L or 3708 Scanner when it is powered up with the scanner attached It is easy to tell if the instrument has recognized the scanner because the word Scan and the active channel number appear on the top line of the display When the scanner is attached operation related to the resistance measurement input changes slightly This paragraph summarizes scanner operation although most of the operation is addressed in the individual feature paragraphs 4 8 1 Channel On or Off If any of the 16 scanner channels are skipped during channel selection or other selection operations those channels may have been turned off To turn on the channels refer to Paragraph 4 6 6 4 12 Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 4 8 2 Manual Scanner Channel Selection To change the active scanner channel manually use the Scan Channel AW keys The new channel number will appear on the top display line immediately but it can take several seconds for a valid reading to appear on the display The first valid reading will be ready after the hardware settling time change pause time and filter settling The hardware settling time is fixed but the change pause Paragraph 4 6 3 and the filter settling time Paragraph 4 9 can be set by the user If the instrument is used to control temperature and scan channels at the same time the control algorithm can periodically override the scan channel 4 8 3 Automatic Scanni
64. low ohmic and low resistance Hall effect measurement capability by a factor of two over the Model 3716L Like the Model 3716L unused leads are left open to facilitate Hall effect measurement sequencing With DC bias current of 55 pA however it is not recommended for subkelvin temperature measurements These measurements require very low DC bias current to prevent measurement errors as a result of self heating Specifications for resistance range accuracy and resolution noise are different than the standalone Model 370 1 4 Introduction Lake Shore Model 370 AC Resistance Bridge User s Manual 1 2 SENSOR PERFORMANCE Lake Shore Germanium GR 200A 30 Sensor Properties Excitation and Instrumentation Temperature Resistance dR dT Thermal Resistance Excitation Excitation Power K Resistance Range Voltage Limit Current 0 05 25kQ 3 5 MQ K 200 mK nW 63 2 KQ 63 2 UV 1nA 25 fW 0 1 2kQ 60 kQ K 20 mK nW 6 32kQ 63 2 uV 10nA 200 fW 0 3 1720 890 Q K 4 mK nW 632 Q 200 uV 316 nA 17 pW 1 420 36 Q K 0 1 mk nW 200 Q 200 uV 1 A 42 pW Instrument Performance Overall Performance Temperature Measurement Electronic Calibration Self Heating Interpolation Overall K Resolution Accuracy Accuracy Errors Error Accuracy 0 05 6 Q 1 7 uK 13 8 Q 3 9 uK 5 mK 5 uK 0 2 mK 5 2 mK 0 1 300 mQ 5 uk 1 13 Q 19 uK 5 mK 4 uK 0
65. of the controlled quantity closer to a desired value Also known as closed loop control system four lead measurement technique where one pair of excitation leads and an independent pair of measurement leads are used to measure a sensor This method reduces the effect of lead resistance on the measurement gamma A cgs unit of low level flux density where 100 000 gamma equals one oersted or 1 gamma equals 10 oersted gauss G The cgs unit for magnetic flux density B 1 gauss 10 tesla Named for Karl Fredrich Gauss 1777 1855 a German mathematician astronomer and physicist gaussian system units A system in which centimeter gram second units are used for electric and magnetic qualities general purpose interface bus GPIB Another term for the IEEE 488 bus germanium Ge A common temperature sensing material fabricated from doped germanium to make the Lake Shore GR family of resistance temperature sensor elements gilbert Gb A cgs electromagnetic unit of the magnetomotive force required to produce one maxwell of magnetic flux in a magnetic circuit of unit reluctance One gilbert is equal to 10 4m ampere turn Named for William Gilbert 1540 1603 an English physicist hypothesized that the Earth is a magnet gilbert per centimeter Practical cgs unit of magnet intensity Gilberts per cm are the same as oersteds Greek alphabet The Greek alphabet is defined as follows Alpha a A Iota 1 I Rho p P Beta B B Kappa K K Sigma 6
66. off time indicates firmware settling including pause and filter settling times During these first two phases autorange Max Min alarm analog output and temperature control functions are frozen so they are not affected by transitional data The display update indicator blinks on 90 and off 10 when the instrument is reporting stable readings and all features are active If a Model 3716 Scanner is attached to the Model 370 the channel indicators take the place of the display update indicator If the active scan channel is chosen for one of the display locations the channel number blinks while the channel is being read The blink pattern indicates both range and channel changes 4 2 3 Range Excitation and Power The lower right corner of the display provides feedback from the active resistance measurement and is continually updated for both autorange and scanner operation The present resistance range is shown first followed by the chosen excitation Excitation is displayed in volts if voltage mode is active and amps if current mode is active Excitation power is shown on the bottom display line Power is continuously calculated using the instruments nominal current source excitation and the resistance measurement after each update Power from stray noise sources is not included in the calculation 4 2 4 Display and LED Annunciators The top line of the display is reserved for feature annunciators that are shown as necessary to indicate the status of
67. option is Model 8001 370 and should be ordered with a calibrated sensor A CalCurve can be done in the field when additional or replacement sensors are installed Curve data 1s loaded into some type of non volatile memory that is installed into the instrument by the user In the case of the Model 370 the curve is loaded into a non volatile memory which can be installed into the instrument The field installed version is a Model 8002 05 370 and should be ordered with a calibrated sensor Customers that have a PC compatible computer with an RS 232C or IEEE 488 interface have another option The Model 8000 is included with the calibrated sensor and can be loaded by the user Lake Shore offers a free curve loading program that is compatible with Model 8000 CalCurves 5 2 FRONT PANEL CURVE ENTRY OPERATIONS There are three operations associated with front panel curve entry Edit curve Copy curve and Erase curve as detailed below Similar operations are available over computer interface Edit Curve Edit allows the user to enter or edit a curve at any user curve location Refer to Paragraph 5 2 1 Erase Curve Erase allows the user to delete a curve from any user curve location Refer to Paragraph 5 2 2 Copy allows the user to copy a curve from any location to any other a locaton Refer to Paragraph 5 2 3 Copy Curve 5 2 1 Edit Curve The Edit Curve operation is used to enter a new curve or edit an existing user curve A
68. or epoxy used to hold the sensor act as both gasket and adhesive Permanent mounting is not a good solution for everyone because it limits flexibility and can potentially damage sensors Much care should be taken not to over heat or mechanically stress sensor packages Less permanent mountings require some pressure to hold the sensor to its mounting surface Pressure greatly improves the thermal resistance of joints and reduces thermal gradients A spring clamp is recommended so that different rates of thermal expansion do not increase or decrease pressure with temperature change 2 9 3 6 Lead Wire Different types of sensors come with different types and lengths of electrical leads In general a significant length of lead wire must be added to the sensor for proper heat sinking and connecting to a bulk head connector at the vacuum boundary The lead wire must be an electrical conductor but should not be a good thermal conductor or heat will transfer down the leads and change the temperature reading of the sensor Small 30 to 40 AWG wire made of an alloy like Manganin or phosphor bronze is much better than copper wire Super conducting wire is a good choice if wire temperature remains below the transition temperature Thin wire insulation is preferred for better heat sinking Twisted wire is recommended if electrical noise is present The wire used on the room temperature side of the vacuum boundary is not critical so shielded copper cable is normally used
69. or loop when that loop contains changing current or magnetic field In measurements this problem is commonly referred to as a ground loop and generally results when improperly grounded instruments are exposed to AC line current Isolation reduces ground loop noise by breaking the loop or circuit The Model 370 has a well designed power supply and uses optical isolation between the instrument chassis and measurement circuits Careful installation is necessary to take advantage of the isolation Ground loop noise can be kept as low as a few microvolts and usually appears as line frequency or one of its low harmonics 2 5 3 Analog Circuit Noise The current source and amplifiers inside the Model 370 make a contribution to measurement noise Circuits were carefully chosen to have the maintain good noise performance over a wide resistance range without generating high leakage currents that could self heat resistors Leakage current limits the practical current excitation range of the Model 370 to a few picoamps Most analog circuit noise is distributed over a broad frequency band It is often referred to as 1 f noise because there is equal noise power in each decade of frequency Instrument noise specified in Table 1 1 is dominated by amplifier noise for low resistances and by resistor thermal noise for high resistances Noise from both sources can be further reduced with longer filter settling time 2 5 4 Digital Circuit Noise All digitally controlled ins
70. power supply of almost any size A resistor must be placed across the heater output to convert the output current into a programming voltage An appropriate resistor value can be calculated as R V I where R is the resistor value in ohms V is the full scale programming input voltage of the power supply and I is the full scale current in amps of the control heater output range For example if the supply requires a 0 to 10 V programming input a 100 Q resistor is required for the 100 mA 0 1 A heater output range 100 Q 10 V 0 1 A The resistor must be rated for the power dissipated in it P VI which would be I W Lower heater output ranges can be used to lower the power dissipated in the programming resistor The maximum output current of 100 mA and the maximum compliance voltage of 10 V cannot be exceeded in this configuration The analog voltage outputs can be used to directly control a programmable DC power supply in open loop control mode but the closed loop algorithm is not available for those outputs 3 8 ANALOG OUTPUT AND STILL HEATER OUTPUT The Model 370 has two analog voltage outputs They have the same hardware configuration but Analog Output 2 has additional firmware features that make it more useful as a still heater output for a dilution refrigerator Both signal outputs are referenced to their common that is internally tied to chassis ground ANALOG ANALOG OUT 2 STILL OUT 1 Analog Outputs bmp Figure 3 15 Analog Output Conne
71. provided in case a cable must be made for a special application Both instrument and scanner have the same pin numbering but the instrument has a connector receptacle and scanner a connector plug Scanner Control Connector Definition SCAN CONTROL 4 Scan Control bmp Pin Name Description 1 Scan Present Signal to Model 370 that a scanner is installed 2 Digital Common Common or ground for digital control signals 3 Digital Common 4 Unused 5 Analog Common Common or ground for analog signals 6 Analog Common 7 Analog Common 8 15 V 15 V referenced to scanner analog common 9 Scan 1 Scan channel select line 1 10 Scan 2 Scan channel select line 2 11 Scan 3 Scan channel select line 3 12 Scan 4 Scan channel select line 4 13 Scan Gain Selects scanner gain of 1 or 100 14 5 V 5 V referenced to scanner digital common 15 15 V 15 V referenced to scanner analog common Body Chassis Do Not Use NOTES 1 Both scanner analog and scanner digital common are tied to isolated measurement circuits inside the instrument and must remain isolated from the instrument chassis and Earth ground 2 The connector body should not be used for shielding the scan control cable because the chassis of the instrument is at Earth ground and the chassis of the scanner is at isolated measurement common Figure 3 10 Scanner SCAN CONTROL Connector 3 5 7 Scanner Control Cabling A scanner control cable is included wit
72. ps 2 15 2 9 2 Sensor Calibration s szy ZZ R PE OE EE EEK BEA EE d slate teeta Je e 2 16 2 9 3 Sensor Installation cid BE AE ET EE tdi 2 16 2 9 3 1 Mounting Materials marianne OO AGO ee E RA E 2 16 2 9 3 2 DENSON LOCATON iii Ai ea 2 16 2 9 3 3 Thermal Cond ctiVity ereis eerta iaag O Z Pa a ata paee kanaat aa STAT Eins 2 16 2 9 3 4 CORtACEATEA enen aerat aaa Ko z a a ea aa aeaea atp Paea brete 2 17 2 9 3 5 Go tactePreSSUre TEE AE Z da TEE E T GENUA 2 17 2 9 3 6 Lead Wi A iio 2 17 2 9 3 7 Lead Soldering iii A Ada 2 17 2 9 3 8 Heat Sinking Leads iii iaa titi 2 17 2 9 3 9 Thermal Radiation aiii 2 17 2 10 TEMPERATURE CONTROL 00 cccescceeececeeeeeeeeeceaeeeeaeecaaeeeeeeecaaesseeeecaaeseeeesaaeseeeeesaaeseeeeesaaeeeeneeaees 2 17 2 10 1 Considerations For Good Control eeeeeee see eee nono aaa aaa nn rra nara rra 2 18 2 10 1 1 Thermal Gonductivity ass az A AO OO A O AYO 2 18 2 10 1 2 Thermal Lag iia 2 18 2 10 1 3 Two Sensor Approach ue zessi swojo ian owego are ie tec 2 18 2 10 1 4 Thermal MaSS cirio 2 18 2 10 1 5 System Nonlinearity cooooocionerrr rieron LEWA EEE 2 18 2 10 2 Closed Loop PID Controlan Adi 2 19 2 10 2 1 Pr poriional Puccini edad 2 19 2 10 2 2 Integral I zza Ot WE e e O tent ese Soad AOS 2 19 2 10 2 3 Derivative Diario AA A vada 2 19 2 10 2 4 Manual OUfIpUt e aoi ce cate EE A a 2 19 2 10 3 Tuning a Closed Loop PID Controller e eee eee aaa aaa aaa aaa aaa aaa aaa
73. ranges Note This value is not filtered CALSAVE Issue the CALSAVE command to save any new calibration coefficients in non volatile memory It is always recommended to read out old calibration coefficients using the CALZ and CALG interface queries before attempting to calibrate This will give the operator experience with the interface command data formatting and typical values If the old values are saved they can be reloaded in the case of accidental loss of data during calibration New calculated calibration coefficients should be very similar to the old values Discrepancy between the old and new values of more than 0 1 of gain calibration coefficients or 0 1 of range for zero coefficients could indicate an error in the calibration procedure or a hardware failure Do not attempt to recalibrate a damaged instrument The instrument will use the new calibration coefficients as soon as they are sent with the either the CALZ or CALG interface command but they are not saved permanently until the CALSAVE command is issued If a mistake is made in the calibration process turn the instrument power off and on again before CALSAVE is issued to restore the old calibration constants Once CALSAVE is issued old values cannot be retrieved from the instrument If calibration coefficients are left at default or are outside of the normal calibration range the following error message will appear in the instrument display when the instrument is turned on
74. runs outside of the experimental Dewar Cables with signals sharing the same ground or common should be routed together to reduce ground loop area Cables with signals that differ in ground or signal size should be routed separately to reduce induced noise Cables should be kept away from noise generators like fluorescent lights and vacuum pump motors Cables should be supported to minimize stress on connectors and terminations Cable support should also limit movement and vibration of the cables during measurements because cable motion can generate electrical noise Model 370 370 Case to Earth Ground through the Power Cord Shield Tied To Scanner Connector Long Cables Short Cables Shielded Shielded and Guarded Only Scanner Case Electrically Connected To Dewar Case NOTE Resistor leads from non selected channels are shorted to measurement common to control noise in the Model 3716 scanner In the Model 3716L and 3708 scanners these are left open Please take this into account when installing Dewar wiring Dewar Connected To Earth Ground Instrument Grounding eps Figure 3 6 Example of Instrument Grounding 3 6 Installation Lake Shore Model 370 AC Resistance Bridge User s Manual 3 4 6 Temperature Sensors Temperature sensors should be treated as any other resistor being measured One difference is that four lead devices are normally calibrated with a specific arrangement of current and volta
75. s Manual 4 6 1 Channel Selection To begin input setup press the Input Setup key If a scanner is installed the first input setup screen appears as a prompt for the channel number All of the parameter changes that follow will act on the selected channel If no scanner is installed the sequence will begin below with change pause Use the A or V key to select a channel number between 1 and 16 Choose the selection All and the changes will be applied to all channels at the same time Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display 4 6 2 Scanner Dwell Time The scanner dwell time parameter allows the user to select how long the measurement will dwell on any channel when the Autoscan feature is on Each channel can have a different dwell time from 1 to 200 seconds Scan dwell time begins after both the change pause time and firmware filter is settled Dwell time will be interrupted if temperature control and Autoscan are active at the same time The next input setup screen appears as a prompt for the scanner dwell time Use the data entry keys to enter a dwell time between 1 and 200 seconds Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal disp
76. shoes are recommended C5 0 RECOMMENDED FIRST AID Every site that stores and uses LHe and LN should have an appropriate Material Safety Data Sheet MSDS present The MSDS may be obtained from the manufacturer distributor The MSDS will specify the symptoms of overexposure and the first aid to be used A typical summary of these instructions is provided as follows If symptoms of asphyxia such as headache drowsiness dizziness excitation excess salivation vomiting or unconsciousness are observed remove the victim to fresh air If breathing is difficult give oxygen If breathing has stopped give artificial respiration Call a physician immediately If exposure to cryogenic liquids or cold gases occurs restore tissue to normal body temperature 98 6 F as rapidly as possible then protect the injured tissue from further damage and infection Call a physician immediately Rapid warming of the affected parts is best achieved by bathing it in warm water The water temperature should not exceed 105 F 40 C and under no circumstances should the frozen part be rubbed either before or after rewarming If the eyes are involved flush them thoroughly with warm water for at least 15 minutes In case of massive exposure remove clothing while showering with warm water The patient should not drink alcohol or smoke Keep warm and rest Call a physician immediately C 2 Handling LHe and LN
77. signals 8 7 2 2 Noise Signals In actual installation the Monitor output can be used to determine the character of offending noise Observed signals will include the measurement signal 13 7 Hz sine wave and any noise signals combined Front panel selection refer to Paragraph 4 7 4 allows monitoring of different points along the analog circuit path The following describes noise characteristics backing up through the measurement circuit from the most processed signal to the least processed signal The VAD signal is the output of the phase sensitive detector circuit and is a DC representation range 2 50VDC of the measurement It is the signal that is directly converted by the Analog to Digital converter for digital processing and display This point is conditioned by a 4 pole low pass filter TC 0 5 sec so the fundamental 13 7 Hz and harmonics along with any 50 60 Hz related signals and other noise signals are greatly attenuated Any DC offset at this point is subtracted in the digital domain This means that a small DC voltage offset several mV or so may be present at this monitoring point A DC voltage exceeding 2 50VDC triggers a R OVER message Bear in mind that this signal is capable of going negative directly indicating a reversal of current or voltage leads to the measured resistor A negative resistance display will occur in this condition A DC voltage exceeding 2 50VDC in the negative direction triggers a R
78. speed two lead Measurement technique where one pair of leads is used for both excitation and measurement of a sensor This method will not reduce the effect of lead resistance on the measurement Underwriters Laboratories UL An independent laboratory that establishes standards for commercial and industrial products unit magnetic pole A pole with a strength such that when it is placed 1 cm away from a like pole the force between the two is 1 dyne volt V The difference of electric potential between two points of a conductor carrying a constant current of one ampere when the power dissipated between these points is equal to one watt volt ampere VA The SI unit of apparent power The volt ampere is the apparent power at the points of entry of a single phase two wire system when the product of the RMS value in amperes of the current by the RMS value in volts of the voltage is equal to one watt W The SI unit of power The watt is the power required to do work at the rate of 1 joule per second References 1 Sybil P Parker Editor McGraw Hill Dictionary of Scientific and Technical Terms Fifth Edition New York McGraw Hill 1994 IBSN 0 07 113584 7 2 Christopher J Booth Editor The New IEEE Standard Dictionary of Electrical and Electronic Terms IEEE Std 100 1992 Fifth Edition New York Institute of Electrical and Electronics Engineers 1993 IBSN 1 55937 240 0 3 Nelson Robert A Guide For Metric Practice Page B
79. still heater load nominal 100 Q by way of one of the instrument s analog outputs to enhance temperature control The best control stability is achieved using only one sensor but the Model 370 can control temperature based on one of multiple scanned sensors Because the Model 370 alternates control with scanned sensor readings and the alternating update rate is slower than operation based on a single sensor control stability may degrade in some systems when multiple sensors are in use Computer Interfaces The Model 370 includes IEEE 488 2 parallel and RS 232C serial interfaces Both use the instrument chassis as ground while measurement input is optically isolated from the chassis to minimize interface noise and ground loops Both interfaces can accommodate data transmission at the maximum reading rate of the Model 370 for automated data recording All instrument parameters all available status information and almost every instrument function including analog voltage outputs and relays can be accessed by computer interface Analog Outputs With two analog voltage outputs and two relays the Model 370 can perform functions that might otherwise require additional hardware and system complexity Configured for use as resistance monitors the analog voltage outputs provide a voltage proportional to measured resistance that can be used to monitor resistance error AR temperature or temperature error Analog outputs can be controlled manually from t
80. that defines the temperature response of a temperature sensor It is used to convert the signal from the sensor to temperature decibels dB A unit for describing the ratio of two powers or intensities or the ratio of a power to a reference power equal to one tenth bel if P and P are two amounts of power the first is said to be n decibels greater where n 10 logo Py P degree An incremental value in the temperature scale i e there are 100 degrees between the ice point and the boiling point of water in the Celsius scale and 180 degrees between the same two points in the Fahrenheit scale demagnetization when a sample is exposed to an applied field H poles are induced on the surface of the sample Some of the returned flux from these poles is inside of the sample This returned flux tends to decrease the net magnetic field strength internal to the sample yielding a true internal field Hjn given by Hin Ha DM where M is the volume magnetization and D is the demagnetization factor D is dependent on the sample geometry and orientation with respect to the field A 2 Glossary of Terminology Lake Shore Model 370 AC Resistance Bridge User s Manual deviation The difference between the actual value of a controlled variable and the desired value corresponding to the setpoint Dewar flask A vessel having double walls the space between being evacuated to prevent the transfer of heat and the surfaces facing the vacuum being h
81. time and can eliminate control offset or droop Derivative rate acts on the rate of change in error to dampen the system reducing overshoot quench A condition where the superconducting magnet goes normal i e becomes non superconductive When this happens the magnet becomes resistive heat is generated liquid cryogen is boiled off and the magnet power supply is shut down due to the sudden increase in current demand rack mount An instrument is rack mountable when it has permanent or detachable brackets that allow it to be securely mounted in an instrument rack The standard rack mount is 19 inches wide A full rack instrument requires the entire width of the rack Two half rack instruments fit horizontally in one rack width relief valve A type of pressure relief device which is designed to relieve excessive pressure and to reclose and reseal to prevent further flow of gas from the cylinder after reseating pressure has been achieved remanence The remaining magnetic induction in a magnetic material when the material is first saturated and then the applied field is reduced to zero The remanence would be the upper limit to values for the remanent induction Note that no strict convention exists for the use of remanent induction and remanence and in some contexts the two terms may be used interchangeably remanent induction The remaining magnetic induction in a magnetic material after an applied field is reduced to zero Also see remane
82. to 300 mK Even at such relatively warm temperatures the benefits of a well designed resistance bridge must be considered when selecting instrumentation Figure 2 2 illustrates the basic features Helium Dewar 4 z He Reservoir Vacuum Space 1 K Pot He Evaporator Pumped reservoir of He maintained at 1 K Used in the process of condensing He gas returning from pumping system Also used as a heat sink for measurement leads Good location for a temperature sensor for cool down monitoring and troubleshooting Heat Exchangers He Evaporator Vacuum insulated Dewar large enough to contain the following components Sometimes includes a nitrogen outer jacket to improve efficiency Reservoir of liquid He at atmospheric pressure that provides bulk cooling down to 4 2 K Pre cools gasses entering the vacuum space and acts as a radiation shield for refrigeration components Insulates refrigeration components and provides a clear workspace One or more heat exchangers serve to cool returning He liquid to near the temperature of the He evaporator Also used Lake Shore Model 370 AC Resistance Bridge User s Manual to heat sink measurement leads Helium Dewar Vacuum Insulation 4He Reservoir Vacuum Space 1K Pot Radiation Shield Heat Exchangers 3He Evaporator Sample Holder Vacuum Chamber Cryostat_Block eps Figure 2 2 HE Cryostat Block Diagram Chamber where co
83. to reach the setpoint It can also see the error decreasing rapidly when the temperature nears the setpoint and reduce the output for less overshoot The derivative term can be useful in fast changing systems but it is often turned off during steady state control because it reacts too strongly to small disturbances or noise The derivative setting D is related to the dominant time constant of the load similar to the integral I and is also set in seconds but a smaller setting creates a less active derivative 2 10 2 4 Manual Output The Model 370 has a control parameter that is not a normal part of a PID control loop Manual output can be used for open loop control meaning feedback is ignored and the heater output stays at the users manual setting This is a good way to put constant heating power into a load when needed The manual output term can also be added to the PID output Some users prefer to set an output value near that necessary to control at a setpoint and let the closed loop make up the small difference NOTE Manual Output should be set to 0 when not in use Theory of Operation 2 19 Lake Shore Model 370 AC Resistance Bridge User s Manual 5 E change in setpoint 5 actual temperature response P Only too high a time P Only b P Only too low c P l d P 1 D e PID Control bmp Figure 2 4 Examples of PID Control 2 20 Theory of Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 2 1
84. wave rather than the sloped rolloff of a first order low pass This can create problems when trying to filter periodic noise that has a frequency lower than the excitation frequency The input hardware settles about 1 second after a large resistance change if no range change is required When the firmware filter is turned on that filter settling time must be added to the hardware settling time to approximate overall settling Firmware filter settling time settings are available 1 second to 200 seconds Please remember that filtering does not reduce the electrical noise that reaches the resistor and therefore will not reduce self heating from noise Range change requires additional settling because the internal circuits need time to reestablish their operating point The hardware settles in 2 to 3 seconds after range change Settling may take a little longer especially if resistance is very large or excitation is very low A change pause can be programmed into the instrument to allow for settling without disturbing other instrument firmware features During the pause period features including filter max min capture analog voltage output calculations and temperature control and error testing are suspended The pause time can be set from 3 to 60 seconds and must be added to the filter settling time to approximate overall settling after range change Scan channel change is similar to range change except the new channel may autorange after a channel change cre
85. 0 1 5 System Nonlinearity Because of nonlinearities in the control system a system controlling well at one temperature may not control well at another temperature While nonlinearities exist in all temperature control systems they are most evident at cryogenic temperatures When the operating temperature changes the behavior of the control loop the controller must be retuned As an example a thermal mass acts differently at different temperatures The specific heat of the load material is a major factor in thermal mass and the specific heat of materials like copper change as much as three orders of magnitude when cooled from 100 K to 10 K Changes in cooling power and sensor sensitivity are also sources of nonlinearity The cooling power of most cooling sources also changes with load temperature This is very important when operating at temperatures near the highest or lowest temperature that a system can reach Nonlinearities very close to these high and low temperatures make it very difficult to configure them for stable control If difficulty is encountered it is recommended to gain experience with the system at temperatures further away from the limit and gradually approach it in small steps Keep an eye on temperature sensor sensitivity Sensitivity not only affects control stability but it also contributes to the overall control system gain The large changes in sensitivity that make some sensors so useful may make it necessary to retune the con
86. 0 1 Curve Selection After a temperature response curve is loaded into the instrument it must be assigned to an input channel even if only one channel is active Curve selection is done under the Input Setup key as described in Paragraph 4 6 4 Ifa curve is not properly chosen for a channel any temperature displays for that channel will show the message NOCURV 4 10 2 Temperature Display After a curve is assigned to an input channel the channel resistance temperature or both can be displayed or queried over computer interface Temperature is displayed when Kelvin units are chosen for the display data source under the Display Setup key as described in Paragraph 4 4 1 Refer to Paragraph 4 12 for instructions on displaying Max and Min in temperature refer to Paragraph 4 11 for instructions on displaying a control setpoint in temperature 4 11 TEMPERATURE CONTROL There are many steps involved in configuring a temperature control loop Chapter 2 of this manual describes the principals of closed loop feedback control Chapter 3 describes the necessary hardware installation This paragraph describes how to operate control features and set control parameters Temperature measurement should be operating fully before proceeding to control The action of each control parameter should be considered before enabling temperature control or the instrument may not be able to perform even the most simple control functions The more complex features of se
87. 0 3 Tuning a Closed Loop PID Controller There has been a lot written about tuning closed loop control systems and specifically PID control loops This section does not attempt to compete with control theory experts It describes a few basic rules of thumb to help less experienced users get started This technique will not solve every problem but it has worked for many others in the field This section assumes the user has worked through the operation sections of this manual and has a stable temperature reading from the control sensor It is also a good idea to begin at the center of the temperature range of the cooling system not close to its highest or lowest temperature 2 10 3 1 Setting Heater Range Setting an appropriate heater output range is an important first part of the tuning process The heater range should allow enough heater power to comfortably overcome the cooling power of the cooling system If the heater range will not provide enough power the load will not be able to reach the setpoint temperature If the range is set too high the load may have very large temperature changes that take a long time to settle out Delicate loads can even be damaged by too much power Often there is little information on the cooling power of the cooling system at the desired setpoint If this is the case try the following Allow the load to cool completely with the heater off Set manual heater output to 50 while in Open Loop control mode Turn the h
88. 0 times per second Failure to follow these simple rules will result in inability to establish communication with the instrument or intermittent failures in communication 6 2 7 Serial Interface Example Programs A Visual Basic program is included to illustrate the serial communication functions of the instrument Refer to Section 6 2 7 1 for instructions on how to setup the program The Visual Basic code is provided in Table 6 8 A description of program operation is provided in Section 6 2 7 2 While the hardware and software required to produce and implement these programs not included with the instrument the concepts illustrated apply to most applications Remote Operation 6 13 6 2 7 1 Lake Shore Model 370 AC Resistance Bridge User s Manual Visual Basic Serial Interface Program Setup The serial interface program works with Visual Basic 6 0 VB6 on an IBM PC or compatible with a Pentrum class processor A Pentium 90 or higher is recommended running Windows 95 or better with a serial interface It uses the COM1 communications port at 9600 Baud Use the following procedure to develop the Serial Interface Program in Visual Basic v AO 47 Start VB6 Choose Standard EXE and select Open Resize form window to desired size On the Project Menu click Components to bring up a list of additional controls available in VB6 Scroll through the controls and select Microsoft Comm Control 6 0 Select OK In the toolbar at the left
89. 00 for maximum power Heater Output Gain Accuracy 1 of setting Heater Output Offset at 0 0 02 of range Max Heater Noise Current lt 0 005 of range PID Control Parameters Proportional 0 001 1000 Integral 0 10 000 seconds Derivative 0 10 000 seconds Manual Output 0 100 resolution 0 001 PID Zone Settings 10 zones that include setpoint heater range PID relays and analog outputs Still Still Output Up to 1 W of power available using analog output 2 Heater Protection Short circuit protected disabled with a relay on power up defaults to off range on power up selectable heater range limit open circuit detection Scanner Support Control with scanned input is supported with reduced stability Front Panel Display Type Graphic 8 line by 40 character vacuum fluorescent display Number of Reading Displays 1 to 8 Reading Display Units mQ Q kQ MQ mk K Display Resolution 4 5 or 6 digit user selected Display Update Rate 2 readings per second Reading Display Options Ohms Kelvin Max Min Linear Other Displays Channel number units resistance range excitation voltage or current excitation power control setpoint heater range and heater output Display Annunciators Reading errors CMR Alarm Ramp Zone Remote LED Annunciators Autorange Excitation Mode Autoscan Keypad 36 key numeric and specific functions Interface IEEE 488 2 Interface Capability SH1 AH1 T5 L4
90. 2 mK 5 2 mK 0 3 10 mQ 11 uK 64 2 MQ 72 uk 5 mK 68 uK 0 2 mK 5 3 mK 1 3 mQ 83 uK 16 6 MQ 461 uK 5 mK 5 HK 0 2 mK 5 7 mK Lake Shore 1000 Q Ruthenium Oxide Sensor Properties Excitation and Instrumentation Temperature Resistance dR dT Thermal Resistance Excitation Excitation Power K Resistance Range Voltage Limit Current 0 05 70 kQ 5 0 MQ K 7000 mK nW 200 kQ 63 2 uV 316 pA TW 0 1 193kQ 266 kQ K 800 mK nW 20 KQ 63 2 uV 3 16 nA 193 fW 0 3 5 6kQ 16 6 kQ K 50 mK nW 6 32 KQ 200 uV 31 6 nA 5 6 fW 1 23kQ 1 2kQ K 8 mK nW 6 32 KQ 200 uV 31 6 nA 2 3 pW Instrument Performance Overall Performance Temperature Measurement Electronic Calibration Self Heating Interpolation Overall K Resolution Accuracy Accuracy Errors Error Accuracy 0 05 40 Q 8 uK 35 Q 7 UK 5 mK 49 uK 0 2 mK 5 2 mK 0 1 1 Q 3 8 uK 9 7 Q 36 uK 5 mK 155 uK 0 2 mK 5 2 mK 0 3 0 1 Q 6 uK 2 8 Q 170 uK 5 mK 280 uK 0 2 mK 5 4 mK 1 0 1 Q 83 uK 0 7 Q 580 uK 5 mK 18 uK 0 2 mK 5 8 mK NOTES 1 Recommended operating range of GR 200A 30 is from 50 mK to 1 K but it can be used beyond this range 2 Excitation chosen to minimize sensor self heating 3 Typical thermal resistance with minimal heat sinking can be improved with permanent installation 4 Typical sensor characteristics individual sensors vary in resistance and sensitivity Excitation Power Actual Current x Measured Resistance Resolution Temperature Resolution Resistance dR dT Elec
91. 200 Q 63 2 Q 200 316pA 2mQ 60010 20040 130uQ 4010 100pW 324W 10uW 324W 1uWw 6 32 kQ 2 0 KQ 632 Q 200 Q 63 2 Q 100 pA 6mQ 2m0 60040 400uQ 13010 32uW 10uW 32pW 1UW 320nW 20kQ 6 32 kQ 20kQ 632 Q 200 Q 31 6 HA 20 mQ 6mQ 2mQ 1 3mQ 400 uQ 10uW 32uW 1pW 320nW 100nW 63 2 kQ 20kQ 6 32kQ 2 0 ka 632 Q 10 pA 60 mQ 20 MQ 6mQ 4mQ 13mQ 3 2 UW 1 uw 320 nW 100 nw 32 nW 200 kQ 63 2 kQ 20 KQ 6 32 KQ 2 0 ka 3 16HA 200mQ 60 mQ 20 mQ 13 mQ 4mQ 1 uw 320 nW 100 nW 32 nW 10 nW 632 KQ 200 kQ 63 2 kQ 20 KQ 6 32 KQ 1 0 HA 600mQ 200 mQ 60 mQ 40 mQ 13 mQ 320 nW 100 nW 32 nW 10 nW 3 2 nW 2 0 MQ 632 KQ 200 kQ 63 2 KQ 20 ka 316 nA 600mQ 200mQ 130 mQ 40 mQ 32 nW 10 nW 3 2 nW 1nW 2 0 MQ 632 ka 200 ka 63 2 KQ 100 nA 600mQ 400 MQ 130 mQ 32 nW 3 2 nW 1 0 nW 320 pW 20 MQ 2 0 MQ 632 kQ 200 kQ 31 6 nA i 130 400 MQ 10nW 3 2 nW 320 pW 100 pW 63 2 MQ 20 MQ 2 0 MQ 632 kQ 10nA Z sA ds 200 3 2 nW 1nW 320 pW 32 pw 5 63 2 MQ 20 MQ 2 0 MQ 3 16 nA ak ak kk 4 320pW 100pW 32pW gt 63 2MQ 20MQ 4 0nA xk sk 2 3 32 pW 10 pW E E b 63 2 MQ 316 pA sk 3 2 pW 5 200 kQ 63 2 kQ 20 kQ 100 pA 7 i al ij hd 400 Q 200 Q 100 Q 60 Q 60 Q dl fi 320 fW 100 fW 32 fW 3 2 fW 1 0 W 320 aW 100 aw i 4 s il 63 2 MQ 20 MQ 2 0 MQ 632 kQ 200 kQ 63 2 ka 31 6 pA K w dd 10kQ 600 ka 300 Q 200 mQ gt i u 32 fW 10 W 3 2 W 1 0 W 320 aW 100 aw 32 aW 2 E s i E 63 2 MQ 20 MQ 2 0 MQ 632 KQ 200 ka 10pA El gt gt E ls li 3 0
92. 3 O Weighting ee Bit Name Bits 1 3 and 6 are not used The bus controller will only be interrupted with the reports of this register if the bits have been enabled in the Standard Event Status Enable Register and if bit 5 of the Service Request Enable Register has been set The Standard Event Status Enable Register allows the user to enable any of the Standard Event Status Register reports The Standard Event Status Enable command ESE sets the Standard Event Status Enable Register bits If a bit of this register is set then that function is enabled To set a bit send the command ESE with the bit weighting for each bit you want to be set added together See the ESE command discussion for further details 6 4 Remote Operation Lake Shore Model 370 AC Resistance Bridge User s Manual Standard Event Status Register and Standard Event Status Enable Register Continued The Standard Event Status Enable Query ESE reads the Standard Event Status Enable Register kESR reads the Standard Event Status Register Once this register has been read all of the bits are reset to zero Power On PON Bit 7 This bit is set to indicate an instrument off on transition Command Error CME Bit 5 This bit is set if a command error has been detected since the last reading This means that the instrument could not interpret the command due to a syntax error an unrecognized header unrecognized terminators or an unsupported c
93. 5 mA 15 15 V 15 V 25 mA 9 Scan 1 TTL Digital Output 1 3 5 9 Scanner Mounting Figure 3 11 SCAN CONTROL Signals Without A Scanner Bracket A scanner mounting bracket is included with the Model 3716 L and 3708 Scanners so that it can be mounted on or very near the test Dewar Rigidly mounting the scanner will prevent unnecessary cable movement and help maintain a uniform shield connection over time The hole pattern for the bracket is shown below but the bracket is made of soft aluminum and additional holes can be added easily No mounting screws are included ECZ1 TEZ t a a a 2 00 a a a 2 00 H 2 00 2 00 47 a a a 1 19 i ZI I I m 3 07 0 25 0 25 0 25 RADIUS RADIUS HOLES a 0 196 DIA THRU HOLE 9 PLACES b 0 156 DIA THRU HOLE C SINK 82 TO 0 28 DIA 4 PLACES Mounting Bracket bmp Figure 3 12 Scanner Mounting Bracket Footprint Installation 3 11 Lake Shore Model 370 AC Resistance Bridge User s Manual 3 6 MONITOR AND REFERENCE OUTPUTS The monitor and reference voltage outputs are included to aid in troubleshooting especially identification of ground loops and other noise sources during instrument installation Both signal outputs and their common connections are part of the isolated measurement circuitry and should be treated like part of the resistance measurement They should only be us
94. 67 225 67 143 15 387 67 220 140 382 219 67 139 82 380 210 134 44 379 67 209 67 134 26 370 207 67 133 15 369 67 202 130 364 200 128 89 360 199 67 128 71 359 67 190 123 33 351 67 189 67 123 15 350 184 120 349 67 180 117 78 346 179 67 117 59 340 171 67 113 15 339 67 170 112 22 333 67 169 67 112 04 330 166 110 329 67 160 106 67 328 159 67 106 48 320 153 67 103 15 319 67 150 101 11 315 67 149 67 100 93 310 148 100 309 67 140 95 96 300 139 67 95 37 299 67 135 67 93 15 297 67 130 90 B 2 Temperature Scales Lake Shore Model 370 AC Resistance Bridge User s Manual APPENDIX C HANDLING LIQUID HELIUM AND NITROGEN C1 0 GENERAL Use of liquid helium LHe and liquid nitrogen LN gt is often associated with the Model 370 AC Resistance Bridge Although not explosive there are a number of safety considerations to keep in mind in the handling of LHe and LN C2 0 PROPERTIES LHe and LN are colorless odorless and tasteless gases Gaseous nitrogen makes up about 78 percent of the Earth s atmosphere while helium comprises only about 5 ppm Most helium is recovered from natural gas deposits Once collected and isolated the gases will liquefy when properly cooled A quick comparison between LHe and LN is provided in Table C 1 Table C 1 Comparison of Liquid Helium and Liquid Nitrogen PROPERTY LIQUID HELIUM LIQUID NITROGE
95. ANGE AND AUTORANGE The Model 370 uses current source excitation to measure resistance and it can be operated in this mode The instrument firmware is also able to simulate voltage excitation of the measured resistor to create a second operating mode Resistance range selection and autorange behave differently when a different excitation mode is selected These parameter selections can be different for every scanner channel if a scanner is installed The instrument remembers how each channel is configured and will return to that configuration whenever the channel is chosen Refer to Paragraph 4 8 4 for instructions on how to change parameter selections for all channels at the same time 4 6 Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 4 5 1 Voltage Excitation Mode In voltage excitation mode the excitation selection in volts is treated as a not to exceed value The instrument assigns an excitation current to each resistance range that keeps the measurement voltage is as large as possible without exceeding the selected excitation voltage Current is changed to accommodate resistance range or voltage excitation parameter changes To use voltage excitation mode press the Voltage Mode key The LED annunciator above that key will light to indicate the excitation mode The excitation display line in the lower right corner of the display will also change to indicate the maximum excitation in volts Current excitation arrow keys ar
96. CPOL lt polarity gt term n lt polarity gt Specifies control heater output polarity 0 Unipolar 1 Bipolar When the temperature control output heater output is used to control a resistive heater this parameter should always be set to unipolar The bipolar setting should only be used in rare circumstances when the output is used to drive something other than a resistive heater Temperature Control Polarity Query CPOL term lt polarity gt term n Refer to command for description Curve Delete Command CRVDEL lt curve gt term nn lt curve gt Specifies which curve to delete I 20 CRYVDEL 20 term Deletes Curve 20 Curve Header Command CRVHDR lt curve gt lt name gt lt SN gt lt format gt lt setpoint limit gt lt coefficient gt term nn aaaaaaaaaaaaaaa aaaaaaaaaa N NNN NNN N lt curve gt Specifies which curve to configure 1 20 lt name gt Specifies curve name Limited to 15 characters lt SN gt Specifies the curve serial number Limited to 10 characters lt format gt Specifies the curve data format 3 Q K 4 log Q K lt setpoint limit gt Specifies the curve temperature limit in Kelvin lt coefficient gt Specifies the curves temperature coefficient 1 negative 2 positive Configures the user curve header CRVHDR 2 DT 470 00011134 3 3 250 1 term Configures User Curve 2 with a name of DT 470 serial no 00011134 data format of Ohms vs Kelvin setpoint li
97. Check if empty string strReturn RTrim strReturn Remove extra spaces and Terminators Do While Right strReturn 1 Chr 10 Or Right strReturn 1 Chr 13 strReturn Left strReturn Len strReturn 1 Loop Else strReturn No Response Send No Response End If frmIEEE txtResponse Text strReturn Put response in text on main form End If Loop End Sub Remote Operation 6 9 Lake Shore Model 370 AC Resistance Bridge User s Manual 6 1 5 3 Program Operation Once the example program is running try the following commands and observe the response of the instrument Input from the user is shown in bold and terminators are added by the program The word term indicates the required terminators included with the response ENTER COMMAND IDN Identification query Instrument will return a string identifying itself RESPONSE LSCI MODEL370 123456 02152001 term ENTER COMMAND RDGR 1 Ohm reading query Instrument will return a string with the present resistance reading from channel 1 RESPONSE 273 150E 00 term ENTER COMMAND HTRRNG 0 Heater range command Instrument will turn off the heater No response will be sent ENTER COMMAND HTRRNG Heater range query Instrument will return a string with the present heater range setting RESPONSE O term ENTER COMMAND HTRRNG 1 HTRRNG Heater range command followed by a query Instrument will change to heater Low setting then return a string RESPONSE
98. Cmd 1 11 Setpoint Ramp Parameter Query Setpoint Ramp Status Query eee Kelvin Reading Query cccccecceeseeeteeseteeseneeeenes Excitation Power Reading Query Resistance Reading Query oooocicicnconononncnnncnononananos Resistance Range Cmd 2 0uu0 0 04021212 Resistance Range Query Reading Status Query ue eueeoeesaea aaa eaeacacaaea Relay Parameter Cmd euuoeoeaea aaa caenaea Relay Parameter Query Relay Status Query Mininiai aaa a n ann Scanner Parameter Cmd 2 20 0 2 12121112 Scanner Parameter Query Temperature Control Setpoint Cmd Temperature Control Setpoint Query 35 Still Output Cmd 21 111212 Still Output Query Zone Table Cda Zone Table Query Remote Operation 6 19 Lake Shore Model 370 AC Resistance Bridge User s Manual 6 3 1 Interface Commands Alphabetical Listing CLS Clear Interface Command Input CLS term Remarks Clears bits in the Status Byte Register and Standard Event Status Register and terminates all pending operations Clears the interface but not the instrument The related instrument command is RST ESE Event Status Enable Register Command Input ESE lt bit weighting gt term Format nnn Remarks Each bit is assigned a bit weighting and represents the enable disable mask of the corresponding event flag bit in the Standard Event Status
99. Computers DE 9P DTE DB 25P DTE DE 9P DTE Pin Description Pin Description Pin Description 1 No Connection 2 TD out 1 DCD in 2 Receive Data RD in 3 RD in 2 RD in 3 Transmit Data TD out 4 RTS out 3 TD out 4 Data Terminal Ready DTR out 5 CTS in 4 DTR out 5 Ground GND 6 DSR in 5 GND 6 Data Set Ready DSR in 7 GND 6 DSR in 7 Data Terminal Ready DTR out tied to 4 8 DCD in 7 RTS out 8 No Connection 20 DTR out 8 CTS in 9 No Connection 22 Ring in in 9 Ring in in Figure 8 11 RS 232 DTE Connector Details Service 8 13 Lake Shore Model 370 AC Resistance Bridge User s Manual 8 11 1 Serial Interface Cable Wiring The following are suggested cable wiring diagrams for connecting the Model 370 Serial Interface to various Customer Personal Computers PCs Model 370 to PC Serial Interface PC with DE 9P Model 370 DE 9P Standard Null Modem Cable DE 9S to DE 9S PC DE 9P 5 GND at _ 5 GND 2 RD in MN 2 LLL 3 TD Out 3 TD out TD RD in 4 DTR out H 6 DSR in 6 DSR in SCO A DTR out 1 NC GN GE 7 RTS out 7 DTR tied to 4 gt 8 CTS in 8 NC 1 DCD in Model 370 to PC Serial Interface PC with DB 25P Model 370 DE 9P Standard Null Modem Cable DE 9S to DB 25S PC DB 25P 5 GND 7 GND 2 RD in TT 2 TD out 3 TD out 3 RD in 1 NC 3 4 RTS out 7 DTR tie
100. Configuration 6 6 Remote Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 6 1 5 2 Visual Basic IEEE 488 Interface Program Setup This IEEE 488 interface program works with Visual Basic 6 0 VB6 on an IBM PC or compatible with a Pentium class processor A Pentium 90 or higher is recommended running Windows 95 or better It assumes your IEEE 488 GPIB card is installed and operating correctly refer to Paragraph 6 1 5 1 Use the following procedure to develop the IEEE 488 Interface Program in Visual Basic Start VB6 Choose Standard EXE and select Open Resize form window to desired size R U BN On the Project Menu select Add Module select the Existing tab then navigate to the location on your computer to add the following files Niglobal bas and Vbib 32 bas 5 Add controls to form a Add three Label controls to the form b Add two TextBox controls to the form c Add one CommandButton control to the form 6 On the View Menu select Properties Window 7 Inthe Properties window use the dropdown list to select between the different controls of the current project 10 Set the properties of the controls as defined in Table 6 1 11 Save the program Remote Operation 6 7 Lake Shore Model 370 AC Resistance Bridge User s Manual Table 6 1 IEEE 488 Interface Program Control Properties Current Name Property New Value Labell Name IblExitProgram Caption T
101. E E E N Off Scan Dwell Time 1111111111111 10 sec Change Pause Time 111 11 3 sec Curve Number 21112111111 11 No Curve Temperature Coefficient Neg Cha nel oss ooo On Heater Heater RA gE a a iien as Off Instrument Setup Fregen yeon een o w 13 7 Hz Common Mode Reduction On Guard Diyes coccion a evhens Off Keypad Locking Modesto iia Unlocked Lock Code cooooonocococononononinnconnonnnonnnnnnononono 123 Linear and Filter CA R 1 Equation eiennenn ias MX B Slope M evitar 1 0000 X Sour Ennn niin Ohms A niak Value Ofisct Bcninnd at 0 0000 A dd O tte Off PID and Manual Heater Setting Proportional P s 44ss21 422 gt 10 Integral Mei 20 sec Derivative Dj 0 sec Manual wy ias 0 000 Remote Local Remote Local 12 12241241241221122 Local Setpoint Setpoint Value cece 0 0000 Zone Settings All Channels Setpoint Limit 2 44422411 211 12 0 000 K Proportional P c cccsseeseeseeeseeseeeteeees 10 Integral I oooooononoccn nonoconononnnonnnonncnnnnos 20 sec Derivative D 0ooonconconiccnonnnonconnconccnnonnnos 0 sec Manual Output oocoooccnccnconiconocnnonnnnnnnnos 0 000 Heater RaOg6 ooooonococococonocnnonnnonocanocnocnnoo Off Relay LOW ZE Off Relay High ooooooonccnncciconoocnoonconnconoconccnnoo Off Analog Output l ooonnonnccnocnioccnoninnconnos 0 00 Analog Outp
102. G7 8 Physics Today Eleventh Annual Buyer s Guide August 1994 ISSN 0031 9228 coden PHTOAD A 8 Glossary of Terminology Lake Shore Model 370 AC Resistance Bridge User s Manual APPENDIX B TEMPERATURE SCALES B1 0 DEFINITION Temperature is a fundamental unit of measurement which describes the kinetic and potential energies of the atoms and molecules of bodies When the energies and velocities of the molecules in a body are increased the temperature is increased whether the body is a solid liquid or gas Thermometers are used to measure temperature The temperature scale is based on the temperature at which ice liquid water and water vapor are all in equilibrium This temperature is called the triple point of water and is assigned the value 0 C 32 F and 273 15 K These three temperature scales are defined as follows Celsius Abbreviation C A temperature scale that registers the freezing point of water as 0 C and the boiling point as 100 C under normal atmospheric pressure Formerly known as Centigrade Originally devised by Anders Celsius 1701 1744 a Swedish astronomer Fahrenheit Abbreviation F A temperature scale that registers the freezing point of water as 32 F and the boiling point as 212 F under normal atmospheric pressure Originally devised by Gabriel Fahrenheit 1686 1736 a German physicist residing in Holland developed use of mercury in thermometry Kelvin Abbreviation K An absol
103. Hz 6 Read the voltage from the instrument using the CALVIN query Sufficient settling time and filtering must be used to obtain a stable reading Refer to Paragraph 8 12 2 6 for practical filter time constants The CALVIN reading is not filtered with the instrument filter feature 7 Calculate the gain coefficient value for the voltage range by the following CALG Voltage DVM CALVIN The gain coefficient should be 1 01 2 8 Send the gain coefficient to the instrument using the CALG command Example for the highest voltage range CALG 2 12 1 01212E 0 9 Send the CALSAVE command and wait 10 seconds for the calibration to be saved in non volatile memory Alternately this step can be done after all ranges are calibrated 10 This process can be repeated on lower voltage ranges until the accuracy of the DVM falls below the specified accuracy of the instrument With most DVMs the highest range is the only range that can be calibrated in this manner Paragraph 8 12 2 6 gives an example of how to use the instrument itself and known resistance values to accurately calibrate lower voltage ranges using a calibrated current range 8 12 2 3 Current Source Gain To calibrate current source gain perform the following sequence of operations Current settings do not require a zero offset calibration Current source gain can only be calibrated using a voltage range that is fully calibrated Allow the instrument to warm up for 1 hour 2 Select
104. Instrument Uncalibrated This error message must be bypassed to allow calibration of the instrument Press both Enter and Escape keys simultaneously to bypass the error message Operation in this state is possible but at least one calibration is known to be out of proper range and measurement is likely to be erroneous 8 12 2 Resistance Measurement Calibration A complete measurement calibration of the Model 370 is very complex and time consuming Make sure calibration is necessary before returning the instrument for calibration or attempting to calibrate it Verify the instrument accuracy by comparing measurements of resistive standards to the specifications in Chapter 1 of this manual Calibrating only the current settings and voltage ranges that are actually being used can save significant time and effort Lake Shore recommends returning the instrument to the factory whenever an instrument is damaged or a complete calibration is necessary Calibration coefficients for the measurement are used by the instrument firmware to compute an accurate resistance reading The sourced current and voltage at the instrument A D converter do not change as a result of calibration The equation for calculating resistance is as follows Calibrated Resistance RDGR A D Voltage Ideal Gain CALZ V x CALG V deal Current CALG 1 8 16 Service Lake Shore Model 370 AC Resistance Bridge User s Manual Resistance Measurement Calibration Continued
105. N Boiling Point 1 atm in K Thermal Conductivity Gas w em K Latent Heat of Vaporization Btu liter Liquid Density pounds liter C3 0 HANDLING CRYOGENIC STORAGE DEWARS Cryogenic containers Dewars must be operated in accordance with manufacturer instructions Safety instructions will also be posted on the side of each Dewar Cryogenic Dewars must be kept in a well ventilated place where they are protected from the weather and away from any sources of heat A typical cryogenic Dewar is shown in Figure C 1 NON MAGNETIC LIQUID HELIUM NON FLAMMABLE KEEP UPRIGHT Dewar eps Figure C 1 Typical Cryogenic Storage Dewar Handling LHe and LN C 1 Lake Shore Model 370 AC Resistance Bridge User s Manual C4 0 LIQUID HELIUM AND NITROGEN SAFETY PRECAUTIONS Transferring LHe and LN and operation of the storage Dewar controls should be in accordance with the manufacturer supplier s instructions During this transfer 1t is important that all safety precautions written on the storage Dewar and recommended by the manufacturer be followed WARNING Liquid helium and liquid nitrogen are potential asphyxiants and can cause rapid suffocation without warning Store and use in area with adequate ventilation DO NOT vent container in confined spaces DO NOT enter confined spaces where gas may be present unless area has been well ventilated If inhaled remove to fresh air If not breat
106. ONTACTING LAKE SHORE CRYOTRONICS ee eeeeeo eee aaa aaa aaa aaa aaa aaa rr nnnn rra 8 1 8 2 RETURNING PRODUCTS TO LAKE SHORE ee eee eee aaa aaa aaa aaa aaa naar rra rnnnnnns 8 1 8 3 FUSE DRAWER tee E No EA Z a A fo SE SEC 8 2 8 4 LINE VOLTAGE SELECTION uu eee da e 8 2 8 5 FUSE REPLACEMENT eo zc od O OE A wasiak O kesto 8 3 8 6 ERROR MESSAGES germanas 8 3 8 7 TROUBLESHOOTING irran oia ide a dadas 8 4 8 8 ENCLOSURE TOP REMOVAL AND REPLACEMENT coccccccccccocononcnnnnncnonnnnnnnnnonnnnnnnonnnnnnnnnnnnnnnnnnnnnns 8 8 8 9 ELECTROSTATIC DISCHARGE cui A YO ah 8 9 8 10 FIRMWARE REPLACEMENT oset ay reece OSP A PUGET PELE AES 8 9 8 11 CONNECTOR AND CABLE DEFINITIONS e eee eee eau aaaaaaaneanaa 8 11 8 12 MODEL 370 CALIBRATION nc owe pinoia pane an a PEEP Eta dk kt 8 16 8 13 MODEL 3716 SCANNER CALIBRATION eeeeeee eee ee aaa aaa aaa aaa aaa aaa aaa aaa 8 26 APPENDIX A GLOSSARY OF TERMINOLOGY 2 2 2 ssenessomes names omen awa anawa a awa aa reee A 1 APPENDIX B TEMPERATURE SCALES iii iii B 1 APPENDIX C HANDLING OF LIQUID HELIUM AND NITROGENY sssenessmensenananananaanananazzaawznazzaawe C 1 Table of Contents Lake Shore Model 370 AC Resistance Bridge User s Manual LIST OF ILLUSTRATIONS Figure No Title Page 1 1 Model 370 Front A NA 1 1 2 1 Dilution Refrigerator Block Diagram ooocoon
107. Paragraph 3 4 1 in case a cable must be made for a special application Guard pins are not connected inside the scanner but guards can be carried through the cable making them reversible Connectors mate with 6 pin Circular DIN plug Lake Shore P N G 106 233 3 5 5 Scanner Output Cabling A scanner to instrument cable Lake Shore P N 112 374 is included with the Model 3716 L and 3708 Scanner The figure below illustrates the cable wiring This configuration is recommended when making cables for the scanner Shielded twisted pair as described in Paragraph 3 4 3 can be used instead of dual triax for less demanding applications Cable shields should always be tied to connector shield pins on both ends of this cable 3 8 Installation Lake Shore Model 370 AC Resistance Bridge User s Manual Scan Control JosjuoD UBIS Input To Model 370 To Optional Scanner Iindno 112 374a bmp A yndjno Scan Control OI3U09 ULIS Shield Not Attached To Model 370 AC Resistance Bridge To Optional Scanner 13ndjno Cable Wiring bmp Figure 3 9 Scanner to Instrument Measurement Cable 3 9 Installation Lake Shore Model 370 AC Resistance Bridge User s Manual 3 5 6 A 15 pin D subminiature plug provides power supply and digital control input connections to the Model 3716 L and 3708 Scanner The scanner control cable included with the scanner is adequate for most installations A connector pin definition is
108. Paragraph 6 2 6 6 2 4 Character Format A character is the smallest piece of information that can be transmitted by the interface Each character is 10 bits long and contains data bits bits for character timing and an error detection bit The instrument uses 7 bits for data in the ASCII format One start bit and one stop bit are necessary to synchronize consecutive characters Parity is a method of error detection One parity bit configured for odd parity is included in each character ASCII letter and number characters are used most often as character data Punctuation characters are used as delimiters to separate different commands or pieces of data Two special ASCII characters carriage return CR ODH and line feed LF 0AH are used to indicate the end of a message string Table 6 3 Serial Interface Specifications Connector Type Connector Wiring Voltage Levels Transmission Distance 9 pin D style connector plug DTE EIA RS 232C Specified 50 feet maximum Timing Format Asynchronous Transmission Mode Half Duplex Baud Rate 300 1200 9600 Handshake Software timing Character Bits 1 Start 7 Data 1 Parity 1 Stop Parity Odd Terminators CR ODH LF OAH Command Rate 20 commands per second maximum 6 2 5 Message Strings A message string is a group of characters assembled to perform an interface function There are three types of message strings commands queries and responses The computer issues
109. Q REN Remote Puts the Model 370 into remote mode IFC Interface Clear Stops current operation on the bus SRQ Service Request Tells the bus controller that the Model 370 needs interface service 6 2 Remote Operation Lake Shore Model 370 AC Resistance Bridge User s Manual Bus Control Commands Continued A Multiline Command asserts a group of signal lines All devices equipped to implement such commands do so simultaneously upon command transmission These commands transmit with the Attention ATN line asserted low The Model 370 recognizes two Multiline commands LLO Local Lockout Prevents the use of instrument front panel controls DCL Device Clear Clears Model 370 interface activity and puts it into a bus idle state Finally Addressed Bus Control Commands are Multiline commands that must include the Model 370 listen address before the instrument responds Only the addressed device responds to these commands The Model 370 recognizes three of the Addressed Bus Control Commands SDC Selective Device Clear The SDC command performs essentially the same function as the DCL command except that only the addressed device responds GTL Go To Local The GTL command is used to remove instruments from the remote mode With some instruments GTL also unlocks front panel controls if they were previously locked out with the LLO command SPE Serial Poll Enable and SPD Serial Poll Disable Seri
110. R OR NOT LAKE SHORE HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES Your use of the Product is entirely at your own risk Some countries states and provinces do not allow the exclusion of liability for incidental or consequential damages so the above limitation may not apply to you LIMITED WARRANTY STATEMENT Continued 9 EXCEPT TO THE EXTENT ALLOWED BY APPLICABLE LAW THE TERMS OF THIS LIMITED WARRANTY STATEMENT DO NOT EXCLUDE RESTRICT OR MODIFY AND ARE IN ADDITION TO THE MANDATORY STATUTORY RIGHTS APPLICABLE TO THE SALE OF THE PRODUCT TO YOU CERTIFICATION Lake Shore certifies that this product has been inspected and tested in accordance with its published specifications and that this product met its published specifications at the time of shipment The accuracy and calibration of this product at the time of shipment are traceable to the United States National Institute of Standards and Technology NIST formerly known as the National Bureau of Standards NBS FIRMWARE LIMITATIONS Lake Shore has worked to ensure that the Model 370 firmware is as free of errors as possible and that the results you obtain from the instrument are accurate and reliable However as with any computer based software the possibility of errors exists In any important research as when using any laboratory equipment results should be carefully examined and rechecked before final conclusions are drawn Neither Lake Shore nor anyone else involv
111. RELAYS 30V 5A LNO LCOM LNC HNO HCOM HNC into slot Alarm Relays bmp Temrinal Block eps Figure 3 16 Alarm Relays Terminal Block 3 10 RACK MOUNTING The Model 370 can be installed into a 19 inch rack mount cabinet using the optional Lake Shore Model RM 1 Rack Mount Kit The kit contains mounting ears handles and screws that adapt the front of the instrument to fit into a 3 5 inch tall full rack space Additional support may be required in the rear of the instrument and to relieve strain on heavy cables The mounting ears are painted and do not guarantee good electrical contact between the instrument and cabinet They should not be used for ground strapping unless paint is removed from under all screws O Item Description P N Qty 1 Rack Mount Ear 107 440 2 N a A Q a 2 Screw 6 32x3 8 0 033 4 PH FLHD MS SS EN 3 Handles Rack 107 433 2 Fa Mount 3 Black NL 4 Screw 8 32x3 8 0 081 4 o PH FLHD MS SS Remove and discard 4 screws from case Replace with 4 screws from kit Rack Mount eps Figure 3 17 Model RM 1 Rack Mount Kit Assembly Installation 3 15 Lake Shore Model 370 AC Resistance Bridge User s Manual This Page Intentionally Left Blank 3 16 Installation Lake Shore Model 370 AC Resistance Bridge User s Manual CHAPTER 4 OPERATION 4 0 GENERAL This chapter provides operating instructions for most features of the Model 370 AC Resistance Bridg
112. ROM or installed in a Lake Shore instrument Previously called Precision Option calibrate To determine by measurement or comparison with a standard the correct value of each scale reading on a meter or other device or the correct value for each setting of a control knob Carbon Glass A temperature sensing material fabricated from a carbon impregnated glass matrix used to make the Lake Shore CGR family of sensors Celsius C Scale A temperature scale that registers the freezing point of water as 0 C and the boiling point as 100 C under normal atmospheric pressure Celsius degrees are purely derived units calculated from the Kelvin Thermodynamic Scale Formerly known as centigrade See Temperature for conversions Cernox A Lake Shore resistance temperature detector based on a ceramic oxy nitride resistance material CGR Carbon Glass Resistor cgs system of units A system in which the basic units are the centimeter gram and second Chebychev polynomials A family of orthogonal polynomials which solve Chebychev s differential equation Chebychev differential equation A special case of Gauss hypergeometric second order differential equation 1 x f x xf x nf x 0 closed loop See feedback control system coercive force coercive field The magnetic field strength H required to reduce the magnetic induction B in a magnetic material to zero coercivity generally used to designate th
113. Remarks RST Input Remarks SRE Input Format Remarks Example SRE Input Returned Format STB Input Returned Format Remarks TST Input Returned Format Remarks WAI Input Remarks Lake Shore Model 370 AC Resistance Bridge User s Manual Operation Complete Query OPC term 1 term Places a 1 in the instrument output queue upon completion of all pending selected device operations Send as the last command in a command string Not the same as OPC Reset Instrument Command RST term Sets instrument parameters to power up settings Use the DFLT command to set to factory defaults Service Request Enable Register Command SRE lt bit weighting gt term nnn Each bit has a bit weighting and represents the enable disable mask of the corresponding status flag bit in the Status Byte Register To enable a status flag bit send the command SRE with the sum of the bit weighting for each desired bit Refer to Paragraph 6 1 4 1 for a list of status flags To enable status flags 2 3 4 and 6 send the command SRE 92 term 92 is the sum of the bit weighting for each bit Bit Bit Weighting Event Name 2 4 Valid Read 3 8 Alarm 4 16 Error 6 64 SRQ 92 Service Request Enable Register Query SRE term lt bit weighting gt term nnn Refer to Paragraph 6 1 4 1 for a list of status flags Status Byte Query STB term lt bit weighting gt term nnn Acts like
114. Resistance Range Autorange and Scanner Channel are examples of keys that operate this way Setting Selection Allows the user to select from a finite list of parameter values During setting selection the A and Y keys are used to select a parameter value Enter is used to accept the change and advance to the next parameter Escape will cancel the change to that parameter and return to the normal display Setting selection screens always include the message Select with AW Data Entry Allows the user to enter numeric parameter values using the data entry keys that are printed on the key tops Data entry keys include numbers from 0 to 9 sign and decimal point The labels printed above the keys describe the key function during normal operation When one of the keys is pressed and a data entry sequence is started the keys follow the data entry functions printed on the key tops Once the correct parameter value is entered press Enter to accept the change and advance to next parameter Pressing Escape once will clear the new value and restart the setting sequence Pressing Escape again will return to the normal display Data entry screens always include the message Enter a value for Related setting selection and data entry sequences are often chained together under a single key To skip over a parameter without changing its value press Enter before pressing an arrow or number key To return to the normal display in the middle of a setting seq
115. SCAN SCAN SETP SETP STILL STILL ZONE ZONE Function Page Excitation Frequency Cmd 2 42424221111 121 28 Excitation Frequency Query 14242122011111121 28 Guard Drive Cid isis naci 28 Guard Drive Query cococococccncncninonononnnononncncnonanonanacnono 28 Heater Output Query 28 Heater Range Cd zawija i RR E 28 Heater Range Query euuaea aaa aaa enea acne 28 Heater Status Query 45 29 IEEE 488 Interface Parameter Cmd 29 IEEE 488 Interface Parameter Query 29 Input Channel Parameter Cmd Input Channel Parameter Query Keypad Status Query ococicicicoconncnononononcnnnnnononnnnannnano Linear Equation Data Query Linear Equation Parameter Cmd 30 Linear Equation Parameter Query 30 Keypad Lock Cmd Keypad Lock Query Min Max Data Query occcoconoconcnnnnnnnncnononannnoranncncnnnos Min Max Parameter Cmd Min Max Parameter Query 1 112111211212121211 Min Max Reset Cmd euueoeaeaeaa iai IEEE Interface Mode Cmd IEEE Interface Mode Query Monitor Output Selection Cmd 1 Monitor Output Selection Query Manual Heater Output Cd eee Manual Heater Output Query eee Closed Loop PID Parameter Cmd Closed Loop PID Parameter Query Setpoint Ramp Parameter
116. Send the zero offset coefficient to the appropriate output Analog Output 1 CALZ 4 1 Coefficient Analog Output 2 CALZ 4 2 Coefficient 12 Repeat from Step 3 for the second analog output 13 Send the CALSAVE command 8 12 4 Heater Output Calibration The Model 370 has 8 heater output current ranges The heater output is calibrated with individual zero offset and gain coefficients for each range A DC voltmeter with accuracy of I mV on a 10 V range resistors of known value from 20 Q to 60 kQ and cable to connect to the heater output are required Please note that calibration accuracy does not affect closed loop control accuracy Small errors in heater output accuracy are eliminated by the PID control equation Calculation of heater output power does depend on this calibration but it also depends on the heater resistance that is entered into the instrument Heater resistance is often difficult to measure because it changes with temperature and the heater is often installed with small gauge lead wire To calibrate the heater output ranges perform the following sequence of operations Configure the Model 370 for Open Loop control Refer to Paragraph 4 11 2 1 Set the heater output to the highest range of 100 mA Set the heater calibration coefficients to their defaults of 0 and 1 with the following commands CALZ 3 1 0 and CALG 3 1 1 Attach a nominal 20 Q resistor of known value across the heater output 2 Set the Manual Output parameter val
117. The CALVIN query returns only the voltage part of the equation so that current and voltage contributions can be separated for calibration The voltage represented is the voltage at the input terminals V V of the instrument The equation for CALVIN is as follows Calibrated Voltage CALVIN A D Voltage Ideal Gain CALZ V x CALG V The fundamental relationship between current voltage and resistance is straightforward but the interaction of the calibration coefficients should be noted The measurement input allows calibration of 22 individual current gains and 12 voltage ranges with zero offset and gain Calibration can be performed on voltage ranges and current settings shown in Table 1 1 but not all combinations can be selected during operation This results in 46 calibration coefficients for the input alone This approach was chosen because calibrating each possible combination of voltage range and current would require 528 coefficients Even the chosen method has difficulties Some of the voltages and currents are so small that they cannot be measured with conventional calibration equipment After the basic calibration technique is described a method is described in Paragraph 8 12 2 6 for using the Model 370 and known resistances to perform a calibration 8 12 2 1 Voltage Measurement Zero Offset To calibrate voltage zero offset perform the following sequence of operations Voltage zero offset must be calibrated before voltage gain
118. The signal voltage at either monitoring point can be viewed on an oscilloscope by using the Monitor output described in Paragraph 4 7 4 Usual causes of this error include 1 An attempt to measure a resistance that is much greater 3 orders of magnitude than the selected range 2 An open connection of the current drive leads I and or I To resolve this condition first verify that a reasonable resistance range has been selected for the measurement Next verify the integrity of current lead connections to the measured resistance 8 7 1 2 VCM OVL Common Mode Overload Even though differential mode noise may be fully under control by using twisted leads and appropriate shielding it is still possible to have enough common mode noise coupled to the measurement leads to exceed the common mode voltage limits of the input instrumentation amplifier Ifthe combination of signal and common mode noise components exceeds 10 V peak the common mode overload error VCM OVL is displayed The overload can occur at either VCMPOS or VCNEG points in the instrumentation amplifier circuit Refer to Figure 8 3 for the location of VCMPOS and VCNEG monitoring points The signal voltage at either monitoring point can be viewed on an oscilloscope by using the Monitor output described in Paragraph 4 7 4 Usual causes for this error include 1 Improperly shielded measurement leads 2 An improperly grounded Dewar case To resolve this con
119. This section describes not only how to attach a resistor to the Model 370 but also suggests ways to manage measurement noise Understanding the information given in this section before installing the Model 370 can save hours of troubleshooting Additional information on installing the Model 3716 Scanner is given in Paragraph 3 5 3 4 1 Resistor Input Connector Definition Two 6 pin 240 Circular DIN 45322 receptacles are used for the resistance measurement input One connector provides voltage source excitation connections V and the other provides current measurement connections I Both mate with 6 pin Circular DIN plug included with instrument Lake Shore P N G 106 233 Examples of lead arrangements are shown in Figure 3 3 SHIELD SHIELD Input Connectors eps INPUT V INPUT I Pin Name Description Name Description 1 V Measurement voltage lead V I Measurement current lead I 2 VG Guard or shield for voltage lead V IG Guard or shield for current lead I 3 Shield Shield tied to measurement common Shield Shield tied to measurement common 4 VG Guard or shield for voltage lead V IG Guard or shield for current lead I 5 V Measurement voltage lead V I Measurement current lead I 6 Shield tied to measurement common Shield tied to measurement common Body Tied to measurement common Tied to measurement common NOTES 1 The Model 370 makes an AC resistance measurement therefore the and
120. VL message Common mode voltage can be observed at the CMPOS and CMNEG monitoring points The normal signal level is very small at these points and may not be easily observed when measuring at low excitations Although the Model 370 largely ignores common mode noise there may be a large enough signal to cause a problem Ifa VCM OVL message is displayed the signal character of the offending noise can be observed and verified gone after repair This is usually due to poor shielding and or grounding as discussed in Paragraph 3 4 A level exceeding 10 V peak at these points trigger a VCM OVL message The current source drive to the current sampling resistors of the current source section can be observed using CSPOS and CSNEG monitor points This signal shows predominately the controlling sine wave signal applied to the current sampling resistors but will also give good insight in to common mode noise voltages that are reflected back through the current source A level exceeding 10 V peak at these points triggers a CS OVL message 8 7 2 3 Phase Relationship The design of the Model 370 includes a phase sensitive detector that uses a reference signal directly derived from the oscillator section to demodulate the voltage signal from the measured resistor The Reference signal is brought out to the back of the instrument next to the Monitor jack The phase relationship of these two signals is critical to the accuracy of the measurement Co
121. Values Alarm and Relay Add dei Off Alarm Audible oooonnoninoninonicnnonnconcconccnno Off Relay LOW aaa Zoo A Off Relay High a8 cats Off Analog Output Analog Output 1 oo eee eeeeereeeeetees Off Analog Output 2 0 0 ee eeeeeeeeeeeeeees Off Computer Interface BaUd see a st eaten 9600 IEEE Address 2 2 2 2 2 2 24 2212112212 12 IEEE Terminators 11111111111 CR LF Control Setup Control Channel 2 0 2 uussu1eu1a2 1 Setpoint Units oooooocciocnionoconocnnonnnnnos Ohms Control Mode 1 121122112212112111 1 Open Heater Range Limit 100 mA Setpoint Change Pause Time 25 sec Control Filternet Off Heater Display 21121112111111 1 Current Setpoint Ramp oooooconocononcnonononononononinnnonos Off Heater Resistance 100 ohm Heater Polarity 111111111111 Unipolar Display Format Number of Locations 22212211241124112112 1 Location 1 Channel Scan Location 1 Data Source Ohms Display Resolution 11221121112112211 1 6 Brightness cita 15 Excitation and Range Excitation Mode 1 111 111111 1 Voltage EXcitation ccccssecssserseersestsesseens 632 mV Resistance Range 63 2 Mohm Altos Off Input Setup Scan Channel coooococcccnoccnocononcconccnnccnnonnnonoos 1 PaO i 1 A A
122. a closed loop PID control algorithm to balance the continuous cooling power of the refrigeration system and maintain steady load temperature HEATER OUTPUT Heater Output bmp Figure 3 14 Control Heater Output Connector 3 7 1 Heater Output Isolation Heater output circuitry is independently isolated from both chassis ground and measurement circuits Isolation is intended to reduce noise from coupling into the heater output from digital circuits inside the instrument and to eliminate ground loop problems The heater output should not be left completely floating during operation Similar to the isolated measurement circuit shied the heater common should be connected to the test Dewar at a single point Other noise reduction techniques used for the resistor input apply to the heater as well 3 12 Installation Lake Shore Model 370 AC Resistance Bridge User s Manual 3 7 2 Heater Output Connection BNC receptacle provides connections for the temperature control Heater Output Signal is carried on the center conductor and isolated heater output measurement common is on the outer conductor The heater output has a maximum voltage of 10 V and can source 100 mA Mating connector not included 3 7 3 Heater Output Cabling Coaxial cable is recommended for the heater output outside of the test Dewar The center conductor should be large enough to carry the maximum 100 mA current to prevent accidental burn out of the conductor The outer
123. aa aaa aaa aaa aaa ania ca 2 5 2 4 1 Why is the Model 370 called a Bridge eee eau aaa rra 2 5 2 4 2 Resistance Measuremen a tdt erica 2 5 2 4 3 Excitation MOde coran 2 4 4 Resistance RangeS us u324346 ssa cine gueeccoletgadgecgudesdanetencceeegesuusdanledyscsuecteeedeesteeceeveccnesdeedeesleed 2 4 5 Measurement Circuit 2 4 5 1 Shielding and Grounding Theory eee eu aaa eee aaa aaa aaa aaa aaa rare 2 7 2 4 5 2 OT 2 7 2 4 5 3 Electrical ISOlathOn i oe ETZ td ti 2 7 2 4 5 4 Driven Guards az wat A zd ZA ba EZ RE tit 2 7 2 4 5 5 Active Common Mode Reduction oooooccccnnoncccnococnnononcncnono nono aaa aaa aaa n rra 2 8 2 4 5 6 Matched Impedance Current Source eee eee eee eee aaa aaa aaa aaa nr rr anna 2 8 2 4 6 Measurement Speed and Filtering irois vacante ade dias 2 8 2 5 NOISE SOURCES cirio A AYO ie ici 2 9 2 5 1 Induced ElectricalNoiSe cocino a E e a nP e Ea 2 9 2 5 2 Ground LOOp NISC socio dico 2 9 2 5 3 Analog Circuit NOISE iii Ad 2 9 2 5 4 Digital Circuit Noise s h 40 ised ot AE O A A OT AO ea wod oO dA os nee aia a 2 9 2 5 5 Resistor Thermal NoiSe inie at Le ERY dada ca 2 10 2 5 6 Interference Between Two Model 370S o eee aaa aaa aaa aaa aaa aaa nan nn 2 10 2 5 7 Vibration NOIS sa cit TE La aaa doga cidad 2 10 2 6 MEASUREMENT ERROR senapan O ie steer A OE YO PA OWA 2 10 2 6 1 Warm Up and Temperature Drift uuu ee
124. ads as if the noise source and lead are two halves of a capacitor The amount of coupling is related to the voltage change at the source the common area between source and receiver length of lead and dielectric between them distance in air A cable shield is the first line of defense against E field noise No measurement cable of any length should be run from the Model 370 without a shield The shield provides a low impedance path to measurement common that prevents the voltage change at the source from being seen at the measurement leads H field noise inductively couples into measurement leads or other conductive loop Current is induced in the lead as if the noise source and lead are two halves of a transformer The amount of coupling is related to the field change from the source and the loop area Tightly twisted leads inside the shielded cable reduce the total loop area and minimize the effect of H field noise It is also effective to move the instrument and leads away from noise sources because field strength reduces by the inverse cube of distance Even with proper installation it is difficult to reduce induced noise below 1 to 10 UV High resistance ranges are affected more because the leads create a more efficient antenna when they are terminated at a high resistance 2 5 2 Ground Loop Noise Ground loops produce noise that acts similar to induced noise but the cause of the noise is different Current is generated in any conductive circuit
125. al display The next zone setup screen appears as a prompt for the setpoint value that is the top of the zone Use the data entry keys to enter a setpoint value Enter the numeric value first Press Enter to accept the new value and advance to the units multiplier Use the A or Y key to select a units multiplier of m blank k Press Enter to accept the new selection and return to the normal display Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display The next zone setup screen appears as a prompt for the proportional P value Use the data entry keys to enter a proportional value between 0 001 and 1000 Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display Advanced Operation 5 7 Lake Shore Model 370 AC Resistance Bridge User s Manual Entering Zone Parameter Values Continued The next zone setup screen appears as a prompt for the integral I value Use the data entry keys to enter an integral value between 1 and 10 000 or zero for off Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and retur
126. al display If the ramp feature is turned on the next setup screen appears as a prompt for ramp rate Use the data entry keys to enter a ramp rate value between 0 001 and 10 K m Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a different value Press Escape again to cancel the sequence and return to the normal display To begin a ramp enter a new setpoint value To stop an active ramp at its present value press Setpoint and then press Enter twice 4 22 Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 4 11 8 Controlling and Scanning The Model 370 and 3716 Scanner allow both temperature control and data taking at the same time Firmware in the Model 370 manages automatic switching between the control channel and scan channel and will maintain the control heater output when the control channel is not active Scanning during closed loop temperature control will never be as stable as dedicated control instrument but with proper setup acceptable performance can be achieved for many applications 4 11 8 1 Reading Sequence The instrument takes control of the reading sequence when controlling temperature with one scanner channel and measuring on another It will alternate between the control channel and selected scan channel as quickly as possible so that control is not suspended for longer than necessary If autoscan is enable
127. al measurement leads from external noise Obviously this will not shield noise generated within the Dewar In many installations the best noise performance of the Model 370 can be achieved when the cable shield is tied directly to the Dewar or other large conductive mass surrounding the experimental space The problem is that most test Dewars are connected to Earth ground either directly or through the many vacuum and transfer lines running in and out It would be very difficult to isolate a Dewar that was not specifically designed that way therefore the isolation of the instrument must be relied upon to prevent ground loops in this configuration When the cable shield is connected to a Dewar and therefore Earth ground that should be the only connection between the measurement circuits and Earth Installation 3 5 Lake Shore Model 370 AC Resistance Bridge User s Manual Isolation and Grounding Continued Any additional connection between measurement circuits and ground will defeat the isolation built into the Model 370 and allow ground loops to form In the unusual circumstance that one measurement lead is shorted to the cable shield measurement common the Model 370 can still make resistance measurements but the active CMR feature must be turned off 3 4 5 Dressing Cables Cable dressing is often overlooked in crowded laboratories but even neatness can matter when making small signal measurements Consider the following ideas for long cable
128. al polling accesses the Service Request Status Byte Register This status register contains important operational information from the unit requesting service The SPD command ends the polling sequence 6 1 3 2 Common Commands Common Commands are addressed commands which create commonalty between instruments on the bus All instruments that comply with the IEEE 488 1987 standard share these commands and their format Common commands all begin with an asterisk They generally relate to bus and instrument status and identification Common query commands end with a question mark Model 370 common commands are detailed in Paragraph 6 3 and summarized in Table 6 6 6 1 3 3 Device Specific Commands Device specific commands are addressed commands The Model 370 supports a variety of device specific commands to program instruments remotely from a digital computer and to transfer measurements to the computer Most device specific commands perform functions also performed from the front panel Model 370 device specific commands are detailed in Paragraph 6 3 and summarized in Table 6 6 6 1 3 4 Message Strings A message string is a group of characters assembled to perform an interface function There are three types of message strings commands queries and responses The computer issues command and query strings through user programs the instrument issues responses Two or more command strings can be chained together in one communication bu
129. alarms from all channels will be reset 4 28 Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 4 15 RELAYS The Model 370 has two mechanical relays designated high and low The relays are most commonly associated with the high and low alarms but they can also be controlled manually and used for other purposes Both relays default to their normal state when instrument power is turned off The relays have three modes of operation Manual The relay is turned off normal or on active manually from the front panel or over computer interface Alarm The relay is configured to follow the operation of the alarms The user can assign channel number to each relay and tie the action of the relay to the high alarm low alarm or both alarms of that channel The relay is activated any time the associated alarm is active Zone The relay is operated with the zone control feature as described in Paragraph 5 3 Press the Relay key to begin relay setup The following message appears as a prompt for relay selection Use the A or Y key to select the high or low relay Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next relay setup screen appears as a prompt for operating mode Use the A or V key to select the relay mode Manual Off Manual On Alarm or Zone Press Enter to accept th
130. all contents have been accounted for Check off each item on the packing list as it is unpacked Instruments themselves may be shipped as several parts The items included with the Model 370 and optional scanners are listed below Contact Lake Shore immediately if there is a shortage of parts or accessories Lake Shore is not responsible for any missing items if not notified within 60 days of shipment Inspect all items for both visible and hidden damage that occurred during shipment If damage is found contact Lake Shore immediately for instructions on how to file a proper insurance claim Lake Shore products are insured against damage during shipment but a timely claim must be filed before Lake Shore will take further action Procedures vary slightly with shipping companies Keep all shipping materials and damaged contents until instructed to either return or discard them If the instrument must be returned for recalibration replacement or repair a returned goods authorization RGA number must be obtained from a factory representative before it is returned The Lake Shore RGA procedure is given in Paragraph 8 2 Items Included with Model 370 AC Resistance Bridge 1 Model 370 Instrument Model 370 User s Manual Line Power Cord Resistor Input Mating Connectors Terminal Block Mating Connector 6 Position Rh Pa NOR R Fuse Pair for Alternative Voltage Items Included with Model 3716 L and 3708 Scanner Model 3716 L scanner or 3708 scann
131. and repair Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design manufacture and intended instrument use Lake Shore Cryotronics Inc assumes no liability for Customer failure to comply with these requirements The Model 370 protects the operator and surrounding area from electric shock or burn mechanical hazards excessive temperature and spread of fire from the instrument Environmental conditions outside of the conditions below may pose a hazard to the operator and surrounding area Indoor use e Altitude to 2000 meters Temperature for safe operation 5 C to 40 C e Maximum relative humidity 80 for temperature up to 31 C decreasing linearly to 50 at 40 C e Power supply voltage fluctuations not to exceed 10 of the nominal voltage e Overvoltage category II e Pollution degree 2 Ground the Instrument To minimize shock hazard the instrument is equipped with a 3 conductor AC power cable Plug the power cable into an approved three contact electrical outlet or use a three contact adapter with the grounding wire green firmly connected to an electrical ground safety ground at the power outlet The power jack and mating plug of the power cable meet Underwriters Laboratories UL and International Electrotechnical Commission IEC safety standards Ventilation The instrument has ventilation holes in its side covers Do not block these holes w
132. and return to the normal display The next curve entry screen appears as a prompt for the control setpoint limit Set to 9999K if no limit is wanted Use the data entry keys to enter a setpoint limit in Kelvin Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display The next curve entry screen that appears is for entering or editing breakpoint data The operator can scroll up and down the entire list of breakpoints Breakpoints in the middle of the table can be edited or deleted New breakpoints can be added between existing ones but they must be entered at the end of the table and the instrument will put them in the proper place The example below illustrates Curve Number 1 named User Curve 01 Serial Number 0123456789 in log Q K format The first seven breakpoints are shown and the eighth line is ready for the next breakpoint When the breakpoint number is highlighted use the A or Y key to scroll through the breakpoints in the curve Scrolling will stop at the first zero breakpoint at the bottom of the curve so new breakpoints can be entered Press the Enter key when the desired breakpoint number is highlighted and continue to the resistance field When the resistance field is highlighted use the numerical keypad to enter the applicable resistance val
133. anner 15 pin D subminiature receptacle provides power supply output and digital control 2 SEAN CONTROL connections for the Model 3716 Scanner Refer to Paragraph 3 5 6 REFERENCE BNC receptacle provides connections for Frequency Reference Output Refer to Paragraph 3 6 MONITOR BNC receptacle provides connections for Monitor Voltage Output Refer to Paragraph 3 6 BNC receptacle provides current source output connections for the temperature control HEATER CUTEUF Heater Output Refer to Paragraph 3 7 ANALOG OUT 1 BNC receptacles provide connections for analog voltage output 1 and 2 Still The STILL output can be configured to heat the still of a dilution refrigerator Refer to and 2 STILL Paragraph 3 8 O ALARM RELAYS Detachable terminal block provides alarm relay connections Refer to Paragraph 3 9 IEEE 488 compliant interface connector for use with IEEE 488 parallel computer IEEE 488 INTERFACE interface Refer to Paragraph 6 1 and see Figure 8 12 Requires extender Lake Shore P N 4005 if used at the same time as the Alarm Relays connector da iiin Rear Panel bmp Figure 3 1 Model 370 Rear Panel 3 2 Installation Lake Shore Model 370 AC Resistance Bridge User s Manual 3 3 LINE INPUT ASSEMBLY This section describes how to properly connect the Model 370 to line power Please follow these instructions carefully to ensure proper operation of the instrument and the safety of oper
134. aph 3 4 4 8 7 1 5 R OVER Range Over This message indicates that the selected resistance range is lower than that required for the measured resistance To resolve the condition simply increase the resistance range Service 8 5 Lake Shore Model 370 AC Resistance Bridge User s Manual 8 7 1 6 R UNDER Range Under This message indicates two simultaneous problems 1 Either the current leads or the voltage leads are reversed in the and sense 2 The selected range is lower than that required for the measured resistance To resolve the condition correct the sensor wiring and increase the resistance range as needed NOTE As a test just increasing the resistance range will remove the error message but a negative resistance will be displayed 8 7 2 Signal Level Troubleshooting The Model 370 provides user access to all the error monitoring points in the circuit for troubleshooting convenience Access to these points is provided through the MONITOR output BNC connector on the back of the instrument To select the desired monitoring point refer to Paragraph 4 7 4 Monitor Output Voltages The Reference output provides an excellent means to trigger the oscilloscope when looking at the measurement AC signal in the presence of noise Be aware that the attachment of an earth grounded oscilloscope to the Monitor and or Reference output jacks may induce ground loop related noise into the system that is not normally present Converse
135. aph 3 7 4 The temperature control heater output has several hardware and firmware protection features to prevent accidental overheating of loads Hardware protection includes short circuit protection and built in circuitry that limits the transient energy sourced by the output when power is turned on or the range is changed The instrument firmware will turn off the heater when an overload or other error condition is detected on the control channel or if there is no heater load attached The instrument always powers up with the heater off Operation 4 21 Lake Shore Model 370 AC Resistance Bridge User s Manual Heater Output Range and Heater Output Off Continued To change the heater output range press the Heater Range key The old range selection on the display will be highlighted to prompt for the new range setting Use the A or Y key to select one of the eight heater output ranges Press Enter to accept the new selection and complete the setting sequence Press Escape to cancel the new value and leave the setting sequence To turn the heater output off press the Heater Off key The power to the heater will quickly be turned off and the following display will appear 4 11 7 Setpoint Ramp The Model 370 can be configured to ramp the control setpoint from its present value to a new value The ramp can produce a smooth linear transition in temperature rather than the step response normally associated with clos
136. arameters isanne A O YO o AO O A OTO 5 2 5 2 Recommended Curve Parameters vie ie cccecccescacesceedeavasecsscvassaevsthnsecesdecuaseasteessceescesdateesadshedcusaasasectesdaseabaesntdonsaasvadeutageons 5 3 6 1 IEEE 488 Interface Program Control Properties cccccccceceseeeeeeceeeeeaeeeeeeeeeeeeaeeeseaeeeseaeeeeeeeeeseaeeeseaeeeseieeeeseeeeseneeeseaee 6 2 Visual Basic IEEE 488 Interface Program 6 3 Serial Interface Specifications 6 4 Serial Interface Program Control Properties e 6 5 Visual Basic Serial Interface Program eee eee eee aaa aaa aaa aaaaaaacca 6 6 command SUMIMANY Gmin RR a ME a dg EE e A da L 8 1 Instrument Hardware Messages 8 2 Measurement or Compliance Limit Messages 8 3 legal Operation Messages voii inca idilio dede WSP RE EEEE RPG B 1 Temperature Conversion Table swa ace dace cd a aa aaia aeda dera PEELA AE C 1 Comparison of Liquid Helium and Liquid Nitrogen e eee eee eee eee aaa a aaa anawa aaa aaa aaa raaaaaace C 1 vi Table of Contents Lake Shore Model 370 AC Resistance Bridge User s Manual CHAPTER 1 INTRODUCTION 1 0 GENERAL This chapter provides an introduction to the Model 370 AC Resistance Bridge The Model 370 was designed and manufactured in the United States of America by Lake Shore Cryotronics Inc The Model 370 features include the following Model 370 Features e Resistance measurement
137. arge for this option Other country line cords available consult Lake Shore Accessories included with the Model 370 G 106 233 Sensor input mating connector 6 pin DIN plug 2 included 106 737 Terminal block mating connector Calibration certificate MAN 370 Model 370 user s manual Accessories included with the preamps scanners G 106 253 DB 25 plug 4 included G 106 264 DB 25 hood 4 included 107 379 Mounting bracket 112 374 3 m 10 ft cable from scanner to Model 370 Options and Accessories 4005 1 m 3 3 ft IEEE 488 GPIB computer interface cable assembly includes extender required for simultaneous use of IEEE cable and relay terminal block 8000 CalCurve CD ROM calibrated sensor breakpoint table on a CD ROM for customer upload 8001 370 CalCurve factory installed calibrated sensor breakpoint table factory installed into nonvolatile memory 8000 05 370 CalCurve field installed calibrated sensor breakpoint table loaded into nonvolatile memory for customer installation CAL 370 CERT Instrument recalibration with certificate CAL 370 DATA Instrument recalibration with certificate and data RM 1 Kit for mounting one Model 370 in a 482 6 mm 19 in rack mount cabinet Introduction Lake Shore Model 370 AC Resistance Bridge User s Manual 1 4 SAFETY SUMMARY Observe these general safety precautions during all phases of instrument operation service
138. ather than measured Step Range Setting Coefficient Resistor Current Filter DVM CALVIN Coefficient 3 632 mV Gain CALG 2 12 200 31 6 mA 10s Voltage Gain Coefficient Voltage DVM CALVIN Step Range Setting Coefficient Resistor Voltage Filter Actual RDGR Coefficient 4a 31 6mA Gain CALG 1 22 200 632 mV 10s 4b 10 mA Gain CALG 1 21 60 0 632 mV 10s 4c 3 16mA Gain CALG 1 20 200 Q 632 mV 10s 4d I mA Gain CALG 1 19 600 Q 632 mV 10s 4e 316 HA Gain CALG 1 18 2 KQ 632 mV 10s 4f 100 uA Gain CALG 1 17 6 kQ 632 mV 10s 4g 31 6 uA Gain CALG 1 16 20 kQ 632 mV 10s 4h 10 pA Gain CALG 1 15 60 ka 632 mV 10s 4i 3 16 UA Gain CALG 1 14 200 KQ 632 mV 10s 4j 1 UA Gain CALG 1 13 600 kQ 632 mV 10s Current Gain Coefficient Actual Resistance RDGR Step Range Setting Coefficient Resistor Current Filter Actual RDGR Coefficient 5a 200mV Gain CALG2 11 200 kQ 1 uA 10s 5b 63 2 mV Gain CALG 2 10 60kQ l uA 10s 5c 20 mV Gain CALG 2 9 20 kQ I uA 10s 5d 6 32mV Gain CALG 2 8 6 KQ 1 uA 10s Se 2mV Gain CALG 2 7 2 kQ 1 uA 10s 5f 632 uV Gain CALG 2 6 600 Q 1 uA 10s 5g 200 uV Gain CALG 2 5 200 Q l uA 10s 5h 63 2uV Gain CALG 2 4 60 Q l uA 50s 5i 20HV Gain CALG 2 3 200 l uA 50s 5j 6 32uV Gain CALG 2 2 60 l uA 100s 5k 2uV Gain CALG 2 1 20 luA 200s
139. ating a further delay The instruments A D sampling update rate is 10 readings per second and all readings are available over computer interface with an efficiently written program This rate allows the A D to capture any resistance changes that are able to pass through the input hardware The temperature control loop analog voltage outputs and max min capture are updated with each A D reading The display is updated at 2 readings per second 2 8 Theory of Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 2 5 NOISE SOURCES The effect of electrical noise is usually small and can be ignored when making routine measurements but all noise matters when signals fall to the microvolt level The following paragraphs describe the most common noise source encountered with small signal measurements and some techniques that can help reduce their effect More specific installation instructions are given n Chapter 3 The Model 370 offers a monitor voltage output that can be helpful in diagnosing noise related problems The instrument will also display a message if noise overloads the measurement electronics Refer to Chapter 7 for more information on diagnosing noise problems 2 5 1 Induced Electrical Noise Noise from the environment is typically classified as either electric field E field or magnetic field H field E field noise capacitively couples into measurement leads or the resistor being measured Voltage is induced on the le
140. ation b fuses software and non rechargeable batteries c software interfacing parts or other supplies not furnished by Lake Shore d unauthorized modification or misuse e operation outside of the published specifications or f improper site preparation or maintenance 6 TO THE EXTENT ALLOWED BY APPLICABLE LAW THE ABOVE WARRANTIES ARE EXCLUSIVE AND NO OTHER WARRANTY OR CONDITION WHETHER WRITTEN OR ORAL IS EXPRESSED OR IMPLIED LAKE SHORE SPECIFICALLY DISCLAIMS ANY IMPLIED WARRANTIES OR CONDITIONS OF MERCHANTABILITY SATISFACTORY QUALITY AND OR FITNESS FOR A PARTICULAR PURPOSE WITH RESPECT TO THE PRODUCT Some countries states or provinces do not allow limitations on an implied warranty so the above limitation or exclusion might not apply to you This warranty gives you specific legal rights and you might also have other rights that vary from country to country state to state or province to province 7 TO THE EXTENT ALLOWED BY APPLICABLE LAW THE REMEDIES IN THIS WARRANTY STATEMENT ARE YOUR SOLE AND EXCLUSIVE REMEDIES 8 EXCEPT TO THE EXTENT PROHIBITED BY APPLICABLE LAW IN NO EVENT WILL LAKE SHORE OR ANY OF ITS SUBSIDIARIES AFFILIATES OR SUPPLIERS BE LIABLE FOR DIRECT SPECIAL INCIDENTAL CONSEQUENTIAL OR OTHER DAMAGES INCLUDING LOST PROFIT LOST DATA OR DOWNTIME COSTS ARISING OUT OF THE USE INABILITY TO USE OR RESULT OF USE OF THE PRODUCT WHETHER BASED IN WARRANTY CONTRACT TORT OR OTHER LEGAL THEORY AND WHETHE
141. ation point of the current source output to minimize the common mode voltage seen at the measurement input Active common mode reduction allows the system to operate in environments that would otherwise saturate the instrument s differential input amplifiers The effect of active common mode reduction is most evident when excitation current is uA or greater but it should be left on most of the time Active common mode reduction should be turned off any time a measurement lead is tied to measurement common shield The reduction circuit will not operate properly in this configuration because the symmetry of the measurement is lost Measurements can be made with one lead grounded if the active common mode reduction feature is turned off 2 4 5 6 Matched Impedance Current Source The Model 370 takes the idea of a differential input one step further with a unique feature in its current source The two current source output terminals have the same source impedance similar to the voltage input terminals of a differential input that have the same input impedance The resistor being measured and all noise sources in the environment see the same impedance on both current source leads when looking back into the instrument The circuit improves the measurement in two ways First the differential input remains truly differential and performs well measuring any value of resistor Traditional current sources with different impedance on each lead one lead grounde
142. ators Line Cord Power Switch Fuse Input O Off I On Drawer lt 100 120V_O 5AT 250V 5x20mm 220 240V_0 25A T 250V 5x20mm 10 6 Voltage 50 60 Hz 50 VA MAX DN 100 120 220 240 V Power Switch bmp Figure 3 2 Line Input Assembly 3 3 1 Line Voltage The Model 370 has four different AC line voltages configurations so that it can be operated from line power anywhere in the world The nominal voltage and voltage range of each configuration is shown below The recommended setting for 230 V operation is 240 V Nominal Minimum Maximum 100 V 90 V 106 V 120 V 108 V 127 V 220 V 198 V 233 V 240 V 216 V 254 V Verify that the AC line voltage indicator in the fuse drawer window shows the appropriate AC line voltage before turning the instrument on The instrument may be damaged if it is turned on with the wrong voltage selected Instructions for changing the line voltage configuration are given in Paragraph 8 4 3 3 2 Line Fuse and Fuse Holder The line fuse is an important safety feature of the Model 370 If a fuse ever fails it is important to replace it with the value and type indicated on the rear panel for the line voltage setting The letter T on the fuse rating indicates that the instrument requires a time delay or slow blow fuse Fuse values should be verified any time line voltage configuration is changed Instructions for changing and verifying a line fuse are given in Paragrap
143. ature coefficient parameter below is ignored No Curve can be selected for any channel that is not used for temperature readings The next input setup screen appears as a prompt for the curve number Curve name and serial number follow for reference Use the A or Y key to select a curve number between one and twenty or No Curve Only the numbers for curves loaded in the instrument will appear in the selection sequence Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display 4 6 5 Temperature Coefficient If an RTD is used for temperature control and does not have a temperature response curve the user must indicate the temperature coefficient of the device so that the control equation can be calculated correctly This setting sequence will be skipped if a curve is selected for a channel Platinum and Rhodium Iron RTDs are positive temperature coefficient and most thermistors and semiconductor RTDs like Cernox and Germanium are negative temperature coefficient The next input setup screen appears as a prompt for a temperature coefficient Use the A or Y key to select a coefficient of positive and negative Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display 4 6 6 Turning A Channel On or Off If a scanner is
144. ature that will increase the resistance range any time the measured resistance is greater than full scale for the range and decrease the range any time the measured resistance is below 20 of full scale The ranges have an approximate 20 over range capability so in most cases the instrument will increase range before an overload appears Autorange works with either voltage or current excitation It will not change the excitation setting in order to reach a resistance range that is not available for the present selection 2 4 5 Measurement Circuit The idea that small signal measurement instruments require special consideration during installation is expressed frequently in this manual The following paragraphs illustrate how to treat the instrument and its environment as a Measurement Circuit and in general how to integrate instruments like the Model 370 The circuit analogy is appropriate because almost anything electrical or electronic in a laboratory has the potential to induce electric field E field or magnetic field H field noise on the measurement Since electrical noise is everywhere it is better to begin each installation with a good instrumentation strategy than try to deal with noise after measurements are compromised Most parts of a good strategy are obvious some are not and even the best plans do not work all the time The Model 370 offers several features that help control noise and maintain signal integrity The ideas and instrum
145. ayzas danin Nad id dA tlre OR ee 4 2 Beadband Example us Lato a Ani A does Es 4 3 Unipolar Versus Bipolar Analog Output SE 5 1 Record Of Zone Settings isea naa a ri 6 1 GPIB Setting COMM UPA ON a a a a a e A a ad PD 6 2 DEV 12 Device Template Configuration 7 1 Model 112 374 Cable Assembly 8 1 Fuse DraWef a aiz iria tara alta datar ids avseevesveuarscosesueteehusnesa ada A ER a aaa 8 2 Power FUSG ACCESS x s Se o za Od ta eZ ALA dao YE CEA dd SO A 8 3 Signal Flow Block Diagram 8 4 Location Of Important Internal Components 8 5 SCAN CONTROL Connector Details 8 6 Resistor Sensor INPUT Connectors 8 7 REFERENCE and MONITOR Connector Details A 8 8 HEATER OUT PUT Connector Details iss a i A A O ORA R AA aa A OO a Wa aa R 8 12 8 9 ANALOG OUT 1 and 2 STILL Connector Details e eee eee aaa aaa aaa aaa aaa aaa aaa carr ncnnnnns 8 12 8 10 ALARM RELAYS Terminal Block 8 11 RS 232 DTE Connector Details sed 8 12 IEEE 488 Rear Panel Connector Details eee eee eeaeeeeaeeeaeeseeeseaeesaeeseaesseeseaeeseaeecaeeseaeseaeeseeeseeeeeesaes 8 15 B 1 Temperature Scale Compas Ae odia B 1 C 1 Typical CryogenicStorage DeWwar ui Ad E eles C 1 LIST OF TABLES Table No Title Page 1 1 Voltage Range ri W AL kai Aina ka lone cio cz w hehe need i bis 3 1 Heater Resistance 4 1 Defa lt Parameter Valles usa OO aida 4 34 5 1 Curve Header P
146. be erased Erase Curve sets all identification parameters to default and blanks all breakpoint values To erase an existing curve press Curve Entry The first curve entry screen appears as a prompt for the curve function Use the A or Y key to select Erase Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next curve entry screen appears as a prompt for the number of the curve to erase The curve name and serial number are displayed below the curve number when available Use the A or V key to select a curve number from 1 to 20 Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next screen appears to make sure curves are not accidentally erased Press the Escape key to cancel or the Enter key to erase the selected user curve and return to the normal display Advanced Operation 5 5 Lake Shore Model 370 AC Resistance Bridge User s Manual 5 2 3 Copy Curve Temperature curves can be copied from one location inside the Model 370 to another This is a good way to make small changes to an existing curve Curve copy may also be necessary if the user needs the same curve with two different temperature limits The curve that is copied from is always preserved NOTE The copy routine a
147. bed in 8 12 2 3 except use the 200 kQ resistance range instead of the highest resistance range available 7 Calibrate the last two current gains of 3 16 pA and 1 pA Remain on 2 UV range and use 600 kQ and 2 MQ resistors for their respective ranges Follow procedure described in Paragraph 8 12 2 3 8 Calibrate frequency compensation as described in Paragraph 8 12 2 4 9 Calibrate scanner compensation as described in Paragraph 8 12 2 5 10 Send the CALSAVE command and wait 10 seconds for the calibration to be saved in non volatile memory Alternatively this step can be done after all ranges are calibrated 8 20 Service Lake Shore Model 370 AC Resistance Bridge User s Manual Model 370 Resistance Input Calibration Work Sheet for Paragraph 8 12 2 6 Step Range Setting Coefficient Resistor Current Filter CALVIN Coefficient 2a 632mV Zero CALZ 2 12 0 10 nA 10s 2b 200mV Zero CALZ2 11 0 3 16 nA 10s 2c 63 2mV Zero CALZ 2 10 0 1nA 10s 2d 20mV Zero CALZ 2 9 0 316 pA 10s 2e 6 32mV Zero CALZ 2 8 0 100 pA 10s 2f 2mV Zero CALZ 2 7 0 31 6 pA 10s 2g 632 HV Zero CALZ 2 6 0 10 pA 10s 2h 200 UV Zero CALZ 2 5 0 3 16 pA 10s 2i 63 2uV Zero CALZ 2 4 0 1pA 50s 2j 20uV Zero CALZ 2 3 0 1pA 50s 2k 632uV Zero CALZ 2 2 0 1pA 100s 21 2uV Zero CALZ 2 1 0 1pA 200s Voltage Zero Offset Coefficient CALVIN Note Lower ranges can be calculated r
148. capacitance after that point can attenuate measurement signals for large resistors On the other hand the times 100 amplifier in the scanner can actually reduce measurement noise in installations that require the Model 370 to be a significant distance from the test Dewar The built in amplifier in the scanner may improve measurements any time long input leads are required even if only one channel is being used Accuracy is not as big a concern as noise when using the Model 3716 L and 3708 Scanners The Model 370 AC measurement eliminates any voltage offsets that may be present in the scanners relays or amplifier The amplifier gain is calibrated at the factory and can be recalibrated in the field to keep scanner gain error well below overall measurement uncertainty Settling time is a concern any time multiple inputs are multiplexed and must be scanned The channel change settling time of the Model 3716L and 3708 Scanners are virtually the same as the range change settling time of the Model 370 Additional time may be required for the instrument to autorange if that feature is enabled and the resistance on the new channel changes significantly between scans 2 8 2 Controlling Temperature While Scanning The close integration of the Model 370 and Model 3716 Scanner permits fast efficient scan channel changes making it possible to both control temperature and take data with the same instrument Firmware in the Model 370 manages automatic switching bet
149. ccccnonccccnoncccnnonccononnncononocnnnnn e vaaia dataa SE Enae da azaa aaa 2 2 2 2 He Cryostat Block Dd riadas 2 3 2 3 Single Stage ADR Block Diagram sadza A OE AAA A 2 4 2 4 Examples or PID Gontroli rc EAN AA ao lo ct el Le rn e 2 20 3 1 Model 370 Rear Peli ti tt A A tes 3 2 3 2 Line Input Assembly 3 3 Resistance Input Connectors erar en a Ad ROA ee A A tia 3 4 3 4 Twisted Shielded Cables ui a a Ai a tdi 3 4 3 5 Guarded Shielded GabIES ra A A O A O OTO 3 5 3 6 Example of Instrument Grounding eee eee eee eee eee aaa aaa aaea aa aiaa aariaa aaa ka neia iaden aaa 3 6 3 7 Model 3716 Scanner Front and Rear Panels ccscsssiecscecceteoscenseccsatenesatccssosscassnecatcesstecesscedeenssteteceucedsersesscodeanensenieensede 3 7 3 8 Model 3716 Scanner Resistor Input Connector e eee eee eee aaa aaa aaa aaa aaa zaawan eaca 3 8 3 9 Model 3716 Scanner to Instrument Measurement Cable ooo eee eee aaa aaa 3 9 3 10 Model 3716 SCAN CONTROL Connector 3 11 SCAN CONTROL Signals Without A Scanner 3 12 Model 3716 Scanner Mounting Bracket Footprint 3 13 Monitor and Reference Connectors r 3 14 Control Hleater Oufput Connector A ee ele eee A GEY noe daa 3 15 Analog Output Connectors zania zdr ATA APE A OE A OA AE EA ie aide dees opra RYSA 3 16 Alarm Relays Terminal Block 3 17 Model RM 1 Rack Mount Kit Assembly gt 4 1 Model 370 Displ
150. cessary to change the frequency to avoid interaction between this instrument and other measurements This is sometimes necessary when two or more bridges are used in the same system The Model 370 offers two alternate nominal frequencies of 9 8 and 16 2 Hz Frequency can be changed through firmware without turning the instrument off The measurement must be allowed to settle before new readings are valid To begin instrument setup press the Instrument Setup key The first instrument setup screen appears as a prompt for the excitation frequency Use the A or Y key to select one of the three excitation frequencies Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display 4 7 2 Active Common Mode Reduction Active common mode reduction is a special feature of the Model 370 that can reduce noise on measurement leads It is most useful when excitation current is 1 UA or greater but should be left on unless one resistor lead is shorted to measurement common refer to Paragraph 3 4 The next instrument setup screen appears as a prompt for active common mode reduction On Off Use the A or Y key to select active common mode reduction on or off Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display 4 7 3 Guard Drive Driven guard
151. cked out Operation 4 7 Lake Shore Model 370 AC Resistance Bridge User s Manual Current Excitation Mode Continued To change the current excitation use the Current Excitation AW keys Available ranges are listed in Table 1 1 As the range table indicates not every combination of current excitation and resistance range is possible The following messages are displayed in the excitation display to indicate limits Maximum Maximum current excitation has already been chosen Minimum Minimum current excitation has already been chosen Decrease Resistance Resistance range must be decreased before changing current excitation Increase Resistance Resistance range must be increased before changing current excitation Resistance ranges have approximately 20 over range but it is recommended that the range be increased when the reading is at full scale for the range to guarantee enough overhead to account for noise voltages Better measurement resolution can be achieved by decreasing the resistance range when the reading is below 20 of full scale To change the resistance range Resistance Range AW keys The following messages are displayed to indicate range limits Maximum Maximum resistance range has already been chosen Minimum Minimum resistance range has already been chosen Decrease Excitation Current excitation must be decreased before changing resistance range Increase Excitation Current excitation must be increased before changing resistance ra
152. cnnnonn conan nn nc nano nro rnnnnnnnns 4 12 4 8 MODEL 3716 SCANNER OPERATION cceesseeeeseeeeeeeneeeeeeeeeeceeaeeeseeeeeeesnaeeeeeeaeeeeeeeeeeeeneeeesenaes 4 12 4 8 1 Chama Om On Off ii tac 4 12 4 8 2 Manual Scanner Channel Selection ooonnnconnnnnnnnnnnoccnonncacnnnnnannnonornco aaa aaa aaa aaa 4 13 4 8 3 Automatic SCAaNNiNGie eee a A A A toba AE P Pad AO EA 4 13 4 8 4 Settings for Multiple Channels eu ue ea a aa aaa aaa 4 13 4 9 FIRMWARE FILTER zee zz NE ON Z W Z a AN O A TT 4 13 4 10 TEMPERATURE MEASUREMENT 2 eeee s eeee easy aaa aaa rra 4 15 4 10 1 CUIVe Sele CON sere pap ao O A AEG AI dA 4 15 4 10 2 Temperature Display suse ei iii 4 15 4 11 TEMPERATURE CONTROL ge oesie Aygo ddd PE o att 4 15 4 11 1 Temperature Control Modes e eee ee ea aaa aaa aaa aaa aaa aaa aaa aaa ennnen enen 4 15 4 11 2 Control Setup ssa AO Siena ne A are nie 4 16 4 11 2 1 setting Control Mode viii a eine nein een epee eee 4 16 4 11 2 2 Control Channel gs ssa aa eb Ee da dada 4 16 4 11 2 3 Setpoint UNIE sec spo ta ad ide 4 17 4 11 2 4 Setpoint Change Pause Time e eee eee aaa eee aec 4 17 4 11 2 5 Comtrol Finer cst a GA EO ORAN OO ASRR ANIE 4 17 4 11 2 6 Heater Output Display caidos 4 18 4 11 2 7 Heater Resistance out 4 18 4 11 2 8 Heater Range Limitadas is Pai oe hn BAe ide aie ss R 4 18 4 11 2 9 Heater Polarity issii a Sayeed vee a ental edb aa 4 19 4 11 3 GhaNGING S
153. conductor heater output common should be terminated at the bulk head connection or other conductive point on the Dewar to provide the single connection between isolated heater output circuitry and the Earth grounded Dewar 3 7 4 Calculating Control Heater Power The control heater output current source on the Model 370 has eight ranges scaled in approximate one third steps Each range will source one tenth the power of the next largest range with the highest range being 1 W The current source voltage compliance is 10 V and limits the heater resistance to 100 Q when operating at full power The restriction on heater resistance is not as severe on lower power ranges and higher heater resistance is often preferred Heater output can be displayed either as percent of full scale current for a range or as power in watts Power calculation requires an accurate heater resistance be entered into the instrument Some examples of heater resistance and output range are given in Table 3 1 Maximum power for a range is the lower of the two P PR and P V R where P is heater power in watts I is the maximum current for the range and V is the compliance voltage of 10 V Heater output resolution is lost any time the compliance voltage limits power If a permanently installed heater has a very high resistance value a resistor may have to be put in parallel with it on the outside of the Dewar to increase the number of useful output ranges and maintain good control
154. connectors provide several shield pins so that individual shields can be connected separately at the scanner end of the cable If the shield screw on the scanner enclosure is strapped the test Dewar the cable shield can be left unterminated at the Dewar end If the scanner enclosure is not connected to the Dewar the shields of the cable should be terminated at the Installation 3 7 Lake Shore Model 370 AC Resistance Bridge User s Manual bulk head connection or other conductive point on the Dewar to provide the single connection between isolated measurement circuitry and the Earth grounded Dewar 25 Pin Female bmp Pin Name Pin Name 1 Shield 14 Shield 2 Shield 15 1 I 3 1 I 16 1 V 4 1 V 17 Shield 5 Shield 18 2 I 6 21 19 2 V 7 2 V 20 Shield 8 Shield 21 3I 9 3 22 3 V 10 3 V 23 Shield 11 Shield 24 4 12 4 25 4 V 13 4 V Body Shield NOTES 1 Guards are not carried through the scanner 2 Shields are connected to isolated measurement common Figure 3 8 Scanner Input Connector 3 5 4 Scanner Output Connector Definition Two 6 pin 240 Circular DIN 45322 receptacles are used for the resistance measurement output from the scanner These are identical to the connectors used on the resistance input of the Model 370 The cables included with the scanner are adequate for most installations Connector pin definition on the scanner is identical to the input connector definition given in
155. cribed below Latching Operation During latching operation if the reading value is ever greater than the high alarm setpoint the high alarm for that channel is latched in as an active state Similarly if the reading value is ever lower than the low alarm setpoint the low alarm for that channel is latched in The alarm remains active until the alarm condition has been cleared and the alarm feature is reset This mode is often used to detect faults in a system or experiment that require operator intervention The alarm remains visible to the operator for diagnostics even if the alarm condition no longer exists Non latching Operation During non latching operation the high alarm remains active only as long as the reading value is greater than the high alarm setpoint Similarly the low alarm remains active only as long as the reading value 1s lower than the low alarm setpoint No action is required from the operator to reset the alarm state after the alarm condition is cleared The deadband parameter is available to prevent the alarm from chattering turning on and off repeatedly when the reading value is very near the setpoint This mode provides a simple form of control that is often actuated through a relay An example of non latching operation is shown below High Meni neliveted x High Alarm Deactivated 100K High Alarm Setpoint Temperature Reading Alarm Latching Off Deadband 5 K 55K Low Alarm Setpoint 50 K Low Alarm Activat
156. ctors 3 8 1 Analog Output Connections Two BNC receptacles provides connections for Analog Voltage Output 1 and 2 with 10 V range 100 mA maximum current into a 100 2 resistor Signal is carried on the center conductor and common is on the outer conductor The outputs are short circuit protected to prevent them from being damaged but operation into a resistance of less than 100 Q 1s not recommended Mating connector not included 3 8 2 Calculating Still Heater Power The heating power sourced by an analog output can be calculated as P V7 R where P is power in watts V is the output in volts and R is the heater resistance in ohms Output voltage ranges from 0 to 10 V and is capable of driving a resistance of 100 Q or greater The maximum power output of 0 to 1 W can be achieved with a 100 heater Most stills do not require the maximum power Using a larger resistor will reduce the range of power and improve setting resolution For example a 1000 Q resistor will produce a 0 to 100 mW power range with full setting resolution If a small resistive heater is permanently installed in the cooling system lower power can be created by adding a series resistor in a heater lead outside the Dewar 3 8 3 Calculating Still Heater Current The current sourced by an analog output can be calculated as I V R where I is current in amps V is the output in volts and R is the heater resistance in ohms Output voltage ranges from 0 to 10 V and is capable of d
157. d can lessen the common mode rejection of a differential input especially when measuring large resistors Second common mode noise sources have the same effect on both sides of the resistor being measured ensuring that common mode voltages are not turned into normal mode voltages voltages are present between measurement leads Common mode voltage is rejected by the differential input but normal mode error voltage is measured along with the desired signal 2 4 6 Measurement Speed and Filtering The Model 370 has built in hardware low pass filter with a time constant of 200 milliseconds That filter is necessary to convert the changing output of the phase sensitive detector operating at 13 7 Hz to a stable DC signal The minimum filtering is sufficient to measure small resistance values with large excitation but that is not the true purpose of the AC Bridge As resistance increases and excitation decreases additional filtering and longer time constants are necessary to maintain good resolution Additional filtering is done in the instrument firmware as a linear average Linear averaging requires more memory than exponential filter algorithms but it gives the best possible settling time Exponential filters implemented in firmware can take several times as long to settle for the same number of filter points Unfortunately linear filters do not have a well behaved frequency response The frequency response of a linear filter has the shape of a damped sine
158. d the instrument will read the control channel between each active scan channel It forces the scan dwell time to five seconds for the control channel and one second for other reading channels 4 11 8 2 Settling Time Settling time on channel change is very important when alternating between control and other readings Longer settling times suspend active control for longer periods of time increasing instability The change pause must be set appropriately for each active channel and the filter should have as short a settling time as possible Filter should be turned off for control 4 12 MAX MIN CAPTURE The Max Min feature automatically captures the largest and smallest valid readings taken since the last reset The feature operates at the full update rate of 10 readings per second It is suspended during range or channel change and captured readings are lost when power is turned off The maximum and minimum readings can be displayed in any of the 8 reading display locations Paragraph 4 4 1 The Model 370 will only capture Max Min data from one data source per channel The Max Min feature uses filtered reading data when the filter is enabled To select a Max Min data source the Max Min key and press Enter to skip past the reset screen If a scanner is installed the first input setup screen appears as a prompt for the channel number If no scanner is installed the sequence will begin below with source selection Use the A or V ke
159. d active common mode reduction circuitry minimize noise and self heating errors With 16 channels IEEE 488 and serial interfaces and closed loop temperature control the Model 370 offers seamless integration with existing cryogenic systems and is the most complete package on the market today Used with Lake Shore calibrated subkelvin resistance temperature sensors it not only measures and displays but also controls temperature for dilution refrigerators and other cryogenic systems Resistance Measurement With the same attention to precision and detail that helped Lake Shore become the world leader in subkelvin temperature sensors the Model 370 AC Resistance Bridge combines a full range of design strategies which optimize resolution and minimize measurement uncertainty in low power resistance measurement The Model 370 uses four lead AC measurement for the best possible accuracy with the lowest possible excitation current AC coupling at each amplifier stage reduces offsets for higher gain and greater sensitivity than DC techniques allow Phase sensitive detection an AC filtering technique used in lock in amplifiers reclaims small measurement signals from environmental noise A low excitation frequency at 13 7 Hz reduces the effect of lead capacitance on measurement These features in conjunction with innovative lead shielding and active noise reduction circuitry significantly reduce measurement noise and resistor self heating The Model 370 curre
160. d by the standard Refer to Paragraph 8 11 2 Cables can be purchased from Lake Shore or other electronic suppliers A connector extender Model 4005 is required to use the IEEE 488 Interface and the ALARM RELAYS terminal block at the same time Cable lengths are limited to 2 meters for each device and 20 meters for the entire bus The Model 370 can drive bus with up to 10 loads If more instruments or cable length is required a bus expander must be used 6 1 1 Changing IEEE 488 Interface Parameters Two interface parameters address and terminators must be set from the front panel before communication with the instrument can be established Other interface parameters can be set via the interface using the device specific commands provided in Paragraph 6 3 Remote Operation 6 1 Lake Shore Model 370 AC Resistance Bridge User s Manual Changing IEEE 488 Interface Parameters Continued To set the IEEE 488 parameters press the Computer Interface key and press Enter to skip past Serial Interface Baud Rate The following computer interface screen appears as a prompt for the IEEE 488 address Use the A or Y key to select an address between 1 and 30 The default is twelve Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next computer interface screen appears as a prompt for the IEEE 488 terminators Use t
161. d for description RAMPST Setpoint Ramp Status Query Input RAMPST term Returned lt ramp status gt term Format n lt ramp status gt 0 Not ramping 1 Setpoint is ramping RDGK Kelvin Reading Query Input RDGK lt channel gt term Format nn lt channel gt Specifies which channel to query 1 16 Returned lt kelvin value gt term Format 9 nnn nnnE nn Remarks Also see RDGST If no curve is present for the channel the returned reading will be zero 6 32 Remote Operation RDGPWR Input Format Returned Format Remarks RDGR Input Format Returned Format Remarks RDGRNG Input Format Example Lake Shore Model 370 AC Resistance Bridge User s Manual Excitation Power Reading Query RDGPRR lt channel gt term nn lt channel gt Specifies which channel to query 1 16 lt excitation power gt term nnn nnnEtnn Also see RDGST Returns the most recent power calculation for the specified channel Resistance Reading Query RDGR lt channel gt term nn lt channel gt lt ohm value gt term Specifies which channel to query 1 16 nnn nnnE nn Also see the RDGST command Resistance Range Command RDGRNG lt channel gt lt mode gt lt excitation gt lt range gt lt autorange gt lt cs off gt term nn n nn nn lt channel gt Specifies which channel to configure 1 16 0 all channels lt mode gt 0 Voltage Excitation Mode Current Excitation Mode l
162. d for precise process control in data handling communication and manufacturing MKSA System of Units A system in which the basic units are the meter kilogram and second and the ampere is a derived unit defined by assigning the magnitude 4m x 107 to the rationalized magnetic constant sometimes called the permeability of space NBS National Bureau of Standards Now referred to as NIST negative temperature coefficient NTC Refers to the sign of the temperature sensitivity For example the resistance of a NTC sensor decreases with increasing temperature National Institute of Standards and Technology NIST Government agency located in Gaithersburg Maryland and Boulder Colorado that defines measurement standards in the United States noise electrical Unwanted electrical signals that produce undesirable effects in circuits of control systems in which they occur normalized sensitivity For resistors signal sensitivity dR dT is geometry dependent i e dR dT scales directly with R consequently very often this sensitivity is normalized by dividing by the measured resistance to give a sensitivity sr in percent change per Kelvin sy 100 R dR dT K where T is the temp in Kelvin and R is the resistance in ohms normally closed N C A term used for switches and relay contacts Provides a closed circuit when actuator is in the free unenergized position normally open N O A term used for switches and relay contacts Provides
163. d to 4 L gt 5 CTS in 8 NC 8 DCD in 6 DSR in 20 DTR out 4 DTR w gt 6 DSR in Model 370 to PC Interface using Null Modem Adapter Model 370 DE 9P Null Modem Adapter PC DE 9P 5 GND IAH A A XX gt 5 GND 2 RD in TTT TD out 3 TD out RR RD in 1 NC M 4 DTR out 6 DSR in ak 1 DCD in 4 DTR out 6 DSR in 7 DTR tiedto4 8 CTS in 8 NC E 7 RTS out 9 NC 9N NOTE Same as null modem cable design except PC CTS is provided from the Model 370 on DTR 8 14 Service Lake Shore Model 370 AC Resistance Bridge User s Manual 8 11 2 IEEE 488 Interface Connector Connect to the IEEE 488 Interface connector on the Model 370 rear with cables specified in the IEEE 488 1978 standard document The cable has 24 conductors with an outer shield The connectors are 24 way Amphenol 57 Series or equivalent with piggyback receptacles to allow daisy chaining in multiple device systems The connectors are secured in the receptacles by two captive locking screws with metric threads The total length of cable allowed in a system is 2 meters for each device on the bus or 20 meters maximum The Model 370 can drive a bus of up to 10 devices A connector extender is required to use the IEEE 488 Interface and Relay Terminal Block at the same time Figure 8 12 shows the IEEE 488 Interface connector pin location and signal names as viewed from the Model 370 rear panel
164. d without loss of measurement accuracy To calibrate the scanner perform the following sequence of operations 1 Remove the calibration access cover on the front panel of the scanner NOTE The scanner does not have to be disassembled for calibration 2 Attach the scanner to a Model 370 3 Attach a near 20 kQ resistor of known value to the scanner input designated channel 1 4 Configure the Model 370 to read Channel 1 continuously on the 20 k range with 20 mV 1 HA excitation and 5 second filter settling time Allow the instrument and scanner to warm up for 1 hour Read and record the resistance reading Change excitation to 6 32 mV 316 nA Adjust the trim pot behind the scanner front panel through the access hole until the resistance reading is the same as recorded in Step 6 E 9 Replace the calibration access cover on the scanner front panel 8 26 Service Lake Shore Model 370 AC Resistance Bridge User s Manual APPENDIX A GLOSSARY OF TERMINOLOGY absolute zero The temperature of 273 16 C or 459 69 F or 0 K thought to be the temperature at which molecular motion vanishes and a body would have no heat energy accuracy The degree of correctness with which a measured value agrees with the true value electronic accuracy The accuracy of an instrument independent of the sensor sensor accuracy The accuracy of a temperature sensor and its associated calibration or its ability to match a standard curve
165. ded The exact resistor value is not important but the value must be known to 0 01 or better to achieve specified reading accuracy The following equipment is required to perform this calibration procedure e Agilent HP 3458A Digital Multimeter or equivalent Service 8 19 Lake Shore Model 370 AC Resistance Bridge User s Manual e Short 12 30cm Calibration Cable to connect voltage and current leads of the Model 370 to resistive loads or a short e Short 12 30cm Calibration Cable to connect voltage and current leads of the Model 3716 Scanner to resistive loads Test leads for the multimeter to connect across the instrument voltage leads e Resistive Loads of the following nominal values known to 0 01 accuracy Nominal Actual Nominal Actual 20 6Q 20 Q 60 QO 200 Q 600 Q 2kQ 6 kQ 20 kQ 60 ka 200 ka 600 kQ 2MQ NOTE Resistor values of 1 3 multiples can be substituted for the above recommended 2 6 multiples but more averaging may be required to achieve the same results Four terminal connections are necessary for lower value resistors To calibrate both voltage and current perform the following sequence of operations The resulting calibration accuracy from this procedure depends only on the accuracy of the calibration resistors being used and not the DMM The order of operations was chosen to keep known resistance values between 2 Q and 2 MQ make calibrations at the
166. display at the same time but one or more of the locations can be assigned to follow the active scan channel Channel numbers 1 to 16 to indicate the individual scanner channels and the word Scan indicates that the display location should follow the active scan channel 4 4 Operation Lake Shore Model 370 AC Resistance Bridge User s Manual Display Setup and Units Selection Continued To configure the display press Display Setup The first display setup screen appears as a prompt for the number of reading locations Use the A or Y key to select a number of reading locations between 1 and 8 Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next display setup screen appears as a prompt for the display location number The channel source and resolution settings that follow apply to the location selected in this step Use the A or Y key to select a location between 1 and 8 Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display If a scanner is attached the channel selection screen will appear to prompt for the input channel that is to appear in the chosen location If no scanner is attached the instrument will skip to source selection Use the A or Y key to select an input channel bet
167. dition first verify that the experimental Dewar case is properly connected to the instrument measurement common and earth ground Paragraph 3 4 4 Next verify the integrity of the measurement lead shields Use of the Common Mode Reduction CMR feature of the Model 370 reduces this effect electronically It is particularly effective at excitation currents above LA 8 4 Service Lake Shore Model 370 AC Resistance Bridge User s Manual 8 7 1 3 VDIF OVL Differential Overload The input stage of the voltage measurement side of the instrument is a very high input impedance instrumentation amplifier with a high gain of 1000 10 at voltage excitations of 20 mV and above If the combination of signal and differential noise components cause the output of this stage to exceed 10 V peak the differential overload error VDIF OVL is displayed The overload occurs at the VDIF point in the instrumentation amplifier circuit Refer to Figure 8 3 for the location of the VDIF monitoring point The signal voltage can be viewed on an oscilloscope by using the Monitor output described in Paragraph 4 7 4 Usual causes of this error include 1 An attempt to measure a resistance that is much greater 2 orders of magnitude than the selected range 2 An open connection of one or both of the voltage measurement leads 3 Significant differential noise on current and or voltage leads To resolve this condition first verify that a reasonable resis
168. documentation Frame or chassis terminal Background color Yellow Symbol and outline Black On supply Off supply Fuse O H d Introduction 1 11 Lake Shore Model 370 AC Resistance Bridge User s Manual This Page Intentionally Left Blank Introduction Lake Shore Model 370 AC Resistance Bridge User s Manual CHAPTER 2 BACKGROUND 2 0 GENERAL This chapter provides background information related to the Model 370 AC Resistance Bridge and bridge measurements in general It is intended to give the user insight into the benefits and limitations of the instrument and help apply the features of the Model 370 to a variety of experimental challenges For information on how to install the Model 370 please refer to Chapter 3 Instrument operation information is contained in Chapter 4 2 1 USEFUL TEMPERATURE RANGE The Model 370 is not an ordinary ohmmeter As described throughout this manual every effort has been made to measure resistance with as little power as possible entering the resistor The need for such careful attention is most evident at low temperatures particularly below 1 K The instrument is not limited to those applications and is useful in applications above 1 K when excitation power is critical When excitation power is not critical however alternatives may be available that provide faster readings or lower cost The lowest temperature at which the Model 370 can make useful resistance measurements
169. e Instructions for the front panel programming features of curve entry and zone entry are in Chapter 5 Computer interface instructions are in Chapter 6 4 1 TURNING POWER ON Verify that the AC line voltage indicator in the fuse drawer window shows the appropriate AC line voltage before turning the instrument on The instrument may be damaged if it is turned on with the wrong voltage selected Instructions for checking line voltage selection are given in Paragraph 3 3 1 When the Model 370 is turned on the display shows the Lake Shore logo for a few seconds and the alarm beeper sounds briefly to indicate the instrument is initializing itself Most of the instrument setup parameter values are retained when power is off with a few exceptions If the instrument was controlling temperature the heater output is turned off All captured Max and Min data is lost If the resistance reading is set to voltage excitation mode and autorange is on the instrument will select the lowest excitation current available at power up then select the optimal range When the instrument is powered on for the first time parameter values are set to their defaults listed in Table 4 1 When initialization is complete the instrument will begin its normal reading cycle and resistance or temperature readings should appear on the display Messages will appear in the reading locations on the display if the measurement input has not been fully configured Messages listed in Paragraph
170. e still requires a few milliwatts of heat from a resistive heater to sustain evaporation The Still Output can drive the still heater Good location for a diagnostic temperature sensor E 3 r One or more heat exchangers serve to cool returning He rich Helium Dewar Vacuum Insulation He Reservoir Chamber Still Heat Exchangers Mixing Chamber Sample Holder Dilution_Block eps Figure 2 1 Dilution Refrigerator Block Diagram liquid to near the temperature of the mixing chamber Also used to heat sink measurement leads Chamber where cooling action takes place Holds both He rich liquid and dilute He separated by a phase boundary When He is depleted from the mix by the still more crosses the phase boundary to maintain equilibrium In doing so it is expanded into diluted by the He creating cooling of the chamber Best location for temperature control sensor and resistive heater The heater output can control temperature by driving the heater under closed loop PID control or open loop control Load Sample Holder End of the mixing chamber where experiments can take place May either be part of or attached to the mixing chamber and it is often designed for a unique purpose Requires a separate temperature sensor if its temperature is different from the mixing chamber 2 2 Theory of Operation 2 3 2 He Cryostat He Cryostats are the least complex systems capable of reaching temperatures down
171. e with straight through cables Paragraph 8 11 1 illustrates suggested cables that can be used between the instrument and common computers The instrument uses drivers to generate the transmission voltage levels required by the RS 232C standard These voltages are considered safe under normal operating conditions because of their relatively low voltage and current limits The drivers are designed to work with cables up to 50 feet in length Remote Operation 6 11 Lake Shore Model 370 AC Resistance Bridge User s Manual 6 2 3 Hardware Support The Model 370 interface hardware supports the following features Asynchronous timing is used for the individual bit data within a character This timing requires start and stop bits as part of each character so the transmitter and receiver can resynchronized between each character Half duplex transmission allows the instrument to be either a transmitter or a receiver of data but not at the same time Communication speeds of 300 1200 or 9600 Baud are supported The Baud rate is the only interface parameter that can be changed by the user Hardware handshaking is not supported by the instrument Handshaking is often used to guarantee that data message strings do not collide and that no data is transmitted before the receiver is ready In this instrument appropriate software timing substitutes for hardware handshaking User programs must take full responsibility for flow control and timing as described in
172. e A or Y key to select Edit Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next curve entry screen appears as a prompt for the number of the curve to edit The curve name and serial number are displayed below the curve number when available Curve name cannot be changed from the front panel Use the A or Y key to select a curve number from 1 to 20 Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next curve entry screen appears as a prompt for the sensor serial number Only numeric serial numbers can be entered from the front panel Use the data entry keys to enter a serial number of up to 10 numbers Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display Advanced Operation 5 3 Lake Shore Model 370 AC Resistance Bridge User s Manual Edit Curve Continued The next curve entry screen appears as a prompt for the curve format Use the A or V key to select a curve format of Q K or Log Q K Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection
173. e User s Manual Entering Zone Parameter Values Continued The next zone setup screen appears as a prompt for the low relay setting This setting 1s only used if the relay is set to zone mode Use the A or Y key to select on or off Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next zone setup screen appears as a prompt for the high relay setting This setting is only used if the relay is set to zone mode Use the A or Y key to select on or off Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next zone setup screen appears as a prompt for the Analog Output 1 manual output value This setting is only used if the analog output is set to zone mode Use the data entry keys to enter a manual output value 100 and 100 in bipolar mode or 0 and 100 in unipolar mode Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a different value Press Escape again to cancel the sequence and return to the normal display The next zone setup screen appears as a prompt for the Analog Output 2 manual output value This setting is only used if the analog output is set to zone mode Us
174. e and advance to the units multiplier Use the A or V key to select a units multiplier of m blank k or M Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display The next alarm setup screen appears as a prompt for the high alarm setpoint value Use the data entry keys to enter the high setpoint between 1x 10 and 100x 10 depending on selected units Enter the numeric value first Press Enter to accept the new value and advance to the units multiplier Use the A or V key to select a units multiplier of m blank k or M Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display Operation 4 27 Lake Shore Model 370 AC Resistance Bridge User s Manual Alarms Continued The next alarm setup screen appears as a prompt for latching or non latching Use the A or Y key to select latching or non latching Press Enter to accept the new selection If latching is selected the deadband screen will be skipped Press Escape to cancel the new selection and return to the normal display The next alarm setup screen appears as a prompt for the alarm deadband value if the alarm is configured for
175. e been chosen for each temperature range or zone the zone feature can automatically select the correct values each time the setpoint is changed This mode does not help choose control parameter values it helps use chosen values more efficiently Once all zone parameters have been programmed the controller must be placed in zone tuning mode 5 3 1 Entering Zone Parameter Values The Model 370 allows the user to establish up to 10 custom contiguous temperature zones The user should configure the zones using zone 1 for the lowest temperature and zone 10 for the highest temperature Zone boundaries are always specified in Kelvin K and the upper limit of each zone acts as the lower limit of the next zone The bottom of the first zone is always 0 K therefore only the upper setpoint limit is required to identify each zones Make a copy of Figure 5 1 to plan your zones before entering Each zone includes control parameters of P I D Manual Output and heater range Manual settings for both analog voltage outputs and both relays can also be controlled with the zone feature if those features are put into zone mode To enter parameter values into the zone table press Zone Setup The first zone setup screen appears as a prompt for the zone number Use the A or V key to select a zone from 1 to 10 Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the norm
176. e controller in closed loop PID mode proportional P integral I and derivative D parameter values must be properly selected to tune the control loop Chapter 2 of this manual describes the principals of closed loop feedback control including some tuning suggestions PID values may need to change any time the control temperature cooling power or heater range change Once parameter values are determined for several temperatures they can be entered into the zone table and recalled automatically as described in Paragraph 5 3 The manual output parameter can also be used during closed loop control but it should be set to zero for normal tuning 4 11 4 1 Manually Setting Proportional P The proportional parameter also called gain is the P part of the PID equation It has a range of 0 001 to 1000 The default value for P is 10 To begin control PID setting sequence press the PID key The first PID setup screen appears as a prompt for a P value Use the data entry keys to enter a proportional value between 0 001 and 1000 Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display 4 11 4 2 Manually Setting Integral l The integral parameter also called reset is the I part of the PID equation It has a setting range of 1 to 10 000 seconds A setting of zero turns I
177. e default of 13 7 Hz Two additional measurement frequencies are available The analog filters in the Model 370 will attenuate slightly more or less of the measured signal at these frequencies so a compensation factor must be programmed into the instrument Only one compensation factor is needed per frequency because the attenuation is related to frequency alone and does not depend on signal amplitude or measurement range To calculate the frequency compensation coefficients perform the following sequence of operations 1 Configure the Model 370 on the 20 Q resistance range with excitation of 632 mV 31 6 mA frequency set to 13 7 Hz and a filter setting of 20 seconds Allow the instrument to warm up for at least 1 hour 2 Attach a nominal load of 20 Q and allow the reading to stabilize 30 to 60 seconds 3 Record the resistance reading 8 18 Service Lake Shore Model 370 AC Resistance Bridge User s Manual 4 Use CALG 5 1 query to read and record the existing 9 8Hz frequency compensation coefficient A safety step to revert to the original calibration if necessary Set the 9 8 Hz frequency compensation coefficient to its default of 1 with the following command CALG 5 1 1 Change frequency to 9 8 Hz Paragraph 4 7 1 and allow the reading to stabilize 60 seconds Record the resistance reading SALON Calculate the 9 8 Hz frequency compensation coefficient as Coefficient 9 8 Hz Resistance 13 7 Hz Resistance 9 8 Hz The
178. e eee aaa aaa aaa nn rra 2 10 2 6 2 Resistor InstallatioN lt 144125 et EEE AEE 2 10 2 6 3 Resistor Self Heating and Thermal Resistance euee eee aaa aaa ana 2 11 2 6 4 LORENA AEON a a R AE A a SEE 2 11 2 6 5 Current Leakage ii ARA ias endo 2 11 2 6 6 Cable Capacitance umi e i 2 11 2 6 7 Instrument Operation maricon 2 12 2 7 OPERATING TRADEOF PS it iii 2 12 2 7 1 Resistor Self heating vs Excitation uuu ea a aaa aaa aan 2 12 2 7 2 Excitation vs Signal to Noise sareti a paea t a e a E a a aneas 2 13 2 7 3 Signal to Noise vs Measurement Bandwidth o ooocconocccinococcnnnoncncnononcncnono nan n non cnn no nr rr nnnr rr 2 13 2 8 MODEL 3716 L and 3708 SCANNER eee eee eee aaa eau aaa aaa aaa aaa anna ran rrarnnnn 2 13 2 8 1 Scanner Performance rota 2 14 2 8 2 Controlling Temperature While Scanning eeeuue esse ea aaa aaa aaa aaa aaa 2 14 2 9 TEMPERATURE MEASUREMENT uu eeeeee eee eau eee enea aaa aaa aaa aaa aaa aaa aaa aaa ena 2 14 2 9 1 Sensor SeleCtIoN ii AR 2 14 Table of Contents Lake Shore Model 370 AC Resistance Bridge User s Manual TABLE OF CONTENTS Continued Chapter Paragraph Title Page 2 9 1 1 SENSIIIVIEY 5 RZE ee cha TR a A aS ta ai aS PERS 2 14 2 9 1 2 Temperature RANGE Gi 2 15 2 9 1 3 Temperature Resolution comic WO iai a Ea 2 15 2 9 1 4 ACCUTA iia OE OLA AE O A AE e OOOO a 2 15 2 9 1 5 PACKAGING awa tw AA 2 15 2 9 1 6 ENVIO Medusa tii
179. e features include matched impedance current source and ground isolation The scanners provide up to four 25 pin D sub connectors for resistance inputs Each connector accommodates four inputs with four signal and two shield lines for each input Guarding is supported between the instrument and scanner only Interface cables from the Model 370 to the scanner are included with the scanner Supported firmware features include temperature conversion math functions linear equations and in some applications temperature control The Model 370 can store measurement range and temperature conversion data for each resistor Appropriate parameter values are automatically recalled with input changes reducing interface overhead and settling time As with any scanner the resistance reading rate is slower during input changes resulting in longer filter settling times and a longer sample period for temperature control The Model 370 can be used with third party scanners however access to integrated features is lost The Model 3716 Scanner The Model 3716 mirrors the single input of the Model 370 that is optimized for low residual DC bias current Low bias provides the lowest available resistor self heating when excitation currents are in the range of 1 pA to 30 pA It also provides the best available accuracy when resistances are above 200 kQ The tradeoff for these performance features is a slightly greater noise figure 33 nV VHz than the Models 3716L and 3708
180. e guards themselves provide some shielding but they are not as effective as shields at reducing induced noise Pin Name 4 Gt 5 6 Shield Pin Name 2 G 1 3 Shield Driven_Guards eps Figure 3 5 Guarded Shielded Cables Typical for Current and Voltage Leads 3 4 4 Isolation and Grounding The Model 370 has two distinct circuit areas that have different grounding needs the chassis grounded circuits and the isolated measurement circuits The control heater output is a third isolated circuit area that should be considered when controlling temperature Chassis grounded circuits include the instrument chassis digital control computer interface and analog voltage outputs Measurement circuits include the current and voltage leads with their associated guards and shield the monitor voltage the reference voltage and scanner control lines Proper installation of the instrument can enhance the benefits of both isolation and shielding Improper installation can defeat both Some trial and error may be required to determine the best configuration A typical grounding scheme is shown in Figure 3 6 Chassis grounded circuits should have a low impedance path to Earth ground for both performance and safety This is achieved by using a 3 conductor power cord in a properly grounded outlet All instrumentation that is connected to the analog outputs or computer interface or use Earth ground as measurement common should have similar low impedance t
181. e locked out To change the voltage excitation use the Voltage Excitation AW keys Available ranges are listed in Table 1 1 As the table indicates not every combination of voltage excitation and resistance range is possible The following messages are displayed in the excitation display to indicate limits Maximum Maximum voltage excitation has already been chosen Minimum Minimum voltage excitation has already been chosen Increase Resistance Resistance range must be increased before changing voltage excitation Decrease Resistance Resistance range must be decreased before changing voltage excitation Resistance ranges have approximately 20 over range but it is recommended that the range be increased when the reading is at full scale for the range to guarantee enough overhead to account for noise voltages Better measurement resolution can be achieved by decreasing the resistance range when the reading is below 20 of full scale To change the resistance range Resistance Range AW keys Available ranges are listed in Table 1 1 The following messages are displayed to indicate range limits Maximum Maximum resistance range has already been chosen Minimum Minimum resistance range has already been chosen Increase Excitation Voltage excitation must be increased before changing resistance range Decrease Excitation Voltage excitation must be decreased before changing resistance range The autorange feature allows the instrument to automatically choose
182. e magnetic field strength H required to reduce the magnetic induction B in a magnetic material to zero from saturation The coercivity would be the upper limit to the coercive force common mode rejection CMR The ability of an amplifier to cancel a common mode signal while responding to an out of phase signal cryogen See cryogenic fluid cryogenic Refers to the field of low temperatures usually 130 F or below as defined by 173 300 f of Title 49 of the Code of Federal Regulations cryogenic fluid A liquid that boils at temperatures of less than about 110 K at atmospheric pressure such as hydrogen helium nitrogen oxygen air or methane Also known as cryogen cryostat An apparatus used to provide low temperature environments in which operations may be carried out under controlled conditions cryotronics The branch of electronics that deals with the design construction and use of cryogenic devices Curie temperature Tc Temperature at which a magnetized sample is completely demagnetized due to thermal agitation Named for Pierre Curie 1859 1906 a French chemist current source A type of power supply that supplies a constant current through a variable load resistance by automatically varying its compliance voltage A single specification given as compliance voltage means the output current is within specification when the compliance voltage is between zero and the specified voltage curve A set of data
183. e new selection and continue If alarm is selected the next setting screen will appear otherwise the instrument will return to the normal display Press Escape to cancel the new selection and return to the normal display If the relay operation is set to alarm mode and a Model 3716 Scanner is installed the next setting screen will appear to prompt for channel number If no scanner is installed this screen will be skipped Use the A or V key to select a channel number between 1 and 16 or Scan Select Scan and the relay will follow the active scan channel Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display If the relay is in alarm mode the next relay setup screen appears as a prompt for an alarm selection Use the A or Y key to select low high or both alarms Press Enter to accept the new selection and return to the normal display Press Escape to cancel the new selection and return to the normal display Operation 4 29 Lake Shore Model 370 AC Resistance Bridge User s Manual 4 16 ANALOG OUTPUTS The Model 370 has two analog voltage outputs designated ANALOG OUT 1 and ANALOG OUT 2 STILL They are variable DC voltage sources that can vary from 10 V to 10 V The voltages are generated from a 16 bit digital to analog converter DAC with a resolution of 0 3 mV or 0 003 of full scale The o
184. e setpoint in whatever units the setpoint is using SETP 1 5 term Setpoint is changed to 1 5 Kelvin if control units are Kelvin SETP 2 0E 03 term Setpoint is changed to 2 0 kQ if control units are ohms Temperature Control Setpoint Query SETP term lt value gt term tnnn nnnE nn Still Output Command STILL lt value gt term nnnnnn lt value gt Specifies value for the Still output in percent STILL 22 45 term Still output is 22 45 of full scale Command only valid if Still is properly configured in Analog Output 2 Still Output Query STILL term lt value gt term nnnnnn Refer to command for description Remote Operation 6 35 Lake Shore Model 370 AC Resistance Bridge User s Manual ZONE Zone Table Parameter Command Input ZONE lt zone gt lt setpoint limit gt lt P value gt lt I value gt lt D value gt lt mout value gt lt htr range gt lt relay low gt lt relay high gt lt analog 1 gt lt analog 2 gt term Format nn tnnn nnnE nn nnnnnn nnnnnn nnnnnn nnnnnn n n n Annnnnn nnnnnn lt zone gt Specifies which zone in the table to configure 1 10 lt top value gt Specifies the setpoint limit of this zone lt P value gt Specifies the P for this zone 0 001 1000 lt I value gt Specifies the I for this zone 0 10000 lt D value gt Specifies the D for this zone 0 10000 lt mout value gt Specifies the manual ou
185. e the data entry keys to enter a manual output value 100 and 100 in bipolar mode or 0 and 100 in unipolar mode Press Enter to accept the new selection and return to the normal display Press Escape to restart the setting sequence and enter a different value Press Escape again to cancel the sequence and return to the normal display Repeat this process for up to ten zones Not all zones have to be used but they must start at Zone 1 with no blank zones in the middle of the table Advanced Operation 5 9 Lake Shore Model 370 AC Resistance Bridge User s Manual 5 3 2 Using Zone Tuning Once all zone parameters have been programmed the controller must be placed in zone tuning mode When in zone tuning mode the instrument will update the control settings each time the setpoint is changed to a new zone If the settings are changed manually the controller will use the new setting while it is in the same zone and update to the zone table settings when the setpoint is changed to a value outside that zone If a setpoint value is entered that is greater than the setpoint limit in zone 10 the instrument will use the parameter values from zone 10 Select closed loop PID control mode to turn zone tuning off To select zone tuning mode press Control Setup The first control setup screen appears as a prompt for the control mode Use the A or Y key to select Zone Press Enter to accept the new selection and continue to the next sett
186. e was aware of a trademark claim they appear with initial capital letters and the or symbol CalCurve Cernox Duo Twist Quad Lead Quad Twist Rox and SoftCal are trademarks of Lake Shore Cryotronics Inc MS DOS and Windows are trademarks of Microsoft Corp NI 488 2 is a trademark of National Instruments PC XT AT and PS 2 are trademarks of IBM Copyright 2001 2009 by Lake Shore Cryotronics Inc All rights reserved No portion of this manual may be reproduced stored in a retrieval system or transmitted in any form or by any means electronic mechanical photocopying recording or otherwise without the express written permission of Lake Shore Lake Shore Model 370 AC Resistance Bridge User s Manual CE DECLARATION OF CONFORMITY We Lake Shore Cryotronics Inc 575 McCorkle Blvd Westerville OH 43082 8888 USA hereby declare that the equipment specified conforms to the following Directives and Standards Application of Council Directives cccccscsscsesssscesessescssessceccsceseaneees 73 23 EEC 89 336 EEC Standards to which Conformity is declared c cccccccsesesssssseseseseseseees EN61010 1 2001 Overvoltage II Pollution Degree 2 EN61326 A2 2001 Class A Annex B Model Number EE PE AEEA EE E EAE E E 370 El ZZ 7 200H Ed Maloof Printed Name Vice President of Engineering Position Lake Shore Model 370 AC Resistance Bridge User s Manual El
187. eat reflective used to hold liquid gases and to study low temperature phenomena Invented by Sir James Dewar 1842 1923 a Scottish physical chemist differential permeability The slope of a B versus H curve ua dB dH differential susceptibility The slope of a M versus H curve X dM dH digital controller A feedback control system where the feedback device sensor and control actuator heater are joined by a digital processor In Lake Shore controllers the heater output is maintained as a variable DC current source digital data Pertaining to data in the form of digits or interval quantities Contrast with analog data dimensionless sensitivity Sensitivity of a physical quantity to a stimulus expressed in dimensionless terms The dimensionless temperature sensitivity of a resistance temperature sensor is expressed as Sa T R dR dT which is also equal to the slope of R versus T on a log log plot that is Sa d InR d InT Note that the absolute temperature in Kelvin must be used in these expressions drift instrument An undesired but relatively slow change in output over a period of time with a fixed reference input Note Drift is usually expressed in percent of the maximum rated value of the variable being measured dynamic data exchange DDE A method of interprocess communication which passes data between processes and synchronized events DDE uses shared memory to exchange data between applications and a protocol to s
188. eater to the lowest range and write down the temperature rise if any Select the next highest heater range and continue the process until the load warms up through its operating range Do not leave the system unattended the heater may have to be turned off manually to prevent overheating If the load never reaches the top of its operating range some adjustment may be needed in heater resistance or an external power supply may be necessary to boost the output power of the instrument The list of heater range versus load temperature is a good reference for selection the proper heater range It is common for systems to require two or more heater ranges for good control over their full temperature Lower heater ranges are normally needed for lower temperature The Model 370 is of no use controlling at or below the temperature reached when the heater was off 2 10 3 2 Tuning Proportional The proportional setting is so closely tied to heater range that they can be thought of as fine and course adjustments of the same setting An appropriate heater range must be known before moving on to the proportional setting Begin this part of the tuning process by letting the cooling system cool and stabilize with the heater off Place the Model 370 in closed loop PID control mode then turn integral derivative and manual output settings off Enter a setpoint above the cooling systems lowest temperature Enter a low proportional setting of approximately 5 or 10 and t
189. ected the next setting screen will appear to allow the entry of a value for B Use the A or V key to select a source for B as either Value SP or SP Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next linear setup screen appears as a prompt for an offset value B if value was selected on the previous screen Use the data entry keys to enter an offset between 1 x 10 and 100 x 10 depending on selected units Enter the numeric value first Press Enter to accept the new value and advance to the units multiplier LN is the source designator used on the display for the linear equation Use the A or Y key to select a units multiplier of m blank k or M Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and retum to the normal display The offset screen is the last screen in the linear equation setup sequence if temperature control is enabled Operation 4 25 Lake Shore Model 370 AC Resistance Bridge User s Manual Linear Equation Continued A setpoint setting screen appears in the sequence if temperature control is disabled because the Setpoint key is locked out Use the data entry keys to enter a setpoint value Enter the numeric value fir
190. ectromagnetic Compatibility EMC for the Model 370 AC Resistance Bridge Electromagnetic Compatibility EMC of electronic equipment is a growing concern worldwide Emissions of and immunity to electromagnetic interference is now part of the design and manufacture of most electronics To qualify for the CE Mark the Model 370 meets or exceeds the requirements of the European EMC Directive 89 336 EEC as a CLASS A product A Class A product is allowed to radiate more RF than a Class B product and must include the following warning WARNING This is a Class A product In a domestic environment this product may cause radio interference in which case the user may be required to take adequate measures The instrument was tested under normal operating conditions with sensor and interface cables attached If the installation and operating instructions in the User s Manual are followed there should be no degradation in EMC performance This instrument is not intended for use in close proximity to RF Transmitters such as two way radios and cell phones Exposure to RF interference greater than that found in a typical laboratory environment may disturb the sensitive measurement circuitry of the instrument Pay special attention to instrument cabling Improperly installed cabling may defeat even the best EMC protection For the best performance from any precision instrument follow the grounding and shielding instructions in the User s Manual In addition t
191. ed Low Alarm Deactivated Deadband eps Figure 4 2 Deadband Example 4 26 Operation Lake Shore Model 370 AC Resistance Bridge User s Manual Alarms Continued To begin alarm setup press the Alarm key and press Enter to skip past the reset screen If a scanner is installed the alarm setup screen appears as a prompt for the channel number All of the parameter changes that follow will act on the selected channel If no scanner is installed the sequence will begin below with source selection Use the A or Y key to select a channel number between 1 and 16 Choose the selection All and the changes will be applied to all channels at the same time Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next alarm setup screen appears as a prompt for the data source Use the A or V key to select one of three data sources ohms Kelvin or linear result of the linear equation Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next alarm setup screen appears as a prompt for the low alarm setpoint value Use the data entry keys to enter the low setpoint between 1x 10 and 100x 10 depending on selected units Enter the numeric value first Press Enter to accept the new valu
192. ed in the creation or production of this firmware can pay for loss of time inconvenience loss of use of the product or property damage caused by this product or its failure to work or any other incidental or consequential damages Use of our product implies that you understand the Lake Shore license agreement and statement of limited warranty FIRMWARE LICENSE AGREEMENT The firmware in this instrument is protected by United States copyright law and international treaty provisions To maintain the warranty the code contained in the firmware must not be modified Any changes made to the code is at the user s risk Lake Shore will assume no responsibility for damage or errors incurred as result of any changes made to the firmware Under the terms of this agreement you may only use the Model 370 firmware as physically installed in the instrument Archival copies are strictly forbidden You may not decompile disassemble or reverse engineer the firmware If you suspect there are problems with the firmware return the instrument to Lake Shore for repair under the terms of the Limited Warranty specified above Any unauthorized duplication or use of the Model 370 firmware in whole or in part in print or in any other storage and retrieval system is forbidden TRADEMARK ACKNOWLEDGMENT Many manufacturers and sellers claim designations used to distinguish their products as trademarks Where those designations appear in this manual and Lake Shor
193. ed increase in load generally from no load to full load unless otherwise specified lock in amplifier An amplifier that uses some form of automatic synchronization with an external reference signal to detect and measure very weak electromagnetic radiation at radio or optical wavelengths in the presence of very high noise levels M Symbol for magnetization See magnetization magnetic air gap The air space or non magnetic portion of a magnetic circuit magnetic field strength H The magnetizing force generated by currents and magnetic poles For most applications the magnetic field strength can be thought of as the applied field generated for example by a superconducting magnet The magnetic field strength is not a property of materials Measure in SI units of A m or cgs units of oersted magnetic flux density B Also referred to as magnetic induction This is the net magnetic response of a medium to an applied field H The relationship is given by the following equation B Ho H M for SI and B H 47M for cgs where H magnetic field strength M magnetization and uo permeability of free space 4n x 10 H m magnetic hysteresis The property of a magnetic material where the magnetic induction B for a given magnetic field strength H depends upon the past history of the samples magnetization magnetic induction B See magnetic flux density magnetic moment m This is the fundamental magnetic property measured with dc magnet
194. ed loop PID control For the feature to have a beneficial effect the ramp rate must be slower than the time response of the load The ramp feature will only operate if the setpoint units are Kelvin The ramp rate is set in Kelvin per minute with a range of 0 001 to 10 and a resolution of 0 001 Once the feature is turned on and a rate is set ramp action is initiated by a setpoint change When a new setpoint value is entered the instrument changes the setpoint from the old value to the new value in small increments that correspond to the ramp rate The ramp rate is entered as a positive number but it can be used for ramping up and down in temperature When the ramp feature is turned on the Ramp annunciator will be shown on the top line of the display During an active ramp the annunciator will blink When the instrument is waiting for a setpoint change the annunciator is steady If the setpoint units are set to ohms or the heater range is turned off the ramp annunciator will remain steady and the setpoint will go directly to the new value To configure the ramp feature press the Control Setup key and press Enter until the following display appears as a prompt for ramp on off Use the A or Y key to choose ramp on or off Press Enter to accept the new selection and continue to the next setting screen If the ramp feature is set to off the ramp rate parameter will be skipped Press Escape to cancel the new selection and return to the norm
195. ed to drive isolated or differential inputs Please note that most line powered oscilloscopes are Earth ground referenced and can create the exact problems that they are being used to diagnose MONITOR REFERENCE MonitorRefOutputs bmp Figure 3 13 Monitor and Reference Connectors 3 6 1 Monitor Voltage Output BNC receptacle provides connections for Monitor Voltage Output Signal is carried on the center conductor and isolated measurement common is on the outer conductor The signal can be switched internally to one of several diagnostic voltage test points It has a maximum voltage range of 15 V and should be used to source no more than 1 mA of current Mating connector is not included 3 6 2 Reference Voltage Output BNC receptacle provides connections for Frequency Reference Output Signal is carried on the center conductor and isolated measurement common is on the outer conductor The signal is a TTL level digital voltage synchronized to the measurement frequency that can be used for synchronous lock in detection of AC measurement signals The output maintains TTL logic voltage levels when driving no more than 2 TTL loads Mating connector is not included 3 7 CONTROL HEATER OUTPUT The temperature control heater output is a controlled DC current source capable of generating a maximum of 1 W of power into a resistive heater that is attached to the cooled load During control the heater output is adjusted either manually or through
196. el 370 provides the appropriate voltage gain when the resistance range and a fixed excitation current are selected The instrument changes gain when the resistance range is changed In current mode the actual excitation current is continuously displayed 1 2 Introduction Lake Shore Model 370 AC Resistance Bridge User s Manual In voltage mode the Model 370 firmware simulates voltage excitation In voltage mode the instrument changes the current to maintain the voltage limit when the resistance range is changed Because voltage is limited in voltage mode excitation power decreases as resistance increases and temperature decreases when negative temperature coefficient NTC resistance materials are measured As a result voltage mode provides a convenient way to limit excitation power and resistor self heating for NTC resistance temperature sensors In voltage mode the upper input voltage limit but not the actual voltage is continuously displayed Autorange and Manual Range The Model 370 autorange and manual range selection functions of the Model 370 are available for use with both current excitation mode and voltage mode The autorange function increases or decreases the resistance range when measured resistance exceeds or falls below the range in use In current mode the autorange function modifies the voltage gain In voltage mode autorange modifies the current setting Manual range selection provides the option of full user control Wi
197. ent features described in this manual often have to be tried in different combination to get the best results Specific recommendations on installation and lead wiring are given in Chapter 3 2 6 Theory of Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 2 4 5 1 Shielding and Grounding Theory Begin any installation by separating noise sources from their targets Sources of E field and H field noise can be just about anything but the targets are most often signal leads and other wires and cables H field coupling can be dramatically reduced with an inch or less of separation E field noise requires more separation but is also more easily shielded Noise can also travel easily on the power lines If clean power lines are not available for instrumentation place line filters on noisy electrical or electronic devices The line filter in the instrument can help but noise is always better reduced at its source The best defense against E field noise is a conductive shield around the measurement The more complete the shield the better Ideally 1t should extend from the instrument measurement circuits through the cables and down to the measured resistor This ideal approach is impractical in all but a few cryogenic systems The practical approach is to connect the shield from the measurement circuits through the room temperature cables to the conductive Dewar or vacuum chamber This creates a faraday shield around the entire measuremen
198. er Scanner Output to Resistance Input and Control Cable 10 foot Ground Strap for Scanner Enclosure Scanner Input Mating Connectors A M R Scanner Mounting Bracket Included only when purchased with VAC 120 ALL Power Option Installation 3 1 Lake Shore Model 370 AC Resistance Bridge User s Manual 3 2 REAR PANEL DEFINITION This paragraph defines the rear panel of the Model 370 See Figure 3 1 Readers are referred to paragraphs that contain installation instructions and connector pin outs for each feature Definition of the Model 3716 Scanner is given in Paragraph 3 5 A summary of connector pin outs is provided in Paragraph 8 11 CAUTION Verify that the AC line voltage indicator in the fuse drawer window shows the appropriate AC line voltage before turning he instrument on CAUTION Make rear panel connections with the instrument power off Includes the IEC 320 C14 line cord receptacle instrument power switch and line O Einelnpur Assembly voltage selector with line voltage indicator and line fuse holder Refer to Paragraph 3 3 RS 232 DTE 9 pin D subminiature plug wired in DTE configuration for use with RS 232C serial computer interface Refer to Paragraph 6 2 and see Figure 8 11 Two 6 pin 240 Circular DIN 45322 receptacles for resistance measurement input INPUT Refer to Paragraph 3 4 for use with one resistor or Paragraph 3 5 for use with the Model 3716 Sc
199. er s Manual MOUT Manual Heater Output Query Input MOUT term Returned lt value gt term Format nnn nnnEtnn Refer to command for description PID Closed Loop PID Parameter Command Input PID lt P value gt lt I value gt lt D value gt term Format nnnnnn Ennnnnn Ennnnnn lt P value gt The value for control loop Proportional gain 0 001 to 1000 lt I value gt The value for control loop Integral reset O to 10000 lt D value gt The value for control loop Derivative rate O to 2500 Remarks Setting resolution is less than 6 digits indicated Example PID 10 100 20 term P 10 I 100 seconds and D 20 seconds PID Closed Loop PID Parameter Query Input PID term Returned lt P value gt lt I value gt lt D value gt term Format mnnnnn nnnnnn nnnnnn Refer to command for description RAMP Setpoint Ramp Parameter Command Input RAMP lt off on gt lt rate value gt term Format n Ennnnn lt off on gt Specifies whether ramping is on or off 0 Off 1 On lt rate value gt Specifies setpoint ramp rate in Kelvin per minute from 0 001 to 10 The rate is always positive but will respond to ramps up or down Example RAMP 1 1 5 term When the temperature control setpoint is changed it ramps to the target at 1 5 K minute RAMP Setpoint Ramp Parameter Query Input RAMP term Returned lt off on gt lt rate value gt term Format n nnnnn Refer to comman
200. er value cannot be changed Control OFF Temperature control it turned off and control parameter values cannot be accessed Still Not Selected Analog Output 2 is not configured as still heater output NO CURVE No curve is selected for the input so readings cannot be displayed in temperature Negative Resistance Voltage and current lead polarity does not match 8 7 TROUBLESHOOTING It is likely through the course of installation and usage that the Model 370 will meet conditions that cannot yield a valid resistance reading These conditions can usually be traced to environmental noise or misapplied measurement parameters The instrument has been designed with features to simplify diagnosis of problems arising from these sources 8 7 1 Overload Monitoring Atall times there are four analog circuit areas of interest monitored by the firmware of the instrument to ensure that the measurement signals stay within useful limits If any signal exceeds its limit a descriptive overload message is displayed An explanation of the error messages and likely causes follows 8 7 1 1 CS OVL Current Source Overload This message is displayed in the measurement window anytime the current source drive signal s into the current sample resistor s exceeds a 10 V peak limit The overload can occur at either CSPOS or CSNEG points in the current source circuit Refer to Figure 8 3 for the location of CSPOS and CSNEG monitoring points
201. erature Control Setup Parameter Command Input CSET lt channel gt lt filter gt lt units gt lt delay gt lt current power gt lt htr limit gt lt htr resistance gt term Format nn n n nnn n n Ennn nnn lt channel gt Specifies which channel to control from 1 16 lt filter gt Specifies controlling on unfiltered or filtered readings 0 unfiltered 1 filtered lt units gt Specifies setpoint units 1 Kelvin 2 Ohms lt delay gt Delay in seconds for setpoint change during Autoscanning 1 255 lt current power gt Specifies heater output display in current or power 1 current 2 power lt htr limit gt Maximum heater range 1 8 Refer to HTRRNG for detailed range listing lt htr resistance gt Heater load in ohms 1 100 000 Example CSET 16 0 1 30 1 8 100 term The Model 370 controls from channel 16 unfiltered readings The setpoint is in Kelvin with a 30 second setpoint change pause time The heater output is displayed in current The heater maximum range limit is set to 100 mA and the heater load is set to 100 Q CSET Temperature Control Setup Parameter Query Input CSET term Returned lt channel gt lt filter gt lt units gt lt delay gt lt current power gt lt htr limit gt lt htr resistance gt term Format nn n n nnn n n Ennn nnn Refer to command for description DFLT Factory Defaults Command Input DFLT 99 term Remarks Sets all configuration values to factory defaults and resets the instr
202. errous metal softer than lead and extremely malleable and ductile It stays soft and workable down to cryogenic temperatures Indium can be used to create solder bumps for microelectronic chip attachments and also as ID 10 XX gaskets for pressure and vacuum sealing purposes ID 10 31 Indium Disks are 0 312 diameter x 0 005 inches ID 10 56 Indium Disks are 0 562 diameter x 0 005 inches Indium Foil Sheets Quantity 5 When used as a washer between temperature sensors and IF 5 refrigerator cold stages indium foil increases the thermal contact area and prevents the sensor from detaching due to vibration It also may be used as a sealing gasket for covers flanges and windows in cryogenic applications Each sheet is 0 005 x 2 x 2 inches 7 2 Options and Accessories Lake Shore Model 370 AC Resistance Bridge User s Manual Apiezon H Grease 25 gram Tube It is designed for general purposes where operating GAH 25 temperatures necessitate the use of a relatively high melting point grease Melting point is 523 K 250 C Can be removed using Xylene with an Isopropyl alcohol rinse Apiezon N Grease 25 gram Tube General purpose grease well suited for cryogenic use because of its low viscosity It is often used as a means of thermally anchoring cryogenic sensors as well as lubricating joints and o rings Contains high molecular weight polymeric hydrocarbon additive which gives it a tenacious rubbery consiste
203. ersion Supported Sensors Any resistive sensor including NTC resistors e g Germanium Carbon Glass Cernox Ruthenium Oxide Rox and PTC resistors e g Rhodium lron RTD Requirements Requires calibrated sensor and a temperature response curve loaded into the instrument at the factory or by the user Temperature Coefficient Negative or Positive Temperature Units Kelvin Low Temperature Below 20 mK in a well designed system Temperature Resolution Sensor and temperature dependent Refer to Paragraph 1 2 Curve Memory Space for twenty 200 point curves Curve Entry Entered via front panel computer interface or CalCurve option Curve Format Ohm K Log Ohm K Temperature Control Control Type High resolution digital PID Control Modes Closed Loop PID Open Loop Tuning Modes Manual PID Zones Setpoint Units Ohms or Kelvin with temperature calibration curve Setpoint Resolution Same as reading display resolution Control Stability Below 10 mK p p at 50 mK system dependent Heater Output Type Variable DC current source Heater Output Isolation Optically isolated from chassis and measurement ground Output D A Resolution 18 bit Heater Output Ranges 100 mA 31 6 mA 10 mA 3 16 mA 1 mA 316 HA 100 HA 31 6 UA Heater Output Compliance 10 V Max Power of Output Ranges 1 W 100 mW 10 mW 1 mW 100 uW 10 uW 1 UW 0 1 uW nominal 100 Heater Type Resistive Heater Resistance Range 1 100 kQ 1
204. es RELAY Relay Parameter Query Input RELAY lt relay number gt term Format n lt relay number gt Specifies which relay to query 1 2 Returned lt mode gt lt channel alarm gt lt alarm type gt term Format nnnn Refer to command for description RELAYST Relay Status Query Input RELAYST lt relay number gt term Format n lt relay number gt Specifies which relay to query 1 Low relay 2 High relay Returned lt status gt term Format n 0 Off 1 On 6 34 Remote Operation SCAN Input Format Example Remarks SCAN Input Returned Format SETP Input Format Example SETP Input Returned Format STILL Input Format Example Remarks STILL Input Returned Format Lake Shore Model 370 AC Resistance Bridge User s Manual Scanner Parameter Command SCAN lt channel gt lt autoscan gt term nn n lt channel gt Specifies which channel to switch the scanner to 1 16 or 1 8 if a Model 3708 scanner is installed lt autoscan gt 0 autoscan feature off 1 autoscan feature on SCAN 5 1 term Change the scanner to channel 5 and begin autoscanning This command is only active if an optional scanner is installed Scanner Parameter Query SCAN term lt channel gt lt autoscan gt term nnn Refer to command for description Temperature Control Setpoint Command SETP lt value gt term nnn nnnE tnn lt value gt The value for th
205. es analog output channel 1 2 lt bipolar enable gt lt mode gt lt channel gt lt source gt lt high value gt lt low value gt lt manual value gt term 00 00 nnn nnnE nn nnn nnnEtnn tnnnnnn Refer to command for definition Analog Output Data Query AOUT lt analog channel gt term n lt analog channel gt Specifies analog output channel 1 2 lt analog output gt term tnnnnnn Refer to command for definition Returns the percentage of output of the analog control RS 232 Baud Rate Command BAUD lt bps gt term n lt bps gt Specifies Baud rate 0 300 Baud 1 1200 Baud 2 9600 Baud RS 232 Baud Rate Query BAUD term lt bps gt term n Refer to command for description Audible Alarm Beeper Command BEEP lt state gt term n lt state gt 0 Off Audible alarm beeper remains off 1 On Audible alarm beeper is on when any alarm is active Enables or disables system beeper sound when an alarm condition is met Audible Alarm Beeper Query BEEP term lt state gt term n Refer to command for description Remote Operation 6 23 Lake Shore Model 370 AC Resistance Bridge User s Manual BRIGT Display Brightness Command Input BRIGT lt bright gt term Format n lt bright gt Specifies display brightness 0 25 1 50 2 75 3 100 Default 2 BRIGT Display Brightness Query Input BRIGT term Returned lt bright gt term Format n Refer to c
206. etPOIN tes see iaa 4 19 4 11 4 Manually Setting PID Vales x u az ve ee ee i ee A ee 4 20 4 11 4 1 Manually Setting Proportional P ooonnoncnnnnncccnnnanccononananononnncnon cnn n nono cnn nan aaa aaa aaa 4 20 4 11 4 2 Manually Setting Integral l cc cccceeeeeeeeceeeeeeeeeeeaeaeeeeeeeeeceaeaeceeeeeseseeaaeaeeeeeeesecaeeeeeeseeeeeaees 4 20 4 11 4 3 Manually Setting Derivative D oooononccinnnccononcccnnnoncnononancnononcnc aaa nono cnn nan aaa aaa aaa 4 20 4 11 5 Manual Heater Output 50 etacie O Ga oe EP AP aska Edek 4 21 4 11 6 Heater Output Range and Heater Output OfF eee eeeeaa aaa ae aaa aaa aaa aaa aaa aaa 4 21 4 11 7 Setpoint Raimi A aai 4 22 4 11 8 Controlling and Scanning ocio Ai daba 4 23 4 11 8 1 Reading SEQUENCE iia 4 23 4 11 8 2 Settling Timesis a ea e picas 4 23 4 12 MAXMIN CAPTURE ita ios 4 23 4 13 LINEAR EQUATION AR aiani nne ic 4 24 4 14 ALARM O ads 4 26 4 15 RELAYS coronada aiii 4 29 4 16 ANALOG QUTPUTS e a SE a EE a et wi 4 30 4 17 STILL OUTPUT A tas 4 32 4 18 LOCKING THE KEYPAD cocina tit ci dt ad ass 4 33 4 19 DEFAULT PARAMETER VALUES c ow wos as oez owsa ide eon iiae dn oai aparea Nidaa 4 33 5 ADVANCED OPERATION ccoo cia dida 5 1 5 0 GENERAL soii aia ape daa cb 5 1 5 1 CURVE NUMBERS AND STORAGE eee ee ea ae eee ea aaa aaa nan rr rr naar nn nan n nn nnnnr rr ennnn ns 5 1 5 1 1 Curve Header Parameters ecccscecsseeecetececeresseneeseseeeensaseceessuaceesenn
207. etic salt that acts as the refrigerant It is magnetized while in thermal contact with the pre cooling stage so the heat of magnetization is removed and entropy is reduced Then it is demagnetized while thermally isolated and as field is decreased the salt must decrease in temperature to maintain constant entropy Best location for temperature control sensor The heater output can control temperature by trimming current in the magnet with its heater output current source Resistive heating is not desirable Load Sample Holder End of the final cooling stage where experiments can take place Requires a separate temperature sensor if its temperature is different from the final stage 2 4 Theory of Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 2 4 MODEL 370 THEORY OF OPERATION 2 4 1 Why is the Model 370 called a Bridge Historically precision resistance measurements were made using one of several bridge configurations that are a combination of known reference resistors and the unknown being measured Normally in a bridge one of the known resistors is trimmed until a null appears at the measurement point of the circuit and at that time the unknown resistance can be calculated The technique is preferred because it relies only on resistance ratios and a null measurement which eliminates many of the errors normally associated with amplifiers voltage references and current sources As electronic circuits improve fewer and fewer
208. ew selection and continue to the bipolar unipolar setting screen Press Escape to restart the setting sequence and enter a different value Press Escape again to cancel the sequence and return to the normal display If input mode is selected and a scanner is installed the next analog output setup screen appears as a prompt for the input channel If no scanner is installed the sequence skips to source selection Use the A or Y key to select a channel number between 1 and 16 Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next analog output setup screen appears as a prompt for input data source Use the A or Y key to select one of three data sources ohms Kelvin or linear result of the linear equation Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next analog output setup screen appears as a prompt for the low endpoint This example illustrates the low endpoint setting sequence for unipolar operation Use the data entry keys to enter the endpoint value between 1 x10 and 100x 10 depending on selected units Enter the numeric value first Press Enter to accept the new value and advance to the units multiplier Use the A or Y key to select a units multiplier of m blank k or M
209. f is recommended when the control system is seldom changed and data is taken when the load is at steady state The derivative setting is entered into the Model 370 in seconds and a good starting point is one fourth the integral setting in seconds Again do not be afraid to make some small setpoint changes halving or doubling this setting to watch the affect Expect positive setpoint changes to react differently from negative setpoint changes 2 22 Theory of Operation Lake Shore Model 370 AC Resistance Bridge User s Manual CHAPTER 3 INSTALLATION 3 0 GENERAL This chapter provides general installation instructions for the Model 370 AC Resistance Bridge Please read this entire chapter before installing the instrument and powering it on to ensure the best possible performance and maintain operator safety For instrument operating instructions refer to Chapter 4 For computer interface installation and operation refer to Chapter 6 3 1 INSPECTION AND UNPACKING Inspect shipping containers for external damage before opening them Photograph any container that has significant damage before opening it If there is visible damage to the contents of the container contact the shipping company and Lake Shore immediately preferably within 5 days of receipt of goods Keep all damaged shipping materials and contents until instructed to either return or discard them Open the shipping container and keep the container and shipping materials until
210. ft change in reading with change in ambient temperature is small in the Model 370 and seldom a significant factor in absolute accuracy Measurement repeatability is often dominated by temperature drift and temperature change should be considered if the instrument is being installed in a rack If possible precision measurement instruments should not be placed in the same rack with large power supplies or other equipment that can change the rack temperature quickly Instrument racks should be well ventilated to prevent the build up of heat 2 6 2 Resistor Installation Installation is an important part of resistor measurements at cryogenic temperatures The measured resistor must be at the temperature of its surroundings for the measurement to be valid Some installation suggestions for temperature sensors are given in Paragraph 2 9 3 These suggestions are applicable to any resistor 2 10 Theory of Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 2 6 3 Resistor Self Heating and Thermal Resistance A resistor self heats when the electrical power being put into the measured resistor is large enough to warm the resistor above the temperature of its surroundings Electrical power gets into the resistor from both the necessary excitation current and undesirable noise Paragraph 2 5 Errors from self heating are dangerous because they can easily go undetected One strategy for detecting these errors is to understand the thermal resi
211. g measurement bandwidth can help with nearly every continuous noise source including induced noise at discrete frequencies 1 f circuit noise and broadband thermal noise Unfortunately there is a price to pay and it is time AII small signal measurement instruments include some type of bandwidth limiting to improve resolution The band limiting capabilities are so closely tied to instrument performance that most resolution specifications are given at a specified time constant or bandwidth The Model 370 uses a combination of analog band pass and low pass filters and digital averaging The phase sensitive detector PSD in the Model 370 is a very narrow band pass filter The output of the PSD is run through an analog low pass filter with 200 ms time constant to give an equivalent DC voltage The user cannot change these two filters making the widest bandwidth possible on the Model 370 200 ms Once the DC voltage is digitized the user can choose additional digital averaging to reduce bandwidth further and increase resolution Unfortunately reducing bandwidth increases time constant and settling time slows the instrument response to real changes in resistance because those changes are averaged along with the noise Probably the most frustrating aspect of using a resistance bridge is realizing that when noise can be as large as the signal being measured it takes a lot of time averaging to separate the two Back to the thermal noise example above if the noise v
212. ge leads Follow all manufacturer recommendations on installation including lead arrangement or errors in temperature measurement will result 3 5 OPTIONAL SCANNER INSTALLATION The Model 3716 L and 3708 Scanners expands the input capability of the Model 370 and increase the number of input channels to 16 3716 L or 8 3708 It is an extension of the isolated measurement circuitry of the Model 370 so its installation should follow the guidelines described in Paragraph 3 4 as well as this paragraph Cables between the scanner and Dewar should be as short as possible because they carry the smallest signals Cables between the scanner and instruments can be longer e LakeShore SCAN CONTROL 3716 SCANNER a a x OUTPUT el BRED W EL e Front View ua Rear View 3716 FrontRear bmp Figure 3 7 Model 3716 Scanner Front and Rear Panels 3 5 1 Grounding the Scanner It is important to note that the enclosure of the Model 3716 L and 3708 is connected to the input cable shield and therefore tied to measurement common not the chassis of the Model 370 In most installations the shield screw on the enclosure should be ground strapped to the test Dewar creating the single connection between measurement circuits and Earth If a different shield to Earth connection is preferred the scanner enclosure must be electrically insulated from its surroundings 3 5 2 Scanner Resistor Input Connector Definition With the Model 3716 L four 25 pin D
213. h 8 5 3 3 3 Power Cord The Model 370 includes a 3 conductor power cord that mates with the IEC 320 C14 line cord receptacle Line voltage is present on the two outside conductors and the center conductor is a safety ground The safety ground attaches to the instrument chassis and protects the user in case of a component failure A CE approved power cord is included with instruments shipped to Europe a domestic power cord is included with all other instruments unless otherwise specified when ordered Always plug the power cord into a properly grounded receptacle to ensure safe operation of the instrument The delicate nature of measurement being taken with this instrument may necessitate additional grounding including ground strapping of the instrument chassis In these cases the operators safety should remain the highest priority and low impedance from the instrument chassis to safety ground should always be maintained 3 3 4 Power Switch The power switch is part of the line input assembly on the rear panel of the Model 370 and turns line power to the instrument On and Off When the circle is depressed power is Off When the line is depressed power is On Installation 3 3 Lake Shore Model 370 AC Resistance Bridge User s Manual 3 4 RESISTOR INPUT CONNECTION AND CABLING Signals being measured by the Model 370 are so small that the measurement environment must be taken into account when connecting resistors to the instrument input
214. h the Model 3716 L and 3708 Scanner Commercial 15 pin extension cables should not be used because the shield is tied to the body of the connector that is at chassis ground on the instrument Longer cables can be constructed with shield tied to scanner analog common The cable must have one connector plug end to mate with the instrument and one connector receptacle end to mate with the scanner 3 5 8 Using the Scan Control Signals Without A Scanner If an optional scanner is not installed some of the scan control lines are available for other uses These signals remain part of the isolated measurement circuitry and should be treated like part of the resistance measurement They should only be used to drive isolated or differential inputs Scan channel select lines can be controlled by the computer interface command DOUT The power supplies can also be used but external circuits must guarantee that current is limited or performance of the instrument may be degraded 3 10 Installation Lake Shore Model 370 AC Resistance Bridge User s Manual SCAN CONTROL 1 8 zzz Scan Control bmp Pin Name Description Pin Name Description 2 Digital Common Common or ground 5 V 10 Scan 2 TTL Digital Output 2 3 Digital Common 11 Scan 3 TTL Digital Output 3 5 Analog Common Common or ground for 15 V 12 Scan 4 TTL Digital Output 4 6 Analog Common 13 Scan Gain TTL Digital Output 5 7 Analog Common 14 5 V 5 V 50 mA 8 15 V 15 V 02
215. he A or Y key to select one of the following terminators CR LF LF CR LF and EOI The default is Cr Lf Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display 6 1 2 Remote Local Operation Normal operations from the keypad are referred to as Local operations The Model 370 can also be configured for Remote operations via the IEEE 488 interface or the Local key The Local key will toggle between Remote and Local operations During Remote operations the Remote annunciator will appear in the upper right corner of the display and operations from the keypad will be disabled 6 1 3 IEEE 488 Command Structure The Model 370 supports several command types These commands are divided into three groups 1 Bus Control Refer to Paragraph 6 1 3 1 a Universal 1 Uniline 2 Multiline b Addressed Bus Control Common Refer to Paragraph 6 1 3 2 3 Device Specific Refer to Paragraph 6 1 3 3 4 Message Strings Refer to Paragraph 6 1 3 4 6 1 3 1 Bus Control Commands A Universal Command addresses all devices on the bus Universal Commands include Uniline and Multiline Commands A Uniline Command Message asserts only a single signal line The Model 370 recognizes two of these messages from the BUS CONTROLLER Remote REN and Interface Clear IFC The Model 370 sends one Uniline Command Service Request SR
216. he front panel by computer interface and for some functions by the Model 370 internal operating modes Closed loop control is not available for analog output functions Configurable Display The Model 370 includes an eight line by forty character vacuum fluorescent display Input readings can be displayed in m Q kQ MQ mK or K Introduction 1 3 Lake Shore Model 370 AC Resistance Bridge User s Manual Scanners for the Model 370 Three custom scanners are available with the Model 370 the Model 3716 3716L and the 3708 These are designed specifically to increase the Model 370 input capability from 1 to either 8 or up to 16 resistors without sacrificing measurement performance Each scanner is housed in a separate enclosure and can be mounted directly on the cryostat so leads carrying the most sensitive low voltage signals are as short as possible The scanner also allows extension of the Model 370 shield to all resistor leads A preamplifier in the scanner amplifies measurement signals before they travel through the longer leads to the Model 370 Different preamplifiers in the 3716 3716L and 3708 optimize them for different applications The Model 370 supplies power and control to the scanner eliminating additional noise from AC power lines ground loops and computer interface connections Scanner operation is fully integrated so most of the Model 370 hardware and firmware features can be used with the scanner Supported hardwar
217. he installer of the Model 370 should consider the following e Shield measurement and computer interface cables Leave no unused or unterminated cables attached to the instrument e Make cable runs as short and direct as possible Higher radiated emissions is possible with long cables Do not tightly bundle cables that carry different types of signals Lake Shore Model 370 AC Resistance Bridge User s Manual TABLE OF CONTENTS Chapter Paragraph Title Page 1 INTRODUCTION sza ic ROWE dod IE Or W OCE O IE dot 1 1 1 0 GENERA lbs ie sowa p LEJE A ady ESEE 1 1 1 1 DESCRIPTION uc aa zad da OO dd SORY 1 2 1 2 SENSOR PERFORMANCE etiain in n a a a e d aain 1 5 1 3 SPECIFIGATION Sica 1 9 1 4 SAFETY SUMMARY crio enana 1 11 1 5 SAFETY SYMBOLS 2 ati titi EEE 1 11 2 BACKGROUND cui naa ana anaa Aaa saaa MAAA anaana aaia anaana wyw d 2 1 2 0 GENERAL voae eaaet aaaeeeaei aa ea aa aaa Ea ee e Eaa ea aa aE LSP 2 1 2 1 USEFUL TEMPERATURE RANGE ciar in 2 1 2 2 TEMPERATURE SCALES cistitis 2 1 2 3 CRYOGENIC COOLING SYSTEM TERMINOLOGY eeissesiissrriiesriirerrissrrirssriitesirinreeinntrrnent 2 1 2 3 1 Dilution Refrigerator Terminology eee esse aaa aaa aaa aaa aaa aaa ania 2 2 2 3 2 STAG yO da a eck a aa AA ee ees at a lala Data OPRZE 2 3 2 3 3 Adiabatic Demagnetization Refrigerator ADR ccccceeseeeceeeneeeeeeeeeeesnaeeeeeeaeeeseneeeeesnaeeeeeeaaes 2 4 2 4 MODEL 370 THEORY OF OPERATION eee eee eee ea eee aaa a
218. he instrument will respond only to the last query it receives The response can be a reading value status report or the present value of a parameter Response data formats are listed along with the associated queries in Paragraph 6 3 The response is sent as soon as possible after the instrument receives the query Typically it takes 10 ms for the instrument to begin the response Some responses take longer 6 2 6 Message Flow Control It is important to remember that the user program is in charge of the serial communication at all times The instrument can not initiate communication determine which device should be transmitting at a given time or guarantee timing between messages All of this is the responsibility of the user program When issuing commands only the user program should e Properly format and transmit the command including terminators as one string e Guarantee that no other communication is started for 50 ms after the last character is transmitted e Not initiate communication more than 20 times per second When issuing queries or queries and commands together the user program should e Properly format and transmit the query including terminators as one string e Prepare to receive a response immediately e Receive the entire response from the instrument including the terminators e Guarantee that no other communication is started during the response or for 50 ms after it completes e Not initiate communication more than 2
219. hen enter the appropriate heater range as described above The heater display should show a value greater than zero and less than 100 when temperature stabilizes The load temperature should stabilize at a temperature below the setpoint If the load temperature and heater display swing rapidly the heater range or proportional value may be set too high and should be reduced Very slow changes in load temperature that could be described as drifting are an indication of a proportional setting that is too low which is addressed in the next step Gradually increase the proportional setting by doubling it each time At each new setting allow time for the temperature of the load to stabilize As the proportional setting is increased there should be a setting in which the load temperature begins a sustained and predictable oscillation rising and falling in a consistent period of time See Figure 2 4 a The goal is to find the proportional value in which the oscillation begins do not turn the setting so high that temperature and heater output changes become violent In many systems it is difficult to differentiate oscillation and noise Operating the control sensor at higher than normal excitation power can help Record the proportional setting and the amount of time it takes for the load change from one temperature peak to the next The time is called the oscillation period of the load It helps describe the dominant time constant of the load which is used
220. hen the instrument is operating Do Not Operate In An Explosive Atmosphere Do not operate the instrument in the presence of flammable gases or fumes Operation of any electrical instrument in such an environment constitutes a definite safety hazard Keep Away From Live Circuits Operating personnel must not remove instrument covers Refer component replacement and internal adjustments to qualified maintenance personnel Do not replace components with power cable connected To avoid injuries always disconnect power and discharge circuits before touching them Do Not Substitute Parts Or Modify Instrument Do not install substitute parts or perform any unauthorized modification to the instrument Return the instrument to an authorized Lake Shore Cryotronics representative for service and repair to ensure that safety features are maintained Cleaning Do not submerge instrument Clean only with a damp cloth and mild detergent Exterior only 1 5 SAFETY SYMBOLS Direct current power line Equipment protected throughout by double insulation or reinforced insulation equivalent to Class II of Alternating or direct current power line IEC 536 see Annex H Alternating current power line Three phase alternating current power line Caution High voltages danger of electric shock Background color Earth ground terminal Yellow Symbol and outline Black f Caution or Warning See Protective conductor terminal A instrument
221. hing 1 Latching Remarks Configures the alarm parameters for a channel Examples ALARM 1 0 term Turns off alarm checking for Channel 1 All other parameters are set to zero ALARM 5 1 1 270 0 0 0 1 term Turns channel 5 alarm checking on with the data source set to Kelvin Activates high alarm if the reading is over 270 activates low alarm if reading is below 0 and latches the alarm when Kelvin reading falls below 270 ALARM Alarm Parameter Query Input ALARM lt channel gt term Format nn lt channel gt Specifies which channel to query 1 16 Returned lt off on gt lt data source gt lt high value gt lt low value gt lt deadband gt lt latch gt term Format n n tnnn nnnE nn nnn nnnE nn tnnn nnnE nn n Refer to command for description ALARMST Alarm Status Query Input ALARMST lt channel gt term Format nn lt channel gt Specifies which channel to query 1 16 Returned lt high state gt lt low state gt term Format n n lt high state gt 0 Off 1 On lt low state gt 0 Off 1 On ALMRST Reset Alarm Command Input ALMRST term Remarks Clears both the high and low status of all alarms ANALOG Analog Output Parameter Command Input ANALOG lt analog channel gt lt polarity gt lt mode gt lt channel gt lt data source gt lt high value gt lt low value gt lt manual value gt term Format n n n nn n tnnn nnnE nn nnn nnnE nn nnnnnn lt analog channel gt Specifies analog outp
222. hing give artificial respiration If breathing is difficult give oxygen Get medical help WARNING Liquid helium and liquid nitrogen can cause severe frostbite to the eyes or skin DO NOT touch frosted pipes or valves In case of frostbite consult a physician at once If a physician is not readily available warm the affected areas with water that is near body temperature The two most important safety aspects to consider when handling LHe and LN are adequate ventilation and eye and skin protection Although helium and nitrogen gases are non toxic they are dangerous in that they replace the air in a normal breathing atmosphere Liquid products are of an even greater threat since a small amount of liquid evaporates to create a large amount of gas Therefore it is imperative that cryogenic Dewars be stored and the cryogenic system be operated in open and well ventilated areas Persons transferring LHe and LN should make every effort to protect eyes and skin from accidental contact with liquid or the cold gas issuing from it Protect your eyes with full face shield or chemical splash goggles Safety glasses even with side shields are not adequate Always wear special cryogenic gloves Tempshield Cryo Gloves or equivalent when handling anything that is or may have been in contact with the liquid or cold gas or with cold pipes or equipment Long sleeve shirts and cuffless trousers that are of sufficient length to prevent liquid from entering the
223. ic measurements systems such as a vibrating sample magnetometer extraction magnetometer SQUID magnetometer etc The exact technical definition relates to the torque exerted on a magnetized sample when placed in a magnetic field Note that the moment is a total attribute of a sample and alone does not necessarily supply sufficient information in understanding material properties A small highly magnetic sample can have exactly the same moment as a larger weakly magnetic sample see Magnetization Measured in SI units as Am and in cgs units as emu 1 emu 107 Am magnetic units Units used in measuring magnetic quantities Includes ampere turn gauss gilbert line of force maxwell oersted and unit magnetic pole magnetization M This is a material specific property defined as the magnetic moment m per unit volume V M m V Measured in SI units as A m and in cgs units as emu cm 1 emu cm 10 A m Since the mass of a sample is generally much easier to determine than the volume magnetization is often alternately expressed as a mass magnetization defined as the moment per unit mass material safety data sheet MSDS OSHA Form 20 contains descriptive information on hazardous chemicals under the OSHA Hazard Communication Standard HCS These data sheets also provide precautionary information on the safe handling of the gas as well as emergency and first aid procedures microcontroller A microcomputer microprocessor or other equipment use
224. ifferent excitation levels making resistance range selection is somewhat more involved than it initially appears The goal is to use enough excitation to maintain reasonable accuracy resolution and settling time without putting too much heat into the resistor Higher resistance ranges from 6 32 MQ to 63 2 MQ are available but not specified because the effects of lead capacitance can not be eliminated at such high resistances These ranges should be used for relative measurements only The selected resistance range and excitation are always shown on the instrument display Refer to Table 1 1 for resistance and excitation ranges As shown in Table 1 1 not every combination of excitation and resistance range is possible Unusable combinations of range and excitation are automatically locked out The Model 370 also continuously calculates and displays the excitation power being dissipated in the resistor as additional feedback to the operator The excitation current source is shorted inside the instrument to minimize transients during range changes Manual operation allows the user to choose both the excitation and resistance range The instrument does not change ranges automatically If the input resistance is too large for the range an overload condition will be indicated This mode is used when the system is very sensitive to excitation changes or if changes in input resistance are too fast for the auto range algorithms The Model 370 has an autorange fe
225. igned to it Setpoint units will default to ohms if no curve is selected for the control channel If the setpoint units are changed during control the instrument will convert the setpoint value to the new units when possible to minimize disturbance to the load The next control setup screen appears as a prompt for setpoint units Use the A or Y key to select setpoint units of ohms or Kelvin Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display 4 11 2 4 Setpoint Change Pause Time The temperature control loop needs time to settle after any setpoint change If temperature control is active while scanning channels the loop may not get enough time to settle into the new setpoint before the scanner moves away from the control channel The instrument can be forced to read only the control channel for a period of time after any setpoint change to guarantee settling This time period is set with the setpoint change pause parameter Manually changing scanner channel will override the pause and force a reading The next control setup screen appears as a prompt for the setpoint change pause time Use the data entry keys to enter a change pause time from to 255 seconds Press Enter to accept the new selection and continue to the setpoint ramp feature Paragraph 4 11 7 Press Escape to restart the setting sequence and enter a new
226. in setting integral If all has gone well the appropriate proportional setting is one half of the value required for sustained oscillation See Figure 2 4 b Theory of Operation 2 21 Lake Shore Model 370 AC Resistance Bridge User s Manual Tuning Proportional Continued If the load does not oscillate in a controlled manner the heater range could be set too low A constant heater reading of 100 on the display would be an indication of a low range setting The heater range could also be too high indicated by rapid changes in the load temperature or heater output less than 10 when temperature is stable There are a few systems that will stabilize and not oscillate with a very high proportional setting and a proper heater range setting For these systems setting a proportional setting of one half of the highest setting is the best choice 2 10 3 3 Tuning Integral When the proportional setting is chosen and the integral is set to zero off the Model 370 controls the load temperature below the setpoint Setting the integral allows the Model 370 control algorithm to gradually eliminate the difference in temperature by integrating the error over time See Figure 2 4 d An integral setting that is too high causes the load to take too long to reach the setpoint An integral setting that is too low can create instability and cause the load temperature to oscillate Begin this part of the tuning process with the system controlling in proportiona
227. inal packing material should be retained for reshipment If not available a minimum of three inches of shock adsorbent packing material should be placed snugly on all sides of the instrument in a sturdy corrugated cardboard box The RGA number should be included in the mailing label or written prominently on the outside of the box A copy of the customer contact information and RGA number should be included inside the box Consult Lake Shore with questions regarding shipping and packing instructions Service 8 1 Lake Shore Model 370 AC Resistance Bridge User s Manual 8 3 FUSE DRAWER The fuse drawer supplied with the Model 370 holds the instrument line fuses and line voltage selection module The drawer holds two 5 x 20 mm time delay fuses It requires two good fuses of the same rating to operate safely Fuse ratings are different for different line voltages Refer to Paragraph 8 5 for details Fuse Fuse Front View Side View Rear View European Dual Fuse Configuration EU_Fuse eps Figure 8 1 Fuse Drawer 8 4 LINE VOLTAGE SELECTION Use the following procedure to change the instrument line voltage selector Verify the fuse value whenever line voltage is changed WARNING To avoid potentially lethal shocks turn off controller and disconnect it from AC power before performing this procedure 1 Identify the line input assembly on the instrument rear panel See Figure 8 2 2 Turn the line power switch OFF O
228. inear value gt term Format nnn nnnE nn Remarks Also see the RDGST command LINEAR Linear Equation Parameter Command Input LINEAR lt channel gt lt equation gt lt M value gt lt X source gt lt B source gt lt B value gt term Format nn n tnnnnnn n n Ennn nnnE nn lt channel gt Specifies channel to configure 1 16 0 all channels lt equation gt Specifies linear equation to use 1 y mx b 2 y m x b lt M value gt Specifies a value for M in the equation lt X source gt Specifies data source for X 1 Kelvin 2 Ohms lt B source gt Specifies source for B in the equation To use a setpoint set its units to the same type specified in lt X source gt 1 a value 2 SP 3 SP lt B value gt Specifies a value for B in the equation if lt B source gt is 1 Example LINEAR 5 1 1 0 1 3 term The linear data for Channel 5 is calculated using the equation y mx b The value of m is 1 0 x is set to Kelvin and b is set to SP LINEAR Linear Equation Parameter Query Input LINEAR lt channel gt term Format nn lt channel gt Specifies which channel to query 1 16 Returned lt equation gt lt M value gt lt X source gt lt B source gt lt B value gt term Format n nnnnnn n n tnnn nnnEtnn Refer to command for description Remarks Returns channel linear equation parameter values LOCK Keypad Lock Command Input LOCK lt state gt lt code gt term Format n nnn lt state
229. ing screen Press Escape to cancel the new selection and return to the normal display 5 10 Advanced Operation Lake Shore Model 370 AC Resistance Bridge User s Manual Zone Setting Work Sheet Setpoint D Manual Output Heater Low High Analog Out1 Analog Out2 0 100 Range Relay Relay 0 100 0 100 P 0 1 1k 0 1 1k 0 200 On Off On Off Setpoint High Analog Out1 Analog Out2 0 100 0 100 P I D Manual Output Heater Low Range Relay Relay 0 1 1k 0 1 1k 0 200 0 100 On Off On Off Zone 08 Vast Rf AREA a RE ca I D Manual Output Heater Low High Analog Out1 Analog Out2 Range Relay Relay 0 100 0 100 s 1k 0 1 1k 0 200 0 100 On Off On Off Setpoint High Analog Out1 Analog Out2 0 100 0 100 P D Manual Output Heater Low Range Relay Relay 0 1 1k 0 1 1k 0 200 0 100 On Off On Off Zone 06 Se tp ont D Manual Output Heater Low High Analog Out1 Analog Out2 Range Relay Relay 0 100 0 100 P 0 100 0 1 1k 0 1 1k 0 200 On Off On Off Zone 05 NH Setpoint D Manual Output Heater Low High Analog Out1 Analog Out2 Range Relay Relay 0 100 0 100 P l 0 1 1k 0 1 1k 0 200 0 100 On Off On Off Setpoint High Analog Out1 Analog Out2 0 100 0 100 P I D Manual Output Heater Low Range Relay Relay 0 1 1k 0 1 1k 0 200 0 100 On Off On Off Zone 3 gt etpoint
230. installed and not all 16 channels are being used some of the scanner channels can be disabled Disabling a channel does not change any of the measurement hardware but it does allow the firmware to skip the channel in keypad entry sequences and streamline operation This sequence will be skipped if there is no scanner attached The instrument will not allow all channels to be turned off at the same time and will turn on channel 1 as a default The next input setup screen appears as a prompt for channel On Off Use the A or Y key to select channel on or off Press Enter to accept the new selection and return to the normal display Press Escape to cancel the new selection and return to the normal display 4 10 Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 4 7 INSTRUMENT SETUP The input parameters grouped under Instrument Setup apply to all input channels whether or not a scanner is installed The other parameters included in this paragraph do not belong anywhere else so they ended up here 4 7 1 Frequency Selection The Model 370 AC resistance measurement runs at a nominal excitation frequency of 13 7 Hz This frequency was chosen so that it and its harmonics are separated from both 50 Hz and 60 Hz line frequency It was also chosen to balance the effects of settling time and cable capacitance All specifications including resolution and filter settling are given at 13 7 Hz excitation frequency It may be ne
231. ion 24 bit Measurement Resolution Range dependent See Table 1 1 Accuracy Range dependent See Table 1 1 Measurement Temp Co 0 0015 of reading 0 0002 of range C Lead Connections V V I V shield shield individual guards Scanner Lead Connections V V I l for each sensor shield common to all Max Lead Resistance 100 2 10 of resistance range per lead for current lt 3 16 mA 10 Q 10 of resistance range per lead for current gt 10 mA Input Isolation Measurement optically isolated from chassis ground Common Mode Reduction Matched impedance voltage input and current output active CMR Excitation Type Sinusoidal AC current source Excitation Frequency 13 7 Hz 9 8 Hz and 16 2 Hz alternates Excitation Currents 21 ranges from 3 16 pA to 31 6 mA RMS Minimum Excitation Power 10 18 W into 100 kQ see Table 1 1 for other ranges Maximum DC Current 4 pA 1 of excitation current 1 6 x 10 18 W into 100 kQ Current Protection Current leads relay shorted on power up and range change Voltage Ranges 12 ranges from 2 uV to 632 mV RMS Voltage Over Range 20 nominal Input Impedance Approaches 5 x 1012 Q Range Selection Modes Manual Voltage Excitation Current Excitation and Autorange Scanner Modes Manual or Autoscan Filter Settling Times 1 to 200 seconds Additional Software Features Max Min capture Linear equation user programmable pause on range and input change Temperature Conv
232. ions 2 10 1 3 Two Sensor Approach There is a conflict between the best sensor location for measurement accuracy and the best sensor location for control For measurement accuracy the sensor should be very near the sample being measured which is away from the heating and cooling sources to reduce heat flow across the sample and therefore thermal gradients The best control stability is achieved when the feedback sensor is near both the heater and cooling source to reduce thermal lag If both control stability and measurement accuracy are critical it may be necessary to use two sensors one for each function Another reason to use a separate control sensor is that a dedicated control sensor can be run with higher excitation power The additional power will reduce the accuracy of the feedback sensor but create a larger signal for better control stability Accuracy of the control sensor is not important if the temperature sample is measured with a different sensor 2 10 1 4 Thermal Mass Cryogenic designers understandably want to keep the thermal mass of the load as small as possible Small mass can have the advantage of reduced thermal gradients but controlling a very small mass is difficult because there is no buffer to adsorb changes in cooling power Without buffering small disturbances can very quickly create large temperature changes In some systems it is necessary to add a small amount of thermal mass in order to improve control stability 2 1
233. is difficult to define In a practical sense the instrument is limited by its need for four electrically conductive leads to be attached to a resistor and that resistor attached to a cold surface Heat conduction and noise pick up in the resistor leads limit the lowest temperature that can be achieved in this environment even before an instrument is attached The diminishing thermal contact between the resistor and cold surface at lower and lower temperatures only exaggerate the problem In general resistance bridge measurements are considered routine down to 100 mK in a capable cryogenic cooling system With careful installation of a high quality bridge like the Model 370 it is not difficult to achieve 50 mK Below that each detail of the overall system has to be optimized to a low end of between 10 and 20 mK 2 2 TEMPERATURE SCALES Temperature conversions at such low temperatures can raise even more questions than resistance measurements One reason is the delay in getting a recognized temperature standard for very low temperatures As of the writing of this manual the International Temperature Scale of 1990 ITS 90 is the most current effort to define an internationally accepted temperature scale It is based on primary thermometers and fixed temperature points A detailed discussion of ITS 90 including interpolation instruments and equations was published by the National Institute of Standards and Testing NIST as Technical Note 1265 Guideli
234. ith U S power cord VAC 220 Instrument configured for 220 VAC with universal European line cord VAC 240 Instrument configured for 240 VAC with universal European line cord VAC 120 ALL Instrument configured for 120 VAC with U S power cord and universal European line cord and fuses for 220 240 setting Options and Accessories 7 1 Lake Shore Model 370 AC Resistance Bridge User s Manual 7 2 OPTIONS The list of Model 370 options is provided as follows Description Of Options CalCurve on Floppy Disk or E Mail The Model 8000 CalCurve is offered on floppy disk or via e mail free of charge at time of order to any customer who orders a calibrated sensor The Model 8000 consists of calibration breakpoint interpolation data stored in ASCII format Also 8000 included is a PC executable program to load the data into a Lake Shore Instrument via IEEE 488 or RS 232 Interface Once loaded the instrument uses the data to calculate and display temperature The following information is included raw data coefficients interpolation table instrument breakpoints a curve loading program and Readme txt describing the file formats CalCurve Factory Installed Provides users with a convenient method of storing sensor 8001 370 calibrations within Lake Shore Instruments Calibration data breakpoint interpolation table for a specific sensor is stored into a nonvolatile memory CalCurve Field Installation For users who already own a M
235. iving an organism of oxygen autotuning In Lake Shore instruments the Autotuning algorithm automatically determines the proper settings for Gain Proportional Reset Integral and Rate Derivative by observing the time response of the system upon changes in setpoint B Symbol for magnetic flux density See Magnetic Flux Density Glossary of Terminology A 1 Lake Shore Model 370 AC Resistance Bridge User s Manual bar Unit of pressure equal to 10 pascal or 0 98697 standard atmosphere Baud A unit of signaling speed equal to the number of discrete conditions or signal events per second or the reciprocal of the time of the shortest signal element in a character bel B A dimensionless unit expressing the ration of two powers or intensities or the ratio of a power to a reference power such that the number of bels is the common logarithm of this ratio bifilar windings A winding consisting of two insulated wires side by side with currents traveling through them in opposite directions bit A contraction of the term binary digit a unit of information represented by either a zero or a one BNC Bayonet Nut Connector boiling point The temperature at which a substance in the liquid phase transforms to the gaseous phase commonly refers to the boiling point at sea level and standard atmospheric pressure CalCurve Service The service of storing a mathematical representation of a calibration curve on an EEP
236. ka 2 0 ka 1 0kQ i i hi a 3 2 W 1 0 fW 320 aW 100 aw 32 aW 10 aW i i ja 7 j 63 2 MQ 20 MQ 6 32 MQ 20MQ 632 KQ 316pA il gt u i dl i hs 10kQ 6 0 ka i i j 5 320 aW 100 aw 32 aW 10aW 3 2 aW i 63 2MQ 20MQ 632MQ 1 0 pA kk kk ak i 7 A z Al s 32 aW 10 aw 3 2 aW 1 0 aw Accuracy 0 5 1 0 1 6 Introduction Lake Shore Model 370 AC Resistance Bridge User s Manual Accuracy 0 03 o 9 9 Table 1 2 Model 370 3716L Voltage Range e hd 632 mV 200 mV 63 2 mV 20 mV 6 32 mV 2 0 mV 632 uV 200 uV 63 2 UV 20 uV 6 32 UV 2 0 uV 200 6 32 Q 200 632 mQ 200mQ 63 2 MQ 20mQ 6 32 mQ 2 0 mQ 632 pO 63 2 uQ 31 6 mA 20 yQ 6 yQ 2 0 Q 1 3 Q 400nQ 130 nQ 40 nQ 100 nQ 20nQ 20nQ 20nQ 10mW 32mW 10mW 320uW 100uW 32uWw 10pWw 32uW 10uw 320nw 32 nW 63 2 Q 200 6 32 Q 20Q 632 mQ 200mQ 63 2 MQ 20mQ 6 32 MQ 2 0 mQ 200 uQ 10mA 60pQ 20uQ 60402 40uQ 1340 400nQ 130nQ 60nQ 60nQ 60nQ 60nQ 3 2mW 1mW 320pW 1004W 32uWw 104W 32uW 10pW 320nW 100nW 10nW 200 Q 63 2 Q 200 6 32 Q 200 632 mQ 200mQ 63 2 MQ 20 mQ 6 32 MQ 632 3 16 mA 200 Q 60 yQ 20 yQ 13 HQ 4 0 uQ 1 3 Q 400nQ 200 nQ 200 nQ 200nQ 200nQ 1mW 320uW 100pW 324W 10uW 32uW 10uW 320nW 100nW 32nW 3 2 nW 632 Q 200 Q 63 2 Q 200 6 32 Q
237. l One of the input channels must be selected as the control channel if a scanner is attached and a closed loop control mode is selected The closed loop control equation gets feedback from the selected channel Reading channels other than the control channel by manually changing scanner channel or with autoscan presents a special case for the control algorithm described in Paragraph 4 11 8 If the control channel is changed during control the instrument will attempt to minimize disturbance to the load by choosing the reading value from the new channel as the setpoint when controlling in ohms NOTE If there is not a curve assigned to the control channel the appropriate temperature coefficient must be selected in the input setup sequence Paragraph 4 6 4 or the PID control algorithm will not operate properly The next control setup screen appears as a prompt for control channel If no scanner is installed this screen will be skipped Use the A or Y key to select a control channel from 1 to 16 Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display 4 16 Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 4 11 2 3 Setpoint Units Closed loop temperature control will operate with a setpoint in resistance units for any input channel and a setpoint in Kelvin for any channel that has a temperature response curve ass
238. l 370 has 21 current settings down to 3 16 pA These currents have low noise and almost no DC component to self heat the measured resistor Two excitation modes are available to best utilize the features of the current source in different applications One lower current setting 1 pA is available but not specified because leakage and bias currents in the measurement circuits can be a large percentage of the excitation current The 1 pA current setting should be used for relative measurements only Precision resistance measurements use a four lead method to eliminate the effect of nominal lead resistance Excitation current is driven on one pair of leads while the resulting signal voltage is measured on the second pair Measurement errors from lead resistance are proportional to current flowing in the voltage leads Instrument voltage inputs are expected to have high impedance to prevent the unwanted current from flowing The Model 370 has very high input impedance to both normal mode voltages created by current source excitation and common mode voltages created by induced noise on the leads Offset voltages occur in every electronic circuit both in signal amplifiers and in lead wire connections that create thermal EMF voltages Offset voltages affect accuracy more when the excitation is small because they are a larger percentage of the measured voltage AC measurements allow AC filters that greatly reduce amplifier offsets at each stage allowing higher
239. l 3716 Scanner is being used When excitation is turned off the instrument sets its current source to the lowest output possible and then shunts across the output with a mechanical relay The relay provides a low resistance path for any remaining leakage current and prevents the current source from climbing to its compliance voltage When excitation is turned back on the process is reversed to prevent transients from heating the measured resistor To turn excitation off press the Voltage Mode and Current Mode keys at the same time for one second Both excitation mode LED annunciators will turn off and the excitation display location will show Off To restore excitation press any of the range or excitation keys The first key press will return the instrument to its previous excitation without making a change 4 6 INPUT SETUP There are several parameters besides resistance range and excitation that directly affect the measurement input The parameters grouped under Input Setup are specific to individual scanner channels if a scanner is installed Channel selection scan pause time and channel enable are skipped if a scanner is not installed Change pause and curve or temperature coefficient should be set even if the Model 370 is configured for only one channel The parameters grouped under Instrument Setup Paragraph 4 7 apply to all channels whether or not a scanner is installed 4 8 Operation Lake Shore Model 370 AC Resistance Bridge User
240. l Number Up to a 10 characters can be entered as a sensor serial number Both numbers and letters can be entered over computer interface only numbers can be entered from the front panel Format The format parameter tells the instrument what breakpoint data format to expect Different sensor types require different formats Formats for Lake Shore sensors are Log Q K Log Resistance vs Kelvin for germanium or other NTC sensors Q K Resistance vs Kelvin for platinum RTD or other PTC sensors Limit Temperature control setpoint limit in Kelvin Default is 375 K Enter a setting of 9999 K if no limit is needed Temperature Coefficient The unit derives the temperature coefficient from the first two breakpoints The user does not enter this setting Check for proper entry of the first two breakpoints if the temperature coefficient is not correct A positive coefficient P indicates that the sensor signal increases with increasing temperature A negative coefficient N indicates that the sensor signal decreases with increasing temperature 5 1 2 Curve Breakpoints Temperature response data of a calibrated sensor must be reduced to a table of breakpoints before entering it into the instrument Each breakpoint consists of one sensor value in ohms or log ohms and one temperature value in Kelvin The instrument uses linear interpolation to calculate temperature between breakpoints From 2 to 200 breakpoints can be entered as a curve The instrument will
241. l germanium RTD changing from 3 5 MQ K at 50 mK to 36 Q K at 1 K An associated change in resolution is also shown as 1 7 UK at 50 mK to 83 uK at I K The reason that the drop in resolution is not as severe as the drop in sensitivity is that excitation current is increased approximately 1 000 times by the Model 370 to maintain practical measurement parameters Sensitivity and therefore resolution of most sensors changes with temperature so it is important to choose a sensor with high sensitivity in the most critical temperature range Two or more sensors may be necessary to cover a large range of temperature with high resolution 2 9 1 4 Accuracy Resistive thermometers serve well as secondary standards and are much easier to use than most primary thermometers Accuracy of a resistive temperature sensor depends on how closely it is compared to a primary standard how the temperature scale is interpreted how much the sensor drifts over time and how well the conditions used in calibration are duplicated when the sensor is being used General rules to achieve best accuracy choose a sensor with good stability and high sensitivity in the most critical temperature range calibrate it carefully or purchase a calibration from a reputable company install the sensor according to manufacture recommendations and use the same excitation that was used during calibration Do not forget to include instrumentation errors when determining overall accuracy and recalibrate
242. l only mode Use the oscillation period of the load that was measured above in seconds as the integral setting Enter the integral setting into the Model 370 and watch the load temperature approach the setpoint If the temperature does not stabilize and begins to oscillate around the setpoint the integral setting is too low and should be doubled If the temperature is stable but never reaches the setpoint the integral setting is too high and should be decreased by half To verify the integral setting make a few small 2 to 5 degree changes in setpoint and watch the load temperature react Trial and error can help improve the integral setting by optimizing for experimental needs Faster integrals for example get to the setpoint more quickly at the expense of greater overshoot In most systems setpoint changes that raise the temperature act differently than changes that lower the temperature If it was not possible to measure the oscillation period of the load during proportional setting start with an integral setting of 50 If the load becomes unstable double the setting If the load is stable make a series of small setpoint changes and watch the load react Continue to decrease the integral setting until the desired response is achieved 2 10 3 4 Tuning Derivative If an experiment requires frequent changes in setpoint or data taking between changes in the setpoint derivative should be considered See Figure 2 4 e A derivative setting of zero of
243. lation The maximum steady state amount that the output voltage or current changes as result of a specified change in input line voltage usually for a step change between 105 125 or 210 250 volts unless otherwise specified line voltage The RMS voltage of the primary power source to an instrument liquid helium LHe Used for low temperature and superconductivity research minimum purity 99 998 Boiling point at 1 atm 4 2 K Latent heat of vaporization 2 6 kilojoules per liter Liquid density 0 125 kilograms per liter EPA Hazard Categories Immediate Acute Health and Sudden Release of Pressure Hazards DOT Name Helium Refrigerated Liquid DOT Label Nonflammable Gas DOT Class Nonflammable Gas DOT ID No UN 1963 A 4 Glossary of Terminology Lake Shore Model 370 AC Resistance Bridge User s Manual liquid nitrogen LN gt Also used for low temperature and superconductivity research and for its refrigeration properties such as in freezing tissue cultures minimum purity 99 998 O 8 ppm max Boiling point at 1 atm 77 4 K Latent heat of vaporization 160 kilojoules per liter Liquid density 0 81 kilograms per liter EPA Hazard Categories Immediate Acute Health and Sudden Release of Pressure Hazards DOT Name Nitrogen Refrigerated Liquid DOT Label Nonflammable Gas DOT Class Nonflammable Gas DOT ID No UN 1977 load regulation A steady state decrease of the value of the specified variable resulting from a specifi
244. lay 4 6 3 Change Pause Time The input change pause parameter allows the user to select how long the instrument waits after a resistance range or channel change before it begins using readings for autorange Max Min alarm analog output and temperature control functions The default of 3 seconds is recommended for typical applications but the pause time should be increased for large resistances or very low excitations The next input setup screen appears as a prompt for the change pause time Use the data entry keys to enter a pause time between 3 and 200 seconds Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display Operation 4 9 Lake Shore Model 370 AC Resistance Bridge User s Manual 4 6 4 Selecting a Temperature Response Curve If the Model 370 is being used to read resistive temperature detectors RTDs a temperature response curve can be entered into the instrument so that it can calculate and display temperature as well as measured resistance This paragraph instructs how to select a curve that is already installed in the instrument Refer to Paragraph 5 2 1 for information on how to enter curves into the Model 370 Ifa curve is selected for a channel the temperature coefficient used for temperature control is taken from that curve and the temper
245. llows you to overwrite an existing user curve Please ensure the curve number you are writing to is correct before proceeding with curve copy To copy a curve press Curve Entry The first curve entry screen appears as a prompt for the curve function Use the A or Y key to select Copy Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next curve entry screen appears as a prompt for the number of the curve to copy from The curve name and serial number are displayed below the curve number when available Use the A or V key to select a curve number from 1 to 20 Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next curve entry screen appears as a prompt for the number of the curve to copy to The curve name and serial number are displayed below the curve number when available Use the A or V key to select a curve number from 1 to 20 Press Enter to accept the new selection and copy the curve Press Escape to cancel the operation and return to the normal display 5 6 Advanced Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 5 3 ZONE TUNING Optimal closed loop PID control parameters values are often different at varied temperatures within a system Once values hav
246. lter is an integral part of the Model 370 resistance measurement It is directly related to measurement resolution and frequency response The instrument takes A D readings 10 times per second no matter how the filter configured The filter algorithm takes a running linear average of from 10 to 2000 readings depending on settling time value New filtered reading is available over the interface ten times a second and displayed at the normal display update rate The linear average filter gives faster settling time than other filter methods but its frequency response is not well behaved The characteristics of the filter should be well understood before attempting to track periodic changes in resistance with filtered readings To begin filter setup press the Filter key If a scanner is installed the first filter setup screen appears as a prompt for the channel number All of the parameter changes that follow will act on the selected channel If no scanner is installed the sequence will begin below with filter On Off Use the A or V key to select a channel number between 1 and 16 If you choose the selection All changes will be applied to all channels at the same time Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The firmware filter can be turned on or off without changing the settling time value When the firmware filter is di
247. lvin 10 k Q 0 001 Q log R K log Resistance vs Kelvin 7 log R 0 00001 log R Setpoint A setpoint temperature limit can be included with every curve When controlling in temperature the ik it setpoint cannot exceed the limit entered with the curve for the control sensor The default is 375 K Set to 9999 K if no limit is required Towoskaie The instrument derives the temperature coefficient from the first two breakpoints If set improperly Gages t check the first 2 breakpoints A positive coefficient indicates the sensor signal increases with increasing temperature A negative coefficient indicates the sensor signal decreases with increasing temperature 5 2 Advanced Operation Lake Shore Model 370 AC Resistance Bridge User s Manual Table 5 2 Recommended Curve Parameters Type bea Shore Units Format Limit K Coefficient rea Platinum 100 PT 100 Ohms Q K 800 Positive 0 001 Q Platinum 1000 PT 100 Ohms Q K 800 Positive 0 01 Q Rhodium Iron RF 100 Ohms Q K 325 Positive 0 001 Q Carbon Glass CGR 1 1000 Ohms logR K 325 Negative 0 00001 logR Cernox CX 1030 Ohms logR K 325 Negative 0 00001 logR Germanium GR 200A 100 Ohms logR K 325 Negative 0 00001 logR Rox RX 102A Ohms logR K 40 Negative 0 00001 logR To enter a new curve or edit an existing curve press Curve Entry The first curve entry screen appears as a prompt for the curve function Use th
248. ly it is also possible that the addition of the grounding path through the oscilloscope may actually quiet an improperly grounded system If in doubt a differential input to the oscilloscope should be used A battery operated oscilloscope is another option 8 7 2 1 Nominal Signal Levels Measuring a 100 kQ resistor at a voltage excitation of 632 mV resistance range of 200 kQ and CMR on measuring a balanced or floating resistance the following AC sine wave signals will be observed at the appropriate signal points CSPOS 4 5 VAC Peak CSNEG 4 5 VAC Peak VCMPOS 0 224 VAC Peak VCMNEG 0 224 VAC Peak VDIF 4 5 VAC Peak VMIX 1 4 VAC Peak VAD 1 VDC OFF No signal The given levels may vary by or 2 At the above levels the observed sine wave should be very clean and noise free These signal levels are representative of a balanced resistance measurement If one side of the measured resistance is common to measurement common large deviations from these voltages will be seen Also be reminded that CMR must be turned off in that condition for an accurate measurement As excitation is reduced environmental noise will be seen on the signals At voltage excitations below 200 pV the measurement signal levels at CSPOS CSNEG VCMPOS and VCMNEG get small enough that a common oscilloscope my not adequately see the measurement signal However common mode noise may be quite visible on the CSPOS and CSNEG
249. m Format n nnn nn Refer to command for description FREQ Excitation Frequency Command Input FREQ lt frequency gt term Format n lt frequency gt Specifies the instruments excitation frequency 1 9 8 Hz 2 13 7 Hz 3 16 2 Hz Remarks Changes the excitation frequency of the instrument Takes several seconds to complete FREQ Excitation Frequency Query Input FREQ term Returned lt frequency gt term Format n Refer to command for description GUARD Guard Drive Command Input GUARD lt state gt term Format n lt state gt 0 Off 1 On Remarks Enables or disables the guard drive GUARD Guard Drive Query Input GUARD term Returned lt state gt term Format n Refer to command for description HTR Heater Output Query Input HTR term Returned lt heater value gt term Format nnn nnnEtnn lt heater value gt Heater output in percent of current or in actual power depending on the heater output selection HTRRNG Heater Range Command Input HTRRNG lt range gt term Format n lt range gt Specifies heater current range 0 Off 1 31 6 pA 2 100 pA 3 316 pA 4 1 00 mA 5 3 16 mA 6 10 0 mA 7 31 6 mA 8 100 mA HTRRNG Heater Range Query Input HTRRNG term Returned lt range gt term Format n Refer to command for description 6 28 Remote Operation Lake Shore Model 370 AC Resistance Bridge User s Manual HTRST Heater Status Query Input HTRST term Returned lt error code
250. make resistance measurements as long as the common mode reduction feature is turned off This is not a recommended operating mode and accuracy is reduced Increase in measurement error will not exceed 0 1 in a typical installation The Model 370 is calibrated and specified at 1ts default measurement frequency of 13 7 Hz An additional calibration is done at each of the two alternate measurement frequencies to compensate for the frequency response of the hardware filters This calibration eliminates instrument errors only and will not account for attenuation caused by lead capacitance that is different at each frequency The autorange algorithm includes a hysteresis band to prevent instability when the measured resistor is at a range boundary It is possible for the instrument to choose a different excitation current when approaching a value from below than it chooses when approaching from above Autorange should not be used when a specific excitation current is critical 2 7 OPERATING TRADEOFFS Any measurement must be optimized to provide the best practical result and tradeoffs must be made to favor the most important measurement characteristics Some of the common tradeoffs of low temperature resistance measurement are described below 2 7 1 Resistor Self heating vs Excitation Even after leads are properly heat sunk and noise is effectively shielded there can be a temperature error in resistance measurements that comes from within Self heating
251. mal conductivity is important in any part of a cryogenic system that is intended to be at the same temperature Most systems begin with materials that have good conductivity themselves but as sensors heaters sample holders etc are added to an ever more crowded space the junctions between parts are often overlooked In order for control to work well junctions between the elements of the control loop must be in close thermal contact and have good thermal conductivity Gasket materials should always be used along with reasonable pressure 2 10 1 2 Thermal Lag Poor thermal conductivity causes thermal gradients that reduce accuracy and also cause thermal lag that make it difficult for controllers to do their job Thermal lag is the time it takes for a change in heating or cooling power to propagate through the load and get to the feedback sensor Because the feedback sensor is the only thing that lets the controller know what is happening in the system slow information to the sensor slows the response time For example if the temperature at the load drops slightly below the setpoint the controller gradually increases heating power If the feedback information is slow the controller puts too much heat into the system before it is told to reduce heat The excess heat causes a temperature overshoot which degrades control stability The best way to improve thermal lag is to pay close attention to thermal conductivity both in the parts used and their junct
252. ment interfacing The Model 370 IEEE 488 Interface complies with the IEEE 488 2 1987 standard and incorporates its functional electrical and mechanical specifications unless otherwise specified in this manual All instruments on the interface bus perform one or more of the interface functions of TALKER LISTENER or BUS CONTROLLER A TALKER transmits data onto the bus to other devices A LISTENER receives data from other devices through the bus The BUS CONTROLLER designates to the devices on the bus which function to perform The Model 370 performs the functions of TALKER and LISTENER but cannot be a BUS CONTROLLER The BUS CONTROLLER is the digital computer which tells the Model 370 which functions to perform Below are Model 370 IEEE 488 interface capabilities e SHl Source handshake capability e RL1 Complete remote local capability e DC1 Full device clear capability DTO No device trigger capability e CO No system controller capability T5 Basic TALKER serial poll capability talk only unaddressed to talk if addressed to listen L4 Basic LISTENER unaddressed to listen if addressed to talk e SRl Service request capability AHl Acceptor handshake capability e PPO No parallel poll capability El Open collector electronics NOTE The Model 370 IEEE 488 Interface requires that repeat addressing be enabled on the bus controller Instruments are connected to the IEEE 488 bus by a 24 conductor connector cable as specifie
253. ment accuracy Many cooling systems include a radiation shield The purpose of the shield is to surround the load sample and sensor with a surface that is at or near their temperature to minimize radiation The shield is exposed to the room temperature surface of the vacuum shroud on its outer surface so some cooling power must be directed to the shield to keep it near the load temperature If the cooling system does not include an integrated radiation shield or one cannot be easily made one alternative is to wrap several layers of super insulation aluminized Mylar loosely between the vacuum shroud and load This reduces radiation transfer to the sample space 2 10 TEMPERATURE CONTROL The Model 370 includes the necessary electronics to perform closed loop temperature control in a dilution refrigerator or other low power cryogenic cooling system The load must be equipped with a temperature sensor for feedback and a resistive heater to add heating power into the system The instrument sensor and heater play important roles in establishing stable temperature control Theory of Operation 2 17 Lake Shore Model 370 AC Resistance Bridge User s Manual 2 10 1 Considerations For Good Control Most of the techniques discussed above to improve cryogenic temperature measurement apply to control as well There is an obvious exception in sensor location and a compromise is suggested in Paragraph 2 10 3 2 10 1 1 Thermal Conductivity Good ther
254. mit of 3 25 K and negative coefficient Curve Header Query CRVHDR lt curve gt term nn lt curve gt Specifies which curve to query 1 20 lt name gt lt SN gt lt format gt lt setpoint limit gt lt coefficient gt term aaaaaaaaaaaaaaa aaaaaaaaaa n tnnn nnn n Refer to command for description Curve Data Point Command CRVPT lt curve gt lt index gt lt units value gt lt temp value gt term nn nnn nnn nnnE nn tnnn nnnE nn lt curve gt Specifies which curve to configure 1 20 lt index gt Specifies the points index in the curve 1 200 lt units value gt Specifies resistance value for this point to 6 digits in ohms or log ohms depending on curve data format lt temp value gt Specifies the corresponding temperature in Kelvin for this point to 6 digits Configures a user curve data point CRVPT 1 2 0 10191 0 470 term Sets User Curve 1 second data point to 0 10191 resistance units and 470 000 mK Remote Operation 6 25 Lake Shore Model 370 AC Resistance Bridge User s Manual CRVPT Curve Data Point Query Input CRVPT lt curve gt lt index gt term Format nn nnn lt curve gt Specifies which curve to query 1 20 lt index gt Specifies the points index in the curve 1 200 Returned lt units value gt lt temp value gt term Format nnn nnnE nn tnnn nnnEtnn Refer to command for description Remarks Returns a standard or user curve data point CSET Temp
255. mmary Command CLS KESE KESE XESR XIDN OPC OPC KRST SRE SRE STB XTST KWAI ALARM ALARM ALARMST ALMRST ANALOG ANALOG AOUT BAUD BAUD BEEP BEEP BRIGT BRIGT CHGALL CHGALL CMODE CMODE CMR CMR CPOL CPOL CRVDEL CRVHDR CRVHDR CRVPT CRVPT CSET CSET DFLT DISPLAY DISPLAY DISPLOC DISPLOC DOUT DOUT FILTER FILTER Function Page Clear Interface Cmd 0 cc eeceeeceseseeeeseeseseeeeneeseeeees 20 Event Status Enable Cmd 2 24 4 2121 41 2 20 Event Status Enable Query ccs 20 Event Status Register Query Identification Query ccceceesceetseseteeseeeeseeseseeeeneees Operation Complete Cmdr Operation Complete Query Reset Instrument Cmd ou eeeeeeeeeeeeneees Service Request Enable Cmd 1 1 1 21 Service Request Enable Query 21 Status Byte QUE Y e O O WA 21 Self Test Query ud OE RA 21 Wait To Continue Cmd u cients 21 Alarm Parameter Cid eee eeseseeeeeeeneees 22 Alarm Parameter Query 22 Alarm Status Query c ooocococonononncnonononononacnnnononennrnnaonos 22 Alarm Reset md ad ato 22 Analog Output Parameter Cmd 22 Analog Output Parameter Query 1 23 Analog Output Data Query 23 RS 232 Baud Rate Cmd 67 29 RS 232 Baud Rate Query 1 1111112121112121 23 Audible Alarm Beeper Cmd 11111 211 23
256. mmon installations at resistance levels of 100k or less are not likely to have difficulty with the phase relationship of the two signals As resistance and or cable length increases so should concern for this relationship Cable capacitance is the single biggest cause of this error Refer to Paragraph 2 6 6 The Model 370 has a means to observe the phase shift caused by cable capacitance by observing the VMIX monitor point and comparing it against the Reference output The Reference output can either be used to sync an oscilloscope for observation of the VMIX signal or the two signals can be x y plotted to observe the aperture width of the pattern as an indicator of phase A simple application of the feature is the observation of phase improvement by the use of driven guards on the measurement leads The change can be observed immediately as the guard drive is turned on or off Judicious use of this feature with external instrumentation can even extend the range of the Model 370 measurement into reactive capacitive inductive areas 8 7 3 Considerations for the Model 3708 Scanner The Model 3708 Scanner was released after the Model 370 was released The firmware in older Model 370 s may need to be updated to properly work with the Model 3708 Scanner Refer to section 4 19 to check with the firmware revision In order for the Model 3708 Scanner to function properly the main firmware needs to have a date of 4 10 2008 or newer and the input firmware needs
257. most stable operating point available for the current range or voltage setting and minimize the number of times a calibrated range is used to calibrate another range Refer to the work sheet below for range dependant information 1 Allow the Model 370 to warm up to 1 hour Disconnect the scanner if present and attach a resistive load during warm up Select the default frequency of 13 7 Hz 2 Calibrate all voltage zero offsets Follow the procedure described in 8 12 2 1 Be sure to set all voltage gain coefficients to default 1 3 Calibrate the 632 mV voltage gain only Follow the procedure described in 8 12 2 2 Small measurement errors in this step will not contribute to the final resistance measurement accuracy 4 Calibrate current gains from 31 6 mA through 1 UA Remain on the 632 mV range that now has both voltage coefficients Follow the procedure described in 8 12 2 3 and repeat down through the I UA setting 5 Calibrate all remaining voltage gains from 200 mV to 2 UV Remain on the HA current setting which is now calibrated e Attach a near full scale resistor of known value to the input of the instrument Read and record the RDGR resistance value e Calculate the voltage gain coefficient CALG Actual Resistance RDGR Send the coefficient to the instrument e Repeat for all voltage ranges from 200 mV to 2 LV 6 Calibrate current gains from 316 nA to 10 pA Remain on the 200 KQ resistance range Follow procedure descri
258. n the positive side of the differential input amplifier AC voltage output of the differential input amplifier Conditioned AC input voltage at input of the product detector The next instrument setup screen appears as a prompt for the monitor output voltage setting Use the A or Y key to select one of the monitor output voltages Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display 4 7 5 Range and Excitation Keys for One or All Channels When a scanner is attached the range and excitation keys are normally used to change parameters for the active scanner channel If all of the resistors connected to the scanner are the same 1t is convenient for these changes to affect all channels at the same time The Model 370 allows the user to configure the range and excitation keys to change one channel at a time or all channels at once Similar operation is available for input parameters under the Input Setup key by selecting the channel designator All The next instrument setup screen appears as a prompt for the input change keys setting Use the A or Y key to select an input change keys setting of One or All Press Enter to accept the new selection and return to the normal display Press Escape to cancel the new selection and return to the normal display 4 8 OPTIONAL SCANNER OPERATION The Model 370 automatically
259. n to the normal display The next zone setup screen appears as a prompt for the derivative D value Use the data entry keys to enter a derivative value between 1 and 2500 or zero for off Press Enter to accept the new selection and return to the normal display Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display The next zone setup screen appears as a prompt for the manual heater output value Manual output must be entered in percent of full scale current Power units are not available within the zone feature Use the data entry keys to enter a manual output value between 0 and 100 The units multiplier is determined by the heater range and can not be changed Press Enter to accept the new selection and return to the normal display Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display The next zone setup screen appears as a prompt for the heater range Heater ranges are entered by their full scale current Power units are not available within the zone feature Use the A or Y key to select on or off Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display 5 8 Advanced Operation Lake Shore Model 370 AC Resistance Bridg
260. nce repeatability The closeness of agreement among repeated measurements of the same variable under the same conditions resistance temperature detector RTD Resistive sensors whose electrical resistance is a known function of the temperature made of e g carbon glass germanium platinum or rhodium iron A 6 Glossary of Terminology Lake Shore Model 370 AC Resistance Bridge User s Manual resolution The degree to which nearly equal values of a quantity can be discriminated display resolution The resolution of the physical display of an instrument This is not always the same as the measurement resolution of the instrument Decimal display resolution specified as n digits has 10 possible display values A resolution of n and one half digits has 2 x 10 possible values measurement resolution The ability of an instrument to resolve a measured quantity For digital instrumentation this is often defined by the analog to digital converter being used A n bit converter can resolve one part in 2 The smallest signal change that can be measured is the full scale input divided by 2 for any given range Resolution should not be confused with accuracy RhFe Rhodium iron Rhodium alloyed with less than one atomic percent iron is used to make the Lake Shore RF family of sensors Rhodium iron is a spin fluctuation alloy which has a significant temperature coefficient of resistance below 20 K where most metals rapidly lose sensitivity
261. ncy allowing the grease to form a cushion between mating surfaces Melting point is 316 K 43 C Use Xylene with Isopropyl alcohol rinse to remove it MAN 370 Model 370 AC Resistance Bridge User s Manual Rack Mounting Kit for Model 370 AC Resistance Bridge Hardware to attach one Model 370 to RM i a 483 mm 19 inch rack mount cabinet IMI 7031 Varnish formerly GE 7031 Varnish 1 pint can IMI 7031 Insulating Varnish and Adhesive possesses electrical and bonding properties which when combined with its chemical resistance and good saturating properties make it an excellent material for cryogenic temperatures VGE 7031 As an adhesive IMI 7031 bonds a variety of materials has fast tack time and may be air dried or GAN 25 baked It is also an electrically insulating adhesive at cryogenic temperatures and is often used as a calorimeter cement When soaked into cigarette paper it makes a good high thermal conductivity low electrical conductivity heat sinking layer Maximum operating temperature 423 K 150 C Lake Shore Cryogenic Wire Lake Shore sells the following types of cryogenic wire DT Duo Twist MN Single Strand MW Manganin NC Nichrome Heater Wire ND Heavy Duty QL Quad Lead and QT Quad Twist Lake Shore Coaxial Cable Lake Shore sells the following types of coaxial cable CC Ultra Miniature Coaxial Cable SR Semi Rigid Coaxial Cable CRYC CryoCableTM Accessories incl
262. nd use proper syntax Old Installation No Longer Working 7 Power instrument off then on again to see if it is a soft failure 8 Power computer off then on again to see if communication port is locked up 9 Verify that Baud rate has not been changed on the instrument during a memory reset 10 Check all cable connections Intermittent Lockups 11 Check cable connections and length 12 Increase delay between all commands to 100 ms to make sure instrument is not being over loaded Remote Operation 6 17 Lake Shore Model 370 AC Resistance Bridge User s Manual 6 3 COMMAND SUMMARY This paragraph provides a listing of the IEEE 488 and Serial Interface Commands A summary of all the commands is provided in Table 6 6 All the commands are detailed in Paragraph 6 3 1 which is presented in alphabetical order Sample Command Format FILTER Filter Parameter Command Input FILTER lt channel gt lt off on gt lt settle time gt lt window gt term Format nn n nnn nn lt channel gt Specifies which channel to configure 1 16 0 all channels lt ofi on gt Specifies the filter operating state 0 Off 1 On lt settle time gt Specifies filter settle time 1 200 seconds lt window gt Specifies what percent of full scale reading limits the filtering function Reading changes greater than this percentage reset the filter 1 80 Example FILTER 5 1 10 2 term Filter channel 5 data through a filter with a settling time of
263. ned lt mode gt lt excitation gt lt range gt lt autorange gt lt cs off gt term Format n nn nn n n Refer to command for description RDGST Reading Status Query Input RDGST lt channel gt term Format nn lt channel gt Specifies which channel to query 1 16 Returned lt status bit weighting gt term Format nnn Remarks The integer returned represents the sum of the bit weighting of the status flag bits A 000 response indicates a valid reading is present If the channel being queried is not the actively scanning channel the status will represent the last state prior to the channel change Bit Bit Weighting Status Indicator 0 1 CS OVL 1 2 VCM OVL 2 4 VMIX OVL 3 8 VDIF OVL 4 16 R OVER 5 32 R UNDER 6 64 T OVER 7 128 T UNDER RELAY Relay Parameter Command Input RELAY lt relay number gt lt mode gt lt channel alarm gt lt alarm type gt term Format n n nn n lt relay number gt Specifies which relay to configure 1 low relay 2 high relay lt mode gt Specifies relay mode 0 Off 1 On 2 Alarms 3 Zone lt channel alarm gt Specifies which channel alarm activates the relay when the relay is in alarm mode 1 16 0 scan channel lt alarm type gt Specifies the channel alarm type that activates the relay when the relay is in alarm mode 0 Low Alarm I High Alarm 2 Both Alarms Example RELAY 1 2 2 0 term Relay 1 is setup in Alarms mode and activates when Channel 2 low alarm activat
264. nes for Realizing the International Temperature Scale of 1990 ITS 90 by B W Magnum and G T Furukawa Unfortunately for those working at lower temperature ITS 90 is only defined down to 650 mK In the absence of a recognized international standard those knowledgeable in the field have created a working standard below 650 mK Lake Shore has worked closely with NIST to create a means of interpolating between superconducting fixed points using a secondary standard germanium resistance thermometer GRT The process has been verified by other laboratories to match accepted working standards to within a few millikelvin down to a temperature of 48 mK Lake Shore presently calibrates temperature sensors down to 50 mK using this method We are looking forward to an extended low temperature standard in 2002 2 3 CRYOGENIC COOLING SYSTEM TERMINOLOGY There are many different cryogenic cooling systems used in laboratories today and it would be impossible to describe them all in this manual This section briefly illustrates the three most common cooling systems capable of reaching temperatures below 1 K This is an attempt to demonstrate how the Model 370 can be integrated into these systems for resistance measurement temperature measurement and temperature control It is hoped that from these examples those knowledgeable in cryogenic techniques will become familiar with the terminology used throughout this manual and the intended operation of many of the in
265. ng Minimum and Maximum Parameter Query MNMX lt channel gt term nn lt channel gt Specifies which channel to query 1 16 lt source gt term n Refer to command for description Minimum and Maximum Reset Command MNMXRST term Resets the minimum and maximum data for all channels Active scan channel will reset immediately while all other channels will reset with the first valid reading they receive IEEE Interface Mode Command MODE lt mode gt term n lt mode gt 0 local I remote 2 remote with local lockout MODE 2 term Places the Model 370 into remote mode with local lockout IEEE Interface Mode Query MODE term lt mode gt term n Refer to command for description Monitor Output Selection Command MONITOR lt source gt term n lt source gt 0 Off 1 CS NEG 2 CS POS 3 VAD 4 VCM NEG 5 VCM POS 6 VDIF 7 VMIX Selects the source of the Monitor Output Monitor Output Selection Query MONITOR term lt source gt term n Refer to command for description Manual Heater Output Command MOUT lt value gt term nnn nnnEtnn lt value gt Specifies value for manual output in percent current or power Depending on the heater output selection MOUT 22 45 term Control Loop manual output is 22 45 of full scale current MOUT 1 0E 03 term Control Loop manual output is 1 mW Remote Operation 6 31 Lake Shore Model 370 AC Resistance Bridge Us
266. ng The Model 370 has an autoscan feature that will sequence through all enabled scanner channels In addition to setting an appropriate change pause time and filter time constant the user must set the scan dwell time for each channel as described in Paragraph 4 6 2 The scan dwell time is the length of time the instrument will take valid readings on the active channel before moving on to the next channel The scan dwell time takes place after the hardware settling time change pause time and filter settling If the instrument is used to control temperature and autoscan at the same time the control algorithm will override the autoscan dwell times To turn on the autoscan feature press the Autoscan key The LED annunciator above that key will light to indicate autoscan is on To turn off the autoscan feature press the Autoscan key again 4 8 4 Settings for Multiple Channels Most often parameters related to the measurement input are changed for the active scanner channel only If all of the resistors connected to the scanner are the same it is convenient for these changes to affect all channels at the same time The Model 370 also allows the user to make changes to the input parameters for all channels All channels can be selected for parameters under the Input Setup key by selecting the channel designator All Paragraph 4 7 5 describes how to configure the range and excitation keys to change all channels 4 9 FIRMWARE FILTER The firmware fi
267. nge The autorange feature allows the instrument to automatically choose an appropriate resistance range The instrument will increase the resistance range decrease input gain if the measurement reaches full scale Normally no over range error is created because of the approximate 20 over range capability of the input Over range errors are possible when resistance changes very quickly or when resistance increases beyond the limit of the current excitation selection The autorange algorithm will not change the excitation setting to accommodate larger resistances The instrument will decrease the resistance range increase input gain if the measurement drops below 20 of full scale Autorange is not recommended when resistance changes are very fast measurements are very noisy or when data acquisition cannot tolerate a range change delay in the middle of a data set To turn autorange on or off press the Autorange key The LED annunciator above the key lights to indicate autorange is on 4 5 3 Turning Excitation Off The Model 370 offers many excitation levels with the lowest being very low in power It is still sometimes necessary to turn excitation off Excitation should be turned off any time the current source leads are open circuited and then attached to a cold resistor with the instrument turned on This condition may occur when the input is scanned or multiplexed externally Excitation is turned off automatically for channel changes when the Mode
268. non latching Operation Use the data entry keys to enter the deadband value between I x 10 and 100x100 or zero depending on selected units Enter the numeric value first Press Enter to accept the new value and advance to the units multiplier Use the A or Y key to select a units multiplier of m blank k or M Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display The next alarm setup screen appears as a prompt for alarm On Off If the alarm feature is turned off this will be the first setting screen to appear Use the A or V key to select alarm on or off Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next alarm setup screen appears as a prompt for audible alarm beeper On Off The audible alarm is turned on or off for all channels at the same time Use the A or Y key to select audible alarm on or off Press Enter to accept the new selection and return to the normal display Press Escape to cancel the new selection and return to the normal display To reset latched alarms clear the alarm condition and press the Alarm key twice The following message appears as a prompt after the key is pressed the first time Latched
269. nt excitation source provides stable reliable low power excitation current Twenty one AC current levels from 3 16 pA to 31 6 mA RMS offer low noise with no significant DC component to contribute to sensor self heating Two operating modes provide excitation control options to meet user and application needs Performance Enhancement Innovative features enhance the performance of the Model 370 from excitation to output Common mode voltage can come from many sources including external noise coupling into the lead wires The Model 370 provides a unique patented matched impedance current source as its first defense against common mode noise Just as voltage input terminals for a differential input have the same input impedance the two current source output terminals of the Model 370 have the same source impedance This matched impedance ensures that common mode voltages do not become normal mode voltages With this strategy the differential input remains truly differential for accurate resistance measurement To further reduce the effect of common mode voltage the Model 370 includes an active common mode reduction circuit This circuit dynamically adjusts the current source output operation point to minimize common mode voltage at the measurement input Active common mode reduction allows the Model 370 to operate in environments that would otherwise saturate the differential input amplifiers Optocouplers isolate the analog front end of the Model 370
270. ntered for these display values to be meaningful A 100 heater is required for maximum power output but other values can be used If a heater resistance is not known the heater output display should be configured for current The next control setup screen appears as a prompt for the heater resistance Use the data entry keys to enter a heater resistance between 1 and 100 000 ohms Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display 4 11 2 8 Heater Range Limit The one watt maximum heater output power available from the Model 370 can be far too much for some loads The heater range limit parameter allows the user to lock out higher heater ranges so they are not accidentally selected from the front panel The limit setting represents the highest allowable heater range and is always set in current To disable the control output entirely refer to Paragraph 4 11 1 The next control setup screen appears as a prompt for the heater range limit Use the A or Y key to select a heater range limit from 100 mA to 31 6 uA as limit Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel and return to the normal display 4 18 Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 4 11 2
271. ny user curves 1 to 20 can be changed Entering the identification parameters associated with the curve is as important as entering the breakpoints Curve header parameters are listed in Table 5 1 Typical parameters for common sensors are listed in Table 5 2 Read this section completely and gather all necessary data before beginning the process NOTE If the curve you wish to enter has similar parameters to an existing curve first copy the similar curve as described in Paragraph 5 2 3 to a new location then edit the curve to the desired parameters Table 5 1 Curve Header Parameters The curve name cannot be changed from the front panel Curve names can only be entered over the Name computer interface up to 15 characters The default curve name is User Curve xx where xx is the curve number Serial Specific sensors serial numbers up to 10 characters The serial number field accepts both numbers and Number letters over computer interface but the instrument front panel enters only numbers The default is blank The instrument must know the data format of the curve breakpoints Different sensor types use different data formats The sensor inputs require one of the formats below The range and resolution specified are not always available at the same time Practical range and resolution depend on the sensor type aa Sensor Units Sensor Units orma Format Description Full Scale Range Maximum Resolution Q K Resistance vs Ke
272. o the same ground Good grounding of all chassis in a group of instruments prevents voltage differences between their chassis usually at line frequency from inducing noise on the measurement If good quality Earth ground is not available the chassis of line powered instruments should be strapped together with heavy gauge copper wire or ground strap Signal leads in communication cables should not be relied on for ground strapping Measurement circuits should be as well isolated from Earth ground as possible with one important exception Measurement leads outside the Dewar should be shielded and in general protected from contact with Earth ground Inside care must be taken to prevent electrical shorts between leads and conductive parts of the cooling system During installation it is important to check for these shorts as well as shorts between the leads themselves Guards should be terminated at the instrument end of cables only and not at the resistor end Monitor and reference voltage outputs and scanner control lines should be treated like part of the measurement and connected to isolated or differential measurement inputs The only unanswered question is Where should cable shields be terminated Shield termination is the exception to the rule that isolated measurement circuits should be isolated from Earth ground In an ideal installation the shield of the instrument can be joined with the natural shielding of a metal test Dewar to shield all intern
273. odel 370 When ordering 8002 05 370 please specify your instrument serial number and calibrated sensor model and serial number A new NOVRAM will be sent for Customer installation 7 33 ACCESSORIES Accessories are devices that perform a secondary duty as an aid or refinement to the primary unit Refer to the Lake Shore Temperature Measurement and Control Catalog for details A list of accessories for the Model 370 follows Description Of Accessories Sensor Input Mating Connector 6 pin DIN plugs for Input connections 2 included DB 25 Plug Quantity of 4 included DB 25 Hood Quantity of 4 included Terminal Block Mating Connector 6 pin terminal block for the Alarm Relays output Scanner Box Mounting Bracket Used to mount the Model 3716 Scanner 112 374 Cable Assembly Connects Model 370 to optional scanner 3 meters 10 feet long See Figure 7 1 ES IEEE 488 Cable Kit One meter 3 feet long IEEE 488 GPIB computer interface cable 4005 assembly Includes extender required to use both IEEE cable and relay terminal block simultaneously Stycast Epoxy 2850 FT Catalyst 9 20 packets 2 grams each Stycast is a common highly versatile nonconductive epoxy resin system for cryogenic use The primary use for Stycast is for vacuum feedthroughs or permanent thermal anchors Stycast is an alternative to Apiezon N Grease when permanent sensor mountings are desired Indium Solder Disks Quantity 10 Indium is a semi precious non f
274. odel 370 AC Resistance Bridge User s Manual 4 11 2 6 Heater Output Display The heater output or control output is a variable DC current source with several ranges to accommodate different loads The Model 370 can be configured to display the output range as its full scale current and the output as a percent of that current This configuration is helpful in setup and tuning of a control system but does not provide good information on power entering the load The instrument can also be configured to display the output range by its full scale power and the output as power in watts This configuration is better for predicting the power entering the load but it relies on an accurate heater resistance being entered into the instrument refer to Paragraph 4 11 2 7 The manual output parameter also follows this units setting The next control setup screen appears as a prompt for the heater output display to be in current or power Use the A or Y key to select heater output display as current or power Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display 4 11 2 7 Heater Resistance The Model 370 does not need to know heater resistance to perform control its control functions Heater resistance is only required to accurately display the heater output and manual output in power The instrument does not measure heater resistance so one must be e
275. oeeeeseeesecuansesnepentessenessaseders 5 1 5 1 2 Curve Breakpoints iii A A W OLA 5 1 5 1 3 CalC rv Mia iii 5 2 5 2 FRONT PANEL CURVE ENTRY OPERATIONS oococococccnconcncnononcnononnnnnononnncnnnoncnn non cnn nan nn nc naar nr rnnnn nos 5 2 5 2 1 Edit CUPE iia iii lili 5 2 5 2 2 Erase GUNVE rta cie 5 5 5 2 3 Copy Curve a e Miah tie a aan a ened 5 6 5 3 ZONE TUNING site een ea PAWEL a A i ee 5 7 5 3 1 Entering Zone Parameter Values sieisen ieena kia niania NE aana inai 5 7 5 3 2 Using Z ne TUNING vecinos ars eraispena wt ea eiaa pe ia eii 5 10 iv Table of Contents Lake Shore Model 370 AC Resistance Bridge User s Manual TABLE OF CONTENTS Continued Chapter Paragraph Title Page 6 COMPUTER INTERFACE OPERATION ccssssseecssseesseeesseeesseeseseeeeseeseseeeaseeseseeesseeseseeseeeseseeeeseeeseseesseesaseeeasees 6 1 6 0 GENERA lu seta eee vee fa eee I Az A A A Wd Stee yea 6 1 6 1 IEEE 488 INTERFACE iii oa 6 1 6 1 1 Changing IEEE 488 Interface Parameters eee aaa aaa aaa aaa aaa 6 1 6 1 2 Remote Local OperaliOmi ssh OE a AE O A Eo dd ad 6 2 6 1 3 IEEE 488 Command Structure eee aeaea aaa aaa aaa aaa aaa aaa aaa ana 6 2 6 1 4 Status RegiSsterS EE ER OWE AE O EAP do A Z ERA E LA 6 4 6 1 4 1 Status Byte Register and Service Request Enable Register eeaao aaa nwa 6 4 6 1 4 2 Standard Event Status Register and Standard Event Status Enable Register
276. of the screen the Comm Control will have appeared as a telephone icon Select the Comm control and add it to the form 7 Add controls to form a b c d Add three Label controls to the form Add two TextBox controls to the form Add one CommandButton control to the form Add one Timer control to the form 8 On the View Menu select Properties Window 9 In the Properties window use the dropdown list to select between the different controls of the current project Label Command1 im Serial Interface Program Label3 Label2 10 Set the properties of the controls as defined in Table 6 4 11 Save the program 6 14 Remote Operation Lake Shore Model 370 AC Resistance Bridge User s Manual Table 6 4 Serial Interface Program Control Properties Current Name Property New Value Labell Name IblExitProgram Caption Type exit to end program Label2 Name IblCommand Caption Command Label3 Name IblResponse Caption Response Textl Name txtCommand Text lt blank gt Text2 Name txtResponse Text lt blank gt Command l Name cmdSend Caption Send Default True Forml Name frmSerial Caption Serial Interface Program Timerl Enabled False Interval 10 12 Add code provided in Table 6 5 a In the Code Editor window under the Object dropdown list select General Add the statement Public gSend as Boolean Double Click on cmdSend Add code segmen
277. off The default value for I is 20 seconds The next PID setup screen appears as a prompt for an I value Use the data entry keys to enter an integral value between 1 and 10 000 or zero for off Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display 4 11 4 3 Manually Setting Derivative D The derivative parameter also called rate is the D part of the PID equation It has a setting range of 1 to 2500 seconds A setting of zero turns D off The default value for D is zero off The next PID setup screen appears as a prompt for a D value Use the data entry keys to enter a derivative value between 1 and 2500 or zero for off Press Enter to accept the new selection and return to the normal display Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display 4 20 Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 4 11 5 Manual Heater Output Manual heater output is most often used during open loop control but it is also active during closed loop PID control Operation of the parameter is different during the two control modes During open loop control the instrument uses the manual output value to calculate the heater output on a given heate
278. og output selection Use the A or Y key to select analog output 1 or 2 Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next analog output setup screen appears as a prompt for bipolar unipolar Use the A or V key to select between bipolar and unipolar Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next analog output setup screen appears as a prompt for the analog output mode Use the A or Y key to select one of the following modes Input Manual Zone Still for Analog Out 2 only or Off Press Enter to accept the new selection and continue to the next setting screen If the manual mode is selected the next setting screen will be for manual output value If input is selected the next setting screen will be for input channel Press Escape to cancel the new selection and return to the normal display 4 30 Operation Lake Shore Model 370 AC Resistance Bridge User s Manual Analog Outputs Continued If manual mode is selected the next analog output setup screen appears as a prompt for the manual output value Use the data entry keys to enter a manual output value 100 and 100 in bipolar mode or 0 and 100 in unipolar mode Press Enter to accept the n
279. ol heater can be turned off but not changed in range and latched alarms can be reset when the keypad is locked A 3 digit code must be used to lock and unlock the keypad The factory default code is 123 and it can only be changed using a computer interface If the instrument parameters are set to default values the code is reset to the factory default The instrument parameters cannot be reset to default values from the front panel when the keypad is locked The following message will appear on the display if the user attempts to change a locked parameter NOTE The computer interface has a remote operation mode that may be mistaken for a locked keypad If the display shows the Remote annunciator on the top line press the Local key to regain local control of the instrument To lock the instrument keypad press and hold the Enter key for 5 seconds The following screen appears as a prompt for the 3 digit lock code Use the data entry keys to enter the 3 digit lock code default 123 The instrument returns to the normal display with the keypad locked To unlock the instrument keypad press and hold the Enter key for 5 seconds The following screen appears as a prompt for the 3 digit lock code Use the data entry keys to enter the 3 digit lock code default 123 The instrument returns to the normal display with the keypad unlocked 4 19 DEFAULT PARAMETER VALUES It is sometimes desirable to reset in
280. oling action takes place Liquid He is pumped under vacuum to reduce its boiling point to approximately 300 mK Best location for temperature control sensor and resistive heater The heater output can control temperature by driving the heater under closed loop PID control or open loop control Load Sample Holder End of the He evaporator where experiments can take place May either be part of or attached to the evaporator and is often designed for a unique purpose Requires a separate temperature sensor if its temperature is different from the He evaporator Theory of Operation 2 3 Lake Shore Model 370 AC Resistance Bridge User s Manual 2 3 3 Adiabatic Demagnetization Refrigerator ADR The principals of ADR have been known for some time and many very capable systems have been built over the years There has recently been growing interest in the technology and it is finding its way into many new applications The Model 370 has many features that can help automate and control these systems in temperature ranges where leaded resistance measurement is practical Figure 2 3 illustrates a typical first stage of an ADR but other stages can be added Helium Dewar Vacuum insulated Dewar large enough to contain the following components Sometimes includes a nitrogen outer jacket to improve efficiency Contains Helium superconducting magnet if the Dewar first magnetic cooling stage has 1 5 8 stag Vacuum a TRASS Insulation He Rese
281. oltage of a 100 kQ resistor at room temperature using a 1 second filter is approximately 80 nV PP reducing it by a decade to 8 nV PP would require increasing time constant 100 s bandwidth to 10 ms This reality is unappealing to most investigators because actual changes in resistance that take less than a few hundred seconds could easily be missed 2 8 MODEL 3716 L and 3708 SCANNER The Model 3716 L Scanners expand the input capability of the Model 370 from 1 to 16 channels The 3716L provides different measurement circuitry to allow better low resistance measurements The 3708 expands the input capability of the Model 370 from 1 to 8 channels and provides an even lower noise low resistance measurement Refer to section 1 1 for details on the differences between the available scanners The scanners are made up of relay switches for current and voltage signals of each channel a times 100 amplifier for the voltage signals going back to the Model 370 and necessary control circuits The scanner is housed in a separate shielded enclosure so that it can be placed near the test Dewar Power for the scanner is taken directly from the regulated DC power supplies of the Model 370 Mounting bracket and 10 foot cables to join the Model 370 are included with the scanner Mating connectors are supplied for the scanner inputs AII scanner circuits including its power supplies and enclosure are referenced to the isolated measurement common in the instrument This
282. oltage output 2 None of the analog output hardware changes when it is configured for still operation Paragraph 3 8 3 describes how to estimate the full scale current sourced to a resistive heater When analog output 2 is in still mode the Still key gives direct access to the still output parameter which is used to set a percentage of the full scale current The still output parameter is the same as the manual output parameter in manual mode but the setting can be made without sequencing through the analog output menu To use Analog Output 2 as a still heater output configure the analog output in still mode as described in Paragraph 4 16 Unipolar operation is recommended for this application Press the Still key and the following screen appears as a prompt for the output value Still Use the data entry keys to enter a still output value 100 and 100 in bipolar mode or 0 and 100 in unipolar mode Press Enter to accept the new selection and return to the normal display Press Escape to restart the setting sequence and enter a different value Press Escape again to cancel the sequence and return to the normal display 4 32 Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 4 18 LOCKING THE KEYPAD The keypad lock feature prevents accidental changes to parameter values When the keypad is locked most parameter values may be viewed but only a few can be changed from the front panel The temperature contr
283. oltage selector and fuses Remove and discard both existing fuses Replace with proper Slow Blow time delay fuse ratings as follows Qu R wer 100 120V 0 5AT250V 5x20 mm 220 240V 0 25AT250V 5x20 mm Re assemble line input assembly in reverse order Verify voltage indicator in the line input assembly window Connect instrument power cord Turn power switch On o 0 Ja 8 6 ERROR MESSAGES The following messages appear on the instrument display when it identifies a problem during operation The messages are divided into three groups Instrument hardware messages are related to the instruments internal circuits or non volatile memory If one of these messages persists after power is cycled the instrument requires repair or recalibration Limit messages are most often associated with over voltage conditions caused by an improperly selected resistance range or excessive noise on the measurement leads Paragraph 8 7 can be helpful in diagnosing these problems If these messages persist after the input or output is configured properly the instrument may require repair Illegal operation messages remind the operator when a feature is locked out or the instrument is not configured to support a feature Instrument Hardware Defective NOVRAM NOVRAM is defective and must be replaced Invalid NOVRAM Contents NOVRAM contents are corrupt and parameter values will be reset to default Reoccurrence may indicate defective NOVRAM Calibra
284. olts cannot be seen felt or heard 8 9 1 Identification of Electrostatic Discharge Sensitive Components The following are various industry symbols used to label components as ESDS so A Bee 8 9 2 Handling Electrostatic Discharge Sensitive Components Observe all precautions necessary to prevent damage to ESDS components before attempting installation Bring the device and everything that contacts it to ground potential by providing a conductive surface and discharge paths As a minimum observe these precautions De energize or disconnect all power and signal sources and loads used with unit Place unit on a grounded conductive work surface Ground technician through a conductive wrist strap or other device using 1 MQ series resistor to protect operator BO DND Ground any tools such as soldering equipment that will contact unit Contact with operator s hands provides a sufficient ground for tools that are otherwise electrically isolated 5 Place ESDS devices and assemblies removed from a unit on a conductive work surface or in a conductive container An operator inserting or removing a device or assembly from a container must maintain contact with a conductive portion of the container Use only plastic bags approved for storage of ESD material 6 Do not handle ESDS devices unnecessarily or remove from the packages until actually used or tested 8 10 FIRMWARE REPLACEMENT There are three integrated circuits ICs that may po
285. ommand Execution Error EXE Bit 4 This bit is set if an execution error has been detected This occurs when the instrument is instructed to do something not within its capabilities Query Error QYE Bit 2 This bit indicates a query error It occurs rarely and involves loss of data because the output queue is full Operation Complete OPO Bit 0 This bit is generated in response to the OPC common command It indicates when the Model 370 has completed all selected pending operations It is not related to the OPC command which is a separate interface feature 6 1 5 IEEE Interface Example Programs A Visual Basic program is included to illustrate the IEEE 488 communication functions of the instrument Instructions for setting up the IEEE 488 board is included in Section 6 1 5 1 Refer to Section 6 1 5 2 for instructions on how to setup the program The Visual Basic code is provided in Table 6 2 A description of program operation is provided in Section 6 1 5 3 While the hardware and software required to produce and implement these programs is not included with the instrument the concepts illustrated apply to most applications 6 1 5 1 IEEE 488 Interface Board Installation for Visual Basic Program This procedure works for Plug and Play GPIB Hardware and Software for Windows 98 95 This example uses the AT GPIB TNT GPIB card Install the GPIB Plug and Play Software and Hardware using National Instruments instructions
286. ommand for description CHGALL Input Change All or One Command Input CHGALL lt state gt term Format n lt state gt Specifies if Range and Excitation keys affect all channels together or individually 0 One 1 All Remarks Only active when the Model 3716 Scanner is installed CHGALL Input Change All or One Query Input CHGALL term Returned lt state gt term Format n Refer to command for description CMODE Temperature Control Mode Command Input CMODE lt mode gt term Format n lt mode gt Specifies control mode 1 Closed Loop PID 2 Zone Tuning 3 Open Loop 4 Off Example CMODE 3 term Set control mode to Open Loop control CMODE Temperature Control Mode Query Input CMODE term Returned lt mode gt term Format n Refer to command for description CMR Common Mode Reduction Command Input CMR lt state gt term Format n lt state gt 0 O0ff 1 On Remarks Enables or disables common mode reduction CMR Common Mode Reduction Query Input CMR term Returned lt state gt term Format n Refer to command for description 6 24 Remote Operation CPOL Input Format Remarks CPOL Input Returned Format CRVDEL Input Format Example CRVHDR Input Format Remarks Example CRVHDR Input Format Returned Format CRVPT Input Format Remarks Example Lake Shore Model 370 AC Resistance Bridge User s Manual Temperature Control Polarity Command
287. onvenient guide interpolation table A table listing the output and sensitivity of a sensor at regular or defined points which may be different from the points at which calibration data was taken intrinsic coercivity The magnetic field strength H required to reduce the magnetization M or intrinsic induction in a magnetic material to zero intrinsic induction The contribution of the magnetic material B to the total magnetic induction B B B u4H SD B B H cgs IPTS 68 International Practical Temperature Scale of 1968 Also abbreviated as Tss isolated neutral system A system that has no intentional connection to ground except through indicating measuring or protective devices of very high impedance ITS 90 International Temperature Scale of 1990 Also abbreviated as Too This scale was designed to bring into as close a coincidence with thermodynamic temperatures as the best estimates in 1989 allowed Kelvin K The unit of temperature on the Kelvin Scale It is one of the base units of SI The word degree and its symbol are omitted from this unit See Temperature Scale for conversions Kelvin Scale The Kelvin Thermodynamic Temperature Scale is the basis for all international scales including the ITS 90 It is fixed at two points the absolute zero of temperature 0 K and the triple point of water 273 16 K the equilibrium temperature that pure water reaches in the presence of ice and its own vapor line regu
288. oop area but also guarantees that both leads are exposed to the same common mode effects which are rejected by the measurement input 2 4 5 2 Shields Shields provide a low impedance path to measurement common to block E field noise from getting to the signal leads Shield pins are provided in the input connectors as attachment points for the shield conductors in lead cables Shielding individual leads inside a cryostat is difficult because of limited space and the potential for heat leak Cable shields are usually connected to the experimental Dewar to create a faraday shield around the measurement 2 4 5 3 Electrical Isolation The precision analog front end is electrically isolated from digital circuitry and chassis of the instrument by optocouplers This can improve measurements in two ways First isolation makes it more difficult for digital noise from control circuits and computer interfaces to affect the measurement leads Second isolation breaks ground loops It is often desirable to extend shielding from the measurement leads to the experimental Dewar If the instrument shield and the Dewar both have low impedance to Earth ground a ground loop will be created Isolation separates the instrument measurement from Earth ground to break the loop 2 4 5 4 Driven Guards For installations where lead length is greater than 10 feet or measured resistance is greater than 100 kQ four separate driven guards are provided The driven guards follow the
289. or example the resistance of a PTC sensor increases with increasing temperature pounds per square inch psi A unit of pressure 1 psi 6 89473 kPa Variations include psi absolute psia measured relative to vacuum zero pressure where one atmosphere pressure equals 14 696 psia and psi gauge psig where gauge measured relative to atmospheric or some other reference pressure ppm Parts per million e g 4 x 10 is four parts per million precision Careful measurement under controlled conditions which can be repeated with similar results See repeatability Also means that small differences can be detected and measured with confidence See resolution prefixes SI prefixes used throughout this manual are as follows Factor Prefix Symbol Factor Prefix Symbol 10 yotta Y 107 deci d 107 zetta Z 107 centi c 105 exa E 107 milli m 10 peta P 10 micro u 10 tera T 10 nano n 10 giga G 10 pico p 10 mega M 1075 femto f 10 kilo k 1078 atto a 10 hecto h 10 zepto Z 10 deka da 10 yocto y probe A long thin body containing a sensing element which can be inserted into a system in order to make measurements Typically the measurement is localized to the region near the tip of the probe proportional integral derivative PID A control function where output is related to the error signal in three ways Proportional gain acts on the instantaneous error as a multiplier Integral reset acts on the area of error with respect to
290. ort Wait loop Give up processor to other events Loop until Send button pressed Set Flag as false Get Command Clear response display Set all characters to upper case Get out on EXIT Send command to instrument Check to see if query Wait for response Add 1 to timeout if no character Wait for 10 millisecond timer False Timeout at 2 seconds Reset timeout for each character Read in one character Add next character to string Get characters until terminators Check if string empty Term 1 Strip terminators Send No Response Put response in textbox on main form Reset holding string Reset timeout counter Private Sub Timerl Timer frmSerial Timerl Enabled False End Sub Routine to handle Timer interrupt Turn off timer 6 16 Remote Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 6 2 7 2 Program Operation Once the example program is running try the following commands and observe the response of the instrument Input from the user is shown in bold and terminators are added by the program The word term indicates the required terminators included with the response ENTER COMMAND IDN Identification query Instrument will return a string identifying itself RESPONSE LSCI MODEL370 123456 02152001 term ENTER COMMAND RDGR 1 Ohm reading query Instrument will return a string with the present resistance reading from channel 1
291. ove existing IC from the socket 4 Noting orientation of new IC use an IC insertion tool to place new device into socket PU at t wy de DALLAS A 4 XXXXXXXXX ESD XXXXXXXXX P Sensitive A O 1 Match notch on Match notch on IC to notch Typical IC NOVRAM to notch in socket in socket Typical NOVRAM Eprom eps 5 Follow the top of enclosure INSTALLATION procedure in Paragraph 8 8 8 11 CONNECTOR AND CABLE DEFINITIONS The SCAN CONTROL INPUT REFERENCE MONITOR HEATER OUTPUT ANALOG OUTPUT 1 and 2 STILL ALARM RELAYS and RS 232 DTE connectors are defined in Figures 8 5 thru 8 11 t SCAN CONTROL iD gt Scan Control bmp Pin Name Description 1 Scan Present Signal to Model 370 that a scanner is installed 2 Digital Common Common or ground for digital control signals 3 Digital Common 4 Unused 5 Analog Common Common or ground for analog signals 6 Analog Common 7 Analog Common 8 15 V 15 V referenced to scanner analog common 9 Scan 1 Scan channel select line 1 10 Scan 2 Scan channel select line 2 11 Scan 3 Scan channel select line 3 12 Scan 4 Scan channel select line 4 13 Scan Gain Selects scanner gain of 1 or 100 14 5 V 5 V referenced to scanner digital common 15 15 V 15 V referenced to scanner analog common Body Chassis Do Not Use Figure 8 5 SCAN CONTROL Connector Details Service 8 11 Lake Shore Model 370 AC Resistance Bridge User s Manual SHIELD
292. pA i 7 i i gt s 200 Q 200 Q 600 t 3 3 2 fW 1 0 W 320 aW 100 aW 32 aW x ss T 6 32 MQ 2 0 MQ 632 kQ 200 KQ 10 pA kk kk 600 Q 600 Q hi ki hi i i i i 320 aW 100 aW 32 aW 10 aW hi i hi 7 3 6 32 MQ 2 0 MQ 632 kQ 3 16 pA kk kk ak i ld 8 S 32 aW 10 aW 3 2 aW 1 0 pA Accuracy 0 5 Introduction 1 7 Lake Shore Model 370 AC Resistance Bridge User s Manual Accuracy 0 03 Table 1 3 Model 370 3708 Voltage Range 0 05 6 32 mV 31 6 mA 10 mA 3 16 mA 1 0 mA 316 pA 100 pA 31 6 pA 10 pA 3 16 A 1 0 pA 316 nA 100 nA 31 6 nA 10 nA 3 16 nA 10nA 316 pA 200k0 632K0 20kQ 100 pA 60 0 32 0 200 1 0fW 320aW 100aW 632kQ 200kQ 632k0 31 6 pA 1900 2009 630 320aW 100aW 32aW 20MQ 632k0 200k0 dd E 10 pA 20kQ 6300 6000 Accuracy 0 5 100aW 32aW 10aW 20M0 632kQ 3 16 pA 6 0kQ 19kO 1 0 10aW 32aW 1 8 Introduction Lake Shore Model 370 AC Resistance Bridge User s Manual 1 2 SPECIFICATIONS Measurement Type AC 4 lead differential resistance Number of Inputs 1 16 with Model 3716 Scanner Measurement Units Ohms Kelvin with temperature response curve Resistance Ranges 2 mQ to 2 MQ excitation dependent Reading Rate 10 readings per second same range and channel Range Change Settling 3 s filter settling Channel Change Scan Settling 3 s filter settling A D Resolut
293. perature Copper and aluminum are examples of metals that have good thermal conductivity while stainless steel does not Metals that super conduct tend to be better thermal conductors when normal and poor thermal conductors below their transition temperature Non metallic electrically insulating materials like alumina oxide and similar ceramics have good thermal conductivity while G 10 epoxy impregnated fiberglass does not Sensor leads should have poor thermal conductivity Sensor packages cooling loads and sample holders should have good thermal conductivity to reduce temperature gradients Surprisingly the connections between thermally conductive mounting surfaces often have very poor thermal conductivity 2 16 Theory of Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 2 9 3 4 Contact Area Thermal contact area greatly affects thermal conduction because a larger area has more opportunity to transfer heat Even when the size of a sensor package is fixed thermal contact area can be improved with the use of a gasket material A soft gasket material forms into the rough mating surface to increase the area of the two surfaces that are in contact Good gasket materials are soft thin and have good thermal conductivity They must also withstand the environmental extremes Indium foil when not super conducting and cryogenic grease are good examples 2 9 3 5 Contact Pressure When sensors are permanently mounted the solder
294. pper and lower limit to its temperature range The upper limit of an NTC resistor is normally the point at which sensitivity falls so low that the sensor is no longer useful The exact temperature where this happens is determined by the sensor itself instrumentation capabilities and required resolution Sensors that have a few micro kelvin resolution at 50 mK may have a resolution of several kelvin at room temperature The sensor may be useful to monitor warm up and cool down but has little experimental value The lower limit of an NTC resistor 1s the temperature at which self heating makes the measurement impractical As temperature decreases the thermal resistance between sensor and load increases and eventually the heat of excitation can no longer be dissipated into the load A nice feature of NTC resistors is that resistance increases as temperature decreases and if the voltage across the sensor is limited excitation power in the sensor decreases to offset the increase in thermal resistance Unfortunately there is always a temperature at which self heating wins and it is usually between 10 mK and 50 mK The exact temperature depends on the sensor construction mounting and instrumentation used for readout 2 9 1 3 Temperature Resolution Temperature measurement resolution is controlled to a large extent by sensor sensitivity when using a resistance bridge The temperature example given in the instrument specifications shows the sensitivity of a typica
295. r The instrument will respond only to the last query it receives The response can be a reading value status report or the present value of a parameter Response data formats are listed along with the associated queries in Paragraph 6 3 Remote Operation 6 3 Lake Shore Model 370 AC Resistance Bridge User s Manual 6 1 4 Status Registers There are two status registers the Status Byte Register described in Paragraph 6 1 4 1 and the Standard Event Status Register in Paragraph 6 1 4 2 6 1 4 1 Status Byte Register and Service Request Enable Register The Status Byte Register contains six bits of information about the operation of the Model 370 STATUS BYTE REGISTER FORMAT Bit A AO O Weighting KREM INC WE W ATI ER SZM Bit Name Ramp a Alarm Valid Not Not Done Read Used Used If Service Request is enabled any of these bits being set will cause the Model 370 to pull the SRQ management line low to signal the BUS CONTROLLER These bits are reset to zero upon a serial poll of the Status Byte Register These reports can be inhibited by turning their corresponding bits in the Service Request Enable Register to off The Service Request Enable Register allows the user to inhibit or enable any of the status reports in the Status Byte Register The SRE command is used to set the bits If a bit in the Service Request Enable Register is set 1 then that function is enabled Refer to the SRE command discussion Ramp Done Bit
296. r are calibrated properly the calibrations do not interact so scanners and instruments can be interchanged without loss of accuracy To calculate the scanner compensation coefficients perform the following sequence of operations 1 Read and record the voltage gain coefficient for 632 mV range Range 12 with the following command CALG 2 12 2 Read and record the voltage gain coefficient for the 6 32 mV range Range 8 with the following command CALG 2 8 3 Use CALG 6 1 query to read and record the existing scanner coefficient A safety step to revert to the original calibration if necessary 4 Calculate the scanner compensation coefficient as Scanner Coefficient CALG 12 CALG 8 x 100 101 The coefficient should be 1 0 3 5 Set the scanner compensation coefficient command CALG 6 1 Coefficient 6 Send the CALSAVE command and wait 10 seconds for the calibration to be saved in non volatile memory Alternatively this step can be done after all ranges are calibrated 8 12 2 6 Example Measurement Input Calibration Procedure This calibration procedure was developed because traditional laboratory test equipment cannot accurately measure small AC signals The procedure relies on resistors of known value and the Model 370 being calibrated The resistors are all between 2 Q and 2 MQ in value because it is difficult to purchase stable resistors outside of that range Resistors with temperature stability of 25 ppm C are recommen
297. r as power in watts The user must enter an accurate heater resistance for the power calculation to be meaningful 4 3 KEYPAD DEFINITION The Model 370 has 36 keys separated into 3 groups on the instrument front panel The group of keys closest to the display act directly on the resistance measurement the center group contains temperature control features and the far right group combines instrument setup and data entry 4 3 1 Key Descriptions Autorange On Off Resistance Range AV Voltage Mode Voltage excitation AV Current Mode Current excitation AV Autoscan On Off Scanner Channel AV Zone Setup Manual Output PID Setpoint Control Setup Still Output Heater Range Heater Off Input Setup Instrument Setup Filter Max Min Display Setup Curve Entry Linear AR Alarm Local Computer Interface Analog Output Relay Setup 0 9 Escape A Up Y Down Enter Turn Autorange on or off for resistance range Manually increase or decrease resistance range Enable voltage excitation mode Increase or decrease voltage excitation Enable current excitation mode Increase or decrease current excitation Turn Autoscan on or off if a scanner is installed Manually increment or decrement scanner channel Setup temperature control zones for closed loop control Change open loop control parameter Manual Output Change closed loop control parameters P I and D Change the temperature control setpoint Setup temperature control and se
298. r gain compensation lt value gt Gain calibration constant value Set value to 1 for default setting Remarks Gain calibration constant is always a number near 1 Use CALSAVE to permanently save the gain calibration constants Calibrated Reading Uncalibrated Reading CALZ x CALG 8 24 Service Lake Shore Model 370 AC Resistance Bridge User s Manual Calibration Commands Continued CALG Input Format Returned Format CALSAVE Input Remarks CALVIN Input Returned Format Remarks CALZ Input Format Remarks CALZ Input Format Returned Format Gain Calibration Constant Query CALG lt input gt lt range gt term n nn lt input gt Specifies input or output to query 1 6 lt range gt Dependent on lt input gt lt value gt term nnn nnnE nn Refer to command for description Calibration Save Command CALSAVE term Saves the serial number and all CALZ and CALG calibration constants to the EEPROM Input Voltage Query CALVIN term lt voltage gt term nnn nnnE nn Value returned represents the most recent voltage at the input of the Model 370 for the active channel Zero Offset Calibration Constant Command CALZ lt input gt lt range gt lt value gt term nnn nnn nnnE nn lt input gt Specifies the input or output to assign the zero offset calibration constant 1 Not Used Reserved for Current Excitation 2 Voltage
299. r range This allows the user direct control of the output During closed loop control the manual output value is added to the result of the PID control equation In effect the PID algorithm operates about the manual output setting Manual output should be set to zero for most closed loop control applications Manual output units follow the heater output display units of current or power In current units the manual output parameter is set in percent of full scale current for the selected heater range The setting range is from 0 to 100 with a setting resolution of 0 001 The actual output resolution is close to the setting resolution In power units the manual output parameter is set in watts An accurate heater resistance must be entered under control setup for the power calculation to be accurate The setting range is from 0 to full scale power for the selected heater range with a setting resolution of six digits The actual resolution is limited to the 0 001 resolution of the heater output current source Manual heater output value may be changed by the instrument if the control mode or heater range are changed during active control To enter a manual heater output value press the Manual Output key The following screen appears as a prompt for the manual output value This example illustrates the manual output setting sequence with heater output display units set to power on the 10 mW 10 mA and 100 heater range Use the data
300. ranges from 2 mQ to 2 MQ e 21 excitation levels from 3 16 pA to 31 6 mA e Displays real time sensor excitation power e One sensor input up to 8 or 16 with scanner e PID temperature control e JEEE 488 and RS 232C interfaces alarms relays and analog outputs e Unique noise reduction elements o Patented current source preserves common mode noise rejection o Optically isolated measurement electronics eliminates the potential for ground loops Two 16 channel scanners e Model 3716 scanner is optimized for low DC bias current e Model 3716L scanner is optimized for low noise 8 channel preamp scanner e Model 3708 scanner is optimized for ultra low noise AC resistance measurements 370 AC Resistance Bri s BE S886 a o A n E ZH EG U EW M Front bmp Figure 1 1 Model 370 Front Panel U S Patent 6 501 255 Dec 2002 Differential current source with active common mode reduction Lake Shore Cryotronics Inc Introduction 1 1 Lake Shore Model 370 AC Resistance Bridge User s Manual 1 1 DESCRIPTION The Model 370 AC Resistance Bridge is designed for precise accurate low noise low excitation power AC resistance measurement Its primary application is the measurement of resistance materials in cryogenic environments from 20 mK to 1 K Fully integrated the Model 370 includes features to reduce and control noise at every step of the resistance measurement process A unique patented matched impedance current source an
301. ration within the phase sensitive detector This is likely to be the most common error message displayed It is not uncommon to see this error temporarily when changing measurement parameters Usual causes for this error are include 1 A temporary transient during current voltage or range change 2 Anattempt to measure a resistance that is greater 1 order of magnitude than the selected range 3 A differential noise level that is lower than the VDIF overload level but high enough after the gain stages to exceed the dynamic range of the phase sensitive detector To resolve this condition first verify that a reasonable resistance range has been selected for the measurement Next verify the integrity of the voltage lead connections to the measured resistance If the error persists verify that the voltage pair of leads is twisted and shielded Likewise the current pair of leads should be twisted and shielded Twisted pair wiring reduces the differential signals potentially created by environmental noise However there is no advantage gained by twisting the current pair of leads together with the voltage pair of leads When twisted leads are not an option inside a Dewar the shielding must be verified adequate for the task The Dewar itself acts as a shield when the case is properly connected Ensure that the experimental Dewar case is properly connected to the measurement common of the instrument and earth ground Refer to Installation Paragr
302. rences The first equation is most often used to scale a sensor reading that has a linear response but requires offset and gain correction The second can easily produce a scaled error value when the control setpoint is used for B A summary of parameter selections is listed as follows Equation M X B MX B Value Ohms Value M X B Kelvin SP SP where M Slope of line X Reading data B Offset of line Value Number value entered by user and SP Temperature control setpoint value The instrument will calculate meaningless numbers if this feature is not configured properly For example the setpoint units and reading units for X must match if SP is used in the equation The display resolution can also be misleading For example the scaled error value of 1OAR does offer more measurement resolution than the normal display reading However increasing the slope M to 100 or 1000 will only increase the display resolution and not the measurement resolution To begin linear equation setup press the Linear AR key If a scanner is installed the first filter setup screen appears as a prompt for the channel number for X All of the parameter changes that follow will act on the selected channel If no scanner is installed the sequence will begin below with equation selection Use the A or Y key to select a channel number between 1 and 16 Choose the selection All and the changes will be applied
303. resolution Table 3 1 Heater Resistance Heater Resistance 10Q 100 Q 1000 Q 5000 Q 100 mA 100mW 1W 100mW 20mW 31 6 mA 10mW 100mW 100mW 20mW 10 mA 1mW 10mW 100mW 20mW Output 3 16mA 100 WW 1 mW 10mW 20mW Range 1 mA 10 uW 100 uW 1 mW 5 mW 316 LA 1 uW 10u W 100 uW 500 uW 100uA 100nW 1 uW 10 uW 50 uW 31 6 uA 10 nW 100 nW 1 uW SUW ndicates maximum power range and resistance Indicates power limited by compliance voltage 3 7 5 Heater Output Protection The heater output hardware is internally protected against open circuit and short circuit conditions The output should never be connected to an external power source There are additional firmware features in the Model 370 to help prevent damage to the load The heater will automatically turn itself off if an open circuit condition is detected The instrument will not allow the control setpoint to be larger than the setpoint temperature limit stored with the temperature response curve The heater range can be limited so it is not accidentally set too high from the front panel Installation 3 13 Lake Shore Model 370 AC Resistance Bridge User s Manual 3 7 6 Boosting Heater Power Above 1W An external power supply can be used to boost the power output of a Model 370 if a temperature control application requires more than the 1 W of heater power The control heater output can be used to program a voltage programmable DC
304. ress V Terminate Read on EDS H E Mouse Primary jl E W National Instruments GPIB Interfaces o H IV Set EO with EOS on Write AT GPIB TNT Plug and Play I Sbit EOS Compare E 7 Network adapters Secondary 4 Ports COM amp LPT 10 EOS Byt NONE gt Bite System devices 1 0 Timeout p 10sec X Properties Refresh Remove IV System Controller OK Cancel Figure 6 1 GPIBO Setting Configuration ystem Properties 2 x General Device Manager Hardware Profiles Performance I View devices by type ew da National Instruments GPIB Interfaces Properties HBE Computer General Device Templates CDROM 4 6 Disk drives Y National Instruments GPIB Interfaces w Display adapters Floppy disk controllers Hard disk controllers Device Name 8 Keyboard Monitor 2 Mouse Y N ational Instruments GPIB Interface y 54ND Network adapters DEV12 Attributes y Ports COM amp LPT Interface m Termination Methods Timeouts EH System devi E System devices GPIBO y SendEOlatendotwite O M Terminate Read on EOS 10sec El GPIB Address Serial Poll Primary Y Set EO with EDS on Write A isec y 12 Properties Refresh A I 8 bit EOS Compare Secondary NONE gt fio EOS Byte M Readdress Figure 6 2 DEV 12 Device Template
305. rial communication The original RS 232C standard specifies 25 pins but both 9 and 25 pin connectors are commonly used in the computer industry Many third party cables exist for connecting the instrument to computers with either 9 or 25 pin connectors Paragraph 8 11 1 gives the most common pin assignments for 9 and 25 pin connectors Please note that not all pins or functions are supported by the Model 370 The instrument serial connector is the plug half of a mating pair and must be matched with a socket on the cable Ifa cable has the correct wiring configuration but also has a plug end a gender changer can be used to mate two plug ends together The letters DTE near the interface connector stand for Data Terminal Equipment and indicate the pin connection of the directional pins such as transmit data TD and receive data RD Equipment with Data Communications Equipment DCE wiring can be connected to the instrument with a straight through cable As an example Pin 3 of the DTE connector holds the transmit line and Pin 3 of the DCE connector holds the receive line so the functions complement It is likely both pieces of equipment are wired in the DTE configuration In this case Pin 3 on one DTE connector used for transmit must be wired to Pin 2 on the other used for receive Cables that swap the complementing lines are called null modem cables and must be used between two DTE wired devices Null modem adapters are also available for us
306. riving a resistance of 100 or greater The maximum current output of 0 to 100 mA can be achieved with a 100 Q heater Most stills do not require the maximum current Using a larger resistor will reduce the full scale current and maintain good setting resolution For example a 1000 Q resistor will produce a 0 to 10 mA current range with full setting resolution If a small resistive heater is permanently installed in the cooling system lower current can be created by adding a series resistor in a heater lead outside the Dewar 3 14 Installation Lake Shore Model 370 AC Resistance Bridge User s Manual 3 9 RELAY CONNECTION The Model 370 has two relays normally associated with the high and low alarm features Each relay has a normally open NO common COM and normally closed NC contact for a total of 6 All of these connections are electrically isolated from all other circuits in the Model 370 The relay is in its normal state when it is Off and in its active state when it is On When power to the instrument is turned off the relay stays in the normal state A detachable terminal block provides alarm relay connections which are rated at 30V DC 5A The mate is a 6 position terminal block included with the instrument Lake Shore P N 106 737 The terminal block accepts up to 12 gauge wire but smaller wire is adequate for most applications Slides into ALARM RELAYS slot in rear of Model 370 Use screwdriver to lock or unlock wires ALARM
307. rmation on the Model 370 AC Resistance Bridge models options and accessories 7 1 MODELS The list of Model 370 and optional scanner model numbers is provided as follows Model Description Of Models 3708 Standard AC Resistance Bridge Includes a Lake Shore Model 370 AC Resistance Bridge with a Lake Shore Model 3716 16 Chamnel Scanner Power is configured as shown below 370L Standard AC Resistance Bridge Includes a Lake Shore Model 370 AC Resistance Bridge with a Lake Shore Model 3716 L 16 Channel Scanner Power is configured as shown below 370U Standard AC Resistance Bridge Includes a Lake Shore Model 370 AC Resistance Bridge with a Lake Shore Model 3708 8 Channel Scanner Power is configured as shown below AC Resistance Bridge Only Includes a Lake Shore Model 370 AC Resistance Bridge only 370N i f with no scanner Power is configured as shown below 3716 16 channel scanner for Model 370 3716L Low resistance 16 channel scanner for Model 370 3708 Ultra low resistance 8 channel preamp scanner for Model 370 3708 ARW Ultra low resistance 8 channel preamp scanner for Model 370 Includes 370 upgrade for instruments with main version 9 27 2005 and input version 1 3 or earlier firmware Power Configurations The instrument is configured at the factory for customer selected power as follows VAC 100 Instrument configured for 100 VAC with U S power cord VAC 120 Instrument configured for 120 VAC w
308. rvoir Reservoir of liquid He at i atmospheric pressure that He Reservoir provides bulk cooling down to 1K 4 2 K Pre cools leads entering the vacuum space and acts as a Precooled radiation shield for refrigeration Mass or Previous Stage components Vacuum Space Insulates refrigeration components and provides a clear workspace Pre Cooling Stage Brings the system to a low enough temperature for magnetic cooling to become practical Also acts as a heat sink for the first magnetic conducting cooling stage This can be a Magnet cryogenic refrigerator He bath pumped He cryostat or pumped Sample He cryostat Good location for Holder a temperature sensor for cool down monitoring and ADR Blockeps troubleshooting Figure 2 3 Single Stage ADR Block Diagram Heat Switch Allows thermal contact between the pre cooling stage and salt pill during magnetization then provides thermal isolation during demagnetization cooling The construction of the switch varies depending on temperature but could be superconducting exchange gas gas gap or mechanical Several types of heat switches can be controlled by the still heater or analog output Superconducting Magnet Magnetizes the salt pill while in thermal contact with the pre cooling stage then demagnetizes it when thermally isolated The magnet power supply can be programmed by the analog output and controlled via computer interface Salt Pill Mass of paramagn
309. ry 1 8 nn n n Refer to command for description Digital Output Command DOUT lt bit weighting gt term nn lt bit weighting gt Specifies the manual digital output settings 0 31 Bit Bit Weighting Name 0 1 DI 1 2 D2 2 4 D3 3 8 D4 4 16 D5 This command is only active if no Model 3716 Scanner is installed DOUT 21 term Manually sets digital output lines D1 D3 and DS to logic high and D2 and D4 to logic low Digital Output Query DOUT term lt bit weighting gt term nn Refer to command for description Filter Parameter Command FILTER lt channel gt lt off on gt lt settle time gt lt window gt term nn n nnn nn lt channel gt Specifies which channel to configure 1 16 0 all channels lt off on gt Specifies the filter operating state 0 Off 1 On lt settle time gt Specifies filter settle time 1 200 seconds lt window gt Specifies what percent of full scale reading limits the filtering function Reading changes greater than this percentage reset the filter 1 80 FILTER 5 1 10 2 term Filter channel 5 data through a filter with a settling time of 10 seconds and a 2 of full scale window Remote Operation 6 27 Lake Shore Model 370 AC Resistance Bridge User s Manual FILTER Filter Parameter Query Input FILTER lt channel gt term Format nn lt channel gt Specifies which channel to query 1 16 Returned lt off on gt lt settle time gt lt window gt ter
310. s reduce the effect of cable capacitance on the resistance measurement Cable capacitance is normally not large enough to cause a measurable effect until the measured resistor is 100 kQ or greater Guards should be turned off when measuring lower resistances When guards are turned off they remain at measurement common potential Guards are not a substitute for appropriate cable shielding The next instrument setup screen appears as a prompt for guard drive On Off Use the A or Y key to select guard drive on or off Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display Operation 4 11 Lake Shore Model 370 AC Resistance Bridge User s Manual 4 7 4 Monitor Output Selection The monitor output on the Model 370 is a tool to help the user diagnose measurement problems The output can be set to monitor test voltages from several points inside the instrument The monitor voltages are often used with the reference output An explanation of these voltage is given with other troubleshooting information in Paragraph 8 7 Monitor Output Voltages Output grounded to measurement common AC voltage driving negative side of current source programming resistor AC voltage driving positive side of current source programming resistor DC voltage into the AD converter AC voltage on the negative side of the differential input amplifier AC voltage o
311. sabled the hardware filter remains active with an approximate 200 ms time constant Operation 4 13 Lake Shore Model 370 AC Resistance Bridge User s Manual Firmware Filter Continued The next filter setup screen appears as a prompt for filter On Off Use the A or Y key to choose filter on or off Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display Settling time is the most important filter setting Filter settling time can be set from 1 to 200 seconds The table below shows how settling time is related to number of readings averaged and nominal noise attenuation Filter settling time begins after the pause time finishes on range or channel change Because the Model 370 uses a linear average the measurement bandwidth is not a simple function of frequency Settling Time s Number of Readings Attenuation of Noise 6 60 10 dB 12 120 16 dB 25 250 22 dB 50 500 28 dB 100 1000 34 dB 200 2000 40 dB The next filter setup screen appears as a prompt for filter settling time Use the data entry keys to enter a filter settling time between 1 and 200 seconds Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal di
312. se 4kTRAf where Vnoise is the RMS noise voltage k is Boltzmann s constant 1 38 x 10 J K Tis temperature in Kelvin and Af is the noise bandwidth in Hz Some considerations when using the equation peak to peak noise will be within 5 times RMS noise more than 99 of the time Af can be calculated several ways and a common method is 1 2 f34 where f3 is the 3 dB bandwidth of the dominant low pass filter in the measuring circuit As a practical example the noise voltage of a 100 KQ resistor at room temperature using a 1 second filter is approximately 16 nV RMS or 80 nV PP which is 8 of a measurement made at I uV When the resistor is cooled the thermal noise drops accordingly Please note that thermal noise is normally the biggest noise source in resistor being measured but not the only one Metallic conductors approach the theoretical limit for thermal noise but other materials have additional noise sometimes referred to as flicker noise that has a 1 f spectrum Flicker noise is different for every material is proportional to current and can change with construction techniques including lead attachment Shot noise which is a product of the discrete nature of electron flowing through resistive materials is insignificant in most measurement but can become measurable when current drops into the picoamp range 2 5 6 Interference Between Two Model 370s The phase sensitive detector in the Model 370 effectively cancels noise from external
313. show an error message on the display if the sensor input is outside the range of the breakpoints No special endpoints are required Sensor units are defined by the format setting in Table 5 2 Breakpoint setting resolution is six digits in temperature Most temperature values are entered with 0 001 resolution Temperature values of 1000 K and greater can be entered to 0 01 resolution Temperature values below 10 K can be entered with 0 0001 resolution Temperature range for curve entry is 0 to 1500 K Setting resolution is also six digits in sensor units The curve format parameter defines the range and resolution for sensor units data entry The sensor type determines the practical setting resolution Table 5 2 lists recommended sensor units resolutions for some popular Lake Shore sensors The breakpoints should be entered with the resistance value increasing as point number increases There should not be any breakpoint locations left blank in the middle of a curve The search routine in the Model 370 interprets a blank breakpoint as the end of the curve Advanced Operation 5 1 Lake Shore Model 370 AC Resistance Bridge User s Manual 5 1 3 CalCurveTM The CalCurve service provides the user with a convenient way get the temperature response curve from Lake Shore calibrated sensors into instruments like the Model 370 It can be performed at the factory when calibrated sensors and instruments are ordered together The factory installed CalCurve
314. so see magnetic hysteresis LD Inner diameter IEC International Electrotechnical Commission IEEE Institute of Electrical and Electronics Engineers IEEE 488 An instrumentation bus with hardware and programming standards designed to simplify instrument interfacing The addressable parallel bus specification is defined by the IEEE initial permeability The permeability determined at H 0 and B 0 initial susceptibility The susceptibility determined at H 0 and M 0 infrared IR For practical purposes any radiant energy within the wavelength range 770 to 10 nanometers is considered infrared energy The full range is usually divided into three sub ranges near IR far IR and sub millimeter input card Electronics on a printed circuit board card that plug into an instrument main frame Used by configurable instruments to allow for different sensor types or interface options interchangeability Ability to exchange one sensor or device with another of the same type without a significant change in output or response international system of units SI A universal coherent system of units in which the following seven units are considered basic meter kilogram second ampere Kelvin mole and candela The International System of Units or Systeme International d Unit s SI was promulgated in 1960 by the Eleventh General Conference on Weights and Measures For definition spelling and protocols see Reference 3 for a short c
315. sources but it does have some difficulty canceling noise at or very near its own measurement frequency An odd operating frequency was intentionally chosen to minimize the possibility that outside noise sources would be the same frequency as the instrument One source of noise that can be very close to the measurement frequency of a Model 370 is another Model 370 When two or more instruments are operated with leads closely coupled in the same Dewar it is possible for them to beat against each other and cause small oscillations If this happens the measurement frequency of one instrument can be changed slightly to avoid interference 2 5 7 Vibration Noise Noise can be generated in lead cables when they are moved or vibrated This noise can be related to the triboelectric effect or changes in capacitance between conductors in the cable A sudden movement of measurement lead cables can overload the instrument momentarily when it is on the most sensitive ranges Cables carrying small signals should be fastened down without putting strain on the connections 2 6 MEASUREMENT ERROR 2 6 1 Warm Up and Temperature Drift When the Model 370 is first turned on it should be allowed to warm up for at least 30 minutes before use to allow its internal temperature to stabilize Warm up is an exaggerated form of temperature drift because the temperature change inside the enclosure is larger and faster than would be experienced in a laboratory setting Temperature dri
316. splay Filter window is an extra feature of the filter algorithm If a single reading is different from the filtered value by more than the window limit the algorithm will assume the change was intentional and restart the filter The filter window is set in percent of full scale resistance range The next filter setup screen appears as a prompt for filter window Use the data entry keys to enter a filter window between 1 and 80 percent Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display 4 14 Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 4 10 TEMPERATURE MEASUREMENT One application of the Model 370 is measuring temperature of resistive temperature detectors RTDs The resistance range and excitation power of the instrument are ideal for temperature measurements using any common cryogenic RTD Resistance of an RTD can be measured and displayed at any time The instrument can convert resistance to temperature and display temperature in Kelvin if a temperature response curve is loaded into the instrument and assigned to the sensor Information on how to load curves into the instrument is given in Paragraph 5 2 This paragraph summarizes other operations related to temperature measurement Paragraph 4 11 describes temperature control 4 1
317. st Press Enter to accept the new value and advance to the units multiplier Use the A or Y key to select a units multiplier of m blank k or M Press Enter to accept the new selection and return to the normal display Press Escape to restart the setting sequence and enter a new value Press Escape again to cancel the sequence and return to the normal display 4 14 ALARMS The Model 370 can compare reading data from each channel to a high and low alarm setpoint Alarm setpoints and most other alarm parameters can be set independently for each channel If the Model 3716 Scanner is installed the alarm operates for the active scan channel only The alarm feature uses filtered reading data when the filter is enabled When any alarm is enabled the Alarm annunciator will be shown on the top line of the display The annunciator remains steady if neither alarm is in the active alarm state It blinks when either alarm is active The alarming channel does not have to be displayed for the alarm annunciator to indicate status If a Model 3716 Scanner is installed the alarm annunciator follows the active scan channel to indicate the alarm status of that channel The instrument has an audible alarm beeper that can be programmed to sound is any of the alarms are active The relays described in Paragraph 4 15 can also be programmed to follow the alarm feature The alarm feature has two modes of operation latching and non latching The operation of these modes is des
318. stance of the resistor and predict its reaction to excitation and noise Self heating ATsh is a product of the resistors thermal resistance Rt expressed in kelvin per watt K W and excitation power Ps expressed in watts ATsh Rt Ps Ps can be calculated as excitation current squared times measured resistance Ps PR A familiar form of self heating is sensor self heating in resistive temperature detectors RTDs but the same principles apply to any resistor being measured Specifications for the Model 370 include typical sensor characteristics for a Germanium RTD that estimates the thermal resistance of the sensor This example demonstrates how excitation current must be reduced to compensate for the increase in thermal resistance at lower temperature Otherwise self heating would become a significant part of the overall accuracy Calculations like this are difficult to do for two reasons first manufactures do not always publish specifications on thermal resistance and second the actual thermal resistance is very dependent on installation An explanation of how to calculate thermal resistance of a temperature sensor is given in the Lake Shore Temperature Measurement and Control Catalog Appendix A With an estimate of thermal resistance the effect of increased excitation current can be predicted If self heating behavior does not match the prediction noise power may be high enough to heat the resistor 2 6 4 Lead Resistance Measuremen
319. strument features Theory of Operation 2 1 Lake Shore Model 370 AC Resistance Bridge User s Manual 2 3 1 Dilution Refrigerator Terminology Dilution refrigerators remain the workhorse of low temperature research down to approximately 5 10 mK So many variations of factory and home built units exist that it is unlikely that any two are identical Figure 2 1 illustrates only the most basic features Helium Dewar 4 i He Reservoir Vacuum Space Vacuum insulated Dewar large enough to contain the following components Sometimes includes a nitrogen outer jacket to improve efficiency or superconducting magnet coil when needed Reservoir of liquid He at atmospheric pressure that provides bulk cooling down to 4 2 K Pre cools gas entering the vacuum space and acts as a radiation shield for refrigeration components Insulates refrigeration components and provides a clear workspace 1 K Pot He Evaporator Pumped reservoir of He maintained at K Used in the process of condensing He rich gas returning from pumping system Also used as a heat sink for measurement leads Good location for a temperature sensor for cool down monitoring and troubleshooting Still Heat Exchanger Chamber used to evaporate Heat Exchangers Mixing Chamber He out of a dilute He mix liquid He containing some He sometimes called mash It is the first stage below 1 K and is also used to liquefy returning He Th
320. strument parameters to their default values or erase the contents of user curve memory This data is stored in nonvolatile memory called NOVRAM The instrument allows parameter and curve data locations in NOVRAM to be cleared independently Instrument calibration is not affected by this operation Firmware version dates for both main and input firmware are also displayed during this sequence To clear NOVRAM memory or view firmware revision dates press and hold the Escape key for 5 seconds The following screen appears to show the main firmware revision date and as a prompt for returning the instrument parameters to default values Default parameter values are listed in Table 4 1 Use the AV keys to select Yes for default values and No to continue without changing the parameter values Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display Operation 4 33 Lake Shore Model 370 AC Resistance Bridge User s Manual Default Parameter Values Continued The next screen appears to show the input firmware revision date and as a prompt for clearing user curve memory Use the AV keys to select Yes to clear curves and No to continue without clearing Press Enter to accept the new selection and return to the normal display Press Escape to cancel the new selection and return to the normal display Table 4 1 Default Parameter
321. subminiature receptacles hold connections for up to sixteen 4 lead resistor inputs in four sets of four With the 3708 two 25 pin DIN receptacles hold connections for up to eight 4 lead resistor inputs They mate with 25 pin D subminiature plugs included with the scanner Lake Shore P N 106 253 Unused inputs can either be left open or all four measurement pins can be shorted together Pin connections for Channels 1 4 are shown in Figure 3 8 Channels 5 8 9 12 and 13 16 are wired the same way 3 5 3 Scanner Resistor Input Cabling The cable between scanner and Dewar should be kept as short as possible without causing strain on the cable or connections See Figure 3 6 Short cables can improve the measurement in several ways They minimize the potential for induced noise on the smallest voltage signals limit cable capacitance that cannot be guarded because the guards are not carried through the scanner and reduce the potential for ground loops if shield lines accidentally touch the Dewar Individually shielded twisted pair cable is recommended for wiring the resistor inputs Cable containing up to 8 shielded pairs of wire is available form electronics distributors so one cable can provide enough wires to connect the four channels of one input connector If multiple smaller cables are used they should be wrapped together to minimize ground loops The two current leads of a channel should be twisted and the two voltage leads should be twisted The input
322. t Lockups 12 Check cable connections and length 13 Increase delay between all commands to 50 ms to make sure instrument is not being over loaded 6 10 Remote Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 6 2 SERIAL INTERFACE OVERVIEW The serial interface used in the Model 370 is commonly referred to as an RS 232C interface RS 232C is a standard of the Electronics Industries Association EIA that describes one of the most common interfaces between computers and electronic equipment The RS 232C standard is quite flexible and allows many different configurations However any two devices claiming RS 232C compatibility cannot necessarily be plugged together without interface setup The remainder of this paragraph briefly describes the key features of a serial interface that are supported by the instrument A customer supplied computer with similarly configured interface port is required to enable communication 6 2 1 Changing Baud Rate To select the Serial Interface Baud Rate press the Computer Interface key The first computer interface screen appears as a prompt for Baud Use the A or Y key to select 300 1200 or 9600 Baud The default is 9600 Baud Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display 6 2 2 Physical Connection The Model 370 has a 9 pin D Subminiature plug on the rear panel for se
323. t circuit The shield can only divert noise energy away from measurement leads if that energy has a place to go Connecting the shield to measurement common shield pin in the connector is a good starting point but may not be the final answer The measurement common is isolated from Earth ground and has by design a relatively low mass Most noise sources are referenced to Earth ground and if they are large enough can swing the entire measurement shield and all The shield may require a low impedance connection to Earth ground to effectively divert noise energy but connecting to Earth ground can cause new problems If the shield or any other part of the measurement circuit is connected to Earth ground it is necessary to prevent ground loops that couple line frequency noise into the measurement A ground loop is created when measurement common is tied to Earth ground at more than one point The isolation of the instrument breaks the loop between the instrument power cord and the measurement common No loop is formed if measurement common is tied to Earth ground at only one point Loops are also the primary means of coupling H field noise Conductive shields do not reduce the effects of H fields Two ways to minimize H field coupling are to reduce the loop area or break the loop as described with ground loops Loop area can be minimized by tightly twisting lead pairs both voltage and current within the shielded cables This not only reduces l
324. t deal must be known about the load sensor and controller to compute a proportional setting P Most often the proportional setting is determined by trial and error The proportional setting is part of the overall control loop gain and so are the heater range and cooling power The proportional setting will need to change if either of these change 2 10 2 2 Integral I In the control loop the integral term also called reset looks at error over time to build the integral contribution to the output Output I P YA tear By adding the integral to proportional contributions the error that is necessary in a proportional only system can be eliminated When the error is at zero controlling at the setpoint the output is held constant by the integral contribution The integral setting I is more predictable than the proportional setting It is related to the dominant time constant of the load As discussed in Paragraph 2 7 3 measuring this time constant allows a reasonable calculation of the integral setting In the Model 370 the integral term is set in seconds and a smaller setting creates a more active integrator 2 10 2 3 Derivative D The derivative term also called rate acts on the change in error with time to make its contribution to the output Output D BDE dt By reacting to a fast changing error signal the derivative can work to boost the output when the setpoint changes quickly reducing the time it takes for temperature
325. t excitation gt Excitation range number 1 22 for Current Excitation 1 12 for Voltage Excitation lt range gt Resistance range number 1 22 lt autorange gt 0 autorange off 1 autorange on lt cs off gt 0 excitation on excitation off RDGRNG 8 1 9 14 0 0 term Channel 8 in current excitation mode with a excitation of 9 10 0 nA The resistance range is set to 14 which is the 6 32 kQ range Turn autorange off and excitation on Voltage Range Current Range Resistance Range 1 2 00 UV 1 1 00 pA 1 2 00 mQ 2 6 32 UV 2 3 16 pA 2 6 32 mQ 3 20 0 LV 3 10 0 pA 3 20 0 mQ 4 63 2 UV 4 31 6 pA 4 63 2 mQ 3 200 uV 5 100 pA 5 200 mQ 6 632 uV 6 316 pA 6 632 mQ 7 2 00 mV 7 1 00 nA 7 2 00 Q 8 6 32 mV 8 3 16 nA 8 6 32 Q 9 20 0 mV 9 10 0 nA 9 20 0 Q 10 63 2 mV 10 31 6 nA 10 63 2 0 11 200 mV 11 100 nA 11 200 Q 12 632 mV 12 316nA 12 632 Q 13 1 00 uA 13 2 00 kQ 14 3 16 WA 14 6 32 KQ 15 10 0 pA 15 20 0 kQ 16 31 6 WA 16 63 2 KQ 17 100 WA 17 200 kQ 18 316 WA 18 632 KQ 19 1 00 mA 19 2 00 MQ 20 3 16 mA 20 6 32 MQ 21 10 0 mA 21 20 0 MQ 22 31 6 mA 22 63 2 MQ Ranges not available with the Model 3708 Scanner Remote Operation 6 33 Lake Shore Model 370 AC Resistance Bridge User s Manual RDGRNG Resistance Range Query Input RDGRNG lt channel gt term Format nn lt channel gt Specifies which channel to query 1 16 Retur
326. t leads in cryogenic systems routinely have resistance of tens or even hundreds of ohms Lead wire is chosen for its thermal properties first and electrical resistance is a secondary consideration The Model 370 measurement circuits can tolerate significant lead resistance without measurement error A lead resistance specification of 100 Q plus 10 of resistance range for each of the four leads is given for most excitation currents This lead resistance will give a measurement error that is small compared to the specified accuracy on the resistance range and in most cases the error is not measurable The lead resistance specification is reduced to 10 Q plus 10 of resistance range for the highest current excitations At the higher currents lead resistance of 100 Q can create drops of several volts in each lead These large voltage swings will measurably affect the millvolt and microvolt resistor measurements If lead resistance exceeds the specified values the instrument will measure with reduced accuracy until the compliance voltage of approximately 7 V RMS is reached Keep the resistance in all of the leads similar whenever possible to maintain measurement balance 2 6 5 Current Leakage Lead wiring cannot be taken for granted when determining accuracy of cryogenic resistance measurements Cryostats are notorious for having bundles of long thinly insulated leads tightly wrapped around every available heat sink area Shorts are inevitable Lead wires sho
327. t limit in Kelvin can be entered in the temperature response curve This limit is in force for the control channel only and only when operating with setpoint units of Kelvin Setpoint resolution matches the 6 digit display resolution of the measurement Maximum and minimum setpoint range and resolution are shown below Maximum Resistance 100 000 MQ Minimum Resistance 0 000001 mQ Maximum Temperature 1500 00 K Minimum Temperature 00 0001 mK To change the control setpoint press the Setpoint key The setpoint value on the normal display will be highlighted to prompt for the numeric part of the new setpoint value Use the data entry keys to enter the numeric setpoint value Press Enter to accept the new value and advance to the units multiplier Press Escape to restart the setting sequence and enter a different value Press Escape again to leave the setting sequence The setpoint units multiplier on the normal display will be highlighted to prompt for that portion of the new setpoint value Use the A or Y key to select a units multiplier of m blank k or M Press Enter to accept the new selection and complete the setting sequence Press Escape to restart the setting sequence and enter a different value Press Escape again to leave the setting sequence Operation 4 19 Lake Shore Model 370 AC Resistance Bridge User s Manual 4 11 4 Manually Setting PID Values When operating the Model 370 as a temperatur
328. t they must be separated by a semi colon Only one query is permitted per communication but it can be chained to the end of a command The total communication string must not exceed 64 characters in length A command string is issued by the computer and instructs the instrument to either perform a function or change a parameter setting When a command is issued the computer is acting as talker and the instrument as listener The format is lt command mnemonic gt lt space gt lt parameter data gt lt terminators gt Command mnemonics and parameter data necessary for each one is described in Paragraph 6 3 Terminators must be sent with every message string A query string is issued by the computer and instructs the instrument which response to send Queries are issued similar to commands with the computer acting as talker and the instrument as listener The query format is lt query mnemonic gt lt gt lt space gt lt parameter data gt lt terminators gt Query mnemonics are often the same as commands with the addition of a question mark Parameter data is often unnecessary when sending queries Query mnemonics and parameter data if necessary is described in Paragraph 6 3 Terminators must be sent with every message string Issuing a query does not initiate a response from the instrument A response string is sent by the instrument only when it is addressed as a talker and the computer becomes the listene
329. t this Lake Shore product the Product will be free from defects in materials and workmanship for the Warranty Period specified above the Warranty Period If Lake Shore receives notice of any such defects during the Warranty Period and the Product is shipped freight prepaid Lake Shore will at its option either repair or replace the Product if it is so defective without charge to the owner for parts service labor or associated customary return shipping cost Any such replacement for the Product may be either new or equivalent in performance to new Replacement or repaired parts will be warranted for only the unexpired portion of the original warranty or 90 days whichever is greater 2 Lake Shore warrants the Product only if it has been sold by an authorized Lake Shore employee sales representative dealer or original equipment manufacturer OEM 3 The Product may contain remanufactured parts equivalent to new in performance or may have been subject to incidental use 4 The Warranty Period begins on the date of delivery of the Product or later on the date of installation of the Product if the Product is installed by Lake Shore provided that if you schedule or delay the Lake Shore installation for more than 30 days after delivery the Warranty Period begins on the 31st day after delivery 5 This limited warranty does not apply to defects in the Product resulting from a improper or inadequate maintenance repair or calibr
330. t under Private Sub cmdSend_Click as shown in Table 6 5 c Inthe Code Editor window under the Object dropdown list select Form Make sure the Procedure dropdown list is set at Load The Code window should have written the segment of code Private Sub Form_Load Add the code to this subroutine as shown in Table 6 5 d Double Click on the Timer control Add code segment under Private Sub Timerl_Timer as shown in Table 6 5 e Make adjustments to code if different Com port settings are being used 13 Save the program 14 Run the program The program should resemble the following im Serial Interface Program Type exit to end program Command Response m m lel Es 15 Type in a command or query in the Command box as described in Paragraph 6 2 7 3 16 Press Enter or select the Send button with the mouse to send command 17 Type Exit and press Enter to quit Remote Operation 6 15 Lake Shore Model 370 AC Resistance Bridge User s Manual Table 6 5 Visual Basic Serial Interface Program Public gSend As Boolean Global used for Send button state Private Sub cmdSend Click gSend True End Sub Routine to handle Send button press Set Flag to True Private Sub Form Load Dim strReturn As String Dim strHold As String Dim Term As String Dim ZeroCount As Integer Dim strCommand As String frmSerial Show Term Chr 13 amp Chr 10 ZeroCount 0 strReturn s
331. tance range has been selected for the measurement Next verify the integrity of the voltage lead connections to the measured resistance If the error persists verify that the voltage pair of leads is twisted and shielded Likewise the current pair of leads should be twisted and shielded Twisted pair wiring reduces the differential signals potentially created by environmental noise However there is no advantage gained by twisting the current pair of leads together with the voltage pair of leads When twisted leads are not an option inside a Dewar the shielding must be verified adequate for the task The Dewar itself can act as a shield when the case is properly connected Ensure that the experimental Dewar case is properly connected to the measurement common of the instrument and earth ground Refer to Installation Paragraph 3 4 4 8 7 1 4 VMIX OVL Mixer Overload Between the instrumentation amplifier section and the mixer portion of the phase sensitive detector are several gain stages and a band pass filter If the combination of signal and all differential noise components cause the input to the mixer to exceed 10 V peak the mixer overload error is displayed The overload occurs at the VMIX point in the instrumentation amplifier circuit Refer to Figure 8 3 for the location of the VMIX monitoring point The signal voltage can be viewed on an oscilloscope by using the Monitor output described in Paragraph 4 7 4 This ensures linear ope
332. temperatures The Model 370 is well equipped to measure and control temperature using these devices 2 9 1 Sensor Selection Lake Shore offers a full line of cryogenic temperature sensors including several models that are well suited for low temperature work to 50 mK and below Two of the most popular are Germanium and Ruthenium Oxide RTDs The Lake Shore Temperature Measurement and Control catalog gives detailed specifications of these sensors and should be consulted as a reference when choosing temperature sensors The Model 370 will work with resistive temperature sensors from Lake Shore or any other manufacture A few important selection criteria are discussed below 2 9 1 1 Sensitivity Sensitivity is often the most important criteria when selecting a temperature sensor because it dictates many of the other specifications including range resolution and accuracy Many NTC resistive materials like germanium have a beneficial characteristic of temperature sensitivity that makes them particularly useful at low temperatures The resistance and sensitivity of these materials increase exponentially as temperature decreases The sensitivity of the sensors is best at the 2 14 Theory of Operation Lake Shore Model 370 AC Resistance Bridge User s Manual lowest temperatures where signal to noise is worst making them useful at lower temperatures than a linear device of the same resistance 2 9 1 2 Temperature Range Every type of sensor has an u
333. ten are assigned their own keys 4 11 2 1 Setting Control Mode Control mode is the first parameter that appears in the control setup sequence because many of the parameters that follow depend on that choice If a parameter does not appear in the sequence it is not used in the chosen control mode Control mode is most often changed when the heater is turned off but it can be changed during active control The instrument will attempt to maintain a stable control output if control mode is changed during active control When control mode is changed from closed loop PID to open loop the instrument will copy the closed loop heater output value to the manual output parameter Open loop control will then maintain the same operating point if the system was stable when the mode change was made When control mode is changed from open loop to closed loop PID the instrument will copy the manual output value into the closed loop integrator hold memory Closed loop control will then begin controlling to the setpoint without needing to recharge the integrator from zero To begin control setup press the Control Setup key The first control setup screen appears as a prompt for the control mode Use the A or V key to select a control mode from the list in Paragraph 4 11 1 Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display 4 11 2 2 Control Channe
334. tentially require replacement See Figure 8 4 for IC locations Input Microcontroller U14 Contains software that configures the inputs takes readings and performs control functions Has a sticker on top labeled M370IF HEX and a version number e Main Firmware Erasable Programmable Read Only Memory EPROM U104 Contains the user interface software Has a sticker on top labeled M370MF HEX and a date e Non Volatile Random Access Memory NOVRAM U107 Contains instrument settings and user curves The NOVRAM is replaced when the customer purchases a Model 8002 05 370 CalCurve Refer to Paragraph 7 2 Use the following procedure to replace any of these ICs 1 Follow the top of enclosure REMOVAL procedure in Paragraph 8 8 2 Locate the IC on the main circuit board See Figure 8 4 Note orientation of existing IC Service 8 9 Lake Shore Model 370 AC Resistance Bridge User s Manual U104 Main Firmware EEPROM U107 NOVRAM U14 Input Microcontroller PCB Layout bmp Figure 8 4 Location Of Important Internal Components 8 10 Service Lake Shore Model 370 AC Resistance Bridge User s Manual Firmware Replacement Continued CAUTION The ICs are Electrostatic Discharge Sensitive ESDS devices Wear shock proof wrist straps resistor limited to lt 5 mA to prevent injury to service personnel and to avoid inducing an Electrostatic Discharge ESD into the device 3 Use IC puller to rem
335. tep can be done after all ranges are calibrated 9 Select the next lowest voltage range and repeat from Step 3 until all ranges are calibrated Refer to Paragraph 8 12 2 6 for voltage range numbers NOTE Lower voltage ranges especially below 200 LV require long filter time constants to obtain a stable zero offset measurement Zero offsets for ranges below the highest range can be calculated instead of being measured To calculate zero offsets of lower ranges scale the last measured offset by the difference in ideal gain on the consecutive ranges For example divide by 3 16228 10 31 6228 100 and so on 8 12 2 2 Voltage Measurement Gain To calibrate voltage gain perform the following sequence of operations Voltage Gain must be calibrated after voltage zero offset because the voltage zero offset procedure sets the gain coefficient to default Attach a 20 Q resistor to the input of the instrument with short leads Allow the instrument to warm up for 1 hour Select the highest voltage range 632 mV and the 20 resistance range Aw NS Use the CALG query to read and verify that the gain coefficient for the selected voltage range has been set to default during zero calibration Service 8 17 Lake Shore Model 370 AC Resistance Bridge User s Manual 5 Read the AC RMS voltage across the 20 2 resistor with a digital volt meter DVM and record the measured voltage reading The DVM must be capable of accurate measurements at 13 7
336. terials with linear expansion coefficients that differ by more than three A flexible mounting scheme should be used or the parts will break apart potentially damaging them The thermal expansion or contraction of rigid clamps or holders could crush fragile samples or sensors that do not have the same coefficient Thermal conductivity is a property of materials that can change with temperature Do not assume that materials that work well at room temperature will do the same job at low temperatures 2 9 3 2 Sensor Location Finding a good place to mount a sensor in an already crowded cryostat is never easy There are fewer problems if the entire load and sample holder are at the same temperature unfortunately this not the case in many systems Temperature gradients differences in temperature exist because there is seldom perfect balance between the cooling source and heat sources Even in a well controlled system unwanted heat sources like thermal radiation and heat conducting through mounting structures can cause gradients For best accuracy sensors should be positioned near the sample so that little or no heat flows between the sample and sensor This may not however be the best location for temperature control as discussed below 2 9 3 3 Thermal Conductivity The ability of heat to flow through a material is called thermal conductivity Good thermal conductivity is important in any part of a cryogenic system that is intended to be the same tem
337. th manual range selection the user selects excitation as well as resistance range ranges do not change automatically If input resistance exceeds the range an overload message appears With both autorange and manual range selection the excitation current source is shorted inside the Model 370 during range changes to minimize transients Temperature Conversion The Model 370 temperature conversion function converts measured resistance to temperature for calibrated resistance thermometers Temperature as well as resistance values can be displayed temperature and resistance values are also available for interface query Conversion is based on temperature response curve data loaded into the instrument for each calibrated resistance thermometer in use Up to twenty 200 point curves can be entered into nonvolatile memory via computer interface or the instrument front panel Lake Shore CalCurvesTM are available for Lake Shore calibrated sensors users can also generate their own curves as desired Temperature Control The Model 370 provides all of the circuitry and software for digital proportional integral derivative PID control Heater output is a variable DC current source with multiple power ranges from 0 1 uW to 1 W to drive a resistive heater nominal 100 Heater output is designed for low noise with circuits to eliminate power surges during range changes or at instrument start up A still heater function can also supply up to 1 W of power into a
338. the highest current excitation 31 6 mA and the highest resistance highest voltage range available for that current 20 9 632 mV for the highest current 3 Attach a near full scale 20 Q for the highest current resistor of known value to the input of the instrument with short leads 4 Use the CALG query to read and record the gain coefficient for the selected current A safety step to revert to the original calibration if necessary 5 Set the CALG coefficient for the selected current to default 1 Read the resistance from the instrument with the RDGR query Calculate the gain coefficient value for the current by the following CALG Actual Resistance RDGR The gain coefficient should be 1 42 8 Send the gain coefficient to the instrument using the CALG command Example for the highest current range CALG 1 22 1 01212E 0 9 Send the CALSAVE command and wait 10 seconds for the calibration to be saved in non volatile memory Alternately this step can be done after all ranges are calibrated 10 This process is repeated for lower current settings The voltage range must be reduced to calibrate currents below 1 UA in order to maintain the use of reasonable test resistance values Paragraph 8 12 2 6 gives an example order of voltage and current calibrations that maintains reasonable resistance values 8 12 2 4 Frequency Compensation The Model 370 is normally operated and calibrated with the measurement frequency set to th
339. tion is not affected Input Not Responding Input microprocessor is not responding to main microprocessor If the condition persists after power is cycled the instrument must be repaired Instrument Uncalibrated Calibration memory has been corrupted Recalibration is required for accurate measurements See Paragraph 8 12 for calibration details Measurement or Compliance Limit CS OVL Current source overload possibly caused by a broken current source lead VCM OVL Common mode overload at the first amplifier stage of the voltage input VDIF OVL Overload at the differential amplifier stage of the voltage input VMIX OVL Overload at the mixer phase sensitive detector stage of the voltage input R OVER Resistance range overload R UNDER Resistance range underload equivalent of R OVER in the negative direction normally a result of improper lead wiring T OVER Measured resistance is outside of the temperature response curve on the high temperature end T UNDER Measured resistance is outside of the temperature response curve on the low temperature end Heater Open Temperature control heater is too large or open circuited Service 8 3 Lake Shore Model 370 AC Resistance Bridge User s Manual Error Messages Continued Illegal Operation Disabled Scanner channel has been turned off and readings cannot be displayed Locked Instrument keypad is locked and paramet
340. to all channels at the same time Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next linear setup screen appears as a prompt for the equation Use the A or Y key to select M X B or MX B Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display 4 24 Operation Lake Shore Model 370 AC Resistance Bridge User s Manual Linear Equation Continued The next linear setup screen appears as a prompt for a slope value M Use the data entry keys to enter a slope between 0 001 and 1000 Press Enter to accept the new selection and continue to the next setting screen Press Escape to restart the setting sequence and enter a different value Press Escape again to cancel the sequence and return to the normal display The next linear setup screen appears as a prompt for data source for the selected input X Use the A or Y key to select between Kelvin and ohms Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next linear setup screen appears as a prompt for the source of the offset B If either setpoint variable is selected the next setting screen will be skipped If value is sel
341. to be version 2 0 or newer Contact Lake Shore service if the firmware needs to be updated Service 8 7 Lake Shore Model 370 AC Resistance Bridge User s Manual CS POS Positive Current Source Drive V O VCM POS Current Sample Resistor VDIFF Y Measured X 3 to Resistor 1000 Current Sample Resistor E gt O VCMNEG V O Oscillator Reference 9 O Low Pass Q Analog Filter RI Mixer Signal Flow eps Figure 8 3 Signal Flow Block Diagram 8 8 ENCLOSURE TOP REMOVAL AND REPLACEMENT WARNING To avoid potentially lethal shocks turn off controller and disconnect it from AC power line before performing this procedure Only qualified personnel should perform this procedure REMOVAL 1 Set power switch to Off O and disconnect power cord from rear of unit 2 Ifattached remove 19 inch rack mounting brackets 3 Use a Phillips screwdriver to remove 4 flat head screws centered front amp rear on the top amp bottom of the enclosure 4 Use 5 64 hex key to remove 4 screws attaching top panel to unit 5 Use 5 64 hex key to loosen 4 screws attaching bottom panel to unit 6 Carefully remove the back bezel by sliding it straight back away from the unit 7 Slide the top panel back and remove it from the unit INSTALLATION POM Ge ee A Slide the top panel forward in the track provided on each side of the unit Carefully replace the back bezel by sliding it straight into the unit Use a
342. to connect to the analog output are required A zero offset and gain coefficient must be calculated for each output The method used for calculating these coefficients provides the most accurate calibration of the DAC used in the instrument To calibrate the analog voltage outputs perform the following sequence of operations 1 Configure the analog voltage output for manual operation Paragraph 4 16 Set the manual output value to 100 Attach the voltmeter and allow the instrument to warm up for 1 hour 2 Set the zero offset and gain calibration coefficients to their defaults Analog Output 1 CALZ 4 1 0 CALG 4 1 1 Analog Output 2 CALZ 4 2 0 CALG 4 2 1 Measure and record the 100 output voltage Set the manual output value to 100 Measure and record the 100 voltage Calculate the gain coefficient as Coefficient 20 V Voltage 100 Voltage 100 The gain should be 0 995 1 7 Send the gain coefficient for the appropriate output to the instrument Analog Output 1 CALG 4 1 Coefficient Analog Output 2 CALG 4 2 Coefficient PL tz gt Set the manual output value to 0 9 Measure and record the output voltage 8 22 Service Lake Shore Model 370 AC Resistance Bridge User s Manual Analog Voltage Output Calibration Continued 10 Calculate the zero offset coefficient in percent as Coefficient Voltage 0 10 V x 100 Voltage 0 x 10 The zero offset should be between 1 11
343. tpoint ramp feature Change the still heater output value Select a heater current source output range for temperature control Turn the heater current source output off Setup parameters related to each input channel Setup instrument parameters common to all channels Setup firmware digital filter Select Source for Max Min data and reset captured values Setup display including display units Enter a temperature response curve from the front panel Setup linear equation or AR Setup alarm feature and reset latched alarms Returns the instrument to local operation if in remote Setup computer interface parameters Setup analog voltage output and still heater output Setup relay feature Numeric data entry within a setting sequence Exits from parameter setting sequence without changing the parameter value press and hold to reset parameters to default values Increments a parameter selection or value Decrements a parameter selection or value Accepts a new parameter value press and hold to lock keypad Operation 4 3 Lake Shore Model 370 AC Resistance Bridge User s Manual 4 3 2 General Keypad Operation The Model 370 uses three basic keypad operations direct operation setting selection and data entry for the majority of operator interface A few specialized keypad operations such as curve entry are described in the individual operation paragraphs Direct Operation Key functions occur immediately when the key is pressed
344. tpoint ramping and zone control should only be used after the basic functions are operating 4 11 1 Temperature Control Modes The Model 370 offers three temperature control modes or the control output can be disabled Each mode requires different control parameters so the control mode must be selected before other parameter values Open Loop Open loop control is less complicated than closed loop control but is also less powerful Open loop control mode allows the user to directly set the heater output value using only the Manual Output parameter During Open Loop control only the heater range and Manual Output parameters are active the setpoint control channel and PID parameters are ignored This type of control guarantees constant power to the load but it does not actively control temperature Any change in the characteristics of the cooling system or load will cause a change in temperature e Closed Loop PID Closed loop PID control often called feedback control is the control mode most often associated with temperature controllers In this mode the controller attempts to keep the load at exactly the user entered setpoint that can be entered in resistance or temperature To do this it uses feedback from the control sensor to calculate and actively adjust the control heater output The Model 370 uses a control algorithm called PID that refers to the three terms used to tune the controller for each unique system Manual output can also be active
345. tput for this zone 0 100 lt htr range gt Specifies the heater range for this zone 0 8 Refer to HTRRNG for listing lt relay low gt Specifies the Low relay state 0 off 1 on lt relay high gt Specifies the High relay state 0 off 1 on lt analog 1 gt Specifies the Analog 1 output for this zone 100 to 100 lt analog 2 gt Specifies the Analog 2 output for this zone 100 to 100 Remarks Configures the control loop limit band parameters used in conjunction with the Status Byte Register Refer to Paragraph 6 1 4 1 Example ZONE 1 0 25 10 20 0 0 2 0 0 10 50 term Zone 1 is valid to 0 250 K with P 10 I 20 D 0 a manual out of 0 and a heater range of 2 Both relays will be off and Analog output 1 will be at 10 and Analog output 2 will be at 50 ZONE Control Loop Zone Table Parameter Query Input ZONE lt zone gt term Format n lt zone gt Specifies which zone in the table to query 1 10 Returned lt setpoint limit gt lt P value gt lt I value gt lt D value gt lt mout value gt lt htr range gt lt relay low gt lt relay high gt lt analog 1 gt lt analog 2 gt term Format nnn nnnE nn Ennnnnn Ennnnnn Ennnnnn Ennnnnn N n n nnnnnn Ennnnnn Refer to command for description 6 36 Remote Operation 7 0 GENERAL Lake Shore Model 370 AC Resistance Bridge User s Manual CHAPTER 7 OPTIONS AND ACCESSORIES This chapter provides info
346. trHold If frmSerial MSComml PortOpen True Then frmSerial MSComml PortOpen False End If frmSerial MSComm1 CommPort 1 frmSerial MSComml Settings 9600 0 7 1 frmSerial MSComml InputLen 1 frmSerial MSComm1 PortOpen True Do Do DoEvents Loop Until gSend True gSend False strCommand frmSerial txtCommand Text strReturn strCommand UCase strCommand If strCommand EXIT Then End End If frmSerial MSComm1 Output strCommand amp Term If InStr strCommand lt gt 0 Then While ZeroCount lt 20 And strHold lt gt Chr 10 If frmSerial MSComml InBufferCount frmSerial Timerl Enabled True Do DoEvents Loop Until frmSerial Timerl Enabled ZeroCount ZeroCount 1 Else ZeroCount 0 strHold frmSerial MSComml Input strReturn strReturn strHold End If Wend If strReturn lt gt Then strReturn Mid strReturn 1 InStr strReturn Else strReturn No Response End If frmSerial txtResponse Text strReturn strHold ZeroCount 0 End If Loop End Sub Main code section Used to return response Temporary character space Terminators Counter used for Timing out Data string sent to instrument Show main window Terminators are lt CR gt lt LF gt Initialize counter Clear return string Clear holding string Close serial port to change settings Example of Comm 1 Example of 9600 Baud Parity Data Stop Read one character at a time Open p
347. trol loop more often 2 18 Theory of Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 2 10 2 Closed Loop PID Control Closed loop PID control often called feedback control is the control mode most often associated with temperature controllers In this mode the controller attempts to keep the load at exactly the user entered setpoint that can be entered in resistance or temperature To do this it uses feedback from the control sensor to calculate and actively adjust the control heater output The Model 370 uses a control algorithm called PID that refers to the three terms used to tune the controller for each unique system The PID control equation has three variable terms proportional P integral I and derivative D See Figure 2 4 Changing these variables for best control of a system is called tuning The PID equation in the Model 370 is HeaterOutput pe I e at D E where the error e is defined as e Setpoint Feedback Reading 2 10 2 1 Proportional P The Proportional term also called gain must have a value greater than zero for the control loop to operate The value of the proportional term is multiplied by the error e which is defined as the difference between the setpoint and feedback temperatures to generate the proportional contribution to the output Output P Pe If proportional is acting alone with no integral there must always be an error or the output will go to zero A grea
348. tronic Accuracy Temperature Electronic Accuracy Resistance dR dT Self Heating Excitation Power x Thermal Resistance Table 1 1 2 and 3 Information Accuracy Percent reading specified on facing page 0 005 of range Precision Dominated by measurement temperature coefficient 0 0015 of reading 0 0002 of range C Resistance Full scale resistance range nominal 20 over range Resolution RMS noise with 18 s filter settling time approximates 3 s analog time constant 200 KQ resistance range 1000 resolution 1 0 fw power Power Excitation power at one half full scale resistance Range not available Range available not specified Introduction 1 5 Accuracy 0 03 632 mV 63 2 mV Lake Shore Model 370 AC Resistance Bridge User s Manual Table 1 1 Model 370 3716 Voltage Range 20 mV 6 32 mV 200 ma 2 0 mV 632 uV 0 1 0 3 200 uV 6 32 uv 2 0 pV 316mA 2040 640 210 1340 400nQ 10 mW 3 2 mW 1mw 320 uW 100 uW 63 2 Q 200 6 32 Q 200 632 MQ i i 10 mA 60 Q 20 yQ 6 HQ 4 uQ 1 3 yQ cj 32mW 1mW 320uW 100uW 32uW 200 Q 63 2 Q 200 6 32 Q 200 3 16 mA 20042 60uQ 2040 134Q 410 1mW 320uW 100uW 324W 10pW 632 Q 200 Q 63 2 Q 200 6 32 Q 1mA 600uQ 200u0 60uQ 40pQ 1310 320 4W 100uWw 324W 10uW 32uW 2 0 ka 632 Q
349. truments emit some high frequency noise both through the air and on their leads Emission from a Model 370 is very low and will not affect resistance measurements or heat measured resistors if the instrument is installed properly The high frequency noise can interfere with very sensitive measurements being made in the same Dewar Interference can be minimized by separating the instruments and their leads outside the Dewar and placing ferrite beads on the Model 370 measurement leads near the Dewar to stop the noise from getting inside Wrapping leads through the large beads more than one time improves their attenuation of unwanted signals Lake Shore sells a clamp on ferrite bead as part number 2071 A more common source of digital noise than the instrument itself is digital computers used for data acquisition Computers with their high clock seeds generate a wide spectrum of noise and transmit it through line cords and interface cables Isolation in the Model 370 separates the measurement circuits from interface circuits but computers must still be treated carefully during installation of a system Theory of Operation 2 9 Lake Shore Model 370 AC Resistance Bridge User s Manual 2 5 5 Resistor Thermal Noise Thermal noise or Johnson noise is the most common noise associated with resistors This noise is present in all resistors 1s independent of excitation and has a flat frequency spectrum The equation for Johnson noise is often expressed as Vnoi
350. tup screen appears as a prompt for display resolution on the chosen display location Use the A or V key to select a resolution from 4 to 6 Press Enter to accept the new selection and continue to other display features Press Escape to cancel the new selection and return to the normal display 4 4 3 Display Brightness The vacuum fluorescent VF display on the Model 370 has four brightness settings between 100 and 25 that can be changed from the front panel The brightness setting changes the entire VF display but does not affect the LED annunciators to the right of the display Continuous use of the instrument at 100 brightness will reduce the operating life of the display None of the brightness settings will turn the display off completely The instrument will operate with the display unplugged for applications that require absolute darkness but both the digital control and power supply cables must be unplugged with the instrument power turned off or damage to the display will result To change display brightness continue from the display resolution screen or press Display Setup then Enter until the following display setup screen appears as a prompt for display brightness Use the A or Y key to select brightness from 25 to 100 Press Enter to accept the new selection and return to the normal display Press Escape to cancel the new selection and return to the normal display 4 5 EXCITATION MODES RESISTANCE R
351. uded with a new Model 370 Y Accessories included with a new Model 3716 3716 L and 3708 Scanner Options and Accessories 7 3 Lake Shore Model 370 AC Resistance Bridge User s Manual Scan Control 01JUuOJ ues Input I To Model 370 Bridge To Optional Scanner Iindno 112 374a bmp Andino Scan Control OIJU09 ULIS Shield Not Attached To Model 370 AC Resistance Bridge To Optional Scanner 13ndjno Cable Wiring bmp Figure 7 1 Model 112 374 Cable Assembly Options and Accessories 7 4 Lake Shore Model 370 AC Resistance Bridge User s Manual CHAPTER 8 SERVICE 8 0 GENERAL This chapter provides basic service information for the Model 370 AC Resistance Bridge Customer service of the product is limited to the information presented in this chapter Factory trained service personnel should be consulted if the instrument requires repair 8 1 CONTACTING LAKE SHORE CRYOTRONICS If a Lake Shore product was purchased through a dealer or representative please use that resource for prompt sales or service information When contacting Lake Shore directly please specify the name of a department if do not know the name of an individual Questions regarding product applications price availability and shipments should be directed to sales Questions regarding instrument calibration or repair should be directed to instrument service Do not return a product to Lake Shore without an RGA number
352. udes magnetic field dependence for most temperature sensors Other specific data on environmental factors is included when available Lake Shore Cernox RTDs are an excellent choice for many of these critical applications Theory of Operation 2 15 Lake Shore Model 370 AC Resistance Bridge User s Manual 2 9 2 Sensor Calibration The Model 370 can convert the resistance of a sensor to temperature if a temperature response curve is loaded into the instrument Most curves are generated from an individual sensor calibration because few sensors follow a standard curve at low temperatures The instrument can use curve options called CalCurves available with Lake Shore calibrated sensors or curves generated by the user The Lake Shore Temperature Measurement and Control catalog gives complete specifications for calibrated sensors and gives the calibration range available for each sensor When calibrating sensors in house remember to account for error caused by interpolation between curve breakpoints The Model 370 allows curves with up to 200 breakpoints and uses linear interpolation between breakpoints Linear interpolation is considered by some to be less accurate than polynomial interpolation but it is more predictable If linear interpolation errors are larger than desired in a specific temperature range more breakpoints can be added to reduce the error NTC resistor breakpoints are entered as Log 62 vs K which acts to linearize temperature response
353. ue in ohms or log ohms Press Enter to accept the new selection and continue to the temperature field Press Escape to restart the setting sequence and enter a new value Press Escape again to highlight the breakpoint number When the temperature field is highlighted use numerical keypad to enter the applicable temperature in Kelvin Press Enter to accept the new selection and continue to the next breakpoint Press Escape to restart the setting sequence and enter a new value Press Escape again to highlight the breakpoint number Similarly enter or edit all breakpoints When complete press the Escape key while the breakpoint number field is highlighted to exit the curve edit screen and return to the normal display 5 4 Advanced Operation Lake Shore Model 370 AC Resistance Bridge User s Manual Edit Curve Continued NOTES 1 Typing over an existing resistance or temperature will replace that value when you press the Enter key 2 To add a new breakpoint to an existing curve go to the end of the curve and enter the new values The new point is automatically put into its proper place in breakpoint sequence 3 To delete a breakpoint enter all zeros for both the resistance and temperature 4 When curve entry is complete the user must assign the new curve to an input channel The Model 370 does not automatically assign the new curve to a channel even if there is only one 5 2 2 Erase Curve User curves that are no longer needed can
354. ue to 0 With a DVM read and record the voltage across the resistor as Voltage 0 A A R Calculate the heater range zero offset coefficient in percent as Coefficient Voltage 0 Actual Resistance Nominal Current x 1 07 x 100 The zero offset should be 0 1 0 Send the zero offset coefficient to the instrument with the following command CALZ 3 1 Coefficient 9 Change the Manual Output parameter value to 100 10 With a DVM read and record the voltage across the resistor as Voltage 100 11 Calculate the heater range gain coefficient as Coefficient Nominal Current Voltage 100 Actual Resistance The gain should be 0 936 2 12 End the gain coefficient to the instrument with the following command CALG 3 1 Coefficient 13 End the CALSAVE command and wait 10 seconds for the calibration to be saved in non volatile memory Alternatively this step can be done after all ranges are calibrated 14 Select the next lowest heater output range and repeat from Step 2 until all ranges are calibrated Refer to the worksheet below for resistor values and command forms Service 8 23 Lake Shore Model 370 AC Resistance Bridge User s Manual Model 370 Heater Output Calibration Worksheet for Paragraph 8 12 4 Step Range Setting Coefficient Resistor Actual Voltage 0 Voltage 100 Coefficient 1 100 mA Zero CALZ 3 8 200 1 100 mA Gain CALG 3 8 200 2 31
355. uence press Escape before pressing an arrow or number key Changes entered before Escape is pressed are kept 4 4 DISPLAY SETUP The display setup sequence for the Model 370 may seem complicated at first but the steps are necessary to provide flexibility to the operator 4 4 1 Display Setup and Units Selection Display setup begins with choosing the number of reading locations on the display One to four readings are displayed with large characters and five through eight readings are displayed with standard characters Reading locations are assigned the following numbers Display Locations Location 1 Location 2 Location 3 Location 4 Location 5 Location 6 Location 7 Location 8 If no scanner is installed only a data source must be chosen for each reading location Choosing units of resistance in ohms or temperature in Kelvin for a live reading is part of the source selection Units for Max Min and linear equation are selected during setup of those features Setting of display resolution follows in the display setup sequence if necessary Data Sources Ohms Live resistance reading in ohms Kelvin Live temperature reading in Kelvin if curve installed Max Maximum reading captured Min Minimum reading captured Linear Result of linear equation If a scanner is installed both channel and source must be chosen for each reading location When scanning more that 8 channels it is not possible to have a reading for all channels on the
356. uld be ohmed out on a regular basis to test for short circuits that develop over time and thermal cycling 2 6 6 Cable Capacitance Cable capacitance and measured resistance act together as a low pass filter to attenuate the desired measurement signal and shift its phase Low excitation frequency was chosen for the Model 370 to minimize this effect but capacitance can still contribute to measurement error when resistance is high Driven guards should be used when measuring resistance above 100 kQ but they often cannot be used inside of the cryostat and are not available for scanner input leads The effect of cable capacitance on unguarded leads can be estimated by the following equation This equation is only an estimate because it models the lead capacitance as a single capacitor across the measured resistor In a practical installation there is capacitance from each lead to shield or ground in addition to capacitance across the resistor Actual results depend on cable construction and installation techniques The equation does account for both magnitude attenuation caused by capacitor impedance in parallel with the resistor and phase attenuation caused by phase sensitive detection of the input signal Theory of Operation 2 11 Lake Shore Model 370 AC Resistance Bridge User s Manual Rattenuated R 1 0 C R Where Rattenuatea Resistance seen by the instrument R Actual resistance in ohms 0 2nf f Frequency in Hz default 13 7Hz
357. ument The 99 is included to prevent accidentally setting the unit to defaults DISPLAY Display Number Command Input DISPLAY lt locations gt term Format n lt locations gt Specifies number of locations to display 1 8 Example DISPLAY 7 term Display 7 locations DISPLAY Display Number Query Input DISPLAY term Returned lt locations gt term Format n Refer to command for description 6 26 Remote Operation DISPLOC Input Format Example DISPLOC Input Format Returned Format DOUT Input Format Remarks Example DOUT Input Returned Format FILTER Input Format Example Lake Shore Model 370 AC Resistance Bridge User s Manual Displayed Location Parameter Command DISPLOC lt location gt lt channel gt lt data source gt lt resolution gt term n nn n n lt location gt Specifies location to configure 1 8 lt channel gt Specifies channel to display in the field 1 16 0 Scan channel Specifies data for display 1 Kelvin 2 Ohms 3 linear data 4 minimum data and 5 maximum data lt resolution gt Display resolution in number of digits 4 6 DISPLOC 2 11 1 6 term Setup display location 2 to display the channel 11 Kelvin readings with 6 digits of resolution lt data source gt Displayed Location Query DISPLOC lt location gt term n lt location gt lt channel gt lt data source gt lt resolution gt term Specifies location to que
358. ut 2 ooooonconnciocnnocconncnnconnos 0 00 4 34 Operation Lake Shore Model 370 AC Resistance Bridge User s Manual CHAPTER 5 ADVANCED OPERATION 5 0 GENERAL This chapter provides information on advanced data entry operations for the Model 370 including curve entry and the zone feature 5 1 CURVE NUMBERS AND STORAGE The Model 370 has 20 user curve locations numbered 1 thru 20 Each location can hold from 2 to 200 data pairs breakpoints including a value in ohms or log ohms and a corresponding temperature value in Kelvin The curves are stored in nonvolatile memory and retained when power is turned off Using fewer than 200 breakpoints will not increase the number of available curve locations CalCurves can be stored as user curves for Lake Shore calibrated sensors The Model 370 does not include any standard temperature response curves If temperature sensors that follow a standard curve are used with the instrument that curve must be entered as a user curve 5 1 1 Curve Header Parameters Each curve has a set of parameters used for identification and to allow the instrument to use the curve effectively The parameters must be set correctly before a curve can be used for temperature conversion or temperature control Curve Number 1 thru 20 Name Defaults to the name User Curve for front panel entry When entering a user curve over the computer interface a unique curve name of up to 15 characters can be entered Seria
359. ut channel 1 2 lt polarity gt Specifies analog output polarity 0 unipolar 1 bipolar lt mode gt Specifies data the analog output monitors 0 off 1 channel 2 manual 3 zone Analog 2 also has mode 4 still lt channel gt Specifies which channel to monitor if lt mode gt 1 1 16 lt source gt Specifies channel data 1 Kelvin 2 Ohms 3 Linear Data 6 22 Remote Operation Lake Shore Model 370 AC Resistance Bridge User s Manual ANALOG Continued Example ANALOG Input Format Returned Format AOUT Input Format Returned Format Remarks BAUD Input Format BAUD Input Returned Format BEEP Input Format Remarks BEEP Input Returned Format lt high value gt If lt mode gt is 1 this parameter represents the data at which the analog output reaches 100 output If lt mode gt is 1 this parameter represents the data at which the analog output reaches 100 output if bipolar or 0 output if positive only lt low value gt lt manual value gt ANALOG 1 0 1 15 1 100 0 0 0 term Configure analog output as unipolar channel mode Set to monitor channel 15 Kelvin reading with 100 0 K at 100 output 10 0 V and 0 0 K at 0 output 0 0 V Manual value is ignored If lt mode gt is 2 this parameter is the value of the analog output Analog Output Parameter Query ANALOG lt analog channel gt term n lt analog channel gt Specifi
360. ute scale of temperature the zero point of which is approximately 273 15 C Scale units are equal in magnitude to Celsius degrees Originally devised by Lord Kelvin William Thompson 1824 1907 a British physicist mathematician and inventor B2 0 COMPARISON The three temperature scales are graphically compared in Figure B 1 Boiling point of water 373 15 K 100 C 212 F Freezing point of water 273 15 K 0 C 32 F Absolute zero OK 273 15 C 459 67 F kelvin Celsius Fahrenheit Figure B 1 Temperature Scale Comparison B3 0 CONVERSIONS To convert Fahrenheit to Celsius subtract 32 from F then divide by 1 8 or C F 32 1 8 To convert Celsius to Fahrenheit multiply C by 1 8 then add 32 or F 1 8 x C 32 To convert Fahrenheit to Kelvin first convert F to C then add 273 15 To convert Celsius to Kelvin add 273 15 Temperature Scales B 1 Lake Shore Model 370 AC Resistance Bridge User s Manual Table B 1 Temperature Conversion Table 292 180 129 67 290 178 89 120 289 67 178 71 119 67 280 173 33 117 67 279 67 173 15 112 274 170 110 270 167 78 109 67 436 269 67 167 59 100 430 261 67 163 15 99 67 429 67 260 162 22 94 423 67 259 67 162 04 90 420 256 160 419 67 250 156 67 418 00 249 67 156 48 410 243 67 153 15 409 67 240 151 11 405 67 239 67 150 93 400 238 150 399 67 230 145 56 390 229 67 145 37 389
361. utputs are short circuit protected but should not be used to drive a resistance of less than 100 Q They are capable of sourcing up to 1 W of power making them useful as auxiliary heater supplies Built in circuitry limits the transient energy sourced by the outputs when power is turned on to the instrument Analog Output 2 can be configured as a still heater output for a dilution refrigerator as described in Paragraph 4 17 The analog voltage outputs have four modes of operation described below In addition to these operating modes the outputs can be turned off or configured for unipolar only operation Manual The analog output is controlled manually from the instrument front panel or over computer interface Setting range is 100 to 100 with a setting resolution of 0 001 The true output resolution is limited to 0 003 Input The analog output is configured to follow measured reading of an input channel The user can select the channel if a Model 3716 Scanner is installed the data source and scale the output by defining the reading values that correspond to the maximum and minimum output values An example of this mode is given below Zone The analog output is operated with the zone control feature as described in Paragraph 5 3 Still Analog Output 2 is operated manually using the Still Output key as described in Paragraph 4 17 Press the Analog Out key to begin analog output setup The following setup screen appears as a prompt for anal
362. ves lt tempco gt Sets the temperature coefficient that will be used for temperature control if no curve is selected 1 negative 2 positive Example INSET 5 1 10 3 0 2 term Enables channel 5 with 10 second autoscan dwell 3 second pause time and no curve with a positive temperature coefficient INSET Input Channel Parameter Query Input INSET lt channel gt term Format nn lt channel gt Specifies which channel to query 1 16 Returned lt off on gt lt dwell gt lt pause gt lt curve number gt lt tempco gt term Format n nnn nnn n Refer to command for description KEYST Keypad Status Query Input KEYST term Returned lt keypad status gt term Format n lt keypad status gt 5 40 key pressed 0 no key pressed Remarks Returns keypad status since the last KEYST KEYST returns 1 after initial power up Is set to 0 after query If a number other than 0 is returned that number represents a mapping of the actual key pressed Mapping is determined by counting the keys until the key being pressed is located beginning with row 1 of the keypad If the key is not found in the row being mapped counting will continue in the next row of the keypad Remote Operation 6 29 Lake Shore Model 370 AC Resistance Bridge User s Manual LDAT Linear Equation Data Query Input LDAT lt channel gt term Format nn lt channel gt Specifies which channel to query 1 16 Returned lt l
363. voltage on each lead to maintain the lowest possible voltage difference between the lead and its environment This technique reduces the effect of cable capacitance Without the guards cable capacitance and measured resistance act together as a low pass filter and can attenuate the desired measurement signal Guards are not necessary when measured resistance is low or cable capacitance is low The guards on the Model 370 can be turned on or off by the user as required by the application Cables between the instrument and cryostat can be guarded even if it is impractical to continue guarding inside the cryostat Driven guards are not available for scanned inputs but they are recommended for cables that run between the Model 370 and optional scanners Theory of Operation 2 7 Lake Shore Model 370 AC Resistance Bridge User s Manual 2 4 5 5 Active Common Mode Reduction It is common for precision measurement equipment to use a differential input as the first stage of signal conditioning The differential input offers high impedance on both input voltage terminals to reduce the effect of common mode voltage voltage that acts on both leads at the same time Common mode voltage can come from many sources including external noise coupling into the lead wires and the instrument s own current source To reduce the effect of common mode voltage further the Model 370 includes an active common mode reduction circuit This circuit dynamically adjusts the oper
364. ween 1 and 16 Select Scan for the display to follow the active scan channel Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next display setup screen appears as a prompt for the source of data for the chosen display location Units for live readings can be chosen from this screen Units for Max Min and linear equation are selected during setup of those features Use the A or Y key to select one of the data sources ohms Kelvin Max Min or linear result of the linear equation Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display Operation 4 5 Lake Shore Model 370 AC Resistance Bridge User s Manual 4 4 2 Display Resolution Each display location can be assigned a resolution from 4 to 6 digits The resolution selection only affects the display readings sent over computer interface are always 6 digits It is not necessary to change display resolution from the default of 6 digits because that selection gives the maximum resolution available on the instrument If a resolution of less than 6 is chosen additional digits are truncated to make measurement noise less distracting To change display resolution continue from the display source screen or press Display Setup then Enter until the following display se
365. ween the control channel and scan channel and it will freeze the control heater output when the control channel is not active Even with these features scanning during closed loop temperature control will never be as stable as a dedicated control instrument The biggest causes of control instability are the subtle changes in cooling power that are present in low temperature cooling systems These changes are offset by changes in heater output during closed loop control to keep temperature constant When the instrument is scanning other channels it cannot read the control feedback sensor for several seconds or tens of seconds at a time When the control channel is inactive the control output freezes creating open loop control and changes in cooling power result in changing temperature The amount of instability depends on speed and mass of the cooled load Scanning while controlling should be reserved for the least critical temperature control applications All instrument specifications for temperature control are for continuous closed loop control without scanning 2 9 TEMPERATURE MEASUREMENT Temperature measurement is a popular use of resistance bridges because most of the temperature sensors that can operate at low temperatures are negative temperature coefficient resistive temperature detectors NTC RTDs and need the flexibility of a bridge Resistive temperature sensors must be treated with the same care as any other resistor being measured at low
366. will occur when the power from excitation current is large enough to warm the resistor above the temperature of its surroundings The effect is most noticeable at low temperatures because the thermal conduction at material boundaries thermal boundary resistance becomes very poor Proper attachment of the resistor to the cooling source is critical but cannot eliminate the problem entirely If noise is not suspected as a potential problem it is may be easier to measure the effect of self heating using the many current ranges available in the Model 370 than try to calculate thermal resistance Begin by cooling the resistor to the desired temperature and measure its value using current excitation and select a current in the middle of the available choices Step the current down a range approximately a decade in power and measure resistance again A change in measured value indicates self heating was present on the higher range Continue reducing current until the change is no longer significant to the measurement If the first step down did not indicate self heating step current up to improve resolution Thermal resistance changes with temperature so this process must be repeated at several temperatures 2 12 Theory of Operation Lake Shore Model 370 AC Resistance Bridge User s Manual 2 7 2 Excitation vs Signal to Noise Since excitation power has such a dramatic effect on self heating why doesn t everyone use the lowest excitation range The
367. x Gamma Y r Lambda A A Tau T T Delta A Mu u M Upsilon v Y Epsilon e E Nu v N Phi o p Zeta 6 Z Xi g Chi x X Eta n H Omicron o o Psi y Y Theta 8 Pi T II Omega 0 Q Glossary of Terminology A 3 Lake Shore Model 370 AC Resistance Bridge User s Manual ground A conducting connection whether intentional or accidental by which an electric circuit or equipment is connected to the Earth or to some conducting body of relatively large extent that serves in place of the Earth Note It is used for establishing and maintaining the potential of the Earth or of the conducting body or approximately that potential on conductors connected to it and for conducting ground current to and from the Earth or of the conducting body H Symbol for magnetic field strength See Magnetic Field Strength Hall effect The generation of an electric potential perpendicular to both an electric current flowing along a thin conducting material and an external magnetic field applied at right angles to the current Named for Edwin H Hall 1855 1938 an American physicist hazard communication standard HCS The OSHA standard cited in 29 CFR 1910 1200 requiring communication of risks from hazardous substances to workers in regulated facilities hertz Hz A unit of frequency equal to one cycle per second hysteresis The dependence of the state of a system on its previous history generally in the form of a lagging of a physical effect behind its cause Al
368. y to select a channel number between 1 and 16 Choose the selection All and the changes will be applied to all channels at the same time Press Enter to accept the new selection and continue to the next setting screen Press Escape to cancel the new selection and return to the normal display The next display setup screen appears as a prompt for the source of Max Min data for the selected channel Changing the data source will reset captured values for the selected channel Use the A or V key to select one of three data sources ohms Kelvin or linear result of the linear equation Press Enter to accept the new selection and return to the normal display Press Escape to cancel the new selection and return to the normal display Operation 4 23 Lake Shore Model 370 AC Resistance Bridge User s Manual Max Min Capture Continued To reset the Max Min feature press the Max Min key twice The following message appears as a prompt after the key is pressed the first time Captured values for all inputs will be reset at the same time 4 13 LINEAR EQUATION AR The Model 370 can scale one source of reading data from each channel with either of two linear equations Scaled readings can be displayed or sent to the analog voltage output as a control signal The feature can be configured to calculate an error value AR or a scaled error value such as 10AR The two equations MX B and M X B have subtle diffe
369. ynchronize the passing of data dynamic link library DLL A module that contains code data and Windows resources that multiple Windows programs can access electromagnet A device in which a magnetic field is generated as the result of electrical current passing through a helical conducting coil It can be configured as an iron free solenoid in which the field is produced along the axis of the coil or an iron cored structure in which the field is produced in an air gap between pole faces The coil can be water cooled copper or aluminum or superconductive electrostatic discharge ESD A transfer of electrostatic charge between bodies at different electrostatic potentials caused by direct contact or induced by an electrostatic field error Any discrepancy between a computed observed or measured quantity and the true specified or theoretically correct value or condition excitation Either an AC or DC input to a sensor used to produce an output signal Common excitations include constant current constant voltage or constant power Fahrenheit F Scale A temperature scale that registers the freezing point of water as 32 F and the boiling point as 212 F under normal atmospheric pressure See Temperature for conversions feedback control system A system in which the value of some output quantity is controlled by feeding back the value of the controlled quantity and using it to manipulate an input quantity so as to bring the value
370. ype exit to end program Label2 Name IblCommand Caption Command Label3 Name IblResponse Caption Response Textl Name txtCommand Text lt blank gt Text2 Name txtResponse Text lt blank gt Command1 Name cmdSend Caption Send Default True Forml Name frmIEEE Caption IEEE Interface Program 12 Add code provided in Table 6 2 a In the Code Editor window under the Object dropdown list select General Add the statement Public gSend as Boolean Double Click on cmdSend Add code segment under Private Sub cmdSend_Click as shown in Table 6 2 In the Code Editor window under the Object dropdown list select Form Make sure the Procedure dropdown list is set at Load The Code window should have written the segment of code Private Sub Form _Load Add the code to this subroutine as shown in Table 6 2 13 Save the program 14 Run the program The program should resemble the following u IEEE Interface Program Type exit to end program Command Bil ES Response 15 Type in a command or query in the Command box as described in Paragraph 6 1 5 5 16 Press Enter or select the Send button with the mouse to send command 17 Type Exit and press Enter to quit 6 8 Remote Operation Lake Shore Model 370 AC Resistance Bridge User s Manual Table 6 2 Visual Basic IEEE 488 Interface Program Public gSend As Boolean Global used for Send button state Private Sub cmdSend Click