Home
        ARES User Manual
         Contents
1.       gt    13 7 gmcm  Ay 277 19       Figure 2 6  ARES Front Panel LCD Display    ARES User Manual    Actuator Bearing Locks  Standard Transducers  2K STD and 10K STD     The standard transducers do not require a bearing lock     FRT without Normal Force  100 FRT and 200 FRT     The 100 and 200 FRTs use a bearing lock consisting of an aluminum block  Figure 2 7  that is fastened to the  transducer by two machine screws  The block is machined at an angle that allows the bearing to be locked and  unlocked by sliding the block         LOCKED POSITION    SCREWS  2   UNLOCKED POSITION    Figure 2 7  Bearing Lock  100 FRT and 200 FRT    Procedure for Locking and Unlocking Air Bearings  100 FRT and 200 FRT    Refer to Figure 2 7 while performing the following procedures     To Lock the FRT air bearing   1  Read the Caution on page 11   2  Ensure that instrument power is off and air is applied to the FRT   3  Loosen both screws and slide the block to the locked position   4  Tighten both screws     To Unlock the FRT bearing   1  Read the Caution on page 11   2  Ensure that instrument power is off and air is applied to the FRT   3  Loosen both screws and slide the block to the unlocked position   4  Tighten both screws     ARES User Manual    FRT with Normal Force  FRTN     The FRIN transducers utilize a bearing lock consisting of a steel pin with a clamp on both ends  Figure 2 8   that is inserted through the transducer housing  On most 2K FRTN1 transducers  the pin is installed suc
2.      3     4        Ensure that Orchestrator online indicators do  not indicate a problem  if so go to 7     Ensure that the commanded temperature is  correct and reasonable  if not  command a  suitable temperature     Ensure that the SET USER TEMPERATURE LIMIT  is set to a reasonable value  if not  set  accordingly     Ensure that the temperature reported by  Orchestrator is close to actual temperature   If the temperature is 650  C  the one or both  of the PRTs are open     Visually inspect PRTs for damage  If  damaged  go to 7     An oven fuse may be open   Call Technical Service     Ensure that the temperature reported by  Orchestrator is close to actual temperature  if  not  change temperature control to Mode 3  If  reported temperature is now correct  then  Tool PRT or upper oven PRT control loop is  malfunctioning  If reported temperature is  incorrect then lower oven PRT control loop is  malfunctioning     Visually inspect PRTs for damage  If  damaged  go to 4     An oven fuse may be open   Call Technical Service     Ensure that LN  supply pressure is adequate   i e   20 30 psi   if not  check supply lines   Ensure that LN  filter is not clogged  if so  go  to 5     Ensure that  when the LN2 Controller is on   vent gas exits from the muffler on the LN2  Controller  if not  go to 5    Ensure that there is no leaking around  dewar     Call Technical Service     ARES User Manual    Table 6 2  Instrument Operation Troubleshooting Guide  Continued     PROBLEM CORRECTIVE AC
3.      Figure 5 11  Setup Instrument Temperature Options Showing the Adjustable Temperature  Calibration Table    ARES User Manual 225    ARES User Manual       Chapter 6    Maintenance  General Information    This chapter contains the following information     e Routine Maintenance  Routine maintenance consists of tasks that we recommend you perform on a  periodic basis     e Special Maintenance  Special maintenance tasks can be performed only by qualified electronic  technicians     e Troubleshooting Guide  A troubleshooting guide is supplied to assist you in diagnosing selected  problems     Routine Maintenance    Cable and Hose Inspection    Damage to the AC power cords can cause a safety hazard  Periodically inspect these items as follows     AC Power Cords    Remove AC power to the instrument as follows     1  Push the Main Power Switch to the OFF  O  position    2  Remove the POWER IN plug from the AC main source    3  Inspect all the cords for frayed insulation or exposed bare copper wire  especially in the immediate  vicinity of the plugs on either end  If any damage is found  notify TA Instruments Technical Service     Apply AC power to the instrument as follows     1  Install the POWER IN plug in the AC main  line voltage source    2  Push the Main Power Switch to the ON  I  position     Air Hoses  Remove AC power to the instrument as follows     Push the Main Power Switch to the OFF  O  position    Remove the POWER IN plug from the AC main  line voltage source    Lo
4.      lg Rig       13 97mm rn  INNER CUP DIAMETER   27 94 mm         PEL 0     Marry 14 75 mm i  INNER BOB DIAMETER   29 50         R     16 00 mm    OUTER BOB DIAMETER   32 00 mm      R     17 00 mm    OUTER CUP DIAMETER   34 00 mm     Figure 4 16  Double Wall Couette     Original fluid bath set up showing inner and outer cup as well as tool dimensions       ARES User Manual    Tool Installation  Fluid Bath 2    The Double Wall Couette lower tool  cup  mounts into the Fluid Bath 2  see Chapter 2  to ensure precise  thermal control  Install the bath on the test station prior to mounting the tool in the bath     a  CAUTION  Never place any lower tool into the bath if the temperature      of the lower tool is cooler than that of the bath  Placing a tool into a  warmer bath will result in expansion of the tool during use  After  expansion  the tool may not be removable without damaging your bath        We suggest that you partially insert the tool by placing the tool loosely  into the bath  Allow the lower tool temperature to match that of the bath   then fully thread the lower tool into the bath     1  Select the Set GaplInstrument Control function under the Control menu in Orchestrator  Use the   Send to Top  button to raise the stage to the loading position    2  Verify that the motor is on  then mount PRT and lower tool  Cup  into the Fluid Bath 2  and the upper  tool  Bob  on the transducer shaft    3  Using the Set Gap Instrument Control function in Orchestrator  zero the n
5.     Operating range is defined as the region bounded by the maximum and minimum complex modulus G that  can be measured by each transducer type using a specific geometry  For each geometry there are specific  factors that will affect the operating range for that geometry  Additionally  the following instrument specific  factors affect the operating range of all geometries     e Minimum torque that can be measured by the transducer    e Transducer compliance  e Maximum strain that can be generated by the motor    To calculate the minimum and maximum G that can be measured by each transducer type  use the following  equation     x   K   A       where  A    Stress Constant for the specific geometry    K    Strain Constant for the specific geometry and C is computed from the following     C for G  MAXIMUM C for G  MINIMUM    2K FRTN1    2K FRTN1E 7 J 1 15e 06  rad gecm  Za  see note below     Transducer    1K FRTN1 J 4 9 e 06    see note below   J    2K STD  10K STD F J   2 60e 06  see note below     100 FRT for    100  J   2 60e 05    200 FRT a ee  100 FRTN1 for   10  J   2 60e 06    200 FRTN1    a  see note below     NOTE  The values for M  gecm  and 6  rad  are found in the specification tables in Chapter 1   Table 1 6 through Table 1 13   Pick the correct values for your specific transducer and  motor combination        ARES User Manual 151    To determine the maximum or minimum complex viscosity 1   that can be measured at a given frequency  use  the following formula   G   
6.     Raise the stage to provide sufficient room for sample loading     Sample Loading  Original Fluid Bath    The nominal sample volume is 8 ml  In general  pour the sample into the cup  then lower the bob until the gap  again reads zero  and the sample appears as shown in Figure 4 15C  If the gap is set as described in the  previous section this will ensure that the nominal bob length is   31 90 mm  Note that the upper surface of the bob must be as flush as possible to the upper surface of the cup     Tool Maintenance    To facilitate cleaning of the lower tool  the inner cup can be removed from the outer cup when necessary   Inspect the o ring for cuts or other damage and replace it if necessary     ARES User Manual      STRAIGHT EDGE       1 BRING UPPER SURFACE OF BOB  FLUSH WITH UPPER SURFACE OF CUP  BY RAISING BOB UP TO STRAIGHT EDGE        B    UPPER SURFACE OF BOB  UPPER SURFACE OF CUP    FILL SAMPLE  TO HERE UPPER SURFACE OF BOB      MUST BE FLUSH WITH  a  UPPER SURFACE OF CUP  BOB    T EFFECTIVE BOB LENGTH   31 90 mm  WITH SAMPLE LOADED AS SHOWN  SAMPLE    SAMPLE VOLUME   8 ml  cur    MAN Ua                CUP INNER WALL       C    Figure 4 15  Double Wall Couette for the Original Fluid Bath   Set Up and Use    A and B   Setting the Gap  C   With Sample Loaded     ARES User Manual        THESE TWO SURFACES MUST  BE ALIGNED TO OBTAIN  EFFECTIVE BOB LENGTH SHOWN    MOUNTING    SCREW    a       INNER    CUP L 31 90 mm   EFFECTIVE BOB LENGTH   BOB  he  OUTER OO  CUP      
7.     THIS DISTANCE  IS THE  SAMPLE LENGTH    SAMPLE       LOWER FIXTURE       SLIDING CLAMP    SAMPLE    NOTE    Center sample in  fixture using Centering  Lines as guides     Figure 4 11  Torsion Rectangular with Sample Loaded    ARES User Manual    Torsion Rectangular   Original Design    Strain Constant Stress Constant    2 T  1  0 378 T 3  18      W K   1000    WT     Variables  Gc   Gravitational constant    980 7  cgs  or 98 07  SI   T   Thickness of sample  mm   W   Width of sample  mm     L   Length of sample  mm           Options    Insets to accommodate thicknesses from 0 76 mm to  6 35 mm    Environmental Systems Torsion Rectangular Tool  Ambient  original design   Oven       General Information    The Torsion Rectangular tool is used for testing solid materials with high modulus  including thermosets   thermoplastics and elastomers  The sample is held in tension between the upper and lower tool  Several  inserts are provided to accommodate samples of varying thicknesses     Sample Dimensions    To prepare samples that fit within the physical constraints of the tool  use the following guidelines     e Maximum Sample Width  12 7 millimeters  e Typical Sample Length  45 millimeters  e Sample Thickness depends on the size of the insert used     INSERT DESIGNATION MAXIMUM SAMPLE THICKNESS  MM        184 ARES User Manual    Always use the correct size insert for the sample thickness  If the sample does not fit tightly in the insert   erroneous data may result  Shims can 
8.   10 A  60 Hz  Europe   220 VAC  10 A  50 Hz  Japan     100 VAC  10 A  50 Hz   60 Hz with boost transformer     Standard   Mechanical Chiller 220 VAC  10 A  60 Hz  AC PO Option   200 VAC  10 A  50 Hz       ARES User Manual    Table 1 5  Operating Specifications  Environmental Controller    Operating Parameter Forced Convection Fluids Bath  Oven  Temperature Range  Standard Ambient to  600   C  10   C to 150   C  30   C to 150   C  Optional  150   C to  600   C    S a Depends on 30   C min  at 20   C  empero ramprals OE N Circulator used circulator fluid temp   Temperature Stability   05  C  0 01   C  0 1  C  at thermal equilibrium o aa a    Motor Performance Specifications       Table 1 6  Specifications  High Resolution Motor  HR     PARAMETER DYNAMIC MODE STEADY MODE    Angular Displacement Range 0 005 to 500 milliradians Not Applicable    Frequency Range 1E 5 to 500 rad sec Not Applicable  Rotational Rate Range Not Applicable 0 001 to 100 rad sec          Table 1 7  Specifications  High Torque Motor  HT     Angular Displacement Range 0 005 to 500 milliradians Not Applicable  Frequency Range 1E 5 to 100 rad sec Not Applicable  Rotational Rate Range Not Applicable 0 001 to 100 rad sec       Table 1 8 Specifications  Low Shear Motor  LS     Angular Displacement Range 0 005 to 500 milliradians Not Applicable    Frequency Range 1E 5 to 100 rad sec Not Applicable  Rotational Rate Range Not Applicable 1 X 10   to 200 rad sec    ARES User Manual       Transducer Operating Specific
9.   94  96  150   Strain Controlled Steady Test Methods  116  Strain Controlled Transient Test Methods  121  Stress  93  94  96   Stress Ramp Test  139   Stress Relaxation  100  123    Suspensions  149    y    TA Instruments Offices  237  Tan 6  95  Technical Support  14  Temperature Calibration  224  Tensile Strain  93  Tensile Stress  93  Test Tools  general guidelines  154  160  upper tool installation  154  Test Modes  100  Test Station  35  Thermal Expansion  160  Thermopastics  149  Thermosets  149  Thermosetting Resins  150  Thin Films  149  Thixotropic Loop  130    Torque  Normal Relaxation  132    ARES User Manual    Torsion Rectangular  New Design  test tool   constants  177  tool installation  180  general information  177  operating range  178  sample loading  181    Torsion Rectangular  Original Design  test tool   constants  184  tool installation  187  general information  184  operating range  185  sample loading  187  trademarks  5  Transducer  description  37  principles of operation  26    Transient Step Strain  123    Troubleshooting Guide  232    V    Vane Tool  See Couette  Viscoelastic  93   Viscosity  93   Viscous Modulus  E         94    Viscous Stress  t          94    ARES User Manual          Appendix A    Complex Modulus Limits    This appendix provides complex modulus limits for various tools     Appendix Table A1 1  Complex Modulus Limits for Parallel Plate   2K FRTN1 and 2K FRTN1E Transducers     PLATE DIAMETER  mm  G  Maximum  dynes cm     G  Mi
10.   C       ARES User Manual    Circulator Options    The Peltier must be connected to a circulating fluid source  Any manually controllable circulator with  heating cooling capabilities can be used to operate the Peltier system over its entire specified temperature  range  This can also be the same circulator used with the Fluid Bath  The standard fittings installed onto the  ARES Peltier Assembly fluid hoses accommodate computer controlled circulators that are sold by TA  Instruments     The Peltier Circulator  manufactured by TA Instruments  is an effective yet inexpensive means of providing  non temperature controlled circulating fluid  Instructions for use of the Peltier Circulator are printed on the  label that is affixed to the circulator  If using the Peltier Circulator  you must remove the standard fluid fittings  from the ARES Peltier Assembly fluid hoses  then install the special pair of fittings that are supplied for use  with the Peltier Circulator    selecting a Thermal Operating Range    The temperature of the circulating fluid determines the Thermal Operating Range of the Peltier system  To  select a thermal operating range  fluid temperature must be set in accordance with the following guideline     Assuming that the ambient temperature in the vicinity of the heat sink is  20  C  a low end differential  ATL  of  approximately  40  C exists between the fluid temperature and the lower limit of the thermal operating range   A high end differential  ATy  of approxi
11.   Please refer to the preceding documents  as well as this manual  prior to  and during test setup and execution     ARES User Manual 99      Test Modes    The ARES is capable of performing dynamic  steady and transient  time based  mechanical tests  These are    organized in Orchestrator under the following categories     e Strain Controlled   Dynamic    Dynamic  oscillatory  mechanical measurements where strain is controlled    and stress is measured     e Strain Controlled   Steady    Steady mechanical measurements where strain is controlled and stress is  measured  After the sample is allowed to reach  steady state   measurements are averaged over a period    of time     e Strain Controlled   Transient    Static time based tests where strain is controlled and stress is measured   Data are collected rapidly  to look at the sample response over time to an applied deformation   e Stress Controlled   Transient    Similar to Strain Controlled Transient tests  but where stress is controlled    and strain is measured     Specific categories and tests are selected from within the Edit Start Test function of Orchestrator  The    available tests used with the ARES are     Strain Controlled   Dynamic   e Dynamic Single Point Measurement  e Frequency Sweep   e Dynamic Temperature Step   e Frequency Temperature Sweep   e Dynamic Strain Sweep   e Dynamic Time Sweep   e Dynamic Temperature Ramp   e Mulitwave Single Point   e Multiwave Temperature Ramp    Strain Controlled     Steady  e 
12.   S    PARALLEL PLATE  PRT  PN 700 03469        BATH 2 COUETTE   lower fixture       BATH 2   DOUBLE WALL COUETTE     lower fixture     Figure 2 24  Fluid Bath 2 Lower Tool and PRT options     Lower Tool Installation and Removal    The lower tool is mounted into the bath well  The well is threaded to mate with the threads on the tool  The  tool should be tightened clockwise  by hand only  Before installing the lower tool ensure that the correct PRT  is installed and operational     To remove the tool  use the supplied wrenches  If necessary  to loosen the tool from the bath well  The  wrench used on the tool is labeled  FOR TOOL  and the wrench used to hold the bath is labeled  FOR  COVER   Keeping the lower tool and bath well threads clean  and free of damage  will help ensure easy  installation and removal     Filling the Circulator    Depending on the type of circulator in use and the desired operating range of the circulator  fill the circulator  with fluid as specified in Table 2 6  The Julabo FS 18 circulator requires about 2 liters of fluid  and should be  filled to within 5 mm from the top  However  refer to your actual Bath documentation for specific circulator  filling and operating guidelines  as well as other bath fluid options for your application  Also refer to the bath  fluid MSD for guidelines regarding the safe handling of your particular bath fluid     ARES User Manual    Because of the construction of the Fluid Bath 2  the seals are very delicate  It is imp
13.   and  transducer  mount the pulley wheel  or hub  appropriately    Install the calibration tool and pulley as shown in Figure 5 2    a  For 10K STD Transducers  Two pulleys are supplied  Referring to Figure 5 2  insert a pulley  into each side of the Test Station frame  ensuring that the flat machined into each pulley shaft  faces the access hole  Please note that Figure 5 2 shows the pulley on the right side only    b  For all other transducers  Insert the pulley into the Test Station frame as shown in Figure 5 2   ensuring that the flat machined into the pulley shaft faces the access hole    Secure the pulley s  by tightening the setscrew in the access hole  Figure 5 2  using a 1 5 mm hex  wrench  Ensure that the calibration tool and pulley are installed as shown in Figure 5 3  If you have a  10K STD transducer  an additional pulley should be installed on the left    Using the Set Gap Instrument Control function in Orchestrator  zero the normal force and torque on  the motor using the  Offset Normal Force to Zero  and  Offset Torque to Zero  buttons  Exit the form  when finished    Prepare the calibration line specified in Table 5 2  depending on the transducer in use  Prepare the  calibration line s  by making a loop at each end     ARES User Manual            INSTRUMENT       ACCESS HOLE    TRANSDUCER  COVER    HUB PULLEY     ONE EITHER SIDE        CALIBRATION FIXTURE    Figure 5 2  Installation of the Calibration Tool and Single Pulley       Figure 5 3  Calibration Tool
14.   n   4 2     where  n   Complex viscosity  Poise     G   Complex Modulus  dynes  cm          Frequency  rad sec     Using a spreadsheet application such as Microsoft  Excel     you can use the equations above to plot the range  of complex viscosity that can be tested for a given geometry transducer combination as follows     1  Calculate G  MAXIMUM and G  MINIMUM using equation 4 1    2  Substitute the G  MAXIMUM value into equation 4 2  then determine the upper limit of operation by  calculating 1  at each O  0 values chosen to be from the lowest to highest frequencies within the  transducer operating range     3  Substitute the G  MINIMUM value into equation 4 2  then determine the lower limit of operation by  calculating N  at each       4  Generate an X Y scatter plot of complex viscosity N   Y axis  versus frequency    X axis      The region between the upper and lower limits of operation is the range of complex viscosity that can be  tested  Please note that the values for J above are nominal values  and will vary slightly between transducers   Accordingly  and per good standard practice  care should be taken  to ensure the data are valid  when testing  near the upper or lower limits for a given system     Possible Sources and Causes for Error    This section contains information describing errors that can occur as a result of tool and sample limitations     Discrepancies in Sample Geometry    The force generated by a sample for a given amount of strain is a function o
15.   use  the following formula     nea  4 2   a    162 ARES User Manual    where  n   Complex viscosity  Poise     G   Complex Modulus  dynes  cm         Frequency  rad sec     Using a spreadsheet application such as Microsoft  Excel     you can use the equations above to plot the range  of complex viscosity that can be tested for a given geometry  transducer combination as follows     1   2     4     Calculate G  MAXIMUM and G  MINIMUM using equation 4 1    Substitute the G  MAXIMUM value into equation 4 2  then determine the upper limit of operation by  calculating N  at each O     values chosen to be from the lowest to highest frequencies within the  transducer operating range     Substitute the G  MINIMUM value into equation 4 2  then determine the lower limit of operation by    calculating 1  at each      Generate an X Y scatter plot of complex viscosity 1   Y axis  versus frequency     X axis      The region between the upper and lower limits of operation is the range of complex viscosity that can be  tested  Appendix 1 contains tables of G  values for some tool combinations  transducers and a standard motor     Tool Installation    Select the Set Gap Instrument Control function under the Control menu in Orchestrator  Use the   Send to Top  button to raise the stage to the loading position    Verify that the motor is on then mount the upper and lower tools on the actuator shafts    Using the Set Gap Instrument Control function in Orchestrator  zero the normal force and to
16.  0  C  Paints Per    one  200    Max 350 Min 20       Zone Number 1 2 4 4    SOWIE TIE ee ceeeeeeeseeeees  s or hems   200 i 00  50  200  Extension Value  See Mode for Units   0 05  0 007  50 0  50 0    Extension Mode ooo   Rate  mms      Hencky  Vis     Force  am        End Test y        Options  Delay OF Analog Of Ratebiar Ut    Options   End of Test   Save   s   Help   Cancel      Figure 3 16  Multiple Extension Mode Test Set up Screen    Ok         Test Options  Multiple Extension Mode Test options are as follows     e Delay Before Test   e Analog Data Input   e Motor Control Gain   Manual or Default  only for the Constant Force Mode   o Proportional Factor gain settings  manual or default of 1 0   o Integral Factor gain settings  manual or default of 0 1     The motor control gain can be adjusted to change the response of the control loop controlling stress level in  constant stress mode     ARES User Manual 125    Table 3 4  Summary of Multiple Extension Mode Test Options     Extension Description  Mode    Rate  mm s  Apply a constant rate of linear Can be used to determine the range of  displacement  Extension rate in the linear behavior in a material by plotting  sample is not constant due to change Force as a function of displacement   in sample length     Hencky  1 s  Apply a constant rate of sample strain    Used to measure the extensional modulus  Linear displacement rate is adjusted to   and properties in samples     maintain a constant sample strain rate     Rim
17.  02  High range   4 10E 03  Low range        100 FRT    Appendix Table A1 10  Complex Modulus Limits for  Double Wall Couette  Geometry  Cup OD   34 mm   Cup ID   27 95 mm  Bob OD   32 mm  Bob ID   29 5 mm      G  MINIMUM  dynes cm        TRANSDUCER TYPE G  Maximum  dynes cm      at Frequency     rad sec   5 39E 02    5 39E 03 0    100 FRTN1    2  8E 05    9 35E 02 Q   9 3DE 03      1 11E 02  High range   1 11E 03  Low range     5 57E 03  High range   5 57E 04  Low range        100 FRT    ARES User Manual    2  8E 05    ARES User Manual    
18.  115    G    Gap  automatic zero  158  enabling the gap control panel  158  manual zero  157  max force allowed option  159  read text tool gap  159  set gap  instrument control function  38  setting  general  157    Gas    input options  53  pressure specifications  53    ARES User Manual    H    Hastelloy Tool  tool installation  171  general information  165  liquid seal  168  operating ranges  165  parallel plate tool  165  use with oven  169   Hencky Strain Rate  125   high voltage warning  207   Hookean Region  93   Hooke s Law  93   Host Computer  35    Humidity Cover  83  84    Instrument Control Panel  36  65    L    LCD Display  40    LEDs  diagnostic  230    License Agreement  3  Linear Region  93  Linear Strain Sweep  107    LN   Controller   description  61   location  33   physical specifications  28  Logarithmic Strain Sweep  107    Loss Modulus  See Viscous Modulus    ARES User Manual    Main Power Switch  51    Maintenance  cleaning  228  routine  227  special  230    Manual Delay  141  Modulus  93    Motor  air pressure requirements  53  description  35  location  34  principles of operation  26  rapid shut off  36  specifications  30  turning on and off  36    Multiple Extension Mode  125  MultiWave Single Point  112    MultiWave Temperature Ramp  112    N    Newton s Law  93    O    One Cycle Correlation  146    Oven  description  58  gas selection  62  location  34  operating requirements  63  operation  65  rapid shutdown  36  signal connection to test s
19.  500 0 Min 1 00e 05  Initial Temp   25 0 PC  Max 600 0  C Min  150  0         Strain Limite         Mas 312 5000 Min 0 0031 25     one Number 1 E    Final Temp  occ Peon  00  Ramp Rate    PCminlao po  Computed Ramp Time           hi rr      58 20  Soak Time After Ramp  s or hom  foo  Time Per Measure          s or Fires  od    MM O   2  fi    0 0  PT    il       Uptions  FreShear O Delay  Of AutoTens  OF Analogi Ot Autost O MeasOps 01       Options   End of Test   Save As   Help   Cancel      Figure 3 7  Dynamic Temperature Ramp Test Set Up Screen    Suggested Uses  Suggested uses for temperature ramp tests     e To quickly analyze the behavior of samples as a function of temperature  and to determine test parameters  for subsequent Temperature Step tests    e To simulate processing conditions that a material may experience in use by programming the relevant  time  temperature profile    e To study the response of a material to changing strain levels as a function of time and or temperature    e The Temperature Ramp test is one of the most flexible of all of the test modes due to its ability to handle  complex time  temperature   strain profiles     ARES User Manual    Test Options  The following test options can be selected for use with the temperature ramp     e Steady PreShear  e Delay Before Test  e AutoTension  e Analog Data Input  e AutoStrain  e Measurement Options  o Delay Settings  o Strain Amplitude Control    ARES User Manual    MultiWave Single Point  MultiWave Te
20.  AC power to the instrument as follows   a  Push the Main Power Switch to the OFF  O  position   b  Remove the POWER IN plug from the AC main  line voltage source    2  Apply some cleaning solution onto a cotton cloth  then wring out the cloth to discharge excess water   the cloth must be damp  but not wet   3  While ensuring that excess fluid from the cloth does not enter any crevice of the instrument  use the  cloth to gently clean the desired external surfaces   4  Ensure that all surfaces of the instrument are dry   5  Apply AC power to the instrument as follows   a  Install the POWER IN plug in the AC main  line voltage source    b  Push the Main Power Switch to the ON  I  position     ARES User Manual    Lifting and Carrying the Instrument  Test Station    The Test Station weighs 275 pounds  125 kilograms   It can be safely lifted and carried only by a fork lift that is  rated to carry such weight  However  since there is no surface that offers adequate contact points for a fork lift   the Test Station is designed to be lifted only when it is mounted onto a pallet  this is the shipping  configuration  by placing the forks in the pallet     The handles on the side of the test station can be used to slide the test station on the workbench surface  You  may need to apply some lift while sliding the instrument  but do not attempt to completely lift and carry the  Test Station by the handles  as balancing the instrument is difficult  Be careful not to damage the feet on the 
21.  ARES User Manual                                                                                                                                           Torque Calibration  for All Transducers     Torque Calibration ensures that the transducer is accurately measuring torque  The calibration involves  hanging a precision weight on the calibration tool  a 2 5 centimeter moment arm that is mounted on the  transducer during calibration  The applied torque is therefore the product of the weight and the 2 5   centimeter moment  For example  hanging a 500 gram weight applies a torque of  500 g  2 5 cm    1250 gecm     NOTE  For this  and all subsequent calibration procedures  the instructions  assume that the Host Computer is connected to the ARES test station  and  that Orchestrator is running     Procedure    1          If calibrating an FRT  ensure that the transducer is set to High Range before proceeding  To set the  transducer to High Range do the following    a  Access the Set Transducer Characteristics form  Figure 5 7  by selecting the Transducer option  from the Service function of the Utilities pull down menu    b  Using the  Transducer Selected  menu  select the high range transducer then Click  Ok     Remove any test tools    Turn on the motor    Raise the stage to bring the bottom of the stainless steel transducer cover about 2 inches below the  instrument cover  Figure 5 2     Determine which calibration tool you have based upon Figure 5 1  Depending upon your tool
22.  Calibration form is within the limits  for the CALIBRATED FULL SCALE VALUE shown in Table 5 4     Table 5 4  Normal Force Calibration Weights  Applied Normal Forces  and Calibrated Full Scale  Values    APPLIED NORMAL CALIBRATED FULL  TRANSDUCER WEIGHT NORMAL FORCE SCALE VALUE  FORCE VALUE    a ay E 1000 grams ont 1000   2 gmf 2100 gmf   5    p n 613 01222  gm  998 to 1002   1995 to 2205     1K FRTN1    100 FRTN1 100 grams 100 gmf 100   0 1 gmf 105 gmf   5    200 ERTN1  p n 613 02060   99 9 to 100 1   99 75 to 110 25   2K STD 1000 grams 00H 1000   5 gmf 1575 gmf   5    10K STD  p n 613 01222  gm  995 to 1005   1496 25 to 1653 75        ARES User Manual    13  Select the  ACCEPT  button  Control returns to the Transducer Characteristics form     14  Verify that the Normal Calibration Value now displayed in the form  the high range value if the  transducer is an FRT  is the same as the Current Normal Cal value just displayed in the Transducer  Calibration form  Press  Ok      15  If the transducer in use is an FRT  remain in the Transducer Setup form and perform the following  step  If the transducer is a STD transducer  go to step 16     a  Access the Set Transducer Characteristics form  Figure 5 7  by selecting the Transducer option  from the Service function of the Utilities pull down menu    b  Copy the high range transducer Normal Calibration Value into the column containing the  transducer settings for the low range transducer     16  Click Ok     This concludes the Norm
23.  Controller is currently ready for use  In this case  the LN  level in the  Dewar flask is between 50 and 75  of the flask capacity     LN2 Off The LN2 Controller is currently disabled        66 ARES User Manual    Fluid Bath 2    Description    The Fluid Bath 2 offers precise control of sample temperature using a closed fluid re circulant system  The  operational range of the Fluid Bath 2 is  10   C to 140   C  Thermally controlled fluid  supplied by a circulator   flows through the bath  The lower test tool is mounted within the Bath Well  and attains thermal equilibrium  with the surrounding bath  The temperature of the lower tool is measured by the bath PRT  which mounts  through the Bath Well into the Motor  You can choose to control the temperature of either the lower tool or the  circulator fluid itself     The Circulator regulates the temperature of the bath fluid  and pumps the fluid through the Fluid Bath  The  circulator  as supplied by TA Instruments   is connected to  and is under the control of  the test station and  software  The fluid circulated through the bath is maintained at the temperature selected in Orchestrator  The  circulator also has its own fluid temperature regulation  which can be used as the temperature control loop for  the Bath by selecting manual temperature control Orchestrator  Presently  the standard circulator supplied by  TA Instruments is the Julabo FS 18     Installation of Fluid Bath    The fluid bath is mounted onto the motor using a
24.  Linear    Points Per Zone  300    Max 350 Min 20       Shear Rate Limits  1s     Max 200  0000 Min 0  007 000    one Mumber 1 2 4 4    Shear Rate           Meza foo foo foo  Zone Time  sor LS o oc o    Direction i  f Clockwise C Counterclockwise       Options  Delay  OFf    Options   End of Test   Save As   Help   Cancel         Figure 3 14  Step Shear Rate Test Set Up Form    ARES User Manual 121    Suggested Uses  Step Shear Rate can be used to examine the following sample characteristics   e Stress growth and relaxation at constant temperature    e Time required to reach steady state flow behavior   e Relaxation after steady shear  see Shear Rate in the description of Parameters     Test Options  The following test options are available when using Step Rate with the specified instrument types     e Delay Before Test    ARES User Manual    Stress Relaxation  Transient Step Strain     Functional Description    Stress Relaxation  transient step strain  monitors sample stress relaxation by taking measurements following a  single upward or downward step of the motor to the selected strain  Four independent measurement  zones    with a maximum of 350 data points per zone  are available  The force response  relaxation profile or G t   to  the step strain is measured in either logarithmic or linear sampling mode     Log  Logarithmic     Logarithmic sampling takes data at logarithmically incremented intervals  As an example  selecting 5 points  per zone during a 100 second zone di
25.  Mode     Outputs a DC voltage that is proportional to motor strain  actual motor angular  deflection      o Ee Scaling is  9 VDC   0 radians   5 VDC   0 5 radians    Steady Mode     Outputs a DC voltage that is proportional to transducer normal  axial  force   Scaling is  0 VDC   0 gmf    5 VDC     full scale normal force    As the instrument is shipped  STRAIN NORMAL IN is connected to  and outputs  STRAIN   NORMAL   the same signal as  STRAIN NORMAL OUT  This connector is normally used by  IN TA service personel for diagnostic purposes     Outputs a DC voltage that is proportional to the selected strain  which drives the  COMMAND OUT motor in dynamic mode   Scaling is     VDC   0 radians  10 VDC   0 5 radians    As the instrument is shipped  COMMAND IN is connected to  and outputs the  COMMAND IN same signal as  COMMAND OUT  This connector is normally used by TA service  personel for diagnostic purposes     Outputs a DC voltage that is proportional to transducer normal  axial  force   Scaling is  VDC   0 gmf   5 VDC     full scale normal force       NORMAL OUT    ANALOG 1 IN Accepts a   10 VDC input signal that can be sampled at 1 Hz and stored in the  data file  This is the input for the Analog Data input feature in Orchestrator    LN2 Communications interface between the optional LN2 controler and the  instrument  Connected to LN2 controller     Communications interface between the Host Computer and the instrument     Unless you have selected a different port using Orche
26.  OF CUP  f z        COM as mm    NO LOWER THAN 5 mm BELOW  BOB UPPER SURFACE OF CUP    SAMPLE      CUP                      l    Figure 4 13  Couette with Sample Loaded    Vane Tool    The Vane Tool is designed to replace the bob under certain circumstances  Itis primarily used with materials  that are highly structured  such as foams and lotions  which may tend to slip with a normal bob  This slippage  could mistakenly be interpreted as a yield stress using a normal bob  For these fluids  when using the vane  tool  the material within the vanes moves as a solid plug  However  for less shear thinning fluids  there will be  secondary flows between the vanes  which will result in incorrect viscosity shear rate data  so it is important to  use the correct tool for the material  The vane tool is also useful when testing  chunky  fluids  or that contain  larger particles  Many food items would fall into this category     The vane tool is used to do Creep Recovery  to measure Yield Stress  or low speed Steady shear testing  The  Vane Tool should only be used for low speed steady testing     The vane tool is handled similarly to the standard bob  The Couette Geometry should still be selected in  Orchestrator The tool dimensions are determined as shown in Figure 4 10  The vane tool is designed for use  with the 34 mm cup  The vane tool should be lowered carefully into the sample to minimize disturbing the  sample     ARES User Manual 193    194       Figure 4 14  Vane Tool    ARES Us
27.  Options Form Used to Input the Fluid Bath System Configuration    NOTES    1  Make sure the following are selected   e Instrument Setup   TEMPERATURE CONTROL    e Temperature Control   BATH  INSTRUMENT CONTROLLED ONLY     ARES User Manual    NOTES  continued       2  Ensure that the maximum and minimum temperatures corresponding to  the desired circulator operating range and fluid  See the previous topic  Filling the Circulator  are entered in the Max and Min Temperature fields     3  Select either Temperature Loop Control option  using the following  information as a guideline     e Circulator Temperature  When selected  the temperature of the  fluid in the circulator is maintained at the commanded temperature   The lower tool PRT is independent of the circulator temperature  and  due to its location  will report temperatures that are slightly different  than the commanded  circulator  temperature  The circulator  temperature control should be used when running step or ramp type  temperature studies where a controlled  steady  temperature change   or speed  is more critical than actual sample temperature    e Tool Temperature  When selected  the temperature of the lower test  tool is maintained at the commanded temperature  Tool Temperature  control is the recommended setting for isothermal  and some step   testing  where controlling the actual sample temperature is most  important  When the commanded temperature has been achieved   wait 20 to 30 minutes at the commanded te
28.  STRAIN 100         TEMPERATURE Current ambient temperature    Ensure that the phase angle  6  is between 87   and 92   throughout the frequency range  If the values obtained  from this test are outside this range  please contact Technical Services for further assistance     This concludes the phase angle check     ARES User Manual 219    Strain Calibration Check    Strain Calibration Check allows you to determine if actual strain  the angular deflection of the motor   corresponds to a commanded strain  In general  this procedure involves selecting the parallel plate geometry   then modifying the tool dimensions to achieve a strain constant of 1  This action results in commanded strain  being identical to actual motor deflection  1 e   if you command 25  strain  the motor should deflect   0 25  radians from dynamic zero position     While running a Dynamic Time Sweep  a calibration pointer attached to the motor allows you to visually  examine motor position in relation to calibration marks etched onto the motor cover  The calibration marks  are graduated in increments of 0 1  0 25  and 0 5 radians from either side of dynamic zero position     Additionally  you can check the STRAIN reported by the Orchestrator online parameter display  which should  indicate 25      Procedure  1  Remove any test tools   2  Turn on the motor and ensure that it is in dynamic mode   3  Install the pointer onto the motor as shown in Figure 5 8     4  Use the Strain Offset to align the pointer with
29.  Shear  1 s  Apply a constant shear rate to the Measure the squeeze flow properties of a  sample in squeeze flow  material     Rate corresponds to the shear rate at  the edge of the sample     Force  gmf  Apply and maintain a constant force on   This can be used to provide information on   the sample  the creep behavior of a material   End Test Used to indicate that a zone is not  used and that the test should end        ARES User Manual    Arbitrary Waveshape Test    Functional Description    The Arbitrary Waveshape test allows you to define the strain history used to deform the sample by supplying  one or more equations for strain as a function of time  Up to four different equations  each with a specified  time period can be used  There are also 4  Zones  used for data collection  with each Zone capable of sampling  over a time interval that is independent of the time specified for the waveform     You should first define the time intervals for each of the data collection Zones  Figure 3 17   and the number  of points  samples  to collect in each Zone  The second part of the form is used to enter equations for the  strain as a function of time  as well as a    playback    time  Wave Time   which determines the period of time  over which the equation is to be played back  Note that the strain is always given in strain units  not percent  strain   and the limits  based on the current geometry  are given below the equation fields  The waveform  cannot be built if the strain v
30.  User Manual    Lower Tool Installation    To install a lower tool into the fluid bath  place the tool into the bath  Figure 2 36  and apply downward  pressure until the tool is seated flush against the bath     of the lower tool is cooler than that of the bath    Placing a tool into a warmer bath will result in expansion of the tool  during use  After expansion  the tool may not be removable without  damaging your bath     O CAUTION  Never place any lower tool into the bath if the temperature    We suggest that you partially insert the tool by placing a phenolic  spacer between the upper lip of the lower tool  and the surface of the  bath  Allow the lower tool temperature to match that of the bath  then  remove the spacer and fully insert the lower tool     rem Lower Fixture                l a Bath Well    Pi        ls    Fi       NOTE     install the bath PRT prior to  installing the fixture     p Fluids Bath       Figure 2 36  Installation of Lower tool into Fluids Bath    ARES User Manual    Circulator Connections    Figure 2 37 shows fluid connections between the bath and Test Station  The circulator should be positioned on  the floor below the work area  Make sure all hoses are installed completely onto their respective hose barbs  and that the hose clamps are tight  Connect the supplied cable between the Circulator RS 232 connector  and  the Test Station Signal Panel CIRCULATOR connector     Inlet Line Clamp    A clamp should be installed on the hose running from the c
31.  Using a spreadsheet application such as Microsoft  Excel     you can use the equations above to plot the range  of complex viscosity that can be tested for a given geometry  transducer combination as follows     1  Calculate G  MAXIMUM and G  MINIMUM  using equation 4 1  for a sample of fixed length and width  and the minimum and maximum thickness the tool can accommodate     2  Generate an X Y scatter plot of sample thickness  Y axis  versus complex modulus  G   X axis      The region between the upper and lower limits of operation is the range of complex modulus that can be  tested     ARES User Manual    Coefficient of Thermal Expansion  a     When testing at other than ambient temperatures  the coefficient of thermal expansion for Torsion Rectangular  geometry is defined as     AL 1    At Lo  where   Coefficient of Thermal Expansion a    At   Change in temperature    C   Lo   Original length of sample  mm   AL   Change in length of sample  mm   Positive AL indicates increasing sample length    Tool Installation    1  Select the Set Gap Instrument Control function under the Control menu in Orchestrator  Use the   Send to Top  button to raise the stage to the loading position    2  Verify that the motor is on  and in dynamic mode    3  Mount the upper and lower tools on the actuator shafts    4  Using the Set Gap Instrument Control function in Orchestrator  zero the normal force and torque on  the motor using the  Offset Normal Force to Zero  and  Offset torque to Zero  bu
32.  VALUE is nominally about 5   higher than the maximum force that is  measurable by the transducer in use  This  value should not be zero  nor should it be a  very large number such as 1E 5  If it is   enter the correct full scale value for the  transducer in use  see  nstrument  Specifications   then repeat the calibration     2  Exit the calibration operation and cycle the  instrument main power  turn off  then on      Repeat the calibration   3  Call Technical Service    PHASE ANGLE CALIBRATION    During Phase Angle Calibration with a steel Repeat the calibration  only once    sample  the phase angle is not within the limits    specified in the calibration procedure  An eee con echon aie    offset are reasonable   Ensure sample is loaded correctly     Verify static force is correct and in proper  direction     Remove the sample  exit the calibration  operation  then cycle the instrument main  power  turn off  then on   Re load the  sample and repeat the calibration     Call Technical Service       ARES User Manual    Operation    Table 6 2  Instrument Operation Troubleshooting Guide    PROBLEM CORRECTIVE ACTIONS    MOTOR    Motor does not engage or respond to command  Ensure that the motor is on     Ensure that the instrument is not in an OVERLOAD  condition  OVERLOAD is indicated by an online  indicator  If an OVERLOAD is indicated  reset the  test using the function END TEST RESET     LS Motor Only  Ensure that motor air pressure is  60 psi   Call Technical Service   Moto
33.  and Single Pulley Installed    ARES User Manual    9     10     11     12     13     14     Table 5 2  Calibration Lines    TRANSDUCER   CALIBRATION LINE      2K FRTN1 Cut one length monofilament line    2K FRTN1E  part number 613 01075   1K FRTN1    2K STD  100 FRTN1    200 FRTN1 Cut one length of thread    100 FRT  part number 613 00716     200 FRT  Cut two separate lengths of monofilament  10K STD line  part number 613 01075     Access the Transducer Characteristics form using the Orchestrator function Calibrate Instrument  under the Utilities pull down menu        Select the  XducerCal  button  Establish a zero torque reference value by selecting the  Zero  button   Wait about 30 seconds  during which time the instrument takes several readings to establish a zero  normal reference  When zeroing is completed  the Transducer Calibration form is displayed  Figure  5 4   The zero value displayed in this form should be less than 0 1  of the full scale Torque value     If  after selecting the  ZERO  button  the TORQUE value displayed is either very high  such as 1E 5  or  exactly zero  refer to the Troubleshooting Guide     NOTE  Do not hang any weights until after at least one  1  zero reading has  been taken     Apply a calibrated torque  Figure 5 5   depending on your transducer  as follows   a  Place one end of the line over the hub on the calibration tool  and place the line in the groove  of the pulley     b  Hang the weight specified in Table 5 3  depending on the trans
34.  and cups using various  transducers and a standard motor     Tool Installation    If you are installing the lower tool into either of the fluid baths  please refer to the appropriate section of  Chapter 2  for your specific bath  for additional information before installing the tool into the bath     1  Select the Set Gap Instrument Control function under the Control menu in Orchestrator  Use the   Send to Top  button to raise the stage to the loading position    2  Verify that the motor is on then mount the upper tools on the transducer shaft and lower tool into the  fluid bath or actuator shaft depending upon which environmental control system is being used    3  Using the Set Gap Instrument Control function in Orchestrator  zero the normal force and torque on  the motor using the  Offset Normal Force to Zero  and  Offset Torque to Zero    buttons     Sample Loading    In general  pour the sample into the cup then lower the bob until the upper surface of the bob is 1 to 2 mm  below the surface of the sample  Figure 4 13   Note that the upper surface of the bob must be between zero  and five millimeters below the upper surface of the cup  If this is not the case  adjust the sample volume  accordingly  Nominal sample volumes are as follows     Bob   Cup size Sample Volume  16 5 mm   17 mm  25 mm   27 mm    32 mm   34 mm  Vane   34mm       192 ARES User Manual    UPPER SURFACE OF CUP  UPPER SURFACE OF BOB        UPPER SURFACE OF BOB MUST BE        NO HIGHER THAN UPPER SURFACE
35.  are desired  Unlike Parallel Plates  the gap for a Cone and Plate is fixed  and  defined by the cone geometry  Because of this  the Cone and Plate is normally used for isothermal testing only   as temperature changes would lead to changes in gap due to thermal expansion  For special cases  however   Invar tools can be used  Because of Invar s exceptional thermal expansion properties  Invar tools are used for  applications where temperature ramps or sweeps are necessary     ARES User Manual    Operating Ranges    Operating range is defined as the region bounded by the maximum and minimum complex modulus G that  can be measured by each transducer type using the cone and plate geometry  The following geometry specific  factors affect the operating range of cone and plate geometry     e Plate diameter  strain constant K  and stress constant Kz     e Cone angle  strain constant K       Additionally  the following instrument specific factors affect the operating range of all geometries     e Minimum torque that can be measured by the transducer  e Transducer compliance  e Maximum strain that can be generated by the motor    To calculate the minimum and maximum G that can be measured by each transducer type using the cone and  plate geometry  use the following equation     E C  41     Y     where K    Stress Constant  K    Strain Constant    and C is computed from the following     C for G  MAXIMUM C for G  MINIMUM    2K FRTN1    2K FRTN1E l J 1 15e 06  rad gecm     a  see note bel
36.  bottom of the Test Station     During installation  the Test Station will be removed from the pallet and placed on the workbench  Retain the  pallet for future use in case the Test Station requires moving  To place the Test Station back on the pallet  use a  forklift to position the pallet flush with the workbench top and  slide  the Test Station onto the pallet     WARNING  Do not attempt to lift or carry the Test Station by hand   Use a fork lift that is rated to carry the weight of the Test Station   Attempting to lift or carry the Test Station by hand can result in serious  personal injury or damage to the Test Station        ARES User Manual 229    Special Maintenance    Service and Repair of the Test Station    Other than the routine maintenance listed in the previous section  there are no other repairs or service that  you  as the customer  can perform  Contact TA Instruments regarding service or repairs  as well as the  availability of service contracts and plans     Diagnostic LEDs    Each removable circuit board on the Test Station Mother Board is equipped with a bank of diagnostic LEDs   light emitting diodes  that indicate the operational status of key electrical signals  Figure 6 1 shows the  location of the LEDs  which can be viewed by opening the access door  The electronics are protected by a metal  shield that allows inspection of the LEDs during basic troubleshooting     The electrical signal monitored by each LED is labeled  Except as noted in Figure 6 1  
37.  calibration menu screens  or  procedures  we provide the following option in Orchestrator for adjusting the ARES for temperature error  effects  Please note that this option is only available for ARES firmware version 5 xx and above     Obtain a table of calibration temperatures relative to PRT temperatures using either a reference thermometer  or known samples transition points as described above  If using material transition points  more than one  material should be used for greater precision across a broad range of temperatures     Under the Utilities pull down menu select the Instrument Configuration function from the Service function  table  Select TEMPERATURE CONTROL  Figure 5 11   Near the bottom of the form  click  Adjustable  for the  Temperature Calibration Table  A table will be displayed with up to 20 windows to input the calibration  values  Enter the measured and theoretical values for each calibration points obtained  leaving the remaining  windows unaltered  The ARES will now adjust temperatures measured by the PRT according to the calibration  table  linearly interpolating between table values     It is recommended that you enter the same commanded and calibrated values for room temperature so that  there is no offset at room temperature     ARES User Manual    Setup Instrument Options     Temperature Control  e  Oven  Air Chiller or LW2 Dewar            500 0   150 0    Oven Air Temperature Ea    0 0 100 0 150 0 200 0 3595 0  2 0 101 5 151 0   00 5 3999 0  
38.  current required to drive back to  null position is proportional to the amount of force applied  This current is converted to DC voltage  and  scaled to become the force  torque  output of the transducer     All Force Rebalance Transducers are dual range  Selection of range is performed using Orchestrator software   Firmware versions 5 00 00 and higher automatically switch ranges during a test  FRT transducers are also  available with normal force measurement capability  For the normal force option  the measurement range in  tension  downward  is 60  that of compression  upward  due to the weight of the transducer shaft being  supported by the normal force servo     Standard Transducer  STD     The Standard transducer provides high frequency response with the ruggedness required by QC labs  They  are generally used for solids and melts testing since they typically lack the low end sensitivity needed for  fluids measurements  The Standard transducer utilizes a shaft that is supported by a torsion bar  Mounted to  this is a moment arm  A position sensor on each end of the moment arm produces rotational position  information  In response to rotation in a given direction  the output of one sensor increases while the other  decreases  A torque signal is then derived by taking the differential between these two outputs  The torsion  bar and moment arm are axially supported by a flat spring  allowing axial compliance  A third position  sensor  mounted to the top of the spring assem
39.  data gathered over many cycles     Strain Amplitude Control    Measurement does not begin until the strain that you command during a test is within a certain percentage of  the strain actually applied to the sample  This feature allows you to specify this percentage     Default    Does not use strain amplitude control  The actual sample strain is simply recorded   Adjustable    Measurement begins when sample strain is within the entered percentage of initial commanded strain  This  will increase the test time as the motor movement increases to reach the desired sample strain     ARES User Manual    ARES User Manual    Dynamic Frequency Sweep Test  Strain Control        Figure 3 29  Measurement Options Set up Screen     ARES User Manual       Chapter 4    Test Geometries and Formulas    Introduction    This chapter provides information on the various test geometries and formulas that are used with the ARES  instrument  The following material is covered     e General Test Tool Information  e General Test Tool Installation  e Zeroing and Setting the Gap   e Specific Test Tool Geometries  e Test Formulas    General Test Tool Information    General Recommendations for Geometry Selection    Although the physical properties of the sample generally dictate appropriate sample geometry  it is sometimes  possible to test a given sample using more than one geometry  Ideally  the test results should be identical in  the different geometries  However  there exist experimental limitations 
40.  dynamic zero position  Figure 5 9      5  Select the parallel plate geometry as the current geometry  and enter the following tool dimensions     6  Setup a Dynamic Time Sweep with the following conditions        Current ambient temperature    7  Run the Dynamic Time Sweep        8  Ensure that the following conditions exist     a  The STRAIN value reported is between 24 90 and 25 10   b  The pointer deflects   0 25 radians  as shown in Figure 5 9     If either condition is not met  contact TA Instruments Service group for further assistance   9  Stop the Dynamic Time Sweep     This concludes the Strain Calibration Check     ARES User Manual    POINTER          DYNAMIC  wo  POSITION LS           0 25 RADIANS       Figure 5 9  Strain Calibration Check with Pointer    ARES User Manual    System Check Using PDMS    Included in the calibration kit  supplied with the instrument  is a jar of PDMS  PDMS  polydimethy  siloxane   is a rheological reference material that is used to verify the correct operation of the test station  A PDMS test  should be run periodically to ensure proper operation of the instrument  It should also be run as a preliminary  diagnostic any time there is a question regarding instrument performance     Procedure  1  Turn on the motor and ensure that it is in dynamic mode     2  Ensure that the transducer range is in high range  use the Orchestrator function  Utilities   Service Transducer      3  Install the 25 mm parallel plate test tool  see Chapter 4    
41.  geometry     The following geometry specific factors affect the operating range of torsion rectangular geometry   e Thickness of sample  stress constant K  and strain constant K     e Width of sample  stress constant K  and strain constant K   e Length of sample  strain constant K     Additionally  the following instrument specific factors affect the operating range of all geometries   e Minimum torque that can be measured by the transducer    e Transducer compliance  e Maximum strain that can be generated by the motor    178 ARES User Manual    To calculate the minimum and maximum G that can be measured by each transducer type using the torsion  rectangular geometry  use the following equation         Ke lc  4 1   r    Y     where  K     Stress Constant    K y   Strain Constant    and C is computed from the following        Transducer C for G  MAXIMUM C for G  MINIMUM  2K FRTN1 1 Cz a min  2K FRTNIE C 0 1 F J 1 15e 06  rad gecm  a  see note below     This transducer is not generally recommended  1K FRTN1 for use with this tool  However  it may be used  for some limited applications     2K STD  10K STD   0  1  J   2 60e 06 C              see note below     100 FRT    ace a These transducers are not recommended    200 FRTN1 for use with the torsion rectangular tool     NOTE  The values for M  gecm  and 6  rad  are found in the specification tables in Chapter 1     Table 1 6 through Table 1 13   Pick the correct values for your specific transducer and  motor combination       
42.  gram centimeter of applied torque   Stiffness is the reciprocal of compliance  Transducer compliance is the transducer shaft displacement resulting  from torque applied to the transducer  Shaft displacement is measured by a position sensor on the transducer  shaft  Sample compliance is sample displacement resulting from force applied to the sample     Because the transducer is not an infinitely stiff device  both the transducer and sample exhibit compliance   Since the transducer is being deformed along with the sample  some of the strain that is commanded deforms  the sample  and some of the strain deforms the transducer  This leads to errors in sample moduli  which  becomes larger as sample stiffness increases     ARES utilizes an on line hardware correction scheme to adjust for transducer compliance  The system  determines sample deformation  strain  by taking the difference between the  measured  motor and  transducer displacement signals  The measurement is sensitive to limits in strain resolution  as well as  variations in motor  and transducer  calibration and linearity  Under  ideal  conditions  the sample  deformation is relatively large  and as such  the transducer displacement is much smaller than the motor  displacement applying the torque  The difference between the two deformations  which is used to obtain  sample displacement  is therefore a large number  and the relative error associated with the measurement is  small  However  this error becomes significant 
43.  is not at the limit of its travel  if it is  adjust  accordingly   Ensure that the AutoTension Window is set  to a reasonable value  if not  re set     Call Technical Service        ARES User Manual 235    Table 6 2  Instrument Operation Troubleshooting Guide  Continued     PROBLEM CORRECTIVE ACTIONS    RUNNING A TEST AND COLLECTING DATA    Erratic data points or missing data  For FRT Transducers Only  Ensure that the  correct air pressure is supplied to the  transducer in use  see instrument  specifications   if pressure is incorrect  go to    2  Ensure that AutoTension is functioning  correctly  if not  go to 8     3  Ensure that the oven is not in contact with  the upper and lower test tools  if so  go to 8     4  Ensure that the sample is loaded correctly     5  Cycle the instrument main power  turn off   then on      6  Ensure on line indicator shows ON TEMP     7  Run confidence check with steel shim   8  Perform a torque calibration    9  Perform a strain check    10  Check force level for overload condition   11  Check that pretension is correct     12  Make sure the sample stiffness dimensions  are reasonable for the test tool        13  Call Technical Service     ARES User Manual    TA Instruments Offices    For information on our latest products and more  see our web site at  www tainst com     TA Instruments  Inc    109 Lukens Drive   New Castle  DE 19720   Telephone  1 302 427 4000 or  1 302 427 4040   Fax  1 302 427 4001    HELPLINE   U S A    For assistance  p
44.  one of the following depending on the transducer type   e 2K FRIN1 or 2K FRINIE  delivered after January 2002  see note below   Push the pin toward  the right of the instrument until the left clamp contacts the transducer housing   e 100 FRTN1 200 FRTN1 or 1K FRTN1  Push the pin toward the front of the instrument until the  rear clamp contacts the transducer housing     To Unlock the FRTN1 bearing     1  Read the Caution on page 11   2  Ensure that instrument power is off and air is applied to the transducer   3  Do one of the following  depending on the transducer type   e 2K FRTNI1 or 2K FRTNIE  delivered after January 2002  see note below   Push the pin toward  the left of the instrument until the right clamp contacts the transducer housing   e 100 FRTN1  200 FRTN1 or 1K FRTN1  Push the pin toward the rear of the instrument until the  front clamp contacts the transducer housing     NOTE  2KFRTN1 Bearing Lock Orientation    During design revisions to the 2KFRIN1 transducer  the bearing lock direction was changed  Transducers  delivered before January 2002 have the bearing lock arranged so that the locked position is with pin pushed to  the left and the  right clamp  touching the transducer housing  Transducers delivered after January 2002   approximately  are just the opposite  the locked position is with the pin pushed to the right and the  left  clamp  touching the housing  It is very important to know which locking arrangement your transducer has     In general  instrum
45.  proprietary technology patented by TA Instruments   Waters LLC  U S   Patent No  5 710 426      Thermogravimetric Apparatus describes proprietary technology patented by TA Instruments   Waters LLC   U S  Patent No  5 321 719      Power Compensation Differential Scanning Calorimeter  Tzero  describes proprietary technology patented by  TA Instruments   Waters LLC  U S  Patent No  6 428 203      Differential Scanning Calorimeter  Tzero  describes proprietary technology patented by TA Instruments     Waters LLC  U S  Patent No  6 488 406      Apparatus and Method for Measuring Viscoelastic Properties of Materials describes proprietary technology    patented by Rheometric Scientific  Inc   acquired by TA Instruments   Waters LLC  January 2003   U S  Patent  No  4 601 195      Other Trademarks    Windows   NT  2000  XP  98  985E  Me  Microsoft Excel   and Microsoft Word 97   are registered trademarks  of the Microsoft Corporation     Adobe   Acrobat   Reader   are registered trademarks of Adobe Systems Incorporated    Oracle   and Oracle9i    are trademarks or registered trademarks of Oracle Corporation   TrueMetrix    and Scanning Tip Technology    are registered trademarks of ThermoMicroscopes  Inc   CHROMEL   and ALUMEL   are registered trademarks of Hoskins Manufacturing Company   Teflon   is a registered trademark of E  I  du Pont de Nemours and Company     Loctite   is a registered trademark of the Loctite Corporation      continued on next page     6 ARES User Manual    Ot
46.  range   use the Orchestrator function Utilities   Service Transducer      3  Install the Torsion Rectangular tool  see Chapter 4      4  Load the calibration steel sample  part number 400 02589  that is supplied with the tool  see Chapter 4  for details      5  Conduct a Dynamic Frequency Sweep using the following parameters     FREQUENCY 0 1 to 100 rad sec log sweep  STRAIN 0 02         TEMPERATURE Current ambient temperature    Ensure that the phase angle  6  is   0 25   throughout the frequency range  If the values obtained from this test  are outside this range  please contact Technical Services for further assistance     ARES User Manual    Procedure for 100 FRTN1  200 FRTN1  100 FRT  200 FRT and 1K FRTN1    1     2     Turn on the motor and ensure that it is in dynamic mode     In Orchestrator enter the Edit Start Instrument Test function  Select Parallel Plate Geometry and then  press the  Edit Geometry  button  Select the  Options  button and enter a fluid density of 1 g cm    in  the displayed form     Ensure that the transducer range is in high range  use the Orchestrator function  Utilities  Service Transducer   When checking the 1K FRTN1 only  set the transducer to the low  20  gecm  range     Install the 50 mm parallel plate test tool see Chapter 4    Load the 1000 cP Newtonian Calibration fluid  part number 700 01016  with a gap of 1 millimeter     Conduct a Dynamic Frequency Sweep using the following parameters     p     FREQUENCY 0 1 to 100 rad sec log sweep 
47.  remain constant for the life of the system  The phase  angle correction is computed   and entered into Orchestrator  at the factory before the instrument is shipped   and should not have to be adjusted under normal operating conditions     It is recommended that the phase angle be checked periodically using the following procedures to ensure that  the entire system is functioning properly  If the values obtained from this test are abnormal  please contact  Technical Services for further assistance     To check the phase angle using the 2K FRTN1  2K FRTNI1E  2K STD  or 10K STD Transducer  a Dynamic  Frequency Sweep is run on a steel sample that is loaded into a torsion rectangular tool  Recall that a purely  elastic sample has a phase angle of zero degrees  The phase angle of steel is near zero     If using any of the 100 FRTN1  200 FRTN1  100 FRT or 200 FRT transducers  a Dynamic Frequency Sweep is  run on a 1000 cP Newtonian fluid  calibration fluid  that is loaded onto a parallel plate tool  Again  recall that a  purely viscous sample has a phase angle of 90 degrees  The phase angle of the Newtonian fluid is near 90    degrees     To check the phase angle for the 1K FRTN1 transducer either the steel sample or oil sample may be used     Procedure for 1K FRTN1  2K FRIN1  2K FRTN1E  2K STD  and 10K STD  Transducers    1  Turn on the motor and ensure that it is in dynamic mode     2  Ifthe transducer is a 2K FRTN1  2K FRTN1E  or 1K FRTN 1  ensure that the transducer is in high
48.  shown in Figure 2 38     Setup Instrument Options Ei Ed         Figure 2 38  Setup Instrument Options Form    Used to Input the Fluid Bath System Configuration    NOTES    1  Make sure the following are selected  e Instrument Setup   TEMPERATURE CONTROL    e Temperature Control   BATH  INSTRUMENT CONTROLLED ONLY     2  Ensure that the maximum and minimum temperatures corresponding to  the desired circulator operating range and fluid  See the previous topic  Filling the Circulator  are entered in the Max and Min Temperature fields     3  Select either Temperature Loop Control option  using the following  information as a guideline    e Circulator Temperature  When selected  the temperature of the  fluid in the circulator is maintained at the commanded temperature   The lower tool PRT is independent of the circulator temperature  and  due to its location  will report temperatures that are slightly different  than the commanded  circulator  temperature  The circulator  temperature control should be used when running step or ramp type  temperature studies where a controlled  steady  temperature change   or speed  is more critical than actual sample temperature     90 ARES User Manual    NOTE  continued     e Tool Temperature  When selected  the temperature of the lower  test tool is maintained at the commanded temperature  Tool  Temperature control is the recommended setting for isothermal   and some step  testing  where controlling the actual sample  temperature is most importa
49.  step increases in strain  while holding frequency and  temperature constant  Two types of strain sweeps can be executed  Figure 3 6   as follows     Logarithmic Strain Sweep    The logarithmic strain sweep uses the entered strain values  in dimensionless strain units  as the upper and  lower strain limits of the sweep  An increment or step size  also in strain units  must be entered  Increments  are performed in logarithmic steps  The number of points per decade includes the initial strain  but excludes  the final strain     As an example  consider a sweep conducted over a single decade of strain  between 10  and 100   Selecting  five data points to be measured per decade divides the difference of the endpoint logarithms into five equally   spaced fractional exponents  so that six discrete strains are generated     e Initial Frequency   10   e Decade Frequencies   15 9   25 1   39 8   63 1   e Final Frequency   100     One data point is measured at each of the strains     Up to 500 data points can be measured per each decade of strain  Logarithmic strain sweeps can be run in  ascending or descending order  and can begin or end at any strain within the range of the instrument     Linear Strain Sweep  The linear strain sweep uses the entered strain limits  The test starts at the initial strain and continues until the    final strain is reached  adding the strain increment to each subsequent strain if the final strain is greater than  the initial value  or subtracting the incre
50.  suggest that you partially insert the tool by placing a phenolic  spacer between the upper lip of the lower tool  and the surface of the  bath  Allow the lower tool temperature to match that of the bath  then  remove the spacer and fully insert the lower tool     ARES User Manual 197    Assemble the lower tool  cup  by installing the inner cup into the outer cup using the mounting screw  provided  Figure 4 16   Make sure the o ring is present  clean  and free of defects    Select the Set Gap Instrument Control function under the Control menu in Orchestrator  Use the   Send to Top  button to raise the stage to the loading position    Verify that the motor is on then mount the upper tool on the transducer shaft and lower tool and PRT  into the properly installed fluid bath    Using the Set Gap Instrument Control function in Orchestrator  zero the normal force and torque on  the motor using the  Offset Normal Force to Zero  and  Offset Torque to Zero  buttons    Using the manual stage control buttons  lower the stage until the upper surface of the bob is below the  upper surface of the cup  Figure 4 15A     Place a straight edge across the upper surface of the cup  While monitoring the normal force  raise the  stage slowly until the upper surface of the bob touches the straight edge  There will be a visible  increase in normal force when the bob contacts the straight edge  Figure 4 15B     Zero the gap using the  Zero Indicator  button in the Set Gap Instrument Control function
51.  test station input  or air dryer  and purge  the lines  before re powering your instrument     To quickly determine if there is an interruption in the air supply  we have found that it is helpful to install a  pressure gauge before the air dryer     Pneumatic Connections    Figure 2 15 shows the location where all the pneumatic connections are made  Connections are made using  standard fittings  The connections to the air supply MAIN and N2 GAS ports use quick disconnect fittings   Table 2 3 identifies and describes Pneumatic connections  on Pneumatics Panel  necessary for test station  operation     ARES User Manual    Connector    AIR SUPPLY   MAIN    AIR SUPPLY      N2 GAS    GUN HEATERS    OVEN PRESSURE  SENSOR    ARES User Manual    Table 2 3  Pneumatic Connections    Connected to the  external   air dryer     Connected to an external  nitrogen  or other  gas    supply     Connected to the oven     Connected to the oven     Connected to the oven     Pneumatic input to the Test Station  Used to  supply air to the oven and throughout the Test  Station  Supply pressure should be 80 psi     Pneumatic input to the Test Station  Supplies  nitrogen  or other  gas to the oven  thus  allowing Nitrogen gas to be used as the  heating medium  Supply pressure should be  60 to 70 psi     Pneumatic output from the Test Station to the  oven  Supplies the oven with air or N  gas    depending upon position of Gas Supply to  Oven selector switch  black knob below  pressure gauges      Pne
52.  the sine of the phase angle gives the out of phase component of the  stress  G        which is proportional to the amount of energy lost to viscous dissipation     9  ARES User Manual    Data Correlation    All dynamic mechanical data correlation is performed on between 1 and 64 cycles of oscillation  depending  upon the testing frequency  as shown in Table 3 3     Table 3 3  Single Point Measurement Data Correlation    Frequency  rad sec  Cycles of Correlation       The control computer samples 2 048 data points regardless of the test frequency  The number of cycles of  correlation determines the number of data points sampled per cycle  For example  at frequencies less than 2  rad sec  the computer samples all 2 048 data points during one cycle of strain and force  At 500 rad sec  only  32 data points are sampled  At higher frequencies fewer points per cycle are used  however the data is  collected over multiple cycles  so the actual number of data samples still totals 2048     At least one complete cycle is required for correlation  resulting in measurement time being inversely  proportional to frequency  This is an important consideration when observing materials that change rapidly  over time  such as cures involving gels and thermosets where sample material properties may differ  dramatically between test start and completion     ARES User Manual    4    SE         REFERENCE SINE WAVE  X     REFERENCE  90  Y          B FORCE  S ANN    SZ   STRAIN    A  Dynamic Measureme
53.  threaded collar     Three screws fasten the rotating Shaft of the Fluid Bath  Figure 2 21  to the Test Station Motor Anvil  A  threaded collar secures the Fluid Bath Body to the Test Station Motor Housing     Two hoses supply fluid between the Fluid Bath and the fluid source  which is typically a computer controlled  circulator     Prior to mounting the Fluid Bath  perform the following actions on the ARES instrument     e Raise the Stage to maximum height    e Remove all Upper and Lower Test Tools  and loosen the Anvil Tightening Knob on the Motor Anvil    e Thoroughly inspect the Test Tool mounting surfaces  i e   the transducer anvil and the motor anvil and  clean off any material that may interfere with the mounting of the Fluid Bath  This is essential to ensure  proper mechanical mating between the bath and the instrument    e Turn off the Motor  using the Instrument Control Panel     Figure 2 22 while performing the following steps to install the bath   1  Remove the protective plastic base from the Fluid Bath Threaded Collar by placing the two pins on the  Threaded Collar Spanner Wrench  provided  into two of the holes machined into the Collar  and  rotating the wrench counterclockwise     2  Gain access to the Shaft by sliding the cover of the Fluid Bath fully upward     3  Hold the Fluid Bath above the ARES Motor Anvil with the Bath Hoses facing toward the right of the  instrument     ARES User Manual    4  Rotate the Fluid Bath Shaft to align the flat portion of th
54.  transducer     x  WARNING  This is a high torque motor  Turning on the motor while in       Never turn on the motor while a sample is loaded     Keep hands clear of the motor     ARES User Manual    Motor On Off Control    Power to the Motor is controlled using the Instrument Control Panel function available in Orchestrator  software  To turn the Motor on or off  open the Instrument Control Panel by either clicking the  Control  Panel  button  Figure 2 3   which appears on the Tool Bar  or from the Control pull down menu  Choose either  MOTOR POWER ON or OFF  then select the  Ok  button       l Control Panel Button    Instrument Control Panel ki E4    Environmental Control Settings    Temperature                    i 00 0 PC  Max 500 0  C Min  1 50  0      Temperature Control f Oyen  Air  Chiller or LN2 Dewar        Environmental Controller 1 Off   On       Motor Control Settings       Ok Help   Cancel         Figure 2 3  Instrument Control Panel    NOTE  If the Environmental Controller option is set  on   the Control Panel will  appear as in Figure 2 20     Motor Oven Stop    The Motor Stop and Oven Stop Buttons on the front panel of the Test Station will quickly and unconditionally  power down their associated component  To turn the motor or oven back on  use the normal software    controls     ARES User Manual    Transducer    The transducer  Figure 2 2  measures force generated by the sample during deformation by the motor  The  sample is mounted between the motor and tr
55. 13 01222   4980 to 5020   9975 to 11 025        ARES User Manual    Normal Force Calibration  for All Transducers     Normal Force Calibration ensures that the transducer is properly measuring normal force  The calibration  involves hanging a precision weight on the calibration tool  which is mounted on the transducer during  calibration  The applied normal force is the amount of weight applied to the calibration tool  For example   hanging a 1000 gram weight applies a normal force of 1000 gmf     Normal Force Calibration Procedure    1  If calibrating an FRT  ensure that the transducer is set to High Range before proceeding  To set the  transducer to High Range do the following   a  Access the Set Transducer Characteristics form  Figure 5 7  by selecting the Transducer option  from the Service function of the Utilities pull down menu   b  Using the  Transducer Selected  menu  select the high range transducer then click  Ok    2  Turn on the motor   3  Raise the stage to maximum height and remove any test tools   4  Install the calibration tool as shown in Figure 5 2  the pulley need not be installed      5  Using the Set Gap Instrument Control function in Orchestrator  zero the normal force on the motor  using the  Offset Normal Force to Zero  buttons     6  Access the Transducer Characteristics form using the Orchestrator function Calibrate Instrument  under the Utilities pull down menu     7  Select the  XducerCal  button  Establish a zero torque reference value by selecti
56. 170  degree C  activates the absorbent qualities of the    getters        NOTE  Operate the LN2 controller for 15 minutes at least once every 60 days  to maintain the insulating properties of the LN2 transfer line     Oven Gas Selection    Either gaseous nitrogen  N   or air can be flowed through the heaters during a test  To switch between N  and  air  use the Gas Supply to Oven Selector Switch  located below the pressure gauges on the Pneumatics side  panel  Figure 2 15   Rotate the valve to the desired position  positions are labeled      When using the optional LN2 Controller you can select to flow gases through the heaters in either of the  following two ways through software control     1  Use evaporated liquid nitrogen  LN   at all times   2  Use liquid nitrogen until a specified temperature is reached  then use either nitrogen gas or air  thereafter  depending upon which gas input port is selected with the Gas Supply to Oven Selector  Switch    3  When using the gas switching option  the input temperature for switching from LN  to Gas should be  approximately 10 degrees higher than the input to switch from Gas back to LN   This will prevent the  system from switching back and forth between the two sources at the switch temperature since  when  switching from liquid nitrogen to gas  there is a momentary change in temperature as the switch takes  place     Enable the Gas switching option  and set the switch point temperatures  using the Set Test Conditions function  in 
57. 34 93 600 93 00  Fax  34 93 325 98 96    SWEDEN NORWAY   Waters Sverige AB   TA Instruments Division   PO Box 485 Turebergsvagen 3  SE 191 24 Sollentuna   Sweden   Phone  46 8 59 46 92 00   Fax  46 8 59 46 92 09    AUSTRALIA   TA Instruments   C O Waters Australia Pty  Ltd   Unit 3  38 46 South Street  Rydalmere NSW 2116  Australia   Phone  613 9553 0813   Fax  61 3 9553 0813    ARES User Manual          A    AC Power  main power switch  51  specifications  28   Air  See Gas  input options   Air Dryer  33  54  55  228   Air Quality  54   Analog Data Input  125  144   Arbitrary Waveshape Test  127    AutoStrain  145    AutoTension  142    B    Bearing Locks  FRT with normal force  42  FRT without normal force  41  low shear  LS  motor  45  standard motors  45  Standard Transducer  41    Boltzmann Superposition Principle  112    C    Calibration  normal force  215  phase angle check  218  strain check  220  temperature  224  torque  210    Calibration Intervals  208    calibration procedures  207    ARES User Manual    Index    Chiller  description  59  diagram  60  electrical specifications  29  physical specifications  28    circulator  27  30  48  67  70  77  88  Cleaning the Instrument  228  COM Port  48   Complex Modulus  E     94    Compliance  definition  150  determination of operational range  151    Cone and Plate  test tool   constants  173  tool installation  175  general information  173  operating range  174  sample loading  176    Constant Stress Test  137  Corre
58. 4  Load the PDMS  part number 700 01011  using a gap of 2 millimeter     5  Conduct a Dynamic Frequency Sweep using the following parameters     5   for 100 or 200 FRT FRTN1  use 1      STRAIN    SWEEP MODE Log    INITIAL FREQUENCY 0 1 rad s    FINAL FREQUENCY 100 rad s    6  Set up the plot to show G   G   viscosity  and phase angle  The displayed data should look similar to  that shown in Figure 5 10        7  At the completion of the test  select the GG    Crossover function from the Analysis pull down menu   The frequency  and value  where G  and G  cross  crossover point  is computed and displayed   Compare the computed frequency and G  G  crossover to the values labeled on the PDMS jar  Please  note that on the PDMS jar  crossover frequency  in rad s  is labeled Wc  and the crossover value is  labeled Gc  no exponent is shown   The computed crossover point should be within the error limits  provided on the jar  Press the  Stamp  button to place the computed crossover data on the plot     This concludes the PDMS test     ARES User Manual    ARES User Manual    Dynamic Frequency Sweep of PDMS    G   G  Crossover Point  4 9045 2 438x10    Error 1 54  for Freq  6 94  for G       Freq  rad s     Figure 5 10  Typical Results of PDMS Tests    90 0    80 0    30 0    20 0              ajbuyaseyd    Temperature Calibration    The temperature of the ARES is monitored by a Platinum Resistance Thermometer   PRT  which is  automatically calibrated upon power up  This calibration is 
59. 6 5 mm bob   17mm cup  for fluid bath or fluid bath 2     Enviromental Systems O D  Ambient BATH MOUNT  Oven  Fluid Bath  Fluid Bath 2       see Chapter 2 for more details regarding lower tool Couette Tool       General Information    The Couette tool is used for testing lower viscosity fluids that would not generate enough torque using  parallel plates  Itis also used where containment of the fluid would be difficult using other tools     Operating Ranges    Operating range is defined as the region bounded by the maximum and minimum complex modulus G that    can be measured by each transducer type using the Couette  The following geometry specific factors affect  the operating range of the Couette     e Length of bob  stress constant K      e Radius of bob  strain constant K  and stress constant K      e Radius of cup  strain constant K       ARES User Manual    Additionally  the following instrument specific factors affect the operating range of all geometries     e Minimum torque that can be measured by the transducer  e Transducer compliance  e Maximum strain that can be generated by the motor    To calculate the minimum and maximum G  that can be measured by each transducer type using the Couette   use the following equation       Ks  o  4 1   cx    Y     where K    Stress Constant  K    Strain Constant  and C is computed from the following     C for G  MAXIMUM C for G  MINIMUM    2K FRTN1 1   c Minin  2K FRTN1E   C 0 1 E J  1 15e 06  rad gecm  a     Transducer     see n
60. 76  D Ea T a iria E E A A did 76  C tor OPTIONS ore ira E o A N A T nase E EE E E EO A 77  esqui a Thermal Opera Fe NS Ao E o A 77  Tiss alla ork Pee and cena E AE dues E 78  Caireculator Cone cons ii tios 80  iaa o EE 80  Peltier Configuration in Orchestrator cocine 81  a AAA Po O e ee ee 82  Ester Op  ra CCU CICS pri S 83  PCO ceca sees o E sonia eaennssescuesun E E E 83    Lei e A e o E em enn en ner rn A ee tel er eee te 83  A A A 85  TESS CUM a e A E A T E E E T E 85  ERO I A EA Ea EEA E II EAN TE 85  Lowe Toole aa Oaa E ee 87  S Een a era 88  Mc PP A TT E T T A A A T 88  AAA e o 88  Fluid Bath Configuration in Orchestrator        cscsssssssssssssssssssssssssscsssecssssessssscssssesssseessnecessscsssnecssnecessecesseeseaneessn 90  A o rer re tere nee eee eee eer ee eee 91  Marl Bat OPeratie INCA UU ete in 6S   enero inn pta a i 97  P Datl Operan oia 92  Chapter   Test DescOpuoAS AAA AA 93  Iba Vago e hULeiaLo   q A o Pre o O ter er ee ee ee 93  Sesso Hal  a  a Viscoclas uc sad is ceo 93   DS ciao ho al aa e T o Tee err ee ee ee 93  Hooke sand IN GW LOM e LAWS israel NN 93  PCAC  SIN FOS UUM  T A e CE A 94  Dinamme Meenanical TESTO pato 94  Pe in OTA oi Wi el NS APA OO pcan nasa E EET E S aceecunyeadeenestoe 94  Measurement MeMO paar odo 96  DET Ae ay PU a EE O O or PEI E E E E EE N gents 97    ARES User Manual    Documents Describing Instrument Software Opera On sins 99    TET M  016   hc e e E E EE E E A E E E 100  Stran Controlled Dynamic Test Memo dsp 101  Dynamic Sin
61. ARES    Rheometer          PN 902 30026 Rev D  June 2003       1999 by Rheometric Scientific Inc       2003 by TA Instruments   Waters LLC  109 Lukens Drive   New Castle  DE 19720    Notice    The material contained in this manual  and in the online help for the software used to support this instrument   is believed adequate for the intended use of the instrument  If the instrument or procedures are used for  purposes other than those specified herein  confirmation of their suitability must be obtained from TA  Instruments  Otherwise  TA Instruments does not guarantee any results and assumes no obligation or  liability  TA Instruments also reserves the right to revise this document and to make changes without notice     TA Instruments may have patents  patent applications  trademarks  copyrights  or other intellectual property  covering subject matter in this document  Except as expressly provided in written license agreement from TA  Instrument  the furnishing of this document does not give you any license to these patents  trademarks   copyrights  or other intellectual property     TA Instruments Operating Software  as well as Module  Data Analysis  and Utility Software and their  associated manuals and online help  are proprietary and copyrighted by TA Instruments  Purchasers are  granted a license to use these software programs on the module and controller with which they were  purchased  These programs may not be duplicated by the purchaser without the prior written cons
62. Analogli Df Autost OF MeasOps 04       Ok   Options   End of Test   Save As   Help   Cancel      Figure 3 3  Dynamic Time Sweep Test Set up Screen    Suggested Use    Time sweep provides a means of monitoring the time dependent behavior of a material  for example thermal  degradation at temperature  curing in thermosetting systems  or the build up or breakdown of network  structure     Test Options    The following test options can be selected for use with the time sweep   e Steady PreShear  e Delay Before Test  e AutoTension  e Analog Data Input  e AutoStrain  e Measurements Options  o Delay Settings  o Strain Amplitude Control    102 ARES User Manual    Frequency Sweep    Functional Description    The Frequency Sweep test takes successive measurements at selectable frequencies while holding a constant  strain and temperature  Ideally  the selected strain should be within the linear viscoelastic region of the  sample  The time it takes to complete the test is highly dependent on the low end frequency selected  For  example  a frequency sweep from 0 1 to 100 rad sec  taking 10 data points per decade  will take  approximately 12 minutes  Changing the range to 0 01 to 100 rad sec increases the test time to around 2  hours  and running the test from 0 001 to 100 rad sec may take over 20 hours     Three types of frequency sweeps can be executed  Figure 3 4   as follows   Logarithmic Frequency Sweep    The logarithmic frequency sweep uses the entered frequency values as the upper 
63. C Min  20 0  C    Temperature Control         Bath  Instrument Controlled Only             Motor Control Settings  Motor Power     Off tf Or    Cancel      Figure 2 40  Instrument Control Panel Showing Fluid Bath Control Options       ARES User Manual       Chapter 3    Test Descriptions    Introduction    Stress  Strain  and Viscoelasticity    Definition of Terms    Elasticity is the ability of a material to store deformational energy  and can be viewed as the capacity of a  material to regain its original shape after being deformed  Viscosity is a measure of the ability of a material to  resist flow  and reflects the ability of the material to dissipate deformational energy through flow  Material  will respond to an applied force by exhibiting either elastic or viscous behavior  or more commonly  a  combination of both mechanisms  The combined behavior is termed viscoelasticity     In rheological measurements  the deformational force is expressed as the stress  or force per unit area  The  degree of deformation applied to a material is called the strain  Strain may also be expressed as sample  displacement  after deformation  relative to pre deformation sample dimensions  Sample deformations can be  in the form of either simple shear  where the material is deformed in a plane while confined between two  surfaces   or linear deformations  where the material is either compressed or extended      Hooke s and Newton s Laws    Hooke s law describes the mechanical behavior of an 
64. Commanded stress  which is maintained by varying shear rate  Positive Stress results in clockwise rotation   Negative Stress results in counterclockwise rotation     Time    Duration that shear stress is commanded  unless Max Allowed Strain  see previous item  is reached first   When either Time or Max Allowed Strain is reached  zero rate  and zero stress  is commanded     Constant Stress Test El x        Estimated Viscosity f3 00e 05  Pas   Max Allowed Strain  1 00e 06         Points Per Zane        200    Max 350 Min 20            fea Stress Limits  Pa     Max tbs929 42 Min  6  392942  ht ee Zone Number 1 2 de Ed  ao Ses ee  Palfso0o foo ooo o ae sais  ore Time  sor homes  EM lo  E ee intense    Options  Delay  Off MotorPID On ae E   ee E Hien      a  Options   End of Test   Save As      Figure 3 22  Constant Stress Test Set Up Screen       ARES User Manual 137    Suggested Uses    This test is used to measure the creep response of materials  which provides transient information of the  measured strain as a function of the commanded stress     Options  The following test options can be selected for use with the Constant Stress test   e Delay Before Test    e Steady PreShear  e Motor PID    ARES User Manual    Stress Ramp Test    Description    Stress Ramp commands a steady stress level from an initial to a final stress at a selectable  linear rate that is  based upon zone time  Positive Final Stress values result in clockwise rotation of the actuator  stress head    negativ
65. D SPECIFICATION 10K STD  Measurement Range 0 2 to 2000 gecm 1 to 10 000 gecm  Maximum Operating Frequency   500 rad sec  80 Hz     NORMAL FORCE SPECIFICATION 2K STD AND 10K STD    Measurement Range 2 to 1500 gmf 2 1500 gmf       ARES User Manual       Chapter 2    Instrument Components   Identification and Operation    Component Identification and Placement    The base system  Figure 2 1  consists of the Test Station with oven  and Host Computer  Additional optional  components include LN2 Controller and an air filter dryer  Figure 2 1 shows the recommended arrangement     WARNING  If this instrument is used in a manner not intended or  specified in this manual  the protection provided by the instrument may  be impaired     AIR DRYER   Mount to wall behind test station   LN2 CONTROLLER From Your Compressed    Air Supply        From Your LN2 Supply    ARES  TEST STATION    HOST COMPUTER  Figure 2 1  Base System    ARES User Manual            FRONT PANEL  LCD DISPLAY    STAGE  TRANSDUCER    OVEN OPTION    MOTOR   ACTUATOR     MOTOR   OVEN    STOP BUTTONS  MANUAL  STAGE CONTROL       FLUID BATH OPTION    PELTIER OPTION    Figure 2 2  Test Station  Front View  Including Environmental Control System Options     ARES User Manual    Host Computer    The Host Computer allows the human operator to control the test station  and to monitor  display  and analyze  data during the test  Control is achieved through Orchestrator software  While some basic references to the  software will be 
66. Duration     2600  h m        Figure 3 21  Steady Step Rate Temperature Ramp Set Up Screen    ARES User Manual 135    Suggested Uses  Steady Step rate Temperature Ramp can be used to examine the following     e Process simulation  e Transient material response to changing shear rates and temperatures     Options  The following test option is available when using Steady Step Rate Temperature Ramp     e Delay Before Test    ARES User Manual    Stress Controlled Transient Test Methods    Constant Stress Test    Functional Description    Constant Stress is a transient test that applies a stress  in a selected direction for a specified time period  in up  to two zones  In each zone  rate is varied to maintain stress until either a strain or time limit is reached  at  which time zero stress is commanded     When setting up the Constant Stress test  the desired temperature to conduct the test and the sample estimated  viscosity  calculated by dividing the shear stress by the shear rate  are first entered  Figure 3 22   Then the  Maximum Allowed Strain  which is the maximum displacement of the actuator throughout the test  is entered   When either the maximum allowed strain or time  see next item  is reached  zero rate  and zero stress  is  commanded  The estimated viscosity value is used to modify the gain term in the closed loop control  algorithm  which adjusts the motor s rotational rate to generate the desired stress level     For each zone the following are entered   Stress    
67. ES User Manual    Delay Before Test    Manual Delay  A manually entered  Delay Before Test  is the time period between the start of the test and the first    measurement  1 through 65 000 seconds   The time period selected in  Delay Before Test   Figure 3 25  allows  the instrument to equilibrate  or the sample to relax  prior to imposing the deformation     Apply AutoTension  The    Apply AutoTension at the End of Delay    check box determines the point at which AutoTension is    applied  If left unchecked  AutoTension is applied at the start of the Delay period  Checking the box causes it  to not be applied until the delay period has ended     Automatically Start Test When On Temperature    When checked  the start of the test is delayed until the Environmental Control System has stabilized at the  Commanded Temperature        Figure 3 25  Delay Before Test Options Set Up Screen    ARES User Manual 141    Auto Tension Adjustment    AutoTension maintains a specified axial static force on the sample  AutoTension can apply tensile force to  keep the sample taut  prevent sample buckling   or compressive force to maintain a compressive load and  prevent loss of test tool contact  During dynamic testing using the Torsion Rectangular test tool  AutoTension  can be used to compensate for the change in sample length that occurs as a result of thermal expansion  Using  AutoTension with tools such as Parallel Plate can prevent loss of contact between sample and tool     Following is a 
68. Gap in millimeters   b  Max Allowed Force   Enter the maximum Normal Force  in grams force  that will be  generated while the sample is being compressed during the gap setting operation  When  Normal Force exceeds this value  the stage stops descending until Normal Force drops below  this value  Stage movement then resumes    3  Click Set Gap  An information form appears  indicating the time elapsed since the button was clicked   The Stage should descend relatively quickly  typically 5 mm sec  until the Upper Tool is 3 mm from  the Commanded Gap Position  at which time it will slow its rate of descent until the Gap is achieved   The gap is set when the information form is no longer displayed     Comments Concerning the Gap Control Panel    Enabling the Gap Control Panel    The Gap Control Panel function can only operate if the Stepper Motor and Remote Gap Monitoring options  are enabled  These options have been enabled at our manufacturing facility  and should require no  modification  However  for your reference  these options are set by using the Instrument Configuration  function located under the Service function of the Utilities pull down menu  Set the  Stepper or Linear Motor   and  Remote Gap Monitoring  options to yes  as shown in Figure 4 3     ARES User Manual    Read Test Tool Gap Checkbox    The  Read Test Fixture Gap  checkbox is located in the form used to edit a specific geometry  accessed by  clicking the  Edit Geometry  button in the Edit Start Test dialog bo
69. Member States relating to  electrical equipment designed for use within certain voltage limits      EN 61010 1  1993 Safety requirements for electrical equipment for measurement  control  and laboratory use   Part 1  General Requirements   Amendments     EN 61010 2 010  1994 Particular requirements for laboratory equipment for the heating of materials    Amendments     Electromagnetic Compatibility Standards    For the European Economic Area   In accordance with Council Directive 89 336 EEC of 3 May 1989  amended  by Council Directive 93 68 EEC  on the approximation of the laws of the Member States relating to  electromagnetic compatibility      Immunity  EN 50082 1 1997 Electromagnetic compatibility   Generic immunity standard  Part 1  Residential   commercial and light industry     Emissions  EN 55011 1998  Class A    Technical Support    For technical support concerning TA Instruments Rheometric Series  please call     1 302 427 4070  8 a m  to 5 p m  USA ET  Monday through Friday    For technical support concerning Microsoft Windows  please call Microsoft Corp  at 1 206 635 7000  USA  6    a m  to 6 p m  USA PT  Monday through Friday  In other countries  call the relevant Microsoft international  subsidiary office     ARES User Manual    Table of Contents    TA Instruments End User License Ac eC in iit erpressen a E a E E AS 3  Traiasmarksiand Pati A E Ie reer 5  TA Instruments Trademarks ai 5   SUE EEE E AAA a A O E TEE 5  Otier Trademarks ern rey ee TOT Re Tm ais apta T
70. N2 Controller  The main power cord  220V IN  must be  disconnected from the Power Panel to completely remove AC power  from the system     ARES User Manual    WARNING  HIGH VOLTAGE is used in the operation of this  instrument  DEATH ON CONTACT may result if operating personnel fail  to observe safety precautions  Learn the areas of high voltage  connections  and exercise care not to contact these areas when  performing instrument calibration  Prior to working inside the  instrument  remove all jewelry  turn off the power  and ground points of  high voltage before touching them  Make adjustments using an  insulated electronic adjustment tool  Do not make physical contact with  any component inside the instrument while power is applied to the  instrument        Instrument Labels    Label found on the Rear of Power Chassis    WARNING   FOR CONTINUED PROTECTION  AGAINST FIRE HAZARD     REPLACE ONLY WITH SAME  TYPE AND RATING OF FUSE        Label found on the Front of Transducer  FRT only   INSTALL REMOVE LOCK  WITH AIR PRESSURE ON    Label found on the Front of Motor Mount    WARNING   KEEP HANDS AND LOOSE OBJECTS  AWAY FROM MOTOR DURING  TURN ON AND OPERATION        Label found on the Front of Oven    DANGER    HOT TOOLS AND SURFACES       ARES User Manual      Regulatory Compliance  Safety Standards    For the European Economic Area   In accordance with Council Directive 73 23 EEC of 19 February 1973   amended by Council Directive 93 68 EEC  on the harmonization of the laws of 
71. O Ao A  115  TEO EE O 115  Strain Controlled Steady Test MethodS     eeseeeeeseseesesesersrsesesesereeseseserersesereseseeseseseseseressesesesessesesesesseseseseseeseseseseee 116  oead iomele Pola E E E 116  EUncuona DES   0  O en E A E E E tre 116  Pi o o E E seas yseeennanstosetssacateaaenins 116  Sa a A E E E N T E E sen stsecenae  117  OPONE A o e o o E EA OO AES 117  SPE WE oa 118  Puncuona  DESEO erasmo ita olaa 118  Data Coleco MOS api E EE E A E E EAE E E 118  Dato A E E RO 119  roses de 6 US ata 119  OPUS o 120  Stralm Controlled Transient Test Methods eiiean a S 121  AP Ea 1 AE a o E E E E eco aceenue ae 121    ARES User Manual    Funcional ISSO MOM nata tencia 121    BN OC  SOS O nantes E EA O E E E E E E osa anaeees 122  TESCOP HONG eieae a a A N E rr rere E 122  Stress Relax ton  Iransieni Step Stali  ni pelo AAA EEA 123  Funciona  CS iON  r E E st cantecsansteo wuiaeeeadsaunsanedesnedemarcancsesoer 123  BS 0 IN IE   OP O o E octet ep A E E E T 124  TES TOPS ie 124  PATO Te SiO MOS aaa 125  Funcuona FEDO ET PO iS 125  AREE ONO 11S AEA EEE E EEE E E E O E E E 125  oa AAA PO ee ee 127  ancora DESC D O a E A E T E E E E E O E 127  PCa OO yay ida 128  T D a E A E N E E E E E E E E E S 128  Topo 1G Loop  Rale Ramp a oes a een ee  130   B el alle  asa A A NA tne een rrerTe 130  ES Se Usas 130  OIA PP pease E ASS 131  Torque Normal Relato senti aia oia 132  Funcional Peger pO ienien E 132  Sre e O MN  EU y EPR O E E E E E E E ses errno ees 132  dano erorii a E S E E sea
72. O nee erty E mre rn ny rr 187  rote Laat    EOR AAA e ee iO O N E 187   A A II PUNA E E T EE seu emia  190  Cial dal Ml Sauces earn LO  qee OR PUN  A 190  dera AAA in A o 190  Took llo A e E Terr A ne rere irerrrre terre rrsr errr rere ery re 192  Sample LOIS anti toas 192  Va Vol is 193   TOM ca LO a AP o E A ETE 195  Eire EO AON AAA o e E a aE OENE 195  Opera NS KINE OS srr o ices inane re tinsaue i ntessaied E ous tuietegnatsotghescea nennes mmattentestiess 196  ToolInstallation Original WC Dri 197  sample Loading Orie mal Fluid Dalton oio ae eins 198  dio Melia iaa A o A 198  Tool installation  Fluid Dat Zas TEE 201  Sample Oa End Dalila pops 201  Tool IV ar bees Elda Dat 2 eere EA eactessiersadscoesseaes 202   A o o E PE UU O O ere a ener rer oi 204  Dynamic Measurement Portas pistas 204  steady and Transient Measurement FOU AS sesspreicseneai ici 205   Chapters  all Dra Mrs 207   A O reer N eter 207   A     A 208   Calbraton Intel E E T A AE EE A E 208   Torque Calibration  Or All Trans Quecers Muspnirain ita 210  IXOCS O so 210   Normal Force Calibration  for All Transducers            cccccscccccssssecesessececeesseceeessseeeecseseeeecsseeesessseeeeeneesees 215    ARES User Manual    Normal Force Calibration Procedure            c  ccccccccccceccccccccccccccucccccccausceusceuesccucceuucceusccauseecscesceeseauseeeecens 215    A e o eee eee eee ee eee 218   I TOMY WI CAPA Po ie E A ve secede T A O 218  Procedure for 1K FRTN1  2K FRTN1  2K FRTNIE  2K STD  and 10K STD Tra
73. Ops  Oft       Options   End of Test   Save As   Help   Cancel      Figure 3 10  Frequency Temperature Sweep Test Set up Screen    Suggested Use    Frequency  temperature sweep combines the frequency and temperature step methods to generate a group of  curves which can be shifted using time temperature superposition  TTS  along the frequency axis to extend  the range of frequency characterization beyond that which is experimentally practical at the chosen reference  temperature  From TTS data  a master curve can be generated for the sample     Test Options  The following test options can be selected for use with the frequency temperature sweep     e Steady Preshear  e Delay Before Test  e Analog Data Input  e AutoTension  e Measurements Options  o Delay Settings  o Strain Amplitude Control  ARES User Manual 115    Strain Controlled Steady Test Methods  Steady Single Point    Functional Description    Steady Single Point takes a single measurement while applying a steady shear deformation at a chosen shear  rate  Data are taken at the commanded shear rate and temperature  Figure 3 12   Data can be collected using  either of two modes     Time Based    Time Based data collection takes a single measurement  Following the start of the test  the Delay Before  Measure is the time period between the beginning of motor motion and the beginning of data collection  This  allows time for the material to reach steady state  The Measurement Time is the period during which data are  actual
74. ROLLER uses the COMMAND signal  and the SERVO SPEED and SERVO DISPLACEMENT feedback signals   to drive the motor  SERVO DRIVE   applying deformation to the sample  A feedback displacement signal  SERVO  DISPLACEMENT  is derived from a sensor on the motor shaft  SERVO DISPLACEMENT is conditioned and sent to the  A D input as the strain deformation signal  STRAIN      Transducer    The FRT Transducer is also configured as a position servo  The transducer shaft moves as a result of the  sample deformation that is applied by the motor  A feedback displacement signal is derived from a sensor on  the transducer shaft  The torque applied to the transducer is proportional to the energy required to hold the  transducer shaft at a known position  This energy  TORQUE RESPONSE  is conditioned and sent to the A D input  as the force applied to the sample  FORCE      The Standard transducer  not shown in Figure 1 1  generates a signal proportional to the movement of the  internal torsion bar  without any active feedback control     ARES User Manual    Environmental Control System    This instrument can subject the sample to a number of thermal environments using several different  environmental systems     When using the forced air convection oven  two resistive heater guns  mounted on the left side of the oven  are  used to control the sample test temperature  An optional liquid nitrogen controller allows testing at sub   ambient temperatures  The power to the heaters is directed and moni
75. RTN1   1 92E 02  Low range  200 FRTN1        Appendix Table A1 8  Complex Modulus Limits for Cone and Plate  FRT Transducer     CONE ANGLE G  MAXIMUM  dynes cm     G  MINIMUM   rad  at Frequency   rad sec   dynes cm        PLATE  DIAMETER    9 22E 04  9 22E 05  4 79E 07    1 84E 04  1 84E 05  9 59E 06    3 69E 04  3 69E 05  1 92E 07         100  w  10  w  lt  10       100  w  10  w lt  10       100  w  10  w lt  10    2 31E 03 w   2 31E 04 w     1 20E 06    4 61E 03 w   4 61E 04 w     2 40E 06    9 59E 01  High range  100 FRT   1 92E 00  High range  200 FRT   9 59E 02  Low range  100 FRT   1 92E 01  Low range  200 FRT   1 92E 01  High range  100 FRT   3 84E 01  High range  200 FRT   1 92E 02  Low range  100 FRT   3 84E 02  Low range  200 FRT   3 84E 01  High range  100 FRT   7 68E 01  High range  200 FRT   3 84E 02  Low range  100 FRT   7 68E 02  Low range  200 FRT   2 40E 02  High range  100 FRT   4 80E 02  High range  200 FRT   2 40E 03  Low range  100 FRT   4 80E 03  Low range  200 FRT   4 79E 02  High range  100 FRT   9 58E 02  High range  200 FRT   4 79E 03  Low range  100 FRT   9 58E 03  Low range  200 FRT        ARES User Manual    Appendix Table A1 9  Complex Modulus Limits for    Couette  geometry  2R   34mm  2Rg 32mm  L 34mm    TRANSDUCER TYPE G  Maximum  dynes cm      at Frequency     rad sec   100 FRTN1    3 94E 03 w   3 94E 04 w     2 05E 06    3 94E 03 w   3 94E 04  0    2 05E 06    G  MINIMUM  dynes cm        8 19E 02  High range   8 19E 01  Low range     4 10E
76. Rapid Cooling is  displayed on the form     Oven Pressure Source Selects the gas input to the oven  Selecting Gas supplies  either air or gaseous Nz  Selecting LN2 supplies evaporated  LN   from the LN2 Controller     LN2 Rapid Cooling Turns on off rapid cooling  which evaporates LN2 at the  maximum possible rate  Rapid Cooling is always on at  temperatures lower than  124  C        Oven System Status Indicators    Two special purpose neon lamps  located on the front of the oven cover  inform you of the status of the two  gun heaters  The top lamp shows the output level of the top heater  When the lamps are at their brightest  the  heater guns are full ON     Orchestrator Online Help provides instructions for choosing Online  instrument status  Indicators to be  displayed along the bottom of the screen  The following indicators can be displayed  which show the status of  the oven environmental system     Environment On Off Oven Open   Oven Open The oven is not closed  This condition disables the oven environmental system   Env On The oven environmental system is currently enabled    Env Off The oven environmental system is currently disabled     Gas LN2 Status  Oven Only   Gas    Gas Either air or gaseous N2 is currently input to the oven     LN2 Fill The LN2 Controller Dewar flask is currently filling with liquid nitrogen  and the LN    level is less than 50  of the flask capacity  Temperature control will not be active  while the Dewar flask is filling     LN2 Ready The LN2
77. Reps  Miri          k Help   Cancel      Figure 2 39  PID setup form showing Fluid Bath settings     The values listed in Table 2 11 should be used as guidelines  and will work for most applications  However   you may need to experiment somewhat to determine their baths best PID settings for their system  or specific  applications  The Orchestrator Online Help has a complete description of how to determine and tune the PID  coefficients     ARES User Manual    Table 2 11  PID Values for Various Circulators and Fluids    Circulator Julabo   Julabo   NesLab V5 NesLab V4  or  FS 18 FS 18 firmware below  firmware    50  Water  Bath Fluid 100  Water 50  Ethylene Glycol 100   Water 100  Water       Fluid Bath Operating Requirements    The Fluid Bath can operate only if the following conditions are met     e The Fluid Bath is selected as the current environmental system    e The circulator must be filled  on  and circulating fluid through the bath    e The circulator must be connected to the test station via the correct RS 232 cable   e The clamp is set properly     Fluid Bath Operation    The Fluid Bath is operated using the Instrument Control Panel  Figure 2 40   The desired temperature is set in  the Temperature input field  When testing at higher than ambient temperature  the circulator set point must  always be set to a temperature greater than the desired bath temperature     Instrument Control Panel Ei E    Environmental Control Settings    Temperature     5 0 CC  Max 150 0  
78. S     Oven Operation    3  RAA Oven   Oven PRTs Lower Oven Oven gas temperature is maintained as in Mode 1   Air Temp  both upper PRT Temperature is reported using the lower Oven PRT   and lower  Designed for use when lower tool does not support a       The oven is operated using the Instrument Control Panel  Figure 2 20   The options displayed on the  Instrument Control Panel will be a reflection of the Environmental System Configuration entered using the  Instrument Configuration function  Table 2 5 describes the features of each option     Instrument Control Panel ix    Environmental Control Settings    Temperature                    100 0  PC  Max 500 0  C Min  150 0 C  Temperature Control        Oven  Sir  Chiller or LN2 Dewar          Environmental Controller    off   On  Liquid Nitrogen Dewar    Off   On  Oven Pressure Source     C Gas   LN     LN2 Rapid Cooling           C Off    Or       Motor Control Settings  Motor Power    Off   On       Figure 2 20  Control Panel showing Environmental Control Options    ARES User Manual 65      Table 2 5  Description of Oven Control Options in the Instrument Control Panel    OPTION DESCRIPTION    Temperature Desired temperature to conduct test  commanded  temperature     Temperature Control Selects the environmental system  oven  bath  in use  Set this  to   Oven  Air  Chiller or LN2 Dewar     Environmental Controller Turns on off the oven environmental system     Liquid Nitrogen Dewar Turns on off the LN2 Controller  If on  LN2 
79. STION    Figure 2 9  Bearing Lock Orientation  2K FRTN1 FRTNTE     The view shown is looking from the front of the instrument   A  Systems delivered before January 2002   B  Systems delivered after January 2002     ARES User Manual    High Resolution  HR  and High Torque  HT  Motor    The HT and HR motor do not require a bearing lock     Low Shear  LS  Motor    The LS motor bearing lock  Figure 2 10  consists of a formed metal plate that is fastened to the motor anvil  with 3 machine screws and one Phillips head screw  To install this lock the motor cover must be removed     Procedure for Locking and Unlocking Air Bearings  LS Motor    Refer to Figure 2 10 while performing the following procedures     To Lock the LS Motor air bearing     1     oy PTA ly    Read the Caution on page 11    Ensure that instrument power is off and air is applied to the Motor    Remove the Motor cover  Figure 2 21  page 68   retaining 1 of the Phillips head screws for the lock   Place the lock on the motor  aligning the 4 screw holes    Inset the 3 machine screws and the Phillips head screw and tighten    Retain the Motor cover and remaining Phillips head screws for re installation when the lock is  removed     To Unlock the LS Motor bearing     i oS    Read the Caution on page 11    Ensure that instrument power is off and air is applied to the Motor   Loosen and remove the 3 machine screws and the Phillips head screw   Remove the lock    Install the Motor Cover     NOTE  the Phillips head screw re
80. Steady Single Point  e Steady Rate Sweep    Strain Controlled   Transient  e Stress Relaxation  e Arbitrary Waveshape  e Step Shear Rate  e  Thixotropic Loop  e Force Gap Test  e Torque Normal Relaxation  e Multiple Extension Mode  o Constant Force mode  Creep Recovery   o Strain Rate mode  o Hencky Strain Rate mode  o Rim Shear  Squeeze Flow  mode  e Steady Step Rate Temperature Ramp    Stress Controlled Transient  e Constant Stress Test  e tress Ramp Test    ARES User Manual    Strain Controlled Dynamic Test Methods    Dynamic Single Point Measurement  Functional Description    The Single Point Measurement test makes a single measurement at a fixed frequency  strain and temperature     Dynamic Single Point Ei Ed  Frequency fi 0 0  rads  Max 500 0 Min 1  00e 05  SUMP    fo    Max 312 5000 Min 0 003125  Temperature  25 0  CE  Mas 600 0  C Min  150 0  C       Options  PreShear 0f Delay  Df Analoga O MeasOps  Oft    Options   End of Test   Save 4s   Help   Cancel         Figure 3 2  Dynamic Single Point Test Set Up Screen    Suggested Uses  Suggested uses for single point measurement are as follows     e Determination of unknown range response limitations of a new sample material   e Determination of parameters for the design of new tests   e Determination of force generated at various conditions and temperatures for the purpose of keeping force  within the range of the transducer during sweeps     Test Options    The following test options can be selected for use with the sing
81. T   WFL    1  00  G   1 8756TR    l  00 G   1 8756TW    Couette   1000XG      2TL R  Y   Double Wall Couette   1000 G     27L  R  Y t  R      Variables     length  mm  W   width  mm  T  thickness  mm  R  radius  mm  H  height  mm     Ge   gravitational constant   980 7 cm s        cone angle  Rpg   radius of the bob Rc  radius of cup  For double wall Couette refer to Figure 4 16 for definition of Ry  R2  R3  R4       ARES User Manual       Chapter 5    Calibration  Introduction    This chapter contains calibration procedures that you can performed  Do not attempt any calibration  procedure unless you are thoroughly familiar with the operation of both the instrument and Orchestrator  software  Calibration procedures are given for Orchestrator version 6 5 6  which is the software released with  the instrument at the time of this manual publication  However  menu and function names may change  without notice during subsequent software releases     WARNING  HIGH VOLTAGE is used in the operation of this  instrument  DEATH ON CONTACT may result if operating personnel fail  to observe safety precautions  Learn the areas of high voltage  connections  and exercise care not to contact these areas when  performing instrument calibration  Prior to working inside the  instrument  remove all jewelry  turn off the power  and ground points of  high voltage before touching them  Make adjustments using an  insulated electronic adjustment tool  Do not make physical contact with  any component i
82. TIONS  TRANSDUCER    Torque or Normal Force either does not respond    Applies to FRT transducers only   or is beyond full scale when no force is applied  1  Ensure that the correct air pressure is    supplied to the transducer in use  see  instrument specifications   if pressure is  incorrect  go to 3     Ensure that the transducer is unlocked   Call Technical Service        Stage does not move  NOTE  Manual AutoTension is activated by  selecting the HOLD button  see Orchestrator  Online Help for details      1  Make sure the stage is not at its limit of its  travel     2  Ensure that there is not a force overload  condition     3  Ensure that manual AutoTension functions  correctly  if not  go to 5     4  Ensure that manual AutoTension is not on  when trying to move stage using the Stage  Control  if it is on  turn it off     5  Ensure it is activated in firmware     6  Cycle the instrument main power  turn off   then on      7  Ensure that the oven is all the way to the left  or all the way to the right     8  Call Technical Service     Stage moves in only one direction when Ensure that a Normal Force Overload  AutoTension is not in use  condition does not exist  if it does  determine  the cause   Ensure that the stage is not at the limit of its  travel  if it is  adjust accordingly   Cycle the instrument main power  turn off   then on      Call Technical Service     Stage moves in only one direction when During sample loading  ensure that the stage  AutoTension is in use 
83. Ts  to ensure that they do  not collide with the tool  causing potential damage to the PRT     To adjust the oven clearance  do the following     1  Mount the lower tool and try and slide the oven past to tool   2  Note any clearance problems     3  Shave the foam using a razor knife or fine rasp  Cover the motor with a paper towel or cloth to keep the  shavings from getting into the motor     4  Slide the oven past the tool and again note any clearance problems     5  Continue shaving the oven until it slides past the tool     Once the oven clears the tool  close the oven almost all the way while noting the position of the lower PRT  relative to the tool  If the PRT will contact the tool  adjust its position  There is a mounting bracket that holds  the PRT located on the top of the oven under the outermost cover  Remove the outer cover to access the  bracket  Loosen the two screws and adjust the PRT position as necessary     ARES User Manual 169    170         FLUID SEAL    SAMPLE  OVERFLOW    WELL GLASS CHAMBER    SAMPLE    O RING    LOWER TOOL PRT  LOWER    TOOL    Figure 4 6  Cross Section Showing Fluid Seal    Adjust PRT Position   if needed    using 2 Screws on  PRT Top Mounting       Plate  OVEN  INSULATION  FOAM  Shave Down  This Edge As  LOWER Needed  OVEN PRT          Figure 4 7  Oven Modifications    ARES User Manual    Tool Installation    Select the Set Gap Instrument Control function under the Control menu in Orchestrator  Use the  Send to  Top  button to raise 
84. User Manual 201    Tool Maintenance  Fluid Bath 2    To facilitate cleaning of the lower tool  the inner cup can be removed from the outer cup  Unscrew the inner  cup mounting screw and remove the inner cup and O ring  Figure 4 17   Clean the cup as necessary   Reassemble the lower tool  cup  by installing the inner cup into the outer cup using the inner cup mounting  screw  Before reassembling the tool  inspect the O ring for cuts or other damage  and replace it if necessary     Inner Cup  Mounting Screw       1mm Gap         i AN Gap must be 1mm to obtain  A Effective Bob Length       Outer Cup A  L 32 mm     Effective Bob Length   Bob      rf tm i  Inner Cup   ii  F m l     iit a rae    1   T  Me      i  la R    gt    2R    INNER CUP DIAMETER   13 97mm     1  R     2R    INNER BOB DIAMETER   14 75mmi    R3 a   2R    OUTER BOB DIAMETER   16 00mm      I  Ry p k    OUTER CUP DIAMETER   17 00mm    A    cu    Figure 4 17  Fluid Bath 2 Double Wall Couette Setup Showing Inner and  Outer Cup as Well as Tool Dimensions    202 ARES User Manual        Zero the Gap        F Set Gap to 1mm          Raise Stage and  Fill Cup with Sample    Lower Stage  Back to 1mm Gap     Sample Should Slightly Over Flow Lip   Sample Volume 8 to 9 ml       PA Minimum Sample Fill Level    D     Figure 4 18  Fluid Bath 2 Double Wall Couette Sample Loading Procedures  A   Zero the gap  B   Set the gap to 1 mm   C   Raise the stage and fill the cup  D   With sample loaded     ARES User Manual 203    Test Fo
85. Zone  200    Max 350 Min 20       Zone Humber 1 J 3 4  Zone Time  s or kmp Eo m m    Direction ian     Clockwise ote       Options Preshear Of Delay  Of Analogln Oft      Options   End of Test   Save As   Help   Cancel         Figure 3 19  Torque Normal Relaxation Test Set Up Form    Suggested Uses  e Determination of time required for a sample material to relax after a deformation  as in sample material    loading    e Analysis of time dependent behavior of a sample     ARES User Manual    Options  The following test options are available when using Stress Relaxation     e Delay Before Test  e Steady PreShear    ARES User Manual    Force Gap Test    Description    Executing the Force Gap test adjusts the sample gap to a specified value over a specified time  Data can be  collected during the time that the gap is changing  The following parameters are set through the Force Gap  Test Set Up Screen  Figure 3 20     Force Sample Gap to    In this field  enter the sample gap that is desired at the end of the Force Gap test  The test concludes when this  gap is reached  regardless of the Gap Adjustment Time  next item      Gap Adjustment Time   This field determines the rate at which the gap command changes  Enter the time period during which the  gap will change from initial position  at the start of the test  to the gap specified in the Force Sample Gap to  field     Save Test Data    When checked  saves data that have been collected while the gap was changing  The data are repo
86. accomplished by switching in precision resistors  in place of the PRT and using the known resistance to adjust the offset and gain of the electronics  However   some errors can still exist due to PRT errors  or temperature gradients that exist within the oven  which cause  the temperature at the PRT to truly be different than the temperature at the sample  Although these errors are  generally small  they can lead to discrepancies between the temperature measured by the PRT and the actual  temperature of the sample at any given time     For studies that require the most exacting temperature accuracy  it can be helpful to input a temperature  calibration table to account for differences in measured temperature versus actual sample temperature   However  keep in mind that performing the calibration incorrectly could actually make the errors worse than  doing nothing  Of prime consideration is providing a temperature reference that is of sufficient accuracy to be  used as a calibration reference  One option is to mount an accurate thermometer in place of the sample and  compare its output to that of the PRT at several temperature steps  Another method used is to run a  temperature ramp study on a material with well defined and well known thermal transitions  such as the glass  transition point  and using the reported values to adjust the PRT temperatures     Temperature Calibration using Orchestrator       While temperature calibration is not handled under the standard instrument
87. al Force Calibration  Remove the weight and calibration hook from the transducer   and store both in the calibration kit     Low Range High Range             Set Transducer Characteristics      E        Transducer Selected    a lTransducer    Transducer 3  Transducer Description                     NN      Maximum Torque                  lacmlf200    fio000 Mo  Minimum Torque                   lg cm   0 004 ho oo Do o  Torque Calibration Value               1 20 42   1021 0 loo 80  Torque Compliance                         J6 50e 06   650e06  00     Inertia Constant                   lger lf  160 po 80S  Phase Adjust Constant  rads rad   1 086 04  1 08e 04  00           w                                                 Maximum Normal Force              lalf2000 0  f2o000 foo foo    Minimum Normal Force uwus O20 f20  oo foo    Normal Calibration Value            faro   2101 56  foo foo SC  Normal Compliance             umkg   0 0 fo  foo foo  Phase Offset ccs sense fred   0 0 fo bo Joao         OK   XducerCal   Help   Cancel      Figure 5 7  Set Transducer Characteristics Form    Shaded windows show input cells for Torque and Normal Calibration values     ARES User Manual 217    Phase Angle Check  Principle    The phase angle correction compensates for possible phase shifts that may be added to torque and strain due  to A D filtering by the instrument electronics  It is a function of the electronics and not the Motor or  Transducer  Once the correction is determined  it should
88. alue exceeded the limit at any point     If the sum of the wave times is less than the sum of the data collection times  then the motor will remain  stationary once the entire wave has been played     Arbitrary Waveshape Test ki    Temperature          25 0 PC  Max 0  0   Min 0 0  C  Points Per Zone    200    Max 350 Min 40  Aone Number 1 2 3 4    Time    or hms   jp oo ct ct ct       Wave Time Wave Equation straini    Ft  where t ig time    ze El Isnt   l2 El BO  ze El int   fz    ial Isnt     Strain Limits       Max 0  008905 Min 0  000000       Options  Delay Of  Options   End of Test   Wave   Save Ag   Help   Cancel         Figure 3 17  Arbitrary Waveshape Test Set up Screen    In evaluating each equation  the time for each equation is based on the running time since the start of the first  equation  For example  if two equations are used  each 2 seconds in length  the time variable is evaluated from  0 2 seconds for the first equation  and 2 4 seconds in the second  If there are discontinuities in the strain value  between the end of one equation and the start of another  a jump in position will occur during measurement    If this is not desired  then the strain value should be contiguous from equation to equation in order to avoid  such transients     ARES User Manual 127    Once the strain equations and sampling times have been specified  the waveform must be sent to the  instrument  The    Wave    button is used to compute the resulting waveform  display the results g
89. ample Loading esmero olor onneani si ia Sa a aN 160  ROTC HG lme eae A E E 160  A e Beene E ere E e IU E E A E 161  E A na D o tai A E E A A N N 161  OPen Ene TN a A A ES 161  oa i o o A 163   pee aa     LI 07 10 T 0 OCIO o O ee 163  Hlastelloy POOL tara lle Rates portones tino aaa cel 165  General Mora ON reo ocaciones 165  Uperitns NOS tia 165    Mts Re AAA     A E E E EA once E E O aa eee ey apeteyecuiases 168   Sy AE Ue Vea adas 169  A OA 171    ARES User Manual    Sample LOMO cesar 171    EP    A soe  173  General doo iio 173  perane Nao iitis 174  i Deve  Miatie 0l ciate q sip S   A A Teer TTS 175  Oa OS a OCA E T e ate secu Er cose E A T 175  Testing at Other than Ambient Temperatures    ocooconococcnonnnnnnnnananinonnnnnancnnononanononcnnnnnnanononnnnnann conc na nonncnnancanos 175  noe 1 Ak 6  Cal 8 2  6 o a pms ore eee E eR ener re E ee ee rere ee re ee eee 176   Torsion Kectansular New Destinos 177  A re tea UIT OT VAL ees a  Go oaceamerp EE ease ssacetcuded atecneseeaiesadieoenshecteradaueet spe  177  sample DINOS catas 178  Clip TU an 178  Operas lea acia 178  Coefficient of Thermal Expansion  O  PP PI e o iaia aea nn aa aai 180  AAA E A E Un E errr E mn Ter arr 180  Sample GO io 181   Torsion Rectansulas  Diana DE pandeo 184  Sanar o AAS Po mE ener earn ener ee sere ene et ene errr rere ererenn terrae 184  Dat Plo Pe MeO esa pi tee anes 184  da suit CEE o e in EE E 185  Costicientor Thermal Expansion osa anotaron ope nipona poi 186  dle ratio  hala  a OOOO o Ue 
90. an illuminated LED  indicates proper operation     ARES User Manual                       FRI BENDIX SPRING   15  15y     pr   HOTCRON SLAVE STEPPER BD         SLAVE TEMP 1 BD         A D FUNC GEN 2 BD  LS MOTOR          E Blinks at 1Hz      On when link to Host      Computer is active         All other LEDs are         15V   15V  PHASE LOCK   5V    0000    INDUCTOSYN  amp          O000 0000    Figure 6 1  Location of Diagnostic LEDs    ARES User Manual    Troubleshooting Guide    Table 6 1 and Table 6 2 list some problems that may arise during instrument calibration and operation  and  offers the corresponding corrective actions  Each corrective action is listed by number  Each action constitutes  a single troubleshooting operation  For example  first try corrective action 1  If the problem persists  try 2  and  so on until you reach Call Technical Service     Before calling technical service have the following information ready     e The instrument model and serial number  e The software version you are running     e Any troubleshooting steps you followed to diagnose the problem     Calibration  Table 6 1  Calibration Troubleshooting Guide    PROBLEM CORRECTIVE ACTIONS  NORMAL FORCE CALIBRATION    During either Torque or Normal Force 1  Exit the calibration operation and enter the  Calibration  the first value of force that is TRANSDUCER SETUP  UTILITIES  SERVICE   displayed is either very high  such as 1E 5  or TRANSDUCER   The TORQUE  or NORMAL   exactly zero  CALIBRATION
91. and lower frequency limits   of the sweep  Logarithmic frequency sweeps can be run in ascending or descending order  They can begin or   end at any frequency from 1x10e    to 80 Hz with a maximum of 500 data points measured during each decade  of frequency     Frequencies are selected by specifying initial and final frequencies  and the number of data points to measure  between each decade of frequency  points per decade   The number of points per decade includes the initial  frequency  but excludes the final frequency  As an example  consider a sweep conducted over a single decade  of frequency  between 10 and 100 radians per second  Selecting five data points to be measured per decade  divides the difference of the endpoint logarithms into five equally spaced fractional exponents  so that six  discrete frequencies are generated     e Initial Frequency   10 rad sec  e Decade Frequencies   15 9  25 1  39 8  63 1 rad sec  e Final Frequency   100 rad sec    One data point is measured at each of the frequencies    Linear Frequency Sweep   The linear frequency sweep uses the entered frequency limits  The test starts at the initial frequency and  continues until the final frequency is reached  adding the frequency increment to each subsequent frequency if  the final frequency is greater than the initial value  or subtracting the increment if the final frequency is less    than the initial value     As an example  a linear sweep from 10 to 50 rad sec  in increments of 10 rad sec  g
92. aneously  That is  the  collar should be tightened a small amount  then the screws started  then the collar some more  to ensure that  all components are aligned and mating properly  If at any time a part does not seem to fit well  or tighten  easily  stop and ascertain the reason for the problem  Do not force anything     ARES User Manual    6  Tighten the Peltier Assembly Collar by placing the two pins on the spanner wrench  provided  into  two of the holes machined into the Collar  and rotating the wrench clockwise until the Collar is snug   do not overtighten      7  Tighten the three screws in the Peltier Assembly Shaft using the supplied 5 5 mm open end wrench   do not overtighten          ame  WE    a PELTIER or BATH  HOUSING        CROSS SECTION SHOWN FOR VISIBILITY     Figure 2 30  Peltier Alignment Pin    Circulator Connections    Connect the Circulator hoses to the Peltier Assembly as follows     Fluid Output from Circulator Lower Hose of Peltier Assembly    Fluid Input to Circulator Upper Hose of Peltier Assembly    The circulator should be positioned on the floor below the work area  Make sure all hoses are installed  completely onto their respective hose barbs and that the hose clamps are tight        Filling the Circulator  Regardless of the type of circulator used  we recommend that the fluid used be a mixture of 50  water and    50  ethylene glycol  Use of pure water or any other mixture will ultimately result in damage to the Peltier  Assembly  specifically 
93. ansducer using the various test tools described in Chapter 4     CAUTION  Force Rebalance Transducers contain a precision air  bearing that is equipped with a bearing lock  which prevents movement  of the air bearing when no air is applied  To avoid damaging your  transducer  familiarize yourself with the operation of the bearing lock   see the next topic   and observe the following cautions        Do not apply power to the instrument when the bearing is locked   Do not unlock the bearing unless air is applied to the transducer     If the air supply must be intentionally interrupted  turn off instrument  power and lock the bearing prior to removing air     If the air supply is interrupted while the bearing is unlocked  do not  touch the anvil until air is restored     Maintain air flow to the transducer at all times to prevent contamination  of the air bearing     Failure to observe these cautions will result in damage to the  transducer     Stage    The stage  Figure 2 2  is a motorized platform that supports the transducer  The stage can be raised and  lowered to facilitate sample loading using either the manual stage control buttons on the right side of the test  station  or through Orchestrator software control  The rate at which the stage moves can be adjusted through  software     Manual Stage Control    The Manual Stage Control uses 3 push button actuators located on the lower right side panel of the Test  Station  Figure 2 2   It allows movement of the stage during op
94. anvil and insert the tool into the anvil  pulling apart the retainers if necessary   Hand tighten the knob  do not over torque      Lower Tool Installation  Fluid Bath Mount    To install a lower tool into either of the Fluids Baths  please refer to Chapter 2 for specific installation  instructions  For either bath  the correct lower PRT should be inserted first  followed by the tool     ARES User Manual    RETAINERS _        KNOB    UPPER FIXTURE NX    1  HOLD FIXTURE    HERE WHILE INSTALLING  OR REMOVING          RETAINER    ANVIL               a    KNOB    FIXTURE    2  PUSH FIXTURE FULLY UPWARD INTO  ANVIL AND HOLD FIXTURE IN PLACE           3  TIGHTEN THE KNOB        4  FIXTURE INSTALLED    Figure 4 1  Upper Tool Installation    ARES User Manual 155       LOWER FIXTURE       TOOL PRT       Sareea  ee  e  id  Ce      Va ALIGNMENT KEY    KNOB       Figure 4 2  Lower Tool Installation  Motor Mount    ARES User Manual    Setting the Gap    Zeroing the gap between upper and lower test tools is a prerequisite to setting the gap during sample loading   Zeroing the gap requires lowering the stage until the upper and lower tools touch  then setting the GAP value  to read zero millimeters  After that  moving the stage results in an accurate gap indication as displayed by the  GAP value     The following procedures are general for all test tools  Specific details regarding each tool are described under  that tool s individual section  Two methods are available for zeroing and setti
95. as expressly set forth in this  LICENSE     3  OTHER RESTRICTIONS  This LICENSE is your proof of license to exercise the rights granted herein and  must be retained by you  The following also apply to the software described above as covered by this  LICENSE    a  You may not rent or lease this software  but you may transfer your rights under this LICENSE on a  permanent basis provided you transfer this LICENSE  this software  security key s   and all  accompanying written materials and retain no copies  and the recipient agrees to the terms of this  LICENSE  Any transfer of this software must include the most recent update and all prior versions      continued     ARES User Manual    b  You may not reverse engineer  decompile  or disassemble this software  except and only to the extent  that such activity is expressly permitted by applicable law notwithstanding this limitation    c  This software is licensed as a single product and its component parts may not be separated for use on  more than one computer    d  Without prejudice to any other rights  TA Instruments may terminate this agreement if you fail to  comply with the terms and conditions of this agreement  In such event  you must destroy all copies of  this software and all of its component parts     4  NOWARRANTY  The software described above and documentation are provided on an    as is    basis and  all risk is with you  TA Instruments makes no warranties  express  implied  or statutory  as to any matter  whatsoeve
96. ated by a viscoelastic material can be separated into two    components  an elastic stress  T      that is in phase with strain  and a viscous stress  T       that is in phase with    the strain rate  AY dt  but 90   out of phase with strain  The elastic and viscous stresses are sometimes referred  to as the in phase and out of phase stresses  respectively     The elastic stress is a measure of the degree to which the material behaves as an elastic solid  The viscous  stress is a measure of the degree to which the material behaves as an ideal fluid  By separating the stress into  these components  both strain amplitude and strain rate dependence of a material can be simultaneously  measured     The viscous and elastic stresses can be related to material properties through the ratio of stress to strain  or  modulus  Thus  the ratio of the elastic stress to strain is referred to as the elastic  or storage  modulus  G       which represents the ability of a material to store energy elastically  The ratio of viscous stress to strain is  referred to as the viscous  or loss  modulus  G         and is the measure of a material s ability to dissipate energy   The complex modulus  G   is a measure of the overall resistance of a material to deformation  If these  measurements are made using a linear geometry instead of a shear geometry then the letter E is used to  represent the modulus  instead of G     ARES User Manual    In some cases it is useful to define the ratio of sample strai
97. ations    Table 1 9  Specifications  Force Rebalance Transducers 2K FRTN1 and 2K FRTN1E  TORQUE    Measurement Range High Range  2 to 2000 gecm  Low Range  0 02 to 200 gecm    500 rad sec version V8 x and above  Maximum Operating Frequency 100 rad sec older than V8 x  Measurement Range 2 to 2000 gmf       Table 1 10 Specifications  Force Rebalance Transducer 1K FRTN1    Jp     TORQUE SPECIFICATION    Measurement Range High Range  1 to 1000 gecm  Low Range  0 002 to 20 gecm    200 rad sec versionV8 x and above  Maximum Operating Frequency 100 rad sec older than V8 x    NORMAL FORCE SPECIFICATION    Measurement Range 2 to 2000 gmf       Table 1 11  Specifications  Force Rebalance Transducers 100 FRTN1 and 200 FRTN1    SPECIFICATION 100 FRTN1 SPECIFICATION 200 FRTN1  Measurement Range High Range  0 04 to 100 gecm   High Range  0 08 to 200 gecm  Low Range  0 004 to 10 gecm   Low Range  0 008 to 20 gecm    ml perang 100 rad sec  16 Hz   Frequency    SPECIFICATION 100 FRTN1 AND 200 FRTN1  Measurement Range 0 1 to 100 gmf       ARES User Manual    Table 1 12  Specifications  Force Rebalance Transducers 100 FRT and 200 FRT    TORQUE SPECIFICATION 100 FRT SPECIFICATION 200 FRT  Measurement Range High Range  0 02 to 100 gecm   High Range  0 04 to 200 gecm  Low Range  0 002 to 10 gecm   Low Range  0 004 to 20 gecm    Maximum Operating Frequency 100 rad sec  16 Hz     Not Used for Measurement       Table 1 13  Specifications  Standard Transducers 2K STD and 10K STD    SPECIFICATION 2K ST
98. be used for very thin samples  with the stipulation that an identical  number of equally sized shims must be used on either side of the sample     NOTE  Loading soft samples or samples that do not properly fit the inserts  can result in inaccurate data     Operating Ranges    Operating range is defined as the region bounded by the maximum and minimum complex modulus G that  can be measured by each transducer type using the torsion rectangular geometry     The following geometry specific factors affect the operating range of torsion rectangular geometry     e Thickness of sample  stress constant K  and strain constant K   e Width of sample  stress constant K  and strain constant K   e Length of sample  strain constant K     Additionally  the following instrument specific factors affect the operating range of all geometries     e Minimum torque that can be measured by the transducer  e Transducer compliance  e Maximum strain that can be generated by the motor    To calculate the minimum and maximum G that can be measured by each transducer type using the torsion  rectangular geometry  use the following equation     x K          where K    Stress Constant  K    Strain Constant    and C is computed from the following     ARES User Manual 185    C for G  MAXIMUM C for G  MINIMUM    2K FRTN1  2K FRIN1E    This transducer is not generally recommended  4K FRTN1 for use with this tool  However  it may be used  for some limited applications     2K STD B  10K STD A J   2 60e 06 r  see n
99. bly  is used to sense axial force     Environmental Control Options    This instrument can be configured to subject samples to various thermal environments  These options are  a  Forced Convection oven  a re circulating fluid bath  or a Peltier system  See Chapter 2 for details     ARES User Manual    HOST COMPUTER kk RS 2320       TRANSDUCER SUBSYSTEM    TORQUE RESPONSE    I   I   l   I   I  ETTE TRANSDUCER    A ss SIGNAL  TORQUE    eae oe CONDITIONER       CPU AND RAM       eee           m       FUNCTION    MOTOR GENERATOR    CONTROLLER  TEMP  CONTROLLER       ee A ee ee ee ee ee ee ee A A ee ee                ii TEMP CMD    TOOL TEMPERATURE   AND    MEASURE        eee eee                   i  i          TEMP CONTROL  gt  A ooann nnn    i  ee ance ENVIRONMENTAL CONTROLLER e   ARES  I    scan e TEST STATION      TEMP    ae CONTROLLER    Mee TERT UN a          from  CPU BOARD    TEMP CMD        AND      CIRCULATOR MEASURE    TEMP CONTROL  ANG FEEDBACK PELTIER OPTION    TOOL TEMPERATURE FLUID BATH  OPTION       Figure 1 1  ARES  Functional Block Diagram    ARES User Manual    Principles of Operation    Figure 1 1 is a simplified block diagram of the ARES  The following discussion offers a brief explanation of  ARES operation  using Figure 1 1 for reference  In Figure 1 1  terminology printed in upper case refers to the  various components or subsystems  and the signals generated by the system are printed in italics     The ARES consists of the following subsystems     e Contro
100. ceptacle  Do not use excessive force if resistance is encountered  while lowering  as this may damage the PRT Plug and Receptacle  In  this case  raise the Peltier Assembly  check the alignment described in  the next step  and try again        ARES User Manual    UPPER HOSE    FLOW my             COVER   RAISED POSITION        SHAFT    SCREWS  3        at serio  COLLAR p    i       LOWER HOSE    q FLOW                      PRT PLUG    Oq RED DOT    VIEW FACING FRONT OF INSTRUMENT    Figure 2 29  ARES Peltier Assembly    5  Slowly lower the Peltier Assembly onto the ARES Motor Housing  ensuring that     e The three screws in the Peltier Assembly Shaft align with the three threaded holes machined into  the ARES motor anvil     e The Pin  Figure 2 30  machined into the bottom of the Peltier Assembly is seated into the notch in  the ARES Motor Housing  the Pin and notch should be located toward the rear of the instrument    It may be necessary to rotate the Peltier Assembly back and forth until the Pin falls into the notch     e The threaded Collar of the Peltier Assembly rests in the threaded portion of the ARES Motor  Housing     e Ensure that the spring  Figure 2 29  is positioned behind the Anvil Tightening Knob on the ARES  Motor  Rotate the Motor Anvil to gain access to the Anvil tightening knob then tighten the knob  a  flat head screwdriver may be used if you do not overtighten      While the next steps are listed sequentially  they should be performed more or less simult
101. chined into the Collar  and rotating the wrench clockwise  Do not overtighten the Collar    2  Tighten the three screws in the Shaft using the supplied 5 5 mm open end wrench  Again  do not  overtighten the screws     VIEW FACING FRONT OF INSTRUMENT    UPPER HOSE    FLOW ml        COVER   RAISED POSITION           SHAFT  SCREWS  3   COLLAR E  LOWER HOSE    T                Figure 2 21  Fluid Bath 2 Lower Portion Details    68 ARES User Manual           LOWER FIXTURE       LOWER FIXTURE  WRENCH   Fixture Removal Only     BATH WRENCH     Lower Fixture  Removal Only     COVER   RAISED POSITION     THREADED COLLAR    SPANNER WRENCH UPPER HOSE     FLOW OUT     LOWER HOSE   FLOW IN     THREADED  COLLAR    a    Y a  Ps    MOTOR COVER a    MOTOR ANVIL    KNOB    Figure 2 22  Fluid Bath 2 Components and Installation    ARES User Manual    Circulator Connections    Connect the circulator to the bath using the supplied connectors  The lower hose on the bath is for flow into  the bath  out of the circulator  and the upper hose is for flow out of the bath  Figure 2 21   The circulator  should be positioned on the floor below the work area  Make sure all hoses are installed completely onto their  respective hose barbs and that the hose clamps are tight  Connect cable 707 00750 between the Circulator RS   232 connector  and the Test Station Signal Panel CIRCULATOR connector     PRT Installation    There are two PRIs available  Figure 2 24   each designed to be used with a specific lower too
102. ck the Transducer and Motor bearings    Remove Air supply to instrument     Po ES    ARES User Manual 227    Inspect the air hoses for cracks and other damage that could result in leaks  especially in the vicinity of the  bend radii  If any damage is found  notify TA Instruments for service     There are also internal filters inside the test station that should be inspected and serviced on an approximately  6 month to 1 year basis  Only qualified service personal should perform this maintenance     If no leaks are found  apply AC power to the instrument as follows     1  Establish airflow through the air dryer  but do not connect the air output to the test station  Allow the  air to purge for 3 to 5 minutes    Connect the air supply to the test station    Unlock the transducer and motor bearings    Install the POWER IN plug in the AC main  line voltage source     Push the Main Power Switch to the ON  I  position     a ly    Verify that the air pressure to the motor  transducer  and oven are correct  Note  to check the oven air pressure  the oven must be all the way to the right and closed     Air Dryer    Inspect and service the air dryer according to its manual  This should include draining and cleaning the pre   filters  Replace them if necessary     Cleaning the Instrument    If the exterior plastic or metallic surfaces of the instrument require cleaning  use only a solution consisting of a  non abrasive household dish detergent and water     Clean as follows     1  Remove
103. complex viscosity N   Y axis  versus frequency     X axis      The region between the upper and lower limits of operation is the range of complex viscosity that can be  tested     ARES User Manual 167           COVER    UPPER COVER CLIP  TOOL  SEALING FLUID  WELL  UPPER  PLATE  GLASS CHAMBER  SAMPLE OVERFLOW  LOWER WELL  PLATE  O RING  LOWER  TOOL       Figure 4 5  Hastelloy Tool  Liquid Seal    The upper tool is designed with a liquid seal  The upper plate has a  well   machined in it  that holds the  sealing fluid  When the upper cover is slid into place on top of the glass chamber  a lip on the under surface of  the cover extends into the fluid  completing the seal  Figure 4 6   The bottom of the glass chamber is sealed  with a special O ring  During use  the lower tool  glass chamber  and cover move as one piece  effectively one  large lower tool     The liquid provides a good seal without the drag associated with conventional seals  To ensure optimal  performance  the sealing fluid should be of low viscosity  The sealing fluid should also be  chemically  appropriate  for the sample material to avoid any dangerous reactions  or interactions between the sample  and sealing fluid     For the best  and most consistent results  the volume of sealing fluid should be kept constant for each sample  run  Some experimentation may be necessary to determine the optimal fluid volume for a particular tool gap  and sample  A pipette  or other accurate fluid delivery system  should be u
104. concerning the operation of  Orchestrator software  see the Orchestrator  Online Help     Description of  Instrument    The Advanced Rheometric Expansion System   ARES  is a mechanical spectrometer that is  capable of subjecting a sample to either a  dynamic  sinusoidal  or steady shear strain  deformation  and then measuring the  resultant torque expended by the sample in  response to this shear strain  Shear strain is  applied by the motor  torque is measured by  the transducer  Strain amplitude and  frequency are set by the operator  with the actual sample deformation determined by the measured motor  and  transducer  displacement        Motors    There are 3 basic motors available for use in ARES  The High Resolution  HR  motor is a ball bearing  direct  drive motor  It provides very precise and accurate motion in both dynamic and steady modes  The applied  strain is essentially instantaneous     The High Torque  HT  motor is similar to the HR Motor  but is designed to deliver much higher torque  The  HT motor is intended for use in conjunction with the 10K STD transducer     The Low Shear  LS  is most commonly used for fluids testing  The motor shaft is supported axially by a    precision air bearing  This minimizes axial runout  providing the smoothest normal force data  It also has a  special controller that allows extremely low shear rates to be applied during steady shear tests     ARES User Manual    Motor Modes    In Dynamic Mode  the motion is oscillatory  The mot
105. ctive zone  The Ramp Rate is the Rate of thermal change that the sample material undergoes during   the test  Orchestrator reports the difference between the Initial Temperature and Final Temperature setpoints   divided by the Ramp Rate that you specified as the computed Ramp time     Ramp direction is set using the Initial and Final Temperatures in a zone  The Ramp Rate is in units of   C per  minute    C min    and the entry field accepts any value within the range of the instrument  with a display  resolution of 0 1  C  If temperature control is not needed  and the deformation is the only desired change  the  Ramp Rate should be set to O  C min  A Ramp Rate of 0  C min  makes the test a time sweep  hence  this test  is sometimes referred to as a time cure sweep     Following the Ramp  the Soak Time After Ramp is the time period during which temperature is held at the  Final Temperature  Orchestrator reports the total time required to run the test  including all zones and soak  times as the Computed Test Duration     Steady Step Hate Temperature Ramp Test Ea Ed  Initial Temp           25 0 PC  Ma  500  0   Min  150  0      sampling Mode  Log Linear  Points Per Zone   200    Max 350 Min 20       Shear Rate Limits  1 s     Max 400 5737 Min 0 004005  Zone Number 1 2    3 4  Final Temp  cono  Poo foo   foo fo  Ramp Rate oe PCAmin yao lao ooo po 8 foo  Computed Ramp Time           ms  230 pza      Soak Time After Ramp  s okma   fo Mo   fo  O Mero Poo     no foo o  Computed Test 
106. d Orly    Li    Instrument Control Panel Showing Fluid Bath Control Options    Peltier    Description    The Peltier environmental control system utilizes a solid state heat pump  thermopile  to control the  temperature of the lower tool  which is an integral part of the Peltier Assembly  In general  the solid state heat  pump consists of multiple semiconductor devices in series  alternating P type and N type  The devices are  placed between the lower tool and a heat sink  Figure 2 28                      e ELECTRICAL INSULATOR   HEAT CONDUCTOR        gt   a    ELECTRICAL CONDUCTOR          N  N TYPE SEMICONDUCTOR             P P TYPE SEMICONDUCTOR          Figure 2 28  Solid State Heat Pump  Peltier Element  Schematic    DC voltage is applied across the semiconductors to either heat or cool the lower tool  depending on the  polarity of the voltage  The semiconductors transfer thermal energy between the lower tool and the heat sink   which is essentially a heat exchanger through which fluid  typically water  is circulated  The fluid  temperature determines the amount of thermal energy that the heat sink can source to or sink from the lower  tool  The fluid temperature thereby modifies the actual operational temperature range within the overall  System Specifications  Table 2 8      Table 2 8  Peltier System Specifications      30  C to  150  C  depending on circulator fluid  temperature     30  C per minute from 0  C to  100  C   at 20  C circulator fluid temperature      0 1
107. d appears at the top seam between the  bath cup and body  Figure 2 35   the clamp should be tightened immediately     Filling the Circulator   Depending on the type of circulator in use and the desired operating range of the circulator  fill the circulator  with fluid as specified in Table 2 10  Refer to your Bath documentation for specific circulator filling and  operating guidelines  as well as other bath fluid options for your application  Also refer to the bath fluid MSD  for guidelines regarding the safe handling of your particular bath fluid     Table 2 10  Circulator Operating Ranges Using Various Bath Fluids    Desired Operating Range of Circulator Fluid    bath temperature range is slightly less    20  C to  30  C 100  Dow Corning   Syltherm   XLT   1  C to  99  C 100  water     5  C to  100  C 50  ethylene glycol   50  water     20  C to  150  C Julabo   Thermal H10S     40  C to  110  C 100  ethylene glycol       ARES User Manual    LABEL       ACCESS PORT       BATH OUTLET    Q  EN    N    Figure 2 37  Fluid Bath  Fluid Connections    ARES User Manual    Fluid Bath Configuration in Orchestrator     When you command a temperature  the Orchestrator software uses the Instrument Configuration to determine  the environmental system currently in use  and establish operating conditions  Prior to operating the fluid  bath  access the Instrument Configuration function  located under the Service function of the Utilities menu   and set up the instrument using the guidelines
108. d temperature     When the Peltier system is used  temperature control and monitoring is accomplished through the lower tool  PRT connector  TOOL TEMPERATURE   and special electronics in the ARES Test Station     ARES User Manual    Specifications    Table 1 1  Physical Specifications    Chassis Width Depth Height Weight    Test Station 26 in 21 in 36 in 275 lbs  LN   Controller 11 in 9 5 in 22 in 38 lbs    Mechanical Chiller 29 in 20 in 46 in 260 lbs  Fluid Bath Circulator 12 in 17 in 295 in 74 lb    Table 1 2  Test Station Operating Specifications    Operating Parameter Specification    180 to 264 VAC   20 A  50 60 Hz single phase  Transient Over Voltages Installation over voltage category II       Temperature Range  Ambient 5   to 40   C    Relative Humidity 80  up to 35   C  then decreasing linearly to 50  at  40  C       ARES User Manual    Table 1 3  Pneumatic Ratings       AIR INLET RATINGS    AIR PRESSURE FLOW AIR QUALITY    Instrument quality air having the  with Oven  following characteristics   5 5 scfm  156 l min  e Any particles present in the air  Instrument 80 psi  5 5 bar  must be smaller than 5 microns in  without Oven  diameter   3 scfm  85 l min  e Relative humidity   35 to 70   Dewpoint   10   C    with Oven   7 5 scfm  213 l min        mea       Air Dryer 100 psi  6 8 bar   without Oven   5 scfm  142 l min        Table 1 4  Accessories Power Specifications    NOMINAL RATED    COMPONENT VOLTAGE  CURRENT  AND FREQUENCY    USA   Fluid Bath Circulator 115 VAC
109. d valve to control  liquid nitrogen flow into the Dewar  and hardware to control the boiling of the liquid nitrogen to produce very  cold nitrogen gas within the flask  A second solenoid valve regulates the flow of the cold nitrogen gas out of  the Dewar flask for use by the Test Station     Dewar Flask    A Dewar flask is a container specifically designed to efficiently store liquid nitrogen  Dewar flasks help  prevent evaporation due to their double wall construction  Within the Dewar flask  a heater immersed in the  LN  provides controlled boiling  The Dewar is equipped with a pressure relief valve to prevent damage to the  system  It is common for the relief valve to open  loud pop  several minutes after the LN2 controller is shut off   as the residual liquid nitrogen in the Dewar evaporates to gas     Connection    The LN2 output connects to the Forced Convection Oven  Connect the signal cable from the LN2 controller to  the LN2 port on the Signal Panel  Figure 2 12   All connections to the ARES are normally made during system  installation by a TA service technician  Connect the LN2 controller filler hose to your liquid nitrogen source   The external liquid nitrogen tank should have a low pressure   lt 30 psi  valve  The hose between the nitrogen  tank and the LN2 controller should be kept as short as possible to maximize the LN2 controller performance     Software Configuration    The LN2 operates using Orchestrator software  Chapter 2 discusses the set up and operatio
110. description of the AutoTension parameters that must be set in the AutoTension Adjustment Set   Up Screen  Fig 3 26      AutoTension Direction    When Tension is selected  tensile static force is applied  When Compression is selected  compressive static  force is applied     Initial Static Force    This is defined as the static force that is maintained throughout the test     AutoTension Sensitivity  Minimum change in normal force that results in an adjustment to maintain the Initial Static Force   Switch AutoTension to Programmed Extension When Sample Modulus        When sample modulus decreases below the value entered in this field  the last valid coefficient of thermal  expansion  0  of the sample is used to determine an AutoTension static force that best maintains the sample  length  The coefficient of thermal expansion  01  is given by     AL  am  AT  where  AL   change in sample length  L   sample length  AT   change in sample temperature    This feature is useful when running temperature  sweeps   to prevent excess sample stretching with  increasing temperature  It is important to apply this adjustment when sampling through the glass transition  point  where there will be a significant  rapid drop in G   Without this adjustment  the now  softer  sample   beyond the glass transition point  would rapidly stretch and ultimately be pulled part  A test run may be  necessary to determine where the glass transition point occurs before setting this value     142 ARES User Manua
111. ducer     WARNING  Do not attempt to lift or carry the Test Station alone   Attempting to lift or carry the Test Station can result in serious personal  injury or damage to the Test Station     CAUTION  Never place any lower tool into the bath if the temperature  of the lower tool is cooler than that of the bath     Placing a tool into a warmer bath will result in expansion of the tool  during use  After expansion  the tool may not be removable without  damaging your bath     We suggest that you partially insert the tool by placing a phenolic  spacer between the upper lip of the lower tool  and the surface of the  bath  Allow the lower tool temperature to match that of the bath  then  remove the spacer and fully insert the lower tool              WARNING  Fuses must be changed by a qualified electronic  technician only     WARNING  Prior to changing a fuse  ensure that AC power to the Test  Station is OFF  Changing a fuse on a live electrical circuit can cause  serious personal injury or death     WARNING  For continued protection against fire hazard  replace only  with a fuse of the same type and rating     WARNING  The following procedure must be performed only by a  qualified electronic technician     WARNING  Ensure that AC power to the Test Station is OFF before  attempting the following procedure  see step a   Changing a fuse ona  live electrical circuit can cause serious personal injury or death     WARNING  The Main Power Switch does not remove power to the  oven or L
112. ducer in use  from the loop in  the line     c  Ensure that the weight s  is  are  free to hang without obstruction  and that the weight is  steady  not swinging from side to side    Select the  Torque Cal  button  The Torque Calibration form is displayed     Enter the Calibration Torque specified in Table 5 3  depending on the transducer in use     Select the  Calibrate Now  button  When calibration is complete  the Transducer Calibration form is  again displayed     ARES User Manual    Transducer Calibration       Figure 5 4  Transducer Calibration Form       Figure 5 5  Applying Torque Using the Calibration Weight    ARES User Manual 213    15  Verify that the Torque value displayed on the Transducer Calibration Form  Figure 5 4  is within the  limits for the TORQUE VALUE shown in Table 5 3  depending on the transducer in use  Verify that the  Current Torque Cal value displayed in the Transducer Calibration Form is within the limits for the  CALIBRATED FULL SCALE VALUE shown in Table 5 3     If the Torque values are not within the specified limits  contact TA Instruments Service   16  Select the  Accept  button  Control returns to the Transducer Characteristics form     17  Verify that the Torque Calibration Value displayed in the form  the high range value if the  transducer is an FRT  is the same as the Current Torque Cal value just displayed in the Transducer  Calibration form  Press  Ok      18  If the transducer in use is an FRT perform the following step  If the tran
113. dulus Limits for Cone and Plate   2K and 10K Standard Transducer     PLATE DIAMETER CONE ANGLE Gx MAXIMUM Gx MINIMUM   mm   rad   dynes cm    dynes cm       0 1 9 22E 05 9 59E 01  2K    4 80E 02  10K     1 84E 05 1 92E 01  2K   9 60E 01  10K     3 69E 05 3 84E 01  2K    1 92E 02  10K    0 02 2 31E 04 2 40E 00  2K   1 20E 01  10K    0 04 4 61E 04 4 79E 00  2K   2 40E 01  10K        ARES User Manual    Appendix Table A1 7  Complex Modulus Limits for Cone and Plate  FRTN1 Transducer     PLATE DIAMETER   CONE ANGLE G  Maximum  dynes cm     Gx MINIMUM   mm   rad  at Frequency     rad sec   dynes cm        9 22E 04  9 22E 05  4 79E 07    1 84E 04  1 84E 05  9 59E 06    3 69E 04  3 69E 05  1 92E 07     wm   100  w  10  w  lt  10       100  w  10  w  lt  10       100  w  10  w lt  10    2 31E 03 w   2 31E 04 w     1 20E 06    4 61E 03 w   4 61E 04 w     2 40E 06    1 92E 00  High range  100 FRTN1   3 84E 00  High range  200 FRTN1   1 92E 01  Low range  100 FRTN1   3 84E 01  Low range  200 FRTN1   3 84E 01  High range  100 FRTN1   7 68E 01  High range  200 FRTN1   3 84E 02  Low range  100 FRTN1   7 68E 02  Low range  200 FRTN1   7 67E 01  High range  100 FRTN1   1 53E 00  High range  200 FRTN1   7 67E 02  Low range  100 FRTN1   1 53E 01  Low range  200 FRTN1   4 79E 02  High range  100 FRTN1   9 58E 02  High range  200 FRTN1   4 79E 03  Low range  100 FRTN1   9 58E 03  Low range  200 FRTN1   9 59E 02  High range  100 FRTN1   1 92E 01  High range  200 FRTN1   9 59E 03  Low range  100 F
114. e    Tighten the lower and upper collars       Raise the stage until a force of approximately 10  of full scale is generated  Please note that this    tension level is a general recommendation only  and you should set the tension level according to the  sample characteristics  When using the AutoTension feature  adjust the stage so that the normal force  is Zero    Using the  Motor Position Offset  slider in the Set Gap Instrument Control function in Orchestrator  adjust the motor position until the displayed torque is zero    DO NOT use the  Offset torque to Zero  button    Read the gap and record this distance as the sample length  2K FRT transducers only     Use Hold function under the Control pull down menu when changing temperature     ARES User Manual    UPPER FIXTURE                         NON  y ya  A 4  A Wau  6 05  A      SAMPLE  ENTER THIS  DIMENSION AS         INSERT    SAMPLE LENGTH  ONE EACH   UPPER AND LOWER                   COLLAR  S  ONE EACH   UPPER AND LOWER   v E l  N 4  N F A   a        STOP PIN    LOWER FIXTURE       Figure 4 12  Torsion Rectangular with Sample Loaded    ARES User Manual    Couette    Strain Constant Stress Constant     1000   G        on L  Rg     MOTOR MOUNT  Variables  ambient testing     Ge   Gravitational constant    980 7 cgs  or 98 07 Sl   L   Length of bob  mm     Rp   Radius of bob  mm     Rc  Radius of cup  mm     Options  25 mm bob   27 mm cup  ambient testing only     32 mm bob   34mm cup  for fluid bath or fluid bath 2   1
115. e PID loop settings  Figure 2 32   affect how the Peltier settles at a set point temperature  As the Peltier reaches a setpoint there will typically be  a few degrees of overshoot  after which the temperature will oscillate a few times and then stabilize  The  Peltier will typically reach stabilization within a few minutes  By modifying the Proportional Band    P coefficient  less overshoot can be achieved  but there will typically be more oscillations before the bath  stabilizes  so the overall stabilization time will be longer     PID Setup Form ki Ei    PID Temperature Control Setup        Peltier    Proportional Band  2 5  4   peset o ae i 0 87  Reps Mmin   cle eae  0 01 F  Miri        Ok Help   Cancel      Figure 2 32  PID Setup Form Showing Peltier Settings       Table 2 9  Default PID Loop Values for Various ARES Firmware Versions    PID Loo Version 4 Version 5 and 6  P Firmware Firmware    Proportional 5 2 5   Band    Reset 10 67reps min 10 67reps min       Ras  osm CI    ARES User Manual    Peltier Operating Requirements    The Peltier can operate only if the following conditions are met   e The Peltier is selected as the current environmental system   e The environmental system is turned ON in the Instrument Control Panel     e Fluid must be flowing through the Peltier     pumped through it  At all times while operating the Peltier system   ensure that the circulator is connected to the Peltier Assembly  and the  circulator pump is ON  Failure to heed this warning w
116. e Shaft with the flat portion of the ARES  Motor Anvil  both flats should be facing toward the right as you face the instrument      5  Slowly lower the Bath onto the ARES Motor Housing  ensuring that     e The three screws in the Shaft align with the three threaded holes machined into the ARES motor  anvil     e The Pin  Figure 2 23  machined into the bottom of the Fluid Bath is seated into the notch in the  ARES Motor Housing  the Pin and notch should be located toward the rear of the instrument   It  may be necessary to rotate the bath back and forth until the Pin falls into the notch     e The Threaded Collar of the Fluid Bath rests in the threaded portion of the ARES Motor Housing     e Ensure that the Spring  Figure 2 21  is positioned behind the Anvil Tightening Knob on the ARES  Motor  Rotate the Motor Anvil to gain access to the Anvil tightening knob then tighten the knob   A flat head screwdriver may be used if you do not overtighten     While the next steps are listed sequentially  they should be  in practice  performed more or less simultaneously   That is  the collar should be tightened a small amount  then screws started  then the collar some more  to  ensure that all components are aligned and mating properly  If at any time a part does not seem to fit well  or  tighten easily  stop and ascertain the reason for the problem  Do not force anything     1  Tighten the Threaded Collar by placing the two pins on the spanner wrench  provided  into two of the  holes ma
117. e optional LN2 controller  operation to  150  C  or the mechanical air Chiller  operation to  60  C  can  be used to extend the operational range of the oven  The air convection oven is recommended for polymer  melts and solids  and can accommodate Cone and Plate  Parallel Plate  Torsion Rectangular and Torsion  Cylindrical geometries     The recirculating Fluid Bath 2  or original Fluid Bath  is appropriate for liquid samples that may evaporate  under the influence of circulating air  The temperature range of the bath is approximately  20   to 140   C   depending upon the Circulator and bath fluid used   and can be used with Parallel Plates  Cone and Plate   Couette  and Double Wall Couette geometries     Our patented rotating  oscillating Peltier system has a temperature range of  30   to 150   C  with heating rates  as high as 30   C min  The Peltier system can be used with Cone and Plate or Parallel Plate geometries     Refer to the following sections for information regarding each specific environmental option     OVEN OPTION       FLUID BATH OPTION    ARES User Manual    Oven    The oven is a forced air convection environmental chamber that encloses the sample  Mounted in the oven are  two resistive heaters  gun heaters  that are used to control the temperature of the gas that is input to them   During testing at or above ambient temperature  either air  or nitrogen gas  can be input to the heaters  If test  temperatures must be extended to below ambient  the input to t
118. e or negative  that the sample material undergoes during the test   Ramp direction is set using the starting and final temperatures in a zone  The ramp rate is in units of  degrees minute  The entry field accepts any positive value for ramping rate  The actual maximum ramping  rate that can be accurately maintained is a function of the size and thermal properties of the sample  If  temperature control is not needed  and strain or sampling rate is the only desired change  the ramp rate  should be set to 0  C min     Time Per Measurement  Time Per Measurement establishes the sampling rate  When entering this value  consider the time per  measurement cannot be shorter than the physical time required to make a dynamic measurement  If the time    entered is less than physical time required to make a dynamic measurement the instrument will sample as fast  as it can     ARES User Manual 109    Strain    Strain can be set to any value within the range of the instrument  as shown in the boundary window   but  should not exceed the linear viscoelastic region of the sample material  If a value of zero is entered then this  zone inherits the strain level from the previous zone  Note also that this value can be overridden if the  AutoStrain function is being used     Soak Time After Ramp    Following the Ramp  the time period during which temperature is held at the Final Temperature before  proceeding to the next zone     Dynamic Temperature Ramp Test Ei Ed    Frequency  f 0 0  rad s  Max
119. e regulator  if necessary  to achieve 60 psi by turning the knob  clockwise to decrease the pressure and counterclockwise to increase pressure     During extended use  parts of the Chiller may freeze inside  blocking gas flow  To help prevent damage to the  gun heaters  due to insufficient gas flow   the Chiller is equipped with an emergency shut off that will shut  down the compressor  and turn off the heaters  if the airflow through the Chiller drops  If this happens turn  off the Chiller and allow it to thaw before continuing  Freezing will generally happen if the dew point of the  input gas is greater than  80   C  The air dryer is designed to provide dry gas with a dew point sufficient to  prevent freezing  The Chiller should be turned off when it is not actually in use  and the air dryers should be  properly maintained and inspected     Maintenance    There are no user serviceable parts inside the refrigeration unit  The compressor and condenser are lifetime  lubricated and do not require oiling  The condenser fins and grill should be kept clean to ensure good airflow   Clean them as needed by vacuuming or blowing with compressed air  The air filters should be inspected and  cleaned as necessary  Also  any moisture should be drained from the filter housings regularly  Refer to the air  dryer  air filter  and refrigeration unit documentation that came with the Chiller for further maintenance  information     The balance pressure should also be periodically monitored as d
120. e values result in counterclockwise rotation  In each zone the data can be sampled at a logarithmic or  linear interval  Figure 3 23      Logarithmic sampling takes data at logarithmically incremented intervals  resulting in equally spaced data  points when plotted as a function of logarithmically scaled time  the number of points taken is inversely  proportional to zone time   Linear sampling takes data at linearly incremented intervals  resulting in equally  spaced data points when plotted as a function of linearly scaled time  This technique is useful for relatively  short zone times  where linear time scaling is practical  The estimated viscosity value is used to modify the  gain term in the closed loop control algorithm  which adjusts the motor s rotational rate to generate the  desired stress level     Stress Ramp Test ki E3    Temperature         25 0 PC  Max 600 0  C Min  150  0    sampling Mode  Log    Linear    Paints Per Zone  200    Max 250 Min 20       Stress Limits  Pa     Max  tb3929  42 Min  6  392942    Estimated Viscosity        i coon   Pers        one Number 1 2 3 4    Final Stress            Palfiooo foo foo foo  Zone Time  s or homes   300 a COM CI          Options      La  omis   ndortest_   Saveds   tee   coce    Figure 3 23  Stress Ramp Test Set Up Screen    suggested Use    Stress Ramp is used to determine material hysteresis by deforming the material by linearly accelerating or  decelerating the shear stress command  The test is similar to the thixo
121. ed ooo    f No       ves       Help   Cancel      Figure 4 3  Instrument Configuration Function    Instrument Options   Set the   Stepper or Linear Motor  and  Remote Gap Monitoring    options  to yes to enable the Gap Control Panel functions     ARES User Manual 159    Test Tool Installation and Sample Loading    General Guidelines    The next several sections contain specific procedures to install the test tools used in each geometry  Specific  sample loading guidelines are given for each tool  However  the following general sample loading guidelines  pertain to all tools     e Make sure the tools are clean and free from damage    e Install the tools correctly as described  In general the tools should be easy to install  Having to use undue  force is a sign that something is wrong  Stop and ascertain the problem before continuing    e Make sure the tools align with each other properly and that the Gap is properly zeroed    e When using the oven close the oven door carefully to ensure that the sample or tool does not interfere  with the oven door     While loading the sample onto the tool  ensure that the sample is centered  as well as is possible  between the  tool mounting surfaces  Off center loading may cause misalignment of the transducer motor shafts   Additionally  misalignment may also affect the accuracy of the data  If misalignment does occur  the sample  should be removed from the tool and carefully reloaded     When testing at temperatures below ambient temperatu
122. ee errr Torre 6  y en  CON UB UO  np 8  e o     O 10  Kequiea axe iae   apoco rat  n loo aio 10  ANCL N UIC So dono oro 10  iae Gr enn PA CIS APP In e E E AE E A EAEE ET 13  Resulatory E e E E A ica 14  E T e E E E E A OEE 14  Electromagnetic Compatibility Standards       concnicnncnncnnnnnoncnnnnnaninonacanoncannannnonncnconannononnonccnncnonan cr nannc nac nn nan ncncns 14  Technical 16   0  POR paa eer ene eee ree ae 14  Lable alg OIE S reer Me TTT NT Ree A ii cad 15  Ch  pler e Ntro dic austin imine anisinn ha arenes 23    AL Y LE POCA O nn o E T IE slg val pda urns cena saad uae gn se ET A ben eo oem canes ners 23  Pesci nonoke N a aos 23  DVN WOM q CA o OE OO EEE mm 23  la APP Cen eee ee ee ee 24  Pii ee a A Teen Over or rene ete E A A nr A ET Tere eer Ree rer oer rrr ttn errr er rrr rt err re 24  Force Rebalance Transducer with Normal Force  FRTN1          cee cccesscccesesseeeeeseseeeeeeeaeeeeeseseeeeseeeeeeesaes 24  Standard Transducer  STD   2ss ccsapsonacsannsasonssasussonsatand neekin r oran E insi nieki 24  Environmental ControboOpuodns traes titi 24  earar de a aD e E T E E E A E E E 26  CI E p ads 26  Aa 07 a E E E E E E E E E A E E T 26  Tran de e A E E E E E S 26  FE POU aE oo Ea een oe EE A 27  See E O E O O E 28  Motor Tr OL aCe OPECIN aOR S  Ns 30  Transducer Operant Speci nea GONS ai 31  Chapter 2  Instrument Components   Identification and Operation           e seesssesseosseesseessossseeseoessesssessseesesssseesess 33  Component Identification and Place
123. een    Test Options  The following test options can be selected for use with the temperature ramp     Steady PreShear  Delay Before Test  AutoTension  Analog Data Input  Measurement Options  o Delay Settings  o Strain Amplitude Control    ARES User Manual    Frequency Temperature Sweep    Functional Description    The Frequency   Temperature Sweep takes successive measurements over a range of selected frequencies at a  series of constant temperature  steps   At each temperature step  a frequency sweep is run  while the  temperature is held constant   The temperature is then changed to the next step  allowed to equilibrate  and  the frequency sweep is run again     In a frequency  temperature sweep  entry fields for strain value and initial temperature are displayed on the  host computer  Figure 3 10   You can select either a linear  logarithmic  or discrete frequency sweep then enter  the frequency parameters  Thermal step size and soak time are also selectable     Frequency Temperature Sweep Test    i x     Strain  0 1    Maz 1 2 5000 Min 0 003125    Sweep Mode               cy Log O Discrete 0 Linear  Initial Frequency           0 1  rads  Max 500 0 Min 1 00e 05    Final Frequency           fi 00 0  rads     Points Per Decade       5    Max 500       Initial Temp            25 0 PC  Max 600 0  C Min  150  0      Final Temp             35 0  E   Temp  Increment  2 0  PC   Soak Time            fi  00  e or him        Options  PreShear Of Delay Of AutoTens Off Analogin Of Meas
124. enerates frequencies of 10   20  30  40  and 50 rad sec  One data point is measured at each of the frequencies     Discrete Frequency Sweep  The discrete frequency sweep takes a measurement at each of up to ten selected frequencies  Discrete    frequency sweeps can be run in any order  and can begin or end at any frequency within the range of the  instrument     ARES User Manual 103    Dynamic Frequency Sweep Test  Strain Control        Figure 3 4  Frequency Sweep Test Set up Screen    Suggested Use    The frequency sweep is used to analyze frequency and time dependent behavior of samples  In general  high  frequencies correspond to short time scales  and low frequencies correspond to long time scales     Test Options  The following test options can be selected for use with the frequency sweep     e Steady PreShear   e Delay Before Test   e Analog Data Input   e AutoStrain Adjustment   e Correlation  Measurement  Options    104 ARES User Manual    Dynamic Temperature Step    Functional Description    The Temperature Step test takes successive measurements at selected temperatures while holding a constant  frequency and strain     Temperature is automatically incremented or decremented from selected lower and upper temperature limits  by user selected steps  A unique set of test conditions can be entered in up to four temperature ranges  or  zones  A selected thermal  soak time  at each temperature ensures temperature equilibrium prior to  measurement     When setting up a tem
125. ent of TA  Instruments  Each licensed program shall remain the exclusive property of TA Instruments  and no rights or  licenses are granted to the purchaser other than as specified above     ARES User Manual    TA Instruments End User License Agreement    This License Agreement is your proof of license  Please treat it as valuable property     IMPORTANT   READ CAREFULLY  This is a legal agreement between you  either an individual or an entity   and TA Instruments for this TA Instruments instrument control and or data analysis software product  By  breaking the seal on your disk CD ROM package and or installing  copying or otherwise using this  SOFTWARE  you agree to be bound by the terms and conditions of this agreement  Failure to comply with the  terms of this license terminates the agreement and your rights to use this software     TA Instruments Software License    1  GRANT OF LICENSE  This TA Instruments End User Software License Agreement  LICENSE  grants  you the following rights    a  Instrument Control Software  This LICENSE permits you to use one copy of the specified version of  the TA Instruments instrument control software  provided this software is in use on only one  computer at any time  If you have multiple LICENSES for this software  then at any time you may  have as many copies of this software in use as you have LICENSES  If the anticipated number of users  of this software will exceed the number of applicable LICENSES  then you must have a reasonable  mec
126. ents delivered before January 2002 should have the older locking pin alignment  When  locking and unlocking the bearings on these instruments  the directions described above should be reversed as  shown in Figure 2 9     If you are unsure as to which locking orientation your system has  the following procedure will help you  determine it     1  Turn off the power to instrument  Ensure that air is supplied to the transducer    2  Holding the upper tool mount  manually try to move the transducer both axially  up and down  and  rotationally    3  Move the bearing lock to the opposite position and try to move the transducer again     Note how each orientation feels as you try to move the transducer   In the locked position  there will be no movement axially or rotationally  In the unlocked position  you should    feel some movement in both directions  Make a note in your manual as to the locked and unlocked positions  for future reference     ARES User Manual    F LEFT CLAMP 2K FRTN1  or  2K FRTN1E        RIGHT CLAMP         __ BALL PLUNGER  Factory Set    ONE EACH SIDE     LOCKED POSITION       UNLOCKED POSITION    Bearing Lock orientation for  ARES SHIPPED BEFORE JANUARY 2002               A   po LEFT CLAMP  2K FRTN1 i  or  ZK FRINIE RRR BERS See MES RIGHT CLAMP    E E BALL PLUNGER              Factory Set   i ee M   Oi E ai  5    Pin      a      Ome Each SIDE        i    a   t    UNLOCKED POSITION    Bearing Lock orientation for  ARES SHIPPED AFTER JANUARY 2002       LOCKED PO
127. er Manual    Double Wall Couette    Strain Constant Stress Constant   1000   G    qe 2 2  27 L  R    Rs       Variables    Ge   Gravitational constant    980 7 cgs  or 98 07 Sl     L   Length of bob  mm        R1 R2 R3 R4 equal the following radii   BOB  gt    CUP                                   lt  gt                                                                  Options  CUP BOB  34 mm OD  27 95 ID 32 mm OD  29 5 ID       Enviromental Systems  Fluid Bath  Fluid Bath 2   Double Wall Couette Tool      see Chapter 2 for more details regarding lower tool       General Information    The double wall Couette tool is used for testing lower viscosity fluids that would not generate enough torque  using parallel plates  It is also used where small sample volumes are necessary     ARES User Manual 195    Operating Ranges    Operating range is defined as the region bounded by the maximum and minimum complex modulus G that  can be measured by each transducer type using the double wall Couette geometry  The following geometry   specific factors affect the operating range of the double wall Couette    e Inner diameter of cup  R   strain constant K    e Outer diameter of cup  R   strain constant K    e Inner diameter of bob  R   stress constant K and strain constant K    e Outer diameter of bob  R   stress constant K and strain constant K    e Length of bob  stress constant K_     Additionally  the following instrument specific factors affect the operating range of all geometries   e M
128. er limits of operation is the range of complex viscosity that can be  tested  Appendix 1 contains tables of G  values for some tool combinations  transducers and a standard motor     Tool Installation  Install a tool as follows     1  Select the Set Gap Instrument Control function under the Control menu in Orchestrator  Use the   Send to Top  button to raise the stage to the loading position    2  Verify that the motor is on then mount the upper and lower tools on the actuator shafts    3  Using the Set Gap Instrument Control function in Orchestrator  zero the normal force and torque on  the motor using the  Offset Normal Force to Zero  and  Offset torque to Zero  buttons    4  Using the stepper control buttons  lower the stage to a point where the tools are close but not  touching    5  Use the  Zero Fixture  button in the Set Gap Instrument Control function to determine the zero point  for the test tools    6  Raise the stage to provide sufficient room for sample loading     Gap and Cone Angle    The actual gap setting and cone angle for each cone and plate tool is stated on the Certificate of Calibration that  is included in the tool kit  If no Certificate of Calibration is available  contact our Technical Service department     ARES User Manual    Testing at Other than Ambient Temperatures    Use the standard cone and plate tools for isothermal testing only  Transitions to elevated temperatures cause  expansion of the tool  For example  with a 50 micron nominal gap settin
129. erations such as sample loading  The  movement of the stage is defined by the following modes of operation     1  Step   When stepped  the stage moves very slowly  in single steps of the stage motor   To step the stage downward  press and hold the bottom button   To step the stage upward  press and hold the top button    2  Slew   When slewed  the stage moves very quickly and smoothly   To slew the stage downward  press and hold both center and bottom buttons   To slew the stage upward  press and hold both center and top buttons     ARES User Manual    Software Stage Control    Movement of the stage can also be controlled from the Set Gap Instrument Control function in the  Orchestrator software  Click the  Set Gap  button  Figure 2 4  located on the tool bar  The Set Gap Instrument  Control function can also be opened from the Control pull down menu  The Set Gap Instrument Control  function has displays that show Torque  Normal Force and the current Gap  In addition there are several  buttons that control instrument features  The  Send to Top  button can be used to move the stage all the way  to the top of its motion  The  Set Gap  button can be used to move the stage to the entered commanded gap   The force used to move the tools together while setting the gap will not exceed the value entered in the  Max  Allowed Force  box  The  Zero Fixture  button is used to bring the tools together to establish a zero gap  reference  from which the gap is then set     EE   Set Gap Inst
130. erative that a mixture of  50  ethylene glycol   50  water be used in the bath  Alternatively  Julabo   Thermal H10S can be used  Using  pure water is not recommended  as it will significantly reduce the lifetime of the seals  In the event of a seal  failure  the bath must be returned to TA Instruments for service and repair     Table 2 6  Circulator Operating Ranges using Various Bath Fluids    Desired Operating Range of Circulator Fluid   bath temperature range is slightly less      5  C to  100  C 50  ethylene glycol   50  water     40  C to  110  C 100  ethylene glycol   20  C to  150  C Julabo   Thermal H10S    Fluid Bath 2 Configuration in Orchestrator          When you command a temperature  Orchestrator software uses the Instrument Configuration to determine the  environmental system currently in use  and establish operating conditions  Prior to operating the fluid bath   access the Instrument Configuration function  located under the Service function of the Utilities menu  and set  up the instrument using the guidelines shown in Figure 2 25     Setup Instrument Options fx     Instrument Testing Limits    Intr  mem S MUD eeaeee neeaae aeaee etene    Temperature Control      Temperabure Conto ee   Bath  Instrument Controlled Only       Ma aa fi 50 0 fe    MEMO GMa terete eee A   20 0 E   Temperature Loop Control                          Ji pe       Temp Calibration Table                                Default C Adjustable       Cancel      Figure 2 25  Setup Instrument
131. es in the opposite direction     3  The data are averaged to supply the single data point that is reported by Orchestrator     Using two directions per measurement can give more accurate results at low rates     ARES User Manual    Steady Single Point Test       Figure 3 12  Steady Single Point Test Set Up Screen    Suggested Uses   Steady Single Point can be used to determine the following    e Unknown range response limitations of a new sample material  e Parameters for the design of new tests    e Time required for a sample material to reach steady state at a given shear rate when torque  stress signal  is manually monitored  e g  with an external strip chart recorder     Options  The following test options are available for use with Steady Single Point     e Delay Before Test  e Analog Data Input    ARES User Manual 117    Steady Rate Sweep    Functional Description    Steady Rate Sweep applies varying magnitudes of steady shear deformation  the magnitude of each  deformation depending on selected shear rates  Figure 3 13   Shear rates can be generated as follows     Logarithmic   The logarithmic rate sweep commands rates that are logarithmically incremented  resulting in equally spaced  data points when plotted as a function of logarithmically scaled shear rate  in reciprocal seconds  1 s   Shear  rates are selected by specifying initial and final rates  and the number of data points to measure between each  decade of rate    As an example  consider a sweep conducted ove
132. escribed in the Polycold documentation  included with your Chiller  Upon receiving the instrument  you should note and record the pressure reading  on the suction gauge  This gauge is located on the opposite side of the refrigeration unit as the air dryer  air  filter and J box  The Chiller should be off for 24 hours before taking a balance pressure reading  The initial  reading should be over 100 psig  Subsequent reading should be within 7 psig of the original value  If the  balance pressure drops more than 7 psig  it would indicate a leak in the refrigeration system     ARES User Manual 59      NOTE  0  Polycold suction gauge  and on off switch    are located on this panel      l    INPUT TO  REFRIGERATION UNIT             COLD AIR OUT  PRESSURE REGULATOR   ADJUST TO 60 PSI   AIR DRYER  AIR INPUT  SEPARATOR  ah rt AIR FILTERS  a ia  TO  CUSTOMER  FURNISHED  a  AIR FLOW oe  Rea CONTROLLER  o  CONDENSER 1  i ON OFF    GRILL SWITCH  I I  J BOX SIGNAL CONNECTION    i  to TEST STATION I    LN2 CONNECTOR  i  AC POWER IN  i I  I I  i I    AC POWER IN Y     J BOX SIDE PANEL    Figure 2 17  Chiller  Showing Various Components    60 ARES User Manual    LN2 Controller    The LN2 controller extends the lower range of the oven to  150   C using liquid nitrogen  The LN2 controller is  connected between an external liquid nitrogen source and the Test Station  Figure 2 1   Controlled by the test  station through Orchestrator software  the LN2 controller consists of a Dewar flask  a solenoi
133. f the material s modulus and the  size of the sample  Therefore  discrepancies in sample dimensions will obviously affect the accuracy of test  results  Referring to the equations for each geometries strain and stress constants  note the dependence of  compliance on each of the dimensions  Due to this dependence  a small error in sample dimension may  propagate into a large error in compliance  and therefore modulus     Testing outside the Linear Region    Most of the tests available are designed to be conducted within the samples linear region  Conducting tests  outside the linear region may result in erratic or incorrect data  It is critical that for a given material the linear  region is determined first  and then subsequent test performed based on the findings  To determine a  materials linear region typically a Dynamic Strain Sweep is run as a preliminary test     152 ARES User Manual    Temperature Variations    Changes in sample temperature during measurement can have a significant effect on the measured dynamic  data  particularly in the loss component of stress  and in tan     As a result  temperature steps  where  temperature is held constant during a measurement  may yield better tan resolution when studying the  temperature dependence of a material  as opposed to a temperature ramp  where temperature is changing  during a measurement  Also  make sure that sufficient equilibration time is provided for the entire sample to  reach the test temperature     Some sample
134. g for a 50 mm cone and plate  geometry  thermal expansion can cause contact between tools  resulting in erroneous test data  The system  should be equilibrated at the desired working temperature and then the final gap set     For special applications  Invar tools can be used  Invar has a coefficient of thermal expansion that is 10 times  less than steel between 0 and 230   C  Outside this range  Invar s thermal expansion properties are similar to  steel s  If temperature ramps or sweeps must be performed using cone and plate tools  Invar must be used     sample Loading    1   ze  3     176    Ensure that the gap has been zeroed    Place the sample on the lower plate  Ensure that the sample is centered on the tool    Using the stepper motor buttons on the right side of the test station adjust the sample gap until the  upper plate is close to the specimen  Set the gap using the Set Gap Instrument Control function   entering the appropriate parameters  The initial gap should be set approximately   0 05 mm above the final gap to facilitate sample trimming  For softer materials  using the normal force  limits helps to avoid damaging the sample as the sample may be rapidly compressed in a manual  loading procedure  with insufficient time for sample relaxation    Trim the sample flush with the edges of the plates    Lower the stage to achieve the gap specified on the Certificate of Calibration  Again  this can be done  manually  or using the Set Gap Instrument Control function  chan
135. ging the entered gap to the  specified final value  The sample should bulge slightly as shown in Figure 4 9    Allow samples to relax before beginning testing by monitoring the force and waiting for it to decay to  close to zero    Enter the sample dimension in the Orchestrator Test Geometry screen  Note that if the tools were  correctly zeroed the actual sample gap can either be read from the Gap real time parameter  or  measured automatically at the start of the test by selecting the  Read Test Fixture Gap  check box in  the Geometry screen        UPPER FIXTURE      SAMPLE    LOWER FIXTURE    Figure 4 9  Cone and Plate with Sample Loaded    ARES User Manual    Torsion Rectangular  New Design    Strain Constant Stress Constant    K  TOTO COW   TT J    WAL    3  18      K   1000    WT    Variables    Gc   Gravitational constant    980 7 cgs  or 98 07 Sl     T   Thickness of sample  mm    W   Width of sample  mm    L   Length of sample  mm   Options    Clamps to accommodate thicknesses up to 6 5 mm    Environmental Systems    Ambient Torsion Rectangular Tool  Oven  new design        General Information    The re designed Torsion Rectangular tool is used for testing solid materials with high modulus  including  thermosets  thermoplastics and elastomers  The sample is held in tension between the upper and lower tool   Three setting anvils  Figure 4 10  are provided to accommodate samples of varying thickness  Each setting  anvil is designed to provide clamping for two different 
136. gle RomtMeasure enana 101  FCO IND Eee o e EA e En o UU oo O E E E N eae  101  U Usas 101  TOPON PAP     E E E E E S 101  Byna E N O a a E E A E I E EE re rer eee 102  Funciona Dese O a E E E R E E 102  Seea O E E T eRe E EE E E ner ere een ae eee eer 102  MS NO a E A A ES o A AOE E E  102  FOC SIN E E A o A cesar conn eceeosaueacsuace cece yerrameradeas 103  Funciona I DOSCH AON oa EEE A A 103  pro io UL AAA 5 o In 104  iO     a  ahs AP o Pe A O eres 104  aria io a qshabia do 1    OR ER A o ce ee TE 105  Functional Descritores 105  er O AP Ei o A e doneaacenamnnea eaeenaueupye  106  MSS ori o o gts 0  o nn o no E O T N E ETE 106  Basa tea eo q 10 ro y e  lt a OOO PA Un e A 107  Funciona TICS CHOI O I nta r R E EEE A 107  Sue 24 chet  216 Usos E ee ere 108  ici A o E     E eee 108  Dynamic Temperature Raimi p gt  Lest sisiisstasvsansnisvincaasansisnansacvasiasiansdiaansasueieiiiwanesabiagasesenasnanndyermaaslsinwetaansabacaaanens 109  PU COMA Descrip o AAA o oe stcgestipenses scaenatvacauaneiants 109   pe  ORS 24 etc   S16 il Okc  PO OR o eC eee o o O oe et ere ar ere eer Tee 110  MSS  INO A o o o O eo masa siagetatoaecounebanses 111  MultiWave Single Point  MultiWave Temperature Ramp  occoooonncnnccncnioninnnonanncnnanininannanccnnancnnn non nnnnanncnnanannnnso 112  Funcion WB ccs git OE a eceener re eee treme ne cer err ty eter ren rea rere eee ee reer 112  TES COPIE pcia 114  Frequency   Temperature Sweet 115  tiale corn Sei 0 510  q E OPE OP UE E E O A E A N A E 115  oe a
137. h that  it points toward the left front corner of the instrument  However  some 2K FRTN1 transducers may be  equipped with a pin that is rotated slightly from this position  The bearing locking and unlocking instructions  for these transducers directs you to push the pin toward the left or right  Regardless of the orientation of your  pin     left    means toward the left hemisphere of the transducer  as viewed facing the front of the instrument   On low range transducers  100 FRTN1  200 FRTN1 and 1K FRIN1  the pin is installed so it points toward the  front of the transducer housing  as shown in the lower half of Figure 2 8     LEFT CLAMP        RIGHT CLAMP    BALL PLUNGER  Factory Set    ONE EsncH SIDE   PIN    UNLOCKED POSITION    2K FRTN1  or  2K FRTNIE       LOCKED POSITION    LOCKED POSITION  a UNLOCKED POSITION         FRONT CLAMP   Rear CLAMP IS  ON OPPOSITE S 0        BALL PLUNGER 100 FRTN1   Factory Set  or  nents  200 FRTN1    Figure 2 8  Bearing Locks  2K FRTN1 FRTN1E and 100 200 FRTN1     The view shown for the 2K FRTN1 transducers is looking from the front of the instrument  The lock  on the 1K FRTN1 is similar to that shown for the 100 200 FRTN1 transducer     ARES User Manual    Procedure for Locking and Unlocking Air Bearings  FRTN1 and FRTN1E    Refer to Figure 2 8 while performing the following procedures     To Lock the FRTN1 air bearing     1  Read the Caution on page 11   2  Ensure that instrument power is off and air is applied to the transducer   3  Do
138. hanism or process in place to assure that the number of persons using this software concurrently  does not exceed the number of LICENSES     You may also store or install a copy of the TA Instruments instrument control software on a storage  device  such as a network server  used only to install or run this software on your other computers  over an internal network  however  you must acquire and dedicate a LICENSE for each separate  computer on which this software is installed or run from the storage device  A LICENSE for this  software may not be shared or used concurrently on different computers     b  Data Analysis Software  TA Instruments Thermal and Rheology data analysis software is not  restricted to use on only one computer  These programs may be installed on multiple computers for  use provided it is for analysis of data generated by the LICENSED instrument control software     2  OWNERSHIP RIGHTS  The software described above is owned by TA Instruments or its suppliers and is  protected by United States copyright laws and international treaty provisions  TA Instruments and its  suppliers own and retain all right  title and interest in and to this software  including all copyrights   patents  trade secret rights  trademarks and other intellectual property rights therein  Your possession   installation  or use of this software does not transfer to you any title to the intellectual property in this  software  and you will not acquire any rights to this software except 
139. he following     C for G  MAXIMUM C for G  MINIMUM    2K FRTN1    2K FRTN1E i J 1 15e 06  rad gecm  TA  see note below     1K FRTN1 l J   4 9 e 06  see note below     2K STD  10K STD   0  J   2 60e 06  see note below     100 FRT for 0 100  J   2 60e 05  200 FRT    100 ERTN1 for 0 10  J  2 60e 06 E  200 FRTN1 a  see note below     for w lt 10   see note below     NOTE  The values for M  gecm  and 6  rad  are found in the  specification tables in Chapter 1   Pick the correct values for  your specific transducer and motor combination        ARES User Manual    To determine the maximum or minimum complex viscosity N  that can be measured at a given frequency  use  the following formula     nea  4 2   q    where   n   Complex viscosity  Poise     G   Complex Modulus  dynes  cm             Frequency  radians second     Using a spreadsheet application such as Microsoft   Excel     you can use the equations above to plot the range  of complex viscosity that can be tested for a given geometry  transducer combination as follows   1  Calculate G  MAXIMUM and G  MINIMUM using equation 4 1     2  Substitute the G  MAXIMUM value into equation 4 2  then determine the upper limit of operation by    calculating N  at each O  0 values chosen to be from the lowest to highest frequencies within the  transducer operating range      3  Substitute the G  MINIMUM value into equation 4 2  then determine the lower limit of operation by  calculating N  at each        4  Generate an X Y scatter plot of 
140. he heaters must be evaporated liquid nitrogen   LN   supplied by the optional cryogenic LN2 Controller or very cold gas supplied by the optional Chiller        Upper  Gun Heater  ata fe Upper  Lower Oven PRT  Oven PRT  Lower  Gun Heater       Figure 2 16  Oven Chamber Showing Gun Heaters and PRTs     Oven Temperature Control    Oven temperature is maintained by a control loop that is closed around a Platinum Resistance Thermometer   PRT   The ARES has two control loops  Oven PRTs  to minimize vertical temperature gradients  each of which  is located in front of the respective gun heater being monitored  The control system determines actual oven  temperature by supplying a constant current to the PRT and measuring the voltage developed across it  The  difference between commanded and actual oven temperature is continuously driven to minimum     You can choose to control the temperature of either the oven environment  or in some cases  the lower tool   The choice is made using Orchestrator software  The temperature of the lower tool can be monitored using the  lower tool PRT  which is mounted on the motor  Installation of the lower tool PRT is covered as part of the  lower tool installation in Chapter 3  Under normal operating conditions the oven PRTs are used to control  oven temperature and the tool PRT is used to report sample temperature  However  the tool PRT can be used  to control the oven temperature as well     Figure 2 16 shows the inside of the oven chamber and the l
141. he measurement by clicking either the  Toggle Steady Measure  button  Figure 3 11   on the toolbar or the Toggle Steady Measure function  accessed from the Control pull down menu    e When desired  stop the measurement by clicking either the  Toggle Steady Measure  button on the toolbar  or the Toggle Steady Measure function     Direction    For positive Rate values  Direction specifies the rotational direction of the actuator at the first commanded  shear rate  Selecting Directions per Measurements of  One  results in data collection while the actuator rotates  in the specified direction  Selecting Directions Per Measurement of  Two  results in the following sequence of  events     1  Data are collected while the actuator rotates in the specified direction  2  Data are collected while the actuator rotates in the opposite direction   3  The data are averaged to supply the single data point that is reported by Orchestrator     Using two directions per measurement is necessary if the transducer changes range during the test   Steady Rate Sweep Test El Ei    Temperature  25 0 PC  Mas 600 0  C Min  150  0      Sweep Mode           icy Log f Discrete  Initial Rate               f 0  123  Maz 525  0000 Min 0  006250    Final Rate                fi 00 0  ls     Points Per Decade E    Mas 500       Data Collection Mode              amp  Time Based    Manual    Delay Before Measure             20      Measurement Time                  30  z     Director ooo  eee eres   Clockwise C Cou
142. heaters in the oven  It is very important that good quality air is supplied to the ARES test station at all times  during use  otherwise significant damage to the instrument can result     The optional N2 GAS input is available for using other gases as the oven heating medium when the sample  being tested would react with normal air  Nitrogen is commonly used in this case  Use the Gas Supply to Oven  Selector Switch  black knob below gauges  to choose which gas input port is used to supply gas to the oven     Gas Pressure Specifications    Air should be supplied to the ARES at 5 5 scfm  156 1 min  at a pressure of 80 psi  5 5 bar   Gas connected  through the N2 GAS port should be supplied at 60 to 70 psi  4 8 bar   The gas supplied to the various  components is regulated using the Air Pressure Adjust Knobs on the backside of the panel  Pull the knob out  and rotate it clockwise to increase the pressure to the desired component  Pressure is monitored using the  associated gauges on the side of the Pneumatics panel     Pressure to the various components should be adjusted as follows     REQUIRED    COMPONENT PRESSURE    Transducer 35 psi    40 psi   with oven on     Oven       ARES User Manual    Air Quality Specifications    It is critical that high quality  clean dry air is supplied to the test station at all times  Any particles present in  the air must be smaller than 5 microns  Since the motor and transducer use air bearings  larger particles in the  air can easily damage 
143. her Trademarks  continued     Swagelok   is a registered trademark of the Swagelok Company   Inconel   is a registered trademark of Inco Alloys Special Metals   X acto   is a registered trademark of Hunt Corporation    TYGON    is a registered trademark of NORTON Co     TA Instruments Q Series modules contain proprietary embedded operating system software copyrighted by  Mentor Graphics     SILICON SOFTWARE    1989 97 Mentor Graphics Corporation  Microtec Division  All rights reserved  Unpublished rights reserved under the copyright laws of  the United States     RESTRICTED RIGHTS LEGEND  Use duplication  or disclosure by the U S  Government or a U S  Government subcontractor is subject to the restrictions set forth in the  license agreement provided with the Software pursuant to DFARS 227 7202 3 a  or as set forth in subparagraph  c   1  and  2  of the  Commercial Computer Software Restricted Rights clause at FAR 52 227 19 as applicable     MENTOR GRAPHICS CORPORATION  MICROTEC DIVISION   880 RIDDER PARK DRIVE  SAN JOSE  CA 95131 2440    ARES User Manual    System Configuration    This table should be filled out during system installation with assistance from the TA Instruments  service  installation engineer  Please refer to the information below when contacting TA Instruments for  customer support or service     fnstument faes OO S  ICC In  Cres S S  mesoi  O  AA  INS IO EE  AA      O on oo S   soionmental Control  Tag Bata  Peltier   EN2conroer a   chile oo S S Cds   Crua
144. ideal solid  Hooke s law for shear deformations states    that if a shear strain T is applied to an ideal solid  a shear stress Y develops in the material in direct proportion  to the strain     T GY    The proportionality constant in equations for shear  G  is known as the modulus of the material  The modulus  of a material is a measure of its stiffness  or ability to resist deformation  Linear stress strain behavior is  characterized by the modulus remaining unchanged as strain is varied  The region where the modulus  remains constant as strain is changed is called the linear region  The linear region is also called the Hookean  region of the material     Newton s law describes the mechanical behavior of an ideal viscous fluid  When a fluid moves by virtue of  being pushed through a pipe  or dragged through a screw in an extruder  etc   the movement is termed shear     Newton s law relates the shear stress  T  to the rate of strain  or shear rate  dy dt     T  nit where  N   coefficient of viscosity    A fluid is said to be Newtonian if the viscosity does not depend upon the strain rate  An analogous equation  can be written for tensile testing  where the tensile stress  0  is related to the tensile strain  e  by     ARES User Manual    sa  dt    In a non Newtonian fluid  the viscosity is not constant  but is a function of strain rate  Many polymer  solutions are non Newtonian in behavior because their viscosity decreases as shear rate is increased  This is  called shear th
145. ighly corrosive or  noxious materials  such as sulfuric acid  The tool is made form Hastelloy  which is highly resistant to  corrosion  The tool is designed with a sealed glass chamber surrounding the sample  The top tool and cover  are designed to provide a  liquid seal  for the top  After the sample is loaded  the cover is put in place  and an  appropriate liquid is poured into the  well  on the back of the top plate  This seals the chamber  containing  any sample fumes during testing     Operating Ranges    Operating range is defined as the region bounded by the maximum and minimum complex modulus G that  can be measured by each transducer type using the parallel plate geometry  The following geometry specific  factors affect the operating range of parallel plate geometry     e Plate diameter  strain constant K  and stress constant Kx   e Gap between plates  strain constant K       e The viscosity of the sealing fluid may also affect the operating range     ARES User Manual 165    Additionally  the following instrument specific factors affect the operating range of all geometries     e Minimum torque that can be measured by the transducer  e Transducer compliance  e Maximum strain that can be generated by the motor     To calculate the minimum and maximum G that can be measured by each transducer type using the parallel  plate geometry  use the following equation     i K lc  4 1   OTK    Y     where     Kz   Stress Constant  K    Strain Constant    and C is computed from t
146. iia a endesa etc 51  Mant Ower WC AAA T POE o II N 51  Enecumaties Tinei ri EE S ii EA 53  E T anes E N A E A T A A E E T E 53  Cas Prenu cae  9 216   602 MON Sansa 53  Ae ahy OCC iga ON e n E E E N 54  Focumauc k onnec hona cracca aaa 54  Environmental Control Systends usais  57  a E EE E A E A AO 58  Oven Temperature COn Ora 58  l E E E E EE T T T E I 59  CS a E Mm o y ene eee ae 59  S LE TT L E T E EE E T E O E ee ee 59  AO o E CAPA A TT A A T E E O E 59  MAIN E EE ta aa E E N E E E 59  EN OTO E na E ENE O A T T E T are ren 61  Dew T E epa n rasta vd E E IE E A E E E EE E EEE oncia 61  CO O E E T E E E 61  OBES  CONC Ura LION spinanie E EE E E 61  ENSEon tolera 61  Mie INC    sitio dpto 62  RC CUO IO OPIO E E E T E A E T 62  Oven Operatine Requirements surrender R E A 63  Oven Configuration in Orchestrator     sccssssscsssssssssesssssesssscssssssssnscsssscsssscsssnsessnecssssesssneesssecsssesssecessneessneesseeee 64    ARES User Manual    Si o UI A 65    dio berto oo lolo  Lars 1 101 N E E O A E E RT 66  FNC Me ke ia a e A oo o E E A A E A a AR 67  Descrip MOM AAA     EAEE E E E E EOE A E 67  lstallation of Fhud oy  A A oO E Ena an rr 67  Circula lor COMMS CHIOIS sario tii 70  el Bal bg Yo  A O PORO E E eee er ener eee err re 70  Lower Tool installation and  Removal persia ina 71  iras o E o     71  TM  Elia bat 2Conteuration ln Orchestrator ctricos pere a EN AREA aereas Te  PID Loop Cp PEPPER 73  Pid Ban Oporne Requirements oi 74  Pe Tani z Or Oi paa 74  Poe ae tc lee RN 
147. ilities pull down menu  to reflect the correct COM port  You must tell the software  the correct COM port you are using in order for communication with the Test Station to take place  If you have  a serial mouse  more than likely it is using COM1  and you will need to use another serial port  COM2 for  example  to communicate with the test station     The oven position and temperature sensors are connected to the test station through the OVEN connector  The  optional LN2 Controller is interfaced to the LN2 connector if cooling option is required  If the Fluid Bath is  used connect the RS 232 output from the circulator to the CIRCULATOR connector     Many of the connectors are not used under standard operation of the ARES  However  they are available for  more customized applications when needed  Examples might be to plot applied strain on an external strip  chart recorder  or to input an external voltage to be digitized with the test data  Table 2 1 contains more  detailed information regarding these additional connectors        Figure 2 12  Test Station Signal Panel    ARES User Manual    Table 2 1  Signal Panel Connectors    TORQUE OUT Outputs a DE voltage that is proportional to transducer output  torque    Scaling is  Q VDC   0 g cm torque   5 VDC     full scale torque    As the instrument is shipped  TORQUE IN is connected to  and outputs the same  TORQUE IN signal as  TORQUE OUT  This connector is normally used by TA service  personel for diagnostic purposes     Dynamic
148. ill result in rapid  evaporation of fluid trapped in the Peltier Assembly hoses  causing  hose rupture  damage to the Peltier Assembly  and possible personal    injury     ES WARNING  Do not operate the Peltier system unless fluid is being       Peltier Operation    Turn on the Circulator and set it to the desired temperature  The Peltier is operated using the Instrument  Control Panel  Figure 2 33   Set the environmental control option to On  The desired temperature is set in the  Temperature input field     Humidity Cover    After you have loaded the sample  you may install the two piece Humidity Cover  Figure 2 34  by simply  placing each half of the Cover onto the top of the Peltier Assembly and pressing the halves together  The  Humidity Cover helps to create a thermally isolated chamber  thus insuring that the Peltier specifications are  met  It also prevents sample evaporation during testing  The  pads  on the inside of the cover should be  saturated with sample before each use to minimize evaporation     WARNING  During and after testing at other than ambient   x  temperature  the sample test surface can reach temperatures that cause  personal injury  Allow the Peltier System to return to ambient  temperature BEFORE you touch the sample test surface or the sample     ARES User Manual    Instrument Control Panel    Figure 2 33     p50  Peltier  T hermopile heat pump         OF       Instrument Control Panel Showing Peltier Control Options       Figure 2 34  Humidity Co
149. inimum torque that can be measured by the transducer   e Transducer compliance   e Maximum strain that can be generated by the motor    To calculate the minimum and maximum G that can be measured by each transducer type using the double  wall Couette  use the following equation       Ks lc  4 1   ae    e   where K    Stress Constant  K    Strain Constant    and C is computed from the following     C for G  MAXIMUM C for G  MINIMUM     see note below     Transducer    2K FRTN1  2K FRIN1E     see note below     1K FRTN1 Ce    0   2K STD These Ge are not recommended  10K STD for use with the Double Wall Couette     100 FRT for   100  J   2 60e 05    200 FRT ae  100 FRTN1 for 0 10  J  2 60e 06    200 FRTN1   mm  see note below     NOTE  The values for M  gecm and 6  rad  are found in the specification tables in Chapter 1   Table 1 6 through Table 1 13   Pick the correct values for your specific transducer and  motor combination        ARES User Manual    To determine the maximum or minimum complex viscosity N  that can be measured at a given frequency  use  the following formula     nea  4 2   q    where n   Complex viscosity  Poise     G   Complex Modulus  dynes  cm           Frequency  rad sec     Using a spreadsheet application such as Microsoft   Excel     you can use the equations above to plot the range  of complex viscosity that can be tested for a given geometry  transducer combination as follows     1  Calculate G  MAXIMUM and G  MINIMUM using equation 4 1    2  Substitu
150. inning or pseudoplasticity  In the opposite effect  shear thickening or dilatancy  the viscosity  increases with increasing shear rate  This is seen in some concentrated aqueous dispersions of clays and  sands     Steady Shear Testing    Steady Shear Testing uses continuous rotation to apply strain  When a steady shear rate is reached  the shear  stress  T  is measured as a function of the shear rate  dy dt   The stress to shear rate ratio yields the steady    shear viscosity  N   Measurements are typically made over a wide range of shear rates to study the shear rate  dependence of the sample     Dynamic Mechanical Testing    Definition of Variables    Dynamic mechanical testing involves the application of an oscillatory strain to a sample  The resulting  sinusoidal stress is measured and correlated against the input strain  and the viscous and elastic properties of  the sample are simultaneously measured  If the sample behaves as an ideal elastic solid  then the resulting  stress is proportional to the strain amplitude  Hooke s Law   and the stress and strain signals are in phase  If  the sample behaves as an ideal fluid  then the stress is proportional to the strain rate  or the first derivative of  the strain  Newton s Law   In this case  the stress signal is out of phase with the strain  leading it by 90       For viscoelastic materials  the phase angle shift  6  between stress and strain occurs somewhere between the  elastic and viscous extremes  The stress signal gener
151. irculator pump outlet to the fluid bath inlet  This  clamp is critical to adjust the flow to the bath  If the clamp is not installed  or the flow not adjusted properly  it  is possible to overflow the Fluid Bath and cause significant damage to the test station motor     The clamp should be set by placing it on the inlet hose  circulator off  and completely closing the clamp as  tight as it will go  finger tight   Make an index mark on the clamp knob  and then using the index for  reference  open the clamp two to four complete turns  The number of turns the clamp is opened will depend  upon the make and model of circulator used  We have found that when using the Julabo FS 18   supplied with  new systems  the clamp should be opened four turns  For older NesLab chillers two turns was effective   However  these values should be taken as guidelines only  and you should monitor the flow  and bath  to  ensure that the flow rate to the bath is correct for their specific system     When the clamp is set correctly you can feel  by holding the outlet hose  strong fluid flow through the bath  outlet hose  it will pulsate somewhat   and the bath will stabilize at the desired temperature in about 30  minutes  If the clamp is closed too much  the flow through the bath outlet hose will feel weak or nonexistent   and the temperature will continuously oscillate without stabilizing  If the clamp is opened too much the bath  will eventually overflow  The bath should be monitored  and if any flui
152. ish playing  and the motor will stop  Data will be  collected for an additional 5 seconds in Zone 2  after which the test will end     Equation Syntax   Equations for Waveshape are entered as a function of time using the variable    t     Standard arithmetic  operators an be used                  as well as a variety of mathematical functions such as sin    cos    exp    etc   Standard rules for operator precedence are used  and parenthesis may be used to change the order of  evaluation  Table 3 5 contains a list of available functions     Test Options    e Delay Before Test  e Analog Data input    ARES User Manual    Table 3 5  Available functions for the Arbitrary Waveshape Test  Note  The arguments to  trigonometric functions are in radians     Operation    Exponentiation   Highest      Multiptication  1  7  Division fT     Addition fT    Subtraction Lowest    Functions     cosht   asin    Arccosine   atan     sat    abs         ARES User Manual    Thixotropic Loop  Rate Ramp     Description    In each of up to four zones  Thixotropic Loop commands a steady shear rate that linearly accelerates or  decelerates to a final shear rate  The time allotted to achieve the final shear rate is selectable in each zone   offering control of actuator acceleration  Figure 3 18      In each zone the following parameters are set   Final Shear Rate    Shear rate to which the initial rate is accelerated or decelerated  In the first zone  the initial rate is zero  In  subsequent zones  the i
153. l    ARES User Manual    Dynamic Temperature Step Test    Figure 3 26  AutoTension Adjustment Set Up Screen       Analog Data Input    Analog Data Input allows you to apply an external   10 VDC analog input signal to the Analog Input BNC   ANALOG 1 IN  receptacle on the Test Station Rear Panel  and measure  and record with the rest of the data     the input signal during any test  Data are sampled at 1 Hz  and only one measurement is recorded for each  data point     Apply an external   10 VDC input signal  then check the Analog Data Collection checkbox  Figure 3 27    During the test  Orchestrator reports the analog input as Analog Input Data  which can be displayed in the  spreadsheet  plot  or as an on line parameter  The reported values are normalized by dividing input voltage by  10  Thus  an input voltage of  10 volts is recorded in the data as  1 00        Figure 3 27  Analog Data Input Set Up Screen    144 ARES User Manual    AutoStrain    AutoStrain is used to prevent the dynamic force signal from going above or below the specified force limits of  the transducer  When using AutoStrain control  the actual sample strain may differ from that commanded     Following is a description of AutoStrain parameters that must be set  Figure 3 28    Maximum Applied Strain    This value represents the maximum allowed strain to be used in the AutoStrain adjustment  The strain  amplitude will never exceed this value irrespective of the measured force     Maximum Allowed Force    Whe
154. l  Each PRT is  designed to contact the lower tool  and provides the temperature of that tool     Note that the PRT used with the different Couettes is shorter than the PRT used with the Parallel Plate tool   Because of its length  the Couette PRT can be difficult to remove by hand  An extraction tool is provided to aid  in the removal of this PRT  When removing the short PRI  slip the tool over the PRT end and engage the slot  in the tool over the pin on the PRT as shown the detail of Figure 2 24  Pull the PRT straight up and out of the  bath  DO NOT rotate the PRT in its mount since damage to the connector can result     Install the lower tool PRT as follows     1  Ensure that the proper PRT is selected from the lower tool PRTs available for use with the ARES Fluid  Bath 2    2  Using very little force  place the PRT plug  keyed end  through the hole in the bath  and onto the  receptacle that is installed in the motor    3  Gently push down on the PRT and rotate it until the PRT slips into the receptacle  the keyed end of  the PRT then slips into the key slot machined into the receptacle   Seat the PRT completely into the  receptacle  When properly installed  the displayed temperature parameter should indicate ambient  temperature             BATH  HOUSING    S COLLAR     CROSS SECTION SHOWN FOR VISIBILITY     Figure 2 23  Fluid Bath Alignment Pin    ARES User Manual    COUETTE PRT  EXTRACTION TOOL    BATH 2 PARALLEL PLATE   lower fixture        COUETTE PRT     PN 700 03484
155. l Computer  e Motor and Transducer  e Environmental Control    Control Computer    The control computer synchronizes  generates and directs test instructions to  and processes raw data from  all  subsystems  The control computer central processing unit  CPU  commands and monitors the control  computer  maintaining communications via the bi directional BUS  The input output controller  1 0  controls  all hardware switching within the instrument  The analog to digital converter  A D  acquires analog signals  from the motor and transducer  scales the signals for optimum gain  and converts them to digital form for  processing by the CPU  The temperature controller  TEMP CONTROLLER  sends electronic commands to operate  the environmental control subsystem  TEMP CMD AND MEASURE  based upon your input  It also acquires and  processes temperature data  TEMP CONTROL AND FEEDBACK  for environmental control and display     The CPU and random access memory  RAM  contain memory devices programmed to execute test sequences   The RAM circuitry stores data received from all subsystems for further processing by the HOST COMPUTER  IBM    or 100  IBM  compatible PC   that communicates with the control computer through the RS 232C data  communications link     Motor    In dynamic mode  the Motor is configured as a position servo  In steady mode it is configured as a rate servo   Following the start of a test  the FUNCTION GENERATOR determines the waveform to be applied  COMMAND   The  MOTOR CONT
156. lation Delay  146  Couette  test tool   constants  190  tool installation  192  general information  190  operating range  190  sample loading  192    vane tool  193    Creep  126   D   Delay Before Test  141  Dewar Flask  61    Diagnostic LEDs  230    Dilatancy  94    ARES User Manual    Double Wall Couette  test tool    constants  195   tool installation  197  201   general information  195   operating range  196   sample loading  198  201  Dynamic Measurement Formulas  204  Dynamic Single Point Measurment  101  Dynamic Strain Sweep  100  107  Dynamic Temperature Ramp  109  110  Dynamic Temperature Step  105    Dynamic Time Sweep  100  102    E    E     See Elastic Modulus  E     See Viscous Modulus  E   See Complex Modulus  Elastic Modulus  E      94  Elastic Stress  t       94  Elasticity  93  Elastomers  149  Emulsions  149  Environmental Control System  configuring  64  description  57  principles of operation  27  specifications  30  Errors  discrepancies in sample geometry  152    temperature variations  153  testing outsidde linear region  152    ARES User Manual    F    Fluid Bath  circulator connections  87  description  85  inet line clamp  88  installation  85  lower tool  87  operating specifications  30  operation  92    Fluid Bath 2  circulator connections  70  description  67  installation  67  lower tools  71  operation  74  PRT installation  70    Force Gap test  134  Force Gap Test  134  Frequency Sweep  100  103  104    Frequency   Temperature Sweep  100 
157. le point measurement  see section on Testing  Options for further details      e Steady PreShear   e Delay Before Test   e Analog Data Input   e Correlation  Measurement  Options    ARES User Manual 101    Dynamic Time Sweep    Functional Description    The Time Sweep test takes successive measurements at constant temperature  frequency  and strain at a  selected interval     The time required to make a measurement is frequency dependent  At frequencies less than 2 0 rad sec  0 3  Hz  it is equal to the time required to complete approximately 1 5 cycles of oscillation  At high frequencies the  limit is approximately 2 second per measurement  Adjusting the Correlation Delay  see  Measurement  Options   can also have an effect on the time required to make a measurement     If a  Time Per Measure  is entered that is less than the amount of time required to complete the measurement   Figure 3 3   the control computer will take data points at its maximum speed  in accordance with frequency   If many points are to be taken over a short time period  a short time measurement  such as 2 0 sec  could be  used  This sets the instrument to take measurements as fast as possible     Dynamic Time Sweep Test ki El  Frequency fl 0 0  rads  Max 500 0 Min 1 00e 05  Ml  7 0    Max 372 5000 Min 0 0031 25  Temperature  25 0  CE  Max 600 0  C Min  150 0  C       Total Time                      fi  010  s or hems   Time Per Measurement f  s or Airs        Options PreShear Ot Delay  Off SutoTens  Of 
158. lease call the Thermal Analysis Help Desk at 1 302 427 4070   SERVICE   U S A    For instrument service and repairs  please call 1 302 427 4050     BELGIUM   LUXEMBOURG   TA Instruments a Division of Waters N V  S A   Raketstraat 60 Rue de la Fus  e   1130 Brussel   Bruxelles   Belgium   Phone  32 2 706 00 80   Fax  32 2 706 00 81    EUROPE   TA Instruments Ltd   Cleeve Road   Leatherhead  Surrey KT22 7UQ  United Kingdom   Phone  44 1372 360363   Fax  44 1372 360135    FRANCE   TA Instruments Division de Waters SA  1 3  Rue Jacques Monod   78280 Guyancourt   France   Phone  33 1 30 48 94 60   Fax  33 1 30 48 94 51    GERMANY   TA Instruments Germany  Max Planck Strasse 11  63755 ALZENAU  Germany   Phone  49 6023 96470  Fax  49 6023 964777    ARES User Manual    ITALY   Waters S p A    Via Achille Grandi  27   20090 Vimodrone  Milano   Italy  Phone  39 02 27421 283   Fax  39 02 250 1827    JAPAN   TA Instruments Japan   No  5 Koike Bldg    1 3 12 Kitashinagawa   Shinagawa Ku  Tokyo 140   Japan   Phone  813 5479 8418  Sales  amp  Application    Fax  813 5479 7488  Sales  amp  Application    Phone  813 3450 0981  For Service  amp  Accounting   Fax  813 3450 1322  For Service  amp  Accounting     THE NETHERLANDS   TA Instruments   A Division of Waters Chromatography bv  Postbus 379   Florijnstraat 19   4870 AJ Etten Leur   The Netherlands   Phone  31 76 508 72 70   Fax  31 76 508 72 80    SPAIN   Waters Cromatografia S A   Entenza 24 Planta Baja  08015 Barcelona   Spain   Phone  
159. lowing the guidelines  discussed in the section on the liquid seal     ARES User Manual 171    172    Enter the sample dimension in the Orchestrator Test Geometry screen  Note that if the tools were  correctly zeroed the actual sample gap can either be read from the Gap real time parameter  or measured  automatically at the start of the test by selecting the  Read Test Fixture Gap  check box in the Geometry  screen          FILL PORT     for sealing fluid  eee ees    UPPER a    OL   Ae SAMPLE    GLASS  CHAMBER    LOWER TOOL    Figure 4 8  Hastelloy Tool with Sample Loaded    ARES User Manual    Cone and Plate          Strain Constant Stress Constant  3000 G       A ta  2TR    Normal Stress Constant   First Normal Stress  Difference  _ 200 P    K  Z aR     Go N   K   F     Variables    Gc   Gravitational constant    980 7 cgs  or 98 07 Sl     R   Radius of plate  mm        B   Cone angle  rad     F    Normal force  g     Options  25  50 mm Diameter  0 02  0 04  0 1 rad Cone Angle  Disposable cones and plates  Invar cones and plates   Environmental Systems    Ambient  Oven  Fluid Bath  Fluid Bath 2      Peltier  Cone and Plate Tool         see Chapter 2 for more details regarding lower tool                      General Information    The Cone and Plate tool is used in the testing of polymer melts  as well as suspensions and emulsions  By  design  there is no velocity gradient across its diameter during steady shear testing  It is also used when shear  normal stress measurements
160. lt in extensive damage to this instrument     WARNING  If this instrument is used in a manner not intended or  specified in this manual  the protection provided by the instrument may  be impaired        WARNING  This is a high torque motor  Turning on the motor while in  dynamic mode causes the motor to snap to dynamic zero position at a  high velocity  This can cause severe damage to the transducer and or  personal injury  To avoid damaging yourself and the transducer        Never turn on the motor while a sample is loaded     Keep hands clear of the motor     ARES User Manual    ARES User Manual       Q          CAUTION  Force Rebalance Transducers contain a precision air  bearing that is equipped with a bearing lock  which prevents movement  of the air bearing when no air is applied  To avoid damaging your  transducer  familiarize yourself with the operation of the bearing lock   see the next topic   and observe the following cautions     Do not apply power to the instrument when the bearing is locked   Do not unlock the bearing unless air is applied to the transducer     If the air supply must be intentionally interrupted  turn off instrument  power and lock the bearing prior to removing air     If the air supply is interrupted while the bearing is unlocked  do not  touch the anvil until air is restored     Maintain air flow to the transducer at all times to prevent contamination  of the air bearing     Failure to observe these cautions will result in damage to the  trans
161. ly collected     Manual Mode    Following the start of the test  Manual Mode data collection takes a single measurement when commanded to   do so  Manual Mode operation is as follows    e Start the test    e When desired  start the measurement by clicking either the  Toggle Steady Measure  button  Figure 3 11   on the toolbar  or the Toggle Steady Measure function  accessed from the Control pull down menu    e When desired  stop the measurement by clicking either the  Toggle Steady Measure  button on the  toolbar  or the Toggle Steady Measure function     Manual mode is designed to allow you to acquire data in the steady state region by manually monitoring the  torque signal  e 2   by using an external strip chart recorder   which is proportional to the sample stress  For  accurate steady state data  measurements should be made at a point where all of the transients in the torque  signal have disappeared  and the torque value is relatively constant     aa    Figure 3 11  Toggle Steady Measure Button  Direction    For positive Rate values  Direction specifies the rotational direction of the actuator at the first commanded  shear rate  Selecting Directions per Measurements of  One  results in data collection while the actuator rotates  in the specified Direction  Selecting Directions Per Measurement of  Two  results in the following sequence of  events     1  Data are collected while the actuator rotates in the specified Direction  2  Data are collected while the actuator rotat
162. made in the following sections  you should refer to the Orchestrator online help and other  software references for a complete description of how to use the software     Test Station    The Test Station is the main instrument component that houses the motor and transducer  between which the  sample to be tested is placed  It also houses the environmental controller  as well as the other electronic  subsystems used in powering and controlling the Test Station  Figure 2 2 and Figure 2 11 show Test Station  assemblies that are described in more detail within this chapter     Test Station Front Assemblies and Controls    The front of the test station  Figure 2 2  is where the motor  transducer and oven are located  In addition  the  basic controls necessary to operate the test station are located on the front panel  All basic mechanical  operations associated with the ARES are performed from the front of the test station  The following sections  describe front panel systems in more detail     Motor    The Motor  Figure 2 2   also referred to as the actuator  deforms the sample by applying a shear strain to the  sample  The Motor can be operated in either dynamic  sinusoidal  mode  or steady  constant rotational rate   mode  You can control the amplitude and frequency of the Motor movement     dynamic mode causes the motor to snap to dynamic zero position at a  high velocity  This can cause severe damage to the transducer and or  personal injury  To avoid damaging yourself and the
163. mately  80  C exists between the fluid temperature and the upper limit  of the thermal operating range     To determine thermal operating range for a given fluid temperature  add these differentials to the fluid  temperature  For example  at a fluid temperature of  20  C  the thermal operating range of the Peltier system  is approximately  20  C to  100  C  calculated as follows     Lower Limit Upper Limit      AT    Fluid Temp   ATy   Fluid Temp         40  C     20  C     80  C     20  C      20  C    100  C    At a fluid temperature of  40  C  the thermal operating range of the Peltier system is approximately 0  C to   120  C  calculated as follows     Lower Limit Upper Limit    AT    Fluid Temp   ATy   Fluid Temp      40  C     40  C     80  C     40  C     0C    120  C    ARES User Manual       Installation of Peltier    Three screws fasten the rotating Shaft of the Peltier Assembly  Figure 2 29  to the Test Station Motor Anvil  A  threaded collar secures the Peltier Assembly Body to the Test Station Motor Housing     Electrical contact to the Test Station is established by the PRT Plug  a seven pin male Lemo    plug that is  attached to the rotating Shaft of the Peltier Assembly  During mounting of the Peltier Assembly  the PRT Plug  is automatically pushed downward into the PRT Receptacle at the center of the Motor Anvil     Two hoses supply fluid between the Peltier Assembly and the fluid source  which is typically a temperature   controlled circulator     Prior to mo
164. ment if the final strain is less than the initial value     Dynamic Strain Sweep Test Ei Es  Frequency al 0 0  rad s  Max 500 0 Min 1 00e 05  Temperature  25 0 PC  Ma  500  0   Min  150  0         Sweep Mode             Log C Linear    Initial Strain               0 1    Max s12 5000 Min 0 0031 25    Final Strain               i 00 0       Points Per Decade  5    Mas 500       Options  PreShear Off DelawOth MeasUps  Off    Options   End of Test   Save As   Help   Cancel         Figure 3 6  Dynamic Strain Sweep Test Set up Screen    ARES User Manual 107    Suggested Uses  Suggested uses of the strain sweep are as follows     e Determination of the limits of linear viscoelasticity and torque levels   e Characterization of samples that exhibit extreme nonlinear behavior  such as filled thermoplastics   thermoplastic blends  etc     Test Options  The following test options can be selected for use with the strain sweep     e Steady PreShear  e Delay Before Test  e Analog Data Input  e AutoStrain  e Measurements Options  o Delay Settings  o Strain Amplitude Control    ARES User Manual    Dynamic Temperature Ramp Test    Functional Description    Temperature Ramp testing takes measurements at selectable temperature ramp rates while holding a constant  frequency and strain     Temperature is automatically incremented or decremented from selectable lower and upper temperature limits  at selected rates  A unique set of test conditions can be entered in up to eight temperature ranges  o
165. moved from the lock is one of 4 screws used  to secure the cover     ARES User Manual    MACHINE SCREWS  3  MOTOR LOCK          MOTOR   COVER REMOVED     PHILLIPS HEAD  SCREW    Figure 2 10  LS Motor Bearing Lock    ARES User Manual    Test Station Rear Input Panel    The rear of the test station is where all electrical  pneumatic  and signal connections are made  Figure 2 11    The main power switch is also located at the rear of the test station     PNEUMATICS  PANEL OO    SIGNAL PANEL          y                    gt                   P             a e e       MAIN POWER SWITCH    Figure 2 11  Test Station  Rear View    Signal Panel    The Signal Panel is the input output interface for electrical signals entering and leaving the Test Station   Figure 2 12 and Table 2 1 identify and describe Signal Panel connectors  including the basic connections for  Test Station use     Signal Connections   There are two basic signal connections required for test station operation  They are a connection to the host  computer  and a connection to the oven  or bath circulator  depending upon the environmental control system  used  Connect the test station to the host computer through the HOST port on the signal panel  and the  appropriate COM port on the host computer  By default  COM1 is selected by the software when it is run for    ARES User Manual    the first time  If you are using a different COM port  change the software configuration  using the Instrument  Set Up function in the Ut
166. mperature Ramp    Functional Description    The MultiWave test is a dynamic test method that superimposes up to 7 harmonic frequencies on a selected  fundamental frequency  This allows you to acquire data at several frequencies simultaneously  in a fraction of  the time required to run a conventional frequency sweep    MultiWave is based upon the Boltzmann Superposition Principle  which states that two or more simultaneous  strain deformations are linearly independent of each other  Because each wave acts independently  the  displacement  strain  of a point in the material is the sum of the strains caused by each wave  providing the  total strain is within the linear viscoelastic region of the material  Data correlation takes place at each of the  applied frequencies  using the same set of raw data  and the algorithm mathematically extracts the torque and  displacement signals at the desired frequency from the total combination of signals     In order to program this test you must provide information on the fundamental frequency     Frequency    The frequency of the fundamental  lowest value  used  This value must be below 2 0 rad sec because of the  way that the data correlation algorithm works     Strain    This is defined as the strain amplitude of the fundamental frequency   Temperature   This is the test temperature that will be used in your experiments   Harmonic information is given in the next section of the test setup screen   Harmonic    This is an integer value by 
167. mperature before starting  the test to allow the bath and sample to stabilize      4  For normal applications  set the Temperature Calibration Table to Default   For critical temperature work  the adjustable option can be used to enter  calibration  offsets  for specific temperature  setpoints   Refer to the  Orchestrator online help for details on how to set up this feature     PID Loop Setup    When using Tool Temperature control  errors between commanded and actual temperature are driven to zero  by a PID  Proportional   Integral   Derivative  loop control system  The PID loop settings  Figure 2 26  affect  how the bath settles at a set point temperature  The values are affected by circulator fluid used  as well as the  circulator specifications  As the bath reaches a setpoint there will typically be a few degrees of overshoot   after which the temperature will oscillate a few times and then stabilize  Typical stabilization times are 15  minutes to reach 0 1   C and 35 minutes to reach 0 01   C  By modifying the Proportional Band  P coefficient   less overshoot can be achieved  but there will typically be more oscillations before the bath stabilizes  so the  overall stabilization time will be longer     ARES User Manual    PID Setup Form Ei E3    PID Temperature Control Setup    Choose PID            Bath   Tool Temp     Proportional Band  0 75         Cancel      Figure 2 26  PID Setup Form       The values listed in Table 2 7 should be used as guidelines  and will work f
168. mple dimensions     e Width   e Thickness   e Length  STD transducers only   length will be determined from gap setting for 2K FRT  transducers     2  Select an insert and place the sample between the inserts    3  Place the sample  with inserts  into the lower tool then place both collars  one above the other  over  the sample  Ensure that the lower collar rests flush against the four stop pins in the lower tool    4  Lower the stage until the upper tool is about 1 4 inch from the sample    5  Open the Set Gap Instrument Control function  Use the  Motor Position Offset  button to radially  align the sample with the upper tool if necessary     ARES User Manual 187    11     12   13     CAUTION  In the next step  do not generate a Torque or Normal Force  greater than 50  of full scale  Failure to observe this caution may result  in damage to the transducer     While confirming the upper insert  and sample  fits into the upper tool  lower the stage until a  compressive  downward  Normal Force of about 10  of full scale is generated  If the sample is not  aligned properly  re raise the stage and realign the sample and tool using the  Motor Position Offset   button    Ensure  visually  that the insert is resting on the four small pins that are located directly below the  lower  and above the upper  insert    Slide the upper collar up onto the upper tool  ensuring that the collar rests flush against the four stop  pins in the tool  Tighten the collar just enough to hold it in plac
169. mpressed in a manual loading procedure with insufficient time for sample relaxation  In  either case the initial gap should be about 0 05 mm higher than the final desired gap     ARES User Manual 163    164    Trim the sample flush with the edges of the plates    Lower the stage to the final gap setting  Again  this can be done manually  or using the Set  Gap Instrument Control function  changing the entered gap to the desired final value  The sample  should bulge slightly as shown in Figure 4 4    Allow samples to relax before beginning dynamic testing by monitoring the force and waiting for it to  decay to close to zero    Enter the sample dimension in the Orchestrator Test Geometry screen  Note that if the tools were  correctly zeroed the actual sample gap can either be read from the Gap real time parameter  or  measured automatically at the start of the test by selecting the  Read Test Fixture Gap  check box in  the Geometry screen     UPPER FIXTURE  SAMPLE    LOWER FIXTURE    Figure 4 4  Parallel Plates with Sample Loaded    ARES User Manual    Hastelloy Tool   Parallel Plates     Strain Constant Stress Constant  2000 G  es g    2006    H TR        Variables  G    Gravitational Constant    980 7  cgs  or 98 07  SI   R   Radius of plates  mm   H   Gap between plates  mm     Options    40 mm size    Environmental Systems    Ambient  Oven    Hastelloy Tool    General Information    The Hastelloy Tool is a specialized set of 40 mm Parallel Plates  This tool is used to test h
170. n 149  iii e E nc I E E A E A N E E oe errr 149  Fluids  Suspensions and EMUlSIONS nar Rd 149  Solid Samples  Including Thermosets  Thermoplastics  and Elastomers     oooocnnccncnncnnnnoninnnnncnnanncnnannnnnos 149  POT EME E O O ias 150  Inermos  ttine Resins and other Curing SUIS ss narnia mene 150  Testing Limits and  Completa 150  Derr O nl SOL aC SAIC GM so resina n inean acre asp 150  Determination ot Operational Rad dia 151  Possible SOU ees and Causes OLEO ap 152  Discrepancies im Sample Geometry sansirnir eai arn oma RTN 152  Tests Ouis ide the near RE ION easing E E E E pcled 152  Ler E Y O a E E A EE N OE I E N 153  OET E ClO PP T AS 153  Eo o o Mi ea a a E E EE E A E EE E E ra  154  Upper Too In lonas 154  Lower Tool Installation  Motor Mount  Oven or Ambient        cccooocccnnnoccccnonononnnnnonnnnnnnnanononononnnnnnnnnnnnnonnnnnos 154  Lower Tool Installation  Fluid Bath MoUDE   ooooonccncciccnoconinannnononacnnaninonnnnnanonanncnnoncnnnnonnannancnnn non cnn canncna cnn noncannos 154   lt a bateas A n a On E wari bausdonaads hess desdldauneameomuacsanertenntte  157  MATA AO e aaa 157  Automate Zeroing and Gap SOLANO erario tilde 158  FTO 19 Len sy o PAPA o E A O 158   rs DEN TM E o DOOR O OU o o O yay wee E E E E crus 158  Comments Concenime the Gap Control Pane rss icicii 158  Enable the Gap Control Ponent pio 158  Read Test POOH Cap GCC DO aye E OE E E E S E 159   DONS play Me the Atunes 159  Max Allowed Force While Seitine ii 159  Test Tool Installation and S
171. n of the oven   including use of either the LN2 controller or Chiller     LN2 Controller Operation    Open the valve on the liquid nitrogen source  All remaining operation of the LN2 controller is performed  through Orchestrator  In order to fill the Dewar  the LN2 Controller Dewar must be turned ON using the  Orchestrator Instrument Control Panel  While the Dewar is filling the software system status will indicate  LN2 filling  and temperature control is not active  Once the Dewar is filled  the software status will indicate  LN2 READY  at which point temperature control will be active     ARES User Manual    Maintenance    There are no user serviceable parts inside the LN2 controller  Periodic servicing is performed in conjunction  with Test Station maintenance visits by Technical Services     LN2 Transfer Line Maintenance    This transfer line utilizes an insulating vacuum jacket  The stainless steel used to manufacturer this transfer  line    outgases    under vacuum conditions  This outgasing degrades the transfer line   s vacuum jacket  and  therefore its insulating properties  Absorbents or    getters    are added in the vacuum space to absorb the gases  and prevent the metal   s outgasing from destroying the transfer line   s insulating vacuum     The    getters    must be renewed periodically to maintain their absorbent qualities  This is simply done by using  the transfer line to transport nitrogen gas from the dewar to the oven  The low temperature of the N2 gas   
172. n the dynamic force rises above this value  strain is decreased  This should be set this to the maximum  dynamic force that you wish to maintain during a measurement     Minimum Allowed Force    When the dynamic force drops below this value  strain is increased  This should be set to the minimum  dynamic force that you wish to maintain during a measurement     Strain Adjustment    This is the percentile by which strain is increased or decreased when the measured dynamic force is below the  entered Minimum Allowed Force or above the entered Maximum Allowed Force  The percentage of  commanded strain entered depends upon how fast the dynamic force is decreasing between measurements     Dynamic Temperature Ramp Test EAE    Options    Steady PreShear  f Delay Before Test  f AutoTension Adjustment    f Analog Data Input  f    AutoStrain Adjustment    f Measurement Options       W AutoStrain  Strain Limits         Max 312 5000 Min 0 003125    Max Applied Strain             i D     Max Allowed Torque          i 800 0  g em   Min Allowed Torque            5 0  g em     Strain Adjustment                50 0   of Current Strain        Figure 3 28  AutoStrain Set Up Screen    ARES User Manual 145    Measurement Options    The Measurement Options in a test can control the delay settings  The Default Delay setting gives 0 5 cycles  and 3 seconds  whichever is longer  delay before data collection  For some samples this may need to be  increased to allow the material to reach a steady state 
173. n to sample stress  This is known as a compliance   and is represented by a J in shear testing  and a D in linear testing  As in case of modulus values it is possible  to define both elastic  J  or D   and viscous  J  or D   components to the complex compliance  J  or D       The ratio of the viscous modulus to the elastic modulus is the tangent of the phase angle shift between stress  and strain  or tan     Tan 6 is a measure of the damping property of the material     Table 3 1 and Table 3 2 summarize the variables frequently used in dynamic mechanical testing for both linear  and shear testing geometries     Table 3 1  Dynamic Mechanical Variables    Shear Geometries    Complex Stress Amplitude a    Complex Strain Amplitude    G    Complex Modulus     Complex Modulus      Elastic Modulus G  a  Storage Modulus    Viscous Modulus G  sin 0     a Modulus    ue   Complex Compliance ye  Elastic Compliance J  cos A   Storage Compliance  Viscous Compliance J  sin A       a   Glo    nt   Complex Viscosity   Viscosity    a In phase Viscosity n  sin      Out of phase Viscosity n  cos d        ARES User Manual 95      Table 3 2  Dynamic Mechanical Variables    Linear Geometries     or      complex Suess Ampiudo    fe Complex ran Ampiudo    o ees oo  C Lone    Elastic Modulus E  cos d   Storage Modulus    Viscous Modulus E  sin d   Fo Fon Modulus  DY   Complex Compliance   Compliance  Elastic Compliance D  a  Storage Compliance  Viscous Compliance D  sin      z Co E   Bo    pe   Comple
174. ng    Figure 2 23   aligned with the corresponding notch  which should be located toward the rear of the   instrument  machined into the motor housing    6  Mount the bath onto the motor by placing the threaded collar onto the threads machined into the  motor and seating the alignment pin into the notch machined into the motor housing  It may be  necessary to rotate the bath back and forth until the pin falls into the notch    7  Tighten the threaded collar then tighten the knob on the motor anvil   hand tighten only   do not over torque     8  Install the lower tool PRT as follows    a  Ensure that the proper PRT is selected from the three lower tool PRTs available for use with  the ARES Bath     Lower Plate  Parallel Plate or Cone and Plate Geometries    700 02647    EA eS S    Lower Cup  Standard Couette Geometry  700 02647 1  17mm Lower Cup  Shallow Couette Geometry  700 02647 2       b  Using very little force  place the PRT plug  keyed end  through the hole in the bath  and onto  the receptacle that is installed in the motor     ARES User Manual 85      c  Gently push down on the PRT and rotate it until the PRT slips into the receptacle  the keyed  end of the PRT then slips into the key slot machined into the receptacle   When properly  installed  Orchestrator should indicate ambient temperature        CAUTION BATH PRT    CHECK FOR FLUID LEAKS AT THIS SEAM     A BATH WELL    ACCESS PORT    THREADED COLLAR    MOTOR COVER    Figure 2 35  Fluid Bath Installation    86 ARES
175. ng the  Zero  button   Wait about 30 seconds  during which time the instrument takes several readings to establish a zero  torque reference  When zeroing is completed  the Transducer Calibration form is displayed  Figure  5 4      If  after selecting the  Zero  button  the NORMAL FORCE value displayed is either very high  such as    1E 5  or exactly zero  refer to the Troubleshooting Guide  The normal force should be less than 0 1   of the full scale normal force     NOTE  Do not hang any weights until after at least one  1  zero reading has  been taken     8  Hang the weight specified in Table 5 4  depending on the transducer in use  from the hook on the  bottom of the calibration tool  Figure 5 6   Ensure that the weight is free to hang without obstruction   and that the weight is steady  not swinging from side to side      9  Select the  Normal Cal  button  The Normal Calibration form is displayed     10  Enter the APPLIED NORMAL FORCE specified in Table 5 4  depending on the transducer in use     ARES User Manual 215       Figure 5 6  Applying Normal Force Using the Calibration Weight     11  Select the  Calibrate Now  button  When calibration is complete  the Transducer Calibration form is  displayed again     12  Verify that the Normal value displayed on the Transducer Calibration Form  Figure 5 4  is within the  limits for the NORMAL FORCE VALUE shown in Table 5 4  depending on the transducer in use  Verify  that the Current Normal Cal value displayed in the Transducer
176. ng the gap  Manual Zero and  Auto Zero  For most applications  the Auto Zero method is recommended  The Set Gap Instrument Control  function in Orchestrator is used to control the stage movement when zeroing and setting the gap     If testing is to be done at temperatures other than ambient  install the tool and then adjust the temperature to  the desired initial value  Allow the tools to reach thermal equilibrium before zeroing and setting the gap     Manual Zero    To manually zero and set the gap between upper and lower tools  perform the following steps     1  Ensure that the upper and lower tools are clean and install them    2  Select the Set Gap Instrument Control function  The Gap Instrument Control Panel is displayed    3  Zero the Torque and Normal Force by pressing the  Offset Torque To Zero  and  Offset Force To  Zero    buttons    4  Using the slew  rapid  mode of the Manual Stage Control  lower the stage until the distance between  upper and lower tools is about 0 5 mm    5  Using the step  slow  mode of the Stage Control  lower the stage until the Normal Force  indicated on   the Set Gap Instrument Control Panel  just begins to deflect from zero  indicating that the tools are in   contact    Zero the Gap value by pressing the  Zero Indicator  button under the displayed Gap    7  Raise the stage to a height that allows the sample to be loaded     Sh    The sample can now be loaded  When the stage is lowered  the Set Gap Instrument Control Panel displays  the cor
177. nimum  dynes cm      2 05E 02  High range   eee 2 05E 00  Low range     1 02E 02  High range   2 22E 06 1 02E 00  Low range     5 11E 01  High range   ESOO 5 11E 01  Low range     1 28E 01  High range   2 78E 05 1 28E 01  Low range    6 39E 00  High range   1 39E 05 6 39E 02  Low range    3 20E 00  High range   6 95E 04 3 20E 02  Low range        Appendix Table A1 2  Complex Modulus Limits for Parallel Plate  STD Transducer     PLATE DIAMETER  mm  Gx Minima  dwnaclem    1 02E 03  10K STD   eee  5 10E 2  10K STD   2 56E 02  10K STD     6 40E 01  10K STD    4 09E 01  10K STD     8   1 60E 01  10K STD     ARES User Manual 249       Appendix Table A1 3  Complex Modulus Limits for Parallel Plate  100 and 200 FRTN1  Transducers     PLATE G  Maximum  dynes cm      Gap  mm  y G  MINIMUM  dynes cm   DIAMETER  mm  SAP  mm  at Frequency     rad sec   dy    9 84E 04      100 2 05E 00  High range  100 FRTN1   q EOS 010 4 10E 00  High range  200 FRTN1   10 2 05E 01  Low range  100 FRTN1   25 5 11E 0  0 lt  4 10E 01  Low range  200 FRTN1   4 92E 04 w   100 1 02E 00  High range  100 FRTN1   0 5    E 2 04E 00  High range  200 FRTN1   canis Aaa 1 02E 01  Low range  100 FRTN1   2 56E 07    lt  10 2 04E 01  Low range  200 FRTN1   6 15E 03 1 28E 01  High range  100 FRTN1   eae vod 2 66E 01  High range  200 FRTN1   1 28E 02  Low range  100 FRTN1   3 20E 06 2 66E 02  Low range  200 FRTN1        3 07E 03 w  100 6 39E 02  High range  100 FRTN1    05 edo 1 28E 01  High range  200 FRTN1   q 6 39E 03  L
178. nitial rate is the Final Shear Rate from the previous zone     Zone Time   Total time allotted to achieve the Final Shear Rate    Direction   Direction specifies the rotational direction of the actuator for positive Final Shear Rate values     Thixotropic Loop Test Ei x    Sampling Mode  log   Linear    Paints Per    one  200    Max 350 Min 20       Shear Rate Limits  17s     Max 200  0000 Min 0  007 000     one Number 1 2 3 4    Final Shear Rate  Meloo foo foo foo  Zone Time  s or hwag ooo oo mo IN    Direction o    f Clockwise    Counterclockwise       Options  Delay  Off    Options   End of Test   Save As   Help   Cancel         Figure 3 18  Thixotropic Loop Test Set Up Form    Suggested Uses    e  Thixotropic Loop is used to determine material thixotropy  or hysteresis  This refers to the dependence of  the material s response  as measured by shear stress or viscosity  to the previous strain rate history  This  in turn  can be related to the build up or breakdown of structure in complex fluids     ARES User Manual    Options  The following test option is available for use with Thixotropic Loop     e Delay Before Test    ARES User Manual    Torque Normal Relaxation    Functional Description    Torque  Normal Relaxation applies and maintains a single transient deformation  step strain   Data is then  collected during each of four zones  the duration of each can be selected  Figure 3 19   Torque and normal  force are monitored and reported during the test  This is in contras
179. nsducers     occcccccnnnnnnnn  218  Procedure for 100 FRTN1  200 FRTN1  100 FRT  200 FRT and 1K FRIN1 wee eeeeeseeeeeeeees 219   Sua Colorao  ICC ooo EE E noises E Oa 220  Pole UN y IPP no o OOO N EE T 220   yate Che Us MIS a a 222  o geese once ac vsc tae see vec psec yee co E otecaeecceseae eerie  222  Tempera nie NON estan pacers a e E E 224  Temperature Calibration usin es Orchestrator asii a Na 224  khair GO  Dane msn dida 227  Sos Mio e A A   A PP E E A T 221  Route IV a NVC odo escent ects patente eee owe te cocacola ooi cesta 227  Cableand Hose Tns pecuarias 227  roo AP o A O A EN II E E 227   AIE FO SS ease sai POT E E EE E EE AE E E E E A 227   A LU T EEC E E E E E E E E eee 228  Caa N r trio 228  Line anda earra LV SES I Le inepto ae Ee SEARE SAE EA E E 229  EON tee os acta T E AER T T O O E E EE meee essen 229   Specii Mane Ba eea cara o e o E E E Seo cores 230  cervice and Repair ile deco AAA e o E E E 230  asnos LS iento bip rrr Tr cre 230   liga elol  oiqta lo dieing     5 110 l  APA REO o o ae eee ec en E 232  GEN  0  A ON o A ee eee ee E E A 232  Apera MO q Ao carseat der ets ceaaese ts stages E A 233   Daca bate treks ak E TO 239  Appen A  Complex Modulas Tis anna sia 249    ARES User Manual       Chapter I    Introduction  Overview    This manual describes the following  instrument specific features of the Advanced  Rheometric Expansion System  ARES  The  figure to the right shows the test station  equipped with the Forced Convection Oven   For information 
180. nside the instrument while power is applied to the  instrument        ARES User Manual 207    Procedures    Calibration Intervals    Table 5 1 lists required calibration tasks and the recommended calibration interval for each task  The table also  shows when certain calibrations must be performed following repairs     Table 5 1  Calibration Tasks and Recommended Intervals    Torque Calibration Suggested  Monthly    Mandatory  Following transducer replacement   Normal Force Calibration   Suggested  Monthly    Mandatory  Following transducer replacement     Phase Angle Check Suggested  Monthly   Mandatory  Following transducer replacement    When attempting to diagnose system problem   Suggested  Monthly    Mandatory  When attempting to diagnose suspected motor  control system problems      Strain Calibration Check       ARES User Manual    NEW STYLE CALIBRATION FIXTURE    HUB   ONE EITHER SIDE        CALIBRATION FIXTURE     new style     PULLEY SHAFT  new style     Pully Wheel mounts here  for 100 200 FRT 100 200 FRTN1   2K STD and 10K STD transducers     On Pully Wheel mounts here for 1K FRTN1   NOTE  this hole is not present on all Pully Shafts        Pully Wheel mounts here  for 2K FRT and 2K FRTN1 transducers    OLD STYLE CALIBRATION FIXTURE    Mount Hub here for   100 200 FRT   100 200 FRTN1   2K 10K STD  Transducers    PULLEY           Mount Hub here for  HUB    KFRT  2K FRTN1 f  Transducers        CALIBRATION FIXTURE   old style     Figure 5 1  ARES Calibration Tools   
181. nt  When the commanded  temperature has been achieved  wait 20 to 30 minutes at the  commanded temperature before starting the test to allow the bath  and sample to stabilize     4  For normal applications set the Temperature Calibration Table to  Default  For critical temperature work  the adjustable option can be  used to enter calibration  offsets  for specific temperature  setpoints    Refer to the Orchestrator online Help for details on how to set up this  feature     PID Loop Setup    When using environmental control systems  errors between commanded and actual temperature are driven to  zero by a PID  Proportional   Integral   Derivative  loop control system  The PID loop settings  Figure 2 39   affect how the bath settles at a set point temperature  The values are affected by circulator fluid used  as well  as the circulator specifications  As the bath reaches a setpoint there will typically be a few degrees of  overshoot  after which the temperature will oscillate a few times and then stabilize  Typical stabilization times    are 15 minutes to reach 0 1   C and 35 minutes to reach a temperature stable at 0 01   C  By modifying the  Proportional Band  P coefficient  less overshoot can be achieved  but there will typically be more oscillations  before the bath stabilizes  so the overall stabilization time will be longer     PID Setup Form Fla    PID Temperature Control Setup    Choose PID           le ath  Tool Temp      Froportional Bard  0 75       A oo    eee  0 5  
182. nt Waveforms          X   REFERENCE   Y   REFERENCE  90     S   STRAIN VECTOR   F   FORCE VECTOR   Bs   STRAIN PHASE SHIFT  FROM REFERENCE          FORCE PHASE SHIFT  FROM REFERENCE    B  Phaser Diagram    F   FORCE VECTOR  S   STRAIN VECTOR  O   FORCE PHASE SHIFT  Xp   COMPONENT OF FROCE  IN PHASE WITH STRAIN    Ye   COMPONENT OF FORCE  90   OUT OF PHASE WITH STRAIN    X  STRAIN AXIS REFERENCE  Y   STRAIN AXIS  90      xY    C  Phaser Diagram Referenced to Strain       Figure 3 1     Dynamic Waveforms and Vectors  ARES User Manual    Documents Describing Instrument Software Operation    The interface between you and the instrument is the TA Orchestrator software  the latest generation of  instrument operating software  Operational procedures shown this manual are given using Orchestrator  version 6 5 6  which is the software released with the instrument at the time of this manual publication   However  menu and function names may change without notice during subsequent software releases  Please  refer to the Orchestrator Online Help system for specific details regarding the software version you are  currently running     Comprehensive procedures concerning Orchestrator operation are provided in the two TA Instruments  documents     e   Orchestrator    Getting Started Guide  TA Instruments document number 902 30010G  This document  included in the Orchestrator    Setup Kit contains information on the operation of the system software     e Orchestrator    Online Help system   
183. nta pipa apenas 33   ale Bou PM ccna sce o  S 35  lion AA e o E E AA E E N OE A A T 35    ARES User Manual    MIO poo 35  DEO Ch SCO urna pta 36  Motor OV e Patio E E Tne rena ee en ert eer 36  RS 37  no y ea E o cae E can cap E E E A T T A E EE ease 37  e aiie erara ao ie 37  sea ON ON E E E E A E E E A E 38  oae Rae AU EN N eea S T EE SR 39  Re nE ME AAB e A E E E E esa E N E eer  40  Ae aor bese EO EE aa E S A NE 41  Standard Transducers  2K STD and 10K STD     ccccccccssseceeessneceesssecesseeseeeceseaaeeeseseaseeceseseeecsseeeesenseaaeees 41  FRT without Normal Force  100 FRT and 200 FRT     ooooooocccnnonoccnonononnnonononnnonononnnnnnnnonnnononnnnnnnnnnnnnnnnnnonnnnnnnnnnnos 41  Procedure for Locking and Unlocking Air Bearings  100 FRT and 200 FRT      oooncnncccccicanicinnnnninncananacinno 41  FRE vita Normal Force  ERIN estriado 42  Procedure for Locking and Unlocking Air Bearings  FRTN1 and FRIN1E     oocnncnnccnccnccnnncnnonncnanacnnaninns 43  NOTE ORE TIN d Bearing Lock Olen ia WON  esmero nc 43  High Resolution  HR  and High Torque  HT  Motor  ocoococnnccconanoninnanononninnonancnnonanonnananonnn corona nean aS 45  Low henr  ES Mora picas 45  Procedure for Locking and Unlocking Air Bearings  LS MotOT    conccnocnoconnnnnaninnnnnnnonanananccnncnnnonacnanccnnannaso 45  Test Staton Rear Input Late asin tado 47  al A e E ER E E O o open eee cea ingaeensose hance goes 47   gt  A a AA o EE ooo E eee Meese ieeness 47  Nacio APP o o o E A A 50  Power OnneC ONS estancias 
184. nterclockwice    Directions Per Measurement O Onei Two       Options  Delay Ot Analoglr  Off      Options   End of Test   Save Az   Help   Cancel         Figure 3 13  Steady Rate Sweep Test set Up Screen  Suggested Uses    This test is used to generate    flow curves    for samples  by measuring the stress and viscosity as a function of  shear rate  This can be used to characterize the non Newtonian behavior of materials     ARES User Manual 119    Options  The following test options are available when using Steady Rate Sweep     e Delay Before Test  e Analog Data Input    ARES User Manual    Strain Controlled Transient Test Methods  Step Shear Rate    Functional Description  Step Shear Rate is a steady transient test that applies a constant commanded shear rate for a selected time  period  Up to four separate zones can be programmed  each with its own set of parameters  Figure 3 14   A    maximum of 350 data points can be sampled in each zone  The interval between data points can be  incremented either logarithmically or linearly  Within each zone the following are then set     Shear Rate    Commanded shear rate in reciprocal seconds  Entering a Shear Rate of zero prevents actuator movement  during data collection  allowing study of relaxation after steady shear     Zone Time   Total time during which Shear Rate is commanded    Direction   Direction specifies the rotational direction of the actuator for positive Shear Rate values     Step Rate Test El x    Sampling Mode  Log
185. ntering the  appropriate parameters as necessary  If the specimen is a regularly shaped  non flowing material  you can  also manually set the gap by continuing to lower the upper plate until only a slight force is generated  The  initial gap should be set approximately 0 05 mm above the final desired gap to facilitate sample trimming     If the specimen is a gel or flowing material  lowering the upper plate onto the sample will result in the  specimen being distributed across the lower plate into a regular cylindrical geometry  For this type of  sample  using the normal force limits  helps to avoid damaging the sample  as the sample may be rapidly  compressed in a manual loading procedure with insufficient time for sample relaxation  In either case  the  initial gap should be about 0 05 mm higher than the final desired gap     Trim the sample flush with the edges of the plates     NOTE  Given the nature of materials used with this tool  sample trimming may  not be practical or possible  In this case  lower the tool to the final gap  For  liquid samples there is a  catch well  that will hold excess sample     Lower the stage to the final gap setting  Again  this can be done manually  or using the Set  GaplInstrument Control function  changing the entered gap to the desired final value  The sample should  bulge slightly     Remove the cover clip and slide the cover down onto the glass chamber     Fill the  liquid seal well  with an appropriate fluid for the sample material  fol
186. nu   and set up the instrument using the guidelines shown in Figure 2 19     Setup Instrument Options    Temperature Control y  Oven  Air  Chiller or LN  Dewar    7    500 0       Figure 2 19  Setup Instrument Options Form  Used to input the environmental system configuration     NOTES    1  Make sure the following are selected    e Instrument Setup   TEMPERATURE CONTROL   e Temperature Control   OVEN AIR  CHILLER  OR LN2 DEWAR    2  Ensure that the maximum and minimum temperatures correspond to the  desired allowable operating range  Normally these are set to the instruments  environmental system limits    e Be sure to set cooling controller to LN2  if this optional feature is   installed and used   e Select the Temperature Loop Control option based upon Table 2 4     ARES User Manual    Table 2 4  Oven Temperature Control Loop Options    Temperature Temperature is   Temperature Control Method  is controlled reported by     by       1  Oven Air Oven PRTs Tool PRT Gas entering the Oven is maintained at the commanded  Temp  both upper temperature  Temperature is reported by the Lower Tool  and lower  PRT     2  Sample  Tool PRT Tool PRT Oven temperature is maintained at the commanded  Tool Temp temperature using the Lower Tool PRT  Temperature is  reported using the same PRT     Tool PRT  such as Torsional Rectangular test tools    Tool PRT Lower Oven Oven temperature is maintained as in Mode 2   PRT Temperature is reported using the lower Oven PRT   NOT NORMALLY USED WITH ARE
187. o plug on Main Fused 220VAC power input to the test  Power Switch Station     OVEN IN Connected to Oven 220VAC power input to the oven control  circuitry     N2 HEATER Connected to optional LN2 Power for the optional LN2 Controller   Controller    Main Power Switch       The Main Power Switch  Rear Panel  switches power on and off to the main portion of Test Station  Figure  2 14   The AC line fuse to the main portion of the test station is located below the power cord socket     x  WARNING  The Main Power Switch does not remove power to the  oven or LN2 Controller  The main power cord  220V IN  must be  disconnected from the Power Panel to completely remove AC power  from the system     ARES User Manual       220V power  connection from  Power Panel    Figure 2 14  Main Power Switch    ARES User Manual    Pneumatics Panel    The Pneumatics Panel is where all gas connections are made  adjusted  and monitored  This panel is located at  the right rear corner of the Test Station  This panel is unique in that it is a two sided panel  and wraps around  the Test Station from side to back  Figure 2 15 shows both the side and back of the Pneumatics panel     Gas Inputs    Two gas inputs are available  one for standard compressed air  Air Supply  MAIN  and an optional port   N2 GAS  for other compressed gases  The main air supply serves two very important functions  It provides  air to the transducer and motor air bearings  Additionally  it serves as a circulating medium for the gun  
188. ocation of the gun heaters and two PRIs  Table 2 4  contains a complete description of temperature control loop options     ARES User Manual    Chiller    The Air Chiller extends the lower range of the oven to   60   C by use of mechanical refrigeration  The Chiller  is a single integrated package  consisting of air filters  air dryers  and the refrigeration unit  It is designed to  connect directly to Test Stations equipped with a Forced Convection Oven  The Chiller requires a separate air  input line at 85 psi and 4 scfm  Air is sent through the filter and air dryer  and then is input to the refrigeration  unit where it is chilled to  70   C  The cold air is then sent on to the oven     Connection    The Chiller air output connects to the Forced Convection Oven  Connect the signal cable from the Chiller to  the LN2 port on the Signal Panel  Figure 2 12   Plug the Chiller into the appropriate power source   Installation should be performed by a qualified TA Instruments service technician     software Configuration    The Chiller operates using Orchestrator software  The following sections discuss set up and operation of the  oven  including use of either the LN2 controller or Chiller     Chiller Operation    To operate the Chiller turn on the Polycold power switch  located on the opposite side of the refrigeration unit  as the air dryer  J box and air filter  and J box on off switch  Figure 2 17   Verify that the pressure regulator  gauge indicates 60 psi  Adjust the pressur
189. odulus G that  can be measured by each transducer type using the parallel plate geometry  The following geometry specific    factors affect the operating range of parallel plate geometry   e Plate diameter  strain constant K  and stress constant Kx     e Gap between plates  strain constant K       ARES User Manual 161    Additionally  the following instrument specific factors affect the operating range of all geometries     e Minimum torque that can be measured by the transducer  e Transducer compliance  e Maximum strain that can be generated by the motor    To calculate the minimum and maximum G that can be measured by each transducer type using the parallel  plate geometry  use the following equation     a Ks lc  4 1   o  K    Y     where  A    Stress Constant  K    Strain Constant    and C is computed from the following     C for G  MAXIMUM C for G  MINIMUM    2K FRTN1    2K FRTN1E   J 1 15e 06  rad gecm  a an  see note below     1K FRTN1 l J   4 9 e 06  see note below     2K STD    10K STD N J   2 60e 06  see note below     100 FRT for   100  J   2 60e 05    200 FRT a  100 FRTN1 for 0 10  J  2 60e 06    200 FRTN1 Za  see note below     for w lt 10   see note below     NOTE  The values for M  gecm  and 6  rad  are found in the specification tables in Chapter 1     Table 1 6 through Table 1 13   Pick the correct values for your specific transducer and  motor combination        To determine the maximum or minimum complex viscosity n  that can be measured at a given frequency
190. onger soak times to equilibrate  Typical soak times are from 1 to 5  minutes  The minimum soak time is zero seconds  maximum 65000 seconds    Dynamic Temperature Step Test El E    Frequency  f 0 0  rads  Max 500 0 Min 1 00e 05  Initial Temp   25 0 PC  Max B00 0  C Min  1 50  0         strain Limits  2     Max 31 25000 Min 0 003125      one Number 1 2 3 4  Final Temp             E   50 0  200 0  0 0  0 0  Temp  Increment    C1 5 0  10 0  0 0  0 0    Soak Time  sorhms ffag Bon fo fo    e    ipo feo foo foo       Options PreShearcOff Delay Of AutoTens  OF AutoStrn On MeasOpe  Off    Dk   End of Test   Save Ag   Help   Cancel         Figure 3 5  Temperature Step Test Set Up Menu  ARES User Manual 105    Suggested Use    Temperature step tests are generally used to analyze the behavior of a sample  as a function of temperature   Temperature steps are preferred over the temperature ramp method if accurate isothermal data are needed   This is because any sample thermal conductivity effects cause internal temperature gradients until the sample  has had sufficient time to equilibrate at a given temperature     Test Options  The following test options can be selected for use with the temperature sweep     e Steady PreShear  e Delay Before Test  e Analog Data Input  e AutoTension  e AutoStrain  e Measurement Options  o Delay Settings  o Strain Amplitude Control    ARES User Manual    Dynamic Strain Sweep    Functional Description    Strain Sweep takes successive measurements at selected
191. or begins all tests at the motor zero position  and drives  symmetrically about motor zero  at the chosen frequency  to the commanded displacement  strain   The  maximum angular deflection of the motor is 0 5 radians from either side of motor zero  When in dynamic  mode  the instrument measures strain and torque     In Steady Mode  the motor can begin a test from any position  rotating either clockwise or counterclockwise   as specified  at a specific rotational  shear  rate  When in steady mode  the instrument measures rotational  rate  sample torque  and  if equipped with the appropriate transducer  normal force     Transducers    There are two types of transducers available for ARES  The Force Rebalance Transducer  FRT  is an active type  transducer  and provides excellent resolution and temperature stability    The Standard Transducer is a passive  spring type transducer that provides high frequency response with the  ruggedness desired by QC labs  Several different ranges are available for each transducer type     Force Rebalance Transducer with Normal Force  FRTN7     The Force Rebalance Transducer with normal force  FRTN1  consists of independent rotational  torque  and  axial  normal force  servo control systems  each utilizing position feedback to maintain the FRT shaft  mass in  contact with the sample  in a null position when no force is applied  When force is applied to the FRT shaft   the servo control systems drive the shaft back to null position  The electrical
192. or most applications  However   you may need to experiment somewhat to determine their baths best PID settings for their system  or specific  applications  The Orchestrator ON Line Help has a complete description of how to determine and tune the  PID coefficients     Table 2 7  PID Values for the Julabo   Circulator and Selected Fluids    Julabo    Circulator ES 18      50  Water  PAPUG 50  Ethylene Glycol    Proportional Band 0 75       Fluid Bath 2 Operating Requirements  The Fluid Bath can operate only if the following conditions are met     e The Fluid Bath is selected as the current environmental system   e The circulator must be filled  on  and circulating fluid through the bath   e The circulator must be connected to the test station via the correct RS 232 cable     Fluid Bath 2 Operation    The Fluid Bath is operated using the Instrument Control Panel  Figure 2 27   The desired temperature is set in  the Temperature input field  For most applications  especially isothermal testing  or when the actual tool  temperature is required  Tool Temperature control is used  For temperature ramp tests  where a steady ramp  rate is most important  the circulator control gives better results  If using manual temperature control for  isothermal testing  you may have to adjust the circulator fluid temperature somewhat to give the desired  temperature at the tool     ARES User Manual    Instrument Control Panel    Figure 2 27     ARES User Manual       p50  Bath  Instrument Controlle
193. ormal force and torque on  the motor using the  Offset Normal Force to Zero  and  Offset Torque to Zero  buttons    4  Zero the gap using the  Zero Indicator  button in the Set Gap Instrument Control function  Using a    Maximum Allowed Force  of 100 gm works well    Using the Set Gap Instrument Control function  set the Gap to 1 mm    6  Raise the stage to provide sufficient room for sample loading     m    Sample Loading  Fluid Bath 2    The nominal sample volume is between 8 to 9 ml  For lower viscosity fluids  a volume closer to 9 ml is  optimal  For higher viscosity fluids  using a volume closer to 8 ml gives good results  In all cases  the fluid  level must be at least up to the sample fill level  lip   To avoid edge and boundary layer errors  filling slightly  past the  lip  is desirable as shown in Figure 4 18  Overfilling the sample however  especially in the case of  higher viscosity fluids  may result in errors due to an actual wetted bob length longer than the entered  effective length     Carefully pour the sample into the cup  making sure not to spill sample into the recess that holds the inner cup  mounting screw  A small graduated cylinder can be used to transfer material into the Couette  as well as a  syringe or pipette  In any case  a consistent sample volume from sample to sample is desirable for each  material tested     After filling the cup  return the bob back to the 1 mm gap  This will ensure that the nominal bob length is the  correct 32 mm     ARES 
194. ote below     100 FRT    e au These transducers are not recommended    200 FRTN1 for use with the torsion rectangular tool     NOTE  The values for M  gecm  and 6  rad  are found in the specification tables in Chapter 1   Table 1 6 through Table 1 13   Pick the correct values for your specific transducer and  motor combination        Using a spreadsheet application such as Microsoft   Excel     you can use the equations above to plot the range  of complex viscosity that can be tested for a given geometry  transducer combination as follows     1  Calculate G  MAXIMUM and G  MINIMUM  using equation 4 1  for a sample of fixed length and width  and the minimum and maximum thicknesses the tool can accommodate     2  Generate an X Y scatter plot of sample thickness  Y axis  versus complex modulus  G   X axis      The region between the upper and lower limits of operation is the range of complex modulus that can be  tested     Coefficient of Thermal Expansion  a     When testing at other than ambient temperatures  the coefficient of thermal expansion for Torsion Rectangular  geometry is defined as   AL 1    At Ly    where   Coefficient of Thermal Expansion aa    At   Change in temperature    C   Lo   Original length of sample  mm   AL   Change in length of sample  mm   Positive AL indicates increasing sample length    ARES User Manual    Tool Installation    1  Select the Set Gap Instrument Control function under the Control menu in Orchestrator  Use the   Send to Top  button to rai
195. ote below   max    a i au for   100  J   2 60e 05  7     form 10  J 2 608 06    100 FRTN1  200 FRIN1    NOTE  The values for M  gecm  and 6  rad  are found in the specification tables in Chapter 1   Table 1 6 through Table 1 13   Pick the correct values for your specific transducer and  motor combination        To determine the maximum or minimum complex viscosity N  that can be measured at a given frequency  use  the following formula     n   42   a    ARES User Manual 191    where n   Complex viscosity  Poise     G   Complex Modulus  dynes  cm           Frequency  rad sec     Using a spreadsheet application such as Microsoft  Excel     you can use the equations above to plot the range  of complex viscosity that can be tested for a given geometry transducer combination as follows     1  Calculate G  MAXIMUM and G  MINIMUM using equation 4 1    2  Substitute the G  MAXIMUM value into equation 4 2  then determine the upper limit of operation by  calculating N  at each O  O values chosen to be from the lowest to highest frequencies within the  transducer operating range     3  Substitute the G  MINIMUM value into equation 4 2  then determine the lower limit of operation by  calculating 1  at each       4  Generate an X Y scatter plot of complex viscosity N   Y axis  versus frequency    X axis      The region between the upper and lower limits of operation is the range of complex viscosity that can be  tested  Appendix 1 contains tables of G  values for some combinations of bobs
196. ote that this  tension level is a general recommendation only  and you should set the tension level according to the  sample characteristics  with thinner and  or lower modulus materials requiring less axial force  When  using the AutoTension feature  adjust the stage so that the normal force is zero     a    ARES User Manual 181    10  Using the  Motor Position Offset  slider in the Set Gap Instrument Control function in Orchestrator  adjust the motor position until the displayed torque is zero   DO NOT use the  Offset torque to Zero  button    11  Read the gap and record this distance as the sample length  2K FRT transducers only     12  Use the Hold function under the Control pull down menu when changing temperature     Setting Anvil          Sliding Clamp  Centering Lines     Nominal  Clamping  Thickness Shown  on This Face    Adjusting Screw    Figure 4 10  New Torsion Rectangular Tool Details    182 ARES User Manual    SIDE VIEW        NOTE    Each Setting Anvil is  designed to clamp two  different thickness ranges   The Anvil should be mounted  as shown below  with the    desired  nominal  thickness on the  outside face     The number on the face that  actually touches the sample is  NOT the clamp thickness in use        SETTING ANVIL    CENTERING  LINES       Seed  iididiil J      aa    i    TOP VIEW   LOWER FIXTURE             The number shown on this  face of the Setting Anvil is  the  Nominal  thickness the  fixture will properly clamp        on       UPPER FIXTURE
197. ow     1K FRTN1  0  J 4 9 e 06  see note below     2K STD  10K STD i J   2 60e 06  see note below     100 FRT for   100  J   2 60e 05    200 FRT ON  100 ERTN1 for 0 10  J  2 60e 06    200 FRTN1   2  see note below     for w lt 10   see note below     NOTE  The values for M  gecm  and 6  rad  are found in the specification tables in Chapter 1   Table 1 6 through Table 1 13   Pick the correct values for your specific transducer and  motor combination        174 ARES User Manual    To determine the maximum or minimum complex viscosity N  that can be measured at a given frequency  use  the following formula     nea  4 2   q    where  n   Complex viscosity  Poise     G   Complex Modulus  dynes  cm       Frequency  rad sec     Using a spreadsheet application such as Microsoft  Excel     you can use the equations above to plot the range  of complex viscosity that can be tested for a given geometry  transducer combination as follows     1  Calculate G  MAXIMUM and G  MINIMUM using equation 4 1    2  Substitute the G  MAXIMUM value into equation 4 2  then determine the upper limit of operation by  calculating N  at each O  O values chosen to be from the lowest to highest frequencies within the  transducer operating range     3  Substitute the G  MINIMUM value into equation 4 2  then determine the lower limit of operation by  calculating N  at each       4  Generate an X Y scatter plot of complex viscosity N   Y axis  versus frequency    X axis      The region between the upper and low
198. ow range  100 FRTN1    1 60E 06 w lt  10 1 28E 02  Low range  200 FRTN1     Appendix Table A1 4  Complex Modulus Limits for Parallel Plate  100 and 200 FRT Transducers     PLATE G  Maximum  dynes cm      GAP  mm  y G  Minimum  dynes cm   DIAMETER  mm    Gae mm  at Frequency   rad sec  al    9 84E 04 w  100 1 02E 00  High range  100 FRT     1 84E    4 2 04E 00  High range  200 FRT   tii po 1 02E 01  Low range  100 FRT   25 9 11E 07 0  lt  2 04E 01  Low range  200 FRT     4  92E 04 w  100 5 11E 01  High range  100 FRT     1 02E 00  High range  200 FRT   4 92E   1  eo E 5 11E 02  Low range  100 FRT   2 56E 07 w lt  10 1 02E 01  Low range  200 FRT   6 15E 03 6 39E 02  High range  100 FRT   6 15E 04 1 28E 01  High range  200 FRT     6 39E 03  Low range  100 FRT   3 20E 06 1 28E 02  Low range  200 FRT   3 07E 03 3 20E 02  High range  100 FRT     3 07E 04 6 40E 02  High range  200 FRT   3 20E 03  Low range  100 FRT   1 60E 06 6 40E 03 Low range  200 FRT        ARES User Manual    Appendix Table A1 5  Complex Modulus Limits for Cone and Plate   2K FRTN1 and 2K FRTN1E Transducers     PLATE DIAMETER CONE ANGLE G  MAXIMUM G  MINIMUM   mm  e  dynes cm      dynes cm      9 59E 01  high range   2 08E 06   9 59E 01  low range   1 92E 01  high range   4 17E 05   4 996 01  low range     3 84E 01  high range   eile 3 84E 01  low range     2 40E 00  high range    lt a 5 21E 04   5 40E 02  low range    4 79E 00  high range   1 04E 05   4 79    02  low range        Appendix Table A1 6  Complex Mo
199. perature step test  Figure 3 5   you can enter a test frequency and an initial or starting  temperature  Next  you can enter each individual zone final temperature  step size  soak time  and strain   Testing limits are displayed for each of these fields     When setting strain values in this test be aware that the sample modulus can change significantly with  temperature  and the torque signal generated at a given strain may vary dramatically  A strain that provides  good torque and is within the linear viscoelastic region at room temperature may overload the transducer at  lower temperatures  Single point tests at the temperature extremes are a good way to find the appropriate  strains to set for each zone  The AutoStrain adjustment is also a good option for optimizing the torque  generated in this test  Note that when AutoStrain is used it is possible to have the current strain propagate  into a new zone by entering a value of zero for the strain   Inherit  feature      The final temperature can be any temperature within the range of the instrument  and there is no limitation on  step size     Data Measurement Method    Each measurement is equivalent to a data point  For example  changing temperature from 10   to 100   in 10    steps produces 10 data points  Making the same temperature change in 5   steps results in 19 data points  Soak  time should be chosen with consideration given to sample volume and temperature increment  Larger  samples and larger step sizes require l
200. r  In particular  any and all warranties of merchantability  fitness for a particular purpose or  non infringement of third parties rights are expressly excluded  Further  TA Instruments makes no  representations or warranties that this software and documentation provided are free of errors or viruses  or that this software and documentation are suitable for your intended use     LIMITATION OF LIABILITY  In no event shall TA Instruments or its suppliers be liable to you or any  other party for any incidental  special  or consequential damages  loss of data  or data being rendered  inaccurate  loss of profits or revenue  or interruption of business in any way arising out of or related to the  use or inability to use the software described above and or documentation  regardless of the form of  action  whether in contract  tort  including negligence   strict product liability or otherwise  even if any  representative of TA Instruments or its suppliers has been advised of the possibility of such damages     This LICENSE represents the entire agreement concerning the software described above between you and TA  Instruments  It supersedes any prior proposal  representation  or understanding between the parties     The Oracle   software  used in conjunction with the Advantage Integrity software  is not included as part of  the TA Instruments  Advantage Integrity license agreement  Users of Oracle database software must abide by  the terms and conditions as specified by the Oracle Co
201. r  zones    A programmable  thermal soak time  at the end of each zone ensures temperature stability prior to beginning  the next ramp     When setting up a temperature ramp test  you enter a test frequency and an initial or starting temperature   Next  you enter each individual zone final temperature  ramp rate  i e   2   C min  5  C min  etc    soak time   and strain     When setting strain values in this test be aware that the sample modulus can change significantly with  temperature  such that the force generated at a given strain may vary dramatically at different temperatures  A  strain that provides good force and is within the linear viscoelastic region at room temperature may overload  the transducer at lower temperatures  Single point tests at the temperature extremes are a good way to find  the appropriate strains to set for each zone  The AutoStrain adjustment is also a good option for optimizing  the force generated in this test  Note that when AutoStrain is used  it is possible to have the current strain  propagate into a new zone by entering a value of zero for the strain   Inherit  feature      Frequency and initial temperature are entered first  Figure 3 7   and then the following parameters are entered  for each zone     Final Temperature    Final temperature is the temperature at which the instrument stops ramping temperature while in the  respective zone  The final temperature is independent of time     Ramp Rate    Ramp rate is the rate of change  positiv
202. r Displacement is incorrect  Ensure that correct geometry is entered   Ensure that the desired strain is entered   Verify that the diagnostic LEDs are lit   Call Technical Service   Motor oscillates with an accompanying high  Ensure sample is not too stiff     pitched  audible noise when running a test  Call Technical Service     OVEN AND LN2 CONTROLLER    Can not turn on the oven  If air is being used as an input to the heaters   ensure that the Orchestrator AIR LOW indicator  is not on  If itis  the air supply to the oven has  been interrupted  Restore air flow     If using the LN2 Controller  ensure that    The LN2 supply is adequate and the valve is   open    The LN2 READY indicator is on    The LN2 FAULT indicator is off  if on  go to 7    Ensure that the oven door is closed and the  OVEN OPEN indicator is not on     Ensure that the reported temperature is not  constantly above 650   C  If itis  there is an open  in the PRT electrical circuit  Go to 8     Ensure that the SET USER TEMPERATURE LIMIT is  set to a reasonable value     Ensure that the instrument MAXIMUM  TEMPERATURE is set to a reasonable value     An oven or heater fuse may be open  Check and  replace     Call Technical Service        ARES User Manual 233    Table 6 2  Instrument Operation Troubleshooting Guide  Continued        PROBLEM CORRECTIVE ACTIONS    Oven can be turned on  but does not heat     Oven can be turned on  but is not correctly  heating     No LN2 READY indication  Orchestrator      1     2
203. r a single decade of rate  from 10 to 100 reciprocal seconds   1 s   Selecting five data points to be measured per decade divides the difference of the endpoint logarithms  into five equally spaced fractional exponents     Six discrete rates are generated  in succession  by taking the antilogarithm of each exponent     e Initial Rate   10 1 sec  e Decade Rates   15 9  25 1  39 8  63 1 1 sec  e Final Rate   100 1  sec    One data point is measured at each of the rates   Discrete    The discrete rate sweep generates up to five shear rates in succession  Each shear rate is entered into a Zone     Data Collection Mode    Data can be collected in either time based mode  one measurement is taken at each rate  or manual mode  one  measurement is taken at selected rate      Time Based    Following the start of the test  Time Based data collection takes one measurement at each rate  Setting the  Sweep Mode to Log commands logarithmically incremented shear rates  A Discrete Sweep Mode commands  up to five unique shear rates in succession  At each shear rate  Measurement Time is the period during which  data are collected  At each shear rate  Delay Before Measure is the time period between command of the  current rate and the beginning of data collection     ARES User Manual    Manual Mode    Following the start of the test  Manual Mode data collection takes a single measurement when commanded to   do so  Manual Mode operation is as follows    e Start the test    e When desired  start t
204. raphically   and provide the option to download directly into the instruments waveform memory  Any subsequent change  to the equations or Wave Time fields requires recalculating and re sending the waveform  The software keeps  track of any new changes  and when the test is started  reminds you to resend the waveform if necessary     The Zone 1 through Zone 4 times  as displayed on the host computer  refer to data collection time  and are  independent of the wave playback times  That is  the waveforms will be played back  one immediately after  the other  based on their entered Wave Times  and not on the data collection Zone times  It is possible to  playback more than one waveform in a Zone  or play an equation across Zones  Once all the waveforms have  been played  the motor will stop movement  but the software will continue to acquire data until the end of the  last entered data collection Zone time     Consider the following example     Equation One Wave Time   10 seconds  Equation Two Wave Time   15 seconds  Zone 1 acquisition time   20 seconds  Zone 2 acquisition time   10 seconds    Data will be collected in Zone 1 for the entire Equation One playback duration  Wave Time  of 10 seconds  As  soon as Equation One finishes  Equation Two will start playing  with data still being collected in Zone 1  Ten  seconds later  Equation Two still playing  data collection will end in Zone 1 and will data collection will start  in Zone 2  After five more seconds  Equation Two will fin
205. re  or the temperature at which the sample was  loaded   the clamps  on some tools  may loosen as they cool  This is due to difference in the thermal  expansion coefficients of the sample and that of the tools  Immediately prior to initiating testing at the lowest  test temperature  you may wish to open the oven door and verify that the screws securing the clamps are tight   taking care not to touch any surface of the oven or tools which may be at dangerous cryogenic temperatures     ARES User Manual    Parallel Plates    Strain Constant Stress Constant    2000 G  K  2 al    Y H i TR     Variables    G    Gravitational Constant    980 7  cgs  or 98 07  SI   R   Radius of plates  mm     H   Gap between plates  mm     Options  8  25  40  50 mm sizes  Serated plates  Disposable plates  Invar Plates       Environmental Systems    Ambient   Oven   Fluid Bath  Fluid Bath 2    Peltier       Parallel Plate Tool         see Chapter 2 for more details regarding lower tool       General Information    Parallel Plates are used to test polymer melts  soft solids  and higher viscosity fluids  Using disposable plates  they are used for testing thermosetting resins and epoxy curing  The wide range of sizes  use over a wide  range of viscosities  variable gap  and ease of loading  make them a very versatile tool  Additionally  high  shear rates are accessible using a small gap setting     Operating Ranges    Operating range is defined as the region bounded by the maximum and minimum complex m
206. rect distance between tools     ARES User Manual 157    Automatic Zeroing and Gap Setting    The Set Gap Instrument Control Panel function allows you to automatically zero and set the gap between  upper and lower tools     Zeroing the Gap  Establish a zero reference gap between Upper and Lower tools as follows     1  Ensure that a sample is not loaded and the upper and lower tools are clean    2  Using the Manual Stage Control  lower the stage to achieve a Gap of about 1 mm  as judged visually     3  Select the Set Gap Instrument Control Panel function  The Set Gap Instrument Control Panel dialog  box is displayed    4  Zero the Torque and Normal Force by pressing the  Offset Torque To Zero  and  Offset Force To  Zero  buttons    5  Click  Zero Tool   An information form is displayed showing the duration  Elapsed Time  of the  zeroing operation  The present gap is displayed in the Current Gap field  which updates at one   second intervals  The following events occur during the zeroing operation    a  The stage lowers to achieve contact between tools   b  Following contact  the Gap display is zeroed   6  In preparation for loading the sample  raise the stage to maximum height by clicking Send to Top     Remain in the Gap Control Panel and proceed to the next topic     Setting the Gap        Setting the Gap    After zeroing the Gap  set it as follows     1  Place the sample onto the lower tool   2  Enter the following information   a  Commanded Gap Position   Enter the desired 
207. response  Manual settings can be adjusted as shown in  Figure 3 29     Correlation Delay    Correlation delay is the time period between the start of sample deformation and the first measurement in a  dynamic test  During this period the instrument does not collect data  Note that data collection is delayed   not motor movement  Correlation delay allows the sample to achieve equilibrium with the test conditions  before the measurement is taken  and allows development of the phase angle      from the start of stress and  strain sine waves  Internally  the instrument computes the correlation delay in seconds  with a minimum  increment of 0 1 second and a minimum time period of 0 2 second  The number of cycles to which this  corresponds varies depending upon the frequency  The maximum allowable time period is 65 000 seconds     One Cycle Correlation    The One Cycle Correlation option speeds the test by commanding the control computer to use data measured  over one cycle  regardless of test frequency  Normally  at frequencies above 2 rad sec data are collected over  multiple cycles and averaged  One Cycle Correlation forces measurements to be made with only on cycle  worth of data  saving the time that would be required for the subsequent cycles  One Cycle Correlation is  useful for applications requiring fast measurements such as sample curing at high frequencies  The  disadvantage is that  due to the absence of data averaging  resulting data points may contain more noise than 
208. rmulas    All formulas and geometry constants are also listed in Orchestrator Online Help  see Reference Guide under  the Contents tab      Dynamic Measurement Formulas    Table 4 1  Dynamic Measurement Formulas    VARIABLE AND FORMULA DEFINITION OF VARIABLES    204    STRESS   Kz   M     STRAIN   Ky   8     ELASTIC  STORAGE  MODULUS    cos 0 a  Yy    VISCOUS  LOSS  MODULUS    sin    E  Y    COMPLEX MODULUS     BP   6        LOSS TANGENT    REAL PART OF DYNAMIC  COMPLEX     VISCOSITY  G     0    IMAGINARY PART OF DYNAMIC  COMPLEX     VISCOSITY  G  0       Stress Constant  Torque  g   cm   Strain Constant    Shearing angle of motor  radians     Phase angle  phase shift between stress  and strain vectors     Frequency  angular  in rad sec     ARES User Manual    Table 4 1  Dynamic Measurement Formulas  Continued     VARIABLE AND FORMULA DEFINITION OF VARIABLES    n    DYNAMIC  COMPLEX  VISCOSITY       Steady and Transient Measurement Formulas    Table 4 2  Steady and Transient Measurement Formulas    STRESS K    Stress Constant   Kz  M  Torque  g   cm   STRAIN Strain Constant     K   0  Shearing angle of motor  radians     y   STRAIN RATE  SHEAR RATE  Strain Constant    Angular velocity of motor  radians sec     n   VISCOSITY    N    NORMAL STRESS Kz   Normal stress constant  Ger F    Normal force  g     ARES User Manual 205       Table 4 3  Strain and Stress Constants    Strain Constant K  Stress Constant K     Parallel Plates    Cone and Plate    Feroe canghlar TOR IATIW LW L
209. rporation     ARES User Manual    Trademarks and Patents    The following references apply to the information presented in this document     TA Instruments Trademarks  O Series    is a trademark of TA Instruments Waters   LLC  109 Lukens Drive  New Castle  DE 19720     Advantage Integrity    is a trademark of TA Instruments Waters   LLC  109 Lukens Drive  New Castle  DE  19720     Modulated DSC   and MDSC   are registered trademarks of TA Instruments Waters   LLC  109 Lukens  Drive  New Castle  DE 19720    Tzero    is a trademark of TA Instruments Waters   LLC  109 Lukens Drive  New Castle  DE 19720    pTA   is a registered trademark of TA Instruments Waters   LLC  109 Lukens Drive  New Castle  DE 19720   Smart Swap    is a trademark of TA Instruments Waters   LLC  109 Lukens Drive  New Castle  DE 19720   Hi Res    is a trademark of TA Instruments Waters   LLC  109 Lukens Drive  New Castle  DE 19720   Mobius Drive    is a trademark of TA Instruments Waters   LLC  109 Lukens Drive  New Castle  DE 19720     Orchestrator    is a trademark of TA Instruments Waters   LLC  109 Lukens Drive  New Castle  DE 19720     TA Instruments Patents    Method and Apparatus for Modulated Differential Analysis  MDSC    describes proprietary technology  patented by TA Instruments Waters   LLC  U S  Patent Nos  5 224 775  5 248 199  5 346 306  Additional Patent  Nos  CA 2 089 225  JP 2 966 691  and BE  DE  EP  GB  IT  NL 0559362      Heat Flux Differential Scanning Calorimeter Sensor  Tzero     de
210. rque on  the motor using the  Offset Normal Force to Zero  and  Offset Torque to Zero  buttons    Using the stepper control buttons  lower the stage to a point where the tools are close but not  touching    Use the  Zero Fixture    button in the Set Gap Instrument Control function to determine the zero point  for the test tools    Raise the stage to provide sufficient room for sample loading     Sample Loading    The recommended gap setting for parallel plates is between 0 5 and 2 millimeters  Figure 4 4 shows parallel  plates with sample loaded     1   2     Place the sample on the lower plate  Ensure that the sample is centered on the tool    Using the stepper motor buttons on the right side of the test station  adjust the sample gap until the  upper plate is close to the specimen  Set the gap using the Set Gap Instrument Control function   entering the appropriate parameters  If the specimen is a regularly shaped  non flowing material  you  can also manually set the gap by continuing to lower the upper plate until only a slight force is  generated  The initial gap should be set approximately 0 05 mm above the final desired gap to  facilitate sample trimming    If the specimen is a gel or flowing material  lowering the upper plate onto the sample will result in the  specimen being distributed across the lower plate into a regular cylindrical geometry  For this type of  sample  using the normal force limits helps to avoid damaging the sample  as the sample may be  rapidly co
211. rted as gap   Instrument Gap Measurement  versus time         Force Gap Test E E          Force Sample Gap tofo   mrn   Gap Adjustment Time i    ad  e or Fem  s     Temperature                 25 0 PC  Ma   500  0   Min  1 50 0      Save Test Data      Save As   Help   EA      Figure 3 20  Force Gap Test Set up Screen    Suggested Uses    This test is useful during test sequences  as it can be inserted into the sequence to effect a controlled gap  adjustment between tests     This test is also good for sampling loading  It allows for documentable  consistent sample loading  and will  minimize damage to  or axial force on a sample during loading     Test Options    There are no options available with this test     ARES User Manual    Steady Step Rate Temperature Ramp    Description    Steady Rate Temperature Ramp monitors material stress and viscosity as a function of time at a series of  specified shear rates within up to eight independently programmable zones  Within each zone  you can ramp  temperature  at a selectable rate  upward or downward from the initial temperature  Figure 3 21   In each  zone  specifying Log sampling takes the specified number of data points at logarithmically incremented  intervals  Selecting Linear takes the specified number of data points at linearly incremented intervals  A  maximum of 350 data points can be taken in each zone     In Each Zone the Final Temp is the Temperature at which the instrument stops ramping temperature while in  the respe
212. rument Control Button    Gap Instrument Control 3       Gap  mm       117 47  Send to Top     eno Indicator    Gap Command    Commanded Gap  mm  i 050  Max allowed Force  gm  li FOO    Set Gap     ero Fixture      Torque  gr cm           Offset Force To  Zero           Status  Ton Motor Wn      Reset eid    Remove ero  Offset           Motor Position Offset  4 F 5      k 0 00           Advanced      Help   Exit         Figure 2 4  Set Gap Instrument Control function main input form    ARES User Manual    Stage Rate Adjustment    Both the operator controlled step rate and slew rate of the stage can be adjusted through the Advanced Control  form  shown in Figure 2 5  Click the  Advanced  button in the Set Gap Instrument Control function main  form to access the Advanced Control form  Adjust the step speed and slew speed as desired  Automatic stage  movement rates  initiated using the  Send to Top  and  Set Gap  buttons  cannot be adjusted  The Front Panel  LCD contrast is also adjusted from this dialog box     Advanced Control       Figure 2 5  Advanced Control Form of the Gap Instrument Control Function     ARES User Manual 39      Front Panel LCD Display    The front Panel LCD Display  Figure 2 6  is used to display certain instrument status information  For the  ARES  temperature  gap  torque  and normal force are displayed  The LCD display can be turned on and off  from the Advanced Control form of the Set Gap Instrument Control function  Figure 2 5      ARES  TEMP   22 4 C
213. s may undergo relatively large dimensional fluctuations as a result of temperature changes   AutoTension or AutoStrain may be necessary to adjust for temperature related changes     Other Factors    Other factors that will cause errors include sample edge failure  sample buckling  and sample slippage or  clamping problems  Care should be taken to load the sample properly  determine it s proper testing range   and  when needed  apply AutoTension correctly     ARES User Manual 153    General Test Tool Installation    WARNING  If this instrument is used in a manner not intended or  specified in this manual  the protection provided by the instrument may  be impaired        Upper Tool Installation    To install an upper tool  loosen the knob on the anvil  Figure 4 1  and insert the tool into the anvil  pulling  apart the retainers if necessary  Tighten the knob  Hand tighten the knobs   do not over torque them     Lower Tool Installation  Motor Mount  Oven or Ambient     To install a lower tool onto the motor  refer to Figure 4 2 while performing the following steps     1  If the tool accepts a PRT  install the tool PRT as follows    a  Place  do not push  the plug onto the electrical jack mounted in the motor anvil    b  While applying light downward force  rotate the tool PRT until it slips into place  indicating  that the PRT has aligned with the keyway in the electrical jack  When properly installed   Orchestrator will indicate ambient temperature    2  Loosen the knob on the 
214. sample thickness ranges  which are listed on the next    page     ARES User Manual 177    Sample Dimensions    To prepare samples that fit within the physical constraints of the tool  use the following guidelines     e Maximum Sample Width  12 7 millimeters  e Typical Sample Length  45 millimeters  e Sample Thickness depends on the size of the setting anvil used     SETTING ANVIL NOMINAL THICKNESS ACTUAL SAMPLE THICKNESS  STAMPED ON FACE CLAMPING RANGE    1 5 to 2  5mm       Clamping Torque    Always use the correct size setting anvil for the sample thickness  If the sample does not fit properly in the  tool  erroneous data may result  Adjusting the sliding clamps to the proper tightness is imperative  A torque  screwdriver is included with this tool  For each material  some experimentation may be required to find the  best clamp torque value to obtain good results  Under tightening  or for softer materials  over tightening the  clamps will result in erratic data  Once a good torque value is obtained for a specific sample  material and  thickness   all subsequent samples should be tightened to the same value  Also  both clamps should always be  tightened to the same torque value     NOTE  Loading soft samples or samples that do not properly fit the clamps  can result in inaccurate data     Operating Ranges    Operating range is defined as the region bounded by the maximum and minimum complex modulus G that  can be measured by each transducer type using the torsion rectangular
215. scribes proprietary technology patented by  TA Instruments   Waters LLC  U S  Patent No  6 431 747      Method and Apparatus of Modulated Temperature Thermogravimetry  MTGA M5  describes proprietary  technology patented by TA Instruments   Waters LLC  U S  Patent Nos  6 336 741 and 6 113 261      Modulated Temperature Thermomechanical Analysis describes proprietary technology patented by TA  Instruments   Waters LLC  U S  Patent No  6 007 240      Method and Apparatus for Parsed Dynamic Differential Analysis describes proprietary technology patented  by TA Instruments   Waters LLC  U S  Patent No  5 474 385 and EP Patent No  0701122      Method and Apparatus for AC Differential Thermal Analysis describes proprietary technology patented by  TA Instruments   Waters LLC  U S  Patent No  5 439 291       continued on next page     ARES User Manual 5      TA Instruments Patents  continued     Method and Apparatus for High Resolution Analysis of the Composition of a Material describes proprietary  technology patented by TA Instruments   Waters LLC  U S  Patent No  5 368 391 and 5 165 792  Additional  Patent Nos  CA 2 051 578 and DE  EP  FR  GB  IT 0494492      Method and Apparatus for Thermal Conductivity Measurements describes proprietary technology patented  by TA Instruments   Waters LLC  U S  Patent No  5 335 993 and EP Patent No  0634649      Dynamic and Thermal Mechanical Analyzer Having an Optical Encoder with Diffraction Grating and a Linear  Permanent Magnet Motor describes
216. sducer is a STD transducer  go  to step 19    a  Access the Set Transducer Characteristics form  Figure 5 7  by selecting the Transducer option  from the Service function of the Utilities pull down menu    b  Divide the displayed Torque Calibration Value for the high range transducer by 10 then enter  this value  into the form  as the Torque Calibration Value for the low range transducer  For  the 1K FRTN1 transducer only  divide the high range Torque Calibration Value by 50 and  enter this value for the low range transducer setting     19  Click Ok  Proceed to the Normal Force Calibration if desired  If Normal Force Calibration is not to be  performed  remove the calibration tool  weight  and pulley  and store them in the calibration kit     This concludes the Torque Calibration     Table 5 3  Torque Calibration Weights  Applied Torques  and Full Scale Values    CALIBRATION TORQUE CALIBRATED FULL  TRANSDUCER WEIGHT TORQUE VALUE SCALE VALUE   applied   displayed   computed     2K FRIN1 200 grams 500 gecm 500   2 gecm 2100   5   2K aaa E  p n 613 01221   498 to 502   1995 to 2205   2K STD    1K FRTN1 200 grams 500 gecm 500   1 gecm 1050   5    p n 613 01221   499 to 501   997 5 to 1102 5     100 FRTN1 20 grams 50 gecm 50   0 1 gecm 105   5   100 FRT  p n 613 02775   49 9 to 50 1   99 75 to 110 25   200 FRTN1 20 grams 50 gecm 50   0 2 gecm 210   5   200 FRT  p n 613 02775   49 8 to 50 2   199  5 to 220  5   10K STD Two   1000 gram 5000 gecm 5000   10 gecm   10 500   5     p n 6
217. se the stage to the loading position    2  Verify that the motor is on  and in dynamic mode    3  Mount the upper and lower tools on the actuator shafts    4  Using the Set Gap Instrument Control function in Orchestrator  zero the normal force and torque on  the motor using the  Offset Normal Force to Zero  and  Offset torque to Zero  buttons    5  Using the stepper control buttons  lower the stage to a point where the tools are close but not  touching    6  Use the  Motor Position Offset  button in the Set Gap Instrument Control function to ensure that the  upper and lower tool openings are aligned    7  For 2K FRT transducers only  skip this for STD transducers  use the  Zero Fixture  button in the Set  Gap Instrument Control function to bring the tools together and determine the zero point for the test  tools    8  Raise the stage to provide sufficient room for sample loading     Sample Loading    This procedure is for use with 2K FRT and STD transducers only  Refer to Figure 4 12 during the following  procedure     0  WARNING   POSSIBLE PERSONAL INJURY    POSSIBLE DAMAGE TO INSTRUMENT    This is a high torque motor  Turning on the motor while in dynamic  mode causes the motor to snap to dynamic zero position at a high  velocity  This can cause severe damage to the transducer and or  personal injury  To avoid damaging yourself and the transducer     Never turn on the motor while a sample is loaded     Keep hands clear of the motor     1  Measure and record the following sa
218. seal degradation and corrosion     80 ARES User Manual    Peltier Configuration in Orchestrator     When you command a temperature  Orchestrator software uses the Instrument Configuration to determine the  environmental system currently in use  and establish operating conditions  Prior to operating the fluid bath   access the Instrument Configuration function  located under the Service function of the Utilities menu  and set  up the instrument using the guidelines shown in Figure 2 31     Setup Instrument Options    Temperature Control na    Peltier  Thermopile heat pump    E    150 0   30 0    ol       Figure 2 31  Setup Instrument Options form used to input the Peltier system configuration     NOTES    1  Make sure the following are selected  e Instrument Setup   TEMPERATURE CONTROL  e Temperature Control   PELTIER  THERMOPILE HEAT PUMP    2  Ensure that the maximum and minimum temperatures are set as follows   e Maximum Temp    150  C  e Minimum Temp    30  C   3  For normal applications set the Temperature Calibration Table to Default    4  For critical temperature work  the adjustable option can be used to enter  calibration  offsets  for specific temperature  setpoints   Refer to the  Orchestrator online help for details on how to set up this feature     ARES User Manual    PID Loop Setup    When using environmental control systems  errors between commanded and actual temperature are driven to  zero by a PID  Proportional   Integral   Derivative  loop control system  Th
219. sed to ensure consistent fluid  volumes from sample to sample  The  well  is filled through the hole in the center of the cover  fill port in  Figure 4 8      Even though a low viscosity fluid is used  it can affect the test  Basically  the sealing fluid can act like a sample   For the most demanding work  it may be desirable to run preliminary tests on just the sealing fluid  to    ARES User Manual    determine its  characteristics  Once the sealing fluid influences are known  they can be removed for the sample  data     To determine the effect of the sealing fluid  set up the tool exactly as if you were running a sample  but without  actually loading a sample  For these tests  air will be the sample  Set the gap to the intended setting  and fill the  sealing well with the desired volume of the chosen sealing fluid  Run the actual tests that you are going to use  on your real samples  The results are then used to correct the real sample data     Use with the Oven    When using the Hastelloy tool with the oven some minor modifications to the oven may be necessary to  facilitate sliding the oven past the tool  Referring to Figure 4 7  shave the edge of the oven foam to allow the  oven to slide past the tool during opening and closing  Only a small amount of material should need to be  removed  approximately 1mm   While adjusting the clearance  be careful that you do not force the oven  or  tool  to the point that either is damaged  Also  pay attention to the position of the PR
220. st Set up Screen    ARES User Manual 123    Suggested Uses  Suggested uses of strain relaxation are as follows   e Determination of time required for a sample material to relax after a deformation  as in sample material    loading    e Analysis of time dependent behavior of a sample     Test Options  e Delay Before Test    e Steady Preshear  e Analog Data Input    ARES User Manual    Multiple Extension Mode    Functional Description    Multiple Extension mode offers four zones in which a variety of separate extensional test types can be  performed  The mode of extension can be based on linear rate  Hencky strain rate  related to the specimen  geometry   force imposed to create the extension  or as a compressive Rim Shear mode  also known as squeeze  flow      The set up screen for multiple extension mode  Figure 3 16  requires you to select a zone time for each zone   This is the time in seconds minutes hours over which the test type will occur in each zone  The extensional  value units are dependent upon the test mode selected     Several different test options are available in this mode  Any test type can be combined in any sequence  during a multiple extension mode test  A fifth test type selection that is available is    End Test     which halts  the series of extensional mode testing  Table 3 4 provides a summary of the different test types available and  the uses of each of them     Multiple Extension Mode Test kE x     Temperature         25 0 PC  Max 500 0  C Min  150
221. strator software  HOST is  connected to the Host Computer serial communications port COM1     Communications interface between the optional fluid bath environmental control   CIRCULATOR circulator and the instrument    RS 232 Accepts a communication link that uses RS 232 protocol  used with OAM2 and  DETA options     OVEN Electrical interface between the oven switches and temperature sensors and the  instrument  Connected to the oven     ARES User Manual    HOST       Power Panel    For Test Stations equipped with an oven  the Power Panel serves three functions   1  Electrical power interface for the Test Station   2  Electrical power interface to the oven   3  Electrical power interface to the optional LN2 Controller     For Test Stations without an oven  the power panel is not required  Main power plugs directly into the Main  Power Switch  shown in the next section      Figure 2 13 identifies the Power Panel connectors            220V  IN        0     O oO  OVEN N   IN HEATER  o o O    o       000000000000000000000000  000000000000000000000000  00 000000000000 0000000000  000000000000000000000000  00000000 0000000000000000          O       Figure 2 13  Power Panel     ARES User Manual    Power Connections    Power connections for the system are made as outlined in Table 2 2  Input power requirements are 220 VAC   20A  50 60 Hz     Table 2 2  Power Panel Connectors    220V IN 220 Volt input power supplied   Main power to the system        by customer    220V OUT Connected t
222. t into the  linearity of the measurement     After all frequency components are specified  the implementation of single point measurement test and  temperature ramp differ  The MultiWave Temperature Ramp test requires additional input to specify up to  eight zones of temperature control  See the section on the Temperature Ramp Test for more information     Once the strains and frequencies have been specified the resulting waveform must then be downloaded into  the instrument  The    Wave    button is used to compute the resulting waveform  display the results  graphically  and provide the option to download directly into the instruments waveform memory  Any  subsequent change to the strain or frequency parameters requires recalculating and re sending the waveform   The software keeps track of any additional changes  and when the test is started  reminds you to resend the  waveform  if necessary            Hult ave Single Point Test H E    Fundamental Frequency    Frequency sof      rad s  Max 2 0 Min 1 00e 05  Strain     0 1    Max 31 2  5000 Min 0 003125  Temperature  25 0 PC  Max 600 0  C Min  150 0  C       Harmonics 1 2 3 4  Harmonic    3  5  o  o  strain     0 2  0 2  0 0  0 0       Options PreShear Off Dela OF Analogin OH MeasOps  Otf      Ok   Options   End of Test   d   Save As   Help   Cancel      Figure 3 8  MultiWave Single Point Test Set up Screen    ARES User Manual 113    Hult ave Temperature Ramp Test       Figure 3 9  MultiWave Temperature Ramp Test Set up Scr
223. t to the Stress Relaxation experiment  which  uses the same type of deformation  but monitors torque and strain instead of torque and normal force  The  direction selected is the rotational direction of the actuator for positive Strain values  Data can be taken in  either of two modes     Log  Logarithmic     Logarithmic sampling takes data at logarithmically incremented intervals  resulting in equally spaced data  points when plotted as a function of logarithmically scaled time  the number of points taken is inversely  proportional to zone time   As an example  selecting 5 points per zone during a 100 second zone divides the  difference of the endpoint times  logarithms  into five equally spaced fractional exponents  Six data points are  measured  in succession  at times determined by taking the antilogarithm of each exponent  10  15 9  25 1   39 8  63 1  and 100 seconds     Linear    Linear sampling takes data at linearly incremented intervals  resulting in equally spaced data points when  plotted as a function of linearly scaled time  This technique is useful for relatively short zone times  where  linear time scaling is practical  As an example  selecting 5 points per zone during a 100 second zone results in  five data points measured at linear increments  20  40  60  80  and 100 seconds     Torque Normal Relaxation Test H ES  SUTIN occ foa    Mas 31 2 5000 Min 0 0031 25  Temperature         25 0 PC  Max 600 0  C Min  150 0      sampling Mode    Log    Linear  Points Per 
224. tation  48  temperature control  58    Oven PRT  65    244    ARES User Manual    P    Parallel Plates  test tool   constants  161  165  tool installation  163  171  general information  161  165  operating range  161  165  sample loading  163  171    patents  5    Peltier  description  76  humidity cover  83  installation  78  operating specifications  30  operation  83  PID loop setup  82  selecting operating range  77    Phase Angle  description  94  shift  9   94  Platinum Resistance Thermometer  See PRT  Pneumatics Panel  connections  54  description  53  location  47  Polymer Melts  150  Power Panel  connections  51  description  50  location  47  PreTension  See AutoTension    PRI  27  58  65  224  233    Pseudoplasticity  94    R    registered trademarks  5    Regulatory Compliance  14    ARES User Manual    Rim Shear  125    S    Sample PRT  65  Service and Repair  230  Shear Rate  93  Shear Stress  93  Shear Thickening  See Dilatancy  Shear Thinning  See Pseudoplasticity  Signal Panel  connectors  49  description  47  general connections  47    location  47    software  license  3    Squeeze Flow  125  Stage  description  37  location  34  Stage Control  manual  37  rate adjustment  39  software  38  Steady and Transient Measurement Formulas  205  Steady PreShear  140  Steady Rate Sweep  118  Steady Single Point  116  Steady Step Rate Temperature Ramp  135  Step Shear Rate  121  Stiffness  150    Storage Modulus  See Elastic Modulus    ARES User Manual    Strain  93
225. te the G  MAXIMUM value into equation 4 2  then determine the upper limit of operation by  calculating N  at each O     values chosen to be from the lowest to highest frequencies within the  transducer operating range     3  Substitute the G  MINIMUM value into equation 4 2  then determine the lower limit of operation by  calculating N  at each       4  Generate an X Y scatter plot of complex viscosity N   Y axis  versus frequency     X axis      The region between the upper and lower limits of operation is the range of complex viscosity that can be  tested  Appendix 1 contains a table of G  values for the double wall Couette for some transducers and a  standard motor     NOTE The cup installation and use procedures are different for the original  fluid bath and the fluid bath 2  Please ensure that you identify which bath and  Couette cup you have  and follow the correct procedures for your specific  bath cup arrangement     Tool Installation  Original Fluid Bath    The Double Wall Couette lower tool  cup  mounts into the fluid bath or fluid bath 2  see next section  to ensure  precise thermal control  Install the bath on the test station prior to mounting the tool in the bath        _    CAUTION  Never place any lower tool into the bath if the temperature  0  of the lower tool is cooler than that of the bath  Placing a tool into a  warmer bath will result in expansion of the tool during use  After  expansion  the tool may not be removable without damaging your bath        We
226. that may make testing in one  geometry preferable to testing in another  Additionally  factors such as anisotropy and differences in strain  dependence may yield inconsistent results for different geometries     Recommendations for selection of a geometry based upon sample type are as follows     Thin Films or Fibers    Test thin films such as magnetic recording tape  or fibers  using the fiber film tool  Enter the test geometry  dimensions into the appropriate Tension geometry screen     Fluids  Suspensions and Emulsions    Low viscosity fluids  or suspensions of limited stability can be tested using either the Couette or Double Wall  Couette geometry  Higher viscosity fluids  and thicker suspensions and emulsions can be tested using parallel  plates  or cone and plate geometries     Solid Samples  Including Thermosets  Thermoplastics  and Elastomers    These materials can be tested using the Torsion Rectangular tool  Several inserts are available to accommodate  a variety of sample thicknesses     ARES User Manual 149    Polymer Melts and Soft Solids    Melts can be tested using the parallel plate or cone and plate geometries     Thermosetting Resins and other Curing Studies    These materials are best tested in a parallel plate tool  Disposable plates are available for curing studies  or  other tool destructive materials     Testing Limits and Compliance    Definition of Compliance and Stiffness    For this discussion  compliance is defined as displacement  in radians  per
227. the Control pull down menu of Orchestrator  Figure 2 18      ARES User Manual    Set Test Conditions Form k E    Change Test Parameters    Temperature oo    eeeeseseseees  5 0 PC  Masw 500  0   Min  1 50  0    AutoTension Sensitivity             0 0  g   Enable N2Gas Switch             C No  amp  Yez    Switch from Nz to Gas Above  45 0  FE   Switch from Gas to We Below  30 0  FC        Figure 2 18  Set Test Conditions Form Showing Gas Switching Enabled  Note that N2 in the form refers to liquid nitrogen     Oven Operating Requirements    The oven can operate only if the following conditions are met     e The oven is selected as the current environmental system    e The environmental controller is turned ON in the Instrument Control Panel    e The oven must be positioned all the way to the right  and the oven must be closed    e Either air or N  must be supplied to the oven at a pressure greater than 35 psi    e If using the LN2 Controller  the liquid nitrogen  LN2  level in the Dewar Flask must be between 50 and  75  of capacity  When it is  the Orchestrator Online Indicator LN2 Ready is displayed     ARES User Manual    Oven Configuration in Orchestrator     When you command a temperature  Orchestrator software uses the Instrument Configuration to determine the  environmental system currently in use  and establish operating conditions  Prior to operating the oven  access  the Instrument Configuration function  located under the Service function of the Utilities pull down me
228. the delicate high precision bearing surfaces of these components  Damage to the bearing  surfaces will result in faulty measurements  and if significant enough  the damage will require the replacement  of the entire motor or transducer     The relative humidity of the supplied air should be 35  to 70  with a dew point of 10 deg C  We highly  recommend installing the optional air dryer filter between the test station and the air supply  If your air  supply has excessive moisture levels  which result in the immediate condensation into water  an additional  water trap will be required before the air dryer    Excessive moisture in the air supply will damage the test station     CAUTION  Read the operating and maintenance instructions that were  supplied with your air dryer  Failure to properly operate and maintain  your air dryer will result in extensive damage to this instrument        Because of the critical nature of the air supplied to the instrument  and the potential for expensive damage  through mishandling  we highly recommend that you inform your laboratory manager  or compressor  maintenance personnel  of your instrument requirements in detail  You should also ask to be informed before  any air supply interruption  or compressor maintenance  so that you can properly shut down and protect the  test station  Typically after any compressor maintenance  there will be some residual particulates and moisture  present in the supply lines  You should disconnect the air from the
229. the stage to the loading position     Place the top cover onto the upper tool shaft and hold it in place using the clip  The cover should be held  near the top of the tool to allow access to the upper plate for mounting and sample loading     Verify that the motor is on  Mount the lower tool PRT  as described in your main instrument manual  if  desired  Mount the upper and lower tools on the actuator shafts     Using the Set Gap Instrument Control function in Orchestrator  zero the normal force and torque on the  motor using the  Offset Normal Force to Zero  and  Offset Torque to Zero  buttons     Using the stepper control buttons  lower the stage to a point where the tools are close but not touching     Use the  Zero Fixture    button in the Set Gap Instrument Control function to determine the zero point for  the test tools     Raise the stage to provide sufficient room for sample loading     Sample Loading    The recommended gap setting for parallel plates is between 0 5 and 2 millimeters  For the Hastelloy tool  the  maximum sample gap is 2mm  Beyond that  the upper toll will rub on the cover  Figure 4 8 shows the  Hastelloy tool with a sample loaded and the cover installed     1   2     Place the sample on the lower plate  Ensure that the sample is centered on the tool     Using the stepper motor buttons on the right side of the test station  adjust the sample gap until the upper  plate is close to the specimen  Set the gap using the Set Gap Instrument Control function  e
230. tor S SS       EN ARES User Manual    Notes  Cautions  and Warnings    Throughout this manual  the following terms and symbols are used to draw attention to specific situations     NOTE  A NOTE highlights important information about equipment or  procedures     0  CAUTION  A CAUTION emphasizes a procedure that may damage  equipment or cause loss of data if not followed correctly          WARNING  A WARNING indicates a procedure that may be hazardous    to the operator or to the environment if not followed correctly     ARES User Manual 9      Safety    Do Not Attempt Service    Do not attempt to service this instrument  as it contains no user serviceable components     Required Equipment    While operating this instrument  you must wear eye protection that either meets or exceeds ANSI Z87 1  standards  Additionally  wear protective clothing that has been approved for protection against the materials  under test and the test temperatures     Safety Notices    The following notices are intended to draw your attention to situations that pose a risk to either your personal  safety or the safety of the instrument  Although these notices appear at relevant points throughout this  manual  they are repeated here for emphasis  Additionally  all safety notices that appear on the instrument are  reproduced on page 13     01  CAUTION  Read the operating and maintenance instructions that were  supplied with your air dryer  Failure to properly operate and maintain  your air dryer will resu
231. tored by the ENVIRONMENTAL  CONTROLLER  that receives its commands from the temperature controller  TEMP CONTROLLER  located within  the control computer     When a test temperature is selected at the HOST COMPUTER  the CPU configures the TEMP CONTROLLER to output  a command  TEMP CMD AND MEASURE  to the ENVIRONMENTAL CONTROLLER  This command combines with the  heater power feedback signals to increase or decrease heater power as necessary to control oven temperature   To monitor actual oven temperature Platinum Resistance Thermometers  PRT  are installed in the oven  Two  PRIs monitor the air temperature by each heater and a third PRT  TOOL TEMPERATURE  monitors the  temperature of the lower sample tool in use  The temperature controller electronics monitor PRT resistance to  determine the actual internal oven temperature  A difference between the actual temperature as sensed by the  PRIs  and the commanded temperature that is entered by you at the HOST COMPUTER  results in the generation  of an error signal by the ENVIRONMENTAL CONTROLLER electronics  TEMP CONTROL AND FEEDBACK   This error  signal either raises or lowers the internal oven temperature until the PRT sensed temperature and the  commanded temperature agree     When using the Fluid Bath  the circulator is under RS 232 control from the Test Station  Either the fluid in the  circulator  using the circulator s internal PRT   or the lower test tool itself  using the lower tool PRT  can be    maintained at the desire
232. tropic loop test available on Shear  Strain Controlled instruments     Options  The following test options are available for use with Stress Ramp     e Delay Before Test    ARES User Manual 139    Test Options    This section describes the following test options that you can select while programming selected dynamic    tests     e Steady PreShear  e Delay Before Test    e Auto Tension Adjustment    e Analog Data Input    e Auto Strain Adjustment    e Measurement Options    Steady PreShear    Steady PreShear allows you to subject the sample to a steady shear deformation prior to the start of a dynamic  test  If the Delay Before Test option is selected  Steady PreShear occurs prior to the delay  No data are taken  while the pre shear is being applied     The following parameters are set through the Steady PreShear Set Up Screen  Figure 3 24      PreShear Rate    Shear rate of the PreShear stress  Positive PreShear Rates result in clockwise actuator rotation  Negative  PreShear Rates result in counterclockwise actuator rotation     PreShear Time    The length of time the PreShear is applied            Dynamic Temperature Step Test    Uptions     Steady PreShear  Delay Before Test  AutoTension Adjustment    AutoStrain Adjustment    pa Mio Ha     Measurement Options    PreShear Rate                   fa  T  1 3  Max 2500 000 Min 0  025000    Pres hear Time                    45  e or hom      Positive Rate  Cf Negative Rate  CEW     Figure 3 24  Steady PreShear Set Up Screen    AR
233. ts es eaaeeeaoe ona meanaeets 133  Fore a AA tE E err rr rere ere 134  PSS a o AAA E ct E Eo O T A S 134  PS o E E ness  134  USO TONG inte fisio  134  oleada tep kale Tempera i e Nai 135  Descrip Otesi onin oro E E E E T E E EEO 135  Saue o MOE A oo 5 E E corse pucainosonsaleases oe  136  a mo 191  A NE E E E E E ane ca eee taenaaeese 136  Sess ontrolled Transient Test Melodia 137  Sonae CSS CSU E E E E T E A E 137  Funcuonal D S WOM ea a E E R E E E E R EAA E opa 137  Saue Us nio noise 138  OPUS testi 138  MESS INN  TEO ias 139  Descanso A E E E ai E E sahersbanate Rndaaansets 139  Sa D E E E A E A E E E A E E 139  Deon tcc o o 139  T EOD mapeo E E E E E E A E AE EEN E EE EEEE eipiro adoro moros 140  US My as  140  Delay STOR CS E E E ae eee eee 141  Mantel Dean ici 141  APP AO Ens icon 141  Automatically Start Test When On Temperature   oooncnocnocnncnnononnnnnanncnnanonancnnnnanancnnnnnnnonncnnancanncnnnonnancancananins 141  AU Wiis Ol AU O aa  142  ae e E RE 144  PUT ue ON OP aps soem E EEA desea E E EO AE EA E AAEE EE A TA 145  Iie Ri mont Op 11019 a E E E EA 146  Sea EN EAE E E O 146    ARES User Manual    ES Cy Cl Comi stereo 146    SO A ie o o AAA EU OO A A 146  Chapters  Test Geometries and FOIS Sun d one 149  TRU cig creat cts ts apart c o e rn ht aatuntanen eagancnrtenees 149  General Test Tool Morna UON AAA o O aa AARTE NE a a O 149  General Recommendations for Geometry Selection           ssessssessssessssessrsessssesesstsersrsesrtstsresrsrrnenesnrsesresenessese
234. ttons    5  Using the stepper control buttons  lower the stage to a point where the tools are close but not  touching    6  Use the  Motor Position Offset  button in the Set Gap Instrument Control function to ensure that the  upper and lower tool openings are aligned    7  For 2K FRT transducers only  skip this for STD transducers   use the  Zero Fixture  button in the Set  Gap Instrument Control function to bring the tools together and determine the zero point for the test  tools    8  Raise the stage to provide sufficient room for sample loading     ARES User Manual    Sample Loading    Refer to Figure 4 10 and Figure 4 11 during the following procedure     WARNING  This is a high torque motor  Turning on the motor while in  ES dynamic mode causes the motor to snap to dynamic zero position at a  high velocity  This can cause severe damage to the transducer and or  personal injury  To avoid damaging yourself and the transducer        Never turn on the motor while a sample is loaded     Keep hands clear of the motor     1  Measure and record the following sample dimensions     e Width   e Thickness   e Length  STD transducers only   length will be determined from gap setting for 2K FRT  transducers     2  Select a matching pair of setting anvils  based upon sample thickness  and secure them in the upper  and lower tool  Please note that each setting anvil is machined in such a way as to provide mounting  for two different sample thickness ranges  A  nominal  thickness  roughl
235. umatic input to the Oven Pressure  Sensor  which is used to monitor oven gas  pressure  The gas pressure must be above  35 psi for the oven to function     Pneumatic input to the Oven that accepts air   which is circulated throughout the Oven in  order to maintain positive pressure and  minimize frost when using LNo        ee ee      Gas Supply to      OVEN      Selector Switch    Position      for AIR   i    SSeS SSeS SS SS    ses SS SSS    Fosition for i   gt  N2AGAS    SS SSS SPS SRS SS SPSS SSS SSeS eS ee ee     Panel Side     ADUCER  RSA 111    40 PSI    XDUCER  BENDIXFRT    35 PSI    MOTOR    60 PSI    MAIN  50 PSI        SZ    5 5 BAR        Panel Back     Figure 2 15  Pneumatics Panel     TRANSDUCER  Air Pressure  Adjust Knob    MOTOR  Air Pressure  Adjust Knob    OVEN    Air Pressure  Adjust Knob  OVEN    PRESSURE  SENSOR    N2 GAS  70 PSI    YN    This panel has 2 sides  and wraps around the right rear corner of the Test Station     ARES User Manual    Environmental Control Systems    There are three environmental control systems available for use with ARES  Each system is used to precisely  control sample temperature  The three systems are a forced air convection oven  Oven   a re circulating fluid  bath  and a rotating  oscillating Peltier     The air convection oven has a dual element heater with counter rotating air flow for a wide temperature range   ambient to 600  C  and maximum temperature stability  If temperatures below ambient are required then  either th
236. unting the Peltier Assembly  perform the following actions on the ARES instrument     e Raise the Stage to maximum height    e Remove all Upper and Lower Test Tools  and loosen the Anvil Tightening Knob on the Motor Anvil    e Thoroughly inspect the Test Tool mounting surfaces  i e   inspect the transducer anvil and the motor anvil  and clean off any material that may interfere with the mounting of the Peltier  This is essential to ensure  proper mechanical mating between the Peltier Assembly and the instrument    e Turn off the Motor     Refer to Figure 2 29 during the following installation procedures     1  Remove the protective plastic base from the Peltier Assembly Collar by placing the two pins on the  spanner wrench  provided  into two of the holes machined into the Collar  and rotating the wrench  counterclockwise     2  Gain access to the Peltier Assembly Shaft by sliding the cover of the Peltier Assembly fully upward     3  Hold the Peltier Assembly above the ARES Motor Anvil with the Bath Hoses facing toward the right  of the instrument     4  Rotate the Peltier Assembly Shaft to align the flat portion of the Shaft with the flat portion of the ARES  Motor Anvil  both flats should be facing toward the right as you face the instrument   At this time   the red dot on the PRT Plug should be facing toward the front of the instrument     CAUTION  As you lower the Peltier Assembly in the next step  the  01  Peltier Assembly PRT Plug will be inserted into the ARES PRT   Re
237. ver    ARES User Manual    Fluid Bath    Description    The Fluid Bath offers precise control of sample temperature using an open fluid re circulant system  The  lower test tool is mounted within the Bath Well  around which flows thermally controlled fluid supplied by a  circulator  The temperature of the lower tool is measured by the bath PRT  which mounts through the Bath  Well into the Motor  You can choose to control the temperature of either the lower tool or circulator fluid     The Circulator regulates the temperature of the bath fluid  and pumps the fluid through the Fluid Bath  The  circulator  as supplied by TA Instruments   is connected to  and is under the control of  the test station and  software  The fluid circulated through the bath is maintained at the temperature selected in Orchestrator  The  circulator has its own fluid temperature regulation  which can optionally be used as the temperature control  loop for the Bath     Installation of Fluid Bath    The fluid bath is mounted onto the motor using a threaded collar   Refer to Figure 2 35 while performing the following steps to install the bath     Raise the stage to maximum height and remove the upper test tool    Slide the Oven all the way to the left    Turn off the Motor    Position the motor anvil with the knob facing the front of the instrument    Position the bath with the access port facing the front of the instrument  and the alignment pin  which   is located on the inside diameter of the bath housi
238. vides the difference of the endpoint times  logarithms  into five equally   spaced fractional exponents  Six data points are measured  in succession  at times determined by taking the  antilogarithm of each exponent  10  15 9  25 1  39 8  63 1  and 100 seconds     Linear    Linear sampling takes data at linearly incremented intervals  As an example  selecting 5 points per zone  during a 100 second zone results in five data points measured at linear increments  20  40  60  80  and 100  seconds     Select a strain  and direction  which is maintained throughout the entire test  Figure 3 15   and then a sample  time  zone time  for each of the four zones  The times can be anywhere from 2 0 to 1 6x10  seconds in length   Usually  the first zones are set at very short time periods  typically from 2 0 to 10 seconds   as most of the  relaxation happens very quickly  with subsequent zones being set to longer times  Please note  that regardless  of the field inputs  the fastest the instrument will take data is 1 data point every 3 msec     Stress Relaxation Test E ES       SEAMOS sossoscoooscso0ss fico    Max 31 2 5000 Min 0 003125  Temperature        IET PC  Max 600 0  C Min  150 0  C  Sampling Mode  Log     Linear  Points Per Zone f200    Max 350 Min 20  Zone Number 1 2 3 4  Zone Time  s or e aa fc IS  Directora f Clockwise O Counterclockwise       Options  FreShear Ot Delay Ot Analogln OFf      Options   End of Test   Save As   Help   Cancel         Figure 3 15  Stress Relaxation Te
239. when very stiff samples are tested  and the transducer  displacement becomes close to the motor displacement  In this case  because the difference between the two  displacements is small  the resulting relative error is large and of similar magnitude as the measurement     If the measured strain value is significantly smaller than the commanded strain value  the data are likely  affected by transducer compliance  As a practical guideline  measured sample strain should be at least 30   the commanded strain  Although measurements can be taken below these limits  you are cautioned that  accuracy may be affected  Measurements that are affected by transducer compliance typically report modulus  values that are lower than the true modulus  One method of determining if transducer compliance is affecting  the data is to switch to a different geometry and compare the results to the first tests  If the data are unaffected  by compliance  the results from the two geometries should be nearly identical     Sample compliance  or stiffness  is related to both the modulus and geometry of the sample  Since the  modulus is fixed  the sample dimensions are normally adjusted  to make the sample less stiff   or the geometry  is changed altogether  to obtain the desired sample compliance  It is critical that the sample compliance is  within the operational range of the instrument otherwise inconsistent or incorrect results will be obtained     ARES User Manual    Determination of Operational Range
240. which the fundamental is multiplied by to determine the frequency for a given  harmonic  A value of 0 0  zero  is used to indicate that that zone is not used     Strain    This is defined as the strain amplitude of the given harmonic  Note that a value of 0 0  zero  can be used to  signal the instrument to make a measurement at the given harmonic without any applied strain at that  frequency  which can be used to monitor distortions in the stress sine wave that can indicate nonlinear  behavior      The MultiWave setup menu is displayed following selection of either single point  Figure 3 8  or temperature  ramp  Figure 3 9  test methods  Here  you must select a fundamental frequency between 2 rad sec and 1x10e  rad sec  0 318 and 1 59x10e   Hz   The strain level for the fundamental frequency must also be entered     5    ARES User Manual    You can select up to seven harmonics  multiples of the fundamental  and assign a strain to each harmonic  As  a general rule  the sum of all strains should not exceed the linear viscoelastic region of the material  A strain  sweep can be used to determine the limits of linear behavior if that information is not known  If a value of  zero is entered for both the stain and harmonic  then that column is not used  If a nonzero frequency is  entered  with a strain level of zero  data will still be acquired at that frequency  This provides a means for  measuring the amount of harmonic stress generated by other frequencies  and can provide insigh
241. x   When the checkbox is selected  the  instrument reads  and stores  the actual gap immediately prior to the start of a test  It then rechecks strain  error limits using the actual gap  overriding the gap entered in the specific geometry form   Out of range  limits are then reported     For Stored Geometries  the  Read Test Fixture Gap  checkbox is displayed only if you saved a Stored Geometry  while the  Remote Gap Monitoring  option was enabled  using the Instrument Configuration function      Displaying the Instrument Gap    The gap read from the instrument can be displayed as the online parameter CurrGap   Max Allowed Force While Setting the Gap    The  Max Allowed Force  option  when setting the gap  provides two benefits  First  it allows an operator to  set the gap in a repeatable  documentable way  Loading the sample the same way each time leads to more  reproducible results between different operators  or with different geometries  Additionally  it helps avoid  internal pre stressing a sample during loading  which again  could lead to erratic or inconsistent results  This  feature is also helpful when loading soft samples that may squeeze out from between the test tool if to much  force is applied     Setup Instrument Options El ES    Instrument Testing Limits    Instrument Setup       Stepper or Linear Motor    Ho 0 Yez  Autoranging Transducer    f No       Yes  Normal Force Measured ooo     No f amp   Yez  Remote Gap Monitoring    f No 0    ves  AOA Option Connect
242. x Viscosity Viscosity       In phase Viscosity u  sin d   Out of phase Viscosity u  cos 0        Measurement Method    During dynamic mechanical testing  the ARES control computer makes a digital cross correlation of measured  strain and force by comparing the amplitude and phase shift between the imposed motion  strain  and the  force  stress   When a test is started  the computer measures strain and force 2 048 times to determine the  average amplitude and phase shift of both  The measurements are made relative to two reference sine waves   command  of fixed amplitude  and having phase angles of 0   and 90    as shown in Figure 3 14     The result is a strain and force phaser relative to the reference as shown in Figure 3 1B     Using fundamental geometric techniques  the phasers in Figure 3 1B can be rotated so that strain becomes the  reference axis as shown in Figure 3 1C  The force vector can now be defined in terms of an in phase and out   of phase component of force proportional to angle 6     Using equations appropriate to the geometry under test  the average force phaser is converted to stress and the  average angular strain phaser is converted to percent strain  Dividing the stress by the strain produces the  complex modulus  G   which indicates the total energy required to deform the material  Multiplying G  by the  cosine of the phase angle gives the in phase component of the stress  G     which is proportional to the energy  stored elastically  Multiplying G  by
243. y the center of the clamping  range  is stamped on two of the anvil faces  opposite sides   The setting anvil should be mounted such  that the desired  nominal  thickness is visible from the outside  back  of the tool  opposite the actual  sample     3  Place the sample into the lower tool  Center the sample in the tool using the reference lines scribed in  the setting anvil and sliding clamp  Partially tighten the clamp  using the adjusting screw  to hold the  sample    4  Lower the stage until the upper tool is about 1  4 inch from the sample    5  Use the  Motor Position Offset    button in the Set Gap Instrument Control function in Orchestrator  to  radially align the sample with the upper tool if necessary     CAUTION  In the next step  do not generate a Torque or Normal Force  greater than 50  of full scale  Failure to observe this caution may result  in damage to the transducer     6  While confirming the sample fits into the upper tool  lower the stage until a compressive  downward   Normal Force of about 10  of full scale is generated  If the sample is not aligned properly  re raise the  stage and carefully realign the sample and tool using the  Motor Position Offset  button    Ensure  visually  that the sample is completely inserted into the tools    8  Tighten the lower and upper sliding clamps  using the adjusting screw  to the desired torque using  the torque screwdriver    9  Raise the stage until a force of approximately 10  of full scale is generated  Please n
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Samsung 2433BW Εγχειρίδιο χρήσης  MANUAL 2256010 Walther PPQ AUG12 WR.indd    TOSHIBA MINI NB500  Ingenico IPP320  Samsung Galaxy Grand Neo Duos Εγχειρίδιο χρήσης  Trullo - Laviers et Muraillers de Bourgogne  DeLOCK 0.3m SATA III    Copyright © All rights reserved. 
   Failed to retrieve file