Home
        IS4K User Manual
         Contents
1.            503   J l i L i    06 00 00 PM 06  00 50PM 06 01 40 PM 06 02 30PM 06 03 20PM 06 04 10 PM 06 05 18 PM    pe ae    EET   Communication Initialized O   LED OFF  LoggmgOFF jenn    Figure 18     Temperature Window       Main Menu Navigation  Minimize or close window  Shows signal value    Main charting area    fo e Noa    Status Bar     shows if communication has been initialized  number of data packages via the  COM port  whether the LED illuminators on the optical head are ON or OFF  and if Logging is  ON or OFF     Before using this instrument you need to perform other setup steps  see the section on Data    SAVA T    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 20 of 51           Engines for Thin Film Innovation PROPRIETARY INFORMATION  Managing Graphs and Windows    Each of the instrument windows may display a strip chart graph showing the recent history of the  measured data        Chart data can be analyzed using the    Graph Pallet        To    zoom into    the data displayed  right click in the graph area and de select    AutoScale X    and or     AutoScale Y    This will cause the display to stop displaying new data  making it easier to zoom into the  data of interest     Note  Remember to turn AutoScale back on 
2.     Reflectance Calibration Window     Low                SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 30 of 51           Engines for Thin Film Innovation PROPRIETARY INFORMATION    SS    Cover the LED   s so the LED reflection is no longer visible on the video image  Wait at least 3 seconds  for the reading to stabilize        Note  Add time of wobble period plus a safety factor if the Wobble Filter is turned on   Perform the    Calibrate Reflectance     Low Value    option   1  Click the    Signal Value    button and select Low    2  Enter 0 0 for the reflectance in    Low   Reflect    for both the 950 nm and 470 nm reflectometer as  the reflectance is zero with no substrate reflection     3  Click    Get Current Data    Set button to set the Low Signal reflectance values  This will place the  current measured reflectance values into the    Low   Reflect    windows   Remove the cover from the LED   s to show the LED reflection in the video image  Wait at least 3 seconds  for the reading to stabilize     Note  Add time of wobble period plus a safety factor if the Wobble Filter is turned on     Reflect 951 30    Reflect 485 00    Syal Reflect Sya Reflect  Low 0 00 0 Low 10 00 0  High 1000 00 TED High 1000 00 e    Current
3.    If manual gain control is active  the gain may be conveniently changed on the main Engine window by  selecting settings from the pull down menus     SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 25 of 51           Engines for Thin Film Innovation PROPRIETARY INFORMATION    Temperature LUT Calibration    This procedure loads the proper translation table for generating temperatures from the ratio of the two  pyrometry intensities     Go to File to ilies Look Up Table    JJ Engine Temperature Ref  Log File  Gain Files    Pyrometer Look Up Table       o The Pyrometer Look Up Table window will appear    SVT Associates   Acculemp    ne Temperature Reflectometer Configuration   Help    Ratio Be ase Look up Table File          Say E phd   my ya i Ale N  400 0 500 0 6000 7000 600 0 92900 0 1000 0 1100 0  Temperature   C      4 C  Code Lab  iew I54K8 Growth_LV Supportltemp_4012 lut a       omnc 0 LED OFF Logg    Figure 23     Pyrometer Look Up Table Window       The factory provided LUT file is generated for a    Silicon Substrate    radiator from a calibrated silicon  substrate target     Click Browse and navigate to the temperature LUT calibration file found on the factory calibration DVD   See Appendix LUT for further informa
4.    the aluminum silicon eutectic point at 577   C is often used to sense the substrate temperature  To  calibrate the single channel temperature method by other means     Go to Configuration to Data Acquisition Setup      Configuration   Help    Reflectance Calibration    Data Acquisition Setup  Wobble Filter  Debug          o The Data Acquisition Setup window will appear    SVT Associates   Acculemp      File Engine Temperature Reflectometer Configuration Help    Data Acquistion Setup   Geometry Factor    Averaging   1 Method of temperature Reference    y User defined Temperature  Scan Timeims   250        Com Port      1 User Defined Reference Temperature  C   0       Comm Delay  Sec      0 050 Calculated Geometry Factor    i g 2 100E 5  Gain Blank Delay    2 000  Set Geometry Factor 7 983E 7  Fixed Geometry Factor     Fixed Emissivity  amp  Geometry Factor    Current Emissivity 0 998    Set Emissivity 0 613    Fixed Emissivity         Communication Initialized io  LED   OFF Lagging OFF    Figure 32     Data Acquisition Setup Window       In the Geometry Factor box    1  Under Method of temperature Reference select    User defined Temperature       2  Click Calculate    3  If you know the geometry factor number you want to use  enter it in the box and click    Fixed    Geometry Factor     Box to activate it     SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934 2737   Em
5.   Communication Initialized O   LED OFF LoggingOFF      Figure 31     Data Acquisition Setup Window    In the Geometry Factor box  1  Under Method of temperature Reference select    Two Channel Temperature     2  Click Calculate    The single channel temperature channel is now calibrated by means of the ratio or two color channel     Note  Make sure there is sufficient signal strength available to make a reliable ratio calculation  i e   temperature  gt  550   C      SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 34 of 51           Engines for Thin Film Innovation PROPRIETARY INFORMATION    However  there are instances where the ratio method is not applicable  If the heated surface has a  significant difference in the emissivity between the two pyrometry wavelengths  then the ratio calculation  will be in error  For example  the ratio measurement for a silicon substrate may be different than a GaAs  substrate at the same temperature  To properly handle these material differences  the ratio to  temperature conversion is performed by a look up table  LUT  that can be altered  See Appendix for  more information  Page 40         Alternately  the single channel pyrometry calibration may be performed via other means  For example
6.   MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 45 of 51       Engines for Thin Film Innovation    S    gt  Te    Some typical values for well known semiconductors are listed in the table below                VEIGHEN E  0  eV 10E    6          Accurate Temperature Measurement    As consequence of above mentioned facts and phenomena a more versatile instrument that combines  both measurement methods is desirable  The Accutemp Process Monitor provides high end  performance over the entire temperature range     Below 400   C emissivity corrected pyrometry cannot be used due to lack of black body radiation   Bandgap based temperature monitoring is the method of choice  At high temperatures bandgap based  measurements are impossible due to carrier concentration and radiation intensities  Pyrometry is the  method of choice  In the intermediate temperature regime emissivity corrected pyrometry is used for best  results on opaque substrates  Precise calibration and repeatability can be achieved with band gap based  calibration  For transparent substrates bandgap based monitoring can provide a good alternative  Once  layers become opaque pyrometry can take over for best performance     The Accutemp Process Monitor combines both techniques in a single instrument  In addition it allows  accurate grow
7.   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 4 of 51          Engines for Thin Film Innovation PROPRIETARY INFORMATION    SS ann       Introduction  Theory of Operation  Pyrometry    Pyrometry is a method to calculate temperature by measurement of the electromagnetic radiation  emitted by surfaces of opaque objects at finite temperatures  This effect is described by Planck   s Law  which describes the spectral radiance R  A T  as a function of wavelength and temperature      1       2ehc    E    Fi  R a T       Where e is the surface emissivity  h is Planck   s constant  c is the speed of light     is the wavelength  T is  the absolute temperature  and k is Boltzmann   s constant  Thus if one knows the surface emissivity and  has a well calibrated radiometer  the surface temperature may be obtained  A black body is an ideal  radiator and has e   1 for all temperatures and wavelengths  but real materials have emissivities less  than one and may not have a constant value for all wavelengths  An example of a black body radiance  curve for 700   C is shown in Figure 1     4       3 5 5    wo  l    Y  o    EE  ol  l    Radiance  x10 9 W m 3 sr   N         li       ol  l          O       6 8 10 12  Wavelength  um     O  N 4  D    Figure 1     Black Body Radiance Curve    Traditional pyrometry selects a fixed range of wavelengths and measures the emitted radiation intensity  over that range  The resulting intensity v
8.   y   o      Figure 25     Temperature Window    1  Ratio Temp    2  Select if you want to plot this temperature    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 28 of 51           PROPRIETARY INFORMATION    ee  n    3  If this is red there is a problem and the instrument cannot plot  if it is green the instrument is  getting data     Engines for Thin Film Innovation    Emissivity Corrected Temp  Select if you want to plot this temperature    If this is red there is a problem and the instrument cannot plot  if it is green the instrument is  getting data   Reflectometer Menu    Reflectometer Window  Go to Reflectometer to Reflectometer Window    Configuration       o The Reflectometer Window will appear       Reflectometer Window        Reflect 950 80 nm Reflect 470 10 nm  0 0652 0 0838      Plot OFF ON Plot OFF ON  Cursors   x l Caas      5 B GursorO 09 09 49 aM    0 55172 Atine AP 4 Refl9so 7  B cursor1 09 09 59 AM  0 51724   9 80518    Ew  Reia   gt               oH f EEE  j   i                           3m 1  09 09 19 AM       09 15 00 AM 09 20 00 AM 09 25 00 AM 09 31 01 AM  Figure 26     Reflectometer Window     Click on the cursors icon in the graph pallet    Right click on either Cursor 0 or Cursor 1 in the Cursor L
9.  0        e cee ecececececececececececececeeeeeaeaeaees 8 4  Color  RS170  Input  24 V DC    OPTIONAL ACCESSORIES   220 V power supply   Alternate reflectometer wavelengths available   Remote External illuminator reflectometer available for non normal illumination  Analog output for feedback control    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 38 of 51           Engines for Thin Film Innovation PROPRIETARY INFORMATION    Appendix    Index of Refraction Values    Below is a list of useful index of refraction values for common substrate material  These may be used for  reflectometer calibration  See Reflectance Calibration on page 29     MEIGIEN 950 nm Index 470 nm Index  Silicon 3 67  i0 005 4 32   i0 07    GaAs 3 54   0 08 4 33   i0 44  Ge 4 65   i0 3 4 3   12 4  AlAs 2 974   10 00 3 51   10 00  GaSb 4 3   10 34 4 3   12 3  InAs 3 7   10 43 4 3   i1 8  InP 3 37   i0 2 3 818   i0 5  InSb 4 42   i0 64 3 57   12 2  GaP 3 17  i0 00 3 6   i0 01    GaN 2 37   10 00 2 47   10 00    These values are obtained from the Palik database  Handbook of Optical Constants of Solids   Edward D  Palik  Academic Press  First edition 1985  2nd Edition 1991  and the Landolt Bornstein  tables  Semiconductors  Group IV Elements and III V comp
10.  13 Page 51 of 51       
11.  2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 32 of 51           Engines for Thin Film Innovation PROPRIETARY INFORMATION    SS n    3  Com Port     The RS232 port numbers that will be used to communicate with the In Situ 4000  Process Monitor     4  Comm Delay  Sec      Sets the polling rate for reading data at the RS232 port     5  Gain Blank Delay     Sets the timeout after a gain change is made  AGC or Manual   This allows  the signal to stabilize     6  Save     Saves Data Acquisition values        Calibration of the Temperature Instrument    The temperature instrument can compute temperature in two different ways  by the    two color    or    ratio     method and single color radiometry  The intended use of the In Situ 4000 is to use the two color    ratio     method to provide the reference calibration for the single color radiometry method  The reason for this is  three fold     1  There is more noise in the ratio method compared to the radiometric method     2  The two color method only applies to    gray body    surfaces and is not correct when multiple layer  dielectrics are deposited on the surface of the substrate  Multiple layer optical interference alters  the emissivity at the two different wavelengths and the ratio calculation is then in error  Thus the  two color method is only useful before thin films are deposited     3  The single color method may be used even 
12.  934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 24 of 51          Engines for Thin Film Innovation PROPRIETARY INFORMATION    Se  DL                 1  Click on the Pyrometer C1 950 nm Gain channel File Browse button and browse to the location  containing the gain files and load P950_4XXxX gnc  where 4XXX is the serial number of your  particular optical head     2  Click on the Pyrometer C2 850 nm Gain channel File Browse button and browse to the location  containing the gain files and load P850_4XXxX gnc  where 4XXX is the serial number of your  particular optical head     3  Click on the Reflect C1 950 nm Gain channel File Browse button and browse to the location  containing the gain files and load R950 _4XXxX gnc  where 4XXxX is the serial number of your  particular optical head     4  Click on the Reflect C2 470 nm Gain channel File Browse button and browse to the location  containing the gain files and load R470_4XXxX gnc  where 4XXX is the serial number of your  particular optical head     5  The Gain Files Setup window allows you to enable    Automatic Gain Control    where the computer  determines the need to switch from one of four hardware gain states depending on the intensity  of the signals being received  Check the box next to Use Automatic Gain Control to activate  the  system will use the threshold intensity values to cause the software to swit
13.  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 44 of 51          Engines for Thin Film Innovation PROPRIETARY INFORMATION  Theory of Operation    Bandedge or bandgap temperature calculation is based on the principle of a semiconductor material s    energy gap  E   dependence on temperature  The energy gap  bandgap  is calculated by the following  empirical equation        2    aT    E  T doping    E   0    Bat   E  doping   Where    E   0    Energy Gap at 0K  a   Absorption Coefficient    P   Fitting parameter forPhonons  E   doping    Fitting parameter for doping    E  doping  is different for different type of materials and must be known in order to correctly find the  Energy gap  This is done empirically using a LUT similar to the ratio temperature method  This look up  table tabulates the relationship between bandgap and temperature for determination of temperature  The  software fits the materials using above equation over the temperature range of interest     An example of temperature dependence of the Energy gap of GaAs is shown in Figure 37     Temperature Dependence of Energy Gap  1 6       ing  n     an  A        GaAs    Energy Bandgap  Eg   X    es  NO    1 1    300 400 500 600 700 800 900 1000    Temperature  K   Figure 37     Temperature Dependence of the Energy gap for GaAs    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie
14.  Reflect 951 30 nm Current Reflect 485 00 nm    5 30 7 28       MAXIMIZE the signals by Channel Wavelengths  aoina optical head     ly    f Pyrometer Channel 1   950 nm  aoe Ve P D Pyrometer Channel 2  850 nm  Get Current Data   Set Reflectometer Charewi 1  951 3 nm   gt   Reflectometer Channel 2  455 nm    ave    Communication Intigined 0  LED   OFF Logging OFF                        Figure 29     Reflectance Calibration Window     High  4  Click the Signal Value button and select High    Click    Get Current Data    Set button to set the High Signal reflectance values  This will place the  current measured reflectance values into the    High   Reflect    windows     6  Enter the values for reflect 950 nm and 470 nm respectively  See Appendix Page 40   Use  known reflectance value or data from literature     7  Select    Save Calibration    to save these settings     Keep in mind that viewport coating will affect the reflectometer calibration  so routine calibration of the  reflectance is advised     SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 31 of 51           Engines for Thin Film Innovation PROPRIETARY INFORMATION    Channel Wavelengths    The Channel Wavelengths box provides information on your optical head config
15.  chamber  into the optical head and may cause an errant reading  This may be evident when using vacuum  instruments with glowing filaments like ion gauges  mass spectrometers  etc  Also  radiation from  electron beam or effusion sources may also be scattered into the optical head and affect the  pyrometry adversely  Solutions to this problem may be to use substrates polished on both sides   or ensuring pyrometry is performed at high enough temperatures where the semiconductor  substrate is sufficiently opaque     3  Ifthe transparent wafer is optically polished on both sides  then the normal incidence specular  reflectance must be known prior to loading in the vacuum chamber  This is because there are  now two surfaces  which reflect light back to the reflectometer  and the back surface  if coated  with an opaque film  will have a much different reflectance than that determined by the substrate  index of refraction     PROPRIETARY INFORMATION    SS n    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 16 of 51       Engines for Thin Film Innovation PROPRIETARY INFORMATION       Focusing Tips   The optical head has two different detection branches for detecting both pyrometric radiometry and  reflectometry  Both of these syst
16.  each  Pyrometry can measure  temperatures from 400   C and higher  band gap thermometry is limited to temperatures typically below  650   C due to free carrier absorption and black body radiation  Since pyrometry relies on absolute  intensity measurement it is affected by viewport coating and stray light from hot sources  Using ratio or  2 wavelength pyrometry can eliminate some of the interference problems but during growth is of limited  use since emissivity changes are not properly accounted for  Transparent substrates such as SiC  ZnO  and others cannot be measured by pyrometry unless a backside opaque coating is applied to the  substrate  Bandgap based monitoring relies on the spectral change in intensity as a relative  characteristic to determine bandgap  As such it is not limited and can be used in these cases  However     SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved       In Situ Process Monitor Version 2 0 4 13 Page 43 of 51       Engines for Thin Film Innovation PROPRIETARY INFORMATION    ee              different substrates may require different spectral ranges and  detector configurations  Double side polished substrates also cannot be measured well with the bandgap  based technique in reflection mode due to low light scattering efficiency     Pyrometry 
17.  grey body     the ratio  of the radiance equation above becomes        hc       oMkT _     R A T  Ay     R Ay T  Ay  TAAT            and the ratio is independent of emissivity and the absolute radiometric gain  Thus the temperature may  be computed from knowledge of only the measurement wavelengths and the intensity ratio  An example  curve of radiometric ratio vs  temperature is shown below in Figure 3 for the wavelengths of 950 nm and  850 nm     This feature of ratio pyrometry is very convenient in correcting errors  which affect both wavelengths  equally such as optical misalignment  viewport window coating  or an unknown emissivity     SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 6 of 51       Engines for Thin Film Innovation PROPRIETARY INFORMATION                         Radiance Ratio                                     400 500 600 700 800 900 1000  Temperature  C     Figure 3     Ratio of Radiance vs  Temperature for 950 and 850 nm Wavelengths    The intensity ratio vs  substrate temperature will differ for varying substrate materials  so the In Situ 4000  system allows the flexibility of a look up table  See Appendix     LUT section  to translate from intensity  ratio to reported temperature  This look up tabl
18.  laser scattering illuminators   Viewport Requirement                ccccccecceeeeeeeeeeeeees Minimum vacuum viewport size is 2 75 inch Conflat  flange  at normal incidence to wafer  Minimum clear  aperture is 33 mm diameter  Transmission range    470 950 nm  FOCUS Adjustment              cc ceecceecceeeeeeeseeseeeeeeenees Manual  Operating Ambient Temperature                   000  18   C to 25   C   lt 2   C thermal drift during use  Storage Ambient Temperature                   00  00008  40   C to 50   C  Relative Humidity 2 0 0    cece eecceeeeeeeeeeeeeeeeeenees Non condensing  Power Requirements               ccceceecceeceeeeeeeeeeeeeees 120 V 220 V 50 60 Hz wide input  Power Cable Length              ccccccccsecceeeeeaeeeeeeeeeees 2m  NN GIGI cats caceacessitten terns E 3 2 kg  not including external power supply   POUSING spesies EEE EREE Eana Black anodized aluminum  Dimensions            ccccceecceeecseeceeccueceeesueccueeseeseesaes 100 x 140 x 90 mm  optical head   COMPUTER REQUIREMENTS  Operating System        o  nonnnnneennennnnnnnnnennnnennnnnnnnne Windows XP or later  Computer Interface              ccccccccceeeseeeseeeeeseeeeeees RS232  9 pin D male connector  AT standard   OPTICAL ALIGNMENT VIDEO SYSTEM  Video Camera Output               cccccceeceseeceeeeeeeeeaees B W RS170 signal at 60 Hz  75 Q BNC connector   CCIR at 50 Hz is also available   Video Cable oo    ccc cecececececeeeecececeeeseceeaeaeaeaneees 75 QO BNC coaxial cable  2 m length  Video Monitor 2
19.  the measured temperature is by inserting the temperature value as a comment in  the log file  After many different temperatures have been mapped out  the log file is then easily analyzed  to extract a table of ratio and temperature values  The new LUT file may then be generated according  the following file format     The first line in the LUT file must be   B74B9D7A 30B7 11D2 AEE0 244611C10000  LUT  1  This signifies to the program that the file is a LUT file     Comment lines are inserted with the    notation  The next non comment line in the file must be the  material label line     Material  Silicon    where the two fields are separated by a tab character  Next  the pyrometer gain range for which this LUT  is valid is specified by a line    Gain Range  1    where the gain range may be 1 through 4 and the two fields are again separated by a tab  You must be  aware of the gain range when you perform the temperature mapping measurement     The rest of the LUT is a sorted table of ratio values and temperatures values in degrees C separated by  tab characters  There is no explicit limit to the number of entries in the table     Load in the new LUT text file from the    Load    button in the ratio LUT property page of the Temperature    instrument  The new LUT will be saved to the default configuration file on the hard disk so it will  automatically be loaded each time the program is run     SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   U
20.  the two virtual substrate indices may be limited in  this fashion  Therefore  if the user must has some a priori knowledge of the film parameters fitting results  will benefit significantly     Minimum Rate Allowed  It is possible for the fitting algorithm to    flat line    or settle on a seemingly  satisfactory fit with a very small   lt 0 001 nm sec  growth rate  The solution is to    kick    the growth rate to a  more reasonable value if it falls below a certain threshold  However  if the actual growth rate is lower  than the threshold  the fit becomes ineffective     Maximum Parameter Variation  These values set boundaries in parameter space beyond which the fitter  is not allowed to search  This region is defined as absolute values on either side of user s initial guess     Performing a Stand alone Fit    Standalone fits are useful for performing deposition rate calibrations or monitoring a very thick film  growth  Prior to start the fit the user needs to select the wave length in the pull down menu Wave Length  Fit  After that reasonable guesses for the optical parameters of the substrate  the film and the expected  growth rate are required  It is possible to open a previously created setup file to reuse parameters for  convenience  The fitter uses all data selected between the two cursers Cursor 0 and Cursor 1 in the  reflectance chart for its calculations  A click on the Run Growth Fitter button starts the instrument and  outputs are generated within less than a sec
21.  view  Figure 15  orange color  whereas the reflectometer detectors have a 3 6 degree field of view   Figure 15 blue color   The video image views the reflectometer view so the dark  boundary outside of the central video image corresponds to the limits of the  reflectometer field  The pyrometer field is approximately 55  of the reflectometer  and is centered on the video image  Note  The inset in Figure 15 shows the  alignment when only a quarter of a wafer is used        Note  there is a fourth electrical connector present on the rear panel which supplies  a TTL compatible sync pulse signal for synchronization of external  non normal Figure 15  incidence illuminators     If you ordered the optional external illuminator follow instructions to install it  Locate  the external illuminator and the external illuminator cable  Mount the external  illuminator to a 2 75  viewport  The illuminator has one screw to clamp it to the  viewport  Using the external illuminator ribbon cable attach the illuminator to the  In Situ 4000  The DBM9 on the cable attaches to the DBM 9 on the front of the In   Situ 4000  and the 10 pin ribbon connector goes to the illuminator  The illuminator  connector is keyed for proper electrical connection        Figure 16     RVA    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  Al
22.  view the substrate in the video monitor  Additional illumination applied  to the substrate may be required  for example by a flashlight shining through a spare viewport or the  glow from an ionization gauge     Focus the image on the video monitor by loosening the objective lens set screw and i Set  crew    sliding the lens in and out to obtain a focused image of the substrate  When trying to  view a large polished substrate with no features to observe in the image  you may need  to perform a focus operation before the substrate is in place and instead focus the  image when viewing the substrate heater  Ensure that the focus adjustment set screw is  secure before continuing        Once the substrate area is centered in the video image  the reflection of the LED  illuminator must be visible in the image  Perform additional tilt and translation  adjustments of the optical head and or the substrate to achieve a normal incidence  reflection  Center the image of the LEDs in the video image     Note  The LEDs used to illuminate the substrate are quite bright and may saturate the  video image producing unusual patterns on the screen  Be assured this saturation will  not damage the video camera electronics and that this image is normal        Figure 14  Once aligned  the instrument is prepared for temperature and reflectance measurement     There are two different fields of view for the two different types of measurements   The standard pyrometer detectors have a two degree field of
23.  works based upon the Software  You will not  reverse engineer  decompile  disassemble or otherwise attempt to discover the source code of the  Software except to the extent you may be expressly permitted to reverse engineer or decompile under  applicable law     SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 2 of 51       Engines for Thin Film Innovation    SS     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA  Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com    Table of Contents    POCO ct arrctncutttatnactiinestuandseuenset antuuahssanedenerinnexaneis stadscuPereadiuiactiabiaenednacet neue Asinsstaentmice EAA 5  EON OF OPE raO e A E tone Soesiaveaasumeneals 5  IN a E E E A E T E A E E 5  TWO COOr F WOME UY rE T TE a A A N E 6  ESSIE eI   E E EEE RTN AE A TE PAE A EE E EE E EA N E E 7  Emissivity Compensated Pyrometry              cccceccccseeeeceeeeeeeeeeeeeeeeeeeeeeeeaeeeeeseeeeeseeeeeseaeeeseeeeesneeeseeeeeseeeess 8  Problems with Pyrometry Of Semiconductors             ceecccceeeeeceeeeeeeeeeeeeeeeeseeeesseeeeeseeeeeseeeeesaeeeeseeeeeseneeess 9  Unpacking ANG INSIANAUOM s c ccsc Gecereates8Qocauened iastanmcentautecetatat a E E EE E DARRE ra a 12  MIT AG aae PEE EE E E E E E EE
24. 21   Gain Files     select and change the Gain settings  Page 24   Pyrometer Look Up Table     select the temperature LUT calibration file  Page 26  Exit     exits the software application    Engine Menu  Engine Window     displays measurement values and operates the In Situ 4000 unit  Page 27    Temperature Menu  Temperature Window     view temperature signals  Page 28    Reflectometer Menu  Reflectometer Window     view reflectance signals and calculate growth rate  Page 29    Configuration Menu  Reflectance Calibration     calibrate reflectometer intensity signals  Page 31    Data Acquisition Setup     set up COM port settings and calibrate geometry factor  Page 33    Wobble Filter     correct for wafer tilt during rotation  Page 35    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 19 of 51           Engines for Thin Film Innovation PROPRIETARY INFORMATION        De    Debug     service mode of the In Situ 4000       Help Menu  View help manual    Start Up Screen    When you first start up the software you will see the Temperature window     SVT Associates   Acculemp    e Fie Enone Temperature Refectoneter Configuration   Hep          ik ee  S RVA T Ratio Temp Commcted Temp     C  Plot OFF ON      Plot OFF ON           
25. E E ETE E E 12  K esta oea A E E E E EEO A E EOR E EA EE T EEEN E ER 13  Mounting Using SVT Associates    FIANGC            ccccccccseeeeeeeeeeeseeeeeseeeeeseeeeeeeeeeeseeeeseeseesaeeesseeeeeaeeeesaees 14  Electrical Connections             ccccccccceececeeecceeeceeeeceeeeeeeeseeeeseeeeseeeeseeeesaeeesaeesea ease eesaueesauecsaeecsusessusessueesueesaeeeaaaes 15  Other Recommended Usage Notes               cccccccccceccceeeeeseeeeeeeeeecceeeueeeeeesecseusessesesseueesseesesseusesseneesseeeessensessegs 16  Window Glare SOU CUO aaa es csc etdiek neat a douse E Aa E a cepeouardemeceneesadceuten usecase  16  Use With   Smal VIEW DOWNS ssri sisri eaei rar ein N dena reiii arain esindada 16  Use With Transparent Substrates              ccccccccccceececeeeeeeeeeeeeeeeeeeeseeeeeeceeae cess cess eeseeeeseeeesaeessneeeeeeneeeees 16  POCUNG NS ace te ete  atie tse ear ree era ensaie tania s Seem AEE AAE EE EEA 17  Film Thickness Non Uniformity ssiicietiaersaidindadadwsaarsns ince odaondea Versa cnsaandadiaacendamstrsndlandnirownisbennataner 17  Optical Head lI IAS MM snstvinatraicsd na tedamtalasttiandaawet poansata soutigaitced aha dayan ialustitoaedetateddummbaiatGatceniatctushalaeaiteden tstiubaiiabond 18  SONNIE se epee cin E ei  Sc es nos pec eco am tae 9 ore es eas ates patibesenee aut e bercee aes entee dene teaneeaet 19  Software Installation 2 2    ec cece cee e cece eee cece eeee ance sae ee Ae ee Ge eeAAGeesA A eesaa See A ees G sees Gs eesGaeesaaeesaeeEaeEeEaaEREEES 19  PGES 
26. ETARY INFORMATION        n       G  Goods being returned for other than warranty repair shall  be subject to a restocking charge of twenty  20  percent of the original sales price of the  returned item     Returning Equipment for Repair or Servicing  Before shipping equipment for repair or servicing  obtain a Return Authorization Number assigned by  SVT Associates     Liability Disclaimer   SVT Associates  Inc  takes steps to assure that its published specifications and manuals are correct   however  errors do occur  SVT Associates  Inc  reserves the right to correct any such errors and  disclaims liability resulting therefrom     No Liability for Consequential Damage   In no event shall SVT Associates  Inc  or anyone else involved in the creation  production  or delivery of  the accompanying product  including hardware and software  be liable for any damages whatsoever   including  without limitation  damages for loss of business profits  business interruption  loss of business  information  or other pecuniary loss  arising out of the use of or the results of use of or inability to use  such product  even if SVT Associates  Inc  has been advised of the possibility of such damages     SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4
27. F  Logging OFF      Figure 6     Temperature Measurement during GaN Growth using Emissivity Compensation    Problems with Pyrometry of Semiconductors    Pyrometry is a well established measurement technique and is successfully used in many applications  It  has one main feature that makes it attractive for substrate temperature measurement during MBE or  MOCVD  it is a non contact and vacuum compatible temperature sensing method  However as those  who have used pyrometry for semiconductor applications know  this technique can be fraught with    problems  which greatly limit its usefulness     SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 9 of 51          Engines for Thin Film Innovation PROPRIETARY INFORMATION    Sj LL          SVT Associates has recognized the problems with pyrometry of   semiconductors and has designed the In Situ 4000 specifically to solve many of the problems preventing  pyrometry from being a useful tool for both MBE and MOCVD processes  This system combines  traditional pyrometry with specular reflectometry to provide a single instrument to monitor both substrate  temperature and film thickness in real time  Listed below are some of the problems and how the In Situ  4000 overcomes them        1  Substrate t
28. JV    SVT Associates  Inc Engines for Thin Film Innovation       Model In Situ 4000 Process Monitor  User Manual       THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION OF SVT ASSOCIATES  INC  AND  SHALL NOT BE USED  DISCLOSED OR REPRODUCED  IN WHOLE OR IN PART  WITHOUT THE  PRIOR WRITTEN CONSENT OF SVT ASSOCIATES  INC     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13          Engines for Thin Film Innovation PROPRIETARY INFORMATION    OTT       Notice  This manual and the software it describes are considered proprietary intellectual property of  SVT Associates  Inc  and are not to be disclosed to any third party without permission from  SVT Associates     software License Agreement   The software application supplied with the In Situ 4000 Process Monitor is licensed to the user only and  may not be transferred to other parties without the expressed written agreement of SVT Associates  Inc     This program is protected by copyright law and international treaties     Unauthorized reproduction or distribution of this program  or any portion of it  may result in severe civil  and criminal penalties  and will be prosecuted to the maximum extent possible under law     No Modification or Reverse Engineering    You may not modify  adapt  translate or create derivative
29. PEIE E EE T E TE E EE E E E E E EET 19  EUD S E SN e E E E E ean asneseeeonseees 20  Managing Graphs and WINDOWS wii sciscsseisecnciavecsvacexanccsesdecctsviedaedsaseesaaneveadcsvnntenigeeesdieal EEES EEEE enren nErenn een 21  PIE NICU ee A E E E E Supenachaastantonduass 21  Bate lel  gle DaO F  cho srera E E EA R A E eter ee ERR 21  Electronics Calibration and Gain Filter           ccc ccccccceecceeeeeeeeeeseeeeeeeeeaeeeseeeeseeeeseeeeseeeeseeesseeseeeseneees 24  Manual Gain C OM ON ascot ccnneceduicesitcecueceeacecexesankaideaedevensinndel eee dennecadcecetaelicecuchuidtousdacendundsudeeseteecedeeseeexsee 25  Temperature LUT Calibration               cccccccccsccccseecceeecceeeecececeaeeesaeecsaceseeeseusesaueeseuessasessusessusessueeseeeeaaaes 26  EAONE MOME ieee en er eee E EE ee ee ee ee 27  Operation of the Engine INStrUMeNt                cc cccccecce ec eeeeeceeeeeeeeeeee ae eeeeaeeeeeaeeeeeaeeeeeaaeeeeaaeeeesaneeeseeeeesneeees 27  Temperature WIEN scr tuascncexcuesnsueeneeesssnaniankeuacendenneimexcusirnarenembecpenseunanbashunanesdaciatacmiseixanernietseransmunendaeialeeuigenedas 28  Temperature WINGOW              ccccccesccesecceeceseceueeceecaueccueceaeceueesaecsaeesaeesaessaeesseesaessaeeceeetaeeseeeseessaeessensaeeses 28  FREME COMETS  NWS AN saoer ea eiaeia eran kaia eiaa reirei eiriaa 29  Reflectometer VV IN GOW wis csedo sete aides ansiundessxeawvenseegandacoueudauettedsosvasduiiesadeasdwi genes bend deniedenedtied eameaneeceteisels 29  CO MTOUE AOR NG GOs i
30. SA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 40 of 51       Engines for Thin Film Innovation PROPRIETARY INFORMATION          0 6  I I l 1 I 1 1  400 0 500 0 6000 7000 800 0 9000 1000 0 1100 0  Temperature   C    ffHeader for file of type LUT  f GUID Nice Name Version Number   B74B9D74 38B67 11D02 AEE6 244611016066   LUT  2  f tab delimited lookup table file  First data lines must be   f Material   lt tab material name  f Gain Range  lt tab valid range    f The column headings are   ff PI956 PI856 Temp Deg C  Material  Silicon  Gain Range  14  6 29214 696 4  6 29216 674 1  6 29215 674 2  6 29216 674 3  6 29213 674 4  6 29267 674 4  6 29268 674 5  8 29522 669 8  6 31779 629 8  6 33363 684 6  6 34578 566 8  6 36415 575 7  6 36985 566 5  6 37272 563 5  6 37537 559 8  6 37726 557 2  8 37688 555 1  0 38003 553 5  0 38089 552 2  6 38164 551 2  6 35416 549 6  6 41289 522 9  6 44261 494 7  6 47364 472 8    6 48899 458 6  6 56196 447 2 S AVA T  0 50930 438 6 X    0 51122 431 8    6 51642 426 3 SVT Associates  Inc   0 52255 422 1  AYAS CNG 6 52685 418 6 SVT Associates Proprietary Information  mAlOi MEceYanceee 0 52232 415 3   2013 SVT Associates  Inc  All Rights Reserved  DESIOEN  H  52443 413 2 Page 41 of 51  6 51766 410 7    A Canca AA  N A           Engines for Thin Film Innovation PROPRIETARY INFORMATION    TT 
31. The pull down menu Fit Type selects the scope of fit  Fix None  Fix S  Fix  N  Fix K  Fix N S Fix K S  Fix N  K and Fix N K S  Section of the type of fit impacts its results and  convergence and is describes later in this chapter  Controlling the Fit   Another factor influencing the  fitter instrument is the Fit Tolerance which acts as a stop measure once the error in fit is below this  threshold value     In addition  the user must provide initial guesses to give the fit algorithm a satisfactory starting place to  search for a best fit and a minimum and maximum value  The initial values used in the fit can be the  values determined from the last fit or any other guess by the user     One limitation of interpreting reflectance oscillations is that the fit parameters are not independent from  each other for very thin films  For example if a film is being deposited with a current thickness  corresponding to less than 1   4 optical wave  the fit algorithm can find a satisfactory fit for many different  combinations of the six possible parameters which leaves little useful information for the user  As the film  becomes thicker  there is more information available to the fit algorithm and the fit parameters     decorrelate    and allow meaningful interpretation of the results  The Growth Fitter algorithm has an  adaptive fitting feature  which allows the user to still gain meaningful results in the early stages of  growth  The Growth Fitter holds the film n and k parameters consta
32. ___    Figure 35     Example LUT File    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 42 of 51       Engines for Thin Film Innovation PROPRIETARY INFORMATION       Bandgap Based Temperature Measurements  Introduction    Emissivity corrected pyrometry relies upon measurement of surface black body radiation as long as  temperature is high enough and the substrate is opaque  To overcome this limitation the In Situ 4000  Process Monitor provides bandgap based thermometry as add on  Band edge based thermometry  utilizes the temperature dependence of a semiconductor bandgap to determine substrate temperature  A  typical setup is shown in figure XX               Lighted Source       Diffuse Reflected                  A  AccuTemp    N  Receiving Head ta    JBandgap Detector    Figure 36    For higher temperatures substrate heater radiation can be used to determine bandgap  for low  temperatures an auxiliary light source needs to be used  In the first case scattered transmitted light and  the latter case diffuse reflected light is analyzed in the Receiving Head     Both emissivity corrected pyrometry and band gap thermometry work well for GaAs  InP  and Si  substrates  however  the general usable temperature range is limited with
33. adiated light from a substrate heater located behind the substrate will be transmitted  through the substrate and enter the radiometer  The radiometer cannot distinguish between  radiation from the substrate and that transmitted through the substrate and a pyrometry  measurement error will result  One solution to this problem is to coat the backside of the  substrate with a sufficiently opaque metal coating     1  Semiconductor substrates have varying transparency with temperature  Many semiconductor  substrates are partially transparent at 850 and 950 nm at low temperatures  At higher  temperatures  typically the band gap shifts to longer wavelengths and the substrate becomes  acceptably opaque  The wavelengths used in the In Situ 4000 were chosen to be opaque for  silicon  GaAs  and InP substrates at elevated temperatures  Silicon  for example  is partially  transparent at 950 nm below about 550   C and so pyrometry measurements may be adversely  affected at lower temperatures  Depending on doping levels  GaAs substrates can be partially  transparent up to 600   C  When the substrate is transparent  the effective emissivity will be low   yet the emissivity measurement will not be correct since the emissivity calculation is based upon  the assumption of opaque substrates   See Figure 7     2  Ifa substrate is transparent  the backside texture of the substrate is visible to the pyrometric  radiometer  A non polished back surface will then scatter stray light in the deposition
34. ail  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 35 of 51          Engines for Thin Film Innovation PROPRIETARY INFORMATION    SS Ann       Fixed Emissivity and Geometry Factor box    4  The In Situ 4000 can be operated as pyrometer without emissivity correction  In this case enter a  value for set emissivity     5  Select the    Fixed Emissivity     check box     Now the temperature is calculated only by using the 950 nm pyrometer channel  This method is  useful for non specular substrates or when the refectometer cannot be calibrated     The single channel temperature channel is now calibrated by means of the    User defined Temperature      Other methods such as Oxide desorption can also be used     Wobble Filter    Many thin film processes utilize substrate rotation to help produce more uniform thin films  This  mechanical motion can produce strong errors in the reflectance measurement if the surface tilts when  rotating  If the substrate is mounted such that the axis of rotation deviates from the surface normal by  more than 0 5 degrees  then the reflection of the LED illuminators will appear to    move    around the video  image and the measured reflectance will vary  If the tilt is large enough  then the LED reflection will  completely leave the field of view and the reflectance measurement will fall to Zero        Substrate    A special software filter has been implemented to address th
35. amount of light  the       1   R relation is no longer true  If the deposited film  is not uniform  then the reflectance measurement  made at a single point on the substrate  may not be  well correlated to the pyrometry measurement  made over a larger area of the substrate      The three techniques of single color pyrometry  ratio pyrometry  and specular reflectance are combined  in the In Situ 4000 temperature instrument  The single color pyrometry temperature measurement is the  preferred method because it has better noise performance than the ratio method  The 950 nm  reflectometry allows calculation of the varying 950 nm emissivity resulting in emissivity compensated  pyrometry  Since there are instrumental and alignment errors in measuring absolute radiometry  the  single color method actually calculates temperature based on previously measured radiation at a Known  temperature  Future calculations of temperature are made relative to the single reference measurement   In the In Situ 4000  the reference brabration may be made using the ratio method  or through some other  method such as oxide desorption or a eutectic temperature measurement   See Calibration of the  Temperature Instrument chapter     SVT Associates   AcculTemp    File Engine Temperature Reflecbometer Configuration Help    Emissivity  S RA T Ratio Temp Corrected Temp Red   Without Correction    s ESGLAIS a   P Blue   With Correction     C Plot OFF ON Plot OFF ON          Communication Initialized O  LED OF
36. and the  emissivity is different between the 950 nm and 850 nm wavelengths  In fact  not only is the emissivity  different  but the emissivity ratio for those channels will be different for different substrate materials and  will even change with temperature  Rather than trying to compute the emissivity ratio to temperature  response from first principles  a more practical method is to empirically measure the ratio for each  substrate of interest  The resulting measurement is captured in a LUT     The ratio LUT is a series of discrete ratio   temperature data points from which the Temperature  instrument linearly interprets temperature  A factory calibrated ratio LUT for a silicon substrate is  provided with the instrument  Alternate substrate materials will require the user to generate a LUT  specific for that material  Making a new LUT requires that the user perform a pyrometry ratio  measurement simultaneous with an independent temperature measurement such the aluminum melting  point measurement   See page 33     Calibration of the Temperature Instrument  Other temperatures can  then be used with the pyrometry 950 nm single channel temperature     Mapping a new LUT    You may generate a new LUT by recording the ratio measured by the Temperature instrument and  measuring the substrate temperature through some independent means such as a thermocouple  The  log file of the Temperature instrument provides the ratio measurement to the user and a convenient  method for recording
37. ate is high enough to provide a measurable  signal  Typical substrates such as silicon  GaAs and InP are sufficiently opaque at 950 nm for  substrate temperatures typical of epitaxial growth  The condition of substrate opacity may break  down for some low temperature processes where the band edge may shift close enough to  950 nm to cause some problems  Therefore  the user must be careful to ensure opacity of the  substrate at the temperatures of interest     2  Band gap shifts  The band gap of a semiconductor substrate is a strong function of temperature  and so the basic assumption of constant emissivity made by standard pyrometers is not valid    See Band Gap based AccuTemp in the Appendix  An example of In Gaj _  N Al Gaj   N is shown  in Figure 8  Using emissivity corrected pyrometer allows compensation for emissivity layers     SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 10 of 51          Engines for Thin Film Innovation PROPRIETARY INFORMATION    Se L_LLLL             Indium Molefraction  0 0 2 0 4 0 6 0 8 1    Band Gap Energy In Ga     N  ev   Band Gap Energy In Ga     N  ev           0 0 2 0 4 0 6 0 8  Aluminium  Molefraction    Figure 8     Dependence of band gap on Al and In mole fractions    Viewport coati
38. ay light introduces an error into the  measurement usually causing a higher than actual reading     Multilayer interference effects  When thin films having different index of refraction from the  substrate are deposited  the resulting interference effects cause the radiated light to rise and fall  with the film thickness even if the actual temperature is constant  This causes temperature  measurement errors during film growth and can cause actual temperature fluctuations due to  non constant radiative cooling of the substrate     SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 11 of 51       Engines for Thin Film Innovation PROPRIETARY INFORMATION       Unpacking and Installation  Unpacking    The In Situ 4000 Process Monitor consists of the following components     Part Number   Description               5038705 Optical sensing head   5034399 Wall mount power supply or a dedicated rack mount controller  5022756 Computer interface cable   500834 Video cable        Software application DVD    5036126 Optional remote illuminator    Unpack all components and remove any protective wrapping  Please save the shipping material for  future use if the unit must be returned to SVT Associates     Caution  The optical head is construct
39. can wcneaetacsntanteteantadacensndectansdnbisuntendehestanlecatusnetaddeasens E a E 30  Calibration of the Reflectometer Instrument               cccceccceececeeeeceeeeeeeeeeeeeeseeeeaeeeese cess eesaeeeseeeeseeeeaeeeeaes 30  Reflectance Callibration             ccccccccccscccceccceececneeceeeeseeeeseueeseueeseeessaeessaeessueessaeesaueesaeeesaeeesaeesgeeseusesneess 30  Channel VV  AVC SUMS osatanctrcandantetndseKeauntciesnantaieiea cvagsaeatinan Vasari caaidssenmannameniaaapians snipe iapabienont saatuada wane 32  Data Acquisition Setup  i cderesccrnsnctecd cncstecacnadalnwtaateiadite xbeveesaiestiteedonsssaenawsbdadeeicdcaiadeabedenscoedteieedeteatedeevee 32  Calibration of the Temperature Instrument               cccccceccceeccceeeeeeeeeeeeeeseeeeeeeeeaeeeeseeesaeeesseeeseeeeseeeeaeeeeaes 33  WOO FS Maer tarncttrceatiee ee AEAEE AEAEE A AEE raS EEES SA 36  SPeciNCaUONS ccs ccttesexadcetecscseanesedeiecansisabsceundrachscethcstencatehtacccadubecantsanbinenndnsatinadsensasnosdsaedenndbbeceedasehacsnedncosiasceniss  38  PRC es erates nena eete cate ence E E colenedcadunmiuieediantdcuubdetebigabustanceltixadend E E 39       PROPRIETARY INFORMATION    SAAT    SVT Associates  Inc     In Situ Process Monitor Version 2 0 4 13    SVT Associates Proprietary Information    2013 SVT Associates  Inc  All Rights Reserved    Page 3 of 51       Engines for Thin Film Innovation PROPRIETARY INFORMATION       Index Of Refraction ValueS             ccccccccsecccsececeeeececeececeeeeeecsecesen
40. ch to a higher or lower  gain  Changing from one gain state to another takes a few seconds and will prevent the  acquisition of data during that time  Automatic gain control is recommended for the Pyrometry  channels as the range of detected signal varies by several orders of magnitude as the  temperature changes  Manual gain control is often sufficient for the reflectometry channels as  the expected changes in reflectance are not as large  Since the signal intensity for the pyrometry  channels are always near each other  the gain control for both the 950 nm and 850 nm channels  are forced to be the same at all times     Manual Gain Control    Each of the four detection channels allows varying ranges of electronic gain depending on the intensity  of the signals reaching the instrument  The two pyrometry channels  channels 1 and 2  and channel  number 4 each allow four different gain settings labeled 1 to 1000 and channel 3 allows two gain settings  labeled 1 and 10  The In Situ 4000 Engine instrument allows the user to manipulate the gain ranges of  each channel by clicking on the handle and selecting the desired gain  See Engine window     The signal values displayed in the Engine Instrument window are the measured intensities in mV  For  optimum signal to noise ratio  the following gain ranges are recommended     Intensity  mV  Gain Setting       If the gain range is set too high  then the electronic amplifiers may saturate  so one must use manual  gain settings with care  
41. debug mode is enabled  the 950 nm pyrometry channel displays the intensity of    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 27 of 51           Engines for Thin Film Innovation PROPRIETARY INFORMATION    the outgoing 950 nm LED  the 850 nm pyrometry channel  displays the intensity of the outgoing 470 nm LED  the 950 nm reflectance channel displays the    internal temperature of the optical head in   C  and the 470 nm reflectance channel displays the  amplitude of the optical chopper  Please consult SVT Associates for diagnostic help if needed     6  Reset Communication button    In case the software lost communication to the optical head click the    Reset Communication    button to  reset the COM port  It will flash the COM buffer and restarts communication  You can also toggle the  power handle to achieve the same result     Temperature Menu    Temperature Window  Go to Temperature to Temperature Window    SVT Associates  Acculemp    re stoke IGS  ASRS CEST ER       Emissivity    S R7 oe Ratio Temp Corrected Temp  saa foo Nal ie    2m  C  Plot OFF ON er    Plot OFF ON       03 I   j l L   I  06 00 00 PM  06 00 50PM 06 01 40 PM 06 02 30 PM 06 03 20 PM 06 04 10 PM 06 05 18 PM           Communication Initiazed OO OOO  
42. e  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 22 of 51           Engines for Thin Film Innovation PROPRIETARY INFORMATION    __SYT Associates   AccuTemp         PATEN TA STATAN ie Fe ee T Td    ve A L  X     m m mm  File Engine Temperature Reflectometer Configuration   Help  Ui z pao   ON ot Fon Pa    aa ee pen       gt           Log Fie Path oneni     3    Desktop       ios Log Interval    My Documents          Communication Iritialized tee   Ay     Figure 21     Log File Window  Enter a name for the file  The file will be saved as a text compatible file     txt   Select where you want the file saved  Click OK    Logging is now active  To turn logging off click the green Logging button  The Status Bar shows  if logging is activated or not  ON OFF        i oe    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 23 of 51          Engines for Thin Film Innovation PROPRIETARY INFORMATION    TT _____    Electronics Calibration and Gain Filter    When installing and running the software on the computer for the first time  the factory calibra
43. e can be changed for different substrates  or updated to  compensate for any calibration errors or instrument drift  The ratio pyrometry feature of the In Situ 4000  provides the user with a repeatable day to day calibration  which is independent of viewport coating  effects     Reflectance    In addition to accurate pyrometry  the reflectometer feature of the In Situ 4000 Process Monitor  measures specular reflectance at two wavelengths  950 nm and 470 nm     The reflectometer uses two LEDs housed in the optical head to       illuminate the wafer through the vacuum viewport window and Heater     detects the light with the same optical system as the pyrometer    me War    Thus the system needs only a single  normal incidence viewport   J      that is provided in many vacuum systems  The viewport must be   chamber      large enough to allow room for both the illumination light to enter         and the reflected and pyrometric radiated light to exit the chamber  e Pyrometry      The In Situ 4000 requires a 2 75 inch Conflat viewport or larger to   accomplish this measurement  Another requirement is that the Hezd   viewport window is located at normal incidence to the substrate so    Video Image B   that the illumination beam is specularly reflected back to the a   instrument  This places some limits on the substrate holders  which   need to maintain normal incidence alignment during use and especially during wafer rotation  The In Situ   4000 system has special software feature
44. ed down by the time specified in the filter period  The entered value is the period of rotation of the  substrate  After several rotations the unit automatically calculates the period        Note  Press enter after entering a value to activate the wobble filter     SAA T    SVT Associates Proprietary Information    7620 Executive Drive   Eden Prairie  MN 55344 3677   USA    2013 SVT Associates  Inc  All Rights Reserved    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com  In Situ Process Monitor Version 2 0 4 13 Page 37 of 51          Engines for Thin Film Innovation PROPRIETARY INFORMATION    SS n       Specifications  Temperature Range              ccccccceceeeeeeeeceeeeeeaeeees 450     850   C  or 500     1200   C  others on request   Compatible Substrates                ccccccecceeeeeeeeeeeeeees Silicon  GaAs  GaSb  InP  Sapphire  backside coated   Radiometer Wavelengths                   ccccceeeseeeeeeees 950 nm  850 nm  Temperature Equivalent Noise                   ccceceeee  lt  0 5   C RMS at 450   C  0 05   C RMS at 600   C  Target Distance Range             cccccecccceeeeeeeseeeeeeees 400 mm to infinity  Pyrometry measurement Spot on Target              20 mm diameter at 500 mm target distance  Reflectometer Field of VieW               ccccceeceeeeeeeees 3 6 degrees  Auxiliary Output           cece eececeeeeeeeeeeeeeeeeseeeeaees TTL signal of internal chopper waveform to allow    synchronization to external reflectance illuminators or 
45. ed using a strong outer enclosure and is quite  rugged  However it is a precise optical instrument and must be treated with care   Dropping the instrument on the floor or other rough treatment may cause damage   which is not covered under warranty        SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 12 of 51       Engines for Thin Film Innovation    SS Aan    Mounting    Typically you can use the flange mount provided by SVT Associates  to mount the In Situ 4000 unit  However  the optical head has two  standard tripod mounting holes located on the bottom of the unit   Either  or both  of these holes may be used to mount the optical  head on a tripod or other mounting bracket  These holes are  machined with 74 20 inch threads  which are also compatible with  standard opto mechanical mounting hardware from suppliers of    optical hardware        PROPRIETARY INFORMATION       Tri Pod Mounting Holes  e       Mechanical stability is very important in using this system  so ensure   that the optical head is rigidly mounted and will not move or vibrate during the entire intended use of the  product  Also  the instrument requires a precise pointing alignment to the wafer under measurement so  the mounting system should include adj
46. eeseeeeseeessaeessesecsuseseueeseueeseueesaeessesesseeeees 39  EUM NERIO E10 i iemeerse nes ener eee E eee ne ee ee ee ee ee 40  Mapping a new LU W iscccecranincnetueshesencdunenscearaiteacametenderaditud setemabiaahcdusatanaumnasinonatieeeieaciuascsurehanentiedandenste a 40  Bandgap Based Temperature MEaSUreMe nts                cccceecceeeceecceeeeeeeeeeeeeeeeeeeeseeeeaeeseeeseesaeeseeesaeeseeeseeeseeeanes 43  IPOCIICUION sessar arr E EE E EE aA 43  WAGON On Op ra UON ai A E E E A 45  Accurate Temperature Measurement                cccccccceecescccececeeceececeeeceecueecaeesaeessuesaeeseeecsueseeecseesaeeseeeseas 46  POW RAE FIET petted E E E N e E 47  SOUD er E E E E casein Se eaeee eee 47  Controlling te F iersrere pren RE AE AEE ASE AEREE nid seausuaasted 49  Performing a Stand alone Fl cissstccsetinduicandenicetaindatatandeuasiuiaddovsnbawndsaniisidedieeidaduiwidunandendtirdbeeastsentonns 49  Warranty and Limitations Of REMEIES               ccceccccceeeeeeneeeeeee cesses eeesaeeeesaeeeesaeeeeseeeeeseeeeeseeneesseeeesseeeesaneeesaaaes 50  PST UNG EH OUGY eE E E E E enue nmcstnrnienienceeahicesatsb Asean tetas levee taanecasteents 50  Returning Equipment for Repair or Servicing             ccccccceeccceeeeceeeese cece eeeeeeeseeeeeeeeeeeeeeseeeeseeeeaeeeeeeeeeaeeesaeeesaeees 51    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com 
47. egend and select    Bring to center        Click and drag the cursor to the point of interest         amp  YS    When both cursors are positioned in the window  the    A Time    window will display relative time  between the cursors     Note  Calibration is required prior to use  See Reflectance Calibration     SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com    In Situ Process Monitor Version 2 0 4 13    SVT Associates Proprietary Information    2013 SVT Associates  Inc  All Rights Reserved  Page 29 of 51           Engines for Thin Film Innovation PROPRIETARY INFORMATION    Configuration Menu    Calibration of the Reflectometer Instrument    The specular reflectance is computed from the raw reflectometer data and is displayed in the  Reflectometer instrument graph  Both the 950 nm and 470 nm channels are displayed simultaneously  and the graph axes are manipulated in the same manner as previously discussed  The reflectometer  measures reflected light in millivolts and two calibration steps must be performed to allow the translation  to specular reflectance     The instrument must be provided two cases of Known reflectance to compute specular reflectance   Typically this is performed with a zero reflectance case and the known reflectance of the substrate under  measurement  The zero reflectance case can be performed with the substrate missing  i e  the  re
48. ems focus at the same distance  however there are two different cases  one may choose to optimize focus  1  at the substrate or 2  on the LED illuminators which is twice the  substrate distance  Focusing at the substrate creates the smallest possible measurement area for  pyrometry  however the LED image is out of focus and    blooms    to a relatively large size  This large size  may cause the reflectometer to become more sensitive to substrate tilt errors  Tilt error effects are  minimized when the system is focused at the LEDs to make the LED image as small as possible  This  method causes the pyrometry sensor to be defocused and will sample an area larger than the 2 degree  field obtained when focusing at the substrate   See Optical Head Alignment Chapter        Optimize Focus     Substrate Optimize Focus     LED    Thus for large area substrates the best performance is obtained by focusing at the LEDs  because the  larger than normal pyrometry sensing area will still be contained within a large diameter substrate   However  for small area substrates the best performance will likely occur when focusing at the substrate  to ensure the pyrometry is measured within the boundaries of the substrate surface     Film Thickness Non Uniformity    The computation of emissivity based on the measurement of reflectance is made based on the  assumption of uniform film properties across the substrate  If the film thickness  for example  varies  substantially across the substrate  then t
49. end of the serial cable to the 9 pin Computer Interface connector on the rear of the optical  head  Connect the other end of the serial cable to an available serial port  RS232  on your computer  Be  sure to fully tighten the connector screws on both ends to ensure a reliable data connection     Locate the video monitor and set it at a convenient location to the optical head  Connect the coaxial  video cable from the BNC video signal connector on the rear of the optical head to the video input  connector on the video monitor  Plug the video monitor power cord into the wall power outlet     Make sure the power switch on the optical head is in the off position  then plug the round power cable  connector into the power connector of the optical head and plug the power supply into the wall power  outlet  A dedicated rack mount controller might be provided     The optical head may now be powered on and is ready for use     SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 15 of 51       Engines for Thin Film Innovation       Other Recommended Usage Notes    Window Glare Reduction    The LEDs that direct light at the substrate for the reflectometry measurement can also generate  unwanted glare from the viewport window through which 
50. es  shall be subject to  the following conditions     A  The CUSTOMER must make notification to SVT Associates within thirty  80  days of original  shipping date     B  A   RETURN GOODS AUTHORIZATION    number must be assigned to and accompany all  goods or materials being returned by the CUSTOMER to SVT Associates  SVT Associates  must assign said number prior to any and all returns  Goods not accompanied by a    RETURN  GOODS AUTHORIZATION    number will be refused by SVT Associates and returned at the  CUSTOMER S expense     CUSTOMER shall prepay shipping charges for products being returned to SVT Associates     D  Products being returned to SVT Associates should be properly crated for shipment  and the  CUSTOMER shall bear the risk of loss until delivered to SVT Associates     E  Products being returned to SVT Associates must be returned in the condition originally received  by the CUSTOMER and free from damage  use  or modification  which would render the  product unusable for resale  by SVT Associates     F  All applicable taxes  duties  insurance  and shipping charges shall be the sole responsibility of    SAVA T    SVT Associates  Inc     O    7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 50 of 51          Engines for Thin Film Innovation PROPRI
51. f an opaque substrate  a perfectly specular substrate front  surface  and perfect uniformity of the deposited films  the emissivity may be measured via reflectometry  at the same wavelength as pyrometry  The relation linking emissivity       and reflectometry  R  under  these conditions is        1   R  The In Situ 4000 provides a 950 nm reflectometer which is matched to the  950 nm pyrometry system thus allowing this    emissivity compensation    to provide an emissivity  independent temperature measurement  Thus  if the film being deposited has a different optical  absorption at 950 nm  and or has a different index of refraction as the substrate producing interference    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 8 of 51          Engines for Thin Film Innovation PROPRIETARY INFORMATION    LL              effects  reflectometry will observe these changes and the software   will correct the error in the temperature calculation  The success of this technique depends upon how  closely the material system conforms to the assumptions listed above  namely substrate opacity  surface  specularity  smoothness   and film uniformity  If the substrate is not sufficiently opaque or the surface is  rough and scatters significant 
52. f refraction     The Growth Rate Fitter instrument fits a model    of the oscillations to the data in real time to determine  the film   s optical parameters and growth rate  The fitter can analyze the reflectance data from both  reflectometer channels  The acquired data  the model data and the parameter values are all displayed in  the instrument s window     1 See for example  W G  Breiland and K P  Killeen  J  Appl  Phys  V 78  p  6726  1995      Setup    Reflectometer Window    Reflect 951 30 nm Reflect 485 00    0 0000  0 0000           Plot OFF ON       12 46 52 PM me pain SF    12 47 41 PM X 24  omi Refl 470 AZ    I 1 i  06 00 00 PM 06 00 01 PM      Growth Fitter Outputs  Growth Fitt     pial i    Fit Tolerance              Figure 39     Growth Fitter Window    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 47 of 51          Engines for Thin Film Innovation PROPRIETARY INFORMATION    SS n       The Growth Rate Fitter instrument is installed on the computer   during the In Situ 4000 installation procedure  It will appear in the lower window of the Reflectometer  Window  In order to make use of the Fitter several input parameters are required to control the fitting  algorithm  These parameters are located on the left 
53. fects in material and workmanship and conform to COMPANY  specifications  No warranty is provided by SVT Associates for products sold hereunder which are not  manufactured or processed by SVT Associates  but the manufacturer   s warranty for such products  if  any  shall be assigned to the CUSTOMER without recourse to SVT Associates The foregoing warranties  are in lieu of and exclude all other warranties not expressly set forth herein  whether expressed or  implied by law or otherwise  including without limitation any warranty of merchantability or fitness for a  particular purpose  In no event will SVT Associates be liable for any consequential damages     IN THE EVENT OF SVT ASSOCIATES LIABLITY  WHETHER BASED ON CONTRACT  TORT   INCLUDING BUT NOT LIMITED TO NEGLIGENCE AND STRICT LIABLITY  OR OTHERWISE  THE  CUSTOMER S SOLE AND EXCLUSIVE REMEDY WILL BE LIMITED  SVT ASSOCIATES HAS THE  FOLLOWING OPTIONS  TO REPAIR OR REPLACEMENT  F 0 B  SVT ASSOCIATES  MANUFACTURING PLANT  BY THE COMPANY OF ANY NON CONFORMING ITEM FOR WHICH  CLAIM IS MADE BY THE CUSTOMER OR TO REPAYMENT OF THE PORTION OF THE PURCHASE  PRICE PAID BY THE CUSTOMER ATTRIBUTABLE TO THE NON CONFORMING ITEM    SVT ASSOCIATES WILL NOT BE LIABLE FOR ANY OTHER DAMAGES  WHETHER DIRECT   INCIDENTAL  CONSEQUENTIAL OR OTHERWISE     Return Policy   Any request by the CUSTOMER for return of standard products other than for warranty claims under  warranty hereof  for all or any part of purchase order accepted by SVT Associat
54. flectometer measuring a diffusing scattering surface  or preferable tilt the substrate slightly so the  reflected image of the LED is no longer visible in the video image  Another option is to cover the LED   s  so they cannot be seen in the video monitor  This case represents a baseline measurement for which  stray LED light is accounted in the calibration  The substrate is then restored to its normal tilt position   the LED reflection is visible in the center of the video image or the LED   s are uncovered  You can  calibrate the reflectance under this condition because the substrate reflectance is assumed to be known  such as for a bare semiconductor wafer     Reflectance Calibration  Go to Configuration to Reflectance Calibration    Reflectance Calibration       Data Acquisition Setup  Wobble Filter  Debug    o The Reflectance Calibration window will appear     SVT Associates   Acculemp    File Engine Temperature Reflectometer Configuration Help    Reflect 951 30 Reflect 485 00    Signal Reflect Signal Reflect       Low 0 00  gt   Low 0 00  High 1000 00 High 1000 00    Current Reflect 951 30 nm Current Reflect 485 00 nm  5 30 7 28          MINIMIZE the signals by Channel Wavelengths  aligning optical head    Pyrometer Channel 1  950 nm  Pa Vo Low fl D gt  Pyrometer Channel 2  850 nm  Get Current Data  set  gt  gt  Reflectometer Channel 1  951 3 nm    Reflectometer Channel 2  485 nm    Save Calibration AEN    Communication Initialized 0  LED   OFF Logging OFF    Figure 28 
55. he pyrometry measurement and the emissivity measurement  may no longer be correlated since they may be measured at different places on the substrate and hence  difference thicknesses  If that is the case then the temperature compensation may be compromised and  a temperature error may result  Substrate rotation or other means of ensuring film uniformity is strongly  recommended     SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 17 of 51       Engines for Thin Film Innovation PROPRIETARY INFORMATION       Optical Head Alignment   The optical head must be positioned so that the LED illuminator and the objective lens on the front of the  instrument have a clear and unobstructed view of the substrate under measurement  In addition  the  optical head must be oriented perpendicular to the substrate so that the reflected light from the  illuminator is directed back into the objective lens  If this alignment cannot be achieved  then the  reflectance measurement is not possible and the pyrometry performance is compromised     Alignment of the optical head using the In Situ 4000 internal light source   With the optical head power on  start the software application and turn on the LED illuminator     Point the optical head so you can
56. ions can  be analyzed in real time to provide both film thickness and film index of refraction  The In Situ 4000 has  powerful analysis algorithms  which can provide accurate growth rate  thickness  and index information   This valuable information can even be fed back to the growth system controller to achieve layer  thickness precision improvements over traditional open loop timed deposition control  The best growth  rate and index of refraction measurements are obtained with films of at least 4 4 in thickness where    is  the reflectometer measurement wavelength  i e  950 nm or 470 nm   See Growth Rate Fitter Chapter        r AYN  MART  TTI    Y    ro  T a      B GaN Growth AlGaN Growth a    0 Ooo oo T ee  0 2000 _ 4000 6000  Time  s   Figure 5     Growth Rate Measurement during GaN AlGaN Growth  using Two Wavelengths  950 nm and 470 nm           Reflectance                               2  The specular reflectance of a thin film may change with deposition of the film due to multiple  layer interference if the film has an index of refraction that is different from the underlying  material  If the index of refraction of the depositing film is known  at elevated temperature  then  the film thickness and growth rate may be calculated from the periodic undulations in specular  reflectance with time     Emissivity Compensated Pyrometry    The problem of unknown or shifting emissivity is solved using information from the accompanying  reflectometer system  Under the conditions o
57. is not affected by material properties of the grown layer  However  the signal for band gap  thermometry is attenuated by materials having smaller bandgap energy  This degradation depends on  layer thickness   lt 1 micron  but deteriorates measurement performance over time  Once an opaque layer  has been deposited   lt 100 nm  pyrometry can in such case take over for uninterrupted temperature  monitoring and calibration of the pyrometer  This technique is superior to eutectic melting point  calibration  For materials with higher band gap both techniques can be used                 aa S     9   amp  ie        D    O 400 v  zZ       Q  a       Pyrometer      amp  Eutectic Al Si  Ag Si         Bonded TC       100 200 300 400 500 600 700  800 900 1000  Temperature    C     Figure 38     Combination of pyrometry and band gap thermometry for GaN temperature  monitoring over extended temperature range and for optimum accuracy    During heterostructure growth interference oscillations due to multi layers occur  see Application note  1401   Emissivity corrected pyrometry accounts for these changes and automatically corrects the  temperature readings  However  bandgap thermometry is affected by these oscillations and accurate  determination of the bandgap is affected  By using sophisticated derivative analysis this effect can be at  least reduced     SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone 
58. is problem  It calculates the approximate  rotation rate of the substrate and looks back one rotation period in history to find and return the highest  reflectance value observed  Thus the reflection can completely leave the field of view for a short time  and the reflectance value can be preserved during the interruption     To activate the Wobble Filter  Go to Configuration to Wobble Filter    Help    Reflectance Calibration  Data Acquisition Setup  Wobble Filter       Debug    o The Wobble Filter Window will open    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 36 of 51           Engines for Thin Film Innovation PROPRIETARY INFORMATION       SVT Associates   AccuTemp       Fik e Engi Eng ine Temperat  2 Reflectometer Coni figuration   Wobble Filter  Purpose  To minimize the effect of substrate rotation on reflectance signals   C Wobble Filter on off Enter Estimated Wobble Period  S  10 BED                 200 0         j   1                       1  0 0 10 20 30 40 50 6 0 70 8 0 taps 11 0 12 0 13 0 14 0 15 0 16 0 17 0 18 0 19 0   Communication Iniiakzed o  LED OFF Logging OFF  Figure 33     Wobble Filter Window  Enter an estimated value for the filter keeping in mind that the response time of the reflectometer is  slow
59. l Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 18 of 51          Engines for Thin Film Innovation PROPRIETARY INFORMATION    SS n       Software    Software Installation    The software is installed under Windows  XP  or Windows 7 operating system by executing the program     Setup exe    on the supplied DVD  Follow the directions given by the installation program  The DVD  provides calibration information  please use this calibration information when prompted by the installation  program     At this point the software installation has the factory calibration values installed on the computer hard  disk and will remember these values each time the RoboMBE program is executed     Note  This procedure to load the factory calibration values must be performed after each installation of  the RoboMBE program     Note  The first time the application is run  the serial RS232 communication port  or COM Port  used by  the computer must be assigned and the communication values specified  The correct values to use are     Data Rate 9600 bps  Parity None  See Data Acquisition Setup   Stop Bits 1    Menus  Below is an overview of the software menus with a short description     SVT Associates   Acculemp    Engine Temperature Reflectometer         Engine Windows   Temperature Window Reflectometer Window    Gain Files   Data Acquisition Setup  Pyrometer Look Up Table Wobble Filter    Debu  g    Exit ea  Figure 17  File Menu    Log File     turn file logging on or off  Page 
60. ng  Standard pyrometry computes temperature from the intensity of emitted  radiation  If the viewport window through which the pyrometer measures the radiation becomes  coated over time  the pyrometer temperature calibration will drift accordingly  This problem is  addressed in the In Situ 4000 through the use of    two color    or    ratio    pyrometry  which  measures optical radiation at two wavelengths and takes the ratio of the measured intensities  If  the film coating of the viewport is spectrally neutral then the temperature may still be calculated  since these errors affect both channels equally  Since many of the viewport coating problems  deal with    metallic    type films  e g  Ga  In  As   these films attenuate the light equally between  the two wavelengths  the ratio calculation is still valid  If the viewport becomes coated with a  dielectric film causing interference effects  then the ratio measurement becomes imbalanced and  may not be valid     Alignment and optical obscuration errors  This problem is similar to viewport coating in that there  are errors introduced into the collection of radiated intensity by misalignment of the optics to the  viewport window or other obscuration problems     Pickup of stray light  Any hot surface in the chamber can radiate light  If that light makes its way  into the pyrometer  the pyrometer will not be able to distinguish between    black body    radiation  of the substrate itself and that of the stray light  Thus the str
61. nt at the user supplied    guesses     until it determines that there is enough data to allow them to vary  If the user has specified that the film n  and k be allowed to be fit  then once the film has reached sufficient thickness the fitter frees them and  allows the algorithm to modify the values  If the user has specified either of the parameters as fixed  then    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 48 of 51          Engines for Thin Film Innovation PROPRIETARY INFORMATION    SS        they will remain fixed throughout the current film deposition  The fit  status of the scaling factor  s  is not changed by the adaptive fitting     Controlling the Fit    The fit algorithm is fairly sensitive to the initial guess provided by the user  The reason for this sensitivity  is that the fit error    surface    used to fit the model to the data may contain several local minima  Some of  these local minima are almost as    deep    as the true global minimum     The In situ 4000 Growth Rate Fitter solves this problem by limiting the parameter space in which the fit  algorithm searches  The user specifies how far the fit algorithm may deviate from the initial guess in its  search for the true minimum  All parameters except
62. of the flange  mount to hold it in place  Make sure  both 14 20 screws on the bottom are  fastened to hold the optical head on  the platform     e Make sure the objective lens is  approximately centered in the  mechanical holder to start  See  Optical Head Alignment section on  page 18 of this manual for detailed  information     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com  In Situ Process Monitor Version 2 0 4 13        n    Figure 12    PROPRIETARY INFORMATION       Figure 10       Figure 11       SAAT    SVT Associates  Inc     SVT Associates Proprietary Information    2013 SVT Associates  Inc  All Rights Reserved  Page 14 of 51          Engines for Thin Film Innovation PROPRIETARY INFORMATION    SS       Electrical Connections  The In Situ 4000 optical head provides a panel of connectors on the rear of the unit for connection of  electrical cables  The connectors are from left to right  video signal  sync signal  computer interface  and  power supply   See Figure 13   Three cables  supplied  are needed for normal operation    1  Power supply cable 5034399   2  Computer serial interface cable 5022756    3  Video signal cable 5005834    Video Signal Sync Pulse Computer Interface Power Connector Power Switch    Me    www svta com Process Monitor       2  SVT Associates    lt    In Situ4000          Figure 13     Rear view of the optical head showing connector panel     Connect one 
63. ond duration  In addition to the growth rate is calculated  using the time between cursors and growth rate from the fit  The results of the fitter are stored ina  dedicated results file     SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 49 of 51          Engines for Thin Film Innovation PROPRIETARY INFORMATION    SS       Warranty and Limitations of Remedies   SVT Associates warrants that all equipment manufactured by it shall be free from defects in materials  and workmanship under normal use and service for a period of twelve  12  months from the date of  shipment from SVT Associates manufacturing facility  This warranty is subject to SVT Associates  equipment being installed  maintained  and operated in accordance with the operating and maintenance  instructions accompanying each item manufactured by SVT Associates  Warranty shall be void if   SVT Associates equipment is modified by the CUSTOMER or used in other than the recommended  manner or applications  Purchased equipment incorporated into any item supplied by SVT Associates  will be covered by said manufacturer s warranty     SVT Associates warrants that  at the time of delivery  any other products processed or manufactured  and sold by it hereunder are free of de
64. ounds  O  Madelung ed   Springer   Verlag  1991   The data for GaAs is based on ellipsometry measurements by Aspnes et  al   Phys   Rev  B 27  p  985  1983      Reflectance of an assortment of substrate materials  at normal incidence and at room temperature     Substrate 950 nm Reflectance 470 nm Reflectance  Silicon 0 326 0 419    Ge 0 419 0 492  GaAs 0 313 0 438  GaSb 0 39   InAs 0 33 0 454  InP 0 305 0 349  InSb 0 41  Sapphire 0 075 0 078    These values are obtained from the Palik database  Handbook of Optical Constants of Solids   Edward D  Palik  Academic Press  First edition 1985  2nd Edition 1991  and the Landolt Bornstein  tables  Semiconductors  Group IV Elements and III V compounds  O  Madelung ed   Springer   Verlag  1991   The data for GaAs is based on ellipsometry measurements by Aspnes et  al   Phys   Rev  B 27  p  985  1983             IN   SJ         iN  CO    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 39 of 51          Engines for Thin Film Innovation PROPRIETARY INFORMATION    SS       Editing the Ratio LUT    The translation from pyrometry intensity ratio to temperature is performed using a look up table  LUT   A  LUT was chosen because many semiconductor substrates are not ideal grey body emitters 
65. ransparency  Semiconductors by nature are partially transparent for wavelengths  below the band gap that is usually in the infrared portion of the spectrum  This transparency  causes two severe limitations in pyrometric measurements  low substrate emissivity and optical    interference from other hot elements in the chamber   1 0        Substrate Transparent  for NIR wavelength    0 5  0 5 mm GaAs    SPECRAL EMISSIVITY    0 5 1 0 1 5  WAVELENGTH  microns   Figure 7     Substrate Transparency of GaAs requires small spectral detection range  and limits Temperature Measurement to 370   C   950 nm    The problem of low emissivity and substrate transparency at infrared wavelengths is addressed  by careful selection of the pyrometry wavelength  The In Situ 4000 performs single wavelength  pyrometry at 950 nm  which is short enough to ensure that the substrate is opaque  and long  enough to ensure a measurable optical radiation at growth temperatures  Ensuring substrate  Opacity is a key element in addressing several of the previously listed difficulties  An opaque  substrate prevents radiation from a filament heater from being transmitted though to the  pyrometer  It also prevents other radiation sources in the chamber  effusion cells  ion gauge  filaments  etc   from scattering their light from the rough back side of the substrate into the  pyrometer  Since optical absorption and black body radiation are closely coupled  substrate  Opacity ensures that the emissivity of the substr
66. s  temperature relationship is exponential and allows a direct  mapping of intensity to temperature  An example radiance curve vs  temperature at 950 nm is shown in  Figure 2     SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 5 of 51          Engines for Thin Film Innovation PROPRIETARY INFORMATION    A n       100000                               10000                               1000                               100                         Radiance  x10 4 W m 3 st                                                              400 500 600 700 800 900 1000  Temperature  C     Figure 2     Radiance vs  Temperature at 950 nm    Real surfaces have emissivity values less than 1 so the pyrometer instrument used to measure a surface  temperature must have knowledge of the particular surface emissivity     Two color Pyrometry    For surfaces where the emissivity is not known or the radiometer cannot be calibrated with absolute  values  a technique known as    ratio    or    two color    pyrometry may be used  Here  a single optical system  is used to measure the radiance at two distinct wavelengths and the ratio of the radiant intensities is  calculated  Assuming that the emissivity is the same for the two wavelengths  i e  a   
67. s  which can allow the measurement to function well even if the  substrate tilts strongly during wafer rotation   See Wobble Filter Chapter     e Reflectometry       Reflectance is used in this instrument for two reasons     1  When single or multiple layer films are deposited  optical interference will alter the emissivity of  the substrate as a function of time and the single color pyrometry will be in error  Measurement  of specular reflectance at the same wavelength as the single color pyrometry will allow the  calculation of emissivity through the relation       1   R which is based on conservation of energy  considerations and Kirchhoffs Law  Note that this relation is valid only for opaque substrates     SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 7 of 51       Engines for Thin Film Innovation PROPRIETARY INFORMATION       Principle of Thin Film Interference  2n2   mA for maximum reflectance   2n2    m      A for minimum reflectance  n index of refraction   k extinction coefficient       d layer thickness   Figure 4  Reflectometry is obtained in real time during deposition so that films that have differing index of  refraction from the substrate produce reflectance oscillations in time  These reflectance oscillat
68. side of the screen  see Figure 39      The fitter uses a Simplex algorithm to minimize the error function of the model  The analytical model  used treats any material underneath the currently growing film as a    virtual substrate    described only by  a single complex index of refraction  The virtual substrate can be a simple material like the physical  substrate itself  or it can be a complex combination of thin film layers on a physical substrate  In any  case  the virtual substrate is modeled with a real substrate index of refraction  Ngyp  and imaginary  substrate index of refraction  Ksub     The model contains the following fit parameters     1  G Growth rate in nm s    2  N Real part of the film   s index of refraction    3  k Imaginary part of the film s index of refraction    4  Neup Real part of the effective substrate s index of refraction    5  Keup Imaginary part of the effective substrate s index of refraction    6  s A scaling factor used to compensate for reflectance attenuation by unintended coatings  deposited on the vacuum viewport window    7  Load Setup File button     Load any setup files that have been saved     Save Setup File button     Saves current settings to a setup file to be available for loading at a  later date     9  Open Log File Growth Fitter Results     Opens the log file for the Growth Rate Fitter     The fit algorithm always adjusts the G  ngyp and Ksub parameters  The user can specify whether or not to  fit the other three parameters  
69. ssociates  Inc     SVT Associates Proprietary Information    2013 SVT Associates  Inc  All Rights Reserved  Page 13 of 51           Engines for Thin Film Innovation    Mounting Using SVT Associates    Flange  e Locate the In Situ 4000 Flange Mount     The mount should be fully assembled     4 20 screws  4     e There are four 14 20 socket head set  screws around the circumference of  the mount  Using a 7 32 Allen wrench  back the set screws out until they  clear the inner diameter of the mount    See Figure 10  The flange mount has  an ID tolerance of approximately  015  inches  Nothing  such as bolt heads or  washers  may be protruding beyond  the flange  Use socket head screws if  necessary on your flange     e Install the mount over the flange with  the platform down  Tighten all four set  screws equally until the mount is firmly  attached to the flange     e Looking at the mount find the four 8   32 screws that attach the platform to  the side gussets  Note that the  gussets are slotted  These slots allow  the platform to be moved vertically  and also to be tilted fore and aft     e The optical head is attached to the  platform through two slotted cutouts   These slots allow the optical head to  be moved left and right and also allow  for a rotational adjustment   See  Figure 11     e Locate the optical head and mount it  to the platform using two 14 20 hex  head bolts     e Mount the IS4K head using provided  flange mount to the chamber viewport   Use all four setscrews 
70. th rate monitoring using its multi wavelength reflectometer  This enables precise thickness  monitoring  rate monitoring and surface roughness analysis     Combining emissivity corrected pyrometry and band gap thermometry in a single instrument provides  optimum performance for temperature monitoring during heterostructure growth  The AccuTemp process  monitor provides dedicated hardware  an easy to use software interface and ready to use growth  recipes  Addition of its spectral reflectometer enables superior process monitoring and control        PROPRIETARY INFORMATION    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 46 of 51           Engines for Thin Film Innovation PROPRIETARY INFORMATION    SS nn    Growth Rate Fitter   The Growth Rate Fitter instrument utilizes data from the reflectometer instrument  When a film is  deposited on a substrate of differing index of refraction  the optical reflectance generally oscillates in  time due to optical interference effects  see page 7     Reflectance   The period of the oscillation is  directly dependent on the growth rate of the film and the film refractive index  The amplitude  phase and  damping characteristics of the reflectance are dependent on the film s complex index o
71. the instrument is viewing the substrate  If the  optical head is separated from the vacuum viewport window by more than 100 mm  there will be a  reflection from that window back into the reflectometer  This glare can be accommodated in the  reflectance calibration  however the glare signal is often larger than the reflectance from the substrate  under measurement and has the effect of significantly reducing the signal quality  i e  more noise  and  making the instrument more sensitive to misalignment errors and mechanical motion  Use of the optical  head with less than 100 mm of distance to the viewport is highly recommended     Use With Small Viewports    The optical head is designed to view through vacuum viewports with at least a 33 mm diameter window   which is found  for example  on a standard 2  inch Conflat Pyrex viewport flange  Smaller viewport  windows are not recommended because either the outgoing LED light or the returning reflected light  or  the black body radiated light will be obstructed by the window frame and the measurement results may  be incorrect  Contact SVT Associates for further information on using smaller viewports     Use With Transparent Substrates    The In Situ 4000 is designed for use with substrates that are optically opaque at 470  850  and 950 nm   Care must be exercised when attempting to use the instrument with transparent or partially transparent  semiconductor substrates  Substrate transparency causes the following problems     1  The r
72. tion  Page 40      SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 26 of 51           Engines for Thin Film Innovation PROPRIETARY INFORMATION    Engine Menu    Operation of the Engine Instrument    To open the Engine Window  Go to Engine to Engine Window       ay Temperature    Engine Window       o A separate window will appear    o This window acts as the main interface to collect data and operate the hardware     Engine Window KBS   950nm Pyrometer T am Pe A agint eee         Plot OFF ON      Plot OFF ON    Plot OFF ON    Plot OFF ON    Pyrometery Gain Reflectance Gain Reflectance Gain     10 N Ea Increase E    oa i    400 5  06 00 00 PM       Figure 24     Engine Window    1  Pyrometery Gain  these fields allow manual gain adjustment of the signal channels   See  Page 20     Reflectance Gain Buttons    To start data collection from the optical head  click on the Power toggle switch  Turn it to the     ON    position  The Status Bar shows the status of communication     4  Turn on the LED illuminators on the optical head by clicking the LED switch  The Status Bar  shows the status of the LEDs  ON OFF      5  The Debug Normal switch provides diagnostic information on the internal operation of the optical  head  When the 
73. tion values  for electronic and optical gain will be transferred from the supplied DVD to the computer hard disk but  will not be used by the program until instructed to do so by the user  There are 5 files to be loaded by the  user  and once they have been loaded  they are automatically saved by the program and need not be  reloaded again on that computer     To reset the hardware amplifier gain and offset values back to factory defaults  follow the instructions  below     To open the gain files control screen  Go to File to Gain Files     eg Engine Temperature Ref    Gairm Files      Pyrometer Look Up Table    Exit       o The Gain Files window will appear    SVT Associates   Acc culenp                Pyrometer C2   850 nm  io  ens Oe  Mie     3  TE N  5 eal  tooo E    C  aaa mo a    Support p850_4012 gnc                  C   CodelLabViewlIS4K amp Grawth LY  Lv       See a 4012  gnc    Reflect C1   950 nm Reflect C2   470 nm   ee a  Sica on   ee Ge  igs Sy ore pe a Peet  1000 se   i     ooo  929 433 Eao  p cod abenitsKB Groh LV Wy i 3   CA CodetLabviewiIs4K amp Growth Lv     a 4      i Supportir950_4012 gnc q Supportir470_4012  gnc                   Automatic Gain Control 5   Muse Automatic Gain Control             Communication Initialized o  LED OFF LoggngOFF   a  Ria es           amp                 Figure 22     Gain Files Window    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952
74. uration  The actual  wavelengths used for computation are  950 nm  850 nm  650 nm  and 470 nm  Refer to your In Situ  4000 calibration sheet for values specific to your setup  Save the values after entering for future use     Data Acquisition Setup    Go to Configuration to Data Acquisition Setup    Reflectance Calibration    Data Acquisition Setup  Wobble Filter  Debug       o The Data Acquisition Setup window will appear    SVT Associates   Acculemp      File Engine Temperature Reflectometer Configu       Data Acquistion Setup Geometry Factor  Averaging  1   Method of temperature Reference    User defined Temperature      Scan Timers   250   A      Com Port Ji   User Defined Reference Temperature  C   0  Comm Delay  Sec  t  0 050   Calculated Geometry Factor           Sateen   Gain Blank Delay     2 000   calculate   i 2 100E 5    Set Geometry Factor 7 983E 7  Fixed Geometry Factor  C    Current Emissivity 10 998  Set Emissivity 0 613    Fixed Emissivity   C     ae     Communication Initialized LED  OFF Logging OFF    n    E         Figure 30     Data Acquisition Setup Window    1  Averaging     Sets the number of data points to be averaged to minimize noise in graphs typically  values from 1     4 are used     2  Scan Time     Sets the S232 port numbers that will be used to communicate with the In Situ 4000    SAVA T    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934
75. ustment capability for precise pointing of the instrument        Specular  Geometry    Receiving Head    Remote    Illuminator                AccuTemp     Receiving Head            Normal Incidence Geomerty          Figure 9     Geometry    7620 Executive Drive   Eden Prairie  MN 55344 3677   USA    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com  In Situ Process Monitor Version 2 0 4 13    The optical head should be mounted within 4  inches of the vacuum viewport window for  best results  Mounting the instrument close to  the window reduces the LED illuminator    s light  from being reflected back directly into the  instrument     The vacuum viewport flanges compatible with  this instrument are 2 75  Conflat    Flange or  larger  Ensure that neither the LED  illuminators on the front of the optical head   nor the objective lens are blocked from  viewing the substrate     The instrument is designed to view the  substrate from a normal or specular incidence  so that the light from the illuminator is  specularly reflected back toward the  instrument  If this geometry cannot be  achieved  then the reflectometry portion of the  instrument will not produce useful information  and the pyrometry performance will be limited  by no longer allowing emissivity  measurement  If the user requires a non   normal incidence  a remote illuminator may  be purchased which will provide a specular  reflection of the LED light into the optical  head     SAAT    SVT A
76. when multiple layer interference occurs because the  In Situ 4000 has the ability to measure the emissivity at the same wavelength as the single  channel pyrometry     The Temperature instrument has a dialog box that allows you to use the ratio temperature method to  calibrate the single channel 950 nm pyrometer when it is valid to do so     Go to Configuration to Data Acquisition Setup      Configuration   Help    Reflectance Calibration    Data Acquisition Setup  Wobble Filter  Debug          o The Data Acquisition Setup window will appear    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information    Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved  In Situ Process Monitor Version 2 0 4 13 Page 33 of 51       Engines for Thin Film Innovation PROPRIETARY INFORMATION        SVT Associates   Acculemp       x mar aA E H  ere ii JratIgI nel  ture Reflectometer Configuration Help                       Method of temperature Reference    Two channa Terperature    Scan Time ms   250   ESD  Com Port Hfi User Defined Reference Temperature  C   0  Comm Delay  Sec     0 050 Calculated Geometry Factor        Gain Blank Delay     2 000     cantare   100    ro  1  Set Geometry Factor 7   7    Fixed Geometry Factor  C               Fixed Emissivity  amp  Geome    Current Ernissivity 0 998  Set Emissivity 0 613    Fixed Emissivity   C          
77. when you are done  No data is lost while AutoScale is off   the screen is just not updated during that time     By clicking on the magnifier icon you can   Zoom into a selected area  top left   Zoom Horizontally  top middle     Zoom vertically  top right         Zoom back out to original  bottom left  Figure 19     Zoom Features  Zoom in around the cursor location  bottom middle     Zoom out around cursor location  bottom right      File Menu    Logging Data to Files    To open the Data Logging window   Go to File to Log Files    Gain Files  Pyrometer Look Up Table       Exit    o The Log File window will appear    SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairie  MN 55344 3677   USA SVT Associates Proprietary Information  Phone  952 934 2100   Fax  952 934 2737   Email  sales svta com   www svta com   2013 SVT Associates  Inc  All Rights Reserved    In Situ Process Monitor Version 2 0 4 13 Page 21 of 51           Engines for Thin Film Innovation PROPRIETARY INFORMATION    SVT Associates   Acculemp          Log File Settings  Log File Size  Click to Log Data   f0 bytes    Log File Path      10sec Log Interval    Indicator of data packages via COM port   typically 16 34     I Communication Initialized LEC  Figure 20     Log File Window    To activate Data Logging  click the    Click to Log Data    button    o Anew window will appear asking you to name and select where to save the log file     SAAT    SVT Associates  Inc     7620 Executive Drive   Eden Prairi
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
2711-6.0.4, DeviceNet User Guide  Engines for Biogas (1988) (GTZ-Gate) (introductory      Page 1 de 2  QTP 03 - Grifo  MARKETLINE    Copyright © All rights reserved. 
   Failed to retrieve file