        Human Parainfluenza Virus Type 2 Real Time RT-PCR Kit
         Contents
1.     Liferiver       Revision No   ZJO003  Issue Date  Jul 1    2012    Human Parainfluenza Virus Type 2 Real Time RT PCR Kit C        User Manual   20   C  Z I    For In Vitro Diagnostic Use Only   RR 0157 01  rw N ZJ Bio Tech Co   Ltd   www liferiver com cn Tel   86 21 34680596    For use with LightC ycler1 0 2 0 Instrument  trade  liferiver com cn Fax   86 21 34680595    pec   rer  Obelis S A   Boulevard G  n  ral Wahis 53   1030 Brussels  BELGIUM   Tel    32  2 732 59 54 2  floor No 15 Building No 188 Xinjunhuan road   Fax    32  2 732 60 03 PuJiang Hi tech Park Shanghai China   E Mail   mail  obelis net    1  Intended Use  Human Parainfluenza Virus Type 2 real time RI PCR kit is used for the detection of Human  Parainfluenza Virus Type 2 in nasal and pharyngeal secretions by using real time PCR systems   2 Principle of Real Time PCR  The principle of the real time detection is based on the fluorogenic 5   nuclease assay  During the  PCR reaction  the DNA polymerase cleaves the probe at the 5    end and separates the reporter dye  from the quencher dye only when the probe hybridizes to the target DNA  This cleavage results in  the fluorescent signal generated by the cleaved reporter dye  which is monitored real time by the  PCR detection system  The PCR cycle at which an increase in the fluorescence signal is detected  initially  Ct  is proportional to the amount of the specific PCR product  Monitoring the  fluorescence intensities during Real Time allows the detection of 
2.  e All reagents should be stored at  20  C  Storage at  4  C is not recommended    e All reagents can be used until the expiration date indicated on the kit label    e Repeated thawing and freezing   gt  3x  should be avoided  as this may reduce the sensitivity of the assay   e Cool all reagents during the working steps    e Super Mix should be stored in the dark     6  Additionally Required Materials and Devices  e Biological cabinet  e Real time PCR system  e Desktop microcentrifuge for    eppendorf    type tubes  RCF max  16 000 x g   e Vortex mixer  e RNA extraction kit  e Real time PCR reaction tubes plates  e Cryo container  e Pipets  0 5 ul     1000 ul   e Sterile filter tips for micro pipets  e Sterile microtubes  e Disposable gloves  powderless  e Biohazard waste container  e Refrigerator and freezer  e Tube racks   7  Warnings and Precaution  Carefully read this instruction before starting the procedure   e For in vitro diagnostic use only   e This assay needs to be carried out by skilled personnel   e Clinical samples should be regarded as potentially infectious materials and  should be prepared in a laminar flow hood   e This assay needs to be run according to Good Laboratory Practice   e Do not use the kit after its expiration date   e Avoid repeated thawing and freezing of the reagents  this may reduce the sensitivity of the  test   e Once the reagents have been thawed  vortex and centrifuge briefly the tubes before use   e Prepare quickly the Reaction mix on ic
3. 38 40 Re test  If it is still 38 40  report as 1   PCR Inhibition  No diagnosis can be concluded     For further questions or problems  please contact our technical support at trade  liferiver com cn      i  Positive  and the software displays the quantitative value       
4. e or in the cooling block   e Set up two separate working areas  1  Isolation of the RNA  DNA and 2  Amplification        detection of amplification products   e Pipets  vials and other working materials should not circulate among working units   e Use always sterile pipette tips with filters   e Wear separate coats and gloves in each area   e Do not pipette by mouth  Do not eat  drink  smoke in laboratory   e Avoid aerosols  8  Sample Collection  Storage and transport  e Collected samples in sterile tubes   e Specimens can be extracted immediately or frozen at  20  C to  80  C   e Transportation of clinical specimens must comply with local regulations for the transport of  etiologic agents  9  Procedure  9 1 RNA Extraction  Different brand RNA extraction kits are available  You may use your own extraction systems  or the commercial kit based on the yield  For the RNA extraction  please comply with the  manufacturer   s instructions  The recommended Extraction kit is as follows     Nucleic Acid Isolation Kit Cat  Number  RNA Isolation Kit ME 0010 ME 0012 ZJ Biotech       9 2 Internal Control   It is necessary to add internal control  IC  in the reaction mix  Internal Control  IC  allows the user  to determine and control the possibility of PCR inhibition    Add the internal control  IC  1ul rxn and the result will be shown in the 560nm Channel    9 3 Quantitation   The kit can be used for quantitative or qualitative real time RT PCR  A positive control defined    as 1x10 copi
5. es ml is supplied in the kit    For performance of quantitative real time PCR  Standard dilutions must prepare first as  follows  Molecular Grade Water is used for dilution    Dilution is not needed for qualitative real time PCR detection     Take positive control  1x10 copies ml  as the starting high standard in the first tube  Respectively  pipette 36ul Molecular Grade Water into next three tubes  Do three dilutions as the following  figures    Dilution of Standards    Aul Aul 4ul    Y VY V Y    1X107 1X10 1X10  1X 104 copiesim  To generate a standard curve on the real time system  all four dilution standards should be used and  defined as standard with specification of the corresponding concentrations   Attention   A  Mix thoroughly before next transfer     B  The positive control  1x10 copies ml  contains high concentration of the target DNA  Therefore   be careful during the dilution in order to avoid contamination    9 4 RT PCR Protocol   The Master Mix volume for each reaction should be pipetted as follows     134l 1 yl 1y  Super Mix Enzyme Mix Internal Control  Sul 15l  Extraction RNA Master Mix  Reaction  Plate  Tube    i    PCR Instrument    XPCR system without 560nm channel may be treated with 1ul Molecular Grade Water instead of 1ul IC    1  The volumes of Super Mix and Enzyme Mix per reaction multiply with the number of  samples  which includes the number of controls  standards  and sample prepared  Molecular  Grade Water is used as the negative control  For reas
6. ons of unprecise pipetting  always add  an extra virtual sample  Mix completely then spin down briefly in a centrifuge    2  Pipet 15ul Master Mix with micropipets of sterile filter tips to each of the Real time PCR  reaction plate tubes  Separately add 5ul RNA sample  positive and negative controls to  different reaction plate tubes  Immediately close the plate tubes to avoid contamination    3  Spin down briefly in order to collect the Master Mix in the bottom of the reaction tubes    4  Perform the following protocol in the instrument       Fluorescence measured at 60  C   10  Threshold setting  Choose Arithmetic as back ground and none as Noise Band method  then  adjust the Noise band just above the maximum level of molecular grade water  and adjust the  threshold just under the minimum of the positive control   11 Calibration for quantitative detection  Input each concentration of standard controls at the  end of run  and a standard curve will be automatically formed     12 Quality control Negative control  positive control  internal control and QS curve must be  performed correctly  otherwise the sample results is invalid     530nm   Molecular Grade Water   Positive Contol qualiaiveasayy   35    QS  quantitative detection     13  Data Analysis and Interpretation   The following sample results are possible     Crossing point value l  Result Anal  O Tam T sam OO Patam    5  1  25 35 Below the detection limit or negative  25     Selection of fluorescence channels          
7. the accumulating product without  having to re open the reaction tube after the amplification   3  Product Description  Human parainfluenza viruses are second to respiratory syncytial virus  RSV  as a common cause of  lower respiratory tract disease in young children  Similar to RSV  HPIVs can cause repeated  infections throughout life  usually manifested by an upper respiratory tract illness  e g   a cold  and or sore throat   HPIVs are negative sense  single stranded RNA viruses that possess fusion and  hemagglutinin neuraminidase glycoprotein  spikes  on their surface  There are four serotypes types  of HPIV  1 through 4  and two subtypes  4a and 4b   Each of the four HPIVs has different clinical  and epidemiologic features  The most distinctive clinical feature of HPIV 1 and HPIV 2 is croup   i e   laryngotracheobronchitis   HPIV 1 is the leading cause of croup in children  whereas HPIV 2  is less frequently detected  Both HPIV 1 and  2 can cause other upper and lower respiratory tract  illnesses  HPIV 2 causes annual or biennial fall outbreaks   The Human Parainfluenza Virus Type 2 real time RT PCR kit contains a specific ready to use  system for the detection of the Human Parainfluenza Virus Type 2 using RT PCR  Reverse  Transcription Polymerase Chain Reaction  in the real time PCR system  The master contains a  Super Mix for the specific amplification of the Human Parainfluenza Virus Type 2 RNA  The  reaction is done in one step real time RT PCR  The first step is a re
8. verse transcription  RT   during  which the Human Parainfluenza Virus Type 2 RNA is transcribed into cDNA  Afterwards  a  thermostable DNA polymerase is used to amplify the specific gene fragments by means of PCR   polymerase chain reaction   Fluorescence is emitted and measured by the real time systems     optical unit during the PCR  The detection of amplified Human Parainfluenza Virus Type 2 DNA  fragment is performed in fluorimeter channel 530nm with the fluorescent quencher BHQ1  In  addition  the kit contains a system to identify possible PCR inhibition by measuring the 560nm  fluorescence of the internal control  IC   An external positive control defined as 1x10   copies ml is  supplied which allow the determination of the gene load  For further information  please refer to  section 9 3 Quantitation   4  Kit Contents   HPIV 2 Super Mix 1 vial  350u1   RT PCR Enzyme Mix 1 vial  28ul   Molecular Grade Water 1 vial  400u1   Internal Control  IC  1 vial  30u1   HPIV 1 Positive Control  1x10   copies ml    1 vial  30ul  Analysis sensitivity  1X 10   copies ml  LOQ  2X10   1X 10 copies ml  Note  Analysis sensitivity depends on the sample volume  elution volume  nucleic acid extraction  methods and other factors  If you use the RNA extraction kits recommended  the analysis  sensitivity is the same as it declares  However  when the sample volume is dozens or even  hundreds of times greater than elution volume by some concentrating method  it can be much  higher   5  Storage   
    