1.    BETTER FUTURE     variant LIGHT     graphs showing the functions of the device     User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    The device ACUCOMB AC V2 edition    BETTER FUTURE   variant  LIGHT is fully controlled by microprocessors  which primarily means  even data interpreted by the analogue measuring instrument is first  processed by the computer  and evaluated in such a way that it is  completely accurate and always provides the maximum necessary  information  The analogue measuring instrument itself excels with its  sufficient size  high level of accuracy  and suitable hand movement  dynamics  For operating the device  a single revolving and push   button control is used  Using the device is therefore very simple and  problem free  Besides its visual output  the device also provides a  microprocessor synthesized sound indication  Acupuncture methods  often require that the relevant electronic devices for seeking out and  measuring in acupuncture points are equipped with metal bowls or  plates connected to the negative pole of the power supply of the  device  For this reason  the device ACUCOMB AC V2 edition     BETTER FUTURE  variant LIGHT is above standard  equipped with  one built in metal bowl of large dimensions  The device is powered  by a built in battery with a long life span  This battery may be  charged from the USB port of any computer  or with the help of a  mains adapter with a USB port     While the device is being charged fro
2.   adapter     Belkin  model F8Z222ea or another  equivalent  type    Warning  The charger  mains adapter  is not a  medical device        Electrotechnical  standard used    CSN EN 60 601     1  or IEC 601     1        Classification of the  auxiliary parts of the  electrical medical  device    BF  in accordance with standard CSN EN 60 601 1  or ICE 601 1        Classification  regarding protection  against penetration by  water or solids    IP 40  in accordance with IEC 60 529        Classification for  operation  in  premises           permanent   indoors  normal  without increased risk       12       User manual   ACUCOMB AC V2 edition       BETTER FUTURE    variant LIGHT             Temp  rat  r Operating  5 to  45  C  without condensation   emperaire Storage   10 to  50  C    ee Operating  15 to 90    without condensation   Relative humidity Storage  5 to 95    F Operating  425 to 809 mmHg  AtmoSpnerie pressuie Storage  120 to 809 mmHg       Sterilization     The device is not intended for sterilization        Power input    30 mW during operation       Securing the power  supply    A secure internal battery source as well as the creation  of an electrical fuse by a meander in the printed circuit        Electrical voltage  produced by the  device and applied to  the intact skin of the  patient client    Max  2 4 V       Electrical current  produced by the  device and applied to  the intact skin of the  patient client    Max  25 micro amperes       Weight    The whole devi
3.  ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT      which do not contain alcohol      which do not leave behind traces of colour  or cause  discolouration or damage to the outer surfaces of the  device      which allow gentle application     Examples of suitable disinfectants are Sanosil super Ag25  or Incidur  Foam     During longer idle periods  the device should be covered or packed  in protective foil packaging  During normal operation  it is necessary  to regularly wipe or vacuum clean dust which clings to the device or  any of its parts  using gentle and dry instruments  If more substantial  fouling needs to be removed  a soft mull  cloth or sponge should be  used  slightly moistened with water or a mild cleaning agent   cleaning agents containing alcohol  or abrasive cleaning agents   must never be used     Important notes   During cleaning or disinfection of the device       the device must be switched off and the built in battery  charger must be disconnected from the mains by pulling  it out of the electrical socket  or disconnecting it from  the USB port of the computer which in this case serves  as a power source for charging the batteries       it must be ensured that no liquids or particles penetrate  the closed  sheathed parts of the device     12  Environmental protection and disposal of the  device    he device ACUCOMB AC V2 edition    BETTER FUTURE     variant LIGHT does not have a directly negative effect on the  environment  Operating the device 
4.  Oncological illnesses and serious pre cancerous conditions  in  most of these cases the localized site of the disease can be  identified  so the contraindication is local and relative  Also  it is  unlikely that such a small electrical voltage and current could  by  some physical mechanism  affect healthy or damaged tissue  cells     e Serious ischemic diseases of vital organs  again we are dealing  with relative contraindication  associated with a hypothetical and  highly unlikely possibility of undesirable biological effects from  such negligible values of electrical voltage and current     e Cardiac rhythm disorders  disorders of the electrical conduction  system of the heart and any associated heart disease  again  the  application of such a small electrical voltage and current should  not result in a disruption of the electrical conduction system of  the heart  Also  it is a relative and hypothetical contraindication of  a local character     e Serious neurological and paroxysmic diseases  here  in theory   but this is highly unlikely  there may be  as a result of applying  voltage and current to an area of the head  interference with the  central nervous system  CNS   This is a local contraindication   The electrodes can be applied a safe distance away from the  localization of the CNS     e Recent thrombosis  this is a relative  local and temporary  contraindication     e Menstruation  this is a relative  local and time limited  contraindication     Using the device i
5.  a person not suitably qualified and knowledgeable  about using the device  The warranty also does not cover damage  caused by fire  water  static electricity  power surges  accidents   mechanical damage  unauthorized tampering with the device   improper installation  empty or leaking batteries  the results of normal  wear and tear  or consequences of force majeure     Consumable material and auxiliary equipment for the device are not  covered by the stipulated warranty conditions     Other conditions of the validity of the warranty are specific to medical  devices  medical devices should be subject to regular functionality  and safety inspections  these inspections must be carried out  exclusively by the manufacturer or by an authorized service agent   The manufacturer stipulates the time period of these inspections in  accordance with the anticipated life span of the built in batteries  i e   a time period of 3 years  Any servicing of the device must be carried  out only by the manufacturer  or by an authorized service agent   From the point of view of the legal requirement of complete  traceability  vigilance  of medical devices  it is essential that the  manufacturer of the devices knows the end user of the device  or  if  applicable  the relevant responsible organization     The persons servicing the device must be familiarized with the use of  the device at least to the extent of this User Manual  There are no  special qualifications required for the persons servicing t
6.  by the legal regulations for all  medical devices  The life span of the built in battery is dependent on  regular and sufficient charging  under normal conditions it should be  3 years     Warning     While the integrated battery is being charged  either from a  computer or from the mains adapter  the device must not be  used     Normal routine maintenance by the operator consists only of regular  and continuous inspection of the integrity of those parts of the device  which are important to its functionality and safety  as well as keeping  the device clean and essential disinfection of the electrodes  if  required  The operator must also heed all safety instructions listed in  this manual  At the same time  the operator of the device must pay  attention to any perceptible changes in the appearance and function  of the device  or its individual parts  If any serious irregularities are  noticed  an authorized service agent must be called     11  Cleaning and disinfection    The device  the whole device set  must be kept in a clean condition   and excessive fouling or contamination of it must be effectively  prevented     If  in spite of all precautions  contamination of the applied part  i e   the electrode  or any other part of the device results  or if it is  necessary to effectively disinfect the device before using it on  another patient client  we can carry out a disinfection of any part of  the device by using normal  but recognized disinfectants     24    User manual  
7.  connection which is required by  some health care professionals for reasons which the manufacturer  of the device cannot explain  The cable shown is used  fig  3   number 18   its banana slots into another black socket  fig  3   number 11  labelled PC database  and the other end of the cable  fig   3  number 18  in the form of a USB connector slots into any USB port  on the computer  Warning  this method does not create any  galvanized or electronic data connection  The cables in the black  sockets are not interchangeable     Working with the device     The device ACUCOMB AC V2 edition    BETTER FUTURE    variant  LIGHT is  in spite of its high technical level  easy to operate  The  device is switched on by pressing the large revolving button  fig  3   number 4  on the front panel of the device  Briefly press and release  the button  and the device switches on  the power on indicator lights  up   Now the device is ready for operation  The operator holds the  differential probe in their hand  and the person being analyzed holds  21    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    the non differential electrode in their hand    If the device is left unused for a longer period of time  it will switch  itself off  But if we want to switch the device off immediately  which   due to the minimal energy consumption from the built in batteries   and the automatic switching off  is not usually necessary  then we  have to press the RESET button  which is foun
8.  represent insufficient pressure  perhaps the ball shaped tip  had not been placed on an active skin point  Values coloured red to  crimson  80 and above  mean there is too much pressure on the tip  of the probe  and possibly indicate skin that is too damp and perhaps  even damaged     in this case the work with the indicator must  terminate immediately     ap 49 50 60 7  wn    aluda 7 why 2  ny       w Q 2   Vy wN  NY ly    mis ca    Figure 5  The indicator scale     10  Maintenance of the device    The device contains an internal battery which must be regularly  recharged  If the battery is empty  the empty battery indicator will  begin to flash  The batteries are charged using the charging cable   fig  3  number 15  which slots into the charging socket  fig  3   number 10   The other end slots into the USB port of any computer   or into a mains charger  fig  3  number 14     Completely discharging the built in battery is always harmful  and  shortens its life span  On the other hand  it is impossible to charge  the battery too much  so there should not be any fear of charging it  If  the built in battery cannot be charged  then it has probably reached    23    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    the end of its life span  and must be changed  This operation must be  carried out exclusively by an authorized service agent  who will at the  same time carry out an overall inspection of the functionality and  safety of the device  as required
9.  with a negligible electrical  input of 830mW  does not significantly affect the environment   Special attention  in accordance with local legal regulations in  force  must be paid only to the disposal of the built in battery   The device must not be disposed of as municipal waste     25    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    At the end of its life span  when the device is decommissioned  it is  recommended that it be dismantled into       the metal electrodes and other metal parts  bowls       the battery      the plastic parts      the built in electronic component itself  the printed circuit  boards  the cables and the measuring device      The dismantling is easily done by disconnecting the metal electrodes  and cables  and then dismantling the device unit and removing the  battery and other embedded parts     The metals are disposed of according to local rules and legal  regulations regarding aluminium  brass  copper and silver  the  batteries are subject to specific disposal requirements and the other  parts are disposed of as electrical equipment  The plastic parts are  made of ABS plastic     No part of the device constitutes hazardous waste  The device does  not produce any waste which would unduly burden the environment     Important notice    The life span of the device  when taking into account normal  wear and tear  ageing of materials and most importantly its  ethical ageing  to which all electronic devices are subject  is se
10.  without an electrical  connection      17    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT       BETTER FUTURE       Fig  3  The device and its accessories    18    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    Fig  3     continued  The device and its accessories       19    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    Tab 1  Description of parts of the device                                                                   No  Description   1 Plastic sheathing of the device unit   2 Hand held analogue measuring device   3 Metal bowl galvanically connected with the non differential electrode  9  13   and therefore with the negative pole of the device power supply   4 Control push button for setting the sensitivity of the device  and the sound  volume  Power on indicator  Low battery indicator  Change in sound volume indicator   8 Socket for differential probe  red      hand held probe ending in a ball shaped  tip  see no  12    9 Socket for non differential electrode     a silver plated cylinder held in the  client s patient s hand  see no  13    10 Socket for the mains adapter charger cable  see nos  14  15   Socket for connecting the external metal bowls     labelled Test  see nos  16   11 and 17  and for symbolic connection to PC     labelled PC database  see no   18    12 Differential probe ending in a ball shaped tip   13 Non differential hand held cylindrical electrode   14 Mains adapter ch
11. USER MANUAL     Operating instructions     ACUCOMB AC V2 edition    BETTER FUTURE       variant LIGHT    MFN    NE    Indicator of areas of reduced electrical resistance  of intact skin    J    Manufacturer     EMBITRON s r o   Borsk   55   301 00 Plze       esk   republika   tel    420  377 421 607   377 423 556  embitron embitron eu  www embitron eu    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    Contents   The stated purpose and basic description of the device  The manufacturer of the device   The purpose of the manual   Warranties  technical service and operation of the device  Safety instructions   Technical characteristics of the device   Contraindications associated with the use of the device  Description of the device    Operating the device      Maintenance of the device     Cleaning and disinfection     Environmental protection and disposal of the device    Packaging     Storage      Transportation    21    23    24    25    27    28    28    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    1  The stated purpose and basic description of  the device    The purpose of the device ACUCOMB AC V2 edition    BETTER  FUTURE    variant LIGHT  fig  1  is to provide technical conditions  for the electrical seeking out and indication of small areas     points  on intact skin  usually called    active skin points    and mostly  identical to acupuncture points  which are distinguished by  specific electrical properties  primaril
12. arger   15 Charging cable for the mains adapter battery charger   16 External metal bowl   17 Cable for connecting external metal bowls   18 Cable for symbolic connection to PC  without an electrical connection    19 Demonstration panel connection  above standard equipment           20       User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT  9  Operating the device    Preparation of the device prior to measurement     The device itself must first be connected to the necessary cables and  probes  The connectors for this can be found on the rear panel of the  device  The red socket  fig  3  number 8  is for connecting the red  banana of the differential probe  fig  3  number 12   The black socket   which is found beside the red socket  fig  3  number 9  is for  connecting the black banana of the non differential probe  fig  3   number 13   Another black socket  fig  3  number 11  is for  connecting the external bowl  fig  3  number 16  using the connecting  cable  fig  3  number 17   The connecting cable is connected to the  device using the safety banana  and afterwards it slots into one of  the two side openings in the external metal bowl  Furthermore  it is  possible to symbolically connect the device to a PC computer  It  must be stressed that this is a purely symbolic connection  which is  not an electrical connection and therefore no electrical current or  electrical data flows through it  The connection is also not  galvanized  It is a purely symbolic
13. cal  Connection nieee eee teen et dat eee eee 1 piece    27    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT  14  Storage    The following conditions apply for storing the device     Temperature  10 to   50  C  without any sudden changes   Relative air humidity max  95   at 25  C   Atmospheric pressure 120 to 809 mmHg     15  Transportation    The transportation box containing the device set can be transported  in conventional covered vehicles  but it must be secured in such a  way that shocks  impacts and vibrations are not transmitted to the  device  The packaged device in the transport crate must be treated  as a fragile consignment  it must not be turned upside down  and the  required storage conditions must be adhered to     Because of the pointer analogue measuring instrument  embedded in the device  the device must not be subjected to  mechanical shocks  impacts and vibrations     Last revised  30 5 2012    28    
14. ce set including packaging  1 5 kg  The whole device set including packaging  1 2 kg       Dimensions    Device unit  190 x 137 x 103 mm       Plastic sheathing  material    ABS  acrylonitrile butadiene styrene           Surfaces of the metal  parts       Brass  silver plated brass  silver  anodized aluminium       13       User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    7  Contraindications associated with the use of the  device    It is absolutely essential to avoid any diagnostic conclusions  which were not unambiguously verified by standard medical  diagnostic procedures  and which could lead to any harm  whatsoever to the patient client  incorrect treatment  omission  or delay of necessary medical procedures  causing unfounded  fears etc       Using the device ACUCOMB AC V2 edition    BETTER FUTURE     variant LIGHT must never become a reason for neglecting or  delaying other necessary medical care procedures     There are no direct risks from using the device ACUCOMB AC V2  edition    BETTER FUTURE    varianta LIGHT  but from the perspective  of precaution it is necessary to consider the following aspects     e Using the device is absolutely contraindicated in a case when  the subjectively evaluated diagnostic conclusions cannot be  unambiguously confirmed by standard medical diagnostic  procedures  which might result in various types of harm to the  patient client  neglecting necessary care  improper subsequent  care  omission or delay of neces
15. d on the bottom of the  device  near its production label   see figure 4     Manulactured  EMBITRON       Fig  4  The bottom part of the device  with the production label and a non   arresting button     If necessary  it is possible to change the sensitivity of the device  while it is switched on  Turning the control button to the right  increases the sensitivity  and turning it to the left decreases the  sensitivity     In order to better indicate the measured value  the device generates  a tone which corresponds to the value shown on the analogue  measuring device  The intensity of the sound can be regulated  We  press the buttom  and the device begins to emit a constant tone  By  turning the button right left while holding it down we can  increase decrease the intensity of the sound     To facilitate and visually simplify the indication of the active skin  point  the device is equipped with a quality hand held measuring  device  The physical meaning of the indicated value is expressed in    22    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    chapter 1 of this User Manual  Ideally  the sensitivity of the device  should be set in such a way that  during an application of the  electrode tip  the probe  of a pressure of around 10 N  to a well  indicated active skin point  the hand of the indicator moves in the  area of green  to yellow coloured decades  being between 50 and 80  degrees     of the indicator scale  Lower values  coloured blue   usually
16. he device     7    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    but there is a presumption that they are capable of using electrical  medical devices  and that they are working under the supervision of  a health care professional     The life span of the device is set as 10 years     5  Safety instructions    The device ACUCOMB AC V2 edition    BETTER FUTURE    variant  LIGHT has been designed and manufactured for the safest possible  use  Users of this device must have available the information  provided by this user manual  and other information provided during  operator training or in connection with the execution of the relevant  related methods  as well as legal provisions     The device must not be used     e for any other purpose than that for which it is intended     e if the device is being charged from the USB port of a  computer  or from a USB charger connected to the mains     e in an outdoor environment during adverse weather conditions  if  the required operating conditions cannot be met  temperature 5  to 45  C  maximum humidity 90  at 25  C  without an increase in  air flow or dust  in accordance with CSN 33 2000 3  a Czech  electrotechnical regulation      e in an environment where there is a high risk of injury from  electric shock  e g  a damp or wet environment  or an  environment with moving  unsecured mechanical objects      e inan environment where there is a high risk of explosion   e inan environment where there is a high r
17. ice to the user  The faulty  device must be sent  insured  by post or by using a registered courier  service  including any auxiliary equipment  in appropriate packaging  which sufficiently protects the sensitive measuring device against  impact and vibration  the package must be marked    Fragile  In a  warranty claim  it is necessary to provide the requisite claim  information  the address of the user or  if applicable  the sender of  the device  phone and also possibly mail details  a precise  description of the defect  proof of purchase of the device with the  manufacturing code highlighted and the date of sale shown  as well  as the stamp and the legible signature of the seller   Any device  which shows a defect demonstrably covered by the warranty will be  6    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    repaired free of charge and immediately sent back     Any post warranty service must of course be paid for  otherwise it is  subject to the same conditions     The warranty is extended for the period of time during which the  device was under repair  The warranty is invalid if the device was not  used in accordance with the accompanying documentation and in  accordance with the related applicable legal provisions  regulations   standards and normal  generally accepted ways of using electronic  devices  Similarly  the warranty does not cover any possible damage  caused by misuse or improper use of the device  or damage caused  by servicing by
18. isk of fire    e during unauthorized technical modifications    e in unsuitable power supply conditions     e in the absence of a fully responsible user  the proper execution  of the appropriate methods  or in the absence of qualified and    8    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    knowledgeable service staff     in a situation where the device or the relevant related  method may endanger the patient client  or their legitimate  interests  or if applicable the operator or any third party  or  where there is a risk of any other harm     The skin of the body parts which come into contact with the  device must be fully intact  so that contact with the device or any  parts thereof  the electrodes  does not pose any risk to the  patient or other persons coming into contact with the equipment     The device must be used only in a sufficiently stable position   The falling of any part of the device must be effectively  prevented     It is essential to regularly monitor the integrity of all parts of the  device  especially electric cables and the sheathed parts of the  device     Only original parts must be used as part of the equipment  and  no modifications must be carried out  other than those explicitly  carried out or authorized by the manufacturer     The device ACUCOMB AC V2 edition    BETTER FUTURE     variant LIGHT is a medical device  therefore it should be subject  to regular functionality and safety inspections  These inspections  are 
19. m the USB port of a    computer  or from a USB charger connected to the mains  it  must not be used     2  The manufacturer of the device    EMBITRON s r o   Borsk   55  301 00 Plze     Ceska republika  ICO 26361175    www embitron eu    embitron embitron eu    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    3  The purpose of the user manual    The purpose of this manual is to provide the users of the device  and  the relevant responsible organizations  with basic essential  information necessary for the service and safe operation of the  device  If the user or responsible person has any doubts or  uncertainties  it is important that they contact the manufacturer  or an  authorised distributor  without delay     Recipients of this manual are all users of the device  as well as  responsible organizations which use the device in any way  or are  responsible for its operation     This user manual must accompany each device  and must be  accessible to all persons whose work is in any way connected with  the device in question     4  Warranties  technical service and operation  of the device    The device ACUCOMB AC V2 edition    BETTER FUTURE    variant  LIGHT is provided with a warranty  in compliance with the legal  regulations in effect  for a period of 24 months from the day of sale   unless otherwise stated in the relevant purchase contract     Any warranty claim will be realized through the agent in the relevant  country and region  who sold the dev
20. o any harm  whatsoever to the patient client  incorrect treatment  omission  or delay of necessary medical procedures  causing unfounded  fears etc       Using the device ACUCOMB AC V2 edition    BETTER FUTURE     variant LIGHT must never become a reason for neglecting or  delaying other necessary medical care procedures     The manufacturer of the device ACUCOMB AC V2 edition    BETTER  FUTURE    variant LIGHT is not liable for any damage arising from  failure to comply with all the listed requirements  or from a disregard  for the principles of medical ethics  qualifications or the findings of  medical science     Il    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    6  The technical characteristics of the device       Medical classification    A class lla medical device in accordance with  Council Directive MDD93 42 EEC   equivalent to NV CR 336 2004 Coll         Electrotechnical  classification    A device with an internal power source    battery Ni MH 2 x 1 2 VDC  2 100 mAh     which is charged  using a charger  adapter   from an  external power source    mains AC power supply 230 V   50 Hz  after adjustment  by the manufacturer also 115 V   60 Hz    or charged from the USB port of any computer     Use of the device while it is connected to the mains  using a mains adapter   charger  or using a PC  is  prohibited     Regarding protection against injury by electric shock   the mains adapter   charger itself is a class II device        Charger  USB
21. on until it is completely dry     The covers of the device must be removed only by the  manufacturer or by a person appointed by them  who is  authorized to carry out a technical service  Similarly  no one may  dismantle  repair  modify or otherwise interfere with the device   By ignoring this warning  the user of the device not only stands  to make the warranty void  but also puts themselves in danger of  wilful injury or other harm     Before cleaning or disinfection of the device  or any parts  thereof  the device must be switched off  The detachable mains  supply of the adapter     the charger of the built in bateries     must  be completely disconnected  both from the mains and from the  device     During cleaning and disinfection  only gentle and tried and tested  cleaning and disinfection products must be used  In no case may  any of these liquids be allowed to penetrate the device  or any of  its closed  sheathed  parts     The detachable mains supply of the adapter  charger  of the  device ACUCOMB AC V2 edition    BETTER FUTURE    variant  LIGHT must not be located in a place where it could be broken  or forcibly pulled out of the mains socket  or from the socket on  the rear panel of the device unit     10    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    Special warning     It is absolutely essential to avoid any diagnostic conclusions  which were not unambiguously verified by standard medical  diagnostic procedures  and which could lead t
22. required in all EU countries by the relevant legal provisions   The user  or if applicable the responsible organization  is in  charge of the proper implementation of these inspections  These  functionality and safety inspections must be carried out only by  the manufacturer of the device  or by a person authorized by  them who is suitable qualified  trained and equipped with the  necessary calibrated measuring equipment  The manufacturer  stipulates the time period of these inspections in accordance with  the anticipated life span of the built in batteries  i e  a time period  of 3 years    The user of the device ACUCOMB AC V2 edition    BETTER  FUTURE  variant LIGHT is obliged to protect the device against  the risk of fire  in a manner similar to any other electrical devices   Any excessive increase in the temperature of the device  or any    9    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    parts thereof  is reason for its immediate shutdown and  disconnection from the mains  Immediately afterwards  a  professional repair service must be arranged     During operation  the device must not be sealed in an airtight  manner     Entry of foreign objects and liquids into the device must be  prevented  likewise regarding unwanted shocks and sudden  changes in temperature     Where there is a risk of condensation  an accumulation of  airborne water vapours as a result of moving from a cool  environment to a warmer one  the device must not be switched  
23. s prohibited in cases of immediate danger to  life  It is also prohibited in states of vegetative instability and or  serious mental illness  where the application of a non standard  method may cause an undesirable reaction in the patient client     16    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT  8  Description of the device    The device ACUCOMB AC V2 edtion    BETTER FUTURE    variant  LIGHT is composed of the device unit itself  fig  3  legend in table 1   and auxiliary equipment  fig  3      which includes     e a grey coiled cable with an integrated hand held differential  probe ending in a silver spherical electrode with a diameter of 3  mm  the other end of the cable ends in a red coloured safety  banana which is designed to be plugged into the socket of the  same colour on the rear panel of the device  also see fig  3      e a black  highly flexible straight cable with an integrated non   differential silver plated cylindrical electrode  the other end of the  cable ends in a black coloured safety banana which is designed  to be plugged into the socket of the same colour on the rear  panel of the device  fig  3      e a USB charger   adapter  230V 50Hz  5 VA   fig  3  designed for  charging the built in batteries directly from the mains  fig  3      e a cable for charging the built in batteries through the USB port of  any computer  fig  3      e An outer metal bowl with a supply cable  fig  3      e A cable for symbolic connection to a PC 
24. sary medical procedures   causing unfounded fears etc       e Using the device is absolutely contraindicated in all cases where  there would be a risk to the patient associated with the  mechanical application of the metal electrode with the ball   shaped tip on to the skin     e Using the device is absolutely contraindicated in all cases where  the patient s skin is in any way damaged in the places where the  electrode is to be applied     e Use of the device is absolutely contraindicated in cases of risk of  a serious allergic or toxico allergic reaction by the patient client  to the metal of the electrodes  or in cases of individual  intolerance to them     14    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    Use of the device is absolutely contraindicated if complete  disinfection of the electrodes to be applied to the patient has not  been ensured  Disinfection must be carried out with recognized  and suitable disinfectants     Use of the device is relatively contraindicated in the case of  patients with an implanted electrical stimulator  most often a  pacemaker  or any other electrical substitute which affects vital  bodily functions  e g  the electronically controlled administration  of insulin   Use of the device is also contraindicated in the case  of patients whose important bodily functions are monitored or  controlled by electronic devices connected to the patient s body   e g  in an intensive care unit   The electrical voltage and  elec
25. t  as 10 years     For this time period  technical service  as well as any necessary  delivery of spare parts  is guaranteed     26    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    13  Packaging    The device and its auxiliary equipment is packaged in individual inert  plastic protective foil packaging  For normal transportation  the  device is also placed into a paper transportation box  in which the  device  as well as the auxiliary equipment  is fixed in conventional  reinforcing plastic or paper packaging material  chosen especially to  protect the device from shock  The transportation box must be  suitably labelled  The transportation box with the device set must  contain       The device ACUCOMB AC V2 edition    BETTER FUTURE     variant LIGHT  basic unit  fig  1  aeee 1 piece      A differential hand held probe with a silver plated spherical  electrode on its tip  and a coiled supply cable                   1 piece      A non differential cylindrical silver plated probe with a straight   highly flexible Cable    eeeeeceeeeeeeeeeeeeeeeeeeeseeeeeeeeeeeenaeees 1 piece      A cable for connecting the device to an external metal bowl  a aes E E E E AE TOA 1 piece      Acable for linking the connector designed for charging the built   in battery  ending in a USB COnneCtOl             ccceeeeeeeeeeeee 1 piece      AUSB charger   mains adaptel              c cessceeeeteeeesteeeees 1 piece      A cable for symbolic connection to PC  without an electri
26. trical current produced by the device do not exceed 2 4 Volts  and 25 micro amperes  but even so it is not possible to  completely statistically exclude the possibility of a fault in the  functioning of the electronic devices or vitally important implants     Use of the device is absolutely contraindicated in the case of a  patient who has a serious medical condition which requires  completely different treatment  These are  for example   circulatory uncompensated cardiac patients  patients with  coronary insufficiency  a pre coronary state  recent myocardial  infarction  serious acute cardiac rhythm disorders  patients with  an acute central brain injury  patients who are in shock as a  result of injury and or are bleeding heavily  patients with serious  infections  patients in an acute paroxysmal state  patients with  hyper tonic or hypo tonic crisis  patients with a glycemic crisis   patients in anaphylactic shock or with serious symptoms of  allergic reactions  acutely intoxicated patients in a serious state   etc     Pregnancy  here it is a local and relative contraindication  related  only to the woman s abdominal area where the foetus is  developing  The possibility of it being affected by a voltage of 2 4  V applied to intact skin  and a maximum current of 25 micro   amperes  is extremely unlikely  especially if most of the contact  measurements take place away from the abdominal area     15    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    e
27. y reduced electrical  resistance and reduced electrical impedance compared to its  surroundings   The seeking out of these small skin areas  points   is carried out by sufficient  non damaging pressure by an  attached electrode which ends in a ball shaped tip  The result of  the combined mechanical and  subliminal  electrical irritation of  the relevant small skin area  point  with reduced electrical  resistance is its reliable and objective localization  At the same  time  on a 100 degree measuring scale  the value is indicated  which is the result of the effect of the mechanical and electrical  quantities at the given point of reduced resistance of the skin  see  the following graphs in fig  2   and it is conversely higher as the  the resistance of the indicated point of skin is lower  the indication  of the point   but likewise it is higher  the greater the pressure on  the ball shaped tip electrode  which is laid on the active skin  point        Fig 1  Device ACUCOMB AC V2 edition    BETTER FUTURE    variant  LIGHT     picture   3    User manual   ACUCOMB AC V2 edition    BETTER FUTURE    variant LIGHT    Symbolic   certainty of finding an area of skin with reduced  electrical resistance    sess sesasee 8    Electrical voltage and electrical current on the skin area   during given electrical resistance    5 00    0 00  5 100 150 200 250 300 35 400    Bes tricky odpor R  ohm       Electrical voltage  V     Electrical current  uA        Fig  2  Device ACUCOMB AC V2 edition 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Horizon Fitness T40 Exploded Diagram  4-in-1 portable evaporative air conditioner and heat pump  MV-TILT3B INSTALLATION INSTRUCTIONS  APP521 User manual  1 How to De-install the Demo-Version  Remote Control Center Professional version user manual for Android  CFOx61/x91/432 video links User manual      Copyright © All rights reserved. 